Scanning electron microscope

A scanning electron microscope (SEM) is a type of electron microscope that produces images of a sample by scanning the surface with a focused beam of electrons. The electrons interact with atoms in the sample, producing various signals that contain information about the surface topography and composition of the sample. The electron beam is scanned in a raster scan pattern, and the position of the beam is combined with the intensity of the detected signal to produce an image. In the most common SEM mode, secondary electrons emitted by atoms excited by the electron beam are detected using a secondary electron detector (Everhart–Thornley detector). The number of secondary electrons that can be detected, and thus the signal intensity, depends, among other things, on specimen topography. Some SEMs can achieve resolutions better than 1 nanometer.

Image of pollen grains taken on a SEM shows the characteristic depth of field of SEM micrographs
M. von Ardenne's first SEM
Operating principle of a Scanning Electron Microscope (SEM)
SEM with opened sample chamber
Analog type SEM

Specimens are observed in high vacuum in a conventional SEM, or in low vacuum or wet conditions in a variable pressure or environmental SEM, and at a wide range of cryogenic or elevated temperatures with specialized instruments.[1]

History Edit

An account of the early history of scanning electron microscopy has been presented by McMullan.[2][3] Although Max Knoll produced a photo with a 50 mm object-field-width showing channeling contrast by the use of an electron beam scanner,[4] it was Manfred von Ardenne who in 1937 invented[5] a microscope with high resolution by scanning a very small raster with a demagnified and finely focused electron beam. Ardenne applied scanning of the electron beam in an attempt to surpass the resolution of the transmission electron microscope (TEM), as well as to mitigate substantial problems with chromatic aberration inherent to real imaging in the TEM. He further discussed the various detection modes, possibilities and theory of SEM,[6] together with the construction of the first high resolution SEM.[7] Further work was reported by Zworykin's group,[8] followed by the Cambridge groups in the 1950s and early 1960s[9][10][11][12] headed by Charles Oatley, all of which finally led to the marketing of the first commercial instrument by Cambridge Scientific Instrument Company as the "Stereoscan" in 1965, which was delivered to DuPont.

Principles and capacities Edit

 
Schottky-emitter electron source
 
Electron–matter interaction volume and types of signal generated

The signals used by a SEM to produce an image result from interactions of the electron beam with atoms at various depths within the sample. Various types of signals are produced including secondary electrons (SE), reflected or back-scattered electrons (BSE), characteristic X-rays and light (cathodoluminescence) (CL), absorbed current (specimen current) and transmitted electrons. Secondary electron detectors are standard equipment in all SEMs, but it is rare for a single machine to have detectors for all other possible signals.[citation needed]

Secondary electrons have very low energies on the order of 50 eV, which limits their mean free path in solid matter. Consequently, SEs can only escape from the top few nanometers of the surface of a sample. The signal from secondary electrons tends to be highly localized at the point of impact of the primary electron beam, making it possible to collect images of the sample surface with a resolution of below 1 nm. Back-scattered electrons (BSE) are beam electrons that are reflected from the sample by elastic scattering. Since they have much higher energy than SEs, they emerge from deeper locations within the specimen and, consequently, the resolution of BSE images is less than SE images. However, BSE are often used in analytical SEM, along with the spectra made from the characteristic X-rays, because the intensity of the BSE signal is strongly related to the atomic number (Z) of the specimen. BSE images can provide information about the distribution, but not the identity, of different elements in the sample. In samples predominantly composed of light elements, such as biological specimens, BSE imaging can image colloidal gold immuno-labels of 5 or 10 nm diameter, which would otherwise be difficult or impossible to detect in secondary electron images.[13] Characteristic X-rays are emitted when the electron beam removes an inner shell electron from the sample, causing a higher-energy electron to fill the shell and release energy. The energy or wavelength of these characteristic X-rays can be measured by Energy-dispersive X-ray spectroscopy or Wavelength-dispersive X-ray spectroscopy and used to identify and measure the abundance of elements in the sample and map their distribution.

Due to the very narrow electron beam, SEM micrographs have a large depth of field yielding a characteristic three-dimensional appearance useful for understanding the surface structure of a sample.[14] This is exemplified by the micrograph of pollen shown above. A wide range of magnifications is possible, from about 10 times (about equivalent to that of a powerful hand-lens) to more than 500,000 times, about 250 times the magnification limit of the best light microscopes.

Sample preparation Edit

 
A spider sputter-coated in gold, having been prepared for viewing with an SEM
 
Low-voltage micrograph (300 V) of distribution of adhesive droplets on a Post-it note. No conductive coating was applied: such a coating would alter this fragile specimen.

SEM samples have to be small enough to fit on the specimen stage, and may need special preparation to increase their electrical conductivity and to stabilize them, so that they can withstand the high vacuum conditions and the high energy beam of electrons. Samples are generally mounted rigidly on a specimen holder or stub using a conductive adhesive. SEM is used extensively for defect analysis of semiconductor wafers, and manufacturers make instruments that can examine any part of a 300 mm semiconductor wafer. Many instruments have chambers that can tilt an object of that size to 45° and provide continuous 360° rotation.[citation needed]

Nonconductive specimens collect charge when scanned by the electron beam, and especially in secondary electron imaging mode, this causes scanning faults and other image artifacts. For conventional imaging in the SEM, specimens must be electrically conductive, at least at the surface, and electrically grounded to prevent the accumulation of electrostatic charge. Metal objects require little special preparation for SEM except for cleaning and conductively mounting to a specimen stub. Non-conducting materials are usually coated with an ultrathin coating of electrically conducting material, deposited on the sample either by low-vacuum sputter coating, electroless deposition[citation needed] or by high-vacuum evaporation. Conductive materials in current use for specimen coating include gold, gold/palladium alloy, platinum, iridium, tungsten, chromium, osmium,[13] and graphite. Coating with heavy metals may increase signal/noise ratio for samples of low atomic number (Z). The improvement arises because secondary electron emission for high-Z materials is enhanced.[citation needed]

An alternative to coating for some biological samples is to increase the bulk conductivity of the material by impregnation with osmium using variants of the OTO staining method (O-osmium tetroxide, T-thiocarbohydrazide, O-osmium).[15][16]

Nonconducting specimens may be imaged without coating using an environmental SEM (ESEM) or low-voltage mode of SEM operation. In ESEM instruments the specimen is placed in a relatively high-pressure chamber and the electron optical column is differentially pumped to keep vacuum adequately[clarification needed] low at the electron gun. The high-pressure region around the sample in the ESEM neutralizes charge and provides an amplification of the secondary electron signal.[citation needed] Low-voltage SEM is typically conducted in an instrument with a field emission guns (FEG) which is capable of producing high primary electron brightness and small spot size even at low accelerating potentials. To prevent charging of non-conductive specimens, operating conditions must be adjusted such that the incoming beam current is equal to sum of outgoing secondary and backscattered electron currents, a condition that is most often met at accelerating voltages of 0.3–4 kV.[citation needed]

Embedding in a resin with further polishing to a mirror-like finish can be used for both biological and materials specimens when imaging in backscattered electrons or when doing quantitative X-ray microanalysis.

The main preparation techniques are not required in the environmental SEM outlined below, but some biological specimens can benefit from fixation.

Biological samples Edit

Conventionally, a SEM specimen is required to be completely dry, since the specimen chamber is under high vacuum. Hard, dry materials such as wood, bone, feathers, dried insects, or shells (including egg shells[17]) can be examined with little further treatment, but living cells and tissues and whole, soft-bodied organisms require chemical fixation to preserve and stabilize their structure.

Fixation is usually performed by incubation in a solution of a buffered chemical fixative, such as glutaraldehyde, sometimes in combination with formaldehyde[18][19][20] and other fixatives,[21] and optionally followed by postfixation with osmium tetroxide.[18] The fixed tissue is then dehydrated. Because air-drying causes collapse and shrinkage, this is commonly achieved by replacement of water in the cells with organic solvents such as ethanol or acetone, and replacement of these solvents in turn with a transitional fluid such as liquid carbon dioxide by critical point drying.[22] The carbon dioxide is finally removed while in a supercritical state, so that no gas–liquid interface is present within the sample during drying.

The dry specimen is usually mounted on a specimen stub using an adhesive such as epoxy resin or electrically conductive double-sided adhesive tape, and sputter-coated with gold or gold/palladium alloy before examination in the microscope. Samples may be sectioned (with a microtome) if information about the organism's internal ultrastructure is to be exposed for imaging.

If the SEM is equipped with a cold stage for cryo microscopy, cryofixation may be used and low-temperature scanning electron microscopy performed on the cryogenically fixed specimens.[18] Cryo-fixed specimens may be cryo-fractured under vacuum in a special apparatus to reveal internal structure, sputter-coated and transferred onto the SEM cryo-stage while still frozen.[23] Low-temperature scanning electron microscopy (LT-SEM) is also applicable to the imaging of temperature-sensitive materials such as ice[24][25] and fats.[26]

Freeze-fracturing, freeze-etch or freeze-and-break is a preparation method particularly useful for examining lipid membranes and their incorporated proteins in "face on" view. The preparation method reveals the proteins embedded in the lipid bilayer.

Materials Edit

Back-scattered electron imaging, quantitative X-ray analysis, and X-ray mapping of specimens often requires grinding and polishing the surfaces to an ultra-smooth surface. Specimens that undergo WDS or EDS analysis are often carbon-coated. In general, metals are not coated prior to imaging in the SEM because they are conductive and provide their own pathway to ground. Fractography is the study of fractured surfaces that can be done on a light microscope or, commonly, on an SEM. The fractured surface is cut to a suitable size, cleaned of any organic residues, and mounted on a specimen holder for viewing in the SEM. Integrated circuits may be cut with a focused ion beam (FIB) or other ion beam milling instrument for viewing in the SEM. The SEM in the first case may be incorporated into the FIB, enabling high-resolution imaging of the result of the process. Metals, geological specimens, and integrated circuits all may also be chemically polished for viewing in the SEM. Special high-resolution coating techniques are required for high-magnification imaging of inorganic thin films.

Scanning process and image formation Edit

 
Schematic of an SEM

In a typical SEM, an electron beam is thermionically emitted from an electron gun fitted with a tungsten filament cathode. Tungsten is normally used in thermionic electron guns because it has the highest melting point and lowest vapor pressure of all metals, thereby allowing it to be electrically heated for electron emission, and because of its low cost. Other types of electron emitters include lanthanum hexaboride (LaB
6
) cathodes, which can be used in a standard tungsten filament SEM if the vacuum system is upgraded, or field emission guns (FEG), which may be of the cold-cathode type using tungsten single crystal emitters or the thermally assisted Schottky type, that use emitters of tungsten single crystals coated in zirconium oxide.

The electron beam, which typically has an energy ranging from 0.2 keV to 40 keV, is focused by one or two condenser lenses to a spot about 0.4 nm to 5 nm in diameter. The beam passes through pairs of scanning coils or pairs of deflector plates in the electron column, typically in the final lens, which deflect the beam in the x and y axes so that it scans in a raster fashion over a rectangular area of the sample surface.

