Infrared spectroscopy (IR spectroscopy or vibrational spectroscopy) is the measurement of the interaction of infrared radiation with matter by absorption, emission, or reflection. It is used to study and identify chemical substances or functional groups in solid, liquid, or gaseous forms. It can be used to characterize new materials or identify and verify known and unknown samples. The method or technique of infrared spectroscopy is conducted with an instrument called an infrared spectrometer (or spectrophotometer) which produces an infrared spectrum. An IR spectrum can be visualized in a graph of infrared light absorbance (or transmittance) on the vertical axis vs. frequency, wavenumber or wavelength on the horizontal axis. Typical units of wavenumber used in IR spectra are reciprocal centimeters, with the symbol cm−1. Units of IR wavelength are commonly given in micrometers (formerly called "microns"), symbol μm, which are related to the wavenumber in a reciprocal way. A common laboratory instrument that uses this technique is a Fourier transform infrared (FTIR) spectrometer. Two-dimensional IR is also possible as discussed below.

OVIRS instrument of the OSIRIS-REx probe is a visible and infrared spectrometer

The infrared portion of the electromagnetic spectrum is usually divided into three regions; the near-, mid- and far- infrared, named for their relation to the visible spectrum. The higher-energy near-IR, approximately 14,000–4,000 cm−1 (0.7–2.5 μm wavelength) can excite overtone or combination modes of molecular vibrations. The mid-infrared, approximately 4,000–400 cm−1 (2.5–25 μm) is generally used to study the fundamental vibrations and associated rotational–vibrational structure. The far-infrared, approximately 400–10 cm−1 (25–1,000 μm) has low energy and may be used for rotational spectroscopy and low frequency vibrations. The region from 2–130 cm−1, bordering the microwave region, is considered the terahertz region and may probe intermolecular vibrations.[1] The names and classifications of these subregions are conventions, and are only loosely based on the relative molecular or electromagnetic properties.

Theory Edit

 
Sample of an IR spec. reading; this one is from bromomethane (CH3Br), showing peaks around 3000, 1300, and 1000 cm−1 (on the horizontal axis).

Infrared spectroscopy exploits the fact that molecules absorb frequencies that are characteristic of their structure. These absorptions occur at resonant frequencies, i.e. the frequency of the absorbed radiation matches the vibrational frequency. The energies are affected by the shape of the molecular potential energy surfaces, the masses of the atoms, and the associated vibronic coupling.

 
3D animation of the symmetric stretch-compress mode of the C–H bonds of bromomethane

In particular, in the Born–Oppenheimer and harmonic approximations (i.e. when the molecular Hamiltonian corresponding to the electronic ground state can be approximated by a harmonic oscillator in the neighborhood of the equilibrium molecular geometry), the resonant frequencies are associated with the normal modes of vibration corresponding to the molecular electronic ground state potential energy surface.

The resonant frequencies are also related to the strength of the bond and the mass of the atoms at either end of it. Thus, the frequency of the vibrations are associated with a particular normal mode of motion and a particular bond type.

Number of vibrational modes Edit

In order for a vibrational mode in a sample to be "IR active", it must be associated with changes in the molecular dipole moment. A permanent dipole is not necessary, as the rule requires only a change in dipole moment.[2]

A molecule can vibrate in many ways, and each way is called a vibrational mode. For molecules with N number of atoms, geometrically linear molecules have 3N – 5 degrees of vibrational modes, whereas nonlinear molecules have 3N – 6 degrees of vibrational modes (also called vibrational degrees of freedom). As examples linear carbon dioxide (CO2) has 3 × 3 – 5 = 4, while non-linear water (H2O), has only 3 × 3 – 6 = 3.

 
Stretching and bending oscillations of the CO2 carbon dioxide molecule. Upper left: symmetric stretching. Upper right: antisymmetric stretching. Lower line: degenerate pair of bending modes.

Simple diatomic molecules have only one bond and only one vibrational band. If the molecule is symmetrical, e.g. N2, the band is not observed in the IR spectrum, but only in the Raman spectrum. Asymmetrical diatomic molecules, e.g. carbon monoxide (CO), absorb in the IR spectrum. More complex molecules have many bonds, and their vibrational spectra are correspondingly more complex, i.e. big molecules have many peaks in their IR spectra.

The atoms in a CH2X2 group, commonly found in organic compounds and where X can represent any other atom, can vibrate in nine different ways. Six of these vibrations involve only the CH2 portion: two stretching modes (ν): symmetrics) and antisymmetricas); and four bending modes: scissoring (δ), rocking (ρ), wagging (ω) and twisting (τ), as shown below. Structures that do not have the two additional X groups attached have fewer modes because some modes are defined by specific relationships to those other attached groups. For example, in water, the rocking, wagging, and twisting modes do not exist because these types of motions of the H atoms represent simple rotation of the whole molecule rather than vibrations within it. In case of more complex molecules, out-of-plane (γ) vibrational modes can be also present.[3]

Symmetry

Direction
Symmetric Antisymmetric
Radial  
Symmetric stretching (νs)
 
Antisymmetric stretching (νas)
Latitudinal  
Scissoring (δ)
 
Rocking (ρ)
Longitudinal  
Wagging (ω)
 
Twisting (τ)

These figures do not represent the "recoil" of the C atoms, which, though necessarily present to balance the overall movements of the molecule, are much smaller than the movements of the lighter H atoms.

The simplest and most important or fundamental IR bands arise from the excitations of normal modes, the simplest distortions of the molecule, from the ground state with vibrational quantum number v = 0 to the first excited state with vibrational quantum number v = 1. In some cases, overtone bands are observed. An overtone band arises from the absorption of a photon leading to a direct transition from the ground state to the second excited vibrational state (v = 2). Such a band appears at approximately twice the energy of the fundamental band for the same normal mode. Some excitations, so-called combination modes, involve simultaneous excitation of more than one normal mode. The phenomenon of Fermi resonance can arise when two modes are similar in energy; Fermi resonance results in an unexpected shift in energy and intensity of the bands etc.[citation needed]

Practical IR spectroscopy Edit

The infrared spectrum of a sample is recorded by passing a beam of infrared light through the sample. When the frequency of the IR is the same as the vibrational frequency of a bond or collection of bonds, absorption occurs. Examination of the transmitted light reveals how much energy was absorbed at each frequency (or wavelength). This measurement can be achieved by scanning the wavelength range using a monochromator. Alternatively, the entire wavelength range is measured using a Fourier transform instrument and then a transmittance or absorbance spectrum is generated using a dedicated procedure.

This technique is commonly used for analyzing samples with covalent bonds. Simple spectra are obtained from samples with few IR active bonds and high levels of purity. More complex molecular structures lead to more absorption bands and more complex spectra.

 
Typical IR solution cell. The windows are CaF2.

Sample preparation Edit

Gas samples Edit

Gaseous samples require a sample cell with a long pathlength to compensate for the diluteness. The pathlength of the sample cell depends on the concentration of the compound of interest. A simple glass tube with length of 5 to 10 cm equipped with infrared-transparent windows at both ends of the tube can be used for concentrations down to several hundred ppm. Sample gas concentrations well below ppm can be measured with a White's cell in which the infrared light is guided with mirrors to travel through the gas. White's cells are available with optical pathlength starting from 0.5 m up to hundred meters.

Liquid samples Edit

Liquid samples can be sandwiched between two plates of a salt (commonly sodium chloride, or common salt, although a number of other salts such as potassium bromide or calcium fluoride are also used).[4] The plates are transparent to the infrared light and do not introduce any lines onto the spectra. With increasing technology in computer filtering and manipulation of the results, samples in solution can now be measured accurately (water produces a broad absorbance across the range of interest, and thus renders the spectra unreadable without this computer treatment).

Solid samples Edit

Solid samples can be prepared in a variety of ways. One common method is to crush the sample with an oily mulling agent (usually mineral oil Nujol). A thin film of the mull is applied onto salt plates and measured. The second method is to grind a quantity of the sample with a specially purified salt (usually potassium bromide) finely (to remove scattering effects from large crystals). This powder mixture is then pressed in a mechanical press to form a translucent pellet through which the beam of the spectrometer can pass.[4] A third technique is the "cast film" technique, which is used mainly for polymeric materials. The sample is first dissolved in a suitable, non-hygroscopic solvent. A drop of this solution is deposited on the surface of a KBr or NaCl cell. The solution is then evaporated to dryness and the film formed on the cell is analysed directly. Care is important to ensure that the film is not too thick otherwise light cannot pass through. This technique is suitable for qualitative analysis. The final method is to use microtomy to cut a thin (20–100 μm) film from a solid sample. This is one of the most important ways of analysing failed plastic products for example because the integrity of the solid is preserved.

