An ion source is a device that creates atomic and molecular ions.[1] Ion sources are used to form ions for mass spectrometers, optical emission spectrometers, particle accelerators, ion implanters and ion engines.

Mass spectrometer EI/CI ion source

Electron ionization Edit

 
Electron ionization source schematic

Electron ionization is widely used in mass spectrometry, particularly for organic molecules. The gas phase reaction producing electron ionization is

 

where M is the atom or molecule being ionized,   is the electron, and   is the resulting ion.

The electrons may be created by an arc discharge between a cathode and an anode.

An electron beam ion source (EBIS) is used in atomic physics to produce highly charged ions by bombarding atoms with a powerful electron beam.[2][3] Its principle of operation is shared by the electron beam ion trap.

Electron capture ionization Edit

Electron capture ionization (ECI) is the ionization of a gas phase atom or molecule by attachment of an electron to create an ion of the form A−•. The reaction is

 

where the M over the arrow denotes that to conserve energy and momentum a third body is required (the molecularity of the reaction is three).

Electron capture can be used in conjunction with chemical ionization.[4]

An electron capture detector is used in some gas chromatography systems.[5]

Chemical ionization Edit

Chemical ionization (CI) is a lower energy process than electron ionization because it involves ion/molecule reactions rather than electron removal.[6] The lower energy yields less fragmentation, and usually a simpler spectrum. A typical CI spectrum has an easily identifiable molecular ion.[7]

In a CI experiment, ions are produced through the collision of the analyte with ions of a reagent gas in the ion source. Some common reagent gases include: methane, ammonia, and isobutane. Inside the ion source, the reagent gas is present in large excess compared to the analyte. Electrons entering the source will preferentially ionize the reagent gas. The resultant collisions with other reagent gas molecules will create an ionization plasma. Positive and negative ions of the analyte are formed by reactions with this plasma. For example, protonation occurs by

  (primary ion formation),
  (reagent ion formation),
  (product ion formation, e.g. protonation).

Charge exchange ionization Edit

Charge-exchange ionization (also known as charge-transfer ionization) is a gas phase reaction between an ion and an atom or molecule in which the charge of the ion is transferred to the neutral species.[8]

 

Chemi-ionization Edit

Chemi-ionization is the formation of an ion through the reaction of a gas phase atom or molecule with an atom or molecule in an excited state.[9][10] Chemi-ionization can be represented by

 

where G is the excited state species (indicated by the superscripted asterisk), and M is the species that is ionized by the loss of an electron to form the radical cation (indicated by the superscripted "plus-dot").

Associative ionization Edit

Associative ionization is a gas phase reaction in which two atoms or molecules interact to form a single product ion.[11][12][13] One or both of the interacting species may have excess internal energy.

For example,

 

where species A with excess internal energy (indicated by the asterisk) interacts with B to form the ion AB+.

Penning ionization Edit

Penning ionization is a form of chemi-ionization involving reactions between neutral atoms or molecules.[14][15] The process is named after the Dutch physicist Frans Michel Penning who first reported it in 1927.[16] Penning ionization involves a reaction between a gas-phase excited-state atom or molecule G* and a target molecule M resulting in the formation of a radical molecular cation M+., an electron e, and a neutral gas molecule G:[17]

 

Penning ionization occurs when the target molecule has an ionization potential lower than the internal energy of the excited-state atom or molecule.

Associative Penning ionization can proceed via

 

Surface Penning ionization (also known as Auger deexcitation) refers to the interaction of the excited-state gas with a bulk surface S, resulting in the release of an electron according to

 .

Ion attachment Edit

Ion-attachment ionization is similar to chemical ionization in which a cation is attached to the analyte molecule in a reactive collision:

 

Where M is the analyte molecule, X+ is the cation and A is a non-reacting collision partner.[18]

In a radioactive ion source, a small piece of radioactive material, for instance 63Ni or 241Am, is used to ionize a gas.[citation needed] This is used in ionization smoke detectors and ion mobility spectrometers.

Gas-discharge ion sources Edit

 
NASA's NEXT (ion thruster) space craft propulsion system

These ion sources use a plasma source or electric discharge to create ions.

Inductively-coupled plasma Edit

Ions can be created in an inductively coupled plasma, which is a plasma source in which the energy is supplied by electrical currents which are produced by electromagnetic induction, that is, by time-varying magnetic fields.[19]

Microwave-induced plasma Edit

Microwave induced plasma ion sources are capable of exciting electrodeless gas discharges to create ions for trace element mass spectrometry.[20][21] A microwave plasma is a type of plasma, that has high frequency electromagnetic radiation in the GHz range. It is capable of exciting electrodeless gas discharges. If applied in surface-wave-sustained mode, they are especially well suited to generate large-area plasmas of high plasma density. If they are both in surface-wave and resonator mode, they can exhibit a high degree of spatial localization. This allows to spatially separate the location of plasma generations from the location of surface processing. Such a separation (together with an appropriate gas-flow scheme) may help reduce the negative effect, that particles released from a processed substrate may have on the plasma chemistry of the gas phase.

ECR ion source Edit

The ECR ion source makes use of the electron cyclotron resonance to ionize a plasma. Microwaves are injected into a volume at the frequency corresponding to the electron cyclotron resonance, defined by the magnetic field applied to a region inside the volume. The volume contains a low pressure gas.

Glow discharge Edit

Ions can be created in an electric glow discharge. A glow discharge is a plasma formed by the passage of electric current through a low-pressure gas. It is created by applying a voltage between two metal electrodes in an evacuated chamber containing gas. When the voltage exceeds a certain value, called the striking voltage, the gas forms a plasma.

A duoplasmatron is a type of glow discharge ion source that consists of a cathode (hot cathode or cold cathode) that produces a plasma that is used to ionize a gas.[1][22] Duoplasmatrons can produce positive or negative ions.[23] Duoplasmatrons are used for secondary ion mass spectrometry.,[24][25] ion beam etching, and high-energy physics.[26]

Flowing afterglow Edit

In a flowing afterglow, ions are formed in a flow of inert gas, typically helium or argon.[27][28][29] Reagents are added downstream to create ion products and study reaction rates. Flowing-afterglow mass spectrometry is used for trace gas analysis [30] for organic compounds.[31]

Spark ionization Edit

Electric spark ionization is used to produce gas phase ions from a solid sample. When incorporated with a mass spectrometer the complete instrument is referred to as a spark ionization mass spectrometer or as a spark source mass spectrometer (SSMS).[32]

A closed drift ion source uses a radial magnetic field in an annular cavity in order to confine electrons for ionizing a gas. They are used for ion implantation and for space propulsion (Hall-effect thrusters).

Photoionization Edit

Photoionization is the ionization process in which an ion is formed from the interaction of a photon with an atom or molecule.[33]

Multi-photon ionization Edit

In multi-photon ionization (MPI), several photons of energy below the ionization threshold may actually combine their energies to ionize an atom.

Resonance-enhanced multiphoton ionization (REMPI) is a form of MPI in which one or more of the photons accesses a bound-bound transition that is resonant in the atom or molecule being ionized.

Atmospheric pressure photoionization Edit

Atmospheric pressure photoionization (APPI) uses a source of photons, usually a vacuum UV (VUV) lamp, to ionize the analyte with single photon ionization process. Analogous to other atmospheric pressure ion sources, a spray of solvent is heated to relatively high temperatures (above 400 degrees Celsius) and sprayed with high flow rates of nitrogen for desolvation. The resulting aerosol is subjected to UV radiation to create ions. Atmospheric-pressure laser ionization uses UV laser light sources to ionize the analyte via MPI.

Desorption ionization Edit

Field desorption Edit

 
Field desorption schematic

Field desorption refers to an ion source in which a high-potential electric field is applied to an emitter with a sharp surface, such as a razor blade, or more commonly, a filament from which tiny "whiskers" have formed.[34] This results in a very high electric field which can result in ionization of gaseous molecules of the analyte. Mass spectra produced by FI have little or no fragmentation. They are dominated by molecular radical cations   and less often, protonated molecules  .

Particle bombardment Edit

Fast atom bombardment Edit

Particle bombardment with atoms is called fast atom bombardment (FAB) and bombardment with atomic or molecular ions is called secondary ion mass spectrometry (SIMS).[35] Fission fragment ionization uses ionic or neutral atoms formed as a result of the nuclear fission of a suitable nuclide, for example the Californium isotope 252Cf.

In FAB the analytes is mixed with a non-volatile chemical protection environment called a matrix and is bombarded under vacuum with a high energy (4000 to 10,000 electron volts) beam of atoms.[36] The atoms are typically from an inert gas such as argon or xenon. Common matrices include glycerol, thioglycerol, 3-nitrobenzyl alcohol (3-NBA), 18-crown-6 ether, 2-nitrophenyloctyl ether, sulfolane, diethanolamine, and triethanolamine. This technique is similar to secondary ion mass spectrometry and plasma desorption mass spectrometry.

Secondary ionization Edit

Secondary ion mass spectrometry (SIMS) is used to analyze the composition of solid surfaces and thin films by sputtering the surface of the specimen with a focused primary ion beam and collecting and analyzing ejected secondary ions. The mass/charge ratios of these secondary ions are measured with a mass spectrometer to determine the elemental, isotopic, or molecular composition of the surface to a depth of 1 to 2 nm.

In a liquid metal ion source (LMIS), a metal (typically gallium) is heated to the liquid state and provided at the end of a capillary or a needle. Then a Taylor cone is formed under the application of a strong electric field. As the cone's tip get sharper, the electric field becomes stronger, until ions are produced by field evaporation. These ion sources are particularly used in ion implantation or in focused ion beam instruments.

Plasma desorption ionization Edit

 
Schematic representation of a plasama desorption time-of-flight mass spectrometer

Plasma desorption ionization mass spectrometry (PDMS), also called fission fragment ionization, is a mass spectrometry technique in which ionization of material in a solid sample is accomplished by bombarding it with ionic or neutral atoms formed as a result of the nuclear fission of a suitable nuclide, typically the californium isotope 252Cf.[37][38]

Laser desorption ionization Edit

 
Diagram of a MALDI ion source

Matrix-assisted laser desorption/ionization (MALDI) is a soft ionization technique. The sample is mixed with a matrix material. Upon receiving a laser pulse, the matrix absorbs the laser energy and it is thought that primarily the matrix is desorbed and ionized (by addition of a proton) by this event. The analyte molecules are also desorbed. The matrix is then thought to transfer proton to the analyte molecules (e.g., protein molecules), thus charging the analyte.

