Angular resolution

Angular resolution describes the ability of any image-forming device such as an optical or radio telescope, a microscope, a camera, or an eye, to distinguish small details of an object, thereby making it a major determinant of image resolution. It is used in optics applied to light waves, in antenna theory applied to radio waves, and in acoustics applied to sound waves. The colloquial use of the term "resolution" sometimes causes confusion; when an optical system is said to have a high resolution or high angular resolution, it means that the perceived distance, or actual angular distance, between resolved neighboring objects is small. The value that quantifies this property, θ, which is given by the Rayleigh criterion, is low for a system with a high resolution. The closely related term spatial resolution refers to the precision of a measurement with respect to space, which is directly connected to angular resolution in imaging instruments. The Rayleigh criterion shows that the minimum angular spread that can be resolved by an image forming system is limited by diffraction to the ratio of the wavelength of the waves to the aperture width. For this reason, high resolution imaging systems such as astronomical telescopes, long distance telephoto camera lenses and radio telescopes have large apertures.

A series of images representing the magnification of M87* with an angular size of some microarcseconds, comparable to viewing a tennis ball on the Moon (magnification from top left corner counter−clockwise to the top right corner).

Definition of terms edit

Resolving power is the ability of an imaging device to separate (i.e., to see as distinct) points of an object that are located at a small angular distance or it is the power of an optical instrument to separate far away objects, that are close together, into individual images. The term resolution or minimum resolvable distance is the minimum distance between distinguishable objects in an image, although the term is loosely used by many users of microscopes and telescopes to describe resolving power. As explained below, diffraction-limited resolution is defined by the Rayleigh criterion as the angular separation of two point sources when the maximum of each source lies in the first minimum of the diffraction pattern (Airy disk) of the other. In scientific analysis, in general, the term "resolution" is used to describe the precision with which any instrument measures and records (in an image or spectrum) any variable in the specimen or sample under study.

The Rayleigh criterion edit

 
Airy diffraction patterns generated by light from two point sources passing through a circular aperture, such as the pupil of the eye. Points far apart (top) or meeting the Rayleigh criterion (middle) can be distinguished. Points closer than the Rayleigh criterion (bottom) are difficult to distinguish.

The imaging system's resolution can be limited either by aberration or by diffraction causing blurring of the image. These two phenomena have different origins and are unrelated. Aberrations can be explained by geometrical optics and can in principle be solved by increasing the optical quality of the system. On the other hand, diffraction comes from the wave nature of light and is determined by the finite aperture of the optical elements. The lens' circular aperture is analogous to a two-dimensional version of the single-slit experiment. Light passing through the lens interferes with itself creating a ring-shape diffraction pattern, known as the Airy pattern, if the wavefront of the transmitted light is taken to be spherical or plane over the exit aperture.

The interplay between diffraction and aberration can be characterised by the point spread function (PSF). The narrower the aperture of a lens the more likely the PSF is dominated by diffraction. In that case, the angular resolution of an optical system can be estimated (from the diameter of the aperture and the wavelength of the light) by the Rayleigh criterion defined by Lord Rayleigh: two point sources are regarded as just resolved when the principal diffraction maximum (center) of the Airy disk of one image coincides with the first minimum of the Airy disk of the other,[1][2] as shown in the accompanying photos. (In the bottom photo on the right that shows the Rayleigh criterion limit, the central maximum of one point source might look as though it lies outside the first minimum of the other, but examination with a ruler verifies that the two do intersect.) If the distance is greater, the two points are well resolved and if it is smaller, they are regarded as not resolved. Rayleigh defended this criterion on sources of equal strength.[2]

Considering diffraction through a circular aperture, this translates into:

 

where θ is the angular resolution (radians), λ is the wavelength of light, and D is the diameter of the lens' aperture. The factor 1.22 is derived from a calculation of the position of the first dark circular ring surrounding the central Airy disc of the diffraction pattern. This number is more precisely 1.21966989... (OEISA245461), the first zero of the order-one Bessel function of the first kind   divided by π.