In photoacoustic spectroscopy the need for sample treatment is minimal. The sample, liquid or solid, is placed into the sample cup which is inserted into the photoacoustic cell which is then sealed for the measurement. The sample may be one solid piece, powder or basically in any form for the measurement. For example, a piece of rock can be inserted into the sample cup and the spectrum measured from it.

A useful way of analyzing solid samples without the need for cutting samples uses ATR or attenuated total reflectance spectroscopy. Using this approach, samples are pressed against the face of a single crystal. The infrared radiation passes through the crystal and only interacts with the sample at the interface between the two materials.

Comparing to a reference Edit

 
Schematics of a two-beam absorption spectrometer. A beam of infrared light is produced, passed through an interferometer (not shown), and then split into two separate beams. One is passed through the sample, the other passed through a reference. The beams are both reflected back towards a detector, however first they pass through a splitter, which quickly alternates which of the two beams enters the detector. The two signals are then compared and a printout is obtained. This "two-beam" setup gives accurate spectra even if the intensity of the light source drifts over time.

It is typical to record spectrum of both the sample and a "reference". This step controls for a number of variables, e.g. infrared detector, which may affect the spectrum. The reference measurement makes it possible to eliminate the instrument influence.

The appropriate "reference" depends on the measurement and its goal. The simplest reference measurement is to simply remove the sample (replacing it by air). However, sometimes a different reference is more useful. For example, if the sample is a dilute solute dissolved in water in a beaker, then a good reference measurement might be to measure pure water in the same beaker. Then the reference measurement would cancel out not only all the instrumental properties (like what light source is used), but also the light-absorbing and light-reflecting properties of the water and beaker, and the final result would just show the properties of the solute (at least approximately).

A common way to compare to a reference is sequentially: first measure the reference, then replace the reference by the sample and measure the sample. This technique is not perfectly reliable; if the infrared lamp is a bit brighter during the reference measurement, then a bit dimmer during the sample measurement, the measurement will be distorted. More elaborate methods, such as a "two-beam" setup (see figure), can correct for these types of effects to give very accurate results. The Standard addition method can be used to statistically cancel these errors.

Nevertheless, among different absorption-based techniques which are used for gaseous species detection, Cavity ring-down spectroscopy (CRDS) can be used as a calibration-free method. The fact that CRDS is based on the measurements of photon life-times (and not the laser intensity) makes it needless for any calibration and comparison with a reference [5]

Some instruments also automatically identify the substance being measured from a store of thousands of reference spectra held in storage.

FTIR Edit

 
An interferogram from an FTIR measurement. The horizontal axis is the position of the mirror, and the vertical axis is the amount of light detected. This is the "raw data" which can be Fourier transformed to get the actual spectrum.

Fourier transform infrared (FTIR) spectroscopy is a measurement technique that allows one to record infrared spectra. Infrared light is guided through an interferometer and then through the sample (or vice versa). A moving mirror inside the apparatus alters the distribution of infrared light that passes through the interferometer. The signal directly recorded, called an "interferogram", represents light output as a function of mirror position. A data-processing technique called Fourier transform turns this raw data into the desired result (the sample's spectrum): light output as a function of infrared wavelength (or equivalently, wavenumber). As described above, the sample's spectrum is always compared to a reference.

An alternate method for acquiring spectra is the "dispersive" or "scanning monochromator" method. In this approach, the sample is irradiated sequentially with various single wavelengths. The dispersive method is more common in UV-Vis spectroscopy, but is less practical in the infrared than the FTIR method. One reason that FTIR is favored is called "Fellgett's advantage" or the "multiplex advantage": The information at all frequencies is collected simultaneously, improving both speed and signal-to-noise ratio. Another is called "Jacquinot's Throughput Advantage": A dispersive measurement requires detecting much lower light levels than an FTIR measurement.[6] There are other advantages, as well as some disadvantages,[6] but virtually all modern infrared spectrometers are FTIR instruments.

Infrared microscopy Edit

Various forms of infrared microscopy exist. These include IR versions of sub-diffraction microscopy[7][8] such as IR NSOM,[9] photothermal microspectroscopy, Nano-FTIR and atomic force microscope based infrared spectroscopy (AFM-IR).

Other methods in molecular vibrational spectroscopy Edit

Infrared spectroscopy is not the only method of studying molecular vibrational spectra. Raman spectroscopy involves an inelastic scattering process in which only part of the energy of an incident photon is absorbed by the molecule, and the remaining part is scattered and detected. The energy difference corresponds to absorbed vibrational energy.

The selection rules for infrared and for Raman spectroscopy are different at least for some molecular symmetries, so that the two methods are complementary in that they observe vibrations of different symmetries.

Another method is electron energy loss spectroscopy (EELS), in which the energy absorbed is provided by an inelastically scattered electron rather than a photon. This method is useful for studying vibrations of molecules adsorbed on a solid surface.

Recently, high-resolution EELS (HREELS) has emerged as a technique for performing vibrational spectroscopy in a transmission electron microscope (TEM).[10] In combination with the high spatial resolution of the TEM, unprecedented experiments have been performed, such as nano-scale temperature measurements,[11][12] mapping of isotopically labeled molecules,[13] mapping of phonon modes in position- and momentum-space,[14][15] vibrational surface and bulk mode mapping on nanocubes,[16] and investigations of polariton modes in van der Waals crystals.[17] Analysis of vibrational modes that are IR-inactive but appear in inelastic neutron scattering is also possible at high spatial resolution using EELS.[18] Although the spatial resolution of HREELs is very high, the bands are extremely broad compared to other techniques.[10]

Computational infrared microscopy Edit

By using computer simulations and normal mode analysis it is possible to calculate theoretical frequencies of molecules.[19]

Absorption bands Edit

IR spectroscopy is often used to identify structures because functional groups give rise to characteristic bands both in terms of intensity and position (frequency). The positions of these bands are summarized in correlation tables as shown below.

 
List of main IR spectroscopy bands. For example, the carboxyl group will contain a C = O band at 1700 cm−1 and an OH band at 3500 cm−1 (total group -COOH). Wavenumbers listed in cm−1.

Regions Edit

A spectrograph is often interpreted as having two regions.[20]

  • functional group region  

In the functional region there are one to a few troughs per functional group.[20]

  • fingerprint region  

In the fingerprint region there are many troughs which form an intricate pattern which can be used like a fingerprint to determine the compound.[20]

Badger's rule Edit

For many kinds of samples, the assignments are known, i.e. which bond deformation(s) are associated with which frequency. In such cases further information can be gleaned about the strength on a bond, relying on the empirical guideline called Badger's rule. Originally published by Richard McLean Badger in 1934,[21] this rule states that the strength of a bond (in terms of force constant) correlates with the bond length. That is, increase in bond strength leads to corresponding bond shortening and vice versa.

Uses and applications Edit

 
Infrared spectrophotometer used to analyze the diethyltoluamide insect repellent, 1960
 
US Food and Drug Administration scientist uses portable near infrared spectroscopy device to detect potentially illegal substances

Infrared spectroscopy is a simple and reliable technique widely used in both organic and inorganic chemistry, in research and industry. In catalysis research it is a very useful tool to characterize the catalyst,[22][23][24] as well as to detect intermediates[25] and products during the catalytic reaction. It is used in quality control, dynamic measurement, and monitoring applications such as the long-term unattended measurement of CO2 concentrations in greenhouses and growth chambers by infrared gas analyzers.

It is also used in forensic analysis in both criminal and civil cases, for example in identifying polymer degradation. It can be used in determining the blood alcohol content of a suspected drunk driver.

IR spectroscopy has been successfully used in analysis and identification of pigments in paintings[26] and other art objects[27] such as illuminated manuscripts.[28]

Infrared spectroscopy is also useful in measuring the degree of polymerization in polymer manufacture. Changes in the character or quantity of a particular bond are assessed by measuring at a specific frequency over time. Modern research instruments can take infrared measurements across the range of interest as frequently as 32 times a second. This can be done whilst simultaneous measurements are made using other techniques. This makes the observations of chemical reactions and processes quicker and more accurate.

Infrared spectroscopy has also been successfully utilized in the field of semiconductor microelectronics:[29] for example, infrared spectroscopy can be applied to semiconductors like silicon, gallium arsenide, gallium nitride, zinc selenide, amorphous silicon, silicon nitride, etc.