Surface-assisted laser desorption/ionization Edit

Surface-assisted laser desorption/ionization (SALDI) is a soft laser desorption technique used for analyzing biomolecules by mass spectrometry.[39][40] In its first embodiment, it used graphite matrix.[39] At present, laser desorption/ionization methods using other inorganic matrices, such as nanomaterials, are often regarded as SALDI variants. A related method named "ambient SALDI" - which is a combination of conventional SALDI with ambient mass spectrometry incorporating the DART ion source - has also been demonstrated.[41]

Surface-enhanced laser desorption/ionization Edit

Surface-enhanced laser desorption/ionization (SELDI) is a variant of MALDI that is used for the analysis of protein mixtures that uses a target modified to achieve biochemical affinity with the analyte compound.[42]

Desorption ionization on silicon Edit

Desorption ionization on silicon (DIOS) refers to laser desorption/ionization of a sample deposited on a porous silicon surface.[43]

Smalley source Edit

A laser vaporization cluster source produces ions using a combination of laser desorption ionization and supersonic expansion.[44] The Smalley source (or Smalley cluster source)[45] was developed by Richard Smalley at Rice University in the 1980s and was central to the discovery of fullerenes in 1985.[46][47]

Aerosol ionization Edit

In aerosol mass spectrometry with time-of-flight analysis, micrometer sized solid aerosol particles extracted from the atmosphere are simultaneously desorbed and ionized by a precisely timed laser pulse as they pass through the center of a time-of-flight ion extractor.[48][49]

Spray ionization Edit

 
Atmospheric-pressure chemical ionization source

Spray ionization methods involve the formation of aerosol particles from a liquid solution and the formation of bare ions after solvent evaporation.[50]

Solvent-assisted ionization (SAI) is a method in which charged droplets are produced by introducing a solution containing analyte into a heated inlet tube of an atmospheric pressure ionization mass spectrometer. Just as in Electrospray Ionization (ESI), desolvation of the charged droplets produces multiply charged analyte ions. Volatile and nonvolatile compounds are analyzed by SAI, and high voltage is not required to achieve sensitivity comparable to ESI.[51] Application of a voltage to the solution entering the hot inlet through a zero dead volume fitting connected to fused silica tubing produces ESI-like mass spectra, but with higher sensitivity.[52] The inlet tube to the mass spectrometer becomes the ion source.

Matrix-Assisted Ionization Edit

Matrix-Assisted Ionization [MAI] is similar to MALDI in sample preparation, but a laser is not required to convert analyte molecules included in a matrix compound into gas-phase ions. In MAI, analyte ions have charge states similar to electrospray ionization but obtained from a solid matrix rather than a solvent. No voltage or laser is required, but a laser can be used to obtain spatial resolution for imaging. Matrix-analyte samples are ionized in the vacuum of a mass spectrometer and can be inserted into the vacuum through an atmospheric pressure inlet. Less volatile matrices such as 2,5-dihydroxybenzoic acid require a hot inlet tube to produce analyte ions by MAI, but more volatile matrices such as 3-nitrobenzonitrile require no heat, voltage, or laser. Simply introducing the matrix:analyte sample to the inlet aperture of an atmospheric pressure ionization mass spectrometer produces abundant ions. Compounds at least as large as bovine serum albumin [66 kDa] can be ionized with this method.[53] In this simple, low cost and easy to use ionization method, the inlet to the mass spectrometer can be considered the ion source.

Atmospheric-pressure chemical ionization Edit

Atmospheric-pressure chemical ionization is a form of chemical ionization using a solvent spray at atmospheric pressure.[54] A spray of solvent is heated to relatively high temperatures (above 400 degrees Celsius), sprayed with high flow rates of nitrogen and the entire aerosol cloud is subjected to a corona discharge that creates ions with the evaporated solvent acting as the chemical ionization reagent gas. APCI is not as "soft" (low fragmentation) an ionization technique as ESI.[55] Note that atmospheric pressure ionization (API) should not be used as a synonym for APCI.[56]

Thermospray ionization Edit

Thermospray ionization is a form of atmospheric pressure ionization in mass spectrometry. It transfers ions from the liquid phase to the gas phase for analysis. It is particularly useful in liquid chromatography-mass spectrometry.[57]

 
Electrospray ion source

Electrospray ionization Edit

In electrospray ionization, a liquid is pushed through a very small, charged and usually metal, capillary.[58] This liquid contains the substance to be studied, the analyte, dissolved in a large amount of solvent, which is usually much more volatile than the analyte. Volatile acids, bases or buffers are often added to this solution too. The analyte exists as an ion in solution either in its anion or cation form. Because like charges repel, the liquid pushes itself out of the capillary and forms an aerosol, a mist of small droplets about 10 μm across. The aerosol is at least partially produced by a process involving the formation of a Taylor cone and a jet from the tip of this cone. An uncharged carrier gas such as nitrogen is sometimes used to help nebulize the liquid and to help evaporate the neutral solvent in the droplets. As the solvent evaporates, the analyte molecules are forced closer together, repel each other and break up the droplets. This process is called Coulombic fission because it is driven by repulsive Coulombic forces between charged molecules. The process repeats until the analyte is free of solvent and is a bare ion. The ions observed are created by the addition of a proton (a hydrogen ion) and denoted  , or of another cation such as sodium ion,  , or the removal of a proton,  . Multiply charged ions such as   are often observed. For large macromolecules, there can be many charge states, occurring with different frequencies; the charge can be as great as  , for example.

Probe electrospray ionization Edit

Probe electrospray ionization (PESI) is a modified version of electrospray, where the capillary for sample solution transferring is replaced by a sharp-tipped solid needle with periodical motion.[59]

Contactless atmospheric pressure ionization Edit

Contactless atmospheric pressure ionization is a technique used for analysis of liquid and solid samples by mass spectrometry.[60] Contactless API can be operated without an additional electric power supply (supplying voltage to the source emitter), gas supply, or syringe pump. Thus, the technique provides a facile means for analyzing chemical compounds by mass spectrometry at atmospheric pressure.

Sonic spray ionization Edit

Sonic spray ionization is method for creating ions from a liquid solution, for example, a mixture of methanol and water.[61] A pneumatic nebulizer is used to turn the solution into a supersonic spray of small droplets. Ions are formed when the solvent evaporates and the statistically unbalanced charge distribution on the droplets leads to a net charge and complete desolvation results in the formation of ions. Sonic spray ionization is used to analyze small organic molecules and drugs and can analyze large molecules when an electric field is applied to the capillary to help increase the charge density and generate multiple charged ions of proteins.[62]

Sonic spray ionization has been coupled with high performance liquid chromatography for the analysis of drugs.[63][64] Oligonucleotides have been studied with this method.[65][66] SSI has been used in a manner similar to desorption electrospray ionization[67] for ambient ionization and has been coupled with thin-layer chromatography in this manner.[68]

Ultrasonication-assisted spray ionization Edit

Ultrasonication-assisted spray ionization (UASI) involves ionization through the application of ultrasound.[69][70]

Thermal ionization Edit

Thermal ionization (also known as surface ionization, or contact ionization) involves spraying vaporized, neutral atoms onto a hot surface, from which the atoms re-evaporate in ionic form. To generate positive ions, the atomic species should have a low ionization energy, and the surface should have a high work function. This technique is most suitable for alkali atoms (Li, Na, K, Rb, Cs) which have low ionization energies and are easily evaporated.[71]

To generate negative ions, the atomic species should have a high electron affinity, and the surface should have a low work function. This second approach is most suited for halogen atoms Cl, Br, I, At.[72]

Ambient ionization Edit

 
Direct analysis in real time ambient ionization ion source

In ambient ionization, ions are formed outside the mass spectrometer without sample preparation or separation.[73][74][75] Ions can be formed by extraction into charged electrospray droplets, thermally desorbed and ionized by chemical ionization, or laser desorbed or ablated and post-ionized before they enter the mass spectrometer.

Solid-liquid extraction based ambient ionization uses a charged spray to create a liquid film on the sample surface.[74][76] Molecules on the surface are extracted into the solvent. The action of the primary droplets hitting the surface produces secondary droplets that are the source of ions for the mass spectrometer. Desorption electrospray ionization (DESI) uses an electrospray source to create charged droplets that are directed at a solid sample a few millimeters to a few centimeters away. The charged droplets pick up the sample through interaction with the surface and then form highly charged ions that can be sampled into a mass spectrometer.[77]

Plasma-based ambient ionization is based on an electrical discharge in a flowing gas that produces metastable atoms and molecules and reactive ions. Heat is often used to assist in the desorption of volatile species from the sample. Ions are formed by chemical ionization in the gas phase. A direct analysis in real time source operates by exposing the sample to a dry gas stream (typically helium or nitrogen) that contains long-lived electronically or vibronically excited neutral atoms or molecules (or "metastables"). Excited states are typically formed in the DART source by creating a glow discharge in a chamber through which the gas flows. A similar method called atmospheric solids analysis probe [ASAP] uses the heated gas from ESI or APCI probes to vaporize sample placed on a melting point tube inserted into an ESI/APCI source.[78] Ionization is by APCI.

Laser-based ambient ionization is a two-step process in which a pulsed laser is used to desorb or ablate material from a sample and the plume of material interacts with an electrospray or plasma to create ions. Electrospray-assisted laser desorption/ionization (ELDI) uses a 337 nm UV laser[79] or 3 μm infrared laser[80] to desorb material into an electrospray source. Matrix-assisted laser desorption electrospray ionization (MALDESI)[81] is an atmospheric pressure ionization source for generation of multiply charged ions. An ultraviolet or infrared laser is directed onto a solid or liquid sample containing the analyte of interest and matrix desorbing neutral analyte molecules that are ionized by interaction with electrosprayed solvent droplets generating multiply charged ions. Laser ablation electrospray ionization (LAESI) is an ambient ionization method for mass spectrometry that combines laser ablation from a mid-infrared (mid-IR) laser with a secondary electrospray ionization (ESI) process.

Applications Edit

Mass spectrometry Edit

In a mass spectrometer a sample is ionized in an ion source and the resulting ions are separated by their mass-to-charge ratio. The ions are detected and the results are displayed as spectra of the relative abundance of detected ions as a function of the mass-to-charge ratio. The atoms or molecules in the sample can be identified by correlating known masses to the identified masses or through a characteristic fragmentation pattern.

Particle accelerators Edit

 
Surface ionization source at the Argonne Tandem Linear Accelerator System (ATLAS)
 
Ion source used in the Cockcroft-Walton pre-accelerator at Fermilab[82]

In particle accelerators an ion source creates a particle beam at the beginning of the machine, the source. The technology to create ion sources for particle accelerators depends strongly on the type of particle that needs to be generated: electrons, protons, H ion or a Heavy ions.

Electrons are generated with an electron gun, of which there are many varieties.

Protons are generated with a plasma-based device, like a duoplasmatron or a magnetron.

H ions are generated with a magnetron or a Penning source. A magnetron consists of a central cylindrical cathode surrounded by an anode. The discharge voltage is typically greater than 150 V and the current drain is around 40 A. A magnetic field of about 0.2 tesla is parallel to the cathode axis. Hydrogen gas is introduced by a pulsed gas valve. Caesium is often used to lower the work function of the cathode, enhancing the amount of ions that are produced. Large caesiated H sources are also used for plasma heating in nuclear fusion devices.

For a Penning source, a strong magnetic field parallel to the electric field of the sheath guides electrons and ions on cyclotron spirals from cathode to cathode. Fast H-minus ions are generated at the cathodes as in the magnetron. They are slowed down due to the charge exchange reaction as they migrate to the plasma aperture. This makes for a beam of ions that is colder than the ions obtained from a magnetron.

Heavy ions can be generated with an electron cyclotron resonance ion source. The use of electron cyclotron resonance (ECR) ion sources for the production of intense beams of highly charged ions has immensely grown over the last decade. ECR ion sources are used as injectors into linear accelerators, Van-de-Graaff generators or cyclotrons in nuclear and elementary particle physics. In atomic and surface physics ECR ion sources deliver intense beams of highly charged ions for collision experiments or for the investigation of surfaces. For the highest charge states, however, Electron beam ion sources (EBIS) are needed. They can generate even bare ions of mid-heavy elements. The Electron beam ion trap (EBIT), based on the same principle, can produce up to bare uranium ions and can be used as an ion source as well.

Heavy ions can also be generated with an Ion Gun which typically uses the thermionic emission of electrons to ionize a substance in its gaseous state. Such instruments are typically used for surface analysis.