 
Mechanisms of emission of secondary electrons, backscattered electrons, and characteristic X-rays from atoms of the sample

When the primary electron beam interacts with the sample, the electrons lose energy by repeated random scattering and absorption within a teardrop-shaped volume of the specimen known as the interaction volume, which extends from less than 100 nm to approximately 5 µm into the surface. The size of the interaction volume depends on the electron's landing energy, the atomic number of the specimen, and the specimen's density. The energy exchange between the electron beam and the sample results in the reflection of high-energy electrons by elastic scattering, the emission of secondary electrons by inelastic scattering, and the emission of electromagnetic radiation, each of which can be detected by specialized detectors. The beam current absorbed by the specimen can also be detected and used to create images of the distribution of specimen current. Electronic amplifiers of various types are used to amplify the signals, which are displayed as variations in brightness on a computer monitor (or, for vintage models, on a cathode-ray tube). Each pixel of computer video memory is synchronized with the position of the beam on the specimen in the microscope, and the resulting image is, therefore, a distribution map of the intensity of the signal being emitted from the scanned area of the specimen. Older microscopes captured images on film, but most modern instruments collect digital images.

 
Low-temperature SEM magnification series for a snow crystal. The crystals are captured, stored, and sputter-coated with platinum at cryogenic temperatures for imaging.

Magnification Edit

Magnification in an SEM can be controlled over a range of about 6 orders of magnitude from about 10 to 3,000,000 times.[27] Unlike optical and transmission electron microscopes, image magnification in an SEM is not a function of the power of the objective lens. SEMs may have condenser and objective lenses, but their function is to focus the beam to a spot, and not to image the specimen. Provided the electron gun can generate a beam with a sufficiently small diameter, an SEM could in principle work entirely without condenser or objective lenses. However, it might not be very versatile or achieve very high resolution. In an SEM, as in scanning probe microscopy, magnification results from the ratio of the raster on the display device and dimensions of the raster on the specimen. Assuming that the display screen has a fixed size, higher magnification results from reducing the size of the raster on the specimen, and vice versa. Magnification is therefore controlled by the current supplied to the x, y scanning coils, or the voltage supplied to the x, y deflector plates, and not by objective lens power.

Detection of secondary electrons Edit

The most common imaging mode collects low-energy (<50 eV) secondary electrons that are ejected from conduction or valence bands of the specimen atoms by inelastic scattering interactions with beam electrons. Due to their low energy, these electrons originate from within a few nanometers below the sample surface.[14] The electrons are detected by an Everhart–Thornley detector,[28] which is a type of collector-scintillator-photomultiplier system. The secondary electrons are first collected by attracting them towards an electrically biased grid at about +400 V, and then further accelerated towards a phosphor or scintillator positively biased to about +2,000 V. The accelerated secondary electrons are now sufficiently energetic to cause the scintillator to emit flashes of light (cathodoluminescence), which are conducted to a photomultiplier outside the SEM column via a light pipe and a window in the wall of the specimen chamber. The amplified electrical signal output by the photomultiplier is displayed as a two-dimensional intensity distribution that can be viewed and photographed on an analogue video display, or subjected to analog-to-digital conversion and displayed and saved as a digital image. This process relies on a raster-scanned primary beam. The brightness of the signal depends on the number of secondary electrons reaching the detector. If the beam enters the sample perpendicular to the surface, then the activated region is uniform about the axis of the beam and a certain number of electrons "escape" from within the sample. As the angle of incidence increases, the interaction volume increases and the "escape" distance of one side of the beam decreases, resulting in more secondary electrons being emitted from the sample. Thus steep surfaces and edges tend to be brighter than flat surfaces, which results in images with a well-defined, three-dimensional appearance. Using the signal of secondary electrons image resolution less than 0.5 nm is possible.

Detection of backscattered electrons Edit

 
Comparison of SEM techniques:
Top: backscattered electron analysis – composition
Bottom: secondary electron analysis – topography

Backscattered electrons (BSE) consist of high-energy electrons originating in the electron beam, that are reflected or back-scattered out of the specimen interaction volume by elastic scattering interactions with specimen atoms. Since heavy elements (high atomic number) backscatter electrons more strongly than light elements (low atomic number), and thus appear brighter in the image, BSEs are used to detect contrast between areas with different chemical compositions.[14] The Everhart–Thornley detector, which is normally positioned to one side of the specimen, is inefficient for the detection of backscattered electrons because few such electrons are emitted in the solid angle subtended by the detector, and because the positively biased detection grid has little ability to attract the higher energy BSE. Dedicated backscattered electron detectors are positioned above the sample in a "doughnut" type arrangement, concentric with the electron beam, maximizing the solid angle of collection. BSE detectors are usually either of scintillator or of semiconductor types. When all parts of the detector are used to collect electrons symmetrically about the beam, atomic number contrast is produced. However, strong topographic contrast is produced by collecting back-scattered electrons from one side above the specimen using an asymmetrical, directional BSE detector; the resulting contrast appears as illumination of the topography from that side. Semiconductor detectors can be made in radial segments that can be switched in or out to control the type of contrast produced and its directionality.

Backscattered electrons can also be used to form an electron backscatter diffraction (EBSD) image that can be used to determine the crystallographic structure of the specimen.

Beam-injection analysis of semiconductors Edit

The nature of the SEM's probe, energetic electrons, makes it uniquely suited to examining the optical and electronic properties of semiconductor materials. The high-energy electrons from the SEM beam will inject charge carriers into the semiconductor. Thus, beam electrons lose energy by promoting electrons from the valence band into the conduction band, leaving behind holes.

In a direct bandgap material, recombination of these electron-hole pairs will result in cathodoluminescence; if the sample contains an internal electric field, such as is present at a p-n junction, the SEM beam injection of carriers will cause electron beam induced current (EBIC) to flow. Cathodoluminescence and EBIC are referred to as "beam-injection" techniques, and are very powerful probes of the optoelectronic behavior of semiconductors, in particular for studying nanoscale features and defects.

Cathodoluminescence Edit

 
Color cathodoluminescence overlay on SEM image of an InGaN polycrystal. The blue and green channels represent real colors, the red channel corresponds to UV emission.

Cathodoluminescence, the emission of light when atoms excited by high-energy electrons return to their ground state, is analogous to UV-induced fluorescence, and some materials such as zinc sulfide and some fluorescent dyes, exhibit both phenomena. Over the last decades, cathodoluminescence was most commonly experienced as the light emission from the inner surface of the cathode-ray tube in television sets and computer CRT monitors. In the SEM, CL detectors either collect all light emitted by the specimen or can analyse the wavelengths emitted by the specimen and display an emission spectrum or an image of the distribution of cathodoluminescence emitted by the specimen in real color.

X-ray microanalysis Edit

Characteristic X-rays that are produced by the interaction of electrons with the sample may also be detected in an SEM equipped for energy-dispersive X-ray spectroscopy or wavelength dispersive X-ray spectroscopy. Analysis of the x-ray signals may be used to map the distribution and estimate the abundance of elements in the sample.

Resolution of the SEM Edit

A video illustrating a typical practical magnification range of a scanning electron microscope designed for biological specimens. The video starts at 25×, about 6 mm across the whole field of view, and zooms in to 12000×, about 12 μm across the whole field of view. The spherical objects are glass beads with a diameter of 10 μm, similar in diameter to a red blood cell.

SEM is not a camera and the detector is not continuously image-forming like a CCD array or film. Unlike in an optical system, the resolution is not limited by the diffraction limit, fineness of lenses or mirrors or detector array resolution. The focusing optics can be large and coarse, and the SE detector is fist-sized and simply detects current. Instead, the spatial resolution of the SEM depends on the size of the electron spot, which in turn depends on both the wavelength of the electrons and the electron-optical system that produces the scanning beam. The resolution is also limited by the size of the interaction volume, the volume of specimen material that interacts with the electron beam. The spot size and the interaction volume are both large compared to the distances between atoms, so the resolution of the SEM is not high enough to image individual atoms, as is possible with a transmission electron microscope (TEM). The SEM has compensating advantages, though, including the ability to image a comparatively large area of the specimen; the ability to image bulk materials (not just thin films or foils); and the variety of analytical modes available for measuring the composition and properties of the specimen. Depending on the instrument, the resolution can fall somewhere between less than 1 nm and 20 nm. As of 2009, The world's highest resolution conventional (≤30 kV) SEM can reach a point resolution of 0.4 nm using a secondary electron detector.[29]

Environmental SEM Edit

Conventional SEM requires samples to be imaged under vacuum, because a gas atmosphere rapidly spreads and attenuates electron beams. As a consequence, samples that produce a significant amount of vapour, e.g. wet biological samples or oil-bearing rock, must be either dried or cryogenically frozen. Processes involving phase transitions, such as the drying of adhesives or melting of alloys, liquid transport, chemical reactions, and solid-air-gas systems, in general cannot be observed with conventional high-vacuum SEM. In environmental SEM (ESEM), the chamber is evacuated of air, but water vapor is retained near its saturation pressure, and the residual pressure remains relatively high. This allows the analysis of samples containing water or other volatile substances. With ESEM, observations of living insects have been possible.[30]

The first commercial development of the ESEM in the late 1980s[31][32] allowed samples to be observed in low-pressure gaseous environments (e.g. 1–50 Torr or 0.1–6.7 kPa) and high relative humidity (up to 100%). This was made possible by the development of a secondary-electron detector[33][34] capable of operating in the presence of water vapour and by the use of pressure-limiting apertures with differential pumping in the path of the electron beam to separate the vacuum region (around the gun and lenses) from the sample chamber. The first commercial ESEMs were produced by the ElectroScan Corporation in USA in 1988. ElectroScan was taken over by Philips (who later sold their electron-optics division to FEI Company) in 1996.[35]

ESEM is especially useful for non-metallic and biological materials because coating with carbon or gold is unnecessary. Uncoated plastics and elastomers can be routinely examined, as can uncoated biological samples. This is useful because coating can be difficult to reverse, may conceal small features on the surface of the sample and may reduce the value of the results obtained. X-ray analysis is difficult with a coating of a heavy metal, so carbon coatings are routinely used in conventional SEMs, but ESEM makes it possible to perform X-ray microanalysis on uncoated non-conductive specimens; however some specific for ESEM artifacts are introduced in X-ray analysis. ESEM may be the preferred for electron microscopy of unique samples from criminal or civil actions, where forensic analysis may need to be repeated by several different experts. It is possible to study specimens in liquid with ESEM or with other liquid-phase electron microscopy methods.[36]

Transmission SEM Edit

The SEM can also be used in transmission mode by simply incorporating an appropriate detector below a thin specimen section.[37] Detectors are available for bright field, dark field, as well as segmented detectors for mid-field to high angle annular dark-field. Despite the difference in instrumentation, this technique is still commonly referred to as scanning transmission electron microscopy (STEM).

SEM in Forensic Science Edit

The SEM is used often in Forensic Science for magnified analysis of microscopic things such as diatoms and gunshot residue. Because SEM is a nondestructive force on the sample, it can be used to analyze evidence without damaging it. The SEM shoots a beam of high energy electrons to the sample which bounce off of the sample without changing or destroying it. This is great when it comes to analyzing diatoms. When a person dies by drowning, they inhale the water which causes what is in the water (diatoms) to get in the blood stream, brain, kidneys, and more. These diatoms in the body can be magnified with the SEM to determine the type of diatoms which aid in understanding how and where the person died. By using the images produced by the SEM, forensic scientists can compare diatoms types to confirm the body of water a person died in.[38]

Gunshot residue (GSR) analysis can be done with many different analytical instruments,[39] but SEM is a common way to analyze inorganic compounds because of the way it can closely analyze the types of elements (mostly metals) through its three detectors: backscatter electron detector, secondary electron detector, and X-ray detector. GSR can be collected from the crime scene, victim, or shooter and analyzed with the SEM. This can help scientists determine proximity and or contact with the discharged firearm.[39]

Color in SEM Edit

Electron microscopes do not naturally produce color images, as an SEM produces a single value per pixel; this value corresponds to the number of electrons received by the detector during a small period of time of the scanning when the beam is targeted to the (x, y) pixel position.

