fbpx
Wikipedia

Electron microscope

An electron microscope is a microscope that uses a beam of electrons as a source of illumination. They use electron optics that are analogous to the glass lenses of an optical light microscope to control the electron beam, for instance focusing them to produce magnified images or electron diffraction patterns. As the wavelength of an electron can be up to 100,000 times smaller than that of visible light, electron microscopes have a much higher resolution of about 0.1 nm, which compares to about 200 nm for light microscopes. Electron microscope may refer to:

A transmission electron microscope from the 2000s
An image of an ant in a scanning electron microscope

Additional details can be found in the above links. This article contains some general information mainly about transmission electron microscopes.

History edit

 
Reproduction of an early electron microscope constructed by Ernst Ruska in the 1930s

Many developments laid the groundwork of the electron optics used in microscopes.[1] One significant step was the work of Hertz in 1883[2] who made a cathode-ray tube with electrostatic and magnetic deflection, demonstrating manipulation of the direction of an electron beam. Others were focusing of the electrons by an axial magnetic field by Emil Wiechert in 1899,[3] improved oxide-coated cathodes which produced more electrons by Arthur Wehnelt in 1905[4] and the development of the electromagnetic lens in 1926 by Hans Busch.[5] According to Dennis Gabor, the physicist Leó Szilárd tried in 1928 to convince him to build an electron microscope, for which Szilárd had filed a patent.[6]

To this day the issue of who invented the transmission electron microscope is controversial.[7][8][9][10] In 1928, at the Technical University of Berlin, Adolf Matthias (Professor of High Voltage Technology and Electrical Installations) appointed Max Knoll to lead a team of researchers to advance research on electron beams and cathode-ray oscilloscopes. The team consisted of several PhD students including Ernst Ruska. In 1931, Max Knoll and Ernst Ruska[11][12] successfully generated magnified images of mesh grids placed over an anode aperture. The device, a replicate of which is shown in the figure, used two magnetic lenses to achieve higher magnifications, the first electron microscope. (Max Knoll died in 1969, so did not receive a share of the 1986 Nobel prize for the invention of electron microscopes.)

Apparently independent of this effort was work at Siemens-Schuckert by Reinhold Rüdenberg. According to patent law (U.S. Patent No. 2058914[13] and 2070318,[14] both filed in 1932), he is the inventor of the electron microscope, but it is not clear when he had a working instrument. He stated in a very brief article in 1932[15] that Siemens had been working on this for some years before the patents were filed in 1932, claiming that his effort was parallel to the university development. He died in 1961, so similar to Max Knoll, was not eligible for a share of the 1986 Nobel prize.

In the following year, 1933, Ruska and Knoll built the first electron microscope that exceeded the resolution of an optical (light) microscope.[16] Four years later, in 1937, Siemens financed the work of Ernst Ruska and Bodo von Borries, and employed Helmut Ruska, Ernst's brother, to develop applications for the microscope, especially with biological specimens.[16][17] Also in 1937, Manfred von Ardenne pioneered the scanning electron microscope.[18] Siemens produced the first commercial electron microscope in 1938.[19] The first North American electron microscopes were constructed in the 1930s, at the Washington State University by Anderson and Fitzsimmons [20] and at the University of Toronto by Eli Franklin Burton and students Cecil Hall, James Hillier, and Albert Prebus. Siemens produced a transmission electron microscope (TEM) in 1939.[21] Although current transmission electron microscopes are capable of two million times magnification, as scientific instruments they remain similar but with improved optics.

Wavelength edit

Operating principle of a transmission electron microscope

In a typical electron gun, individual electrons, which have an elementary charge   (about   coulombs) and a mass   (about   kg), with a potential of   volts, have an energy amount of   joules. The wavelength is[22]

 ,

where   is the speed of light in vacuum (about   m/s). See electron diffraction for a full explanation.

Types edit

Transmission electron microscope (TEM) edit

 
Diagram of a transmission electron microscope

The original form of the electron microscope, the transmission electron microscope (TEM), uses a high voltage electron beam to illuminate the specimen and create an image. An electron beam is produced by an electron gun, with the electrons typically at 40 to 400 keV, focused by electromagnetic lenses, and transmitted through the specimen. When it emerges from the specimen, the electron beam carries information about the structure of the specimen that is magnified by lenses of the microscope. The spatial variation in this information (the "image") may be viewed by projecting the magnified electron image onto a detector. For example, the image may be viewed directly by an operator using a fluorescent viewing screen coated with a phosphor or scintillator material such as zinc sulfide. A high-resolution phosphor may also be coupled by means of a lens optical system or a fibre optic light-guide to the sensor of a digital camera. Direct electron detectors have no scintillator and are directly exposed to the electron beam, which addresses some of the limitations of scintillator-coupled cameras.[23]

The resolution of TEMs is limited primarily by spherical aberration, but a new generation of hardware correctors can reduce spherical aberration to increase the resolution in high-resolution transmission electron microscopy (HRTEM) to below 0.5 angstrom (50 picometres),[24] enabling magnifications above 50 million times.[25] The ability of HRTEM to determine the positions of atoms within materials is useful for nano-technologies research and development.[26]

Transmission electron microscopes are often used in electron diffraction mode. The advantages of electron diffraction over X-ray crystallography are that the specimen need not be a single crystal or even a polycrystalline powder.[citation needed]

Scanning transmission electron microscope (STEM) edit

The STEM rasters a focused incident probe across a specimen. The high resolution of the TEM is thus possible in STEM. The focusing action (and aberrations) occur before the electrons hit the specimen in the STEM, but afterward in the TEM. The STEMs use of SEM-like beam rastering simplifies annular dark-field imaging, and other analytical techniques, but also means that image data is acquired in serial rather than in parallel fashion.[citation needed]

Scanning electron microscope (SEM) edit

Operating principle of a scanning electron microscope
 
Image of Bacillus subtilis taken with a 1960s electron microscope

The SEM produces images by probing the specimen with a focused electron beam that is scanned across the specimen (raster scanning). When the electron beam interacts with the specimen, it loses energy by a variety of mechanisms. The lost energy is converted into alternative forms such as heat, emission of low-energy secondary electrons and high-energy backscattered electrons, light emission (cathodoluminescence) or X-ray emission, all of which provide signals carrying information about the properties of the specimen surface, such as its topography and composition.[citation needed] The image displayed by an SEM maps the varying intensity of any of these signals into the image in a position corresponding to the position of the beam on the specimen when the signal was generated. In the SEM image of an ant shown, the image was constructed from signals produced by a secondary electron detector, the normal or conventional imaging mode in most SEMs.[citation needed]

Generally, the image resolution of an SEM is lower than that of a TEM. However, because the SEM images the surface of a sample rather than its interior, the electrons do not have to travel through the sample. This reduces the need for extensive sample preparation to thin the specimen to electron transparency. The SEM also has a great depth of field, and so can produce images that are good representations of the three-dimensional surface shape of the sample.[citation needed]

In their most common configurations, electron microscopes produce images with a single brightness value per pixel, with the results usually rendered in greyscale.[27] However, often these images are then colourized through the use of feature-detection software, or simply by hand-editing using a graphics editor. This may be done to clarify structure or for aesthetic effect and generally does not add new information about the specimen.[28]

Sample preparation for TEM edit

 
An insect coated in gold for viewing with a scanning electron microscope

Materials to be viewed in a transmission electron microscope may require processing to produce a suitable sample. The technique required varies depending on the specimen and the analysis required:

  • Cryofixation – freezing a specimen so that the water forms vitreous (non-crystalline) ice. This preserves the specimen in a snapshot of its native state. Methods to achieve this vitrification include plunge freezing rapidly in liquid ethane, and high pressure freezing. An entire field called cryo-electron microscopy has branched from this technique. With the development of cryo-electron microscopy of vitreous sections (CEMOVIS)[30] and cryo-focused ion beam milling of lamellae,[31] it is now possible to observe samples from virtually any biological specimen close to its native state.
  • Dehydration – replacement of water with organic solvents such as ethanol or acetone, followed by critical point drying or infiltration with embedding resins. See also freeze drying.[citation needed]
  • Embedding, biological specimens – after dehydration, tissue for observation in the transmission electron microscope is embedded so it can be sectioned ready for viewing. To do this the tissue is passed through a 'transition solvent' such as propylene oxide (epoxypropane) or acetone and then infiltrated with an epoxy resin such as Araldite, Epon, or Durcupan;[32] tissues may also be embedded directly in water-miscible acrylic resin. After the resin has been polymerized (hardened) the sample is sectioned by ultramicrotomy and stained.[citation needed]
  • Embedding, materials – after embedding in resin, the specimen is usually ground and polished to a mirror-like finish using ultra-fine abrasives.[citation needed]
  • Freeze-fracture or freeze-etch – a preparation method[33][34][35] particularly useful for examining lipid membranes and their incorporated proteins in "face on" view.[36][37][38]
     
