Liquid chromatography–mass spectrometry (LC–MS) is an analytical chemistry technique that combines the physical separation capabilities of liquid chromatography (or HPLC) with the mass analysis capabilities of mass spectrometry (MS). Coupled chromatography - MS systems are popular in chemical analysis because the individual capabilities of each technique are enhanced synergistically. While liquid chromatography separates mixtures with multiple components, mass spectrometry provides spectral information that may help to identify (or confirm the suspected identity of) each separated component.[1] MS is not only sensitive, but provides selective detection, relieving the need for complete chromatographic separation.[2] LC-MS is also appropriate for metabolomics because of its good coverage of a wide range of chemicals.[3] This tandem technique can be used to analyze biochemical, organic, and inorganic compounds commonly found in complex samples of environmental and biological origin. Therefore, LC-MS may be applied in a wide range of sectors including biotechnology, environment monitoring, food processing, and pharmaceutical, agrochemical, and cosmetic industries.[4][5] Since the early 2000s, LC-MS (or more specifically LC-MS-MS) has also begun to be used in clinical applications.[6]

Liquid chromatography–mass spectrometry
Ion trap LCMS system with ESI interface
AcronymLCMS
ClassificationChromatography
Mass spectrometry
Analytesorganic molecules
biomolecules
ManufacturersAgilent
Bruker
PerkinElmer
SCIEX
Shimadzu Scientific
Thermo Fisher Scientific
Waters Corporation
Other techniques
RelatedGas chromatography–mass spectrometry

In addition to the liquid chromatography and mass spectrometry devices, an LC-MS system contains an interface that efficiently transfers the separated components from the LC column into the MS ion source.[5][7] The interface is necessary because the LC and MS devices are fundamentally incompatible. While the mobile phase in a LC system is a pressurized liquid, the MS analyzers commonly operate under high vacuum. Thus, it is not possible to directly pump the eluate from the LC column into the MS source. Overall, the interface is a mechanically simple part of the LC-MS system that transfers the maximum amount of analyte, removes a significant portion of the mobile phase used in LC and preserves the chemical identity of the chromatography products (chemically inert). As a requirement, the interface should not interfere with the ionizing efficiency and vacuum conditions of the MS system.[5] Nowadays, most extensively applied LC-MS interfaces are based on atmospheric pressure ionization (API) strategies like electrospray ionization (ESI), atmospheric-pressure chemical ionization (APCI), and atmospheric pressure photoionization (APPI). These interfaces became available in the 1990s after a two decade long research and development process.[8][7]

History of LC-MS

The coupling of chromatography with MS is a well developed chemical analysis strategy dating back from the 1950s. Gas chromatography (GC)MS was originally introduced in 1952, when A. T. James and A. J. P. Martin were trying to develop tandem separation - mass analysis techniques.[9] In GC, the analytes are eluted from the separation column as a gas and the connection with electron ionization (EI) or chemical ionization (CI) ion sources in the MS system was a technically simpler challenge. Because of this, the development of GC-MS systems was faster than LC-MS and such systems were first commercialized in the 1970s.[7] The development of LC-MS systems took longer than GC-MS and was directly related to the development of proper interfaces. V. L. Tal'roze and collaborators started the development of LC-MS in the late 1960s,[10][11] when they first used capillaries to connect an LC columns to an EI source.[12][8] A similar strategy was investigated by McLafferty and collaborators in 1973 who coupled the LC column to a CI source,[13] which allowed a higher liquid flow into the source. This was the first and most obvious way of coupling LC with MS, and was known as the capillary inlet interface. This pioneer interface for LC-MS had the same analysis capabilities of GC-MS and was limited to rather volatile analytes and non-polar compounds with low molecular mass (below 400 Da). In the capillary inlet interface, the evaporation of the mobile phase inside the capillary was one of the main issues. Within the first years of development of LC-MS, on-line and off-line alternatives were proposed as coupling alternatives. In general, off-line coupling involved fraction collection, evaporation of solvent, and transfer of analytes to the MS using probes. Off-line analyte treatment process was time consuming and there was an inherent risk of sample contamination. Rapidly, it was realized that the analysis of complex mixtures would require the development of a fully automated on-line coupling solution in LC-MS.[8]

The key to the success and wide-spread adoption of LC-MS as a routine analytical tool lies in the interface and ion source between the liquid-based LC and the vacuum-base MS. The following interfaces were stepping-stones on the way to the modern atmospheric-pressure ionization interfaces, and are described for historical interest.

Moving Belt Interface

The moving-belt interface (MBI) was developed by McFadden et al in 1977 and commercialized by Finnigan.[14] This interface consisted of an endless moving belt onto which the LC column effluent was deposited in a band. On the belt, the solvent was evaporated by gently heating and efficiently exhausting the solvent vapours under reduced pressure in two vacuum chambers. After the liquid phase was removed, the belt passed over a heater which flash desorbed the analytes into the MS ion source. One of the significant advantages of the MBI was its compatibility with a wide range of chromatographic conditions.[8] MBI was successfully used for LC-MS applications between 1978 and 1990 because it allowed coupling of LC to MS devices using EI, CI, and fast-atom bombardment (FAB) ion sources. The most common MS systems connected by MBI interfaces to LC columns were magnetic sector and quadrupole instruments. MBI interfaces for LC-MS allowed MS to be widely applied in the analysis of drugs, pesticides, steroids, alkaloids, and polycyclic aromatic hydrocarbons. This interface is no longer used because of its mechanical complexity and the difficulties associated with belt renewal as well as its inability to handle very labile biomolecules.

Direct liquid introduction interface

The direct liquid introduction (DLI) interface was developed in 1980. This interface was intended to solve the problem of evaporation of liquid inside the capillary inlet interface. In DLI, a small portion of the LC flow was forced through a small aperture or diaphragm (typically 10um in diameter) to form a liquid jet composed of small droplets that were subsequently dried in a desolvation chamber.[11] The analytes were ionized using a solvent assisted chemical ionization source, where the LC solvents acted as reagent gases. To use this interface, it was necessary to split the flow coming out of the LC column because only a small portion of the effluent (10 to 50 μl/min out of 1 ml/min) could be introduced into the source without raising the vacuum pressure of the MS system too high. Alternately, Henion at Cornell University had success with using micro-bore LC methods so that the entire (low) flow of the LC could be used. One of the main operational problems of the DLI interface was the frequent clogging of the diaphragm orifices. The DLI interface was used between 1982 and 1985 for the analysis of pesticides, corticosteroids, metabolites in horse urine, erythromycin, and vitamin B12. However, this interface was replaced by the thermospray interface, which removed the flow rate limitations and the issues with the clogging diaphragms.[5][8]

A related device was the particle beam interface (PBI), developed by Willoughby and Browner in 1984.[15] Particle beam interfaces took over the wide applications of MBI for LC-MS in 1988.[8][16] The PBI operated by using a helium gas nebulizer to spray the eluant into the vacuum, drying the droplets and pumping away the solvent vapour (using a jet separator) while the stream of monodisperse dried particles containing the analyte entered the source.[11] Drying the droplets outside of the source volume, and using a jet separator to pump away the solvent vapour, allowed the particles to enter and be vapourized in a low-pressure EI source. As with the MBI, the ability to generate library-searchable EI spectra was a distinct advantage for many applications. Commercialized by Hewlett Packard, and later by VG and Extrel, it enjoyed moderate success, but has been largely supplanted by the atmospheric pressure interfaces such as electrospray and APCI which provide a broader range of compound coverage and applications.

Thermospray interface

The thermospray (TSP) interface was developed in 1980 by Marvin Vestal and co-workers at the University of Houston.[17] It was commercialized by Vestec and several of the major mass spectrometer manufacurers. The interface resulted from a long term research project intended to find a LC-MS interface capable of handling high flow rates (1 ml/min) and avoiding the flow split in DLI interfaces. The TSP interface was composed of a heated probe, a desolvation chamber, and an ion focusing skimmer. The LC effluent passed through the heated probe and emerged as a jet of vapor and small droplets flowing into the desolvation chamber at low pressure. Initially operated with a filament or discharge as the source of ions (thereby acting as a CI source for vapourized analyte), it was soon discovered that ions were also observed when the filament or discharge was off. This could be attributed to either direct emission of ions from the liquid droplets as they evaporated in a process related to electrospray ionization or ion evaporation, or to chemical ionization of vapourized analyte molecules from buffer ions (such as ammonium acetate). The fact that multiply-charged ions were observed from some larger analytes suggests that direct analyte ion emission was occurring under at least some conditions.[11] The interface was able to handle up to 2 ml/min of eluate from the LC column and would efficiently introduce it into the MS vacuum system. TSP was also more suitable for LC-MS applications involving reversed phase liquid chromatography (RT-LC). With time, the mechanical complexity of TSP was simplified, and this interface became popular as the first ideal LC-MS interface for pharmaceutical applications comprising the analysis of drugs, metabolites, conjugates, nucleosides, peptides, natural products, and pesticides. The introduction of TSP marked a significant improvement for LC-MS systems and was the most widely applied interface until the beginning of the 1990s, when it began to be replaced by interfaces involving atmospheric pressure ionization (API).[5][7][16]