The formal Rayleigh criterion is close to the empirical resolution limit found earlier by the English astronomer W. R. Dawes, who tested human observers on close binary stars of equal brightness. The result, θ = 4.56/D, with D in inches and θ in arcseconds, is slightly narrower than calculated with the Rayleigh criterion. A calculation using Airy discs as point spread function shows that at Dawes' limit there is a 5% dip between the two maxima, whereas at Rayleigh's criterion there is a 26.3% dip.[3] Modern image processing techniques including deconvolution of the point spread function allow resolution of binaries with even less angular separation.

Using a small-angle approximation, the angular resolution may be converted into a spatial resolution, Δ, by multiplication of the angle (in radians) with the distance to the object. For a microscope, that distance is close to the focal length f of the objective. For this case, the Rayleigh criterion reads:

 .

This is the radius, in the imaging plane, of the smallest spot to which a collimated beam of light can be focused, which also corresponds to the size of smallest object that the lens can resolve.[4] The size is proportional to wavelength, λ, and thus, for example, blue light can be focused to a smaller spot than red light. If the lens is focusing a beam of light with a finite extent (e.g., a laser beam), the value of D corresponds to the diameter of the light beam, not the lens.[Note 1] Since the spatial resolution is inversely proportional to D, this leads to the slightly surprising result that a wide beam of light may be focused to a smaller spot than a narrow one. This result is related to the Fourier properties of a lens.

A similar result holds for a small sensor imaging a subject at infinity: The angular resolution can be converted to a spatial resolution on the sensor by using f as the distance to the image sensor; this relates the spatial resolution of the image to the f-number, f/#:

 .

Since this is the radius of the Airy disk, the resolution is better estimated by the diameter,  

Specific cases edit

 
Log-log plot of aperture diameter vs angular resolution at the diffraction limit for various light wavelengths compared with various astronomical instruments. For example, the blue star shows that the Hubble Space Telescope is almost diffraction-limited in the visible spectrum at 0.1 arcsecs, whereas the red circle shows that the human eye should have a resolving power of 20 arcsecs in theory, though normally only 60 arcsecs.

Single telescope edit

Point-like sources separated by an angle smaller than the angular resolution cannot be resolved. A single optical telescope may have an angular resolution less than one arcsecond, but astronomical seeing and other atmospheric effects make attaining this very hard.

The angular resolution R of a telescope can usually be approximated by

 

where λ is the wavelength of the observed radiation, and D is the diameter of the telescope's objective. The resulting R is in radians. For example, in the case of yellow light with a wavelength of 580 nm, for a resolution of 0.1 arc second, we need D=1.2 m. Sources larger than the angular resolution are called extended sources or diffuse sources, and smaller sources are called point sources.

This formula, for light with a wavelength of about 562 nm, is also called the Dawes' limit.

Telescope array edit

The highest angular resolutions for telescopes can be achieved by arrays of telescopes called astronomical interferometers: These instruments can achieve angular resolutions of 0.001 arcsecond at optical wavelengths, and much higher resolutions at x-ray wavelengths. In order to perform aperture synthesis imaging, a large number of telescopes are required laid out in a 2-dimensional arrangement with a dimensional precision better than a fraction (0.25x) of the required image resolution.

The angular resolution R of an interferometer array can usually be approximated by

 

where λ is the wavelength of the observed radiation, and B is the length of the maximum physical separation of the telescopes in the array, called the baseline. The resulting R is in radians. Sources larger than the angular resolution are called extended sources or diffuse sources, and smaller sources are called point sources.

For example, in order to form an image in yellow light with a wavelength of 580 nm, for a resolution of 1 milli-arcsecond, we need telescopes laid out in an array that is 120 m × 120 m with a dimensional precision better than 145 nm.

Microscope edit

The resolution R (here measured as a distance, not to be confused with the angular resolution of a previous subsection) depends on the angular aperture  :[5]

  where  .