This single number is usually represented, for each pixel, by a grey level, forming a monochrome image.[40] However, several ways have been used to get color electron microscopy images.[41]

False color using a single detector Edit

  • On compositional images of flat surfaces (typically BSE):

The easiest way to get color is to associate to this single number an arbitrary color, using a color look-up table (i.e. each grey level is replaced by a chosen color). This method is known as false color. On a BSE image, false color may be performed to better distinguish the various phases of the sample.[42]

  • On textured-surface images:

As an alternative to simply replacing each grey level by a color, a sample observed by an oblique beam allows researchers to create an approximative topography image (see further section "Photometric 3D rendering from a single SEM image"). Such topography can then be processed by 3D-rendering algorithms for a more natural rendering of the surface texture.

SEM image coloring Edit

Very often, published SEM images are artificially colored.[42] This may be done for aesthetic effect, to clarify structure or to add a realistic appearance to the sample and generally does not add information about the specimen.[43]

Coloring may be performed manually with photo-editing software, or semi-automatically with dedicated software using feature-detection or object-oriented segmentation.[44]

Color built using multiple electron detectors Edit

In some configurations more information is gathered per pixel, often by the use of multiple detectors.[45]

As a common example, secondary electron and backscattered electron detectors are superimposed and a color is assigned to each of the images captured by each detector,[46][47] with a result of a combined color image where colors are related to the density of the components. This method is known as density-dependent color SEM (DDC-SEM). Micrographs produced by DDC-SEM retain topographical information, which is better captured by the secondary electrons detector and combine it to the information about density, obtained by the backscattered electron detector.[48][49]

Analytical signals based on generated photons Edit

Measurement of the energy of photons emitted from the specimen is a common method to get analytical capabilities. Examples are the energy-dispersive X-ray spectroscopy (EDS) detectors used in elemental analysis and cathodoluminescence microscope (CL) systems that analyse the intensity and spectrum of electron-induced luminescence in (for example) geological specimens. In SEM systems using these detectors it is common to color code these extra signals and superimpose them in a single color image, so that differences in the distribution of the various components of the specimen can be seen clearly and compared. Optionally, the standard secondary electron image can be merged with the one or more compositional channels, so that the specimen's structure and composition can be compared. Such images can be made while maintaining the full integrity of the original signal data, which is not modified in any way.

3D in SEM Edit

SEMs do not naturally provide 3D images contrary to SPMs. However 3D data can be obtained using an SEM with different methods as follows.

3D SEM reconstruction from a stereo pair Edit

  • photogrammetry is the most metrologically accurate method to bring the third dimension to SEM images.[42] Contrary to photometric methods (next paragraph), photogrammetry calculates absolute heights using triangulation methods. The drawbacks are that it works only if there is a minimum texture, and it requires two images to be acquired from two different angles, which implies the use of a tilt stage. (Photogrammetry is a software operation that calculates the shift (or "disparity") for each pixel, between the left image and the right image of the same pair. Such disparity reflects the local height).

Photometric 3D SEM reconstruction from a four-quadrant detector by "shape from shading" Edit

This method typically uses a four-quadrant BSE detector (alternatively for one manufacturer, a 3-segment detector). The microscope produces four images of the same specimen at the same time, so no tilt of the sample is required. The method gives metrological 3D dimensions as far as the slope of the specimen remains reasonable.[42] Most SEM manufacturers now (2018) offer such a built-in or optional four-quadrant BSE detector, together with proprietary software to calculate a 3D image in real time.[51]

Other approaches use more sophisticated (and sometimes GPU-intensive) methods like the optimal estimation algorithm and offer much better results[52] at the cost of high demands on computing power.

In all instances, this approach works by integration of the slope, so vertical slopes and overhangs are ignored; for instance, if an entire sphere lies on a flat, little more than the upper hemisphere is seen emerging above the flat, resulting in wrong altitude of the sphere apex. The prominence of this effect depends on the angle of the BSE detectors with respect to the sample, but these detectors are usually situated around (and close to) the electron beam, so this effect is very common.

Photometric 3D rendering from a single SEM image Edit

This method requires an SEM image obtained in oblique low angle lighting. The grey-level is then interpreted as the slope, and the slope integrated to restore the specimen topography. This method is interesting for visual enhancement and the detection of the shape and position of objects; however the vertical heights cannot usually be calibrated, contrary to other methods such as photogrammetry.[42]

Other types of 3D SEM reconstruction Edit

  • Inverse reconstruction using electron-material interactive models[53][54]
  • Multi-Resolution reconstruction using single 2D File: High-quality 3D imaging may be an ultimate solution for revealing the complexities of any porous media, but acquiring them is costly and time-consuming. High-quality 2D SEM images, on the other hand, are widely available. Recently, a novel three-step, multiscale, multiresolution reconstruction method is presented that directly uses 2D images in order to develop 3D models. This method, based on a Shannon Entropy and conditional simulation, can be used for most of the available stationary materials and can build various stochastic 3D models just using a few thin sections.[55][56][57]
  • Ion-abrasion SEM (IA-SEM) is a method of nanoscale 3D imaging that uses a focused beam of gallium to repeatedly abrade the specimen surface 20 nanometres at a time. Each exposed surface is then scanned to compile a 3D image.[58][59]

Applications of 3D SEM Edit

One possible application is measuring the roughness of ice crystals. This method can combine variable-pressure environmental SEM and the 3D capabilities of the SEM to measure roughness on individual ice crystal facets, convert it into a computer model and run further statistical analysis on the model.[60] Other measurements include fractal dimension, examining fracture surface of metals, characterization of materials, corrosion measurement, and dimensional measurements at the nano scale (step height, volume, angle, flatness, bearing ratio, coplanarity, etc.).[citation needed]

SEM is also used by art conservationists to discern threats to paintings' surface stability due to aging, such as the formations of complexes of zinc ions with fatty acids.[61] Forensic scientists use SEM to detect art forgeries.

Gallery of SEM images Edit

The following are examples of images taken using an SEM.