 
Ion beam deposition system with mass separator

Gas flows through the ion source between the anode and the cathode. A positive voltage is applied to the anode. This voltage, combined with the high magnetic field between the tips of the internal and external cathodes allow a plasma to start. Ions from the plasma are repelled by the anode electric field. This creates an ion beam.[83]

Surface modification Edit

See also Edit

References Edit

  1. ^ a b Bernhard Wolf (31 August 1995). Handbook of Ion Sources. CRC Press. ISBN 978-0-8493-2502-1.
  2. ^ Ian G. Brown (6 March 2006). The Physics and Technology of Ion Sources. John Wiley & Sons. ISBN 978-3-527-60454-8.
  3. ^ Heinrich Beyer; Heinrich F. Beyer; H.-Jürgen Kluge; H.-J. Kluge; Vi͡acheslav Petrovich Shevelʹko (14 August 1997). X-Ray Radiation of Highly Charged Ions. Springer Science & Business Media. ISBN 978-3-540-63185-9.
  4. ^ Donald F. Hunt; Frank W. Crow (1978), "Electron capture negative ion chemical ionization mass spectrometry", Analytical Chemistry, 50 (13): 1781–1784, doi:10.1021/ac50035a017
  5. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "electron capture detector (in gas chromatography)". doi:10.1351/goldbook.E01981
  6. ^ Munson, M. S. B.; Field, F. H. (1966). "Chemical Ionization Mass Spectrometry. I. General Introduction". Journal of the American Chemical Society. 88 (12): 2621–2630. doi:10.1021/ja00964a001.
  7. ^ de Hoffmann, Edmond; Vincent Stroobant (2003). Mass Spectrometry: Principles and Applications (Second ed.). Toronto: John Wiley & Sons, Ltd. p. 14. ISBN 978-0-471-48566-7.
  8. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "charge-exchange ionization". doi:10.1351/goldbook.C00989
  9. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "chemi-ionization". doi:10.1351/goldbook.C01044 C01044
  10. ^ Klucharev, A. N. (1993), "Chemi-ionization processes", Physics-Uspekhi, 36 (6): 486–512, Bibcode:1993PhyU...36..486K, doi:10.1070/PU1993v036n06ABEH002162
  11. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "associative ionization". doi:10.1351/goldbook.A00475
  12. ^ *Jones DM, Dahler JS (April 1988). "Theory of associative ionization". Physical Review A. 37 (8): 2916–2933. Bibcode:1988PhRvA..37.2916J. doi:10.1103/PhysRevA.37.2916. PMID 9900022.
  13. ^ Cohen, James S. (1976). "Multistate curve-crossing model for scattering: Associative ionization and excitation transfer in helium". Physical Review A. 13 (1): 99–114. Bibcode:1976PhRvA..13...99C. doi:10.1103/PhysRevA.13.99.
  14. ^ Arango CA, Shapiro M, Brumer P (2006). "Cold atomic collisions: coherent control of penning and associative ionization". Phys. Rev. Lett. 97 (19): 193202. arXiv:physics/0610131. Bibcode:2006PhRvL..97s3202A. doi:10.1103/PhysRevLett.97.193202. PMID 17155624. S2CID 1480148.
  15. ^ Hiraoka K, Furuya H, Kambara S, Suzuki S, Hashimoto Y, Takamizawa A (2006). "Atmospheric-pressure Penning ionization of aliphatic hydrocarbons". Rapid Commun. Mass Spectrom. 20 (21): 3213–22. Bibcode:2006RCMS...20.3213H. doi:10.1002/rcm.2706. PMID 17016831.
  16. ^ Penning, F. M. Die Naturwissenschaften, 1927, 15, 818. Über Ionisation durch metastabile Atome.
  17. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "Penning gas mixture". doi:10.1351/goldbook.P04476
  18. ^ Selvin, P. Christopher; Fujii, Toshihiro (2001). "Lithium ion attachment mass spectrometry: Instrumentation and features". Review of Scientific Instruments. 72 (5): 2248. Bibcode:2001RScI...72.2248S. doi:10.1063/1.1362439.
  19. ^ A. Montaser and D. W. Golightly, eds. Inductively Coupled Plasmas in Analytical Atomic Spectrometry, VCH Publishers, Inc., New York, 1992.
  20. ^ Okamoto, Yukio (1994). "High-sensitivity microwave-induced plasma mass spectrometry for trace element analysis". Journal of Analytical Atomic Spectrometry. 9 (7): 745. doi:10.1039/ja9940900745. ISSN 0267-9477.
  21. ^ Douglas, D. J.; French, J. B. (1981). "Elemental analysis with a microwave-induced plasma/quadrupole mass spectrometer system". Analytical Chemistry. 53 (1): 37–41. doi:10.1021/ac00224a011. ISSN 0003-2700.
  22. ^ Lejeune, C. (1974). "Theoretical and experimental study of the duoplasmatron ion source". Nuclear Instruments and Methods. 116 (3): 417–428. Bibcode:1974NucIM.116..417L. doi:10.1016/0029-554X(74)90821-0. ISSN 0029-554X.
  23. ^ Aberth, William; Peterson, James R. (1967). "Characteristics of a Low Energy Duoplasmatron Negative Ion Source". Review of Scientific Instruments. 38 (6): 745. Bibcode:1967RScI...38..745A. doi:10.1063/1.1720882. ISSN 0034-6748.
  24. ^ Coath, C. D.; Long, J. V. P. (1995). "A high-brightness duoplasmatron ion source for microprobe secondary-ion mass spectrometry". Review of Scientific Instruments. 66 (2): 1018. Bibcode:1995RScI...66.1018C. doi:10.1063/1.1146038. ISSN 0034-6748.
  25. ^ Christine M. Mahoney (9 April 2013). Cluster Secondary Ion Mass Spectrometry: Principles and Applications. John Wiley & Sons. pp. 65–. ISBN 978-1-118-58925-0.
  26. ^ Stanley Humphries (25 July 2013). Charged Particle Beams. Dover Publications. pp. 309–. ISBN 978-0-486-31585-0.
  27. ^ Ferguson, E. E.; Fehsenfeld, F. C.; Schmeltekopf, A. L. (1969). "Ion-Molecule Reaction Rates Measured in a Discharge Afterglow". Chemical Reactions in Electrical Discharges. Advances in Chemistry. Vol. 80. pp. 83–91. doi:10.1021/ba-1969-0080.ch006. ISBN 978-0-8412-0081-4. ISSN 0065-2393.
  28. ^ Ferguson, Eldon E. (1992). "A Personal history of the early development of the flowing afterglow technique for ion-molecule reaction studies". Journal of the American Society for Mass Spectrometry (Submitted manuscript). 3 (5): 479–486. doi:10.1016/1044-0305(92)85024-E. ISSN 1044-0305. PMID 24234490.
  29. ^ Bierbaum, Veronica M. (2014). "Go with the flow: Fifty years of innovation and ion chemistry using the flowing afterglow". International Journal of Mass Spectrometry. 377: 456–466. Bibcode:2015IJMSp.377..456B. doi:10.1016/j.ijms.2014.07.021. ISSN 1387-3806.
  30. ^ Smith, David; Španěl, Patrik (2005). "Selected ion flow tube mass spectrometry (SIFT-MS) for on-line trace gas analysis". Mass Spectrometry Reviews. 24 (5): 661–700. Bibcode:2005MSRv...24..661S. doi:10.1002/mas.20033. ISSN 0277-7037. PMID 15495143.
  31. ^ Dhooghe, Frederik; Vansintjan, Robbe; Schoon, Niels; Amelynck, Crist (2012-08-30). "Studies in search of selective detection of isomeric biogenic hexen-1-ols and hexanal by flowing afterglow tandem mass spectrometry using [H3O]+ and [NO]+ reagent ions". Rapid Communications in Mass Spectrometry. 26 (16): 1868–1874. doi:10.1002/rcm.6294. ISSN 1097-0231. PMID 22777789.
  32. ^ H. E. Beske; A. Hurrle; K. P. Jochum (1981). "Part I. Principles of spark source mass spectrometry (SSMS)". Fresenius' Journal of Analytical Chemistry. 309 (4): 258–261. doi:10.1007/BF00488596. S2CID 92433014.
  33. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "photoionization". doi:10.1351/goldbook.P04620
  34. ^ Beckey, H.D. (1969). "Field ionization mass spectrometry". Research/Development. 20 (11): 26.
  35. ^ Williams, Dudley H.; Findeis, A. Frederick; Naylor, Stephen; Gibson, Bradford W. (1987). "Aspects of the production of FAB and SIMS mass spectra". Journal of the American Chemical Society. 109 (7): 1980–1986. doi:10.1021/ja00241a013. ISSN 0002-7863.
  36. ^ Morris HR, Panico M, Barber M, Bordoli RS, Sedgwick RD, Tyler A (1981). "Fast atom bombardment: a new mass spectrometric method for peptide sequence analysis". Biochem. Biophys. Res. Commun. 101 (2): 623–31. doi:10.1016/0006-291X(81)91304-8. PMID 7306100.
  37. ^ Macfarlane, R.; Torgerson, D. (1976). "Californium-252 plasma desorption mass spectroscopy". Science. 191 (4230): 920–925. Bibcode:1976Sci...191..920M. doi:10.1126/science.1251202. ISSN 0036-8075. PMID 1251202.
  38. ^ Hilf, E.R. (1993). "Approaches to plasma desorption mass spectrometry by some theoretical physics concepts". International Journal of Mass Spectrometry and Ion Processes. 126: 25–36. Bibcode:1993IJMSI.126...25H. doi:10.1016/0168-1176(93)80067-O. ISSN 0168-1176.
  39. ^ a b Sunner, Jan.; Dratz, Edward.; Chen, Yu-Chie. (1995). "Graphite surface-assisted laser desorption/ionization time-of-flight mass spectrometry of peptides and proteins from liquid solutions". Analytical Chemistry. 67 (23): 4335–4342. doi:10.1021/ac00119a021. ISSN 0003-2700. PMID 8633776.
  40. ^ Dattelbaum, Andrew M; Iyer, Srinivas (2006). "Surface-assisted laser desorption/ionization mass spectrometry". Expert Review of Proteomics (Submitted manuscript). 3 (1): 153–161. doi:10.1586/14789450.3.1.153. ISSN 1478-9450. PMID 16445359. S2CID 39538990.
  41. ^ Zhang, Jialing; Li, Ze; Zhang, Chengsen; Feng, Baosheng; Zhou, Zhigui; Bai, Yu; Liu, Huwei (2012). "Graphite-Coated Paper as Substrate for High Sensitivity Analysis in Ambient Surface-Assisted Laser Desorption/Ionization Mass Spectrometry". Analytical Chemistry. 84 (7): 3296–3301. doi:10.1021/ac300002g. ISSN 0003-2700. PMID 22380704.
  42. ^ Tang N, Tornatore P, Weinberger SR (2004). "Current developments in SELDI affinity technology". Mass Spectrometry Reviews. 23 (1): 34–44. Bibcode:2004MSRv...23...34T. doi:10.1002/mas.10066. PMID 14625891.
  43. ^ Buriak, Jillian M.; Wei, Jing; Siuzdak, Gary (1999). "Desorption-ionization mass spectrometry on porous silicon". Nature. 399 (6733): 243–246. Bibcode:1999Natur.399..243W. doi:10.1038/20400. ISSN 0028-0836. PMID 10353246. S2CID 4314372.
  44. ^ Duncan, Michael A. (2012). "Invited Review Article: Laser vaporization cluster sources". Review of Scientific Instruments. 83 (4): 041101–041101–19. Bibcode:2012RScI...83d1101D. doi:10.1063/1.3697599. ISSN 0034-6748. PMID 22559508.