Another important application of infrared spectroscopy is in the food industry to measure the concentration of various compounds in different food products.[30][31]

The instruments are now small, and can be transported, even for use in field trials.

Infrared spectroscopy is also used in gas leak detection devices such as the DP-IR and EyeCGAs.[32] These devices detect hydrocarbon gas leaks in the transportation of natural gas and crude oil.

In February 2014, NASA announced a greatly upgraded database,[33] based on IR spectroscopy, for tracking polycyclic aromatic hydrocarbons (PAHs) in the universe. According to scientists, more than 20% of the carbon in the universe may be associated with PAHs, possible starting materials for the formation of life. PAHs seem to have been formed shortly after the Big Bang, are widespread throughout the universe, and are associated with new stars and exoplanets.[34]

Infrared spectroscopy is an important analysis method in the recycling process of household waste plastics, and a convenient stand-off method to sort plastic of different polymers (PET, HDPE, ...).[35]

Other developments include a miniature IR-spectrometer that's linked to a cloud based database and suitable for personal everyday use,[36] and NIR-spectroscopic chips[37] that can be embedded in smartphones and various gadgets.

Isotope effects Edit

The different isotopes in a particular species may exhibit different fine details in infrared spectroscopy. For example, the O–O stretching frequency (in reciprocal centimeters) of oxyhemocyanin is experimentally determined to be 832 and 788 cm−1 for ν(16O–16O) and ν(18O–18O), respectively.

By considering the O–O bond as a spring, the frequency of absorbance can be calculated as a wavenumber [= frequency/(speed of light)]

 

where k is the spring constant for the bond, c is the speed of light, and μ is the reduced mass of the A–B system:

 

(  is the mass of atom  ).

The reduced masses for 16O–16O and 18O–18O can be approximated as 8 and 9 respectively. Thus

 

The effect of isotopes, both on the vibration and the decay dynamics, has been found to be stronger than previously thought. In some systems, such as silicon and germanium, the decay of the anti-symmetric stretch mode of interstitial oxygen involves the symmetric stretch mode with a strong isotope dependence. For example, it was shown that for a natural silicon sample, the lifetime of the anti-symmetric vibration is 11.4 ps. When the isotope of one of the silicon atoms is increased to 29Si, the lifetime increases to 19 ps. In similar manner, when the silicon atom is changed to 30Si, the lifetime becomes 27 ps.[38]

Two-dimensional IR Edit

Two-dimensional infrared correlation spectroscopy analysis combines multiple samples of infrared spectra to reveal more complex properties. By extending the spectral information of a perturbed sample, spectral analysis is simplified and resolution is enhanced. The 2D synchronous and 2D asynchronous spectra represent a graphical overview of the spectral changes due to a perturbation (such as a changing concentration or changing temperature) as well as the relationship between the spectral changes at two different wavenumbers.

 
Pulse Sequence used to obtain a two-dimensional Fourier transform infrared spectrum. The time period   is usually referred to as the coherence time and the second time period   is known as the waiting time. The excitation frequency is obtained by Fourier transforming along the   axis.

Nonlinear two-dimensional infrared spectroscopy[39][40] is the infrared version of correlation spectroscopy. Nonlinear two-dimensional infrared spectroscopy is a technique that has become available with the development of femtosecond infrared laser pulses. In this experiment, first a set of pump pulses is applied to the sample. This is followed by a waiting time during which the system is allowed to relax. The typical waiting time lasts from zero to several picoseconds, and the duration can be controlled with a resolution of tens of femtoseconds. A probe pulse is then applied, resulting in the emission of a signal from the sample. The nonlinear two-dimensional infrared spectrum is a two-dimensional correlation plot of the frequency ω1 that was excited by the initial pump pulses and the frequency ω3 excited by the probe pulse after the waiting time. This allows the observation of coupling between different vibrational modes; because of its extremely fine time resolution, it can be used to monitor molecular dynamics on a picosecond timescale. It is still a largely unexplored technique and is becoming increasingly popular for fundamental research.

As with two-dimensional nuclear magnetic resonance (2DNMR) spectroscopy, this technique spreads the spectrum in two dimensions and allows for the observation of cross peaks that contain information on the coupling between different modes. In contrast to 2DNMR, nonlinear two-dimensional infrared spectroscopy also involves the excitation to overtones. These excitations result in excited state absorption peaks located below the diagonal and cross peaks. In 2DNMR, two distinct techniques, COSY and NOESY, are frequently used. The cross peaks in the first are related to the scalar coupling, while in the latter they are related to the spin transfer between different nuclei. In nonlinear two-dimensional infrared spectroscopy, analogs have been drawn to these 2DNMR techniques. Nonlinear two-dimensional infrared spectroscopy with zero waiting time corresponds to COSY, and nonlinear two-dimensional infrared spectroscopy with finite waiting time allowing vibrational population transfer corresponds to NOESY. The COSY variant of nonlinear two-dimensional infrared spectroscopy has been used for determination of the secondary structure content of proteins.[41]

