Wikipedia

Microscopy

Microscopy is the technical field of using microscopes to view objects and areas of objects that cannot be seen with the naked eye (objects that are not within the resolution range of the normal eye).[1] There are three well-known branches of microscopy: optical, electron, and scanning probe microscopy, along with the emerging field of X-ray microscopy.[citation needed]

Scanning electron microscope image of pollen (false colors)
Microscopic examination in a biochemical laboratory

Optical microscopy and electron microscopy involve the diffraction, reflection, or refraction of electromagnetic radiation/electron beams interacting with the specimen, and the collection of the scattered radiation or another signal in order to create an image. This process may be carried out by wide-field irradiation of the sample (for example standard light microscopy and transmission electron microscopy) or by scanning a fine beam over the sample (for example confocal laser scanning microscopy and scanning electron microscopy). Scanning probe microscopy involves the interaction of a scanning probe with the surface of the object of interest. The development of microscopy revolutionized biology, gave rise to the field of histology and so remains an essential technique in the life and physical sciences. X-ray microscopy is three-dimensional and non-destructive, allowing for repeated imaging of the same sample for in situ or 4D studies, and providing the ability to "see inside" the sample being studied before sacrificing it to higher resolution techniques. A 3D X-ray microscope uses the technique of computed tomography (microCT), rotating the sample 360 degrees and reconstructing the images. CT is typically carried out with a flat panel display. A 3D X-ray microscope employs a range of objectives, e.g., from 4X to 40X, and can also include a flat panel.

History edit

 
Antonie van Leeuwenhoek (1632–1723)

The field of microscopy (optical microscopy) dates back to at least the 17th-century. Earlier microscopes, single lens magnifying glasses with limited magnification, date at least as far back as the wide spread use of lenses in eyeglasses in the 13th century[2] but more advanced compound microscopes first appeared in Europe around 1620[3][4] The earliest practitioners of microscopy include Galileo Galilei, who found in 1610 that he could close focus his telescope to view small objects close up[5][6] and Cornelis Drebbel, who may have invented the compound microscope around 1620[7][8] Antonie van Leeuwenhoek developed a very high magnification simple microscope in the 1670s and is often considered to be the first acknowledged microscopist and microbiologist.[9][10]

Optical microscopy edit

 
Stereo microscope

Optical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single lens or multiple lenses to allow a magnified view of the sample.[11] The resulting image can be detected directly by the eye, imaged on a photographic plate, or captured digitally. The single lens with its attachments, or the system of lenses and imaging equipment, along with the appropriate lighting equipment, sample stage, and support, makes up the basic light microscope. The most recent development is the digital microscope, which uses a CCD camera to focus on the exhibit of interest. The image is shown on a computer screen, so eye-pieces are unnecessary.[12]

Limitations edit

Limitations of standard optical microscopy (bright field microscopy) lie in three areas;

  • The technique can only image dark or strongly refracting objects effectively.
  • There is a diffraction-limited resolution depending on incident wavelength; in visible range, the resolution of optical microscopy is limited to approximately 0.2  micrometres (see: microscope) and the practical magnification limit to ~1500x.[13]
  • Out-of-focus light from points outside the focal plane reduces image clarity.[14]

Live cells in particular generally lack sufficient contrast to be studied successfully, since the internal structures of the cell are colorless and transparent. The most common way to increase contrast is to stain the structures with selective dyes, but this often involves killing and fixing the sample.[15] Staining may also introduce artifacts, which are apparent structural details that are caused by the processing of the specimen and are thus not features of the specimen. In general, these techniques make use of differences in the refractive index of cell structures. Bright-field microscopy is comparable to looking through a glass window: one sees not the glass but merely the dirt on the glass. There is a difference, as glass is a denser material, and this creates a difference in phase of the light passing through. The human eye is not sensitive to this difference in phase, but clever optical solutions have been devised to change this difference in phase into a difference in amplitude (light intensity).[citation needed]

Techniques edit

To improve specimen contrast or highlight structures in a sample, special techniques must be used. A huge selection of microscopy techniques are available to increase contrast or label a sample.

Bright field edit

Bright field microscopy is the simplest of all the light microscopy techniques. Sample illumination is via transmitted white light, i.e. illuminated from below and observed from above. Limitations include low contrast of most biological samples and low apparent resolution due to the blur of out-of-focus material. The simplicity of the technique and the minimal sample preparation required are significant advantages.[citation needed]

Oblique illumination edit

The use of oblique (from the side) illumination gives the image a three-dimensional appearance and can highlight otherwise invisible features. A more recent technique based on this method is Hoffmann's modulation contrast, a system found on inverted microscopes for use in cell culture. Oblique illumination enhances contrast even in clear specimens; however, because light enters off-axis, the position of an object will appear to shift as the focus is changed. This limitation makes techniques like optical sectioning or accurate measurement on the z-axis impossible.

Dark field edit

Dark field microscopy is a technique for improving the contrast of unstained, transparent specimens.[16] Dark field illumination uses a carefully aligned light source to minimize the quantity of directly transmitted (unscattered) light entering the image plane, collecting only the light scattered by the sample. Dark field can dramatically improve image contrast – especially of transparent objects – while requiring little equipment setup or sample preparation. However, the technique suffers from low light intensity in the final image of many biological samples and continues to be affected by low apparent resolution.

 
A diatom under Rheinberg illumination

Rheinberg illumination is a variant of dark field illumination in which transparent, colored filters are inserted just before the condenser so that light rays at high aperture are differently colored than those at low aperture (i.e., the background to the specimen may be blue while the object appears self-luminous red). Other color combinations are possible, but their effectiveness is quite variable.[17]

Dispersion staining edit

Dispersion staining is an optical technique that results in a colored image of a colorless object. This is an optical staining technique and requires no stains or dyes to produce a color effect. There are five different microscope configurations used in the broader technique of dispersion staining. They include brightfield Becke line, oblique, darkfield, phase contrast, and objective stop dispersion staining.

Phase contrast edit

 
Phase-contrast light micrograph of undecalcified hyaline cartilage showing chondrocytes and organelles, lacunae and extracellular matrix
In electron microscopy: Phase-contrast imaging

More sophisticated techniques will show proportional differences in optical density. Phase contrast is a widely used technique that shows differences in refractive index as difference in contrast. It was developed by the Dutch physicist Frits Zernike in the 1930s (for which he was awarded the Nobel Prize in 1953). The nucleus in a cell for example will show up darkly against the surrounding cytoplasm. Contrast is excellent; however it is not for use with thick objects. Frequently, a halo is formed even around small objects, which obscures detail. The system consists of a circular annulus in the condenser, which produces a cone of light. This cone is superimposed on a similar sized ring within the phase-objective. Every objective has a different size ring, so for every objective another condenser setting has to be chosen. The ring in the objective has special optical properties: it, first of all, reduces the direct light in intensity, but more importantly, it creates an artificial phase difference of about a quarter wavelength. As the physical properties of this direct light have changed, interference with the diffracted light occurs, resulting in the phase contrast image. One disadvantage of phase-contrast microscopy is halo formation (halo-light ring).

Differential interference contrast edit

Superior and much more expensive is the use of interference contrast. Differences in optical density will show up as differences in relief. A nucleus within a cell will actually show up as a globule in the most often used differential interference contrast system according to Georges Nomarski. However, it has to be kept in mind that this is an optical effect, and the relief does not necessarily resemble the true shape. Contrast is very good and the condenser aperture can be used fully open, thereby reducing the depth of field and maximizing resolution.

The system consists of a special prism (Nomarski prism, Wollaston prism) in the condenser that splits light in an ordinary and an extraordinary beam. The spatial difference between the two beams is minimal (less than the maximum resolution of the objective). After passage through the specimen, the beams are reunited by a similar prism in the objective.

In a homogeneous specimen, there is no difference between the two beams, and no contrast is being generated. However, near a refractive boundary (say a nucleus within the cytoplasm), the difference between the ordinary and the extraordinary beam will generate a relief in the image. Differential interference contrast requires a polarized light source to function; two polarizing filters have to be fitted in the light path, one below the condenser (the polarizer), and the other above the objective (the analyzer).

Note: In cases where the optical design of a microscope produces an appreciable lateral separation of the two beams we have the case of classical interference microscopy, which does not result in relief images, but can nevertheless be used for the quantitative determination of mass-thicknesses of microscopic objects.

Interference reflection edit

An additional technique using interference is interference reflection microscopy (also known as reflected interference contrast, or RIC). It relies on cell adhesion to the slide to produce an interference signal. If there is no cell attached to the glass, there will be no interference.

Interference reflection microscopy can be obtained by using the same elements used by DIC, but without the prisms. Also, the light that is being detected is reflected and not transmitted as it is when DIC is employed.

Fluorescence edit

 
Images may also contain artifacts. This is a confocal laser scanning fluorescence micrograph of thale cress anther (part of stamen). The picture shows among other things a nice red flowing collar-like structure just below the anther. However, an intact thale cress stamen does not have such collar, this is a fixation artifact: the stamen has been cut below the picture frame, and epidermis (upper layer of cells) of stamen stalk has peeled off, forming a non-characteristic structure. Photo: Heiti Paves from Tallinn University of Technology.

When certain compounds are illuminated with high energy light, they emit light of a lower frequency. This effect is known as fluorescence. Often specimens show their characteristic autofluorescence image, based on their chemical makeup.

This method is of critical importance in the modern life sciences, as it can be extremely sensitive, allowing the detection of single molecules. Many fluorescent dyes can be used to stain structures or chemical compounds. One powerful method is the combination of antibodies coupled to a fluorophore as in immunostaining. Examples of commonly used fluorophores are fluorescein or rhodamine.

The antibodies can be tailor-made for a chemical compound. For example, one strategy often in use is the artificial production of proteins, based on the genetic code (DNA). These proteins can then be used to immunize rabbits, forming antibodies which bind to the protein. The antibodies are then coupled chemically to a fluorophore and used to trace the proteins in the cells under study.

Highly efficient fluorescent proteins such as the green fluorescent protein (GFP) have been developed using the molecular biology technique of gene fusion, a process that links the expression of the fluorescent compound to that of the target protein. This combined fluorescent protein is, in general, non-toxic to the organism and rarely interferes with the function of the protein under study. Genetically modified cells or organisms directly express the fluorescently tagged proteins, which enables the study of the function of the original protein in vivo.

Growth of protein crystals results in both protein and salt crystals. Both are colorless and microscopic. Recovery of the protein crystals requires imaging which can be done by the intrinsic fluorescence of the protein or by using transmission microscopy. Both methods require an ultraviolet microscope as protein absorbs light at 280 nm. Protein will also fluorescence at approximately 353 nm when excited with 280 nm light.[18]

Since fluorescence emission differs in wavelength (color) from the excitation light, an ideal fluorescent image shows only the structure of interest that was labeled with the fluorescent dye. This high specificity led to the widespread use of fluorescence light microscopy in biomedical research. Different fluorescent dyes can be used to stain different biological structures, which can then be detected simultaneously, while still being specific due to the individual color of the dye.

To block the excitation light from reaching the observer or the detector, filter sets of high quality are needed. These typically consist of an excitation filter selecting the range of excitation wavelengths, a dichroic mirror, and an emission filter blocking the excitation light. Most fluorescence microscopes are operated in the Epi-illumination mode (illumination and detection from one side of the sample) to further decrease the amount of excitation light entering the detector.

See also: total internal reflection fluorescence microscope Neuroscience

Confocal edit

Confocal laser scanning microscopy uses a focused laser beam (e.g. 488 nm) that is scanned across the sample to excite fluorescence in a point-by-point fashion. The emitted light is directed through a pinhole to prevent out-of-focus light from reaching the detector, typically a photomultiplier tube. The image is constructed in a computer, plotting the measured fluorescence intensities according to the position of the excitation laser. Compared to full sample illumination, confocal microscopy gives slightly higher lateral resolution and significantly improves optical sectioning (axial resolution). Confocal microscopy is, therefore, commonly used where 3D structure is important.

A subclass of confocal microscopes are spinning disc microscopes which are able to scan multiple points simultaneously across the sample. A corresponding disc with pinholes rejects out-of-focus light. The light detector in a spinning disc microscope is a digital camera, typically EM-CCD or sCMOS.

