A sector instrument is a general term for a class of mass spectrometer that uses a static electric (E) or magnetic (B) sector or some combination of the two (separately in space) as a mass analyzer.[1] Popular combinations of these sectors have been the EB, BE (of so-called reverse geometry), three-sector BEB and four-sector EBEB (electric-magnetic-electric-magnetic) instruments. Most modern sector instruments are double-focusing instruments (first developed by Francis William Aston, Arthur Jeffrey Dempster, Kenneth Bainbridge and Josef Mattauch in 1936[2]) in that they focus the ion beams both in direction and velocity.[3]

A five sector mass spectrometer

Theory edit

The behavior of ions in a homogeneous, linear, static electric or magnetic field (separately) as is found in a sector instrument is simple. The physics are described by a single equation called the Lorentz force law. This equation is the fundamental equation of all mass spectrometric techniques and applies in non-linear, non-homogeneous cases too and is an important equation in the field of electrodynamics in general.

 

where E is the electric field strength, B is the magnetic field induction, q is the charge of the particle, v is its current velocity (expressed as a vector), and × is the cross product.

So the force on an ion in a linear homogenous electric field (an electric sector) is:

 ,

in the direction of the electric field, with positive ions and opposite that with negative ions.

 
Electric sector from a Finnigan MAT mass spectrometer (vacuum chamber housing removed)

The force is only dependent on the charge and electric field strength. The lighter ions will be deflected more and heavier ions less due to the difference in inertia and the ions will physically separate from each other in space into distinct beams of ions as they exit the electric sector.

And the force on an ion in a linear homogenous magnetic field (a magnetic sector) is:

 ,

perpendicular to both the magnetic field and the velocity vector of the ion itself, in the direction determined by the right-hand rule of cross products and the sign of the charge.

The force in the magnetic sector is complicated by the velocity dependence but with the right conditions (uniform velocity for example) ions of different masses will separate physically in space into different beams as with the electric sector.

Classic geometries edit

These are some of the classic geometries from mass spectrographs which are often used to distinguish different types of sector arrangements, although most current instruments do not fit precisely into any of these categories as the designs have evolved further.

Bainbridge–Jordan edit

The sector instrument geometry consists of a 127.30°   electric sector without an initial drift length followed by a 60° magnetic sector with the same direction of curvature. Sometimes called a "Bainbridge mass spectrometer," this configuration is often used to determine isotopic masses. A beam of positive particles is produced from the isotope under study. The beam is subject to the combined action of perpendicular electric and magnetic fields. Since the forces due to these two fields are equal and opposite when the particles have a velocity given by

 

they do not experience a resultant force; they pass freely through a slit, and are then subject to another magnetic field, transversing a semi-circular path and striking a photographic plate. The mass of the isotope is determined through subsequent calculation.

Mattauch–Herzog edit

The Mattauch–Herzog geometry consists of a 31.82° (  radians) electric sector, a drift length which is followed by a 90° magnetic sector of opposite curvature direction.[4] The entry of the ions sorted primarily by charge into the magnetic field produces an energy focussing effect and much higher transmission than a standard energy filter. This geometry is often used in applications with a high energy spread in the ions produced where sensitivity is nonetheless required, such as spark source mass spectrometry (SSMS) and secondary ion mass spectrometry (SIMS).[5] The advantage of this geometry over the Nier–Johnson geometry is that the ions of different masses are all focused onto the same flat plane. This allows the use of a photographic plate or other flat detector array.

Nier–Johnson edit

The Nier–Johnson geometry consists of a 90° electric sector, a long intermediate drift length and a 60° magnetic sector of the same curvature direction.[6][7]

Hinterberger–Konig edit

The Hinterberger–Konig geometry consists of a 42.43° electric sector, a long intermediate drift length and a 130° magnetic sector of the same curvature direction.

Takeshita edit

The Takeshita geometry consists of a 54.43° electric sector, and short drift length, a second electric sector of the same curvature direction followed by another drift length before a 180° magnetic sector of opposite curvature direction.

Matsuda edit

The Matsuda geometry consists of an 85° electric sector, a quadrupole lens and a 72.5° magnetic sector of the same curvature direction.[8] This geometry is used in the SHRIMP and Panorama (gas source, high-resolution, multicollector to measure isotopologues in geochemistry).

