Electrospray ionization

Electrospray ionization (ESI) is a technique used in mass spectrometry to produce ions using an electrospray in which a high voltage is applied to a liquid to create an aerosol. It is especially useful in producing ions from macromolecules because it overcomes the propensity of these molecules to fragment when ionized. ESI is different from other ionization processes (e.g. matrix-assisted laser desorption/ionization, MALDI) since it may produce multiple-charged ions, effectively extending the mass range of the analyser to accommodate the kDa-MDa orders of magnitude observed in proteins and their associated polypeptide fragments.[1][2]

Electrospray (nanoSpray) ionization source

Mass spectrometry using ESI is called electrospray ionization mass spectrometry (ESI-MS) or, less commonly, electrospray mass spectrometry (ES-MS). ESI is a so-called 'soft ionization' technique, since there is very little fragmentation. This can be advantageous in the sense that the molecular ion (or more accurately a pseudo molecular ion) is almost always observed, however very little structural information can be gained from the simple mass spectrum obtained. This disadvantage can be overcome by coupling ESI with tandem mass spectrometry (ESI-MS/MS). Another important advantage of ESI is that solution-phase information can be retained into the gas-phase.

The electrospray ionization technique was first reported by Masamichi Yamashita and John Fenn in 1984,[3] and independently by Lidia Gall and co-workers in Soviet Union, also in 1984.[4] Gall's work was not recognised or translated in the western scientific literature until a translation was published in 2008.[4] The development of electrospray ionization for the analysis of biological macromolecules[5] was rewarded with the attribution of the Nobel Prize in Chemistry to John Bennett Fenn and Koichi Tanaka in 2002.[6] One of the original instruments used by Fenn is on display at the Science History Institute in Philadelphia, Pennsylvania.

History edit

 
Diagram of electrospray ionization in positive mode: under high voltage, the Taylor cone emits a jet of liquid drops. The solvent from the droplets progressively evaporates, leaving them more and more charged. When the charge exceeds the Rayleigh limit the droplet explosively dissociates, leaving a stream of charged (positive) ions

In 1882, Lord Rayleigh theoretically estimated the maximum amount of charge a liquid droplet could carry before throwing out fine jets of liquid.[7] This is now known as the Rayleigh limit.

In 1914, John Zeleny published work on the behaviour of fluid droplets at the end of glass capillaries and presented evidence for different electrospray modes.[8] Wilson and Taylor[9] and Nolan investigated electrospray in the 1920s[10] and Macky in 1931.[11] The electrospray cone (now known as the Taylor cone) was described by Sir Geoffrey Ingram Taylor.[12]

The first use of electrospray ionization with mass spectrometry was reported by Malcolm Dole in 1968.[13][14] John Bennett Fenn was awarded the 2002 Nobel Prize in Chemistry for the development of electrospray ionization mass spectrometry in the late 1980s.[15]

Ionization mechanism edit

 
Fenn's first electrospray ionization source coupled to a single quadrupole mass spectrometer

The liquid containing the analytes of interest (typically 10-6 - 10-4 M needed [16]) is dispersed by electrospray,[17] into a fine aerosol. Because the ion formation involves extensive solvent evaporation (also termed desolvation), the typical solvents for electrospray ionization are prepared by mixing water with volatile organic compounds (e.g. methanol[18] acetonitrile). To decrease the initial droplet size, compounds that increase the conductivity (e.g. acetic acid) are customarily added to the solution. These species also act to provide a source of protons to facilitate the ionization process. Large-flow electrosprays can benefit from nebulization of a heated inert gas such as nitrogen or carbon dioxide in addition to the high temperature of the ESI source.[19] The aerosol is sampled into the first vacuum stage of a mass spectrometer through a capillary carrying a potential difference of approximately 3000 V, which can be heated to aid further solvent evaporation from the charged droplets. The solvent evaporates from a charged droplet until it becomes unstable upon reaching its Rayleigh limit. At this point, the droplet deforms as the electrostatic repulsion of like charges, in an ever-decreasing droplet size, becomes more powerful than the surface tension holding the droplet together.[20] At this point the droplet undergoes Coulomb fission, whereby the original droplet 'explodes' creating many smaller, more stable droplets. The new droplets undergo desolvation and subsequently further Coulomb fissions. During the fission, the droplet loses a small percentage of its mass (1.0–2.3%) along with a relatively large percentage of its charge (10–18%).[21][22]

There are two major theories that explain the final production of gas-phase ions: the ion evaporation model (IEM) and the charge residue model (CRM). The IEM suggests that as the droplet reaches a certain radius the field strength at the surface of the droplet becomes large enough to assist the field desorption of solvated ions.[23][24] The CRM suggests that electrospray droplets undergo evaporation and fission cycles, eventually leading progeny droplets that contain on average one analyte ion or less.[13] The gas-phase ions form after the remaining solvent molecules evaporate, leaving the analyte with the charges that the droplet carried.

 
IEM, CRM and CEM schematic.

A large body of evidence shows either directly or indirectly that small ions (from small molecules) are liberated into the gas phase through the ion evaporation mechanism,[24][25][citation needed][26] while larger ions (from folded proteins for instance) form by charged residue mechanism.[27][28][29]

A third model invoking combined charged residue-field emission has been proposed.[30] Another model called chain ejection model (CEM) is proposed for disordered polymers (unfolded proteins).[31]

The ions observed by mass spectrometry may be quasimolecular ions created by the addition of a hydrogen cation and denoted [M + H]+, or of another cation such as sodium ion, [M + Na]+, or the removal of a hydrogen nucleus, [M − H]. Multiply charged ions such as [M + nH]n+ are often observed. For large macromolecules, there can be many charge states, resulting in a characteristic charge state envelope. All these are even-electron ion species: electrons (alone) are not added or removed, unlike in some other ionization sources. The analytes are sometimes involved in electrochemical processes, leading to shifts of the corresponding peaks in the mass spectrum. This effect is demonstrated in the direct ionization of noble metals such as copper, silver and gold using electrospray.[32]

The efficiency of generating the gas phase ions for small molecules in ESI varies depending on the compound structure, the solvent used and instrumental parameters.[33] The differences in ionization efficiency reach more than 1 million times.

Variants edit

The electrosprays operated at low flow rates generate much smaller initial droplets, which ensure improved ionization efficiency. In 1993 Gale and Richard D. Smith reported significant sensitivity increases could be achieved using lower flow rates, and down to 200 nL/min.[34] In 1994, two research groups coined the name micro-electrospray (microspray) for electrosprays working at low flow rates. Emmett and Caprioli demonstrated improved performance for HPLC-MS analyses when the electrospray was operated at 300–800 nL/min.[35] Wilm and Mann demonstrated that a capillary flow of ~ 25 nL/min can sustain an electrospray at the tip of emitters fabricated by pulling glass capillaries to a few micrometers.[36] The latter was renamed nano-electrospray (nanospray) in 1996.[37][38] Currently the name nanospray is also in use for electrosprays fed by pumps at low flow rates,[39] not only for self-fed electrosprays. Although there may not be a well-defined flow rate range for electrospray, microspray, and nano-electrospray,[40] studied "changes in analyte partition during droplet fission prior to ion release".[40] In this paper, they compare results obtained by three other groups.[41][42][43] and then measure the signal intensity ratio [Ba2+ + Ba+]/[BaBr+] at different flow rates.

Cold spray ionization is a form of electrospray in which the solution containing the sample is forced through a small cold capillary (10–80 °C) into an electric field to create a fine mist of cold charged droplets.[44] Applications of this method include the analysis of fragile molecules and guest-host interactions that cannot be studied using regular electrospray ionization.

Electrospray ionization has also been achieved at pressures as low as 25 torr and termed subambient pressure ionization with nanoelectrospray (SPIN) based upon a two-stage ion funnel interface developed by Richard D. Smith and coworkers.[45] The SPIN implementation provided increased sensitivity due to the use of ion funnels that helped confine and transfer ions to the lower pressure region of the mass spectrometer. Nanoelectrospray emitter is made out of a fine capillary with a small aperture about 1–3 micrometer. For sufficient conductivity this capillary is usually sputter-coated with conductive material, e.g. gold. Nanoelectrospray ionization consumes only a few microliters of a sample and forms smaller droplets.[46] Operation at low pressure was particularly effective for low flow rates where the smaller electrospray droplet size allowed effective desolvation and ion formation to be achieved. As a result, the researchers were later able to demonstrate achieving an excess of 50% overall ionization utilization efficiency for transfer of ions from the liquid phase, into the gas phase as ions, and through the dual ion funnel interface to the mass spectrometer.[47]

Ambient ionization edit

 
Diagram of a DESI ambient ionization source

In ambient ionization, the formation of ions occurs outside the mass spectrometer without sample preparation.[48][49][50] Electrospray is used for ion formation in a number of ambient ion sources.

Desorption electrospray ionization (DESI) is an ambient ionization technique in which a solvent electrospray is directed at a sample.[51][52] The electrospray is attracted to the surface by applying a voltage to the sample. Sample compounds are extracted into the solvent which is again aerosolized as highly charged droplets that evaporate to form highly charged ions. After ionization, the ions enter the atmospheric pressure interface of the mass spectrometer. DESI allows for ambient ionization of samples at atmospheric pressure, with little sample preparation.

