High-performance liquid chromatography

High-performance liquid chromatography (HPLC), formerly referred to as high-pressure liquid chromatography, is a technique in analytical chemistry used to separate, identify, and quantify specific components in mixtures. The mixtures can originate from food, chemicals, pharmaceuticals,[1] biological, environmental and agriculture, etc, which have been dissolved into liquid solutions.

A modern self-contained HPLC.
Schematic representation of an HPLC unit. (1) Solvent reservoirs, (2) Solvent degasser, (3) Gradient valve, (4) Mixing vessel for delivery of the mobile phase, (5) High-pressure pump, (6) Switching valve in "inject position", (6') Switching valve in "load position", (7) Sample injection loop, (8) Pre-column (guard column), (9) Analytical column, (10) Detector (i.e., IR, UV), (11) Data acquisition, (12) Waste or fraction collector.

It relies on high pressure pumps, which deliver mixtures of various solvents, called the mobile phase, which flows through the system, collecting the sample mixture on the way, delivering it into a cylinder, called the column, filled with solid particles, made of adsorbent material, called the stationary phase.

Each component in the sample interacts differently with the adsorbent material, causing different migration rates for each component. These different rates lead to separation as the species flow out of the column into a specific detector such as UV detectors. The output of the detector is a graph, called a chromatogram. Chromatograms are graphical representations of the signal intensity versus time or volume, showing peaks, which represent components of the sample. Each sample appears in its respective time, called its retention time, having area proportional to its amount.

HPLC is widely used for manufacturing (e.g., during the production process of pharmaceutical and biological products),[2][3] legal (e.g., detecting performance enhancement drugs in urine),[4] research (e.g., separating the components of a complex biological sample, or of similar synthetic chemicals from each other), and medical (e.g., detecting vitamin D levels in blood serum) purposes.[5]

Chromatography can be described as a mass transfer process involving adsorption and/or partition. As mentioned, HPLC relies on pumps to pass a pressurized liquid and a sample mixture through a column filled with adsorbent, leading to the separation of the sample components. The active component of the column, the adsorbent, is typically a granular material made of solid particles (e.g., silica, polymers, etc.), 1.5–50 μm in size, on which various reagents can be bonded. The components of the sample mixture are separated from each other due to their different degrees of interaction with the adsorbent particles. The pressurized liquid is typically a mixture of solvents (e.g., water, buffers, acetonitrile and/or methanol) and is referred to as a "mobile phase". Its composition and temperature play a major role in the separation process by influencing the interactions taking place between sample components and adsorbent. These interactions are physical in nature, such as hydrophobic (dispersive), dipole–dipole and ionic, most often a combination.

Operation edit

The liquid chromatograph is complex[6] and has sophisticated and delicate technology. In order to properly operate the system. It is also desirable that there should be a minimum basis for understanding of how the device performs the data Processing to avoid incorrect data and distorted results[7][8][9]

HPLC is distinguished from traditional ("low pressure") liquid chromatography because operational pressures are significantly higher (50–~1400 bar), while ordinary liquid chromatography typically relies on the force of gravity to pass the mobile phase through the packed column. Due to the small sample amount separated in analytical HPLC, typical column dimensions are 2.1–4.6 mm diameter, and 30–250 mm length. Also HPLC columns are made with smaller adsorbent particles (1.5–50 μm in average particle size). This gives HPLC superior resolving power (the ability to distinguish between compounds) when separating mixtures, which makes it a popular chromatographic technique.

The schematic of an HPLC instrument typically includes solvents' reservoirs, one or more pumps, a solvent-degasser, a sampler, a column, and a detector. The solvents are prepared in advance according to the needs of the separation, they pass through the degasser to remove dissolved gasses, mixed to become the mobile phase, then flow through the sampler, which brings the sample mixture into the mobile phase stream, which then carries it into the column. The pumps deliver the desired flow and composition of the mobile phase through the stationary phase inside the column, then directly into a flow-cell inside the detector. The detector generates a signal proportional to the amount of sample component emerging from the column, hence allowing for quantitative analysis of the sample components. The detector also marks the time of emergence, the Retention Time, which serves for initial identification of the component. More advanced detectors, provide also additional information, specific to the analyte's characteristics, such as UV-VIS spectrum or Mass spectrum, which can provide insight on its structural features. These detectors are in common use, such as UV/Vis, Photodiode array (PDA) / diode array detector and mass spectrometry detector.

A digital microprocessor and user software control the HPLC instrument and provide data analysis. Some models of mechanical pumps in an HPLC instrument can mix multiple solvents together at a ratios changing in time, generating a composition gradient in the mobile phase. Most HPLC instruments also have a column oven that allows for adjusting the temperature at which the separation is performed.

The sample mixture to be separated and analyzed is introduced, in a discrete small volume (typically microliters), into the stream of mobile phase percolating through the column. The components of the sample move through the colum, each at a different velocity, which are a function of specific physical interactions with the adsorbent, the stationary phase. The velocity of each component depends on its chemical nature, on the nature of the stationary phase (inside the column) and on the composition of the mobile phase. The time at which a specific analyte elutes (emerges from the column) is called its retention time. The retention time, measured under particular conditions, is an identifying characteristic of a given analyte.

Many different types of columns are available, filled with adsorbents varying in particle size, porosity, and surface chemistry. The use of smaller particle size packing materials requires the use of higher operational pressure ("backpressure") and typically improves chromatographic resolution (the degree of peak separation between consecutive analytes emerging from the column). Sorbent particles may be ionic, hydrophobic or polar in nature.

The most common mode of liquid chromatography is Reversed Phase, whereby the mobile phases used, include any miscible combination of water or buffers with various organic solvents (the most common are acetonitrile and methanol). Some HPLC techniques use water-free mobile phases (see normal-phase chromatography below). The aqueous component of the mobile phase may contain acids (such as formic, phosphoric or trifluoroacetic acid) or salts to assist in the separation of the sample components. The composition of the mobile phase may be kept constant ("isocratic elution mode") or varied ("gradient elution mode") during the chromatographic analysis. Isocratic elution is typically effective in the separation of simple mixtures. Gradient elution is required for complex mixtures, with varying interactions with the stationary and mobile phases. This is the reason why in gradient elution the composition of the mobile phase is varied typically from low to high eluting strength. The eluting strength of the mobile phase is reflected by analyte retention times, as the high eluting strength speeds up the elution (resulting in shortening of retention times). For example, a typical gradient profile in reversed phase chromatography for might start at 5% acetonitrile (in water or aqueous buffer) and progress linearly to 95% acetonitrile over 5–25 minutes. Periods of constant mobile phase composition (plateau) may be also part of a gradient profile. For example, the mobile phase composition may be kept constant at 5% acetonitrile for 1–3 min, followed by a linear change up to 95% acetonitrile.

 
A rotary fraction collector collecting HPLC output. The system is being used to isolate a fraction containing Complex I from E. coli plasma membranes. About 50 litres of bacteria were needed to isolate this amount.[10]

The chosen composition of the mobile phase depends on the intensity of interactions between various sample components ("analytes") and stationary phase (e.g., hydrophobic interactions in reversed-phase HPLC). Depending on their affinity for the stationary and mobile phases, analytes partition between the two during the separation process taking place in the column. This partitioning process is similar to that which occurs during a liquid–liquid extraction but is continuous, not step-wise.

In the example using a water/acetonitrile gradient, the more hydrophobic components will elute (come off the column) later, then, once the mobile phase gets richer in acetonitrile (i.e., in a mobile phase becomes higher eluting solution), their elution speeds up.

The choice of mobile phase components, additives (such as salts or acids) and gradient conditions depends on the nature of the column and sample components. Often a series of trial runs is performed with the sample in order to find the HPLC method which gives adequate separation.

History and development edit

Prior to HPLC, scientists used benchtop column liquid chromatographic techniques. Liquid chromatographic systems were largely inefficient due to the flow rate of solvents being dependent on gravity. Separations took many hours, and sometimes days to complete. Gas chromatography (GC) at the time was more powerful than liquid chromatography (LC), however, it was obvious that gas phase separation and analysis of very polar high molecular weight biopolymers was impossible.[11] GC was ineffective for many life science and health applications for biomolecules, because of they are mostly non-volatile and thermally unstable at the high temperatures of GC.[12] As a result, alternative methods were hypothesized which would soon result in the development of HPLC.

Following on the seminal work of Martin and Synge in 1941, it was predicted by Calvin Giddings,[13] Josef Huber, and others in the 1960s that LC could be operated in the high-efficiency mode by reducing the packing-particle diameter substantially below the typical LC (and GC) level of 150 μm and using pressure to increase the mobile phase velocity.[11] These predictions underwent extensive experimentation and refinement throughout the 60s into the 70s until these very days.[14] Early developmental research began to improve LC particles, for example the historic Zipax, a superficially porous particle.[15]

The 1970s brought about many developments in hardware and instrumentation. Researchers began using pumps and injectors to make a rudimentary design of an HPLC system.[16] Gas amplifier pumps were ideal because they operated at constant pressure and did not require leak-free seals or check valves for steady flow and good quantitation.[12] Hardware milestones were made at Dupont IPD (Industrial Polymers Division) such as a low-dwell-volume gradient device being utilized as well as replacing the septum injector with a loop injection valve.[12]

While instrumentation developments were important, the history of HPLC is primarily about the history and evolution of particle technology.[12][17] After the introduction of porous layer particles, there has been a steady trend to reduced particle size to improve efficiency.[12] However, by decreasing particle size, new problems arose. The practical disadvantages stem from the excessive pressure drop needed to force mobile fluid through the column and the difficulty of preparing a uniform packing of extremely fine materials.[18] Every time particle size is reduced significantly, another round of instrument development usually must occur to handle the pressure.[14][12]

Types edit

Partition chromatography edit

 
HILIC Partition Technique Useful Range

Partition chromatography was one of the first kinds of chromatography that chemists developed, and is barely used these days.[19] The partition coefficient principle has been applied in paper chromatography, thin layer chromatography, gas phase and liquid–liquid separation applications. The 1952 Nobel Prize in chemistry was earned by Archer John Porter Martin and Richard Laurence Millington Synge for their development of the technique, which was used for their separation of amino acids.[20] Partition chromatography uses a retained solvent, on the surface or within the grains or fibers of an "inert" solid supporting matrix as with paper chromatography; or takes advantage of some coulombic and/or hydrogen donor interaction with the stationary phase. Analyte molecules partition between a liquid stationary phase and the eluent. Just as in Hydrophilic Interaction Chromatography (HILIC; a sub-technique within HPLC), this method separates analytes based on differences in their polarity. HILIC most often uses a bonded polar stationary phase and a mobile phase made primarily of acetonitrile with water as the strong component. Partition HPLC has been used historically on unbonded silica or alumina supports. Each works effectively for separating analytes by relative polar differences. HILIC bonded phases have the advantage of separating acidic, basic and neutral solutes in a single chromatographic run.[21]

The polar analytes diffuse into a stationary water layer associated with the polar stationary phase and are thus retained. The stronger the interactions between the polar analyte and the polar stationary phase (relative to the mobile phase) the longer the elution time. The interaction strength depends on the functional groups part of the analyte molecular structure, with more polarized groups (e.g., hydroxyl-) and groups capable of hydrogen bonding inducing more retention. Coulombic (electrostatic) interactions can also increase retention. Use of more polar solvents in the mobile phase will decrease the retention time of the analytes, whereas more hydrophobic solvents tend to increase retention times.

Normal–phase chromatography edit

Normal–phase chromatography was one of the first kinds of HPLC that chemists developed, but has decreased in use over the last decades. Also known as normal-phase HPLC (NP-HPLC) this method separates analytes based on their affinity for a polar stationary surface such as silica, hence it is based on analyte ability to engage in polar interactions (such as hydrogen-bonding or dipole-dipole type of interactions) with the sorbent surface. NP-HPLC uses a non-polar, non-aqueous mobile phase (e.g., Chloroform), and works effectively for separating analytes readily soluble in non-polar solvents. The analyte associates with and is retained by the polar stationary phase. Adsorption strengths increase with increased analyte polarity. The interaction strength depends not only on the functional groups present in the structure of the analyte molecule, but also on steric factors. The effect of steric hindrance on interaction strength allows this method to resolve (separate) structural isomers.

The use of more polar solvents in the mobile phase will decrease the retention time of analytes, whereas more hydrophobic solvents tend to induce slower elution (increased retention times). Very polar solvents such as traces of water in the mobile phase tend to adsorb to the solid surface of the stationary phase forming a stationary bound (water) layer which is considered to play an active role in retention. This behavior is somewhat peculiar to normal phase chromatography because it is governed almost exclusively by an adsorptive mechanism (i.e., analytes interact with a solid surface rather than with the solvated layer of a ligand attached to the sorbent surface; see also reversed-phase HPLC below). Adsorption chromatography is still somewhat used for structural isomer separations in both column and thin-layer chromatography formats on activated (dried) silica or alumina supports.

Partition- and NP-HPLC fell out of favor in the 1970s with the development of reversed-phase HPLC because of poor reproducibility of retention times due to the presence of a water or protic organic solvent layer on the surface of the silica or alumina chromatographic media. This layer changes with any changes in the composition of the mobile phase (e.g., moisture level) causing drifting retention times.

Recently, partition chromatography has become popular again with the development of Hilic bonded phases which demonstrate improved reproducibility, and due to a better understanding of the range of usefulness of the technique.

Displacement chromatography edit

The use of displacement chromatography is rather limited, and is mostly used for preparative chromatography. The basic principle is based on a molecule with a high affinity for the chromatography matrix (the displacer) which is used to compete effectively for binding sites, and thus displace all molecules with lesser affinities.[22] There are distinct differences between displacement and elution chromatography. In elution mode, substances typically emerge from a column in narrow, Gaussian peaks. Wide separation of peaks, preferably to baseline, is desired in order to achieve maximum purification. The speed at which any component of a mixture travels down the column in elution mode depends on many factors. But for two substances to travel at different speeds, and thereby be resolved, there must be substantial differences in some interaction between the biomolecules and the chromatography matrix. Operating parameters are adjusted to maximize the effect of this difference. In many cases, baseline separation of the peaks can be achieved only with gradient elution and low column loadings. Thus, two drawbacks to elution mode chromatography, especially at the preparative scale, are operational complexity, due to gradient solvent pumping, and low throughput, due to low column loadings. Displacement chromatography has advantages over elution chromatography in that components are resolved into consecutive zones of pure substances rather than "peaks". Because the process takes advantage of the nonlinearity of the isotherms, a larger column feed can be separated on a given column with the purified components recovered at significantly higher concentration.

Reversed-phase Liquid chromatography (RP-LC) edit

 
A chromatogram of complex mixture (perfume water) obtained by reversed phase HPLC

Reversed phase HPLC (RP-HPLC)[23] is the most widespread mode of chromatography. It has a non-polar stationary phase and an aqueous, moderately polar mobile phase. In the Reversed Phase methods, the substances are retained in the system the more hydrophobic they are.  For the retention of organic materials, the stationary phases, packed inside the columns, are consisted mainly of porous granules of silica gel in various shapes, mainly spherical, at different  diameters (1.5, 2, 3, 5, 7, 10 um), with varying pore diameters (60, 100, 150, 300, A), on whose surface are chemically bound various hydrocarbon ligands such as C3, C4, C8,C18  There are also polymeric hydrophobic particles that serve as stationary phases, when solutions at extreme pH are needed, or hybrid silica, polymerized with organic substances.  The longer the hydrocarbon ligand on the stationary phase, the longer the sample components can be retained. Most of the current methods of separation of biomedical materials use C-18 type of columns, sometimes called by a trade names such as ODS (octadecylsilane) or RP-18 (Reversed Phase 18).

The most common RP stationary phases are based on a silica support, which is surface-modified by bonding RMe2SiCl, where R is a straight chain alkyl group such as C18H37 or C8H17.

With such stationary phases, retention time is longer for lipophylic molecules, whereas polar molecules elute more readily (emerge early in the analysis). A chromatographer can increase retention times by adding more water to the mobile phase; thereby making the interactions of the hydrophobic analyte with the hydrophobic stationary phase relatively stronger. Similarly, an investigator can decrease retention time by adding more organic solvent to the mobile phase. RP-HPLC is so commonly used among the biologists and life science users, therefore it is often incorrectly referred to as just "HPLC" without further specification. The pharmaceutical industry also regularly employs RP-HPLC to qualify drugs before their release.

