An electron microscope is a microscope that uses a beam of accelerated electrons as a source of illumination. As the wavelength of an electron can be up to 100,000 times shorter than that of visible light photons, electron microscopes have a higher resolving power than light microscopes and can reveal the structure of smaller objects. A scanning transmission electron microscope has achieved better than 50 pm resolution in annular dark-field imaging mode[1] and magnifications of up to about 10,000,000× whereas most light microscopes are limited by diffraction to about 200 nm resolution and useful magnifications below 2000×.

A modern transmission electron microscope
An image of an ant in a scanning electron microscope

Electron microscopes use shaped magnetic fields to form electron optical lens systems that are analogous to the glass lenses of an optical light microscope.

Electron microscopes are used to investigate the ultrastructure of a wide range of biological and inorganic specimens including microorganisms, cells, large molecules, biopsy samples, metals, and crystals. Industrially, electron microscopes are often used for quality control and failure analysis. Modern electron microscopes produce electron micrographs using specialized digital cameras and frame grabbers to capture the images.

History

Theoretical foundations

 
Diagram of a transmission electron microscope

In 1924, physicist Louis de Broglie (Nobel Prize laureate, 1929) asserted that moderately accelerated electrons must show an associated wave, and calculated its wavelength, which would be in the order of the X rays in the electromagnetic spectrum.[2] This was later confirmed by the Davisson–Germer experiment in 1927.[3] All they provided the theoretical principles that make the electron microscope possible.

While X-rays cannot be diverted by optical means, moving electrons can be, by using electromagnetic fields as a sort of lenses, that may be arranged as in a standard optical microscope. A properly built electronic device could, then, be able to focus the electron beam onto a sample to study it.

Practical developments

 
Electron microscope constructed by Ernst Ruska in 1933

In 1926, Hans Busch developed the electromagnetic lens.

According to Dennis Gabor, the physicist Leó Szilárd tried in 1928 to convince him to build an electron microscope, for which he had filed a patent.[4] The first prototype electron microscope, capable of four-hundred-power magnification, was developed in 1931 by the physicist Ernst Ruska and the electrical engineer Max Knoll at the Berlin Technische Hochschule or Berlin Technical University.[5] The apparatus was the first practical demonstration of the principles of electron microscopy.[6] In May of the same year, Reinhold Rudenberg, the scientific director of Siemens-Schuckertwerke, obtained a patent for an electron microscope. In 1932, Ernst Lubcke of Siemens & Halske built and obtained images from a prototype electron microscope, applying the concepts described in Rudenberg's patent.[7]

In the following year, 1933, Ruska built the first electron microscope that exceeded the resolution attainable with an optical (light) microscope.[6] Four years later, in 1937, Siemens financed the work of Ernst Ruska and Bodo von Borries, and employed Helmut Ruska, Ernst's brother, to develop applications for the microscope, especially with biological specimens.[6][8] Also in 1937, Manfred von Ardenne pioneered the scanning electron microscope.[9] Siemens produced the first commercial electron microscope in 1938.[10] The first North American electron microscopes were constructed in the 1930, at the Washington State University by Anderson and Fitzsimmons [11] and the University of Toronto, by Eli Franklin Burton and students Cecil Hall, James Hillier, and Albert Prebus. Siemens produced a transmission electron microscope (TEM) in 1939.[clarification needed][12] Although current transmission electron microscopes are capable of two million-power magnification, as scientific instruments, they remain based upon Ruska's prototype.[citation needed]

Physical principle

 
Diagram illustrating the phenomena resulting from the interaction of highly energetic electrons with matter

In a typical electron gun, individual electrons, which have an elementary charge   (about   coulombs) and a mass   (about   kg), under a potential difference of   volts, an energy amount of   joules is transferred to every of them, which they absorb entirely as kinetic energy,   (assuming non-relativistic mechanics) and start to move at a speed  . Thus, being

 ,

then  .

For a voltage of  , this gives a speed of about   m/s.[13]

According to De Broglie, the (non-relativistic) associated wavelength   for a momentum   is

 ,

being   the Planck constant (about   J·s), thus

 ,

so for the said voltage it gives a wavelength of about   metres, or 0.1228 nm, well below that of the visible violet wavelength (380 nm), in the soft X-rays domain.

Note that higher voltages accelerate electrons up to noticiable relativistic effects. Then, the following expression should be used instead:[14]

 ,

where   is the speed of light in vacuum (about   m/s). See the theory of electron diffraction (relativistic) for a full explanation.

Types

Transmission electron microscope (TEM)

Operating principle of a transmission electron microscope

The original form of the electron microscope, the transmission electron microscope (TEM), uses a high voltage electron beam to illuminate the specimen and create an image. The electron beam is produced by an electron gun, commonly fitted with a tungsten filament cathode as the electron source. The electron beam is accelerated by an anode typically at +100 keV (40 to 400 keV) with respect to the cathode, focused by electrostatic and electromagnetic lenses, and transmitted through the specimen that is in part transparent to electrons and in part scatters them out of the beam. When it emerges from the specimen, the electron beam carries information about the structure of the specimen that is magnified by the objective lens system of the microscope. The spatial variation in this information (the "image") may be viewed by projecting the magnified electron image onto a fluorescent viewing screen coated with a phosphor or scintillator material such as zinc sulfide. Alternatively, the image can be photographically recorded by exposing a photographic film or plate directly to the electron beam, or a high-resolution phosphor may be coupled by means of a lens optical system or a fibre optic light-guide to the sensor of a digital camera. The image detected by the digital camera may be displayed on a monitor or computer.

The resolution of TEMs is limited primarily by spherical aberration, but a new generation of hardware correctors can reduce spherical aberration to increase the resolution in high-resolution transmission electron microscopy (HRTEM) to below 0.5 angstrom (50 picometres),[1] enabling magnifications above 50 million times.[15] The ability of HRTEM to determine the positions of atoms within materials is useful for nano-technologies research and development.[16]

Transmission electron microscopes are often used in electron diffraction mode. The advantages of electron diffraction over X-ray crystallography are that the specimen need not be a single crystal or even a polycrystalline powder, and also that the Fourier transform reconstruction of the object's magnified structure occurs physically and thus avoids the need for solving the phase problem faced by the X-ray crystallographers after obtaining their X-ray diffraction patterns.

One major disadvantage of the transmission electron microscope is the need for extremely thin sections of the specimens, typically about 100 nanometers. Creating these thin sections for biological and materials specimens is technically very challenging. Semiconductor thin sections can be made using a focused ion beam. Biological tissue specimens are chemically fixed, dehydrated and embedded in a polymer resin to stabilize them sufficiently to allow ultrathin sectioning. Sections of biological specimens, organic polymers, and similar materials may require staining with heavy atom labels in order to achieve the required image contrast.

Serial-section electron microscope (ssEM)

One application of TEM is serial-section electron microscopy (ssEM), for example in analyzing the connectivity in volumetric samples of brain tissue by imaging many thin sections in sequence.[17] This may be achieved by introducing a milling method into the imaging pipeline, by which successive slices of a 3D volume are exposed to the beam and imaged. These methods include Serial Block Face SEM (SB-SEM)[18] and Focused Ion Beam-SEM (FIB-SEM).[19][20] Pre-processing of volumes to create many slices which are imaged in an automated fashion have recently achieved high-throughput imaging of volumes up to 1 mm3.[21] Using this method, entire local neuronal microcircuits may be resolved, though the equipment and time requirements for this are still significant: imaging a 1 mm3 block of brain tissue required 6 months of near-continuous imaging by six TEM running in parallel.

Scanning transmission electron microscope (STEM)

The STEM rasters a focused incident probe across a specimen that (as with the TEM) has been thinned to facilitate detection of electrons scattered through the specimen. The high resolution of the TEM is thus possible in STEM. The focusing action (and aberrations) occur before the electrons hit the specimen in the STEM, but afterward in the TEM. The STEMs use of SEM-like beam rastering simplifies annular dark-field imaging, and other analytical techniques, but also means that image data is acquired in serial rather than in parallel fashion. Often TEM can be equipped with the scanning option and then it can function both as TEM and STEM.

Scanning electron microscope (SEM)

Operating principe of a scanning electron microscope
 
Image of Bacillus subtilis taken with a 1960s electron microscope

The SEM produces images by probing the specimen with a focused electron beam that is scanned across a rectangular area of the specimen (raster scanning). When the electron beam interacts with the specimen, it loses energy by a variety of mechanisms. The lost energy is converted into alternative forms such as heat, emission of low-energy secondary electrons and high-energy backscattered electrons, light emission (cathodoluminescence) or X-ray emission, all of which provide signals carrying information about the properties of the specimen surface, such as its topography and composition. The image displayed by an SEM maps the varying intensity of any of these signals into the image in a position corresponding to the position of the beam on the specimen when the signal was generated. In the SEM image of an ant shown below and to the right, the image was constructed from signals produced by a secondary electron detector, the normal or conventional imaging mode in most SEMs.

Generally, the image resolution of an SEM is lower than that of a TEM. However, because the SEM images the surface of a sample rather than its interior, the electrons do not have to travel through the sample. This reduces the need for extensive sample preparation to thin the specimen to electron transparency. The SEM is able to image bulk samples that can fit on its stage and still be maneuvered, including a height less than the working distance being used, often 4 millimeters for high-resolution images. The SEM also has a great depth of field, and so can produce images that are good representations of the three-dimensional surface shape of the sample. Another advantage of SEMs comes with environmental scanning electron microscopes (ESEM) that can produce images of good quality and resolution with hydrated samples or in low, rather than high, vacuum or under chamber gases. This facilitates imaging unfixed biological samples that are unstable in the high vacuum of conventional electron microscopes.

Reflection electron microscope (REM)

In the reflection electron microscope (REM) as in the TEM, an electron beam is incident on a surface but instead of using the transmission (TEM) or secondary electrons (SEM), the reflected beam of elastically scattered electrons is detected. This technique is typically coupled with reflection high energy electron diffraction (RHEED) and reflection high-energy loss spectroscopy (RHELS).[citation needed] Another variation is spin-polarized low-energy electron microscopy (SPLEEM), which is used for looking at the microstructure of magnetic domains.[22]

Colour

In their most common configurations, electron microscopes produce images with a single brightness value per pixel, with the results usually rendered in greyscale.[23] However, often these images are then colourized through the use of feature-detection software, or simply by hand-editing using a graphics editor. This may be done to clarify structure or for aesthetic effect and generally does not add new information about the specimen.[24]

In some configurations information about several specimen properties is gathered per pixel, usually by the use of multiple detectors.[25] In SEM, the attributes of topography and material contrast can be obtained by a pair of backscattered electron detectors and such attributes can be superimposed in a single colour image by assigning a different primary colour to each attribute.[26] Similarly, a combination of backscattered and secondary electron signals can be assigned to different colours and superimposed on a single colour micrograph displaying simultaneously the properties of the specimen.[27]

Some types of detectors used in SEM have analytical capabilities, and can provide several items of data at each pixel. Examples are the energy-dispersive X-ray spectroscopy (EDS) detectors used in elemental analysis and cathodoluminescence microscope (CL) systems that analyse the intensity and spectrum of electron-induced luminescence in (for example) geological specimens. In SEM systems using these detectors, it is common to colour code the signals and superimpose them in a single colour image, so that differences in the distribution of the various components of the specimen can be seen clearly and compared. Optionally, the standard secondary electron image can be merged with the one or more compositional channels, so that the specimen's structure and composition can be compared. Such images can be made while maintaining the full integrity of the original signal, which is not modified in any way.