    Freeze-fracturing helps to peel open membranes to allow visualization of what is inside
     
    External face of bakers yeast membrane showing the small holes where proteins are fractured out, sometimes as small ring patterns.
    The fresh tissue or cell suspension is frozen rapidly (cryofixation), then fractured by breaking[39] (or by using a microtome)[38] while maintained at liquid nitrogen temperature. The cold fractured surface (sometimes "etched" by increasing the temperature to about −100 °C for several minutes to let some ice sublime)[38] is then shadowed with evaporated platinum or gold at an average angle of 45° in a high vacuum evaporator. The second coat of carbon, evaporated perpendicular to the average surface plane is often performed to improve the stability of the replica coating. The specimen is returned to room temperature and pressure, then the extremely fragile "pre-shadowed" metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids, hypochlorite solution or SDS detergent. The still-floating replica is thoroughly washed free from residual chemicals, carefully fished up on fine grids, dried then viewed in the TEM.[citation needed]
  • Freeze-fracture replica immunogold labeling (FRIL) – the freeze-fracture method has been modified to allow the identification of the components of the fracture face by immunogold labeling. Instead of removing all the underlying tissue of the thawed replica as the final step before viewing in the microscope the tissue thickness is minimized during or after the fracture process. The thin layer of tissue remains bound to the metal replica so it can be immunogold labeled with antibodies to the structures of choice. The thin layer of the original specimen on the replica with gold attached allows the identification of structures in the fracture plane.[40] There are also related methods which label the surface of etched cells[41] and other replica labeling variations.[42]
  • Ion beam milling – thins samples until they are transparent to electrons by firing ions (typically argon) at the surface from an angle and sputtering material from the surface. A subclass of this is focused ion beam milling, where gallium ions are used to produce an electron transparent membrane or 'lamella' in a specific region of the sample, for example through a device within a microprocessor or a focused ion beam SEM. Ion beam milling may also be used for cross-section polishing prior to analysis of materials that are difficult to prepare using mechanical polishing.[citation needed]
  • Negative stain – suspensions containing nanoparticles or fine biological material (such as viruses and bacteria) are briefly mixed with a dilute solution of an electron-opaque solution such as ammonium molybdate, uranyl acetate (or formate), or phosphotungstic acid.[citation needed] This mixture is applied to an EM grid, pre-coated with a plastic film such as formvar, blotted, then allowed to dry. Viewing of this preparation in the TEM should be carried out without delay for best results. The method is important in microbiology for fast but crude morphological identification, but can also be used as the basis for high-resolution 3D reconstruction using EM tomography methodology when carbon films are used for support. Negative staining is also used for observation of nanoparticles.[citation needed]
  • Sectioning – produces thin slices of the specimen, semitransparent to electrons. These can be cut using ultramicrotomy on an ultramicrotome with a glass or diamond knife to produce ultra-thin sections about 60–90 nm thick. Disposable glass knives are also used because they can be made in the lab and are much cheaper. Sections can also be created in situ by milling in a focused ion beam SEM, where the section is known as a lamella.[31]
  • Staining – uses heavy metals such as lead, uranium or tungsten to scatter imaging electrons and thus give contrast between different structures, since many (especially biological) materials are nearly "transparent" to electrons (weak phase objects). In biology, specimens can be stained "en bloc" before embedding and also later after sectioning. Typically thin sections are stained for several minutes with an aqueous or alcoholic solution of uranyl acetate followed by aqueous lead citrate.[43]

EM workflows edit

Early electron microscopy of biological specimens was often descriptive, making use of the newly available higher resolution.[44] This is still the case for various applications, such as diagnostic electron microscopy.

However, electron microscopes are now frequently used in more complex workflows, with each workflow typically using multiple technologies to enable more complex and/or more quantitative analyses of a sample. A few examples are outlined below, but this should not be considered an exhaustive list. The choice of workflow will be highly dependent on the application and the requirements of the corresponding scientific questions, such as resolution, volume, nature of the target molecule, etc.

For example, images from light and electron microscopy of the same region of a sample can be overlaid to correlate the data from the two modalities. This is commonly used to provide higher resolution contextual EM information about a fluorescently labelled structure. This correlative light and electron microscopy (CLEM)[45] is one of a range of correlative workflows now available. Another example is high resolution mass spectrometry (ion microscopy), which has been used to provide correlative information about subcellular antibiotic localisation,[46] data that would be difficult to obtain by other means.[citation needed]

The initial role of electron microscopes in imaging two-dimensional slices (TEM) or a specimen surface (SEM with secondary electrons) has also increasingly expanded into the depth of samples.[47] An early example of these ‘volume EM’ workflows was simply to stack TEM images of serial sections cut through a sample. The next development was virtual reconstruction of a thick section (200-500 nm) volume by backprojection of a set of images taken at different tilt angles - TEM tomography.[48]

Serial imaging for volume EM edit

To acquire volume EM datasets of larger depths than TEM tomography (micrometers or millimeters in the z axis), a series of images taken through the sample depth can be used. For example, ribbons of serial sections can be imaged in a TEM as described above, and when thicker sections are used, serial TEM tomography can be used to increase the z-resolution. More recently, back scattered electron (BSE) images can be acquired of a larger series of sections collected on silicon wafers, known as SEM array tomography.[49][50] An alternative approach is to use BSE SEM to image the block surface instead of the section, after each section has been removed. By this method, an ultramicrotome installed in an SEM chamber can increase automation of the workflow; the specimen block is loaded in the chamber and the system programmed to continuously cut and image through the sample. This is known as serial block face SEM.[51] A related method uses focused ion beam milling instead of an ultramicrotome to remove sections. In these serial imaging methods, the output is essentially a sequence of images through a specimen block that can be digitally aligned in sequence and thus reconstructed into a volume EM dataset. The increased volume available in these methods has expanded the capability of electron microscopy to address new questions,[47] such as mapping neural connectivity in the brain,[52] and membrane contact sites between organelles.[53]

Disadvantages edit

 
JEOL transmission and scanning electron microscope made in the mid-1970s

Electron microscopes are expensive to build and maintain. Microscopes designed to achieve high resolutions must be housed in stable buildings (sometimes underground) with special services such as magnetic field canceling systems.[citation needed]

The samples largely have to be viewed in vacuum, as the molecules that make up air would scatter the electrons. An exception is liquid-phase electron microscopy[54] using either a closed liquid cell or an environmental chamber, for example, in the environmental scanning electron microscope, which allows hydrated samples to be viewed in a low-pressure (up to 20 Torr or 2.7 kPa) wet environment. Various techniques for in situ electron microscopy of gaseous samples have been developed.[55]

Scanning electron microscopes operating in conventional high-vacuum mode usually image conductive specimens; therefore non-conductive materials require conductive coating (gold/palladium alloy, carbon, osmium, etc.). The low-voltage mode of modern microscopes makes possible the observation of non-conductive specimens without coating. Non-conductive materials can be imaged also by a variable pressure (or environmental) scanning electron microscope.[citation needed]

Small, stable specimens such as carbon nanotubes, diatom frustules and small mineral crystals (asbestos fibres, for example) require no special treatment before being examined in the electron microscope. Samples of hydrated materials, including almost all biological specimens, have to be prepared in various ways to stabilize them, reduce their thickness (ultrathin sectioning) and increase their electron optical contrast (staining). These processes may result in artifacts, but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods. Since the 1980s, analysis of cryofixed, vitrified specimens has also become increasingly used by scientists, further confirming the validity of this technique.[56][57][58]