See also edit

References edit

  1. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "electric sector". doi:10.1351/goldbook.E01938
  2. ^ Arthur Jeffrey Dempster (American physicist) at the Encyclopædia Britannica
  3. ^ Burgoyne, Thomas W.; Gary M. Hieftje (1996). "An introduction to ion optics for the mass spectrograph". Mass Spectrometry Reviews. 15 (4): 241–259. Bibcode:1996MSRv...15..241B. CiteSeerX 10.1.1.625.841. doi:10.1002/(SICI)1098-2787(1996)15:4<241::AID-MAS2>3.0.CO;2-I. PMID 27082712. Archived from the original (abstract) on 2012-12-10.
  4. ^ Klemm, Alfred (1946). "Zur Theorie der für alle Massen doppelfokussierenden Massenspektrographen" [The theory of a mass-spectrograph with double focus independent of mass]. Zeitschrift für Naturforschung A. 1 (3): 137–141. Bibcode:1946ZNatA...1..137K. doi:10.1515/zna-1946-0306. S2CID 94043005.
  5. ^ Schilling GD; Andrade FJ; Barnes JH; Sperline RP; Denton MB; Barinaga CJ; Koppenaal DW; Hieftje GM (2006). "Characterization of a second-generation focal-plane camera coupled to an inductively coupled plasma Mattauch–Herzog geometry mass spectrograph". Anal. Chem. 78 (13): 4319–25. doi:10.1021/ac052026k. PMID 16808438.
  6. ^ De Laeter; J. & Kurz; M. D. (2006). "Alfred Nier and the sector field mass spectrometer". Journal of Mass Spectrometry. 41 (7): 847–854. Bibcode:2006JMSp...41..847D. doi:10.1002/jms.1057. PMID 16810642.
  7. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "Nier-Johnson geometry". doi:10.1351/goldbook.N04141
  8. ^ US 4553029, Matsuda, Hisashi, "Mass spectrometer", published 1985-11-12, assigned to Jeol Ltd. 

Further reading edit

  • Thomson, J. J.: Rays of Positive Electricity and their Application to Chemical Analyses; Longmans Green: London, 1913