FAB based interfaces

The frit fast atom bombardment (FAB) and continuous flow-FAB (CF-FAB) interfaces were developed in 1985 and 1986 respectively.[16] Both interfaces were similar, but they differed in that the first used a porous frit probe as connecting channel, while CF-FAB used a probe tip. From these, the CF-FAB was more successful as a LC-MS interface and was useful to analyze non-volatile and thermally labile compounds. In these interfaces, the LC effluent passed through the frit or CF-FAB channels to form a uniform liquid film at the tip. There, the liquid was bombarded with ion beams or high energy atoms (fast atoms). For stable operation, the FAB based interfaces were able to handle liquid flow rates of only 1–15 μl and were also restricted to microbore and capillary columns. In order to be used in FAB MS ionization sources, the analytes of interest had to be mixed with a matrix (e.g., glycerol) that could be added before or after the separation in the LC column. FAB based interfaces were extensively used to characterize peptides, but lost applicability with the advent of electrospray based interfaces in 1988.[5][8]

Liquid chromatography

 
Diagram of an LC-MS system

Liquid chromatography is a method of physical separation in which the components of a liquid mixture are distributed between two immiscible phases, i.e., stationary and mobile. The practice of LC can be divided into five categories, i.e., adsorption chromatography, partition chromatography, ion-exchange chromatography, size-exclusion chromatography, and affinity chromatography. Among these, the most widely used variant is the reverse-phase (RP) mode of the partition chromatography technique, which makes use of a nonpolar (hydrophobic) stationary phase and a polar mobile phase. In common applications, the mobile phase is a mixture of water and other polar solvents (e.g., methanol, isopropanol, and acetonitrile), and the stationary matrix is prepared by attaching long-chain alkyl groups (e.g., n-octadecyl or C18) to the external and internal surfaces of irregularly or spherically shaped 5 μm diameter porous silica particles.[5]

In HPLC, typically 20 μl of the sample of interest are injected into the mobile phase stream delivered by a high pressure pump. The mobile phase containing the analytes permeates through the stationary phase bed in a definite direction. The components of the mixture are separated depending on their chemical affinity with the mobile and stationary phases. The separation occurs after repeated sorption and desorption steps occurring when the liquid interacts with the stationary bed.[8] The liquid solvent (mobile phase) is delivered under high pressure (up to 400 bar or 5800 psi) into a packed column containing the stationary phase. The high pressure is necessary to achieve a constant flow rate for reproducible chromatography experiments. Depending on the partitioning between the mobile and stationary phases, the components of the sample will flow out of the column at different times.[16] The column is the most important component of the LC system and is designed to withstand the high pressure of the liquid. Conventional LC columns are 100–300 mm long with outer diameter of 6.4 mm (1/4 inch) and internal diameter of 3.04.6 mm. For applications involving LC-MS, the length of chromatography columns can be shorter (30–50 mm) with 3–5 μm diameter packing particles. In addition to the conventional model, other LC columns are the narrow bore, microbore, microcapillary, and nano-LC models. These columns have smaller internal diameters, allow for a more efficient separation, and handle liquid flows under 1 ml/min (the conventional flow-rate).[8] In order to improve separation efficiency and peak resolution, ultra performance liquid chromatography (UHPLC) can be used instead of HPLC. This LC variant uses columns packed with smaller silica particles (~1.7 μm diameter) and requires higher operating pressures in the range of 310000 to 775000 torr (6000 to 15000 psi, 400 to 1034 bar).[5]

Mass spectrometry

 
LC-MS spectrum of each resolved peak

Mass spectrometry (MS) is an analytical technique that measures the mass-to-charge ratio (m/z) of charged particles (ions). Although there are many different kinds of mass spectrometers, all of them make use of electric or magnetic fields to manipulate the motion of ions produced from an analyte of interest and determine their m/z.[18] The basic components of a mass spectrometer are the ion source, the mass analyzer, the detector, and the data and vacuum systems. The ion source is where the components of a sample introduced in a MS system are ionized by means of electron beams, photon beams (UV lights), laser beams or corona discharge. In the case of electrospray ionization, the ion source moves ions that exist in liquid solution into the gas phase. The ion source converts and fragments the neutral sample molecules into gas-phase ions that are sent to the mass analyzer. While the mass analyzer applies the electric and magnetic fields to sort the ions by their masses, the detector measures and amplifies the ion current to calculate the abundances of each mass-resolved ion. In order to generate a mass spectrum that a human eye can easily recognize, the data system records, processes, stores, and displays data in a computer.[5]

The mass spectrum can be used to determine the mass of the analytes, their elemental and isotopic composition, or to elucidate the chemical structure of the sample.[5] MS is an experiment that must take place in gas phase and under vacuum (1.33 * 10−2 to 1.33 * 10−6 pascal). Therefore, the development of devices facilitating the transition from samples at higher pressure and in condensed phase (solid or liquid) into a vacuum system has been essential to develop MS as a potent tool for identification and quantification of organic compounds like peptides.[19] MS is now in very common use in analytical laboratories that study physical, chemical, or biological properties of a great variety of compounds. Among the many different kinds of mass analyzers, the ones that find application in LC-MS systems are the quadrupole, time-of-flight (TOF), ion traps, and hybrid quadrupole-TOF (QTOF) analyzers.[7]

Interfaces

The interface between a liquid phase technique (HPLC) with a continuously flowing eluate, and a gas phase technique carried out in a vacuum was difficult for a long time. The advent of electrospray ionization changed this. Currently, the most common LC-MS interfaces are electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and atmospheric pressure photo-ionization (APPI). These are newer MS ion sources that facilitate the transition from a high pressure environment (HPLC) to high vacuum conditions needed at the MS analyzer.[20][7] Although these interfaces are described individually, they can also be commercially available as dual ESI/APCI, ESI/APPI, or APCI/APPI ion sources.[8] Various deposition and drying techniques were used in the past (e.g., moving belts) but the most common of these was the off-line MALDI deposition.[21][22] A new approach still under development called direct-EI LC-MS interface, couples a nano HPLC system and an electron ionization equipped mass spectrometer.[23][24]

Electrospray ionization (ESI)

ESI interface for LC-MS systems was developed by Fenn and collaborators in 1988.[25] This ion source/ interface can be used for the analysis of moderately polar and even very polar molecules (e.g., metabolites, xenobiotics, peptides, nucleotides, polysaccharides). The liquid eluate coming out of the LC column is directed into a metal capillary kept at 3 to 5 kV and is nebulized by a high-velocity coaxial flow of gas at the tip of the capillary, creating a fine spray of charged droplets in front of the entrance to the vacuum chamber. To avoid contamination of the vacuum system by buffers and salts, this capillary is usually perpendicularly located at the inlet of the MS system, in some cases with a counter-current of dry nitrogen in front of the entrance through which ions are directed by the electric field. In some sources, rapid droplet evaporation and thus maximum ion emission is achieved by mixing an additional stream of hot gas with the spray plume in front of the vacuum entrance. In other sources, the droplets are drawn through a heated capillary tube as they enter the vacuum, promoting droplet evaporation and ion emission. These methods of increasing droplet evaporation now allow the use of liquid flow rates of 1 - 2 mL/min to be used while still achieving efficient ionisation[26] and high sensitivity. Thus while the use of 1 - 3 mm microbore columns and lower flow rates of 50 - 200 μl/min was commonly considered necessary for optimum operation, this limitation is no longer as important, and the higher column capacity of larger bore columns can now be advantageously employed with ESI LC-MS systems. Positively and negatively charged ions can be created by switching polarities, and it is possible to acquire alternate positive and negative mode spectra rapidly within the same LC run . While most large molecules (greater than MW 1500-2000) produce multiply charged ions in the ESI source, the majority of smaller molecules produce singly charged ions.[7]

Atmospheric pressure chemical ionization (APCI)

The development of the APCI interface for LC-MS started with Horning and collaborators in the early 1973.[27] However, its commercial application was introduced at the beginning of the 1990s after Henion and collaborators improved the LC-APCI-MS interface in 1986.[8] The APCI ion source/ interface can be used to analyze small, neutral, relatively non-polar, and thermally stable molecules (e.g., steroids, lipids, and fat soluble vitamins). These compounds are not well ionized using ESI. In addition, APCI can also handle mobile phase streams containing buffering agents. The liquid from the LC system is pumped through a capillary and there is also nebulization at the tip, where a corona discharge takes place. First, the ionizing gas surrounding the interface and the mobile phase solvent are subject to chemical ionization at the ion source. Later, these ions react with the analyte and transfer their charge. The sample ions then pass through small orifice skimmers by means of or ion-focusing lenses. Once inside the high vacuum region, the ions are subject to mass analysis. This interface can be operated in positive and negative charge modes and singly-charged ions are mainly produced.[7] APCI ion source can also handle flow rates between 500 and 2000 μl/min and it can be directly connected to conventional 4.6 mm ID columns.[16]

Atmospheric pressure photoionization (APPI)