See also Edit

References Edit

  1. ^ Zeitler JA, Taday PF, Newnham DA, Pepper M, Gordon KC, Rades T (February 2007). "Terahertz pulsed spectroscopy and imaging in the pharmaceutical setting--a review". The Journal of Pharmacy and Pharmacology. 59 (2): 209–23. doi:10.1211/jpp.59.2.0008. PMID 17270075. S2CID 34705104.
  2. ^ Atkins P, de Paula J (2002). Physical Chemistry (7th ed.). New York: W.H.Freeman. p. 513. ISBN 0-7167-3539-3.|quote=the electric dipole moment of the molecule must change when the atoms are displaced relative to one another
  3. ^ Schrader B (1995). Infrared and Raman Spectroscopy: Methods and Applications. New York: VCH, Weinheim. p. 787. ISBN 978-3-527-26446-9.
  4. ^ a b Harwood LM, Moody CJ (1989). Experimental organic chemistry: Principles and Practice (Illustrated ed.). Wiley-Blackwell. p. 292. ISBN 978-0-632-02017-1.
  5. ^ Shadman S, Rose C, Yalin AP (2016). "Open-path cavity ring-down spectroscopy sensor for atmospheric ammonia". Applied Physics B. 122 (7): 194. Bibcode:2016ApPhB.122..194S. doi:10.1007/s00340-016-6461-5. S2CID 123834102.
  6. ^ a b Chromatography/Fourier transform infrared spectroscopy and its applications, by Robert White, p7
  7. ^ H M Pollock and S G Kazarian, Microspectroscopy in the Mid-Infrared, in Encyclopedia of Analytical Chemistry (Robert A. Meyers, Ed, 1-26 (2014), John Wiley & Sons Ltd,
  8. ^ Pollock Hubert M (2014). "Microspectroscopy in the Mid-Infrared". Encyclopedia of Analytical Chemistry. pp. 1–26. doi:10.1002/9780470027318.a5609.pub2. ISBN 9780470027318.
  9. ^ H M Pollock and D A Smith, The use of near-field probes for vibrational spectroscopy and photothermal imaging, in Handbook of vibrational spectroscopy, J.M. Chalmers and P.R. Griffiths (eds), John Wiley & Sons Ltd, Vol. 2, pp. 1472 - 1492 (2002)
  10. ^ a b Krivanek OL, Lovejoy TC, Dellby N, Aoki T, Carpenter RW, Rez P, et al. (October 2014). "Vibrational spectroscopy in the electron microscope". Nature. 514 (7521): 209–12. Bibcode:2014Natur.514..209K. doi:10.1038/nature13870. PMID 25297434. S2CID 4467249.
  11. ^ Idrobo JC, Lupini AR, Feng T, Unocic RR, Walden FS, Gardiner DS, et al. (March 2018). "Temperature Measurement by a Nanoscale Electron Probe Using Energy Gain and Loss Spectroscopy". Physical Review Letters. 120 (9): 095901. Bibcode:2018PhRvL.120i5901I. doi:10.1103/PhysRevLett.120.095901. PMID 29547334.
  12. ^ Lagos MJ, Batson PE (July 2018). "Thermometry with Subnanometer Resolution in the Electron Microscope Using the Principle of Detailed Balancing". Nano Letters. 18 (7): 4556–4563. Bibcode:2018NanoL..18.4556L. doi:10.1021/acs.nanolett.8b01791. PMID 29874456. S2CID 206748146.
  13. ^ Hachtel JA, Huang J, Popovs I, Jansone-Popova S, Keum JK, Jakowski J, et al. (February 2019). "Identification of site-specific isotopic labels by vibrational spectroscopy in the electron microscope". Science. 363 (6426): 525–528. Bibcode:2019Sci...363..525H. doi:10.1126/science.aav5845. PMID 30705191.
  14. ^ Hage FS, Nicholls RJ, Yates JR, McCulloch DG, Lovejoy TC, Dellby N, et al. (June 2018). "Nanoscale momentum-resolved vibrational spectroscopy". Science Advances. 4 (6): eaar7495. Bibcode:2018SciA....4.7495H. doi:10.1126/sciadv.aar7495. PMC 6018998. PMID 29951584.
  15. ^ Senga R, Suenaga K, Barone P, Morishita S, Mauri F, Pichler T (September 2019). "Position and momentum mapping of vibrations in graphene nanostructures". Nature. 573 (7773): 247–250. arXiv:1812.08294. Bibcode:2019Natur.573..247S. doi:10.1038/s41586-019-1477-8. PMID 31406319. S2CID 118999071.
  16. ^ Lagos MJ, Trügler A, Hohenester U, Batson PE (March 2017). "Mapping vibrational surface and bulk modes in a single nanocube". Nature. 543 (7646): 529–532. Bibcode:2017Natur.543..529L. doi:10.1038/nature21699. PMID 28332537. S2CID 4459728.
  17. ^ Govyadinov AA, Konečná A, Chuvilin A, Vélez S, Dolado I, Nikitin AY, et al. (July 2017). "Probing low-energy hyperbolic polaritons in van der Waals crystals with an electron microscope". Nature Communications. 8 (1): 95. arXiv:1611.05371. Bibcode:2017NatCo...8...95G. doi:10.1038/s41467-017-00056-y. PMC 5522439. PMID 28733660.
  18. ^ Venkatraman K, Levin BD, March K, Rez P, Crozier PA (2019). "Vibrational spectroscopy at atomic resolution with electron impact scattering". Nature Physics. 15 (12): 1237–1241. arXiv:1812.08895. Bibcode:2019NatPh..15.1237V. doi:10.1038/s41567-019-0675-5. S2CID 119452520.
  19. ^ Henschel H, Andersson AT, Jespers W, Mehdi Ghahremanpour M, van der Spoel D (May 2020). "Theoretical Infrared Spectra: Quantitative Similarity Measures and Force Fields". Journal of Chemical Theory and Computation. 16 (5): 3307–3315. doi:10.1021/acs.jctc.0c00126. PMC 7304875. PMID 32271575.
  20. ^ a b c Smith JG (2011). . In Hodge T, Nemmers D, Klein J (eds.). Organic chemistry (3rd ed.). New York, NY: McGraw-Hill. pp. 463–488. ISBN 978-0-07-337562-5. Archived from the original (Book) on 2018-06-28. Retrieved 2018-06-30.
  21. ^ Badger R (1934). "A Relation Between Internuclear Distances and Bond Force Constants" (PDF). The Journal of Chemical Physics. 2 (3): 128. Bibcode:1934JChPh...2..128B. doi:10.1063/1.1749433.
  22. ^ Boosa, Venu; Varimalla, Shirisha; Dumpalapally, Mahesh; Gutta, Naresh; Velisoju, Vijay Kumar; Nama, Narender; Akula, Venugopal (2021-09-05). "Influence of Brønsted acid sites on chemoselective synthesis of pyrrolidones over H-ZSM-5 supported copper catalyst". Applied Catalysis B: Environmental. 292: 120177. doi:10.1016/j.apcatb.2021.120177. ISSN 0926-3373.
  23. ^ Fajdek-Bieda, Anna; Wróblewska, Agnieszka; Miądlicki, Piotr; Tołpa, Jadwiga; Michalkiewicz, Beata (2021-08-01). "Clinoptilolite as a natural, active zeolite catalyst for the chemical transformations of geraniol". Reaction Kinetics, Mechanisms and Catalysis. 133 (2): 997–1011. doi:10.1007/s11144-021-02027-3. ISSN 1878-5204. S2CID 236149569.
  24. ^ Azancot, Lola; Bobadilla, Luis F.; Centeno, Miguel A.; Odriozola, José A. (2021-05-15). "IR spectroscopic insights into the coking-resistance effect of potassium on nickel-based catalyst during dry reforming of methane". Applied Catalysis B: Environmental. 285: 119822. doi:10.1016/j.apcatb.2020.119822. hdl:10261/251396. ISSN 0926-3373. S2CID 232627637.
  25. ^ Nuguid, Rob Jeremiah G.; Elsener, Martin; Ferri, Davide; Kröcher, Oliver (2021-12-05). "Operando diffuse reflectance infrared detection of cyanide intermediate species during the reaction of formaldehyde with ammonia over V2O5/WO3-TiO2". Applied Catalysis B: Environmental. 298: 120629. doi:10.1016/j.apcatb.2021.120629. ISSN 0926-3373.
  26. ^ "Infrared Spectroscopy". ColourLex. 12 March 2021. Retrieved 19 January 2022.
  27. ^ Derrick MR, Stulik D, Landry JM (2000). "Infrared Spectroscopy in Conservation Science, Scientific Tools for Conservation" (PDF). Getty Publications. Retrieved 11 December 2015. {{cite journal}}: Cite journal requires |journal= (help)
  28. ^ Ricciardi P. "Unlocking the secrets of illuminated manuscripts". Retrieved 11 December 2015.
  29. ^ Lau WS (1999). Infrared characterization for microelectronics. World Scientific. ISBN 978-981-02-2352-6.
  30. ^ Osborne BG (2006). "Near-Infrared Spectroscopy in Food Analysis". Encyclopedia of Analytical Chemistry. John Wiley & Sons. doi:10.1002/9780470027318.a1018. ISBN 9780470027318.
  31. ^ Villar A, Gorritxategi E, Aranzabe E, Fernández S, Otaduy D, Fernández LA (December 2012). "Low-cost visible-near infrared sensor for on-line monitoring of fat and fatty acids content during the manufacturing process of the milk". Food Chemistry. 135 (4): 2756–60. doi:10.1016/j.foodchem.2012.07.074. PMID 22980869.
  32. ^ www.TRMThemes.com, TRM Theme by. "Infrared (IR) / Optical Based Archives - Heath Consultants". Heath Consultants. Retrieved 2016-04-12.
  33. ^ "NASA Ames PAH IR Spectroscopic Database". The Astrophysics & Astrochemistry Laboratory. 12 September 2019. Retrieved 19 January 2022.
  34. ^ Hoover R (February 21, 2014). "Need to Track Organic Nano-Particles Across the Universe? NASA's Got an App for That". NASA. Retrieved February 22, 2014.
  35. ^ Becker, Wolfgang; Sachsenheimer, Kerstin; Klemenz, Melanie (2017). "Detection of Black Plastics in the Middle Infrared Spectrum (MIR) Using Photon Up-Conversion Technique for Polymer Recycling Process". Polymers. MDPI. 9 (9). doi:10.3390/polym9090435. PMC 6418689. PMID 30965736.
  36. ^ Perry, Tekla S. (2017-03-14). "What Happened When We Took the SCiO Food Analyzer Grocery Shopping". IEEE Spectrum: Technology, Engineering, and Science News. Retrieved 2017-03-23.
  37. ^ Coates, John (18 June 2014). "A Review of New Small-Scale Technologies for Near Infrared Measurements". American Pharmaceutical Review. Coates Consulting LLC. Retrieved 2017-03-23.
  38. ^ Kohli KK, Davies G, Vinh NQ, West D, Estreicher SK, Gregorkiewicz T, et al. (June 2006). "Isotope dependence of the lifetime of the vibration of oxygen in silicon". Physical Review Letters. 96 (22): 225503. Bibcode:2006PhRvL..96v5503K. doi:10.1103/PhysRevLett.96.225503. PMID 16803320.
  39. ^ Hamm P, Lim MH, Hochstrasser RM (1998). "Structure of the amide I band of peptides measured by femtosecond nonlinear-infrared spectroscopy". J. Phys. Chem. B. 102 (31): 6123–6138. doi:10.1021/jp9813286.
  40. ^ Mukamel S (2000). "Multidimensional femtosecond correlation spectroscopies of electronic and vibrational excitations". Annual Review of Physical Chemistry. 51 (1): 691–729. Bibcode:2000ARPC...51..691M. doi:10.1146/annurev.physchem.51.1.691. PMID 11031297. S2CID 31230696.
  41. ^ Demirdöven N, Cheatum CM, Chung HS, Khalil M, Knoester J, Tokmakoff A (June 2004). "Two-dimensional infrared spectroscopy of antiparallel beta-sheet secondary structure" (PDF). Journal of the American Chemical Society. 126 (25): 7981–90. doi:10.1021/ja049811j. PMID 15212548.

