fbpx
Wikipedia

Gas chromatography–mass spectrometry

Gas chromatography–mass spectrometry (GC-MS) is an analytical method that combines the features of gas-chromatography and mass spectrometry to identify different substances within a test sample.[1] Applications of GC-MS include drug detection, fire investigation, environmental analysis, explosives investigation, and identification of unknown samples, including that of material samples obtained from planet Mars during probe missions as early as the 1970s. GC-MS can also be used in airport security to detect substances in luggage or on human beings. Additionally, it can identify trace elements in materials that were previously thought to have disintegrated beyond identification. Like liquid chromatography–mass spectrometry, it allows analysis and detection even of tiny amounts of a substance.[2]

Example of a GC-MS instrument

GC-MS has been regarded as a "gold standard" for forensic substance identification because it is used to perform a 100% specific test, which positively identifies the presence of a particular substance. A nonspecific test merely indicates that any of several in a category of substances is present. Although a nonspecific test could statistically suggest the identity of the substance, this could lead to false positive identification. However, the high temperatures (300°C) used in the GC-MS injection port (and oven) can result in thermal degradation of injected molecules,[3] thus resulting in the measurement of degradation products instead of the actual molecule(s) of interest.

History

The first on-line coupling of gas chromatography to a mass spectrometer was reported in the late 1950s.[4][5] An interest in coupling the methods had been suggested as early as December 1954.[6] The development of affordable and miniaturized computers has helped in the simplification of the use of this instrument, as well as allowed great improvements in the amount of time it takes to analyze a sample. In 1964, Electronic Associates, Inc. (EAI), a leading U.S. supplier of analog computers, began development of a computer controlled quadrupole mass spectrometer under the direction of Robert E. Finnigan.[7] By 1966 Finnigan and collaborator Mike Uthe's EAI division had sold over 500 quadrupole residual gas-analyzer instruments.[7] In 1967, Finnigan left EAI to form the Finnigan Instrument Corporation along with Roger Sant, T. Z. Chou, Michael Story, Lloyd Friedman, and William Fies.[8] In early 1968, they delivered the first prototype quadrupole GC/MS instruments to Stanford and Purdue University.[7] When Finnigan Instrument Corporation was acquired by Thermo Instrument Systems (later Thermo Fisher Scientific) in 1990, it was considered "the world's leading manufacturer of mass spectrometers".[9]

Instrumentation

 
The insides of the GC-MS, with the column of the gas chromatograph in the oven on the right.

The GC-MS is composed of two major building blocks: the gas chromatograph and the mass spectrometer. The gas chromatograph utilizes a capillary column whose properties regarding molecule separation depend on the column's dimensions (length, diameter, film thickness) as well as the phase properties (e.g. 5% phenyl polysiloxane). The difference in the chemical properties between different molecules in a mixture and their relative affinity for the stationary phase of the column will promote separation of the molecules as the sample travels the length of the column. The molecules are retained by the column and then elute (come off) from the column at different times (called the retention time), and this allows the mass spectrometer downstream to capture, ionize, accelerate, deflect, and detect the ionized molecules separately. The mass spectrometer does this by breaking each molecule into ionized fragments and detecting these fragments using their mass-to-charge ratio.

 
GC-MS schematic

These two components, used together, allow a much finer degree of substance identification than either unit used separately. It is not possible to make an accurate identification of a particular molecule by gas chromatography or mass spectrometry alone. The mass spectrometry process normally requires a very pure sample while gas chromatography using a traditional detector (e.g. Flame ionization detector) cannot differentiate between multiple molecules that happen to take the same amount of time to travel through the column (i.e. have the same retention time), which results in two or more molecules that co-elute. Sometimes two different molecules can also have a similar pattern of ionized fragments in a mass spectrometer (mass spectrum). Combining the two processes reduces the possibility of error, as it is extremely unlikely that two different molecules will behave in the same way in both a gas chromatograph and a mass spectrometer. Therefore, when an identifying mass spectrum appears at a characteristic retention time in a GC-MS analysis, it typically increases certainty that the analyte of interest is in the sample.