Here NA is the numerical aperture,   is half the included angle   of the lens, which depends on the diameter of the lens and its focal length,   is the refractive index of the medium between the lens and the specimen, and   is the wavelength of light illuminating or emanating from (in the case of fluorescence microscopy) the sample.

It follows that the NAs of both the objective and the condenser should be as high as possible for maximum resolution. In the case that both NAs are the same, the equation may be reduced to:

 

The practical limit for   is about 70°. In a dry objective or condenser, this gives a maximum NA of 0.95. In a high-resolution oil immersion lens, the maximum NA is typically 1.45, when using immersion oil with a refractive index of 1.52. Due to these limitations, the resolution limit of a light microscope using visible light is about 200 nm. Given that the shortest wavelength of visible light is violet ( ),

 

which is near 200 nm.

Oil immersion objectives can have practical difficulties due to their shallow depth of field and extremely short working distance, which calls for the use of very thin (0.17 mm) cover slips, or, in an inverted microscope, thin glass-bottomed Petri dishes.

However, resolution below this theoretical limit can be achieved using super-resolution microscopy. These include optical near-fields (Near-field scanning optical microscope) or a diffraction technique called 4Pi STED microscopy. Objects as small as 30 nm have been resolved with both techniques.[6][7] In addition to this Photoactivated localization microscopy can resolve structures of that size, but is also able to give information in z-direction (3D).

List of telescopes and arrays by angular resolution edit

Name Image Angular resolution (arc seconds) Wavelength Type Site Year
Global mm-VLBI Array (successor to the Coordinated Millimeter VLBI Array) 0.000012 (12 μas) radio (at 1.3 cm) very long baseline interferometry array of different radio telescopes a range of locations on Earth and in space[8] 2002 - 
Very Large Telescope/PIONIER   0.001 (1 mas) light (1-2 micrometre)[9] largest optical array of 4 reflecting telescopes Paranal Observatory, Antofagasta Region, Chile 2002/2010 -
Hubble Space Telescope   0.04 light (near 500 nm)[10] space telescope Earth orbit 1990 -
James Webb Space Telescope   0.1[11] infrared (at 2000 nm)[12] space telescope Sun–Earth L2 2022 -

See also edit

Notes edit

  1. ^ In the case of laser beams, a Gaussian Optics analysis is more appropriate than the Rayleigh criterion, and may reveal a smaller diffraction-limited spot size than that indicated by the formula above.

References edit

  1. ^ Born, M.; Wolf, E. (1999). Principles of Optics. Cambridge University Press. p. 461. ISBN 0-521-64222-1.
  2. ^ a b Lord Rayleigh, F.R.S. (1879). "Investigations in optics, with special reference to the spectroscope". Philosophical Magazine. 5. 8 (49): 261–274. doi:10.1080/14786447908639684.
  3. ^ Michalet, X. (2006). "Using photon statistics to boost microscopy resolution". Proceedings of the National Academy of Sciences. 103 (13): 4797–4798. Bibcode:2006PNAS..103.4797M. doi:10.1073/pnas.0600808103. PMC 1458746. PMID 16549771.
  4. ^ (PDF). Melles Griot Optics Guide. Melles Griot. 2002. Archived from the original (PDF) on 2011-07-08. Retrieved 2011-07-04.
  5. ^ Davidson, M. W. "Resolution". Nikon’s MicroscopyU. Nikon. Retrieved 2017-02-01.
  6. ^ Pohl, D. W.; Denk, W.; Lanz, M. (1984). "Optical stethoscopy: Image recording with resolution λ/20". Applied Physics Letters. 44 (7): 651. Bibcode:1984ApPhL..44..651P. doi:10.1063/1.94865.
  7. ^ Dyba, M. "4Pi-STED-Microscopy..." Max Planck Society, Department of NanoBiophotonics. Retrieved 2017-02-01.
  8. ^ "Images at the Highest Angular Resolution in Astronomy". Max Planck Institute for Radio Astronomy. 2022-05-13. Retrieved 2022-09-26.
  9. ^ de Zeeuw, Tim (2017). "Reaching New Heights in Astronomy - ESO Long Term Perspectives". The Messenger. 166: 2. arXiv:1701.01249.
  10. ^ "Hubble Space Telescope". NASA. 2007-04-09. Retrieved 2022-09-27.
  11. ^ Dalcanton, Julianne; Seager, Sara; Aigrain, Suzanne; Battel, Steve; Brandt, Niel; Conroy, Charlie; Feinberg, Lee; Gezari, Suvi; Guyon, Olivier; Harris, Walt; Hirata, Chris; Mather, John; Postman, Marc; Redding, Dave; Schiminovich, David; Stahl, H. Philip; Tumlinson, Jason (2015). "From Cosmic Birth to Living Earths: The Future of UVOIR Space Astronomy". arXiv:1507.04779 [astro-ph.IM].
  12. ^ "FAQ Full General Public Webb Telescope/NASA". jwst.nasa.gov. 2002-09-10. Retrieved 2022-09-27.