 
Diagram of a SESI ambient ionization source

Extractive electrospray ionization is a spray-type, ambient ionization method that uses two merged sprays, one of which is generated by electrospray.[49]

Laser-based electrospray-based ambient ionization is a two-step process in which a pulsed laser is used to desorb or ablate material from a sample and the plume of material interacts with an electrospray to create ions.[49] For ambient ionization, the sample material is deposited on a target near the electrospray. The laser desorbs or ablates material from the sample which is ejected from the surface and into the electrospray which produces highly charged ions. Examples are electrospray laser desorption ionization, matrix-assisted laser desorption electrospray ionization, and laser ablation electrospray ionization.

 
SESI-MS SUPER SESI coupled with Thermo Fisher Scientific-Orbitrap

Electrostatic spray ionization (ESTASI) involved the analysis of samples located on a flat or porous surface, or inside a microchannel. A droplet containing analytes is deposited on a sample area, to which a pulsed high voltage to is applied. When the electrostatic pressure is larger than the surface tension, droplets and ions are sprayed.

Secondary electrospray ionization (SESI) is an spray type, ambient ionization method where charging ions are produced by means of an electrospray. These ions then charge vapor molecules in the gas phase when colliding with them.[53][54]

In paper spray ionization, the sample is applied to a piece of paper, solvent is added, and a high voltage is applied to the paper, creating ions.

Applications edit

 
The outside of the electrospray interface on an LTQ mass spectrometer.

Electrospray is used to study protein folding.[55][56][57]

Liquid chromatography–mass spectrometry edit

Electrospray ionization is the ion source of choice to couple liquid chromatography with mass spectrometry (LC-MS). The analysis can be performed online, by feeding the liquid eluting from the LC column directly to an electrospray, or offline, by collecting fractions to be later analyzed in a classical nanoelectrospray-mass spectrometry setup. Among the numerous operating parameters in ESI-MS,for proteins,[58] the electrospray voltage has been identified as an important parameter to consider in ESI LC/MS gradient elution.[59] The effect of various solvent compositions[60] (such as TFA[61] or ammonium acetate,[22] or supercharging reagents,[62][63][64][65] or derivitizing groups[66]) or spraying conditions[67] on electrospray-LCMS spectra and/or nanoESI-MS spectra.[68] have been studied.

Capillary electrophoresis-mass spectrometry (CE-MS) edit

Capillary electrophoresis-mass spectrometry was enabled by an ESI interface that was developed and patented by Richard D. Smith and coworkers at Pacific Northwest National Laboratory, and shown to have broad utility for the analysis of very small biological and chemical compound mixtures, and even extending to a single biological cell.

Noncovalent gas phase interactions edit

Electrospray ionization is also utilized in studying noncovalent gas phase interactions. The electrospray process is thought to be capable of transferring liquid-phase noncovalent complexes into the gas phase without disrupting the noncovalent interaction. Problems[22][69] such as non specific interactions[70] have been identified when studying ligand substrate complexes by ESI-MS or nanoESI-MS. An interesting example of this is studying the interactions between enzymes and drugs which are inhibitors of the enzyme.[71][72][73] Competition studies between STAT6 and inhibitors[73][74][75] have used ESI as a way to screen for potential new drug candidates.

Electrospray ionization can even be used for studying protein complexes >1 MDa. [76][16]