Purge and trap GC-MS

For the analysis of volatile compounds, a purge and trap (P&T) concentrator system may be used to introduce samples. The target analytes are extracted by mixing the sample with water and purge with inert gas (e.g. Nitrogen gas) into an airtight chamber, this is known as purging or sparging. The volatile compounds move into the headspace above the water and are drawn along a pressure gradient (caused by the introduction of the purge gas) out of the chamber. The volatile compounds are drawn along a heated line onto a 'trap'. The trap is a column of adsorbent material at ambient temperature that holds the compounds by returning them to the liquid phase. The trap is then heated and the sample compounds are introduced to the GC-MS column via a volatiles interface, which is a split inlet system. P&T GC-MS is particularly suited to volatile organic compounds (VOCs) and BTEX compounds (aromatic compounds associated with petroleum).[10]

A faster alternative is the "purge-closed loop" system. In this system the inert gas is bubbled through the water until the concentrations of organic compounds in the vapor phase are at equilibrium with concentrations in the aqueous phase. The gas phase is then analysed directly.[11]

Types of mass spectrometer detectors

The most common type of mass spectrometer (MS) associated with a gas chromatograph (GC) is the quadrupole mass spectrometer, sometimes referred to by the Hewlett-Packard (now Agilent) trade name "Mass Selective Detector" (MSD). Another relatively common detector is the ion trap mass spectrometer. Additionally one may find a magnetic sector mass spectrometer, however these particular instruments are expensive and bulky and not typically found in high-throughput service laboratories. Other detectors may be encountered such as time of flight (TOF), tandem quadrupoles (MS-MS) (see below), or in the case of an ion trap MSn where n indicates the number mass spectrometry stages.

GC-tandem MS

When a second phase of mass fragmentation is added, for example using a second quadrupole in a quadrupole instrument, it is called tandem MS (MS/MS). MS/MS can sometimes be used to quantitate low levels of target compounds in the presence of a high sample matrix background.

The first quadrupole (Q1) is connected with a collision cell (Q2) and another quadrupole (Q3). Both quadrupoles can be used in scanning or static mode, depending on the type of MS/MS analysis being performed. Types of analysis include product ion scan, precursor ion scan, selected reaction monitoring (SRM) (sometimes referred to as multiple reaction monitoring (MRM)) and neutral loss scan. For example: When Q1 is in static mode (looking at one mass only as in SIM), and Q3 is in scanning mode, one obtains a so-called product ion spectrum (also called "daughter spectrum"). From this spectrum, one can select a prominent product ion which can be the product ion for the chosen precursor ion. The pair is called a "transition" and forms the basis for SRM. SRM is highly specific and virtually eliminates matrix background.

Ionization

After the molecules travel the length of the column, pass through the transfer line and enter into the mass spectrometer they are ionized by various methods with typically only one method being used at any given time. Once the sample is fragmented it will then be detected, usually by an electron multiplier, which essentially turns the ionized mass fragment into an electrical signal that is then detected.

The ionization technique chosen is independent of using full scan or SIM.

 
Block diagram for gas chromatography using electron ionization for collecting mass spectrum

Electron ionization

By far the most common and perhaps standard form of ionization is electron ionization (EI). The molecules enter into the MS (the source is a quadrupole or the ion trap itself in an ion trap MS) where they are bombarded with free electrons emitted from a filament, not unlike the filament one would find in a standard light bulb. The electrons bombard the molecules, causing the molecule to fragment in a characteristic and reproducible way. This "hard ionization" technique results in the creation of more fragments of low mass-to-charge ratio (m/z) and few, if any, molecules approaching the molecular mass unit. Hard ionization is considered by mass spectrometrists as the employ of molecular electron bombardment, whereas "soft ionization" is charge by molecular collision with an introduced gas. The molecular fragmentation pattern is dependent upon the electron energy applied to the system, typically 70 eV (electronvolts). The use of 70 eV facilitates comparison of generated spectra with library spectra using manufacturer-supplied software or software developed by the National Institute of Standards (NIST-USA). Spectral library searches employ matching algorithms such as Probability Based Matching[12] and dot-product[13] matching that are used with methods of analysis written by many method standardization agencies. Sources of libraries include NIST,[14] Wiley,[15] the AAFS,[16] and instrument manufacturers.