External links edit

  • "Concepts and Formulas in Microscopy: Resolution" by Michael W. Davidson, Nikon MicroscopyU (website).

angular, resolution, this, article, about, optics, imaging, systems, angular, resolution, graph, drawing, angular, resolution, graph, drawing, this, article, needs, additional, citations, verification, please, help, improve, this, article, adding, citations, r. This article is about optics and imaging systems For angular resolution in graph drawing see angular resolution graph drawing This article needs additional citations for verification Please help improve this article by adding citations to reliable sources Unsourced material may be challenged and removed Find sources Angular resolution news newspapers books scholar JSTOR January 2012 Learn how and when to remove this template message Angular resolution describes the ability of any image forming device such as an optical or radio telescope a microscope a camera or an eye to distinguish small details of an object thereby making it a major determinant of image resolution It is used in optics applied to light waves in antenna theory applied to radio waves and in acoustics applied to sound waves The colloquial use of the term resolution sometimes causes confusion when an optical system is said to have a high resolution or high angular resolution it means that the perceived distance or actual angular distance between resolved neighboring objects is small The value that quantifies this property 8 which is given by the Rayleigh criterion is low for a system with a high resolution The closely related term spatial resolution refers to the precision of a measurement with respect to space which is directly connected to angular resolution in imaging instruments The Rayleigh criterion shows that the minimum angular spread that can be resolved by an image forming system is limited by diffraction to the ratio of the wavelength of the waves to the aperture width For this reason high resolution imaging systems such as astronomical telescopes long distance telephoto camera lenses and radio telescopes have large apertures A series of images representing the magnification of M87 with an angular size of some microarcseconds comparable to viewing a tennis ball on the Moon magnification from top left corner counter clockwise to the top right corner Contents 1 Definition of terms 2 The Rayleigh criterion 3 Specific cases 3 1 Single telescope 3 2 Telescope array 3 3 Microscope 4 List of telescopes and arrays by angular resolution 5 See also 6 Notes 7 References 8 External linksDefinition of terms editResolving power is the ability of an imaging device to separate i e to see as distinct points of an object that are located at a small angular distance or it is the power of an optical instrument to separate far away objects that are close together into individual images The term resolution or minimum resolvable distance is the minimum distance between distinguishable objects in an image although the term is loosely used by many users of microscopes and telescopes to describe resolving power As explained below diffraction limited resolution is defined by the Rayleigh criterion as the angular separation of two point sources when the maximum of each source lies in the first minimum of the diffraction pattern Airy disk of the other In scientific analysis in general the term resolution is used to describe the precision with which any instrument measures and records in an image or spectrum any variable in the specimen or sample under study The Rayleigh criterion edit Rayleigh criterion redirects here Not to be confused with Rayleigh roughness criterion nbsp Airy diffraction patterns generated by light from two point sources passing through a circular aperture such as the pupil of the eye Points far apart top or meeting the Rayleigh criterion middle can be distinguished Points closer than the Rayleigh criterion bottom are difficult to distinguish The imaging system s resolution can be limited either by aberration or by diffraction causing blurring of the image These two phenomena have different origins and are unrelated Aberrations can be explained