See also edit

References edit

  1. ^ Ho, CS; Chan MHM; Cheung RCK; Law LK; Lit LCW; Ng KF; Suen MWM; Tai HL (February 2003). "Electrospray Ionisation Mass Spectrometry: Principles and Clinical Applications". Clin Biochem Rev. 24 (1): 3–12. PMC 1853331. PMID 18568044.
  2. ^ Pitt, James J (February 2009). "Principles and Applications of Liquid Chromatography-Mass Spectrometry in Clinical Biochemistry". Clin Biochem Rev. 30 (1): 19–34. PMC 2643089. PMID 19224008.
  3. ^ Yamashita, Masamichi; Fenn, John B. (September 1984). "Electrospray ion source. Another variation on the free-jet theme". The Journal of Physical Chemistry. 88 (20): 4451–4459. doi:10.1021/j150664a002.
  4. ^ a b Alexandrov, M. L.; Gall, L. N.; Krasnov, N. V.; Nikolaev, V. I.; Pavlenko, V. A.; Shkruov, V. A. (1984). "Extraction of ions from solutions under atmospheric pressure as a method for mass spectrometric analysis of bioorganic compounds". Doklady Akad. SSSR. 277 (3): 379–383. Bibcode:2008RCMS...22..267A. doi:10.1002/rcm.3113. PMID 18181250.
  5. ^ Fenn, J. B.; Mann, M.; Meng, C. K.; Wong, S. F.; Whitehouse, C. M. (1989). "Electrospray ionization for mass spectrometry of large biomolecules". Science. 246 (4926): 64–71. Bibcode:1989Sci...246...64F. CiteSeerX 10.1.1.522.9458. doi:10.1126/science.2675315. PMID 2675315.
  6. ^ Markides, K; Gräslund, A. "Advanced information on the Nobel Prize in Chemistry 2002" (PDF).
  7. ^ Rayleigh, L. (1882). "On the Equilibrium of Liquid Conducting Masses charged with Electricity". Philosophical Magazine. 14 (87): 184–186. doi:10.1080/14786448208628425.
  8. ^ Zeleny, J. (1914). "The electrical discharge from liquid points, and a hydrostatic method of measuring the electric intensity at their surfaces". Physical Review. 3 (2): 69–91. Bibcode:1914PhRv....3...69Z. doi:10.1103/PhysRev.3.69.
  9. ^ Wilson, C. T.; G. I Taylor (1925). "The bursting of soap bubbles in a uniform electric field". Proc. Cambridge Philos. Soc. 22 (5): 728. Bibcode:1925PCPS...22..728W. doi:10.1017/S0305004100009609. S2CID 137905700.
  10. ^ Nolan, J. J. (1926). "Universal scaling laws for the disintegration of electrified drops". Proc. R. Ir. Acad. A. 37: 28.
  11. ^ Macky, W. A. (October 1, 1931). "Some Investigations on the Deformation and Breaking of Water Drops in Strong Electric Fields". Proceedings of the Royal Society A. 133 (822): 565–587. Bibcode:1931RSPSA.133..565M. doi:10.1098/rspa.1931.0168.
  12. ^ Geoffrey Taylor (1964). "Disintegration of Water Droplets in an Electric Field". Proceedings of the Royal Society A. 280 (1382): 383–397. Bibcode:1964RSPSA.280..383T. doi:10.1098/rspa.1964.0151. JSTOR 2415876. S2CID 15067908.
  13. ^ a b Dole M, Mack LL, Hines RL, Mobley RC, Ferguson LD, Alice MB (1968). "Molecular Beams of Macroions". Journal of Chemical Physics. 49 (5): 2240–2249. Bibcode:1968JChPh..49.2240D. doi:10.1063/1.1670391.
  14. ^ Birendra N. Pramanik; A.K. Ganguly; Michael L. Gross (28 February 2002). Applied Electrospray Mass Spectrometry: Practical Spectroscopy Series. CRC Press. pp. 4–. ISBN 978-0-8247-4419-9.
  15. ^ "Press Release: The Nobel Prize in Chemistry 2002". The Nobel Foundation. 2002-10-09. Retrieved 2011-04-02.
  16. ^ a b Gross, Jürgen H. (2017), "Electrospray Ionization", Mass Spectrometry, Cham: Springer International Publishing, pp. 721–778, doi:10.1007/978-3-319-54398-7_12, ISBN 978-3-319-54397-0, retrieved 2024-03-15
  17. ^ Pozniak BP, Cole RB (2007). "Current Measurements within the Electrospray Emitter". J. Am. Soc. Mass Spectrom. 18 (4): 737–748. doi:10.1016/j.jasms.2006.11.012. PMID 17257852.
  18. ^ Olumee; et al. (1998). "Droplet Dynamics Changes in Electrostatic Sprays of Methanol-Water Mixtures". J. Phys. Chem. A. 102 (46): 9154–9160. Bibcode:1998JPCA..102.9154O. CiteSeerX 10.1.1.661.5000. doi:10.1021/jp982027z.
  19. ^ Fernández De La Mora J (2007). "The Fluid Dynamics of Taylor Cones". Annual Review of Fluid Mechanics. 39 (1): 217–243. Bibcode:2007AnRFM..39..217F. doi:10.1146/annurev.fluid.39.050905.110159.
  20. ^ Cole, Richard B (2010). Electrospray and MALDI Mass Spectrometry: Fundamentals, Instrumentation, Practicalities, and Biological Applications (2 ed.). Wiley. p. 4. ISBN 978-0471741077.
  21. ^ Li KY, Tu H, Ray AK (April 2005). "Charge limits on droplets during evaporation". Langmuir. 21 (9): 3786–94. doi:10.1021/la047973n. PMID 15835938.
  22. ^ a b c Kebarle P, Verkerk UH (2009). "Electrospray: from ions in solution to ions in the gas phase, what we know now". Mass Spectrom Rev. 28 (6): 898–917. Bibcode:2009MSRv...28..898K. doi:10.1002/mas.20247. PMID 19551695.
  23. ^ Iribarne JV, Thomson BA (1976). "On the evaporation of small ions from charged droplets". Journal of Chemical Physics. 64 (6): 2287–2294. Bibcode:1976JChPh..64.2287I. doi:10.1063/1.432536.
  24. ^ a b Nguyen S, Fenn JB (January 2007). "Gas-phase ions of solute species from charged droplets of solutions". Proc. Natl. Acad. Sci. USA. 104 (4): 1111–7. Bibcode:2007PNAS..104.1111N. doi:10.1073/pnas.0609969104. PMC 1783130. PMID 17213314.
  25. ^ Gamero-Castaño M (2000). "Direct measurement of ion evaporation kinetics from electrified liquid surfaces". J. Chem. Phys. 113 (2): 815. Bibcode:2000JChPh.113..815G. doi:10.1063/1.481857. S2CID 36112510.
  26. ^ de la Mora Fernandez (2000). "Electrospray ionization of large multiply charged species proceeds via Dole's charged residue mechanism". Analytica Chimica Acta. 406: 93–104. doi:10.1016/S0003-2670(99)00601-7. An evaluation of the electric field on the drop surface at the point when it just ceases to be spherical (yet carries the total ion charge z) indicates that small PEG ions may be formed by ion evaporation. The break observed in the charge distribution may perhaps mean that the shift from the Dole to the ion evaporation mechanism arises at m(unintelligible)104[clarification needed], though this inference is highly hypothetical.
  27. ^ de la Mora Fernandez (2000). "Electrospray ionization of large multiply charged species proceeds via Dole's charged residue mechanism". Analytica Chimica Acta. 406: 93–104. doi:10.1016/S0003-2670(99)00601-7.
  28. ^ de la Mora Fernandez (2000). "Electrospray ionization of large multiply charged species proceeds via Dole's charged residue mechanism". Analytica Chimica Acta. 406: 93–104. doi:10.1016/S0003-2670(99)00601-7. For most published data examined, zmax is between 65% and 110% of zR, providing strong support in favor of Dole's charged residue mechanism, at least for masses from 3.3 kD up to 1.4 MD. Other large but less compact ions from proteins and linear chains of polyethylene glycols (PEGs) have zmax values considerably larger than zR, apparently implying that they also formas charged residues, though from non-spherical drops held together by the polymer backbone.
  29. ^ de la Mora Fernandez (2000). "Electrospray ionization of large multiply charged species proceeds via Dole's charged residue mechanism". Analytica Chimica Acta. 406: 93–104. doi:10.1016/S0003-2670(99)00601-7. The data do show a nearly discontinuous jump in the observed m/z for a mass somewhere between 20,000 and 50,000, and it is tempting to conclude that this is due to a corresponding transition where the ionization mechanism shifts from one type to the other. This would correspond to a critical value of z in the vicinity of 50, with a corresponding electric field of 2.6 V/nm. Of course, this is entirely hypothetical, and there is yet no compelling evidence of any kind indicating that an ion with as many as 30 charges can be formed by field evaporation.
  30. ^ Hogan CJ, Carroll JA, Rohrs HW, Biswas P, Gross ML (January 2009). "Combined charged residue-field emission model of macromolecular electrospray ionization". Anal. Chem. 81 (1): 369–77. doi:10.1021/ac8016532. PMC 2613577. PMID 19117463.
  31. ^ Konermann, Lars (2013). "Unraveling the Mechanism of Electrospray Ionization". Analytical Chemistry. 85 (1): 2–9. doi:10.1021/ac302789c. PMID 23134552.
  32. ^ Li, Anyin; Luo, Qingjie; Park, So-Jung; Cooks, R. Graham (2014). "Synthesis and Catalytic Reactions of Nanoparticles formed by Electrospray Ionization of Coinage Metals". Angewandte Chemie International Edition. 53 (12): 3147–3150. doi:10.1002/anie.201309193. ISSN 1433-7851. PMID 24554582.
  33. ^ Kruve, Anneli; Kaupmees, Karl; Liigand, Jaanus; Leito, Ivo (2014). "Negative Electrospray Ionization via Deprotonation: Predicting the Ionization Efficiency". Analytical Chemistry. 86 (10): 4822–4830. doi:10.1021/ac404066v. PMID 24731109.
  34. ^ Gale DC, Smith RD (1993). "Small Volume and Low Flow Rate Electrospray Ionization Mass Spectrometry for Aqueous Samples". Rapid Commun. Mass Spectrom. 7 (11): 1017–1021. Bibcode:1993RCMS....7.1017G. doi:10.1002/rcm.1290071111.
  35. ^ Emmett MR, Caprioli RM (1994). "Micro-electrospray mass spectrometry: ultra-high-sensitivity analysis of peptides and proteins". J. Am. Soc. Mass Spectrom. 5 (7): 605–613. doi:10.1016/1044-0305(94)85001-1. PMID 24221962.
  36. ^ Wilm MS, Mann M (1994). "Electrospray and Taylor-Cone theory, Dole's beam of macromolecules at last?". Int. J. Mass Spectrom. Ion Process. 136 (2–3): 167–180. Bibcode:1994IJMSI.136..167W. doi:10.1016/0168-1176(94)04024-9.
  37. ^ Wilm M, Mann M (1996). "Analytical properties of the nanoelectrospray ion source". Anal. Chem. 68 (1): 1–8. doi:10.1021/ac9509519. PMID 8779426.
  38. ^ Gibson; Mugo, Samuel M.; Oleschuk, Richard D.; et al. (2009). "Nanoelectrospray emitters: Trends and perspective". Mass Spectrometry Reviews. 28 (6): 918–936. Bibcode:2009MSRv...28..918G. doi:10.1002/mas.20248. PMID 19479726.
  39. ^ Page JS, Marginean I, Baker ES, Kelly RT, Tang K, Smith RD (December 2009). "Biases in ion transmission through an electrospray ionization-mass spectrometry capillary inlet". J. Am. Soc. Mass Spectrom. 20 (12): 2265–72. doi:10.1016/j.jasms.2009.08.018. PMC 2861838. PMID 19815425.
  40. ^ a b Schmidt A, Karas M, Dülcks T (May 2003). "Effect of different solution flow rates on analyte ion signals in nano-ESI MS, or: when does ESI turn into nano-ESI?". J. Am. Soc. Mass Spectrom. 14 (5): 492–500. doi:10.1016/S1044-0305(03)00128-4. PMID 12745218.