External links Edit

infrared, spectroscopy, table, spectroscopy, data, infrared, spectroscopy, correlation, table, spectroscopy, vibrational, spectroscopy, measurement, interaction, infrared, radiation, with, matter, absorption, emission, reflection, used, study, identify, chemic. For a table of IR spectroscopy data see infrared spectroscopy correlation table Infrared spectroscopy IR spectroscopy or vibrational spectroscopy is the measurement of the interaction of infrared radiation with matter by absorption emission or reflection It is used to study and identify chemical substances or functional groups in solid liquid or gaseous forms It can be used to characterize new materials or identify and verify known and unknown samples The method or technique of infrared spectroscopy is conducted with an instrument called an infrared spectrometer or spectrophotometer which produces an infrared spectrum An IR spectrum can be visualized in a graph of infrared light absorbance or transmittance on the vertical axis vs frequency wavenumber or wavelength on the horizontal axis Typical units of wavenumber used in IR spectra are reciprocal centimeters with the symbol cm 1 Units of IR wavelength are commonly given in micrometers formerly called microns symbol mm which are related to the wavenumber in a reciprocal way A common laboratory instrument that uses this technique is a Fourier transform infrared FTIR spectrometer Two dimensional IR is also possible as discussed below OVIRS instrument of the OSIRIS REx probe is a visible and infrared spectrometerThe infrared portion of the electromagnetic spectrum is usually divided into three regions the near mid and far infrared named for their relation to the visible spectrum The higher energy near IR approximately 14 000 4 000 cm 1 0 7 2 5 mm wavelength can excite overtone or combination modes of molecular vibrations The mid infrared approximately 4 000 400 cm 1 2 5 25 mm is generally used to study the fundamental vibrations and associated rotational vibrational structure The far infrared approximately 400 10 cm 1 25 1 000 mm has low energy and may be used for rotational spectroscopy and low frequency vibrations The region from 2 130 cm 1 bordering the microwave region is considered the terahertz region and may probe intermolecular vibrations 1 The names and classifications of these subregions are conventions and are only loosely based on the relative molecular or electromagnetic properties Contents 1 Theory 1 1 Number of vibrational modes 2 Practical IR spectroscopy 2 1 Sample preparation 2 1 1 Gas samples 2 1 2 Liquid samples 2 1 3 Solid samples 2 2 Comparing to a reference 2 3 FTIR 2 4 Infrared microscopy 2 5 Other methods in molecular vibrational spectroscopy 2 6 Computational infrared microscopy 3 Absorption bands 3 1 Regions 3 2 Badger s rule 4 Uses and applications 5 Isotope effects 6 Two dimensional IR 7 See also 8 References 9 External linksTheory Edit Sample of an IR spec reading this one is from bromomethane CH3Br showing peaks around 3000 1300 and 1000 cm 1 on the horizontal axis Infrared spectroscopy exploits the fact that molecules absorb frequencies that are characteristic of their structure These absorptions occur at resonant frequencies i e the frequency of the absorbed radiation matches the vibrational frequency The energies are affected by the shape of the molecular potential energy surfaces the masses of the atoms and the associated vibronic coupling 3D animation of the symmetric stretch compress mode of the C H bonds of bromomethaneIn particular in the Born Oppenheimer and harmonic approximations i e when the molecular Hamiltonian corresponding to the electronic ground state can be approximated by a harmonic oscillator in the neighborhood of the equilibrium molecular geometry the resonant frequencies are associated with the normal modes of vibration corresponding to the molecular electronic ground state potential energy surface The resonant frequencies are also related to the strength of the bond and the mass of the atoms at either end of it Thus the frequency of the vibrations are associated with a particular normal mode of motion and a particular bond type Number of vibrational modes Edit In order for a vibrational mode in a sample to be IR active it must be associated with changes in the molecular dipole moment A permanent dipole is not necessary as the rule requires only a change in dipole moment 2 A molecule can vibrate in many ways and each way is called a vibrational mode For molecules with N number of atoms geometrically linear molecules have 3N 5 degrees of vibrational modes whereas nonlinear molecules have 3N 6 degrees of vibrational modes also called vibrational degrees of freedom As examples linear carbon dioxide CO2 has 3 3 5 4 while non linear water H2O has only 3 3 6 3 Stretching and bending oscillations of the CO2 carbon dioxide molecule Upper left symmetric stretching Upper right antisymmetric stretching Lower line degenerate pair of bending modes Simple diatomic molecules have only one bond and only one vibrational band If the molecule is symmetrical e g N2 the band is not observed in the IR spectrum but only in the Raman spectrum Asymmetrical diatomic molecules e g carbon monoxide CO absorb in the IR spectrum More complex molecules have many bonds and their vibrational spectra are correspondingly more complex i e big molecules have many peaks in their IR spectra The atoms in a CH2X2 group commonly found in organic compounds and where X can represent any other atom can vibrate in nine different ways Six of these vibrations involve only the CH2 portion two stretching modes n symmetric ns and antisymmetric nas and four bending modes scissoring d rocking r wagging w and twisting t as shown below Structures that do not have the two additional X groups attached have fewer modes because some modes are defined by specific relationships to those other attached groups For example in water the rocking wagging and twisting modes do not exist because these types of motions of the H atoms represent simple rotation of the whole molecule rather than vibrations within it In case of more complex molecules out of plane g vibrational modes can be also present 3 SymmetryDirection Symmetric AntisymmetricRadial Symmetric stretching ns Antisymmetric stretching nas Latitudinal Scissoring d Rocking r Longitudinal Wagging w Twisting t These figures do not represent the recoil of the C atoms which though necessarily present to balance the overall movements of the molecule are much smaller than the movements of the lighter H atoms The simplest and most important or fundamental IR bands arise from the excitations of normal modes the simplest distortions of the molecule from the ground state with vibrational quantum number v 0 to the first excited state with vibrational quantum number v 1 In some cases overtone bands are observed An overtone band arises from the absorption of a photon leading to a direct transition from the ground state to the second excited vibrational state v 2 Such a band appears at approximately twice the energy of the fundamental band for the same normal mode Some excitations so called combination modes involve simultaneous excitation of more than one normal mode The phenomenon of Fermi resonance can arise when two modes are similar in energy Fermi resonance results in an unexpected shift in energy and intensity of the bands etc citation needed Practical IR spectroscopy EditThe infrared spectrum of a sample is recorded by passing a beam of infrared light through the sample When the frequency of the IR is the same as the vibrational frequency of a bond or collection of bonds absorption occurs Examination of the transmitted light reveals how much energy was absorbed at each frequency or wavelength This measurement can be achieved by scanning the wavelength range using a monochromator Alternatively the entire wavelength range is measured using a Fourier transform instrument and then a transmittance or absorbance spectrum is generated using a dedicated procedure This technique is commonly used for analyzing samples with covalent bonds Simple spectra are obtained from samples with few IR active bonds and high levels of purity More complex molecular structures lead to more absorption bands and more complex spectra Typical IR solution cell The windows are CaF2 Sample preparation Edit Gas samples Edit Gaseous samples require a sample cell with a long pathlength to compensate for the diluteness The pathlength of the sample cell depends on the concentration of the compound of interest A simple glass tube with length of 5 to 10 cm equipped with infrared transparent windows at both ends of the tube can be used for concentrations down to several hundred ppm Sample gas concentrations well below ppm can be measured with a White s cell in which the infrared light is guided with mirrors to travel through the gas White s cells are available with optical pathlength starting from 0 5 m up to hundred meters Liquid samples Edit Liquid samples can be sandwiched between two plates of a salt commonly sodium chloride or common salt although a number of other salts such as potassium bromide or calcium fluoride are also used 4 The plates are transparent to the infrared light and do not introduce any lines