See also edit

References edit

  1. ^ Calbick, C. J. (1944). "Historical Background of Electron Optics". Journal of Applied Physics. 15 (10): 685–690. Bibcode:1944JAP....15..685C. doi:10.1063/1.1707371. ISSN 0021-8979.
  2. ^ Hertz, Heinrich (2019), "Introduction to Heinrich Hertz's Miscellaneous Papers (1895) by Philipp Lenard", Heinrich Rudolf Hertz (1857-1894), Routledge, pp. 87–88, doi:10.4324/9780429198960-4, ISBN 978-0-429-19896-0, S2CID 195494352, retrieved 2023-02-24
  3. ^ Wiechert, E. (1899). "Experimentelle Untersuchungen über die Geschwindigkeit und die magnetische Ablenkbarkeit der Kathodenstrahlen". Annalen der Physik und Chemie (in German). 305 (12): 739–766. Bibcode:1899AnP...305..739W. doi:10.1002/andp.18993051203.
  4. ^ Wehnelt, A. (1905). "X. On the discharge of negative ions by glowing metallic oxides, and allied phenomena". The London, Edinburgh, and Dublin Philosophical Magazine and Journal of Science. 10 (55): 80–90. doi:10.1080/14786440509463347. ISSN 1941-5982.
  5. ^ Busch, H. (1926). "Berechnung der Bahn von Kathodenstrahlen im axialsymmetrischen elektromagnetischen Felde". Annalen der Physik (in German). 386 (25): 974–993. Bibcode:1926AnP...386..974B. doi:10.1002/andp.19263862507.
  6. ^ Dannen, Gene (1998) Leo Szilard the Inventor: A Slideshow (1998, Budapest, conference talk). dannen.com
  7. ^ Mulvey, T (1962). "Origins and historical development of the electron microscope". British Journal of Applied Physics. 13 (5): 197–207. doi:10.1088/0508-3443/13/5/303. ISSN 0508-3443.
  8. ^ Tao, Yaping (2018). "A Historical Investigation of the Debates on the Invention and Invention Rights of Electron Microscope". Proceedings of the 3rd International Conference on Contemporary Education, Social Sciences and Humanities (ICCESSH 2018). Atlantis Press. pp. 1438–1441. doi:10.2991/iccessh-18.2018.313. ISBN 978-94-6252-528-3. {{cite book}}: |journal= ignored (help)
  9. ^ Freundlich, Martin M. (1963). "Origin of the Electron Microscope: The history of a great invention, and of a misconception concerning the inventors, is reviewed". Science. 142 (3589): 185–188. doi:10.1126/science.142.3589.185. ISSN 0036-8075. PMID 14057363.
  10. ^ Rüdenberg, Reinhold (2010), Origin and Background of the Invention of the Electron Microscope, Advances in Imaging and Electron Physics, vol. 160, Elsevier, pp. 171–205, doi:10.1016/s1076-5670(10)60005-5, ISBN 9780123810175, retrieved 2023-02-11.
  11. ^ Knoll, M.; Ruska, E. (1932). "Beitrag zur geometrischen Elektronenoptik. I". Annalen der Physik. 404 (5): 607–640. Bibcode:1932AnP...404..607K. doi:10.1002/andp.19324040506. ISSN 0003-3804.
  12. ^ Knoll, M.; Ruska, E. (1932). "Das Elektronenmikroskop". Zeitschrift für Physik (in German). 78 (5–6): 318–339. Bibcode:1932ZPhy...78..318K. doi:10.1007/BF01342199. ISSN 1434-6001. S2CID 186239132.
  13. ^ Rüdenberg, Reinhold. "Apparatus for producing images of objects". Patent Public Search Basic. Retrieved 24 February 2023.
  14. ^ Rüdenberg, Reinhold. "Apparatus for producing images of objects". Patent Public Search Basic. Retrieved 24 February 2023.
  15. ^ Rodenberg, R. (1932). "Elektronenmikroskop". Die Naturwissenschaften (in German). 20 (28): 522. Bibcode:1932NW.....20..522R. doi:10.1007/BF01505383. ISSN 0028-1042. S2CID 263996652.
  16. ^ a b Ruska, Ernst (1986). "Ernst Ruska Autobiography". Nobel Foundation. Retrieved 2010-01-31.
  17. ^ Kruger, DH; Schneck, P; Gelderblom, HR (May 2000). "Helmut Ruska and the visualisation of viruses". The Lancet. 355 (9216): 1713–1717. doi:10.1016/S0140-6736(00)02250-9. PMID 10905259. S2CID 12347337.
  18. ^ Ardenne, M. Von; Beischer, D. (1940). "Untersuchung von Metalloxyd-Rauchen mit dem Universal-Elektronenmikroskop" [Investigation of metal oxide smoking with the universal electron microscope]. Zeitschrift für Elektrochemie und Angewandte Physikalische Chemie (in German). 46 (4): 270–277. doi:10.1002/bbpc.19400460406. S2CID 137136299.
  19. ^ History of electron microscopy, 1931–2000. Authors.library.caltech.edu (2002-12-10). Retrieved on 2017-04-29.
  20. ^ "North America's first electron microscope".
  21. ^ . Inventor of the Week: Archive. 2003-05-01. Archived from the original on 2003-08-23. Retrieved 2010-01-31.
  22. ^ Kirkland, Earl (2010). Advanced computing in electron microscopy. New York: Springer. ISBN 978-1-4419-6533-2. OCLC 668095602.
  23. ^ Cheng Y, Grigorieff N, Penczek PA, Walz T (April 2015). "A primer to single-particle cryo-electron microscopy". Cell. 161 (3): 438–449. doi:10.1016/j.cell.2015.03.050. PMC 4409659. PMID 25910204.
  24. ^ Erni, Rolf; Rossell, MD; Kisielowski, C; Dahmen, U (2009). "Atomic-Resolution Imaging with a Sub-50-pm Electron Probe". Physical Review Letters. 102 (9): 096101. Bibcode:2009PhRvL.102i6101E. doi:10.1103/PhysRevLett.102.096101. PMID 19392535.
  25. ^ . Office of Basic Energy Sciences, U.S. Department of Energy. 2006-05-26. Archived from the original on 2010-02-01. Retrieved 2010-01-31.
  26. ^ O'Keefe MA; Allard LF (2004-01-18). "Sub-Ångstrom Electron Microscopy for Sub-Ångstrom Nano-Metrology" (PDF). Information Bridge: DOE Scientific and Technical Information – Sponsored by OSTI. {{cite journal}}: Cite journal requires |journal= (help)
  27. ^ Burgess, Jeremy (1987). Under the Microscope: A Hidden World Revealed. CUP Archive. p. 11. ISBN 978-0-521-39940-1.
  28. ^ "Introduction to Electron Microscopy" (PDF). FEI Company. p. 15. Retrieved 12 December 2012.
  29. ^ Humbel, Bruno M; Schwarz, Heinz; Tranfield, Erin M; Fleck, Roland A (February 15, 2019). "Chapter 10: Chemical Fixation". In Fleck, Roland A; Humbel, Bruno M (eds.). Biological Field Emission Scanning Electron Microscopy, First Edition. John Wiley & Sons Ltd. pp. 191–221. doi:10.1002/9781118663233.ch10. ISBN 9781118663233. S2CID 243064180.
  30. ^ Al-Amoudi A, Norlen LP, Dubochet J (October 2004). "Cryo-electron microscopy of vitreous sections of native biological cells and tissues". J Struct Biol. 148 (1): 131–5. doi:10.1016/j.jsb.2004.03.010. PMID 15363793.
  31. ^ a b Wagner FR, Watanabe R, Schampers R, Singh D, Persoon H, Schaffer M, Fruhstorfer P, Plitzko J, Villa E (June 2020). "Preparing samples from whole cells using focused-ion-beam milling for cryo-electron tomography". Nat Protoc. 15 (6): 2041–2070. doi:10.1038/s41596-020-0320-x. PMC 8053421. PMID 32405053.
  32. ^ Luft, J.H. (1961). "Improvements in epoxy resin embedding methods". The Journal of Biophysical and Biochemical Cytology. Vol. 9, no. 2. p. 409. PMC 2224998. PMID 13764136.
  33. ^ Meryman H.T. and Kafig E. (1955). The study of frozen specimens, ice crystals and ices crystal growth by electron microscopy. Naval Med. Res. Ints. Rept NM 000 018.01.09 Vol. 13 pp 529–544