Sample preparation

 
An insect coated in gold for viewing with a scanning electron microscope

Materials to be viewed in an electron microscope may require processing to produce a suitable sample. The technique required varies depending on the specimen and the analysis required:

  • Chemical fixation – for biological specimens aims to stabilize the specimen's mobile macromolecular structure by chemical crosslinking of proteins with aldehydes such as formaldehyde and glutaraldehyde, and lipids with osmium tetroxide.
  • Negative stain – suspensions containing nanoparticles or fine biological material (such as viruses and bacteria) are briefly mixed with a dilute solution of an electron-opaque solution such as ammonium molybdate, uranyl acetate (or formate), or phosphotungstic acid. This mixture is applied to a suitably coated EM grid, blotted, then allowed to dry. Viewing of this preparation in the TEM should be carried out without delay for best results. The method is important in microbiology for fast but crude morphological identification, but can also be used as the basis for high-resolution 3D reconstruction using EM tomography methodology when carbon films are used for support. Negative staining is also used for observation of nanoparticles.
  • Cryofixation – freezing a specimen so rapidly, in liquid ethane that the water forms vitreous (non-crystalline) ice. This preserves the specimen in a snapshot of its solution state. An entire field called cryo-electron microscopy has branched from this technique. With the development of cryo-electron microscopy of vitreous sections (CEMOVIS), it is now possible to observe samples from virtually any biological specimen close to its native state.[citation needed]
  • Dehydration – or replacement of water with organic solvents such as ethanol or acetone, followed by critical point drying or infiltration with embedding resins. Also freeze drying.
  • Embedding, biological specimens – after dehydration, tissue for observation in the transmission electron microscope is embedded so it can be sectioned ready for viewing. To do this the tissue is passed through a 'transition solvent' such as propylene oxide (epoxypropane) or acetone and then infiltrated with an epoxy resin such as Araldite, Epon, or Durcupan;[28] tissues may also be embedded directly in water-miscible acrylic resin. After the resin has been polymerized (hardened) the sample is thin sectioned (ultrathin sections) and stained – it is then ready for viewing.
  • Embedding, materials – after embedding in resin, the specimen is usually ground and polished to a mirror-like finish using ultra-fine abrasives. The polishing process must be performed carefully to minimize scratches and other polishing artifacts that reduce image quality.
  • Metal shadowing – Metal (e.g. platinum) is evaporated from an overhead electrode and applied to the surface of a biological sample at an angle.[29] The surface topography results in variations in the thickness of the metal that are seen as variations in brightness and contrast in the electron microscope image.
  • Replication – A surface shadowed with metal (e.g. platinum, or a mixture of carbon and platinum) at an angle is coated with pure carbon evaporated from carbon electrodes at right angles to the surface. This is followed by removal of the specimen material (e.g. in an acid bath, using enzymes or by mechanical separation[30]) to produce a surface replica that records the surface ultrastructure and can be examined using transmission electron microscopy.
  • Sectioning – produces thin slices of the specimen, semitransparent to electrons. These can be cut on an ultramicrotome with a glass or diamond knife to produce ultra-thin sections about 60–90 nm thick. Disposable glass knives are also used because they can be made in the lab and are much cheaper.
  • Staining – uses heavy metals such as lead, uranium or tungsten to scatter imaging electrons and thus give contrast between different structures, since many (especially biological) materials are nearly "transparent" to electrons (weak phase objects). In biology, specimens can be stained "en bloc" before embedding and also later after sectioning. Typically thin sections are stained for several minutes with an aqueous or alcoholic solution of uranyl acetate followed by aqueous lead citrate.[31]
  • Freeze-fracture or freeze-etch – a preparation method[32][33][34] particularly useful for examining lipid membranes and their incorporated proteins in "face on" view.[35][36][37]
     
    Freeze-fracturing helps to peel open membranes to allow visualization of what is inside
     
    External face of bakers yeast membrane showing the small holes where proteins are fractured out, sometimes as small ring patterns.
    The fresh tissue or cell suspension is frozen rapidly (cryofixation), then fractured by breaking[38] (or by using a microtome)[37] while maintained at liquid nitrogen temperature. The cold fractured surface (sometimes "etched" by increasing the temperature to about −100 °C for several minutes to let some ice sublime)[37] is then shadowed with evaporated platinum or gold at an average angle of 45° in a high vacuum evaporator. The second coat of carbon, evaporated perpendicular to the average surface plane is often performed to improve the stability of the replica coating. The specimen is returned to room temperature and pressure, then the extremely fragile "pre-shadowed" metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids, hypochlorite solution or SDS detergent. The still-floating replica is thoroughly washed free from residual chemicals, carefully fished up on fine grids, dried then viewed in the TEM.
  • Freeze-fracture replica immunogold labeling (FRIL) – the freeze-fracture method has been modified to allow the identification of the components of the fracture face by immunogold labeling. Instead of removing all the underlying tissue of the thawed replica as the final step before viewing in the microscope the tissue thickness is minimized during or after the fracture process. The thin layer of tissue remains bound to the metal replica so it can be immunogold labeled with antibodies to the structures of choice. The thin layer of the original specimen on the replica with gold attached allows the identification of structures in the fracture plane.[39] There are also related methods which label the surface of etched cells[40] and other replica labeling variations.[41]
  • Ion beam milling – thins samples until they are transparent to electrons by firing ions (typically argon) at the surface from an angle and sputtering material from the surface. A subclass of this is focused ion beam milling, where gallium ions are used to produce an electron transparent membrane in a specific region of the sample, for example through a device within a microprocessor. Ion beam milling may also be used for cross-section polishing prior to SEM analysis of materials that are difficult to prepare using mechanical polishing.
  • Conductive coating – an ultrathin coating of electrically conducting material, deposited either by high vacuum evaporation or by low vacuum sputter coating of the sample. This is done to prevent the accumulation of static electric fields at the specimen due to the electron irradiation required during imaging. The coating materials include gold, gold/palladium, platinum, tungsten, graphite, etc.
  • Earthing – to avoid electrical charge accumulation on a conductively coated sample, it is usually electrically connected to the metal sample holder. Often an electrically conductive adhesive is used for this purpose.

Disadvantages

 
JEOL transmission and scanning electron microscope made in the mid-1970s

Electron microscopes are expensive to build and maintain, but the capital and running costs of confocal light microscope systems now overlaps with those of basic electron microscopes. Microscopes designed to achieve high resolutions must be housed in stable buildings (sometimes underground) with special services such as magnetic field canceling systems.

The samples largely have to be viewed in vacuum, as the molecules that make up air would scatter the electrons. An exception is liquid-phase electron microscopy[42] using either a closed liquid cell or an environmental chamber, for example, in the environmental scanning electron microscope, which allows hydrated samples to be viewed in a low-pressure (up to 20 Torr or 2.7 kPa) wet environment. Various techniques for in situ electron microscopy of gaseous samples have been developed as well.[43]

Scanning electron microscopes operating in conventional high-vacuum mode usually image conductive specimens; therefore non-conductive materials require conductive coating (gold/palladium alloy, carbon, osmium, etc.). The low-voltage mode of modern microscopes makes possible the observation of non-conductive specimens without coating. Non-conductive materials can be imaged also by a variable pressure (or environmental) scanning electron microscope.

Small, stable specimens such as carbon nanotubes, diatom frustules and small mineral crystals (asbestos fibres, for example) require no special treatment before being examined in the electron microscope. Samples of hydrated materials, including almost all biological specimens, have to be prepared in various ways to stabilize them, reduce their thickness (ultrathin sectioning) and increase their electron optical contrast (staining). These processes may result in artifacts, but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods. Since the 1980s, analysis of cryofixed, vitrified specimens has also become increasingly used by scientists, further confirming the validity of this technique.[44][45][46]