  45. ^ Laser Ablation and Desorption. Academic Press. 10 December 1997. pp. 628–. ISBN 978-0-08-086020-6.
  46. ^ Smalley, Richard (1997). "Discovering the fullerenes". Reviews of Modern Physics. 69 (3): 723–730. Bibcode:1997RvMP...69..723S. doi:10.1103/RevModPhys.69.723. ISSN 0034-6861.
  47. ^ Roy L. Johnston (25 April 2002). Atomic and Molecular Clusters. CRC Press. pp. 150–. ISBN 978-1-4200-5577-1.
  48. ^ Carson, P; Neubauer, K; Johnston, M; Wexler, A (1995). "On-line chemical analysis of aerosols by rapid single-particle mass spectrometry". Journal of Aerosol Science. 26 (4): 535–545. Bibcode:1995JAerS..26..535C. doi:10.1016/0021-8502(94)00133-J.
  49. ^ Guazzotti, S; Coffee, K; Prather, K (2000). "Real time monitoring of size-resolved single particle chemistry during INDOEX-IFP 99". Journal of Aerosol Science. 31: 182–183. Bibcode:2000JAerS..31..182G. doi:10.1016/S0021-8502(00)90189-7.
  50. ^ Chhabil Dass (11 May 2007). Fundamentals of Contemporary Mass Spectrometry. John Wiley & Sons. pp. 45–57. ISBN 978-0-470-11848-1.
  51. ^ Pagnotti VS, Chubatyi ND, McEwen CN (2011). "Solvent Assisted Inlet Ionization: an Ultrasensitive New Liquid Introduction Ionization Method for Mass Spectrometry". Anal. Chem. 83 (11): 3981–3985. doi:10.1021/ac200556z. PMID 21528896.
  52. ^ Pagnotti VS, Chakrabarty S, Harron AF, McEwen CN (2012). "Increasing the Sensitivity of Liquid Introduction Mass Spectrometry by Combining Electrospray Ionization and Solvent Assisted Inlet Ionization". Anal. Chem. 84 (15): 6828–6832. doi:10.1021/ac3014115. PMID 22742705.
  53. ^ Trimpin S, Wang B, Lietz CB, Marshall DD, Richards AL, Inutan ED. "New Ionization Processes and Applications for Use in Mass Spectrometry". Rev. Biochem. Mol. Biol. 2013. 5: 409–429.
  54. ^ Prakash C, Shaffer CL, Nedderman A (2007). "Analytical strategies for identifying drug metabolites". Mass Spectrometry Reviews. 26 (3): 340–69. Bibcode:2007MSRv...26..340P. doi:10.1002/mas.20128. PMID 17405144.
  55. ^ Zaikin VG, Halket JM (2006). "Derivatization in mass spectrometry--8. Soft ionization mass spectrometry of small molecules". European Journal of Mass Spectrometry. 12 (2): 79–115. doi:10.1255/ejms.798. PMID 16723751. S2CID 34838846.
  56. ^ "Atmospheric pressure ionization in mass spectrometry". IUPAC Compendium of Chemical Terminology. 2009. doi:10.1351/goldbook.A00492. ISBN 978-0-9678550-9-7.
  57. ^ Blakley, C. R.; Carmody, J. J.; Vestal, M. L. (1980). "Liquid Chromatograph-Mass Spectrometer for Analysis of Nonvolatile Samples". Analytical Chemistry. 1980 (52): 1636–1641. doi:10.1021/ac50061a025.
  58. ^ Fenn, J. B.; Mann, M.; Meng, C. K.; Wong, S. F.; Whitehouse, C. M. (1990). "Electrospray Ionization-Principles and Practice". Mass Spectrometry Reviews. 9 (1): 37–70. Bibcode:1990MSRv....9...37F. doi:10.1002/mas.1280090103.
  59. ^ Hiraoka K.; Nishidate K.; Mori K.; Asakawa D.; Suzuki S. (2007). "Development of probe electrospray using a solid needle". Rapid Communications in Mass Spectrometry. 21 (18): 3139–3144. Bibcode:2007RCMS...21.3139H. doi:10.1002/rcm.3201. PMID 17708527.
  60. ^ Hsieh, Cheng-Huan; Chang, Chia-Hsien; Urban, Pawel L.; Chen, Yu-Chie (2011). "Capillary Action-Supported Contactless Atmospheric Pressure Ionization for the Combined Sampling and Mass Spectrometric Analysis of Biomolecules". Analytical Chemistry. 83 (8): 2866–2869. doi:10.1021/ac200479s. ISSN 0003-2700. PMID 21446703.
  61. ^ Hirabayashi A, Sakairi M, Koizumi H (1995). "Sonic spray mass spectrometry". Anal. Chem. 67 (17): 2878–82. doi:10.1021/ac00113a023. PMID 8779414.
  62. ^ Chen, Tsung-Yi; Lin, Jia-Yi; Chen, Jen-Yi; Chen, Yu-Chie (2011-11-22). "Ultrasonication-assisted spray ionization mass spectrometry for the analysis of biomolecules in solution". Journal of the American Society for Mass Spectrometry. 21 (9): 1547–1553. doi:10.1016/j.jasms.2010.04.021. ISSN 1044-0305. PMID 20547459.
  63. ^ Arinobu T, Hattori H, Seno H, Ishii A, Suzuki O (2002). "Comparison of SSI with APCI as an interface of HPLC-mass spectrometry for analysis of a drug and its metabolites". J. Am. Soc. Mass Spectrom. 13 (3): 204–208. doi:10.1016/S1044-0305(01)00359-2. PMID 11908800.
  64. ^ Dams R, Benijts T, Günther W, Lambert W, De Leenheer A (2002). "Sonic spray ionization technology: performance study and application to a LC/MS analysis on a monolithic silica column for heroin impurity profiling". Anal. Chem. 74 (13): 3206–3212. doi:10.1021/ac0112824. PMID 12141684.
  65. ^ Huang M, Hirabayashi A, Okumura A, Hirabayashi Y (2001). "Matrix effect on the analysis of oligonucleotides by using a mass spectrometer with a sonic spray ionization source". Anal Sci. 17 (10): 1179–1182. doi:10.2116/analsci.17.1179. PMID 11990592.
  66. ^ Huang M, Hirabayashi A (2002). "Multi-charged oligonucleotide ion formation in sonic spray ionization". Anal Sci. 18 (4): 385–390. doi:10.2116/analsci.18.385. PMID 11999509.
  67. ^ Haddad R, Sparrapan R, Eberlin MN (2006). "Desorption sonic spray ionization for (high) voltage-free ambient mass spectrometry". Rapid Commun. Mass Spectrom. 20 (19): 2901–2905. Bibcode:2006RCMS...20.2901H. doi:10.1002/rcm.2680. PMID 16941547.
  68. ^ Haddad R, Milagre HM, Catharino RR, Eberlin MN (2008). "Easy Ambient Sonic-Spray Ionization Mass Spectrometry Combined with Thin-Layer Chromatography". Anal. Chem. 80 (8): 2744–2750. doi:10.1021/ac702216q. PMID 18331004.
  69. ^ Chen, Tsung-Yi; Lin, Jia-Yi; Chen, Jen-Yi; Chen, Yu-Chie (2010). "Ultrasonication-assisted spray ionization mass spectrometry for the analysis of biomolecules in solution". Journal of the American Society for Mass Spectrometry. 21 (9): 1547–1553. doi:10.1016/j.jasms.2010.04.021. PMID 20547459.
  70. ^ Chen, Tsung-Yi; Chao, Chin-Sheng; Mong, Kwok-Kong Tony; Chen, Yu-Chie (4 November 2010). "Ultrasonication-assisted spray ionization mass spectrometry for on-line monitoring of organic reactions". Chemical Communications. 46 (44): 8347–9. doi:10.1039/C0CC02629H. PMID 20957254. Retrieved 4 November 2011.
  71. ^ Alton, G. D. (1988). "Characterization of a cesium surface ionization source with a porous tungsten ionizer. I". Review of Scientific Instruments (Submitted manuscript). 59 (7): 1039. Bibcode:1988RScI...59.1039A. doi:10.1063/1.1139776. ISSN 0034-6748.
  72. ^ (PDF). Archived from the original (PDF) on 2004-12-18. Retrieved 2014-01-20.
  73. ^ Cooks, R. Graham; Ouyang, Zheng; Takats, Zoltan; Wiseman, Justin M. (2006). "Ambient Mass Spectrometry". Science. 311 (5767): 1566–70. Bibcode:2006Sci...311.1566C. doi:10.1126/science.1119426. PMID 16543450. S2CID 98131681.
  74. ^ a b Monge, María Eugenia; Harris, Glenn A.; Dwivedi, Prabha; Fernández, Facundo M. (2013). "Mass Spectrometry: Recent Advances in Direct Open Air Surface Sampling/Ionization". Chemical Reviews. 113 (4): 2269–2308. doi:10.1021/cr300309q. ISSN 0009-2665. PMID 23301684.
  75. ^ Huang, Min-Zong; Yuan, Cheng-Hui; Cheng, Sy-Chyi; Cho, Yi-Tzu; Shiea, Jentaie (2010). "Ambient Ionization Mass Spectrometry". Annual Review of Analytical Chemistry. 3 (1): 43–65. Bibcode:2010ARAC....3...43H. doi:10.1146/annurev.anchem.111808.073702. ISSN 1936-1327. PMID 20636033.
  76. ^ Badu-Tawiah, Abraham K.; Eberlin, Livia S.; Ouyang, Zheng; Cooks, R. Graham (2013). "Chemical Aspects of the Extractive Methods of Ambient Ionization Mass Spectrometry". Annual Review of Physical Chemistry. 64 (1): 481–505. Bibcode:2013ARPC...64..481B. doi:10.1146/annurev-physchem-040412-110026. ISSN 0066-426X. PMID 23331308.
  77. ^ Takáts Z, Wiseman JM, Cooks RG (2005). "Ambient mass spectrometry using desorption electrospray ionization (DESI): instrumentation, mechanisms and applications in forensics, chemistry, and biology". Journal of Mass Spectrometry. 40 (10): 1261–75. Bibcode:2005JMSp...40.1261T. doi:10.1002/jms.922. PMID 16237663.
  78. ^ McEwen CN, McKay RG, Larsen BS (2005). "Analysis of Solids, Liquids, and Biological Tissues Using Solids Probe Introduction at Atmospheric Pressure on Commercial LC/MS Instruments". Anal. Chem. 77 (23): 7826–7831. doi:10.1021/ac051470k. PMID 16316194.
  79. ^ Shiea J, Huang MZ, Hsu HJ, Lee CY, Yuan CH, Beech I, Sunner J (2005). "Electrospray-assisted laser desorption/ionization mass spectrometry for direct ambient analysis of solids". Rapid Commun. Mass Spectrom. 19 (24): 3701–4. Bibcode:2005RCMS...19.3701S. doi:10.1002/rcm.2243. PMID 16299699.
  80. ^ Peng, Ivory X.; Ogorzalek Loo, Rachel R.; Margalith, Eli; Little, Mark W.; Loo, Joseph A. (2010). "Electrospray-assisted laser desorption ionization mass spectrometry (ELDI-MS) with an infrared laser for characterizing peptides and proteins". The Analyst. 135 (4): 767–72. Bibcode:2010Ana...135..767P. doi:10.1039/b923303b. ISSN 0003-2654. PMC 3006438. PMID 20349541.
  81. ^ Sampson, JS; Hawkridge, AM; Muddiman, DC (2006). "Generation and detection of multiply charged peptides and proteins by matrix-assisted laser desorption electrospray ionization (MALDESI) Fourier transform ion cyclotron resonance mass spectrometry". J. Am. Soc. Mass Spectrom. 17 (12): 1712–6. doi:10.1016/j.jasms.2006.08.003. PMID 16952462.
  82. ^ 35 years of H- ions at Fermilab (PDF). Fermilab. p. 12. Retrieved 12 August 2015.
  83. ^ Cooks, R. G; Ouyang, Z; Takats, Z; Wiseman, J. M (2006). (PDF). Science. 311 (5767): 1566–70. Bibcode:2006Sci...311.1566C. doi:10.1126/science.1119426. PMID 16543450. S2CID 98131681. Archived from the original (PDF) on 2006-10-18. Retrieved 2006-12-14.
  84. ^ . Advanced Energy. Archived from the original on October 18, 2006. Retrieved 2006-12-14.

