A computed tomography scan (usually abbreviated to CT scan; formerly called computed axial tomography scan or CAT scan) is a medical imaging technique used to obtain detailed internal images of the body.[2] The personnel that perform CT scans are called radiographers or radiology technologists.[3][4]

CT scan
Modern CT scanner (2021), photon-counting CT (Siemens NAEOTOM Alpha)
Other namesX-ray computed tomography (X-ray CT), computerized axial tomography scan (CAT scan),[1] computer aided tomography, computed tomography scan
ICD-10-PCSB?2
ICD-9-CM88.38
MeSHD014057
OPS-301 code3–20...3–26
MedlinePlus003330
[edit on Wikidata]
CT scan

CT scanners use a rotating X-ray tube and a row of detectors placed in a gantry to measure X-ray attenuations by different tissues inside the body. The multiple X-ray measurements taken from different angles are then processed on a computer using tomographic reconstruction algorithms to produce tomographic (cross-sectional) images (virtual "slices") of a body. CT scan can be used in patients with metallic implants or pacemakers, for whom magnetic resonance imaging (MRI) is contraindicated.

Since its development in the 1970s, CT scanning has proven to be a versatile imaging technique. While CT is most prominently used in medical diagnosis, it can also be used to form images of non-living objects. The 1979 Nobel Prize in Physiology or Medicine was awarded jointly to South African-American physicist Allan MacLeod Cormack and British electrical engineer Godfrey Hounsfield "for the development of computer-assisted tomography".[5][6]

Types edit

On the basis of image acquisition and procedures various type of scanners are available in the market.

Sequential CT edit

Sequential CT also known as step-and-shoot CT is a type of scanning method in which the CT table moves stepwise. The table increments to a particular location and then stops which is followed by the X-ray tube rotation and acquisition of a slice. Then the table increments again and another slice is taken. The table has to make a stop in taking slices. This results in an increased time of scanning.[7]

Spiral CT edit

 
Drawing of CT fan beam and patient in a CT imaging system
 
CT scan of the thorax. The axial slice (right) is the image that corresponds to number 33 (left).

Spinning tube, commonly called spiral CT, or helical CT, is an imaging technique in which an entire X-ray tube is spun around the central axis of the area being scanned. These are the dominant type of scanners on the market because they have been manufactured longer and offer a lower cost of production and purchase. The main limitation of this type of CT is the bulk and inertia of the equipment (X-ray tube assembly and detector array on the opposite side of the circle) which limits the speed at which the equipment can spin. Some designs use two X-ray sources and detector arrays offset by an angle, as a technique to improve temporal resolution.[8][9]

Electron beam tomography edit

Electron beam tomography (EBT) is a specific form of CT in which a large enough X-ray tube is constructed so that only the path of the electrons, travelling between the cathode and anode of the X-ray tube, are spun using deflection coils.[10] This type had a major advantage since sweep speeds can be much faster, allowing for less blurry imaging of moving structures, such as the heart and arteries.[11] Fewer scanners of this design have been produced when compared with spinning tube types, mainly due to the higher cost associated with building a much larger X-ray tube and detector array and limited anatomical coverage.[12]

Dual Energy CT edit

Dual Energy CT also known as Spectral CT is an advancement of Computed Tomography in which two energies are used to create two sets of data.[13] A Dual Energy CT may employ Dual source, Single source with dual detector layer, Single source with energy switching methods to get two different sets of data.[14]

  1. Dual source CT is an advanced scanner with a two X-ray tube detector system, unlike conventional single tube systems.[15][16] These two detector systems are mounted on a single gantry at 90° in the same plane.[17] Dual Source CT scanners allow fast scanning with higher temporal resolution by acquiring a full CT slice in only half a rotation. Fast imaging reduces motion blurring at high heart rates and potentially allowing for shorter breath-hold time. This is particularly useful for ill patients having difficulty holding their breath or unable to take heart-rate lowering medication.[17][18]
  2. Single Source with Energy switching is another mode of Dual energy CT in which a single tube is operated at two different energies by switching the energies frequently.[19][20]

CT perfusion imaging edit

 

CT perfusion imaging is a specific form of CT to assess flow through blood vessels whilst injecting a contrast agent.[21] Blood flow, blood transit time, and organ blood volume, can all be calculated with reasonable sensitivity and specificity.[21] This type of CT may be used on the heart, although sensitivity and specificity for detecting abnormalities are still lower than for other forms of CT.[22] This may also be used on the brain, where CT perfusion imaging can often detect poor brain perfusion well before it is detected using a conventional spiral CT scan.[21][23] This is better for stroke diagnosis than other CT types.[23]

PET CT edit

 
PET-CT scan of chest

Positron emission tomography–computed tomography is a hybrid CT modality which combines, in a single gantry, a positron emission tomography (PET) scanner and an x-ray computed tomography (CT) scanner, to acquire sequential images from both devices in the same session, which are combined into a single superposed (co-registered) image. Thus, functional imaging obtained by PET, which depicts the spatial distribution of metabolic or biochemical activity in the body can be more precisely aligned or correlated with anatomic imaging obtained by CT scanning.[24]

PET-CT gives both anatomical and functional details of an organ under examination and is helpful in detecting different type of cancers.[25][26]

Medical use edit

Since its introduction in the 1970s,[27] CT has become an important tool in medical imaging to supplement conventional X-ray imaging and medical ultrasonography. It has more recently been used for preventive medicine or screening for disease, for example, CT colonography for people with a high risk of colon cancer, or full-motion heart scans for people with a high risk of heart disease. Several institutions offer full-body scans for the general population although this practice goes against the advice and official position of many professional organizations in the field primarily due to the radiation dose applied.[28]

The use of CT scans has increased dramatically over the last two decades in many countries.[29] An estimated 72 million scans were performed in the United States in 2007 and more than 80 million in 2015.[30][31]

Head edit

 
Computed tomography of human brain, from base of the skull to top. Taken with intravenous contrast medium.
Commons: Scrollable computed tomography images of a normal brain

CT scanning of the head is typically used to detect infarction (stroke), tumors, calcifications, haemorrhage, and bone trauma.[32] Of the above, hypodense (dark) structures can indicate edema and infarction, hyperdense (bright) structures indicate calcifications and haemorrhage and bone trauma can be seen as disjunction in bone windows. Tumors can be detected by the swelling and anatomical distortion they cause, or by surrounding edema. CT scanning of the head is also used in CT-guided stereotactic surgery and radiosurgery for treatment of intracranial tumors, arteriovenous malformations, and other surgically treatable conditions using a device known as the N-localizer.[33][34][35][36][37][38]

Neck edit

Contrast CT is generally the initial study of choice for neck masses in adults.[39] CT of the thyroid plays an important role in the evaluation of thyroid cancer.[40] CT scan often incidentally finds thyroid abnormalities, and so is often the preferred investigation modality for thyroid abnormalities.[40]

Lungs edit

A CT scan can be used for detecting both acute and chronic changes in the lung parenchyma, the tissue of the lungs.[41] It is particularly relevant here because normal two-dimensional X-rays do not show such defects. A variety of techniques are used, depending on the suspected abnormality. For evaluation of chronic interstitial processes such as emphysema, and fibrosis,[42] thin sections with high spatial frequency reconstructions are used; often scans are performed both on inspiration and expiration. This special technique is called high resolution CT that produces a sampling of the lung, and not continuous images.[43]

 
HRCT images of a normal thorax in axial, coronal and sagittal planes, respectively.
Click here to scroll through the image stacks.
 
Bronchial wall thickness (T) and diameter of the bronchus (D)

Bronchial wall thickening can be seen on lung CTs and generally (but not always) implies inflammation of the bronchi.[44]

An incidentally found nodule in the absence of symptoms (sometimes referred to as an incidentaloma) may raise concerns that it might represent a tumor, either benign or malignant.[45] Perhaps persuaded by fear, patients and doctors sometimes agree to an intensive schedule of CT scans, sometimes up to every three months and beyond the recommended guidelines, in an attempt to do surveillance on the nodules.[46] However, established guidelines advise that patients without a prior history of cancer and whose solid nodules have not grown over a two-year period are unlikely to have any malignant cancer.[46] For this reason, and because no research provides supporting evidence that intensive surveillance gives better outcomes, and because of risks associated with having CT scans, patients should not receive CT screening in excess of those recommended by established guidelines.[46]

Angiography edit

 
Example of a CTPA, demonstrating a saddle embolus (dark horizontal line) occluding the pulmonary arteries (bright white triangle)

Computed tomography angiography (CTA) is a type of contrast CT to visualize the arteries and veins throughout the body.[47] This ranges from arteries serving the brain to those bringing blood to the lungs, kidneys, arms and legs. An example of this type of exam is CT pulmonary angiogram (CTPA) used to diagnose pulmonary embolism (PE). It employs computed tomography and an iodine-based contrast agent to obtain an image of the pulmonary arteries.[48][49][50]

Cardiac edit

A CT scan of the heart is performed to gain knowledge about cardiac or coronary anatomy.[51] Traditionally, cardiac CT scans are used to detect, diagnose, or follow up coronary artery disease.[52] More recently CT has played a key role in the fast-evolving field of transcatheter structural heart interventions, more specifically in the transcatheter repair and replacement of heart valves.[53][54][55]

The main forms of cardiac CT scanning are:

  • Coronary CT angiography (CCTA): the use of CT to assess the coronary arteries of the heart. The subject receives an intravenous injection of radiocontrast, and then the heart is scanned using a high-speed CT scanner, allowing radiologists to assess the extent of occlusion in the coronary arteries, usually to diagnose coronary artery disease.[56][57]
  • Coronary CT calcium scan: also used for the assessment of severity of coronary artery disease. Specifically, it looks for calcium deposits in the coronary arteries that can narrow arteries and increase the risk of a heart attack.[58] A typical coronary CT calcium scan is done without the use of radiocontrast, but it can possibly be done from contrast-enhanced images as well.[59]

To better visualize the anatomy, post-processing of the images is common.[52] Most common are multiplanar reconstructions (MPR) and volume rendering. For more complex anatomies and procedures, such as heart valve interventions, a true 3D reconstruction or a 3D print is created based on these CT images to gain a deeper understanding.[60][61][62][63]

Abdomen and pelvis edit

 
CT scan of a normal abdomen and pelvis, in sagittal plane, coronal and axial planes, respectively.
Click here to scroll through the image stacks.

CT is an accurate technique for diagnosis of abdominal diseases like Crohn's disease,[64] GIT bleeding, and diagnosis and staging of cancer, as well as follow-up after cancer treatment to assess response.[65] It is commonly used to investigate acute abdominal pain.[66]

Non-enhanced computed tomography is today the gold standard for diagnosing urinary stones.[67] The size, volume and density of stones can be estimated to help clinicians guide further treatment; size is especially important in predicting spontaneous passage of a stone.[68]

Axial skeleton and extremities edit

For the axial skeleton and extremities, CT is often used to image complex fractures, especially ones around joints, because of its ability to reconstruct the area of interest in multiple planes. Fractures, ligamentous injuries, and dislocations can easily be recognized with a 0.2 mm resolution.[69][70] With modern dual-energy CT scanners, new areas of use have been established, such as aiding in the diagnosis of gout.[71]

Biomechanical use edit

CT is used in biomechanics to quickly reveal the geometry, anatomy, density and elastic moduli of biological tissues.[72][73]

Other uses edit

Industrial use edit

Industrial CT scanning (industrial computed tomography) is a process which utilizes X-ray equipment to produce 3D representations of components both externally and internally. Industrial CT scanning has been utilized in many areas of industry for internal inspection of components. Some of the key uses for CT scanning have been flaw detection, failure analysis, metrology, assembly analysis, image-based finite element methods[74] and reverse engineering applications. CT scanning is also employed in the imaging and conservation of museum artifacts.[75]

Aviation security edit

CT scanning has also found an application in transport security (predominantly airport security) where it is currently used in a materials analysis context for explosives detection CTX (explosive-detection device)[76][77][78][79] and is also under consideration for automated baggage/parcel security scanning using computer vision based object recognition algorithms that target the detection of specific threat items based on 3D appearance (e.g. guns, knives, liquid containers).[80][81][82] Its usage in airport security pioneered at Shannon Airport in March 2022 has ended the ban on liquids over 100 ml there, a move that Heathrow Airport plans for a full roll-out on 1 December 2022 and the TSA spent $781.2 million on an order for over 1,000 scanners, ready to go live in the summer.[83]

Geological use edit

X-ray CT is used in geological studies to quickly reveal materials inside a drill core.[84] Dense minerals such as pyrite and barite appear brighter and less dense components such as clay appear dull in CT images.[85]

Cultural heritage use edit

X-ray CT and micro-CT can also be used for the conservation and preservation of objects of cultural heritage. For many fragile objects, direct research and observation can be damaging and can degrade the object over time. Using CT scans, conservators and researchers are able to determine the material composition of the objects they are exploring, such as the position of ink along the layers of a scroll, without any additional harm. These scans have been optimal for research focused on the workings of the Antikythera mechanism or the text hidden inside the charred outer layers of the En-Gedi Scroll. However, they are not optimal for every object subject to these kinds of research questions, as there are certain artifacts like the Herculaneum papyri in which the material composition has very little variation along the inside of the object. After scanning these objects, computational methods can be employed to examine the insides of these objects, as was the case with the virtual unwrapping of the En-Gedi scroll and the Herculaneum papyri.[86] Micro-CT has also proved useful for analyzing more recent artifacts such as still-sealed historic correspondence that employed the technique of letterlocking (complex folding and cuts) that provided a "tamper-evident locking mechanism".[87][88] Further examples of use cases in archaeology is imaging the contents of sarcophagi or ceramics.[89]

Recently, CWI in Amsterdam has collaborated with Rijksmuseum to investigate art object inside details in the framework called IntACT.[90]

Micro organism research edit

Varied types of fungus can degrade wood to different degrees, one Belgium research group has been used X-ray CT 3 dimension with sub-micron resolution unveiled fungi can penetrate micropores of 0.6 μm[91] under certain conditions.

Timber sawmill edit

Sawmills use industrial CT scanners to detect round defects, for instance knots, to improve total value of timber productions. Most sawmills are planning to incorporate this robust detection tool to improve productivity in the long run, however initial investment cost is considerably high.

Microtec is one log scanner manufacturer, with headquarters in Italy.

Interpretation of results edit

Presentation edit

 
Types of presentations of CT scans:
- Average intensity projection
- Maximum intensity projection
- Thin slice (median plane)
- Volume rendering by high and low threshold for radiodensity

The result of a CT scan is a volume of voxels, which may be presented to a human observer by various methods, which broadly fit into the following categories:

Technically, all volume renderings become projections when viewed on a 2-dimensional display, making the distinction between projections and volume renderings a bit vague. The epitomes of volume rendering models feature a mix of for example coloring and shading in order to create realistic and observable representations.[96][97]

Two-dimensional CT images are conventionally rendered so that the view is as though looking up at it from the patient's feet.[98] Hence, the left side of the image is to the patient's right and vice versa, while anterior in the image also is the patient's anterior and vice versa. This left-right interchange corresponds to the view that physicians generally have in reality when positioned in front of patients.[99]

Grayscale edit

Pixels in an image obtained by CT scanning are displayed in terms of relative radiodensity. The pixel itself is displayed according to the mean attenuation of the tissue(s) that it corresponds to on a scale from +3,071 (most attenuating) to −1,024 (least attenuating) on the Hounsfield scale. A pixel is a two dimensional unit based on the matrix size and the field of view. When the CT slice thickness is also factored in, the unit is known as a voxel, which is a three-dimensional unit.[100] Water has an attenuation of 0 Hounsfield units (HU), while air is −1,000 HU, cancellous bone is typically +400 HU, and cranial bone can reach 2,000 HU.[101] The attenuation of metallic implants depends on the atomic number of the element used: Titanium usually has an amount of +1000 HU, iron steel can completely block the X-ray and is, therefore, responsible for well-known line-artifacts in computed tomograms. Artifacts are caused by abrupt transitions between low- and high-density materials, which results in data values that exceed the dynamic range of the processing electronics.[102]

Windowing edit

CT data sets have a very high dynamic range which must be reduced for display or printing. This is typically done via a process of "windowing", which maps a range (the "window") of pixel values to a grayscale ramp. For example, CT images of the brain are commonly viewed with a window extending from 0 HU to 80 HU. Pixel values of 0 and lower, are displayed as black; values of 80 and higher are displayed as white; values within the window are displayed as a gray intensity proportional to position within the window.[103] The window used for display must be matched to the X-ray density of the object of interest, in order to optimize the visible detail.[104] Window width and window level parameters are used to control the windowing of a scan.[105]

Multiplanar reconstruction and projections edit

 
Typical screen layout for diagnostic software, showing one volume rendering (VR) and multiplanar view of three thin slices in the axial (upper right), sagittal (lower left), and coronal planes (lower right)
 
Special planes are sometimes useful, such as this oblique longitudinal plane in order to visualize the neuroforamina of the vertebral column, showing narrowing at two levels, causing radiculopathy. The smaller images are axial plane slices.

Multiplanar reconstruction (MPR) is the process of converting data from one anatomical plane (usually transverse) to other planes. It can be used for thin slices as well as projections. Multiplanar reconstruction is possible as present CT scanners provide almost isotropic resolution.[106]

MPR is used almost in every scan. The spine is frequently examined with it.[107] An image of the spine in axial plane can only show one vertebral bone at a time and cannot show its relation with other vertebral bones. By reformatting the data in other planes, visualization of the relative position can be achieved in sagittal and coronal plane.[108]

New software allows the reconstruction of data in non-orthogonal (oblique) planes, which help in the visualization of organs which are not in orthogonal planes.[109][110] It is better suited for visualization of the anatomical structure of the bronchi as they do not lie orthogonal to the direction of the scan.[111]

Curved-plane reconstruction (or curved planar reformation = CPR) is performed mainly for the evaluation of vessels. This type of reconstruction helps to straighten the bends in a vessel, thereby helping to visualize a whole vessel in a single image or in multiple images. After a vessel has been "straightened", measurements such as cross-sectional area and length can be made. This is helpful in preoperative assessment of a surgical procedure.[112]

For 2D projections used in radiation therapy for quality assurance and planning of external beam radiotherapy, including digitally reconstructed radiographs, see Beam's eye view.

Examples of different algorithms of thickening multiplanar reconstructions[113]
Type of projection Schematic illustration Examples (10 mm slabs) Description Uses
Average intensity projection (AIP)     The average attenuation of each voxel is displayed. The image will get smoother as slice thickness increases. It will look more and more similar to conventional projectional radiography as slice thickness increases. Useful for identifying the internal structures of a solid organ or the walls of hollow structures, such as intestines.
Maximum intensity projection (MIP)     The voxel with the highest attenuation is displayed. Therefore, high-attenuating structures such as blood vessels filled with contrast media are enhanced. Useful for angiographic studies and identification of pulmonary nodules.
Minimum intensity projection (MinIP)     The voxel with the lowest attenuation is displayed. Therefore, low-attenuating structures such as air spaces are enhanced. Useful for assessing the lung parenchyma.

Volume rendering edit

 
3D human skull from computed tomography data

A threshold value of radiodensity is set by the operator (e.g., a level that corresponds to bone). With the help of edge detection image processing algorithms a 3D model can be constructed from the initial data and displayed on screen. Various thresholds can be used to get multiple models, each anatomical component such as muscle, bone and cartilage can be differentiated on the basis of different colours given to them. However, this mode of operation cannot show interior structures.[114]

Surface rendering is limited technique as it displays only the surfaces that meet a particular threshold density, and which are towards the viewer. However, In volume rendering, transparency, colours and shading are used which makes it easy to present a volume in a single image. For example, Pelvic bones could be displayed as semi-transparent, so that, even viewing at an oblique angle one part of the image does not hide another.[115]

Image quality edit

Low-dose CT scan of the thorax
Standard-dose CT scan of the thorax

Dose versus image quality edit

An important issue within radiology today is how to reduce the radiation dose during CT examinations without compromising the image quality. In general, higher radiation doses result in higher-resolution images,[116] while lower doses lead to increased image noise and unsharp images. However, increased dosage raises the adverse side effects, including the risk of radiation-induced cancer – a four-phase abdominal CT gives the same radiation dose as 300 chest X-rays.[117] Several methods that can reduce the exposure to ionizing radiation during a CT scan exist.[118]

  1. New software technology can significantly reduce the required radiation dose. New iterative tomographic reconstruction algorithms (e.g., iterative Sparse Asymptotic Minimum Variance) could offer super-resolution without requiring higher radiation dose.[119]
  2. Individualize the examination and adjust the radiation dose to the body type and body organ examined. Different body types and organs require different amounts of radiation.[120]
  3. Higher resolution is not always suitable, such as detection of small pulmonary masses.[121]

Artifacts edit

Although images produced by CT are generally faithful representations of the scanned volume, the technique is susceptible to a number of artifacts, such as the following:[122][123]Chapters 3 and 5

Streak artifact
Streaks are often seen around materials that block most X-rays, such as metal or bone. Numerous factors contribute to these streaks: under sampling, photon starvation, motion, beam hardening, and Compton scatter. This type of artifact commonly occurs in the posterior fossa of the brain, or if there are metal implants. The streaks can be reduced using newer reconstruction techniques.[124] Approaches such as metal artifact reduction (MAR) can also reduce this artifact.[125][126] MAR techniques include spectral imaging, where CT images are taken with photons of different energy levels, and then synthesized into monochromatic images with special software such as GSI (Gemstone Spectral Imaging).[127]
Partial volume effect
This appears as "blurring" of edges. It is due to the scanner being unable to differentiate between a small amount of high-density material (e.g., bone) and a larger amount of lower density (e.g., cartilage).[128] The reconstruction assumes that the X-ray attenuation within each voxel is homogeneous; this may not be the case at sharp edges. This is most commonly seen in the z-direction (craniocaudal direction), due to the conventional use of highly anisotropic voxels, which have a much lower out-of-plane resolution, than in-plane resolution. This can be partially overcome by scanning using thinner slices, or an isotropic acquisition on a modern scanner.[129]
Ring artifact
Probably the most common mechanical artifact, the image of one or many "rings" appears within an image. They are usually caused by the variations in the response from individual elements in a two dimensional X-ray detector due to defect or miscalibration.[130] Ring artifacts can largely be reduced by intensity normalization, also referred to as flat field correction.[131] Remaining rings can be suppressed by a transformation to polar space, where they become linear stripes.[130] A comparative evaluation of ring artefact reduction on X-ray tomography images showed that the method of Sijbers and Postnov can effectively suppress ring artefacts.[132]
Noise
This appears as grain on the image and is caused by a low signal to noise ratio. This occurs more commonly when a thin slice thickness is used. It can also occur when the power supplied to the X-ray tube is insufficient to penetrate the anatomy.[133]
Windmill
Streaking appearances can occur when the detectors intersect the reconstruction plane. This can be reduced with filters or a reduction in pitch.[134][135]
Beam hardening
This can give a "cupped appearance" when grayscale is visualized as height. It occurs because conventional sources, like X-ray tubes emit a polychromatic spectrum. Photons of higher photon energy levels are typically attenuated less. Because of this, the mean energy of the spectrum increases when passing the object, often described as getting "harder". This leads to an effect increasingly underestimating material thickness, if not corrected. Many algorithms exist to correct for this artifact. They can be divided into mono- and multi-material methods.[124][136][137]

Advantages edit

CT scanning has several advantages over traditional two-dimensional medical radiography. First, CT eliminates the superimposition of images of structures outside the area of interest.[138] Second, CT scans have greater image resolution, enabling examination of finer details. CT can distinguish between tissues that differ in radiographic density by 1% or less.[139] Third, CT scanning enables multiplanar reformatted imaging: scan data can be visualized in the transverse (or axial), coronal, or sagittal plane, depending on the diagnostic task.[140]

The improved resolution of CT has permitted the development of new investigations. For example, CT angiography avoids the invasive insertion of a catheter. CT scanning can perform a virtual colonoscopy with greater accuracy and less discomfort for the patient than a traditional colonoscopy.[141][142] Virtual colonography is far more accurate than a barium enema for detection of tumors and uses a lower radiation dose.[143]

CT is a moderate-to-high radiation diagnostic technique. The radiation dose for a particular examination depends on multiple factors: volume scanned, patient build, number and type of scan protocol, and desired resolution and image quality.[144] Two helical CT scanning parameters, tube current and pitch, can be adjusted easily and have a profound effect on radiation. CT scanning is more accurate than two-dimensional radiographs in evaluating anterior interbody fusion, although they may still over-read the extent of fusion.[145]

Adverse effects edit

Cancer edit

The radiation used in CT scans can damage body cells, including DNA molecules, which can lead to radiation-induced cancer.[146] The radiation doses received from CT scans is variable. Compared to the lowest dose x-ray techniques, CT scans can have 100 to 1,000 times higher dose than conventional X-rays.[147] However, a lumbar spine x-ray has a similar dose as a head CT.[148] Articles in the media often exaggerate the relative dose of CT by comparing the lowest-dose x-ray techniques (chest x-ray) with the highest-dose CT techniques. In general, a routine abdominal CT has a radiation dose similar to three years of average background radiation.[149]

Large scale population-based studies have consistently demonstrated that low dose radiation from CT scans has impacts on cancer incidence in a variety of cancers.[150][151][152][153] For example, in a large population-based cohort it was found that up to 4% of brain cancers were caused by CT scan radiation.[154] Some experts project that in the future, between three and five percent of all cancers would result from medical imaging.[147] An Australian study of 10.9 million people reported that the increased incidence of cancer after CT scan exposure in this cohort was mostly due to irradiation. In this group, one in every 1,800 CT scans was followed by an excess cancer. If the lifetime risk of developing cancer is 40% then the absolute risk rises to 40.05% after a CT. The risks of CT scan radiation are especially important in patients undergoing recurrent CT scans within a short time span of 1 to 5 years.[155][156][157]

Some experts note that CT scans are known to be "overused," and "there is distressingly little evidence of better health outcomes associated with the current high rate of scans."[147] On the other hand, a recent paper analyzing the data of patients who received high cumulative doses showed a high degree of appropriate use.[158] This creates an important issue of cancer risk to these patients. Moreover, a highly significant finding that was previously unreported is that some patients received >100 mSv dose from CT scans in a single day,[156] which counteracts existing criticisms some investigators may have on the effects of protracted versus acute exposure.