The APPI interface for LC-MS was developed simultaneously by Bruins and Syage in 2000.[28][8] APPI is another LC-MS ion source/ interface for the analysis of neutral compounds that cannot be ionized using ESI.[7] This interface is similar to the APCI ion source, but instead of a corona discharge, the ionization occurs by using photons coming from a discharge lamp. In the direct-APPI mode, singly charged analyte molecular ions are formed by absorption of a photon and ejection of an electron. In the dopant-APPI mode, an easily ionizable compound (Dopant) is added to the mobile phase or the nebulizing gas to promote a reaction of charge-exchange between the dopant molecular ion and the analyte. The ionized sample is later transferred to the mass analyzer at high vacuum as it passes through small orifice skimmers.[8]

Applications

The coupling of MS with LC systems is attractive because liquid chromatography can separate delicate and complex natural mixtures, which chemical composition needs to be well established (e.g., biological fluids, environmental samples, and drugs). Further, LC-MS has applications in volatile explosive residue analysis.[29] Nowadays, LC-MS has become one of the most widely used chemical analysis techniques because more than 85% of natural chemical compounds are polar and thermally labile and GC-MS cannot process these samples.[5] As an example, HPLC-MS is regarded as the leading analytical technique for proteomics and pharmaceutical laboratories.[7][5] Other important applications of LC-MS include the analysis of food, pesticides, and plant phenols.[8]

Pharmacokinetics

LC-MS is widely used in the field of bioanalysis and is specially involved in pharmacokinetic studies of pharmaceuticals. Pharmacokinetic studies are needed to determine how quickly a drug will be cleared from the body organs and the hepatic blood flow. MS analyzers are useful in these studies because of their shorter analysis time, and higher sensitivity and specificity compared to UV detectors commonly attached to HPLC systems. One major advantage is the use of tandem MS-MS, where the detector may be programmed to select certain ions to fragment. The measured quantity is the sum of molecule fragments chosen by the operator. As long as there are no interferences or ion suppression in LC-MS, the LC separation can be quite quick.[30]

Proteomics/metabolomics

LC-MS is used in proteomics as a method to detect and identify the components of a complex mixture. The bottom-up proteomics LC-MS approach generally involves protease digestion and denaturation using trypsin as a protease, urea to denature the tertiary structure, and iodoacetamide to modify the cysteine residues. After digestion, LC-MS is used for peptide mass fingerprinting, or LC-MS/MS (tandem MS) is used to derive the sequences of individual peptides.[31] LC-MS/MS is most commonly used for proteomic analysis of complex samples where peptide masses may overlap even with a high-resolution mass spectrometry. Samples of complex biological (e.g., human serum) may be analyzed in modern LC-MS/MS systems, which can identify over 1000 proteins. However, this high level of protein identification is possible only after separating the sample by means of SDS-PAGE gel or HPLC-SCX.[30] Recently, LC-MS/MS has been applied to search peptide biomarkers. Examples are the recent discovery and validation of peptide biomarkers for four major bacterial respiratory tract pathogens (Staphylococcus aureus, Moraxella catarrhalis; Haemophilus influenzae and Streptococcus pneumoniae) and the SARS-CoV-2 virus.[32] [33]

LC-MS has emerged as one of the most commonly used techniques in global metabolite profiling of biological tissue (e.g., blood plasma, serum, urine).[34] LC-MS is also used for the analysis of natural products and the profiling of secondary metabolites in plants.[35] In this regard, MS-based systems are useful to acquire more detailed information about the wide spectrum of compounds from a complex biological samples. LC-Nuclear magnetic resonance (NMR) is also used in plant metabolomics, but this technique can only detect and quantify the most abundant metabolites. LC-MS has been useful to advance the field of plant metabolomics, which aims to study the plant system at molecular level providing a non-biased characterization of the plant metabolome in response to its environment.[36] The first application of LC-MS in plant metabolomics was the detection of a wide range of highly polar metabolites, oligosaccharides, amino acids, amino sugars, and sugar nucleotides from Cucurbita maxima phloem tissues.[37] Another example of LC-MS in plant metabolomics is the efficient separation and identification of glucose, sucrose, raffinose, stachyose, and verbascose from leaf extracts of Arabidopsis thaliana.[38]

Drug development

LC-MS is frequently used in drug development because it allows quick molecular weight confirmation and structure identification. These features speed up the process of generating, testing, and validating a discovery starting from a vast array of products with potential application. LC-MS applications for drug development are highly automated methods used for peptide mapping, glycoprotein mapping, lipodomics, natural products dereplication, bioaffinity screening, in vivo drug screening, metabolic stability screening, metabolite identification, impurity identification, quantitative bioanalysis, and quality control.[39]