Cold electron ionization

The "hard ionization" process of electron ionization can be softened by the cooling of the molecules before their ionization, resulting in mass spectra that are richer in information.[17][18] In this method named cold electron ionization (cold-EI) the molecules exit the GC column, mixed with added helium make up gas and expand into vacuum through a specially designed supersonic nozzle, forming a supersonic molecular beam (SMB). Collisions with the make up gas at the expanding supersonic jet reduce the internal vibrational (and rotational) energy of the analyte molecules, hence reducing the degree of fragmentation caused by the electrons during the ionization process.[17][18] Cold-EI mass spectra are characterized by an abundant molecular ion while the usual fragmentation pattern is retained, thus making cold-EI mass spectra compatible with library search identification techniques. The enhanced molecular ions increase the identification probabilities of both known and unknown compounds, amplify isomer mass spectral effects and enable the use of isotope abundance analysis for the elucidation of elemental formulas.[19]

Chemical ionization

In chemical ionization (CI) a reagent gas, typically methane or ammonia is introduced into the mass spectrometer. Depending on the technique (positive CI or negative CI) chosen, this reagent gas will interact with the electrons and analyte and cause a 'soft' ionization of the molecule of interest. A softer ionization fragments the molecule to a lower degree than the hard ionization of EI. One of the main benefits of using chemical ionization is that a mass fragment closely corresponding to the molecular weight of the analyte of interest is produced.

In positive chemical ionization (PCI) the reagent gas interacts with the target molecule, most often with a proton exchange. This produces the species in relatively high amounts.

In negative chemical ionization (NCI) the reagent gas decreases the impact of the free electrons on the target analyte. This decreased energy typically leaves the fragment in great supply.

Analysis

A mass spectrometer is typically utilized in one of two ways: full scan or selective ion monitoring (SIM). The typical GC-MS instrument is capable of performing both functions either individually or concomitantly, depending on the setup of the particular instrument.

The primary goal of instrument analysis is to quantify an amount of substance. This is done by comparing the relative concentrations among the atomic masses in the generated spectrum. Two kinds of analysis are possible, comparative and original. Comparative analysis essentially compares the given spectrum to a spectrum library to see if its characteristics are present for some sample in the library. This is best performed by a computer because there are a myriad of visual distortions that can take place due to variations in scale. Computers can also simultaneously correlate more data (such as the retention times identified by GC), to more accurately relate certain data. Deep learning was shown to lead to promising results in the identification of VOCs from raw GC-MS data[20]

Another method of analysis measures the peaks in relation to one another. In this method, the tallest peak is assigned 100% of the value, and the other peaks being assigned proportionate values. All values above 3% are assigned. The total mass of the unknown compound is normally indicated by the parent peak. The value of this parent peak can be used to fit with a chemical formula containing the various elements which are believed to be in the compound. The isotope pattern in the spectrum, which is unique for elements that have many natural isotopes, can also be used to identify the various elements present. Once a chemical formula has been matched to the spectrum, the molecular structure and bonding can be identified, and must be consistent with the characteristics recorded by GC-MS. Typically, this identification is done automatically by programs which come with the instrument, given a list of the elements which could be present in the sample.

A “full spectrum” analysis considers all the “peaks” within a spectrum. Conversely, selective ion monitoring (SIM) only monitors selected ions associated with a specific substance. This is done on the assumption that at a given retention time, a set of ions is characteristic of a certain compound. This is a fast and efficient analysis, especially if the analyst has previous information about a sample or is only looking for a few specific substances. When the amount of information collected about the ions in a given gas chromatographic peak decreases, the sensitivity of the analysis increases. So, SIM analysis allows for a smaller quantity of a compound to be detected and measured, but the degree of certainty about the identity of that compound is reduced.

Full scan MS

When collecting data in the full scan mode, a target range of mass fragments is determined and put into the instrument's method. An example of a typical broad range of mass fragments to monitor would be m/z 50 to m/z 400. The determination of what range to use is largely dictated by what one anticipates being in the sample while being cognizant of the solvent and other possible interferences. A MS should not be set to look for mass fragments too low or else one may detect air (found as m/z 28 due to nitrogen), carbon dioxide (m/z 44) or other possible interference. Additionally if one is to use a large scan range then sensitivity of the instrument is decreased due to performing fewer scans per second since each scan will have to detect a wide range of mass fragments.

Full scan is useful in determining unknown compounds in a sample. It provides more information than SIM when it comes to confirming or resolving compounds in a sample. During instrument method development it may be common to first analyze test solutions in full scan mode to determine the retention time and the mass fragment fingerprint before moving to a SIM instrument method.

Selective ion monitoring

In selective ion monitoring (SIM) certain ion fragments are entered into the instrument method and only those mass fragments are detected by the mass spectrometer. The advantages of SIM are that the detection limit is lower since the instrument is only looking at a small number of fragments (e.g. three fragments) during each scan. More scans can take place each second. Since only a few mass fragments of interest are being monitored, matrix interferences are typically lower. To additionally confirm the likelihood of a potentially positive result, it is relatively important to be sure that the ion ratios of the various mass fragments are comparable to a known reference standard.