There are contrarian views and the debate ongoing. Some studies have shown that publications indicating an increased risk of cancer from typical doses of body CT scans are plagued with serious methodological limitations and several highly improbable results,[159] concluding that no evidence indicates such low doses cause any long-term harm.[160][161][162] One study estimated that as many as 0.4% of cancers in the United States resulted from CT scans, and that this may have increased to as much as 1.5 to 2% based on the rate of CT use in 2007.[146] Others dispute this estimate,[163] as there is no consensus that the low levels of radiation used in CT scans cause damage. Lower radiation doses are used in many cases, such as in the investigation of renal colic.[164]

A person's age plays a significant role in the subsequent risk of cancer.[165] Estimated lifetime cancer mortality risks from an abdominal CT of a one-year-old is 0.1%, or 1:1000 scans.[165] The risk for someone who is 40 years old is half that of someone who is 20 years old with substantially less risk in the elderly.[165] The International Commission on Radiological Protection estimates that the risk to a fetus being exposed to 10 mGy (a unit of radiation exposure) increases the rate of cancer before 20 years of age from 0.03% to 0.04% (for reference a CT pulmonary angiogram exposes a fetus to 4 mGy).[166] A 2012 review did not find an association between medical radiation and cancer risk in children noting however the existence of limitations in the evidences over which the review is based.[167] CT scans can be performed with different settings for lower exposure in children with most manufacturers of CT scans as of 2007 having this function built in.[168] Furthermore, certain conditions can require children to be exposed to multiple CT scans.[146]

Current recommendations are to inform patients of the risks of CT scanning.[169] However, employees of imaging centers tend not to communicate such risks unless patients ask.[170]

Contrast reactions edit

In the United States half of CT scans are contrast CTs using intravenously injected radiocontrast agents.[171] The most common reactions from these agents are mild, including nausea, vomiting, and an itching rash. Severe life-threatening reactions may rarely occur.[172] Overall reactions occur in 1 to 3% with nonionic contrast and 4 to 12% of people with ionic contrast.[173] Skin rashes may appear within a week to 3% of people.[172]

The old radiocontrast agents caused anaphylaxis in 1% of cases while the newer, low-osmolar agents cause reactions in 0.01–0.04% of cases.[172][174] Death occurs in about 2 to 30 people per 1,000,000 administrations, with newer agents being safer.[173][175] There is a higher risk of mortality in those who are female, elderly or in poor health, usually secondary to either anaphylaxis or acute kidney injury.[171]

The contrast agent may induce contrast-induced nephropathy.[176] This occurs in 2 to 7% of people who receive these agents, with greater risk in those who have preexisting kidney failure,[176] preexisting diabetes, or reduced intravascular volume. People with mild kidney impairment are usually advised to ensure full hydration for several hours before and after the injection. For moderate kidney failure, the use of iodinated contrast should be avoided; this may mean using an alternative technique instead of CT. Those with severe kidney failure requiring dialysis require less strict precautions, as their kidneys have so little function remaining that any further damage would not be noticeable and the dialysis will remove the contrast agent; it is normally recommended, however, to arrange dialysis as soon as possible following contrast administration to minimize any adverse effects of the contrast.

In addition to the use of intravenous contrast, orally administered contrast agents are frequently used when examining the abdomen.[177] These are frequently the same as the intravenous contrast agents, merely diluted to approximately 10% of the concentration. However, oral alternatives to iodinated contrast exist, such as very dilute (0.5–1% w/v) barium sulfate suspensions. Dilute barium sulfate has the advantage that it does not cause allergic-type reactions or kidney failure, but cannot be used in patients with suspected bowel perforation or suspected bowel injury, as leakage of barium sulfate from damaged bowel can cause fatal peritonitis.[178]

Side effects from contrast agents, administered intravenously in some CT scans, might impair kidney performance in patients with kidney disease, although this risk is now believed to be lower than previously thought.[179][176]

Scan dose edit

Examination Typical effective
dose
(mSv)
to the whole body
Typical absorbed
dose
(mGy)
to the organ in question
Annual background radiation 2.4[180] 2.4[180]
Chest X-ray 0.02[181] 0.01–0.15[182]
Head CT 1–2[165] 56[183]
Screening mammography 0.4[166] 3[146][182]
Abdominal CT 8[181] 14[183]
Chest CT 5–7[165] 13[183]
CT colonography 6–11[165]
Chest, abdomen and pelvis CT 9.9[183] 12[183]
Cardiac CT angiogram 9–12[165] 40–100[182]
Barium enema 15[146] 15[182]
Neonatal abdominal CT 20[146] 20[182]

The table reports average radiation exposures; however, there can be a wide variation in radiation doses between similar scan types, where the highest dose could be as much as 22 times higher than the lowest dose.[165] A typical plain film X-ray involves radiation dose of 0.01 to 0.15 mGy, while a typical CT can involve 10–20 mGy for specific organs, and can go up to 80 mGy for certain specialized CT scans.[182]

For purposes of comparison, the world average dose rate from naturally occurring sources of background radiation is 2.4 mSv per year, equal for practical purposes in this application to 2.4 mGy per year.[180] While there is some variation, most people (99%) received less than 7 mSv per year as background radiation.[184] Medical imaging as of 2007 accounted for half of the radiation exposure of those in the United States with CT scans making up two thirds of this amount.[165] In the United Kingdom it accounts for 15% of radiation exposure.[166] The average radiation dose from medical sources is ≈0.6 mSv per person globally as of 2007.[165] Those in the nuclear industry in the United States are limited to doses of 50 mSv a year and 100 mSv every 5 years.[165]

Lead is the main material used by radiography personnel for shielding against scattered X-rays.

Radiation dose units edit

The radiation dose reported in the gray or mGy unit is proportional to the amount of energy that the irradiated body part is expected to absorb, and the physical effect (such as DNA double strand breaks) on the cells' chemical bonds by X-ray radiation is proportional to that energy.[185]

The sievert unit is used in the report of the effective dose. The sievert unit, in the context of CT scans, does not correspond to the actual radiation dose that the scanned body part absorbs but to another radiation dose of another scenario, the whole body absorbing the other radiation dose and the other radiation dose being of a magnitude, estimated to have the same probability to induce cancer as the CT scan.[186] Thus, as is shown in the table above, the actual radiation that is absorbed by a scanned body part is often much larger than the effective dose suggests. A specific measure, termed the computed tomography dose index (CTDI), is commonly used as an estimate of the radiation absorbed dose for tissue within the scan region, and is automatically computed by medical CT scanners.[187]

The equivalent dose is the effective dose of a case, in which the whole body would actually absorb the same radiation dose, and the sievert unit is used in its report. In the case of non-uniform radiation, or radiation given to only part of the body, which is common for CT examinations, using the local equivalent dose alone would overstate the biological risks to the entire organism.[188][189][190]

Effects of radiation edit

Most adverse health effects of radiation exposure may be grouped in two general categories:

  • deterministic effects (harmful tissue reactions) due in large part to the killing/malfunction of cells following high doses;[191]
  • stochastic effects, i.e., cancer and heritable effects involving either cancer development in exposed individuals owing to mutation of somatic cells or heritable disease in their offspring owing to mutation of reproductive (germ) cells.[192]

The added lifetime risk of developing cancer by a single abdominal CT of 8 mSv is estimated to be 0.05%, or 1 one in 2,000.[193]

Because of increased susceptibility of fetuses to radiation exposure, the radiation dosage of a CT scan is an important consideration in the choice of medical imaging in pregnancy.[194][195]

Excess doses edit

In October, 2009, the US Food and Drug Administration (FDA) initiated an investigation of brain perfusion CT (PCT) scans, based on radiation burns caused by incorrect settings at one particular facility for this particular type of CT scan. Over 256 patients were exposed to radiations for over 18-month period. Over 40% of them lost patches of hair, and prompted the editorial to call for increased CT quality assurance programs. It was noted that "while unnecessary radiation exposure should be avoided, a medically needed CT scan obtained with appropriate acquisition parameter has benefits that outweigh the radiation risks."[165][196] Similar problems have been reported at other centers.[165] These incidents are believed to be due to human error.[165]

Procedure edit

CT scan procedure varies according to the type of the study and the organ being imaged. The patient is made to lie on the CT table and the centering of the table is done according to the body part. The IV line is established in case of Contrast enhanced CT. After selecting proper and rate of contrast from the pressure injector, the scout is taken to localize and plan the scan. Once the plan is selected, The contrast is given. The raw data is processed according to the study and proper windowing is done to make scans easy to diagnose.[197]

Preparation edit

Patient preparation may vary according to the type of scan. The general patient preparation includes.[197]

  1. Signing the informed consent.
  2. Removal of metallic objects and jewelry from the region of interest.
  3. Changing to the hospital gown according to hospital protocol.
  4. Checking of KFT rate (in case of CECT).[198]

Mechanism edit

 
CT scanner with cover removed to show internal components. Legend:
T: X-ray tube
D: X-ray detectors
X: X-ray beam
R: Gantry rotation
 
Left image is a sinogram which is a graphic representation of the raw data obtained from a CT scan. At right is an image sample derived from the raw data.[199]

Computed tomography operates by using an X-ray generator that rotates around the object; X-ray detectors are positioned on the opposite side of the circle from the X-ray source.[200] As the X-rays pass through the patient, they are attenuated differently by various tissues according to the tissue density.[201] A visual representation of the raw data obtained is called a sinogram, yet it is not sufficient for interpretation.[202] Once the scan data has been acquired, the data must be processed using a form of tomographic reconstruction, which produces a series of cross-sectional images.[203] These cross-sectional images are made up of small units of pixels or voxels.[204]

Pixels in an image obtained by CT scanning are displayed in terms of relative radiodensity. The pixel itself is displayed according to the mean attenuation of the tissue(s) that it corresponds to on a scale from +3,071 (most attenuating) to −1,024 (least attenuating) on the Hounsfield scale. A pixel is a two dimensional unit based on the matrix size and the field of view. When the CT slice thickness is also factored in, the unit is known as a voxel, which is a three-dimensional unit.[204]

Water has an attenuation of 0 Hounsfield units (HU), while air is −1,000 HU, cancellous bone is typically +400 HU, and cranial bone can reach 2,000 HU or more (os temporale) and can cause artifacts. The attenuation of metallic implants depends on the atomic number of the element used: Titanium usually has an amount of +1000 HU, iron steel can completely extinguish the X-ray and is, therefore, responsible for well-known line-artifacts in computed tomograms. Artifacts are caused by abrupt transitions between low- and high-density materials, which results in data values that exceed the dynamic range of the processing electronics. Two-dimensional CT images are conventionally rendered so that the view is as though looking up at it from the patient's feet.[98] Hence, the left side of the image is to the patient's right and vice versa, while anterior in the image also is the patient's anterior and vice versa. This left-right interchange corresponds to the view that physicians generally have in reality when positioned in front of patients.

Initially, the images generated in CT scans were in the transverse (axial) anatomical plane, perpendicular to the long axis of the body. Modern scanners allow the scan data to be reformatted as images in other planes. Digital geometry processing can generate a three-dimensional image of an object inside the body from a series of two-dimensional radiographic images taken by rotation around a fixed axis.[122] These cross-sectional images are widely used for medical diagnosis and therapy.[205]

Contrast edit

Contrast media used for X-ray CT, as well as for plain film X-ray, are called radiocontrasts. Radiocontrasts for CT are, in general, iodine-based.[206] This is useful to highlight structures such as blood vessels that otherwise would be difficult to delineate from their surroundings. Using contrast material can also help to obtain functional information about tissues. Often, images are taken both with and without radiocontrast.[207]

History edit

The history of X-ray computed tomography goes back to at least 1917 with the mathematical theory of the Radon transform.[208][209] In October 1963, William H. Oldendorf received a U.S. patent for a "radiant energy apparatus for investigating selected areas of interior objects obscured by dense material".[210] The first commercially viable CT scanner was invented by Godfrey Hounsfield in 1972.[211]

It is often claimed that revenues from the sales of The Beatles records in the 1960s helped fund the development of the first CT scanner at EMI. The first production X-ray CT machine were in fact called "EMI-Scanner".[212]

Etymology edit

The word "tomography" is derived from the Greek tome (slice) and graphein (to write).[213] Computed tomography was originally known as the "EMI scan" as it was developed in the early 1970s at a research branch of EMI, a company best known today for its music and recording business.[214] It was later known as computed axial tomography (CAT or CT scan) and body section röntgenography.[215]

The term "CAT scan" is no longer used because current CT scans enable for multiplanar reconstructions. This makes "CT scan" the most appropriate term, which is used by radiologists in common vernacular as well as in textbooks and scientific papers.[216][217][218]

In Medical Subject Headings (MeSH), "computed axial tomography" was used from 1977 to 1979, but the current indexing explicitly includes "X-ray" in the title.[219]

The term sinogram was introduced by Paul Edholm and Bertil Jacobson in 1975.[220]

Society and culture edit

Campaigns edit

In response to increased concern by the public and the ongoing progress of best practices, the Alliance for Radiation Safety in Pediatric Imaging was formed within the Society for Pediatric Radiology. In concert with the American Society of Radiologic Technologists, the American College of Radiology and the American Association of Physicists in Medicine, the Society for Pediatric Radiology developed and launched the Image Gently Campaign which is designed to maintain high-quality imaging studies while using the lowest doses and best radiation safety practices available on pediatric patients.[221] This initiative has been endorsed and applied by a growing list of various professional medical organizations around the world and has received support and assistance from companies that manufacture equipment used in Radiology.

Following upon the success of the Image Gently campaign, the American College of Radiology, the Radiological Society of North America, the American Association of Physicists in Medicine and the American Society of Radiologic Technologists have launched a similar campaign to address this issue in the adult population called Image Wisely.[222]

The World Health Organization and International Atomic Energy Agency (IAEA) of the United Nations have also been working in this area and have ongoing projects designed to broaden best practices and lower patient radiation dose.[223][224]

Prevalence edit

Number of CT scanners by country (OECD)
as of 2017[225]
(per million population)
Country Value
  Japan 111.49
  Australia 64.35
  Iceland 43.68
  United States 42.64
  Denmark 39.72
   Switzerland 39.28
  Latvia 39.13
  South Korea 38.18
  Germany 35.13
  Italy 34.71
  Greece 34.22
  Austria 28.64
  Finland 24.51
  Chile 24.27
  Lithuania 23.33
  Ireland 19.14
  Spain 18.59
  Estonia 18.22
  France 17.36
  Slovakia 17.28
  Poland 16.88
  Luxembourg 16.77
  New Zealand 16.69
  Czech Republic 15.76
  Canada 15.28
  Slovenia 15.00
  Turkey 14.77
  Netherlands 13.48
  Russia 13.00
  Israel 9.53
  Hungary 9.19
  Mexico 5.83
  Colombia 1.24

Use of CT has increased dramatically over the last two decades.[29] An estimated 72 million scans were performed in the United States in 2007,[30] accounting for close to half of the total per-capita dose rate from radiologic and nuclear medicine procedures.[226] Of the CT scans, six to eleven percent are done in children,[166] an increase of seven to eightfold from 1980.[165] Similar increases have been seen in Europe and Asia.[165] In Calgary, Canada, 12.1% of people who present to the emergency with an urgent complaint received a CT scan, most commonly either of the head or of the abdomen. The percentage who received CT, however, varied markedly by the emergency physician who saw them from 1.8% to 25%.[227] In the emergency department in the United States, CT or MRI imaging is done in 15% of people who present with injuries as of 2007 (up from 6% in 1998).[228]

The increased use of CT scans has been the greatest in two fields: screening of adults (screening CT of the lung in smokers, virtual colonoscopy, CT cardiac screening, and whole-body CT in asymptomatic patients) and CT imaging of children. Shortening of the scanning time to around 1 second, eliminating the strict need for the subject to remain still or be sedated, is one of the main reasons for the large increase in the pediatric population (especially for the diagnosis of appendicitis).[146] As of 2007, in the United States a proportion of CT scans are performed unnecessarily.[168] Some estimates place this number at 30%.[166] There are a number of reasons for this including: legal concerns, financial incentives, and desire by the public.[168] For example, some healthy people avidly pay to receive full-body CT scans as screening. In that case, it is not at all clear that the benefits outweigh the risks and costs. Deciding whether and how to treat incidentalomas is complex, radiation exposure is not negligible, and the money for the scans involves opportunity cost.[168]

Manufacturers edit

Major manufacturers of CT Scanners Devices and Equipment are:[229]

Research edit

Photon-counting computed tomography is a CT technique currently under development. Typical CT scanners use energy integrating detectors; photons are measured as a voltage on a capacitor which is proportional to the x-rays detected. However, this technique is susceptible to noise and other factors which can affect the linearity of the voltage to x-ray intensity relationship.[230] Photon counting detectors (PCDs) are still affected by noise but it does not change the measured counts of photons. PCDs have several potential advantages, including improving signal (and contrast) to noise ratios, reducing doses, improving spatial resolution, and through use of several energies, distinguishing multiple contrast agents.[231][232] PCDs have only recently become feasible in CT scanners due to improvements in detector technologies that can cope with the volume and rate of data required. As of February 2016, photon counting CT is in use at three sites.[233] Some early research has found the dose reduction potential of photon counting CT for breast imaging to be very promising.[234] In view of recent findings of high cumulative doses to patients from recurrent CT scans, there has been a push for scanning technologies and techniques that reduce ionising radiation doses to patients to sub-milliSievert (sub-mSv in the literature) levels during the CT scan process, a goal that has been lingering.[235][156][157][158]