See also

References

  1. ^ de Hoffmann, Edmond; Stroobant, Vincent (2002). Mass Spectrometry (Principles and Applications) (2nd ed.). Wiley. pp. 157–158. ISBN 0-471-48566-7.
  2. ^ Pitt, James J (February 2009). "Principles and Applications of Liquid Chromatography-Mass Spectrometry in Clinical Biochemistry". Clin Biochem Rev. 30 (1): 19–34. PMC 2643089. PMID 19224008.
  3. ^ Zhou, Bin; Xiao, Jun Feng; Tuli, Leepika; Ressom, Habtom W (Feb 2012). "LC-MS-based metabolomics". Mol. Biosyst. 8 (2): 470–481. doi:10.1039/c1mb05350g. PMC 3699692. PMID 22041788.
  4. ^ Chaimbault, Patrick (2014-01-01). "The Modern Art of Identification of Natural Substances in Whole Plants". In Jacob, Claus; Kirsch, Gilbert; Slusarenko, Alan; Winyard, Paul G.; Burkholz, Torsten (eds.). Recent Advances in Redox Active Plant and Microbial Products. Springer Netherlands. pp. 31–94. doi:10.1007/978-94-017-8953-0_3. ISBN 9789401789523.
  5. ^ a b c d e f g h i j k l Dass, Chhabil (2007-01-01). "Hyphenated Separation Techniques". Fundamentals of Contemporary Mass Spectrometry. John Wiley & Sons, Inc. pp. 151–194. doi:10.1002/9780470118498.ch5. ISBN 9780470118498.
  6. ^ Seger, Christoph; Salzmann, Linda (2020-08-01). "After another decade: LC–MS/MS became routine in clinical diagnostics". Clinical Biochemistry. Advancement and Applications of Mass Spectrometry in Laboratory Medicine. 82: 2–11. doi:10.1016/j.clinbiochem.2020.03.004. ISSN 0009-9120. PMID 32188572. S2CID 213186669.
  7. ^ a b c d e f g h i j Pitt, James J (2017-03-12). "Principles and Applications of Liquid Chromatography-Mass Spectrometry in Clinical Biochemistry". The Clinical Biochemist Reviews. 30 (1): 19–34. ISSN 0159-8090. PMC 2643089. PMID 19224008.
  8. ^ a b c d e f g h i j k l m n Niessen, Wilfried M. A (2006). Liquid Chromatography-Mass Spectrometry, Third Edition. Boca Raton: CRC Taylor & Francis. pp. 50–90. ISBN 9780824740825. OCLC 232370223.
  9. ^ James, A. T.; Martin, A. J. P. (1952-03-01). "Gas-liquid partition chromatography: the separation and micro-estimation of volatile fatty acids from formic acid to dodecanoic acid". Biochemical Journal. 50 (5): 679–690. doi:10.1042/bj0500679. ISSN 0264-6021. PMC 1197726. PMID 14934673.
  10. ^ Tal'roze, V. L; Karpov, G. V.; Gordetskii, I. G.; Skurat, V. E. (1968). "Capillary Systems for the Introduction of Liquid Mixtures into an Analytical Mass Spectrometer". Russ. J. Phys. Chem. 42: 1658–1664.
  11. ^ a b c d Arpino, Patrick (2006). "History of LC-MS Development and Interfacing". Encyclopedia of Mass Spectrometry. Vol. 8. Elsevier. ISBN 978-0080438474.
  12. ^ Tal'roze, V.L.; Gorodetskii, I.G.; Zolotoy, N.B; Karpov, G.V.; Skurat, V.E.; Maslennikova, V.Ya. (1978). "Capillary system for continuous introducing of volatile liquids into analytical MS and its application". Adv. Mass Spectrom. 7: 858.
  13. ^ Baldwin, M. A.; McLafferty, F. W. (1973). "Liquid chromatography-mass spectrometry interface-I: The direct introduction of liquid solutions into a chemical ionization mass spectrometer". Organic Mass Spectrometry. 7 (9): 1111–1112. doi:10.1002/oms.1210070913. ISSN 0030-493X.
  14. ^ Pullen, Franl (2010). "The fascinating history of the development of LC-MS; a personal perspective". Chromatography Today (February/March): 4–6.
  15. ^ de Koster, Chris G.; Schoenmakers, Peter J. (2020). "Chapter 3.1 - History of Liquid Chromatography-Mass Spectrometry". In Tranchida, Peter Q.; Mondello, Luigi (eds.). Hyphenations of Capillary Chromatography With Mass Spectrometry. Elsevier. pp. 279–295. ISBN 978-0-12-809638-3.
  16. ^ a b c d e Ardrey, Robert E. (2003-01-01). "Introduction". Liquid Chromatography – Mass Spectrometry: An Introduction. Analytical Techniques in the Sciences (AnTS). John Wiley & Sons, Ltd. pp. 1–5. doi:10.1002/0470867299.ch1. ISBN 9780470867297.
  17. ^ Blakley, C. R.; Carmody, J. J.; Vestal, M. L. (1980). "A New Soft Ionization Technique for Mass Spectrometry of Complex Samples". J. Am. Chem. Soc. 102: 5931–5933. doi:10.1021/ja00538a050.
  18. ^ Roberts, Gordon (2013). Roberts, Gordon C. K (ed.). Encyclopedia of Biophysics - Springer. doi:10.1007/978-3-642-16712-6. ISBN 978-3-642-16711-9. S2CID 44856071.
  19. ^ Sharp, Thomas R. (2009-01-01). "Mass Spectrometry". In Nassar, Ala F.; Collegiateessor, Paul F. Hollenberg; VP, JoAnn Scatina (eds.). Drug Metabolism Handbook. John Wiley & Sons, Inc. pp. 167–227. doi:10.1002/9780470439265.ch8. ISBN 9780470439265.
  20. ^ Arpino, Patrick (1992). "Combined liquid chromatography mass spectrometry. Part III. Applications of thermospray". Mass Spectrometry Reviews. 11 (1): 3–40. Bibcode:1992MSRv...11....3A. doi:10.1002/mas.1280110103.
  21. ^ Arpino, Patrick (1989). "Combined liquid chromatography mass spectrometry. Part I. Coupling by means of a moving belt interface". Mass Spectrometry Reviews. 8 (1): 35–55. Bibcode:1989MSRv....8...35A. doi:10.1002/mas.1280080103.
  22. ^ Murray, Kermit K. (1997). "Coupling matrix-assisted laser desorption/ionization to liquid separations". Mass Spectrometry Reviews. 16 (5): 283–299. Bibcode:1997MSRv...16..283M. doi:10.1002/(SICI)1098-2787(1997)16:5<283::AID-MAS3>3.0.CO;2-D.
  23. ^ Cappiello, Achille; Famiglini, Giorgio; Palma, Pierangela; Pierini, Elisabetta; Termopoli, Veronica; Trufelli, Helga (2008-12-01). "Overcoming Matrix Effects in Liquid Chromatography−Mass Spectrometry". Analytical Chemistry. 80 (23): 9343–9348. doi:10.1021/ac8018312. ISSN 0003-2700. PMID 19551950.
  24. ^ Cappiello, Achille; Famiglini, Giorgio; Mangani, Filippo; Palma, Pierangela (2002-03-01). "A simple approach for coupling liquid chromatography and electron ionization mass spectrometry". Journal of the American Society for Mass Spectrometry. 13 (3): 265–273. doi:10.1016/S1044-0305(01)00363-4. ISSN 1044-0305. PMID 11908806.
  25. ^ Fenn, J. B.; Mann, M.; Meng, C. K.; Wong, S. F.; Whitehouse, C. M. (1989-10-06). "Electrospray ionization for mass spectrometry of large biomolecules". Science. 246 (4926): 64–71. Bibcode:1989Sci...246...64F. CiteSeerX 10.1.1.522.9458. doi:10.1126/science.2675315. ISSN 0036-8075. PMID 2675315.
  26. ^ Covey, Tom (2022-08-30). "Where have all the ions gone, long time passing? Tandem quadrupole mass spectrometers with atmospheric pressure ionization sensitivity gains since the mid‐1970s. A perspective". Rapid Communications in Mass Spectrometry: e9354. doi:10.1002/rcm.9354. ISSN 0951-4198. PMID 35830299. S2CID 250491726.
  27. ^ Horning, E. C.; Horning, M. G.; Carroll, D. I.; Dzidic, I.; Stillwell, R. N. (1973-05-01). "New picogram detection system based on a mass spectrometer with an external ionization source at atmospheric pressure". Analytical Chemistry. 45 (6): 936–943. doi:10.1021/ac60328a035. ISSN 0003-2700.
  28. ^ Robb, null; Covey, null; Bruins, null (2000-08-01). "Atmospheric pressure photoionization: an ionization method for liquid chromatography-mass spectrometry". Analytical Chemistry. 72 (15): 3653–3659. doi:10.1021/ac0001636. ISSN 1520-6882. PMID 10952556.
  29. ^ Widmer, Leo; Watson, Stuart; Schlatter, Konrad; Crowson, Andrew (2002). "Development of an LC/MS method for the trace analysis of triacetone triperoxide (TATP)". The Analyst. 127 (12): 1627–1632. Bibcode:2002Ana...127.1627W. doi:10.1039/b208350g. ISSN 0003-2654. PMID 12537371.
  30. ^ a b Sudhakar, P.; Latha, P.; Reddy, P. V. (2016-04-05). Phenotyping Crop Plants for Physiological and Biochemical Traits. Academic Press. ISBN 9780128041109.
  31. ^ Wysocki VH, Resing KA, Zhang Q, Cheng G (2005). "Mass spectrometry of peptides and proteins". Methods. 35 (3): 211–22. doi:10.1016/j.ymeth.2004.08.013. PMID 15722218.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  32. ^ Karlsson, Roger; Thorsell, Annika; Gomila, Margarita; Salvà-Serra, Francisco; Jakobsson, Hedvig E.; Gonzales-Siles, Lucia; Jaén-Luchoro, Daniel; Skovbjerg, Susann; Fuchs, Johannes; Karlsson, Anders; Boulund, Fredrik (2020-03-01). "Discovery of Species-unique Peptide Biomarkers of Bacterial Pathogens by Tandem Mass Spectrometry-based Proteotyping". Molecular & Cellular Proteomics. 19 (3): 518–528. doi:10.1074/mcp.RA119.001667. ISSN 1535-9476. PMC 7050107. PMID 31941798.
  33. ^ Van Puyvelde, B; Van Uytfanghe, K; Tytgat, O; Van Oudenhove, L; Gabriels, R; Bouwmeester, R; Daled, S; Van Den Bossche, T; Ramasamy, P; Verhelst, S; De Clerck, L; Corveleyn, L; Willems, S; Debunne, N; Wynendaele, E; De Spiegeleer, B; Judak, P; Roels, K; De Wilde, L; Van Eenoo, P; Reyns, T; Cherlet, M; Dumont, E; Debyser, G; t'Kindt, R; Sandra, K; Gupta, S; Drouin, N; Harms, A; Hankemeier, T; Jones, DJL; Gupta, P; Lane, D; Lane, CS; El Ouadi, S; Vincendet, JB; Morrice, N; Oehrle, S; Tanna, N; Silvester, S; Hannam, S; Sigloch, FC; Bhangu-Uhlmann, A; Claereboudt, J; Anderson, NL; Razavi, M; Degroeve, S; Cuypers, L; Stove, C; Lagrou, K; Martens, GA; Deforce, D; Martens, L; Vissers, JPC; Dhaenens, M (28 June 2021). "Cov-MS: A Community-Based Template Assay for Mass-Spectrometry-Based Protein Detection in SARS-CoV-2 Patients". JACS Au. 1 (6): 750–765. doi:10.1021/jacsau.1c00048. PMC 8230961. PMID 34254058.
  34. ^ Gika, Helen G.; Theodoridis, Georgios A.; Plumb, Robert S.; Wilson, Ian D. (January 2014). "Current practice of liquid chromatography–mass spectrometry in metabolomics and metabonomics". Journal of Pharmaceutical and Biomedical Analysis. 87: 12–25. doi:10.1016/j.jpba.2013.06.032. ISSN 0731-7085. PMID 23916607.
  35. ^ Stobiecki, M.; Skirycz, A.; Kerhoas, L.; Kachlicki, P.; Muth, D.; Einhorn, J.; Mueller-Roeber, B. (2006). "Profiling of phenolic glycosidic conjugates in leaves of Arabidopsis thaliana using LC/MS". Metabolomics. 2 (4): 197–219. doi:10.1007/s11306-006-0031-5. S2CID 39140266.
  36. ^ Jorge, Tiago F.; Rodrigues, João A.; Caldana, Camila; Schmidt, Romy; van Dongen, Joost T.; Thomas-Oates, Jane; António, Carla (2016-09-01). "Mass spectrometry-based plant metabolomics: Metabolite responses to abiotic stress". Mass Spectrometry Reviews. 35 (5): 620–649. Bibcode:2016MSRv...35..620J. doi:10.1002/mas.21449. ISSN 1098-2787. PMID 25589422.
  37. ^ Tolstikov, Vladimir V.; Fiehn, Oliver (2002). "Analysis of Highly Polar Compounds of Plant Origin: Combination of Hydrophilic Interaction Chromatography and Electrospray Ion Trap Mass Spectrometry". Analytical Biochemistry. 301 (2): 298–307. doi:10.1006/abio.2001.5513. PMID 11814300. S2CID 3156968.
  38. ^ Antonio, Carla; Larson, Tony; Gilday, Alison; Graham, Ian; Bergström, Ed; Thomas-Oates, Jane (2008). "Hydrophilic interaction chromatography/electrospray mass spectrometry analysis of carbohydrate-related metabolites from Arabidopsis thaliana leaf tissue". Rapid Communications in Mass Spectrometry. 22 (9): 1399–1407. Bibcode:2008RCMS...22.1399A. doi:10.1002/rcm.3519. PMID 18384194.
  39. ^ Lee, Mike S.; Kerns, Edward H. (1999). "LC/MS applications in drug development". Mass Spectrometry Reviews. 18 (3–4): 187–279. Bibcode:1999MSRv...18..187L. doi:10.1002/(SICI)1098-2787(1999)18:3/4<187::AID-MAS2>3.0.CO;2-K. PMID 10568041.