  41. ^ Wilm M. S.; Mann M. (1994). "Electrospray and Taylor-Cone Theory, Dole's Beam of Macromolecules at Last?". Int. J. Mass Spectrom. Ion Process. 136 (2–3): 167–180. Bibcode:1994IJMSI.136..167W. doi:10.1016/0168-1176(94)04024-9.
  42. ^ Fernandez de la Mora J., Loscertales I. G. (2006). "The Current Emitted by Highly Conducting Taylor Cones". J. Fluid Mech. 260: 155–184. Bibcode:1994JFM...260..155D. doi:10.1017/S0022112094003472. S2CID 122935117.
  43. ^ Pfeifer RJ, Hendricks (1968). "Parametric Studies of Electrohydrodynamic Spraying". AIAA J. 6 (3): 496–502. Bibcode:1968AIAAJ...6..496H. doi:10.2514/3.4525.
  44. ^ RSC Chemical Methods Ontology, Cold-spray ionisation mass spectrometry
  45. ^ Page JS, Tang K, Kelly RT, Smith RD (2008). "A subambient pressure ionization with nanoelectrospray (SPIN) source and interface for improved sensitivity in mass spectrometry". Analytical Chemistry. 80 (5): 1800–1805. doi:10.1021/ac702354b. PMC 2516344. PMID 18237189.
  46. ^ Karas, M.; Bahr, U.; Dülcks, T. (2000-03-01). "Nano-electrospray ionization mass spectrometry: addressing analytical problems beyond routine". Fresenius' Journal of Analytical Chemistry. 366 (6–7): 669–676. doi:10.1007/s002160051561. ISSN 0937-0633. PMID 11225778. S2CID 24730378.
  47. ^ I. Marginean; J. S. Page; A. V. Tolmachev; K. Tang; R. D. Smith (2010). "Achieving 50% Ionization Efficiency in Subambient Pressure Ionization with Nanoelectrospray". Analytical Chemistry. 82 (22): 9344–9349. doi:10.1021/ac1019123. PMC 2982749. PMID 21028835.
  48. ^ Cooks, R. Graham; Ouyang, Zheng; Takats, Zoltan; Wiseman, Justin M. (2006). "Ambient Mass Spectrometry". Science. 311 (5767): 1566–70. Bibcode:2006Sci...311.1566C. doi:10.1126/science.1119426. PMID 16543450. S2CID 98131681.
  49. ^ a b c Monge, María Eugenia; Harris, Glenn A.; Dwivedi, Prabha; Fernández, Facundo M. (2013). "Mass Spectrometry: Recent Advances in Direct Open Air Surface Sampling/Ionization". Chemical Reviews. 113 (4): 2269–2308. doi:10.1021/cr300309q. ISSN 0009-2665. PMID 23301684.
  50. ^ Huang, Min-Zong; Yuan, Cheng-Hui; Cheng, Sy-Chyi; Cho, Yi-Tzu; Shiea, Jentaie (2010). "Ambient Ionization Mass Spectrometry". Annual Review of Analytical Chemistry. 3 (1): 43–65. Bibcode:2010ARAC....3...43H. doi:10.1146/annurev.anchem.111808.073702. ISSN 1936-1327. PMID 20636033.
  51. ^ Z. Takáts; J.M. Wiseman; B. Gologan; R.G. Cooks (2004). "Mass Spectrometry Sampling Under Ambient Conditions with Desorption Electrospray Ionization". Science. 306 (5695): 471–473. Bibcode:2004Sci...306..471T. doi:10.1126/science.1104404. PMID 15486296. S2CID 22994482.
  52. ^ Takáts Z, Wiseman JM, Cooks RG (2005). "Ambient mass spectrometry using desorption electrospray ionization (DESI): instrumentation, mechanisms and applications in forensics, chemistry, and biology". Journal of Mass Spectrometry. 40 (10): 1261–75. Bibcode:2005JMSp...40.1261T. doi:10.1002/jms.922. PMID 16237663.
  53. ^ Vidal-de-Miguel, G.; Macía, M.; Pinacho, P.; Blanco, J. (2012-10-16). "Low-Sample Flow Secondary Electrospray Ionization: Improving Vapor Ionization Efficiency". Analytical Chemistry. 84 (20): 8475–8479. doi:10.1021/ac3005378. ISSN 0003-2700. PMID 22970991.
  54. ^ Barrios-Collado, César; Vidal-de-Miguel, Guillermo; Martinez-Lozano Sinues, Pablo (February 2016). "Numerical modeling and experimental validation of a universal secondary electrospray ionization source for mass spectrometric gas analysis in real-time". Sensors and Actuators B: Chemical. 223: 217–225. doi:10.1016/j.snb.2015.09.073. hdl:20.500.11850/105470.
  55. ^ Konermann, L; Douglas, DJ (1998). "Equilibrium unfolding of proteins monitored by electrospray ionization mass spectrometry: Distinguishing two-state from multi-state transitions". Rapid Communications in Mass Spectrometry. 12 (8): 435–442. Bibcode:1998RCMS...12..435K. doi:10.1002/(SICI)1097-0231(19980430)12:8<435::AID-RCM181>3.0.CO;2-F. PMID 9586231.
  56. ^ Nemes; Goyal, Samita; Vertes, Akos; et al. (2008). "Conformational and Noncovalent Complexation Changes in Proteins during Electrospray Ionization". Analytical Chemistry. 80 (2): 387–395. doi:10.1021/ac0714359. PMID 18081323.
  57. ^ Sobott; Robinson (2004). "Characterising electrosprayed biomolecules using tandem-MS—the noncovalent GroEL chaperonin assembly". International Journal of Mass Spectrometry. 236 (1–3): 25–32. Bibcode:2004IJMSp.236...25S. doi:10.1016/j.ijms.2004.05.010.
  58. ^ Vaidyanathan S.; Kell D.B.; Goodacre R. (2004). "Selective detection of proteins in mixtures using electrospray ionization mass spectrometry: influence of instrumental settings and implications for proteomics". Analytical Chemistry. 76 (17): 5024–5032. doi:10.1021/ac049684+. PMID 15373437.
  59. ^ Marginean I, Kelly RT, Moore RJ, Prior DC, LaMarche BL, Tang K, Smith RD (April 2009). "Selection of the optimum electrospray voltage for gradient elution LC-MS measurements". J. Am. Soc. Mass Spectrom. 20 (4): 682–8. doi:10.1016/j.jasms.2008.12.004. PMC 2692488. PMID 19196520.
  60. ^ Iavarone; Jurchen, John C.; Williams, Evan R.; et al. (2000). "Effects of solvent on the maximum charge state and charge state distribution of protein ions produced by electrospray ionization". J. Am. Soc. Mass Spectrom. 11 (11): 976–985. doi:10.1016/S1044-0305(00)00169-0. PMC 1414794. PMID 11073261.
  61. ^ Garcia (2005). "The effect of the mobile phase additives on sensitivity in the analysis of peptides and proteins by high-performance liquid chromatography–electrospray mass spectrometry". Journal of Chromatography B. 825 (2): 111–123. doi:10.1016/j.jchromb.2005.03.041. PMID 16213445.
  62. ^ Teo CA, Donald WA (May 2014). "Solution additives for supercharging proteins beyond the theoretical maximum proton-transfer limit in electrospray ionization mass spectrometry". Anal. Chem. 86 (9): 4455–62. doi:10.1021/ac500304r. PMID 24712886.
  63. ^ Lomeli SH, Peng IX, Yin S, Loo RR, Loo JA (January 2010). "New reagents for increasing ESI multiple charging of proteins and protein complexes". J. Am. Soc. Mass Spectrom. 21 (1): 127–31. doi:10.1016/j.jasms.2009.09.014. PMC 2821426. PMID 19854660.
  64. ^ Lomeli SH, Yin S, Ogorzalek Loo RR, Loo JA (April 2009). "Increasing charge while preserving noncovalent protein complexes for ESI-MS". J. Am. Soc. Mass Spectrom. 20 (4): 593–6. doi:10.1016/j.jasms.2008.11.013. PMC 2789282. PMID 19101165.
  65. ^ Yin S, Loo JA (March 2011). "Top-Down Mass Spectrometry of Supercharged Native Protein-Ligand Complexes". Int J Mass Spectrom. 300 (2–3): 118–122. Bibcode:2011IJMSp.300..118Y. doi:10.1016/j.ijms.2010.06.032. PMC 3076692. PMID 21499519.
  66. ^ Krusemark CJ, Frey BL, Belshaw PJ, Smith LM (September 2009). "Modifying the charge state distribution of proteins in electrospray ionization mass spectrometry by chemical derivatization". J. Am. Soc. Mass Spectrom. 20 (9): 1617–25. doi:10.1016/j.jasms.2009.04.017. PMC 2776692. PMID 19481956.
  67. ^ Nemes P, Goyal S, Vertes A (January 2008). "Conformational and noncovalent complexation changes in proteins during electrospray ionization". Anal. Chem. 80 (2): 387–95. doi:10.1021/ac0714359. PMID 18081323.
  68. ^ Ramanathan R, Zhong R, Blumenkrantz N, Chowdhury SK, Alton KB (October 2007). "Response normalized liquid chromatography nanospray ionization mass spectrometry". J. Am. Soc. Mass Spectrom. 18 (10): 1891–9. doi:10.1016/j.jasms.2007.07.022. PMID 17766144.
  69. ^ Gabelica V, Vreuls C, Filée P, Duval V, Joris B, Pauw ED (2002). "Advantages and drawbacks of nanospray for studying noncovalent protein-DNA complexes by mass spectrometry". Rapid Commun. Mass Spectrom. 16 (18): 1723–8. Bibcode:2002RCMS...16.1723G. doi:10.1002/rcm.776. hdl:2268/322. PMID 12207359.
  70. ^ Daubenfeld T, Bouin AP, van der Rest G (September 2006). "A deconvolution method for the separation of specific versus nonspecific interactions in noncovalent protein-ligand complexes analyzed by ESI-FT-ICR mass spectrometry". J. Am. Soc. Mass Spectrom. 17 (9): 1239–48. doi:10.1016/j.jasms.2006.05.005. PMID 16793278.
  71. ^ Rosu F, De Pauw E, Gabelica V (July 2008). "Electrospray mass spectrometry to study drug-nucleic acids interactions". Biochimie. 90 (7): 1074–87. doi:10.1016/j.biochi.2008.01.005. PMID 18261993.
  72. ^ Wortmann A, Jecklin MC, Touboul D, Badertscher M, Zenobi R (May 2008). "Binding constant determination of high-affinity protein-ligand complexes by electrospray ionization mass spectrometry and ligand competition". J Mass Spectrom. 43 (5): 600–8. Bibcode:2008JMSp...43..600W. doi:10.1002/jms.1355. PMID 18074334.
  73. ^ a b Jecklin MC, Touboul D, Bovet C, Wortmann A, Zenobi R (March 2008). "Which electrospray-based ionization method best reflects protein-ligand interactions found in solution? a comparison of ESI, nanoESI, and ESSI for the determination of dissociation constants with mass spectrometry". J. Am. Soc. Mass Spectrom. 19 (3): 332–43. doi:10.1016/j.jasms.2007.11.007. hdl:20.500.11850/9214. PMID 18083584.
  74. ^ Touboul D, Maillard L, Grässlin A, Moumne R, Seitz M, Robinson J, Zenobi R (February 2009). "How to deal with weak interactions in noncovalent complexes analyzed by electrospray mass spectrometry: cyclopeptidic inhibitors of the nuclear receptor coactivator 1-STAT6". J. Am. Soc. Mass Spectrom. 20 (2): 303–11. doi:10.1016/j.jasms.2008.10.008. hdl:20.500.11850/15377. PMID 18996720.
  75. ^ Czuczy N, Katona M, Takats Z (February 2009). "Selective detection of specific protein-ligand complexes by electrosonic spray-precursor ion scan tandem mass spectrometry". J. Am. Soc. Mass Spectrom. 20 (2): 227–37. doi:10.1016/j.jasms.2008.09.010. PMID 18976932.
  76. ^ Ishii, Kentaro; Zhou, Min; Uchiyama, Susumu (2018-02-01). "Native mass spectrometry for understanding dynamic protein complex". Biochimica et Biophysica Acta (BBA) - General Subjects. Biophysical Exploration of Dynamical Ordering of Biomolecular Systems. 1862 (2): 275–286. doi:10.1016/j.bbagen.2017.09.019. ISSN 0304-4165.