by geometrical optics and can in principle be solved by increasing the optical quality of the system On the other hand diffraction comes from the wave nature of light and is determined by the finite aperture of the optical elements The lens circular aperture is analogous to a two dimensional version of the single slit experiment Light passing through the lens interferes with itself creating a ring shape diffraction pattern known as the Airy pattern if the wavefront of the transmitted light is taken to be spherical or plane over the exit aperture The interplay between diffraction and aberration can be characterised by the point spread function PSF The narrower the aperture of a lens the more likely the PSF is dominated by diffraction In that case the angular resolution of an optical system can be estimated from the diameter of the aperture and the wavelength of the light by the Rayleigh criterion defined by Lord Rayleigh two point sources are regarded as just resolved when the principal diffraction maximum center of the Airy disk of one image coincides with the first minimum of the Airy disk of the other 1 2 as shown in the accompanying photos In the bottom photo on the right that shows the Rayleigh criterion limit the central maximum of one point source might look as though it lies outside the first minimum of the other but examination with a ruler verifies that the two do intersect If the distance is greater the two points are well resolved and if it is smaller they are regarded as not resolved Rayleigh defended this criterion on sources of equal strength 2 Considering diffraction through a circular aperture this translates into 8 1 22 l D considering that sin 8 8 displaystyle theta approx 1 22 frac lambda D quad text considering that sin theta approx theta nbsp where 8 is the angular resolution radians l is the wavelength of light and D is the diameter of the lens aperture The factor 1 22 is derived from a calculation of the position of the first dark circular ring surrounding the central Airy disc of the diffraction pattern This number is more precisely 1 21966989 OEIS A245461 the first zero of the order one Bessel function of the first kind J 1 x displaystyle J 1 x nbsp divided by p The formal Rayleigh criterion is close to the empirical resolution limit found earlier by the English astronomer W R Dawes who tested human observers on close binary stars of equal brightness The result 8 4 56 D with D in inches and 8 in arcseconds is slightly narrower than calculated with the Rayleigh criterion A calculation using Airy discs as point spread function shows that at Dawes limit there is a 5 dip between the two maxima whereas at Rayleigh s criterion there is a 26 3 dip 3 Modern image processing techniques including deconvolution of the point spread function allow resolution of binaries with even less angular separation Using a small angle approximation the angular resolution may be converted into a spatial resolution Dℓ by multiplication of the angle in radians with the distance to the object For a microscope that distance is close to the focal length f of the objective For this case the Rayleigh criterion reads D ℓ 1 22 f l D displaystyle Delta ell approx 1 22 frac f lambda D nbsp This is the radius in the imaging plane of the smallest spot to which a collimated beam of light can be focused which also corresponds to the size of smallest object that the lens can resolve 4 The size is proportional to wavelength l and thus for example blue light can be focused to a smaller spot than red light If the lens is focusing a beam of light with a finite extent e g a laser beam the value of D corresponds to the diameter of the light beam not the lens Note 1 Since the spatial resolution is inversely proportional to D this leads to the slightly surprising result that a wide beam of light may be focused to a smaller spot than a narrow one This result is related to the Fourier properties of a lens A similar result holds for a small sensor imaging a subject at infinity The angular resolution can be converted to a spatial resolution on the sensor by using f as the distance to the image sensor this relates the