Further reading

  • Thurman, E. M.; Ferrer, Imma (2003). Liquid chromatography/mass spectrometry, MS/MS and time of flight MS: analysis of emerging contaminants. Columbus, OH: American Chemical Society. ISBN 978-0-8412-3825-1.
  • Ferrer, Imma; Thurman, E. M. (2009). Liquid chromatography-Time of Flight Mass Spectrometry: Principles, Tools and Applications for Accurate Mass Analysis. New York, NJ: Wiley. ISBN 978-0-470-13797-0.
  • McMaster, Marvin C. (2005). LC/MS: a practical user's guide. New York: John Wiley. ISBN 978-0-471-65531-2.
  • Yergey, Alfred L. (1990). Liquid chromatography/mass spectrometry: techniques and applications. New York: Plenum Press. ISBN 978-0-306-43186-9.

liquid, chromatography, mass, spectrometry, redirects, here, other, uses, lcms, disambiguation, analytical, chemistry, technique, that, combines, physical, separation, capabilities, liquid, chromatography, hplc, with, mass, analysis, capabilities, mass, spectr. LC MS redirects here For other uses see LCMS disambiguation Liquid chromatography mass spectrometry LC MS is an analytical chemistry technique that combines the physical separation capabilities of liquid chromatography or HPLC with the mass analysis capabilities of mass spectrometry MS Coupled chromatography MS systems are popular in chemical analysis because the individual capabilities of each technique are enhanced synergistically While liquid chromatography separates mixtures with multiple components mass spectrometry provides spectral information that may help to identify or confirm the suspected identity of each separated component 1 MS is not only sensitive but provides selective detection relieving the need for complete chromatographic separation 2 LC MS is also appropriate for metabolomics because of its good coverage of a wide range of chemicals 3 This tandem technique can be used to analyze biochemical organic and inorganic compounds commonly found in complex samples of environmental and biological origin Therefore LC MS may be applied in a wide range of sectors including biotechnology environment monitoring food processing and pharmaceutical agrochemical and cosmetic industries 4 5 Since the early 2000s LC MS or more specifically LC MS MS has also begun to be used in clinical applications 6 Liquid chromatography mass spectrometryIon trap LCMS system with ESI interfaceAcronymLCMSClassificationChromatographyMass spectrometryAnalytesorganic moleculesbiomoleculesManufacturersAgilent Bruker PerkinElmer SCIEX Shimadzu Scientific Thermo Fisher Scientific Waters CorporationOther techniquesRelatedGas chromatography mass spectrometryIn addition to the liquid chromatography and mass spectrometry devices an LC MS system contains an interface that efficiently transfers the separated components from the LC column into the MS ion source 5 7 The interface is necessary because the LC and MS devices are fundamentally incompatible While the mobile phase in a LC system is a pressurized liquid the MS analyzers commonly operate under high vacuum Thus it is not possible to directly pump the eluate from the LC column into the MS source Overall the interface is a mechanically simple part of the LC MS system that transfers the maximum amount of analyte removes a significant portion of the mobile phase used in LC and preserves the chemical identity of the chromatography products chemically inert As a requirement the interface should not interfere with the ionizing efficiency and vacuum conditions of the MS system 5 Nowadays most extensively applied LC MS interfaces are based on atmospheric pressure ionization API strategies like electrospray ionization ESI atmospheric pressure chemical ionization APCI and atmospheric pressure photoionization APPI These interfaces became available in the 1990s after a two decade long research and development process 8 7 Contents 1 History of LC MS 1 1 Moving Belt Interface 1 2 Direct liquid introduction interface 1 3 Thermospray interface 1 4 FAB based interfaces 2 Liquid chromatography 3 Mass spectrometry 4 Interfaces 4 1 Electrospray ionization ESI 4 2 Atmospheric pressure chemical ionization APCI 4 3 Atmospheric pressure photoionization APPI 5 Applications 5 1 Pharmacokinetics 5 2 Proteomics metabolomics 5 3 Drug development 6 See also 7 References 8 Further readingHistory of LC MS EditThe coupling of chromatography with MS is a well developed chemical analysis strategy dating back from the 1950s Gas chromatography GC MS was originally introduced in 1952 when A T James and A J P Martin were trying to develop tandem separation mass analysis techniques 9 In GC the analytes are eluted from the separation column as a gas and the connection with electron ionization EI or chemical ionization CI ion sources in the MS system was a technically simpler challenge Because of this the development of GC MS systems was faster than LC MS and such systems were first commercialized in the 1970s 7 The development of LC MS systems took longer than GC MS and was directly related to the development of proper interfaces V L Tal roze and collaborators started the development of LC MS in the late 1960s 10 11 when they first used capillaries to connect an LC columns to an EI source 12 8 A similar strategy was investigated by McLafferty and collaborators in 1973 who coupled the LC column to a CI source 13 which allowed a higher liquid flow into the source This was the first and most obvious way of coupling LC with MS and was known as the capillary inlet interface This pioneer interface for LC MS had the same analysis capabilities of GC MS and was limited to rather volatile analytes and non polar compounds with low molecular mass below 400 Da In the capillary inlet interface the evaporation of the mobile phase inside the capillary was one of the main issues Within the first years of development of LC MS on line and off line alternatives were proposed as coupling alternatives In general off line coupling involved fraction collection evaporation of solvent and transfer of analytes to the MS using probes Off line analyte treatment process was time consuming and there was an inherent risk of sample contamination Rapidly it was realized that the analysis of complex mixtures would require the development of a fully automated on line coupling solution in LC MS 8 The key to the success and wide spread adoption of LC MS as a routine analytical tool lies in the interface and ion source between the liquid based LC and the vacuum base MS The following interfaces were stepping stones on the way to the modern atmospheric pressure ionization interfaces and are described for historical interest Moving Belt Interface Edit The moving belt interface MBI was developed by McFadden et al in 1977 and commercialized by Finnigan 14 This interface consisted of an endless moving belt onto which the LC column effluent was deposited in a band On the belt the solvent was evaporated by gently heating and efficiently exhausting the solvent vapours under reduced pressure in two vacuum chambers After the liquid phase was removed the belt passed over a heater which flash desorbed the analytes into the MS ion source One of the significant advantages of the MBI was its compatibility with a wide range of chromatographic conditions 8 MBI was successfully used for LC MS applications between 1978 and 1990 because it allowed coupling of LC to MS devices using EI CI and fast atom bombardment FAB ion sources The most common MS systems connected by MBI interfaces to LC columns were magnetic sector and quadrupole instruments MBI interfaces for LC MS allowed MS to be widely applied in the analysis of drugs pesticides steroids alkaloids and polycyclic aromatic hydrocarbons This interface is no longer used because of its mechanical complexity and the difficulties associated with belt renewal as well as its inability to handle very labile biomolecules Direct liquid introduction interface Edit The direct liquid introduction DLI interface was developed in 1980 This interface was intended to solve the problem of evaporation of liquid inside the capillary inlet interface In DLI a small portion of the LC flow was forced through a small aperture or diaphragm typically 10um in diameter to form a liquid jet composed of small droplets that were subsequently dried in a desolvation chamber 11 The analytes were ionized using a solvent assisted chemical ionization source where the LC solvents acted as reagent gases To use this interface it was necessary to split the flow coming out of the LC column because only a small portion of the effluent 10 to 50 ml min out of 1 ml min could be introduced into the source without raising the vacuum pressure of the MS system too high Alternately Henion at Cornell University had success with using micro bore LC methods so that the entire low flow of the LC could be used One of the main operational problems of the DLI interface was the frequent clogging of the diaphragm orifices The DLI interface was used between 1982 and 1985 for the analysis of pesticides corticosteroids metabolites in horse urine erythromycin and vitamin B12 However this interface was replaced by the thermospray interface which removed the flow rate limitations and the issues with the clogging diaphragms 5 8 A related device was the particle beam interface PBI developed by Willoughby and Browner in 1984 15 Particle beam interfaces took over the wide applications of MBI for LC MS in 1988 8 16 The PBI operated by using a helium gas nebulizer to spray the eluant into the vacuum drying the droplets and pumping away the solvent vapour using a jet separator while the stream of monodisperse dried particles containing the analyte entered the source 11 Drying the droplets outside of the source volume and using a jet separator