Two-photon microscopy edit

A two-photon microscope is also a laser-scanning microscope, but instead of UV, blue or green laser light, a pulsed infrared laser is used for excitation. Only in the tiny focus of the laser is the intensity high enough to generate fluorescence by two-photon excitation, which means that no out-of-focus fluorescence is generated, and no pinhole is necessary to clean up the image.[19] This allows imaging deep in scattering tissue, where a confocal microscope would not be able to collect photons efficiently.[20] Two-photon microscopes with wide-field detection are frequently used for functional imaging, e.g. calcium imaging, in brain tissue.[21] They are marketed as Multiphoton microscopes by several companies, although the gains of using 3-photon instead of 2-photon excitation are marginal.

Single plane illumination microscopy and light sheet fluorescence microscopy edit

Using a plane of light formed by focusing light through a cylindrical lens at a narrow angle or by scanning a line of light in a plane perpendicular to the axis of objective, high resolution optical sections can be taken.[22][23][24] Single plane illumination, or light sheet illumination, is also accomplished using beam shaping techniques incorporating multiple-prism beam expanders.[25][26] The images are captured by CCDs. These variants allow very fast and high signal to noise ratio image capture.

Wide-field multiphoton microscopy edit

Wide-field multiphoton microscopy[27][28][29][30] refers to an optical non-linear imaging technique in which a large area of the object is illuminated and imaged without the need for scanning. High intensities are required to induce non-linear optical processes such as two-photon fluorescence or second harmonic generation. In scanning multiphoton microscopes the high intensities are achieved by tightly focusing the light, and the image is obtained by beam scanning. In wide-field multiphoton microscopy the high intensities are best achieved using an optically amplified pulsed laser source to attain a large field of view (~100 µm).[27][28][29] The image in this case is obtained as a single frame with a CCD camera without the need of scanning, making the technique particularly useful to visualize dynamic processes simultaneously across the object of interest. With wide-field multiphoton microscopy the frame rate can be increased up to a 1000-fold compared to multiphoton scanning microscopy.[28] In scattering tissue, however, image quality rapidly degrades with increasing depth.

Deconvolution edit

Fluorescence microscopy is a powerful technique to show specifically labeled structures within a complex environment and to provide three-dimensional information of biological structures. However, this information is blurred by the fact that, upon illumination, all fluorescently labeled structures emit light, irrespective of whether they are in focus or not. So an image of a certain structure is always blurred by the contribution of light from structures that are out of focus. This phenomenon results in a loss of contrast especially when using objectives with a high resolving power, typically oil immersion objectives with a high numerical aperture.

 
Mathematically modeled Point Spread Function of a pulsed THz laser imaging system.[31]

However, blurring is not caused by random processes, such as light scattering, but can be well defined by the optical properties of the image formation in the microscope imaging system. If one considers a small fluorescent light source (essentially a bright spot), light coming from this spot spreads out further from our perspective as the spot becomes more out of focus. Under ideal conditions, this produces an "hourglass" shape of this point source in the third (axial) dimension. This shape is called the point spread function (PSF) of the microscope imaging system. Since any fluorescence image is made up of a large number of such small fluorescent light sources, the image is said to be "convolved by the point spread function". The mathematically modeled PSF of a terahertz laser pulsed imaging system is shown on the right.

The output of an imaging system can be described using the equation:

 

Where n is the additive noise.[32] Knowing this point spread function[33] means that it is possible to reverse this process to a certain extent by computer-based methods commonly known as deconvolution microscopy.[34] There are various algorithms available for 2D or 3D deconvolution. They can be roughly classified in nonrestorative and restorative methods. While the nonrestorative methods can improve contrast by removing out-of-focus light from focal planes, only the restorative methods can actually reassign light to its proper place of origin. Processing fluorescent images in this manner can be an advantage over directly acquiring images without out-of-focus light, such as images from confocal microscopy, because light signals otherwise eliminated become useful information. For 3D deconvolution, one typically provides a series of images taken from different focal planes (called a Z-stack) plus the knowledge of the PSF, which can be derived either experimentally or theoretically from knowing all contributing parameters of the microscope.

Sub-diffraction techniques edit

 
Example of super-resolution microscopy. Image of Her3 and Her2, target of the breast cancer drug Trastuzumab, within a cancer cell.

A multitude of super-resolution microscopy techniques have been developed in recent times which circumvent the diffraction limit.

This is mostly achieved by imaging a sufficiently static sample multiple times and either modifying the excitation light or observing stochastic changes in the image. The deconvolution methods described in the previous section, which removes the PSF induced blur and assigns a mathematically 'correct' origin of light, are used, albeit with slightly different understanding of what the value of a pixel mean. Assuming most of the time, one single fluorophore contributes to one single blob on one single taken image, the blobs in the images can be replaced with their calculated position, vastly improving resolution to well below the diffraction limit.

To realize such assumption, Knowledge of and chemical control over fluorophore photophysics is at the core of these techniques, by which resolutions of ~20 nanometers are obtained.[35][36]

Serial time-encoded amplified microscopy edit

Serial time encoded amplified microscopy (STEAM) is an imaging method that provides ultrafast shutter speed and frame rate, by using optical image amplification to circumvent the fundamental trade-off between sensitivity and speed, and a single-pixel photodetector to eliminate the need for a detector array and readout time limitations[37] The method is at least 1000 times faster than the state-of-the-art CCD and CMOS cameras. Consequently, it is potentially useful for scientific, industrial, and biomedical applications that require high image acquisition rates, including real-time diagnosis and evaluation of shockwaves, microfluidics, MEMS, and laser surgery.[38]

Extensions edit

Most modern instruments provide simple solutions for micro-photography and image recording electronically. However such capabilities are not always present and the more experienced microscopist may prefer a hand drawn image to a photograph. This is because a microscopist with knowledge of the subject can accurately convert a three-dimensional image into a precise two-dimensional drawing. In a photograph or other image capture system however, only one thin plane is ever in good focus.[citation needed]

The creation of accurate micrographs requires a microscopical technique using a monocular eyepiece. It is essential that both eyes are open and that the eye that is not observing down the microscope is instead concentrated on a sheet of paper on the bench besides the microscope. With practice, and without moving the head or eyes, it is possible to accurately trace the observed shapes by simultaneously "seeing" the pencil point in the microscopical image.[citation needed]

It is always less tiring to observe with the microscope focused so that the image is seen at infinity and with both eyes open at all times.[citation needed]

Other enhancements edit

Microspectroscopy:spectroscopy with a microscope

X-ray edit

As resolution depends on the wavelength of the light. Electron microscopy has been developed since the 1930s that use electron beams instead of light. Because of the much smaller wavelength of the electron beam, resolution is far higher.

Though less common, X-ray microscopy has also been developed since the late 1940s. The resolution of X-ray microscopy lies between that of light microscopy and electron microscopy.

Electron microscopy edit

Until the invention of sub-diffraction microscopy, the wavelength of the light limited the resolution of traditional microscopy to around 0.2 micrometers. In order to gain higher resolution, the use of an electron beam with a far smaller wavelength is used in electron microscopes.

  • Transmission electron microscopy (TEM) is quite similar to the compound light microscope, by sending an electron beam through a very thin slice of the specimen. The resolution limit in 2005 was around 0.05[dubious ] nanometer and has not increased appreciably since that time.
  • Scanning electron microscopy (SEM) visualizes details on the surfaces of specimens and gives a very nice 3D view. It gives results much like those of the stereo light microscope. The best resolution for SEM in 2011 was 0.4 nanometer.

Electron microscopes equipped for X-ray spectroscopy can provide qualitative and quantitative elemental analysis. This type of electron microscope, also known as analytical electron microscope, can be a very powerful tool for investigation of nanomaterials.[39]

Scanning probe microscopy edit

This is a sub-diffraction technique. Examples of scanning probe microscopes are the atomic force microscope (AFM), the scanning tunneling microscope, the photonic force microscope and the recurrence tracking microscope. All such methods use the physical contact of a solid probe tip to scan the surface of an object, which is supposed to be almost flat.

Ultrasonic force edit

Ultrasonic force microscopy (UFM) has been developed in order to improve the details and image contrast on "flat" areas of interest where AFM images are limited in contrast. The combination of AFM-UFM allows a near field acoustic microscopic image to be generated. The AFM tip is used to detect the ultrasonic waves and overcomes the limitation of wavelength that occurs in acoustic microscopy. By using the elastic changes under the AFM tip, an image of much greater detail than the AFM topography can be generated.

Ultrasonic force microscopy allows the local mapping of elasticity in atomic force microscopy by the application of ultrasonic vibration to the cantilever or sample. To analyze the results of ultrasonic force microscopy in a quantitative fashion, a force-distance curve measurement is done with ultrasonic vibration applied to the cantilever base, and the results are compared with a model of the cantilever dynamics and tip-sample interaction based on the finite-difference technique.

Ultraviolet microscopy edit

Ultraviolet microscopes have two main purposes. The first is to use the shorter wavelength of ultraviolet electromagnetic energy to improve the image resolution beyond that of the diffraction limit of standard optical microscopes. This technique is used for non-destructive inspection of devices with very small features such as those found in modern semiconductors. The second application for UV microscopes is contrast enhancement where the response of individual samples is enhanced, relative to their surrounding, due to the interaction of light with the molecules within the sample itself. One example is in the growth of protein crystals. Protein crystals are formed in salt solutions. As salt and protein crystals are both formed in the growth process, and both are commonly transparent to the human eye, they cannot be differentiated with a standard optical microscope. As the tryptophan of protein absorbs light at 280 nm, imaging with a UV microscope with 280 nm bandpass filters makes it simple to differentiate between the two types of crystals. The protein crystals appear dark while the salt crystals are transparent.

Infrared microscopy edit

The term infrared microscopy refers to microscopy performed at infrared wavelengths. In the typical instrument configuration, a Fourier Transform Infrared Spectrometer (FTIR) is combined with an optical microscope and an infrared detector. The infrared detector can be a single point detector, a linear array or a 2D focal plane array. FTIR provides the ability to perform chemical analysis via infrared spectroscopy and the microscope and point or array detector enable this chemical analysis to be spatially resolved, i.e. performed at different regions of the sample. As such, the technique is also called infrared microspectroscopy[40][41] An alternative architecture called Laser Direct Infrared (LDIR) Imaging involves the combination of a tuneable infrared light source and single point detector on a flying objective. This technique is frequently used for infrared chemical imaging, where the image contrast is determined by the response of individual sample regions to particular IR wavelengths selected by the user, usually specific IR absorption bands and associated molecular resonances. A key limitation of conventional infrared microspectroscopy is that the spatial resolution is diffraction-limited. Specifically the spatial resolution is limited to a figure related to the wavelength of the light. For practical IR microscopes, the spatial resolution is limited to 1-3x the wavelength, depending on the specific technique and instrument used. For mid-IR wavelengths, this sets a practical spatial resolution limit of ~3-30 μm.

IR versions of sub-diffraction microscopy also exist.[40][41] These include IR Near-field scanning optical microscope (NSOM),[42] photothermal microspectroscopy and atomic force microscope based infrared spectroscopy (AFM-IR), as well as scattering-type Scanning Near-field Optical Microscopy (s-SNOM)[43] & nano-FTIR that provide nanoscale spatial resolution at IR wavelengths.

Digital holographic microscopy edit

 
Human cells imaged by DHM phase shift (left) and phase contrast microscopy (right)

In digital holographic microscopy (DHM), interfering wave fronts from a coherent (monochromatic) light-source are recorded on a sensor. The image is digitally reconstructed by a computer from the recorded hologram. Besides the ordinary bright field image, a phase shift image is created.

DHM can operate both in reflection and transmission mode. In reflection mode, the phase shift image provides a relative distance measurement and thus represents a topography map of the reflecting surface. In transmission mode, the phase shift image provides a label-free quantitative measurement of the optical thickness of the specimen. Phase shift images of biological cells are very similar to images of stained cells and have successfully been analyzed by high content analysis software.

A unique feature of DHM is the ability to adjust focus after the image is recorded, since all focus planes are recorded simultaneously by the hologram. This feature makes it possible to image moving particles in a volume or to rapidly scan a surface. Another attractive feature is The ability of DHM to use low cost optics by correcting optical aberrations by software.

Digital pathology (virtual microscopy) edit

Digital pathology is an image-based information environment enabled by computer technology that allows for the management of information generated from a digital slide. Digital pathology is enabled in part by virtual microscopy, which is the practice of converting glass slides into digital slides that can be viewed, managed, and analyzed.