RP-HPLC operates on the principle of hydrophobic interactions, which originates from the high symmetry in the dipolar water structure and plays the most important role in all processes in life science. RP-HPLC allows the measurement of these interactive forces. The binding of the analyte to the stationary phase is proportional to the contact surface area around the non-polar segment of the analyte molecule upon association with the ligand on the stationary phase. This solvophobic effect is dominated by the force of water for "cavity-reduction" around the analyte and the C18-chain versus the complex of both. The energy released in this process is proportional to the surface tension of the eluent (water: 7.3×10−6 J/cm2, methanol: 2.2×10−6 J/cm2) and to the hydrophobic surface of the analyte and the ligand respectively. The retention can be decreased by adding a less polar solvent (methanol, acetonitrile) into the mobile phase to reduce the surface tension of water. Gradient elution uses this effect by automatically reducing the polarity and the surface tension of the aqueous mobile phase during the course of the analysis.

Structural properties of the analyte molecule can play an important role in its retention characteristics. In theory, an analyte with a larger hydrophobic surface area (C–H, C–C, and generally non-polar atomic bonds, such as S-S and others) can be retained longer as it does not interact with the water structure. On the other hand, analytes with higher polar surface area (as a result of the presence of polar groups, such as -OH, -NH2, COO or -NH3+ in their structure) are less retained, as they are better integrated into water. The interactions with the stationary phase can also affected by steric effects, or exclusion effects, whereby a component of very large molecule may have only restricted access to the pores of the stationary phase, where the interactions with surface ligands (alkyl chains) take place. Such surface hindrance typically results in less retention.

Retention time increases with more hydrophobic (non-polar) surface area of the molecules. For example, branched chain compounds can elute more rapidly than their corresponding linear isomers because their overall surface area is lower. Similarly organic compounds with single C–C bonds frequently elute later than those with a C=C or even triple bond, as the double or triple bond makes the molecule more compact than a single C–C bond.

Another important factor is the mobile phase pH since it can change the hydrophobic character of the ionizable analyte. For this reason most methods use a buffering agent, such as sodium phosphate, to control the pH. Buffers serve multiple purposes: control of pH which affects the ionization state of the ionizable analytes, affect the charge upon the ionizable silica surface of the stationary phase in between the bonded phase linands, and in some cases even act as ion pairing agents to neutralize analyte charge. Ammonium formate is commonly added in mass spectrometry to improve detection of certain analytes by the formation of analyte-ammonium adducts. A volatile organic acid such as acetic acid, or most commonly formic acid, is often added to the mobile phase if mass spectrometry is used to analyze the column effluents.

Trifluoroacetic acid as additive to the mobile phase is widely used for complex mixtures of biomedical samples, mostly peptides and proteins, using mostly a UV based detectors. They are used rarely used in mass spectrometry methods, due to its residues it can leave in the detector and solvent delivery system, which interfere with the analysis and detection. However it can be highly effective in improving retention of analytes such as carboxylic acids, in applications utilizing other detectors such as UV-VIS, as it is a fairly strong organic acid. The effects of acids and buffers vary by application but generally improve chromatographic resolution when dealing with ionizable components.

Reversed phase columns are quite difficult to damage compared to normal silica columns, thanks to the shielding effect of the bonded hydrophobic ligands; however, most reversed phase columns consist of alkyl derivatized silica particles, and are prone to hydrolysis of the silica at extreme pH conditions in the mobile phase. Most types of RP columns should not be used with aqueous bases as these will hydrolyze the underlying silica particle and dissolve it. There are selected brands of hybrid or enforced silica based particles of RP columns which can be used at extreme pH conditions. The use of extreme acidic conditions is also not recommended, as they also might hydrolyzed as well as corrode the inside walls of the metallic parts of the HPLC equipment.

As a rule, in most cases RP-HPLC columns should be flushed with clean solvent after use to remove residual acids or buffers, and stored in an appropriate composition of solvent. Some biomedical applications require non metallic environment for the optimal separation. For such sensitive cases there is a test for the metal content of a column is to inject a sample which is a mixture of 2,2'- and 4,4'-bipyridine. Because the 2,2'-bipy can chelate the metal, the shape of the peak for the 2,2'-bipy will be distorted (tailed) when metal ions are present on the surface of the silica.[citation needed]..

Size-exclusion chromatography edit

Size-exclusion chromatography (SEC)[24] separates polymer molecules and biomolecules based on differences in their molecular size (actually by a particle's Stokes radius). The separation process is based on the ability of sample molecules to permeate through the pores of gel spheres, packed inside the column, and is dependent on the relative size of analyte molecules and the respective pore size of the absorbent. The process also relies on the absence of any interactions with the packing material surface.

Two types of SEC are usually termed:

1. Gel permeation chromatography (GPC)—separation of synthetic polymers (aqueous or organic soluble). GPC is a powerful technique for polymer characterization using primarily organic solvents.

2. Gel filtration chromatography (GFC)—separation of water-soluble biopolymers. GFC uses primarily aqueous solvents (typically for aqueous soluble biopolymers, such as proteins, etc.).

The separation principle in SEC is based on the fully, or partially penetrating of the high molecular weight substances of the sample into the porous stationary-phase particles during their transport through column. The mobile-phase eluent is selected in such a way that it totally prevents interactions with the stationary phase's surface. Under these conditions, the smaller the size of the molecule, the more it is able to penetrate inside the pore space and the movement through the column takes longer. On the other hand, the bigger the molecular size, the higher the probability the molecule will not fully penetrate the pores of the stationary phase, and even travel around them, thus, will be eluted earlier. The molecules are separated in order of decreasing molecular weight, with the largest molecules eluting from the column first and smaller molecules eluting later. Molecules larger than the pore size do not enter the pores at all, and elute together as the first peak in the chromatogram and this is called total exclusion volume which defines the exclusion limit for a particular column. Small molecules will permeate fully through the pores of the stationary phase particles and will be eluted last, marking the end of the chromatogram, and may appear as a total penetration marker.

In biomedical sciences it is generally considered as a low resolution chromatography and thus it is often reserved for the final, "polishing" step of the purification. It is also useful for determining the tertiary structure and quaternary structure of purified proteins. SEC is used primarily for the analysis of large molecules such as proteins or polymers. SEC works also in a preparative way by trapping the smaller molecules in the pores of a particles. The larger molecules simply pass by the pores as they are too large to enter the pores. Larger molecules therefore flow through the column quicker than smaller molecules, that is, the smaller the molecule, the longer the retention time.

This technique is widely used for the molecular weight determination of polysaccharides. SEC is the official technique (suggested by European pharmacopeia) for the molecular weight comparison of different commercially available low-molecular weight heparins.[25]

Ion-exchange chromatography edit

Ion-exchange chromatography (IEC) or Ion chromatography (IC)[26] is an analytical technique for the separation and determination of ionic solutes in aqueous samples from environmental and industrial origins such as metal industry, industrial waste water, in biological systems, pharmaceutical samples, food, etc. Retention is based on the attraction between solute ions and charged sites bound to the stationary phase. Solute ions charged the same as the ions on the column are repulsed and elute without retention, while solute ions charged oppositely to the charged sites of the column are retained on it. Solute ions that are retained on the column can be eluted from it by changing the mobile phase composition, such as increasing its salt concentration and pH or increasing the column temperature, etc..

Types of ion exchangers include polystyrene resins, cellulose and dextran ion exchangers (gels), and controlled-pore glass or porous silica gel. Polystyrene resins allow cross linkage, which increases the stability of the chain. Higher cross linkage reduces swerving, which increases the equilibration time and ultimately improves selectivity. Cellulose and dextran ion exchangers possess larger pore sizes and low charge densities making them suitable for protein separation

In general, ion exchangers favor the binding of ions of higher charge and smaller radius.

An increase in counter ion (with respect to the functional groups in resins) concentration reduces the retention time, as it creates a strong competition with the solute ions. A decrease in pH reduces the retention time in cation exchange while an increase in pH reduces the retention time in anion exchange. By lowering the pH of the solvent in a cation exchange column, for instance, more hydrogen ions are available to compete for positions on the anionic stationary phase, thereby eluting weakly bound cations.

This form of chromatography is widely used in the following applications: water purification, preconcentration of trace components, ligand-exchange chromatography, ion-exchange chromatography of proteins, high-pH anion-exchange chromatography of carbohydrates and oligosaccharides, and others.

Bioaffinity chromatography or edit

High Performance Affinity Chromatography, HPAC,[27] works by passing a sample solution through a column packed with a stationary phase that contains an immobilized biologically active ligand. The ligand is in fact a substrate that has a specific binding affinity for the target molecule in the sample solution. The target molecule binds to the ligand, while the other molecules in the sample solution pass through the column, having little or no retention. The target molecule is then eluted from the column using a suitable elution buffer.

This chromatographic process relies on the capability of the bonded active substances to form stable, specific, and reversible complexes thanks to their biological recognition of certain specific sample components. The formation of these complexes involves the participation of common molecular forces such as the Van der Waals interaction, electrostatic interaction, dipole-dipole interaction, hydrophobic interaction, and the hydrogen bond. An efficient, biospecific bond is formed by a simultaneous and concerted action of several of these forces in the complementary binding sites.

Aqueous normal-phase chromatography edit

Aqueous normal-phase chromatography (ANP) is also called Hydrophilic Interaction Liquid Chromatography, HILIC.[28] This is a chromatographic technique which encompasses the mobile phase region between reversed-phase chromatography (RP) and organic normal phase chromatography (ONP). HILIC is used to achieve unique selectivity for hydrophilic compounds,[29] showing normal phase elution order, using "reversed-phase solvents", i.e., relatively polar mostly non-aqueous solvents in the mobile phase.[28] Many biological molecules, especially those found in biological fluids, are small polar compounds that do not retain well by Reversed Phase-HPLC. This has made hydrophilic interaction LC (HILIC) an attractive alternative and useful approach for analysis of polar molecules. Additionally, because HILIC is routinely used with traditional aqueous mixtures with polar organic solvents such as ACN and methanol, it can be easily coupled to MS.[30]

Isocratic and gradient elution edit

 
At the ARS Natural Products Utilization Research Unit in Oxford, MS., a support scientist (r) extracts plant pigments that will be analyzed by a plant physiologist (l) using an HPLC system.

A separation in which the mobile phase composition remains constant throughout the procedure is termed isocratic (meaning constant composition). The word was coined by Csaba Horvath who was one of the pioneers of HPLC.[31][32]

The mobile phase composition does not have to remain constant. A separation in which the mobile phase composition is changed during the separation process is described as a gradient elution.[33][34] For example, a gradient can start at 10% methanol in water, and end at 90% methanol in water after 20 minutes. The two components of the mobile phase are typically termed "A" and "B"; A is the "weak" solvent which allows the solute to elute only slowly, while B is the "strong" solvent which rapidly elutes the solutes from the column. In reversed-phase chromatography, solvent A is often water or an aqueous buffer, while B is an organic solvent miscible with water, such as acetonitrile, methanol, THF, or isopropanol.

In isocratic elution, peak width increases with retention time linearly according to the equation for N, the number of theoretical plates. This can be a major disadvantage when analyzing a sample that contains analytes with a wide range of retention factors. Using a weaker mobile phase, the runtime is lengthened and results in slowly eluting peaks to be broad, leading to reduced sensitivity. A stronger mobile phase would improve issues of runtime and broadening of later peaks but results in diminished peak separation, especially for quickly eluting analytes which may have insufficient time to fully resolve. This issue is addressed through the changing mobile phase composition of gradient elution.

 
A schematic of gradient elution. Increasing mobile phase strength sequentially elutes analytes having varying interaction strength with the stationary phase.

By starting from a weaker mobile phase and strengthening it during the runtime, gradient elution decreases the retention of the later-eluting components so that they elute faster, giving narrower (and taller) peaks for most components, while also allowing for the adequate separation of earlier-eluting components. This also improves the peak shape for tailed peaks, as the increasing concentration of the organic eluent pushes the tailing part of a peak forward. This also increases the peak height (the peak looks "sharper"), which is important in trace analysis. The gradient program may include sudden "step" increases in the percentage of the organic component, or different slopes at different times – all according to the desire for optimum separation in minimum time.

In isocratic elution, the retention order does not change if the column dimensions (length and inner diameter) change – that is, the peaks elute in the same order. In gradient elution, however, the elution order may change as the dimensions or flow rate change. if they are no scaled down or up according to the change[35]

The driving force in reversed phase chromatography originates in the high order of the water structure. The role of the organic component of the mobile phase is to reduce this high order and thus reduce the retarding strength of the aqueous component.

Parameters edit

The parameters are largely derived from two sets of chromatographic theory: plate theory (as part of Partition chromatography), and the rate theory of chromatography / Van Deemter equation. Of course, they can be put in practice through analysis of HPLC chromatograms, although rate theory is considered the more accurate theory.

They are analogous to the calculation of retention factor for a paper chromatography separation, but describes how well HPLC separates a mixture into two or more components that are detected as peaks (bands) on a chromatogram. The HPLC parameters are the: efficiency factor(N), the retention factor (kappa prime), and the separation factor (alpha). Together the factors are variables in a resolution equation, which describes how well two components' peaks separated or overlapped each other. These parameters are mostly only used for describing HPLC reversed phase and HPLC normal phase separations, since those separations tend to be more subtle than other HPLC modes (e.g., ion exchange and size exclusion).

Void volume is the amount of space in a column that is occupied by solvent. It is the space within the column that is outside of the column's internal packing material. Void volume is measured on a chromatogram as the first component peak detected, which is usually the solvent that was present in the sample mixture; ideally the sample solvent flows through the column without interacting with the column, but is still detectable as distinct from the HPLC solvent. The void volume is used as a correction factor.

Efficiency factor (N) practically measures how sharp component peaks on the chromatogram are, as ratio of the component peak's area ("retention time") relative to the width of the peaks at their widest point (at the baseline). Peaks that are tall, sharp, and relatively narrow indicate that separation method efficiently removed a component from a mixture; high efficiency. Efficiency is very dependent upon the HPLC column and the HPLC method used. Efficiency factor is synonymous with plate number, and the 'number of theoretical plates'.

Retention factor (kappa prime) measures how long a component of the mixture stuck to the column, measured by the area under the curve of its peak in a chromatogram (since HPLC chromatograms are a function of time). Each chromatogram peak will have its own retention factor (e.g., kappa1 for the retention factor of the first peak). This factor may be corrected for by the void volume of the column.

Separation factor (alpha) is a relative comparison on how well two neighboring components of the mixture were separated (i.e., two neighboring bands on a chromatogram). This factor is defined in terms of a ratio of the retention factors of a pair of neighboring chromatogram peaks, and may also be corrected for by the void volume of the column. The greater the separation factor value is over 1.0, the better the separation, until about 2.0 beyond which an HPLC method is probably not needed for separation. Resolution equations relate the three factors such that high efficiency and separation factors improve the resolution of component peaks in an HPLC separation.

Internal diameter edit

 
Tubing on a nano-liquid chromatography (nano-LC) system, used for very low flow capacities.

The internal diameter (ID) of an HPLC column is an important parameter[39].. It can influences the detection response, when reduced, thanks to the reduced lateral diffusion of the solute band. It can also affect the separation selectivity, when flow rate and injection volumes are not scaled down or up proportionally to the smaller or larger diameter used, both in the isocratic and in gradient modes.[40] It determines the quantity of analyte that can be loaded onto the column. Larger diameter columns are usually seen in preparative applications, such as the purification of a drug product for later use.[41] Low-ID columns have improved sensitivity and lower solvent consumption in the recent Ultra-High Performance Liquid Chromatography (UHPLC).[42]

Larger ID columns (over 10 mm) are used to purify usable amounts of material because of their large loading capacity.

Analytical scale columns (4.6 mm) have been the most common type of columns, though narrower columns[42] are rapidly gaining in popularity. They are used in traditional quantitative analysis of samples and often use a UV-Vis absorbance detector.

Narrow-bore columns (1–2 mm) are used for applications when more sensitivity is desired either with special UV-vis detectors, fluorescence detection or with other detection methods like liquid chromatography-mass spectrometry

Capillary columns (under 0.3 mm) are used almost exclusively with alternative detection means such as mass spectrometry. They are usually made from fused silica capillaries, rather than the stainless steel tubing that larger columns employ.