See also Edit

References Edit

  1. ^ Stokes, Debbie J. (2008). Principles and Practice of Variable Pressure Environmental Scanning Electron Microscopy (VP-ESEM). Chichester: John Wiley & Sons. ISBN 978-0470758748.
  2. ^ McMullan, D. (2006). "Scanning electron microscopy 1928–1965". Scanning. 17 (3): 175–185. doi:10.1002/sca.4950170309. PMC 2496789.
  3. ^ McMullan, D. (1988). "Von Ardenne and the scanning electron microscope". Proc Roy Microsc Soc. 23: 283–288.
  4. ^ Knoll, Max (1935). "Aufladepotentiel und Sekundäremission elektronenbestrahlter Körper". Zeitschrift für Technische Physik. 16: 467–475.
  5. ^ GB 511204, von Ardenne, Manfred, "Improvements in electron microscopes", published 1939-08-15 
  6. ^ von Ardenne, Manfred (1938). "Das Elektronen-Rastermikroskop. Theoretische Grundlagen". Zeitschrift für Physik (in German). 109 (9–10): 553–572. Bibcode:1938ZPhy..109..553V. doi:10.1007/BF01341584. S2CID 117900835.
  7. ^ von Ardenne, Manfred (1938). "Das Elektronen-Rastermikroskop. Praktische Ausführung". Zeitschrift für Technische Physik (in German). 19: 407–416.
  8. ^ Zworykin VA, Hillier J, Snyder RL (1942) A scanning electron microscope. ASTM Bull 117, 15–23.
  9. ^ McMullan, D. (1953). "An improved scanning electron microscope for opaque specimens". Proceedings of the IEE - Part II: Power Engineering. 100 (75): 245–256. doi:10.1049/pi-2.1953.0095.
  10. ^ Oatley CW, Nixon WC, Pease RFW (1965) Scanning electron microscopy. Adv Electronics Electron Phys 21, 181–247.
  11. ^ Smith, KCA; Oatley, CW (1955). "The scanning electron microscope and its fields of application". British Journal of Applied Physics. 6 (11): 391–399. Bibcode:1955BJAP....6..391S. doi:10.1088/0508-3443/6/11/304.
  12. ^ Wells OC (1957) The construction of a scanning electron microscope and its application to the study of fibres. PhD Dissertation, Cambridge University.
  13. ^ a b Suzuki, E. (2002). "High-resolution scanning electron microscopy of immunogold-labelled cells by the use of thin plasma coating of osmium". Journal of Microscopy. 208 (3): 153–157. doi:10.1046/j.1365-2818.2002.01082.x. PMID 12460446. S2CID 42452027.
  14. ^ a b c Goldstein, G. I.; Newbury, D. E.; Echlin, P.; Joy, D. C.; Fiori, C.; Lifshin, E. (1981). Scanning electron microscopy and x-ray microanalysis. New York: Plenum Press. ISBN 978-0-306-40768-0.
  15. ^ Seligman, Arnold M.; Wasserkrug, Hannah L.; Hanker, Jacob S. (1966). "A new staining method for enhancing contrast of lipid-containing membranes and droplets in osmium tetroxide-fixed tissue with osmiophilic thiocarbohydrazide (TCH)". Journal of Cell Biology. 30 (2): 424–432. doi:10.1083/jcb.30.2.424. PMC 2106998. PMID 4165523.
  16. ^ Malick, Linda E.; Wilson, Richard B.; Stetson, David (1975). "Modified Thiocarbohydrazide Procedure for Scanning Electron Microscopy: Routine use for Normal, Pathological, or Experimental Tissues". Biotechnic & Histochemistry. 50 (4): 265–269. doi:10.3109/10520297509117069. PMID 1103373.
  17. ^ Conrad, Cyler; Jones, Emily Lena; Newsome, Seth D.; Schwartz, Douglas W. (2016). "Bone isotopes, eggshell and turkey husbandry at Arroyo Hondo Pueblo". Journal of Archaeological Science: Reports. 10: 566–574. Bibcode:2016JArSR..10..566C. doi:10.1016/j.jasrep.2016.06.016.
  18. ^ a b c Jeffree, C. E.; Read, N. D. (1991). "Ambient- and Low-temperature scanning electron microscopy". In Hall, J. L.; Hawes, C. R. (eds.). Electron Microscopy of Plant Cells. London: Academic Press. pp. 313–413. ISBN 978-0-12-318880-9.
  19. ^ Karnovsky, M. J. (1965). "A formaldehyde-glutaraldehyde fixative of high osmolality for use in electron microscopy" (PDF). Journal of Cell Biology. 27 (2): 1A–149A. JSTOR 1604673.
  20. ^ Kiernan, J. A. (2000). "Formaldehyde, formalin, paraformaldehyde and glutaraldehyde: What they are and what they do". Microscopy Today. 2000 (1): 8–12. doi:10.1017/S1551929500057060. S2CID 100881495.
  21. ^ Russell, S. D.; Daghlian, C. P. (1985). "Scanning electron microscopic observations on deembedded biological tissue sections: Comparison of different fixatives and embedding materials". Journal of Electron Microscopy Technique. 2 (5): 489–495. doi:10.1002/jemt.1060020511.
  22. ^ Chandler, Douglas E.; Roberson, Robert W. (2009). Bioimaging : current concepts in light and electron microscopy. Sudbury, Mass.: Jones and Bartlett Publishers. ISBN 9780763738747.
  23. ^ Faulkner, Christine; et al. (2008). "Peeking into Pit Fields: A Multiple Twinning Model of Secondary Plasmodesmata Formation in Tobacco". Plant Cell. 20 (6): 1504–18. doi:10.1105/tpc.107.056903. PMC 2483367. PMID 18667640.
  24. ^ Wergin, W. P.; Erbe, E. F. (1994). . Scanning. 16 (Suppl. IV): IV88. Archived from the original on 17 February 2013. Retrieved 15 December 2012.
  25. ^ Barnes, P. R. F.; Mulvaney, R.; Wolff, E. W.; Robinson, K. A. (2002). "A technique for the examination of polar ice using the scanning electron microscope". Journal of Microscopy. 205 (2): 118–124. doi:10.1046/j.0022-2720.2001.00981.x. PMID 11879426. S2CID 35513404.
  26. ^ Hindmarsh, J. P.; Russell, A. B.; Chen, X. D. (2007). "Fundamentals of the spray freezing of foods—microstructure of frozen droplets". Journal of Food Engineering. 78 (1): 136–150. doi:10.1016/j.jfoodeng.2005.09.011.
  27. ^ "Ultra-high Resolution Scanning Electron Microscope SU9000".
  28. ^ Everhart, T. E.; Thornley, R. F. M. (1960). "Wide-band detector for micro-microampere low-energy electron currents" (PDF). Journal of Scientific Instruments. 37 (7): 246–248. Bibcode:1960JScI...37..246E. doi:10.1088/0950-7671/37/7/307.
  29. ^ Hitachi Launches World's Highest Resolution FE-SEM. Nanotech Now. 31 May 2011.
  30. ^ Takaku, Yasuharu; Suzuki, Hiroshi; Ohta, Isao; Tsutsui, Takami; Matsumoto, Haruko; Shimomura, Masatsugu; Hariyama, Takahiko (7 March 2015). "A 'NanoSuit' surface shield successfully protects organisms in high vacuum: observations on living organisms in an FE-SEM". Proceedings of the Royal Society of London B: Biological Sciences. 282 (1802): 20142857. doi:10.1098/rspb.2014.2857. ISSN 0962-8452. PMC 4344158. PMID 25631998.
  31. ^ Danilatos, G. D. (1988). "Foundations of environmental scanning electron microscopy". Advances in Electronics and Electron Physics Volume 71. Advances in Electronics and Electron Physics. Vol. 71. pp. 109–250. doi:10.1016/S0065-2539(08)60902-6. ISBN 9780120146710.
  32. ^ US patent 4823006, Danilatos, Gerasimos D. and Lewis, George C., "Integrated electron optical/differential pumping/imaging signal detection system for an environmental scanning electron microscope", issued 18 April 1989 
  33. ^ Danilatos, G. D. (1990). Theory of the Gaseous Detector Device in the ESEM. Advances in Electronics and Electron Physics. Vol. 78. pp. 1–102. doi:10.1016/S0065-2539(08)60388-1. ISBN 9780120146789.
  34. ^ US patent 4785182, Mancuso, James F.; Maxwell, William B. and Danilatos, Gerasimos D., "Secondary Electron Detector for Use in a Gaseous Atmosphere", issued 15 November 1988 
  35. ^ History of Electron Microscopy 1990s Archived 4 March 2007 at archive.today. sfc.fr
  36. ^ de Jonge, N.; Ross, F.M. (2011). "Electron microscopy of specimens in liquid". Nature Nanotechnology. 6 (8): 695–704. Bibcode:2003NatMa...2..532W. doi:10.1038/nmat944. PMID 12872162. S2CID 21379512.
  37. ^ Klein, Tobias; Buhr, Egbert; Frase, Carl G. (2012). TSEM: A Review of Scanning Electron Microscopy in Transmission Mode and Its Applications. Advances in Imaging and Electron Physics. Vol. 171. pp. 297–356. doi:10.1016/B978-0-12-394297-5.00006-4. ISBN 9780123942975.
  38. ^ "Forensic Applications of the Phenom Desktop Scanning Electron Microscope (SEM)". AZoNano.com. 21 February 2014. Retrieved 11 May 2023.
  39. ^ a b Shrivastava, Priya; Jain, V. K.; Nagpal, Suman (1 June 2021). "Gunshot residue detection technologies—a review". Egyptian Journal of Forensic Sciences. 11 (1): 11. doi:10.1186/s41935-021-00223-9. ISSN 2090-5939.
  40. ^ Burgess, Jeremy (1987). Under the Microscope: A Hidden World Revealed. CUP Archive. p. 11. ISBN 978-0521399401.
  41. ^ "Showing your true colors, 3D and color in electron microscopy in Lab News magazine".
  42. ^ a b c d e Mignot, Christophe (2018). "Color (and 3D) for Scanning Electron Microscopy". Microscopy Today. 26 (3): 12–17. doi:10.1017/S1551929518000482.
  43. ^ "Introduction to Electron Microscopy" (PDF). FEI Company. p. 15. Retrieved 12 December 2012.
  44. ^ "Next Monday, Digital Surf to Launch Revolutionary SEM Image Colorization". AZO Materials. 22 January 2016. Retrieved 23 January 2016.
  45. ^ Antonovsky, A. (1984). "The application of colour to SEM imaging for increased definition". Micron and Microscopica Acta. 15 (2): 77–84. doi:10.1016/0739-6260(84)90005-4.
  46. ^ Danilatos, G.D. (1986). "Colour micrographs for backscattered electron signals in the SEM". Scanning. 9 (3): 8–18. doi:10.1111/j.1365-2818.1986.tb04287.x. S2CID 96315383.
  47. ^ Danilatos, G.D. (1986). "Environmental scanning electron microscopy in colour". Journal of Microscopy. 142: 317–325. doi:10.1002/sca.4950080104.
  48. ^ Bertazzo, S.; Gentleman, E.; Cloyd, K. L.; Chester, A. H.; Yacoub, M. H.; Stevens, M. M. (2013). "Nano-analytical electron microscopy reveals fundamental insights into human cardiovascular tissue calcification". Nature Materials. 12 (6): 576–583. Bibcode:2013NatMa..12..576B. doi:10.1038/nmat3627. hdl:10044/1/21901. PMC 5833942. PMID 23603848.
  49. ^ Bertazzo, Sergio; Maidment, Susannah C. R.; Kallepitis, Charalambos; Fearn, Sarah; Stevens, Molly M.; Xie, Hai-nan (9 June 2015). "Fibres and cellular structures preserved in 75-million–year-old dinosaur specimens". Nature Communications. 6: 7352. Bibcode:2015NatCo...6.7352B. doi:10.1038/ncomms8352. PMC 4468865. PMID 26056764.
  50. ^ Stereo SEM reconstruction using MountainsMap SEM version 7.4 on i7 2600 CPU at 3.4 GHz
  51. ^ Butterfield, Nicholas; Rowe, Penny M.; Stewart, Emily; Roesel, David; Neshyba, Steven (16 March 2017). "Quantitative three-dimensional ice roughness from scanning electron microscopy". Journal of Geophysical Research: Atmospheres. 122 (5): 3023–3025. Bibcode:2017JGRD..122.3023B. doi:10.1002/2016JD026094.
  52. ^ Butterfield, Nicholas; Rowe, Penny M.; Stewart, Emily; Roesel, David; Neshyba, Steven (16 March 2017). "Quantitative three-dimensional ice roughness from scanning electron microscopy". Journal of Geophysical Research: Atmospheres. 122 (5): 3025–3041. Bibcode:2017JGRD..122.3023B. doi:10.1002/2016JD026094.
  53. ^ Baghaei Rad, Leili (2007). Computational Scanning Electron Microscopy. International Conference on Frontiers of Characterization and Metrology. Vol. 931. p. 512. Bibcode:2007AIPC..931..512R. doi:10.1063/1.2799427.
  54. ^ Baghaei Rad, Leili; Downes, Ian; Ye, Jun; Adler, David; Pease, R. Fabian W. (2007). "Economic approximate models for backscattered electrons". Journal of Vacuum Science and Technology. 25 (6): 2425. Bibcode:2007JVSTB..25.2425B. doi:10.1116/1.2794068.
  55. ^ Tahmasebi, Pejman; Javadpour, Farzam; Sahimi, Muhammad (2015). "Multiscale and multiresolution modeling of shales and their flow and morphological properties". Scientific Reports. 5: 16373. Bibcode:2015NatSR...516373T. doi:10.1038/srep16373. PMC 4642334. PMID 26560178.
  56. ^ Tahmasebi, Pejman; Javadpour, Farzam; Sahimi, Muhammad (2015). "Three-Dimensional Stochastic Characterization of Shale SEM Images". Transport in Porous Media. 110 (3): 521–531. doi:10.1007/s11242-015-0570-1. S2CID 20274015.
  57. ^ Tahmasebi, Pejman; Sahimi, Muhammad (2012). "Reconstruction of three-dimensional porous media using a single thin section". Physical Review E. 85 (6): 066709. Bibcode:2012PhRvE..85f6709T. doi:10.1103/PhysRevE.85.066709. PMID 23005245. S2CID 24307267.
  58. ^ Murphy, GE; Lowekamp, BC; Zerfas, PM (August 2010). "Ion-abrasion scanning electron microscopy reveals distorted liver mitochondrial morphology in murine methylmalonic acidemia". Journal of Structural Biology. 171 (2): 125–32. doi:10.1016/j.jsb.2010.04.005. PMC 2885563. PMID 20399866.
  59. ^ "Multimedia Gallery - 3-D Imaging of Mammalian Cells With Ion-Abrasion SEM | NSF - National Science Foundation". www.nsf.gov.
  60. ^ Butterfield, Nicholas; Rowe, Penny M.; Stewart, Emily; Roesel, David; Neshyba, Steven (16 March 2017). "Quantitative three-dimensional ice roughness from scanning electron microscopy". Journal of Geophysical Research: Atmospheres. 122 (5): 3023–3041. Bibcode:2017JGRD..122.3023B. doi:10.1002/2016JD026094.
  61. ^ Hermans, Joen; Osmond, Gillian; Loon, Annelies van; Iedema, Piet; Chapman, Robyn; Drennan, John; Jack, Kevin; Rasch, Ronald; Morgan, Garry; Zhang, Zhi; Monteiro, Michael (June 2018). "Electron Microscopy Imaging of Zinc Soaps Nucleation in Oil Paint". Microscopy and Microanalysis. 24 (3): 318–322. Bibcode:2018MiMic..24..318H. doi:10.1017/S1431927618000387. ISSN 1431-9276. PMID 29860951. S2CID 44166918.