See also edit

References edit

  1. ^ "CT scan – Mayo Clinic". mayoclinic.org. from the original on 15 October 2016. Retrieved 20 October 2016.
  2. ^ Hermena, Shady; Young, Michael (2022), "CT-scan Image Production Procedures", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 34662062, retrieved 2023-11-24
  3. ^ . ARRT – The American Registry of Radiologic Technologists. Archived from the original on 9 November 2014.
  4. ^ "Individual State Licensure Information". American Society of Radiologic Technologists. from the original on 18 July 2013. Retrieved 19 July 2013.
  5. ^ "The Nobel Prize in Physiology or Medicine 1979". NobelPrize.org. Retrieved 2019-08-10.
  6. ^ "The Nobel Prize in Physiology or Medicine 1979". NobelPrize.org. Retrieved 2023-10-28.
  7. ^ Terrier, F.; Grossholz, M.; Becker, C. D. (2012-12-06). Spiral CT of the Abdomen. Springer Science & Business Media. p. 4. ISBN 978-3-642-56976-0.
  8. ^ Fishman, Elliot K.; Jeffrey, R. Brooke (1995). Spiral CT: Principles, Techniques, and Clinical Applications. Raven Press. ISBN 978-0-7817-0218-8.
  9. ^ Hsieh, Jiang (2003). Computed Tomography: Principles, Design, Artifacts, and Recent Advances. SPIE Press. p. 265. ISBN 978-0-8194-4425-7.
  10. ^ Stirrup, James (2020-01-02). Cardiovascular Computed Tomography. Oxford University Press. ISBN 978-0-19-880927-2.
  11. ^ Talisetti, Anita; Jelnin, Vladimir; Ruiz, Carlos; John, Eunice; Benedetti, Enrico; Testa, Giuliano; Holterman, Ai-Xuan L.; Holterman, Mark J. (December 2004). "Electron beam CT scan is a valuable and safe imaging tool for the pediatric surgical patient". Journal of Pediatric Surgery. 39 (12): 1859–1862. doi:10.1016/j.jpedsurg.2004.08.024. ISSN 1531-5037. PMID 15616951.
  12. ^ Retsky, Michael (31 July 2008). "Electron beam computed tomography: Challenges and opportunities". Physics Procedia. 1 (1): 149–154. Bibcode:2008PhPro...1..149R. doi:10.1016/j.phpro.2008.07.090.
  13. ^ Johnson, Thorsten; Fink, Christian; Schönberg, Stefan O.; Reiser, Maximilian F. (2011-01-18). Dual Energy CT in Clinical Practice. Springer Science & Business Media. ISBN 978-3-642-01740-7.
  14. ^ Johnson, Thorsten; Fink, Christian; Schönberg, Stefan O.; Reiser, Maximilian F. (2011-01-18). Dual Energy CT in Clinical Practice. Springer Science & Business Media. p. 8. ISBN 978-3-642-01740-7.
  15. ^ Carrascosa, Patricia M.; Cury, Ricardo C.; García, Mario J.; Leipsic, Jonathon A. (2015-10-03). Dual-Energy CT in Cardiovascular Imaging. Springer. ISBN 978-3-319-21227-2.
  16. ^ Schmidt, Bernhard; Flohr, Thomas (2020-11-01). "Principles and applications of dual source CT". Physica Medica. 125 Years of X-Rays. 79: 36–46. doi:10.1016/j.ejmp.2020.10.014. ISSN 1120-1797. PMID 33115699. S2CID 226056088.
  17. ^ a b Seidensticker, Peter R.; Hofmann, Lars K. (2008-05-24). Dual Source CT Imaging. Springer Science & Business Media. ISBN 978-3-540-77602-4.
  18. ^ Schmidt, Bernhard; Flohr, Thomas (2020-11-01). "Principles and applications of dual source CT". Physica Medica: European Journal of Medical Physics. 79: 36–46. doi:10.1016/j.ejmp.2020.10.014. ISSN 1120-1797. PMID 33115699. S2CID 226056088.
  19. ^ Mahmood, Usman; Horvat, Natally; Horvat, Joao Vicente; Ryan, Davinia; Gao, Yiming; Carollo, Gabriella; DeOcampo, Rommel; Do, Richard K.; Katz, Seth; Gerst, Scott; Schmidtlein, C. Ross; Dauer, Lawrence; Erdi, Yusuf; Mannelli, Lorenzo (May 2018). "Rapid Switching kVp Dual Energy CT: Value of Reconstructed Dual Energy CT Images and Organ Dose Assessment in Multiphasic Liver CT Exams". European Journal of Radiology. 102: 102–108. doi:10.1016/j.ejrad.2018.02.022. ISSN 0720-048X. PMC 5918634. PMID 29685522.
  20. ^ Johnson, Thorsten R. C. (November 2012). "Dual-Energy CT: General Principles". American Journal of Roentgenology. 199 (5_supplement): S3–S8. doi:10.2214/AJR.12.9116. ISSN 0361-803X. PMID 23097165.
  21. ^ a b c Wittsack, H.-J.; Wohlschläger, A.M.; Ritzl, E.K.; Kleiser, R.; Cohnen, M.; Seitz, R.J.; Mödder, U. (2008-01-01). "CT-perfusion imaging of the human brain: Advanced deconvolution analysis using circulant singular value decomposition". Computerized Medical Imaging and Graphics. 32 (1): 67–77. doi:10.1016/j.compmedimag.2007.09.004. ISSN 0895-6111. PMID 18029143.
  22. ^ Williams, M.C.; Newby, D.E. (2016-08-01). "CT myocardial perfusion imaging: current status and future directions". Clinical Radiology. 71 (8): 739–749. doi:10.1016/j.crad.2016.03.006. ISSN 0009-9260. PMID 27091433.
  23. ^ a b Donahue, Joseph; Wintermark, Max (2015-02-01). "Perfusion CT and acute stroke imaging: Foundations, applications, and literature review". Journal of Neuroradiology. 42 (1): 21–29. doi:10.1016/j.neurad.2014.11.003. ISSN 0150-9861. PMID 25636991.
  24. ^ Blodgett, Todd M.; Meltzer, Carolyn C.; Townsend, David W. (February 2007). "PET/CT: form and function". Radiology. 242 (2): 360–385. doi:10.1148/radiol.2422051113. ISSN 0033-8419. PMID 17255408.
  25. ^ Ciernik, I.Frank; Dizendorf, Elena; Baumert, Brigitta G; Reiner, Beatrice; Burger, Cyrill; Davis, J.Bernard; Lütolf, Urs M; Steinert, Hans C; Von Schulthess, Gustav K (November 2003). "Radiation treatment planning with an integrated positron emission and computer tomography (PET/CT): a feasibility study". International Journal of Radiation Oncology, Biology, Physics. 57 (3): 853–863. doi:10.1016/s0360-3016(03)00346-8. ISSN 0360-3016. PMID 14529793.
  26. ^ Ul-Hassan, Fahim; Cook, Gary J (August 2012). "PET/CT in oncology". Clinical Medicine. 12 (4): 368–372. doi:10.7861/clinmedicine.12-4-368. ISSN 1470-2118. PMC 4952129. PMID 22930885.
  27. ^ Curry, Thomas S.; Dowdey, James E.; Murry, Robert C. (1990). Christensen's Physics of Diagnostic Radiology. Lippincott Williams & Wilkins. p. 289. ISBN 978-0-8121-1310-5.
  28. ^ (PDF). hps.org. Archived from the original (PDF) on 13 October 2016. Retrieved 1 May 2018.
  29. ^ a b Smith-Bindman R, Lipson J, Marcus R, Kim KP, Mahesh M, Gould R, Berrington de González A, Miglioretti DL (December 2009). "Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer". Arch. Intern. Med. 169 (22): 2078–86. doi:10.1001/archinternmed.2009.427. PMC 4635397. PMID 20008690.
  30. ^ a b Berrington de González A, Mahesh M, Kim KP, Bhargavan M, Lewis R, Mettler F, Land C (December 2009). "Projected cancer risks from computed tomographic scans performed in the United States in 2007". Arch. Intern. Med. 169 (22): 2071–7. doi:10.1001/archinternmed.2009.440. PMC 6276814. PMID 20008689.
  31. ^ "Dangers of CT Scans and X-Rays – Consumer Reports". Retrieved 16 May 2018.
  32. ^ Surgeons (AAOS), American Academy of Orthopaedic; Physicians (ACEP), American College of Emergency; UMBC (2017-03-20). Critical Care Transport. Jones & Bartlett Learning. p. 389. ISBN 978-1-284-04099-9.
  33. ^ Galloway, RL Jr. (2015). "Introduction and Historical Perspectives on Image-Guided Surgery". In Golby, AJ (ed.). Image-Guided Neurosurgery. Amsterdam: Elsevier. pp. 3–4. ISBN 978-0-12-800870-6.
  34. ^ Tse, VCK; Kalani, MYS; Adler, JR (2015). "Techniques of Stereotactic Localization". In Chin, LS; Regine, WF (eds.). Principles and Practice of Stereotactic Radiosurgery. New York: Springer. p. 28. ISBN 978-0-387-71070-9.
  35. ^ Saleh, H; Kassas, B (2015). "Developing Stereotactic Frames for Cranial Treatment". In Benedict, SH; Schlesinger, DJ; Goetsch, SJ; Kavanagh, BD (eds.). Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy. Boca Raton: CRC Press. pp. 156–159. ISBN 978-1-4398-4198-3.
  36. ^ Khan, FR; Henderson, JM (2013). "Deep Brain Stimulation Surgical Techniques". In Lozano, AM; Hallet, M (eds.). Brain Stimulation. Handbook of Clinical Neurology. Vol. 116. Amsterdam: Elsevier. pp. 28–30. doi:10.1016/B978-0-444-53497-2.00003-6. ISBN 978-0-444-53497-2. PMID 24112882.
  37. ^ Arle, J (2009). "Development of a Classic: the Todd-Wells Apparatus, the BRW, and the CRW Stereotactic Frames". In Lozano, AM; Gildenberg, PL; Tasker, RR (eds.). Textbook of Stereotactic and Functional Neurosurgery. Berlin: Springer-Verlag. pp. 456–461. ISBN 978-3-540-69959-0.
  38. ^ Brown RA, Nelson JA (June 2012). "Invention of the N-localizer for stereotactic neurosurgery and its use in the Brown-Roberts-Wells stereotactic frame". Neurosurgery. 70 (2 Supplement Operative): 173–176. doi:10.1227/NEU.0b013e318246a4f7. PMID 22186842. S2CID 36350612.
  39. ^ Daniel G Deschler, Joseph Zenga. "Evaluation of a neck mass in adults". UpToDate. This topic last updated: Dec 04, 2017.
  40. ^ a b Bin Saeedan, Mnahi; Aljohani, Ibtisam Musallam; Khushaim, Ayman Omar; Bukhari, Salwa Qasim; Elnaas, Salahudin Tayeb (2016). "Thyroid computed tomography imaging: pictorial review of variable pathologies". Insights into Imaging. 7 (4): 601–617. doi:10.1007/s13244-016-0506-5. ISSN 1869-4101. PMC 4956631. PMID 27271508.
  41. ^ Computed Tomography of the Lung. Springer Berlin Heidelberg. 2007. pp. 40, 47. ISBN 978-3-642-39518-5.
  42. ^ High-resolution CT of the Lung. Lippincott Williams & Wilkins. 2009. pp. 81, 568. ISBN 978-0-7817-6909-9.
  43. ^ Martínez-Jiménez, Santiago; Rosado-de-Christenson, Melissa L.; Carter, Brett W. (2017-07-22). Specialty Imaging: HRCT of the Lung E-Book. Elsevier Health Sciences. ISBN 978-0-323-52495-7.
  44. ^ Yuranga Weerakkody. . Radiopaedia. Archived from the original on 2018-01-06. Retrieved 2018-01-05.
  45. ^ Wiener RS, Gould MK, Woloshin S, Schwartz LM, Clark JA (2012). ""What do you mean, a spot?": A qualitative analysis of patients' reactions to discussions with their doctors about pulmonary nodules". Chest. 143 (3): 672–677. doi:10.1378/chest.12-1095. PMC 3590883. PMID 22814873.
  46. ^ a b c American College of Chest Physicians; American Thoracic Society (September 2013), "Five Things Physicians and Patients Should Question", Choosing Wisely, American College of Chest Physicians and American Thoracic Society, from the original on 3 November 2013, retrieved 6 January 2013, which cites
    • MacMahon H, Austin JH, Gamsu G, Herold CJ, Jett JR, Naidich DP, Patz EF, Swensen SJ (2005). "Guidelines for Management of Small Pulmonary Nodules Detected on CT Scans: A Statement from the Fleischner Society1". Radiology. 237 (2): 395–400. doi:10.1148/radiol.2372041887. PMID 16244247. S2CID 14498160.
    • Gould MK, Fletcher J, Iannettoni MD, Lynch WR, Midthun DE, Naidich DP, Ost DE (2007). "Evaluation of Patients with Pulmonary Nodules: When is It Lung Cancer?*". Chest. 132 (3_suppl): 108S–130S. doi:10.1378/chest.07-1353. PMID 17873164. S2CID 16449420.
    • Smith-Bindman R, Lipson J, Marcus R, Kim KP, Mahesh M, Gould R, Berrington de González A, Miglioretti DL (2009). "Radiation Dose Associated with Common Computed Tomography Examinations and the Associated Lifetime Attributable Risk of Cancer". Archives of Internal Medicine. 169 (22): 2078–2086. doi:10.1001/archinternmed.2009.427. PMC 4635397. PMID 20008690.
    • Wiener RS, Gould MK, Woloshin S, Schwartz LM, Clark JA (2012). ""What do you mean, a spot?": A qualitative analysis of patients' reactions to discussions with their doctors about pulmonary nodules". Chest. 143 (3): 672–677. doi:10.1378/chest.12-1095. PMC 3590883. PMID 22814873.
  47. ^ McDermott, M.; Jacobs, T.; Morgenstern, L. (2017-01-01), Wijdicks, Eelco F. M.; Kramer, Andreas H. (eds.), "Chapter 10 – Critical care in acute ischemic stroke", Handbook of Clinical Neurology, Critical Care Neurology Part I, Elsevier, 140: 153–176, doi:10.1016/b978-0-444-63600-3.00010-6, PMID 28187798
  48. ^ "Computed Tomography Angiography (CTA)". www.hopkinsmedicine.org. 19 November 2019. Retrieved 2021-03-21.
  49. ^ Zeman, R K; Silverman, P M; Vieco, P T; Costello, P (1995-11-01). "CT angiography". American Journal of Roentgenology. 165 (5): 1079–1088. doi:10.2214/ajr.165.5.7572481. ISSN 0361-803X. PMID 7572481.
  50. ^ Ramalho, Joana; Castillo, Mauricio (2014-03-31). Vascular Imaging of the Central Nervous System: Physical Principles, Clinical Applications, and Emerging Techniques. John Wiley & Sons. p. 69. ISBN 978-1-118-18875-0.
  51. ^ "Cardiac CT Scan – NHLBI, NIH". www.nhlbi.nih.gov. from the original on 2017-12-01. Retrieved 2017-11-22.
  52. ^ a b Wichmann, Julian L. . radiopaedia.org. Archived from the original on 2017-12-01. Retrieved 2017-11-22.
  53. ^ Marwan, Mohamed; Achenbach, Stephan (February 2016). "Role of Cardiac CT Before Transcatheter Aortic Valve Implantation (TAVI)". Current Cardiology Reports. 18 (2): 21. doi:10.1007/s11886-015-0696-3. ISSN 1534-3170. PMID 26820560. S2CID 41535442.
  54. ^ Moss, Alastair J.; Dweck, Marc R.; Dreisbach, John G.; Williams, Michelle C.; Mak, Sze Mun; Cartlidge, Timothy; Nicol, Edward D.; Morgan-Hughes, Gareth J. (2016-11-01). "Complementary role of cardiac CT in the assessment of aortic valve replacement dysfunction". Open Heart. 3 (2): e000494. doi:10.1136/openhrt-2016-000494. ISSN 2053-3624. PMC 5093391. PMID 27843568.
  55. ^ Thériault-Lauzier, Pascal; Spaziano, Marco; Vaquerizo, Beatriz; Buithieu, Jean; Martucci, Giuseppe; Piazza, Nicolo (September 2015). "Computed Tomography for Structural Heart Disease and Interventions". Interventional Cardiology Review. 10 (3): 149–154. doi:10.15420/ICR.2015.10.03.149. ISSN 1756-1477. PMC 5808729. PMID 29588693.
  56. ^ Passariello, Roberto (2006-03-30). Multidetector-Row CT Angiography. Springer Science & Business Media. ISBN 978-3-540-26984-7.
  57. ^ Radiology (ACR), Radiological Society of North America (RSNA) and American College of. "Coronary Computed Tomography Angiography (CCTA)". www.radiologyinfo.org. Retrieved 2021-03-19.
  58. ^ "Heart scan (coronary calcium scan)". Mayo Clinic. from the original on 5 September 2015. Retrieved 9 August 2015.
  59. ^ van der Bijl, Noortje; Joemai, Raoul M. S.; Geleijns, Jacob; Bax, Jeroen J.; Schuijf, Joanne D.; de Roos, Albert; Kroft, Lucia J. M. (2010). "Assessment of Agatston Coronary Artery Calcium Score Using Contrast-Enhanced CT Coronary Angiography". American Journal of Roentgenology. 195 (6): 1299–1305. doi:10.2214/AJR.09.3734. ISSN 0361-803X. PMID 21098187.
  60. ^ Vukicevic, Marija; Mosadegh, Bobak; Min, James K.; Little, Stephen H. (February 2017). "Cardiac 3D Printing and its Future Directions". JACC: Cardiovascular Imaging. 10 (2): 171–184. doi:10.1016/j.jcmg.2016.12.001. ISSN 1876-7591. PMC 5664227. PMID 28183437.
  61. ^ Wang, D. D.; Eng, M.; Greenbaum, A.; Myers, E.; Forbes, M.; Pantelic, M.; Song, T.; Nelson, C.; Divine, G.; Taylor, A.; Wyman, J.; Guerrero, M.; Lederman, R. J.; Paone, G.; O'Neill, W. (2016). . JACC: Cardiovascular Imaging. 9 (11): 1349–1352. doi:10.1016/j.jcmg.2016.01.017. PMC 5106323. PMID 27209112. Archived from the original on 2017-12-01. Retrieved 2017-11-22.
  62. ^ Wang, Dee Dee; Eng, Marvin; Greenbaum, Adam; Myers, Eric; Forbes, Michael; Pantelic, Milan; Song, Thomas; Nelson, Christina; Divine, George (November 2016). "Predicting LVOT Obstruction After TMVR". JACC: Cardiovascular Imaging. 9 (11): 1349–1352. doi:10.1016/j.jcmg.2016.01.017. ISSN 1876-7591. PMC 5106323. PMID 27209112.
  63. ^ Jacobs, Stephan; Grunert, Ronny; Mohr, Friedrich W.; Falk, Volkmar (February 2008). "3D-Imaging of cardiac structures using 3D heart models for planning in heart surgery: a preliminary study". Interactive Cardiovascular and Thoracic Surgery. 7 (1): 6–9. doi:10.1510/icvts.2007.156588. ISSN 1569-9285. PMID 17925319.
  64. ^ Furukawa, Akira; Saotome, Takao; Yamasaki, Michio; Maeda, Kiyosumi; Nitta, Norihisa; Takahashi, Masashi; Tsujikawa, Tomoyuki; Fujiyama, Yoshihide; Murata, Kiyoshi; Sakamoto, Tsutomu (2004-05-01). "Cross-sectional Imaging in Crohn Disease". RadioGraphics. 24 (3): 689–702. doi:10.1148/rg.243035120. ISSN 0271-5333. PMID 15143222.
  65. ^ CT of the Acute Abdomen. Springer Berlin Heidelberg. 2011. p. 37. ISBN 978-3-540-89232-8.
  66. ^ Jay P Heiken; Douglas S Katz (2014). "Emergency Radiology of the Abdomen and Pelvis: Imaging of the Nontraumatic and Traumatic Acute Abdomen". In J. Hodler; R. A. Kubik-Huch; G. K. von Schulthess; Ch. L. Zollikofer (eds.). Diseases of the Abdomen and Pelvis. Springer Milan. p. 3. ISBN 9788847056596.
  67. ^ Skolarikos, A; Neisius, A; Petřík, A; Somani, B; Thomas, K; Gambaro, G (March 2022). EAU Guidelines on Urolithiasis. Amsterdam: European Association of Urology. ISBN 978-94-92671-16-5.
  68. ^ Miller, Oren F.; Kane, Christopher J. (September 1999). "Time to stone passage for observed ureteral calculi: a guide for patient education". Journal of Urology. 162 (3 Part 1): 688–691. doi:10.1097/00005392-199909010-00014. PMID 10458343.
  69. ^ . orthoinfo.aaos.org. American Association of Orthopedic Surgeons. Archived from the original on 30 May 2010. Retrieved 30 May 2010.
  70. ^ Buckwalter, Kenneth A.; et al. (11 September 2000). "Musculoskeletal Imaging with Multislice CT". American Journal of Roentgenology. 176 (4): 979–986. doi:10.2214/ajr.176.4.1760979. PMID 11264094.
  71. ^ Ramon, André; Bohm-Sigrand, Amélie; Pottecher, Pierre; Richette, Pascal; Maillefert, Jean-Francis; Devilliers, Herve; Ornetti, Paul (2018-03-01). "Role of dual-energy CT in the diagnosis and follow-up of gout: systematic analysis of the literature". Clinical Rheumatology. 37 (3): 587–595. doi:10.1007/s10067-017-3976-z. ISSN 0770-3198. PMID 29350330. S2CID 3686099.
  72. ^ Keaveny, Tony M. (March 2010). "Biomechanical computed tomography-noninvasive bone strength analysis using clinical computed tomography scans". Annals of the New York Academy of Sciences. 1192 (1): 57–65. Bibcode:2010NYASA1192...57K. doi:10.1111/j.1749-6632.2009.05348.x. ISSN 1749-6632. PMID 20392218. S2CID 24132358.
  73. ^ Barber, Asa; Tozzi, Gianluca; Pani, Martino (2019-03-07). Computed Tomography Based Biomechanics. Frontiers Media SA. p. 20. ISBN 978-2-88945-780-9.
  74. ^ Evans, Ll. M.; Margetts, L.; Casalegno, V.; Lever, L. M.; Bushell, J.; Lowe, T.; Wallwork, A.; Young, P.; Lindemann, A. (2015-05-28). "Transient thermal finite element analysis of CFC–Cu ITER monoblock using X-ray tomography data". Fusion Engineering and Design. 100: 100–111. doi:10.1016/j.fusengdes.2015.04.048. from the original on 2015-10-16.
  75. ^ Payne, Emma Marie (2012). "Imaging Techniques in Conservation" (PDF). Journal of Conservation and Museum Studies. 10 (2): 17–29. doi:10.5334/jcms.1021201.
  76. ^ P. Babaheidarian; D. Castanon (2018). "Joint reconstruction and material classification in spectral CT". In Greenberg, Joel A.; Gehm, Michael E.; Neifeld, Mark A.; Ashok, Amit (eds.). Anomaly Detection and Imaging with X-Rays (ADIX) III. p. 12. doi:10.1117/12.2309663. ISBN 978-1-5106-1775-9. S2CID 65469251.
  77. ^ P. Jin; E. Haneda; K. D. Sauer; C. A. Bouman (June 2012). (PDF). Second International Conference on Image Formation in X-Ray Computed Tomography. Archived from the original (PDF) on 2015-04-11. Retrieved 2015-04-05.
  78. ^ P. Jin; E. Haneda; C. A. Bouman (November 2012). (PDF). Signals, Systems and Computers (ASILOMAR), 2012 Conference Record of the Forty Sixth Asilomar Conference on. IEEE. pp. 613–636. Archived from the original (PDF) on 2015-04-11. Retrieved 2015-04-05.
  79. ^ S. J. Kisner; P. Jin; C. A. Bouman; K. D. Sauer; W. Garms; T. Gable; S. Oh; M. Merzbacher; S. Skatter (October 2013). (PDF). Security Technology (ICCST), 2013 47th International Carnahan Conference on. IEEE. Archived from the original (PDF) on 2015-04-10. Retrieved 2015-04-05.
  80. ^ Megherbi, N.; Flitton, G.T.; Breckon, T.P. (September 2010). "A Classifier based Approach for the Detection of Potential Threats in CT based Baggage Screening" (PDF). Proc. International Conference on Image Processing. IEEE. pp. 1833–1836. CiteSeerX 10.1.1.188.5206. doi:10.1109/ICIP.2010.5653676. ISBN 978-1-4244-7992-4. S2CID 3679917. Retrieved 5 November 2013.[permanent dead link]
  81. ^ Megherbi, N.; Han, J.; Flitton, G.T.; Breckon, T.P. (September 2012). "A Comparison of Classification Approaches for Threat Detection in CT based Baggage Screening" (PDF). Proc. International Conference on Image Processing. IEEE. pp. 3109–3112. CiteSeerX 10.1.1.391.2695. doi:10.1109/ICIP.2012.6467558. ISBN 978-1-4673-2533-2. S2CID 6924816. Retrieved 5 November 2013.[permanent dead link]
  82. ^ Flitton, G.T.; Breckon, T.P.; Megherbi, N. (September 2013). "A Comparison of 3D Interest Point Descriptors with Application to Airport Baggage Object Detection in Complex CT Imagery" (PDF). Pattern Recognition. 46 (9): 2420–2436. Bibcode:2013PatRe..46.2420F. doi:10.1016/j.patcog.2013.02.008. hdl:1826/15213. Retrieved 5 November 2013.[permanent dead link]
  83. ^ Russell, Molly (30 March 2022). "How Next Generation Airport Scanners Are Ending The 100ml Liquid Rule". Simple Flying. Retrieved 11 April 2022.
  84. ^ "Laboratory | About Chikyu | The Deep-sea Scientific Drilling Vessel CHIKYU". www.jamstec.go.jp. Retrieved 2019-10-24.
  85. ^ Tonai, Satoshi; Kubo, Yusuke; Tsang, Man-Yin; Bowden, Stephen; Ide, Kotaro; Hirose, Takehiro; Kamiya, Nana; Yamamoto, Yuzuru; Yang, Kiho; Yamada, Yasuhiro; Morono, Yuki (2019). "A New Method for Quality Control of Geological Cores by X-Ray Computed Tomography: Application in IODP Expedition 370". Frontiers in Earth Science. 7. doi:10.3389/feart.2019.00117. ISSN 2296-6463. S2CID 171394807.
  86. ^ Seales, W. B.; Parker, C. S.; Segal, M.; Tov, E.; Shor, P.; Porath, Y. (2016). "From damage to discovery via virtual unwrapping: Reading the scroll from En-Gedi". Science Advances. 2 (9): e1601247. Bibcode:2016SciA....2E1247S. doi:10.1126/sciadv.1601247. ISSN 2375-2548. PMC 5031465. PMID 27679821.
  87. ^ Castellanos, Sara (2 March 2021). "A Letter Sealed for Centuries Has Been Read—Without Even Opening It". The Wall Street Journal. Retrieved 2 March 2021.
  88. ^ Dambrogio, Jana; Ghassaei, Amanda; Staraza Smith, Daniel; Jackson, Holly; Demaine, Martin L. (2 March 2021). "Unlocking history through automated virtual unfolding of sealed documents imaged by X-ray microtomography". Nature Communications. 12 (1): 1184. Bibcode:2021NatCo..12.1184D. doi:10.1038/s41467-021-21326-w. PMC 7925573. PMID 33654094.
  89. ^ Advanced documentation methods in studying Corinthian black-figure vase painting on YouTube showing a Computed Tomography scan and rollout of the aryballos No. G26, archaeological collection, Graz University. The video was rendered using the GigaMesh Software Framework, cf. doi:10.11588/heidok.00025189. Karl, Stephan; Bayer, Paul; Mara, Hubert; Márton, András (2019), "Advanced Documentation Methods in Studying Corinthian Black-figure Vase Painting" (PDF), Proceedings of the 23rd International Conference on Cultural Heritage and New Technologies (CHNT23), Vienna, Austria, ISBN 978-3-200-06576-5, retrieved 2020-01-09
  90. ^ "CT FOR ART". NICAS. Retrieved 2023-07-04.
  91. ^ Bulcke, Jan Van den; Boone, Matthieu; Acker, Joris Van; Hoorebeke, Luc Van (October 2009). "Three-Dimensional X-Ray Imaging and Analysis of Fungi on and in Wood". Microscopy and Microanalysis. 15 (5): 395–402. Bibcode:2009MiMic..15..395V. doi:10.1017/S1431927609990419. hdl:1854/LU-675607. ISSN 1435-8115. PMID 19709462. S2CID 15637414.
  92. ^ Goldman, L. W. (2008). "Principles of CT: Multislice CT". Journal of Nuclear Medicine Technology. 36 (2): 57–68. doi:10.2967/jnmt.107.044826. ISSN 0091-4916. PMID 18483143.
  93. ^ a b Reis, Eduardo Pontes; Nascimento, Felipe; Aranha, Mateus; Mainetti Secol, Fernando; Machado, Birajara; Felix, Marcelo; Stein, Anouk; Amaro, Edson (29 July 2020). "Brain Hemorrhage Extended (BHX): Bounding box extrapolation from thick to thin slice CT images v1.1". PhysioNet. 101 (23): 215–220. doi:10.13026/9cft-hg92.
  94. ^ Park, S.; Chu, L.C.; Hruban, R.H.; Vogelstein, B.; Kinzler, K.W.; Yuille, A.L.; Fouladi, D.F.; Shayesteh, S.; Ghandili, S.; Wolfgang, C.L.; Burkhart, R.; He, J.; Fishman, E.K.; Kawamoto, S. (2020-09-01). "Differentiating autoimmune pancreatitis from pancreatic ductal adenocarcinoma with CT radiomics features". Diagnostic and Interventional Imaging. 101 (9): 555–564. doi:10.1016/j.diii.2020.03.002. ISSN 2211-5684. PMID 32278586. S2CID 215751181.
  95. ^ a b Fishman, Elliot K.; Ney, Derek R.; Heath, David G.; Corl, Frank M.; Horton, Karen M.; Johnson, Pamela T. (2006). "Volume Rendering versus Maximum Intensity Projection in CT Angiography: What Works Best, When, and Why". RadioGraphics. 26 (3): 905–922. doi:10.1148/rg.263055186. ISSN 0271-5333. PMID 16702462.
  96. ^ Silverstein, Jonathan C.; Parsad, Nigel M.; Tsirline, Victor (2008). "Automatic perceptual color map generation for realistic volume visualization". Journal of Biomedical Informatics. 41 (6): 927–935. doi:10.1016/j.jbi.2008.02.008. ISSN 1532-0464. PMC 2651027. PMID 18430609.
  97. ^ Kobbelt, Leif (2006). Vision, Modeling, and Visualization 2006: Proceedings, November 22-24, 2006, Aachen, Germany. IOS Press. p. 185. ISBN 978-3-89838-081-2.
  98. ^ a b Computerized Tomography chapter 2016-03-04 at the Wayback Machine at University of Connecticut Health Center.