Applications

Environmental monitoring and cleanup

GC-MS is becoming the tool of choice for tracking organic pollutants in the environment. The cost of GC-MS equipment has decreased significantly, and the reliability has increased at the same time, which has contributed to its increased adoption in environmental studies.

Criminal forensics

GC-MS can analyze the particles from a human body in order to help link a criminal to a crime. The analysis of fire debris using GC-MS is well established, and there is even an established American Society for Testing and Materials (ASTM) standard for fire debris analysis. GCMS/MS is especially useful here as samples often contain very complex matrices and results, used in court, need to be highly accurate.

Law enforcement

GC-MS is increasingly used for detection of illegal narcotics, and may eventually supplant drug-sniffing dogs.[1] A simple and selective GC-MS method for detecting marijuana usage was recently developed by the Robert Koch-Institute in Germany. It involves identifying an acid metabolite of tetrahydrocannabinol (THC), the active ingredient in marijuana, in urine samples by employing derivatization in the sample preparation.[21] GC-MS is also commonly used in forensic toxicology to find drugs and/or poisons in biological specimens of suspects, victims, or the deceased. In drug screening, GC-MS methods frequently utilize liquid-liquid extraction as a part of sample preparation, in which target compounds are extracted from blood plasma.[22]

Sports anti-doping analysis

GC-MS is the main tool used in sports anti-doping laboratories to test athletes' urine samples for prohibited performance-enhancing drugs, for example anabolic steroids.[23]

Security

A post–September 11 development, explosive detection systems have become a part of all US airports. These systems run on a host of technologies, many of them based on GC-MS. There are only three manufacturers certified by the FAA to provide these systems,[citation needed] one of which is Thermo Detection (formerly Thermedics), which produces the EGIS, a GC-MS-based line of explosives detectors. The other two manufacturers are Barringer Technologies, now owned by Smith 's Detection Systems, and Ion Track Instruments, part of General Electric Infrastructure Security Systems.

Chemical warfare agent detection

As part of the post-September 11 drive towards increased capability in homeland security and public health preparedness, traditional GC-MS units with transmission quadrupole mass spectrometers, as well as those with cylindrical ion trap (CIT-MS) and toroidal ion trap (T-ITMS) mass spectrometers have been modified for field portability and near real-time detection of chemical warfare agents (CWA) such as sarin, soman, and VX.[24] These complex and large GC-MS systems have been modified and configured with resistively heated low thermal mass (LTM) gas chromatographs that reduce analysis time to less than ten percent of the time required in traditional laboratory systems.[25] Additionally, the systems are smaller, and more mobile, including units that are mounted in mobile analytical laboratories (MAL), such as those used by the United States Marine Corps Chemical and Biological Incident Response Force MAL and other similar laboratories, and systems that are hand-carried by two-person teams or individuals, much ado to the smaller mass detectors.[26] Depending on the system, the analytes can be introduced via liquid injection, desorbed from sorbent tubes through a thermal desorption process, or with solid-phase micro extraction (SPME).

Chemical engineering

GC-MS is used for the analysis of unknown organic compound mixtures. One critical use of this technology is the use of GC-MS to determine the composition of bio-oils processed from raw biomass.[27] GC-MS is also utilized in the identification of continuous phase component in a smart material, Magnetorheological (MR) fluid.[28]

Food, beverage and perfume analysis

Foods and beverages contain numerous aromatic compounds, some naturally present in the raw materials and some forming during processing. GC-MS is extensively used for the analysis of these compounds which include esters, fatty acids, alcohols, aldehydes, terpenes etc. It is also used to detect and measure contaminants from spoilage or adulteration which may be harmful and which is often controlled by governmental agencies, for example pesticides.

Astrochemistry

Several GC-MS have left earth. Two were brought to Mars by the Viking program.[29] Venera 11 and 12 and Pioneer Venus analysed the atmosphere of Venus with GC-MS.[30] The Huygens probe of the Cassini–Huygens mission landed one GC-MS on Saturn's largest moon, Titan.[31] The MSL Curiosity rover's Sample analysis at Mars (SAM) instrument contains both a gas chromatograph and quadrupole mass spectrometer that can be used in tandem as a GC-MS.[32] The material in the comet 67P/Churyumov–Gerasimenko was analysed by the Rosetta mission with a chiral GC-MS in 2014.[33]

Medicine

Dozens of congenital metabolic diseases also known as inborn errors of metabolism (IEM) are now detectable by newborn screening tests, especially the testing using gas chromatography–mass spectrometry. GC-MS can determine compounds in urine even in minor concentration. These compounds are normally not present but appear in individuals suffering with metabolic disorders. This is increasingly becoming a common way to diagnose IEM for earlier diagnosis and institution of treatment eventually leading to a better outcome. It is now possible to test a newborn for over 100 genetic metabolic disorders by a urine test at birth based on GC-MS.