spatial resolution of the image to the f number f D ℓ 1 22 f l D 1 22 l f displaystyle Delta ell approx 1 22 frac f lambda D 1 22 lambda cdot f nbsp Since this is the radius of the Airy disk the resolution is better estimated by the diameter 2 44 l f displaystyle 2 44 lambda cdot f nbsp Specific cases edit nbsp Log log plot of aperture diameter vs angular resolution at the diffraction limit for various light wavelengths compared with various astronomical instruments For example the blue star shows that the Hubble Space Telescope is almost diffraction limited in the visible spectrum at 0 1 arcsecs whereas the red circle shows that the human eye should have a resolving power of 20 arcsecs in theory though normally only 60 arcsecs Single telescope edit Point like sources separated by an angle smaller than the angular resolution cannot be resolved A single optical telescope may have an angular resolution less than one arcsecond but astronomical seeing and other atmospheric effects make attaining this very hard The angular resolution R of a telescope can usually be approximated by R l D displaystyle R frac lambda D nbsp where l is the wavelength of the observed radiation and D is the diameter of the telescope s objective The resulting R is in radians For example in the case of yellow light with a wavelength of 580 nm for a resolution of 0 1 arc second we need D 1 2 m Sources larger than the angular resolution are called extended sources or diffuse sources and smaller sources are called point sources This formula for light with a wavelength of about 562 nm is also called the Dawes limit Telescope array edit The highest angular resolutions for telescopes can be achieved by arrays of telescopes called astronomical interferometers These instruments can achieve angular resolutions of 0 001 arcsecond at optical wavelengths and much higher resolutions at x ray wavelengths In order to perform aperture synthesis imaging a large number of telescopes are required laid out in a 2 dimensional arrangement with a dimensional precision better than a fraction 0 25x of the required image resolution The angular resolution R of an interferometer array can usually be approximated by R l B displaystyle R frac lambda B nbsp where l is the wavelength of the observed radiation and B is the length of the maximum physical separation of the telescopes in the array called the baseline The resulting R is in radians Sources larger than the angular resolution are called extended sources or diffuse sources and smaller sources are called point sources For example in order to form an image in yellow light with a wavelength of 580 nm for a resolution of 1 milli arcsecond we need telescopes laid out in an array that is 120 m 120 m with a dimensional precision better than 145 nm Microscope edit The resolution R here measured as a distance not to be confused with the angular resolution of a previous subsection depends on the angular aperture a displaystyle alpha nbsp 5 R 1 22 l N A condenser N A objective displaystyle R frac 1 22 lambda mathrm NA text condenser mathrm NA text objective nbsp where N A n sin 8 displaystyle mathrm NA n sin theta nbsp Here NA is the numerical aperture 8 displaystyle theta nbsp is half the included angle a displaystyle alpha nbsp of the lens which depends on the diameter of the lens and its focal length n displaystyle n nbsp is the refractive index of the medium between the lens and the specimen and l displaystyle lambda nbsp is the wavelength of light illuminating or emanating from in the case of fluorescence microscopy the sample It follows that the NAs of both the objective and the condenser should be as high as possible for maximum resolution In the case that both NAs are the same the equation may be reduced to R 0 61 l N A l 2 N A displaystyle R frac 0 61 lambda mathrm NA approx frac lambda 2 mathrm NA nbsp The practical limit for 8 displaystyle theta nbsp is about 70 In a dry objective or condenser this gives a maximum NA of 0 95 In a high resolution oil immersion lens the maximum NA is typically 1 45 when using immersion oil with a refractive index of 1 52 Due to these limitations the resolution limit