Further reading edit

  • Cole, Richard (1997). Electrospray ionization mass spectrometry: fundamentals, instrumentation, and applications. New York: Wiley. ISBN 978-0-471-14564-6.
  • Gross, Michael; Pramanik, Birendra N.; Ganguly, A. K. (2002). Applied electrospray mass spectrometry. New York, N.Y: Marcel Dekker. ISBN 978-0-8247-0618-0.
  • Snyder, A. Peter (1996). Biochemical and biotechnological applications of electrospray ionization mass spectrometry. Columbus, OH: American Chemical Society. ISBN 978-0-8412-3378-2.
  • Alexandrov, M. L.; L. N. Gall; N. V. Krasnov; V. I. Nikolaev; V. A. Pavlenko; V. A. Shkurov (July 1984). Экстракция ионов из растворов при атмосферном давлении – Метод масс-спектрометрического анализа биоорганических веществ [Extraction of ions from solutions at atmospheric pressure – A method for mass spectrometric analysis of bioorganic substances]. Doklady Akademii Nauk SSSR (in Russian). 277 (2): 379–383.
  • Alexandrov, M. L.; L. N. Gall; N. V. Krasnov; V. I. Nikolaev; V. A. Pavlenko; V. A. Shkurov (2008) [July 1984]. "Extraction of ions from solutions under atmospheric pressure as a method for mass spectrometric analysis of bioorganic compounds". Rapid Communications in Mass Spectrometry. 22 (3): 267–270. Bibcode:2008RCMS...22..267A. doi:10.1002/rcm.3113. PMID 18181250.

External links edit

  • National High Magnetic Field Laboratory
  • Electrospray+Ionization+Mass+Spectrometry at the U.S. National Library of Medicine Medical Subject Headings (MeSH)