Laser microscopy edit

Laser microscopy is a rapidly growing field that uses laser illumination sources in various forms of microscopy.[44] For instance, laser microscopy focused on biological applications uses ultrashort pulse lasers, in a number of techniques labeled as nonlinear microscopy, saturation microscopy, and two-photon excitation microscopy.[45]

High-intensity, short-pulse laboratory x-ray lasers have been under development for several years. When this technology comes to fruition, it will be possible to obtain magnified three-dimensional images of elementary biological structures in the living state at a precisely defined instant. For optimum contrast between water and protein and for best sensitivity and resolution, the laser should be tuned near the nitrogen line at about 0.3 nanometers. Resolution will be limited mainly by the hydrodynamic expansion that occurs while the necessary number of photons is being registered.[46] Thus, while the specimen is destroyed by the exposure, its configuration can be captured before it explodes.[47][48][49][50][51][52][excessive citations]

Scientists have been working on practical designs and prototypes for x-ray holographic microscopes, despite the prolonged development of the appropriate laser.[53][54][55][56][57][58][59][60][excessive citations]

Photoacoustic microscopy edit

 
Photoacoustic micrograph of human red blood cells.

A microscopy technique relying on the photoacoustic effect,[61] i.e. the generation of (ultra)sound caused by light absorption. A focused and intensity modulated laser beam is raster scanned over a sample. The generated (ultra)sound is detected via an ultrasound transducer. Commonly piezoelectric ultrasound transducers are employed.[62]

The image contrast is related to the sample's absorption coefficient  . This is in contrast to bright or dark field microscopy, where the image contrast is due to transmittance or scattering. In principle, the contrast of fluorescence microscopy is proportional to the sample's absorption too. However, in fluorescence microscopy the fluorescence quantum yield   needs to be unequal to zero in order that a signal can be detected. In photoacoustic microscopy, however, every absorbing substance gives a photoacoustic signal   which is proportional to

 

Here   is the Grüneisen coefficient,   is the laser's photon energy and   is the sample's band gap energy. Therefore, photoacoustic microscopy seems well suited as a complementary technique to fluorescence microscopy, as a high fluorescence quantum yield leads to high fluorescence signals and a low fluorescence quantum yield leads to high photoacoustic signals.

Neglecting non-linear effects, the lateral resolution dx is limited by the Abbe diffraction limit:

 

where   is the wavelength of the excitation laser and NA is the numerical aperture of the objective lens. The Abbe diffraction limit holds if the incoming wave front is parallel. In reality, however, the laser beam profile is Gaussian. Therefore, in order to the calculate the achievable resolution, formulas for truncated Gaussian beams have to be used.[63]

Amateur microscopy edit

Amateur Microscopy is the investigation and observation of biological and non-biological specimens for recreational purposes. Collectors of minerals, insects, seashells, and plants may use microscopes as tools to uncover features that help them classify their collected items. Other amateurs may be interested in observing the life found in pond water and of other samples. Microscopes may also prove useful for the water quality assessment for people that keep a home aquarium. Photographic documentation and drawing of the microscopic images are additional pleasures. There are competitions for photomicrograph art. Participants of this pastime may use commercially prepared microscopic slides or prepare their own slides.

While microscopy is a central tool in the documentation of biological specimens, it is often insufficient to justify the description of a new species based on microscopic investigations alone. Often genetic and biochemical tests are necessary to confirm the discovery of a new species. A laboratory and access to academic literature is a necessity. There is, however, one advantage that amateurs have above professionals: time to explore their surroundings. Often, advanced amateurs team up with professionals to validate their findings and possibly describe new species.

In the late 1800s, amateur microscopy became a popular hobby in the United States and Europe. Several 'professional amateurs' were being paid for their sampling trips and microscopic explorations by philanthropists, to keep them amused on the Sunday afternoon (e.g., the diatom specialist A. Grunow, being paid by (among others) a Belgian industrialist). Professor John Phin published "Practical Hints on the Selection and Use of the Microscope (Second Edition, 1878)," and was also the editor of the "American Journal of Microscopy."

Examples of amateur microscopy images:

Application in forensic science edit

Microscopy has applications in the forensic sciences.[64] The microscope can detect, resolve and image the smallest items of evidence, often without any alteration or destruction. The microscope is used to identify and compare fibers, hairs, soils, and dust...etc.

In ink markings, blood stains or bullets, no specimen treatment is required and the evidence shows directly from microscopical examination. For traces of particular matter, the sample preparation must be done before microscopical examination occurs.[clarification needed]

Light microscopes are the most use in forensics, using photons to form images, microscopes which are most applicable for examining forensic specimens are as follows:[65]

1. The compound microscope

2. The comparison microscope

3. The stereoscopic microscope

4. The polarizing microscope

5. The micro spectrophotometer

This diversity of the types of microscopes in forensic applications comes mainly from their magnification ranges, which are (1- 1200X), (50 -30,000X) and (500- 250,000X) for the optical microscopy, SEM and TEM respectively.[65]