Particle size edit

Most traditional HPLC is performed with the stationary phase attached to the outside of small spherical silica particles (very small beads). These particles come in a variety of sizes with 5 µm beads being the most common. Smaller particles generally provide more surface area and better separations, but the pressure required for optimum linear velocity increases by the inverse of the particle diameter squared.[43][44][45]

According to the equations[46] of the column velocity, efficiency and backpressure, reducing the particle diameter by half and keeping the size of the column the same, will double the column velocity and efficiency; but four times increase the backpressure. And the small particles HPLC also can decrease the width broadening.[47] Larger particles are used in preparative HPLC (column diameters 5 cm up to >30 cm) and for non-HPLC applications such as solid-phase extraction.

Pore size edit

Many stationary phases are porous to provide greater surface area. Small pores provide greater surface area while larger pore size has better kinetics, especially for larger analytes. For example, a protein which is only slightly smaller than a pore might enter the pore but does not easily leave once inside.

Pump pressure edit

Pumps vary in pressure capacity, but their performance is measured on their ability to yield a consistent and reproducible volumetric flow rate. Pressure may reach as high as 60 MPa (6000 lbf/in2), or about 600 atmospheres. Modern HPLC systems have been improved to work at much higher pressures, and therefore are able to use much smaller particle sizes in the columns (<2 μm). These "ultra high performance liquid chromatography" systems or UHPLCs, which could also be known as ultra high pressure chromatography systems,[48] can work at up to 120 MPa (17,405 lbf/in2), or about 1200 atmospheres.[49] The term "UPLC"[50] is a trademark of the Waters Corporation, but is sometimes used to refer to the more general technique of UHPLC.

Detectors edit

HPLC detectors fall into two main categories: universal or selective. Universal detectors typically measure a bulk property (e.g., refractive index) by measuring a difference of a physical property between the mobile phase and mobile phase with solute while selective detectors measure a solute property (e.g., UV-Vis absorbance) by simply responding to the physical or chemical property of the solute.[51] HPLC most commonly uses a UV-Vis absorbance detector, however, a wide range of other chromatography detectors can be used. A universal detector that complements UV-Vis absorbance detection is the Charged aerosol detector (CAD). A kind of commonly utilized detector includes refractive index detectors, which provide readings by measuring the changes in the refractive index of the eluant as it moves through the flow cell. In certain cases, it is possible to use multiple detectors, for example LCMS normally combines UV-Vis with a mass spectrometer.

When used with an electrochemical detector (ECD) the HPLC-ECD selectively detects neurotransmitters such as: norepinephrine, dopamine, serotonin, glutamate, GABA, acetylcholine and others in neurochemical analysis research applications.[52] The HPLC-ECD detects neurotransmitters to the femtomolar range. Other methods to detect neurotransmitters include liquid chromatography-mass spectrometry, ELISA, or radioimmunoassays.

Autosamplers edit

Large numbers of samples can be automatically injected onto an HPLC system, by the use of HPLC autosamplers. In addition, HPLC autosamplers have an injection volume and technique which is exactly the same for each injection, consequently they provide a high degree of injection volume precision. It is possible to enable sample stirring within the sampling-chamber, thus promoting homogeneity.[53]

Applications edit

Manufacturing edit

HPLC has many applications in both laboratory and clinical science. It is a common technique used in pharmaceutical development, as it is a dependable way to obtain and ensure product purity.[54] While HPLC can produce extremely high quality (pure) products, it is not always the primary method used in the production of bulk drug materials.[55] According to the European pharmacopoeia, HPLC is used in only 15.5% of syntheses.[56] However, it plays a role in 44% of syntheses in the United States pharmacopoeia.[57] This could possibly be due to differences in monetary and time constraints, as HPLC on a large scale can be an expensive technique. An increase in specificity, precision, and accuracy that occurs with HPLC unfortunately corresponds to an increase in cost.

Legal edit

This technique is also used for detection of illicit drugs in various samples.[58] The most common method of drug detection has been an immunoassay.[59] This method is much more convenient. However, convenience comes at the cost of specificity and coverage of a wide range of drugs, therefore, HPLC as been used as well as an alternative method. As HPLC is a method of determining (and possibly increasing) purity, using HPLC alone in evaluating concentrations of drugs was somewhat insufficient. Therefore, HPLC in this context is often performed in conjunction with mass spectrometry.[60] Using liquid chromatography-mass spectrometry (LC-MS) instead of gas chromatography-mass spectrometry (GC-MS) circumvents the necessity for derivitizing with acetylating or alkylation agents, which can be a burdensome extra step.[61] LC-MS has been used to detect a variety of agents like doping agents, drug metabolites, glucuronide conjugates, amphetamines, opioids, cocaine, BZDs, ketamine, LSD, cannabis, and pesticides.[62][63] Performing HPLC in conjunction with mass spectrometry reduces the absolute need for standardizing HPLC experimental runs.

Research edit

Similar assays can be performed for research purposes, detecting concentrations of potential clinical candidates like anti-fungal and asthma drugs.[64] This technique is obviously useful in observing multiple species in collected samples, as well, but requires the use of standard solutions when information about species identity is sought out. It is used as a method to confirm results of synthesis reactions, as purity is essential in this type of research. However, mass spectrometry is still the more reliable way to identify species.

Medical and Health Sciences edit

Medical use of HPLC typically use Mass Spectrometer (MS) as the detector, so the technique is called LC-MS[65] or LC-MS/MS for tandem MS, where two types of MS are operated sequentially.[66] When HPLC instrument is connected to more than one detector, it is called Hyphenated LC system.[citation needed] Pharmaceutical applications[67] are the major users of HPLC, LC-MS and LC-MS/MS.[68] This include drug development[69] and pharmacology, which is the scientific study of the effects of drugs and chemicals on living organisms,[70] personalized medicine,[71] public health[72][73] and diagnostics.[74] While urine is the most common medium for analyzing drug concentrations, blood serum is the sample collected for most medical analyses with HPLC.[75] One of the most important roles of LC-MS and LC-MS/MS in the clinical lab is the Newborn Screening (NBS) for metabolic disorders[76] and follow-up diagnostics.[77][78] The infants' samples come in the shape of dried blood spot (DBS),[79] which is simple to prepare and transport, enabling safe and accessible diagnostics, both locally and globally.

Other methods of detection of molecules that are useful for clinical studies have been tested against HPLC, namely immunoassays. In one example of this, competitive protein binding assays (CPBA) and HPLC were compared for sensitivity in detection of vitamin D. Useful for diagnosing vitamin D deficiencies in children, it was found that sensitivity and specificity of this CPBA reached only 40% and 60%, respectively, of the capacity of HPLC.[80] While an expensive tool, the accuracy of HPLC is nearly unparalleled.