to pump away the solvent vapour allowed the particles to enter and be vapourized in a low pressure EI source As with the MBI the ability to generate library searchable EI spectra was a distinct advantage for many applications Commercialized by Hewlett Packard and later by VG and Extrel it enjoyed moderate success but has been largely supplanted by the atmospheric pressure interfaces such as electrospray and APCI which provide a broader range of compound coverage and applications Thermospray interface Edit The thermospray TSP interface was developed in 1980 by Marvin Vestal and co workers at the University of Houston 17 It was commercialized by Vestec and several of the major mass spectrometer manufacurers The interface resulted from a long term research project intended to find a LC MS interface capable of handling high flow rates 1 ml min and avoiding the flow split in DLI interfaces The TSP interface was composed of a heated probe a desolvation chamber and an ion focusing skimmer The LC effluent passed through the heated probe and emerged as a jet of vapor and small droplets flowing into the desolvation chamber at low pressure Initially operated with a filament or discharge as the source of ions thereby acting as a CI source for vapourized analyte it was soon discovered that ions were also observed when the filament or discharge was off This could be attributed to either direct emission of ions from the liquid droplets as they evaporated in a process related to electrospray ionization or ion evaporation or to chemical ionization of vapourized analyte molecules from buffer ions such as ammonium acetate The fact that multiply charged ions were observed from some larger analytes suggests that direct analyte ion emission was occurring under at least some conditions 11 The interface was able to handle up to 2 ml min of eluate from the LC column and would efficiently introduce it into the MS vacuum system TSP was also more suitable for LC MS applications involving reversed phase liquid chromatography RT LC With time the mechanical complexity of TSP was simplified and this interface became popular as the first ideal LC MS interface for pharmaceutical applications comprising the analysis of drugs metabolites conjugates nucleosides peptides natural products and pesticides The introduction of TSP marked a significant improvement for LC MS systems and was the most widely applied interface until the beginning of the 1990s when it began to be replaced by interfaces involving atmospheric pressure ionization API 5 7 16 FAB based interfaces Edit The frit fast atom bombardment FAB and continuous flow FAB CF FAB interfaces were developed in 1985 and 1986 respectively 16 Both interfaces were similar but they differed in that the first used a porous frit probe as connecting channel while CF FAB used a probe tip From these the CF FAB was more successful as a LC MS interface and was useful to analyze non volatile and thermally labile compounds In these interfaces the LC effluent passed through the frit or CF FAB channels to form a uniform liquid film at the tip There the liquid was bombarded with ion beams or high energy atoms fast atoms For stable operation the FAB based interfaces were able to handle liquid flow rates of only 1 15 ml and were also restricted to microbore and capillary columns In order to be used in FAB MS ionization sources the analytes of interest had to be mixed with a matrix e g glycerol that could be added before or after the separation in the LC column FAB based interfaces were extensively used to characterize peptides but lost applicability with the advent of electrospray based interfaces in 1988 5 8 Liquid chromatography Edit Diagram of an LC MS system Main article High performance liquid chromatography Liquid chromatography is a method of physical separation in which the components of a liquid mixture are distributed between two immiscible phases i e stationary and mobile The practice of LC can be divided into five categories i e adsorption chromatography partition chromatography ion exchange chromatography size exclusion chromatography and affinity chromatography Among these the most widely used variant is the reverse phase RP mode of the partition chromatography technique which makes use of a nonpolar hydrophobic stationary phase and a polar mobile phase In common applications the mobile phase is a mixture of water and other polar solvents e g methanol isopropanol and acetonitrile and the stationary matrix is prepared by attaching long chain alkyl groups e g n octadecyl or C18 to the external and internal surfaces of irregularly or spherically shaped 5 mm diameter porous silica particles 5 In HPLC typically 20 ml of the sample of interest are injected into the mobile phase stream delivered by a high pressure pump The mobile phase containing the analytes permeates through the stationary phase bed in a definite direction The components of the mixture are separated depending on their chemical affinity with the mobile and stationary phases The separation occurs after repeated sorption and desorption steps occurring when the liquid interacts with the stationary bed 8 The liquid solvent mobile phase is delivered under high pressure up to 400 bar or 5800 psi into a packed column containing the stationary phase The high pressure is necessary to achieve a constant flow rate for reproducible chromatography experiments Depending on the partitioning between the mobile and stationary phases the components of the sample will flow out of the column at different times 16 The column is the most important component of the LC system and is designed to withstand the high pressure of the liquid Conventional LC columns are 100 300 mm long with outer diameter of 6 4 mm 1 4 inch and internal diameter of 3 0 4 6 mm For applications involving LC MS the length of chromatography columns can be shorter 30 50 mm with 3 5 mm diameter packing particles In addition to the conventional model other LC columns are the narrow bore microbore microcapillary and nano LC models These columns have smaller internal diameters allow for a more efficient separation and handle liquid flows under 1 ml min the conventional flow rate 8 In order to improve separation efficiency and peak resolution ultra performance liquid chromatography UHPLC can be used instead of HPLC This LC variant uses columns packed with smaller silica particles 1 7 mm diameter and requires higher operating pressures in the range of 310000 to 775000 torr 6000 to 15000 psi 400 to 1034 bar 5 Mass spectrometry Edit LC MS spectrum of each resolved peak Main article Mass spectrometry Mass spectrometry MS is an analytical technique that measures the mass to charge ratio m z of charged particles ions Although there are many different kinds of mass spectrometers all of them make use of electric or magnetic fields to manipulate the motion of ions produced from an analyte of interest and determine their m z 18 The basic components of a mass spectrometer are the ion source the mass analyzer the detector and the data and vacuum systems The ion source is where the components of a sample introduced in a MS system are ionized by means of electron beams photon beams UV lights laser beams or corona discharge In the case of electrospray ionization the ion source moves ions that exist in liquid solution into the gas phase The ion source converts and fragments the neutral sample molecules into gas phase ions that are sent to the mass analyzer While the mass analyzer applies the electric and magnetic fields to sort the ions by their masses the detector measures and amplifies the ion current to calculate the abundances of each mass resolved ion In order to generate a mass spectrum that a human eye can easily recognize the data system records processes stores and displays data in a computer 5 The mass spectrum can be used to determine the mass of the analytes their elemental and isotopic composition or to elucidate the chemical structure of the sample 5 MS is an experiment that must take place in gas phase and under vacuum 1 33 10 2 to 1 33 10 6 pascal Therefore the development of devices facilitating the transition from samples at higher pressure and in condensed phase solid or liquid into a vacuum system has been essential to develop MS as a potent tool for identification and quantification of organic compounds like peptides 19 MS is now in very common use in analytical laboratories that study physical chemical or biological properties of a great variety of compounds Among the many different kinds of mass analyzers the ones that find application in LC MS systems are the quadrupole time of flight TOF ion traps and hybrid quadrupole TOF QTOF analyzers 7 Interfaces EditThe interface between a liquid phase technique HPLC with a continuously flowing eluate and a gas phase technique carried out in a vacuum was difficult for a long time The advent of electrospray ionization changed this Currently the most common LC MS interfaces are electrospray ionization ESI atmospheric pressure chemical ionization APCI and atmospheric pressure photo ionization APPI These are newer MS ion sources that facilitate the transition from a high pressure environment HPLC to high vacuum conditions needed at the MS analyzer 20 7 Although these interfaces are described individually they can also be commercially available as dual ESI APCI ESI APPI or APCI APPI ion sources 8 Various deposition and drying techniques were used in the past e g moving belts but the most common of these was the off line MALDI deposition 21 22 A new approach still under development called direct EI LC MS interface couples a nano HPLC system and an electron ionization equipped mass spectrometer 23 24 Electrospray ionization ESI Edit