  34. ^ Steere, Russell L. (1957-01-25). "Electron microscopy of structural detail in frozen biological specimens". The Journal of Biophysical and Biochemical Cytology. 3 (1): 45–60. doi:10.1083/jcb.3.1.45. PMC 2224015. PMID 13416310.
  35. ^ Isailović, Tanja M.; Todosijević, Marija N.; Đorđević, Sanela M.; Savić, Snežana D. (2017-01-01), Čalija, Bojan (ed.), "Chapter 7 - Natural Surfactants-Based Micro/Nanoemulsion Systems for NSAIDs—Practical Formulation Approach, Physicochemical and Biopharmaceutical Characteristics/Performances", Microsized and Nanosized Carriers for Nonsteroidal Anti-Inflammatory Drugs, Boston: Academic Press, pp. 179–217, doi:10.1016/b978-0-12-804017-1.00007-8, ISBN 978-0-12-804017-1, retrieved 2020-10-22
  36. ^ Moor H, Mühlethaler K (1963). "Fine structure in frozen-etched yeast cells". The Journal of Cell Biology. 17 (3): 609–628. doi:10.1083/jcb.17.3.609. PMC 2106217. PMID 19866628.
  37. ^ Black, Joel A. (1990-01-01), Conn, P. Michael (ed.), "[20] - Use of Freeze-Fracture in Neurobiology", Methods in Neurosciences, Quantitative and Qualitative Microscopy, Academic Press, 3: 343–360, doi:10.1016/b978-0-12-185255-9.50025-0, ISBN 9780121852559, retrieved 2020-10-22
  38. ^ a b c Stillwell, William (2016-01-01), Stillwell, William (ed.), "Chapter 11 - Long-Range Membrane Properties", An Introduction to Biological Membranes (Second Edition), Elsevier, pp. 221–245, doi:10.1016/b978-0-444-63772-7.00011-7, ISBN 978-0-444-63772-7, retrieved 2020-10-22
  39. ^ Bullivant, Stanley; Ames, Adelbert (1966-06-01). "A simple freeze-fracture replication method for electron microscopy". The Journal of Cell Biology. 29 (3): 435–447. doi:10.1083/jcb.29.3.435. PMC 2106967. PMID 5962938.
  40. ^ Gruijters, W. T.; Kistler, J; Bullivant, S; Goodenough, D. A. (1987-03-01). "Immunolocalization of MP70 in lens fiber 16-17-nm intercellular junctions". The Journal of Cell Biology. 104 (3): 565–572. doi:10.1083/jcb.104.3.565. PMC 2114558. PMID 3818793.
  41. ^ da Silva, Pedro Pinto; Branton, Daniel (1970-06-01). "Membrane splitting in freeze-etching". The Journal of Cell Biology. 45 (3): 598–605. doi:10.1083/jcb.45.3.598. PMC 2107921. PMID 4918216.
  42. ^ Rash, J. E.; Johnson, T. J.; Hudson, C. S.; Giddings, F. D.; Graham, W. F.; Eldefrawi, M. E. (1982-11-01). "Labelled-replica techniques: post-shadow labelling of intramembrane particles in freeze-fracture replicas". Journal of Microscopy. 128 (Pt 2): 121–138. doi:10.1111/j.1365-2818.1982.tb00444.x. PMID 6184475. S2CID 45238172.
  43. ^ Reynolds, E. S. (1963). "The use of lead citrate at high pH as an electron-opaque stain in electron microscopy". Journal of Cell Biology. 17 (1): 208–212. doi:10.1083/jcb.17.1.208. PMC 2106263. PMID 13986422.
  44. ^ Sjostrand FS, Hanzon V (November 1954). "Ultrastructure of Golgi apparatus of exocrine cells of mouse pancreas". Exp Cell Res. 7 (2): 415–29. doi:10.1016/s0014-4827(54)80087-5. PMID 13220587.
  45. ^ "Methods in Cell Biology | Correlative Light and Electron Microscopy III | ScienceDirect.com by Elsevier".
  46. ^ Finin P, Khan RM, Oh S, Boshoff HI, Barry CE (May 2023). "Chemical approaches to unraveling the biology of mycobacteria". Cell Chem Biol. 30 (5): 420–435. doi:10.1016/j.chembiol.2023.04.014. PMC 10201459. PMID 37207631.
  47. ^ a b Peddie CJ, Genoud C, Kreshuk A, Meechan K, Micheva KD, Narayan K, Pape C, Parton RG, Schieber NL, Schwab Y, Titze B, Verkade P, Aubrey A, Collinson LM (July 2022). "Volume electron microscopy". Nat Rev Methods Primers. 2: 51. doi:10.1038/s43586-022-00131-9. PMC 7614724. PMID 37409324.
  48. ^ Crowther RA, Amos LA, Finch JT, De Rosier DJ, Klug A (May 1970). "Three dimensional reconstructions of spherical viruses by fourier synthesis from electron micrographs". Nature. 226 (5244): 421–5. Bibcode:1970Natur.226..421C. doi:10.1038/226421a0. PMID 4314822. S2CID 4217806.
  49. ^ White IJ, Burden JJ (2023). "A practical guide to starting SEM array tomography—An accessible volume EM technique". Chapter 7 - A practical guide to starting SEM array tomography—An accessible volume EM technique. Methods in Cell Biology. Vol. 177. pp. 171–196. doi:10.1016/bs.mcb.2022.12.023. ISBN 9780323916073. PMID 37451766.
  50. ^ Kolotuev I (August 2023). "Work smart, not hard: How array tomography can help increase the ultrastructure data output". J Microsc. doi:10.1111/jmi.13217. PMID 37626455. S2CID 261174348.
  51. ^ Denk W, Horstmann H (November 2004). "Serial block-face scanning electron microscopy to reconstruct three-dimensional tissue nanostructure". PLOS Biol. 2 (11): e329. doi:10.1371/journal.pbio.0020329. PMC 524270. PMID 15514700.
  52. ^ Abbott LF, Bock DD, Callaway EM, Denk W, Dulac C, Fairhall AL, Fiete I, Harris KM, Helmstaedter M, Jain V, Kasthuri N, LeCun Y, Lichtman JW, Littlewood PB, Luo L, Maunsell JH, Reid RC, Rosen BR, Rubin GM, Sejnowski TJ, Seung HS, Svoboda K, Tank DW, Tsao D, Van Essen DC (September 2020). "The Mind of a Mouse". Cell. 182 (6): 1372–1376. doi:10.1016/j.cell.2020.08.010. PMID 32946777. S2CID 221766693.
  53. ^ Prinz WA, Toulmay A, Balla T (January 2020). "The functional universe of membrane contact sites". Nat Rev Mol Cell Biol. 21 (1): 7–24. doi:10.1038/s41580-019-0180-9. PMC 10619483. PMID 31732717. S2CID 208019972.
  54. ^ de Jonge, N.; Ross, F.M. (2011). "Electron microscopy of specimens in liquid". Nature Nanotechnology. 6 (8): 695–704. Bibcode:2003NatMa...2..532W. doi:10.1038/nmat944. PMID 12872162. S2CID 21379512.
  55. ^ Gai, P.L.; Boyes, E.D. (2009). "Advances in atomic resolution in situ environmental transmission electron microscopy and 1A aberration corrected in situ electron microscopy". Microsc Res Tech. 72 (3): 153–164. arXiv:1705.05754. doi:10.1002/jemt.20668. PMID 19140163. S2CID 1746538.
  56. ^ Adrian, Marc; Dubochet, Jacques; Lepault, Jean; McDowall, Alasdair W. (1984). "Cryo-electron microscopy of viruses". Nature (Submitted manuscript). 308 (5954): 32–36. Bibcode:1984Natur.308...32A. doi:10.1038/308032a0. PMID 6322001. S2CID 4319199.
  57. ^ Sabanay, I.; Arad, T.; Weiner, S.; Geiger, B. (1991). "Study of vitrified, unstained frozen tissue sections by cryoimmunoelectron microscopy". Journal of Cell Science. 100 (1): 227–236. doi:10.1242/jcs.100.1.227. PMID 1795028.
  58. ^ Kasas, S.; Dumas, G.; Dietler, G.; Catsicas, S.; Adrian, M. (2003). "Vitrification of cryoelectron microscopy specimens revealed by high-speed photographic imaging". Journal of Microscopy. 211 (1): 48–53. doi:10.1046/j.1365-2818.2003.01193.x. PMID 12839550. S2CID 40058086.
  • An Introduction to Microscopy 2013-07-19 at the Wayback Machine: resources for teachers and students
  • Cell Centered Database – Electron microscopy data
  • Science Aid: Electron Microscopy:By Kaden park