source, this, article, about, ionization, devices, physical, processes, ionization, source, device, that, creates, atomic, molecular, ions, used, form, ions, mass, spectrometers, optical, emission, spectrometers, particle, accelerators, implanters, engines, ma. This article is about ionization devices For the physical processes see Ionization An ion source is a device that creates atomic and molecular ions 1 Ion sources are used to form ions for mass spectrometers optical emission spectrometers particle accelerators ion implanters and ion engines Mass spectrometer EI CI ion source Contents 1 Electron ionization 1 1 Electron capture ionization 2 Chemical ionization 2 1 Charge exchange ionization 2 2 Chemi ionization 2 3 Associative ionization 2 4 Penning ionization 2 5 Ion attachment 3 Gas discharge ion sources 3 1 Inductively coupled plasma 3 2 Microwave induced plasma 3 3 ECR ion source 3 4 Glow discharge 3 5 Flowing afterglow 3 6 Spark ionization 4 Photoionization 4 1 Multi photon ionization 4 2 Atmospheric pressure photoionization 5 Desorption ionization 5 1 Field desorption 5 2 Particle bombardment 5 2 1 Fast atom bombardment 5 2 2 Secondary ionization 5 2 3 Plasma desorption ionization 5 3 Laser desorption ionization 5 3 1 Surface assisted laser desorption ionization 5 3 2 Surface enhanced laser desorption ionization 5 3 3 Desorption ionization on silicon 5 3 4 Smalley source 5 3 5 Aerosol ionization 6 Spray ionization 6 1 Matrix Assisted Ionization 6 2 Atmospheric pressure chemical ionization 6 3 Thermospray ionization 6 4 Electrospray ionization 6 4 1 Probe electrospray ionization 6 5 Contactless atmospheric pressure ionization 6 6 Sonic spray ionization 6 7 Ultrasonication assisted spray ionization 7 Thermal ionization 8 Ambient ionization 9 Applications 9 1 Mass spectrometry 9 2 Particle accelerators 9 3 Surface modification 10 See also 11 ReferencesElectron ionization EditMain article Electron ionization Electron ionization source schematicElectron ionization is widely used in mass spectrometry particularly for organic molecules The gas phase reaction producing electron ionization is M e M 2 e displaystyle ce M e gt M bullet 2e where M is the atom or molecule being ionized e displaystyle ce e is the electron and M displaystyle ce M bullet is the resulting ion The electrons may be created by an arc discharge between a cathode and an anode An electron beam ion source EBIS is used in atomic physics to produce highly charged ions by bombarding atoms with a powerful electron beam 2 3 Its principle of operation is shared by the electron beam ion trap Electron capture ionization Edit Electron capture ionization ECI is the ionization of a gas phase atom or molecule by attachment of an electron to create an ion of the form A The reaction is A e M A displaystyle ce A e gt M A where the M over the arrow denotes that to conserve energy and momentum a third body is required the molecularity of the reaction is three Electron capture can be used in conjunction with chemical ionization 4 An electron capture detector is used in some gas chromatography systems 5 Chemical ionization EditMain article Chemical ionization Chemical ionization CI is a lower energy process than electron ionization because it involves ion molecule reactions rather than electron removal 6 The lower energy yields less fragmentation and usually a simpler spectrum A typical CI spectrum has an easily identifiable molecular ion 7 In a CI experiment ions are produced through the collision of the analyte with ions of a reagent gas in the ion source Some common reagent gases include methane ammonia and isobutane Inside the ion source the reagent gas is present in large excess compared to the analyte Electrons entering the source will preferentially ionize the reagent gas The resultant collisions with other reagent gas molecules will create an ionization plasma Positive and negative ions of the analyte are formed by reactions with this plasma For example protonation occurs by CH 4 e CH 4 2 e displaystyle ce CH4 e gt CH4 2e primary ion formation CH 4 CH 4 CH 5 CH 3 displaystyle ce CH4 CH4 gt CH5 CH3 reagent ion formation M CH 5 CH 4 M H displaystyle ce M CH5 gt CH4 M H product ion formation e g protonation Charge exchange ionization Edit Charge exchange ionization also known as charge transfer ionization is a gas phase reaction between an ion and an atom or molecule in which the charge of the ion is transferred to the neutral species 8 A B A B displaystyle ce A B gt A B Chemi ionization Edit Chemi ionization is the formation of an ion through the reaction of a gas phase atom or molecule with an atom or molecule in an excited state 9 10 Chemi ionization can be represented by G M G M e displaystyle ce G ast M gt G M bullet e where G is the excited state species indicated by the superscripted asterisk and M is the species that is ionized by the loss of an electron to form the radical cation indicated by the superscripted plus dot Associative ionization Edit Associative ionization is a gas phase reaction in which two atoms or molecules interact to form a single product ion 11 12 13 One or both of the interacting species may have excess internal energy For example A B AB e displaystyle ce A ast B gt AB bullet e where species A with excess internal energy indicated by the asterisk interacts with B to form the ion AB Penning ionization Edit Main article Penning ionization Penning ionization is a form of chemi ionization involving reactions between neutral atoms or molecules 14 15 The process is named after the Dutch physicist Frans Michel Penning who first reported it in 1927 16 Penning ionization involves a reaction between a gas phase excited state atom or molecule G and a target molecule M resulting in the formation of a radical molecular cation M an electron e and a neutral gas molecule G 17 G M G M e displaystyle ce G ast M gt G M bullet e Penning ionization occurs when the target molecule has an ionization potential lower than the internal energy of the excited state atom or molecule Associative Penning ionization can proceed via G M MG e displaystyle ce G ast M gt MG bullet e Surface Penning ionization also known as Auger deexcitation refers to the interaction of the excited state gas with a bulk surface S resulting in the release of an electron according to G S G S e displaystyle ce G ast S gt G S e Ion attachment Edit Main article Ion attachment mass spectrometry Ion attachment ionization is similar to chemical ionization in which a cation is attached to the analyte molecule in a reactive collision M X A MX A displaystyle ce M X A gt MX A Where M is the analyte molecule X is the cation and A is a non reacting collision partner 18 In a radioactive ion source a small piece of radioactive material for instance 63Ni or 241Am is used to ionize a gas citation needed This is used in ionization smoke detectors and ion mobility spectrometers Gas discharge ion sources Edit NASA s NEXT ion thruster space craft propulsion systemThese ion sources use a plasma source or electric discharge to create ions Inductively coupled plasma Edit Main article inductively coupled plasma Ions can be created in an inductively coupled plasma which is a plasma source in which the energy is supplied by electrical currents which are produced by electromagnetic induction that is by time varying magnetic fields 19 Microwave induced plasma Edit Microwave induced plasma ion sources are capable of exciting electrodeless gas discharges to create ions for trace element mass spectrometry 20 21 A microwave plasma is a type of plasma that has high frequency electromagnetic radiation in the GHz range It is capable of exciting electrodeless gas discharges If applied in surface wave sustained mode they are especially well suited to generate large area plasmas of high plasma density If they are both in surface wave and resonator mode they can exhibit a high degree of spatial localization This allows to spatially separate the location of plasma generations from the location of surface processing Such a separation together with an appropriate gas flow scheme may help reduce the negative effect that particles released from a processed substrate may have on the plasma chemistry of the gas phase ECR ion source Edit Main article electron cyclotron resonance ECR ion sources The ECR ion source makes use of the electron cyclotron resonance to ionize a plasma Microwaves are injected into a volume at the frequency corresponding to the electron cyclotron resonance defined by the magnetic field applied to a region inside the volume The volume contains a low pressure gas Glow discharge Edit Main article glow discharge Ions can be created in an electric glow discharge A glow discharge is a plasma formed by the passage of electric current through a low pressure gas It is created by applying a voltage between two metal electrodes in an evacuated chamber containing gas When the voltage exceeds a certain value called the striking voltage the gas forms a plasma A duoplasmatron is a type of glow discharge ion source that consists of a cathode hot cathode or cold cathode that produces a plasma that is used to ionize a gas 1 22 Duoplasmatrons can produce positive or negative ions 23 Duoplasmatrons are used for secondary ion mass spectrometry 24 25 ion beam etching and high energy physics 26 Flowing afterglow Edit Main article plasma afterglow In a flowing afterglow ions are formed in a flow of inert gas typically helium or argon 27 28 29 Reagents are added downstream to create ion products and study reaction rates Flowing afterglow mass spectrometry is used for trace gas analysis 30 for organic compounds 31 Spark ionization Edit Main article spark ionization Electric spark ionization is used to produce gas phase ions from a solid sample When incorporated with a mass spectrometer the complete instrument is referred to as a spark ionization mass spectrometer or as a spark source mass spectrometer SSMS 32 A closed drift ion source uses a radial magnetic field in an annular cavity in order to confine electrons for ionizing a gas They are used for ion implantation and for space propulsion Hall effect thrusters Photoionization EditMain article Photoionization Photoionization is the ionization process in which an ion is formed from the interaction of a photon with an atom or molecule 33 Multi photon ionization Edit In multi photon ionization MPI several photons of energy below the ionization threshold may actually combine their energies to ionize an atom Resonance enhanced multiphoton ionization REMPI is a form of MPI in which one or more of the photons accesses a bound bound transition that is resonant in the atom or molecule being ionized Atmospheric pressure photoionization Edit Main article Atmospheric pressure photoionization Atmospheric pressure photoionization APPI uses a source of photons usually a vacuum UV VUV lamp to ionize the analyte with single photon ionization process Analogous to other atmospheric pressure ion sources a spray of solvent is heated to relatively high temperatures above 400 degrees Celsius and sprayed with high flow rates of nitrogen for desolvation The resulting aerosol is subjected to UV radiation to create ions Atmospheric pressure laser ionization uses UV laser light sources to ionize the analyte via MPI Desorption ionization EditField desorption Edit Main article Field desorption Field desorption schematicField desorption refers to an ion source in which a high potential electric field is applied to an emitter with a sharp surface such as a razor blade or more commonly a filament from which tiny whiskers have formed 34 This results in a very high electric field which can result in ionization of gaseous molecules of the analyte Mass spectra produced by FI have little or no fragmentation They are dominated by molecular radical cations