Applications

See also

References

  1. ^ a b Erni, Rolf; Rossell, MD; Kisielowski, C; Dahmen, U (2009). "Atomic-Resolution Imaging with a Sub-50-pm Electron Probe". Physical Review Letters. 102 (9): 096101. Bibcode:2009PhRvL.102i6101E. doi:10.1103/PhysRevLett.102.096101. PMID 19392535.
  2. ^ Feynman, R., QED: The Strange Theory of Light and Matter, Penguin 1990 Edition, p. 84.
  3. ^ Davisson, C. J.; Germer, L. H. (1928). "Reflection of Electrons by a Crystal of Nickel". Proceedings of the National Academy of Sciences of the United States of America. 14 (4): 317–322. Bibcode:1928PNAS...14..317D. doi:10.1073/pnas.14.4.317. PMC 1085484. PMID 16587341.
  4. ^ Dannen, Gene (1998) Leo Szilard the Inventor: A Slideshow (1998, Budapest, conference talk). dannen.com
  5. ^ Mathys, Daniel, Zentrum für Mikroskopie, University of Basel: Die Entwicklung der Elektronenmikroskopie vom Bild über die Analyse zum Nanolabor, p. 8
  6. ^ a b c Ruska, Ernst (1986). "Ernst Ruska Autobiography". Nobel Foundation. Retrieved 2010-01-31.
  7. ^ Rudenberg, H. Gunther; Rudenberg, Paul G. (2010). "Origin and Background of the Invention of the Electron Microscope". Advances in Imaging and Electron Physics. Vol. 160. pp. 207–286. doi:10.1016/S1076-5670(10)60006-7. ISBN 978-0-12-381017-5.
  8. ^ Kruger, DH; Schneck, P; Gelderblom, HR (May 2000). "Helmut Ruska and the visualisation of viruses". The Lancet. 355 (9216): 1713–1717. doi:10.1016/S0140-6736(00)02250-9. PMID 10905259. S2CID 12347337.
  9. ^ Ardenne, M. Von; Beischer, D. (1940). "Untersuchung von Metalloxyd-Rauchen mit dem Universal-Elektronenmikroskop" [Investigation of metal oxide smoking with the universal electron microscope]. Zeitschrift für Elektrochemie und Angewandte Physikalische Chemie (in German). 46 (4): 270–277. doi:10.1002/bbpc.19400460406 (inactive 31 December 2022).{{cite journal}}: CS1 maint: DOI inactive as of December 2022 (link)
  10. ^ History of electron microscopy, 1931–2000. Authors.library.caltech.edu (2002-12-10). Retrieved on 2017-04-29.
  11. ^ "North America's first electron microscope".
  12. ^ . Inventor of the Week: Archive. 2003-05-01. Archived from the original on 2003-08-23. Retrieved 2010-01-31.
  13. ^ "The speed of electrons". IOPSpark. Institute of Physics.
  14. ^ Kirkland, Earl (2010). Advanced computing in electron microscopy. New York: Springer. ISBN 978-1-4419-6533-2. OCLC 668095602.
  15. ^ . Office of Basic Energy Sciences, U.S. Department of Energy. 2006-05-26. Archived from the original on 2010-02-01. Retrieved 2010-01-31.
  16. ^ O'Keefe MA; Allard LF (2004-01-18). "Sub-Ångstrom Electron Microscopy for Sub-Ångstrom Nano-Metrology" (PDF). Information Bridge: DOE Scientific and Technical Information – Sponsored by OSTI. {{cite journal}}: Cite journal requires |journal= (help)
  17. ^ Yoo, Inwan, David GC Hildebrand, Willie F. Tobin, Wei-Chung Allen Lee, and Won-Ki Jeong. "ssEMnet: Serial-section Electron Microscopy Image Registration using a Spatial Transformer Network with Learned Features" In Deep Learning in Medical Image Analysis and Multimodal Learning for Clinical Decision Support, pp. 249-257. Springer, Cham, 2017.
  18. ^ Denk, Winfried; Horstmann, Heinz (2004-10-19). "Serial Block-Face Scanning Electron Microscopy to Reconstruct Three-Dimensional Tissue Nanostructure". PLOS Biology. 2 (11): e329. doi:10.1371/journal.pbio.0020329. ISSN 1545-7885. PMC 524270. PMID 15514700.
  19. ^ Xu, C Shan; Hayworth, Kenneth J; Lu, Zhiyuan; Grob, Patricia; Hassan, Ahmed M; García-Cerdán, José G; Niyogi, Krishna K; Nogales, Eva; Weinberg, Richard J; Hess, Harald F (2017-05-13). Nathans, Jeremy (ed.). "Enhanced FIB-SEM systems for large-volume 3D imaging". eLife. 6: e25916. doi:10.7554/eLife.25916. ISSN 2050-084X. PMC 5476429. PMID 28500755.
  20. ^ Hoffman, David P.; Shtengel, Gleb; Xu, C. Shan; Campbell, Kirby R.; Freeman, Melanie; Wang, Lei; Milkie, Daniel E.; Pasolli, H. Amalia; Iyer, Nirmala; Bogovic, John A.; Stabley, Daniel R. (2020-01-17). "Correlative three-dimensional super-resolution and block-face electron microscopy of whole vitreously frozen cells". Science. 367 (6475). doi:10.1126/science.aaz5357. PMC 7339343. PMID 31949053.
  21. ^ Yin, Wenjing; Brittain, Derrick; Borseth, Jay; Scott, Marie E.; Williams, Derric; Perkins, Jedediah; Own, Christopher S.; Murfitt, Matthew; Torres, Russel M.; Kapner, Daniel; Mahalingam, Gayathri (2020-10-02). "A petascale automated imaging pipeline for mapping neuronal circuits with high-throughput transmission electron microscopy". Nature Communications. 11 (1): 4949. Bibcode:2020NatCo..11.4949Y. doi:10.1038/s41467-020-18659-3. ISSN 2041-1723. PMC 7532165. PMID 33009388.
  22. ^ . National Center for Electron Microscopy (NCEM). Archived from the original on 2010-05-29. Retrieved 2010-01-31.
  23. ^ Burgess, Jeremy (1987). Under the Microscope: A Hidden World Revealed. CUP Archive. p. 11. ISBN 978-0-521-39940-1.
  24. ^ "Introduction to Electron Microscopy" (PDF). FEI Company. p. 15. Retrieved 12 December 2012.
  25. ^ Antonovsky, A. (1984). "The application of colour to sem imaging for increased definition". Micron and Microscopica Acta. 15 (2): 77–84. doi:10.1016/0739-6260(84)90005-4.
  26. ^ Danilatos, G.D. (1986). "Colour micrographs for backscattered electron signals in the SEM". Scanning. 9 (3): 8–18. doi:10.1111/j.1365-2818.1986.tb04287.x. S2CID 96315383.
  27. ^ Danilatos, G.D. (1986). "Environmental scanning electron microscopy in colour". Journal of Microscopy. 142: 317–325. doi:10.1002/sca.4950080104.
  28. ^ Luft, J.H. (1961). "Improvements in epoxy resin embedding methods". The Journal of Biophysical and Biochemical Cytology. Vol. 9, no. 2. p. 409. PMC 2224998. PMID 13764136.
  29. ^ Williams, R. C.; Wyckoff, R. W. (1945-06-08). "Electron shadow micrography of the tobacco mosaic virus protein". Science. 101 (2632): 594–596. Bibcode:1945Sci...101..594W. doi:10.1126/science.101.2632.594. PMID 17739444. S2CID 44930705.
  30. ^ Juniper, B.E.; Bradley, D.E. (1958). "The carbon replica technique in the study of the ultrastructure of leaf surfaces". Journal of Ultrastructure Research. 2 (1): 16–27. doi:10.1016/s0022-5320(58)90045-5.
  31. ^ Reynolds, E. S. (1963). "The use of lead citrate at high pH as an electron-opaque stain in electron microscopy". Journal of Cell Biology. 17 (1): 208–212. doi:10.1083/jcb.17.1.208. PMC 2106263. PMID 13986422.
  32. ^ Meryman H.T. and Kafig E. (1955). The study of frozen specimens, ice crystals and ices crystal growth by electron microscopy. Naval Med. Res. Ints. Rept NM 000 018.01.09 Vol. 13 pp 529–544
  33. ^ Steere, Russell L. (1957-01-25). "Electron microscopy of structural detail in frozen biological specimens". The Journal of Biophysical and Biochemical Cytology. 3 (1): 45–60. doi:10.1083/jcb.3.1.45. PMC 2224015. PMID 13416310.
  34. ^ Isailović, Tanja M.; Todosijević, Marija N.; Đorđević, Sanela M.; Savić, Snežana D. (2017-01-01), Čalija, Bojan (ed.), "Chapter 7 - Natural Surfactants-Based Micro/Nanoemulsion Systems for NSAIDs—Practical Formulation Approach, Physicochemical and Biopharmaceutical Characteristics/Performances", Microsized and Nanosized Carriers for Nonsteroidal Anti-Inflammatory Drugs, Boston: Academic Press, pp. 179–217, doi:10.1016/b978-0-12-804017-1.00007-8, ISBN 978-0-12-804017-1, retrieved 2020-10-22
  35. ^ Moor H, Mühlethaler K (1963). "Fine structure in frozen-etched yeast cells". The Journal of Cell Biology. 17 (3): 609–628. doi:10.1083/jcb.17.3.609. PMC 2106217. PMID 19866628.
  36. ^ Black, Joel A. (1990-01-01), Conn, P. Michael (ed.), "[20] - Use of Freeze-Fracture in Neurobiology", Methods in Neurosciences, Quantitative and Qualitative Microscopy, Academic Press, 3: 343–360, doi:10.1016/b978-0-12-185255-9.50025-0, retrieved 2020-10-22
  37. ^ a b c Stillwell, William (2016-01-01), Stillwell, William (ed.), "Chapter 11 - Long-Range Membrane Properties", An Introduction to Biological Membranes (Second Edition), Elsevier, pp. 221–245, doi:10.1016/b978-0-444-63772-7.00011-7, ISBN 978-0-444-63772-7, retrieved 2020-10-22
  38. ^ Bullivant, Stanley; Ames, Adelbert (1966-06-01). "A simple freeze-fracture replication method for electron microscopy". The Journal of Cell Biology. 29 (3): 435–447. doi:10.1083/jcb.29.3.435. PMC 2106967. PMID 5962938.
  39. ^ Gruijters, W. T.; Kistler, J; Bullivant, S; Goodenough, D. A. (1987-03-01). "Immunolocalization of MP70 in lens fiber 16-17-nm intercellular junctions". The Journal of Cell Biology. 104 (3): 565–572. doi:10.1083/jcb.104.3.565. PMC 2114558. PMID 3818793.
  40. ^ da Silva, Pedro Pinto; Branton, Daniel (1970-06-01). "Membrane splitting in freeze-etching". The Journal of Cell Biology. 45 (3): 598–605. doi:10.1083/jcb.45.3.598. PMC 2107921. PMID 4918216.
  41. ^ Rash, J. E.; Johnson, T. J.; Hudson, C. S.; Giddings, F. D.; Graham, W. F.; Eldefrawi, M. E. (1982-11-01). "Labelled-replica techniques: post-shadow labelling of intramembrane particles in freeze-fracture replicas". Journal of Microscopy. 128 (Pt 2): 121–138. doi:10.1111/j.1365-2818.1982.tb00444.x. PMID 6184475. S2CID 45238172.
  42. ^ de Jonge, N.; Ross, F.M. (2011). "Electron microscopy of specimens in liquid". Nature Nanotechnology. 6 (8): 695–704. Bibcode:2003NatMa...2..532W. doi:10.1038/nmat944. PMID 12872162. S2CID 21379512.