In combination with isotopic labeling of metabolic compounds, the GC-MS is used for determining metabolic activity. Most applications are based on the use of 13C as the labeling and the measurement of 13C-12C ratios with an isotope ratio mass spectrometer (IRMS); an MS with a detector designed to measure a few select ions and return values as ratios.

See also

References

  1. ^ Sparkman DO, Penton Z, Kitson FG (17 May 2011). Gas Chromatography and Mass Spectrometry: A Practical Guide. Academic Press. ISBN 978-0-08-092015-3.
  2. ^ Jones M. "Gas Chromatography-Mass Spectrometry". American Chemical Society. Retrieved 19 Nov 2019.
  3. ^ Fang M, Ivanisevic J, Benton HP, Johnson CH, Patti GJ, Hoang LT, et al. (November 2015). "Thermal Degradation of Small Molecules: A Global Metabolomic Investigation". Analytical Chemistry. 87 (21): 10935–41. doi:10.1021/acs.analchem.5b03003. PMC 4633772. PMID 26434689.
  4. ^ Holmes JC, Morrell FA (1957). "Oscillographic Mass Spectrometric Monitoring of Gas Chromatography". Applied Spectroscopy. 11 (2): 86–87. doi:10.1366/000370257774633394. ISSN 0003-7028.
  5. ^ Gohlke RS (1959). "Time-of-Flight Mass Spectrometry and Gas-Liquid Partition Chromatography". Analytical Chemistry. 31 (4): 535–541. doi:10.1021/ac50164a024. ISSN 0003-2700.
  6. ^ Patton HW, Lewis JS, Kaye WI (1955). "Separation and Analysis of Gases and Volatile Liquids by Gas Chromatography". Analytical Chemistry. 27 (2): 170–174. doi:10.1021/ac60098a002.
  7. ^ a b c Brock DC (2011). "A Measure of Success". Chemical Heritage Magazine. 29 (1). Retrieved 22 March 2018.
  8. ^ Webb-Halpern L (2008). "Detecting Success". Chemical Heritage Magazine. 26 (2): 31.
  9. ^ "Thermo Instrument Systems Inc. History". International Directory of Company Histories (Volume 11 ed.). St. James Press. 1995. pp. 513–514. Retrieved 23 January 2015.
  10. ^ "Optimizing the Analysis of Volatile Organic Compounds – Technical Guide" Restek Corporation, Lit. Cat. 59887A
  11. ^ Wang T, Lenahan R (April 1984). "Determination of volatile halocarbons in water by purge-closed loop gas chromatography". Bulletin of Environmental Contamination and Toxicology. 32 (4): 429–38. doi:10.1007/BF01607519. PMID 6713137. S2CID 992748.
  12. ^ Stauffer DB, McLafferty FW, Ellis RD, Peterson DW (1974). "Probability based matching of mass spectra. Rapid identification of specific compounds in mixtures". Organic Mass Spectrometry. 9 (4): 690–702. doi:10.1002/oms.1210090710.
  13. ^ Stein SE, Scott DR (September 1994). "Optimization and testing of mass spectral library search algorithms for compound identification". Journal of the American Society for Mass Spectrometry. 5 (9): 859–66. doi:10.1016/1044-0305(94)87009-8. PMID 24222034.
  14. ^ Standard Reference Data. nist.gov
  15. ^ Wiley's Scientific, Technical, and Medical Databases: Home. wiley.com
  16. ^ Mass Spectrometry Database Committee. ualberta.ca
  17. ^ a b Amirav A, Gordin A, Poliak M, Fialkov AB (February 2008). "Gas chromatography-mass spectrometry with supersonic molecular beams". Journal of Mass Spectrometry. 43 (2): 141–63. Bibcode:2008JMSp...43..141A. doi:10.1002/jms.1380. PMID 18225851.
  18. ^ a b SMB-MS (Supersonic GC-MS). tau.ac.il