External links Edit

General
  • HowStuffWorks – How Scanning Electron Microscopes Work
  • Learn to use an SEM – An online learning environment for people wanting to use an SEM. Provided by Microscopy Australia
  • Virtual SEM – sparkler – an interactive simulation of a scanning electron microscope (SEM)
  • Multichannel color SEM imaging – and with BSE
  • DDC-SEM image examples
  • Video on the scanning electron microscope, Karlsruhe University of Applied Sciences
  • Animations and explanations on various types of microscopes including electron microscopes (Université Paris Sud)
History
  • Environmental Scanning Electron Microscope (ESEM) history
Images
  • Rippel Electron Microscope Facility 19 March 2007 at the Wayback Machine Many dozens of (mostly biological) SEM images from Dartmouth College.
  • Lanthanoid staining SEM images from Research Institute of Eye Diseases, Moscow.

scanning, electron, microscope, this, article, needs, additional, citations, verification, please, help, improve, this, article, adding, citations, reliable, sources, unsourced, material, challenged, removed, find, sources, news, newspapers, books, scholar, js. This article needs additional citations for verification Please help improve this article by adding citations to reliable sources Unsourced material may be challenged and removed Find sources Scanning electron microscope news newspapers books scholar JSTOR July 2023 Learn how and when to remove this template message Not to be confused with Scanning tunneling microscope A scanning electron microscope SEM is a type of electron microscope that produces images of a sample by scanning the surface with a focused beam of electrons The electrons interact with atoms in the sample producing various signals that contain information about the surface topography and composition of the sample The electron beam is scanned in a raster scan pattern and the position of the beam is combined with the intensity of the detected signal to produce an image In the most common SEM mode secondary electrons emitted by atoms excited by the electron beam are detected using a secondary electron detector Everhart Thornley detector The number of secondary electrons that can be detected and thus the signal intensity depends among other things on specimen topography Some SEMs can achieve resolutions better than 1 nanometer Image of pollen grains taken on a SEM shows the characteristic depth of field of SEM micrographsM von Ardenne s first SEM source source source source source source source Operating principle of a Scanning Electron Microscope SEM SEM with opened sample chamberAnalog type SEMSpecimens are observed in high vacuum in a conventional SEM or in low vacuum or wet conditions in a variable pressure or environmental SEM and at a wide range of cryogenic or elevated temperatures with specialized instruments 1 Contents 1 History 2 Principles and capacities 3 Sample preparation 3 1 Biological samples 3 2 Materials 4 Scanning process and image formation 4 1 Magnification 5 Detection of secondary electrons 6 Detection of backscattered electrons 7 Beam injection analysis of semiconductors 8 Cathodoluminescence 9 X ray microanalysis 10 Resolution of the SEM 11 Environmental SEM 12 Transmission SEM 13 SEM in Forensic Science 14 Color in SEM 14 1 False color using a single detector 14 2 SEM image coloring 14 3 Color built using multiple electron detectors 14 4 Analytical signals based on generated photons 15 3D in SEM 15 1 3D SEM reconstruction from a stereo pair 15 2 Photometric 3D SEM reconstruction from a four quadrant detector by shape from shading 15 3 Photometric 3D rendering from a single SEM image 15 4 Other types of 3D SEM reconstruction 15 5 Applications of 3D SEM 16 Gallery of SEM images 17 See also 18 References 19 External linksHistory EditAn account of the early history of scanning electron microscopy has been presented by McMullan 2 3 Although Max Knoll produced a photo with a 50 mm object field width showing channeling contrast by the use of an electron beam scanner 4 it was Manfred von Ardenne who in 1937 invented 5 a microscope with high resolution by scanning a very small raster with a demagnified and finely focused electron beam Ardenne applied scanning of the electron beam in an attempt to surpass the resolution of the transmission electron microscope TEM as well as to mitigate substantial problems with chromatic aberration inherent to real imaging in the TEM He further discussed the various detection modes possibilities and theory of SEM 6 together with the construction of the first high resolution SEM 7 Further work was reported by Zworykin s group 8 followed by the Cambridge groups in the 1950s and early 1960s 9 10 11 12 headed by Charles Oatley all of which finally led to the marketing of the first commercial instrument by Cambridge Scientific Instrument Company as the Stereoscan in 1965 which was delivered to DuPont Principles and capacities Edit nbsp Schottky emitter electron source nbsp Electron matter interaction volume and types of signal generatedThe signals used by a SEM to produce an image result from interactions of the electron beam with atoms at various depths within the sample Various types of signals are produced including secondary electrons SE reflected or back scattered electrons BSE characteristic X rays and light cathodoluminescence CL absorbed current specimen current and transmitted electrons Secondary electron detectors are standard equipment in all SEMs but it is rare for a single machine to have detectors for all other possible signals citation needed Secondary electrons have very low energies on the order of 50 eV which limits their mean free path in solid matter Consequently SEs can only escape from the top few nanometers of the surface of a sample The signal from secondary electrons tends to be highly localized at the point of impact of the primary electron beam making it possible to collect images of the sample surface with a resolution of below 1 nm Back scattered electrons BSE are beam electrons that are reflected from the sample by elastic scattering Since they have much higher energy than SEs they emerge from deeper locations within the specimen and consequently the resolution of BSE images is less than SE images However BSE are often used in analytical SEM along with the spectra made from the characteristic X rays because the intensity of the BSE signal is strongly related to the atomic number Z of the specimen BSE images can provide information about the distribution but not the identity of different elements in the sample In samples predominantly composed of light elements such as biological specimens BSE imaging can image colloidal gold immuno labels of 5 or 10 nm diameter which would otherwise be difficult or impossible to detect in secondary electron images 13 Characteristic X rays are emitted when the electron beam removes an inner shell electron from the sample causing a higher energy electron to fill the shell and release energy The energy or wavelength of these characteristic X rays can be measured by Energy dispersive X ray spectroscopy or Wavelength dispersive X ray spectroscopy and used to identify and measure the abundance of elements in the sample and map their distribution Due to the very narrow electron beam SEM micrographs have a large depth of field yielding a characteristic three dimensional appearance useful for understanding the surface structure of a sample 14 This is exemplified by the micrograph of pollen shown above A wide range of magnifications is possible from about 10 times about equivalent to that of a powerful hand lens to more than 500 000 times about 250 times the magnification limit of the best light microscopes Sample preparation Edit nbsp A spider sputter coated in gold having been prepared for viewing with an SEM nbsp Low voltage micrograph 300 V of distribution of adhesive droplets on a Post it note No conductive coating was applied such a coating would alter this fragile specimen SEM samples have to be small enough to fit on the specimen stage and may need special preparation to increase their electrical conductivity and to stabilize them so that they can withstand the high vacuum conditions and the high energy beam of electrons Samples are generally mounted rigidly on a specimen holder or stub using a conductive adhesive SEM is used extensively for defect analysis of semiconductor wafers and manufacturers make instruments that can examine any part of a 300 mm semiconductor wafer Many instruments have chambers that can tilt an object of that size to 45 and provide continuous 360 rotation citation needed Nonconductive specimens collect charge when scanned by the electron beam and especially in secondary electron imaging mode this causes scanning faults and other image artifacts For conventional imaging in the SEM specimens must be electrically conductive at least at the surface and electrically grounded to prevent the accumulation of electrostatic charge Metal objects require little special preparation for SEM except for cleaning and conductively mounting to a specimen stub Non conducting materials are usually coated with an ultrathin coating of electrically conducting material deposited on the sample either by low vacuum sputter coating electroless deposition citation needed or by high vacuum evaporation Conductive materials in current use for specimen coating include gold gold palladium alloy platinum iridium tungsten chromium osmium 13 and graphite Coating with heavy metals may increase signal noise ratio for samples of low atomic number Z The improvement arises because secondary electron emission for high Z materials is enhanced citation needed An alternative to coating for some biological samples is to increase the bulk conductivity of the material by impregnation with osmium using variants of the OTO staining method O osmium tetroxide T thiocarbohydrazide O osmium 15 16 Nonconducting specimens may be imaged without coating using an environmental SEM ESEM or low voltage mode of SEM operation In ESEM instruments the specimen is placed in a relatively high pressure chamber and the electron optical column is differentially pumped to keep vacuum adequately clarification needed low at the electron gun The high pressure region around the sample in the ESEM neutralizes charge and provides an amplification of the secondary electron signal citation needed Low voltage SEM is typically conducted in an instrument with a field emission guns FEG which is capable of producing high primary electron brightness and small spot size even at low accelerating potentials To prevent charging of non conductive specimens operating conditions must be adjusted such that the incoming beam current is equal to sum of outgoing secondary and backscattered electron currents a condition that is most often met at accelerating voltages of 0 3 4 kV citation needed Embedding in a resin with further polishing to a mirror like finish can be used for both biological and materials specimens when imaging in backscattered electrons or when doing quantitative X ray microanalysis The main preparation techniques are not required in the environmental SEM outlined below but some biological specimens can benefit from fixation Biological samples Edit Conventionally a SEM specimen is required to be completely dry since the specimen chamber is under high vacuum Hard dry materials such as wood bone feathers dried insects or shells including egg shells 17 can be examined with little further treatment but living cells and tissues and whole soft bodied organisms require chemical fixation to preserve and stabilize their structure Fixation is usually performed by incubation in a solution of a buffered chemical fixative such as glutaraldehyde sometimes in combination with formaldehyde 18 19 20 and other fixatives 21 and optionally followed by postfixation with osmium tetroxide 18 The fixed tissue is then dehydrated Because air drying causes collapse and shrinkage this is commonly achieved by replacement of water in the cells with organic solvents such as ethanol or acetone and replacement of these solvents in turn with a transitional fluid such as liquid carbon dioxide by critical point drying 22 The carbon dioxide is finally removed while in a supercritical state so that no gas liquid interface is present within the sample during drying The dry specimen is usually mounted on a specimen stub using an adhesive such as epoxy resin or electrically conductive double sided adhesive tape and sputter coated with gold or gold palladium alloy before examination in the microscope Samples may be sectioned with a microtome if information about the organism s internal ultrastructure is to be exposed for imaging If the SEM is equipped with a cold stage for cryo microscopy cryofixation may be used and low temperature scanning electron microscopy performed on the cryogenically fixed specimens 18 Cryo fixed specimens may be cryo fractured under vacuum in a special apparatus to reveal internal structure sputter coated and transferred onto the SEM cryo stage while still frozen 23 Low temperature scanning electron microscopy LT SEM is also applicable to the imaging of temperature sensitive materials such as ice 24 25 and fats 26 Freeze fracturing freeze etch or freeze and break is a preparation method particularly useful for examining lipid membranes and their incorporated proteins in face on view The preparation method reveals the proteins embedded in the lipid bilayer Materials Edit This section does not cite any sources Please help improve this section by adding citations to reliable sources Unsourced material may be challenged and removed February 2016 Learn how and when to remove this template message Back scattered electron imaging quantitative X ray analysis and X ray mapping of specimens often requires grinding and polishing the surfaces to an ultra smooth surface Specimens that undergo WDS or EDS analysis are often carbon coated In general metals are not coated prior to