of a light microscope using visible light is about 200 nm Given that the shortest wavelength of visible light is violet l 400 n m displaystyle lambda approx 400 mathrm nm nbsp R 1 22 400 nm 1 45 0 95 203 nm displaystyle R frac 1 22 times 400 mbox nm 1 45 0 95 203 mbox nm nbsp which is near 200 nm Oil immersion objectives can have practical difficulties due to their shallow depth of field and extremely short working distance which calls for the use of very thin 0 17 mm cover slips or in an inverted microscope thin glass bottomed Petri dishes However resolution below this theoretical limit can be achieved using super resolution microscopy These include optical near fields Near field scanning optical microscope or a diffraction technique called 4Pi STED microscopy Objects as small as 30 nm have been resolved with both techniques 6 7 In addition to this Photoactivated localization microscopy can resolve structures of that size but is also able to give information in z direction 3D List of telescopes and arrays by angular resolution editName Image Angular resolution arc seconds Wavelength Type Site YearGlobal mm VLBI Array successor to the Coordinated Millimeter VLBI Array 0 000012 12 mas radio at 1 3 cm very long baseline interferometry array of different radio telescopes a range of locations on Earth and in space 8 2002 Very Large Telescope PIONIER nbsp 0 001 1 mas light 1 2 micrometre 9 largest optical array of 4 reflecting telescopes Paranal Observatory Antofagasta Region Chile 2002 2010 Hubble Space Telescope nbsp 0 04 light near 500 nm 10 space telescope Earth orbit 1990 James Webb Space Telescope nbsp 0 1 11 infrared at 2000 nm 12 space telescope Sun Earth L2 2022 See also editAngular diameter Beam diameter Dawes limit Diffraction limited system Ground sample distance Image resolution Optical resolution Sparrow s resolution limit Visual acuityNotes edit In the case of laser beams a Gaussian Optics analysis is more appropriate than the Rayleigh criterion and may reveal a smaller diffraction limited spot size than that indicated by the formula above References edit Born M Wolf E 1999 Principles of Optics Cambridge University Press p 461 ISBN 0 521 64222 1 a b Lord Rayleigh F R S 1879 Investigations in optics with special reference to the spectroscope Philosophical Magazine 5 8 49 261 274 doi 10 1080 14786447908639684 Michalet X 2006 Using photon statistics to boost microscopy resolution Proceedings of the National Academy of Sciences 103 13 4797 4798 Bibcode 2006PNAS 103 4797M doi 10 1073 pnas 0600808103 PMC 1458746 PMID 16549771 Diffraction Fraunhofer Diffraction at a Circular Aperture PDF Melles Griot Optics Guide Melles Griot 2002 Archived from the original PDF on 2011 07 08 Retrieved 2011 07 04 Davidson M W Resolution Nikon s MicroscopyU Nikon Retrieved 2017 02 01 Pohl D W Denk W Lanz M 1984 Optical stethoscopy Image recording with resolution l 20 Applied Physics Letters 44 7 651 Bibcode 1984ApPhL 44 651P doi 10 1063 1 94865 Dyba M 4Pi STED Microscopy Max Planck Society Department of NanoBiophotonics Retrieved 2017 02 01 Images at the Highest Angular Resolution in Astronomy Max Planck Institute for Radio Astronomy 2022 05 13 Retrieved 2022 09 26 de Zeeuw Tim 2017 Reaching New Heights in Astronomy ESO Long Term Perspectives The Messenger 166 2 arXiv 1701 01249 Hubble Space Telescope NASA 2007 04 09 Retrieved 2022 09 27 Dalcanton Julianne Seager Sara Aigrain Suzanne Battel Steve Brandt Niel Conroy Charlie Feinberg Lee Gezari Suvi Guyon Olivier Harris Walt Hirata Chris Mather John Postman Marc Redding Dave Schiminovich David Stahl H Philip Tumlinson Jason 2015 From Cosmic Birth to Living Earths The Future of UVOIR Space Astronomy arXiv 1507 04779 astro ph IM FAQ Full General Public Webb Telescope NASA jwst nasa gov 2002 09 10 Retrieved 2022 09 27 External links edit Concepts and Formulas in Microscopy Resolution by Michael W Davidson Nikon MicroscopyU website Portals nbsp Science nbsp Astronomy nbsp Stars nbsp Spaceflight nbsp Outer space nbsp Solar System Retrieved from https en wikipedia org w index php title Angular resolution amp oldid 1172331914, wikipedia, wiki, book, books, library,