See also edit

References edit

  1. ^ Kazakevich, Yuri; LoBrutto, Rosario, eds. (2007). HPLC for pharmaceutical scientists. Hoboken, NJ: Wiley-Interscience. ISBN 978-0-471-68162-5.
  2. ^ Levin, Shulamit (January 2004). "Reversed Phase Stationary Phases in Pharmaceutical Sciences". Journal of Liquid Chromatography & Related Technologies. 27 (7–9): 1353–1376. doi:10.1081/JLC-120030606. ISSN 1082-6076. S2CID 97490509.
  3. ^ Gerber, F.; Krummen, M.; Potgeter, H.; Roth, A.; Siffrin, C.; Spoendlin, C. (2004). "Practical aspects of fast reversed-phase high-performance liquid chromatography using 3μm particle packed columns and monolithic columns in pharmaceutical development and production working under current good manufacturing practice". Journal of Chromatography A. 1036 (2): 127–133. doi:10.1016/j.chroma.2004.02.056. PMID 15146913.
  4. ^ Bayne, Shirley; Carlin, Michelle (2017). Forensic Applications of High Performance Liquid Chromatography (1st ed.). CRC Press. ISBN 9780429251962.
  5. ^ Seger, Christoph; Salzmann, Linda (2020-08-01). "After another decade: LC–MS/MS became routine in clinical diagnostics". Clinical Biochemistry. Advancement and Applications of Mass Spectrometry in Laboratory Medicine. 82: 2–11. doi:10.1016/j.clinbiochem.2020.03.004. ISSN 0009-9120. PMID 32188572. S2CID 213186669.
  6. ^ Dong, Michael (2018). "Ten Common-Sense Corollaries in Pharmaceutical Analysis by High Performance Liquid Chromatography". LCGC Europe. LCGC Europe-08-01-2018. 31 (8): 432–436.
  7. ^ Snyder, Lloyd R.; Kirkland, Joseph J.; Glajch, Joseph L. (2012). Practical HPLC Method Development (2nd ed.). John Wiley & Sons.
  8. ^ McMaster, Marvin C. (2007). HPLC: a practical user's guide (2nd ed.). Hoboken, NJ: Wiley-Interscience. ISBN 978-0-471-75401-5.
  9. ^ Hanai, Toshihiko; Hanai, T. (1999). HPLC: a practical guide. RSC chromatography monographs. Royal Society of Chemistry. Cambridge: Royal Society of Chemistry. ISBN 978-0-85404-515-0.
  10. ^ Morgan, David J. (2003-11-19). "Fraction collector (post on Flickr)". Flickr. Retrieved 28 October 2015.
  11. ^ a b Karger, Barry L. (1997). "HPLC: Early and Recent Perspectives". Journal of Chemical Education. 74 (1): 45. Bibcode:1997JChEd..74...45K. doi:10.1021/ed074p45.
  12. ^ a b c d e f Henry, Richard A. (1 February 2009) "The Early Days of HPLC at Dupont". Chromatography Online. Avanstar Communications Inc.
  13. ^ Giddings, Calvin (1965). Dynamics of Chromatography: Principles and Theory. Marcel Dekker.
  14. ^ a b Levin, Shulamit (2017). Grinberg, Nelu; Carr, Peter W. (eds.). Solid-Core or Fully Porous Columns in Ultra High-Performance Liquid Chromatography—Which Way to Go for Better Efficiency of the Separation?. Advances in Chromatography. Vol. 55 (1 ed.). Boca Raton: CRC Press. pp. 185–203. ISBN 9781315158075.
  15. ^ Iler, R.K. (1979) The Chemistry of Silica. John Wiley & Sons. New York.
  16. ^ Karger, B. L.; Berry, L. V. (1971). "Rapid liquid-chromatographic separation of steroids on columns heavily loaded with stationary phase". Clin. Chem. 17 (8): 757–64. doi:10.1093/clinchem/17.8.757. PMID 4254537.
  17. ^ Neue, Uwe D. (1997). HPLC columns: theory, technology, and practice. New York, NY: Wiley VCH. ISBN 978-0-471-19037-0.
  18. ^ Giddings, J. Calvin (1965) Dynamics of Chromatography, Part I. Principles and Theory. Marcel Dekker, Inc., New York. p. 281.
  19. ^ Ettre, C. (2001). "Milestones in Chromatography: The Birth of Partition Chromatography" (PDF). LCGC. 19 (5): 506–512. Retrieved 2016-02-26.
  20. ^ Martin, A J P; Synge, R L M (1941). "Separation of the higher monoamino-acids by counter-current liquid-liquid extraction: the amino-acid composition of wool". Biochemical Journal. 35 (1–2): 91–121. doi:10.1042/bj0350091. PMC 1265473. PMID 16747393.
  21. ^ Lindsay, S.; Kealey, D. (1987). High performance liquid chromatography. Wiley. OSTI 7013902. from review Hung, L. B.; Parcher, J. F.; Shores, J. C.; Ward, E. H. (1988). "Theoretical and experimental foundation for surface-coverage programming in gas–solid chromatography with an adsorbable carrier gas". J. Am. Chem. Soc. 110 (11): 1090–1096. doi:10.1021/ac00162a003.
  22. ^ Displacement Chromatography. Sacheminc.com. Retrieved 2011-06-07. September 15, 2008, at the Wayback Machine
  23. ^ LoBrutto, Rosario; Kazakevich, Yuri (2007-01-22), Kazakevich, Yuri; LoBrutto, Rosario (eds.), "Reversed‐Phase HPLC", HPLC for Pharmaceutical Scientists (1 ed.), Wiley, pp. 139–239, doi:10.1002/9780470087954.ch4, ISBN 978-0-471-68162-5, retrieved 2023-10-10
  24. ^ Kazakevich, Yuri; LoBrutto, Rosario (2007-01-22), Kazakevich, Yuri; LoBrutto, Rosario (eds.), "Size‐Exclusion Chromatography", HPLC for Pharmaceutical Scientists (1 ed.), Wiley, pp. 263–279, doi:10.1002/9780470087954.ch6, ISBN 978-0-471-68162-5, retrieved 2023-10-10
  25. ^ Mulloy, Barbara; Heath, Alan; Shriver, Zachary; Jameison, Fabian; Al Hakim, Ali; Morris, Tina S.; Szajek, Anita Y. (2014-08-01). "USP compendial methods for analysis of heparin: chromatographic determination of molecular weight distributions for heparin sodium". Analytical and Bioanalytical Chemistry. 406 (20): 4815–4823. doi:10.1007/s00216-014-7940-3. hdl:1721.1/104914. ISSN 1618-2650. PMID 24958344. S2CID 492085.
  26. ^ Fritz, James S.; Gjerde, Douglas T. (2000-04-25). Ion Chromatography (1 ed.). Wiley. doi:10.1002/9783527613243. ISBN 978-3-527-29914-0.
  27. ^ Zhang, Chenhua; Rodriguez, Elliott; Bi, Cong; Zheng, Xiwei; Suresh, Doddavenkatana; Suh, Kyungah; Li, Zhao; Elsebaei, Fawzi; Hage, David S. (2018). "High performance affinity chromatography and related separation methods for the analysis of biological and pharmaceutical agents". Analyst. 143 (2): 374–391. Bibcode:2018Ana...143..374Z. doi:10.1039/C7AN01469D. ISSN 1364-5528. PMC 5768458. PMID 29200216.
  28. ^ a b McCalley, David V. (2017-11-10). "Understanding and manipulating the separation in hydrophilic interaction liquid chromatography". Journal of Chromatography. A. 1523: 49–71. doi:10.1016/j.chroma.2017.06.026. ISSN 1873-3778. PMID 28668366.
  29. ^ Buszewski, Bogusław; Noga, Sylwia (2012). "Hydrophilic interaction liquid chromatography (HILIC)—a powerful separation technique". Analytical and Bioanalytical Chemistry. 402 (1): 231–247. doi:10.1007/s00216-011-5308-5. ISSN 1618-2650. PMC 3249561. PMID 21879300.
  30. ^ Buszewski, Bogusław; Noga, Sylwia (2012). "Hydrophilic interaction liquid chromatography (HILIC)—a powerful separation technique". Analytical and Bioanalytical Chemistry. 402 (1): 231–247. doi:10.1007/s00216-011-5308-5. ISSN 1618-2650. PMC 3249561. PMID 21879300.
  31. ^ Schellinger, Adam P.; Carr, Peter W. (2006). "Isocratic and gradient elution chromatography: A comparison in terms of speed, retention reproducibility and quantitation". Journal of Chromatography A. 1109 (2): 253–266. doi:10.1016/j.chroma.2006.01.047.
  32. ^ Ettre, L. S.; Zlatkis, A., eds. (1979), "Csaba Horváth", Journal of Chromatography Library, 75 years of Chromatography a Historical Dialogue, Elsevier, vol. 17, pp. 151–158, doi:10.1016/s0301-4770(08)60645-4, ISBN 9780444417541, retrieved 2023-10-15
  33. ^ Snyder, Lloyd R.; Dolan, John W. (2006). High-Performance Gradient Elution: The Practical Application of the Linear-Solvent-Strength Model. Wiley Interscience. ISBN 978-0470055519.
  34. ^ Schellinger, Adam P.; Carr, Peter W. (2006). "Isocratic and gradient elution chromatography: A comparison in terms of speed, retention reproducibility and quantitation". Journal of Chromatography A. 19th International Symposium on MicroScale Bioseparations. 1109 (2): 253–266. doi:10.1016/j.chroma.2006.01.047. ISSN 0021-9673. PMID 16460742. S2CID 26072994.
  35. ^ Dolan, John W. (2014). "LC Method Scaling, Part II: Gradient Separations". LCGC North America. 32 (3): 188–193.
  36. ^ Martin, A. J. P.; Synge, R. L. M. (1941-12-01). "A new form of chromatogram employing two liquid phases". Biochemical Journal. 35 (12): 1358–1368. doi:10.1042/bj0351358. ISSN 0306-3283. PMC 1265645. PMID 16747422.
  37. ^ Grushka, Eli (1974-05-01). "Solute band spreading in liquid chromatography. Causes and importance". Analytical Chemistry. 46 (6): 510A–518a. doi:10.1021/ac60342a048. ISSN 0003-2700.
  38. ^ Wren, Stephen A. C. (2005-06-15). "Peak capacity in gradient ultra performance liquid chromatography (UPLC)". Journal of Pharmaceutical and Biomedical Analysis. 38 (2): 337–343. doi:10.1016/j.jpba.2004.12.028. ISSN 0731-7085.
  39. ^ Zelenyánszki, Dóra; Felinger, Attila (2020-10-01). "The Impact of Column Hardware on Efficiency in Liquid Chromatography (LC)". LCGC Europe. LCGC Europe-10-01-2020. 33 (10): 498–504.
  40. ^ Dolan, John (2014). "LC Method Scaling, Part II: Gradient Separations". LCGC North America. LCGC North America-03-01-2014. 32 (3): 188–193.
  41. ^ Jensen, Ole Elvang; Kidal, Steffen (2006-03-01). "Using Volumetric Flow to Scaleup Chromatographic Processes". BioPharm International. BioPharm International-03-01-2006. 19 (3).
  42. ^ a b Walter, Thomas H.; Andrews, Richard W. (2014). "Recent innovations in UHPLC columns and instrumentation". Trends in Analytical Chemistry. 63: 14–20. doi:10.1016/j.trac.2014.07.016. ISSN 0165-9936.
  43. ^ Majors, Ronald E.. (2010-09-07) Fast and Ultrafast HPLC on sub-2 μm Porous Particles — Where Do We Go From Here? – LC-GC Europe. Lcgceurope.com. Retrieved 2011-06-07.
  44. ^ Xiang, Y.; Liu Y.; Lee M.L. (2006). "Ultrahigh pressure liquid chromatography using elevated temperature". Journal of Chromatography A. 1104 (1–2): 198–202. doi:10.1016/j.chroma.2005.11.118. PMID 16376355.
  45. ^ Horváth, Cs.; Preiss B.A.; Lipsky S.R. (1967). "Fast liquid chromatography. Investigation of operating parameters and the separation of nucleotides on pellicular ion exchangers". Analytical Chemistry. 39 (12): 1422–1428. doi:10.1021/ac60256a003. PMID 6073805.
  46. ^ Nguyen, Dao T.-T.; Guillarme, Davy; Rudaz, Serge; Veuthey, Jean-Luc (2006). "Fast analysis in liquid chromatography using small particle size and high pressure". Journal of Separation Science. 29 (12): 1836–1848. doi:10.1002/jssc.200600189. ISSN 1615-9306. PMID 16970187.
  47. ^ Gritti, Fabrice; Guiochon, Georges (2013). "The van Deemter equation: Assumptions, limits, and adjustment to modern high performance liquid chromatography". Journal of Chromatography A. 1302: 1–13. doi:10.1016/j.chroma.2013.06.032. PMID 23838304.
  48. ^ Xiang, Yanqiao; Liu, Yansheng; Lee, Milton L. (2006). "Ultrahigh pressure liquid chromatography using elevated temperature". Journal of Chromatography A. 1104 (1–2): 198–202. doi:10.1016/j.chroma.2005.11.118. PMID 16376355.
  49. ^ 1290 Infinity Quaternary Pump 2015-11-20 at the Wayback Machine. Agilent
  50. ^ waters. "Trademarks : Waters". www.waters.com.
  51. ^ K., Robards (1994). Principles and practice of modern chromatographic methods. Haddad, P. R., Jackson, P. E. Amsterdam: Elsevier/Academic Press. ISBN 9780080571782. OCLC 815471219.
  52. ^ "HPLC-ECD (Electrochemical Detection) Fundamentals".
  53. ^ Markovitch, Omer; Ottelé, Jim; Veldman, Obe; Otto, Sijbren (2020). "Automated device for continuous stirring while sampling in liquid chromatography systems". Communications Chemistry. 3 (1): 180. doi:10.1038/s42004-020-00427-5. PMC 9814086. PMID 36703458.
  54. ^ Gerber, Frederic (May 2004). "Practical aspects of fast reversed-phase high-performance liquid chromatography using 3 μm particle packed columns and monolithic columns in pharmaceutical development and production working under current good manufacturing practice". Journal of Chromatography. 1036 (2): 127–33. doi:10.1016/j.chroma.2004.02.056. PMID 15146913.
  55. ^ Siddiqui, Masoom Raza; AlOthman, Zeid A.; Rahman, Nafisur (2013). "Analytical techniques in pharmaceutical analysis: A review". Arabian Journal of Chemistry. 10: S1409–S1421. doi:10.1016/j.arabjc.2013.04.016.
  56. ^ The European Pharmacopoeia, 2002. fourth ed., Council of Europe, Strasbourg.
  57. ^ United States Pharmacopoeia, 2004. 27th ed. The USP Convention Inc., Rockville, MD.
  58. ^ Merone, Giuseppe M.; Tartaglia, Angela; Rossi, Sandra; Santavenere, Francesco; Bassotti, Elisa; D’Ovidio, Cristian; Bonelli, Martina; Rosato, Enrica; de Grazia, Ugo; Locatelli, Marcello; Savini, Fabio (2021). "Fast Quantitative LC-MS/MS Determination of Illicit Substances in Solid and Liquid Unknown Seized Samples". Analytical Chemistry. 93 (49): 16308–16313. doi:10.1021/acs.analchem.1c03310. ISSN 0003-2700. PMC 8674870. PMID 34843645.
  59. ^ Pesce, Amadeo; Rosenthal, Murray; West, Robert; West, Cameron; Crews, Bridgit; Mikel, Charles; Almazan, Perla; Latyshev, Sergey (2010-06-01). "An evaluation of the diagnostic accuracy of liquid chromatography-tandem mass spectrometry versus immunoassay drug testing in pain patients". Pain Physician. 13 (3): 273–281. PMID 20495592.
  60. ^ Tsai, I.-Lin; Weng, Te-I.; Tseng, Yufeng J.; Tan, Happy Kuy-Lok; Sun, Hsiao-Ju; Kuo, Ching-Hua (2013-12-01). "Screening and confirmation of 62 drugs of abuse and metabolites in urine by ultra-high-performance liquid chromatography-quadrupole time-of-flight mass spectrometry". Journal of Analytical Toxicology. 37 (9): 642–651. doi:10.1093/jat/bkt083. PMID 24084874.
  61. ^ Weinmann, W.; Renz, M.; Vogt, S.; Pollak, S. (2000-01-01). "Automated solid-phase extraction and two-step derivatisation for simultaneous analysis of basic illicit drugs in serum by GC/MS". International Journal of Legal Medicine. 113 (4): 229–235. doi:10.1007/s004149900098. PMID 10929239. S2CID 20451772.
  62. ^ Kolmonen, Marjo; Leinonen, Antti; Pelander, Anna; Ojanperä, Ilkka (2007-02-28). "A general screening method for doping agents in human urine by solid phase extraction and liquid chromatography/time-of-flight mass spectrometry". Analytica Chimica Acta. 585 (1): 94–102. doi:10.1016/j.aca.2006.12.028. PMID 17386652.
  63. ^ Pelander, Anna; Ojanperä, Ilkka; Laks, Suvi; Rasanen, Ilpo; Vuori, Erkki (2003-11-01). "Toxicological screening with formula-based metabolite identification by liquid chromatography/time-of-flight mass spectrometry". Analytical Chemistry. 75 (21): 5710–5718. doi:10.1021/ac030162o. PMID 14588010.
  64. ^ Nobilis, Milan; Pour, Milan; Senel, Petr; Pavlík, Jan; Kunes, Jirí; Voprsalová, Marie; Kolárová, Lenka; Holcapek, Michal (2007-06-15). "Metabolic profiling of a potential antifungal drug, 3-(4-bromophenyl)-5-acetoxymethyl-2,5-dihydrofuran-2-one, in mouse urine using high-performance liquid chromatography with UV photodiode-array and mass spectrometric detection". Journal of Chromatography B. 853 (1–2): 10–19. doi:10.1016/j.jchromb.2007.02.045. PMID 17400036.
  65. ^ Gu, Jatin; Patel, Kumar; Shah, Dhiren (2016). "APPLICATION OF LC-MS". PharmaTutor.
  66. ^ Tallam, Anil Kumar; Alapati, Sahithi; Nuli, Mohana Vamsi (2023). "A review on bioanalytical method development and validation of anticancer drugs by using lc/ms/ms and its applications on routine analysis". Journal of Integral Sciences: 4–19. doi:10.37022/jis.v6i1.51. ISSN 2581-5679. S2CID 257295079.
  67. ^ O’Driscoll, Aimee (2021). "HPLC in Pharmaceutical Applications". Lab Manager.
  68. ^ Beccaria, Marco; Cabooter, Deirdre (2020). "Current developments in LC-MS for pharmaceutical analysis". Analyst. 145 (4): 1129–1157. Bibcode:2020Ana...145.1129B. doi:10.1039/C9AN02145K. hdl:11392/2479221. ISSN 1364-5528. PMID 31971527. S2CID 210866236.
  69. ^ Gu, Chunang (Christine); Russell, David; Yehl, Peter (2016). "Application of LCMS in small-molecule drug development". European Pharmaceutical Review. 21 (4): 54–57.
  70. ^ D’Ovidio, Cristian; Locatelli, Marcello; Perrucci, Miryam; Ciriolo, Luigi; Furton, Kenneth G.; Gazioglu, Isil; Kabir, Abuzar; Merone, Giuseppe Maria; de Grazia, Ugo; Ali, Imran; Catena, Antonio Maria; Treglia, Michele; Marsella, Luigi T.; Savini, Fabio (2023). "LC-MS/MS Application in Pharmacotoxicological Field: Current State and New Applications". Molecules. 28 (5): 2127. doi:10.3390/molecules28052127. ISSN 1420-3049. PMC 10004468. PMID 36903374.
  71. ^ Zhou, Juntuo; Zhong, Lijun (2022). "Applications of liquid chromatography-mass spectrometry based metabolomics in predictive and personalized medicine". Frontiers in Molecular Biosciences. 9. doi:10.3389/fmolb.2022.1049016. ISSN 2296-889X. PMC 9669074. PMID 36406271.
  72. ^ Mathias, Patricia I.; Connor, Thomas H.; B’Hymer, Clayton (2017). "A review of high performance liquid chromatographic-mass spectrometric urinary methods for anticancer drug exposure of health care workers". Journal of Chromatography B. 1060: 316–324. doi:10.1016/j.jchromb.2017.06.028. ISSN 1570-0232. PMC 5585056. PMID 28654869.
  73. ^ Hernández, Félix; Sancho, Juan V.; Ibáñez, María; Guerrero, Carlos (2007). "Antibiotic residue determination in environmental waters by LC-MS". TrAC Trends in Analytical Chemistry. Pharmaceutical-residue analysis. 26 (6): 466–485. doi:10.1016/j.trac.2007.01.012. ISSN 0165-9936.
  74. ^ Seger, Christoph; Salzmann, Linda (2020). "After another decade: LC–MS/MS became routine in clinical diagnostics". Clinical Biochemistry. Advancement and Applications of Mass Spectrometry in Laboratory Medicine. 82: 2–11. doi:10.1016/j.clinbiochem.2020.03.004. ISSN 0009-9120. PMID 32188572. S2CID 213186669.
  75. ^ Sundström, Mira; Pelander, Anna; Angerer, Verena; Hutter, Melanie; Kneisel, Stefan; Ojanperä, Ilkka (2013-10-01). "A high-sensitivity ultra-high performance liquid chromatography/high-resolution time-of-flight mass spectrometry (UHPLC-HR-TOFMS) method for screening synthetic cannabinoids and other drugs of abuse in urine". Analytical and Bioanalytical Chemistry. 405 (26): 8463–8474. doi:10.1007/s00216-013-7272-8. PMID 23954996. S2CID 25743579.
  76. ^ Gelb, Michael H.; Basheeruddin, Khaja; Burlina, Alberto; Chen, Hsiao-Jan; Chien, Yin-Hsiu; Dizikes, George; Dorley, Christine; Giugliani, Roberto; Hietala, Amy; Hong, Xinying; Kao, Shu-Min; Khaledi, Hamid; Klug, Tracy; Kubaski, Francyne; Liao, Hsuan-Chieh (2022). "Liquid Chromatography–Tandem Mass Spectrometry in Newborn Screening Laboratories". International Journal of Neonatal Screening. 8 (4): 62. doi:10.3390/ijns8040062. ISSN 2409-515X. PMC 9781967. PMID 36547379.
  77. ^ Wang, Yanyun; Sun, Yun; Jiang, Tao (2019). "Clinical Application of LC–MS/MS in the Follow-Up for Treatment of Children with Methylmalonic Aciduria". Advances in Therapy. 36 (6): 1304–1313. doi:10.1007/s12325-019-00955-0. ISSN 1865-8652. PMID 31049874. S2CID 143432183.
  78. ^ Arunkumar, Nivethitha; Langan, Thomas J.; Stapleton, Molly; Kubaski, Francyne; Mason, Robert W.; Singh, Rajendra; Kobayashi, Hironori; Yamaguchi, Seiji; Suzuki, Yasuyuki; Orii, Kenji; Orii, Tadao; Fukao, Toshiyuki; Tomatsu, Shunji (2020). "Newborn screening of mucopolysaccharidoses: past, present, and future". Journal of Human Genetics. 65 (7): 557–567. doi:10.1038/s10038-020-0744-8. ISSN 1435-232X. PMID 32277174. S2CID 92042115.
  79. ^ Skogvold, Hanne Bendiksen; Rootwelt, Helge; Reubsaet, Léon; Elgstøen, Katja Benedikte Prestø; Wilson, Steven Ray (2023). "Dried blood spot analysis with liquid chromatography and mass spectrometry: Trends in clinical chemistry". Journal of Separation Science. 46 (15): e2300210. doi:10.1002/jssc.202300210. ISSN 1615-9306. PMID 37269205. S2CID 259047202.
  80. ^ Zahedi Rad, Maliheh; Neyestani, Tirang Reza; Nikooyeh, Bahareh; Shariatzadeh, Nastaran; Kalayi, Ali; Khalaji, Niloufar; Gharavi, Azam (2015-01-01). "Competitive Protein-binding assay-based Enzyme-immunoassay Method, Compared to High-pressure Liquid Chromatography, Has a Very Lower Diagnostic Value to Detect Vitamin D Deficiency in 9–12 Years Children". International Journal of Preventive Medicine. 6: 67. doi:10.4103/2008-7802.161069. PMC 4542329. PMID 26330983.

Further reading edit

  • L. R. Snyder, J.J. Kirkland, and J. W. Dolan, Introduction to Modern Liquid Chromatography, John Wiley & Sons, New York, 2009.
  • M.W. Dong, Modern HPLC for practicing scientists. Wiley, 2006.
  • L. R. Snyder, J.J. Kirkland, and J. L. Glajch, Practical HPLC Method Development, John Wiley & Sons, New York, 1997.
  • S. Ahuja and H. T. Rasmussen (ed), HPLC Method Development for Pharmaceuticals, Academic Press, 2007.
  • S. Ahuja and M.W. Dong (ed), Handbook of Pharmaceutical Analysis by HPLC, Elsevier/Academic Press, 2005.
  • Y. V. Kazakevich and R. LoBrutto (ed.), HPLC for Pharmaceutical Scientists, Wiley, 2007.
  • U. D. Neue, HPLC Columns: Theory, Technology, and Practice, Wiley-VCH, New York, 1997.
  • M. C. McMaster, HPLC, a practical user's guide, Wiley, 2007.