  43. ^ Gai, P.L.; Boyes, E.D. (2009). "Advances in atomic resolution in situ environmental transmission electron microscopy and 1A aberration corrected in situ electron microscopy". Microsc Res Tech. 72 (3): 153–164. arXiv:1705.05754. doi:10.1002/jemt.20668. PMID 19140163. S2CID 1746538.
  44. ^ Adrian, Marc; Dubochet, Jacques; Lepault, Jean; McDowall, Alasdair W. (1984). "Cryo-electron microscopy of viruses". Nature (Submitted manuscript). 308 (5954): 32–36. Bibcode:1984Natur.308...32A. doi:10.1038/308032a0. PMID 6322001. S2CID 4319199.
  45. ^ Sabanay, I.; Arad, T.; Weiner, S.; Geiger, B. (1991). "Study of vitrified, unstained frozen tissue sections by cryoimmunoelectron microscopy". Journal of Cell Science. 100 (1): 227–236. doi:10.1242/jcs.100.1.227. PMID 1795028.
  46. ^ Kasas, S.; Dumas, G.; Dietler, G.; Catsicas, S.; Adrian, M. (2003). "Vitrification of cryoelectron microscopy specimens revealed by high-speed photographic imaging". Journal of Microscopy. 211 (1): 48–53. doi:10.1046/j.1365-2818.2003.01193.x. PMID 12839550. S2CID 40058086.
  47. ^ Boehme, L.; Bresin, M.; Botman, A.; Ranney, J.; Hastings, J.T. (2015). "Focused electron beam induced etching of copper in sulfuric acid solutions". Nanotechnology. 26 (49): 495301. Bibcode:2015Nanot..26W5301B. doi:10.1088/0957-4484/26/49/495301. PMID 26567988. S2CID 21286639.
  48. ^ Kacher, J.; Cui, B.; Robertson, I.M. (2015). "In situ and tomographic characterization of damage and dislocation processes in irradiated metallic alloys by transmission electron microscopy". Journal of Materials Research. 30 (9): 1202–1213. Bibcode:2015JMatR..30.1202K. doi:10.1557/jmr.2015.14.
  49. ^ Rai, R.S.; Subramanian, S. (2009). "Role of transmission electron microscopy in the semiconductor industry for process development and failure analysis". Progress in Crystal Growth and Characterization of Materials. 55 (3–4): 63–97. doi:10.1016/j.pcrysgrow.2009.09.002.
  50. ^ Morris, G.J.; Goodrich, M.; Acton, E.; Fonseca, F. (2006). "The high viscosity encountered during freezing in glycerol solutions: Effects on cryopreservation". Cryobiology. 52 (3): 323–334. doi:10.1016/j.cryobiol.2006.01.003. PMID 16499898.
  51. ^ a b von Appen, Alexander; Beck, Martin (May 2016). "Structure Determination of the Nuclear Pore Complex with Three-Dimensional Cryo electron Microscopy". Journal of Molecular Biology. 428 (10): 2001–10. doi:10.1016/j.jmb.2016.01.004. PMC 4898182. PMID 26791760.
  52. ^ Florian, P.E.; Rouillé, Y.; Ruta, S.; Nichita, N.; Roseanu, A. (2016). "Recent advances in human viruses imaging studies". Journal of Basic Microbiology. 56 (6): 591–607. doi:10.1002/jobm.201500575. PMID 27059598. S2CID 12737742.
  53. ^ a b Cushnie, T.P.; O’Driscoll, N.H.; Lamb, A.J. (2016). "Morphological and ultrastructural changes in bacterial cells as an indicator of antibacterial mechanism of action". Cellular and Molecular Life Sciences. 73 (23): 4471–4492. doi:10.1007/s00018-016-2302-2. hdl:10059/2129. PMID 27392605. S2CID 2065821.
  54. ^ Li, M.-H.; Yang, Y.-Q.; Huang, B.; Luo, X.; Zhang, W.; Han, M.; Ru, J.-G. (2014). "Development of advanced electron tomography in materials science based on TEM and STEM". Transactions of Nonferrous Metals Society of China. 24 (10): 3031–3050. doi:10.1016/S1003-6326(14)63441-5.
  55. ^ Li, W.J.; Shao, L.Y.; Zhang, D.Z.; Ro, C.U.; Hu, M.; Bi, X.H.; Geng, H.; Matsuki, A.; Niu, H.Y.; Chen, J.M. (2016). "A review of single aerosol particle studies in the atmosphere of East Asia: morphology, mixing state, source, and heterogeneous reactions". Journal of Cleaner Production. 112 (2): 1330–1349. doi:10.1016/j.jclepro.2015.04.050.
  56. ^ Sousa, R.G.; Esteves, T.; Rocha, S.; Figueiredo, F.; Quelhas, P.; Silva, L.M. (2015). Automatic detection of immunogold particles from electron microscopy images. Image Analysis and Recognition. Lecture Notes in Computer Science. Vol. 9164. pp. 377–384. doi:10.1007/978-3-319-20801-5_41. ISBN 978-3-319-20800-8.
  57. ^ Perkins, G.A. (2014). "The Use of miniSOG in the Localization of Mitochondrial Proteins". Mitochondrial Function. Methods in Enzymology. Vol. 547. pp. 165–179. doi:10.1016/B978-0-12-801415-8.00010-2. ISBN 9780128014158. PMID 25416358.
  58. ^ Chen, X.D.; Ren, L.Q.; Zheng, B.; Liu, H. (2013). "Physics and engineering aspects of cell and tissue imaging systems: microscopic devices and computer assisted diagnosis". Biophotonics in Pathology: Pathology at the Crossroads. 185 (Biophotonics in Pathology): 1–22. doi:10.3233/978-1-61499-234-9-1. PMID 23542929.
  59. ^ Fagerland, J.A.; Wall, H.G.; Pandher, K.; LeRoy, B.E.; Gagne, G.D. (2012). "Ultrastructural analysis in preclinical safety evaluation". Toxicologic Pathology. 40 (2): 391–402. doi:10.1177/0192623311430239. PMID 22215513. S2CID 206458999.
  60. ^ Heider, S.; Metzner, C. (2014). "Quantitative real-time single particle analysis of virions". Virology. 462–463: 199–206. doi:10.1016/j.virol.2014.06.005. PMC 4139191. PMID 24999044.
  61. ^ Tsekouras, G.; Mozer, A.J.; Wallace, G.G. (2008). "Enhanced performance of dye sensitized solar cells utilizing platinum electrodeposit counter electrodes". Journal of the Electrochemical Society. 155 (7): K124–K128. Bibcode:2008JElS..155K.124T. doi:10.1149/1.2919107.
  62. ^ Besenius, P.; Portale, G.; Bomans, P.H.H.; Janssen, H.M.; Palmans, A.R.A.; Meijer, E.W. (2010). "Controlling the growth and shape of chiral supramolecular polymers in water". Proceedings of the National Academy of Sciences of the United States of America. 107 (42): 17888–17893. Bibcode:2010PNAS..10717888B. doi:10.1073/pnas.1009592107. PMC 2964246. PMID 20921365.
  63. ^ Furuya, K. (2008). "Nanofabrication by advanced electron microscopy using intense and focused beam". Science and Technology of Advanced Materials. 9 (1): Article 014110. Bibcode:2008STAdM...9a4110F. doi:10.1088/1468-6996/9/1/014110. PMC 5099805. PMID 27877936.
  64. ^ Kosasih, Felix Utama; Ducati, Caterina (May 2018). "Characterising degradation of perovskite solar cells through in-situ and operando electron microscopy". Nano Energy. 47: 243–256. doi:10.1016/j.nanoen.2018.02.055.
  65. ^ Maloy, Stuart A.; Sommer, Walter F.; James, Michael R.; Romero, Tobias J.; Lopez, Manuel R.; Zimmermann, Eugene; Ledbetter, James M. (13 May 2017). "The Accelerator Production of Tritium Materials Test Program". Nuclear Technology. 132 (1): 103–114. doi:10.13182/nt00-a3132. S2CID 94639273.
  66. ^ Ukraintsev, Vladimir (27 February 2012). "Review of reference metrology for nanotechnology: significance, challenges, and solutions". Journal of Micro/Nanolithography, MEMS, and MOEMS. 11 (1): 011010. doi:10.1117/1.JMM.11.1.011010.
  67. ^ Wilhelmi, O.; Roussel, L.; Faber, P.; Reyntjens, S.; Daniel, G. (June 2010). "Focussed ion beam fabrication of large and complex nanopatterns". Journal of Experimental Nanoscience. 5 (3): 244–250. Bibcode:2010JENan...5..244W. doi:10.1080/17458080903487448. S2CID 283449.
  68. ^ Vogt, E.T.C.; Whiting, G.T.; Chowdhury, A.D.; Weckhuysen, B.M. (2015). Zeolites and zeotypes for oil and gas conversion. Advances in Catalysis. Vol. 58. pp. 143–314. doi:10.1016/bs.acat.2015.10.001. ISBN 9780128021262.
  69. ^ Lai, Shih-En; Hong, Ying-Jhan; Chen, Yu-Ting; Kang, Yu-Ting; Chang, Pin; Yew, Tri-Rung (18 September 2015). "Direct-Writing of Cu Nano-Patterns with an Electron Beam". Microscopy and Microanalysis. 21 (6): 1639–43. Bibcode:2015MiMic..21.1639L. doi:10.1017/S1431927615015111. PMID 26381450. S2CID 30803055.
  70. ^ Sicignano, Angelo; Di Monaco, Rossella; Masi, Paolo; Cavella, Silvana (October 2015). "From raw material to dish: pasta quality step by step". Journal of the Science of Food and Agriculture. 95 (13): 2579–87. doi:10.1002/jsfa.7176. PMID 25783568.
  71. ^ Brożek-Mucha, Zuzanna (2014). "Scanning Electron Microscopy and X-Ray Microanalysis for Chemical and Morphological Characterisation of the Inorganic Component of Gunshot Residue: Selected Problems". BioMed Research International. 2014: 428038. doi:10.1155/2014/428038. PMC 4082842. PMID 25025050.
  72. ^ Carbonell-Verdu, A.; Garcia-Sanoguera, D.; Jorda-Vilaplana, A.; Sanchez-Nacher, L.; Balart, R. (2016). "A new biobased plasticizer for poly(vinyl chloride) based on epoxidized cottonseed oil". Journal of Applied Polymer Science. 33 (27): 43642. doi:10.1002/app.43642. hdl:10251/82834.
  73. ^ Ding, Jie; Zhang, Zhiming (1 May 2015). "Micro-characterization of dissimilar metal weld joint for connecting pipe-nozzle to safe-end in generation III nuclear power plant". Acta Metall Sin. 51 (4): 425–39. doi:10.11900/0412.1961.2014.00299 (inactive 31 December 2022).{{cite journal}}: CS1 maint: DOI inactive as of December 2022 (link)
  74. ^ Tsikouras, Basilios; Pe-Piper, Georgia; Piper, David J.W.; Schaffer, Michael (June 2011). "Varietal heavy mineral analysis of sediment provenance, Lower Cretaceous Scotian Basin, eastern Canada". Sedimentary Geology. 237 (3–4): 150–165. Bibcode:2011SedG..237..150T. doi:10.1016/j.sedgeo.2011.02.011.
  75. ^ Li, Xiang; Jiang, Chuan; Pan, Lili; Zhang, Haoyang; Hu, Lang; Li, Tianxue; Yang, Xinghao (15 July 2014). "Effects of preparing techniques and aging on dissolution behavior of the solid dispersions of NF/Soluplus/Kollidon SR: identification and classification by a combined analysis by FT-IR spectroscopy and computational approaches". Drug Development and Industrial Pharmacy. 41 (1): 2–14. doi:10.3109/03639045.2014.938080. PMID 25026247. S2CID 32460608.