Main article Electrospray ionization ESI interface for LC MS systems was developed by Fenn and collaborators in 1988 25 This ion source interface can be used for the analysis of moderately polar and even very polar molecules e g metabolites xenobiotics peptides nucleotides polysaccharides The liquid eluate coming out of the LC column is directed into a metal capillary kept at 3 to 5 kV and is nebulized by a high velocity coaxial flow of gas at the tip of the capillary creating a fine spray of charged droplets in front of the entrance to the vacuum chamber To avoid contamination of the vacuum system by buffers and salts this capillary is usually perpendicularly located at the inlet of the MS system in some cases with a counter current of dry nitrogen in front of the entrance through which ions are directed by the electric field In some sources rapid droplet evaporation and thus maximum ion emission is achieved by mixing an additional stream of hot gas with the spray plume in front of the vacuum entrance In other sources the droplets are drawn through a heated capillary tube as they enter the vacuum promoting droplet evaporation and ion emission These methods of increasing droplet evaporation now allow the use of liquid flow rates of 1 2 mL min to be used while still achieving efficient ionisation 26 and high sensitivity Thus while the use of 1 3 mm microbore columns and lower flow rates of 50 200 ml min was commonly considered necessary for optimum operation this limitation is no longer as important and the higher column capacity of larger bore columns can now be advantageously employed with ESI LC MS systems Positively and negatively charged ions can be created by switching polarities and it is possible to acquire alternate positive and negative mode spectra rapidly within the same LC run While most large molecules greater than MW 1500 2000 produce multiply charged ions in the ESI source the majority of smaller molecules produce singly charged ions 7 Atmospheric pressure chemical ionization APCI Edit Main article Atmospheric pressure chemical ionization The development of the APCI interface for LC MS started with Horning and collaborators in the early 1973 27 However its commercial application was introduced at the beginning of the 1990s after Henion and collaborators improved the LC APCI MS interface in 1986 8 The APCI ion source interface can be used to analyze small neutral relatively non polar and thermally stable molecules e g steroids lipids and fat soluble vitamins These compounds are not well ionized using ESI In addition APCI can also handle mobile phase streams containing buffering agents The liquid from the LC system is pumped through a capillary and there is also nebulization at the tip where a corona discharge takes place First the ionizing gas surrounding the interface and the mobile phase solvent are subject to chemical ionization at the ion source Later these ions react with the analyte and transfer their charge The sample ions then pass through small orifice skimmers by means of or ion focusing lenses Once inside the high vacuum region the ions are subject to mass analysis This interface can be operated in positive and negative charge modes and singly charged ions are mainly produced 7 APCI ion source can also handle flow rates between 500 and 2000 ml min and it can be directly connected to conventional 4 6 mm ID columns 16 Atmospheric pressure photoionization APPI Edit Main article Atmospheric pressure photoionization The APPI interface for LC MS was developed simultaneously by Bruins and Syage in 2000 28 8 APPI is another LC MS ion source interface for the analysis of neutral compounds that cannot be ionized using ESI 7 This interface is similar to the APCI ion source but instead of a corona discharge the ionization occurs by using photons coming from a discharge lamp In the direct APPI mode singly charged analyte molecular ions are formed by absorption of a photon and ejection of an electron In the dopant APPI mode an easily ionizable compound Dopant is added to the mobile phase or the nebulizing gas to promote a reaction of charge exchange between the dopant molecular ion and the analyte The ionized sample is later transferred to the mass analyzer at high vacuum as it passes through small orifice skimmers 8 Applications EditThe coupling of MS with LC systems is attractive because liquid chromatography can separate delicate and complex natural mixtures which chemical composition needs to be well established e g biological fluids environmental samples and drugs Further LC MS has applications in volatile explosive residue analysis 29 Nowadays LC MS has become one of the most widely used chemical analysis techniques because more than 85 of natural chemical compounds are polar and thermally labile and GC MS cannot process these samples 5 As an example HPLC MS is regarded as the leading analytical technique for proteomics and pharmaceutical laboratories 7 5 Other important applications of LC MS include the analysis of food pesticides and plant phenols 8 Pharmacokinetics Edit LC MS is widely used in the field of bioanalysis and is specially involved in pharmacokinetic studies of pharmaceuticals Pharmacokinetic studies are needed to determine how quickly a drug will be cleared from the body organs and the hepatic blood flow MS analyzers are useful in these studies because of their shorter analysis time and higher sensitivity and specificity compared to UV detectors commonly attached to HPLC systems One major advantage is the use of tandem MS MS where the detector may be programmed to select certain ions to fragment The measured quantity is the sum of molecule fragments chosen by the operator As long as there are no interferences or ion suppression in LC MS the LC separation can be quite quick 30 Proteomics metabolomics Edit LC MS is used in proteomics as a method to detect and identify the components of a complex mixture The bottom up proteomics LC MS approach generally involves protease digestion and denaturation using trypsin as a protease urea to denature the tertiary structure and iodoacetamide to modify the cysteine residues After digestion LC MS is used for peptide mass fingerprinting or LC MS MS tandem MS is used to derive the sequences of individual peptides 31 LC MS MS is most commonly used for proteomic analysis of complex samples where peptide masses may overlap even with a high resolution mass spectrometry Samples of complex biological e g human serum may be analyzed in modern LC MS MS systems which can identify over 1000 proteins However this high level of protein identification is possible only after separating the sample by means of SDS PAGE gel or HPLC SCX 30 Recently LC MS MS has been applied to search peptide biomarkers Examples are the recent discovery and validation of peptide biomarkers for four major bacterial respiratory tract pathogens Staphylococcus aureus Moraxella catarrhalis Haemophilus influenzae and Streptococcus pneumoniae and the SARS CoV 2 virus 32 33 LC MS has emerged as one of the most commonly used techniques in global metabolite profiling of biological tissue e g blood plasma serum urine 34 LC MS is also used for the analysis of natural products and the profiling of secondary metabolites in plants 35 In this regard MS based systems are useful to acquire more detailed information about the wide spectrum of compounds from a complex biological samples LC Nuclear magnetic resonance NMR is also used in plant metabolomics but this technique can only detect and quantify the most abundant metabolites LC MS has been useful to advance the field of plant metabolomics which aims to study the plant system at molecular level providing a non biased characterization of the plant metabolome in response to its environment 36 The first application of LC MS in plant metabolomics was the detection of a wide range of highly polar metabolites oligosaccharides amino acids amino sugars and sugar nucleotides from Cucurbita maxima phloem tissues 37 Another example of LC MS in plant metabolomics is the efficient separation and identification of glucose sucrose raffinose stachyose and verbascose from leaf extracts of Arabidopsis thaliana 38 Drug development Edit LC MS is frequently used in drug development because it allows quick molecular weight confirmation and structure identification These features speed up the process of generating testing and validating a discovery starting from a vast array of products with potential application LC MS applications for drug development are highly automated methods used for peptide mapping glycoprotein mapping lipodomics natural products dereplication bioaffinity screening in vivo drug screening metabolic stability screening metabolite identification impurity identification quantitative bioanalysis and quality control 39 See also EditGas chromatography mass spectrometry Capillary electrophoresis mass spectrometry Ion mobility spectrometry mass spectrometryReferences Edit de Hoffmann Edmond Stroobant Vincent 2002 Mass Spectrometry Principles and Applications 2nd ed Wiley pp 157 158 ISBN 0 471 48566 7 Pitt James J February 2009 Principles and Applications of Liquid Chromatography Mass Spectrometry in Clinical Biochemistry Clin Biochem Rev 30 1 19 34 PMC 2643089 PMID 19224008 Zhou Bin Xiao Jun Feng Tuli Leepika Ressom Habtom W Feb 2012 LC MS based metabolomics Mol Biosyst 8 2 470 481 doi 10 1039 c1mb05350g PMC 3699692 PMID 22041788 Chaimbault Patrick 2014 01 01 The Modern Art of Identification of Natural Substances in Whole Plants In Jacob Claus Kirsch Gilbert Slusarenko Alan Winyard Paul G Burkholz Torsten eds Recent Advances in Redox Active Plant and Microbial Products Springer Netherlands pp 31 94 doi 10 1007 978 94 017 8953 0 3 