imaging in the SEM because they are conductive and provide their own pathway to ground Fractography is the study of fractured surfaces that can be done on a light microscope or commonly on an SEM The fractured surface is cut to a suitable size cleaned of any organic residues and mounted on a specimen holder for viewing in the SEM Integrated circuits may be cut with a focused ion beam FIB or other ion beam milling instrument for viewing in the SEM The SEM in the first case may be incorporated into the FIB enabling high resolution imaging of the result of the process Metals geological specimens and integrated circuits all may also be chemically polished for viewing in the SEM Special high resolution coating techniques are required for high magnification imaging of inorganic thin films Scanning process and image formation EditThis section needs additional citations for verification Please help improve this article by adding citations to reliable sources in this section Unsourced material may be challenged and removed July 2023 Learn how and when to remove this template message nbsp Schematic of an SEMIn a typical SEM an electron beam is thermionically emitted from an electron gun fitted with a tungsten filament cathode Tungsten is normally used in thermionic electron guns because it has the highest melting point and lowest vapor pressure of all metals thereby allowing it to be electrically heated for electron emission and because of its low cost Other types of electron emitters include lanthanum hexaboride LaB6 cathodes which can be used in a standard tungsten filament SEM if the vacuum system is upgraded or field emission guns FEG which may be of the cold cathode type using tungsten single crystal emitters or the thermally assisted Schottky type that use emitters of tungsten single crystals coated in zirconium oxide The electron beam which typically has an energy ranging from 0 2 keV to 40 keV is focused by one or two condenser lenses to a spot about 0 4 nm to 5 nm in diameter The beam passes through pairs of scanning coils or pairs of deflector plates in the electron column typically in the final lens which deflect the beam in the x and y axes so that it scans in a raster fashion over a rectangular area of the sample surface nbsp Mechanisms of emission of secondary electrons backscattered electrons and characteristic X rays from atoms of the sampleWhen the primary electron beam interacts with the sample the electrons lose energy by repeated random scattering and absorption within a teardrop shaped volume of the specimen known as the interaction volume which extends from less than 100 nm to approximately 5 µm into the surface The size of the interaction volume depends on the electron s landing energy the atomic number of the specimen and the specimen s density The energy exchange between the electron beam and the sample results in the reflection of high energy electrons by elastic scattering the emission of secondary electrons by inelastic scattering and the emission of electromagnetic radiation each of which can be detected by specialized detectors The beam current absorbed by the specimen can also be detected and used to create images of the distribution of specimen current Electronic amplifiers of various types are used to amplify the signals which are displayed as variations in brightness on a computer monitor or for vintage models on a cathode ray tube Each pixel of computer video memory is synchronized with the position of the beam on the specimen in the microscope and the resulting image is therefore a distribution map of the intensity of the signal being emitted from the scanned area of the specimen Older microscopes captured images on film but most modern instruments collect digital images nbsp Low temperature SEM magnification series for a snow crystal The crystals are captured stored and sputter coated with platinum at cryogenic temperatures for imaging Magnification Edit Magnification in an SEM can be controlled over a range of about 6 orders of magnitude from about 10 to 3 000 000 times 27 Unlike optical and transmission electron microscopes image magnification in an SEM is not a function of the power of the objective lens SEMs may have condenser and objective lenses but their function is to focus the beam to a spot and not to image the specimen Provided the electron gun can generate a beam with a sufficiently small diameter an SEM could in principle work entirely without condenser or objective lenses However it might not be very versatile or achieve very high resolution In an SEM as in scanning probe microscopy magnification results from the ratio of the raster on the display device and dimensions of the raster on the specimen Assuming that the display screen has a fixed size higher magnification results from reducing the size of the raster on the specimen and vice versa Magnification is therefore controlled by the current supplied to the x y scanning coils or the voltage supplied to the x y deflector plates and not by objective lens power Detection of secondary electrons EditThe most common imaging mode collects low energy lt 50 eV secondary electrons that are ejected from conduction or valence bands of the specimen atoms by inelastic scattering interactions with beam electrons Due to their low energy these electrons originate from within a few nanometers below the sample surface 14 The electrons are detected by an Everhart Thornley detector 28 which is a type of collector scintillator photomultiplier system The secondary electrons are first collected by attracting them towards an electrically biased grid at about 400 V and then further accelerated towards a phosphor or scintillator positively biased to about 2 000 V The accelerated secondary electrons are now sufficiently energetic to cause the scintillator to emit flashes of light cathodoluminescence which are conducted to a photomultiplier outside the SEM column via a light pipe and a window in the wall of the specimen chamber The amplified electrical signal output by the photomultiplier is displayed as a two dimensional intensity distribution that can be viewed and photographed on an analogue video display or subjected to analog to digital conversion and displayed and saved as a digital image This process relies on a raster scanned primary beam The brightness of the signal depends on the number of secondary electrons reaching the detector If the beam enters the sample perpendicular to the surface then the activated region is uniform about the axis of the beam and a certain number of electrons escape from within the sample As the angle of incidence increases the interaction volume increases and the escape distance of one side of the beam decreases resulting in more secondary electrons being emitted from the sample Thus steep surfaces and edges tend to be brighter than flat surfaces which results in images with a well defined three dimensional appearance Using the signal of secondary electrons image resolution less than 0 5 nm is possible Detection of backscattered electrons Edit nbsp Comparison of SEM techniques Top backscattered electron analysis composition Bottom secondary electron analysis topographyBackscattered electrons BSE consist of high energy electrons originating in the electron beam that are reflected or back scattered out of the specimen interaction volume by elastic scattering interactions with specimen atoms Since heavy elements high atomic number backscatter electrons more strongly than light elements low atomic number and thus appear brighter in the image BSEs are used to detect contrast between areas with different chemical compositions 14 The Everhart Thornley detector which is normally positioned to one side of the specimen is inefficient for the detection of backscattered electrons because few such electrons are emitted in the solid angle subtended by the detector and because the positively biased detection grid has little ability to attract the higher energy BSE Dedicated backscattered electron detectors are positioned above the sample in a doughnut type arrangement concentric with the electron beam maximizing the solid angle of collection BSE detectors are usually either of scintillator or of semiconductor types When all parts of the detector are used to collect electrons symmetrically about the beam atomic number contrast is produced However strong topographic contrast is produced by collecting back scattered electrons from one side above the specimen using an asymmetrical directional BSE detector the resulting contrast appears as illumination of the topography from that side Semiconductor detectors can be made in radial segments that can be switched in or out to control the type of contrast produced and its directionality Backscattered electrons can also be used to form an electron backscatter diffraction EBSD image that can be used to determine the crystallographic structure of the specimen Beam injection analysis of semiconductors EditThe nature of the SEM s probe energetic electrons makes it uniquely suited to examining the optical and electronic properties of semiconductor materials The high energy electrons from the SEM beam will inject charge carriers into the semiconductor Thus beam electrons lose energy by promoting electrons from the valence band into the conduction band leaving behind holes In a direct bandgap material recombination of these electron hole pairs will result in cathodoluminescence if the sample contains an internal electric field such as is present at a p n junction the SEM beam injection of carriers will cause electron beam induced current EBIC to flow Cathodoluminescence and EBIC are referred to as beam injection techniques and are very powerful probes of the optoelectronic behavior of semiconductors in particular for studying nanoscale features and defects Cathodoluminescence Edit nbsp Color cathodoluminescence overlay on SEM image of an InGaN polycrystal The blue and green channels represent real colors the red channel corresponds to UV emission Cathodoluminescence the emission of light when atoms excited by high energy electrons return to their ground state is analogous to UV induced fluorescence and some materials such as zinc sulfide and some fluorescent dyes exhibit both phenomena Over the last decades cathodoluminescence was most commonly experienced as the light emission from the inner surface of the cathode ray tube in television sets and computer CRT monitors In the SEM CL detectors either collect all light emitted by the specimen or can analyse the wavelengths emitted by the specimen and display an emission spectrum or an image of the distribution of cathodoluminescence emitted by the specimen in real color X ray microanalysis EditCharacteristic X rays that are produced by the interaction of electrons with the sample may also be detected in an SEM equipped for energy dispersive X ray spectroscopy or wavelength dispersive X ray spectroscopy Analysis of the x ray signals may be used to map the distribution and estimate the abundance of elements in the sample Resolution of the SEM Edit source source source source source source A video illustrating a typical practical magnification range of a scanning electron microscope designed for biological specimens The video starts at 25 about 6 mm across the whole field of view and zooms in to 12000 about 12 mm across the whole field of view The spherical objects are glass beads with a diameter of 10 mm similar in diameter to a red blood cell SEM is not a camera and the detector is not continuously image forming like a CCD array or film Unlike in an optical system the resolution is not limited by the diffraction limit fineness of lenses or mirrors or detector array resolution The focusing optics can be large and coarse and the SE detector is fist sized and simply detects current Instead the spatial resolution of the SEM depends on the size of the electron spot which in turn depends on both the wavelength of the electrons and the electron optical system that produces the scanning beam The resolution is also limited by the size of the interaction volume the volume of specimen material that interacts with the electron beam The spot size and the interaction volume are both large compared to the distances between atoms so the resolution of the SEM is not high enough to image individual atoms as is possible with a transmission electron microscope TEM The SEM has compensating advantages though including the ability to image a comparatively large area of the specimen the ability to image bulk materials not just thin films or foils and the variety of analytical modes available for measuring the composition and properties of the specimen Depending on the instrument the resolution can fall somewhere between less than 1 nm and 20 nm As of 2009 The world s highest resolution conventional 30 kV SEM can reach a point resolution of 0 4 nm using a secondary electron detector 29 Environmental SEM EditMain article Environmental scanning electron microscope Conventional SEM requires samples to be imaged under vacuum because a gas atmosphere rapidly spreads and attenuates electron beams As a consequence samples that produce a significant amount of vapour e g wet biological samples or oil bearing rock must be either dried or cryogenically frozen Processes involving phase transitions such as the drying of adhesives or melting of alloys liquid transport chemical reactions and solid air gas systems in general cannot be observed with conventional high vacuum SEM In environmental SEM ESEM the chamber is evacuated of air but water vapor is retained near its saturation pressure and the residual pressure remains relatively high This allows the analysis of samples containing water or other volatile substances With ESEM observations of living insects have been possible 30 The first commercial development of the ESEM in the late 1980s 31 32 allowed samples to be observed in low pressure gaseous environments e g 1 50 Torr or 0 1 6 7 kPa and high relative humidity up to 100 