See also edit

References edit

  1. ^ The University of Edinburgh (March 6, 2018). "What is Microscopy?". The University of Edinburgh. from the original on April 10, 2018. Retrieved April 9, 2018.
  2. ^ Atti Della Fondazione Giorgio Ronchi E Contributi Dell'Istituto Nazionale Di Ottica, Volume 30, La Fondazione-1975, page 554
  3. ^ Albert Van Helden; Sven Dupré; Rob van Gent (2010). The Origins of the Telescope. Amsterdam University Press. p. 24. ISBN 978-90-6984-615-6. from the original on 15 February 2017.
  4. ^ William Rosenthal, Spectacles and Other Vision Aids: A History and Guide to Collecting, Norman Publishing, 1996, page 391 - 392
  5. ^ Robert D. Huerta, Giants of Delft: Johannes Vermeer and the Natural Philosophers : the Parallel Search for Knowledge During the Age of Discovery, Bucknell University Press - 2003, page 126
  6. ^ A. Mark Smith, From Sight to Light: The Passage from Ancient to Modern Optics, University of Chicago Press - 2014, page 387
  7. ^ Raymond J. Seeger, Men of Physics: Galileo Galilei, His Life and His Works, Elsevier - 2016, page 24
  8. ^ J. William Rosenthal, Spectacles and Other Vision Aids: A History and Guide to Collecting, Norman Publishing, 1996, page 391
  9. ^ Ford, Brian J. (1992). "From Dilettante to Diligent Experimenter: a Reappraisal of Leeuwenhoek as microscopist and investigator". Biology History. 5 (3). from the original on 2021-04-19. Retrieved 2021-04-16.
  10. ^ Lane, Nick (6 March 2015). "The Unseen World: Reflections on Leeuwenhoek (1677) 'Concerning Little Animal'." Philos Trans R Soc Lond B Biol Sci. 2015 Apr; 370 (1666): 20140344. [doi:10.1098/rstb.2014.0344]
  11. ^ Abramowitz M, Davidson MW (2007). "Introduction to Microscopy". Molecular Expressions. from the original on 2021-04-27. Retrieved 2007-08-22.
  12. ^ "Microscope - Compound, Numerical Aperture, and Immersion Microscopy | Britannica". www.britannica.com. from the original on 2023-06-22. Retrieved 2023-06-22.
  13. ^ Abbe, E. (1874) A contribution to the theory of the microscope and the nature of microscopic vision. Proc. Bristol Nat. Soc. 1, 200–261.
  14. ^ Pawley JB (editor) (2006). Handbook of Biological Confocal Microscopy (3rd ed.). Berlin: Springer. ISBN 0-387-25921-X.
  15. ^ Lakowicz, Joseph R. (1999) Principles Of Fluorescence Spectroscopy. New York: Kluwer Academic/Plenum.
  16. ^ Abramowitz M, Davidson MW (2003-08-01). "Darkfield Illumination". from the original on 2015-11-23. Retrieved 2008-10-21.
  17. ^ Abramowitz M, Davidson MW (2003-08-01). "Rheinberg Illumination". from the original on 2008-09-29. Retrieved 2008-10-21.
  18. ^ Gill, Harindarpal (January 2010). "Evaluating the efficacy of tryptophan fluorescence and absorbance as a selection tool for identifying protein crystals". Acta Crystallographica. F66 (Pt 3): 364–372. doi:10.1107/S1744309110002022. PMC 2833058. PMID 20208182.
  19. ^ Denk, Winfried; Svoboda, Karel (March 1997). "Photon Upmanship: Why Multiphoton Imaging Is More than a Gimmick". Neuron. 18 (3): 351–357. doi:10.1016/S0896-6273(00)81237-4. PMID 9115730. S2CID 2414593.
  20. ^ Denk, W.; Delaney, K.R.; Gelperin, A.; Kleinfeld, D.; Strowbridge, B.W.; Tank, D.W.; Yuste, R. (1994). "Anatomical and functional imaging of neurons using 2-photon laser scanning microscopy". Journal of Neuroscience Methods. 54 (2): 151–162. doi:10.1016/0165-0270(94)90189-9. PMID 7869748. S2CID 3772937.
  21. ^ Svoboda, Karel; Denk, Winfried; Kleinfeld, David; Tank, David W. (1997). "In vivo dendritic calcium dynamics in neocortical pyramidal neurons". Nature. 385 (6612): 161–165. Bibcode:1997Natur.385..161S. doi:10.1038/385161a0. ISSN 0028-0836. PMID 8990119. S2CID 4251386. from the original on 2021-05-25. Retrieved 2020-05-20.
  22. ^ Voie, A.H. (1993). "Imaging the intact guinea pig tympanic bulla by orthogonal-plane fluorescence optical sectioning microscopy". Hearing Research. 71 (1–2): 119–128. doi:10.1016/S0378-5955(02)00493-8. ISSN 0378-5955. PMID 12204356. S2CID 12775304.
  23. ^ Greger, K.; J. Swoger; E. H. K. Stelzer (2007). "Basic building units and properties of a fluorescence single plane illumination microscope". Review of Scientific Instruments. 78 (2): 023705–023705–7. Bibcode:2007RScI...78b3705G. doi:10.1063/1.2428277. ISSN 0034-6748. PMID 17578115.
  24. ^ Buytaert, J.A.N.; E. Descamps; D. Adriaens; J.J.J. Dirckx (2012). "The OPFOS Microscopy Family: High-Resolution Optical Sectioning of Biomedical Specimens". Anatomy Research International. 2012: 206238. arXiv:1106.3162. doi:10.1155/2012/206238. ISSN 2090-2743. PMC 3335623. PMID 22567307.
  25. ^ F. J. Duarte, in High Power Dye Lasers (Springer-Verlag, Berlin,1991) Chapter 2.
  26. ^ Duarte FJ (1993), Electro-optical interferometric microdensitometer system, US Patent 5255069 2017-10-13 at the Wayback Machine.
  27. ^ a b Peterson, Mark D.; Hayes, Patrick L.; Martinez, Imee Su; Cass, Laura C.; Achtyl, Jennifer L.; Weiss, Emily A.; Geiger, Franz M. (2011-05-01). "Second harmonic generation imaging with a kHz amplifier [Invited]". Optical Materials Express. 1 (1): 57. Bibcode:2011OMExp...1...57P. doi:10.1364/ome.1.000057.
  28. ^ a b c Macias-Romero, Carlos; Didier, Marie E. P.; Jourdain, Pascal; Marquet, Pierre; Magistretti, Pierre; Tarun, Orly B.; Zubkovs, Vitalijs; Radenovic, Aleksandra; Roke, Sylvie (2014-12-15). "High throughput second harmonic imaging for label-free biological applications". Optics Express. 22 (25): 31102–12. Bibcode:2014OExpr..2231102M. doi:10.1364/oe.22.031102. hdl:10754/563269. PMID 25607059. from the original on 2020-06-20. Retrieved 2019-12-11.
  29. ^ a b Cheng, Li-Chung; Chang, Chia-Yuan; Lin, Chun-Yu; Cho, Keng-Chi; Yen, Wei-Chung; Chang, Nan-Shan; Xu, Chris; Dong, Chen Yuan; Chen, Shean-Jen (2012-04-09). "Spatiotemporal focusing-based widefield multiphoton microscopy for fast optical sectioning". Optics Express. 20 (8): 8939–48. Bibcode:2012OExpr..20.8939C. doi:10.1364/oe.20.008939. PMID 22513605.
  30. ^ Oron, Dan; Tal, Eran; Silberberg, Yaron (2005-03-07). "Scanningless depth-resolved microscopy". Optics Express. 13 (5): 1468–76. Bibcode:2005OExpr..13.1468O. doi:10.1364/opex.13.001468. PMID 19495022.
  31. ^ Ahi, Kiarash; Anwar, Mehdi (2016-05-26). Anwar, Mehdi F; Crowe, Thomas W; Manzur, Tariq (eds.). "Modeling of terahertz images based on x-ray images: a novel approach for verification of terahertz images and identification of objects with fine details beyond terahertz resolution". Terahertz Physics, Devices, and Systems X: Advanced Applications in Industry. Terahertz Physics, Devices, and Systems X: Advanced Applications in Industry and Defense. 9856: 985610. Bibcode:2016SPIE.9856E..10A. doi:10.1117/12.2228685. S2CID 124315172.
  32. ^ Solomon, Chris (2010). Fundamentals of Digital Image Processing. John Wiley & Sons, Ltd. ISBN 978-0-470-84473-1.
  33. ^ Nasse M. J.; Woehl J. C. (2010). "Realistic modeling of the illumination point spread function in confocal scanning optical microscopy". J. Opt. Soc. Am. A. 27 (2): 295–302. Bibcode:2010JOSAA..27..295N. doi:10.1364/JOSAA.27.000295. PMID 20126241.
  34. ^ Wallace W, Schaefer LH, Swedlow JR (2001). "A workingperson's guide to deconvolution in light microscopy". BioTechniques. 31 (5): 1076–8, 1080, 1082 passim. doi:10.2144/01315bi01. PMID 11730015.
  35. ^ Kaufmann Rainer; Müller Patrick; Hildenbrand Georg; Hausmann Michael; Cremer Christoph (2010). "Analysis of Her2/neu membrane protein clusters in different types of breast cancer cells using localization microscopy". Journal of Microscopy. 242 (1): 46–54. CiteSeerX 10.1.1.665.3604. doi:10.1111/j.1365-2818.2010.03436.x. PMID 21118230. S2CID 2119158.
  36. ^ van de Linde S.; Wolter S.; Sauer S. (2011). "Single-molecule Photoswitching and Localization". Aust. J. Chem. 64 (5): 503–511. doi:10.1071/CH10284.
  37. ^ K. Goda; K. K. Tsia; B. Jalali (2009). "Serial time-encoded amplified imaging for real-time observation of fast dynamic phenomena". Nature. 458 (7242): 1145–9. Bibcode:2009Natur.458.1145G. doi:10.1038/nature07980. PMID 19407796. S2CID 4415762.
  38. ^ Jalali, Bahram; Capewell, Dale; Goda, Keisuke; Tsia, Kevin K. (2010-05-10). "Performance of serial time-encoded amplified microscope". Optics Express. 18 (10): 10016–10028. Bibcode:2010OExpr..1810016T. doi:10.1364/OE.18.010016. hdl:10722/91333. ISSN 1094-4087. PMID 20588855. S2CID 8077381.
  39. ^ Kosasih, Felix Utama; Ducati, Caterina (May 2018). "Characterising degradation of perovskite solar cells through in-situ and operando electron microscopy". Nano Energy. 47: 243–256. doi:10.1016/j.nanoen.2018.02.055. from the original on 2021-01-26. Retrieved 2019-06-28.
  40. ^ a b H M Pollock and S G Kazarian, Microspectroscopy in the Mid-Infrared, in Encyclopedia of Analytical Chemistry (Robert A. Meyers, Ed, 1-26 (2014), John Wiley & Sons Ltd,
  41. ^ a b Pollock Hubert M (2014). "Microspectroscopy in the Mid-Infrared". Encyclopedia of Analytical Chemistry. pp. 1–26. doi:10.1002/9780470027318.a5609.pub2. ISBN 9780470027318.
  42. ^ H M Pollock and D A Smith, The use of near-field probes for vibrational spectroscopy and photothermal imaging, in Handbook of vibrational spectroscopy, J.M. Chalmers and P.R. Griffiths (eds), John Wiley & Sons Ltd, Vol. 2, pp. 1472 - 1492 (2002)
  43. ^ Keilmann, Fritz; Hillenbrand, Rainer (2004-04-15). Richards, David; Zayats, Anatoly (eds.). "Near-field microscopy by elastic light scattering from a tip". Philosophical Transactions of the Royal Society of London. Series A: Mathematical, Physical and Engineering Sciences. 362 (1817): 787–805. doi:10.1098/rsta.2003.1347. ISSN 1364-503X. PMID 15306494. S2CID 20211405. from the original on 2020-11-09. Retrieved 2020-11-02.
  44. ^ Duarte FJ (2016). "Tunable laser microscopy". In Duarte FJ (ed.). Tunable Laser Applications (3rd ed.). Boca Raton: CRC Press. pp. 315–328. ISBN 9781482261066.
  45. ^ Thomas JL, Rudolph W (2008). "Biological Microscopy with Ultrashort Laser Pulses". In Duarte FJ (ed.). Tunable Laser Applications (2nd ed.). Boca Raton: CRC Press. pp. 245–80. ISBN 978-1-4200-6009-6.
  46. ^ Solem, J. C. (1983). "X-ray imaging on biological specimens". Proceedings of International Conference on Lasers '83: 635–640. OSTI 5998195.
  47. ^ Solem, J. C. (1982). "High-intensity x-ray holography: An approach to high-resolution snapshot imaging of biological specimens". Los Alamos National Laboratory Technical Report LA-9508-MS. 83: 23581. Bibcode:1982STIN...8323581S. doi:10.2172/7056325. OSTI 7056325. from the original on 2020-08-04. Retrieved 2019-06-28.
  48. ^ Solem, J. C.; Baldwin, G. C. (1982). "Microholography of living organisms". Science. 218 (4569): 229–235. Bibcode:1982Sci...218..229S. doi:10.1126/science.218.4569.229. PMID 17838608. S2CID 32381820.
  49. ^ Solem, J. C.; Chapline, G. F. (1984). "X-ray biomicroholography". Optical Engineering. 23 (2): 193. Bibcode:1984OptEn..23..193S. doi:10.1117/12.7973410.
  50. ^ Solem, J. C. (1985). "Microholography". McGraw-Hill Encyclopedia of Science and Technology.
  51. ^ Solem, J. C. (1984). "X-ray holography of biological specimens". Proceedings of Ninth International Congress on Photobiology, Philadelphia, PA, USA, 3 July 1984 (LA-UR-84-3340, CONF-840783-3): 19. OSTI 6314558.
  52. ^ Solem, J. C. (1986). "Imaging biological specimens with high-intensity soft X-rays". Journal of the Optical Society of America B. 3 (11): 1551–1565. Bibcode:1986JOSAB...3.1551S. doi:10.1364/josab.3.001551.
  53. ^ Haddad, W. S.; Solem, J. C.; Cullen, D.; Boyer, K.; Rhodes, C. K. (1987). "A description of the theory and apparatus for digital reconstruction of Fourier transform holograms", Proceedings of Electronics Imaging '87, Nov. 2-5, 1987, Boston; Journal of Electronic Imaging, (Institute for Graphic Communication, Inc., Boston), p. 693.
  54. ^ Haddad, W. S.; Cullen, D.; Boyer, K.; Rhodes, C. K.; Solem, J. C.; Weinstein, R. S. (1988). "Design for a Fourier-Transform Holographic Microscope". X-Ray Microscopy II. Springer Series in Optical Sciences. Vol. 56. pp. 284–287. doi:10.1007/978-3-540-39246-0_49. ISBN 978-3-662-14490-9.
  55. ^ Haddad, W. S.; Solem, J. C.; Cullen, D.; Boyer, K.; Rhodes, C. K. (1988). "Design for a Fourier-transform holographic microscope incorporating digital image reconstruction". Proceedings of CLEO '88 Conference on Lasers and Electro-Optics, Anaheim, CA, April, 1988; Optical Society of America, OSA Technical Digest. 7: WS4. from the original on 2016-02-02. Retrieved 2016-01-13.
  56. ^ Haddad, W. S.; Cullen, D.; Solem, J. C.; Boyer, K.; Rhodes, C. K. (1988). "X-ray Fourier-transform holographic microscope". Proceedings of the OSA Topical Meeting on Short Wavelength Coherent Radiation: Generation and Applications, September 26–29, 1988, Cape Cod, MA., Falcone, R.; Kirz, J.; Eds. (Optical Society of America, Washington, DC): 284–289. from the original on 2016-01-27. Retrieved 2016-01-13.
  57. ^ Solem, J. C.; Boyer, K.; Haddad, W. S.; Rhodes, C. K. (1990). Prosnitz, Donald (ed.). Prosnitz, D.; ed. SPIE. "Prospects for X-ray holography with free-electron lasers". Free Electron Lasers and Applications. Free-Electron Lasers and Applications. 1227: 105–116. Bibcode:1990SPIE.1227..105S. doi:10.1117/12.18609. S2CID 121807907.
  58. ^ Haddad, W. S.; Cullen, D.; Solem, J. C.; Longworth, J. W.; McPherson, L. A.; Boyer, K.; Rhodes, C. K. (1991). "Fourier-transform holographic microscope". Proceedings of SPIE Electronic Imaging '91, San Jose, CA; International Society for Optics and Photonics. Camera and Input Scanner Systems. 1448 (24): 81–88. Bibcode:1991SPIE.1448...81H. doi:10.1117/12.45347. PMID 20733659. S2CID 120679508.
  59. ^ Haddad, W. S.; Cullen, D.; Solem, J. C.; Longworth, J.; McPherson, A.; Boyer, K.; Rhodes, C. K. (1992). "Fourier-transform holographic microscope". Applied Optics. 31 (24): 4973–4978. Bibcode:1992ApOpt..31.4973H. doi:10.1364/ao.31.004973. PMID 20733659.
  60. ^ Boyer, K.; Solem, J. C.; Longworth, J.; Borisov, A.; Rhodes, C. K. (1996). "Biomedical three-dimensional holographic microimaging at visible, ultraviolet and X-ray wavelengths". Nature Medicine. 2 (8): 939–941. doi:10.1038/nm0896-939. PMID 8705867. S2CID 25231033.
  61. ^ Bell, A. G. (1880). "On the production and reproduction of sound by light". American Journal of Science. s3-20 (118): 305–324. Bibcode:1880AmJS...20..305B. doi:10.2475/ajs.s3-20.118.305. S2CID 130048089. from the original on 2022-11-27. Retrieved 2019-06-28.
  62. ^ Yao, J.; Wang, L. V. (2013). "Photoacoustic Microscopy". Laser Photonics Rev. 7 (5): 1–36. Bibcode:2013LPRv....7..758Y. doi:10.1002/lpor.201200060. PMC 3887369. PMID 24416085.
  63. ^ Langer, G.; Buchegger, B.; Jacak, J.; Klar, T. A.; Berer, T. (2016). "Frequency domain photoacoustic and fluorescence microscopy". Biomedical Optics Express. 7 (7): 2692–702. doi:10.1364/BOE.7.002692. PMC 4948622. PMID 27446698.
  64. ^ Kotrly M (August 2015). "New Possibilities of Using Micrscopic Techniques in Forensic Field". Proceedings of Microscopy & Microanallysis. 21 (S3): 1365–1366. Bibcode:2015MiMic..21S1365K. doi:10.1017/S1431927615007618.
  65. ^ a b Basu S, Millette JR (1986). Electron Microscopy in Forensic Occupational and Environmental Health Sciences. New York: Plenum Press.