External links

  • An Introduction to Electron Microscopy 2013-07-19 at the Wayback Machine: resources for teachers and students
  • Cell Centered Database – Electron microscopy data
  • Science Aid: Electron Microscopy: high school (GCSE, A Level) resource

General

  • Animations and explanations of various types of microscopy including electron microscopy (Université Paris Sud)
  • Environmental Scanning Electron Microscopy (ESEM)
  • : graphics and images illustrating various procedures
  • Eva Nogales's Seminar: "Introduction to Electron Microscopy"
  • FEI Image Contest: FEI has had a microscopy image contest every year since 2008
  • Introduction to electron microscopy by David Szondy
  • Nanohedron.com image gallery: images generated by electron microscopy
  • X-ray element analysis in electron microscopy: information portal with X-ray microanalysis and EDX contents

History

  • John H.L. Watson's recollections at the University of Toronto when he worked with Hillier and Prebus 2019-03-30 at the Wayback Machine
  • (Archives Center, National Museum of American History, Smithsonian Institution)

Other

  • Albert Lleal. Natural history subjects at Scanning Electron Microscope SEM 2019-03-30 at the Wayback Machine
  • EM Images

electron, microscope, confused, with, scanning, tunneling, microscope, electron, microscope, microscope, that, uses, beam, accelerated, electrons, source, illumination, wavelength, electron, times, shorter, than, that, visible, light, photons, electron, micros. Not to be confused with Scanning tunneling microscope An electron microscope is a microscope that uses a beam of accelerated electrons as a source of illumination As the wavelength of an electron can be up to 100 000 times shorter than that of visible light photons electron microscopes have a higher resolving power than light microscopes and can reveal the structure of smaller objects A scanning transmission electron microscope has achieved better than 50 pm resolution in annular dark field imaging mode 1 and magnifications of up to about 10 000 000 whereas most light microscopes are limited by diffraction to about 200 nm resolution and useful magnifications below 2000 A modern transmission electron microscope An image of an ant in a scanning electron microscope Electron microscopes use shaped magnetic fields to form electron optical lens systems that are analogous to the glass lenses of an optical light microscope Electron microscopes are used to investigate the ultrastructure of a wide range of biological and inorganic specimens including microorganisms cells large molecules biopsy samples metals and crystals Industrially electron microscopes are often used for quality control and failure analysis Modern electron microscopes produce electron micrographs using specialized digital cameras and frame grabbers to capture the images Contents 1 History 1 1 Theoretical foundations 1 2 Practical developments 2 Physical principle 3 Types 3 1 Transmission electron microscope TEM 3 1 1 Serial section electron microscope ssEM 3 1 2 Scanning transmission electron microscope STEM 3 2 Scanning electron microscope SEM 3 3 Reflection electron microscope REM 4 Colour 5 Sample preparation 6 Disadvantages 7 Applications 8 See also 9 References 10 External links 10 1 General 10 2 History 10 3 OtherHistory EditSee also Transmission electron microscopy History Theoretical foundations Edit Diagram of a transmission electron microscope In 1924 physicist Louis de Broglie Nobel Prize laureate 1929 asserted that moderately accelerated electrons must show an associated wave and calculated its wavelength which would be in the order of the X rays in the electromagnetic spectrum 2 This was later confirmed by the Davisson Germer experiment in 1927 3 All they provided the theoretical principles that make the electron microscope possible While X rays cannot be diverted by optical means moving electrons can be by using electromagnetic fields as a sort of lenses that may be arranged as in a standard optical microscope A properly built electronic device could then be able to focus the electron beam onto a sample to study it Practical developments Edit Electron microscope constructed by Ernst Ruska in 1933 In 1926 Hans Busch developed the electromagnetic lens According to Dennis Gabor the physicist Leo Szilard tried in 1928 to convince him to build an electron microscope for which he had filed a patent 4 The first prototype electron microscope capable of four hundred power magnification was developed in 1931 by the physicist Ernst Ruska and the electrical engineer Max Knoll at the Berlin Technische Hochschule or Berlin Technical University 5 The apparatus was the first practical demonstration of the principles of electron microscopy 6 In May of the same year Reinhold Rudenberg the scientific director of Siemens Schuckertwerke obtained a patent for an electron microscope In 1932 Ernst Lubcke of Siemens amp Halske built and obtained images from a prototype electron microscope applying the concepts described in Rudenberg s patent 7 In the following year 1933 Ruska built the first electron microscope that exceeded the resolution attainable with an optical light microscope 6 Four years later in 1937 Siemens financed the work of Ernst Ruska and Bodo von Borries and employed Helmut Ruska Ernst s brother to develop applications for the microscope especially with biological specimens 6 8 Also in 1937 Manfred von Ardenne pioneered the scanning electron microscope 9 Siemens produced the first commercial electron microscope in 1938 10 The first North American electron microscopes were constructed in the 1930 at the Washington State University by Anderson and Fitzsimmons 11 and the University of Toronto by Eli Franklin Burton and students Cecil Hall James Hillier and Albert Prebus Siemens produced a transmission electron microscope TEM in 1939 clarification needed 12 Although current transmission electron microscopes are capable of two million power magnification as scientific instruments they remain based upon Ruska s prototype citation needed Physical principle Edit Diagram illustrating the phenomena resulting from the interaction of highly energetic electrons with matter See also Electron diffraction In a typical electron gun individual electrons which have an elementary charge e displaystyle e about 1 6 10 19 displaystyle 1 6 times 10 19 coulombs and a mass m displaystyle m about 9 1 10 31 displaystyle 9 1 times 10 31 kg under a potential difference of V displaystyle V volts an energy amount of e V displaystyle e cdot V joules is transferred to every of them which they absorb entirely as kinetic energy E K m v 2 displaystyle E text K text mv 2 assuming non relativistic mechanics and start to move at a speed v displaystyle v Thus being e V E K displaystyle eV E text K then v 2 e V m displaystyle v sqrt 2eV m For a voltage of V 100 V displaystyle V 100 text V this gives a speed of about v 6 10 6 displaystyle v 6 times 10 6 m s 13 According to De Broglie the non relativistic associated wavelength l displaystyle lambda for a momentum p m v displaystyle p mv is l h p h m v displaystyle lambda frac h p frac h mv being h displaystyle h the Planck constant about 6 626 10 34 displaystyle 6 626 times 10 34 J s thus l h 2 m e V displaystyle lambda frac h sqrt 2meV so for the said voltage it gives a wavelength of about l 1 228 10 10 displaystyle lambda 1 228 times 10 10 metres or 0 1228 nm well below that of the visible violet wavelength 380 nm in the soft X rays domain Note that higher voltages accelerate electrons up to noticiable relativistic effects Then the following expression should be used instead 14 l h c e V 2 m c 2 e V displaystyle lambda frac hc sqrt eV 2mc 2 eV where c displaystyle c is the speed of light in vacuum about c 3 10 8 displaystyle c 3 times 10 8 m s See the theory of electron diffraction relativistic for a full explanation Types EditTransmission electron microscope TEM Edit Main article Transmission electron microscope source source source source source source source source source source source source source source Operating principle of a transmission electron microscope The original form of the electron microscope the transmission electron microscope TEM uses a high voltage electron beam to illuminate the specimen and create an image The electron beam is produced by an electron gun commonly fitted with a tungsten filament cathode as the electron source The electron beam is accelerated by an anode typically at 100 keV 40 to 400 keV with respect to the cathode focused by electrostatic and electromagnetic lenses and transmitted through the specimen that is in part transparent to electrons and in part scatters them out of the beam When it emerges from the specimen the electron beam carries information about the structure of the specimen that is magnified by the objective lens system of the microscope The spatial variation in this information the image may be viewed by projecting the magnified electron image onto a fluorescent viewing screen coated with a phosphor or scintillator material such as zinc sulfide Alternatively the image can be photographically recorded by exposing a photographic film or plate directly to the electron beam or a high resolution phosphor may be coupled by means of a lens optical system or a fibre optic light guide to the sensor of a digital camera The image detected by the digital camera may be displayed on a monitor or computer The resolution of TEMs is limited primarily by spherical aberration but a new generation of hardware correctors can reduce spherical aberration to increase the resolution in high resolution transmission electron microscopy HRTEM to below 0 5 angstrom 50 picometres 1 enabling magnifications above 50 million times 15 The ability of HRTEM to determine the positions of atoms within materials is useful for nano technologies research and development 16 Transmission electron microscopes are often used in electron diffraction mode The advantages of electron diffraction over X ray crystallography are that the specimen need not be a single crystal or even a polycrystalline powder and also that the Fourier transform reconstruction of the object s magnified structure occurs physically and thus avoids the need for solving the phase problem faced by the X ray crystallographers after obtaining their X ray diffraction patterns One major disadvantage of the transmission electron microscope is the need for extremely thin sections of the specimens typically about 100 nanometers Creating these thin sections for biological and materials specimens is technically very challenging Semiconductor thin sections can be made using a focused ion beam Biological tissue specimens are chemically fixed dehydrated and embedded in a polymer resin to stabilize them sufficiently to allow ultrathin sectioning Sections of biological specimens organic polymers and similar materials may require staining with heavy atom labels in order to achieve the required image contrast Serial section electron microscope ssEM Edit One application of TEM is serial section electron microscopy ssEM for example in analyzing the connectivity in volumetric samples of brain tissue by imaging many thin sections in sequence 17 This may be achieved by introducing a milling method into the imaging pipeline by which successive slices of a 3D volume are exposed to the beam and imaged These methods include Serial Block Face SEM SB SEM 18 and Focused Ion Beam SEM FIB SEM 19 20 Pre processing of volumes to create many slices which are imaged in an automated fashion have recently achieved high throughput imaging of volumes up to 1 mm3 21 Using this method entire local neuronal microcircuits