This was made possible by the development of a secondary electron detector 33 34 capable of operating in the presence of water vapour and by the use of pressure limiting apertures with differential pumping in the path of the electron beam to separate the vacuum region around the gun and lenses from the sample chamber The first commercial ESEMs were produced by the ElectroScan Corporation in USA in 1988 ElectroScan was taken over by Philips who later sold their electron optics division to FEI Company in 1996 35 ESEM is especially useful for non metallic and biological materials because coating with carbon or gold is unnecessary Uncoated plastics and elastomers can be routinely examined as can uncoated biological samples This is useful because coating can be difficult to reverse may conceal small features on the surface of the sample and may reduce the value of the results obtained X ray analysis is difficult with a coating of a heavy metal so carbon coatings are routinely used in conventional SEMs but ESEM makes it possible to perform X ray microanalysis on uncoated non conductive specimens however some specific for ESEM artifacts are introduced in X ray analysis ESEM may be the preferred for electron microscopy of unique samples from criminal or civil actions where forensic analysis may need to be repeated by several different experts It is possible to study specimens in liquid with ESEM or with other liquid phase electron microscopy methods 36 Transmission SEM EditThe SEM can also be used in transmission mode by simply incorporating an appropriate detector below a thin specimen section 37 Detectors are available for bright field dark field as well as segmented detectors for mid field to high angle annular dark field Despite the difference in instrumentation this technique is still commonly referred to as scanning transmission electron microscopy STEM SEM in Forensic Science EditThe SEM is used often in Forensic Science for magnified analysis of microscopic things such as diatoms and gunshot residue Because SEM is a nondestructive force on the sample it can be used to analyze evidence without damaging it The SEM shoots a beam of high energy electrons to the sample which bounce off of the sample without changing or destroying it This is great when it comes to analyzing diatoms When a person dies by drowning they inhale the water which causes what is in the water diatoms to get in the blood stream brain kidneys and more These diatoms in the body can be magnified with the SEM to determine the type of diatoms which aid in understanding how and where the person died By using the images produced by the SEM forensic scientists can compare diatoms types to confirm the body of water a person died in 38 Gunshot residue GSR analysis can be done with many different analytical instruments 39 but SEM is a common way to analyze inorganic compounds because of the way it can closely analyze the types of elements mostly metals through its three detectors backscatter electron detector secondary electron detector and X ray detector GSR can be collected from the crime scene victim or shooter and analyzed with the SEM This can help scientists determine proximity and or contact with the discharged firearm 39 Color in SEM EditElectron microscopes do not naturally produce color images as an SEM produces a single value per pixel this value corresponds to the number of electrons received by the detector during a small period of time of the scanning when the beam is targeted to the x y pixel position This single number is usually represented for each pixel by a grey level forming a monochrome image 40 However several ways have been used to get color electron microscopy images 41 False color using a single detector Edit On compositional images of flat surfaces typically BSE The easiest way to get color is to associate to this single number an arbitrary color using a color look up table i e each grey level is replaced by a chosen color This method is known as false color On a BSE image false color may be performed to better distinguish the various phases of the sample 42 On textured surface images As an alternative to simply replacing each grey level by a color a sample observed by an oblique beam allows researchers to create an approximative topography image see further section Photometric 3D rendering from a single SEM image Such topography can then be processed by 3D rendering algorithms for a more natural rendering of the surface texture nbsp Surface of a kidney stone nbsp The same after re processing of the color from the estimated topography nbsp SEM image of a diagenetically altered discoaster nbsp The same image after similar colorizationSEM image coloring Edit Very often published SEM images are artificially colored 42 This may be done for aesthetic effect to clarify structure or to add a realistic appearance to the sample and generally does not add information about the specimen 43 Coloring may be performed manually with photo editing software or semi automatically with dedicated software using feature detection or object oriented segmentation 44 nbsp SEM image of Cobaea scandens pollen nbsp The same after semi automatic coloring Arbitrary colors help identifying the various elements of the structure nbsp Colored SEM image of Tradescantia pollen and stamens nbsp Colored SEM image of native gold and arsenopyrite crystal intergrowthColor built using multiple electron detectors Edit In some configurations more information is gathered per pixel often by the use of multiple detectors 45 As a common example secondary electron and backscattered electron detectors are superimposed and a color is assigned to each of the images captured by each detector 46 47 with a result of a combined color image where colors are related to the density of the components This method is known as density dependent color SEM DDC SEM Micrographs produced by DDC SEM retain topographical information which is better captured by the secondary electrons detector and combine it to the information about density obtained by the backscattered electron detector 48 49 nbsp DDC SEM of calcified particle in cardiac tissue Signal 1 SE nbsp Signal 2 BSE nbsp Colorized image obtained from the two previous Density dependent color scanning electron micrograph SEM DDC SEM of cardiovascular calcification showing in orange a calcium phosphate spherical particle denser material and in green the extracellular matrix less dense material nbsp Same work with a larger view part of a study on human cardiovascular tissue calcificationAnalytical signals based on generated photons Edit Measurement of the energy of photons emitted from the specimen is a common method to get analytical capabilities Examples are the energy dispersive X ray spectroscopy EDS detectors used in elemental analysis and cathodoluminescence microscope CL systems that analyse the intensity and spectrum of electron induced luminescence in for example geological specimens In SEM systems using these detectors it is common to color code these extra signals and superimpose them in a single color image so that differences in the distribution of the various components of the specimen can be seen clearly and compared Optionally the standard secondary electron image can be merged with the one or more compositional channels so that the specimen s structure and composition can be compared Such images can be made while maintaining the full integrity of the original signal data which is not modified in any way 3D in SEM EditSEMs do not naturally provide 3D images contrary to SPMs However 3D data can be obtained using an SEM with different methods as follows 3D SEM reconstruction from a stereo pair Edit photogrammetry is the most metrologically accurate method to bring the third dimension to SEM images 42 Contrary to photometric methods next paragraph photogrammetry calculates absolute heights using triangulation methods The drawbacks are that it works only if there is a minimum texture and it requires two images to be acquired from two different angles which implies the use of a tilt stage Photogrammetry is a software operation that calculates the shift or disparity for each pixel between the left image and the right image of the same pair Such disparity reflects the local height nbsp An SEM stereo pair of microfossils of less than 1 mm in size Ostracoda produced by tilting along the longitudinal axis nbsp From this pair of SEM images the third dimension has been reconstructed by photogrammetry using MountainsSEM software see next image then a series of 3D representations with different angles have been made and assembled into a GIF file to produce this animation nbsp 3D surface reconstruction of a Ra 3 µm roughness calibration sample as used to calibrate profilometers from 2 scanning electron microscope images tilted by 15 top left The calculation of the 3D model bottom right takes about 1 5 second 50 and the error on the Ra roughness value calculated is less than 0 5 Photometric 3D SEM reconstruction from a four quadrant detector by shape from shading Edit This method typically uses a four quadrant BSE detector alternatively for one manufacturer a 3 segment detector The microscope produces four images of the same specimen at the same time so no tilt of the sample is required The method gives metrological 3D dimensions as far as the slope of the specimen remains reasonable 42 Most SEM manufacturers now 2018 offer such a built in or optional four quadrant BSE detector together with proprietary software to calculate a 3D image in real time 51 Other approaches use more sophisticated and sometimes GPU intensive methods like the optimal estimation algorithm and offer much better results 52 at the cost of high demands on computing power In all instances this approach works by integration of the slope so vertical slopes and overhangs are ignored for instance if an entire sphere lies on a flat little more than the upper hemisphere is seen emerging above the flat resulting in wrong altitude of the sphere apex The prominence of this effect depends on the angle of the BSE detectors with respect to the sample but these detectors are usually situated around and close to the electron beam so this effect is very common Photometric 3D rendering from a single SEM image Edit This method requires an SEM image obtained in oblique low angle lighting The grey level is then interpreted as the slope and the slope integrated to restore the specimen topography This method is interesting for visual enhancement and the detection of the shape and position of objects however the vertical heights cannot usually be calibrated contrary to other methods such as photogrammetry 42 nbsp SEM image of a house fly compound eye surface at 450 magnification nbsp Detail of the previous image nbsp SEM 3D reconstruction from the previous using shape from shading algorithms nbsp Same as the previous but with lighting homogenized before applying the shape from shading algorithmsOther types of 3D SEM reconstruction Edit Inverse reconstruction using electron material interactive models 53 54 Multi Resolution reconstruction using single 2D File High quality 3D imaging may be an ultimate solution for revealing the complexities of any porous media but acquiring them is costly and time consuming High quality 2D SEM images on the other hand are widely available Recently a novel three step multiscale multiresolution reconstruction method is presented that directly uses 2D images in order to develop 3D models This method based on a Shannon Entropy and conditional simulation can be used for most of the available stationary materials and can build various stochastic 3D models just using a few thin sections 55 56 57 Ion abrasion SEM IA SEM is a method of nanoscale 3D imaging that uses a focused beam of gallium to repeatedly abrade the specimen surface 20 nanometres at a time Each exposed surface is then scanned to compile a 3D image 58 59 Applications of 3D SEM Edit One possible application is measuring the roughness of ice crystals This method can combine variable pressure environmental SEM and the 3D capabilities of the SEM to measure roughness on individual ice crystal facets convert it into a computer model and run further statistical analysis on the model 60 Other measurements include fractal dimension examining fracture surface of metals characterization of materials corrosion measurement and dimensional measurements at the nano scale step height volume angle flatness bearing ratio coplanarity etc citation needed SEM is also used by art conservationists to discern threats to paintings surface stability due to aging such as the formations of complexes of zinc ions with fatty acids 61 Forensic scientists use SEM to detect art forgeries Gallery of SEM images EditThe following are examples of images taken using an SEM nbsp Colored SEM image of soybean cyst nematode and egg The artificial coloring makes the image easier for non specialists to view and understand the structures and surfaces revealed in micrographs nbsp Compound eye of Antarctic krill Euphausia superba Arthropod eyes are a common subject in SEM micrographs due to the depth of focus that an SEM image can capture Colored picture nbsp Ommatidia of Antarctic krill eye a higher magnification of the krill s eye SEMs cover a range from light microscopy up to the magnifications available with a TEM Colored picture nbsp SEM image of normal circulating human blood This is an older and noisy micrograph of a common subject for SEM micrographs red blood cells nbsp SEM image of a hederelloid from the Devonian of Michigan largest tube diameter is 0 75 mm The SEM is used extensively for capturing detailed images of micro and macro fossils nbsp Backscattered electron BSE image of an antimony rich region in a fragment of ancient glass Museums use SEMs for studying valuable artifacts in a nondestructive manner nbsp SEM image of the corrosion layer on the surface of an ancient glass fragment note the laminar structure of the corrosion