M displaystyle ce M and less often protonated molecules M H displaystyle ce M H Particle bombardment Edit Fast atom bombardment Edit Main article fast atom bombardment Particle bombardment with atoms is called fast atom bombardment FAB and bombardment with atomic or molecular ions is called secondary ion mass spectrometry SIMS 35 Fission fragment ionization uses ionic or neutral atoms formed as a result of the nuclear fission of a suitable nuclide for example the Californium isotope 252Cf In FAB the analytes is mixed with a non volatile chemical protection environment called a matrix and is bombarded under vacuum with a high energy 4000 to 10 000 electron volts beam of atoms 36 The atoms are typically from an inert gas such as argon or xenon Common matrices include glycerol thioglycerol 3 nitrobenzyl alcohol 3 NBA 18 crown 6 ether 2 nitrophenyloctyl ether sulfolane diethanolamine and triethanolamine This technique is similar to secondary ion mass spectrometry and plasma desorption mass spectrometry Secondary ionization Edit Secondary ion mass spectrometry SIMS is used to analyze the composition of solid surfaces and thin films by sputtering the surface of the specimen with a focused primary ion beam and collecting and analyzing ejected secondary ions The mass charge ratios of these secondary ions are measured with a mass spectrometer to determine the elemental isotopic or molecular composition of the surface to a depth of 1 to 2 nm In a liquid metal ion source LMIS a metal typically gallium is heated to the liquid state and provided at the end of a capillary or a needle Then a Taylor cone is formed under the application of a strong electric field As the cone s tip get sharper the electric field becomes stronger until ions are produced by field evaporation These ion sources are particularly used in ion implantation or in focused ion beam instruments Plasma desorption ionization Edit Schematic representation of a plasama desorption time of flight mass spectrometerPlasma desorption ionization mass spectrometry PDMS also called fission fragment ionization is a mass spectrometry technique in which ionization of material in a solid sample is accomplished by bombarding it with ionic or neutral atoms formed as a result of the nuclear fission of a suitable nuclide typically the californium isotope 252Cf 37 38 Laser desorption ionization Edit Diagram of a MALDI ion sourceMatrix assisted laser desorption ionization MALDI is a soft ionization technique The sample is mixed with a matrix material Upon receiving a laser pulse the matrix absorbs the laser energy and it is thought that primarily the matrix is desorbed and ionized by addition of a proton by this event The analyte molecules are also desorbed The matrix is then thought to transfer proton to the analyte molecules e g protein molecules thus charging the analyte Surface assisted laser desorption ionization Edit Surface assisted laser desorption ionization SALDI is a soft laser desorption technique used for analyzing biomolecules by mass spectrometry 39 40 In its first embodiment it used graphite matrix 39 At present laser desorption ionization methods using other inorganic matrices such as nanomaterials are often regarded as SALDI variants A related method named ambient SALDI which is a combination of conventional SALDI with ambient mass spectrometry incorporating the DART ion source has also been demonstrated 41 Surface enhanced laser desorption ionization Edit Main article Surface enhanced laser desorption ionization Surface enhanced laser desorption ionization SELDI is a variant of MALDI that is used for the analysis of protein mixtures that uses a target modified to achieve biochemical affinity with the analyte compound 42 Desorption ionization on silicon Edit Main article Desorption ionization on silicon Desorption ionization on silicon DIOS refers to laser desorption ionization of a sample deposited on a porous silicon surface 43 Smalley source Edit A laser vaporization cluster source produces ions using a combination of laser desorption ionization and supersonic expansion 44 The Smalley source or Smalley cluster source 45 was developed by Richard Smalley at Rice University in the 1980s and was central to the discovery of fullerenes in 1985 46 47 Aerosol ionization Edit In aerosol mass spectrometry with time of flight analysis micrometer sized solid aerosol particles extracted from the atmosphere are simultaneously desorbed and ionized by a precisely timed laser pulse as they pass through the center of a time of flight ion extractor 48 49 Spray ionization Edit Atmospheric pressure chemical ionization sourceSpray ionization methods involve the formation of aerosol particles from a liquid solution and the formation of bare ions after solvent evaporation 50 Solvent assisted ionization SAI is a method in which charged droplets are produced by introducing a solution containing analyte into a heated inlet tube of an atmospheric pressure ionization mass spectrometer Just as in Electrospray Ionization ESI desolvation of the charged droplets produces multiply charged analyte ions Volatile and nonvolatile compounds are analyzed by SAI and high voltage is not required to achieve sensitivity comparable to ESI 51 Application of a voltage to the solution entering the hot inlet through a zero dead volume fitting connected to fused silica tubing produces ESI like mass spectra but with higher sensitivity 52 The inlet tube to the mass spectrometer becomes the ion source Matrix Assisted Ionization Edit Matrix Assisted Ionization MAI is similar to MALDI in sample preparation but a laser is not required to convert analyte molecules included in a matrix compound into gas phase ions In MAI analyte ions have charge states similar to electrospray ionization but obtained from a solid matrix rather than a solvent No voltage or laser is required but a laser can be used to obtain spatial resolution for imaging Matrix analyte samples are ionized in the vacuum of a mass spectrometer and can be inserted into the vacuum through an atmospheric pressure inlet Less volatile matrices such as 2 5 dihydroxybenzoic acid require a hot inlet tube to produce analyte ions by MAI but more volatile matrices such as 3 nitrobenzonitrile require no heat voltage or laser Simply introducing the matrix analyte sample to the inlet aperture of an atmospheric pressure ionization mass spectrometer produces abundant ions Compounds at least as large as bovine serum albumin 66 kDa can be ionized with this method 53 In this simple low cost and easy to use ionization method the inlet to the mass spectrometer can be considered the ion source Atmospheric pressure chemical ionization Edit Main article Atmospheric pressure chemical ionization Atmospheric pressure chemical ionization is a form of chemical ionization using a solvent spray at atmospheric pressure 54 A spray of solvent is heated to relatively high temperatures above 400 degrees Celsius sprayed with high flow rates of nitrogen and the entire aerosol cloud is subjected to a corona discharge that creates ions with the evaporated solvent acting as the chemical ionization reagent gas APCI is not as soft low fragmentation an ionization technique as ESI 55 Note that atmospheric pressure ionization API should not be used as a synonym for APCI 56 Thermospray ionization Edit Main article Thermospray ionization Thermospray ionization is a form of atmospheric pressure ionization in mass spectrometry It transfers ions from the liquid phase to the gas phase for analysis It is particularly useful in liquid chromatography mass spectrometry 57 Electrospray ion sourceElectrospray ionization Edit Main article Electrospray ionization In electrospray ionization a liquid is pushed through a very small charged and usually metal capillary 58 This liquid contains the substance to be studied the analyte dissolved in a large amount of solvent which is usually much more volatile than the analyte Volatile acids bases or buffers are often added to this solution too The analyte exists as an ion in solution either in its anion or cation form Because like charges repel the liquid pushes itself out of the capillary and forms an aerosol a mist of small droplets about 10 mm across The aerosol is at least partially produced by a process involving the formation of a Taylor cone and a jet from the tip of this cone An uncharged carrier gas such as nitrogen is sometimes used to help nebulize the liquid and to help evaporate the neutral solvent in the droplets As the solvent evaporates the analyte molecules are forced closer together repel each other and break up the droplets This process is called Coulombic fission because it is driven by repulsive Coulombic forces between charged molecules The process repeats until the analyte is free of solvent and is a bare ion The ions observed are created by the addition of a proton a hydrogen ion and denoted M H displaystyle ce M H or of another cation such as sodium ion M Na displaystyle ce M Na or the removal of a proton M H displaystyle ce M H Multiply charged ions such as M 2 H 2 displaystyle ce M 2H 2 are often observed For large macromolecules there can be many charge states occurring with different frequencies the charge can be as great as M 25 H 25 displaystyle ce M 25H 25 for example Probe electrospray ionization Edit Main article Probe electrospray ionization Probe electrospray ionization PESI is a modified version of electrospray where the capillary for sample solution transferring is replaced by a sharp tipped solid needle with periodical motion 59 Contactless atmospheric pressure ionization Edit Main article Contactless atmospheric pressure ionization Contactless atmospheric pressure ionization is a technique used for analysis of liquid and solid samples by mass spectrometry 60 Contactless API can be operated without an additional electric power supply supplying voltage to the source emitter gas supply or syringe pump Thus the technique provides a facile means for analyzing chemical compounds by mass spectrometry at atmospheric pressure Sonic spray ionization Edit Main article Sonic spray ionization Sonic spray ionization is method for creating ions from a liquid solution for example a mixture of methanol and water 61 A pneumatic nebulizer is used to turn the solution into a supersonic spray of small droplets Ions are formed when the solvent evaporates and the statistically unbalanced charge distribution on the droplets leads to a net charge and complete desolvation results in the formation of ions Sonic spray ionization is used to analyze small organic molecules and drugs and can analyze large molecules when an electric field is applied to the capillary to help increase the charge density and generate multiple charged ions of proteins 62 Sonic spray ionization has been coupled with high performance liquid chromatography for the analysis of drugs 63 64 Oligonucleotides have been studied with this method 65 66 SSI has been used in a manner similar to desorption electrospray ionization 67 for ambient ionization and has been coupled with thin layer chromatography in this manner 68 Ultrasonication assisted spray ionization Edit Main article Ultrasonication assisted spray ionization Ultrasonication assisted spray ionization UASI involves ionization through the application of ultrasound 69 70 Thermal ionization EditMain article Thermal ionization Thermal ionization also known as surface ionization or contact ionization involves spraying vaporized neutral atoms onto a hot surface from which the atoms re evaporate in ionic form To generate positive ions the atomic species should have a low ionization energy and the surface should have a high work function This technique is most suitable for alkali atoms Li Na K Rb Cs which have low ionization energies and are easily evaporated 71 To generate