electrospray, ionization, technique, used, mass, spectrometry, produce, ions, using, electrospray, which, high, voltage, applied, liquid, create, aerosol, especially, useful, producing, ions, from, macromolecules, because, overcomes, propensity, these, molecul. Electrospray ionization ESI is a technique used in mass spectrometry to produce ions using an electrospray in which a high voltage is applied to a liquid to create an aerosol It is especially useful in producing ions from macromolecules because it overcomes the propensity of these molecules to fragment when ionized ESI is different from other ionization processes e g matrix assisted laser desorption ionization MALDI since it may produce multiple charged ions effectively extending the mass range of the analyser to accommodate the kDa MDa orders of magnitude observed in proteins and their associated polypeptide fragments 1 2 Electrospray nanoSpray ionization source Mass spectrometry using ESI is called electrospray ionization mass spectrometry ESI MS or less commonly electrospray mass spectrometry ES MS ESI is a so called soft ionization technique since there is very little fragmentation This can be advantageous in the sense that the molecular ion or more accurately a pseudo molecular ion is almost always observed however very little structural information can be gained from the simple mass spectrum obtained This disadvantage can be overcome by coupling ESI with tandem mass spectrometry ESI MS MS Another important advantage of ESI is that solution phase information can be retained into the gas phase The electrospray ionization technique was first reported by Masamichi Yamashita and John Fenn in 1984 3 and independently by Lidia Gall and co workers in Soviet Union also in 1984 4 Gall s work was not recognised or translated in the western scientific literature until a translation was published in 2008 4 The development of electrospray ionization for the analysis of biological macromolecules 5 was rewarded with the attribution of the Nobel Prize in Chemistry to John Bennett Fenn and Koichi Tanaka in 2002 6 One of the original instruments used by Fenn is on display at the Science History Institute in Philadelphia Pennsylvania Contents 1 History 2 Ionization mechanism 3 Variants 3 1 Ambient ionization 4 Applications 4 1 Liquid chromatography mass spectrometry 4 2 Capillary electrophoresis mass spectrometry CE MS 4 3 Noncovalent gas phase interactions 5 See also 6 References 7 Further reading 8 External linksHistory edit nbsp Diagram of electrospray ionization in positive mode under high voltage the Taylor cone emits a jet of liquid drops The solvent from the droplets progressively evaporates leaving them more and more charged When the charge exceeds the Rayleigh limit the droplet explosively dissociates leaving a stream of charged positive ions In 1882 Lord Rayleigh theoretically estimated the maximum amount of charge a liquid droplet could carry before throwing out fine jets of liquid 7 This is now known as the Rayleigh limit In 1914 John Zeleny published work on the behaviour of fluid droplets at the end of glass capillaries and presented evidence for different electrospray modes 8 Wilson and Taylor 9 and Nolan investigated electrospray in the 1920s 10 and Macky in 1931 11 The electrospray cone now known as the Taylor cone was described by Sir Geoffrey Ingram Taylor 12 The first use of electrospray ionization with mass spectrometry was reported by Malcolm Dole in 1968 13 14 John Bennett Fenn was awarded the 2002 Nobel Prize in Chemistry for the development of electrospray ionization mass spectrometry in the late 1980s 15 Ionization mechanism edit nbsp Fenn s first electrospray ionization source coupled to a single quadrupole mass spectrometer The liquid containing the analytes of interest typically 10 6 10 4 M needed 16 is dispersed by electrospray 17 into a fine aerosol Because the ion formation involves extensive solvent evaporation also termed desolvation the typical solvents for electrospray ionization are prepared by mixing water with volatile organic compounds e g methanol 18 acetonitrile To decrease the initial droplet size compounds that increase the conductivity e g acetic acid are customarily added to the solution These species also act to provide a source of protons to facilitate the ionization process Large flow electrosprays can benefit from nebulization of a heated inert gas such as nitrogen or carbon dioxide in addition to the high temperature of the ESI source 19 The aerosol is sampled into the first vacuum stage of a mass spectrometer through a capillary carrying a potential difference of approximately 3000 V which can be heated to aid further solvent evaporation from the charged droplets The solvent evaporates from a charged droplet until it becomes unstable upon reaching its Rayleigh limit At this point the droplet deforms as the electrostatic repulsion of like charges in an ever decreasing droplet size becomes more powerful than the surface tension holding the droplet together 20 At this point the droplet undergoes Coulomb fission whereby the original droplet explodes creating many smaller more stable droplets The new droplets undergo desolvation and subsequently further Coulomb fissions During the fission the droplet loses a small percentage of its mass 1 0 2 3 along with a relatively large percentage of its charge 10 18 21 22 There are two major theories that explain the final production of gas phase ions the ion evaporation model IEM and the charge residue model CRM The IEM suggests that as the droplet reaches a certain radius the field strength at the surface of the droplet becomes large enough to assist the field desorption of solvated ions 23 24 The CRM suggests that electrospray droplets undergo evaporation and fission cycles eventually leading progeny droplets that contain on average one analyte ion or less 13 The gas phase ions form after the remaining solvent molecules evaporate leaving the analyte with the charges that the droplet carried nbsp IEM CRM and CEM schematic A large body of evidence shows either directly or indirectly that small ions from small molecules are liberated into the gas phase through the ion evaporation mechanism 24 25 citation needed 26 while larger ions from folded proteins for instance form by charged residue mechanism 27 28 29 A third model invoking combined charged residue field emission has been proposed 30 Another model called chain ejection model CEM is proposed for disordered polymers unfolded proteins 31 The ions observed by mass spectrometry may be quasimolecular ions created by the addition of a hydrogen cation and denoted M H or of another cation such as sodium ion M Na or the removal of a hydrogen nucleus M H Multiply charged ions such as M nH n are often observed For large macromolecules there can be many charge states resulting in a characteristic charge state envelope All these are even electron ion species electrons alone are not added or removed unlike in some other ionization sources The analytes are sometimes involved in electrochemical processes leading to shifts of the corresponding peaks in the mass spectrum This effect is demonstrated in the direct ionization of noble metals such as copper silver and gold using electrospray 32 The efficiency of generating the gas phase ions for small molecules in ESI varies depending on the compound structure the solvent used and instrumental parameters 33 The differences in ionization efficiency reach more than 1 million times Variants editThe electrosprays operated at low flow rates generate much smaller initial droplets which ensure improved ionization efficiency In 1993 Gale and Richard D Smith reported significant sensitivity increases could be achieved using lower flow rates and down to 200 nL min 34 In 1994 two research groups coined the name micro electrospray microspray for electrosprays working at low flow rates Emmett and Caprioli demonstrated improved performance for HPLC MS analyses when the electrospray was operated at 300 800 nL min 35 Wilm and Mann demonstrated that a capillary flow of 25 nL min can sustain an electrospray at the tip of emitters fabricated by pulling glass capillaries to a few micrometers 36 The latter was renamed nano electrospray nanospray in 1996 37 38 Currently the name nanospray is also in use for electrosprays fed by pumps at low flow rates 39 not only for self fed electrosprays Although there may not be a well defined flow rate range for electrospray microspray and nano electrospray 40 studied changes in analyte partition during droplet fission prior to ion release 40 In this paper they compare results obtained by three other groups 41 42 43 and then measure the signal intensity ratio Ba2 Ba BaBr at different flow rates Cold spray ionization is a form of electrospray in which the solution containing the sample is forced through a small cold capillary 10 80 C into an electric field to create a fine mist of cold charged droplets 44 Applications of this method include the analysis of fragile molecules and guest host interactions that cannot be studied using regular electrospray ionization Electrospray ionization has also been achieved at pressures as low as 25 torr and termed subambient pressure ionization with nanoelectrospray SPIN based upon a two stage ion funnel interface developed by Richard D Smith and coworkers 45 The SPIN implementation provided increased sensitivity due to the use of ion funnels that helped confine and transfer ions to the lower pressure region of the mass spectrometer Nanoelectrospray emitter is made out of a fine capillary with a small aperture about 1 3 micrometer For sufficient conductivity this capillary is usually sputter coated with conductive material e g gold Nanoelectrospray ionization consumes only a few microliters of a sample and forms smaller droplets 46 Operation at low pressure was particularly effective for low flow rates where the smaller electrospray droplet size allowed effective desolvation and ion formation to be achieved As a result the researchers were later able to demonstrate achieving an excess of 50 overall ionization utilization efficiency for transfer of ions from the liquid phase into the gas phase as ions and through the dual ion funnel interface to the mass spectrometer 47 Ambient ionization edit nbsp Diagram of a DESI ambient ionization source In ambient ionization the formation of ions occurs outside the mass spectrometer without sample preparation 48 49 50 Electrospray is used for ion formation in a number of ambient ion sources Desorption electrospray ionization DESI is an ambient ionization technique in which a solvent electrospray is directed at a sample 51 52 The electrospray is attracted to the surface by applying a voltage to the sample Sample compounds are extracted into the solvent which is again aerosolized as highly charged droplets that evaporate to form highly charged ions After ionization the ions enter the atmospheric pressure interface of the mass spectrometer DESI allows for ambient ionization of samples at atmospheric pressure with little sample preparation nbsp Diagram of a SESI ambient ionization source Extractive electrospray ionization is a spray type ambient ionization method that uses two merged sprays one of which is generated by electrospray 49 Laser based electrospray based ambient ionization is a two step process in which a pulsed laser is used to desorb or ablate material from a sample and the plume of material