may be resolved though the equipment and time requirements for this are still significant imaging a 1 mm3 block of brain tissue required 6 months of near continuous imaging by six TEM running in parallel Scanning transmission electron microscope STEM Edit Main article Scanning transmission electron microscopy The STEM rasters a focused incident probe across a specimen that as with the TEM has been thinned to facilitate detection of electrons scattered through the specimen The high resolution of the TEM is thus possible in STEM The focusing action and aberrations occur before the electrons hit the specimen in the STEM but afterward in the TEM The STEMs use of SEM like beam rastering simplifies annular dark field imaging and other analytical techniques but also means that image data is acquired in serial rather than in parallel fashion Often TEM can be equipped with the scanning option and then it can function both as TEM and STEM Scanning electron microscope SEM Edit source source source source source source source source source source source source source source Operating principe of a scanning electron microscope Main article Scanning electron microscope Image of Bacillus subtilis taken with a 1960s electron microscope The SEM produces images by probing the specimen with a focused electron beam that is scanned across a rectangular area of the specimen raster scanning When the electron beam interacts with the specimen it loses energy by a variety of mechanisms The lost energy is converted into alternative forms such as heat emission of low energy secondary electrons and high energy backscattered electrons light emission cathodoluminescence or X ray emission all of which provide signals carrying information about the properties of the specimen surface such as its topography and composition The image displayed by an SEM maps the varying intensity of any of these signals into the image in a position corresponding to the position of the beam on the specimen when the signal was generated In the SEM image of an ant shown below and to the right the image was constructed from signals produced by a secondary electron detector the normal or conventional imaging mode in most SEMs Generally the image resolution of an SEM is lower than that of a TEM However because the SEM images the surface of a sample rather than its interior the electrons do not have to travel through the sample This reduces the need for extensive sample preparation to thin the specimen to electron transparency The SEM is able to image bulk samples that can fit on its stage and still be maneuvered including a height less than the working distance being used often 4 millimeters for high resolution images The SEM also has a great depth of field and so can produce images that are good representations of the three dimensional surface shape of the sample Another advantage of SEMs comes with environmental scanning electron microscopes ESEM that can produce images of good quality and resolution with hydrated samples or in low rather than high vacuum or under chamber gases This facilitates imaging unfixed biological samples that are unstable in the high vacuum of conventional electron microscopes Reflection electron microscope REM Edit In the reflection electron microscope REM as in the TEM an electron beam is incident on a surface but instead of using the transmission TEM or secondary electrons SEM the reflected beam of elastically scattered electrons is detected This technique is typically coupled with reflection high energy electron diffraction RHEED and reflection high energy loss spectroscopy RHELS citation needed Another variation is spin polarized low energy electron microscopy SPLEEM which is used for looking at the microstructure of magnetic domains 22 Colour EditIn their most common configurations electron microscopes produce images with a single brightness value per pixel with the results usually rendered in greyscale 23 However often these images are then colourized through the use of feature detection software or simply by hand editing using a graphics editor This may be done to clarify structure or for aesthetic effect and generally does not add new information about the specimen 24 SEM image of Cobaea scandens pollen The same after semi automatic coloring Arbitrary colors help identifying the various elements of the structure In some configurations information about several specimen properties is gathered per pixel usually by the use of multiple detectors 25 In SEM the attributes of topography and material contrast can be obtained by a pair of backscattered electron detectors and such attributes can be superimposed in a single colour image by assigning a different primary colour to each attribute 26 Similarly a combination of backscattered and secondary electron signals can be assigned to different colours and superimposed on a single colour micrograph displaying simultaneously the properties of the specimen 27 Some types of detectors used in SEM have analytical capabilities and can provide several items of data at each pixel Examples are the energy dispersive X ray spectroscopy EDS detectors used in elemental analysis and cathodoluminescence microscope CL systems that analyse the intensity and spectrum of electron induced luminescence in for example geological specimens In SEM systems using these detectors it is common to colour code the signals and superimpose them in a single colour image so that differences in the distribution of the various components of the specimen can be seen clearly and compared Optionally the standard secondary electron image can be merged with the one or more compositional channels so that the specimen s structure and composition can be compared Such images can be made while maintaining the full integrity of the original signal which is not modified in any way Sample preparation Edit An insect coated in gold for viewing with a scanning electron microscope Materials to be viewed in an electron microscope may require processing to produce a suitable sample The technique required varies depending on the specimen and the analysis required Chemical fixation for biological specimens aims to stabilize the specimen s mobile macromolecular structure by chemical crosslinking of proteins with aldehydes such as formaldehyde and glutaraldehyde and lipids with osmium tetroxide Negative stain suspensions containing nanoparticles or fine biological material such as viruses and bacteria are briefly mixed with a dilute solution of an electron opaque solution such as ammonium molybdate uranyl acetate or formate or phosphotungstic acid This mixture is applied to a suitably coated EM grid blotted then allowed to dry Viewing of this preparation in the TEM should be carried out without delay for best results The method is important in microbiology for fast but crude morphological identification but can also be used as the basis for high resolution 3D reconstruction using EM tomography methodology when carbon films are used for support Negative staining is also used for observation of nanoparticles Cryofixation freezing a specimen so rapidly in liquid ethane that the water forms vitreous non crystalline ice This preserves the specimen in a snapshot of its solution state An entire field called cryo electron microscopy has branched from this technique With the development of cryo electron microscopy of vitreous sections CEMOVIS it is now possible to observe samples from virtually any biological specimen close to its native state citation needed Dehydration or replacement of water with organic solvents such as ethanol or acetone followed by critical point drying or infiltration with embedding resins Also freeze drying Embedding biological specimens after dehydration tissue for observation in the transmission electron microscope is embedded so it can be sectioned ready for viewing To do this the tissue is passed through a transition solvent such as propylene oxide epoxypropane or acetone and then infiltrated with an epoxy resin such as Araldite Epon or Durcupan 28 tissues may also be embedded directly in water miscible acrylic resin After the resin has been polymerized hardened the sample is thin sectioned ultrathin sections and stained it is then ready for viewing Embedding materials after embedding in resin the specimen is usually ground and polished to a mirror like finish using ultra fine abrasives The polishing process must be performed carefully to minimize scratches and other polishing artifacts that reduce image quality Metal shadowing Metal e g platinum is evaporated from an overhead electrode and applied to the surface of a biological sample at an angle 29 The surface topography results in variations in the thickness of the metal that are seen as variations in brightness and contrast in the electron microscope image Replication A surface shadowed with metal e g platinum or a mixture of carbon and platinum at an angle is coated with pure carbon evaporated from carbon electrodes at right angles to the surface This is followed by removal of the specimen material e g in an acid bath using enzymes or by mechanical separation 30 to produce a surface replica that records the surface ultrastructure and can be examined using transmission electron microscopy Sectioning produces thin slices of the specimen semitransparent to electrons These can be cut on an ultramicrotome with a glass or diamond knife to produce ultra thin sections about 60 90 nm thick Disposable glass knives are also used because they can be made in the lab and are much cheaper Staining uses heavy metals such as lead uranium or tungsten to scatter imaging electrons and thus give contrast between different structures since many especially biological materials are nearly transparent to electrons weak phase objects In biology specimens can be stained en bloc before embedding and also later after sectioning Typically thin sections are stained for several minutes with an aqueous or alcoholic solution of uranyl acetate followed by aqueous lead citrate 31 Freeze fracture or freeze etch a preparation method 32 33 34 particularly useful for examining lipid membranes and their incorporated proteins in face on view 35 36 37 Freeze fracturing helps to peel open membranes to allow visualization of what is inside External face of bakers yeast membrane showing the small holes where proteins are fractured out sometimes as small ring patterns The fresh tissue or cell suspension is frozen rapidly cryofixation then fractured by breaking 38 or by using a microtome 37 while maintained at liquid nitrogen temperature The cold fractured surface sometimes etched by increasing the temperature to about 100 C for several minutes to let some ice sublime 37 is then shadowed with evaporated platinum or gold at an average angle of 45 in a high vacuum evaporator The second coat of carbon evaporated perpendicular to the average surface plane is often performed to improve the stability of the replica coating The specimen is returned to room temperature and pressure then the extremely fragile pre shadowed metal replica of the fracture surface is released from the underlying biological material by careful chemical digestion with acids hypochlorite solution or SDS detergent The still floating replica is thoroughly washed free from residual chemicals carefully fished up on fine grids dried then viewed in the TEM Freeze fracture replica immunogold labeling FRIL the freeze fracture method has been modified to allow the identification of the components of the fracture face by immunogold labeling Instead of removing all the underlying tissue of the thawed replica as the final step before viewing in the microscope the tissue thickness is