negative ions the atomic species should have a high electron affinity and the surface should have a low work function This second approach is most suited for halogen atoms Cl Br I At 72 Ambient ionization EditMain article Ambient ionization Direct analysis in real time ambient ionization ion sourceIn ambient ionization ions are formed outside the mass spectrometer without sample preparation or separation 73 74 75 Ions can be formed by extraction into charged electrospray droplets thermally desorbed and ionized by chemical ionization or laser desorbed or ablated and post ionized before they enter the mass spectrometer Solid liquid extraction based ambient ionization uses a charged spray to create a liquid film on the sample surface 74 76 Molecules on the surface are extracted into the solvent The action of the primary droplets hitting the surface produces secondary droplets that are the source of ions for the mass spectrometer Desorption electrospray ionization DESI uses an electrospray source to create charged droplets that are directed at a solid sample a few millimeters to a few centimeters away The charged droplets pick up the sample through interaction with the surface and then form highly charged ions that can be sampled into a mass spectrometer 77 Plasma based ambient ionization is based on an electrical discharge in a flowing gas that produces metastable atoms and molecules and reactive ions Heat is often used to assist in the desorption of volatile species from the sample Ions are formed by chemical ionization in the gas phase A direct analysis in real time source operates by exposing the sample to a dry gas stream typically helium or nitrogen that contains long lived electronically or vibronically excited neutral atoms or molecules or metastables Excited states are typically formed in the DART source by creating a glow discharge in a chamber through which the gas flows A similar method called atmospheric solids analysis probe ASAP uses the heated gas from ESI or APCI probes to vaporize sample placed on a melting point tube inserted into an ESI APCI source 78 Ionization is by APCI Laser based ambient ionization is a two step process in which a pulsed laser is used to desorb or ablate material from a sample and the plume of material interacts with an electrospray or plasma to create ions Electrospray assisted laser desorption ionization ELDI uses a 337 nm UV laser 79 or 3 mm infrared laser 80 to desorb material into an electrospray source Matrix assisted laser desorption electrospray ionization MALDESI 81 is an atmospheric pressure ionization source for generation of multiply charged ions An ultraviolet or infrared laser is directed onto a solid or liquid sample containing the analyte of interest and matrix desorbing neutral analyte molecules that are ionized by interaction with electrosprayed solvent droplets generating multiply charged ions Laser ablation electrospray ionization LAESI is an ambient ionization method for mass spectrometry that combines laser ablation from a mid infrared mid IR laser with a secondary electrospray ionization ESI process Applications EditMass spectrometry Edit In a mass spectrometer a sample is ionized in an ion source and the resulting ions are separated by their mass to charge ratio The ions are detected and the results are displayed as spectra of the relative abundance of detected ions as a function of the mass to charge ratio The atoms or molecules in the sample can be identified by correlating known masses to the identified masses or through a characteristic fragmentation pattern Particle accelerators Edit Surface ionization source at the Argonne Tandem Linear Accelerator System ATLAS Ion source used in the Cockcroft Walton pre accelerator at Fermilab 82 In particle accelerators an ion source creates a particle beam at the beginning of the machine the source The technology to create ion sources for particle accelerators depends strongly on the type of particle that needs to be generated electrons protons H ion or a Heavy ions Electrons are generated with an electron gun of which there are many varieties Protons are generated with a plasma based device like a duoplasmatron or a magnetron H ions are generated with a magnetron or a Penning source A magnetron consists of a central cylindrical cathode surrounded by an anode The discharge voltage is typically greater than 150 V and the current drain is around 40 A A magnetic field of about 0 2 tesla is parallel to the cathode axis Hydrogen gas is introduced by a pulsed gas valve Caesium is often used to lower the work function of the cathode enhancing the amount of ions that are produced Large caesiated H sources are also used for plasma heating in nuclear fusion devices For a Penning source a strong magnetic field parallel to the electric field of the sheath guides electrons and ions on cyclotron spirals from cathode to cathode Fast H minus ions are generated at the cathodes as in the magnetron They are slowed down due to the charge exchange reaction as they migrate to the plasma aperture This makes for a beam of ions that is colder than the ions obtained from a magnetron Heavy ions can be generated with an electron cyclotron resonance ion source The use of electron cyclotron resonance ECR ion sources for the production of intense beams of highly charged ions has immensely grown over the last decade ECR ion sources are used as injectors into linear accelerators Van de Graaff generators or cyclotrons in nuclear and elementary particle physics In atomic and surface physics ECR ion sources deliver intense beams of highly charged ions for collision experiments or for the investigation of surfaces For the highest charge states however Electron beam ion sources EBIS are needed They can generate even bare ions of mid heavy elements The Electron beam ion trap EBIT based on the same principle can produce up to bare uranium ions and can be used as an ion source as well Heavy ions can also be generated with an Ion Gun which typically uses the thermionic emission of electrons to ionize a substance in its gaseous state Such instruments are typically used for surface analysis Ion beam deposition system with mass separatorGas flows through the ion source between the anode and the cathode A positive voltage is applied to the anode This voltage combined with the high magnetic field between the tips of the internal and external cathodes allow a plasma to start Ions from the plasma are repelled by the anode electric field This creates an ion beam 83 Surface modification Edit Surface cleaning and pretreatment for large area deposition Thin film deposition Deposition of thick diamond like carbon DLC films Surface roughening of polymers for improved adhesion and or biocompatibility 84 See also EditIon beam RF antenna ion source On Line Isotope Mass SeparatorReferences Edit a b Bernhard Wolf 31 August 1995 Handbook of Ion Sources CRC Press ISBN 978 0 8493 2502 1 Ian G Brown 6 March 2006 The Physics and Technology of Ion Sources John Wiley amp Sons ISBN 978 3 527 60454 8 Heinrich Beyer Heinrich F Beyer H Jurgen Kluge H J Kluge Vi acheslav Petrovich Shevelʹko 14 August 1997 X Ray Radiation of Highly Charged Ions Springer Science amp Business Media ISBN 978 3 540 63185 9 Donald F Hunt Frank W Crow 1978 Electron capture negative ion chemical ionization mass spectrometry Analytical Chemistry 50 13 1781 1784 doi 10 1021 ac50035a017 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 electron capture detector in gas chromatography doi 10 1351 goldbook E01981 Munson M S B Field F H 1966 Chemical Ionization Mass Spectrometry I General Introduction Journal of the American Chemical Society 88 12 2621 2630 doi 10 1021 ja00964a001 de Hoffmann Edmond Vincent Stroobant 2003 Mass Spectrometry Principles and Applications Second ed Toronto John Wiley amp Sons Ltd p 14 ISBN 978 0 471 48566 7 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 charge exchange ionization doi 10 1351 goldbook C00989 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 chemi ionization doi 10 1351 goldbook C01044 C01044 Klucharev A N 1993 Chemi ionization processes Physics Uspekhi 36 6 486 512 Bibcode 1993PhyU 36 486K doi 10 1070 PU1993v036n06ABEH002162 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 associative ionization doi 10 1351 goldbook A00475 Jones DM Dahler JS April 1988 Theory of associative ionization Physical Review A 37 8 2916 2933 Bibcode 1988PhRvA 37 2916J doi 10 1103 PhysRevA 37 2916 PMID 9900022 Cohen James S 1976 Multistate curve crossing model for scattering Associative ionization and excitation transfer in helium Physical Review A 13 1 99 114 Bibcode 1976PhRvA 13 99C doi 10 1103 PhysRevA 13 99 Arango CA Shapiro M Brumer P 2006 Cold atomic collisions coherent control of penning and associative ionization Phys Rev Lett 97 19 193202 arXiv physics 0610131 Bibcode 2006PhRvL 97s3202A doi 10 1103 PhysRevLett 97 193202 PMID 17155624 S2CID 1480148 Hiraoka K Furuya H Kambara S Suzuki S Hashimoto Y Takamizawa A 2006 Atmospheric pressure Penning ionization of aliphatic hydrocarbons Rapid Commun Mass Spectrom 20 21 3213 22 Bibcode 2006RCMS 20 3213H doi 10 1002 rcm 2706 PMID 17016831 Penning F M Die Naturwissenschaften 1927 15 818 Uber Ionisation durch metastabile Atome IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 Penning gas mixture doi 10 1351 goldbook P04476 Selvin P Christopher Fujii Toshihiro 2001 Lithium ion attachment mass spectrometry Instrumentation and features Review of Scientific Instruments 72 5 2248 Bibcode 2001RScI 72 2248S doi 10 1063 1 1362439 A Montaser and D W Golightly eds Inductively Coupled Plasmas in Analytical Atomic Spectrometry VCH Publishers Inc New York 1992 Okamoto Yukio 1994 High sensitivity microwave induced plasma mass spectrometry for trace element analysis Journal of Analytical Atomic Spectrometry 9 7 745 doi 10 1039 ja9940900745 ISSN 0267 9477 Douglas D J French J B 1981 Elemental analysis with a microwave induced plasma quadrupole mass spectrometer system Analytical Chemistry 53 1 37 41 doi 10 1021 ac00224a011 ISSN 0003 2700 Lejeune C 1974 Theoretical and experimental study of the duoplasmatron ion source Nuclear Instruments and Methods 116 3 417 428 Bibcode 1974NucIM 116 417L doi 10 1016 0029 554X 74 90821 0 ISSN 0029 554X Aberth William Peterson James R 1967 Characteristics of a Low Energy Duoplasmatron Negative Ion Source Review of Scientific Instruments 38 6 745 Bibcode 1967RScI 38 745A doi 10 1063 1 1720882 ISSN 0034 6748 Coath C D Long J V P 1995 A high brightness duoplasmatron ion source for microprobe secondary ion mass spectrometry Review of Scientific Instruments 66 2 1018 Bibcode 1995RScI 66 1018C doi 10 1063 1 1146038 ISSN 0034 6748 Christine M Mahoney 9 April 2013 Cluster Secondary Ion Mass Spectrometry Principles and Applications John Wiley amp Sons pp 65 ISBN 978 1 118 58925 0 Stanley Humphries 25 July 2013 Charged Particle Beams Dover Publications pp 309 ISBN 978 0 486 31585 0 Ferguson E E Fehsenfeld F C Schmeltekopf A L 1969 Ion Molecule Reaction Rates Measured in a Discharge Afterglow Chemical Reactions in Electrical Discharges Advances in Chemistry Vol 80 pp 83 91 doi 10 1021 ba 1969 0080 ch006 ISBN 978 0 8412 0081 4 ISSN 0065 2393 Ferguson Eldon E 1992 A Personal history of the early development of the flowing afterglow technique for ion molecule reaction studies Journal of the American Society for Mass Spectrometry Submitted manuscript 3 5 479 486 doi 10 1016 1044 0305 92 85024 E ISSN 1044 0305 PMID 24234490 Bierbaum Veronica M 2014 Go with the flow Fifty years of innovation and ion chemistry using the flowing afterglow International Journal of Mass Spectrometry 377 456 466 Bibcode 2015IJMSp 377 456B doi 10 1016 j ijms 2014 07 021 ISSN 1387 3806 Smith David Spanel Patrik 2005 Selected ion flow tube mass spectrometry SIFT MS for on line trace gas