onto the spectra With increasing technology in computer filtering and manipulation of the results samples in solution can now be measured accurately water produces a broad absorbance across the range of interest and thus renders the spectra unreadable without this computer treatment Solid samples Edit Solid samples can be prepared in a variety of ways One common method is to crush the sample with an oily mulling agent usually mineral oil Nujol A thin film of the mull is applied onto salt plates and measured The second method is to grind a quantity of the sample with a specially purified salt usually potassium bromide finely to remove scattering effects from large crystals This powder mixture is then pressed in a mechanical press to form a translucent pellet through which the beam of the spectrometer can pass 4 A third technique is the cast film technique which is used mainly for polymeric materials The sample is first dissolved in a suitable non hygroscopic solvent A drop of this solution is deposited on the surface of a KBr or NaCl cell The solution is then evaporated to dryness and the film formed on the cell is analysed directly Care is important to ensure that the film is not too thick otherwise light cannot pass through This technique is suitable for qualitative analysis The final method is to use microtomy to cut a thin 20 100 mm film from a solid sample This is one of the most important ways of analysing failed plastic products for example because the integrity of the solid is preserved In photoacoustic spectroscopy the need for sample treatment is minimal The sample liquid or solid is placed into the sample cup which is inserted into the photoacoustic cell which is then sealed for the measurement The sample may be one solid piece powder or basically in any form for the measurement For example a piece of rock can be inserted into the sample cup and the spectrum measured from it A useful way of analyzing solid samples without the need for cutting samples uses ATR or attenuated total reflectance spectroscopy Using this approach samples are pressed against the face of a single crystal The infrared radiation passes through the crystal and only interacts with the sample at the interface between the two materials Comparing to a reference Edit Schematics of a two beam absorption spectrometer A beam of infrared light is produced passed through an interferometer not shown and then split into two separate beams One is passed through the sample the other passed through a reference The beams are both reflected back towards a detector however first they pass through a splitter which quickly alternates which of the two beams enters the detector The two signals are then compared and a printout is obtained This two beam setup gives accurate spectra even if the intensity of the light source drifts over time It is typical to record spectrum of both the sample and a reference This step controls for a number of variables e g infrared detector which may affect the spectrum The reference measurement makes it possible to eliminate the instrument influence The appropriate reference depends on the measurement and its goal The simplest reference measurement is to simply remove the sample replacing it by air However sometimes a different reference is more useful For example if the sample is a dilute solute dissolved in water in a beaker then a good reference measurement might be to measure pure water in the same beaker Then the reference measurement would cancel out not only all the instrumental properties like what light source is used but also the light absorbing and light reflecting properties of the water and beaker and the final result would just show the properties of the solute at least approximately A common way to compare to a reference is sequentially first measure the reference then replace the reference by the sample and measure the sample This technique is not perfectly reliable if the infrared lamp is a bit brighter during the reference measurement then a bit dimmer during the sample measurement the measurement will be distorted More elaborate methods such as a two beam setup see figure can correct for these types of effects to give very accurate results The Standard addition method can be used to statistically cancel these errors Nevertheless among different absorption based techniques which are used for gaseous species detection Cavity ring down spectroscopy CRDS can be used as a calibration free method The fact that CRDS is based on the measurements of photon life times and not the laser intensity makes it needless for any calibration and comparison with a reference 5 Some instruments also automatically identify the substance being measured from a store of thousands of reference spectra held in storage FTIR Edit Main article Fourier transform infrared spectroscopy An interferogram from an FTIR measurement The horizontal axis is the position of the mirror and the vertical axis is the amount of light detected This is the raw data which can be Fourier transformed to get the actual spectrum Fourier transform infrared FTIR spectroscopy is a measurement technique that allows one to record infrared spectra Infrared light is guided through an interferometer and then through the sample or vice versa A moving mirror inside the apparatus alters the distribution of infrared light that passes through the interferometer The signal directly recorded called an interferogram represents light output as a function of mirror position A data processing technique called Fourier transform turns this raw data into the desired result the sample s spectrum light output as a function of infrared wavelength or equivalently wavenumber As described above the sample s spectrum is always compared to a reference An alternate method for acquiring spectra is the dispersive or scanning monochromator method In this approach the sample is irradiated sequentially with various single wavelengths The dispersive method is more common in UV Vis spectroscopy but is less practical in the infrared than the FTIR method One reason that FTIR is favored is called Fellgett s advantage or the multiplex advantage The information at all frequencies is collected simultaneously improving both speed and signal to noise ratio Another is called Jacquinot s Throughput Advantage A dispersive measurement requires detecting much lower light levels than an FTIR measurement 6 There are other advantages as well as some disadvantages 6 but virtually all modern infrared spectrometers are FTIR instruments Infrared microscopy Edit Various forms of infrared microscopy exist These include IR versions of sub diffraction microscopy 7 8 such as IR NSOM 9 photothermal microspectroscopy Nano FTIR and atomic force microscope based infrared spectroscopy AFM IR Other methods in molecular vibrational spectroscopy Edit Infrared spectroscopy is not the only method of studying molecular vibrational spectra Raman spectroscopy involves an inelastic scattering process in which only part of the energy of an incident photon is absorbed by the molecule and the remaining part is scattered and detected The energy difference corresponds to absorbed vibrational energy The selection rules for infrared and for Raman spectroscopy are different at least for some molecular symmetries so that the two methods are complementary in that they observe vibrations of different symmetries Another method is electron energy loss spectroscopy EELS in which the energy absorbed is provided by an inelastically scattered electron rather than a photon This method is useful for studying vibrations of molecules adsorbed on a solid surface Recently high resolution EELS HREELS has emerged as a technique for performing vibrational spectroscopy in a transmission electron microscope TEM 10 In combination with the high spatial resolution of the TEM unprecedented experiments have been performed such as nano scale temperature measurements 11 12 mapping of isotopically labeled molecules 13 mapping of phonon modes in position and momentum space 14 15 vibrational surface and bulk mode mapping on nanocubes 16 and investigations of polariton modes in van der Waals crystals 17 Analysis of vibrational modes that are IR inactive but appear in inelastic neutron scattering is also possible at high spatial resolution using EELS 18 Although the spatial resolution of HREELs is very high the bands are extremely broad compared to other techniques 10 Computational infrared microscopy Edit By using computer simulations and normal mode analysis it is possible to calculate theoretical frequencies of molecules 19 Absorption bands EditIR spectroscopy is often used to identify structures because functional groups give rise to characteristic bands both in terms of intensity and position frequency The positions of these bands are summarized in correlation tables as shown below Main article Infrared spectroscopy correlation table List of main IR spectroscopy bands For example the carboxyl group will contain a C O band at 1700 cm 1 and an OH band at 3500 cm 1 total group COOH Wavenumbers listed in cm 1 Regions Edit A spectrograph is often interpreted as having two regions 20 functional group region 1 500 cm 1 displaystyle geq 1 500 text cm 1 In the functional region there are one to a few