External links edit

  • HPLC Chromatography Principle, Application [Basic Note] – 2020. at Rxlalit.com[permanent dead link]
  • Liquid Chromatography at Curlie

high, performance, liquid, chromatography, hplc, formerly, referred, high, pressure, liquid, chromatography, technique, analytical, chemistry, used, separate, identify, quantify, specific, components, mixtures, mixtures, originate, from, food, chemicals, pharm. High performance liquid chromatography HPLC formerly referred to as high pressure liquid chromatography is a technique in analytical chemistry used to separate identify and quantify specific components in mixtures The mixtures can originate from food chemicals pharmaceuticals 1 biological environmental and agriculture etc which have been dissolved into liquid solutions A modern self contained HPLC Schematic representation of an HPLC unit 1 Solvent reservoirs 2 Solvent degasser 3 Gradient valve 4 Mixing vessel for delivery of the mobile phase 5 High pressure pump 6 Switching valve in inject position 6 Switching valve in load position 7 Sample injection loop 8 Pre column guard column 9 Analytical column 10 Detector i e IR UV 11 Data acquisition 12 Waste or fraction collector It relies on high pressure pumps which deliver mixtures of various solvents called the mobile phase which flows through the system collecting the sample mixture on the way delivering it into a cylinder called the column filled with solid particles made of adsorbent material called the stationary phase Each component in the sample interacts differently with the adsorbent material causing different migration rates for each component These different rates lead to separation as the species flow out of the column into a specific detector such as UV detectors The output of the detector is a graph called a chromatogram Chromatograms are graphical representations of the signal intensity versus time or volume showing peaks which represent components of the sample Each sample appears in its respective time called its retention time having area proportional to its amount HPLC is widely used for manufacturing e g during the production process of pharmaceutical and biological products 2 3 legal e g detecting performance enhancement drugs in urine 4 research e g separating the components of a complex biological sample or of similar synthetic chemicals from each other and medical e g detecting vitamin D levels in blood serum purposes 5 Chromatography can be described as a mass transfer process involving adsorption and or partition As mentioned HPLC relies on pumps to pass a pressurized liquid and a sample mixture through a column filled with adsorbent leading to the separation of the sample components The active component of the column the adsorbent is typically a granular material made of solid particles e g silica polymers etc 1 5 50 mm in size on which various reagents can be bonded The components of the sample mixture are separated from each other due to their different degrees of interaction with the adsorbent particles The pressurized liquid is typically a mixture of solvents e g water buffers acetonitrile and or methanol and is referred to as a mobile phase Its composition and temperature play a major role in the separation process by influencing the interactions taking place between sample components and adsorbent These interactions are physical in nature such as hydrophobic dispersive dipole dipole and ionic most often a combination Contents 1 Operation 2 History and development 3 Types 3 1 Partition chromatography 3 2 Normal phase chromatography 3 3 Displacement chromatography 3 4 Reversed phase Liquid chromatography RP LC 3 5 Size exclusion chromatography 3 6 Ion exchange chromatography 3 7 Bioaffinity chromatography or 3 8 Aqueous normal phase chromatography 4 Isocratic and gradient elution 5 Parameters 5 1 Theoretical 5 2 Internal diameter 5 3 Particle size 5 4 Pore size 5 5 Pump pressure 5 6 Detectors 5 7 Autosamplers 6 Applications 6 1 Manufacturing 6 2 Legal 6 3 Research 6 4 Medical and Health Sciences 7 See also 8 References 9 Further reading 10 External linksOperation editThe liquid chromatograph is complex 6 and has sophisticated and delicate technology In order to properly operate the system It is also desirable that there should be a minimum basis for understanding of how the device performs the data Processing to avoid incorrect data and distorted results 7 8 9 HPLC is distinguished from traditional low pressure liquid chromatography because operational pressures are significantly higher 50 1400 bar while ordinary liquid chromatography typically relies on the force of gravity to pass the mobile phase through the packed column Due to the small sample amount separated in analytical HPLC typical column dimensions are 2 1 4 6 mm diameter and 30 250 mm length Also HPLC columns are made with smaller adsorbent particles 1 5 50 mm in average particle size This gives HPLC superior resolving power the ability to distinguish between compounds when separating mixtures which makes it a popular chromatographic technique The schematic of an HPLC instrument typically includes solvents reservoirs one or more pumps a solvent degasser a sampler a column and a detector The solvents are prepared in advance according to the needs of the separation they pass through the degasser to remove dissolved gasses mixed to become the mobile phase then flow through the sampler which brings the sample mixture into the mobile phase stream which then carries it into the column The pumps deliver the desired flow and composition of the mobile phase through the stationary phase inside the column then directly into a flow cell inside the detector The detector generates a signal proportional to the amount of sample component emerging from the column hence allowing for quantitative analysis of the sample components The detector also marks the time of emergence the Retention Time which serves for initial identification of the component More advanced detectors provide also additional information specific to the analyte s characteristics such as UV VIS spectrum or Mass spectrum which can provide insight on its structural features These detectors are in common use such as UV Vis Photodiode array PDA diode array detector and mass spectrometry detector A digital microprocessor and user software control the HPLC instrument and provide data analysis Some models of mechanical pumps in an HPLC instrument can mix multiple solvents together at a ratios changing in time generating a composition gradient in the mobile phase Most HPLC instruments also have a column oven that allows for adjusting the temperature at which the separation is performed The sample mixture to be separated and analyzed is introduced in a discrete small volume typically microliters into the stream of mobile phase percolating through the column The components of the sample move through the colum each at a different velocity which are a function of specific physical interactions with the adsorbent the stationary phase The velocity of each component depends on its chemical nature on the nature of the stationary phase inside the column and on the composition of the mobile phase The time at which a specific analyte elutes emerges from the column is called its retention time The retention time measured under particular conditions is an identifying characteristic of a given analyte Many different types of columns are available filled with adsorbents varying in particle size porosity and surface chemistry The use of smaller particle size packing materials requires the use of higher operational pressure backpressure and typically improves chromatographic resolution the degree of peak separation between consecutive analytes emerging from the column Sorbent particles may be ionic hydrophobic or polar in nature The most common mode of liquid chromatography is Reversed Phase whereby the mobile phases used include any miscible combination of water or buffers with various organic solvents the most common are acetonitrile and methanol Some HPLC techniques use water free mobile phases see normal phase chromatography below The aqueous component of the mobile phase may contain acids such as formic phosphoric or trifluoroacetic acid or salts to assist in the separation of the sample components The composition of the mobile phase may be kept constant isocratic elution mode or varied gradient elution mode during the chromatographic analysis Isocratic elution is typically effective in the separation of simple mixtures Gradient elution is required for complex mixtures with varying interactions with the stationary and mobile phases This is the reason why in gradient elution the composition of the mobile phase is varied typically from low to high eluting strength The eluting strength of the mobile phase is reflected by analyte retention times as the high eluting strength speeds up the elution resulting in shortening of retention times For example a typical gradient profile in reversed phase chromatography for might start at 5 acetonitrile in water or aqueous buffer and progress linearly to 95 acetonitrile over 5 25 minutes Periods of constant mobile phase composition plateau may be also part of a gradient profile For example the mobile phase composition may be kept constant at 5 acetonitrile for 1 3 min followed by a linear change up to 95 acetonitrile nbsp A rotary fraction collector collecting HPLC output The system is being used to isolate a fraction containing Complex I from E coli plasma membranes About 50 litres of bacteria were needed to isolate this amount 10 The chosen composition of the mobile phase depends on the intensity of interactions between various sample components analytes and stationary phase e g hydrophobic interactions in reversed phase HPLC Depending on their affinity for the stationary and mobile phases analytes partition between the two during the separation process taking place in the column This partitioning process is similar to that which occurs during a liquid liquid extraction but is continuous not step wise In the example using a water acetonitrile gradient the more hydrophobic components will elute come off the column later then once the mobile phase gets richer in acetonitrile i e in a mobile phase becomes higher eluting solution their elution speeds up The choice of mobile phase components additives such as salts or acids and gradient conditions depends on the nature of the column and sample components Often a series of trial runs is performed with the sample in order to find the HPLC method which gives adequate separation History and development editPrior to HPLC scientists used benchtop column liquid chromatographic techniques Liquid chromatographic systems were largely inefficient due to the flow rate of solvents being dependent on gravity Separations took many hours and sometimes days to complete Gas chromatography GC at the time was more powerful than liquid chromatography LC however it was obvious that gas phase separation and analysis of very polar high molecular weight biopolymers was impossible 11 GC was ineffective for many life science and health applications for biomolecules because of they are mostly non volatile and thermally unstable at the high temperatures of GC 12 As a result alternative methods were hypothesized which would soon result in the development of HPLC Following on the seminal work of Martin and Synge in 1941 it was predicted by Calvin Giddings 13 Josef Huber and others in the 1960s that LC could be operated in the high efficiency mode by reducing the packing particle diameter substantially below the typical LC and GC level of 150 mm and using pressure to increase the mobile phase velocity 11 These predictions underwent extensive experimentation and refinement throughout the 60s into the 70s until these very days 14 Early developmental research began to improve LC particles for example the historic Zipax a superficially porous particle 15 The 1970s brought about many developments in hardware and instrumentation Researchers began using pumps and injectors to make a rudimentary design of an HPLC system 16 Gas amplifier pumps were ideal because they operated at constant pressure and did not require leak free seals or check valves for steady flow and good quantitation 12 Hardware milestones were made at Dupont IPD Industrial Polymers Division such as a low dwell volume gradient device being utilized as well as replacing the septum injector with a loop injection valve 12 While instrumentation developments were important the history of HPLC is primarily about the history and evolution of particle technology 12 17 After the introduction of porous layer particles there has been a steady trend to reduced particle size to improve efficiency 12 However by decreasing particle size new problems arose The practical disadvantages stem from the excessive pressure drop needed to force mobile fluid through the column and the difficulty of preparing a uniform packing of extremely fine materials 18 Every time particle size is reduced significantly another round of instrument development usually must occur to handle the pressure 14 12 Types editPartition chromatography edit nbsp HILIC Partition Technique Useful RangePartition chromatography was one of the first kinds of chromatography that chemists developed and is barely used these days 19 The partition coefficient principle has been applied in paper chromatography thin layer chromatography gas phase and liquid liquid separation applications The 1952 Nobel Prize in chemistry was earned by Archer John Porter Martin and Richard Laurence Millington Synge for their development of the technique which was used for their separation of amino acids 20 Partition chromatography uses a retained solvent on the surface or within the grains or fibers of an inert solid supporting matrix as with paper chromatography or takes advantage of some coulombic and or hydrogen donor interaction with the stationary phase Analyte molecules partition between a liquid stationary phase and the eluent Just as in Hydrophilic Interaction Chromatography HILIC a sub technique within HPLC this method separates analytes based on differences in their polarity HILIC most often uses a bonded polar stationary phase and a mobile phase made primarily of acetonitrile with water as the strong component Partition HPLC has been used historically on unbonded silica or alumina supports Each works effectively for separating analytes by relative polar differences HILIC bonded phases have the advantage of separating acidic basic and neutral solutes in a single chromatographic run 21 The polar analytes diffuse into a stationary water layer associated with the polar stationary phase and are thus retained The stronger the interactions between the polar analyte and the polar stationary phase relative to the mobile phase the longer the elution time The interaction strength depends on the functional groups part of the analyte molecular structure with more polarized groups e g hydroxyl and groups capable of hydrogen bonding inducing more retention Coulombic electrostatic interactions can also increase retention Use of more polar solvents in the mobile phase will decrease the retention time of the analytes whereas more hydrophobic solvents tend to increase retention times Normal phase chromatography edit Normal phase chromatography was one of the first kinds of HPLC that chemists developed but has decreased in use over the last decades Also known as normal phase HPLC NP HPLC this method separates analytes based on their affinity for a polar stationary surface such as silica hence it is based on analyte ability to engage in polar interactions such as hydrogen bonding or dipole dipole type of interactions with the sorbent surface NP HPLC uses a non polar non aqueous mobile phase e g Chloroform and works effectively for separating analytes readily soluble in non polar solvents The analyte associates with and is retained by the polar stationary phase Adsorption strengths increase with increased analyte polarity The interaction strength depends not only on the functional groups present in the structure of the analyte molecule but also on steric factors The effect of steric hindrance on interaction strength allows this method to resolve separate structural isomers The use of more polar solvents in the mobile phase will decrease the retention time of analytes whereas more hydrophobic solvents tend to induce slower elution increased retention times Very polar solvents such as traces of water in the mobile phase tend to adsorb to the solid surface of the stationary phase forming a stationary bound water layer which is considered to play an active role in retention This behavior is somewhat peculiar to normal phase chromatography because it is governed almost exclusively by an adsorptive mechanism i e analytes interact with a solid surface rather than with the solvated layer of a ligand attached to the sorbent surface see also reversed phase HPLC below Adsorption chromatography is still somewhat used for structural isomer separations in both column and thin layer chromatography formats on activated dried silica or alumina supports Partition and NP HPLC fell out of favor in the 1970s with the development of reversed phase HPLC because of poor reproducibility of retention times due to the presence of a water or protic organic solvent layer on the surface of the silica or alumina chromatographic media This layer changes with any changes in the composition of the mobile phase e g moisture level causing drifting retention times Recently partition chromatography has become popular again with the development of Hilic bonded phases which demonstrate improved reproducibility and due to a better understanding of the range of usefulness of the technique Displacement chromatography edit The use of displacement chromatography is rather limited and is mostly used for preparative chromatography The basic principle is based on a molecule with a high affinity for the chromatography matrix the displacer which is used to compete effectively for binding sites and thus displace all molecules with lesser affinities 22 There are distinct differences between displacement and elution chromatography In elution mode substances typically emerge from a column in narrow Gaussian peaks Wide separation of peaks preferably to baseline is desired in order to achieve maximum purification The speed at which any component of a mixture travels down the column in elution mode depends on many factors But for two substances to travel at different speeds and thereby be resolved there must be substantial differences in some interaction between the biomolecules and the chromatography matrix Operating parameters are adjusted to maximize the effect of this difference In many cases baseline separation of the peaks can be achieved only with gradient elution and low column loadings Thus two drawbacks to elution mode chromatography especially at the preparative scale are operational complexity due to gradient solvent pumping and low throughput due to low column loadings Displacement chromatography has advantages over elution chromatography in that components are resolved into consecutive zones of pure substances rather than peaks Because the process takes advantage of the nonlinearity of the isotherms a larger column feed can be separated on a given column with the purified components recovered at significantly higher concentration Reversed phase Liquid chromatography RP LC edit nbsp A chromatogram of complex mixture perfume water obtained by reversed phase HPLCFurther information Reversed phase chromatography Reversed phase HPLC RP HPLC 23 is the most widespread mode of chromatography It has a non polar stationary phase and