Further reading edit

  • Pluta, Maksymilian (1988). Advanced Light Microscopy vol. 1 Principles and Basic Properties. Elsevier. ISBN 978-0-444-98939-0.
  • Pluta, Maksymilian (1989). Advanced Light Microscopy vol. 2 Specialised Methods. Elsevier. ISBN 978-0-444-98918-5.
  • Bradbury, S.; Bracegirdle, B. (1998). Introduction to Light Microscopy. BIOS Scientific Publishers. ISBN 978-0-387-91515-9.
  • Inoue, Shinya (1986). Video Microscopy. Plenum Press. ISBN 978-0-306-42120-4.
  • Cremer, C.; Cremer, T. (1978). (PDF). Microscopica Acta. 81 (1): 31–44. PMID 713859. Archived from the original (PDF) on 2016-03-04. Retrieved 2009-12-22. Theoretical basis of super resolution 4Pi microscopy & design of a confocal laser scanning fluorescence microscope
  • Willis, Randall C. (2007). "Portraits of life, one molecule at a time". Analytical Chemistry. 79 (5): 1785–8. doi:10.1021/ac0718795. PMID 17375393., a feature article on sub-diffraction microscopy from the March 1, 2007 issue of Analytical Chemistry

External links edit

General edit

  • Microscopy glossary, Common terms used in amateur light microscopy.
  • Nikon MicroscopyU Extensive information on light microscopy
  • Olympus Microscopy Microscopy Resource center
  • Carl Zeiss "Microscopy from the very beginning", a step by step tutorial into the basics of microscopy.
  • - A resource with many illustrations elaborating the most common microscopy techniques
  • Manawatu Microscopy - first known collaboration environment for Microscopy and Image Analysis.
  • Audio microscope glossary

Techniques edit

  • Examples of Ratiometric Imaging Work on a Microscope
  • Interactive Fluorescence Dye and Filter Database Carl Zeiss Interactive Fluorescence Dye and Filter Database.
  • New approaches to microscopy Eric Betzig: Beyond the Nobel Prize—New approaches to microscopy.