minimized during or after the fracture process The thin layer of tissue remains bound to the metal replica so it can be immunogold labeled with antibodies to the structures of choice The thin layer of the original specimen on the replica with gold attached allows the identification of structures in the fracture plane 39 There are also related methods which label the surface of etched cells 40 and other replica labeling variations 41 Ion beam milling thins samples until they are transparent to electrons by firing ions typically argon at the surface from an angle and sputtering material from the surface A subclass of this is focused ion beam milling where gallium ions are used to produce an electron transparent membrane in a specific region of the sample for example through a device within a microprocessor Ion beam milling may also be used for cross section polishing prior to SEM analysis of materials that are difficult to prepare using mechanical polishing Conductive coating an ultrathin coating of electrically conducting material deposited either by high vacuum evaporation or by low vacuum sputter coating of the sample This is done to prevent the accumulation of static electric fields at the specimen due to the electron irradiation required during imaging The coating materials include gold gold palladium platinum tungsten graphite etc Earthing to avoid electrical charge accumulation on a conductively coated sample it is usually electrically connected to the metal sample holder Often an electrically conductive adhesive is used for this purpose Disadvantages Edit JEOL transmission and scanning electron microscope made in the mid 1970s Electron microscopes are expensive to build and maintain but the capital and running costs of confocal light microscope systems now overlaps with those of basic electron microscopes Microscopes designed to achieve high resolutions must be housed in stable buildings sometimes underground with special services such as magnetic field canceling systems The samples largely have to be viewed in vacuum as the molecules that make up air would scatter the electrons An exception is liquid phase electron microscopy 42 using either a closed liquid cell or an environmental chamber for example in the environmental scanning electron microscope which allows hydrated samples to be viewed in a low pressure up to 20 Torr or 2 7 kPa wet environment Various techniques for in situ electron microscopy of gaseous samples have been developed as well 43 Scanning electron microscopes operating in conventional high vacuum mode usually image conductive specimens therefore non conductive materials require conductive coating gold palladium alloy carbon osmium etc The low voltage mode of modern microscopes makes possible the observation of non conductive specimens without coating Non conductive materials can be imaged also by a variable pressure or environmental scanning electron microscope Small stable specimens such as carbon nanotubes diatom frustules and small mineral crystals asbestos fibres for example require no special treatment before being examined in the electron microscope Samples of hydrated materials including almost all biological specimens have to be prepared in various ways to stabilize them reduce their thickness ultrathin sectioning and increase their electron optical contrast staining These processes may result in artifacts but these can usually be identified by comparing the results obtained by using radically different specimen preparation methods Since the 1980s analysis of cryofixed vitrified specimens has also become increasingly used by scientists further confirming the validity of this technique 44 45 46 Applications EditSemiconductor and data storageCircuit edit 47 Defect analysis 48 Failure analysis 49 Biology and life sciences Cryobiology 50 Cryo electron microscopy 51 Diagnostic electron microscopy 52 Drug research e g antibiotics 53 Electron tomography 54 Particle analysis 55 Particle detection 56 Protein localization 57 Structural biology 51 Tissue imaging 58 Toxicology 59 Virology e g viral load monitoring 60 Materials researchDevice testing and characterization 61 Dynamic materials experiments 62 Electron beam induced deposition 63 In situ characterisation 64 Materials qualification 65 Medical research 53 Nanometrology 66 Nanoprototyping 67 IndustryChemical Petrochemical 68 Direct beam writing fabrication 69 Food science 70 Forensics 71 Fractography 72 Micro characterization 73 Mining mineral liberation analysis 74 Pharmaceutical QC 75 See also EditAcronyms in microscopy Electron diffraction Electron energy loss spectroscopy EELS Electron microscope images Energy filtered transmission electron microscopy EFTEM Environmental scanning electron microscope ESEM Field emission microscope HiRISE Immune electron microscopy In situ electron microscopy Karnovsky fixative Microscope image processing Microscopy Nanoscience Nanotechnology Neutron microscope Quantum microscopy Scanning confocal electron microscopy Scanning electron microscope SEM Scanning tunneling microscope Surface science Transmission Electron Aberration Corrected Microscope X ray diffraction X ray microscope Low energy electron microscopy Hemispherical electron energy analyzerReferences Edit a b Erni Rolf Rossell MD Kisielowski C Dahmen U 2009 Atomic Resolution Imaging with a Sub 50 pm Electron Probe Physical Review Letters 102 9 096101 Bibcode 2009PhRvL 102i6101E doi 10 1103 PhysRevLett 102 096101 PMID 19392535 Feynman R QED The Strange Theory of Light and Matter Penguin 1990 Edition p 84 Davisson C J Germer L H 1928 Reflection of Electrons by a Crystal of Nickel Proceedings of the National Academy of Sciences of the United States of America 14 4 317 322 Bibcode 1928PNAS 14 317D doi 10 1073 pnas 14 4 317 PMC 1085484 PMID 16587341 Dannen Gene 1998 Leo Szilard the Inventor A Slideshow 1998 Budapest conference talk dannen com Mathys Daniel Zentrum fur Mikroskopie University of Basel Die Entwicklung der Elektronenmikroskopie vom Bild uber die Analyse zum Nanolabor p 8 a b c Ruska Ernst 1986 Ernst Ruska Autobiography Nobel Foundation Retrieved 2010 01 31 Rudenberg H Gunther Rudenberg Paul G 2010 Origin and Background of the Invention of the Electron Microscope Advances in Imaging and Electron Physics Vol 160 pp 207 286 doi 10 1016 S1076 5670 10 60006 7 ISBN 978 0 12 381017 5 Kruger DH Schneck P Gelderblom HR May 2000 Helmut Ruska and the visualisation of viruses The Lancet 355 9216 1713 1717 doi 10 1016 S0140 6736 00 02250 9 PMID 10905259 S2CID 12347337 Ardenne M Von Beischer D 1940 Untersuchung von Metalloxyd Rauchen mit dem Universal Elektronenmikroskop Investigation of metal oxide smoking with the universal electron microscope Zeitschrift fur Elektrochemie und Angewandte Physikalische Chemie in German 46 4 270 277 doi 10 1002 bbpc 19400460406 inactive 31 December 2022 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint DOI inactive as of December 2022 link History of electron microscopy 1931 2000 Authors library caltech edu 2002 12 10 Retrieved on 2017 04 29 North America s first electron microscope James Hillier Inventor of the Week Archive 2003 05 01 Archived from the original on 2003 08 23 Retrieved 2010 01 31 The speed of electrons IOPSpark Institute of Physics Kirkland Earl 2010 Advanced computing in electron microscopy New York Springer ISBN 978 1 4419 6533 2 OCLC 668095602 The Scale of Things Office of Basic Energy Sciences U S Department of Energy 2006 05 26 Archived from the original on 2010 02 01 Retrieved 2010 01 31 O Keefe MA Allard LF 2004 01 18 Sub Angstrom Electron Microscopy for Sub Angstrom Nano Metrology PDF Information Bridge DOE Scientific and Technical Information Sponsored by OSTI a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help Yoo Inwan David GC Hildebrand Willie F Tobin Wei Chung Allen Lee and Won Ki Jeong ssEMnet Serial section Electron Microscopy Image Registration using a Spatial Transformer Network with Learned Features In Deep Learning in Medical Image Analysis and Multimodal Learning for Clinical Decision Support pp 249 257 Springer Cham 2017 Denk Winfried Horstmann Heinz 2004 10 19 Serial Block Face Scanning Electron Microscopy to Reconstruct Three Dimensional Tissue Nanostructure PLOS Biology 2 11 e329 doi 10 1371 journal pbio 0020329 ISSN 1545 7885 PMC 524270 PMID 15514700 Xu C Shan Hayworth Kenneth J Lu Zhiyuan Grob Patricia Hassan Ahmed M Garcia Cerdan Jose G Niyogi Krishna K Nogales Eva Weinberg Richard J Hess Harald F 2017 05 13 Nathans Jeremy ed Enhanced FIB SEM systems for large volume 3D imaging eLife 6 e25916 doi 10 7554 eLife 25916 ISSN 2050 084X PMC 5476429 PMID 28500755 Hoffman David P Shtengel Gleb Xu C Shan Campbell Kirby R Freeman Melanie Wang Lei Milkie Daniel E Pasolli H Amalia Iyer Nirmala Bogovic John A Stabley Daniel R 2020 01 17 Correlative three dimensional super resolution and block face electron microscopy of whole vitreously frozen cells Science 367 6475 doi 10 1126 science aaz5357 PMC 7339343 PMID 31949053 Yin Wenjing Brittain Derrick Borseth Jay Scott Marie E Williams Derric Perkins Jedediah Own Christopher S Murfitt Matthew Torres Russel M Kapner Daniel Mahalingam Gayathri 2020 10 02 A petascale automated imaging pipeline for mapping neuronal circuits with high throughput transmission electron microscopy Nature Communications 11 1 4949 Bibcode 2020NatCo 11 4949Y doi 10 1038 s41467 020 18659 3 ISSN 2041 1723 PMC 7532165 PMID 33009388 SPLEEM National Center for Electron Microscopy NCEM Archived from the original on 2010 05 29 Retrieved 2010 01 31 Burgess Jeremy 1987 Under the Microscope A Hidden World Revealed CUP Archive p 11 ISBN 978 0 521 39940 1 Introduction to Electron Microscopy PDF FEI Company p 15 Retrieved 12 December 2012 Antonovsky A 1984 The application of colour to sem imaging for increased definition Micron and Microscopica Acta 15 2 77 84 doi 10 1016 0739 6260 84 90005 4 Danilatos G D 1986 Colour micrographs for backscattered electron signals in the SEM Scanning 9 3 8 18 doi 10 1111 j 1365 2818 1986 tb04287 x S2CID 96315383 Danilatos G D 1986 Environmental scanning electron microscopy in colour Journal of Microscopy 142 317 325 doi 10 1002 sca 4950080104 Luft J H 1961 Improvements in epoxy resin embedding methods The Journal of Biophysical and Biochemical Cytology Vol 9 no 2 p 409 PMC 2224998 PMID 13764136 Williams R C Wyckoff R W 1945 06 08 Electron shadow micrography of the tobacco mosaic virus protein Science 101 2632 594 596 Bibcode 1945Sci 101 594W doi 10 1126 science 101 2632 594 PMID 17739444 S2CID 44930705 Juniper B E Bradley D E 1958 The carbon replica technique in the study of the ultrastructure of leaf surfaces Journal of Ultrastructure Research 2 1 16 27 doi 10 1016 s0022 5320 58 90045 5 Reynolds E S 1963 The use of lead citrate at high pH as an electron opaque stain in electron microscopy Journal of Cell Biology 17 1 208 212 doi 10 1083 jcb 17 1 208 PMC 2106263 PMID 13986422 Meryman H T and Kafig E 1955 The study of frozen specimens ice crystals and ices crystal growth by electron microscopy Naval Med Res Ints Rept NM 000 018 01 09 Vol 13 pp 529 544 Steere Russell L 1957 01 25 Electron microscopy of structural detail in frozen biological specimens The Journal of Biophysical and Biochemical Cytology 3 1 45 60 doi 10 1083 jcb 3 1 45 PMC 2224015 PMID 13416310 Isailovic Tanja M Todosijevic Marija N Đorđevic Sanela M Savic Snezana D 2017 01 01 Calija Bojan ed Chapter 7 Natural Surfactants Based Micro Nanoemulsion Systems for NSAIDs Practical Formulation Approach Physicochemical and Biopharmaceutical Characteristics Performances Microsized and Nanosized Carriers for Nonsteroidal Anti Inflammatory Drugs Boston Academic Press pp 179 217 doi 10 