an aqueous moderately polar mobile phase In the Reversed Phase methods the substances are retained in the system the more hydrophobic they are For the retention of organic materials the stationary phases packed inside the columns are consisted mainly of porous granules of silica gel in various shapes mainly spherical at different diameters 1 5 2 3 5 7 10 um with varying pore diameters 60 100 150 300 A on whose surface are chemically bound various hydrocarbon ligands such as C3 C4 C8 C18 There are also polymeric hydrophobic particles that serve as stationary phases when solutions at extreme pH are needed or hybrid silica polymerized with organic substances The longer the hydrocarbon ligand on the stationary phase the longer the sample components can be retained Most of the current methods of separation of biomedical materials use C 18 type of columns sometimes called by a trade names such as ODS octadecylsilane or RP 18 Reversed Phase 18 The most common RP stationary phases are based on a silica support which is surface modified by bonding RMe2SiCl where R is a straight chain alkyl group such as C18H37 or C8H17 With such stationary phases retention time is longer for lipophylic molecules whereas polar molecules elute more readily emerge early in the analysis A chromatographer can increase retention times by adding more water to the mobile phase thereby making the interactions of the hydrophobic analyte with the hydrophobic stationary phase relatively stronger Similarly an investigator can decrease retention time by adding more organic solvent to the mobile phase RP HPLC is so commonly used among the biologists and life science users therefore it is often incorrectly referred to as just HPLC without further specification The pharmaceutical industry also regularly employs RP HPLC to qualify drugs before their release RP HPLC operates on the principle of hydrophobic interactions which originates from the high symmetry in the dipolar water structure and plays the most important role in all processes in life science RP HPLC allows the measurement of these interactive forces The binding of the analyte to the stationary phase is proportional to the contact surface area around the non polar segment of the analyte molecule upon association with the ligand on the stationary phase This solvophobic effect is dominated by the force of water for cavity reduction around the analyte and the C18 chain versus the complex of both The energy released in this process is proportional to the surface tension of the eluent water 7 3 10 6 J cm2 methanol 2 2 10 6 J cm2 and to the hydrophobic surface of the analyte and the ligand respectively The retention can be decreased by adding a less polar solvent methanol acetonitrile into the mobile phase to reduce the surface tension of water Gradient elution uses this effect by automatically reducing the polarity and the surface tension of the aqueous mobile phase during the course of the analysis Structural properties of the analyte molecule can play an important role in its retention characteristics In theory an analyte with a larger hydrophobic surface area C H C C and generally non polar atomic bonds such as S S and others can be retained longer as it does not interact with the water structure On the other hand analytes with higher polar surface area as a result of the presence of polar groups such as OH NH2 COO or NH3 in their structure are less retained as they are better integrated into water The interactions with the stationary phase can also affected by steric effects or exclusion effects whereby a component of very large molecule may have only restricted access to the pores of the stationary phase where the interactions with surface ligands alkyl chains take place Such surface hindrance typically results in less retention Retention time increases with more hydrophobic non polar surface area of the molecules For example branched chain compounds can elute more rapidly than their corresponding linear isomers because their overall surface area is lower Similarly organic compounds with single C C bonds frequently elute later than those with a C C or even triple bond as the double or triple bond makes the molecule more compact than a single C C bond Another important factor is the mobile phase pH since it can change the hydrophobic character of the ionizable analyte For this reason most methods use a buffering agent such as sodium phosphate to control the pH Buffers serve multiple purposes control of pH which affects the ionization state of the ionizable analytes affect the charge upon the ionizable silica surface of the stationary phase in between the bonded phase linands and in some cases even act as ion pairing agents to neutralize analyte charge Ammonium formate is commonly added in mass spectrometry to improve detection of certain analytes by the formation of analyte ammonium adducts A volatile organic acid such as acetic acid or most commonly formic acid is often added to the mobile phase if mass spectrometry is used to analyze the column effluents Trifluoroacetic acid as additive to the mobile phase is widely used for complex mixtures of biomedical samples mostly peptides and proteins using mostly a UV based detectors They are used rarely used in mass spectrometry methods due to its residues it can leave in the detector and solvent delivery system which interfere with the analysis and detection However it can be highly effective in improving retention of analytes such as carboxylic acids in applications utilizing other detectors such as UV VIS as it is a fairly strong organic acid The effects of acids and buffers vary by application but generally improve chromatographic resolution when dealing with ionizable components Reversed phase columns are quite difficult to damage compared to normal silica columns thanks to the shielding effect of the bonded hydrophobic ligands however most reversed phase columns consist of alkyl derivatized silica particles and are prone to hydrolysis of the silica at extreme pH conditions in the mobile phase Most types of RP columns should not be used with aqueous bases as these will hydrolyze the underlying silica particle and dissolve it There are selected brands of hybrid or enforced silica based particles of RP columns which can be used at extreme pH conditions The use of extreme acidic conditions is also not recommended as they also might hydrolyzed as well as corrode the inside walls of the metallic parts of the HPLC equipment As a rule in most cases RP HPLC columns should be flushed with clean solvent after use to remove residual acids or buffers and stored in an appropriate composition of solvent Some biomedical applications require non metallic environment for the optimal separation For such sensitive cases there is a test for the metal content of a column is to inject a sample which is a mixture of 2 2 and 4 4 bipyridine Because the 2 2 bipy can chelate the metal the shape of the peak for the 2 2 bipy will be distorted tailed when metal ions are present on the surface of the silica citation needed Size exclusion chromatography edit Further information size exclusion chromatography Size exclusion chromatography SEC 24 separates polymer molecules and biomolecules based on differences in their molecular size actually by a particle s Stokes radius The separation process is based on the ability of sample molecules to permeate through the pores of gel spheres packed inside the column and is dependent on the relative size of analyte molecules and the respective pore size of the absorbent The process also relies on the absence of any interactions with the packing material surface Two types of SEC are usually termed 1 Gel permeation chromatography GPC separation of synthetic polymers aqueous or organic soluble GPC is a powerful technique for polymer characterization using primarily organic solvents 2 Gel filtration chromatography GFC separation of water soluble biopolymers GFC uses primarily aqueous solvents typically for aqueous soluble biopolymers such as proteins etc The separation principle in SEC is based on the fully or partially penetrating of the high molecular weight substances of the sample into the porous stationary phase particles during their transport through column The mobile phase eluent is selected in such a way that it totally prevents interactions with the stationary phase s surface Under these conditions the smaller the size of the molecule the more it is able to penetrate inside the pore space and the movement through the column takes longer On the other hand the bigger the molecular size the higher the probability the molecule will not fully penetrate the pores of the stationary phase and even travel around them thus will be eluted earlier The molecules are separated in order of decreasing molecular weight with the largest molecules eluting from the column first and smaller molecules eluting later Molecules larger than the pore size do not enter the pores at all and elute together as the first peak in the chromatogram and this is called total exclusion volume which defines the exclusion limit for a particular column Small molecules will permeate fully through the pores of the stationary phase particles and will be eluted last marking the end of the chromatogram and may appear as a total penetration marker In biomedical sciences it is generally considered as a low resolution chromatography and thus it is often reserved for the final polishing step of the purification It is also useful for determining the tertiary structure and quaternary structure of purified proteins SEC is used primarily for the analysis of large molecules such as proteins or polymers SEC works also in a preparative way by trapping the smaller molecules in the pores of a particles The larger molecules simply pass by the pores as they are too large to enter the pores Larger molecules therefore flow through the column quicker than smaller molecules that is the smaller the molecule the longer the retention time This technique is widely used for the molecular weight determination of polysaccharides SEC is the official technique suggested by European pharmacopeia for the molecular weight comparison of different commercially available low molecular weight heparins 25 Ion exchange chromatography edit Further information Ion exchange chromatography Ion exchange chromatography IEC or Ion chromatography IC 26 is an analytical technique for the separation and determination of ionic solutes in aqueous samples from environmental and industrial origins such as metal industry industrial waste water in biological systems pharmaceutical samples food etc Retention is based on the attraction between solute ions and charged sites bound to the stationary phase Solute ions charged the same as the ions on the column are repulsed and elute without retention while solute ions charged oppositely to the charged sites of the column are retained on it Solute ions that are retained on the column can be eluted from it by changing the mobile phase composition such as increasing its salt concentration and pH or increasing the column temperature etc Types of ion exchangers include polystyrene resins cellulose and dextran ion exchangers gels and controlled pore glass or porous silica gel Polystyrene resins allow cross linkage which increases the stability of the chain Higher cross linkage reduces swerving which increases the equilibration time and ultimately improves selectivity Cellulose and dextran ion exchangers possess larger pore sizes and low charge densities making them suitable for protein separationIn general ion exchangers favor the binding of ions of higher charge and smaller radius An increase in counter ion with respect to the functional groups in resins concentration reduces the retention time as it creates a strong competition with the solute ions A decrease in pH reduces the retention time in cation exchange while an increase in pH reduces the retention time in anion exchange By lowering the pH of the solvent in a cation exchange column for instance more hydrogen ions are available to compete for positions on the anionic stationary phase thereby eluting weakly bound cations This form of chromatography is widely used in the following applications water purification preconcentration of trace components ligand exchange chromatography ion exchange chromatography of proteins high pH anion exchange chromatography of carbohydrates and oligosaccharides and others Bioaffinity chromatography or edit Further information Affinity chromatography High Performance Affinity Chromatography HPAC 27 works by passing a sample solution through a column packed with a stationary phase that contains an immobilized biologically active ligand The ligand is in fact a substrate that has a specific binding affinity for the target molecule in the sample solution The target molecule binds to the ligand while the other molecules in the sample solution pass through the column having little or no retention The target molecule is then eluted from the column using a suitable elution buffer This chromatographic process relies on the capability of the bonded active substances to form stable specific and reversible complexes thanks to their biological recognition of certain specific sample components The formation of these complexes involves the participation of common molecular forces such as the Van der Waals interaction electrostatic interaction dipole dipole interaction hydrophobic interaction and the hydrogen bond An efficient biospecific bond is formed by a simultaneous and concerted action of several of these forces in the complementary binding sites Aqueous normal phase chromatography edit Aqueous normal phase chromatography ANP is also called Hydrophilic Interaction Liquid Chromatography HILIC 28 This is a chromatographic technique which encompasses the mobile phase region between reversed phase chromatography RP and organic normal phase chromatography ONP HILIC is used to achieve unique selectivity for hydrophilic compounds 29 showing normal phase elution order using reversed phase solvents i e relatively polar mostly non aqueous solvents in the mobile phase 28 Many biological molecules especially those found in biological fluids are small polar compounds that do not retain well by Reversed Phase HPLC This has made hydrophilic interaction LC HILIC an attractive alternative and useful approach for analysis of polar molecules Additionally because HILIC is routinely used with traditional aqueous mixtures with polar organic solvents such as ACN and methanol it can be easily coupled to MS 30 Isocratic and gradient elution edit nbsp At the ARS Natural Products Utilization Research Unit in Oxford MS a support scientist r extracts plant pigments that will be analyzed by a plant physiologist l using an HPLC system A separation in which the mobile phase composition remains constant throughout the procedure is termed isocratic meaning constant composition The word was coined by Csaba Horvath who was one of the pioneers of HPLC 31 32 The mobile phase composition does not have to remain constant A separation in which the mobile phase composition is changed during the separation process is described as a gradient elution 33 34 For example a gradient can start at 10 methanol in water and end at 90 methanol in water after 20 minutes The two components of the mobile phase are typically termed A and B A is the weak solvent which allows the solute to elute only slowly while B is the strong solvent which rapidly elutes the solutes from the column In reversed phase chromatography solvent A is often water or an aqueous buffer while B is an organic solvent miscible with water such as acetonitrile methanol THF or isopropanol In isocratic elution peak width increases with retention time linearly according to the equation for N the number of theoretical plates This can be a major disadvantage when analyzing a sample that contains analytes with a wide range of retention factors Using a weaker mobile phase the runtime is lengthened and results in slowly eluting peaks to be broad leading to reduced sensitivity A stronger mobile phase would improve issues of runtime and broadening of later peaks but results in diminished peak separation especially for quickly eluting analytes which may have insufficient time to fully resolve This issue is addressed through the changing mobile phase composition of gradient elution nbsp A schematic of gradient elution Increasing mobile phase strength sequentially elutes analytes having varying interaction strength with the stationary phase By starting from a weaker mobile phase and strengthening it during the runtime gradient elution decreases the retention of the later eluting components so that they elute faster giving narrower and taller peaks for most components while also allowing for the adequate separation of earlier eluting components This also improves the peak shape for tailed peaks as the increasing concentration of the organic eluent pushes the tailing part of a peak forward This also increases the peak height the peak looks sharper which is important in trace analysis The gradient program may include sudden step increases in the percentage of the organic component or different slopes at different times all according to the desire for optimum separation in minimum time In isocratic elution the retention order does not change if the column dimensions length and inner diameter change that is the peaks elute in the same order In gradient elution however the elution order may change as the dimensions or flow rate change if they are no scaled down or up according to the change 35 The driving force in reversed phase chromatography originates in the high order of the water structure The role of the organic component of the mobile phase is to reduce this high order and thus reduce the retarding strength of the aqueous component Parameters editAn editor has launched a copyright investigation involving this section The text under investigation is currently hidden from public view but is accessible in the page history Please do not restore or edit the blanked content until the issue is resolved by an administrator copyright clerk or volunteer response agent The purported copyright violation copies text from http repository tnmgrmu ac in 21269 1 260303221mohammed rizwan pdf Copyvios report as such this page has been listed on the copyright problems page Unless the copyright status of the text of this page or section is clarified and determined to be compatible with Wikipedia s content license the problematic text and revisions or the entire page may be deleted one week after the time of its listing i e after 13 16 31 October 2023 UTC What can I do to resolve the issue If you hold the copyright to this text you can license it in a manner that allows its use on Wikipedia You must permit the use of your material under the terms of the Creative Commons Attribution Sharealike 3 0 Unported License CC BY SA 3 0 and the GNU Free Documentation License GFDL unversioned with no invariant sections front cover texts or back cover texts Explain your intent to license the content on this article s discussion page To confirm your permission you can either display a notice to this effect at the site of original publication or send an e mail from an address associated with the original publication to permissions en nbsp wikimedia org or a postal letter to the Wikimedia Foundation These messages must explicitly permit use under CC BY SA and the GFDL See Wikipedia Donating copyrighted materials Note that articles on Wikipedia must be written from a neutral point of view and must be verifiable in published third party sources consider whether copyright issues aside your text is appropriate for inclusion in