ISBN 9789401789523 a b c d e f g h i j k l Dass Chhabil 2007 01 01 Hyphenated Separation Techniques Fundamentals of Contemporary Mass Spectrometry John Wiley amp Sons Inc pp 151 194 doi 10 1002 9780470118498 ch5 ISBN 9780470118498 Seger Christoph Salzmann Linda 2020 08 01 After another decade LC MS MS became routine in clinical diagnostics Clinical Biochemistry Advancement and Applications of Mass Spectrometry in Laboratory Medicine 82 2 11 doi 10 1016 j clinbiochem 2020 03 004 ISSN 0009 9120 PMID 32188572 S2CID 213186669 a b c d e f g h i j Pitt James J 2017 03 12 Principles and Applications of Liquid Chromatography Mass Spectrometry in Clinical Biochemistry The Clinical Biochemist Reviews 30 1 19 34 ISSN 0159 8090 PMC 2643089 PMID 19224008 a b c d e f g h i j k l m n Niessen Wilfried M A 2006 Liquid Chromatography Mass Spectrometry Third Edition Boca Raton CRC Taylor amp Francis pp 50 90 ISBN 9780824740825 OCLC 232370223 James A T Martin A J P 1952 03 01 Gas liquid partition chromatography the separation and micro estimation of volatile fatty acids from formic acid to dodecanoic acid Biochemical Journal 50 5 679 690 doi 10 1042 bj0500679 ISSN 0264 6021 PMC 1197726 PMID 14934673 Tal roze V L Karpov G V Gordetskii I G Skurat V E 1968 Capillary Systems for the Introduction of Liquid Mixtures into an Analytical Mass Spectrometer Russ J Phys Chem 42 1658 1664 a b c d Arpino Patrick 2006 History of LC MS Development and Interfacing Encyclopedia of Mass Spectrometry Vol 8 Elsevier ISBN 978 0080438474 Tal roze V L Gorodetskii I G Zolotoy N B Karpov G V Skurat V E Maslennikova V Ya 1978 Capillary system for continuous introducing of volatile liquids into analytical MS and its application Adv Mass Spectrom 7 858 Baldwin M A McLafferty F W 1973 Liquid chromatography mass spectrometry interface I The direct introduction of liquid solutions into a chemical ionization mass spectrometer Organic Mass Spectrometry 7 9 1111 1112 doi 10 1002 oms 1210070913 ISSN 0030 493X Pullen Franl 2010 The fascinating history of the development of LC MS a personal perspective Chromatography Today February March 4 6 de Koster Chris G Schoenmakers Peter J 2020 Chapter 3 1 History of Liquid Chromatography Mass Spectrometry In Tranchida Peter Q Mondello Luigi eds Hyphenations of Capillary Chromatography With Mass Spectrometry Elsevier pp 279 295 ISBN 978 0 12 809638 3 a b c d e Ardrey Robert E 2003 01 01 Introduction Liquid Chromatography Mass Spectrometry An Introduction Analytical Techniques in the Sciences AnTS John Wiley amp Sons Ltd pp 1 5 doi 10 1002 0470867299 ch1 ISBN 9780470867297 Blakley C R Carmody J J Vestal M L 1980 A New Soft Ionization Technique for Mass Spectrometry of Complex Samples J Am Chem Soc 102 5931 5933 doi 10 1021 ja00538a050 Roberts Gordon 2013 Roberts Gordon C K ed Encyclopedia of Biophysics Springer doi 10 1007 978 3 642 16712 6 ISBN 978 3 642 16711 9 S2CID 44856071 Sharp Thomas R 2009 01 01 Mass Spectrometry In Nassar Ala F Collegiateessor Paul F Hollenberg VP JoAnn Scatina eds Drug Metabolism Handbook John Wiley amp Sons Inc pp 167 227 doi 10 1002 9780470439265 ch8 ISBN 9780470439265 Arpino Patrick 1992 Combined liquid chromatography mass spectrometry Part III Applications of thermospray Mass Spectrometry Reviews 11 1 3 40 Bibcode 1992MSRv 11 3A doi 10 1002 mas 1280110103 Arpino Patrick 1989 Combined liquid chromatography mass spectrometry Part I Coupling by means of a moving belt interface Mass Spectrometry Reviews 8 1 35 55 Bibcode 1989MSRv 8 35A doi 10 1002 mas 1280080103 Murray Kermit K 1997 Coupling matrix assisted laser desorption ionization to liquid separations Mass Spectrometry Reviews 16 5 283 299 Bibcode 1997MSRv 16 283M doi 10 1002 SICI 1098 2787 1997 16 5 lt 283 AID MAS3 gt 3 0 CO 2 D Cappiello Achille Famiglini Giorgio Palma Pierangela Pierini Elisabetta Termopoli Veronica Trufelli Helga 2008 12 01 Overcoming Matrix Effects in Liquid Chromatography Mass Spectrometry Analytical Chemistry 80 23 9343 9348 doi 10 1021 ac8018312 ISSN 0003 2700 PMID 19551950 Cappiello Achille Famiglini Giorgio Mangani Filippo Palma Pierangela 2002 03 01 A simple approach for coupling liquid chromatography and electron ionization mass spectrometry Journal of the American Society for Mass Spectrometry 13 3 265 273 doi 10 1016 S1044 0305 01 00363 4 ISSN 1044 0305 PMID 11908806 Fenn J B Mann M Meng C K Wong S F Whitehouse C M 1989 10 06 Electrospray ionization for mass spectrometry of large biomolecules Science 246 4926 64 71 Bibcode 1989Sci 246 64F CiteSeerX 10 1 1 522 9458 doi 10 1126 science 2675315 ISSN 0036 8075 PMID 2675315 Covey Tom 2022 08 30 Where have all the ions gone long time passing Tandem quadrupole mass spectrometers with atmospheric pressure ionization sensitivity gains since the mid 1970s A perspective Rapid Communications in Mass Spectrometry e9354 doi 10 1002 rcm 9354 ISSN 0951 4198 PMID 35830299 S2CID 250491726 Horning E C Horning M G Carroll D I Dzidic I Stillwell R N 1973 05 01 New picogram detection system based on a mass spectrometer with an external ionization source at atmospheric pressure Analytical Chemistry 45 6 936 943 doi 10 1021 ac60328a035 ISSN 0003 2700 Robb null Covey null Bruins null 2000 08 01 Atmospheric pressure photoionization an ionization method for liquid chromatography mass spectrometry Analytical Chemistry 72 15 3653 3659 doi 10 1021 ac0001636 ISSN 1520 6882 PMID 10952556 Widmer Leo Watson Stuart Schlatter Konrad Crowson Andrew 2002 Development of an LC MS method for the trace analysis of triacetone triperoxide TATP The Analyst 127 12 1627 1632 Bibcode 2002Ana 127 1627W doi 10 1039 b208350g ISSN 0003 2654 PMID 12537371 a b Sudhakar P Latha P Reddy P V 2016 04 05 Phenotyping Crop Plants for Physiological and Biochemical Traits Academic Press ISBN 9780128041109 Wysocki VH Resing KA Zhang Q Cheng G 2005 Mass spectrometry of peptides and proteins Methods 35 3 211 22 doi 10 1016 j ymeth 2004 08 013 PMID 15722218 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint multiple names authors list link Karlsson Roger Thorsell Annika Gomila Margarita Salva Serra Francisco Jakobsson Hedvig E Gonzales Siles Lucia Jaen Luchoro Daniel Skovbjerg Susann Fuchs Johannes Karlsson Anders Boulund Fredrik 2020 03 01 Discovery of Species unique Peptide Biomarkers of Bacterial Pathogens by Tandem Mass Spectrometry based Proteotyping Molecular amp Cellular Proteomics 19 3 518 528 doi 10 1074 mcp RA119 001667 ISSN 1535 9476 PMC 7050107 PMID 31941798 Van Puyvelde B Van Uytfanghe K Tytgat O Van Oudenhove L Gabriels R Bouwmeester R Daled S Van Den Bossche T Ramasamy P Verhelst S De Clerck L Corveleyn L Willems S Debunne N Wynendaele E De Spiegeleer B Judak P Roels K De Wilde L Van Eenoo P Reyns T Cherlet M Dumont E Debyser G t Kindt R Sandra K Gupta S Drouin N Harms A Hankemeier T Jones DJL Gupta P Lane D Lane CS El Ouadi S Vincendet JB Morrice N Oehrle S Tanna N Silvester S Hannam S Sigloch FC Bhangu Uhlmann A Claereboudt J Anderson NL Razavi M Degroeve S Cuypers L Stove C Lagrou K Martens GA Deforce D Martens L Vissers JPC Dhaenens M 28 June 2021 Cov MS A Community Based Template Assay for Mass Spectrometry Based Protein Detection in SARS CoV 2 Patients JACS Au 1 6 750 765 doi 10 1021 jacsau 1c00048 PMC 8230961 PMID 34254058 Gika Helen G Theodoridis Georgios A Plumb Robert S Wilson Ian D January 2014 Current practice of liquid chromatography mass spectrometry in metabolomics and metabonomics Journal of Pharmaceutical and Biomedical Analysis 87 12 25 doi 10 1016 j jpba 2013 06 032 ISSN 0731 7085 PMID 23916607 Stobiecki M Skirycz A Kerhoas L Kachlicki P Muth D Einhorn J Mueller Roeber B 2006 Profiling of phenolic glycosidic conjugates in leaves of Arabidopsis thaliana using LC MS Metabolomics 2 4 197 219 doi 10 1007 s11306 006 0031 5 S2CID 39140266 Jorge Tiago F Rodrigues Joao A Caldana Camila Schmidt Romy van Dongen Joost T Thomas Oates Jane Antonio Carla 2016 09 01 Mass spectrometry based plant metabolomics Metabolite responses to abiotic stress Mass Spectrometry Reviews 35 5 620 649 Bibcode 2016MSRv 35 620J doi 10 1002 mas 21449 ISSN 1098 2787 PMID 25589422 Tolstikov Vladimir V Fiehn Oliver 2002 Analysis of Highly Polar Compounds of Plant Origin Combination of Hydrophilic Interaction Chromatography and Electrospray Ion Trap Mass Spectrometry Analytical Biochemistry 301 2 298 307 doi 10 1006 abio 2001 5513 PMID 11814300 S2CID 3156968 Antonio Carla Larson Tony Gilday Alison Graham Ian Bergstrom Ed Thomas Oates Jane 2008 Hydrophilic interaction chromatography electrospray mass spectrometry analysis of carbohydrate related metabolites from Arabidopsis thaliana leaf tissue Rapid Communications in Mass Spectrometry 22 9 1399 1407 Bibcode 2008RCMS 22 1399A doi 10 1002 rcm 3519 PMID 18384194 Lee Mike S Kerns Edward H 1999 LC MS applications in drug development Mass Spectrometry Reviews 18 3 4 187 279 Bibcode 1999MSRv 18 187L doi 10 1002 SICI 1098 2787 1999 18 3 4 lt 187 AID MAS2 gt 3 0 CO 2 K PMID 10568041 Further reading EditThurman E M Ferrer Imma 2003 Liquid chromatography mass spectrometry MS MS and time of flight MS analysis of emerging contaminants Columbus OH American Chemical Society ISBN 978 0 8412 3825 1 Ferrer Imma Thurman E M 2009 Liquid chromatography Time of Flight Mass Spectrometry Principles Tools and Applications for Accurate Mass Analysis New York NJ Wiley ISBN 978 0 470 13797 0 McMaster Marvin C 2005 LC MS a practical user s guide New York John Wiley ISBN 978 0 471 65531 2 Yergey Alfred L 1990 Liquid chromatography mass spectrometry techniques and applications New York Plenum Press ISBN 978 0 306 43186 9 Retrieved from https en wikipedia org w index php title Liquid chromatography mass spectrometry amp oldid 1142911706, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.