troughs per functional group 20 fingerprint region lt 1 500 cm 1 displaystyle lt 1 500 text cm 1 In the fingerprint region there are many troughs which form an intricate pattern which can be used like a fingerprint to determine the compound 20 Badger s rule Edit For many kinds of samples the assignments are known i e which bond deformation s are associated with which frequency In such cases further information can be gleaned about the strength on a bond relying on the empirical guideline called Badger s rule Originally published by Richard McLean Badger in 1934 21 this rule states that the strength of a bond in terms of force constant correlates with the bond length That is increase in bond strength leads to corresponding bond shortening and vice versa Uses and applications Edit Infrared spectrophotometer used to analyze the diethyltoluamide insect repellent 1960 US Food and Drug Administration scientist uses portable near infrared spectroscopy device to detect potentially illegal substancesInfrared spectroscopy is a simple and reliable technique widely used in both organic and inorganic chemistry in research and industry In catalysis research it is a very useful tool to characterize the catalyst 22 23 24 as well as to detect intermediates 25 and products during the catalytic reaction It is used in quality control dynamic measurement and monitoring applications such as the long term unattended measurement of CO2 concentrations in greenhouses and growth chambers by infrared gas analyzers It is also used in forensic analysis in both criminal and civil cases for example in identifying polymer degradation It can be used in determining the blood alcohol content of a suspected drunk driver IR spectroscopy has been successfully used in analysis and identification of pigments in paintings 26 and other art objects 27 such as illuminated manuscripts 28 Infrared spectroscopy is also useful in measuring the degree of polymerization in polymer manufacture Changes in the character or quantity of a particular bond are assessed by measuring at a specific frequency over time Modern research instruments can take infrared measurements across the range of interest as frequently as 32 times a second This can be done whilst simultaneous measurements are made using other techniques This makes the observations of chemical reactions and processes quicker and more accurate Infrared spectroscopy has also been successfully utilized in the field of semiconductor microelectronics 29 for example infrared spectroscopy can be applied to semiconductors like silicon gallium arsenide gallium nitride zinc selenide amorphous silicon silicon nitride etc Another important application of infrared spectroscopy is in the food industry to measure the concentration of various compounds in different food products 30 31 The instruments are now small and can be transported even for use in field trials Infrared spectroscopy is also used in gas leak detection devices such as the DP IR and EyeCGAs 32 These devices detect hydrocarbon gas leaks in the transportation of natural gas and crude oil In February 2014 NASA announced a greatly upgraded database 33 based on IR spectroscopy for tracking polycyclic aromatic hydrocarbons PAHs in the universe According to scientists more than 20 of the carbon in the universe may be associated with PAHs possible starting materials for the formation of life PAHs seem to have been formed shortly after the Big Bang are widespread throughout the universe and are associated with new stars and exoplanets 34 Infrared spectroscopy is an important analysis method in the recycling process of household waste plastics and a convenient stand off method to sort plastic of different polymers PET HDPE 35 Other developments include a miniature IR spectrometer that s linked to a cloud based database and suitable for personal everyday use 36 and NIR spectroscopic chips 37 that can be embedded in smartphones and various gadgets Isotope effects EditThe different isotopes in a particular species may exhibit different fine details in infrared spectroscopy For example the O O stretching frequency in reciprocal centimeters of oxyhemocyanin is experimentally determined to be 832 and 788 cm 1 for n 16O 16O and n 18O 18O respectively By considering the O O bond as a spring the frequency of absorbance can be calculated as a wavenumber frequency speed of light n 1 2 p c k m displaystyle tilde nu frac 1 2 pi c sqrt frac k mu where k is the spring constant for the bond c is the speed of light and m is the reduced mass of the A B system m m A m B m A m B displaystyle mu frac m mathrm A m mathrm B m mathrm A m mathrm B m i displaystyle m i is the mass of atom i displaystyle i The reduced masses for 16O 16O and 18O 18O can be approximated as 8 and 9 respectively Thus n 16 O n 18 O 9 8 832 788 displaystyle frac tilde nu 16 mathrm O tilde nu 18 mathrm O sqrt frac 9 8 approx frac 832 788 The effect of isotopes both on the vibration and the decay dynamics has been found to be stronger than previously thought In some systems such as silicon and germanium the decay of the anti symmetric stretch mode of interstitial oxygen involves the symmetric stretch mode with a strong isotope dependence For example it was shown that for a natural silicon sample the lifetime of the anti symmetric vibration is 11 4 ps When the isotope of one of the silicon atoms is increased to 29Si the lifetime increases to 19 ps In similar manner when the silicon atom is changed to 30Si the lifetime becomes 27 ps 38 Two dimensional IR EditTwo dimensional infrared correlation spectroscopy analysis combines multiple samples of infrared spectra to reveal more complex properties By extending the spectral information of a perturbed sample spectral analysis is simplified and resolution is enhanced The 2D synchronous and 2D asynchronous spectra represent a graphical overview of the spectral changes due to a perturbation such as a changing concentration or changing temperature as well as the relationship between the spectral changes at two different wavenumbers Main article Two dimensional infrared spectroscopy Pulse Sequence used to obtain a two dimensional Fourier transform infrared spectrum The time period t 1 displaystyle tau 1 is usually referred to as the coherence time and the second time period t 2 displaystyle tau 2 is known as the waiting time The excitation frequency is obtained by Fourier transforming along the t 1 displaystyle tau 1 axis Nonlinear two dimensional infrared spectroscopy 39 40 is the infrared version of correlation spectroscopy Nonlinear two dimensional infrared spectroscopy is a technique that has become available with the development of femtosecond infrared laser pulses In this experiment first a set of pump pulses is applied to the sample This is followed by a waiting time during which the system is allowed to relax The typical waiting time lasts from zero to several picoseconds and the duration can be controlled with a resolution of tens of femtoseconds A probe pulse is then applied resulting in the emission of a signal from the sample The nonlinear two dimensional infrared spectrum is a two dimensional correlation plot of the frequency w1 that was excited by the initial pump pulses and the frequency w3 excited by the probe pulse after the waiting time This allows the observation of coupling between different vibrational modes because of its extremely fine time resolution it can be used to monitor molecular dynamics on a picosecond timescale It is still a largely unexplored technique and is becoming increasingly popular for fundamental research As with two dimensional nuclear magnetic resonance 2DNMR spectroscopy this technique spreads the spectrum in two dimensions and allows for the observation of cross peaks that contain information on the coupling between different modes In contrast to 2DNMR nonlinear two dimensional infrared spectroscopy also involves the excitation to overtones These excitations result in excited state absorption peaks located below the diagonal and cross peaks In 2DNMR two distinct techniques COSY and NOESY are frequently used The cross peaks in the first are related to the scalar coupling while in the latter they are related to the spin transfer between different nuclei In nonlinear two dimensional infrared spectroscopy analogs have been drawn to these 2DNMR techniques Nonlinear two dimensional infrared spectroscopy with zero waiting time corresponds to COSY and nonlinear two dimensional infrared spectroscopy with finite waiting time allowing vibrational population transfer corresponds to NOESY The COSY variant of nonlinear two dimensional infrared spectroscopy has been used for determination of the secondary structure content of proteins 41 See also EditApplied spectroscopy Astrochemistry Atomic and molecular astrophysics Atomic force microscopy based infrared spectroscopy AFM IR Cosmochemistry Far infrared astronomy Forensic chemistry Forensic engineering Forensic polymer engineering Infrared astronomy Infrared microscopy Infrared multiphoton dissociation Infrared photodissociation spectroscopy Infrared spectroscopy correlation table Infrared spectroscopy of metal carbonyls Near infrared spectroscopy Nuclear resonance vibrational spectroscopy Photothermal microspectroscopy Raman