  19. ^ Alon T, Amirav A (2006). "Isotope abundance analysis methods and software for improved sample identification with supersonic gas chromatography/mass spectrometry". Rapid Communications in Mass Spectrometry. 20 (17): 2579–88. Bibcode:2006RCMS...20.2579A. doi:10.1002/rcm.2637. PMID 16897787.
  20. ^ Skarysz A (July 2018). "Convolutional neural networks for automated targeted analysis of raw gas chromatography-mass spectrometry data". International Joint Conferences on Neural Networks (2018) Rio de Janeiro, Brazil: 1–8. doi:10.1109/IJCNN.2018.8489539. ISBN 978-1-5090-6014-6. S2CID 52989098.
  21. ^ Hübschmann HJ (22 April 2015). Handbook of GC-MS : Fundamentals and Applications (3 ed.). John Wiley & Sons, Incorporated. p. 735. ISBN 9783527674336. Retrieved 22 January 2018.
  22. ^ Hübschmann HJ (22 April 2015). Handbook of GC-MS : Fundamentals and Applications (3 ed.). John Wiley & Sons, Incorporated. p. 731. ISBN 9783527674336. Retrieved 22 January 2018.
  23. ^ Tsivou M, Kioukia-Fougia N, Lyris E, Aggelis Y, Fragkaki A, Kiousi X, et al. (2006). "An overview of the doping control analysis during the Olympic Games of 2004 in Athens, Greece". Analytica Chimica Acta. 555: 1–13. doi:10.1016/j.aca.2005.08.068.
  24. ^ Smith PA, Lepage CJ, Lukacs M, Martin N, Shufutinsky A, Savage PB (2010). "Field-portable gas chromatography with transmission quadrupole and cylindrical ion trap mass spectrometric detection: Chromatographic retention index data and ion/molecule interactions for chemical warfare agent identification". International Journal of Mass Spectrometry. 295 (3): 113–118. Bibcode:2010IJMSp.295..113S. doi:10.1016/j.ijms.2010.03.001.
  25. ^ Sloan KM, Mustacich RV, Eckenrode BA (2001). "Development and evaluation of a low thermal mass gas chromatograph for rapid forensic GC-MS analyses". Field Analytical Chemistry & Technology. 5 (6): 288–301. doi:10.1002/fact.10011.
  26. ^ Patterson GE, Guymon AJ, Riter LS, Everly M, Griep-Raming J, Laughlin BC, et al. (December 2002). "Miniature cylindrical ion trap mass spectrometer". Analytical Chemistry. 74 (24): 6145–53. doi:10.1021/ac020494d. PMID 12510732.
  27. ^ Tekin K, Karagöz S, Bektaş S (2014-12-01). "A review of hydrothermal biomass processing". Renewable and Sustainable Energy Reviews. 40: 673–687. doi:10.1016/j.rser.2014.07.216.
  28. ^ Unuh MH, Muhamad P, Waziralilah NF, Amran MH (2019). "Characterization of Vehicle Smart Fluid using Gas Chromatography-Mass Spectrometry (GCMS)" (PDF). Journal of Advanced Research in Fluid Mechanics and Thermal Sciences. 55 (2): 240–248.
  29. ^ . NASA
  30. ^ Krasnopolsky VA, Parshev VA (1981). "Chemical composition of the atmosphere of Venus". Nature. 292 (5824): 610–613. Bibcode:1981Natur.292..610K. doi:10.1038/292610a0. S2CID 4369293.
  31. ^ Niemann HB, Atreya SK, Bauer SJ, Carignan GR, Demick JE, Frost RL, et al. (December 2005). "The abundances of constituents of Titan's atmosphere from the GCMS instrument on the Huygens probe" (PDF). Nature. 438 (7069): 779–84. Bibcode:2005Natur.438..779N. doi:10.1038/nature04122. hdl:2027.42/62703. PMID 16319830. S2CID 4344046.
  32. ^ . msl-scicorner.jpl.nasa.gov. Archived from the original on 2009-03-20. Retrieved 2019-06-25.
  33. ^ Gösmann F, Rosenbauer H, Roll R, Böhnhardt H (October 2005). "COSAC onboard Rosetta: a bioastronomy experiment for the short-period comet 67P/Churyumov-Gerasimenko". Astrobiology. 5 (5): 622–31. Bibcode:2005AsBio...5..622G. doi:10.1089/ast.2005.5.622. PMID 16225435.