electron, microscope, confused, with, scanning, tunneling, microscope, electron, microscope, microscope, that, uses, beam, electrons, source, illumination, they, electron, optics, that, analogous, glass, lenses, optical, light, microscope, control, electron, b. Not to be confused with Scanning tunneling microscope An electron microscope is a microscope that uses a beam of electrons as a source of illumination They use electron optics that are analogous to the glass lenses of an optical light microscope to control the electron beam for instance focusing them to produce magnified images or electron diffraction patterns As the wavelength of an electron can be up to 100 000 times smaller than that of visible light electron microscopes have a much higher resolution of about 0 1 nm which compares to about 200 nm for light microscopes Electron microscope may refer to Transmission electron microscopy TEM where swift electrons go through a thin sample Scanning transmission electron microscopy STEM which is similar to TEM with a scanned electron probe Scanning electron microscope SEM which is similar to STEM but with thick samples Electron microprobe similar to a SEM but more for chemical analysis Ultrafast scanning electron microscopy version of a SEM that can operate very fast Low energy electron microscopy LEEM used to image surfaces Photoemission electron microscopy PEEM which is similar to LEEM using electrons emitted from surfaces by photonsA transmission electron microscope from the 2000sAn image of an ant in a scanning electron microscopeThis article needs additional citations for verification Please help improve this article by adding citations to reliable sources Unsourced material may be challenged and removed Find sources Electron microscope news newspapers books scholar JSTOR September 2023 Learn how and when to remove this template message Additional details can be found in the above links This article contains some general information mainly about transmission electron microscopes Contents 1 History 2 Wavelength 3 Types 3 1 Transmission electron microscope TEM 3 1 1 Scanning transmission electron microscope STEM 3 2 Scanning electron microscope SEM 4 Sample preparation for TEM 5 EM workflows 5 1 Serial imaging for volume EM 6 Disadvantages 7 See also 8 ReferencesHistory editSee also Transmission electron microscopy History nbsp Reproduction of an early electron microscope constructed by Ernst Ruska in the 1930sMany developments laid the groundwork of the electron optics used in microscopes 1 One significant step was the work of Hertz in 1883 2 who made a cathode ray tube with electrostatic and magnetic deflection demonstrating manipulation of the direction of an electron beam Others were focusing of the electrons by an axial magnetic field by Emil Wiechert in 1899 3 improved oxide coated cathodes which produced more electrons by Arthur Wehnelt in 1905 4 and the development of the electromagnetic lens in 1926 by Hans Busch 5 According to Dennis Gabor the physicist Leo Szilard tried in 1928 to convince him to build an electron microscope for which Szilard had filed a patent 6 To this day the issue of who invented the transmission electron microscope is controversial 7 8 9 10 In 1928 at the Technical University of Berlin Adolf Matthias Professor of High Voltage Technology and Electrical Installations appointed Max Knoll to lead a team of researchers to advance research on electron beams and cathode ray oscilloscopes The team consisted of several PhD students including Ernst Ruska In 1931 Max Knoll and Ernst Ruska 11 12 successfully generated magnified images of mesh grids placed over an anode aperture The device a replicate of which is shown in the figure used two magnetic lenses to achieve higher magnifications the first electron microscope Max Knoll died in 1969 so did not receive a share of the 1986 Nobel prize for the invention of electron microscopes Apparently independent of this effort was work at Siemens Schuckert by Reinhold Rudenberg According to patent law U S Patent No 2058914 13 and 2070318 14 both filed in 1932 he is the inventor of the electron microscope but it is not clear when he had a working instrument He stated in a very brief article in 1932 15 that Siemens had been working on this for some years before the patents were filed in 1932 claiming that his effort was parallel to the university development He died in 1961 so similar to Max Knoll was not eligible for a share of the 1986 Nobel prize In the following year 1933 Ruska and Knoll built the first electron microscope that exceeded the resolution of an optical light microscope 16 Four years later in 1937 Siemens financed the work of Ernst Ruska and Bodo von Borries and employed Helmut Ruska Ernst s brother to develop applications for the microscope especially with biological specimens 16 17 Also in 1937 Manfred von Ardenne pioneered the scanning electron microscope 18 Siemens produced the first commercial electron microscope in 1938 19 The first North American electron microscopes were constructed in the 1930s at the Washington State University by Anderson and Fitzsimmons 20 and at the University of Toronto by Eli Franklin Burton and students Cecil Hall James Hillier and Albert Prebus Siemens produced a transmission electron microscope TEM in 1939 21 Although current transmission electron microscopes are capable of two million times magnification as scientific instruments they remain similar but with improved optics Wavelength edit source source source source source source source source Operating principle of a transmission electron microscopeSee also Electron diffraction In a typical electron gun individual electrons which have an elementary charge e displaystyle e nbsp about 1 6 10 19 displaystyle 1 6 times 10 19 nbsp coulombs and a mass m displaystyle m nbsp about 9 1 10 31 displaystyle 9 1 times 10 31 nbsp kg with a potential of V displaystyle V nbsp volts have an energy amount of e V displaystyle e cdot V nbsp joules The wavelength is 22 l h c e V 2 m c 2 e V displaystyle lambda frac hc sqrt eV 2mc 2 eV nbsp where c displaystyle c nbsp is the speed of light in vacuum about c 3 10 8 displaystyle c 3 times 10 8 nbsp m s See electron diffraction for a full explanation Types editTransmission electron microscope TEM edit Main article Transmission electron microscope nbsp Diagram of a transmission electron microscopeThe original form of the electron microscope the transmission electron microscope TEM uses a high voltage electron beam to illuminate the specimen and create an image An electron beam is produced by an electron gun with the electrons typically at 40 to 400 keV focused by electromagnetic lenses and transmitted through the specimen When it emerges from the specimen the electron beam carries information about the structure of the specimen that is magnified by lenses of the microscope The spatial variation in this information the image may be viewed by projecting the magnified electron image onto a detector For example the image may be viewed directly by an operator using a fluorescent viewing screen coated with a phosphor or scintillator material such as zinc sulfide A high resolution phosphor may also be coupled by means of a lens optical system or a fibre optic light guide to the sensor of a digital camera Direct electron detectors have no scintillator and are directly exposed to the electron beam which addresses some of the limitations of scintillator coupled cameras 23 The resolution of TEMs is limited primarily by spherical aberration but a new generation of hardware correctors can reduce spherical aberration to increase the resolution in high resolution transmission electron microscopy HRTEM to below 0 5 angstrom 50 picometres 24 enabling magnifications above 50 million times 25 The ability of HRTEM to determine the positions of atoms within materials is useful for nano technologies research and development 26 Main article Electron diffraction Transmission electron microscopes are often used in electron diffraction mode The advantages of electron diffraction over X ray crystallography are that the specimen need not be a single crystal or even a polycrystalline powder citation needed Scanning transmission electron microscope STEM edit Main article Scanning transmission electron microscopy The STEM rasters a focused incident probe across a specimen The high resolution of the TEM is thus possible in STEM The focusing action and aberrations occur before the electrons hit the specimen in the STEM but afterward in the TEM The STEMs use of SEM like beam rastering simplifies annular dark field imaging and other analytical techniques but also means that image data is acquired in serial rather than in parallel fashion citation needed Scanning electron microscope SEM edit source source