interacts with an electrospray to create ions 49 For ambient ionization the sample material is deposited on a target near the electrospray The laser desorbs or ablates material from the sample which is ejected from the surface and into the electrospray which produces highly charged ions Examples are electrospray laser desorption ionization matrix assisted laser desorption electrospray ionization and laser ablation electrospray ionization nbsp SESI MS SUPER SESI coupled with Thermo Fisher Scientific Orbitrap Electrostatic spray ionization ESTASI involved the analysis of samples located on a flat or porous surface or inside a microchannel A droplet containing analytes is deposited on a sample area to which a pulsed high voltage to is applied When the electrostatic pressure is larger than the surface tension droplets and ions are sprayed Secondary electrospray ionization SESI is an spray type ambient ionization method where charging ions are produced by means of an electrospray These ions then charge vapor molecules in the gas phase when colliding with them 53 54 In paper spray ionization the sample is applied to a piece of paper solvent is added and a high voltage is applied to the paper creating ions Applications edit nbsp The outside of the electrospray interface on an LTQ mass spectrometer Electrospray is used to study protein folding 55 56 57 Liquid chromatography mass spectrometry edit Main article liquid chromatography mass spectrometry Electrospray ionization is the ion source of choice to couple liquid chromatography with mass spectrometry LC MS The analysis can be performed online by feeding the liquid eluting from the LC column directly to an electrospray or offline by collecting fractions to be later analyzed in a classical nanoelectrospray mass spectrometry setup Among the numerous operating parameters in ESI MS for proteins 58 the electrospray voltage has been identified as an important parameter to consider in ESI LC MS gradient elution 59 The effect of various solvent compositions 60 such as TFA 61 or ammonium acetate 22 or supercharging reagents 62 63 64 65 or derivitizing groups 66 or spraying conditions 67 on electrospray LCMS spectra and or nanoESI MS spectra 68 have been studied Capillary electrophoresis mass spectrometry CE MS edit Capillary electrophoresis mass spectrometry was enabled by an ESI interface that was developed and patented by Richard D Smith and coworkers at Pacific Northwest National Laboratory and shown to have broad utility for the analysis of very small biological and chemical compound mixtures and even extending to a single biological cell Noncovalent gas phase interactions edit Electrospray ionization is also utilized in studying noncovalent gas phase interactions The electrospray process is thought to be capable of transferring liquid phase noncovalent complexes into the gas phase without disrupting the noncovalent interaction Problems 22 69 such as non specific interactions 70 have been identified when studying ligand substrate complexes by ESI MS or nanoESI MS An interesting example of this is studying the interactions between enzymes and drugs which are inhibitors of the enzyme 71 72 73 Competition studies between STAT6 and inhibitors 73 74 75 have used ESI as a way to screen for potential new drug candidates Electrospray ionization can even be used for studying protein complexes gt 1 MDa 76 16 See also editLaser ablation electrospray ionization Probe electrospray ionization Sonic spray ionizationReferences edit Ho CS Chan MHM Cheung RCK Law LK Lit LCW Ng KF Suen MWM Tai HL February 2003 Electrospray Ionisation Mass Spectrometry Principles and Clinical Applications Clin Biochem Rev 24 1 3 12 PMC 1853331 PMID 18568044 Pitt James J February 2009 Principles and Applications of Liquid Chromatography Mass Spectrometry in Clinical Biochemistry Clin Biochem Rev 30 1 19 34 PMC 2643089 PMID 19224008 Yamashita Masamichi Fenn John B September 1984 Electrospray ion source Another variation on the free jet theme The Journal of Physical Chemistry 88 20 4451 4459 doi 10 1021 j150664a002 a b Alexandrov M L Gall L N Krasnov N V Nikolaev V I Pavlenko V A Shkruov V A 1984 Extraction of ions from solutions under atmospheric pressure as a method for mass spectrometric analysis of bioorganic compounds Doklady Akad SSSR 277 3 379 383 Bibcode 2008RCMS 22 267A doi 10 1002 rcm 3113 PMID 18181250 Fenn J B Mann M Meng C K Wong S F Whitehouse C M 1989 Electrospray ionization for mass spectrometry of large biomolecules Science 246 4926 64 71 Bibcode 1989Sci 246 64F CiteSeerX 10 1 1 522 9458 doi 10 1126 science 2675315 PMID 2675315 Markides K Graslund A Advanced information on the Nobel Prize in Chemistry 2002 PDF Rayleigh L 1882 On the Equilibrium of Liquid Conducting Masses charged with Electricity Philosophical Magazine 14 87 184 186 doi 10 1080 14786448208628425 Zeleny J 1914 The electrical discharge from liquid points and a hydrostatic method of measuring the electric intensity at their surfaces Physical Review 3 2 69 91 Bibcode 1914PhRv 3 69Z doi 10 1103 PhysRev 3 69 Wilson C T G I Taylor 1925 The bursting of soap bubbles in a uniform electric field Proc Cambridge Philos Soc 22 5 728 Bibcode 1925PCPS 22 728W doi 10 1017 S0305004100009609 S2CID 137905700 Nolan J J 1926 Universal scaling laws for the disintegration of electrified drops Proc R Ir Acad A 37 28 Macky W A October 1 1931 Some Investigations on the Deformation and Breaking of Water Drops in Strong Electric Fields Proceedings of the Royal Society A 133 822 565 587 Bibcode 1931RSPSA 133 565M doi 10 1098 rspa 1931 0168 Geoffrey Taylor 1964 Disintegration of Water Droplets in an Electric Field Proceedings of the Royal Society A 280 1382 383 397 Bibcode 1964RSPSA 280 383T doi 10 1098 rspa 1964 0151 JSTOR 2415876 S2CID 15067908 a b Dole M Mack LL Hines RL Mobley RC Ferguson LD Alice MB 1968 Molecular Beams of Macroions Journal of Chemical Physics 49 5 2240 2249 Bibcode 1968JChPh 49 2240D doi 10 1063 1 1670391 Birendra N Pramanik A K Ganguly Michael L Gross 28 February 2002 Applied Electrospray Mass Spectrometry Practical Spectroscopy Series CRC Press pp 4 ISBN 978 0 8247 4419 9 Press Release The Nobel Prize in Chemistry 2002 The Nobel Foundation 2002 10 09 Retrieved 2011 04 02 a b Gross Jurgen H 2017 Electrospray Ionization Mass Spectrometry Cham Springer International Publishing pp 721 778 doi 10 1007 978 3 319 54398 7 12 ISBN 978 3 319 54397 0 retrieved 2024 03 15 Pozniak BP Cole RB 2007 Current Measurements within the Electrospray Emitter J Am Soc Mass Spectrom 18 4 737 748 doi 10 1016 j jasms 2006 11 012 PMID 17257852 Olumee et al 1998 Droplet Dynamics Changes in Electrostatic Sprays of Methanol Water Mixtures J Phys Chem A 102 46 9154 9160 Bibcode 1998JPCA 102 9154O CiteSeerX 10 1 1 661 5000 doi 10 1021 jp982027z Fernandez De La Mora J 2007 The Fluid Dynamics of Taylor Cones Annual Review of Fluid Mechanics 39 1 217 243 Bibcode 2007AnRFM 39 217F doi 10 1146 annurev fluid 39 050905 110159 Cole Richard B 2010 Electrospray and MALDI Mass Spectrometry Fundamentals Instrumentation Practicalities and Biological Applications 2 ed Wiley p 4 ISBN 978 0471741077 Li KY Tu H Ray AK April 2005 Charge limits on droplets during evaporation Langmuir 21 9 3786 94 doi 10 1021 la047973n PMID 15835938 a b c Kebarle P Verkerk UH 2009 Electrospray from ions in solution to ions in the gas phase what we know now Mass Spectrom Rev 28 6 898 917 Bibcode 2009MSRv 28 898K doi 10 1002 mas 20247 PMID 19551695 Iribarne JV Thomson BA 1976 On the evaporation of small ions from charged droplets Journal of Chemical Physics 64 6 2287 2294 Bibcode 1976JChPh 64 2287I doi 10 1063 1 432536 a b Nguyen S Fenn JB January 2007 Gas phase ions of solute species from charged droplets of solutions Proc Natl Acad Sci USA 104 4 1111 7 Bibcode 2007PNAS 104 1111N doi 10 1073 pnas 0609969104 PMC 1783130 PMID 17213314 Gamero Castano M 2000 Direct measurement of ion evaporation kinetics from electrified liquid surfaces J Chem Phys 113 2 815 Bibcode 2000JChPh 113 815G doi 10 1063 1 481857 S2CID 36112510 de la Mora Fernandez 2000 Electrospray ionization of large multiply charged species proceeds via Dole s charged residue mechanism Analytica Chimica Acta 406 93 104 doi 10 1016 S0003 2670 99 00601 7 An evaluation of the electric field on the drop surface at the point when it just ceases to be spherical yet carries the total ion charge z indicates that small PEG ions may be formed by ion evaporation The break observed in the charge distribution may perhaps mean that the shift from the Dole to the ion evaporation mechanism arises at m unintelligible 104 clarification needed though this inference is highly hypothetical de la Mora Fernandez 2000 Electrospray ionization of large multiply charged species proceeds via Dole s charged residue mechanism Analytica Chimica Acta 406 93 104 doi 10 1016 S0003 2670 99 00601 7 de la Mora Fernandez 2000 Electrospray ionization of large multiply charged species proceeds via Dole s charged residue mechanism Analytica Chimica Acta 406 93 104 doi 10 1016 S0003 2670 99 00601 7 For most published data examined zmax is between 65 and 110 of zR providing strong support in favor of Dole s charged residue mechanism at least for masses from 3 3 kD up to 1 4 MD Other large but less compact ions from proteins and linear chains of polyethylene glycols PEGs have zmax values considerably larger than zR apparently implying that they also formas charged residues though from non spherical drops held together by the polymer backbone de la Mora Fernandez 2000 Electrospray ionization of large multiply charged species proceeds via Dole s charged residue mechanism Analytica Chimica Acta 406 93 104 doi 10 1016 S0003 2670 99 00601 7 The data do show a nearly discontinuous jump in the observed m z for a mass somewhere between 20 000 and 50 000 and it is tempting to conclude that this is due to a corresponding transition where the ionization mechanism shifts from one type to the other This would correspond to a critical value of z in the vicinity of 50 with a corresponding electric field of 2 6 V nm Of course this is entirely hypothetical and there is yet no compelling evidence of any kind indicating that an ion with as many as 30 charges can be formed by field evaporation Hogan CJ Carroll JA Rohrs HW Biswas P Gross ML January 2009 Combined charged residue field emission model of macromolecular electrospray ionization Anal Chem 81 1 369 77 doi 10 1021 ac8016532 PMC 2613577 PMID 19117463 Konermann Lars 2013 Unraveling the Mechanism of Electrospray Ionization Analytical Chemistry 85 1 2 9 doi 10 1021 ac302789c PMID 23134552 Li Anyin Luo Qingjie Park So Jung Cooks R Graham 2014 Synthesis and Catalytic Reactions of Nanoparticles formed by Electrospray Ionization of Coinage Metals Angewandte Chemie International Edition 53 12 3147 3150 doi 10 1002 anie 201309193 ISSN 1433 7851 PMID 24554582 Kruve Anneli Kaupmees Karl Liigand Jaanus Leito Ivo 2014 Negative Electrospray Ionization via Deprotonation Predicting the Ionization Efficiency Analytical Chemistry 86 10 4822 4830 doi 10 1021 ac404066v PMID 24731109 Gale DC Smith RD 1993 Small Volume and Low Flow Rate Electrospray Ionization Mass Spectrometry for Aqueous Samples Rapid Commun Mass Spectrom 7 11 1017 1021 Bibcode 1993RCMS 7 1017G doi 10 1002 rcm 1290071111 Emmett MR Caprioli RM 1994 Micro electrospray mass spectrometry ultra high sensitivity analysis of peptides and proteins J Am Soc Mass Spectrom 5 7 605 613 doi 10 1016 1044 