analysis Mass Spectrometry Reviews 24 5 661 700 Bibcode 2005MSRv 24 661S doi 10 1002 mas 20033 ISSN 0277 7037 PMID 15495143 Dhooghe Frederik Vansintjan Robbe Schoon Niels Amelynck Crist 2012 08 30 Studies in search of selective detection of isomeric biogenic hexen 1 ols and hexanal by flowing afterglow tandem mass spectrometry using H3O and NO reagent ions Rapid Communications in Mass Spectrometry 26 16 1868 1874 doi 10 1002 rcm 6294 ISSN 1097 0231 PMID 22777789 H E Beske A Hurrle K P Jochum 1981 Part I Principles of spark source mass spectrometry SSMS Fresenius Journal of Analytical Chemistry 309 4 258 261 doi 10 1007 BF00488596 S2CID 92433014 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 photoionization doi 10 1351 goldbook P04620 Beckey H D 1969 Field ionization mass spectrometry Research Development 20 11 26 Williams Dudley H Findeis A Frederick Naylor Stephen Gibson Bradford W 1987 Aspects of the production of FAB and SIMS mass spectra Journal of the American Chemical Society 109 7 1980 1986 doi 10 1021 ja00241a013 ISSN 0002 7863 Morris HR Panico M Barber M Bordoli RS Sedgwick RD Tyler A 1981 Fast atom bombardment a new mass spectrometric method for peptide sequence analysis Biochem Biophys Res Commun 101 2 623 31 doi 10 1016 0006 291X 81 91304 8 PMID 7306100 Macfarlane R Torgerson D 1976 Californium 252 plasma desorption mass spectroscopy Science 191 4230 920 925 Bibcode 1976Sci 191 920M doi 10 1126 science 1251202 ISSN 0036 8075 PMID 1251202 Hilf E R 1993 Approaches to plasma desorption mass spectrometry by some theoretical physics concepts International Journal of Mass Spectrometry and Ion Processes 126 25 36 Bibcode 1993IJMSI 126 25H doi 10 1016 0168 1176 93 80067 O ISSN 0168 1176 a b Sunner Jan Dratz Edward Chen Yu Chie 1995 Graphite surface assisted laser desorption ionization time of flight mass spectrometry of peptides and proteins from liquid solutions Analytical Chemistry 67 23 4335 4342 doi 10 1021 ac00119a021 ISSN 0003 2700 PMID 8633776 Dattelbaum Andrew M Iyer Srinivas 2006 Surface assisted laser desorption ionization mass spectrometry Expert Review of Proteomics Submitted manuscript 3 1 153 161 doi 10 1586 14789450 3 1 153 ISSN 1478 9450 PMID 16445359 S2CID 39538990 Zhang Jialing Li Ze Zhang Chengsen Feng Baosheng Zhou Zhigui Bai Yu Liu Huwei 2012 Graphite Coated Paper as Substrate for High Sensitivity Analysis in Ambient Surface Assisted Laser Desorption Ionization Mass Spectrometry Analytical Chemistry 84 7 3296 3301 doi 10 1021 ac300002g ISSN 0003 2700 PMID 22380704 Tang N Tornatore P Weinberger SR 2004 Current developments in SELDI affinity technology Mass Spectrometry Reviews 23 1 34 44 Bibcode 2004MSRv 23 34T doi 10 1002 mas 10066 PMID 14625891 Buriak Jillian M Wei Jing Siuzdak Gary 1999 Desorption ionization mass spectrometry on porous silicon Nature 399 6733 243 246 Bibcode 1999Natur 399 243W doi 10 1038 20400 ISSN 0028 0836 PMID 10353246 S2CID 4314372 Duncan Michael A 2012 Invited Review Article Laser vaporization cluster sources Review of Scientific Instruments 83 4 041101 041101 19 Bibcode 2012RScI 83d1101D doi 10 1063 1 3697599 ISSN 0034 6748 PMID 22559508 Laser Ablation and Desorption Academic Press 10 December 1997 pp 628 ISBN 978 0 08 086020 6 Smalley Richard 1997 Discovering the fullerenes Reviews of Modern Physics 69 3 723 730 Bibcode 1997RvMP 69 723S doi 10 1103 RevModPhys 69 723 ISSN 0034 6861 Roy L Johnston 25 April 2002 Atomic and Molecular Clusters CRC Press pp 150 ISBN 978 1 4200 5577 1 Carson P Neubauer K Johnston M Wexler A 1995 On line chemical analysis of aerosols by rapid single particle mass spectrometry Journal of Aerosol Science 26 4 535 545 Bibcode 1995JAerS 26 535C doi 10 1016 0021 8502 94 00133 J Guazzotti S Coffee K Prather K 2000 Real time monitoring of size resolved single particle chemistry during INDOEX IFP 99 Journal of Aerosol Science 31 182 183 Bibcode 2000JAerS 31 182G doi 10 1016 S0021 8502 00 90189 7 Chhabil Dass 11 May 2007 Fundamentals of Contemporary Mass Spectrometry John Wiley amp Sons pp 45 57 ISBN 978 0 470 11848 1 Pagnotti VS Chubatyi ND McEwen CN 2011 Solvent Assisted Inlet Ionization an Ultrasensitive New Liquid Introduction Ionization Method for Mass Spectrometry Anal Chem 83 11 3981 3985 doi 10 1021 ac200556z PMID 21528896 Pagnotti VS Chakrabarty S Harron AF McEwen CN 2012 Increasing the Sensitivity of Liquid Introduction Mass Spectrometry by Combining Electrospray Ionization and Solvent Assisted Inlet Ionization Anal Chem 84 15 6828 6832 doi 10 1021 ac3014115 PMID 22742705 Trimpin S Wang B Lietz CB Marshall DD Richards AL Inutan ED New Ionization Processes and Applications for Use in Mass Spectrometry Rev Biochem Mol Biol 2013 5 409 429 Prakash C Shaffer CL Nedderman A 2007 Analytical strategies for identifying drug metabolites Mass Spectrometry Reviews 26 3 340 69 Bibcode 2007MSRv 26 340P doi 10 1002 mas 20128 PMID 17405144 Zaikin VG Halket JM 2006 Derivatization in mass spectrometry 8 Soft ionization mass spectrometry of small molecules European Journal of Mass Spectrometry 12 2 79 115 doi 10 1255 ejms 798 PMID 16723751 S2CID 34838846 Atmospheric pressure ionization in mass spectrometry IUPAC Compendium of Chemical Terminology 2009 doi 10 1351 goldbook A00492 ISBN 978 0 9678550 9 7 Blakley C R Carmody J J Vestal M L 1980 Liquid Chromatograph Mass Spectrometer for Analysis of Nonvolatile Samples Analytical Chemistry 1980 52 1636 1641 doi 10 1021 ac50061a025 Fenn J B Mann M Meng C K Wong S F Whitehouse C M 1990 Electrospray Ionization Principles and Practice Mass Spectrometry Reviews 9 1 37 70 Bibcode 1990MSRv 9 37F doi 10 1002 mas 1280090103 Hiraoka K Nishidate K Mori K Asakawa D Suzuki S 2007 Development of probe electrospray using a solid needle Rapid Communications in Mass Spectrometry 21 18 3139 3144 Bibcode 2007RCMS 21 3139H doi 10 1002 rcm 3201 PMID 17708527 Hsieh Cheng Huan Chang Chia Hsien Urban Pawel L Chen Yu Chie 2011 Capillary Action Supported Contactless Atmospheric Pressure Ionization for the Combined Sampling and Mass Spectrometric Analysis of Biomolecules Analytical Chemistry 83 8 2866 2869 doi 10 1021 ac200479s ISSN 0003 2700 PMID 21446703 Hirabayashi A Sakairi M Koizumi H 1995 Sonic spray mass spectrometry Anal Chem 67 17 2878 82 doi 10 1021 ac00113a023 PMID 8779414 Chen Tsung Yi Lin Jia Yi Chen Jen Yi Chen Yu Chie 2011 11 22 Ultrasonication assisted spray ionization mass spectrometry for the analysis of biomolecules in solution Journal of the American Society for Mass Spectrometry 21 9 1547 1553 doi 10 1016 j jasms 2010 04 021 ISSN 1044 0305 PMID 20547459 Arinobu T Hattori H Seno H Ishii A Suzuki O 2002 Comparison of SSI with APCI as an interface of HPLC mass spectrometry for analysis of a drug and its metabolites J Am Soc Mass Spectrom 13 3 204 208 doi 10 1016 S1044 0305 01 00359 2 PMID 11908800 Dams R Benijts T Gunther W Lambert W De Leenheer A 2002 Sonic spray ionization technology performance study and application to a LC MS analysis on a monolithic silica column for heroin impurity profiling Anal Chem 74 13 3206 3212 doi 10 1021 ac0112824 PMID 12141684 Huang M Hirabayashi A Okumura A Hirabayashi Y 2001 Matrix effect on the analysis of oligonucleotides by using a mass spectrometer with a sonic spray ionization source Anal Sci 17 10 1179 1182 doi 10 2116 analsci 17 1179 PMID 11990592 Huang M Hirabayashi A 2002 Multi charged oligonucleotide ion formation in sonic spray ionization Anal Sci 18 4 385 390 doi 10 2116 analsci 18 385 PMID 11999509 Haddad R Sparrapan R Eberlin MN 2006 Desorption sonic spray ionization for high voltage free ambient mass spectrometry Rapid Commun Mass Spectrom 20 19 2901 2905 Bibcode 2006RCMS 20 2901H doi 10 1002 rcm 2680 PMID 16941547 Haddad R Milagre HM Catharino RR Eberlin MN 2008 Easy Ambient Sonic Spray Ionization Mass Spectrometry Combined with Thin Layer Chromatography Anal Chem 80 8 2744 2750 doi 10 1021 ac702216q PMID 18331004 Chen Tsung Yi Lin Jia Yi Chen Jen Yi Chen Yu Chie 2010 Ultrasonication assisted spray ionization mass spectrometry for the analysis of biomolecules in solution Journal of the American Society for Mass Spectrometry 21 9 1547 1553 doi 10 1016 j jasms 2010 04 021 PMID 20547459 Chen Tsung Yi Chao Chin Sheng Mong Kwok Kong Tony Chen Yu Chie 4 November 2010 Ultrasonication assisted spray ionization mass spectrometry for on line monitoring of organic reactions Chemical Communications 46 44 8347 9 doi 10 1039 C0CC02629H PMID 20957254 Retrieved 4 November 2011 Alton G D 1988 Characterization of a cesium surface ionization source with a porous tungsten ionizer I Review of Scientific Instruments Submitted manuscript 59 7 1039 Bibcode 1988RScI 59 1039A doi 10 1063 1 1139776 ISSN 0034 6748 A Negative Surface Ionization for Generation of Halogen Radioactive Ion Beams PDF Archived from the original PDF on 2004 12 18 Retrieved 2014 01 20 Cooks R Graham Ouyang Zheng Takats Zoltan Wiseman Justin M 2006 Ambient Mass Spectrometry Science 311 5767 1566 70 Bibcode 2006Sci 311 1566C doi 10 1126 science 1119426 PMID 16543450 S2CID 98131681 a b Monge Maria Eugenia Harris Glenn A Dwivedi Prabha Fernandez Facundo M 2013 Mass Spectrometry Recent Advances in Direct Open Air Surface Sampling Ionization Chemical Reviews 113 4 2269 2308 doi 10 1021 cr300309q ISSN 0009 2665 PMID 23301684 Huang Min Zong Yuan Cheng Hui Cheng Sy Chyi Cho Yi Tzu Shiea Jentaie 2010 Ambient Ionization Mass Spectrometry Annual Review of Analytical Chemistry 3 1 43 65 Bibcode 2010ARAC 3 43H doi 10 1146 annurev anchem 111808 073702 ISSN 1936 1327 PMID 20636033 Badu Tawiah Abraham K Eberlin Livia S Ouyang Zheng Cooks R Graham 2013 Chemical Aspects of the Extractive Methods of Ambient Ionization Mass Spectrometry Annual Review of Physical Chemistry 64 1 481 505 Bibcode 2013ARPC 64 481B doi 10 1146 annurev physchem 040412 110026 ISSN 0066 426X PMID 23331308 Takats Z Wiseman JM Cooks RG 2005 Ambient mass spectrometry using desorption electrospray ionization DESI instrumentation mechanisms and applications in forensics chemistry and biology Journal of Mass Spectrometry 40 10 1261 75 Bibcode 2005JMSp 40 1261T doi 10 1002 jms 922 PMID 16237663 McEwen CN McKay RG Larsen BS 2005 Analysis of Solids Liquids and Biological Tissues Using Solids Probe Introduction at Atmospheric Pressure on Commercial LC MS Instruments Anal Chem 77 23 7826 7831 doi 10 1021 ac051470k PMID 16316194 Shiea J Huang MZ Hsu HJ Lee CY Yuan CH Beech I Sunner J 2005 Electrospray assisted laser desorption ionization mass spectrometry for direct ambient analysis of solids Rapid Commun Mass Spectrom 19 24 3701 4 Bibcode 2005RCMS 19 3701S doi 10 1002 rcm 2243 PMID 16299699 Peng Ivory X Ogorzalek Loo Rachel R Margalith Eli Little Mark W Loo Joseph A 2010 Electrospray assisted laser desorption ionization mass spectrometry ELDI MS with an infrared laser for characterizing peptides and proteins The Analyst 135 4 767 72 Bibcode 2010Ana 135 767P doi 10 1039 b923303b ISSN 0003 2654 PMC 3006438 PMID 20349541 Sampson JS Hawkridge AM Muddiman DC 2006 Generation and detection of multiply charged peptides and proteins by matrix assisted laser desorption electrospray ionization MALDESI Fourier transform ion cyclotron resonance mass spectrometry J Am Soc Mass Spectrom 17 12 1712 6 doi 10 1016 j jasms 2006 08 003 PMID 16952462 35 years of H ions at Fermilab PDF Fermilab p 12 Retrieved 12 August 2015 Cooks R G Ouyang Z Takats Z Wiseman J M 2006 Ion Beam Sources PDF Science 311 5767 1566 70 Bibcode 2006Sci 311 1566C doi 10 1126 science 1119426 PMID 16543450 S2CID 98131681 Archived from the original PDF on 2006 10 18 Retrieved 2006 12 14 Ion Beam Source Technology Advanced Energy Archived from the original on October 18 2006 Retrieved 2006 12 14 Retrieved from https en wikipedia org w index php title Ion source amp oldid 1172251871 Sonic spray ionization, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.