microscopy, confused, with, microscopic, microscope, technical, field, using, microscopes, view, objects, areas, objects, that, cannot, seen, with, naked, objects, that, within, resolution, range, normal, there, three, well, known, branches, microscopy, optica. Not to be confused with Microscopic or Microscope Microscopy is the technical field of using microscopes to view objects and areas of objects that cannot be seen with the naked eye objects that are not within the resolution range of the normal eye 1 There are three well known branches of microscopy optical electron and scanning probe microscopy along with the emerging field of X ray microscopy citation needed Scanning electron microscope image of pollen false colors Microscopic examination in a biochemical laboratoryOptical microscopy and electron microscopy involve the diffraction reflection or refraction of electromagnetic radiation electron beams interacting with the specimen and the collection of the scattered radiation or another signal in order to create an image This process may be carried out by wide field irradiation of the sample for example standard light microscopy and transmission electron microscopy or by scanning a fine beam over the sample for example confocal laser scanning microscopy and scanning electron microscopy Scanning probe microscopy involves the interaction of a scanning probe with the surface of the object of interest The development of microscopy revolutionized biology gave rise to the field of histology and so remains an essential technique in the life and physical sciences X ray microscopy is three dimensional and non destructive allowing for repeated imaging of the same sample for in situ or 4D studies and providing the ability to see inside the sample being studied before sacrificing it to higher resolution techniques A 3D X ray microscope uses the technique of computed tomography microCT rotating the sample 360 degrees and reconstructing the images CT is typically carried out with a flat panel display A 3D X ray microscope employs a range of objectives e g from 4X to 40X and can also include a flat panel Contents 1 History 2 Optical microscopy 2 1 Limitations 2 2 Techniques 2 2 1 Bright field 2 2 2 Oblique illumination 2 2 3 Dark field 2 2 4 Dispersion staining 2 2 5 Phase contrast 2 2 6 Differential interference contrast 2 2 7 Interference reflection 2 2 8 Fluorescence 2 2 9 Confocal 2 2 10 Two photon microscopy 2 2 11 Single plane illumination microscopy and light sheet fluorescence microscopy 2 2 12 Wide field multiphoton microscopy 2 2 13 Deconvolution 2 3 Sub diffraction techniques 2 4 Serial time encoded amplified microscopy 2 5 Extensions 2 6 Other enhancements 2 7 X ray 3 Electron microscopy 4 Scanning probe microscopy 4 1 Ultrasonic force 5 Ultraviolet microscopy 6 Infrared microscopy 7 Digital holographic microscopy 8 Digital pathology virtual microscopy 9 Laser microscopy 10 Photoacoustic microscopy 11 Amateur microscopy 12 Application in forensic science 13 See also 14 References 15 Further reading 16 External links 16 1 General 16 2 TechniquesHistory edit nbsp Antonie van Leeuwenhoek 1632 1723 The field of microscopy optical microscopy dates back to at least the 17th century Earlier microscopes single lens magnifying glasses with limited magnification date at least as far back as the wide spread use of lenses in eyeglasses in the 13th century 2 but more advanced compound microscopes first appeared in Europe around 1620 3 4 The earliest practitioners of microscopy include Galileo Galilei who found in 1610 that he could close focus his telescope to view small objects close up 5 6 and Cornelis Drebbel who may have invented the compound microscope around 1620 7 8 Antonie van Leeuwenhoek developed a very high magnification simple microscope in the 1670s and is often considered to be the first acknowledged microscopist and microbiologist 9 10 Optical microscopy editSee also Optical microscope nbsp Stereo microscopeOptical or light microscopy involves passing visible light transmitted through or reflected from the sample through a single lens or multiple lenses to allow a magnified view of the sample 11 The resulting image can be detected directly by the eye imaged on a photographic plate or captured digitally The single lens with its attachments or the system of lenses and imaging equipment along with the appropriate lighting equipment sample stage and support makes up the basic light microscope The most recent development is the digital microscope which uses a CCD camera to focus on the exhibit of interest The image is shown on a computer screen so eye pieces are unnecessary 12 Limitations edit Limitations of standard optical microscopy bright field microscopy lie in three areas The technique can only image dark or strongly refracting objects effectively There is a diffraction limited resolution depending on incident wavelength in visible range the resolution of optical microscopy is limited to approximately 0 2 micrometres see microscope and the practical magnification limit to 1500x 13 Out of focus light from points outside the focal plane reduces image clarity 14 Live cells in particular generally lack sufficient contrast to be studied successfully since the internal structures of the cell are colorless and transparent The most common way to increase contrast is to stain the structures with selective dyes but this often involves killing and fixing the sample 15 Staining may also introduce artifacts which are apparent structural details that are caused by the processing of the specimen and are thus not features of the specimen In general these techniques make use of differences in the refractive index of cell structures Bright field microscopy is comparable to looking through a glass window one sees not the glass but merely the dirt on the glass There is a difference as glass is a denser material and this creates a difference in phase of the light passing through The human eye is not sensitive to this difference in phase but clever optical solutions have been devised to change this difference in phase into a difference in amplitude light intensity citation needed Techniques edit Main article Optical microscope To improve specimen contrast or highlight structures in a sample special techniques must be used A huge selection of microscopy techniques are available to increase contrast or label a sample Four examples of transillumination techniques used to generate contrast in a sample of tissue paper 1 559 mm pixel nbsp Bright field illumination sample contrast comes from absorbance of light in the sample nbsp Cross polarized light illumination sample contrast comes from rotation of polarized light through the sample nbsp Dark field illumination sample contrast comes from light scattered by the sample nbsp Phase contrast illumination sample contrast comes from interference of different path lengths of light through the sampleBright field edit Main article Bright field microscopy Bright field microscopy is the simplest of all the light microscopy techniques Sample illumination is via transmitted white light i e illuminated from below and observed from above Limitations include low contrast of most biological samples and low apparent resolution due to the blur of out of focus material The simplicity of the technique and the minimal sample preparation required are significant advantages citation needed Oblique illumination edit The use of oblique from the side illumination gives the image a three dimensional appearance and can highlight otherwise invisible features A more recent technique based on this method is Hoffmann s modulation contrast a system found on inverted microscopes for use in cell culture Oblique illumination enhances contrast even in clear specimens however because light enters off axis the position of an object will appear to shift as the focus is changed This limitation makes techniques like optical sectioning or accurate measurement on the z axis impossible Dark field edit Main article Dark field microscopy Dark field microscopy is a technique for improving the contrast of unstained transparent specimens 16 Dark field illumination uses a carefully aligned light source to minimize the quantity of directly transmitted unscattered light entering the image plane collecting only the light scattered by the sample Dark field can dramatically improve image contrast especially of transparent objects while requiring little equipment setup or sample preparation However the technique suffers from low light intensity in the final image of many biological samples and continues to be affected by low apparent resolution nbsp A diatom under Rheinberg illuminationRheinberg illumination is a variant of dark field illumination in which transparent colored filters are inserted just before the condenser so that light rays at high aperture are differently colored than those at low aperture i e the background to the specimen may be blue while the object appears self luminous red Other color combinations are possible but their effectiveness is quite variable 17 Dispersion staining edit Main article Dispersion staining Dispersion staining is an optical technique that results in a colored image of a colorless object This is an optical staining technique and requires no stains or dyes to produce a color effect There are five different microscope configurations used in the broader technique of dispersion staining They include brightfield Becke line oblique darkfield phase contrast and objective stop dispersion staining Phase contrast edit nbsp Phase contrast light micrograph of undecalcified hyaline cartilage showing chondrocytes and organelles lacunae and extracellular matrixMain article Phase contrast microscopy In electron microscopy Phase contrast imagingMore sophisticated techniques will show proportional differences in optical density Phase contrast is a widely used technique that shows differences in refractive index as difference in contrast It was developed by the Dutch physicist Frits Zernike in the 1930s for which he was awarded the Nobel Prize in 1953 The nucleus in a cell for example will show up darkly against the surrounding cytoplasm Contrast is excellent however it is not for use with thick objects Frequently a halo is formed even around small objects which obscures detail The system consists of a circular annulus in the condenser which produces a cone of light This cone is superimposed on a similar sized ring within the phase objective Every objective has a different size ring so for every objective another condenser setting has to be chosen The ring in the objective has special optical properties it first of all reduces the direct light in intensity but more importantly it creates an artificial phase difference of about a quarter wavelength As the physical properties of this direct light have changed interference with the diffracted light occurs resulting in the phase contrast image One disadvantage of phase contrast microscopy is halo formation halo light ring Differential interference contrast edit Main article Differential interference contrast microscopy Superior and much more expensive is the use of interference contrast Differences in optical density will show up as differences in relief A nucleus within a cell will actually show up as a globule in the most often used differential interference contrast system according to Georges Nomarski However it has to be kept in mind that this is an optical effect and the relief does not necessarily resemble the true shape Contrast is very good and the condenser aperture can be used fully open thereby reducing the depth of field and maximizing resolution The system consists of a special prism Nomarski prism Wollaston prism in the condenser that splits light in an ordinary and an extraordinary beam The spatial difference between the two beams is minimal less than the maximum resolution of the objective After passage through the specimen the beams are reunited by a similar prism in the objective In a homogeneous specimen there is no difference between the two beams and no contrast is being generated However near a refractive boundary say a nucleus within the cytoplasm the difference between the ordinary and the extraordinary beam will generate a relief in the image Differential interference contrast requires a polarized light source to function two polarizing filters have to be fitted in the light path one below the condenser the polarizer and the other above the objective the analyzer Note In cases where the optical design of a microscope produces an appreciable lateral separation of the two beams we have the case of classical interference microscopy which does not result in relief images but can nevertheless be used for the quantitative determination of mass thicknesses of microscopic objects Interference reflection edit Main article Interference reflection microscopy An additional technique using interference is interference reflection microscopy also known as reflected interference contrast or RIC It relies on cell adhesion to the slide to produce an interference signal If there is no cell attached to the glass there will be no interference Interference reflection microscopy can be obtained by using the same elements used by DIC but without the prisms Also the light that is being detected is reflected and not transmitted as it is when DIC is employed Fluorescence edit nbsp Images may also contain artifacts This is a confocal laser scanning fluorescence micrograph of thale cress anther part of stamen The picture shows among other things a nice red flowing collar like structure just below the anther However an intact thale cress stamen does not have such collar this is a fixation artifact the stamen has been cut below the picture frame and epidermis upper layer of cells of stamen stalk has peeled off forming a non characteristic structure Photo Heiti Paves from Tallinn University of Technology Main article Fluorescence microscopy When certain compounds are illuminated with high energy light they emit light of a lower frequency This effect is known as fluorescence Often specimens show their characteristic autofluorescence image based on their chemical makeup This method is of critical importance in the modern life sciences as it can be extremely sensitive allowing the detection of single molecules Many fluorescent dyes can be used to stain structures or chemical compounds One powerful method is the combination of antibodies coupled to a fluorophore as in immunostaining Examples of commonly used fluorophores are fluorescein or rhodamine The antibodies can be tailor made for a chemical compound For example one strategy often in use is the artificial production of proteins based on the genetic code DNA These proteins can then be used to immunize rabbits forming antibodies which bind to the protein The antibodies are then coupled chemically to a fluorophore and used to trace the proteins in the cells under study Highly efficient fluorescent proteins such as the green fluorescent protein GFP have been developed using the molecular biology technique of gene fusion a process that links the expression of the fluorescent compound to that of the target protein This combined fluorescent protein is in general non toxic to the organism and rarely interferes with the function of the protein under study Genetically modified cells or organisms directly express the fluorescently tagged proteins which enables the study of the function of the original protein in vivo Growth of protein crystals results in both protein and salt crystals Both are colorless and microscopic Recovery of the protein crystals requires imaging which can be done by the intrinsic fluorescence of the protein or by using transmission microscopy Both methods require an ultraviolet microscope as protein absorbs light at 280 nm Protein will also fluorescence at approximately 353 nm when excited with 280 nm light 18 Since fluorescence emission differs in wavelength color from the excitation light an ideal fluorescent image shows only the structure of interest that was labeled with the fluorescent dye This high specificity led to the widespread use of fluorescence light microscopy in biomedical research Different fluorescent dyes can be used to stain different biological structures which can then be detected simultaneously while still being specific due to the individual color of the dye To block the excitation light from reaching the observer or the detector filter sets of high quality are needed These typically consist of an excitation filter selecting the range of excitation wavelengths a dichroic mirror and an emission filter blocking the excitation light Most fluorescence microscopes are operated in the Epi illumination mode illumination and detection from one side of the sample to further decrease the amount of excitation light entering the detector See also total internal reflection fluorescence microscope Neuroscience Confocal edit Main article Confocal microscopy Confocal laser scanning microscopy uses a focused laser beam e g 488 nm that is scanned across the sample to excite fluorescence in a point by point fashion The emitted light is directed through a pinhole to prevent out of focus light from reaching the detector typically a photomultiplier tube The image is constructed in a computer plotting the measured fluorescence intensities according to the position of the excitation laser Compared to full sample illumination confocal microscopy gives slightly higher lateral resolution and significantly improves optical sectioning axial resolution Confocal microscopy is therefore commonly used where 3D structure is important A subclass of confocal microscopes are spinning disc microscopes which are able to scan multiple points simultaneously across the sample A corresponding disc with pinholes rejects out of focus light The light detector in a spinning disc microscope is a digital camera typically EM CCD or sCMOS Two photon microscopy edit A two photon microscope is also a laser scanning microscope but instead of UV blue or green laser light a pulsed infrared laser is used for excitation Only in the tiny focus of the laser is the intensity high enough to generate fluorescence by two photon excitation which means that no out of focus fluorescence is generated and no pinhole is necessary to clean up the image 19 This allows imaging deep in scattering tissue where a confocal microscope would not be able to collect photons efficiently 20 Two photon microscopes with wide field detection are frequently used for functional imaging e g calcium imaging in brain tissue 21 They are marketed as Multiphoton microscopes by several companies although the gains of using 3 photon instead of 2 photon excitation are marginal Single plane