Bibliography

  • Adams RP (2007). Identification of Essential Oil Components By Gas Chromatography/Mass Spectrometry. Allured Pub Corp. ISBN 978-1-932633-21-4.
  • Adlard ER, Handley AJ (2001). Gas chromatographic techniques and applications. London: Sheffield Academic. ISBN 978-0-8493-0521-4.
  • Barry EF, Grob RE (2004). Modern practice of gas chromatography. New York: Wiley-Interscience. ISBN 978-0-471-22983-4.
  • Eiceman GA (2000). "Gas Chromatography". In Meyers RA (ed.). Encyclopedia of Analytical Chemistry: Applications, Theory, and Instrumentation. Chichester: Wiley. p. 10627. ISBN 0-471-97670-9.
  • Giannelli PC, Imwinkelried EJ (1999). "Drug Identification: Gas Chromatography.". Scientific Evidence. Vol. 2. Charlottesville: Lexis Law Publishing. p. 362. ISBN 0-327-04985-5.
  • McEwen CN, Kitson FG, Larsen BS (1996). Gas chromatography and mass spectrometry: a practical guide. Boston: Academic Press. ISBN 978-0-12-483385-2.
  • McMaster C, McMaster MC (1998). GC/MS: a practical user's guide. New York: Wiley. ISBN 978-0-471-24826-2.
  • Message GM (1984). Practical aspects of gas chromatography/mass spectrometry. New York: Wiley. ISBN 978-0-471-06277-6.
  • Niessen WM (2001). Current practice of gas chromatography–mass spectrometry. New York, N.Y: Marcel Dekker. ISBN 978-0-8247-0473-5.
  • Weber A, Maurer HW, Pfleger K (2007). Mass Spectral and GC Data of Drugs, Poisons, Pesticides, Pollutants and Their Metabolites. Weinheim: Wiley-VCH. ISBN 978-3-527-31538-3.

External links

  • Gas+chromatography-mass+spectrometry at the US National Library of Medicine Medical Subject Headings (MeSH)
  • Golm Metabolome Database, a mass spectral reference database of plant metabolites