0305 94 85001 1 PMID 24221962 Wilm MS Mann M 1994 Electrospray and Taylor Cone theory Dole s beam of macromolecules at last Int J Mass Spectrom Ion Process 136 2 3 167 180 Bibcode 1994IJMSI 136 167W doi 10 1016 0168 1176 94 04024 9 Wilm M Mann M 1996 Analytical properties of the nanoelectrospray ion source Anal Chem 68 1 1 8 doi 10 1021 ac9509519 PMID 8779426 Gibson Mugo Samuel M Oleschuk Richard D et al 2009 Nanoelectrospray emitters Trends and perspective Mass Spectrometry Reviews 28 6 918 936 Bibcode 2009MSRv 28 918G doi 10 1002 mas 20248 PMID 19479726 Page JS Marginean I Baker ES Kelly RT Tang K Smith RD December 2009 Biases in ion transmission through an electrospray ionization mass spectrometry capillary inlet J Am Soc Mass Spectrom 20 12 2265 72 doi 10 1016 j jasms 2009 08 018 PMC 2861838 PMID 19815425 a b Schmidt A Karas M Dulcks T May 2003 Effect of different solution flow rates on analyte ion signals in nano ESI MS or when does ESI turn into nano ESI J Am Soc Mass Spectrom 14 5 492 500 doi 10 1016 S1044 0305 03 00128 4 PMID 12745218 Wilm M S Mann M 1994 Electrospray and Taylor Cone Theory Dole s Beam of Macromolecules at Last Int J Mass Spectrom Ion Process 136 2 3 167 180 Bibcode 1994IJMSI 136 167W doi 10 1016 0168 1176 94 04024 9 Fernandez de la Mora J Loscertales I G 2006 The Current Emitted by Highly Conducting Taylor Cones J Fluid Mech 260 155 184 Bibcode 1994JFM 260 155D doi 10 1017 S0022112094003472 S2CID 122935117 Pfeifer RJ Hendricks 1968 Parametric Studies of Electrohydrodynamic Spraying AIAA J 6 3 496 502 Bibcode 1968AIAAJ 6 496H doi 10 2514 3 4525 RSC Chemical Methods Ontology Cold spray ionisation mass spectrometry Page JS Tang K Kelly RT Smith RD 2008 A subambient pressure ionization with nanoelectrospray SPIN source and interface for improved sensitivity in mass spectrometry Analytical Chemistry 80 5 1800 1805 doi 10 1021 ac702354b PMC 2516344 PMID 18237189 Karas M Bahr U Dulcks T 2000 03 01 Nano electrospray ionization mass spectrometry addressing analytical problems beyond routine Fresenius Journal of Analytical Chemistry 366 6 7 669 676 doi 10 1007 s002160051561 ISSN 0937 0633 PMID 11225778 S2CID 24730378 I Marginean J S Page A V Tolmachev K Tang R D Smith 2010 Achieving 50 Ionization Efficiency in Subambient Pressure Ionization with Nanoelectrospray Analytical Chemistry 82 22 9344 9349 doi 10 1021 ac1019123 PMC 2982749 PMID 21028835 Cooks R Graham Ouyang Zheng Takats Zoltan Wiseman Justin M 2006 Ambient Mass Spectrometry Science 311 5767 1566 70 Bibcode 2006Sci 311 1566C doi 10 1126 science 1119426 PMID 16543450 S2CID 98131681 a b c Monge Maria Eugenia Harris Glenn A Dwivedi Prabha Fernandez Facundo M 2013 Mass Spectrometry Recent Advances in Direct Open Air Surface Sampling Ionization Chemical Reviews 113 4 2269 2308 doi 10 1021 cr300309q ISSN 0009 2665 PMID 23301684 Huang Min Zong Yuan Cheng Hui Cheng Sy Chyi Cho Yi Tzu Shiea Jentaie 2010 Ambient Ionization Mass Spectrometry Annual Review of Analytical Chemistry 3 1 43 65 Bibcode 2010ARAC 3 43H doi 10 1146 annurev anchem 111808 073702 ISSN 1936 1327 PMID 20636033 Z Takats J M Wiseman B Gologan R G Cooks 2004 Mass Spectrometry Sampling Under Ambient Conditions with Desorption Electrospray Ionization Science 306 5695 471 473 Bibcode 2004Sci 306 471T doi 10 1126 science 1104404 PMID 15486296 S2CID 22994482 Takats Z Wiseman JM Cooks RG 2005 Ambient mass spectrometry using desorption electrospray ionization DESI instrumentation mechanisms and applications in forensics chemistry and biology Journal of Mass Spectrometry 40 10 1261 75 Bibcode 2005JMSp 40 1261T doi 10 1002 jms 922 PMID 16237663 Vidal de Miguel G Macia M Pinacho P Blanco J 2012 10 16 Low Sample Flow Secondary Electrospray Ionization Improving Vapor Ionization Efficiency Analytical Chemistry 84 20 8475 8479 doi 10 1021 ac3005378 ISSN 0003 2700 PMID 22970991 Barrios Collado Cesar Vidal de Miguel Guillermo Martinez Lozano Sinues Pablo February 2016 Numerical modeling and experimental validation of a universal secondary electrospray ionization source for mass spectrometric gas analysis in real time Sensors and Actuators B Chemical 223 217 225 doi 10 1016 j snb 2015 09 073 hdl 20 500 11850 105470 Konermann L Douglas DJ 1998 Equilibrium unfolding of proteins monitored by electrospray ionization mass spectrometry Distinguishing two state from multi state transitions Rapid Communications in Mass Spectrometry 12 8 435 442 Bibcode 1998RCMS 12 435K doi 10 1002 SICI 1097 0231 19980430 12 8 lt 435 AID RCM181 gt 3 0 CO 2 F PMID 9586231 Nemes Goyal Samita Vertes Akos et al 2008 Conformational and Noncovalent Complexation Changes in Proteins during Electrospray Ionization Analytical Chemistry 80 2 387 395 doi 10 1021 ac0714359 PMID 18081323 Sobott Robinson 2004 Characterising electrosprayed biomolecules using tandem MS the noncovalent GroEL chaperonin assembly International Journal of Mass Spectrometry 236 1 3 25 32 Bibcode 2004IJMSp 236 25S doi 10 1016 j ijms 2004 05 010 Vaidyanathan S Kell D B Goodacre R 2004 Selective detection of proteins in mixtures using electrospray ionization mass spectrometry influence of instrumental settings and implications for proteomics Analytical Chemistry 76 17 5024 5032 doi 10 1021 ac049684 PMID 15373437 Marginean I Kelly RT Moore RJ Prior DC LaMarche BL Tang K Smith RD April 2009 Selection of the optimum electrospray voltage for gradient elution LC MS measurements J Am Soc Mass Spectrom 20 4 682 8 doi 10 1016 j jasms 2008 12 004 PMC 2692488 PMID 19196520 Iavarone Jurchen John C Williams Evan R et al 2000 Effects of solvent on the maximum charge state and charge state distribution of protein ions produced by electrospray ionization J Am Soc Mass Spectrom 11 11 976 985 doi 10 1016 S1044 0305 00 00169 0 PMC 1414794 PMID 11073261 Garcia 2005 The effect of the mobile phase additives on sensitivity in the analysis of peptides and proteins by high performance liquid chromatography electrospray mass spectrometry Journal of Chromatography B 825 2 111 123 doi 10 1016 j jchromb 2005 03 041 PMID 16213445 Teo CA Donald WA May 2014 Solution additives for supercharging proteins beyond the theoretical maximum proton transfer limit in electrospray ionization mass spectrometry Anal Chem 86 9 4455 62 doi 10 1021 ac500304r PMID 24712886 Lomeli SH Peng IX Yin S Loo RR Loo JA January 2010 New reagents for increasing ESI multiple charging of proteins and protein complexes J Am Soc Mass Spectrom 21 1 127 31 doi 10 1016 j jasms 2009 09 014 PMC 2821426 PMID 19854660 Lomeli SH Yin S Ogorzalek Loo RR Loo JA April 2009 Increasing charge while preserving noncovalent protein complexes for ESI MS J Am Soc Mass Spectrom 20 4 593 6 doi 10 1016 j jasms 2008 11 013 PMC 2789282 PMID 19101165 Yin S Loo JA March 2011 Top Down Mass Spectrometry of Supercharged Native Protein Ligand Complexes Int J Mass Spectrom 300 2 3 118 122 Bibcode 2011IJMSp 300 118Y doi 10 1016 j ijms 2010 06 032 PMC 3076692 PMID 21499519 Krusemark CJ Frey BL Belshaw PJ Smith LM September 2009 Modifying the charge state distribution of proteins in electrospray ionization mass spectrometry by chemical derivatization J Am Soc Mass Spectrom 20 9 1617 25 doi 10 1016 j jasms 2009 04 017 PMC 2776692 PMID 19481956 Nemes P Goyal S Vertes A January 2008 Conformational and noncovalent complexation changes in proteins during electrospray ionization Anal Chem 80 2 387 95 doi 10 1021 ac0714359 PMID 18081323 Ramanathan R Zhong R Blumenkrantz N Chowdhury SK Alton KB October 2007 Response normalized liquid chromatography nanospray ionization mass spectrometry J Am Soc Mass Spectrom 18 10 1891 9 doi 10 1016 j jasms 2007 07 022 PMID 17766144 Gabelica V Vreuls C Filee P Duval V Joris B Pauw ED 2002 Advantages and drawbacks of nanospray for studying noncovalent protein DNA complexes by mass spectrometry Rapid Commun Mass Spectrom 16 18 1723 8 Bibcode 2002RCMS 16 1723G doi 10 1002 rcm 776 hdl 2268 322 PMID 12207359 Daubenfeld T Bouin AP van der Rest G September 2006 A deconvolution method for the separation of specific versus nonspecific interactions in noncovalent protein ligand complexes analyzed by ESI FT ICR mass spectrometry J Am Soc Mass Spectrom 17 9 1239 48 doi 10 1016 j jasms 2006 05 005 PMID 16793278 Rosu F De Pauw E Gabelica V July 2008 Electrospray mass spectrometry to study drug nucleic acids interactions Biochimie 90 7 1074 87 doi 10 1016 j biochi 2008 01 005 PMID 18261993 Wortmann A Jecklin MC Touboul D Badertscher M Zenobi R May 2008 Binding constant determination of high affinity protein ligand complexes by electrospray ionization mass spectrometry and ligand competition J Mass Spectrom 43 5 600 8 Bibcode 2008JMSp 43 600W doi 10 1002 jms 1355 PMID 18074334 a b Jecklin MC Touboul D Bovet C Wortmann A Zenobi R March 2008 Which electrospray based ionization method best reflects protein ligand interactions found in solution a comparison of ESI nanoESI and ESSI for the determination of dissociation constants with mass spectrometry J Am Soc Mass Spectrom 19 3 332 43 doi 10 1016 j jasms 2007 11 007 hdl 20 500 11850 9214 PMID 18083584 Touboul D Maillard L Grasslin A Moumne R Seitz M Robinson J Zenobi R February 2009 How to deal with weak interactions in noncovalent complexes analyzed by electrospray mass spectrometry cyclopeptidic inhibitors of the nuclear receptor coactivator 1 STAT6 J Am Soc Mass Spectrom 20 2 303 11 doi 10 1016 j jasms 2008 10 008 hdl 20 500 11850 15377 PMID 18996720 Czuczy N Katona M Takats Z February 2009 Selective detection of specific protein ligand complexes by electrosonic spray precursor ion scan tandem mass spectrometry J Am Soc Mass Spectrom 20 2 227 37 doi 10 1016 j jasms 2008 09 010 PMID 18976932 Ishii Kentaro Zhou Min Uchiyama Susumu 2018 02 01 Native mass spectrometry for understanding dynamic protein complex Biochimica et Biophysica Acta BBA General Subjects Biophysical Exploration of Dynamical Ordering of Biomolecular Systems 1862 2 275 286 doi 10 1016 j bbagen 2017 09 019 ISSN 0304 4165 Further reading editCole Richard 1997 Electrospray ionization mass spectrometry fundamentals instrumentation and applications New York Wiley ISBN 978 0 471 14564 6 Gross Michael Pramanik Birendra N Ganguly A K 2002 Applied electrospray mass spectrometry New York N Y Marcel Dekker ISBN 978 0 8247 0618 0 Snyder A Peter 1996 Biochemical and biotechnological applications of electrospray ionization mass spectrometry Columbus OH American Chemical Society ISBN 978 0 8412 3378 2 Alexandrov M L L N Gall N V Krasnov V I Nikolaev V A Pavlenko V A Shkurov July 1984 Ekstrakciya ionov iz rastvorov pri atmosfernom davlenii Metod mass spektrometricheskogo analiza bioorganicheskih veshestv Extraction of ions from solutions at atmospheric pressure A method for mass spectrometric analysis of bioorganic substances Doklady Akademii Nauk SSSR in Russian 277 2 379 383 Alexandrov M L L N Gall N V Krasnov V I Nikolaev V A Pavlenko V A Shkurov 2008 July 1984 Extraction of ions from solutions under atmospheric pressure as a method for mass spectrometric analysis of bioorganic compounds Rapid Communications in Mass Spectrometry 22 3 267 270 Bibcode 2008RCMS 22 267A doi 10 1002 rcm 3113 PMID 18181250 External links editElectrospray Ionization Primer National High Magnetic Field Laboratory Electrospray Ionization Mass Spectrometry at the U S National Library of Medicine Medical Subject Headings MeSH Retrieved from https en wikipedia org w index php title Electrospray ionization amp oldid 1216630363, wikipedia, wiki, book, books, library,