  99. ^ Schmidt, Derek; Odland, Rick (September 2004). "Mirror-Image Reversal of Coronal Computed Tomography Scans". The Laryngoscope. 114 (9): 1562–1565. doi:10.1097/00005537-200409000-00011. ISSN 0023-852X. PMID 15475782. S2CID 22320649.
  100. ^ Brant and Helms' fundamentals of diagnostic radiology (Fifth ed.). Lippincott Williams & Wilkins. 2018-07-19. p. 1600. ISBN 978-1-4963-6738-9. Retrieved 24 January 2019.
  101. ^ Arthur W. Toga; John C. Mazziotta, eds. (2002). Brain mapping: the methods (2nd ed.). Amsterdam: Academic Press. ISBN 0-12-693019-8. OCLC 52594824.
  102. ^ Jerrold T. Bushberg; J. Anthony Seibert; Edwin M. Leidholdt; John M. Boone (2002). The essential physics of medical imaging (2nd ed.). Philadelphia: Lippincott Williams & Wilkins. p. 358. ISBN 0-683-30118-7. OCLC 47177732.
  103. ^ Kamalian, Shervin; Lev, Michael H.; Gupta, Rajiv (2016-01-01). Computed tomography imaging and angiography – principles. Handbook of Clinical Neurology. Vol. 135. pp. 3–20. doi:10.1016/B978-0-444-53485-9.00001-5. ISBN 978-0-444-53485-9. ISSN 0072-9752. PMID 27432657.
  104. ^ Stirrup, James (2020-01-02). Cardiovascular Computed Tomography. Oxford University Press. p. 136. ISBN 978-0-19-880927-2.
  105. ^ Carroll, Quinn B. (2007). Practical Radiographic Imaging. Charles C Thomas Publisher. p. 512. ISBN 978-0-398-08511-7.
  106. ^ Udupa, Jayaram K.; Herman, Gabor T. (1999-09-28). 3D Imaging in Medicine, Second Edition. CRC Press. ISBN 978-0-8493-3179-4.
  107. ^ Krupski, Witold; Kurys-Denis, Ewa; Matuszewski, Łukasz; Plezia, Bogusław (2007-06-30). "Use of multi-planar reconstruction (MPR) and 3-dimentional [sic] (3D) CT to assess stability criteria in C2 vertebral fractures". Journal of Pre-Clinical and Clinical Research. 1 (1): 80–83. ISSN 1898-2395.
  108. ^ Tins, Bernhard (2010-10-21). "Technical aspects of CT imaging of the spine". Insights into Imaging. 1 (5–6): 349–359. doi:10.1007/s13244-010-0047-2. ISSN 1869-4101. PMC 3259341. PMID 22347928.
  109. ^ "CT imaging: Where are we going? (Proceedings)". DVM 360. April 2010. Retrieved 2021-03-21.
  110. ^ Wolfson, Nikolaj; Lerner, Alexander; Roshal, Leonid (2016-05-30). Orthopedics in Disasters: Orthopedic Injuries in Natural Disasters and Mass Casualty Events. Springer. ISBN 978-3-662-48950-5.
  111. ^ Laroia, Archana T; Thompson, Brad H; Laroia, Sandeep T; van Beek, Edwin JR (2010-07-28). "Modern imaging of the tracheo-bronchial tree". World Journal of Radiology. 2 (7): 237–248. doi:10.4329/wjr.v2.i7.237. ISSN 1949-8470. PMC 2998855. PMID 21160663.
  112. ^ Gong, Jing-Shan; Xu, Jian-Min (2004-07-01). "Role of curved planar reformations using multidetector spiral CT in diagnosis of pancreatic and peripancreatic diseases". World Journal of Gastroenterology. 10 (13): 1943–1947. doi:10.3748/wjg.v10.i13.1943. ISSN 1007-9327. PMC 4572236. PMID 15222042.
  113. ^ Dalrymple, Neal C.; Prasad, Srinivasa R.; Freckleton, Michael W.; Chintapalli, Kedar N. (September 2005). "Informatics in radiology (infoRAD): introduction to the language of three-dimensional imaging with multidetector CT". Radiographics. 25 (5): 1409–1428. doi:10.1148/rg.255055044. ISSN 1527-1323. PMID 16160120.
  114. ^ Calhoun, Paul S.; Kuszyk, Brian S.; Heath, David G.; Carley, Jennifer C.; Fishman, Elliot K. (1999-05-01). "Three-dimensional Volume Rendering of Spiral CT Data: Theory and Method". RadioGraphics. 19 (3): 745–764. doi:10.1148/radiographics.19.3.g99ma14745. ISSN 0271-5333. PMID 10336201.
  115. ^ van Ooijen, P. M. A.; van Geuns, R. J. M.; Rensing, B. J. W. M.; Bongaerts, A. H. H.; de Feyter, P. J.; Oudkerk, M. (January 2003). "Noninvasive Coronary Imaging Using Electron Beam CT: Surface Rendering Versus Volume Rendering". American Journal of Roentgenology. 180 (1): 223–226. doi:10.2214/ajr.180.1.1800223. ISSN 0361-803X. PMID 12490509.
  116. ^ R. A. Crowther; D. J. DeRosier; A. Klug (1970). "The Reconstruction of a Three-Dimensional Structure from Projections and its Application to Electron Microscopy". Proc. R. Soc. Lond. A. 317 (1530): 319–340. Bibcode:1970RSPSA.317..319C. doi:10.1098/rspa.1970.0119. S2CID 122980366.
  117. ^ Nickoloff, Edward L.; Alderson, Philip O. (August 2001). "Radiation Exposures to Patients from CT: Reality, Public Perception, and Policy". American Journal of Roentgenology. 177 (2): 285–287. doi:10.2214/ajr.177.2.1770285. ISSN 0361-803X. PMID 11461846.
  118. ^ Barkan, O; Weill, J; Averbuch, A; Dekel, S. "Adaptive Compressed Tomography Sensing" 2016-03-13 at the Wayback Machine. In Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition 2013 (pp. 2195–2202).
  119. ^ Proceedings. IEEE. 1995. p. 10. ISBN 9780780324985.
  120. ^ "Radiation – Effects on organs of the body (somatic effects)". Encyclopedia Britannica. Retrieved 2021-03-21.
  121. ^ Simpson G (2009). "Thoracic computed tomography: principles and practice". Australian Prescriber. 32 (4): 4. doi:10.18773/austprescr.2009.049.
  122. ^ a b Hsieh, Jiang (2003). Computed Tomography: Principles, Design, Artifacts, and Recent Advances. SPIE Press. p. 167. ISBN 978-0-8194-4425-7.
  123. ^ Bhowmik, Ujjal Kumar; Zafar Iqbal, M.; Adhami, Reza R. (28 May 2012). "Mitigating motion artifacts in FDK based 3D Cone-beam Brain Imaging System using markers". Central European Journal of Engineering. 2 (3): 369–382. Bibcode:2012CEJE....2..369B. doi:10.2478/s13531-012-0011-7.
  124. ^ a b P. Jin; C. A. Bouman; K. D. Sauer (2013). (PDF). IEEE Nuclear Science Symp. & Medical Imaging Conf., Seoul, Korea, 2013. Archived from the original (PDF) on 2014-06-06. Retrieved 2014-04-23.
  125. ^ Boas FE, Fleischmann D (2011). "Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography". Radiology. 259 (3): 894–902. doi:10.1148/radiol.11101782. PMID 21357521.
  126. ^ Mouton, A.; Megherbi, N.; Van Slambrouck, K.; Nuyts, J.; Breckon, T.P. (2013). "An Experimental Survey of Metal Artefact Reduction in Computed Tomography" (PDF). Journal of X-Ray Science and Technology. 21 (2): 193–226. doi:10.3233/XST-130372. hdl:1826/8204. PMID 23694911.[permanent dead link]
  127. ^ Pessis, Eric; Campagna, Raphaël; Sverzut, Jean-Michel; Bach, Fabienne; Rodallec, Mathieu; Guerini, Henri; Feydy, Antoine; Drapé, Jean-Luc (2013). "Virtual Monochromatic Spectral Imaging with Fast Kilovoltage Switching: Reduction of Metal Artifacts at CT". RadioGraphics. 33 (2): 573–583. doi:10.1148/rg.332125124. ISSN 0271-5333. PMID 23479714.
  128. ^ González Ballester, Miguel Angel; Zisserman, Andrew P.; Brady, Michael (December 2002). "Estimation of the partial volume effect in MRI". Medical Image Analysis. 6 (4): 389–405. doi:10.1016/s1361-8415(02)00061-0. ISSN 1361-8415. PMID 12494949.
  129. ^ Goldszal, Alberto F.; Pham, Dzung L. (2000-01-01). "Volumetric Segmentation". Handbook of Medical Imaging: 185–194. doi:10.1016/B978-012077790-7/50016-3. ISBN 978-0-12-077790-7.
  130. ^ a b Jha, Diwaker (2014). "Adaptive center determination for effective suppression of ring artifacts in tomography images". Applied Physics Letters. 105 (14): 143107. Bibcode:2014ApPhL.105n3107J. doi:10.1063/1.4897441.
  131. ^ Van Nieuwenhove, V; De Beenhouwer, J; De Carlo, F; Mancini, L; Marone, F; Sijbers, J (2015). "Dynamic intensity normalization using eigen flat fields in X-ray imaging" (PDF). Optics Express. 23 (21): 27975–27989. Bibcode:2015OExpr..2327975V. doi:10.1364/oe.23.027975. hdl:10067/1302930151162165141. PMID 26480456.
  132. ^ Sijbers J, Postnov A (2004). "Reduction of ring artefacts in high resolution micro-CT reconstructions". Phys Med Biol. 49 (14): N247–53. doi:10.1088/0031-9155/49/14/N06. PMID 15357205. S2CID 12744174.
  133. ^ Newton, Thomas H.; Potts, D. Gordon (1971). Radiology of the Skull and Brain: Technical aspects of computed tomography. Mosby. pp. 3941–3950. ISBN 978-0-8016-3662-2.
  134. ^ Brüning, R.; Küttner, A.; Flohr, T. (2006-01-16). Protocols for Multislice CT. Springer Science & Business Media. ISBN 978-3-540-27273-1.
  135. ^ Peh, Wilfred C. G. (2017-08-11). Pitfalls in Musculoskeletal Radiology. Springer. ISBN 978-3-319-53496-1.
  136. ^ Van de Casteele E, Van Dyck D, Sijbers J, Raman E (2004). "A model-based correction method for beam hardening artefacts in X-ray microtomography". Journal of X-ray Science and Technology. 12 (1): 43–57. CiteSeerX 10.1.1.460.6487.
  137. ^ Van Gompel G, Van Slambrouck K, Defrise M, Batenburg KJ, Sijbers J, Nuyts J (2011). "Iterative correction of beam hardening artifacts in CT". Medical Physics. 38 (1): 36–49. Bibcode:2011MedPh..38S..36V. CiteSeerX 10.1.1.464.3547. doi:10.1118/1.3577758. PMID 21978116.
  138. ^ Mikla, Victor I.; Mikla, Victor V. (2013-08-23). Medical Imaging Technology. Elsevier. p. 37. ISBN 978-0-12-417036-0.
  139. ^ Radiology for the Dental Professional. Elsevier Mosby. 2008. p. 337. ISBN 978-0-323-03071-7.
  140. ^ Pasipoularides, Ares (November 2009). Heart's Vortex: Intracardiac Blood Flow Phenomena. PMPH-USA. p. 595. ISBN 978-1-60795-033-2.
  141. ^ Heiken, JP; Peterson CM; Menias CO (November 2005). "Virtual colonoscopy for colorectal cancer screening: current status: Wednesday 5 October 2005, 14:00–16:00". Cancer Imaging. International Cancer Imaging Society. 5 (Spec No A): S133–S139. doi:10.1102/1470-7330.2005.0108. PMC 1665314. PMID 16361129.
  142. ^ Bielen DJ, Bosmans HT, De Wever LL, et al. (September 2005). "Clinical validation of high-resolution fast spin-echo MR colonography after colon distention with air". J Magn Reson Imaging. 22 (3): 400–5. doi:10.1002/jmri.20397. PMID 16106357. S2CID 22167728.
  143. ^ "CT Colonography". Radiologyinfo.org.
  144. ^ Žabić S, Wang Q, Morton T, Brown KM (March 2013). "A low dose simulation tool for CT systems with energy integrating detectors". Medical Physics. 40 (3): 031102. Bibcode:2013MedPh..40c1102Z. doi:10.1118/1.4789628. PMID 23464282.
  145. ^ Brian R. Subach M.D., F.A.C.S et al."Reliability and accuracy of fine-cut computed tomography scans to determine the status of anterior interbody fusions with metallic cages" 2012-12-08 at the Wayback Machine
  146. ^ a b c d e f g Brenner DJ, Hall EJ (November 2007). "Computed tomography – an increasing source of radiation exposure" (PDF). N. Engl. J. Med. 357 (22): 2277–84. doi:10.1056/NEJMra072149. PMID 18046031. S2CID 2760372. (PDF) from the original on 2016-03-04.
  147. ^ a b c Redberg, Rita F., and Smith-Bindman, Rebecca. "We Are Giving Ourselves Cancer" 2017-07-06 at the Wayback Machine, New York Times, January 30, 2014
  148. ^ Health, Center for Devices and Radiological. "Medical X-ray Imaging – What are the Radiation Risks from CT?". www.fda.gov. from the original on 5 November 2013. Retrieved 1 May 2018.
  149. ^ (ACR), Radiological Society of North America (RSNA) and American College of Radiology (February 2021). (PDF). acr.org. Archived from the original (PDF) on 1 January 2021. Retrieved 6 April 2021.
  150. ^ Mathews, J. D.; Forsythe, A. V.; Brady, Z.; Butler, M. W.; Goergen, S. K.; Byrnes, G. B.; Giles, G. G.; Wallace, A. B.; Anderson, P. R.; Guiver, T. A.; McGale, P.; Cain, T. M.; Dowty, J. G.; Bickerstaffe, A. C.; Darby, S. C. (2013). "Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians". BMJ. 346 (may21 1): f2360. doi:10.1136/bmj.f2360. ISSN 1756-1833. PMC 3660619. PMID 23694687.
  151. ^ Pearce, MS; Salotti, JA; Little, MP; McHugh, K; Lee, C; Kim, KP; Howe, NL; Ronckers, CM; Rajaraman, P; Sir Craft, AW; Parker, L; Berrington de González, A (4 August 2012). "Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study". Lancet. 380 (9840): 499–505. doi:10.1016/S0140-6736(12)60815-0. PMC 3418594. PMID 22681860.
  152. ^ Meulepas, Johanna M; Ronckers, Cécile M; Smets, Anne M J B; Nievelstein, Rutger A J; Gradowska, Patrycja; Lee, Choonsik; Jahnen, Andreas; van Straten, Marcel; de Wit, Marie-Claire Y; Zonnenberg, Bernard; Klein, Willemijn M; Merks, Johannes H; Visser, Otto; van Leeuwen, Flora E; Hauptmann, Michael (1 March 2019). "Radiation Exposure From Pediatric CT Scans and Subsequent Cancer Risk in the Netherlands". JNCI: Journal of the National Cancer Institute. 111 (3): 256–263. doi:10.1093/jnci/djy104. PMC 6657440. PMID 30020493.
  153. ^ de Gonzalez, Amy Berrington; Salotti, Jane A; McHugh, Kieran; Little, Mark P; Harbron, Richard W; Lee, Choonsik; Ntowe, Estelle; Braganza, Melissa Z; Parker, Louise; Rajaraman, Preetha; Stiller, Charles; Stewart, Douglas R; Craft, Alan W; Pearce, Mark S (February 2016). "Relationship between paediatric CT scans and subsequent risk of leukaemia and brain tumours: assessment of the impact of underlying conditions". British Journal of Cancer. 114 (4): 388–394. doi:10.1038/bjc.2015.415. PMC 4815765. PMID 26882064.
  154. ^ Smoll, Nicolas R; Brady, Zoe; Scurrah, Katrina J; Lee, Choonsik; Berrington de González, Amy; Mathews, John D (14 January 2023). "Computed tomography scan radiation and brain cancer incidence". Neuro-Oncology. 25 (7): 1368–1376. doi:10.1093/neuonc/noad012. PMC 10326490. PMID 36638155.
  155. ^ Sasieni, P D; Shelton, J; Ormiston-Smith, N; Thomson, C S; Silcocks, P B (2011). "What is the lifetime risk of developing cancer?: the effect of adjusting for multiple primaries". British Journal of Cancer. 105 (3): 460–465. doi:10.1038/bjc.2011.250. ISSN 0007-0920. PMC 3172907. PMID 21772332.
  156. ^ a b c Rehani, Madan M.; Yang, Kai; Melick, Emily R.; Heil, John; Šalát, Dušan; Sensakovic, William F.; Liu, Bob (2020). "Patients undergoing recurrent CT scans: assessing the magnitude". European Radiology. 30 (4): 1828–1836. doi:10.1007/s00330-019-06523-y. PMID 31792585. S2CID 208520824.
  157. ^ a b Brambilla, Marco; Vassileva, Jenia; Kuchcinska, Agnieszka; Rehani, Madan M. (2020). "Multinational data on cumulative radiation exposure of patients from recurrent radiological procedures: call for action". European Radiology. 30 (5): 2493–2501. doi:10.1007/s00330-019-06528-7. PMID 31792583. S2CID 208520544.
  158. ^ a b Rehani, Madan M.; Melick, Emily R.; Alvi, Raza M.; Doda Khera, Ruhani; Batool-Anwar, Salma; Neilan, Tomas G.; Bettmann, Michael (2020). "Patients undergoing recurrent CT exams: assessment of patients with non-malignant diseases, reasons for imaging and imaging appropriateness". European Radiology. 30 (4): 1839–1846. doi:10.1007/s00330-019-06551-8. PMID 31792584. S2CID 208520463.
  159. ^ Eckel, Laurence J.; Fletcher, Joel G.; Bushberg, Jerrold T.; McCollough, Cynthia H. (2015-10-01). "Answers to Common Questions About the Use and Safety of CT Scans". Mayo Clinic Proceedings. 90 (10): 1380–1392. doi:10.1016/j.mayocp.2015.07.011. ISSN 0025-6196. PMID 26434964.
  160. ^ "Expert opinion: Are CT scans safe?". ScienceDaily. Retrieved 2019-03-14.
  161. ^ McCollough, Cynthia H.; Bushberg, Jerrold T.; Fletcher, Joel G.; Eckel, Laurence J. (2015-10-01). "Answers to Common Questions About the Use and Safety of CT Scans". Mayo Clinic Proceedings. 90 (10): 1380–1392. doi:10.1016/j.mayocp.2015.07.011. ISSN 0025-6196. PMID 26434964.
  162. ^ "No evidence that CT scans, X-rays cause cancer". Medical News Today. 4 February 2016. Retrieved 2019-03-14.
  163. ^ Kalra, Mannudeep K.; Maher, Michael M.; Rizzo, Stefania; Kanarek, David; Shephard, Jo-Anne O. (April 2004). "Radiation exposure from Chest CT: Issues and Strategies". Journal of Korean Medical Science. 19 (2): 159–166. doi:10.3346/jkms.2004.19.2.159. ISSN 1011-8934. PMC 2822293. PMID 15082885.
  164. ^ Rob, S.; Bryant, T.; Wilson, I.; Somani, B.K. (2017). "Ultra-low-dose, low-dose, and standard-dose CT of the kidney, ureters, and bladder: is there a difference? Results from a systematic review of the literature". Clinical Radiology. 72 (1): 11–15. doi:10.1016/j.crad.2016.10.005. PMID 27810168.
  165. ^ a b c d e f g h i j k l m n o p Whaites, Eric (2008-10-10). Radiography and Radiology for Dental Care Professionals E-Book. Elsevier Health Sciences. p. 25. ISBN 978-0-7020-4799-2.
  166. ^ a b c d e Davies, H. E.; Wathen, C. G.; Gleeson, F. V. (25 February 2011). "The risks of radiation exposure related to diagnostic imaging and how to minimise them". BMJ. 342 (feb25 1): d947. doi:10.1136/bmj.d947. PMID 21355025. S2CID 206894472.
  167. ^ Baysson H, Etard C, Brisse HJ, Bernier MO (January 2012). "[Diagnostic radiation exposure in children and cancer risk: current knowledge and perspectives]". Archives de Pédiatrie. 19 (1): 64–73. doi:10.1016/j.arcped.2011.10.023. PMID 22130615.
  168. ^ a b c d Semelka RC, Armao DM, Elias J, Huda W (May 2007). "Imaging strategies to reduce the risk of radiation in CT studies, including selective substitution with MRI". J Magn Reson Imaging. 25 (5): 900–9. doi:10.1002/jmri.20895. PMID 17457809. S2CID 5788891.
  169. ^ Larson DB, Rader SB, Forman HP, Fenton LZ (August 2007). "Informing parents about CT radiation exposure in children: it's OK to tell them". Am J Roentgenol. 189 (2): 271–5. doi:10.2214/AJR.07.2248. PMID 17646450. S2CID 25020619.
  170. ^ Emmerson, Benjamin; Young, Michael (2023), "Radiology Patient Safety and Communication", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 33620790, retrieved 2023-11-24
  171. ^ a b Namasivayam S, Kalra MK, Torres WE, Small WC (Jul 2006). "Adverse reactions to intravenous iodinated contrast media: a primer for radiologists". Emergency Radiology. 12 (5): 210–5. doi:10.1007/s10140-006-0488-6. PMID 16688432. S2CID 28223134.
  172. ^ a b c Christiansen C (2005-04-15). "X-ray contrast media – an overview". Toxicology. 209 (2): 185–7. doi:10.1016/j.tox.2004.12.020. PMID 15767033.
  173. ^ a b Wang H, Wang HS, Liu ZP (October 2011). "Agents that induce pseudo-allergic reaction". Drug Discov Ther. 5 (5): 211–9. doi:10.5582/ddt.2011.v5.5.211. PMID 22466368. S2CID 19001357.
  174. ^ Drain KL, Volcheck GW (2001). "Preventing and managing drug-induced anaphylaxis". Drug Safety. 24 (11): 843–53. doi:10.2165/00002018-200124110-00005. PMID 11665871. S2CID 24840296.
  175. ^ Castells, Mariana C., ed. (2010-12-09). Anaphylaxis and hypersensitivity reactions. New York: Humana Press. p. 187. ISBN 978-1-60327-950-5.
  176. ^ a b c Hasebroock KM, Serkova NJ (April 2009). "Toxicity of MRI and CT contrast agents". Expert Opinion on Drug Metabolism & Toxicology. 5 (4): 403–16. doi:10.1517/17425250902873796. PMID 19368492. S2CID 72557671.
  177. ^ Rawson, James V.; Pelletier, Allen L. (2013-09-01). "When to Order Contrast-Enhanced CT". American Family Physician. 88 (5): 312–316. ISSN 0002-838X. PMID 24010394.
  178. ^ Thomsen, Henrik S.; Muller, Robert N.; Mattrey, Robert F. (2012-12-06). Trends in Contrast Media. Springer Science & Business Media. ISBN 978-3-642-59814-2.
  179. ^ Davenport, Matthew (2020). "Use of Intravenous Iodinated Contrast Media in Patients with Kidney Disease: Consensus Statements from the American College of Radiology and the National Kidney Foundation". Radiology. 294 (3): 660–668. doi:10.1148/radiol.2019192094. PMID 31961246.
  180. ^ a b c Cuttler JM, Pollycove M (2009). "Nuclear energy and health: and the benefits of low-dose radiation hormesis". Dose-Response. 7 (1): 52–89. doi:10.2203/dose-response.08-024.Cuttler. PMC 2664640. PMID 19343116.
  181. ^ a b "What are the Radiation Risks from CT?". Food and Drug Administration. 2009. from the original on 2013-11-05.
  182. ^ a b c d e f Hall EJ, Brenner DJ (May 2008). "Cancer risks from diagnostic radiology". The British Journal of Radiology. 81 (965): 362–78. doi:10.1259/bjr/01948454. PMID 18440940. S2CID 23348032.
  183. ^ a b c d e Shrimpton, P.C; Miller, H.C; Lewis, M.A; Dunn, M. Doses from Computed Tomography (CT) examinations in the UK – 2003 Review 2011-09-22 at the Wayback Machine
  184. ^ Michael T. Ryan; Poston, John W., eds. (2005). A half century of health physics. Baltimore, Md.: Lippincott Williams & Wilkins. p. 164. ISBN 978-0-7817-6934-1.
  185. ^ Polo SE, Jackson SP (March 2011). "Dynamics of DNA damage response proteins at DNA breaks: a focus on protein modifications". Genes Dev. 25 (5): 409–33. doi:10.1101/gad.2021311. PMC 3049283. PMID 21363960.
  186. ^ The Measurement, Reporting, and Management of Radiation Dose in CT 2017-06-23 at the Wayback Machine "It is a single dose parameter that reflects the risk of a nonuniform exposure in terms of an equivalent whole-body exposure."
  187. ^ Hill B, Venning AJ, Baldock C (2005). "A preliminary study of the novel application of normoxic polymer gel dosimeters for the measurement of CTDI on diagnostic X-ray CT scanners". Medical Physics. 32 (6): 1589–1597. Bibcode:2005MedPh..32.1589H. doi:10.1118/1.1925181. PMID 16013718.
  188. ^ Issa, Ziad F.; Miller, John M.; Zipes, Douglas P. (2019-01-01). "Complications of Catheter Ablation of Cardiac Arrhythmias". Clinical Arrhythmology and Electrophysiology. Elsevier. pp. 1042–1067. doi:10.1016/b978-0-323-52356-1.00032-3. ISBN 978-0-323-52356-1.
  189. ^ "Absorbed, Equivalent, and Effective Dose – ICRPaedia". icrpaedia.org. Retrieved 2021-03-21.
  190. ^ Materials, National Research Council (US) Committee on Evaluation of EPA Guidelines for Exposure to Naturally Occurring Radioactive (1999). Radiation Quantities and Units, Definitions, Acronyms. National Academies Press (US).
  191. ^ Pua, Bradley B.; Covey, Anne M.; Madoff, David C. (2018-12-03). Interventional Radiology: Fundamentals of Clinical Practice. Oxford University Press. ISBN 978-0-19-027624-9.
  192. ^ Paragraph 55 in: "The 2007 Recommendations of the International Commission on Radiological Protection". International Commission on Radiological Protection. from the original on 2012-11-16. Ann. ICRP 37 (2-4)
  193. ^ . Harvard Medical School. March 2013. Archived from the original on 2017-12-09. Retrieved 2017-12-09.
  194. ^ CDC (2020-06-05). "Radiation and pregnancy: A fact sheet for clinicians". Centers for Disease Control and Prevention. Retrieved 2021-03-21.
  195. ^ Yoon, Ilsup; Slesinger, Todd L. (2021), "Radiation Exposure In Pregnancy", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 31869154, retrieved 2021-03-21
  196. ^ Wintermark M, Lev MH (January 2010). "FDA investigates the safety of brain perfusion CT". AJNR Am J Neuroradiol. 31 (1): 2–3. doi:10.3174/ajnr.A1967. PMC 7964089. PMID 19892810.
  197. ^ a b Whitley, Stewart A.; Dodgeon, Jan; Meadows, Angela; Cullingworth, Jane; Holmes, Ken; Jackson, Marcus; Hoadley, Graham; Kulshrestha, Randeep (2020-01-06). Clark's Procedures in Diagnostic Imaging: A System-Based Approach. CRC Press. ISBN 978-1-4987-1552-2.
  198. ^ Tippins, R. B.; Torres, W. E.; Baumgartner, B. R.; Baumgarten, D. A. (August 2000). "Are screening serum creatinine levels necessary prior to outpatient CT examinations?". Radiology. 216 (2): 481–484. doi:10.1148/radiology.216.2.r00au23481. ISSN 0033-8419. PMID 10924574.
  199. ^ Jun, Kyungtaek; Yoon, Seokhwan (2017). "Alignment Solution for CT Image Reconstruction using Fixed Point and Virtual Rotation Axis". Scientific Reports. 7: 41218. arXiv:1605.04833. Bibcode:2017NatSR...741218J. doi:10.1038/srep41218. ISSN 2045-2322. PMC 5264594. PMID 28120881.
  200. ^ "Computed Tomography (CT)". www.nibib.nih.gov. Retrieved 2021-03-20.
  201. ^ Aichinger, Horst; Dierker, Joachim; Joite-Barfuß, Sigrid; Säbel, Manfred (2011-10-25). Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology: Physical Principles and Clinical Applications. Springer Science & Business Media. p. 5. ISBN 978-3-642-11241-6.
  202. ^ Erdoğan, Hakan (1999). Statistical Image Reconstruction Algorithms Using Paraboloidal Surrogates for PET Transmission Scans. University of Michigan. ISBN 978-0-599-63374-2.
  203. ^ Themes, U. F. O. (2018-10-07). "CT Image Reconstruction Basics". Radiology Key. Retrieved 2021-03-20.
  204. ^ a b Stirrup, James (2020-01-02). Cardiovascular Computed Tomography. Oxford University Press. ISBN 978-0-19-880927-2.
  205. ^ "computed tomography – Definition from the Merriam-Webster Online Dictionary". from the original on 19 September 2011. Retrieved 18 August 2009.
  206. ^ Webb, W. Richard; Brant, William E.; Major, Nancy M. (2014). Fundamentals of Body CT. Elsevier Health Sciences. p. 152. ISBN 978-0-323-26358-0.
  207. ^ Webb, Wayne Richard; Brant, William E.; Major, Nancy M. (2006-01-01). Fundamentals of Body CT. Elsevier Health Sciences. p. 168. ISBN 978-1-4160-0030-3.