source source source source source source Operating principle of a scanning electron microscopeMain article Scanning electron microscope nbsp Image of Bacillus subtilis taken with a 1960s electron microscopeThe SEM produces images by probing the specimen with a focused electron beam that is scanned across the specimen raster scanning When the electron beam interacts with the specimen it loses energy by a variety of mechanisms The lost energy is converted into alternative forms such as heat emission of low energy secondary electrons and high energy backscattered electrons light emission cathodoluminescence or X ray emission all of which provide signals carrying information about the properties of the specimen surface such as its topography and composition citation needed The image displayed by an SEM maps the varying intensity of any of these signals into the image in a position corresponding to the position of the beam on the specimen when the signal was generated In the SEM image of an ant shown the image was constructed from signals produced by a secondary electron detector the normal or conventional imaging mode in most SEMs citation needed Generally the image resolution of an SEM is lower than that of a TEM However because the SEM images the surface of a sample rather than its interior the electrons do not have to travel through the sample This reduces the need for extensive sample preparation to thin the specimen to electron transparency The SEM also has a great depth of field and so can produce images that are good representations of the three dimensional surface shape of the sample citation needed In their most common configurations electron microscopes produce images with a single brightness value per pixel with the results usually rendered in greyscale 27 However often these images are then colourized through the use of feature detection software or simply by hand editing using a graphics editor This may be done to clarify structure or for aesthetic effect and generally does not add new information about the specimen 28 Sample preparation for TEM edit nbsp An insect coated in gold for viewing with a scanning electron microscopeSee also TEM Sample preparation Ultramicrotomy Staining Cryofixation Chemical milling and Sputtering Materials to be viewed in a transmission electron microscope may require processing to produce a suitable sample The technique required varies depending on the specimen and the analysis required Chemical fixation for biological specimens this aims to stabilize the specimen s mobile macromolecular structure by chemical crosslinking of proteins with aldehydes such as formaldehyde and glutaraldehyde and lipids with osmium tetroxide 29 Cryofixation freezing a specimen so that the water forms vitreous non crystalline ice This preserves the specimen in a snapshot of its native state Methods to achieve this vitrification include plunge freezing rapidly in liquid ethane and high pressure freezing An entire field called cryo electron microscopy has branched from this technique With the development of cryo electron microscopy of vitreous sections CEMOVIS 30 and cryo focused ion beam milling of lamellae 31 it is now possible to observe samples from virtually any biological specimen close to its native state Dehydration replacement of water with organic solvents such as ethanol or acetone followed by critical point drying or infiltration with embedding resins See also freeze drying citation needed Embedding biological specimens after dehydration tissue for observation in the transmission electron microscope is embedded so it can be sectioned ready for viewing To do this the tissue is passed through a transition solvent such as propylene oxide epoxypropane or acetone and then infiltrated with an epoxy resin such as Araldite Epon or Durcupan 32 tissues may also be embedded directly in water miscible acrylic resin After the resin has been polymerized hardened the sample is sectioned by ultramicrotomy and stained citation needed Embedding materials after embedding in resin the specimen is usually ground and polished to a mirror like finish using ultra fine abrasives citation needed Freeze fracture or freeze etch a preparation method 33 34 35 particularly useful for examining lipid membranes and their incorporated proteins in face on view 36 37 38 nbsp Freeze fracturing helps to peel open membranes to allow visualization of what is inside nbsp External face of bakers yeast membrane showing the small holes where proteins are fractured out sometimes as small ring patterns The fresh tissue or cell suspension is frozen rapidly cryofixation then fractured by breaking 39 or by using a microtome 38 while maintained at liquid nitrogen temperature The cold fractured surface sometimes etched by increasing the temperature to about 100 C for several minutes to let some ice sublime 38 is then shadowed with evaporated platinum or gold at an average angle of 45 in a high vacuum evaporator The second coat of carbon evaporated perpendicular to the average surface plane is often performed to improve the stability of the replica coating The specimen is returned to room temperature and pressure then the extremely fragile pre shadowed metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids hypochlorite solution or SDS detergent The still floating replica is thoroughly washed free from residual chemicals carefully fished up on fine grids dried then viewed in the TEM citation needed Freeze fracture replica immunogold labeling FRIL the freeze fracture method has been modified to allow the identification of the components of the fracture face by immunogold labeling Instead of removing all the underlying tissue of the thawed replica as the final step before viewing in the microscope the tissue thickness is minimized during or after the fracture process The thin layer of tissue remains bound to the metal replica so it can be immunogold labeled with antibodies to the structures of choice The thin layer of the original specimen on the replica with gold attached allows the identification of structures in the fracture plane 40 There are also related methods which label the surface of etched cells 41 and other replica labeling variations 42 Ion beam milling thins samples until they are transparent to electrons by firing ions typically argon at the surface from an angle and sputtering material from the surface A subclass of this is focused ion beam milling where gallium ions are used to produce an electron transparent membrane or lamella in a specific region of the sample for example through a device within a microprocessor or a focused ion beam SEM Ion beam milling may also be used for cross section polishing prior to analysis of materials that are difficult to prepare using mechanical polishing citation needed Negative stain suspensions containing nanoparticles or fine biological material such as viruses and bacteria are briefly mixed with a dilute solution of an electron opaque solution such as ammonium molybdate uranyl acetate or formate or phosphotungstic acid citation needed This mixture is applied to an EM grid pre coated with a plastic film such as formvar blotted then allowed to dry Viewing of this preparation in the TEM should be carried out without delay for best results The method is important in microbiology for fast but crude morphological identification but can also be used as the basis for high resolution 3D reconstruction using EM tomography methodology when carbon films are used for support Negative staining is also used for observation of nanoparticles citation needed Sectioning produces thin slices of the specimen semitransparent to electrons These can be cut using ultramicrotomy on an ultramicrotome with a glass or diamond knife to produce ultra thin sections about 60 90 nm thick Disposable glass knives are also used because they can be made in the lab and are much cheaper Sections can also be created in situ by milling in a focused ion beam SEM where the section is known as a lamella 31 Staining uses heavy metals such as lead uranium or tungsten to scatter imaging electrons and thus give contrast between different structures since many especially biological materials are nearly transparent to electrons weak phase objects In biology specimens can be stained en bloc before embedding and also later after sectioning Typically thin sections are stained for several minutes with an aqueous or alcoholic solution of uranyl acetate followed by aqueous lead citrate 43 EM workflows editEarly electron microscopy of biological specimens was often descriptive making use of the newly available higher resolution 44 This is still the case for various applications such as diagnostic electron microscopy However electron microscopes are now frequently used in more complex workflows