spectroscopy Rotational vibrational spectroscopy Time resolved spectroscopy Vibrational spectroscopy of linear moleculesReferences Edit Zeitler JA Taday PF Newnham DA Pepper M Gordon KC Rades T February 2007 Terahertz pulsed spectroscopy and imaging in the pharmaceutical setting a review The Journal of Pharmacy and Pharmacology 59 2 209 23 doi 10 1211 jpp 59 2 0008 PMID 17270075 S2CID 34705104 Atkins P de Paula J 2002 Physical Chemistry 7th ed New York W H Freeman p 513 ISBN 0 7167 3539 3 quote the electric dipole moment of the molecule must change when the atoms are displaced relative to one another Schrader B 1995 Infrared and Raman Spectroscopy Methods and Applications New York VCH Weinheim p 787 ISBN 978 3 527 26446 9 a b Harwood LM Moody CJ 1989 Experimental organic chemistry Principles and Practice Illustrated ed Wiley Blackwell p 292 ISBN 978 0 632 02017 1 Shadman S Rose C Yalin AP 2016 Open path cavity ring down spectroscopy sensor for atmospheric ammonia Applied Physics B 122 7 194 Bibcode 2016ApPhB 122 194S doi 10 1007 s00340 016 6461 5 S2CID 123834102 a b Chromatography Fourier transform infrared spectroscopy and its applications by Robert White p7 H M Pollock and S G Kazarian Microspectroscopy in the Mid Infrared in Encyclopedia of Analytical Chemistry Robert A Meyers Ed 1 26 2014 John Wiley amp Sons Ltd Pollock Hubert M 2014 Microspectroscopy in the Mid Infrared Encyclopedia of Analytical Chemistry pp 1 26 doi 10 1002 9780470027318 a5609 pub2 ISBN 9780470027318 H M Pollock and D A Smith The use of near field probes for vibrational spectroscopy and photothermal imaging in Handbook of vibrational spectroscopy J M Chalmers and P R Griffiths eds John Wiley amp Sons Ltd Vol 2 pp 1472 1492 2002 a b Krivanek OL Lovejoy TC Dellby N Aoki T Carpenter RW Rez P et al October 2014 Vibrational spectroscopy in the electron microscope Nature 514 7521 209 12 Bibcode 2014Natur 514 209K doi 10 1038 nature13870 PMID 25297434 S2CID 4467249 Idrobo JC Lupini AR Feng T Unocic RR Walden FS Gardiner DS et al March 2018 Temperature Measurement by a Nanoscale Electron Probe Using Energy Gain and Loss Spectroscopy Physical Review Letters 120 9 095901 Bibcode 2018PhRvL 120i5901I doi 10 1103 PhysRevLett 120 095901 PMID 29547334 Lagos MJ Batson PE July 2018 Thermometry with Subnanometer Resolution in the Electron Microscope Using the Principle of Detailed Balancing Nano Letters 18 7 4556 4563 Bibcode 2018NanoL 18 4556L doi 10 1021 acs nanolett 8b01791 PMID 29874456 S2CID 206748146 Hachtel JA Huang J Popovs I Jansone Popova S Keum JK Jakowski J et al February 2019 Identification of site specific isotopic labels by vibrational spectroscopy in the electron microscope Science 363 6426 525 528 Bibcode 2019Sci 363 525H doi 10 1126 science aav5845 PMID 30705191 Hage FS Nicholls RJ Yates JR McCulloch DG Lovejoy TC Dellby N et al June 2018 Nanoscale momentum resolved vibrational spectroscopy Science Advances 4 6 eaar7495 Bibcode 2018SciA 4 7495H doi 10 1126 sciadv aar7495 PMC 6018998 PMID 29951584 Senga R Suenaga K Barone P Morishita S Mauri F Pichler T September 2019 Position and momentum mapping of vibrations in graphene nanostructures Nature 573 7773 247 250 arXiv 1812 08294 Bibcode 2019Natur 573 247S doi 10 1038 s41586 019 1477 8 PMID 31406319 S2CID 118999071 Lagos MJ Trugler A Hohenester U Batson PE March 2017 Mapping vibrational surface and bulk modes in a single nanocube Nature 543 7646 529 532 Bibcode 2017Natur 543 529L doi 10 1038 nature21699 PMID 28332537 S2CID 4459728 Govyadinov AA Konecna A Chuvilin A Velez S Dolado I Nikitin AY et al July 2017 Probing low energy hyperbolic polaritons in van der Waals crystals with an electron microscope Nature Communications 8 1 95 arXiv 1611 05371 Bibcode 2017NatCo 8 95G doi 10 1038 s41467 017 00056 y PMC 5522439 PMID 28733660 Venkatraman K Levin BD March K Rez P Crozier PA 2019 Vibrational spectroscopy at atomic resolution with electron impact scattering Nature Physics 15 12 1237 1241 arXiv 1812 08895 Bibcode 2019NatPh 15 1237V doi 10 1038 s41567 019 0675 5 S2CID 119452520 Henschel H Andersson AT Jespers W Mehdi Ghahremanpour M van der Spoel D May 2020 Theoretical Infrared Spectra Quantitative Similarity Measures and Force Fields Journal of Chemical Theory and Computation 16 5 3307 3315 doi 10 1021 acs jctc 0c00126 PMC 7304875 PMID 32271575 a b c Smith JG 2011 Chapter 13 Mass Spectrometry and Infrared Spectroscopy In Hodge T Nemmers D Klein J eds Organic chemistry 3rd ed New York NY McGraw Hill pp 463 488 ISBN 978 0 07 337562 5 Archived from the original Book on 2018 06 28 Retrieved 2018 06 30 Badger R 1934 A Relation Between Internuclear Distances and Bond Force Constants PDF The Journal of Chemical Physics 2 3 128 Bibcode 1934JChPh 2 128B doi 10 1063 1 1749433 Boosa Venu Varimalla Shirisha Dumpalapally Mahesh Gutta Naresh Velisoju Vijay Kumar Nama Narender Akula Venugopal 2021 09 05 Influence of Bronsted acid sites on chemoselective synthesis of pyrrolidones over H ZSM 5 supported copper catalyst Applied Catalysis B Environmental 292 120177 doi 10 1016 j apcatb 2021 120177 ISSN 0926 3373 Fajdek Bieda Anna Wroblewska Agnieszka Miadlicki Piotr Tolpa Jadwiga Michalkiewicz Beata 2021 08 01 Clinoptilolite as a natural active zeolite catalyst for the chemical transformations of geraniol Reaction Kinetics Mechanisms and Catalysis 133 2 997 1011 doi 10 1007 s11144 021 02027 3 ISSN 1878 5204 S2CID 236149569 Azancot Lola Bobadilla Luis F Centeno Miguel A Odriozola Jose A 2021 05 15 IR spectroscopic insights into the coking resistance effect of potassium on nickel based catalyst during dry reforming of methane Applied Catalysis B Environmental 285 119822 doi 10 1016 j apcatb 2020 119822 hdl 10261 251396 ISSN 0926 3373 S2CID 232627637 Nuguid Rob Jeremiah G Elsener Martin Ferri Davide Krocher Oliver 2021 12 05 Operando diffuse reflectance infrared detection of cyanide intermediate species during the reaction of formaldehyde with ammonia over V2O5 WO3 TiO2 Applied Catalysis B Environmental 298 120629 doi 10 1016 j apcatb 2021 120629 ISSN 0926 3373 Infrared Spectroscopy ColourLex 12 March 2021 Retrieved 19 January 2022 Derrick MR Stulik D Landry JM 2000 Infrared Spectroscopy in Conservation Science Scientific Tools for Conservation PDF Getty Publications Retrieved 11 December 2015 a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help Ricciardi P Unlocking the secrets of illuminated manuscripts Retrieved 11 December 2015 Lau WS 1999 Infrared characterization for microelectronics World Scientific ISBN 978 981 02 2352 6 Osborne BG 2006 Near Infrared Spectroscopy in Food Analysis Encyclopedia of Analytical Chemistry John Wiley amp Sons doi 10 1002 9780470027318 a1018 ISBN 9780470027318 Villar A Gorritxategi E Aranzabe E Fernandez S Otaduy D Fernandez LA December 2012 Low cost visible near infrared sensor for on line monitoring of fat and fatty acids content during the manufacturing process of the milk Food Chemistry 135 4 2756 60 doi 10 1016 j foodchem 2012 07 074 PMID 22980869 www TRMThemes com TRM Theme by Infrared IR Optical Based Archives Heath Consultants Heath Consultants Retrieved 2016 04 12 NASA Ames PAH IR Spectroscopic Database The Astrophysics amp Astrochemistry Laboratory 12 September 2019 Retrieved 19 January 2022 Hoover R February 21 2014 Need to Track Organic Nano Particles Across the Universe NASA s Got an App for That NASA Retrieved February 22 2014 Becker Wolfgang Sachsenheimer Kerstin Klemenz Melanie 2017 Detection of Black Plastics in the Middle Infrared Spectrum MIR Using Photon Up Conversion Technique for Polymer Recycling Process Polymers MDPI 9 9 doi 10 3390 polym9090435 PMC 6418689 PMID 30965736 Perry Tekla S 2017 03 14 What Happened When We Took the SCiO Food Analyzer Grocery Shopping IEEE Spectrum Technology Engineering and Science News Retrieved 2017 03 23 Coates John 18 June 2014 A Review of New Small Scale Technologies for Near Infrared Measurements American Pharmaceutical Review Coates Consulting LLC Retrieved 2017 03 23 Kohli KK Davies G Vinh NQ West D Estreicher SK Gregorkiewicz T et al June 2006 Isotope dependence of the lifetime of the vibration of oxygen in silicon Physical Review Letters 96 22 225503 Bibcode 2006PhRvL 96v5503K doi 10 1103 PhysRevLett 96 225503 PMID 16803320 Hamm P Lim MH Hochstrasser RM 1998 Structure of the amide I band of peptides measured by femtosecond nonlinear infrared spectroscopy J Phys Chem B 102 31 6123 6138 doi 10 1021 jp9813286 Mukamel S 2000 Multidimensional femtosecond correlation spectroscopies of electronic and vibrational excitations Annual Review of Physical Chemistry 51 1 691 729 Bibcode 2000ARPC 51 691M doi 10 1146 annurev physchem 51 1 691 PMID 11031297 S2CID 31230696 Demirdoven N Cheatum CM Chung HS Khalil M Knoester J Tokmakoff A June 2004 Two dimensional infrared spectroscopy of antiparallel beta sheet secondary structure PDF Journal of the American Chemical Society 126 25 7981 90 doi 10 1021 ja049811j PMID 15212548 External links Edit Wikimedia Commons has media related to Infrared spectroscopy Infrared spectroscopy for organic chemists Organic compounds spectrum database Archived 2013 01 14 at the Wayback Machine Retrieved from https en wikipedia org w index php title Infrared spectroscopy amp oldid 1170409530, wikipedia, wiki, book, books, library,