sector, mass, spectrometer, this, article, about, mass, spectrometers, sector, compasses, sector, instrument, sector, instrument, general, term, class, mass, spectrometer, that, uses, static, electric, magnetic, sector, some, combination, separately, space, ma. This article is about mass spectrometers For sector compasses see sector instrument A sector instrument is a general term for a class of mass spectrometer that uses a static electric E or magnetic B sector or some combination of the two separately in space as a mass analyzer 1 Popular combinations of these sectors have been the EB BE of so called reverse geometry three sector BEB and four sector EBEB electric magnetic electric magnetic instruments Most modern sector instruments are double focusing instruments first developed by Francis William Aston Arthur Jeffrey Dempster Kenneth Bainbridge and Josef Mattauch in 1936 2 in that they focus the ion beams both in direction and velocity 3 A five sector mass spectrometer Contents 1 Theory 2 Classic geometries 2 1 Bainbridge Jordan 2 2 Mattauch Herzog 2 3 Nier Johnson 2 4 Hinterberger Konig 2 5 Takeshita 2 6 Matsuda 3 See also 4 References 5 Further readingTheory editThe behavior of ions in a homogeneous linear static electric or magnetic field separately as is found in a sector instrument is simple The physics are described by a single equation called the Lorentz force law This equation is the fundamental equation of all mass spectrometric techniques and applies in non linear non homogeneous cases too and is an important equation in the field of electrodynamics in general F q E v B displaystyle mathbf F q mathbf E mathbf v times mathbf B nbsp where E is the electric field strength B is the magnetic field induction q is the charge of the particle v is its current velocity expressed as a vector and is the cross product So the force on an ion in a linear homogenous electric field an electric sector is F q E displaystyle F qE nbsp in the direction of the electric field with positive ions and opposite that with negative ions nbsp Electric sector from a Finnigan MAT mass spectrometer vacuum chamber housing removed The force is only dependent on the charge and electric field strength The lighter ions will be deflected more and heavier ions less due to the difference in inertia and the ions will physically separate from each other in space into distinct beams of ions as they exit the electric sector And the force on an ion in a linear homogenous magnetic field a magnetic sector is F q v B displaystyle F qvB nbsp perpendicular to both the magnetic field and the velocity vector of the ion itself in the direction determined by the right hand rule of cross products and the sign of the charge The force in the magnetic sector is complicated by the velocity dependence but with the right conditions uniform velocity for example ions of different masses will separate physically in space into different beams as with the electric sector Classic geometries editThese are some of the classic geometries from mass spectrographs which are often used to distinguish different types of sector arrangements although most current instruments do not fit precisely into any of these categories as the designs have evolved further Bainbridge Jordan edit The sector instrument geometry consists of a 127 30 p 2 displaystyle left frac pi sqrt 2 right nbsp electric sector without an initial drift length followed by a 60 magnetic sector with the same direction of curvature Sometimes called a Bainbridge mass spectrometer this configuration is often used to determine isotopic masses A beam of positive particles is produced from the isotope under study The beam is subject to the combined action of perpendicular electric and magnetic fields Since the forces due to these two fields are equal and opposite when the particles have a velocity given by v E B displaystyle v E B nbsp they do not experience a resultant force they pass freely through a slit and are then subject to another magnetic field transversing a semi circular path and striking a photographic plate The mass of the isotope is determined through subsequent calculation Mattauch Herzog edit The Mattauch Herzog geometry consists of a 31 82 p 4 2 displaystyle pi 4 sqrt 2 nbsp radians electric sector a drift length which is followed by a 90 magnetic sector of opposite curvature direction 4 The entry of the ions sorted primarily by charge into the magnetic field produces an energy focussing effect and much higher transmission than a standard energy filter This geometry is often used in applications with a high energy spread in the ions produced where sensitivity is nonetheless required such as spark source mass spectrometry SSMS and secondary ion mass spectrometry SIMS 5 The advantage of this geometry over the Nier Johnson geometry is that the ions of different masses are all focused onto the same flat plane This allows the use of a photographic plate or other flat detector array Nier Johnson edit The Nier Johnson geometry consists of a 90 electric sector a long intermediate drift length and a 60 magnetic sector of the same curvature direction 6 7 Hinterberger Konig edit The Hinterberger Konig geometry consists of a 42 43 electric sector a long intermediate drift length and a 130 magnetic sector of the same curvature direction Takeshita edit The Takeshita geometry consists of a 54 43 electric sector and short drift length a second electric sector of the same curvature direction followed by another drift length before a 180 magnetic sector of opposite curvature direction Matsuda edit The Matsuda geometry consists of an 85 electric sector a quadrupole lens and a 72 5 magnetic sector of the same curvature direction 8 This geometry is used in the SHRIMP and Panorama gas source high resolution multicollector to measure isotopologues in geochemistry See also editMass analyzed ion kinetic energy spectrometry Charge remote fragmentation Kenneth Bainbridge Alfred O C NierReferences edit IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 electric sector doi 10 1351 goldbook E01938 Arthur Jeffrey Dempster American physicist at the Encyclopaedia Britannica Burgoyne Thomas W Gary M Hieftje 1996 An introduction to ion optics for the mass spectrograph Mass Spectrometry Reviews 15 4 241 259 Bibcode 1996MSRv 15 241B CiteSeerX 10 1 1 625 841 doi 10 1002 SICI 1098 2787 1996 15 4 lt 241 AID MAS2 gt 3 0 CO 2 I PMID 27082712 Archived from the original abstract on 2012 12 10 Klemm Alfred 1946 Zur Theorie der fur alle Massen doppelfokussierenden Massenspektrographen The theory of a mass spectrograph with double focus independent of mass Zeitschrift fur Naturforschung A 1 3 137 141 Bibcode 1946ZNatA 1 137K doi 10 1515 zna 1946 0306 S2CID 94043005 Schilling GD Andrade FJ Barnes JH Sperline RP Denton MB Barinaga CJ Koppenaal DW Hieftje GM 2006 Characterization of a second generation focal plane camera coupled to an inductively coupled plasma Mattauch Herzog geometry mass spectrograph Anal Chem 78 13 4319 25 doi 10 1021 ac052026k PMID 16808438 De Laeter J amp Kurz M D 2006 Alfred Nier and the sector field mass spectrometer Journal of Mass Spectrometry 41 7 847 854 Bibcode 2006JMSp 41 847D doi 10 1002 jms 1057 PMID 16810642 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 Nier Johnson geometry doi 10 1351 goldbook N04141 US 4553029 Matsuda Hisashi Mass spectrometer published 1985 11 12 assigned to Jeol Ltd Further reading editThomson J J Rays of Positive Electricity and their Application to Chemical Analyses Longmans Green London 1913 Retrieved from https en wikipedia org w index php title Sector mass spectrometer amp oldid 1170157512, wikipedia, wiki, book, books, library,