scan, this, article, about, computed, tomography, used, medicine, cross, sectional, images, used, industry, industrial, computed, tomography, means, tomography, other, than, tomography, computed, tomography, scan, usually, abbreviated, formerly, called, comput. This article is about X ray computed tomography as used in medicine For cross sectional images used in industry see Industrial computed tomography For means of tomography other than X ray see Tomography A computed tomography scan usually abbreviated to CT scan formerly called computed axial tomography scan or CAT scan is a medical imaging technique used to obtain detailed internal images of the body 2 The personnel that perform CT scans are called radiographers or radiology technologists 3 4 CT scanModern CT scanner 2021 photon counting CT Siemens NAEOTOM Alpha Other namesX ray computed tomography X ray CT computerized axial tomography scan CAT scan 1 computer aided tomography computed tomography scanICD 10 PCSB 2ICD 9 CM88 38MeSHD014057OPS 301 code3 20 3 26MedlinePlus003330 edit on Wikidata source source source source CT scanCT scanners use a rotating X ray tube and a row of detectors placed in a gantry to measure X ray attenuations by different tissues inside the body The multiple X ray measurements taken from different angles are then processed on a computer using tomographic reconstruction algorithms to produce tomographic cross sectional images virtual slices of a body CT scan can be used in patients with metallic implants or pacemakers for whom magnetic resonance imaging MRI is contraindicated Since its development in the 1970s CT scanning has proven to be a versatile imaging technique While CT is most prominently used in medical diagnosis it can also be used to form images of non living objects The 1979 Nobel Prize in Physiology or Medicine was awarded jointly to South African American physicist Allan MacLeod Cormack and British electrical engineer Godfrey Hounsfield for the development of computer assisted tomography 5 6 Contents 1 Types 1 1 Sequential CT 1 2 Spiral CT 1 3 Electron beam tomography 1 4 Dual Energy CT 1 5 CT perfusion imaging 1 6 PET CT 2 Medical use 2 1 Head 2 2 Neck 2 3 Lungs 2 4 Angiography 2 5 Cardiac 2 6 Abdomen and pelvis 2 7 Axial skeleton and extremities 2 8 Biomechanical use 3 Other uses 3 1 Industrial use 3 2 Aviation security 3 3 Geological use 3 4 Cultural heritage use 3 5 Micro organism research 3 6 Timber sawmill 4 Interpretation of results 4 1 Presentation 4 1 1 Grayscale 4 1 2 Windowing 4 1 3 Multiplanar reconstruction and projections 4 1 4 Volume rendering 4 2 Image quality 4 2 1 Dose versus image quality 4 2 2 Artifacts 5 Advantages 6 Adverse effects 6 1 Cancer 6 2 Contrast reactions 6 3 Scan dose 6 3 1 Radiation dose units 6 3 2 Effects of radiation 6 3 3 Excess doses 7 Procedure 7 1 Preparation 8 Mechanism 8 1 Contrast 9 History 9 1 Etymology 10 Society and culture 10 1 Campaigns 10 2 Prevalence 11 Manufacturers 12 Research 13 See also 14 References 15 External linksTypes editOn the basis of image acquisition and procedures various type of scanners are available in the market Sequential CT edit Sequential CT also known as step and shoot CT is a type of scanning method in which the CT table moves stepwise The table increments to a particular location and then stops which is followed by the X ray tube rotation and acquisition of a slice Then the table increments again and another slice is taken The table has to make a stop in taking slices This results in an increased time of scanning 7 Spiral CT edit nbsp Drawing of CT fan beam and patient in a CT imaging system nbsp CT scan of the thorax The axial slice right is the image that corresponds to number 33 left Spinning tube commonly called spiral CT or helical CT is an imaging technique in which an entire X ray tube is spun around the central axis of the area being scanned These are the dominant type of scanners on the market because they have been manufactured longer and offer a lower cost of production and purchase The main limitation of this type of CT is the bulk and inertia of the equipment X ray tube assembly and detector array on the opposite side of the circle which limits the speed at which the equipment can spin Some designs use two X ray sources and detector arrays offset by an angle as a technique to improve temporal resolution 8 9 Electron beam tomography edit Main article Electron beam computed tomography Electron beam tomography EBT is a specific form of CT in which a large enough X ray tube is constructed so that only the path of the electrons travelling between the cathode and anode of the X ray tube are spun using deflection coils 10 This type had a major advantage since sweep speeds can be much faster allowing for less blurry imaging of moving structures such as the heart and arteries 11 Fewer scanners of this design have been produced when compared with spinning tube types mainly due to the higher cost associated with building a much larger X ray tube and detector array and limited anatomical coverage 12 Dual Energy CT edit Dual Energy CT also known as Spectral CT is an advancement of Computed Tomography in which two energies are used to create two sets of data 13 A Dual Energy CT may employ Dual source Single source with dual detector layer Single source with energy switching methods to get two different sets of data 14 Dual source CT is an advanced scanner with a two X ray tube detector system unlike conventional single tube systems 15 16 These two detector systems are mounted on a single gantry at 90 in the same plane 17 Dual Source CT scanners allow fast scanning with higher temporal resolution by acquiring a full CT slice in only half a rotation Fast imaging reduces motion blurring at high heart rates and potentially allowing for shorter breath hold time This is particularly useful for ill patients having difficulty holding their breath or unable to take heart rate lowering medication 17 18 Single Source with Energy switching is another mode of Dual energy CT in which a single tube is operated at two different energies by switching the energies frequently 19 20 CT perfusion imaging edit Main article CT Perfusion nbsp CT perfusion imaging is a specific form of CT to assess flow through blood vessels whilst injecting a contrast agent 21 Blood flow blood transit time and organ blood volume can all be calculated with reasonable sensitivity and specificity 21 This type of CT may be used on the heart although sensitivity and specificity for detecting abnormalities are still lower than for other forms of CT 22 This may also be used on the brain where CT perfusion imaging can often detect poor brain perfusion well before it is detected using a conventional spiral CT scan 21 23 This is better for stroke diagnosis than other CT types 23 PET CT edit Main article PET CT nbsp PET CT scan of chestPositron emission tomography computed tomography is a hybrid CT modality which combines in a single gantry a positron emission tomography PET scanner and an x ray computed tomography CT scanner to acquire sequential images from both devices in the same session which are combined into a single superposed co registered image Thus functional imaging obtained by PET which depicts the spatial distribution of metabolic or biochemical activity in the body can be more precisely aligned or correlated with anatomic imaging obtained by CT scanning 24 PET CT gives both anatomical and functional details of an organ under examination and is helpful in detecting different type of cancers 25 26 Medical use editSince its introduction in the 1970s 27 CT has become an important tool in medical imaging to supplement conventional X ray imaging and medical ultrasonography It has more recently been used for preventive medicine or screening for disease for example CT colonography for people with a high risk of colon cancer or full motion heart scans for people with a high risk of heart disease Several institutions offer full body scans for the general population although this practice goes against the advice and official position of many professional organizations in the field primarily due to the radiation dose applied 28 The use of CT scans has increased dramatically over the last two decades in many countries 29 An estimated 72 million scans were performed in the United States in 2007 and more than 80 million in 2015 30 31 Head edit Main article Computed tomography of the head nbsp Computed tomography of human brain from base of the skull to top Taken with intravenous contrast medium Commons Scrollable computed tomography images of a normal brainCT scanning of the head is typically used to detect infarction stroke tumors calcifications haemorrhage and bone trauma 32 Of the above hypodense dark structures can indicate edema and infarction hyperdense bright structures indicate calcifications and haemorrhage and bone trauma can be seen as disjunction in bone windows Tumors can be detected by the swelling and anatomical distortion they cause or by surrounding edema CT scanning of the head is also used in CT guided stereotactic surgery and radiosurgery for treatment of intracranial tumors arteriovenous malformations and other surgically treatable conditions using a device known as the N localizer 33 34 35 36 37 38 Neck edit Contrast CT is generally the initial study of choice for neck masses in adults 39 CT of the thyroid plays an important role in the evaluation of thyroid cancer 40 CT scan often incidentally finds thyroid abnormalities and so is often the preferred investigation modality for thyroid abnormalities 40 Lungs edit Main article Computed tomography of the chest A CT scan can be used for detecting both acute and chronic changes in the lung parenchyma the tissue of the lungs 41 It is particularly relevant here because normal two dimensional X rays do not show such defects A variety of techniques are used depending on the suspected abnormality For evaluation of chronic interstitial processes such as emphysema and fibrosis 42 thin sections with high spatial frequency reconstructions are used often scans are performed both on inspiration and expiration This special technique is called high resolution CT that produces a sampling of the lung and not continuous images 43 nbsp HRCT images of a normal thorax in axial coronal and sagittal planes respectively Click here to scroll through the image stacks nbsp Bronchial wall thickness T and diameter of the bronchus D Bronchial wall thickening can be seen on lung CTs and generally but not always implies inflammation of the bronchi 44 An incidentally found nodule in the absence of symptoms sometimes referred to as an incidentaloma may raise concerns that it might represent a tumor either benign or malignant 45 Perhaps persuaded by fear patients and doctors sometimes agree to an intensive schedule of CT scans sometimes up to every three months and beyond the recommended guidelines in an attempt to do surveillance on the nodules 46 However established guidelines advise that patients without a prior history of cancer and whose solid nodules have not grown over a two year period are unlikely to have any malignant cancer 46 For this reason and because no research provides supporting evidence that intensive surveillance gives better outcomes and because of risks associated with having CT scans patients should not receive CT screening in excess of those recommended by established guidelines 46 Angiography edit nbsp Example of a CTPA demonstrating a saddle embolus dark horizontal line occluding the pulmonary arteries bright white triangle Main article Computed tomography angiography Computed tomography angiography CTA is a type of contrast CT to visualize the arteries and veins throughout the body 47 This ranges from arteries serving the brain to those bringing blood to the lungs kidneys arms and legs An example of this type of exam is CT pulmonary angiogram CTPA used to diagnose pulmonary embolism PE It employs computed tomography and an iodine based contrast agent to obtain an image of the pulmonary arteries 48 49 50 Cardiac edit A CT scan of the heart is performed to gain knowledge about cardiac or coronary anatomy 51 Traditionally cardiac CT scans are used to detect diagnose or follow up coronary artery disease 52 More recently CT has played a key role in the fast evolving field of transcatheter structural heart interventions more specifically in the transcatheter repair and replacement of heart valves 53 54 55 The main forms of cardiac CT scanning are Coronary CT angiography CCTA the use of CT to assess the coronary arteries of the heart The subject receives an intravenous injection of radiocontrast and then the heart is scanned using a high speed CT scanner allowing radiologists to assess the extent of occlusion in the coronary arteries usually to diagnose coronary artery disease 56 57 Coronary CT calcium scan also used for the assessment of severity of coronary artery disease Specifically it looks for calcium deposits in the coronary arteries that can narrow arteries and increase the risk of a heart attack 58 A typical coronary CT calcium scan is done without the use of radiocontrast but it can possibly be done from contrast enhanced images as well 59 To better visualize the anatomy post processing of the images is common 52 Most common are multiplanar reconstructions MPR and volume rendering For more complex anatomies and procedures such as heart valve interventions a true 3D reconstruction or a 3D print is created based on these CT images to gain a deeper understanding 60 61 62 63 Abdomen and pelvis edit nbsp CT scan of a normal abdomen and pelvis in sagittal plane coronal and axial planes respectively Click here to scroll through the image stacks Main article Computed tomography of the abdomen and pelvis CT is an accurate technique for diagnosis of abdominal diseases like Crohn s disease 64 GIT bleeding and diagnosis and staging of cancer as well as follow up after cancer treatment to assess response 65 It is commonly used to investigate acute abdominal pain 66 Non enhanced computed tomography is today the gold standard for diagnosing urinary stones 67 The size volume and density of stones can be estimated to help clinicians guide further treatment size is especially important in predicting spontaneous passage of a stone 68 Axial skeleton and extremities edit For the axial skeleton and extremities CT is often used to image complex fractures especially ones around joints because of its ability to reconstruct the area of interest in multiple planes Fractures ligamentous injuries and dislocations can easily be recognized with a 0 2 mm resolution 69 70 With modern dual energy CT scanners new areas of use have been established such as aiding in the diagnosis of gout 71 Biomechanical use edit CT is used in biomechanics to quickly reveal the geometry anatomy density and elastic moduli of biological tissues 72 73 Other uses editIndustrial use edit Industrial CT scanning industrial computed tomography is a process which utilizes X ray equipment to produce 3D representations of components both externally and internally Industrial CT scanning has been utilized in many areas of industry for internal inspection of components Some of the key uses for CT scanning have been flaw detection failure analysis metrology assembly analysis image based finite element methods 74 and reverse engineering applications CT scanning is also employed in the imaging and conservation of museum artifacts 75 Aviation security edit CT scanning has also found an application in transport security predominantly airport security where it is currently used in a materials analysis context for explosives detection CTX explosive detection device 76 77 78 79 and is also under consideration for automated baggage parcel security scanning using computer vision based object recognition algorithms that target the detection of specific threat items based on 3D appearance e g guns knives liquid containers 80 81 82 Its usage in airport security pioneered at Shannon Airport in March 2022 has ended the ban on liquids over 100 ml there a move that Heathrow Airport plans for a full roll out on 1 December 2022 and the TSA spent 781 2 million on an order for over 1 000 scanners ready to go live in the summer 83 Geological use edit X ray CT is used in geological studies to quickly reveal materials inside a drill core 84 Dense minerals such as pyrite and barite appear brighter and less dense components such as clay appear dull in CT images 85 Cultural heritage use edit X ray CT and micro CT can also be used for the conservation and preservation of objects of cultural heritage For many fragile objects direct research and observation can be damaging and can degrade the object over time Using CT scans conservators and researchers are able to determine the material composition of the objects they are exploring such as the position of ink along the layers of a scroll without any additional harm These scans have been optimal for research focused on the workings of the Antikythera mechanism or the text hidden inside the charred outer layers of the En Gedi Scroll However they are not optimal for every object subject to these kinds of research questions as there are certain artifacts like the Herculaneum papyri in which the material composition has very little variation along the inside of the object After scanning these objects computational methods can be employed to examine the insides of these objects as was the case with the virtual unwrapping of the En Gedi scroll and the Herculaneum papyri 86 Micro CT has also proved useful for analyzing more recent artifacts such as still sealed historic correspondence that employed the technique of letterlocking complex folding and cuts that provided a tamper evident locking mechanism 87 88 Further examples of use cases in archaeology is imaging the contents of sarcophagi or ceramics 89 Recently CWI in Amsterdam has collaborated with Rijksmuseum to investigate art object inside details in the framework called IntACT 90 Micro organism research edit Varied types of fungus can degrade wood to different degrees one Belgium research group has been used X ray CT 3 dimension with sub micron resolution unveiled fungi can penetrate micropores of 0 6 mm 91 under certain conditions Timber sawmill edit Sawmills use industrial CT scanners to detect round defects for instance knots to improve total value of timber productions Most sawmills are planning to incorporate this robust detection tool to improve productivity in the long run however initial investment cost is considerably high Microtec is one log scanner manufacturer with headquarters in Italy Interpretation of results editPresentation edit nbsp Types of presentations of CT scans Average intensity projection Maximum intensity projection Thin slice median plane Volume rendering by high and low threshold for radiodensityThe result of a CT scan is a volume of voxels which may be presented to a human observer by various methods which broadly fit into the following categories Slices of varying thickness Thin slice is generally regarded as planes representing a thickness of less than 3 mm 92 93 Thick slice is generally regarded as planes representing a thickness between 3 mm and 5 mm 93 94 Projection including maximum intensity projection 95 and average intensity projection Volume rendering VR 95 Technically all volume renderings become projections when viewed on a 2 dimensional display making the distinction between projections and volume renderings a bit vague The epitomes of volume rendering models feature a mix of for example coloring and shading in order to create realistic and observable representations 96 97 Two dimensional CT images are conventionally rendered so that the view is as though looking up at it from the patient s feet 98 Hence the left side of the image is to the patient s right and vice versa while anterior in the image also is the patient s anterior and vice versa This left right interchange corresponds to the view that physicians generally have in reality when positioned in front of patients 99 Grayscale edit Pixels in an image obtained by CT scanning are displayed in terms of relative radiodensity The pixel itself is displayed according to the mean attenuation of the tissue s that it corresponds to on a scale from 3 071 most attenuating to 1 024 least attenuating on the Hounsfield scale A pixel is a two dimensional unit based on the matrix size and the field of view When the CT slice thickness is also factored in the unit is known as a voxel which is a three dimensional unit 100 Water has an attenuation of 0 Hounsfield units HU while air is 1 000 HU cancellous bone is typically 400 HU and cranial bone can reach 2 000 HU 101 The attenuation of metallic implants depends on the atomic number of the element used Titanium usually has an amount of 1000 HU iron steel can completely block the X ray and is therefore responsible for well known line artifacts in computed tomograms Artifacts are caused by abrupt transitions between low and high density materials which results in data values that exceed the dynamic range of the processing electronics 102 Windowing edit CT data sets have a very high dynamic range which must be reduced for display or printing This is typically done via a process of windowing which maps a range the window of pixel values to a grayscale ramp For example CT images of the brain are commonly viewed with a window extending from 0 HU to 80 HU Pixel values of 0 and lower are displayed as black values of 80 and higher are displayed as white values within the window are displayed as a gray intensity proportional to position within the window 103 The window used for display must be matched to the X ray density of the object of interest in order to optimize the visible detail 104 Window width and window level parameters are used to control the windowing of a scan 105 Multiplanar reconstruction and projections edit nbsp Typical screen layout for diagnostic software showing one volume rendering VR and multiplanar view of three thin slices in the axial upper right sagittal lower left and coronal planes lower right nbsp Special planes are sometimes useful such as this oblique longitudinal plane in order to visualize the neuroforamina of the vertebral column showing narrowing at two levels causing radiculopathy The smaller images are axial plane slices Multiplanar reconstruction MPR is the process of converting data from one anatomical plane usually transverse to other planes It can be used for thin slices as well as projections Multiplanar reconstruction is possible as present CT scanners provide almost isotropic resolution 106 MPR is used almost in every scan The spine is frequently examined with it 107 An image of the spine in axial plane can only show one vertebral bone at a time and cannot show its relation with other vertebral bones By reformatting the data in other planes visualization of the relative position can be achieved in sagittal and coronal plane 108 New software allows the reconstruction of data in non orthogonal oblique planes which help in the visualization of organs which are not in orthogonal planes 109 110 It is better suited for visualization of the anatomical structure of the bronchi as they do not lie orthogonal to the direction of the scan 111 Curved plane reconstruction or curved planar reformation CPR is performed mainly for the evaluation of vessels This type of reconstruction helps to straighten the bends in a vessel thereby helping to visualize a whole vessel in a single image or in multiple images After a vessel has been straightened measurements such as cross sectional area and length can be made This is helpful in preoperative assessment of a surgical procedure 112 For 2D projections used in radiation therapy for quality assurance and planning of external beam radiotherapy including digitally reconstructed radiographs see Beam s eye view Examples of different algorithms of thickening multiplanar reconstructions 113 Type of projection Schematic illustration Examples 10 mm slabs Description UsesAverage intensity projection AIP nbsp nbsp The average attenuation of each voxel is displayed The image will get smoother as slice thickness increases It will look more and more similar to conventional projectional radiography as slice thickness increases Useful for identifying the internal structures of a solid organ or the walls of hollow structures such as intestines Maximum intensity projection MIP nbsp nbsp The voxel with the highest attenuation is displayed Therefore high attenuating structures such as blood vessels filled with contrast media are enhanced Useful for angiographic studies and identification of pulmonary nodules Minimum intensity projection MinIP nbsp nbsp The voxel with the lowest attenuation is displayed Therefore low attenuating structures such as air spaces are enhanced Useful for assessing the lung parenchyma Volume rendering edit Main article Volume rendering nbsp 3D human skull from computed tomography dataA threshold value of radiodensity is set by the operator e g a level that corresponds to bone With the help of edge detection image processing algorithms a 3D model can be constructed from the initial data and displayed on screen Various thresholds can be used to get multiple