layer nbsp SEM image of a photoresist layer used in semiconductor manufacturing taken on a field emission SEM These SEMs are important in the semiconductor industry for their high resolution capabilities nbsp SEM image of the surface of a kidney stone showing tetragonal crystals of Weddellite calcium oxalate dihydrate emerging from the amorphous central part of the stone Horizontal length of the picture represents 0 5 mm of the figured original nbsp Two images of the same depth hoar snow crystal viewed through a light microscope left and as an SEM image right Note how the SEM image allows for clear perception of the fine structure details which are hard to fully make out in the light microscope image nbsp Epidermal cells from the inner surface of an onion flake Beneath the shagreen like cell walls one can see nuclei and small organelles floating in the cytoplasm This BSE image of a lanthanoid stained sample was taken without prior fixation dehydration or sputtering nbsp SEM image of stomata on the lower surface of a leaf See also EditApplications for electron microscopy Electron microscopy Energy dispersive X ray spectroscopy Cathodoluminescence microscope Forensic engineering Forensic science List of materials analysis methods Microscopy Scanning helium microscopy Teeny Ted from Turnip Town World s smallest book requires a scanning electron microscope to read Transmission electron microscopy TEM References Edit Stokes Debbie J 2008 Principles and Practice of Variable Pressure Environmental Scanning Electron Microscopy VP ESEM Chichester John Wiley amp Sons ISBN 978 0470758748 McMullan D 2006 Scanning electron microscopy 1928 1965 Scanning 17 3 175 185 doi 10 1002 sca 4950170309 PMC 2496789 McMullan D 1988 Von Ardenne and the scanning electron microscope Proc Roy Microsc Soc 23 283 288 Knoll Max 1935 Aufladepotentiel und Sekundaremission elektronenbestrahlter Korper Zeitschrift fur Technische Physik 16 467 475 GB 511204 von Ardenne Manfred Improvements in electron microscopes published 1939 08 15 von Ardenne Manfred 1938 Das Elektronen Rastermikroskop Theoretische Grundlagen Zeitschrift fur Physik in German 109 9 10 553 572 Bibcode 1938ZPhy 109 553V doi 10 1007 BF01341584 S2CID 117900835 von Ardenne Manfred 1938 Das Elektronen Rastermikroskop Praktische Ausfuhrung Zeitschrift fur Technische Physik in German 19 407 416 Zworykin VA Hillier J Snyder RL 1942 A scanning electron microscope ASTM Bull 117 15 23 McMullan D 1953 An improved scanning electron microscope for opaque specimens Proceedings of the IEE Part II Power Engineering 100 75 245 256 doi 10 1049 pi 2 1953 0095 Oatley CW Nixon WC Pease RFW 1965 Scanning electron microscopy Adv Electronics Electron Phys 21 181 247 Smith KCA Oatley CW 1955 The scanning electron microscope and its fields of application British Journal of Applied Physics 6 11 391 399 Bibcode 1955BJAP 6 391S doi 10 1088 0508 3443 6 11 304 Wells OC 1957 The construction of a scanning electron microscope and its application to the study of fibres PhD Dissertation Cambridge University a b Suzuki E 2002 High resolution scanning electron microscopy of immunogold labelled cells by the use of thin plasma coating of osmium Journal of Microscopy 208 3 153 157 doi 10 1046 j 1365 2818 2002 01082 x PMID 12460446 S2CID 42452027 a b c Goldstein G I Newbury D E Echlin P Joy D C Fiori C Lifshin E 1981 Scanning electron microscopy and x ray microanalysis New York Plenum Press ISBN 978 0 306 40768 0 Seligman Arnold M Wasserkrug Hannah L Hanker Jacob S 1966 A new staining method for enhancing contrast of lipid containing membranes and droplets in osmium tetroxide fixed tissue with osmiophilic thiocarbohydrazide TCH Journal of Cell Biology 30 2 424 432 doi 10 1083 jcb 30 2 424 PMC 2106998 PMID 4165523 Malick Linda E Wilson Richard B Stetson David 1975 Modified Thiocarbohydrazide Procedure for Scanning Electron Microscopy Routine use for Normal Pathological or Experimental Tissues Biotechnic amp Histochemistry 50 4 265 269 doi 10 3109 10520297509117069 PMID 1103373 Conrad Cyler Jones Emily Lena Newsome Seth D Schwartz Douglas W 2016 Bone isotopes eggshell and turkey husbandry at Arroyo Hondo Pueblo Journal of Archaeological Science Reports 10 566 574 Bibcode 2016JArSR 10 566C doi 10 1016 j jasrep 2016 06 016 a b c Jeffree C E Read N D 1991 Ambient and Low temperature scanning electron microscopy In Hall J L Hawes C R eds Electron Microscopy of Plant Cells London Academic Press pp 313 413 ISBN 978 0 12 318880 9 Karnovsky M J 1965 A formaldehyde glutaraldehyde fixative of high osmolality for use in electron microscopy PDF Journal of Cell Biology 27 2 1A 149A JSTOR 1604673 Kiernan J A 2000 Formaldehyde formalin paraformaldehyde and glutaraldehyde What they are and what they do Microscopy Today 2000 1 8 12 doi 10 1017 S1551929500057060 S2CID 100881495 Russell S D Daghlian C P 1985 Scanning electron microscopic observations on deembedded biological tissue sections Comparison of different fixatives and embedding materials Journal of Electron Microscopy Technique 2 5 489 495 doi 10 1002 jemt 1060020511 Chandler Douglas E Roberson Robert W 2009 Bioimaging current concepts in light and electron microscopy Sudbury Mass Jones and Bartlett Publishers ISBN 9780763738747 Faulkner Christine et al 2008 Peeking into Pit Fields A Multiple Twinning Model of Secondary Plasmodesmata Formation in Tobacco Plant Cell 20 6 1504 18 doi 10 1105 tpc 107 056903 PMC 2483367 PMID 18667640 Wergin W P Erbe E F 1994 Snow crystals capturing snow flakes for observation with the low temperature scanning electron microscope Scanning 16 Suppl IV IV88 Archived from the original on 17 February 2013 Retrieved 15 December 2012 Barnes P R F Mulvaney R Wolff E W Robinson K A 2002 A technique for the examination of polar ice using the scanning electron microscope Journal of Microscopy 205 2 118 124 doi 10 1046 j 0022 2720 2001 00981 x PMID 11879426 S2CID 35513404 Hindmarsh J P Russell A B Chen X D 2007 Fundamentals of the spray freezing of foods microstructure of frozen droplets Journal of Food Engineering 78 1 136 150 doi 10 1016 j jfoodeng 2005 09 011 Ultra high Resolution Scanning Electron Microscope SU9000 Everhart T E Thornley R F M 1960 Wide band detector for micro microampere low energy electron currents PDF Journal of Scientific Instruments 37 7 246 248 Bibcode 1960JScI 37 246E doi 10 1088 0950 7671 37 7 307 Hitachi Launches World s Highest Resolution FE SEM Nanotech Now 31 May 2011 Takaku Yasuharu Suzuki Hiroshi Ohta Isao Tsutsui Takami Matsumoto Haruko Shimomura Masatsugu Hariyama Takahiko 7 March 2015 A NanoSuit surface shield successfully protects organisms in high vacuum observations on living organisms in an FE SEM Proceedings of the Royal Society of London B Biological Sciences 282 1802 20142857 doi 10 1098 rspb 2014 2857 ISSN 0962 8452 PMC 4344158 PMID 25631998 Danilatos G D 1988 Foundations of environmental scanning electron microscopy Advances in Electronics and Electron Physics Volume 71 Advances in Electronics and Electron Physics Vol 71 pp 109 250 doi 10 1016 S0065 2539 08 60902 6 ISBN 9780120146710 US patent 4823006 Danilatos Gerasimos D and Lewis George C Integrated electron optical differential pumping imaging signal detection system for an environmental scanning electron microscope issued 18 April 1989 Danilatos G D 1990 Theory of the Gaseous Detector Device in the ESEM Advances in Electronics and Electron Physics Vol 78 pp 1 102 doi 10 1016 S0065 2539 08 60388 1 ISBN 9780120146789 US patent 4785182 Mancuso James F Maxwell William B and Danilatos Gerasimos D Secondary Electron Detector for Use in a Gaseous Atmosphere issued 15 November 1988 History of Electron Microscopy 1990s Archived 4 March 2007 at archive today sfc fr de Jonge N Ross F M 2011 Electron microscopy of specimens in liquid Nature Nanotechnology 6 8 695 704 Bibcode 2003NatMa 2 532W doi 10 1038 nmat944 PMID 12872162 S2CID 21379512 Klein Tobias Buhr Egbert Frase Carl G 2012 TSEM A Review of Scanning Electron Microscopy in Transmission Mode and Its Applications Advances in Imaging and Electron Physics Vol 171 pp 297 356 doi 10 1016 B978 0 12 394297 5 00006 4 ISBN 9780123942975 Forensic Applications of the Phenom Desktop Scanning Electron Microscope SEM AZoNano com 21 February 2014 Retrieved 11 May 2023 a b Shrivastava Priya Jain V K Nagpal Suman 1 June 2021 Gunshot residue detection technologies a review Egyptian Journal of Forensic Sciences 11 1 11 doi 10 1186 s41935 021 00223 9 ISSN 2090 5939 Burgess Jeremy 1987 Under the Microscope A Hidden World Revealed CUP Archive p 11 ISBN 978 0521399401 Showing your true colors 3D and color in electron microscopy in Lab News magazine a b c d e Mignot Christophe 2018 Color and 3D for Scanning Electron Microscopy Microscopy Today 26 3 12 17 doi 10 1017 S1551929518000482 Introduction to Electron Microscopy PDF FEI Company p 15 Retrieved 12 December 2012 Next Monday Digital Surf to Launch Revolutionary SEM Image Colorization AZO Materials 22 January 2016 Retrieved 23 January 2016 Antonovsky A 1984 The application of colour to SEM imaging for increased definition Micron and Microscopica Acta 15 2 77 84 doi 10 1016 0739 6260 84 90005 4 Danilatos G D 1986 Colour micrographs for backscattered electron signals in the SEM Scanning 9 3 8 18 doi 10 1111 j 1365 2818 1986 tb04287 x S2CID 96315383 Danilatos G D 1986 Environmental scanning electron microscopy in colour Journal of Microscopy 142 317 325 doi 10 1002 sca 4950080104 Bertazzo S Gentleman E Cloyd K L Chester A H Yacoub M H Stevens M M 2013 Nano analytical electron microscopy reveals fundamental insights into human cardiovascular tissue calcification Nature Materials 12 6 576 583 Bibcode 2013NatMa 12 576B doi 10 1038 nmat3627 hdl 10044 1 21901 PMC 5833942 PMID 23603848 Bertazzo Sergio Maidment Susannah C R Kallepitis Charalambos Fearn Sarah Stevens Molly M Xie Hai nan 9 June 2015 Fibres and cellular structures preserved in 75 million year old dinosaur specimens Nature Communications 6 7352 Bibcode 2015NatCo 6 7352B doi 10 1038 ncomms8352 PMC 4468865 PMID 26056764 Stereo SEM reconstruction using MountainsMap SEM version 7 4 on i7 2600 CPU at 3 4 GHz Butterfield Nicholas Rowe Penny M Stewart Emily Roesel David Neshyba Steven 16 March 2017 Quantitative three dimensional ice roughness from scanning electron microscopy Journal of Geophysical Research Atmospheres 122 5 3023 3025 Bibcode 2017JGRD 122 3023B doi 10 1002 2016JD026094 Butterfield Nicholas Rowe Penny M Stewart Emily Roesel David Neshyba Steven 16 March 2017 Quantitative three dimensional ice roughness from scanning electron microscopy Journal of Geophysical Research Atmospheres 122 5 3025 3041 Bibcode 2017JGRD 122 3023B doi 10 1002 2016JD026094 Baghaei Rad Leili 2007 Computational Scanning Electron Microscopy International Conference on Frontiers of Characterization and Metrology Vol 931 p 512 Bibcode 2007AIPC 931 512R doi 10 1063 1 2799427 Baghaei Rad Leili Downes Ian Ye Jun Adler David Pease R Fabian W 2007 Economic approximate models for backscattered electrons Journal of Vacuum Science and Technology 25 6 2425 Bibcode 2007JVSTB 25 2425B doi 10 1116 1 2794068 Tahmasebi Pejman Javadpour Farzam Sahimi Muhammad 2015 Multiscale and multiresolution modeling of shales and their flow and morphological properties Scientific Reports 5 16373 Bibcode 2015NatSR 516373T doi 10 1038 srep16373 PMC 4642334 PMID 26560178 Tahmasebi Pejman Javadpour Farzam Sahimi Muhammad 2015 Three Dimensional Stochastic Characterization of Shale SEM Images Transport in Porous Media 110 3 521 531 doi 10 1007 s11242 015 0570 1 S2CID 20274015 Tahmasebi Pejman Sahimi Muhammad 2012 Reconstruction of three dimensional porous media using a single thin section Physical Review E 85 6 066709 Bibcode 2012PhRvE 85f6709T doi 10 1103 PhysRevE 85 066709 PMID 23005245 S2CID 24307267 Murphy GE Lowekamp BC Zerfas PM August 2010 Ion abrasion scanning electron microscopy reveals distorted liver mitochondrial morphology in murine methylmalonic acidemia Journal of Structural Biology 171 2 125 32 doi 10 1016 j jsb 2010 04 005 PMC 2885563 PMID 20399866 Multimedia Gallery 3 D Imaging of Mammalian Cells With Ion Abrasion SEM NSF National Science Foundation www nsf gov Butterfield Nicholas Rowe Penny M Stewart Emily Roesel David Neshyba Steven 16 March 2017 Quantitative three dimensional ice roughness from scanning electron microscopy Journal of Geophysical Research Atmospheres 122 5 3023 3041 Bibcode 2017JGRD 122 3023B doi 10 1002 2016JD026094 Hermans Joen Osmond Gillian Loon Annelies van Iedema Piet Chapman Robyn Drennan John Jack Kevin Rasch Ronald Morgan Garry Zhang Zhi Monteiro Michael June 2018 Electron Microscopy Imaging of Zinc Soaps Nucleation in Oil Paint Microscopy and Microanalysis 24 3 318 322 Bibcode 2018MiMic 24 318H doi 10 1017 S1431927618000387 ISSN 1431 9276 PMID 29860951 S2CID 44166918 External links Edit nbsp Wikibooks has a book on the topic of Nanotechnology nbsp Wikimedia Commons has media related to Scanning electron microscope nbsp Wikimedia Commons has media related to Scanning electron microscopic images GeneralHowStuffWorks How Scanning Electron Microscopes Work Learn to use an SEM An online learning environment for people wanting to use an SEM Provided by Microscopy Australia Virtual SEM sparkler an interactive simulation of a scanning electron microscope SEM Multichannel color SEM imaging and with BSE DDC SEM image examples Video on the scanning electron microscope Karlsruhe University of Applied Sciences Animations and explanations on various types of microscopes including electron microscopes Universite Paris Sud HistoryEnvironmental Scanning Electron Microscope ESEM historyImagesRippel Electron Microscope Facility Archived 19 March 2007 at the Wayback Machine Many dozens of mostly biological SEM images from Dartmouth College Lanthanoid staining SEM images from Research Institute of Eye Diseases Moscow Retrieved from https en wikipedia org w index php title Scanning electron microscope amp oldid 1179717413, wikipedia, wiki, book, books, library,