with each workflow typically using multiple technologies to enable more complex and or more quantitative analyses of a sample A few examples are outlined below but this should not be considered an exhaustive list The choice of workflow will be highly dependent on the application and the requirements of the corresponding scientific questions such as resolution volume nature of the target molecule etc For example images from light and electron microscopy of the same region of a sample can be overlaid to correlate the data from the two modalities This is commonly used to provide higher resolution contextual EM information about a fluorescently labelled structure This correlative light and electron microscopy CLEM 45 is one of a range of correlative workflows now available Another example is high resolution mass spectrometry ion microscopy which has been used to provide correlative information about subcellular antibiotic localisation 46 data that would be difficult to obtain by other means citation needed The initial role of electron microscopes in imaging two dimensional slices TEM or a specimen surface SEM with secondary electrons has also increasingly expanded into the depth of samples 47 An early example of these volume EM workflows was simply to stack TEM images of serial sections cut through a sample The next development was virtual reconstruction of a thick section 200 500 nm volume by backprojection of a set of images taken at different tilt angles TEM tomography 48 Serial imaging for volume EM edit To acquire volume EM datasets of larger depths than TEM tomography micrometers or millimeters in the z axis a series of images taken through the sample depth can be used For example ribbons of serial sections can be imaged in a TEM as described above and when thicker sections are used serial TEM tomography can be used to increase the z resolution More recently back scattered electron BSE images can be acquired of a larger series of sections collected on silicon wafers known as SEM array tomography 49 50 An alternative approach is to use BSE SEM to image the block surface instead of the section after each section has been removed By this method an ultramicrotome installed in an SEM chamber can increase automation of the workflow the specimen block is loaded in the chamber and the system programmed to continuously cut and image through the sample This is known as serial block face SEM 51 A related method uses focused ion beam milling instead of an ultramicrotome to remove sections In these serial imaging methods the output is essentially a sequence of images through a specimen block that can be digitally aligned in sequence and thus reconstructed into a volume EM dataset The increased volume available in these methods has expanded the capability of electron microscopy to address new questions 47 such as mapping neural connectivity in the brain 52 and membrane contact sites between organelles 53 Disadvantages edit nbsp JEOL transmission and scanning electron microscope made in the mid 1970sElectron microscopes are expensive to build and maintain Microscopes designed to achieve high resolutions must be housed in stable buildings sometimes underground with special services such as magnetic field canceling systems citation needed The samples largely have to be viewed in vacuum as the molecules that make up air would scatter the electrons An exception is liquid phase electron microscopy 54 using either a closed liquid cell or an environmental chamber for example in the environmental scanning electron microscope which allows hydrated samples to be viewed in a low pressure up to 20 Torr or 2 7 kPa wet environment Various techniques for in situ electron microscopy of gaseous samples have been developed 55 Scanning electron microscopes operating in conventional high vacuum mode usually image conductive specimens therefore non conductive materials require conductive coating gold palladium alloy carbon osmium etc The low voltage mode of modern microscopes makes possible the observation of non conductive specimens without coating Non conductive materials can be imaged also by a variable pressure or environmental scanning electron microscope citation needed Small stable specimens such as carbon nanotubes diatom frustules and small mineral crystals asbestos fibres for example require no special treatment before being examined in the electron microscope Samples of hydrated materials including almost all biological specimens have to be prepared in various ways to stabilize them reduce their thickness ultrathin sectioning and increase their electron optical contrast staining These processes may result in artifacts but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods Since the 1980s analysis of cryofixed vitrified specimens has also become increasingly used by scientists further confirming the validity of this technique 56 57 58 See also editList of materials analysis methods Electron diffraction Electron energy loss spectroscopy EELS Electron microscope images Energy filtered transmission electron microscopy EFTEM Environmental scanning electron microscope ESEM Immune electron microscopy In situ electron microscopy Low energy electron microscopy Microscope image processing Microscopy Nanotechnology Scanning confocal electron microscopy Scanning electron microscope SEM Thin section Transmission Electron Aberration Corrected MicroscopeReferences edit Calbick C J 1944 Historical Background of Electron Optics Journal of Applied Physics 15 10 685 690 Bibcode 1944JAP 15 685C doi 10 1063 1 1707371 ISSN 0021 8979 Hertz Heinrich 2019 Introduction to Heinrich Hertz s Miscellaneous Papers 1895 by Philipp Lenard Heinrich Rudolf Hertz 1857 1894 Routledge pp 87 88 doi 10 4324 9780429198960 4 ISBN 978 0 429 19896 0 S2CID 195494352 retrieved 2023 02 24 Wiechert E 1899 Experimentelle Untersuchungen uber die Geschwindigkeit und die magnetische Ablenkbarkeit der Kathodenstrahlen Annalen der Physik und Chemie in German 305 12 739 766 Bibcode 1899AnP 305 739W doi 10 1002 andp 18993051203 Wehnelt A 1905 X On the discharge of negative ions by glowing metallic oxides and allied phenomena The London Edinburgh and Dublin Philosophical Magazine and Journal of Science 10 55 80 90 doi 10 1080 14786440509463347 ISSN 1941 5982 Busch H 1926 Berechnung der Bahn von Kathodenstrahlen im axialsymmetrischen elektromagnetischen Felde Annalen der Physik in German 386 25 974 993 Bibcode 1926AnP 386 974B doi 10 1002 andp 19263862507 Dannen Gene 1998 Leo Szilard the Inventor A Slideshow 1998 Budapest conference talk dannen com Mulvey T 1962 Origins and historical development of the electron microscope British Journal of Applied Physics 13 5 197 207 doi 10 1088 0508 3443 13 5 303 ISSN 0508 3443 Tao Yaping 2018 A Historical Investigation of the Debates on the Invention and Invention Rights of Electron Microscope Proceedings of the 3rd International Conference on Contemporary Education Social Sciences and Humanities ICCESSH 2018 Atlantis Press pp 1438 1441 doi 10 2991 iccessh 18 2018 313 ISBN 978 94 6252 528 3 a href Template Cite book html title Template Cite book cite book a journal ignored help Freundlich Martin M 1963 Origin of the Electron Microscope The history of a great invention and of a misconception concerning the inventors is reviewed Science 142 3589 185 188 doi 10 1126 science 142 3589 185 ISSN 0036 8075 PMID 14057363 Rudenberg Reinhold 2010 Origin and Background of the Invention of the Electron Microscope Advances in Imaging and Electron Physics vol 160 Elsevier pp 171 205 doi 10 1016 s1076 5670 10 60005 5 ISBN 9780123810175 retrieved 2023 02 11 Knoll M Ruska E 1932 Beitrag zur geometrischen Elektronenoptik I Annalen der Physik 404 5 607 640 Bibcode 1932AnP 404 607K doi 10 1002 andp 19324040506 ISSN 0003 3804 Knoll M Ruska E 1932 Das Elektronenmikroskop Zeitschrift fur Physik in German 78 5 6 318 339 Bibcode 1932ZPhy 78 318K doi 10 1007 BF01342199 ISSN 1434 6001 S2CID 186239132 Rudenberg Reinhold Apparatus for producing images of objects Patent Public Search Basic Retrieved 24 February 2023 Rudenberg Reinhold Apparatus for producing images of objects Patent Public Search Basic Retrieved 24 February 2023 Rodenberg R 1932 Elektronenmikroskop Die Naturwissenschaften in German 20 28 522 Bibcode 1932NW 20 522R doi 10 1007 BF01505383 ISSN 0028 1042 S2CID 263996652 a b Ruska Ernst 1986 Ernst Ruska Autobiography Nobel Foundation Retrieved 2010 01 31 Kruger DH Schneck P Gelderblom HR May 2000 Helmut Ruska and the visualisation of viruses The Lancet 355 9216 1713 1717 doi 10 1016 S0140 6736 00 02250 9 PMID 10905259 S2CID 12347337 Ardenne M Von Beischer D 1940 Untersuchung von Metalloxyd Rauchen mit dem Universal Elektronenmikroskop Investigation of metal oxide smoking with the universal electron microscope Zeitschrift