models each anatomical component such as muscle bone and cartilage can be differentiated on the basis of different colours given to them However this mode of operation cannot show interior structures 114 Surface rendering is limited technique as it displays only the surfaces that meet a particular threshold density and which are towards the viewer However In volume rendering transparency colours and shading are used which makes it easy to present a volume in a single image For example Pelvic bones could be displayed as semi transparent so that even viewing at an oblique angle one part of the image does not hide another 115 Image quality edit source source source source source source source Low dose CT scan of the thorax source source source source source source source source Standard dose CT scan of the thoraxDose versus image quality edit An important issue within radiology today is how to reduce the radiation dose during CT examinations without compromising the image quality In general higher radiation doses result in higher resolution images 116 while lower doses lead to increased image noise and unsharp images However increased dosage raises the adverse side effects including the risk of radiation induced cancer a four phase abdominal CT gives the same radiation dose as 300 chest X rays 117 Several methods that can reduce the exposure to ionizing radiation during a CT scan exist 118 New software technology can significantly reduce the required radiation dose New iterative tomographic reconstruction algorithms e g iterative Sparse Asymptotic Minimum Variance could offer super resolution without requiring higher radiation dose 119 Individualize the examination and adjust the radiation dose to the body type and body organ examined Different body types and organs require different amounts of radiation 120 Higher resolution is not always suitable such as detection of small pulmonary masses 121 Artifacts edit Although images produced by CT are generally faithful representations of the scanned volume the technique is susceptible to a number of artifacts such as the following 122 123 Chapters 3 and 5 Streak artifact Streaks are often seen around materials that block most X rays such as metal or bone Numerous factors contribute to these streaks under sampling photon starvation motion beam hardening and Compton scatter This type of artifact commonly occurs in the posterior fossa of the brain or if there are metal implants The streaks can be reduced using newer reconstruction techniques 124 Approaches such as metal artifact reduction MAR can also reduce this artifact 125 126 MAR techniques include spectral imaging where CT images are taken with photons of different energy levels and then synthesized into monochromatic images with special software such as GSI Gemstone Spectral Imaging 127 Partial volume effect This appears as blurring of edges It is due to the scanner being unable to differentiate between a small amount of high density material e g bone and a larger amount of lower density e g cartilage 128 The reconstruction assumes that the X ray attenuation within each voxel is homogeneous this may not be the case at sharp edges This is most commonly seen in the z direction craniocaudal direction due to the conventional use of highly anisotropic voxels which have a much lower out of plane resolution than in plane resolution This can be partially overcome by scanning using thinner slices or an isotropic acquisition on a modern scanner 129 Ring artifact Probably the most common mechanical artifact the image of one or many rings appears within an image They are usually caused by the variations in the response from individual elements in a two dimensional X ray detector due to defect or miscalibration 130 Ring artifacts can largely be reduced by intensity normalization also referred to as flat field correction 131 Remaining rings can be suppressed by a transformation to polar space where they become linear stripes 130 A comparative evaluation of ring artefact reduction on X ray tomography images showed that the method of Sijbers and Postnov can effectively suppress ring artefacts 132 Noise This appears as grain on the image and is caused by a low signal to noise ratio This occurs more commonly when a thin slice thickness is used It can also occur when the power supplied to the X ray tube is insufficient to penetrate the anatomy 133 Windmill Streaking appearances can occur when the detectors intersect the reconstruction plane This can be reduced with filters or a reduction in pitch 134 135 Beam hardening This can give a cupped appearance when grayscale is visualized as height It occurs because conventional sources like X ray tubes emit a polychromatic spectrum Photons of higher photon energy levels are typically attenuated less Because of this the mean energy of the spectrum increases when passing the object often described as getting harder This leads to an effect increasingly underestimating material thickness if not corrected Many algorithms exist to correct for this artifact They can be divided into mono and multi material methods 124 136 137 Advantages editCT scanning has several advantages over traditional two dimensional medical radiography First CT eliminates the superimposition of images of structures outside the area of interest 138 Second CT scans have greater image resolution enabling examination of finer details CT can distinguish between tissues that differ in radiographic density by 1 or less 139 Third CT scanning enables multiplanar reformatted imaging scan data can be visualized in the transverse or axial coronal or sagittal plane depending on the diagnostic task 140 The improved resolution of CT has permitted the development of new investigations For example CT angiography avoids the invasive insertion of a catheter CT scanning can perform a virtual colonoscopy with greater accuracy and less discomfort for the patient than a traditional colonoscopy 141 142 Virtual colonography is far more accurate than a barium enema for detection of tumors and uses a lower radiation dose 143 CT is a moderate to high radiation diagnostic technique The radiation dose for a particular examination depends on multiple factors volume scanned patient build number and type of scan protocol and desired resolution and image quality 144 Two helical CT scanning parameters tube current and pitch can be adjusted easily and have a profound effect on radiation CT scanning is more accurate than two dimensional radiographs in evaluating anterior interbody fusion although they may still over read the extent of fusion 145 Adverse effects editCancer edit Main article Radiation induced cancer The radiation used in CT scans can damage body cells including DNA molecules which can lead to radiation induced cancer 146 The radiation doses received from CT scans is variable Compared to the lowest dose x ray techniques CT scans can have 100 to 1 000 times higher dose than conventional X rays 147 However a lumbar spine x ray has a similar dose as a head CT 148 Articles in the media often exaggerate the relative dose of CT by comparing the lowest dose x ray techniques chest x ray with the highest dose CT techniques In general a routine abdominal CT has a radiation dose similar to three years of average background radiation 149 Large scale population based studies have consistently demonstrated that low dose radiation from CT scans has impacts on cancer incidence in a variety of cancers 150 151 152 153 For example in a large population based cohort it was found that up to 4 of brain cancers were caused by CT scan radiation 154 Some experts project that in the future between three and five percent of all cancers would result from medical imaging 147 An Australian study of 10 9 million people reported that the increased incidence of cancer after CT scan exposure in this cohort was mostly due to irradiation In this group one in every 1 800 CT scans was followed by an excess cancer If the lifetime risk of developing cancer is 40 then the absolute risk rises to 40 05 after a CT The risks of CT scan radiation are especially important in patients undergoing recurrent CT scans within a short time span of 1 to 5 years 155 156 157 Some experts note that CT scans are known to be overused and there is distressingly little evidence of better health outcomes associated with the current high rate of scans 147 On the other hand a recent paper analyzing the data of patients who received high cumulative doses showed a high degree of appropriate use 158 This creates an important issue of cancer risk to these patients Moreover a highly significant finding that was previously unreported is that some patients received gt 100 mSv dose from CT scans in a single day 156 which counteracts existing criticisms some investigators may have on the effects of protracted versus acute exposure There are contrarian views and the debate ongoing Some studies have shown that publications indicating an increased risk of cancer from typical doses of body CT scans are plagued with serious methodological limitations and several highly improbable results 159 concluding that no evidence indicates such low doses cause any long term harm 160 161 162 One study estimated that as many as 0 4 of cancers in the United States resulted from CT scans and that this may have increased to as much as 1 5 to 2 based on the rate of CT use in 2007 146 Others dispute this estimate 163 as there is no consensus that the low levels of radiation used in CT scans cause damage Lower radiation doses are used in many cases such as in the investigation of renal colic 164 A person s age plays a significant role in the subsequent risk of cancer 165 Estimated lifetime cancer mortality risks from an abdominal CT of a one year old is 0 1 or 1 1000 scans 165 The risk for someone who is 40 years old is half that of someone who is 20 years old with substantially less risk in the elderly 165 The International Commission on Radiological Protection estimates that the risk to a fetus being exposed to 10 mGy a unit of radiation exposure increases the rate of cancer before 20 years of age from 0 03 to 0 04 for reference a CT pulmonary angiogram exposes a fetus to 4 mGy 166 A 2012 review did not find an association between medical radiation and cancer risk in children noting however the existence of limitations in the evidences over which the review is based 167 CT scans can be performed with different settings for lower exposure in children with most manufacturers of CT scans as of 2007 having this function built in 168 Furthermore certain conditions can require children to be exposed to multiple CT scans 146 Current recommendations are to inform patients of the risks of CT scanning 169 However employees of imaging centers tend not to communicate such risks unless patients ask 170 Contrast reactions edit Further information Iodinated contrast Adverse effects In the United States half of CT scans are contrast CTs using intravenously injected radiocontrast agents 171 The most common reactions from these agents are mild including nausea vomiting and an itching rash Severe life threatening reactions may rarely occur 172 Overall reactions occur in 1 to 3 with nonionic contrast and 4 to 12 of people with ionic contrast 173 Skin rashes may appear within a week to 3 of people 172 The old radiocontrast agents caused anaphylaxis in 1 of cases while the newer low osmolar agents cause reactions in 0 01 0 04 of cases 172 174 Death occurs in about 2 to 30 people per 1 000 000 administrations with newer agents being safer 173 175 There is a higher risk of mortality in those who are female elderly or in poor health usually secondary to either anaphylaxis or acute kidney injury 171 The contrast agent may induce contrast induced nephropathy 176 This occurs in 2 to 7 of people who receive these agents with greater risk in those who have preexisting kidney failure 176 preexisting diabetes or reduced intravascular volume People with mild kidney impairment are usually advised to ensure full hydration for several hours before and after the injection For moderate kidney failure the use of iodinated contrast should be avoided this may mean using an alternative technique instead of CT Those with severe kidney failure requiring dialysis require less strict precautions as their kidneys have so little function remaining that any further damage would not be noticeable and the dialysis will remove the contrast agent it is normally recommended however to arrange dialysis as soon as possible following contrast administration to minimize any adverse effects of the contrast In addition to the use of intravenous contrast orally administered contrast agents are frequently used when examining the abdomen 177 These are frequently the same as the intravenous contrast agents merely diluted to approximately 10 of the concentration However oral alternatives to iodinated contrast exist such as very dilute 0 5 1 w v barium sulfate suspensions Dilute barium sulfate has the advantage that it does not cause allergic type reactions or kidney failure but cannot be used in patients with suspected bowel perforation or suspected bowel injury as leakage of barium sulfate from damaged bowel can cause fatal peritonitis 178 Side effects from contrast agents administered intravenously in some CT scans might impair kidney performance in patients with kidney disease although this risk is now believed to be lower than previously thought 179 176 Scan dose edit Examination Typical effective dose mSv to the whole body Typical absorbed dose mGy to the organ in questionAnnual background radiation 2 4 180 2 4 180 Chest X ray 0 02 181 0 01 0 15 182 Head CT 1 2 165 56 183 Screening mammography 0 4 166 3 146 182 Abdominal CT 8 181 14 183 Chest CT 5 7 165 13 183 CT colonography 6 11 165 Chest abdomen and pelvis CT 9 9 183 12 183 Cardiac CT angiogram 9 12 165 40 100 182 Barium enema 15 146 15 182 Neonatal abdominal CT 20 146 20 182 Further information Template Effective dose by medical imaging typeThe table reports average radiation exposures however there can be a wide variation in radiation doses between similar scan types where the highest dose could be as much as 22 times higher than the lowest dose 165 A typical plain film X ray involves radiation dose of 0 01 to 0 15 mGy while a typical CT can involve 10 20 mGy for specific organs and can go up to 80 mGy for certain specialized CT scans 182 For purposes of comparison the world average dose rate from naturally occurring sources of background radiation is 2 4 mSv per year equal for practical purposes in this application to 2 4 mGy per year 180 While there is some variation most people 99 received less than 7 mSv per year as background radiation 184 Medical imaging as of 2007 accounted for half of the radiation exposure of those in the United States with CT scans making up two thirds of this amount 165 In the United Kingdom it accounts for 15 of radiation exposure 166 The average radiation dose from medical sources is 0 6 mSv per person globally as of 2007 165 Those in the nuclear industry in the United States are limited to doses of 50 mSv a year and 100 mSv every 5 years 165 Lead is the main material used by radiography personnel for shielding against scattered X rays Radiation dose units edit The radiation dose reported in the gray or mGy unit is proportional to the amount of energy that the irradiated body part is expected to absorb and the physical effect such as DNA double strand breaks on the cells chemical bonds by X ray radiation is proportional to that energy 185 The sievert unit is used in the report of the effective dose The sievert unit in the context of CT scans does not correspond to the actual radiation dose that the scanned body part absorbs but to another radiation dose of another scenario the whole body absorbing the other radiation dose and the other radiation dose being of a magnitude estimated to have the same probability to induce cancer as the CT scan 186 Thus as is shown in the table above the actual radiation that is absorbed by a scanned body part is often much larger than the effective dose suggests A specific measure termed the computed tomography dose index CTDI is commonly used as an estimate of the radiation absorbed dose for tissue within the scan region and is automatically computed by medical CT scanners 187 The equivalent dose is the effective dose of a case in which the whole body would actually absorb the same radiation dose and the sievert unit is used in its report In the case of non uniform radiation or radiation given to only part of the body which is common for CT examinations using the local equivalent dose alone would overstate the biological risks to the entire organism 188 189 190 Effects of radiation edit Further information Radiobiology Most adverse health effects of radiation exposure may be grouped in two general categories deterministic effects harmful tissue reactions due in large part to the killing malfunction of cells following high doses 191 stochastic effects i e cancer and heritable effects involving either cancer development in exposed individuals owing to mutation of somatic cells or heritable disease in their offspring owing to mutation of reproductive germ cells 192 The added lifetime risk of developing cancer by a single abdominal CT of 8 mSv is estimated to be 0 05 or 1 one in 2 000 193 Because of increased susceptibility of fetuses to radiation exposure the radiation dosage of a CT scan is an important consideration in the choice of medical imaging in pregnancy 194 195 Excess doses edit In October 2009 the US Food and Drug Administration FDA initiated an investigation of brain perfusion CT PCT scans based on radiation burns caused by incorrect settings at one particular facility for this particular type of CT scan Over 256 patients were exposed to radiations for over 18 month period Over 40 of them lost patches of hair and prompted the editorial to call for increased CT quality assurance programs It was noted that while unnecessary radiation exposure should be avoided a medically needed CT scan obtained with appropriate acquisition parameter has benefits that outweigh the radiation risks 165 196 Similar problems have been reported at other centers 165 These incidents are believed to be due to human error 165 Procedure editCT scan procedure varies according to the type of the study and the organ being imaged The patient is made to lie on the CT table and the centering of the table is done according to the body part The IV line is established in case of Contrast enhanced CT After selecting proper and rate of contrast from the pressure injector the scout is taken to localize and plan the scan Once the plan is selected The contrast is given The raw data is processed according to the study and proper windowing is done to make scans easy to diagnose 197 Preparation edit Patient preparation may vary according to the type of scan The general patient preparation includes 197 Signing the informed consent Removal of metallic objects and jewelry from the region of interest Changing to the hospital gown according to hospital protocol Checking of KFT rate in case of CECT 198 Mechanism edit nbsp CT scanner with cover removed to show internal components Legend T X ray tube D X ray detectors X X ray beam R Gantry rotation nbsp Left image is a sinogram which is a graphic representation of the raw data obtained from a CT scan At right is an image sample derived from the raw data 199 Main article Operation of computed tomography Computed tomography operates by using an X ray generator that rotates around the object X ray detectors are positioned on the opposite side of the circle from the X ray source 200 As the X rays pass through the patient they are attenuated differently by various tissues according to the tissue density 201 A visual representation of the raw data obtained is called a sinogram yet it is not sufficient for interpretation 202 Once the scan data has been acquired the data must be processed using a form of tomographic reconstruction which produces a series of cross sectional images 203 These cross sectional images are made up of small units of pixels or voxels 204 Pixels in an image obtained by CT scanning are displayed in terms of relative radiodensity The pixel itself is displayed according to the mean attenuation of the tissue s that it corresponds to on a scale from 3 071 most attenuating to 1 024 least attenuating on the Hounsfield scale A pixel is a two dimensional unit based on the matrix size and the field of view When the CT slice thickness is also factored in the unit is known as a voxel which is a three dimensional unit 204 Water has an attenuation of 0 Hounsfield units HU while air is 1 000 HU cancellous bone is typically 400 HU and cranial bone can reach 2 000 HU or more os temporale and can cause artifacts The attenuation of metallic implants depends on the atomic number of the element used Titanium usually has an amount of 1000 HU iron steel can completely extinguish the X ray and is therefore responsible for well known line artifacts in computed tomograms Artifacts are caused by abrupt transitions between low and high density materials which results in data values that exceed the dynamic range of the processing electronics Two dimensional CT images are conventionally rendered so that the view is as though looking up at it from the patient s feet 98 Hence the left side of the image is to the patient s right and vice versa while anterior in the image also is the patient s anterior and vice versa This left right interchange corresponds to the view that physicians generally have in reality when positioned in front of patients Initially the images generated in CT scans were in the transverse axial anatomical plane perpendicular to the long axis of the body Modern scanners allow the scan data to be reformatted as images in other planes Digital geometry processing can generate a three dimensional image of an object inside the body from a series of two dimensional radiographic images taken by rotation around a fixed axis 122 These cross sectional images are widely used for medical diagnosis and therapy 205 Contrast edit Main article Contrast CT Contrast media used for X ray CT as well as for plain film X ray are called radiocontrasts Radiocontrasts for CT are in general iodine based 206 This is useful to highlight structures such as blood vessels that otherwise would be difficult to delineate from their surroundings Using contrast material can also help to obtain functional information about tissues Often images are taken both with and without radiocontrast 207 History editMain article History of computed tomography The history of X ray computed tomography goes back to at least 1917 with the mathematical theory of the Radon transform 208 209 In October 1963 William H Oldendorf received a U S patent for a radiant energy apparatus for investigating selected areas of interior objects obscured by dense material 210 The first commercially viable CT scanner was invented by Godfrey Hounsfield in 1972 211 It is often claimed that revenues from the sales of The Beatles records in the 1960s helped fund the development of the first CT scanner at EMI The first production X ray CT machine were in fact called EMI Scanner 212 Etymology edit The word tomography is derived from the Greek tome slice and graphein to write 213 Computed tomography was originally known as the EMI scan as it was developed in the early 1970s at a research branch of EMI a company best known today for its music and recording business 214 It was later known as computed axial tomography CAT or CT scan and body section rontgenography 215 The term CAT scan is no longer used because current CT scans enable for multiplanar reconstructions This makes CT scan the most appropriate term which is used by radiologists in common vernacular as well as in textbooks and scientific papers 216 217 218 In Medical Subject Headings MeSH computed axial tomography was used from 1977 to 1979 but the current indexing explicitly includes X ray in the title 219 The term sinogram was introduced by Paul Edholm and Bertil Jacobson in 1975 220 Society and culture editCampaigns edit In response to increased concern by the public and the ongoing progress of best practices the Alliance for Radiation Safety in Pediatric Imaging was formed within the Society for Pediatric Radiology In concert with the American Society of Radiologic Technologists the American College of Radiology and the American Association of Physicists in Medicine the Society for Pediatric Radiology developed and launched the Image Gently Campaign which is designed to maintain high quality imaging