illumination microscopy and light sheet fluorescence microscopy edit Main article Light sheet fluorescence microscopy Using a plane of light formed by focusing light through a cylindrical lens at a narrow angle or by scanning a line of light in a plane perpendicular to the axis of objective high resolution optical sections can be taken 22 23 24 Single plane illumination or light sheet illumination is also accomplished using beam shaping techniques incorporating multiple prism beam expanders 25 26 The images are captured by CCDs These variants allow very fast and high signal to noise ratio image capture Wide field multiphoton microscopy edit Main article Wide field multiphoton microscopyWide field multiphoton microscopy 27 28 29 30 refers to an optical non linear imaging technique in which a large area of the object is illuminated and imaged without the need for scanning High intensities are required to induce non linear optical processes such as two photon fluorescence or second harmonic generation In scanning multiphoton microscopes the high intensities are achieved by tightly focusing the light and the image is obtained by beam scanning In wide field multiphoton microscopy the high intensities are best achieved using an optically amplified pulsed laser source to attain a large field of view 100 µm 27 28 29 The image in this case is obtained as a single frame with a CCD camera without the need of scanning making the technique particularly useful to visualize dynamic processes simultaneously across the object of interest With wide field multiphoton microscopy the frame rate can be increased up to a 1000 fold compared to multiphoton scanning microscopy 28 In scattering tissue however image quality rapidly degrades with increasing depth Deconvolution edit Fluorescence microscopy is a powerful technique to show specifically labeled structures within a complex environment and to provide three dimensional information of biological structures However this information is blurred by the fact that upon illumination all fluorescently labeled structures emit light irrespective of whether they are in focus or not So an image of a certain structure is always blurred by the contribution of light from structures that are out of focus This phenomenon results in a loss of contrast especially when using objectives with a high resolving power typically oil immersion objectives with a high numerical aperture nbsp Mathematically modeled Point Spread Function of a pulsed THz laser imaging system 31 However blurring is not caused by random processes such as light scattering but can be well defined by the optical properties of the image formation in the microscope imaging system If one considers a small fluorescent light source essentially a bright spot light coming from this spot spreads out further from our perspective as the spot becomes more out of focus Under ideal conditions this produces an hourglass shape of this point source in the third axial dimension This shape is called the point spread function PSF of the microscope imaging system Since any fluorescence image is made up of a large number of such small fluorescent light sources the image is said to be convolved by the point spread function The mathematically modeled PSF of a terahertz laser pulsed imaging system is shown on the right The output of an imaging system can be described using the equation s x y P S F x y o x y n displaystyle s x y PSF x y o x y n nbsp Where n is the additive noise 32 Knowing this point spread function 33 means that it is possible to reverse this process to a certain extent by computer based methods commonly known as deconvolution microscopy 34 There are various algorithms available for 2D or 3D deconvolution They can be roughly classified in nonrestorative and restorative methods While the nonrestorative methods can improve contrast by removing out of focus light from focal planes only the restorative methods can actually reassign light to its proper place of origin Processing fluorescent images in this manner can be an advantage over directly acquiring images without out of focus light such as images from confocal microscopy because light signals otherwise eliminated become useful information For 3D deconvolution one typically provides a series of images taken from different focal planes called a Z stack plus the knowledge of the PSF which can be derived either experimentally or theoretically from knowing all contributing parameters of the microscope Sub diffraction techniques edit Main article Super Resolution microscopy nbsp Example of super resolution microscopy Image of Her3 and Her2 target of the breast cancer drug Trastuzumab within a cancer cell A multitude of super resolution microscopy techniques have been developed in recent times which circumvent the diffraction limit This is mostly achieved by imaging a sufficiently static sample multiple times and either modifying the excitation light or observing stochastic changes in the image The deconvolution methods described in the previous section which removes the PSF induced blur and assigns a mathematically correct origin of light are used albeit with slightly different understanding of what the value of a pixel mean Assuming most of the time one single fluorophore contributes to one single blob on one single taken image the blobs in the images can be replaced with their calculated position vastly improving resolution to well below the diffraction limit To realize such assumption Knowledge of and chemical control over fluorophore photophysics is at the core of these techniques by which resolutions of 20 nanometers are obtained 35 36 Serial time encoded amplified microscopy edit Main article Serial time encoded amplified microscopy Serial time encoded amplified microscopy STEAM is an imaging method that provides ultrafast shutter speed and frame rate by using optical image amplification to circumvent the fundamental trade off between sensitivity and speed and a single pixel photodetector to eliminate the need for a detector array and readout time limitations 37 The method is at least 1000 times faster than the state of the art CCD and CMOS cameras Consequently it is potentially useful for scientific industrial and biomedical applications that require high image acquisition rates including real time diagnosis and evaluation of shockwaves microfluidics MEMS and laser surgery 38 Extensions edit Most modern instruments provide simple solutions for micro photography and image recording electronically However such capabilities are not always present and the more experienced microscopist may prefer a hand drawn image to a photograph This is because a microscopist with knowledge of the subject can accurately convert a three dimensional image into a precise two dimensional drawing In a photograph or other image capture system however only one thin plane is ever in good focus citation needed The creation of accurate micrographs requires a microscopical technique using a monocular eyepiece It is essential that both eyes are open and that the eye that is not observing down the microscope is instead concentrated on a sheet of paper on the bench besides the microscope With practice and without moving the head or eyes it is possible to accurately trace the observed shapes by simultaneously seeing the pencil point in the microscopical image citation needed It is always less tiring to observe with the microscope focused so that the image is seen at infinity and with both eyes open at all times citation needed Other enhancements edit Main article stereomicroscope Microspectroscopy spectroscopy with a microscope X ray edit Main article X ray microscopy As resolution depends on the wavelength of the light Electron microscopy has been developed since the 1930s that use electron beams instead of light Because of the much smaller wavelength of the electron beam resolution is far higher Though less common X ray microscopy has also been developed since the late 1940s The resolution of X ray microscopy lies between that of light microscopy and electron microscopy Electron microscopy editMain article Electron microscope Until the invention of sub diffraction microscopy the wavelength of the light limited the resolution of traditional microscopy to around 0 2 micrometers In order to gain higher resolution the use of an electron beam with a far smaller wavelength is used in electron microscopes Transmission electron microscopy TEM is quite similar to the compound light microscope by sending an electron beam through a very thin slice of the specimen The resolution limit in 2005 was around 0 05 dubious discuss nanometer and has not increased appreciably since that time Scanning electron microscopy SEM visualizes details on the surfaces of specimens and gives a very nice 3D view It gives results much like those of the stereo light microscope The best resolution for SEM in 2011 was 0 4 nanometer Electron microscopes equipped for X ray spectroscopy can provide qualitative and quantitative elemental analysis This type of electron microscope also known as analytical electron microscope can be a very powerful tool for investigation of nanomaterials 39 Scanning probe microscopy editMain article Scanning probe microscopy This is a sub diffraction technique Examples of scanning probe microscopes are the atomic force microscope AFM the scanning tunneling microscope the photonic force microscope and the recurrence tracking microscope All such methods use the physical contact of a solid probe tip to scan the surface of an object which is supposed to be almost flat Ultrasonic force edit Ultrasonic force microscopy UFM has been developed in order to improve the details and image contrast on flat areas of interest where AFM images are limited in contrast The combination of AFM UFM allows a near field acoustic microscopic image to be generated The AFM tip is used to detect the ultrasonic waves and overcomes the limitation of wavelength that occurs in acoustic microscopy By using the elastic changes under the AFM tip an image of much greater detail than the AFM topography can be generated Ultrasonic force microscopy allows the local mapping of elasticity in atomic force microscopy by the application of ultrasonic vibration to the cantilever or sample To analyze the results of ultrasonic force microscopy in a quantitative fashion a force distance curve measurement is done with ultrasonic vibration applied to the cantilever base and the results are compared with a model of the cantilever dynamics and tip sample interaction based on the finite difference technique Ultraviolet microscopy editUltraviolet microscopes have two main purposes The first is to use the shorter wavelength of ultraviolet electromagnetic energy to improve the image resolution beyond that of the diffraction limit of standard optical microscopes This technique is used for non destructive inspection of devices with very small features such as those found in modern semiconductors The second application for UV microscopes is contrast enhancement where the response of individual samples is enhanced relative to their surrounding due to the interaction of light with the molecules within the sample itself One example is in the growth of protein crystals Protein crystals are formed in salt solutions As salt and protein crystals are both formed in the growth process and both are commonly transparent to the human eye they cannot be differentiated with a standard optical microscope As the tryptophan of protein absorbs light at 280 nm imaging with a UV microscope with 280 nm bandpass filters makes it simple to differentiate between the two types of crystals The protein crystals appear dark while the salt crystals are transparent Infrared microscopy editThe term infrared microscopy refers to microscopy performed at infrared wavelengths In the typical instrument configuration a Fourier Transform Infrared Spectrometer FTIR is combined with an optical microscope and an infrared detector The infrared detector can be a single point detector a linear array or a 2D focal plane array FTIR provides the ability to perform chemical analysis via infrared spectroscopy and the microscope and point or array detector enable this chemical analysis to be spatially resolved i e performed at different regions of the sample As such the technique is also called infrared microspectroscopy 40 41 An alternative architecture called Laser Direct Infrared LDIR Imaging involves the combination of a tuneable infrared light source and single point detector on a flying objective This technique is frequently used for infrared chemical imaging where the image contrast is determined by the response of individual sample regions to particular IR wavelengths selected by the user usually specific IR absorption bands and associated molecular resonances A key limitation of conventional infrared microspectroscopy is that the spatial resolution is diffraction limited Specifically the spatial resolution is limited to a figure related to the wavelength of the light For practical IR microscopes the spatial resolution is limited to 1 3x the wavelength depending on the specific technique and instrument used For mid IR wavelengths this sets a practical spatial resolution limit of 3 30 mm IR versions of sub diffraction microscopy also exist 40 41 These include IR Near field scanning optical microscope NSOM 42 photothermal microspectroscopy and atomic force microscope based infrared spectroscopy AFM IR as well as scattering type Scanning Near field Optical Microscopy s SNOM 43 amp nano FTIR that provide nanoscale spatial resolution at IR wavelengths Digital holographic microscopy edit nbsp Human cells imaged by DHM phase shift left and phase contrast microscopy right Main article Digital holographic microscopy In digital holographic microscopy DHM interfering wave fronts from a coherent monochromatic light source are recorded on a sensor The image is digitally reconstructed by a computer from the recorded hologram Besides the ordinary bright field image a phase shift image is created DHM can operate both in reflection and transmission mode In reflection mode the phase shift image provides a relative distance measurement and thus represents a topography map of the reflecting surface In transmission mode the phase shift image provides a label free quantitative measurement of the optical thickness of the specimen Phase shift images of biological cells are very similar to images of stained cells and have successfully been analyzed by high content analysis software A unique feature of DHM is the ability to adjust focus after the image is recorded since all focus planes are recorded simultaneously by the hologram This feature makes it possible to image moving particles in a volume or to rapidly scan a surface Another attractive feature is The ability of DHM to use low cost optics by correcting optical aberrations by software Digital pathology virtual microscopy editMain article Digital pathology Digital pathology is an image based information environment enabled by computer technology that allows for the management of information generated from a digital slide Digital pathology is enabled in part by virtual microscopy which is the practice of converting glass slides into digital slides that can be viewed managed and analyzed Laser microscopy editLaser microscopy is a rapidly growing field that uses laser illumination sources in various forms of microscopy 44 For instance laser microscopy focused on biological applications uses ultrashort pulse lasers in a number of techniques labeled as nonlinear microscopy saturation microscopy and two photon excitation microscopy 45 High intensity short pulse laboratory x ray lasers have been under development for several years When this technology comes to fruition it will be possible to obtain magnified three dimensional images of elementary biological structures in the living state at a precisely defined instant For optimum contrast between water and protein and for best sensitivity and resolution the laser should be tuned near the nitrogen line at about 0 3 nanometers Resolution will be limited mainly by the hydrodynamic expansion that occurs while the necessary number of photons is being registered 46 Thus while the specimen is destroyed by the exposure its configuration can be captured before it explodes 47 48 49 50 51 52 excessive citations Scientists have been working on practical designs and prototypes for x ray holographic microscopes despite the prolonged development of the appropriate laser 53 54 55 56 57 58 59 60 excessive citations Photoacoustic microscopy editMain article Photoacoustic microscopy nbsp Photoacoustic micrograph of human red blood cells A microscopy technique relying on the photoacoustic effect 61 i e the generation of ultra sound caused by light absorption A focused and intensity modulated laser beam is raster scanned over a sample The generated ultra sound is detected via an ultrasound transducer Commonly piezoelectric ultrasound transducers are employed 62 The image contrast is related to the sample s absorption coefficient a displaystyle alpha nbsp This is in contrast to bright or dark field microscopy where the image contrast is due to transmittance or scattering In principle the contrast of fluorescence microscopy is proportional to the sample s absorption too However in fluorescence microscopy the fluorescence quantum yield h f l displaystyle eta fl nbsp needs to be unequal to zero in order that a signal can be detected In photoacoustic microscopy however every absorbing substance gives a photoacoustic signal P A displaystyle PA nbsp which is proportional toP A a G 1 h f l E g E l a s e r E g displaystyle PA propto alpha Gamma 1 eta fl E g E laser E g nbsp Here G displaystyle Gamma nbsp is the Gruneisen coefficient E l a s e r displaystyle E laser nbsp is the laser s photon energy and E g displaystyle E g nbsp is the sample s band gap energy Therefore photoacoustic microscopy seems well suited as a complementary technique to fluorescence microscopy as a high fluorescence quantum yield leads to high fluorescence signals and a low fluorescence quantum yield leads to high photoacoustic signals Neglecting non linear effects the lateral resolution dx is limited by the Abbe diffraction limit d x l 2 N A displaystyle dx lambda 2 NA nbsp where l displaystyle lambda nbsp is the wavelength of the excitation laser and NA is the numerical aperture of the objective lens The Abbe diffraction limit holds if the incoming wave front is parallel In reality however the laser beam profile is Gaussian Therefore in order to the calculate the achievable resolution formulas for truncated Gaussian beams have to be used 63 Amateur microscopy editAmateur Microscopy is the investigation and observation of