fur Elektrochemie und Angewandte Physikalische Chemie in German 46 4 270 277 doi 10 1002 bbpc 19400460406 S2CID 137136299 History of electron microscopy 1931 2000 Authors library caltech edu 2002 12 10 Retrieved on 2017 04 29 North America s first electron microscope James Hillier Inventor of the Week Archive 2003 05 01 Archived from the original on 2003 08 23 Retrieved 2010 01 31 Kirkland Earl 2010 Advanced computing in electron microscopy New York Springer ISBN 978 1 4419 6533 2 OCLC 668095602 Cheng Y Grigorieff N Penczek PA Walz T April 2015 A primer to single particle cryo electron microscopy Cell 161 3 438 449 doi 10 1016 j cell 2015 03 050 PMC 4409659 PMID 25910204 Erni Rolf Rossell MD Kisielowski C Dahmen U 2009 Atomic Resolution Imaging with a Sub 50 pm Electron Probe Physical Review Letters 102 9 096101 Bibcode 2009PhRvL 102i6101E doi 10 1103 PhysRevLett 102 096101 PMID 19392535 The Scale of Things Office of Basic Energy Sciences U S Department of Energy 2006 05 26 Archived from the original on 2010 02 01 Retrieved 2010 01 31 O Keefe MA Allard LF 2004 01 18 Sub Angstrom Electron Microscopy for Sub Angstrom Nano Metrology PDF Information Bridge DOE Scientific and Technical Information Sponsored by OSTI a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help Burgess Jeremy 1987 Under the Microscope A Hidden World Revealed CUP Archive p 11 ISBN 978 0 521 39940 1 Introduction to Electron Microscopy PDF FEI Company p 15 Retrieved 12 December 2012 Humbel Bruno M Schwarz Heinz Tranfield Erin M Fleck Roland A February 15 2019 Chapter 10 Chemical Fixation In Fleck Roland A Humbel Bruno M eds Biological Field Emission Scanning Electron Microscopy First Edition John Wiley amp Sons Ltd pp 191 221 doi 10 1002 9781118663233 ch10 ISBN 9781118663233 S2CID 243064180 Al Amoudi A Norlen LP Dubochet J October 2004 Cryo electron microscopy of vitreous sections of native biological cells and tissues J Struct Biol 148 1 131 5 doi 10 1016 j jsb 2004 03 010 PMID 15363793 a b Wagner FR Watanabe R Schampers R Singh D Persoon H Schaffer M Fruhstorfer P Plitzko J Villa E June 2020 Preparing samples from whole cells using focused ion beam milling for cryo electron tomography Nat Protoc 15 6 2041 2070 doi 10 1038 s41596 020 0320 x PMC 8053421 PMID 32405053 Luft J H 1961 Improvements in epoxy resin embedding methods The Journal of Biophysical and Biochemical Cytology Vol 9 no 2 p 409 PMC 2224998 PMID 13764136 Meryman H T and Kafig E 1955 The study of frozen specimens ice crystals and ices crystal growth by electron microscopy Naval Med Res Ints Rept NM 000 018 01 09 Vol 13 pp 529 544 Steere Russell L 1957 01 25 Electron microscopy of structural detail in frozen biological specimens The Journal of Biophysical and Biochemical Cytology 3 1 45 60 doi 10 1083 jcb 3 1 45 PMC 2224015 PMID 13416310 Isailovic Tanja M Todosijevic Marija N Đorđevic Sanela M Savic Snezana D 2017 01 01 Calija Bojan ed Chapter 7 Natural Surfactants Based Micro Nanoemulsion Systems for NSAIDs Practical Formulation Approach Physicochemical and Biopharmaceutical Characteristics Performances Microsized and Nanosized Carriers for Nonsteroidal Anti Inflammatory Drugs Boston Academic Press pp 179 217 doi 10 1016 b978 0 12 804017 1 00007 8 ISBN 978 0 12 804017 1 retrieved 2020 10 22 Moor H Muhlethaler K 1963 Fine structure in frozen etched yeast cells The Journal of Cell Biology 17 3 609 628 doi 10 1083 jcb 17 3 609 PMC 2106217 PMID 19866628 Black Joel A 1990 01 01 Conn P Michael ed 20 Use of Freeze Fracture in Neurobiology Methods in Neurosciences Quantitative and Qualitative Microscopy Academic Press 3 343 360 doi 10 1016 b978 0 12 185255 9 50025 0 ISBN 9780121852559 retrieved 2020 10 22 a b c Stillwell William 2016 01 01 Stillwell William ed Chapter 11 Long Range Membrane Properties An Introduction to Biological Membranes Second Edition Elsevier pp 221 245 doi 10 1016 b978 0 444 63772 7 00011 7 ISBN 978 0 444 63772 7 retrieved 2020 10 22 Bullivant Stanley Ames Adelbert 1966 06 01 A simple freeze fracture replication method for electron microscopy The Journal of Cell Biology 29 3 435 447 doi 10 1083 jcb 29 3 435 PMC 2106967 PMID 5962938 Gruijters W T Kistler J Bullivant S Goodenough D A 1987 03 01 Immunolocalization of MP70 in lens fiber 16 17 nm intercellular junctions The Journal of Cell Biology 104 3 565 572 doi 10 1083 jcb 104 3 565 PMC 2114558 PMID 3818793 da Silva Pedro Pinto Branton Daniel 1970 06 01 Membrane splitting in freeze etching The Journal of Cell Biology 45 3 598 605 doi 10 1083 jcb 45 3 598 PMC 2107921 PMID 4918216 Rash J E Johnson T J Hudson C S Giddings F D Graham W F Eldefrawi M E 1982 11 01 Labelled replica techniques post shadow labelling of intramembrane particles in freeze fracture replicas Journal of Microscopy 128 Pt 2 121 138 doi 10 1111 j 1365 2818 1982 tb00444 x PMID 6184475 S2CID 45238172 Reynolds E S 1963 The use of lead citrate at high pH as an electron opaque stain in electron microscopy Journal of Cell Biology 17 1 208 212 doi 10 1083 jcb 17 1 208 PMC 2106263 PMID 13986422 Sjostrand FS Hanzon V November 1954 Ultrastructure of Golgi apparatus of exocrine cells of mouse pancreas Exp Cell Res 7 2 415 29 doi 10 1016 s0014 4827 54 80087 5 PMID 13220587 Methods in Cell Biology Correlative Light and Electron Microscopy III ScienceDirect com by Elsevier Finin P Khan RM Oh S Boshoff HI Barry CE May 2023 Chemical approaches to unraveling the biology of mycobacteria Cell Chem Biol 30 5 420 435 doi 10 1016 j chembiol 2023 04 014 PMC 10201459 PMID 37207631 a b Peddie CJ Genoud C Kreshuk A Meechan K Micheva KD Narayan K Pape C Parton RG Schieber NL Schwab Y Titze B Verkade P Aubrey A Collinson LM July 2022 Volume electron microscopy Nat Rev Methods Primers 2 51 doi 10 1038 s43586 022 00131 9 PMC 7614724 PMID 37409324 Crowther RA Amos LA Finch JT De Rosier DJ Klug A May 1970 Three dimensional reconstructions of spherical viruses by fourier synthesis from electron micrographs Nature 226 5244 421 5 Bibcode 1970Natur 226 421C doi 10 1038 226421a0 PMID 4314822 S2CID 4217806 White IJ Burden JJ 2023 A practical guide to starting SEM array tomography An accessible volume EM technique Chapter 7 A practical guide to starting SEM array tomography An accessible volume EM technique Methods in Cell Biology Vol 177 pp 171 196 doi 10 1016 bs mcb 2022 12 023 ISBN 9780323916073 PMID 37451766 Kolotuev I August 2023 Work smart not hard How array tomography can help increase the ultrastructure data output J Microsc doi 10 1111 jmi 13217 PMID 37626455 S2CID 261174348 Denk W Horstmann H November 2004 Serial block face scanning electron microscopy to reconstruct three dimensional tissue nanostructure PLOS Biol 2 11 e329 doi 10 1371 journal pbio 0020329 PMC 524270 PMID 15514700 Abbott LF Bock DD Callaway EM Denk W Dulac C Fairhall AL Fiete I Harris KM Helmstaedter M Jain V Kasthuri N LeCun Y Lichtman JW Littlewood PB Luo L Maunsell JH Reid RC Rosen BR Rubin GM Sejnowski TJ Seung HS Svoboda K Tank DW Tsao D Van Essen DC September 2020 The Mind of a Mouse Cell 182 6 1372 1376 doi 10 1016 j cell 2020 08 010 PMID 32946777 S2CID 221766693 Prinz WA Toulmay A Balla T January 2020 The functional universe of membrane contact sites Nat Rev Mol Cell Biol 21 1 7 24 doi 10 1038 s41580 019 0180 9 PMC 10619483 PMID 31732717 S2CID 208019972 de Jonge N Ross F M 2011 Electron microscopy of specimens in liquid Nature Nanotechnology 6 8 695 704 Bibcode 2003NatMa 2 532W doi 10 1038 nmat944 PMID 12872162 S2CID 21379512 Gai P L Boyes E D 2009 Advances in atomic resolution in situ environmental transmission electron microscopy and 1A aberration corrected in situ electron microscopy Microsc Res Tech 72 3 153 164 arXiv 1705 05754 doi 10 1002 jemt 20668 PMID 19140163 S2CID 1746538 Adrian Marc Dubochet Jacques Lepault Jean McDowall Alasdair W 1984 Cryo electron microscopy of viruses Nature Submitted manuscript 308 5954 32 36 Bibcode 1984Natur 308 32A doi 10 1038 308032a0 PMID 6322001 S2CID 4319199 Sabanay I Arad T Weiner S Geiger B 1991 Study of vitrified unstained frozen tissue sections by cryoimmunoelectron microscopy Journal of Cell Science 100 1 227 236 doi 10 1242 jcs 100 1 227 PMID 1795028 Kasas S Dumas G Dietler G Catsicas S Adrian M 2003 Vitrification of cryoelectron microscopy specimens revealed by high speed photographic imaging Journal of Microscopy 211 1 48 53 doi 10 1046 j 1365 2818 2003 01193 x PMID 12839550 S2CID 40058086 nbsp Wikimedia Commons has media related to Electron microscopes An Introduction to Microscopy Archived 2013 07 19 at the Wayback Machine resources for teachers and students Cell Centered Database Electron microscopy data Science Aid Electron Microscopy By Kaden park Retrieved from https en wikipedia org w index php title Electron microscope amp oldid 1193823911, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.