studies while using the lowest doses and best radiation safety practices available on pediatric patients 221 This initiative has been endorsed and applied by a growing list of various professional medical organizations around the world and has received support and assistance from companies that manufacture equipment used in Radiology Following upon the success of the Image Gently campaign the American College of Radiology the Radiological Society of North America the American Association of Physicists in Medicine and the American Society of Radiologic Technologists have launched a similar campaign to address this issue in the adult population called Image Wisely 222 The World Health Organization and International Atomic Energy Agency IAEA of the United Nations have also been working in this area and have ongoing projects designed to broaden best practices and lower patient radiation dose 223 224 Prevalence edit Number of CT scanners by country OECD as of 2017 225 per million population Country Value nbsp Japan 111 49 nbsp Australia 64 35 nbsp Iceland 43 68 nbsp United States 42 64 nbsp Denmark 39 72 nbsp Switzerland 39 28 nbsp Latvia 39 13 nbsp South Korea 38 18 nbsp Germany 35 13 nbsp Italy 34 71 nbsp Greece 34 22 nbsp Austria 28 64 nbsp Finland 24 51 nbsp Chile 24 27 nbsp Lithuania 23 33 nbsp Ireland 19 14 nbsp Spain 18 59 nbsp Estonia 18 22 nbsp France 17 36 nbsp Slovakia 17 28 nbsp Poland 16 88 nbsp Luxembourg 16 77 nbsp New Zealand 16 69 nbsp Czech Republic 15 76 nbsp Canada 15 28 nbsp Slovenia 15 00 nbsp Turkey 14 77 nbsp Netherlands 13 48 nbsp Russia 13 00 nbsp Israel 9 53 nbsp Hungary 9 19 nbsp Mexico 5 83 nbsp Colombia 1 24 Use of CT has increased dramatically over the last two decades 29 An estimated 72 million scans were performed in the United States in 2007 30 accounting for close to half of the total per capita dose rate from radiologic and nuclear medicine procedures 226 Of the CT scans six to eleven percent are done in children 166 an increase of seven to eightfold from 1980 165 Similar increases have been seen in Europe and Asia 165 In Calgary Canada 12 1 of people who present to the emergency with an urgent complaint received a CT scan most commonly either of the head or of the abdomen The percentage who received CT however varied markedly by the emergency physician who saw them from 1 8 to 25 227 In the emergency department in the United States CT or MRI imaging is done in 15 of people who present with injuries as of 2007 up from 6 in 1998 228 The increased use of CT scans has been the greatest in two fields screening of adults screening CT of the lung in smokers virtual colonoscopy CT cardiac screening and whole body CT in asymptomatic patients and CT imaging of children Shortening of the scanning time to around 1 second eliminating the strict need for the subject to remain still or be sedated is one of the main reasons for the large increase in the pediatric population especially for the diagnosis of appendicitis 146 As of 2007 in the United States a proportion of CT scans are performed unnecessarily 168 Some estimates place this number at 30 166 There are a number of reasons for this including legal concerns financial incentives and desire by the public 168 For example some healthy people avidly pay to receive full body CT scans as screening In that case it is not at all clear that the benefits outweigh the risks and costs Deciding whether and how to treat incidentalomas is complex radiation exposure is not negligible and the money for the scans involves opportunity cost 168 Manufacturers editMajor manufacturers of CT Scanners Devices and Equipment are 229 nbsp GE Healthcare nbsp Siemens Healthineers nbsp Canon Medical Systems Corporation formerly Toshiba Medical Systems nbsp Koninklijke Philips N V nbsp Fujifilm Healthcare formerly Hitachi Medical Systems nbsp Neusoft Medical Systems nbsp United Imaging HealthcareResearch editPhoton counting computed tomography is a CT technique currently under development Typical CT scanners use energy integrating detectors photons are measured as a voltage on a capacitor which is proportional to the x rays detected However this technique is susceptible to noise and other factors which can affect the linearity of the voltage to x ray intensity relationship 230 Photon counting detectors PCDs are still affected by noise but it does not change the measured counts of photons PCDs have several potential advantages including improving signal and contrast to noise ratios reducing doses improving spatial resolution and through use of several energies distinguishing multiple contrast agents 231 232 PCDs have only recently become feasible in CT scanners due to improvements in detector technologies that can cope with the volume and rate of data required As of February 2016 photon counting CT is in use at three sites 233 Some early research has found the dose reduction potential of photon counting CT for breast imaging to be very promising 234 In view of recent findings of high cumulative doses to patients from recurrent CT scans there has been a push for scanning technologies and techniques that reduce ionising radiation doses to patients to sub milliSievert sub mSv in the literature levels during the CT scan process a goal that has been lingering 235 156 157 158 See also editBarium sulfate suspension Dosimetry Tomosynthesis Virtopsy X ray microtomography Xenon enhanced CT scanningReferences edit CT scan Mayo Clinic mayoclinic org Archived from the original on 15 October 2016 Retrieved 20 October 2016 Hermena Shady Young Michael 2022 CT scan Image Production Procedures StatPearls Treasure Island FL StatPearls Publishing PMID 34662062 retrieved 2023 11 24 Patient Page ARRT The American Registry of Radiologic Technologists Archived from the original on 9 November 2014 Individual State Licensure Information American Society of Radiologic Technologists Archived from the original on 18 July 2013 Retrieved 19 July 2013 The Nobel Prize in Physiology or Medicine 1979 NobelPrize org Retrieved 2019 08 10 The Nobel Prize in Physiology or Medicine 1979 NobelPrize org Retrieved 2023 10 28 Terrier F Grossholz M Becker C D 2012 12 06 Spiral CT of the Abdomen Springer Science amp Business Media p 4 ISBN 978 3 642 56976 0 Fishman Elliot K Jeffrey R Brooke 1995 Spiral CT Principles Techniques and Clinical Applications Raven Press ISBN 978 0 7817 0218 8 Hsieh Jiang 2003 Computed Tomography Principles Design Artifacts and Recent Advances SPIE Press p 265 ISBN 978 0 8194 4425 7 Stirrup James 2020 01 02 Cardiovascular Computed Tomography Oxford University Press ISBN 978 0 19 880927 2 Talisetti Anita Jelnin Vladimir Ruiz Carlos John Eunice Benedetti Enrico Testa Giuliano Holterman Ai Xuan L Holterman Mark J December 2004 Electron beam CT scan is a valuable and safe imaging tool for the pediatric surgical patient Journal of Pediatric Surgery 39 12 1859 1862 doi 10 1016 j jpedsurg 2004 08 024 ISSN 1531 5037 PMID 15616951 Retsky Michael 31 July 2008 Electron beam computed tomography Challenges and opportunities Physics Procedia 1 1 149 154 Bibcode 2008PhPro 1 149R doi 10 1016 j phpro 2008 07 090 Johnson Thorsten Fink Christian Schonberg Stefan O Reiser Maximilian F 2011 01 18 Dual Energy CT in Clinical Practice Springer Science amp Business Media ISBN 978 3 642 01740 7 Johnson Thorsten Fink Christian Schonberg Stefan O Reiser Maximilian F 2011 01 18 Dual Energy CT in Clinical Practice Springer Science amp Business Media p 8 ISBN 978 3 642 01740 7 Carrascosa Patricia M Cury Ricardo C Garcia Mario J Leipsic Jonathon A 2015 10 03 Dual Energy CT in Cardiovascular Imaging Springer ISBN 978 3 319 21227 2 Schmidt Bernhard Flohr Thomas 2020 11 01 Principles and applications of dual source CT Physica Medica 125 Years of X Rays 79 36 46 doi 10 1016 j ejmp 2020 10 014 ISSN 1120 1797 PMID 33115699 S2CID 226056088 a b Seidensticker Peter R Hofmann Lars K 2008 05 24 Dual Source CT Imaging Springer Science amp Business Media ISBN 978 3 540 77602 4 Schmidt Bernhard Flohr Thomas 2020 11 01 Principles and applications of dual source CT Physica Medica European Journal of Medical Physics 79 36 46 doi 10 1016 j ejmp 2020 10 014 ISSN 1120 1797 PMID 33115699 S2CID 226056088 Mahmood Usman Horvat Natally Horvat Joao Vicente Ryan Davinia Gao Yiming Carollo Gabriella DeOcampo Rommel Do Richard K Katz Seth Gerst Scott Schmidtlein C Ross Dauer Lawrence Erdi Yusuf Mannelli Lorenzo May 2018 Rapid Switching kVp Dual Energy CT Value of Reconstructed Dual Energy CT Images and Organ Dose Assessment in Multiphasic Liver CT Exams European Journal of Radiology 102 102 108 doi 10 1016 j ejrad 2018 02 022 ISSN 0720 048X PMC 5918634 PMID 29685522 Johnson Thorsten R C November 2012 Dual Energy CT General Principles American Journal of Roentgenology 199 5 supplement S3 S8 doi 10 2214 AJR 12 9116 ISSN 0361 803X PMID 23097165 a b c Wittsack H J Wohlschlager A M Ritzl E K Kleiser R Cohnen M Seitz R J Modder U 2008 01 01 CT perfusion imaging of the human brain Advanced deconvolution analysis using circulant singular value decomposition Computerized Medical Imaging and Graphics 32 1 67 77 doi 10 1016 j compmedimag 2007 09 004 ISSN 0895 6111 PMID 18029143 Williams M C Newby D E 2016 08 01 CT myocardial perfusion imaging current status and future directions Clinical Radiology 71 8 739 749 doi 10 1016 j crad 2016 03 006 ISSN 0009 9260 PMID 27091433 a b Donahue Joseph Wintermark Max 2015 02 01 Perfusion CT and acute stroke imaging Foundations applications and literature review Journal of Neuroradiology 42 1 21 29 doi 10 1016 j neurad 2014 11 003 ISSN 0150 9861 PMID 25636991 Blodgett Todd M Meltzer Carolyn C Townsend David W February 2007 PET CT form and function Radiology 242 2 360 385 doi 10 1148 radiol 2422051113 ISSN 0033 8419 PMID 17255408 Ciernik I Frank Dizendorf Elena Baumert Brigitta G Reiner Beatrice Burger Cyrill Davis J Bernard Lutolf Urs M Steinert Hans C Von Schulthess Gustav K November 2003 Radiation treatment planning with an integrated positron emission and computer tomography PET CT a feasibility study International Journal of Radiation Oncology Biology Physics 57 3 853 863 doi 10 1016 s0360 3016 03 00346 8 ISSN 0360 3016 PMID 14529793 Ul Hassan Fahim Cook Gary J August 2012 PET CT in oncology Clinical Medicine 12 4 368 372 doi 10 7861 clinmedicine 12 4 368 ISSN 1470 2118 PMC 4952129 PMID 22930885 Curry Thomas S Dowdey James E Murry Robert C 1990 Christensen s Physics of Diagnostic Radiology Lippincott Williams amp Wilkins p 289 ISBN 978 0 8121 1310 5 CT Screening PDF hps org Archived from the original PDF on 13 October 2016 Retrieved 1 May 2018 a b Smith Bindman R Lipson J Marcus R Kim KP Mahesh M Gould R Berrington de Gonzalez A Miglioretti DL December 2009 Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer Arch Intern Med 169 22 2078 86 doi 10 1001 archinternmed 2009 427 PMC 4635397 PMID 20008690 a b Berrington de Gonzalez A Mahesh M Kim KP Bhargavan M Lewis R Mettler F Land C December 2009 Projected cancer risks from computed tomographic scans performed in the United States in 2007 Arch Intern Med 169 22 2071 7 doi 10 1001 archinternmed 2009 440 PMC 6276814 PMID 20008689 Dangers of CT Scans and X Rays Consumer Reports Retrieved 16 May 2018 Surgeons AAOS American Academy of Orthopaedic Physicians ACEP American College of Emergency UMBC 2017 03 20 Critical Care Transport Jones amp Bartlett Learning p 389 ISBN 978 1 284 04099 9 Galloway RL Jr 2015 Introduction and Historical Perspectives on Image Guided Surgery In Golby AJ ed Image Guided Neurosurgery Amsterdam Elsevier pp 3 4 ISBN 978 0 12 800870 6 Tse VCK Kalani MYS Adler JR 2015 Techniques of Stereotactic Localization In Chin LS Regine WF eds Principles and Practice of Stereotactic Radiosurgery New York Springer p 28 ISBN 978 0 387 71070 9 Saleh H Kassas B 2015 Developing Stereotactic Frames for Cranial Treatment In Benedict SH Schlesinger DJ Goetsch SJ Kavanagh BD eds Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy Boca Raton CRC Press pp 156 159 ISBN 978 1 4398 4198 3 Khan FR Henderson JM 2013 Deep Brain Stimulation Surgical Techniques In Lozano AM Hallet M eds Brain Stimulation Handbook of Clinical Neurology Vol 116 Amsterdam Elsevier pp 28 30 doi 10 1016 B978 0 444 53497 2 00003 6 ISBN 978 0 444 53497 2 PMID 24112882 Arle J 2009 Development of a Classic the Todd Wells Apparatus the BRW and the CRW Stereotactic Frames In Lozano AM Gildenberg PL Tasker RR eds Textbook of Stereotactic and Functional Neurosurgery Berlin Springer Verlag pp 456 461 ISBN 978 3 540 69959 0 Brown RA Nelson JA June 2012 Invention of the N localizer for stereotactic neurosurgery and its use in the Brown Roberts Wells stereotactic frame Neurosurgery 70 2 Supplement Operative 173 176 doi 10 1227 NEU 0b013e318246a4f7 PMID 22186842 S2CID 36350612 Daniel G Deschler Joseph Zenga Evaluation of a neck mass in adults UpToDate This topic last updated Dec 04 2017 a b Bin Saeedan Mnahi Aljohani Ibtisam Musallam Khushaim Ayman Omar Bukhari Salwa Qasim Elnaas Salahudin Tayeb 2016 Thyroid computed tomography imaging pictorial review of variable pathologies Insights into Imaging 7 4 601 617 doi 10 1007 s13244 016 0506 5 ISSN 1869 4101 PMC 4956631 PMID 27271508 Computed Tomography of the Lung Springer Berlin Heidelberg 2007 pp 40 47 ISBN 978 3 642 39518 5 High resolution CT of the Lung Lippincott Williams amp Wilkins 2009 pp 81 568 ISBN 978 0 7817 6909 9 Martinez Jimenez Santiago Rosado de Christenson Melissa L Carter Brett W 2017 07 22 Specialty Imaging HRCT of the Lung E Book Elsevier Health Sciences ISBN 978 0 323 52495 7 Yuranga Weerakkody Bronchial wall thickening Radiopaedia Archived from the original on 2018 01 06 Retrieved 2018 01 05 Wiener RS Gould MK Woloshin S Schwartz LM Clark JA 2012 What do you mean a spot A qualitative analysis of patients reactions to discussions with their doctors about pulmonary nodules Chest 143 3 672 677 doi 10 1378 chest 12 1095 PMC 3590883 PMID 22814873 a b c American College of Chest Physicians American Thoracic Society September 2013 Five Things Physicians and Patients Should Question Choosing Wisely American College of Chest Physicians and American Thoracic Society archived from the original on 3 November 2013 retrieved 6 January 2013 which cites MacMahon H Austin JH Gamsu G Herold CJ Jett JR Naidich DP Patz EF Swensen SJ 2005 Guidelines for Management of Small Pulmonary Nodules Detected on CT Scans A Statement from the Fleischner Society1 Radiology 237 2 395 400 doi 10 1148 radiol 2372041887 PMID 16244247 S2CID 14498160 Gould MK Fletcher J Iannettoni MD Lynch WR Midthun DE Naidich DP Ost DE 2007 Evaluation of Patients with Pulmonary Nodules When is It Lung Cancer Chest 132 3 suppl 108S 130S doi 10 1378 chest 07 1353 PMID 17873164 S2CID 16449420 Smith Bindman R Lipson J Marcus R Kim KP Mahesh M Gould R Berrington de Gonzalez A Miglioretti DL 2009 Radiation Dose Associated with Common Computed Tomography Examinations and the Associated Lifetime Attributable Risk of Cancer Archives of Internal Medicine 169 22 2078 2086 doi 10 1001 archinternmed 2009 427 PMC 4635397 PMID 20008690 Wiener RS Gould MK Woloshin S Schwartz LM Clark JA 2012 What do you mean a spot A qualitative analysis of patients reactions to discussions with their doctors about pulmonary nodules Chest 143 3 672 677 doi 10 1378 chest 12 1095 PMC 3590883 PMID 22814873 McDermott M Jacobs T Morgenstern L 2017 01 01 Wijdicks Eelco F M Kramer Andreas H eds Chapter 10 Critical care in acute ischemic stroke Handbook of Clinical Neurology Critical Care Neurology Part I Elsevier 140 153 176 doi 10 1016 b978 0 444 63600 3 00010 6 PMID 28187798 Computed Tomography Angiography CTA www hopkinsmedicine org 19 November 2019 Retrieved 2021 03 21 Zeman R K Silverman P M Vieco P T Costello P 1995 11 01 CT angiography American Journal of Roentgenology 165 5 1079 1088 doi 10 2214 ajr 165 5 7572481 ISSN 0361 803X PMID 7572481 Ramalho Joana Castillo Mauricio 2014 03 31 Vascular Imaging of the Central Nervous System Physical Principles Clinical Applications and Emerging Techniques John Wiley amp Sons p 69 ISBN 978 1 118 18875 0 Cardiac CT Scan NHLBI NIH www nhlbi nih gov Archived from the original on 2017 12 01 Retrieved 2017 11 22 a b Wichmann Julian L Cardiac CT Radiology Reference Article Radiopaedia org radiopaedia org Archived from the original on 2017 12 01 Retrieved 2017 11 22 Marwan Mohamed Achenbach Stephan February 2016 Role of Cardiac CT Before Transcatheter Aortic Valve Implantation TAVI Current Cardiology Reports 18 2 21 doi 10 1007 s11886 015 0696 3 ISSN 1534 3170 PMID 26820560 S2CID 41535442 Moss Alastair J Dweck Marc R Dreisbach John G Williams Michelle C Mak Sze Mun Cartlidge Timothy Nicol Edward D Morgan Hughes Gareth J 2016 11 01 Complementary role of cardiac CT in the assessment of aortic valve replacement dysfunction Open Heart 3 2 e000494 doi 10 1136 openhrt 2016 000494 ISSN 2053 3624 PMC 5093391 PMID 27843568 Theriault Lauzier Pascal Spaziano Marco Vaquerizo Beatriz Buithieu Jean Martucci Giuseppe Piazza Nicolo September 2015 Computed Tomography for Structural Heart Disease and Interventions Interventional Cardiology Review 10 3 149 154 doi 10 15420 ICR 2015 10 03 149 ISSN 1756 1477 PMC 5808729 PMID 29588693 Passariello Roberto 2006 03 30 Multidetector Row CT Angiography Springer Science amp Business Media ISBN 978 3 540 26984 7 Radiology ACR Radiological Society of North America RSNA and American College of Coronary Computed Tomography Angiography CCTA www radiologyinfo org Retrieved 2021 03 19 Heart scan coronary calcium scan Mayo Clinic Archived from the original on 5 September 2015 Retrieved 9 August 2015 van der Bijl Noortje Joemai Raoul M S Geleijns Jacob Bax Jeroen J Schuijf Joanne D de Roos Albert Kroft Lucia J M 2010 Assessment of Agatston Coronary Artery Calcium Score Using Contrast Enhanced CT Coronary Angiography American Journal of Roentgenology 195 6 1299 1305 doi 10 2214 AJR 09 3734 ISSN 0361 803X PMID 21098187 Vukicevic Marija Mosadegh Bobak Min James K Little Stephen H February 2017 Cardiac 3D Printing and its Future Directions JACC Cardiovascular Imaging 10 2 171 184 doi 10 1016 j jcmg 2016 12 001 ISSN 1876 7591 PMC 5664227 PMID 28183437 Wang D D Eng M Greenbaum A Myers E Forbes M Pantelic M Song T Nelson C Divine G Taylor A Wyman J Guerrero M Lederman R J Paone G O Neill W 2016 Innovative Mitral Valve Treatment with 3D Visualization at Henry Ford JACC Cardiovascular Imaging 9 11 1349 1352 doi 10 1016 j jcmg 2016 01 017 PMC 5106323 PMID 27209112 Archived from the original on 2017 12 01 Retrieved 2017 11 22 Wang Dee Dee Eng Marvin Greenbaum Adam Myers Eric Forbes Michael Pantelic Milan Song Thomas Nelson Christina Divine George November 2016 Predicting LVOT Obstruction After TMVR JACC Cardiovascular Imaging 9 11 1349 1352 doi 10 1016 j jcmg 2016 01 017 ISSN 1876 7591 PMC 5106323 PMID 27209112 Jacobs Stephan Grunert Ronny Mohr Friedrich W Falk Volkmar February 2008 3D Imaging of cardiac structures using 3D heart models for planning in heart surgery a preliminary study Interactive Cardiovascular and Thoracic Surgery 7 1 6 9 doi 10 1510 icvts 2007 156588 ISSN 1569 9285 PMID 17925319 Furukawa Akira Saotome Takao Yamasaki Michio Maeda Kiyosumi Nitta Norihisa Takahashi Masashi Tsujikawa Tomoyuki Fujiyama Yoshihide Murata Kiyoshi Sakamoto Tsutomu 2004 05 01 Cross sectional Imaging in Crohn Disease RadioGraphics 24 3 689 702 doi 10 1148 rg 243035120 ISSN 0271 5333 PMID 15143222 CT of the Acute Abdomen Springer Berlin Heidelberg 2011 p 37 ISBN 978 3 540 89232 8 Jay P Heiken Douglas S Katz 2014 Emergency Radiology of the Abdomen and Pelvis Imaging of the Nontraumatic and Traumatic Acute Abdomen In J Hodler R A Kubik Huch G K von Schulthess Ch L Zollikofer eds Diseases of the Abdomen and Pelvis Springer Milan p 3 ISBN 9788847056596 Skolarikos A Neisius A Petrik A Somani B Thomas K Gambaro G March 2022 EAU Guidelines on Urolithiasis Amsterdam European Association of Urology ISBN 978 94 92671 16 5 Miller Oren F Kane Christopher J September 1999 Time to stone passage for observed ureteral calculi a guide for patient education Journal of Urology 162 3 Part 1 688 691 doi 10 1097 00005392 199909010 00014 PMID 10458343 Ankle Fractures orthoinfo aaos org American Association of Orthopedic Surgeons Archived from the original on 30 May 2010 Retrieved 30 May 2010 Buckwalter Kenneth A et al 11 September 2000 Musculoskeletal Imaging with Multislice CT American Journal of Roentgenology 176 4 979 986 doi 10 2214 ajr 176 4 1760979 PMID 11264094 Ramon Andre Bohm Sigrand Amelie Pottecher Pierre Richette Pascal Maillefert Jean Francis Devilliers Herve Ornetti Paul 2018 03 01 Role of dual energy CT in the diagnosis and follow up of gout systematic analysis of the literature Clinical Rheumatology 37 3 587 595 doi 10 1007 s10067 017 3976 z ISSN 0770 3198 PMID 29350330 S2CID 3686099 Keaveny Tony M March 2010 Biomechanical computed tomography noninvasive bone strength analysis using clinical computed tomography scans Annals of the New York Academy of Sciences 1192 1 57 65 Bibcode 2010NYASA1192 57K doi 10 1111 j 1749 6632 2009 05348 x ISSN 1749 6632 PMID 20392218 S2CID 24132358 Barber Asa Tozzi Gianluca Pani Martino 2019 03 07 Computed Tomography Based Biomechanics Frontiers Media SA p 20 ISBN 978 2 88945 780 9 Evans Ll M Margetts L Casalegno V Lever L M Bushell J Lowe T Wallwork A Young P Lindemann A 2015 05 28 Transient thermal finite element analysis of CFC Cu ITER monoblock using X ray tomography data Fusion Engineering and Design 100 100 111 doi 10 1016 j fusengdes 2015 04 048 Archived from the original on 2015 10 16 Payne Emma Marie 2012 Imaging Techniques in Conservation PDF Journal of Conservation and Museum Studies 10 2 17 29 doi 10 5334 jcms 1021201 P Babaheidarian D Castanon 2018 Joint reconstruction and material classification in spectral CT In Greenberg Joel A Gehm Michael E Neifeld Mark A Ashok Amit eds Anomaly Detection and Imaging with X Rays ADIX III p 12 doi 10 1117 12 2309663 ISBN 978 1 5106 1775 9 S2CID 65469251 P Jin E Haneda K D Sauer C A Bouman June 2012 A model based 3D multi slice helical CT reconstruction algorithm for transportation security application PDF Second International Conference on Image Formation in X Ray Computed Tomography Archived from the original PDF on 2015 04 11 Retrieved 2015 04 05 P Jin E Haneda C A Bouman November 2012 Implicit Gibbs prior models for tomographic reconstruction PDF Signals Systems and Computers ASILOMAR 2012 Conference Record of the Forty Sixth Asilomar Conference on IEEE pp 613 636 Archived from the original PDF on 2015 04 11 Retrieved 2015 04 05 S J Kisner P Jin C A Bouman K D Sauer W Garms T Gable S Oh M Merzbacher S Skatter October 2013 Innovative data weighting for iterative reconstruction in a helical CT security baggage scanner PDF Security Technology ICCST 2013 47th International Carnahan Conference on IEEE Archived from the original PDF on 2015 04 10 Retrieved 2015 04 05 Megherbi N Flitton G T Breckon T P September 2010 A Classifier based Approach for the Detection of Potential Threats in CT based Baggage Screening PDF Proc International Conference on Image Processing IEEE pp 1833 1836 CiteSeerX 10 1 1 188 5206 doi 10 1109 ICIP 2010 5653676 ISBN 978 1 4244 7992 4 S2CID 3679917 Retrieved 5 November 2013 permanent dead link Megherbi N Han J Flitton G T Breckon T P September 2012 A Comparison of Classification Approaches for Threat Detection in CT based Baggage Screening PDF Proc International Conference on Image Processing IEEE pp 3109 3112 CiteSeerX 10 1 1 391 2695 doi 10 1109 ICIP 2012 6467558 ISBN 978 1 4673 2533 2 S2CID 6924816 Retrieved 5 November 2013 permanent dead link Flitton G T Breckon T P Megherbi N September 2013 A Comparison of 3D Interest Point Descriptors with Application to Airport Baggage Object Detection in Complex CT Imagery PDF Pattern Recognition 46 9 2420 2436 Bibcode 2013PatRe 46 2420F doi 10 1016 j patcog 2013 02 008 hdl 1826 15213 Retrieved 5 November 2013 permanent dead link Russell Molly 30 March 2022 How Next Generation Airport Scanners Are Ending The 100ml Liquid Rule Simple Flying Retrieved 11 April 2022 Laboratory About Chikyu The Deep sea Scientific Drilling Vessel CHIKYU www jamstec go jp Retrieved 2019 10 24 Tonai Satoshi Kubo Yusuke Tsang Man Yin Bowden Stephen Ide Kotaro Hirose Takehiro Kamiya Nana Yamamoto Yuzuru Yang Kiho Yamada Yasuhiro Morono Yuki 2019 A New Method for Quality Control of Geological Cores by X Ray Computed Tomography Application in IODP Expedition 370 Frontiers in Earth Science 7 doi 10 3389 feart 2019 00117 ISSN 2296 6463 S2CID 171394807 Seales W B Parker C S Segal M Tov E Shor P Porath Y 2016 From damage to discovery via virtual unwrapping Reading the scroll from En Gedi Science Advances 2 9 