biological and non biological specimens for recreational purposes Collectors of minerals insects seashells and plants may use microscopes as tools to uncover features that help them classify their collected items Other amateurs may be interested in observing the life found in pond water and of other samples Microscopes may also prove useful for the water quality assessment for people that keep a home aquarium Photographic documentation and drawing of the microscopic images are additional pleasures There are competitions for photomicrograph art Participants of this pastime may use commercially prepared microscopic slides or prepare their own slides While microscopy is a central tool in the documentation of biological specimens it is often insufficient to justify the description of a new species based on microscopic investigations alone Often genetic and biochemical tests are necessary to confirm the discovery of a new species A laboratory and access to academic literature is a necessity There is however one advantage that amateurs have above professionals time to explore their surroundings Often advanced amateurs team up with professionals to validate their findings and possibly describe new species In the late 1800s amateur microscopy became a popular hobby in the United States and Europe Several professional amateurs were being paid for their sampling trips and microscopic explorations by philanthropists to keep them amused on the Sunday afternoon e g the diatom specialist A Grunow being paid by among others a Belgian industrialist Professor John Phin published Practical Hints on the Selection and Use of the Microscope Second Edition 1878 and was also the editor of the American Journal of Microscopy Examples of amateur microscopy images nbsp house bee Mouth 100X nbsp Rice Stem cs 400X nbsp Rabbit Testis 100X nbsp Fern Prothallium 400XApplication in forensic science editMicroscopy has applications in the forensic sciences 64 The microscope can detect resolve and image the smallest items of evidence often without any alteration or destruction The microscope is used to identify and compare fibers hairs soils and dust etc In ink markings blood stains or bullets no specimen treatment is required and the evidence shows directly from microscopical examination For traces of particular matter the sample preparation must be done before microscopical examination occurs clarification needed Light microscopes are the most use in forensics using photons to form images microscopes which are most applicable for examining forensic specimens are as follows 65 1 The compound microscope2 The comparison microscope3 The stereoscopic microscope4 The polarizing microscope5 The micro spectrophotometerThis diversity of the types of microscopes in forensic applications comes mainly from their magnification ranges which are 1 1200X 50 30 000X and 500 250 000X for the optical microscopy SEM and TEM respectively 65 See also editAcronyms in microscopy Digital microscope Digital pathology Imaging cycler microscopy Infinity correction Interferometric microscopy Kohler illumination List of microscopists Microdensitometer Timeline of microscope technology Two photon excitation microscopy USB microscope Van Leeuwenhoek s microscopes Van Leeuwenhoek s microscopic discovery of microbial life microorganisms References edit The University of Edinburgh March 6 2018 What is Microscopy The University of Edinburgh Archived from the original on April 10 2018 Retrieved April 9 2018 Atti Della Fondazione Giorgio Ronchi E Contributi Dell Istituto Nazionale Di Ottica Volume 30 La Fondazione 1975 page 554 Albert Van Helden Sven Dupre Rob van Gent 2010 The Origins of the Telescope Amsterdam University Press p 24 ISBN 978 90 6984 615 6 Archived from the original on 15 February 2017 William Rosenthal Spectacles and Other Vision Aids A History and Guide to Collecting Norman Publishing 1996 page 391 392 Robert D Huerta Giants of Delft Johannes Vermeer and the Natural Philosophers the Parallel Search for Knowledge During the Age of Discovery Bucknell University Press 2003 page 126 A Mark Smith From Sight to Light The Passage from Ancient to Modern Optics University of Chicago Press 2014 page 387 Raymond J Seeger Men of Physics Galileo Galilei His Life and His Works Elsevier 2016 page 24 J William Rosenthal Spectacles and Other Vision Aids A History and Guide to Collecting Norman Publishing 1996 page 391 Ford Brian J 1992 From Dilettante to Diligent Experimenter a Reappraisal of Leeuwenhoek as microscopist and investigator Biology History 5 3 Archived from the original on 2021 04 19 Retrieved 2021 04 16 Lane Nick 6 March 2015 The Unseen World Reflections on Leeuwenhoek 1677 Concerning Little Animal Philos Trans R Soc Lond B Biol Sci 2015 Apr 370 1666 20140344 doi 10 1098 rstb 2014 0344 Abramowitz M Davidson MW 2007 Introduction to Microscopy Molecular Expressions Archived from the original on 2021 04 27 Retrieved 2007 08 22 Microscope Compound Numerical Aperture and Immersion Microscopy Britannica www britannica com Archived from the original on 2023 06 22 Retrieved 2023 06 22 Abbe E 1874 A contribution to the theory of the microscope and the nature of microscopic vision Proc Bristol Nat Soc 1 200 261 Pawley JB editor 2006 Handbook of Biological Confocal Microscopy 3rd ed Berlin Springer ISBN 0 387 25921 X Lakowicz Joseph R 1999 Principles Of Fluorescence Spectroscopy New York Kluwer Academic Plenum Abramowitz M Davidson MW 2003 08 01 Darkfield Illumination Archived from the original on 2015 11 23 Retrieved 2008 10 21 Abramowitz M Davidson MW 2003 08 01 Rheinberg Illumination Archived from the original on 2008 09 29 Retrieved 2008 10 21 Gill Harindarpal January 2010 Evaluating the efficacy of tryptophan fluorescence and absorbance as a selection tool for identifying protein crystals Acta Crystallographica F66 Pt 3 364 372 doi 10 1107 S1744309110002022 PMC 2833058 PMID 20208182 Denk Winfried Svoboda Karel March 1997 Photon Upmanship Why Multiphoton Imaging Is More than a Gimmick Neuron 18 3 351 357 doi 10 1016 S0896 6273 00 81237 4 PMID 9115730 S2CID 2414593 Denk W Delaney K R Gelperin A Kleinfeld D Strowbridge B W Tank D W Yuste R 1994 Anatomical and functional imaging of neurons using 2 photon laser scanning microscopy Journal of Neuroscience Methods 54 2 151 162 doi 10 1016 0165 0270 94 90189 9 PMID 7869748 S2CID 3772937 Svoboda Karel Denk Winfried Kleinfeld David Tank David W 1997 In vivo dendritic calcium dynamics in neocortical pyramidal neurons Nature 385 6612 161 165 Bibcode 1997Natur 385 161S doi 10 1038 385161a0 ISSN 0028 0836 PMID 8990119 S2CID 4251386 Archived from the original on 2021 05 25 Retrieved 2020 05 20 Voie A H 1993 Imaging the intact guinea pig tympanic bulla by orthogonal plane fluorescence optical sectioning microscopy Hearing Research 71 1 2 119 128 doi 10 1016 S0378 5955 02 00493 8 ISSN 0378 5955 PMID 12204356 S2CID 12775304 Greger K J Swoger E H K Stelzer 2007 Basic building units and properties of a fluorescence single plane illumination microscope Review of Scientific Instruments 78 2 023705 023705 7 Bibcode 2007RScI 78b3705G doi 10 1063 1 2428277 ISSN 0034 6748 PMID 17578115 Buytaert J A N E Descamps D Adriaens J J J Dirckx 2012 The OPFOS Microscopy Family High Resolution Optical Sectioning of Biomedical Specimens Anatomy Research International 2012 206238 arXiv 1106 3162 doi 10 1155 2012 206238 ISSN 2090 2743 PMC 3335623 PMID 22567307 F J Duarte in High Power Dye Lasers Springer Verlag Berlin 1991 Chapter 2 Duarte FJ 1993 Electro optical interferometric microdensitometer system US Patent 5255069 Archived 2017 10 13 at the Wayback Machine a b Peterson Mark D Hayes Patrick L Martinez Imee Su Cass Laura C Achtyl Jennifer L Weiss Emily A Geiger Franz M 2011 05 01 Second harmonic generation imaging with a kHz amplifier Invited Optical Materials Express 1 1 57 Bibcode 2011OMExp 1 57P doi 10 1364 ome 1 000057 a b c Macias Romero Carlos Didier Marie E P Jourdain Pascal Marquet Pierre Magistretti Pierre Tarun Orly B Zubkovs Vitalijs Radenovic Aleksandra Roke Sylvie 2014 12 15 High throughput second harmonic imaging for label free biological applications Optics Express 22 25 31102 12 Bibcode 2014OExpr 2231102M doi 10 1364 oe 22 031102 hdl 10754 563269 PMID 25607059 Archived from the original on 2020 06 20 Retrieved 2019 12 11 a b Cheng Li Chung Chang Chia Yuan Lin Chun Yu Cho Keng Chi Yen Wei Chung Chang Nan Shan Xu Chris Dong Chen Yuan Chen Shean Jen 2012 04 09 Spatiotemporal focusing based widefield multiphoton microscopy for fast optical sectioning Optics Express 20 8 8939 48 Bibcode 2012OExpr 20 8939C doi 10 1364 oe 20 008939 PMID 22513605 Oron Dan Tal Eran Silberberg Yaron 2005 03 07 Scanningless depth resolved microscopy Optics Express 13 5 1468 76 Bibcode 2005OExpr 13 1468O doi 10 1364 opex 13 001468 PMID 19495022 Ahi Kiarash Anwar Mehdi 2016 05 26 Anwar Mehdi F Crowe Thomas W Manzur Tariq eds Modeling of terahertz images based on x ray images a novel approach for verification of terahertz images and identification of objects with fine details beyond terahertz resolution Terahertz Physics Devices and Systems X Advanced Applications in Industry Terahertz Physics Devices and Systems X Advanced Applications in Industry and Defense 9856 985610 Bibcode 2016SPIE 9856E 10A doi 10 1117 12 2228685 S2CID 124315172 Solomon Chris 2010 Fundamentals of Digital Image Processing John Wiley amp Sons Ltd ISBN 978 0 470 84473 1 Nasse M J Woehl J C 2010 Realistic modeling of the illumination point spread function in confocal scanning optical microscopy J Opt Soc Am A 27 2 295 302 Bibcode 2010JOSAA 27 295N doi 10 1364 JOSAA 27 000295 PMID 20126241 Wallace W Schaefer LH Swedlow JR 2001 A workingperson s guide to deconvolution in light microscopy BioTechniques 31 5 1076 8 1080 1082 passim doi 10 2144 01315bi01 PMID 11730015 Kaufmann Rainer Muller Patrick Hildenbrand Georg Hausmann Michael Cremer Christoph 2010 Analysis of Her2 neu membrane protein clusters in different types of breast cancer cells using localization microscopy Journal of Microscopy 242 1 46 54 CiteSeerX 10 1 1 665 3604 doi 10 1111 j 1365 2818 2010 03436 x PMID 21118230 S2CID 2119158 van de Linde S Wolter S Sauer S 2011 Single molecule Photoswitching and Localization Aust J Chem 64 5 503 511 doi 10 1071 CH10284 K Goda K K Tsia B Jalali 2009 Serial time encoded amplified imaging for real time observation of fast dynamic phenomena Nature 458 7242 1145 9 Bibcode 2009Natur 458 1145G doi 10 1038 nature07980 PMID 19407796 S2CID 4415762 Jalali Bahram Capewell Dale Goda Keisuke Tsia Kevin K 2010 05 10 Performance of serial time encoded amplified microscope Optics Express 18 10 10016 10028 Bibcode 2010OExpr 1810016T doi 10 1364 OE 18 010016 hdl 10722 91333 ISSN 1094 4087 PMID 20588855 S2CID 8077381 Kosasih Felix Utama Ducati Caterina May 2018 Characterising degradation of perovskite solar cells through in situ and operando electron microscopy Nano Energy 47 243 256 doi 10 1016 j nanoen 2018 02 055 Archived from the original on 2021 01 26 Retrieved 2019 06 28 a b H M Pollock and S G Kazarian Microspectroscopy in the Mid Infrared in Encyclopedia of Analytical Chemistry Robert A Meyers Ed 1 26 2014 John Wiley amp Sons Ltd a b Pollock Hubert M 2014 Microspectroscopy in the Mid Infrared Encyclopedia of Analytical Chemistry pp 1 26 doi 10 1002 9780470027318 a5609 pub2 ISBN 9780470027318 H M Pollock and D A Smith The use of near field probes for vibrational spectroscopy and photothermal imaging in Handbook of vibrational spectroscopy J M Chalmers and P R Griffiths eds John Wiley amp Sons Ltd Vol 2 pp 1472 1492 2002 Keilmann Fritz Hillenbrand Rainer 2004 04 15 Richards David Zayats Anatoly eds Near field microscopy by elastic light scattering from a tip Philosophical Transactions of the Royal Society of London Series A Mathematical Physical and Engineering Sciences 362 1817 787 805 doi 10 1098 rsta 2003 1347 ISSN 1364 503X PMID 15306494 S2CID 20211405 Archived from the original on 2020 11 09 Retrieved 2020 11 02 Duarte FJ 2016 Tunable laser microscopy In Duarte FJ ed Tunable Laser Applications 3rd ed Boca Raton CRC Press pp 315 328 ISBN 9781482261066 Thomas JL Rudolph W 2008 Biological Microscopy with Ultrashort Laser Pulses In Duarte FJ ed Tunable Laser Applications 2nd ed Boca Raton CRC Press pp 245 80 ISBN 978 1 4200 6009 6 Solem J C 1983 X ray imaging on biological specimens Proceedings of International Conference on Lasers 83 635 640 OSTI 5998195 Solem J C 1982 High intensity x ray holography An approach to high resolution snapshot imaging of biological specimens Los Alamos National Laboratory Technical Report LA 9508 MS 83 23581 Bibcode 1982STIN 8323581S doi 10 2172 7056325 OSTI 7056325 Archived from the original on 2020 08 04 Retrieved 2019 06 28 Solem J C Baldwin G C 1982 Microholography of living organisms Science 218 4569 229 235 Bibcode 1982Sci 218 229S doi 10 1126 science 218 4569 229 PMID 17838608 S2CID 32381820 Solem J C Chapline G F 1984 X ray biomicroholography Optical Engineering 23 2 193 Bibcode 1984OptEn 23 193S doi 10 1117 12 7973410 Solem J C 1985 Microholography McGraw Hill Encyclopedia of Science and Technology Solem J C 1984 X ray holography of biological specimens Proceedings of Ninth International Congress on Photobiology Philadelphia PA USA 3 July 1984 LA UR 84 3340 CONF 840783 3 19 OSTI 6314558 Solem J C 1986 Imaging biological specimens with high intensity soft X rays Journal of the Optical Society of America B 3 11 1551 1565 Bibcode 1986JOSAB 3 1551S doi 10 1364 josab 3 001551 Haddad W S Solem J C Cullen D Boyer K Rhodes C K 1987 A description of the theory and apparatus for digital reconstruction of Fourier transform holograms Proceedings of Electronics Imaging 87 Nov 2 5 1987 Boston Journal of Electronic Imaging Institute for Graphic Communication Inc Boston p 693 Haddad W S Cullen D Boyer K Rhodes C K Solem J C Weinstein R S 1988 Design for a Fourier Transform Holographic Microscope X Ray Microscopy II Springer Series in Optical Sciences Vol 56 pp 284 287 doi 10 1007 978 3 540 39246 0 49 ISBN 978 3 662 14490 9 Haddad W S Solem J C Cullen D Boyer K Rhodes C K 1988 Design for a Fourier transform holographic microscope incorporating digital image reconstruction Proceedings of CLEO 88 Conference on Lasers and Electro Optics Anaheim CA April 1988 Optical Society of America OSA Technical Digest 7 WS4 Archived from the original on 2016 02 02 Retrieved 2016 01 13 Haddad W S Cullen D Solem J C Boyer K Rhodes C K 1988 X ray Fourier transform holographic microscope Proceedings of the OSA Topical Meeting on Short Wavelength Coherent Radiation Generation and Applications September 26 29 1988 Cape Cod MA Falcone R Kirz J Eds Optical Society of America Washington DC 284 289 Archived from the original on 2016 01 27 Retrieved 2016 01 13 Solem J C Boyer K Haddad W S Rhodes C K 1990 Prosnitz Donald ed Prosnitz D ed SPIE Prospects for X ray holography with free electron lasers Free Electron Lasers and Applications Free Electron Lasers and Applications 1227 105 116 Bibcode 1990SPIE 1227 105S doi 10 1117 12 18609 S2CID 121807907 Haddad W S Cullen D Solem J C Longworth J W McPherson L A Boyer K Rhodes C K 1991 Fourier transform holographic microscope Proceedings of SPIE Electronic Imaging 91 San Jose CA International Society for Optics and Photonics Camera and Input Scanner Systems 1448 24 81 88 Bibcode 1991SPIE 1448 81H doi 10 1117 12 45347 PMID 20733659 S2CID 120679508 Haddad W S Cullen D Solem J C Longworth J McPherson A Boyer K Rhodes C K 1992 Fourier transform holographic microscope Applied Optics 31 24 4973 4978 Bibcode 1992ApOpt 31 4973H doi 10 1364 ao 31 004973 PMID 20733659 Boyer K Solem J C Longworth J Borisov A Rhodes C K 1996 Biomedical three dimensional holographic microimaging at visible ultraviolet and X ray wavelengths Nature Medicine 2 8 939 941 doi 10 1038 nm0896 939 PMID 8705867 S2CID 25231033 Bell A G 1880 On the production and reproduction of sound by light American Journal of Science s3 20 118 305 324 Bibcode 1880AmJS 20 305B doi 10 2475 ajs s3 20 118 305 S2CID 130048089 Archived from the original on 2022 11 27 Retrieved 2019 06 28 Yao J Wang L V 2013 Photoacoustic Microscopy Laser Photonics Rev 7 5 1 36 Bibcode 2013LPRv 7 758Y doi 10 1002 lpor 201200060 PMC 3887369 PMID 24416085 Langer G Buchegger B Jacak J Klar T A Berer T 2016 Frequency domain photoacoustic and fluorescence microscopy Biomedical Optics Express 7 7 2692 702 doi 10 1364 BOE 7 002692 PMC 4948622 PMID 27446698 Kotrly M August 2015 New Possibilities of Using Micrscopic Techniques in Forensic Field Proceedings of Microscopy amp Microanallysis 21 S3 1365 1366 Bibcode 2015MiMic 21S1365K doi 10 1017 S1431927615007618 a b Basu S Millette JR 1986 Electron Microscopy in Forensic Occupational and Environmental Health Sciences New York Plenum Press Further reading editPluta Maksymilian 1988 Advanced Light Microscopy vol 1 Principles and Basic Properties Elsevier ISBN 978 0 444 98939 0 Pluta Maksymilian 1989 Advanced Light Microscopy vol 2 Specialised Methods Elsevier ISBN 978 0 444 98918 5 Bradbury S Bracegirdle B 1998 Introduction to Light Microscopy BIOS Scientific Publishers ISBN 978 0 387 91515 9 Inoue Shinya 1986 Video Microscopy Plenum Press ISBN 978 0 306 42120 4 Cremer C Cremer T 1978 Considerations on a laser scanning microscope with high resolution and depth of field PDF Microscopica Acta 81 1 31 44 PMID 713859 Archived from the original PDF on 2016 03 04 Retrieved 2009 12 22 Theoretical basis of super resolution 4Pi microscopy amp design of a confocal laser scanning fluorescence microscope Willis Randall C 2007 Portraits of life one molecule at a time Analytical Chemistry 79 5 1785 8 doi 10 1021 ac0718795 PMID 17375393 a feature article on sub diffraction microscopy from the March 1 2007 issue of Analytical ChemistryExternal links edit nbsp Wikimedia Commons has media related to Microscopy General edit Microscopy glossary Common terms used in amateur light microscopy Nikon MicroscopyU Extensive information on light microscopy Olympus Microscopy Microscopy Resource center Carl Zeiss Microscopy from the very beginning a step by step tutorial into the basics of microscopy Microscopy in Detail A resource with many illustrations elaborating the most common microscopy techniques Manawatu Microscopy first known collaboration environment for Microscopy and Image Analysis Audio microscope glossaryTechniques edit Ratio metric Imaging Applications For Microscopes Examples of Ratiometric Imaging Work on a Microscope Interactive Fluorescence Dye and Filter Database Carl Zeiss Interactive Fluorescence Dye and Filter Database New approaches to microscopy Eric Betzig Beyond the Nobel Prize New approaches to microscopy Retrieved from https en wikipedia org w index php title Microscopy amp oldid 1193151316, wikipedia, wiki, book, books, library,