Wikipedia You can demonstrate that this text is in the public domain or is already under a license suitable for Wikipedia Explain this on this article s discussion page with reference to evidence Wikipedia Public domain and Wikipedia Compatibly licensed may assist in determining the status Otherwise you may rewrite this page without copyright infringing material Your rewrite should be placed on this page where it will be available for an administrator or clerk to review it at the end of the listing period Follow this link to create the temporary subpage Please mention the rewrite upon completion on this article s discussion page Simply modifying copyrighted text is not sufficient to avoid copyright infringement if the original copyright violation cannot be cleanly removed or the article reverted to a prior version it is best to write the article from scratch See Wikipedia Close paraphrasing For license compliance any content used from the original article must be properly attributed if you use content from the original please leave a note at the top of your rewrite saying as much You may duplicate non infringing text that you had contributed yourself It is always a good idea if rewriting to identify the point where the copyrighted content was imported to Wikipedia and to check to make sure that the contributor did not add content imported from other sources When closing investigations clerks and administrators may find other copyright problems than the one identified If this material is in the proposed rewrite and cannot be easily removed the rewrite may not be usable Steps to list an article at Wikipedia Copyright problems Add the following to the bottom of Wikipedia Copyright problems 2023 October 24 subst article cv High performance liquid chromatography from a rel nofollow class external free href http repository tnmgrmu ac in 21269 1 260303221mohammed rizwan pdf http repository tnmgrmu ac in 21269 1 260303221mohammed rizwan pdf a Add the following template to the talk page of the contributor of the material subst Nothanks web pg High performance liquid chromatography url a rel nofollow class external free href http repository tnmgrmu ac in 21269 1 260303221mohammed rizwan pdf http repository tnmgrmu ac in 21269 1 260303221mohammed rizwan pdf a Place copyvio bottom at the end of the portion you want to blank If nominating the entire page please place this template at the top of the page and set the fullpage parameter to yes Theoretical edit The theory of High performance liquid chromatography HPLC is basically the same as general chromatography theory which is defined as a separation technique for mixtures of compounds based on the differences in their distribution coefficient between two phases the stationary phase packed inside a column and the liquid mobile phase flowing through the columns by high pressure pumps Components whose distribution into the stationary phase is higher are retained longer and get separated from those with lower distribution into the stationary phase The theoretical and practical foundations of this method were laid by A J P Martin and R L M Synge 36 who received Nobel prize for it The theory of chromatography has been used as the basis for System Suitability tests as can be seen in the USP Phamacopaeia which are a set of quantitative criteria which test the suitability of the HPLC system to the required analysis at any step of it The distribution coefficient of a solute between the stationary and the mobile phase is proportional to the parameter called the Capacity Factor or Capacity Ratio k which is defined as shown in the Figure of the Chromatographic Performance Criteria Where Cs is the concentration of the solute at the stationary phase and Cm is its concentration at the mobile phase while F is the volume ratio of the stationary and mobile phase within the solute s zone he time elapsed between the injection time of the sample component into the column and its detection time is known as the Retention Time RT or tR The retention time is longer when the solute has higher distribution coefficient due to its chemical nature For example in reverse phase chromatography the more lipophilic is a compound the higher is its distribution coefficient between the stationary and the mobile phases so it is retained longer Therefore it is obvious that there is a direct relation between the distribution coefficient and the capacity factor k and the solute retention time This relation is also represented as a normalized unit less factor known as the Retention Factor or Retention parameter which is the experimental measurement of the capacity ratio as shown in the Figure of Performance Criteria as well tR is the retention time of the specific component and t0 is the time it takes for a non retained substance to elute through the system without any retention thus it is called the Void Time The ratio between the Retention Factors k of every two adjacent peaks in the chromatogram is used in the evaluation of the degree of separation between them and is called Selectivity Factor a as shown in the Performance Criteria graph The width of a peak relative to its retention time is a measure of the system s efficiency calculated as N Plate Count The migration through the column causes band broadening due to eddy diffusion lateral diffusion and molecular diffusion mass transfer kinetics between the two phases as well as extra column effects which reduce the efficiency of the chromatographic system 37 The narrower the peak the better is the separation as well as the limit of detection and limit of quantification of the chromatographic system because the detector response is higher for a specific area When the response is higher the signal to noise is better hence the system is capable of detecting lower concentrations At the same time the narrower is the peak for a given retention time the better can be the resolution between every two adjacent peaks The efficiency of the separation N the Plate Count is measured as shown in the Figure of Performance Criteria Where tR is the retention time and w baseline is measured from the baseline peak width calculated using lines tangential to the peak width at 50 height The Plate Count N as a criterion for system efficiency was developed for isocratic conditions i e a constant mobile phase composition throughout the run In gradient conditions where the mobile phase changes with time during the chromatographic run it is more appropriate to use the parameter Peak Capacity Pc as a measure for the system efficiency 38 The definition of Peak Capacity in chromatography is the number of peaks that can be separated within a retention window for a specific pre defined Resolution Factor usually 1 It could also be envisioned as the runtime measured in number of peaks average widths The equation is shown in the Figure of the performance criteria In this equation tg is the gradient time and w ave is the average peaks width at the base nbsp The quantitative parameters and equations which determine the extent of performance of the chromatographic systemThe chromatographic peak is assumed to have a Gaussian shape under ideal conditions describing normal distribution of the velocity of the molecules populating the peak zone migrating through the stationary phase inside the column Any deviation from the normal distribution indicates non ideal distribution during the migration process therefore might jeopardize the integrity of the peak s integration reducing the accuracy of the quantitation There are many experimental and chemical parameters that cause peak asymmetry such as non ideal connections or column s packing insufficient pH buffering solutes overloading etc The deviation from symmetry is measured by the Asymmetry Factor Af or Tailing Factor Tf Both equations are found in the Figure of the Performance Criteria The calculation of Asymmetry Factor Af is calculated from the lines A and B which are sections in the horizontal line parallel to the baseline drawn at 10 of the peak height as shown in the Figure The calculation of Tailing Factor Tf which is more widely used in the pharmaceutical industry as suggested by the pharmacopeias is also calculated from A and B the sections in the horizontal line parallel to the baseline but at 5 of the peak height The USP suggests that Tailing Factor should be in the range of 0 5 up to 2 to assure a precise and accurate quantitative measurement The parameters are largely derived from two sets of chromatographic theory plate theory as part of Partition chromatography and the rate theory of chromatography Van Deemter equation Of course they can be put in practice through analysis of HPLC chromatograms although rate theory is considered the more accurate theory They are analogous to the calculation of retention factor for a paper chromatography separation but describes how well HPLC separates a mixture into two or more components that are detected as peaks bands on a chromatogram The HPLC parameters are the efficiency factor N the retention factor kappa prime and the separation factor alpha Together the factors are variables in a resolution equation which describes how well two components peaks separated or overlapped each other These parameters are mostly only used for describing HPLC reversed phase and HPLC normal phase separations since those separations tend to be more subtle than other HPLC modes e g ion exchange and size exclusion Void volume is the amount of space in a column that is occupied by solvent It is the space within the column that is outside of the column s internal packing material Void volume is measured on a chromatogram as the first component peak detected which is usually the solvent that was present in the sample mixture ideally the sample solvent flows through the column without interacting with the column but is still detectable as distinct from the HPLC solvent The void volume is used as a correction factor Efficiency factor N practically measures how sharp component peaks on the chromatogram are as ratio of the component peak s area retention time relative to the width of the peaks at their widest point at the baseline Peaks that are tall sharp and relatively narrow indicate that separation method efficiently removed a component from a mixture high efficiency Efficiency is very dependent upon the HPLC column and the HPLC method used Efficiency factor is synonymous with plate number and the number of theoretical plates Retention factor kappa prime measures how long a component of the mixture stuck to the column measured by the area under the curve of its peak in a chromatogram since HPLC chromatograms are a function of time Each chromatogram peak will have its own retention factor e g kappa1 for the retention factor of the first peak This factor may be corrected for by the void volume of the column Separation factor alpha is a relative comparison on how well two neighboring components of the mixture were separated i e two neighboring bands on a chromatogram This factor is defined in terms of a ratio of the retention factors of a pair of neighboring chromatogram peaks and may also be corrected for by the void volume of the column The greater the separation factor value is over 1 0 the better the separation until about 2 0 beyond which an HPLC method is probably not needed for separation Resolution equations relate the three factors such that high efficiency and separation factors improve the resolution of component peaks in an HPLC separation Internal diameter edit nbsp Tubing on a nano liquid chromatography nano LC system used for very low flow capacities The internal diameter ID of an HPLC column is an important parameter 39 It can influences the detection response when reduced thanks to the reduced lateral diffusion of the solute band It can also affect the separation selectivity when flow rate and injection volumes are not scaled down or up proportionally to the smaller or larger diameter used both in the isocratic and in gradient modes 40 It determines the quantity of analyte that can be loaded onto the column Larger diameter columns are usually seen in preparative applications such as the purification of a drug product for later use 41 Low ID columns have improved sensitivity and lower solvent consumption in the recent Ultra High Performance Liquid Chromatography UHPLC 42 Larger ID columns over 10 mm are used to purify usable amounts of material because of their large loading capacity Analytical scale columns 4 6 mm have been the most common type of columns though narrower columns 42 are rapidly gaining in popularity They are used in traditional quantitative analysis of samples and often use a UV Vis absorbance detector Narrow bore columns 1 2 mm are used for applications when more sensitivity is desired either with special UV vis detectors fluorescence detection or with other detection methods like liquid chromatography mass spectrometryCapillary columns under 0 3 mm are used almost exclusively with alternative detection means such as mass spectrometry They are usually made from fused silica capillaries rather than the stainless steel tubing that larger columns employ Particle size edit Most traditional HPLC is performed with the stationary phase attached to the outside of small spherical silica particles very small beads These particles come in a variety of sizes with 5 µm beads being the most common Smaller particles generally provide more surface area and better separations but the pressure required for optimum linear velocity increases by the inverse of the particle diameter squared 43 44 45 According to the equations 46 of the column velocity efficiency and backpressure reducing the particle diameter by half and keeping the size of the column the same will double the column velocity and efficiency but four times increase the backpressure And the small particles HPLC also can decrease the width broadening 47 Larger particles are used in preparative HPLC column diameters 5 cm up to gt 30 cm and for non HPLC applications such as solid phase extraction Pore size edit Many stationary phases are porous to provide greater surface area Small pores provide greater surface area while larger pore size has better kinetics especially for larger analytes For example a protein which is only slightly smaller than a pore might enter the pore but does not easily leave once inside Pump pressure edit Pumps vary in pressure capacity but their performance is measured on their ability to yield a consistent and reproducible volumetric flow rate Pressure may reach as high as 60 MPa 6000 lbf in2 or about 600 atmospheres Modern HPLC systems have been improved to work at much higher pressures and therefore are able to use much smaller particle sizes in the columns lt 2 mm These ultra high performance liquid chromatography systems or UHPLCs which could also be known as ultra high pressure chromatography systems 48 can work at up to 120 MPa 17 405 lbf in2 or about 1200 atmospheres 49 The term UPLC 50 is a trademark of the Waters Corporation but is sometimes used to refer to the more general technique of UHPLC Detectors edit HPLC detectors fall into two main categories universal or selective Universal detectors typically measure a bulk property e g refractive index by measuring a difference of a physical property between the mobile phase and mobile phase with solute while selective detectors measure a solute property e g UV Vis absorbance by simply responding to the physical or chemical property of the solute 51 HPLC most commonly uses a UV Vis absorbance detector however a wide range of other chromatography detectors can be used A universal detector that complements UV Vis absorbance detection is the Charged aerosol detector CAD A kind of commonly utilized detector includes refractive index detectors which provide readings by measuring the changes in the refractive index of the eluant as it moves through the flow cell In certain cases it is possible to use multiple detectors for example LCMS normally combines UV Vis with a mass spectrometer When used with an electrochemical detector ECD the HPLC ECD selectively detects neurotransmitters such as norepinephrine dopamine serotonin glutamate GABA acetylcholine and others in neurochemical analysis research applications 52 The HPLC ECD detects neurotransmitters to the femtomolar range Other methods to detect neurotransmitters include liquid chromatography mass spectrometry ELISA or radioimmunoassays Autosamplers edit Large numbers of samples can be automatically injected onto an HPLC system by the use of HPLC autosamplers In addition HPLC autosamplers have an injection volume and technique which is exactly the same for each injection consequently they provide a high degree of injection volume precision It is possible to enable sample stirring within the sampling chamber thus promoting homogeneity 53 Applications editManufacturing edit HPLC has many applications in both laboratory and clinical science It is a common technique used in pharmaceutical development as it is a dependable way to obtain and ensure product purity 54 While HPLC can produce extremely high quality pure products it is not always the primary method used in the production of bulk drug materials 55 According to the European pharmacopoeia HPLC is used in only 15 5 of syntheses 56 However it plays a role in 44 of syntheses in the United States pharmacopoeia 57 This could possibly be due to differences in monetary and time constraints as HPLC on a large scale can be an expensive technique An increase in specificity precision and accuracy that occurs with HPLC unfortunately corresponds to an increase in cost Legal edit This technique is also used for detection of illicit drugs in various samples 58 The most common method of drug detection has been an immunoassay 59 This method is much more convenient However convenience comes at the cost of specificity and coverage of a wide range of drugs therefore HPLC as been used as well as an alternative method As HPLC is a method of determining and possibly increasing purity using HPLC alone in evaluating concentrations of drugs was somewhat insufficient Therefore HPLC in this context is often performed in conjunction with mass spectrometry 60 Using liquid chromatography mass spectrometry LC MS instead of gas chromatography mass spectrometry GC MS circumvents the necessity for derivitizing with acetylating or alkylation agents which can be a burdensome extra step 61 LC MS has been used to detect a variety of agents like doping agents drug metabolites glucuronide conjugates amphetamines opioids cocaine BZDs ketamine LSD cannabis and pesticides 62 63 Performing HPLC in conjunction with mass spectrometry reduces the absolute need for standardizing HPLC experimental runs Research edit Similar assays can be performed for research purposes detecting concentrations of potential clinical candidates like anti fungal and asthma drugs 64 This technique is obviously useful in observing multiple species in collected samples as well but requires the use of standard solutions when information about species identity is sought out It is used as a method to confirm results of synthesis reactions as purity is essential in this type of research However mass spectrometry is still the more reliable way to identify species Medical and Health Sciences edit Medical use of HPLC typically use Mass Spectrometer MS as the detector so the technique is called LC MS 65 or LC MS MS for tandem MS where two types of MS are operated sequentially 66 When HPLC instrument is connected to more than one detector it is called Hyphenated LC system citation needed Pharmaceutical applications 67 are the major users of HPLC LC MS and LC MS MS 68 This include drug development 69 and pharmacology which is the scientific