e1601247 Bibcode 2016SciA 2E1247S doi 10 1126 sciadv 1601247 ISSN 2375 2548 PMC 5031465 PMID 27679821 Castellanos Sara 2 March 2021 A Letter Sealed for Centuries Has Been Read Without Even Opening It The Wall Street Journal Retrieved 2 March 2021 Dambrogio Jana Ghassaei Amanda Staraza Smith Daniel Jackson Holly Demaine Martin L 2 March 2021 Unlocking history through automated virtual unfolding of sealed documents imaged by X ray microtomography Nature Communications 12 1 1184 Bibcode 2021NatCo 12 1184D doi 10 1038 s41467 021 21326 w PMC 7925573 PMID 33654094 Advanced documentation methods in studying Corinthian black figure vase painting on YouTube showing a Computed Tomography scan and rollout of the aryballos No G26 archaeological collection Graz University The video was rendered using the GigaMesh Software Framework cf doi 10 11588 heidok 00025189 Karl Stephan Bayer Paul Mara Hubert Marton Andras 2019 Advanced Documentation Methods in Studying Corinthian Black figure Vase Painting PDF Proceedings of the 23rd International Conference on Cultural Heritage and New Technologies CHNT23 Vienna Austria ISBN 978 3 200 06576 5 retrieved 2020 01 09 CT FOR ART NICAS Retrieved 2023 07 04 Bulcke Jan Van den Boone Matthieu Acker Joris Van Hoorebeke Luc Van October 2009 Three Dimensional X Ray Imaging and Analysis of Fungi on and in Wood Microscopy and Microanalysis 15 5 395 402 Bibcode 2009MiMic 15 395V doi 10 1017 S1431927609990419 hdl 1854 LU 675607 ISSN 1435 8115 PMID 19709462 S2CID 15637414 Goldman L W 2008 Principles of CT Multislice CT Journal of Nuclear Medicine Technology 36 2 57 68 doi 10 2967 jnmt 107 044826 ISSN 0091 4916 PMID 18483143 a b Reis Eduardo Pontes Nascimento Felipe Aranha Mateus Mainetti Secol Fernando Machado Birajara Felix Marcelo Stein Anouk Amaro Edson 29 July 2020 Brain Hemorrhage Extended BHX Bounding box extrapolation from thick to thin slice CT images v1 1 PhysioNet 101 23 215 220 doi 10 13026 9cft hg92 Park S Chu L C Hruban R H Vogelstein B Kinzler K W Yuille A L Fouladi D F Shayesteh S Ghandili S Wolfgang C L Burkhart R He J Fishman E K Kawamoto S 2020 09 01 Differentiating autoimmune pancreatitis from pancreatic ductal adenocarcinoma with CT radiomics features Diagnostic and Interventional Imaging 101 9 555 564 doi 10 1016 j diii 2020 03 002 ISSN 2211 5684 PMID 32278586 S2CID 215751181 a b Fishman Elliot K Ney Derek R Heath David G Corl Frank M Horton Karen M Johnson Pamela T 2006 Volume Rendering versus Maximum Intensity Projection in CT Angiography What Works Best When and Why RadioGraphics 26 3 905 922 doi 10 1148 rg 263055186 ISSN 0271 5333 PMID 16702462 Silverstein Jonathan C Parsad Nigel M Tsirline Victor 2008 Automatic perceptual color map generation for realistic volume visualization Journal of Biomedical Informatics 41 6 927 935 doi 10 1016 j jbi 2008 02 008 ISSN 1532 0464 PMC 2651027 PMID 18430609 Kobbelt Leif 2006 Vision Modeling and Visualization 2006 Proceedings November 22 24 2006 Aachen Germany IOS Press p 185 ISBN 978 3 89838 081 2 a b Computerized Tomography chapter Archived 2016 03 04 at the Wayback Machine at University of Connecticut Health Center Schmidt Derek Odland Rick September 2004 Mirror Image Reversal of Coronal Computed Tomography Scans The Laryngoscope 114 9 1562 1565 doi 10 1097 00005537 200409000 00011 ISSN 0023 852X PMID 15475782 S2CID 22320649 Brant and Helms fundamentals of diagnostic radiology Fifth ed Lippincott Williams amp Wilkins 2018 07 19 p 1600 ISBN 978 1 4963 6738 9 Retrieved 24 January 2019 Arthur W Toga John C Mazziotta eds 2002 Brain mapping the methods 2nd ed Amsterdam Academic Press ISBN 0 12 693019 8 OCLC 52594824 Jerrold T Bushberg J Anthony Seibert Edwin M Leidholdt John M Boone 2002 The essential physics of medical imaging 2nd ed Philadelphia Lippincott Williams amp Wilkins p 358 ISBN 0 683 30118 7 OCLC 47177732 Kamalian Shervin Lev Michael H Gupta Rajiv 2016 01 01 Computed tomography imaging and angiography principles Handbook of Clinical Neurology Vol 135 pp 3 20 doi 10 1016 B978 0 444 53485 9 00001 5 ISBN 978 0 444 53485 9 ISSN 0072 9752 PMID 27432657 Stirrup James 2020 01 02 Cardiovascular Computed Tomography Oxford University Press p 136 ISBN 978 0 19 880927 2 Carroll Quinn B 2007 Practical Radiographic Imaging Charles C Thomas Publisher p 512 ISBN 978 0 398 08511 7 Udupa Jayaram K Herman Gabor T 1999 09 28 3D Imaging in Medicine Second Edition CRC Press ISBN 978 0 8493 3179 4 Krupski Witold Kurys Denis Ewa Matuszewski Lukasz Plezia Boguslaw 2007 06 30 Use of multi planar reconstruction MPR and 3 dimentional sic 3D CT to assess stability criteria in C2 vertebral fractures Journal of Pre Clinical and Clinical Research 1 1 80 83 ISSN 1898 2395 Tins Bernhard 2010 10 21 Technical aspects of CT imaging of the spine Insights into Imaging 1 5 6 349 359 doi 10 1007 s13244 010 0047 2 ISSN 1869 4101 PMC 3259341 PMID 22347928 CT imaging Where are we going Proceedings DVM 360 April 2010 Retrieved 2021 03 21 Wolfson Nikolaj Lerner Alexander Roshal Leonid 2016 05 30 Orthopedics in Disasters Orthopedic Injuries in Natural Disasters and Mass Casualty Events Springer ISBN 978 3 662 48950 5 Laroia Archana T Thompson Brad H Laroia Sandeep T van Beek Edwin JR 2010 07 28 Modern imaging of the tracheo bronchial tree World Journal of Radiology 2 7 237 248 doi 10 4329 wjr v2 i7 237 ISSN 1949 8470 PMC 2998855 PMID 21160663 Gong Jing Shan Xu Jian Min 2004 07 01 Role of curved planar reformations using multidetector spiral CT in diagnosis of pancreatic and peripancreatic diseases World Journal of Gastroenterology 10 13 1943 1947 doi 10 3748 wjg v10 i13 1943 ISSN 1007 9327 PMC 4572236 PMID 15222042 Dalrymple Neal C Prasad Srinivasa R Freckleton Michael W Chintapalli Kedar N September 2005 Informatics in radiology infoRAD introduction to the language of three dimensional imaging with multidetector CT Radiographics 25 5 1409 1428 doi 10 1148 rg 255055044 ISSN 1527 1323 PMID 16160120 Calhoun Paul S Kuszyk Brian S Heath David G Carley Jennifer C Fishman Elliot K 1999 05 01 Three dimensional Volume Rendering of Spiral CT Data Theory and Method RadioGraphics 19 3 745 764 doi 10 1148 radiographics 19 3 g99ma14745 ISSN 0271 5333 PMID 10336201 van Ooijen P M A van Geuns R J M Rensing B J W M Bongaerts A H H de Feyter P J Oudkerk M January 2003 Noninvasive Coronary Imaging Using Electron Beam CT Surface Rendering Versus Volume Rendering American Journal of Roentgenology 180 1 223 226 doi 10 2214 ajr 180 1 1800223 ISSN 0361 803X PMID 12490509 R A Crowther D J DeRosier A Klug 1970 The Reconstruction of a Three Dimensional Structure from Projections and its Application to Electron Microscopy Proc R Soc Lond A 317 1530 319 340 Bibcode 1970RSPSA 317 319C doi 10 1098 rspa 1970 0119 S2CID 122980366 Nickoloff Edward L Alderson Philip O August 2001 Radiation Exposures to Patients from CT Reality Public Perception and Policy American Journal of Roentgenology 177 2 285 287 doi 10 2214 ajr 177 2 1770285 ISSN 0361 803X PMID 11461846 Barkan O Weill J Averbuch A Dekel S Adaptive Compressed Tomography Sensing Archived 2016 03 13 at the Wayback Machine In Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition 2013 pp 2195 2202 Proceedings IEEE 1995 p 10 ISBN 9780780324985 Radiation Effects on organs of the body somatic effects Encyclopedia Britannica Retrieved 2021 03 21 Simpson G 2009 Thoracic computed tomography principles and practice Australian Prescriber 32 4 4 doi 10 18773 austprescr 2009 049 a b Hsieh Jiang 2003 Computed Tomography Principles Design Artifacts and Recent Advances SPIE Press p 167 ISBN 978 0 8194 4425 7 Bhowmik Ujjal Kumar Zafar Iqbal M Adhami Reza R 28 May 2012 Mitigating motion artifacts in FDK based 3D Cone beam Brain Imaging System using markers Central European Journal of Engineering 2 3 369 382 Bibcode 2012CEJE 2 369B doi 10 2478 s13531 012 0011 7 a b P Jin C A Bouman K D Sauer 2013 A Method for Simultaneous Image Reconstruction and Beam Hardening Correction PDF IEEE Nuclear Science Symp amp Medical Imaging Conf Seoul Korea 2013 Archived from the original PDF on 2014 06 06 Retrieved 2014 04 23 Boas FE Fleischmann D 2011 Evaluation of Two Iterative Techniques for Reducing Metal Artifacts in Computed Tomography Radiology 259 3 894 902 doi 10 1148 radiol 11101782 PMID 21357521 Mouton A Megherbi N Van Slambrouck K Nuyts J Breckon T P 2013 An Experimental Survey of Metal Artefact Reduction in Computed Tomography PDF Journal of X Ray Science and Technology 21 2 193 226 doi 10 3233 XST 130372 hdl 1826 8204 PMID 23694911 permanent dead link Pessis Eric Campagna Raphael Sverzut Jean Michel Bach Fabienne Rodallec Mathieu Guerini Henri Feydy Antoine Drape Jean Luc 2013 Virtual Monochromatic Spectral Imaging with Fast Kilovoltage Switching Reduction of Metal Artifacts at CT RadioGraphics 33 2 573 583 doi 10 1148 rg 332125124 ISSN 0271 5333 PMID 23479714 Gonzalez Ballester Miguel Angel Zisserman Andrew P Brady Michael December 2002 Estimation of the partial volume effect in MRI Medical Image Analysis 6 4 389 405 doi 10 1016 s1361 8415 02 00061 0 ISSN 1361 8415 PMID 12494949 Goldszal Alberto F Pham Dzung L 2000 01 01 Volumetric Segmentation Handbook of Medical Imaging 185 194 doi 10 1016 B978 012077790 7 50016 3 ISBN 978 0 12 077790 7 a b Jha Diwaker 2014 Adaptive center determination for effective suppression of ring artifacts in tomography images Applied Physics Letters 105 14 143107 Bibcode 2014ApPhL 105n3107J doi 10 1063 1 4897441 Van Nieuwenhove V De Beenhouwer J De Carlo F Mancini L Marone F Sijbers J 2015 Dynamic intensity normalization using eigen flat fields in X ray imaging PDF Optics Express 23 21 27975 27989 Bibcode 2015OExpr 2327975V doi 10 1364 oe 23 027975 hdl 10067 1302930151162165141 PMID 26480456 Sijbers J Postnov A 2004 Reduction of ring artefacts in high resolution micro CT reconstructions Phys Med Biol 49 14 N247 53 doi 10 1088 0031 9155 49 14 N06 PMID 15357205 S2CID 12744174 Newton Thomas H Potts D Gordon 1971 Radiology of the Skull and Brain Technical aspects of computed tomography Mosby pp 3941 3950 ISBN 978 0 8016 3662 2 Bruning R Kuttner A Flohr T 2006 01 16 Protocols for Multislice CT Springer Science amp Business Media ISBN 978 3 540 27273 1 Peh Wilfred C G 2017 08 11 Pitfalls in Musculoskeletal Radiology Springer ISBN 978 3 319 53496 1 Van de Casteele E Van Dyck D Sijbers J Raman E 2004 A model based correction method for beam hardening artefacts in X ray microtomography Journal of X ray Science and Technology 12 1 43 57 CiteSeerX 10 1 1 460 6487 Van Gompel G Van Slambrouck K Defrise M Batenburg KJ Sijbers J Nuyts J 2011 Iterative correction of beam hardening artifacts in CT Medical Physics 38 1 36 49 Bibcode 2011MedPh 38S 36V CiteSeerX 10 1 1 464 3547 doi 10 1118 1 3577758 PMID 21978116 Mikla Victor I Mikla Victor V 2013 08 23 Medical Imaging Technology Elsevier p 37 ISBN 978 0 12 417036 0 Radiology for the Dental Professional Elsevier Mosby 2008 p 337 ISBN 978 0 323 03071 7 Pasipoularides Ares November 2009 Heart s Vortex Intracardiac Blood Flow Phenomena PMPH USA p 595 ISBN 978 1 60795 033 2 Heiken JP Peterson CM Menias CO November 2005 Virtual colonoscopy for colorectal cancer screening current status Wednesday 5 October 2005 14 00 16 00 Cancer Imaging International Cancer Imaging Society 5 Spec No A S133 S139 doi 10 1102 1470 7330 2005 0108 PMC 1665314 PMID 16361129 Bielen DJ Bosmans HT De Wever LL et al September 2005 Clinical validation of high resolution fast spin echo MR colonography after colon distention with air J Magn Reson Imaging 22 3 400 5 doi 10 1002 jmri 20397 PMID 16106357 S2CID 22167728 CT Colonography Radiologyinfo org Zabic S Wang Q Morton T Brown KM March 2013 A low dose simulation tool for CT systems with energy integrating detectors Medical Physics 40 3 031102 Bibcode 2013MedPh 40c1102Z doi 10 1118 1 4789628 PMID 23464282 Brian R Subach M D F A C S et al Reliability and accuracy of fine cut computed tomography scans to determine the status of anterior interbody fusions with metallic cages Archived 2012 12 08 at the Wayback Machine a b c d e f g Brenner DJ Hall EJ November 2007 Computed tomography an increasing source of radiation exposure PDF N Engl J Med 357 22 2277 84 doi 10 1056 NEJMra072149 PMID 18046031 S2CID 2760372 Archived PDF from the original on 2016 03 04 a b c Redberg Rita F and Smith Bindman Rebecca We Are Giving Ourselves Cancer Archived 2017 07 06 at the Wayback Machine New York Times January 30 2014 Health Center for Devices and Radiological Medical X ray Imaging What are the Radiation Risks from CT www fda gov Archived from the original on 5 November 2013 Retrieved 1 May 2018 ACR Radiological Society of North America RSNA and American College of Radiology February 2021 Patient Safety Radiation Dose in X Ray and CT Exams PDF acr org Archived from the original PDF on 1 January 2021 Retrieved 6 April 2021 Mathews J D Forsythe A V Brady Z Butler M W Goergen S K Byrnes G B Giles G G Wallace A B Anderson P R Guiver T A McGale P Cain T M Dowty J G Bickerstaffe A C Darby S C 2013 Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence data linkage study of 11 million Australians BMJ 346 may21 1 f2360 doi 10 1136 bmj f2360 ISSN 1756 1833 PMC 3660619 PMID 23694687 Pearce MS Salotti JA Little MP McHugh K Lee C Kim KP Howe NL Ronckers CM Rajaraman P Sir Craft AW Parker L Berrington de Gonzalez A 4 August 2012 Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours a retrospective cohort study Lancet 380 9840 499 505 doi 10 1016 S0140 6736 12 60815 0 PMC 3418594 PMID 22681860 Meulepas Johanna M Ronckers Cecile M Smets Anne M J B Nievelstein Rutger A J Gradowska Patrycja Lee Choonsik Jahnen Andreas van Straten Marcel de Wit Marie Claire Y Zonnenberg Bernard Klein Willemijn M Merks Johannes H Visser Otto van Leeuwen Flora E Hauptmann Michael 1 March 2019 Radiation Exposure From Pediatric CT Scans and Subsequent Cancer Risk in the Netherlands JNCI Journal of the National Cancer Institute 111 3 256 263 doi 10 1093 jnci djy104 PMC 6657440 PMID 30020493 de Gonzalez Amy Berrington Salotti Jane A McHugh Kieran Little Mark P Harbron Richard W Lee Choonsik Ntowe Estelle Braganza Melissa Z Parker Louise Rajaraman Preetha Stiller Charles Stewart Douglas R Craft Alan W Pearce Mark S February 2016 Relationship between paediatric CT scans and subsequent risk of leukaemia and brain tumours assessment of the impact of underlying conditions British Journal of Cancer 114 4 388 394 doi 10 1038 bjc 2015 415 PMC 4815765 PMID 26882064 Smoll Nicolas R Brady Zoe Scurrah Katrina J Lee Choonsik Berrington de Gonzalez Amy Mathews John D 14 January 2023 Computed tomography scan radiation and brain cancer incidence Neuro Oncology 25 7 1368 1376 doi 10 1093 neuonc noad012 PMC 10326490 PMID 36638155 Sasieni P D Shelton J Ormiston Smith N Thomson C S Silcocks P B 2011 What is the lifetime risk of developing cancer the effect of adjusting for multiple primaries British Journal of Cancer 105 3 460 465 doi 10 1038 bjc 2011 250 ISSN 0007 0920 PMC 3172907 PMID 21772332 a b c Rehani Madan M Yang Kai Melick Emily R Heil John Salat Dusan Sensakovic William F Liu Bob 2020 Patients undergoing recurrent CT scans assessing the magnitude European Radiology 30 4 1828 1836 doi 10 1007 s00330 019 06523 y PMID 31792585 S2CID 208520824 a b Brambilla Marco Vassileva Jenia Kuchcinska Agnieszka Rehani Madan M 2020 Multinational data on cumulative radiation exposure of patients from recurrent radiological procedures call for action European Radiology 30 5 2493 2501 doi 10 1007 s00330 019 06528 7 PMID 31792583 S2CID 208520544 a b Rehani Madan M Melick Emily R Alvi Raza M Doda Khera Ruhani Batool Anwar Salma Neilan Tomas G Bettmann Michael 2020 Patients undergoing recurrent CT exams assessment of patients with non malignant diseases reasons for imaging and imaging appropriateness European Radiology 30 4 1839 1846 doi 10 1007 s00330 019 06551 8 PMID 31792584 S2CID 208520463 Eckel Laurence J Fletcher Joel G Bushberg Jerrold T McCollough Cynthia H 2015 10 01 Answers to Common Questions About the Use and Safety of CT Scans Mayo Clinic Proceedings 90 10 1380 1392 doi 10 1016 j mayocp 2015 07 011 ISSN 0025 6196 PMID 26434964 Expert opinion Are CT scans safe ScienceDaily Retrieved 2019 03 14 McCollough Cynthia H Bushberg Jerrold T Fletcher Joel G Eckel Laurence J 2015 10 01 Answers to Common Questions About the Use and Safety of CT Scans Mayo Clinic Proceedings 90 10 1380 1392 doi 10 1016 j mayocp 2015 07 011 ISSN 0025 6196 PMID 26434964 No evidence that CT scans X rays cause cancer Medical News Today 4 February 2016 Retrieved 2019 03 14 Kalra Mannudeep K Maher Michael M Rizzo Stefania Kanarek David Shephard Jo Anne O April 2004 Radiation exposure from Chest CT Issues and Strategies Journal of Korean Medical Science 19 2 159 166 doi 10 3346 jkms 2004 19 2 159 ISSN 1011 8934 PMC 2822293 PMID 15082885 Rob S Bryant T Wilson I Somani B K 2017 Ultra low dose low dose and standard dose CT of the kidney ureters and bladder is there a difference Results from a systematic review of the literature Clinical Radiology 72 1 11 15 doi 10 1016 j crad 2016 10 005 PMID 27810168 a b c d e f g h i j k l m n o p Whaites Eric 2008 10 10 Radiography and Radiology for Dental Care Professionals E Book Elsevier Health Sciences p 25 ISBN 978 0 7020 4799 2 a b c d e Davies H E Wathen C G Gleeson F V 25 February 2011 The risks of radiation exposure related to diagnostic imaging and how to minimise them BMJ 342 feb25 1 d947 doi 10 1136 bmj d947 PMID 21355025 S2CID 206894472 Baysson H Etard C Brisse HJ Bernier MO January 2012 Diagnostic radiation exposure in children and cancer risk current knowledge and perspectives Archives de Pediatrie 19 1 64 73 doi 10 1016 j arcped 2011 10 023 PMID 22130615 a b c d Semelka RC Armao DM Elias J Huda W May 2007 Imaging strategies to reduce the risk of radiation in CT studies including selective substitution with MRI J Magn Reson Imaging 25 5 900 9 doi 10 1002 jmri 20895 PMID 17457809 S2CID 5788891 Larson DB Rader SB Forman HP Fenton LZ August 2007 Informing parents about CT radiation exposure in children it s OK to tell them Am J Roentgenol 189 2 271 5 doi 10 2214 AJR 07 2248 PMID 17646450 S2CID 25020619 Emmerson Benjamin Young Michael 2023 Radiology Patient Safety and Communication StatPearls Treasure Island FL StatPearls Publishing PMID 33620790 retrieved 2023 11 24 a b Namasivayam S Kalra MK Torres WE Small WC Jul 2006 Adverse reactions to intravenous iodinated contrast media a primer for radiologists Emergency Radiology 12 5 210 5 doi 10 1007 s10140 006 0488 6 PMID 16688432 S2CID 28223134 a b c Christiansen C 2005 04 15 X ray contrast media an overview Toxicology 209 2 185 7 doi 10 1016 j tox 2004 12 020 PMID 15767033 a b Wang H Wang HS Liu ZP October 2011 Agents that induce pseudo allergic reaction Drug Discov Ther 5 5 211 9 doi 10 5582 ddt 2011 v5 5 211 PMID 22466368 S2CID 19001357 Drain KL Volcheck GW 2001 Preventing and managing drug induced anaphylaxis Drug Safety 24 11 843 53 doi 10 2165 00002018 200124110 00005 PMID 11665871 S2CID 24840296 Castells Mariana C ed 2010 12 09 Anaphylaxis and hypersensitivity reactions New York Humana Press p 187 ISBN 978 1 60327 950 5 a b c Hasebroock KM Serkova NJ April 2009 Toxicity of MRI and CT contrast agents Expert Opinion on Drug Metabolism amp Toxicology 5 4 403 16 doi 10 1517 17425250902873796 PMID 19368492 S2CID 72557671 Rawson James V Pelletier Allen L 2013 09 01 When to Order Contrast Enhanced CT American Family Physician 88 5 312 316 ISSN 0002 838X PMID 24010394 Thomsen Henrik S Muller Robert N Mattrey Robert F 2012 12 06 Trends in Contrast Media Springer Science amp Business Media ISBN 978 3 642 59814 2 Davenport Matthew 2020 Use of Intravenous Iodinated Contrast Media in Patients with Kidney Disease Consensus Statements from the American College of Radiology and the National Kidney Foundation Radiology 294 3 660 668 doi 10 1148 radiol 2019192094 PMID 31961246 a b c Cuttler JM Pollycove M 2009 Nuclear energy and health and the benefits of low dose radiation hormesis Dose Response 7 1 52 89 doi 10 2203 dose response 08 024 Cuttler PMC 2664640 PMID 19343116 a b What are the Radiation Risks from CT Food and Drug Administration 2009 Archived from the original on 2013 11 05 a b c d e f Hall EJ Brenner DJ May 2008 Cancer risks from diagnostic radiology The British Journal of Radiology 81 965 362 78 doi 10 1259 bjr 01948454 PMID 18440940 S2CID 23348032 a b c d e Shrimpton P C Miller H C Lewis M A Dunn M Doses from Computed Tomography CT examinations in the UK 2003 Review Archived 2011 09 22 at the Wayback Machine Michael T Ryan Poston John W eds 2005 A half century of health physics Baltimore Md Lippincott Williams amp Wilkins p 164 ISBN 978 0 7817 6934 1 Polo SE Jackson SP March 2011 Dynamics of DNA damage response proteins at DNA breaks a focus on protein modifications Genes Dev 25 5 409 33 doi 10 1101 gad 2021311 PMC 3049283 PMID 21363960 The Measurement Reporting and Management of Radiation Dose in CT Archived 2017 06 23 at the Wayback Machine It is a single dose parameter that reflects the risk of a nonuniform exposure in terms of an equivalent whole body exposure Hill B Venning AJ Baldock C 2005 A preliminary study of the novel application of normoxic polymer gel dosimeters for the measurement of CTDI on diagnostic X ray CT scanners Medical Physics 32 6 1589 1597 Bibcode 2005MedPh 32 1589H doi 10 1118 1 1925181 PMID 16013718 Issa Ziad F Miller John M Zipes Douglas P 2019 01 01 Complications of Catheter Ablation of Cardiac Arrhythmias Clinical Arrhythmology and Electrophysiology Elsevier pp 1042 1067 doi 10 1016 b978 0 323 52356 1 00032 3 ISBN 978 0 323 52356 1 Absorbed Equivalent and Effective Dose ICRPaedia icrpaedia org Retrieved 2021 03 21 Materials National Research Council US Committee on Evaluation of EPA Guidelines for Exposure to Naturally Occurring Radioactive 1999 Radiation Quantities and Units Definitions Acronyms National Academies Press US Pua Bradley B Covey Anne M Madoff David C 2018 12 03 Interventional Radiology Fundamentals of Clinical Practice Oxford University Press ISBN 978 0 19 027624 9 Paragraph 55 in The 2007 Recommendations of the International Commission on Radiological Protection International Commission on Radiological Protection Archived from the original on 2012 11 16 Ann ICRP 37 2 4 Do CT scans cause cancer Harvard Medical School March 2013 Archived from the original on 2017 12 09 Retrieved 2017 12 09 CDC 2020 06 05 Radiation and pregnancy A fact sheet for clinicians Centers for Disease Control and Prevention Retrieved 2021 03 21 Yoon Ilsup Slesinger Todd L 2021 Radiation Exposure In Pregnancy StatPearls Treasure Island FL StatPearls Publishing PMID 31869154 retrieved 2021 03 21 Wintermark M Lev MH January 2010 FDA investigates the safety of brain perfusion CT AJNR Am J Neuroradiol 31 1 2 3 doi 10 3174 ajnr A1967 PMC 7964089 PMID 19892810 a b Whitley Stewart A Dodgeon Jan Meadows Angela Cullingworth Jane Holmes Ken Jackson Marcus Hoadley Graham Kulshrestha Randeep 2020 01 06 Clark s Procedures in Diagnostic Imaging A System Based Approach CRC Press ISBN 978 1 4987 1552 2 Tippins R B Torres W E Baumgartner B R Baumgarten D A August 2000 Are screening serum creatinine levels necessary prior to outpatient CT examinations Radiology 216 2 481 484 doi 10 1148 radiology 216 2 r00au23481 ISSN 0033 8419 PMID 10924574 Jun Kyungtaek Yoon Seokhwan 2017 Alignment Solution for CT Image Reconstruction using Fixed Point and Virtual Rotation Axis Scientific Reports 7 41218 arXiv 1605 04833 Bibcode 2017NatSR 741218J doi 10 1038 srep41218 ISSN 2045 2322 PMC 5264594 PMID 28120881 Computed Tomography CT www nibib nih gov Retrieved 2021 03 20 Aichinger Horst Dierker Joachim Joite Barfuss Sigrid Sabel Manfred 2011 10 25 Radiation Exposure and Image Quality in X Ray Diagnostic Radiology Physical Principles and Clinical Applications Springer Science amp Business Media p 5 ISBN 978 3 642 11241 6 Erdogan Hakan 1999 Statistical Image Reconstruction Algorithms Using Paraboloidal Surrogates for PET Transmission Scans University of Michigan ISBN 978 0 599 63374 2 Themes U F O 2018 10 07 CT Image Reconstruction Basics Radiology Key Retrieved 2021 03 20 a b Stirrup James 2020 01 02 Cardiovascular Computed Tomography Oxford University Press ISBN 978 0 19 880927 2 computed tomography Definition from the Merriam Webster Online Dictionary Archived from the original on 19 September 2011 Retrieved 18 August 2009 Webb W Richard Brant William E Major Nancy M 2014 Fundamentals of Body CT Elsevier Health Sciences p 152 ISBN 978 0 323 26358 0 Webb Wayne Richard Brant William E Major Nancy M 2006 01 01 Fundamentals of Body CT Elsevier Health Sciences p 168 ISBN 978 1 4160 0030 3 a, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.