1016 b978 0 12 804017 1 00007 8 ISBN 978 0 12 804017 1 retrieved 2020 10 22 Moor H Muhlethaler K 1963 Fine structure in frozen etched yeast cells The Journal of Cell Biology 17 3 609 628 doi 10 1083 jcb 17 3 609 PMC 2106217 PMID 19866628 Black Joel A 1990 01 01 Conn P Michael ed 20 Use of Freeze Fracture in Neurobiology Methods in Neurosciences Quantitative and Qualitative Microscopy Academic Press 3 343 360 doi 10 1016 b978 0 12 185255 9 50025 0 retrieved 2020 10 22 a b c Stillwell William 2016 01 01 Stillwell William ed Chapter 11 Long Range Membrane Properties An Introduction to Biological Membranes Second Edition Elsevier pp 221 245 doi 10 1016 b978 0 444 63772 7 00011 7 ISBN 978 0 444 63772 7 retrieved 2020 10 22 Bullivant Stanley Ames Adelbert 1966 06 01 A simple freeze fracture replication method for electron microscopy The Journal of Cell Biology 29 3 435 447 doi 10 1083 jcb 29 3 435 PMC 2106967 PMID 5962938 Gruijters W T Kistler J Bullivant S Goodenough D A 1987 03 01 Immunolocalization of MP70 in lens fiber 16 17 nm intercellular junctions The Journal of Cell Biology 104 3 565 572 doi 10 1083 jcb 104 3 565 PMC 2114558 PMID 3818793 da Silva Pedro Pinto Branton Daniel 1970 06 01 Membrane splitting in freeze etching The Journal of Cell Biology 45 3 598 605 doi 10 1083 jcb 45 3 598 PMC 2107921 PMID 4918216 Rash J E Johnson T J Hudson C S Giddings F D Graham W F Eldefrawi M E 1982 11 01 Labelled replica techniques post shadow labelling of intramembrane particles in freeze fracture replicas Journal of Microscopy 128 Pt 2 121 138 doi 10 1111 j 1365 2818 1982 tb00444 x PMID 6184475 S2CID 45238172 de Jonge N Ross F M 2011 Electron microscopy of specimens in liquid Nature Nanotechnology 6 8 695 704 Bibcode 2003NatMa 2 532W doi 10 1038 nmat944 PMID 12872162 S2CID 21379512 Gai P L Boyes E D 2009 Advances in atomic resolution in situ environmental transmission electron microscopy and 1A aberration corrected in situ electron microscopy Microsc Res Tech 72 3 153 164 arXiv 1705 05754 doi 10 1002 jemt 20668 PMID 19140163 S2CID 1746538 Adrian Marc Dubochet Jacques Lepault Jean McDowall Alasdair W 1984 Cryo electron microscopy of viruses Nature Submitted manuscript 308 5954 32 36 Bibcode 1984Natur 308 32A doi 10 1038 308032a0 PMID 6322001 S2CID 4319199 Sabanay I Arad T Weiner S Geiger B 1991 Study of vitrified unstained frozen tissue sections by cryoimmunoelectron microscopy Journal of Cell Science 100 1 227 236 doi 10 1242 jcs 100 1 227 PMID 1795028 Kasas S Dumas G Dietler G Catsicas S Adrian M 2003 Vitrification of cryoelectron microscopy specimens revealed by high speed photographic imaging Journal of Microscopy 211 1 48 53 doi 10 1046 j 1365 2818 2003 01193 x PMID 12839550 S2CID 40058086 Boehme L Bresin M Botman A Ranney J Hastings J T 2015 Focused electron beam induced etching of copper in sulfuric acid solutions Nanotechnology 26 49 495301 Bibcode 2015Nanot 26W5301B doi 10 1088 0957 4484 26 49 495301 PMID 26567988 S2CID 21286639 Kacher J Cui B Robertson I M 2015 In situ and tomographic characterization of damage and dislocation processes in irradiated metallic alloys by transmission electron microscopy Journal of Materials Research 30 9 1202 1213 Bibcode 2015JMatR 30 1202K doi 10 1557 jmr 2015 14 Rai R S Subramanian S 2009 Role of transmission electron microscopy in the semiconductor industry for process development and failure analysis Progress in Crystal Growth and Characterization of Materials 55 3 4 63 97 doi 10 1016 j pcrysgrow 2009 09 002 Morris G J Goodrich M Acton E Fonseca F 2006 The high viscosity encountered during freezing in glycerol solutions Effects on cryopreservation Cryobiology 52 3 323 334 doi 10 1016 j cryobiol 2006 01 003 PMID 16499898 a b von Appen Alexander Beck Martin May 2016 Structure Determination of the Nuclear Pore Complex with Three Dimensional Cryo electron Microscopy Journal of Molecular Biology 428 10 2001 10 doi 10 1016 j jmb 2016 01 004 PMC 4898182 PMID 26791760 Florian P E Rouille Y Ruta S Nichita N Roseanu A 2016 Recent advances in human viruses imaging studies Journal of Basic Microbiology 56 6 591 607 doi 10 1002 jobm 201500575 PMID 27059598 S2CID 12737742 a b Cushnie T P O Driscoll N H Lamb A J 2016 Morphological and ultrastructural changes in bacterial cells as an indicator of antibacterial mechanism of action Cellular and Molecular Life Sciences 73 23 4471 4492 doi 10 1007 s00018 016 2302 2 hdl 10059 2129 PMID 27392605 S2CID 2065821 Li M H Yang Y Q Huang B Luo X Zhang W Han M Ru J G 2014 Development of advanced electron tomography in materials science based on TEM and STEM Transactions of Nonferrous Metals Society of China 24 10 3031 3050 doi 10 1016 S1003 6326 14 63441 5 Li W J Shao L Y Zhang D Z Ro C U Hu M Bi X H Geng H Matsuki A Niu H Y Chen J M 2016 A review of single aerosol particle studies in the atmosphere of East Asia morphology mixing state source and heterogeneous reactions Journal of Cleaner Production 112 2 1330 1349 doi 10 1016 j jclepro 2015 04 050 Sousa R G Esteves T Rocha S Figueiredo F Quelhas P Silva L M 2015 Automatic detection of immunogold particles from electron microscopy images Image Analysis and Recognition Lecture Notes in Computer Science Vol 9164 pp 377 384 doi 10 1007 978 3 319 20801 5 41 ISBN 978 3 319 20800 8 Perkins G A 2014 The Use of miniSOG in the Localization of Mitochondrial Proteins Mitochondrial Function Methods in Enzymology Vol 547 pp 165 179 doi 10 1016 B978 0 12 801415 8 00010 2 ISBN 9780128014158 PMID 25416358 Chen X D Ren L Q Zheng B Liu H 2013 Physics and engineering aspects of cell and tissue imaging systems microscopic devices and computer assisted diagnosis Biophotonics in Pathology Pathology at the Crossroads 185 Biophotonics in Pathology 1 22 doi 10 3233 978 1 61499 234 9 1 PMID 23542929 Fagerland J A Wall H G Pandher K LeRoy B E Gagne G D 2012 Ultrastructural analysis in preclinical safety evaluation Toxicologic Pathology 40 2 391 402 doi 10 1177 0192623311430239 PMID 22215513 S2CID 206458999 Heider S Metzner C 2014 Quantitative real time single particle analysis of virions Virology 462 463 199 206 doi 10 1016 j virol 2014 06 005 PMC 4139191 PMID 24999044 Tsekouras G Mozer A J Wallace G G 2008 Enhanced performance of dye sensitized solar cells utilizing platinum electrodeposit counter electrodes Journal of the Electrochemical Society 155 7 K124 K128 Bibcode 2008JElS 155K 124T doi 10 1149 1 2919107 Besenius P Portale G Bomans P H H Janssen H M Palmans A R A Meijer E W 2010 Controlling the growth and shape of chiral supramolecular polymers in water Proceedings of the National Academy of Sciences of the United States of America 107 42 17888 17893 Bibcode 2010PNAS 10717888B doi 10 1073 pnas 1009592107 PMC 2964246 PMID 20921365 Furuya K 2008 Nanofabrication by advanced electron microscopy using intense and focused beam Science and Technology of Advanced Materials 9 1 Article 014110 Bibcode 2008STAdM 9a4110F doi 10 1088 1468 6996 9 1 014110 PMC 5099805 PMID 27877936 Kosasih Felix Utama Ducati Caterina May 2018 Characterising degradation of perovskite solar cells through in situ and operando electron microscopy Nano Energy 47 243 256 doi 10 1016 j nanoen 2018 02 055 Maloy Stuart A Sommer Walter F James Michael R Romero Tobias J Lopez Manuel R Zimmermann Eugene Ledbetter James M 13 May 2017 The Accelerator Production of Tritium Materials Test Program Nuclear Technology 132 1 103 114 doi 10 13182 nt00 a3132 S2CID 94639273 Ukraintsev Vladimir 27 February 2012 Review of reference metrology for nanotechnology significance challenges and solutions Journal of Micro Nanolithography MEMS and MOEMS 11 1 011010 doi 10 1117 1 JMM 11 1 011010 Wilhelmi O Roussel L Faber P Reyntjens S Daniel G June 2010 Focussed ion beam fabrication of large and complex nanopatterns Journal of Experimental Nanoscience 5 3 244 250 Bibcode 2010JENan 5 244W doi 10 1080 17458080903487448 S2CID 283449 Vogt E T C Whiting G T Chowdhury A D Weckhuysen B M 2015 Zeolites and zeotypes for oil and gas conversion Advances in Catalysis Vol 58 pp 143 314 doi 10 1016 bs acat 2015 10 001 ISBN 9780128021262 Lai Shih En Hong Ying Jhan Chen Yu Ting Kang Yu Ting Chang Pin Yew Tri Rung 18 September 2015 Direct Writing of Cu Nano Patterns with an Electron Beam Microscopy and Microanalysis 21 6 1639 43 Bibcode 2015MiMic 21 1639L doi 10 1017 S1431927615015111 PMID 26381450 S2CID 30803055 Sicignano Angelo Di Monaco Rossella Masi Paolo Cavella Silvana October 2015 From raw material to dish pasta quality step by step Journal of the Science of Food and Agriculture 95 13 2579 87 doi 10 1002 jsfa 7176 PMID 25783568 Brozek Mucha Zuzanna 2014 Scanning Electron Microscopy and X Ray Microanalysis for Chemical and Morphological Characterisation of the Inorganic Component of Gunshot Residue Selected Problems BioMed Research International 2014 428038 doi 10 1155 2014 428038 PMC 4082842 PMID 25025050 Carbonell Verdu A Garcia Sanoguera D Jorda Vilaplana A Sanchez Nacher L Balart R 2016 A new biobased plasticizer for poly vinyl chloride based on epoxidized cottonseed oil Journal of Applied Polymer Science 33 27 43642 doi 10 1002 app 43642 hdl 10251 82834 Ding Jie Zhang Zhiming 1 May 2015 Micro characterization of dissimilar metal weld joint for connecting pipe nozzle to safe end in generation III nuclear power plant Acta Metall Sin 51 4 425 39 doi 10 11900 0412 1961 2014 00299 inactive 31 December 2022 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint DOI inactive as of December 2022 link Tsikouras Basilios Pe Piper Georgia Piper David J W Schaffer Michael June 2011 Varietal heavy mineral analysis of sediment provenance Lower Cretaceous Scotian Basin eastern Canada Sedimentary Geology 237 3 4 150 165 Bibcode 2011SedG 237 150T doi 10 1016 j sedgeo 2011 02 011 Li Xiang Jiang Chuan Pan Lili Zhang Haoyang Hu Lang Li Tianxue Yang Xinghao 15 July 2014 Effects of preparing techniques and aging on dissolution behavior of the solid dispersions of NF Soluplus Kollidon SR identification and classification by a combined analysis by FT IR spectroscopy and computational approaches Drug Development and Industrial Pharmacy 41 1 2 14 doi 10 3109 03639045 2014 938080 PMID 25026247 S2CID 32460608 External links Edit Wikimedia Commons has media related to Electron microscopes An Introduction to Electron Microscopy Archived 2013 07 19 at the Wayback Machine resources for teachers and students Cell Centered Database Electron microscopy data Science Aid Electron Microscopy high school GCSE A Level resourceGeneral Edit Animations and explanations of various types of microscopy including electron microscopy Universite Paris Sud Environmental Scanning Electron Microscopy ESEM ETH Zurich website graphics and images illustrating various procedures Eva Nogales s Seminar Introduction to Electron Microscopy FEI Image Contest FEI has had a microscopy image contest every year since 2008 Introduction to electron microscopy by David Szondy Nanohedron com image gallery images generated by electron microscopy X ray element analysis in electron microscopy information portal with X ray microanalysis and EDX contentsHistory Edit John H L Watson s recollections at the University of Toronto when he worked with Hillier and Prebus Archived 2019 03 30 at the Wayback Machine Rubin Borasky Electron Microscopy Collection 1930 1988 Archives Center National Museum of American History Smithsonian Institution Other Edit The Royal Microscopical Society Electron Microscopy Section UK Albert Lleal Natural history subjects at Scanning Electron Microscope SEM Archived 2019 03 30 at the Wayback Machine EM Images Retrieved from https en wikipedia org w index php title Electron microscope amp oldid 1135393329, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.