study of the effects of drugs and chemicals on living organisms 70 personalized medicine 71 public health 72 73 and diagnostics 74 While urine is the most common medium for analyzing drug concentrations blood serum is the sample collected for most medical analyses with HPLC 75 One of the most important roles of LC MS and LC MS MS in the clinical lab is the Newborn Screening NBS for metabolic disorders 76 and follow up diagnostics 77 78 The infants samples come in the shape of dried blood spot DBS 79 which is simple to prepare and transport enabling safe and accessible diagnostics both locally and globally Other methods of detection of molecules that are useful for clinical studies have been tested against HPLC namely immunoassays In one example of this competitive protein binding assays CPBA and HPLC were compared for sensitivity in detection of vitamin D Useful for diagnosing vitamin D deficiencies in children it was found that sensitivity and specificity of this CPBA reached only 40 and 60 respectively of the capacity of HPLC 80 While an expensive tool the accuracy of HPLC is nearly unparalleled See also editHistory of chromatography Capillary electrochromatography Column chromatography Csaba Horvath Ion chromatography Micellar liquid chromatographyReferences edit Kazakevich Yuri LoBrutto Rosario eds 2007 HPLC for pharmaceutical scientists Hoboken NJ Wiley Interscience ISBN 978 0 471 68162 5 Levin Shulamit January 2004 Reversed Phase Stationary Phases in Pharmaceutical Sciences Journal of Liquid Chromatography amp Related Technologies 27 7 9 1353 1376 doi 10 1081 JLC 120030606 ISSN 1082 6076 S2CID 97490509 Gerber F Krummen M Potgeter H Roth A Siffrin C Spoendlin C 2004 Practical aspects of fast reversed phase high performance liquid chromatography using 3mm particle packed columns and monolithic columns in pharmaceutical development and production working under current good manufacturing practice Journal of Chromatography A 1036 2 127 133 doi 10 1016 j chroma 2004 02 056 PMID 15146913 Bayne Shirley Carlin Michelle 2017 Forensic Applications of High Performance Liquid Chromatography 1st ed CRC Press ISBN 9780429251962 Seger Christoph Salzmann Linda 2020 08 01 After another decade LC MS MS became routine in clinical diagnostics Clinical Biochemistry Advancement and Applications of Mass Spectrometry in Laboratory Medicine 82 2 11 doi 10 1016 j clinbiochem 2020 03 004 ISSN 0009 9120 PMID 32188572 S2CID 213186669 Dong Michael 2018 Ten Common Sense Corollaries in Pharmaceutical Analysis by High Performance Liquid Chromatography LCGC Europe LCGC Europe 08 01 2018 31 8 432 436 Snyder Lloyd R Kirkland Joseph J Glajch Joseph L 2012 Practical HPLC Method Development 2nd ed John Wiley amp Sons McMaster Marvin C 2007 HPLC a practical user s guide 2nd ed Hoboken NJ Wiley Interscience ISBN 978 0 471 75401 5 Hanai Toshihiko Hanai T 1999 HPLC a practical guide RSC chromatography monographs Royal Society of Chemistry Cambridge Royal Society of Chemistry ISBN 978 0 85404 515 0 Morgan David J 2003 11 19 Fraction collector post on Flickr Flickr Retrieved 28 October 2015 a b Karger Barry L 1997 HPLC Early and Recent Perspectives Journal of Chemical Education 74 1 45 Bibcode 1997JChEd 74 45K doi 10 1021 ed074p45 a b c d e f Henry Richard A 1 February 2009 The Early Days of HPLC at Dupont Chromatography Online Avanstar Communications Inc Giddings Calvin 1965 Dynamics of Chromatography Principles and Theory Marcel Dekker a b Levin Shulamit 2017 Grinberg Nelu Carr Peter W eds Solid Core or Fully Porous Columns in Ultra High Performance Liquid Chromatography Which Way to Go for Better Efficiency of the Separation Advances in Chromatography Vol 55 1 ed Boca Raton CRC Press pp 185 203 ISBN 9781315158075 Iler R K 1979 The Chemistry of Silica John Wiley amp Sons New York Karger B L Berry L V 1971 Rapid liquid chromatographic separation of steroids on columns heavily loaded with stationary phase Clin Chem 17 8 757 64 doi 10 1093 clinchem 17 8 757 PMID 4254537 Neue Uwe D 1997 HPLC columns theory technology and practice New York NY Wiley VCH ISBN 978 0 471 19037 0 Giddings J Calvin 1965 Dynamics of Chromatography Part I Principles and Theory Marcel Dekker Inc New York p 281 Ettre C 2001 Milestones in Chromatography The Birth of Partition Chromatography PDF LCGC 19 5 506 512 Retrieved 2016 02 26 Martin A J P Synge R L M 1941 Separation of the higher monoamino acids by counter current liquid liquid extraction the amino acid composition of wool Biochemical Journal 35 1 2 91 121 doi 10 1042 bj0350091 PMC 1265473 PMID 16747393 Lindsay S Kealey D 1987 High performance liquid chromatography Wiley OSTI 7013902 from review Hung L B Parcher J F Shores J C Ward E H 1988 Theoretical and experimental foundation for surface coverage programming in gas solid chromatography with an adsorbable carrier gas J Am Chem Soc 110 11 1090 1096 doi 10 1021 ac00162a003 Displacement Chromatography Sacheminc com Retrieved 2011 06 07 Archived September 15 2008 at the Wayback Machine LoBrutto Rosario Kazakevich Yuri 2007 01 22 Kazakevich Yuri LoBrutto Rosario eds Reversed Phase HPLC HPLC for Pharmaceutical Scientists 1 ed Wiley pp 139 239 doi 10 1002 9780470087954 ch4 ISBN 978 0 471 68162 5 retrieved 2023 10 10 Kazakevich Yuri LoBrutto Rosario 2007 01 22 Kazakevich Yuri LoBrutto Rosario eds Size Exclusion Chromatography HPLC for Pharmaceutical Scientists 1 ed Wiley pp 263 279 doi 10 1002 9780470087954 ch6 ISBN 978 0 471 68162 5 retrieved 2023 10 10 Mulloy Barbara Heath Alan Shriver Zachary Jameison Fabian Al Hakim Ali Morris Tina S Szajek Anita Y 2014 08 01 USP compendial methods for analysis of heparin chromatographic determination of molecular weight distributions for heparin sodium Analytical and Bioanalytical Chemistry 406 20 4815 4823 doi 10 1007 s00216 014 7940 3 hdl 1721 1 104914 ISSN 1618 2650 PMID 24958344 S2CID 492085 Fritz James S Gjerde Douglas T 2000 04 25 Ion Chromatography 1 ed Wiley doi 10 1002 9783527613243 ISBN 978 3 527 29914 0 Zhang Chenhua Rodriguez Elliott Bi Cong Zheng Xiwei Suresh Doddavenkatana Suh Kyungah Li Zhao Elsebaei Fawzi Hage David S 2018 High performance affinity chromatography and related separation methods for the analysis of biological and pharmaceutical agents Analyst 143 2 374 391 Bibcode 2018Ana 143 374Z doi 10 1039 C7AN01469D ISSN 1364 5528 PMC 5768458 PMID 29200216 a b McCalley David V 2017 11 10 Understanding and manipulating the separation in hydrophilic interaction liquid chromatography Journal of Chromatography A 1523 49 71 doi 10 1016 j chroma 2017 06 026 ISSN 1873 3778 PMID 28668366 Buszewski Boguslaw Noga Sylwia 2012 Hydrophilic interaction liquid chromatography HILIC a powerful separation technique Analytical and Bioanalytical Chemistry 402 1 231 247 doi 10 1007 s00216 011 5308 5 ISSN 1618 2650 PMC 3249561 PMID 21879300 Buszewski Boguslaw Noga Sylwia 2012 Hydrophilic interaction liquid chromatography HILIC a powerful separation technique Analytical and Bioanalytical Chemistry 402 1 231 247 doi 10 1007 s00216 011 5308 5 ISSN 1618 2650 PMC 3249561 PMID 21879300 Schellinger Adam P Carr Peter W 2006 Isocratic and gradient elution chromatography A comparison in terms of speed retention reproducibility and quantitation Journal of Chromatography A 1109 2 253 266 doi 10 1016 j chroma 2006 01 047 Ettre L S Zlatkis A eds 1979 Csaba Horvath Journal of Chromatography Library 75 years of Chromatography a Historical Dialogue Elsevier vol 17 pp 151 158 doi 10 1016 s0301 4770 08 60645 4 ISBN 9780444417541 retrieved 2023 10 15 Snyder Lloyd R Dolan John W 2006 High Performance Gradient Elution The Practical Application of the Linear Solvent Strength Model Wiley Interscience ISBN 978 0470055519 Schellinger Adam P Carr Peter W 2006 Isocratic and gradient elution chromatography A comparison in terms of speed retention reproducibility and quantitation Journal of Chromatography A 19th International Symposium on MicroScale Bioseparations 1109 2 253 266 doi 10 1016 j chroma 2006 01 047 ISSN 0021 9673 PMID 16460742 S2CID 26072994 Dolan John W 2014 LC Method Scaling Part II Gradient Separations LCGC North America 32 3 188 193 Martin A J P Synge R L M 1941 12 01 A new form of chromatogram employing two liquid phases Biochemical Journal 35 12 1358 1368 doi 10 1042 bj0351358 ISSN 0306 3283 PMC 1265645 PMID 16747422 Grushka Eli 1974 05 01 Solute band spreading in liquid chromatography Causes and importance Analytical Chemistry 46 6 510A 518a doi 10 1021 ac60342a048 ISSN 0003 2700 Wren Stephen A C 2005 06 15 Peak capacity in gradient ultra performance liquid chromatography UPLC Journal of Pharmaceutical and Biomedical Analysis 38 2 337 343 doi 10 1016 j jpba 2004 12 028 ISSN 0731 7085 Zelenyanszki Dora Felinger Attila 2020 10 01 The Impact of Column Hardware on Efficiency in Liquid Chromatography LC LCGC Europe LCGC Europe 10 01 2020 33 10 498 504 Dolan John 2014 LC Method Scaling Part II Gradient Separations LCGC North America LCGC North America 03 01 2014 32 3 188 193 Jensen Ole Elvang Kidal Steffen 2006 03 01 Using Volumetric Flow to Scaleup Chromatographic Processes BioPharm International BioPharm International 03 01 2006 19 3 a b Walter Thomas H Andrews Richard W 2014 Recent innovations in UHPLC columns and instrumentation Trends in Analytical Chemistry 63 14 20 doi 10 1016 j trac 2014 07 016 ISSN 0165 9936 Majors Ronald E 2010 09 07 Fast and Ultrafast HPLC on sub 2 mm Porous Particles Where Do We Go From Here LC GC Europe Lcgceurope com Retrieved 2011 06 07 Xiang Y Liu Y Lee M L 2006 Ultrahigh pressure liquid chromatography using elevated temperature Journal of Chromatography A 1104 1 2 198 202 doi 10 1016 j chroma 2005 11 118 PMID 16376355 Horvath Cs Preiss B A Lipsky S R 1967 Fast liquid chromatography Investigation of operating parameters and the separation of nucleotides on pellicular ion exchangers Analytical Chemistry 39 12 1422 1428 doi 10 1021 ac60256a003 PMID 6073805 Nguyen Dao T T Guillarme Davy Rudaz Serge Veuthey Jean Luc 2006 Fast analysis in liquid chromatography using small particle size and high pressure Journal of Separation Science 29 12 1836 1848 doi 10 1002 jssc 200600189 ISSN 1615 9306 PMID 16970187 Gritti Fabrice Guiochon Georges 2013 The van Deemter equation Assumptions limits and adjustment to modern high performance liquid chromatography Journal of Chromatography A 1302 1 13 doi 10 1016 j chroma 2013 06 032 PMID 23838304 Xiang Yanqiao Liu Yansheng Lee Milton L 2006 Ultrahigh pressure liquid chromatography using elevated temperature Journal of Chromatography A 1104 1 2 198 202 doi 10 1016 j chroma 2005 11 118 PMID 16376355 1290 Infinity Quaternary Pump Archived 2015 11 20 at the Wayback Machine Agilent waters Trademarks Waters www waters com K Robards 1994 Principles and practice of modern chromatographic methods Haddad P R Jackson P E Amsterdam Elsevier Academic Press ISBN 9780080571782 OCLC 815471219 HPLC ECD Electrochemical Detection Fundamentals Markovitch Omer Ottele Jim Veldman Obe Otto Sijbren 2020 Automated device for continuous stirring while sampling in liquid chromatography systems Communications Chemistry 3 1 180 doi 10 1038 s42004 020 00427 5 PMC 9814086 PMID 36703458 Gerber Frederic May 2004 Practical aspects of fast reversed phase high performance liquid chromatography using 3 mm particle packed columns and monolithic columns in pharmaceutical development and production working under current good manufacturing practice Journal of Chromatography 1036 2 127 33 doi 10 1016 j chroma 2004 02 056 PMID 15146913 Siddiqui Masoom Raza AlOthman Zeid A Rahman Nafisur 2013 Analytical techniques in pharmaceutical analysis A review Arabian Journal of Chemistry 10 S1409 S1421 doi 10 1016 j arabjc 2013 04 016 The European Pharmacopoeia 2002 fourth ed Council of Europe Strasbourg United States Pharmacopoeia 2004 27th ed The USP Convention Inc Rockville MD Merone Giuseppe M Tartaglia Angela Rossi Sandra Santavenere Francesco Bassotti Elisa D Ovidio Cristian Bonelli Martina Rosato Enrica de Grazia Ugo Locatelli Marcello Savini Fabio 2021 Fast Quantitative LC MS MS Determination of Illicit Substances in Solid and Liquid Unknown Seized Samples Analytical Chemistry 93 49 16308 16313 doi 10 1021 acs analchem 1c03310 ISSN 0003 2700 PMC 8674870 PMID 34843645 Pesce Amadeo Rosenthal Murray West Robert West Cameron Crews Bridgit Mikel Charles Almazan Perla Latyshev Sergey 2010 06 01 An evaluation of the diagnostic accuracy of liquid chromatography tandem mass spectrometry versus immunoassay drug testing in pain patients Pain Physician 13 3 273 281 PMID 20495592 Tsai I Lin Weng Te I Tseng Yufeng J Tan Happy Kuy Lok Sun Hsiao Ju Kuo Ching Hua 2013 12 01 Screening and confirmation of 62 drugs of abuse and metabolites in urine by ultra high performance liquid chromatography quadrupole time of flight mass spectrometry Journal of Analytical Toxicology 37 9 642 651 doi 10 1093 jat bkt083 PMID 24084874 Weinmann W Renz M Vogt S Pollak S 2000 01 01 Automated solid phase extraction and two step derivatisation for simultaneous analysis of basic illicit drugs in serum by GC MS International Journal of Legal Medicine 113 4 229 235 doi 10 1007 s004149900098 PMID 10929239 S2CID 20451772 Kolmonen Marjo Leinonen Antti Pelander Anna Ojanpera Ilkka 2007 02 28 A general screening method for doping agents in human urine by solid phase extraction and liquid chromatography time of flight mass spectrometry Analytica Chimica Acta 585 1 94 102 doi 10 1016 j aca 2006 12 028 PMID 17386652 Pelander Anna Ojanpera Ilkka Laks Suvi Rasanen Ilpo Vuori Erkki 2003 11 01 Toxicological screening with formula based metabolite identification by liquid chromatography time of flight mass spectrometry Analytical Chemistry 75 21 5710 5718 doi 10 1021 ac030162o PMID 14588010 Nobilis Milan Pour Milan Senel Petr Pavlik Jan Kunes Jiri Voprsalova Marie Kolarova Lenka Holcapek Michal 2007 06 15 Metabolic profiling of a potential antifungal drug 3 4 bromophenyl 5 acetoxymethyl 2 5 dihydrofuran 2 one in mouse urine using high performance liquid chromatography with UV photodiode array and mass spectrometric detection Journal of Chromatography B 853 1 2 10 19 doi 10 1016 j jchromb 2007 02 045 PMID 17400036 Gu Jatin Patel Kumar Shah Dhiren 2016 APPLICATION OF LC MS PharmaTutor Tallam Anil Kumar Alapati Sahithi Nuli Mohana Vamsi 2023 A review on bioanalytical method development and validation of anticancer drugs by using lc ms ms and its applications on routine analysis Journal of Integral Sciences 4 19 doi 10 37022 jis v6i1 51 ISSN 2581 5679 S2CID 257295079 O Driscoll Aimee 2021 HPLC in Pharmaceutical Applications Lab Manager Beccaria Marco Cabooter Deirdre 2020 Current developments in LC MS for pharmaceutical analysis Analyst 145 4 1129 1157 Bibcode 2020Ana 145 1129B doi 10 1039 C9AN02145K hdl 11392 2479221 ISSN 1364 5528 PMID 31971527 S2CID 210866236 Gu Chunang Christine Russell David Yehl Peter 2016 Application of LCMS in small molecule drug development European Pharmaceutical Review 21 4 54 57 D Ovidio Cristian Locatelli Marcello Perrucci Miryam Ciriolo Luigi Furton Kenneth G Gazioglu Isil Kabir Abuzar Merone Giuseppe Maria de Grazia Ugo Ali Imran Catena Antonio Maria Treglia Michele Marsella Luigi T Savini Fabio 2023 LC MS MS Application in Pharmacotoxicological Field Current State and New Applications Molecules 28 5 2127 doi 10 3390 molecules28052127 ISSN 1420 3049 PMC 10004468 PMID 36903374 Zhou Juntuo Zhong Lijun 2022 Applications of liquid chromatography mass spectrometry based metabolomics in predictive and personalized medicine Frontiers in Molecular Biosciences 9 doi 10 3389 fmolb 2022 1049016 ISSN 2296 889X PMC 9669074 PMID 36406271 Mathias Patricia I Connor Thomas H B Hymer Clayton 2017 A review of high performance liquid chromatographic mass spectrometric urinary methods for anticancer drug exposure of health care workers Journal of Chromatography B 1060 316 324 doi 10 1016 j jchromb 2017 06 028 ISSN 1570 0232 PMC 5585056 PMID 28654869 Hernandez Felix Sancho Juan V Ibanez Maria Guerrero Carlos 2007 Antibiotic residue determination in environmental waters by LC MS TrAC Trends in Analytical Chemistry Pharmaceutical residue analysis 26 6 466 485 doi 10 1016 j trac 2007 01 012 ISSN 0165 9936 Seger Christoph Salzmann Linda 2020 After another decade LC MS MS became routine in clinical diagnostics Clinical Biochemistry Advancement and Applications of Mass Spectrometry in Laboratory Medicine 82 2 11 doi 10 1016 j clinbiochem 2020 03 004 ISSN 0009 9120 PMID 32188572 S2CID 213186669 Sundstrom Mira Pelander Anna Angerer Verena Hutter Melanie Kneisel Stefan Ojanpera Ilkka 2013 10 01 A high sensitivity ultra high performance liquid chromatography high resolution time of flight mass spectrometry UHPLC HR TOFMS method for screening synthetic cannabinoids and other drugs of abuse in urine Analytical and Bioanalytical Chemistry 405 26 8463 8474 doi 10 1007 s00216 013 7272 8 PMID 23954996 S2CID 25743579 Gelb Michael H Basheeruddin Khaja Burlina Alberto Chen Hsiao Jan Chien Yin Hsiu Dizikes George Dorley Christine Giugliani Roberto Hietala Amy Hong Xinying Kao Shu Min Khaledi Hamid Klug Tracy Kubaski Francyne Liao Hsuan Chieh 2022 Liquid Chromatography Tandem Mass Spectrometry in Newborn Screening Laboratories International Journal of Neonatal Screening 8 4 62 doi 10 3390 ijns8040062 ISSN 2409 515X PMC 9781967 PMID 36547379 Wang Yanyun Sun Yun Jiang Tao 2019 Clinical Application of LC MS MS in the Follow Up for Treatment of Children with Methylmalonic Aciduria Advances in Therapy 36 6 1304 1313 doi 10 1007 s12325 019 00955 0 ISSN 1865 8652 PMID 31049874 S2CID 143432183 Arunkumar Nivethitha Langan Thomas J Stapleton Molly Kubaski Francyne Mason Robert W Singh Rajendra Kobayashi Hironori Yamaguchi Seiji Suzuki Yasuyuki Orii Kenji Orii Tadao Fukao Toshiyuki Tomatsu Shunji 2020 Newborn screening of mucopolysaccharidoses past present and future Journal of Human Genetics 65 7 557 567 doi 10 1038 s10038 020 0744 8 ISSN 1435 232X PMID 32277174 S2CID 92042115 Skogvold Hanne Bendiksen Rootwelt Helge Reubsaet Leon Elgstoen Katja Benedikte Presto Wilson Steven Ray 2023 Dried blood spot analysis with liquid chromatography and mass spectrometry Trends in clinical chemistry Journal of Separation Science 46 15 e2300210 doi 10 1002 jssc 202300210 ISSN 1615 9306 PMID 37269205 S2CID 259047202 Zahedi Rad Maliheh Neyestani Tirang Reza Nikooyeh Bahareh Shariatzadeh Nastaran Kalayi Ali Khalaji Niloufar Gharavi Azam 2015 01 01 Competitive Protein binding assay based Enzyme immunoassay Method Compared to High pressure Liquid Chromatography Has a Very Lower Diagnostic Value to Detect Vitamin D Deficiency in 9 12 Years Children International Journal of Preventive Medicine 6 67 doi 10 4103 2008 7802 161069 PMC 4542329 PMID 26330983 Further reading editL R Snyder J J Kirkland and J W Dolan Introduction to Modern Liquid Chromatography John Wiley amp Sons New York 2009 M W Dong Modern HPLC for practicing scientists Wiley 2006 L R Snyder J J Kirkland and J L Glajch Practical HPLC Method Development John Wiley amp Sons New York 1997 S Ahuja and H T Rasmussen ed HPLC Method Development for Pharmaceuticals Academic Press 2007 S Ahuja and M W Dong ed Handbook of Pharmaceutical Analysis by HPLC Elsevier Academic Press 2005 Y V Kazakevich and R LoBrutto ed HPLC for Pharmaceutical Scientists Wiley 2007 U D Neue HPLC Columns Theory Technology and Practice Wiley VCH New York 1997 M C McMaster HPLC a practical user s guide Wiley 2007 External links edit nbsp Wikimedia Commons has media related to High performance liquid chromatography HPLC Chromatography Principle Application Basic Note 2020 at Rxlalit com permanent dead link Liquid Chromatography at Curlie Retrieved from https en wikipedia org w index php title High performance liquid chromatography amp oldid 1188368207, wikipedia, wiki, book, books, library,