chromatography, mass, spectrometry, analytical, method, that, combines, features, chromatography, mass, spectrometry, identify, different, substances, within, test, sample, applications, include, drug, detection, fire, investigation, environmental, analysis, e. Gas chromatography mass spectrometry GC MS is an analytical method that combines the features of gas chromatography and mass spectrometry to identify different substances within a test sample 1 Applications of GC MS include drug detection fire investigation environmental analysis explosives investigation and identification of unknown samples including that of material samples obtained from planet Mars during probe missions as early as the 1970s GC MS can also be used in airport security to detect substances in luggage or on human beings Additionally it can identify trace elements in materials that were previously thought to have disintegrated beyond identification Like liquid chromatography mass spectrometry it allows analysis and detection even of tiny amounts of a substance 2 Example of a GC MS instrument GC MS has been regarded as a gold standard for forensic substance identification because it is used to perform a 100 specific test which positively identifies the presence of a particular substance A nonspecific test merely indicates that any of several in a category of substances is present Although a nonspecific test could statistically suggest the identity of the substance this could lead to false positive identification However the high temperatures 300 C used in the GC MS injection port and oven can result in thermal degradation of injected molecules 3 thus resulting in the measurement of degradation products instead of the actual molecule s of interest Contents 1 History 2 Instrumentation 2 1 Purge and trap GC MS 2 2 Types of mass spectrometer detectors 2 3 GC tandem MS 3 Ionization 3 1 Electron ionization 3 1 1 Cold electron ionization 3 2 Chemical ionization 4 Analysis 4 1 Full scan MS 4 2 Selective ion monitoring 5 Applications 5 1 Environmental monitoring and cleanup 5 2 Criminal forensics 5 3 Law enforcement 5 4 Sports anti doping analysis 5 5 Security 5 6 Chemical warfare agent detection 5 7 Chemical engineering 5 8 Food beverage and perfume analysis 5 9 Astrochemistry 5 10 Medicine 6 See also 7 References 8 Bibliography 9 External linksHistory EditThe first on line coupling of gas chromatography to a mass spectrometer was reported in the late 1950s 4 5 An interest in coupling the methods had been suggested as early as December 1954 6 The development of affordable and miniaturized computers has helped in the simplification of the use of this instrument as well as allowed great improvements in the amount of time it takes to analyze a sample In 1964 Electronic Associates Inc EAI a leading U S supplier of analog computers began development of a computer controlled quadrupole mass spectrometer under the direction of Robert E Finnigan 7 By 1966 Finnigan and collaborator Mike Uthe s EAI division had sold over 500 quadrupole residual gas analyzer instruments 7 In 1967 Finnigan left EAI to form the Finnigan Instrument Corporation along with Roger Sant T Z Chou Michael Story Lloyd Friedman and William Fies 8 In early 1968 they delivered the first prototype quadrupole GC MS instruments to Stanford and Purdue University 7 When Finnigan Instrument Corporation was acquired by Thermo Instrument Systems later Thermo Fisher Scientific in 1990 it was considered the world s leading manufacturer of mass spectrometers 9 Instrumentation EditMain articles Gas chromatograph and Mass spectrometer The insides of the GC MS with the column of the gas chromatograph in the oven on the right The GC MS is composed of two major building blocks the gas chromatograph and the mass spectrometer The gas chromatograph utilizes a capillary column whose properties regarding molecule separation depend on the column s dimensions length diameter film thickness as well as the phase properties e g 5 phenyl polysiloxane The difference in the chemical properties between different molecules in a mixture and their relative affinity for the stationary phase of the column will promote separation of the molecules as the sample travels the length of the column The molecules are retained by the column and then elute come off from the column at different times called the retention time and this allows the mass spectrometer downstream to capture ionize accelerate deflect and detect the ionized molecules separately The mass spectrometer does this by breaking each molecule into ionized fragments and detecting these fragments using their mass to charge ratio GC MS schematic These two components used together allow a much finer degree of substance identification than either unit used separately It is not possible to make an accurate identification of a particular molecule by gas chromatography or mass spectrometry alone The mass spectrometry process normally requires a very pure sample while gas chromatography using a traditional detector e g Flame ionization detector cannot differentiate between multiple molecules that happen to take the same amount of time to travel through the column i e have the same retention time which results in two or more molecules that co elute Sometimes two different molecules can also have a similar pattern of ionized fragments in a mass spectrometer mass spectrum Combining the two processes reduces the possibility of error as it is extremely unlikely that two different molecules will behave in the same way in both a gas chromatograph and a mass spectrometer Therefore when an identifying mass spectrum appears at a characteristic retention time in a GC MS analysis it typically increases certainty that the analyte of interest is in the sample Purge and trap GC MS Edit For the analysis of volatile compounds a purge and trap P amp T concentrator system may be used to introduce samples The target analytes are extracted by mixing the sample with water and purge with inert gas e g Nitrogen gas into an airtight chamber this is known as purging or sparging The volatile compounds move into the headspace above the water and are drawn along a pressure gradient caused by the introduction of the purge gas out of the chamber The volatile compounds are drawn along a heated line onto a trap The trap is a column of adsorbent material at ambient temperature that holds the compounds by returning them to the liquid phase The trap is then heated and the sample compounds are introduced to the GC MS column via a volatiles interface which is a split inlet system P amp T GC MS is particularly suited to volatile organic compounds VOCs and BTEX compounds aromatic compounds associated with petroleum 10 A faster alternative is the purge closed loop system In this system the inert gas is bubbled through the water until the concentrations of organic compounds in the vapor phase are at equilibrium with concentrations in the aqueous phase The gas phase is then analysed directly 11 Types of mass spectrometer detectors Edit The most common type of mass spectrometer MS associated with a gas chromatograph GC is the quadrupole mass spectrometer sometimes referred to by the Hewlett Packard now Agilent trade name Mass Selective Detector MSD Another relatively common detector is the ion trap mass spectrometer Additionally one may find a magnetic sector mass spectrometer however these particular instruments are expensive and bulky and not typically found in high throughput service laboratories Other detectors may be encountered such as time of flight TOF tandem quadrupoles MS MS see below or in the case of an ion trap MSn where n indicates the number mass spectrometry stages GC tandem MS Edit When a second phase of mass fragmentation is added for example using a second quadrupole in a quadrupole instrument it is called tandem MS MS MS MS MS can sometimes be used to quantitate low levels of target compounds in the presence of a high sample matrix background The first quadrupole Q1 is connected with a collision cell Q2 and another quadrupole Q3 Both quadrupoles can be used in scanning or static mode depending on the type of MS MS analysis being performed Types of analysis include product ion scan precursor ion scan selected reaction monitoring SRM sometimes referred to as multiple reaction monitoring MRM and neutral loss scan For example When Q1 is in static mode looking at one mass only as in SIM and Q3 is in scanning mode one obtains a so called product ion spectrum also called daughter spectrum From this spectrum one can select a prominent product ion which can be the product ion for the chosen precursor ion The pair is called a transition and forms the basis for SRM SRM is highly specific and virtually eliminates matrix background Ionization EditAfter the molecules travel the length of the column pass through the transfer line and enter into the mass spectrometer they are ionized by various methods with typically only one method being used at any given time Once the sample is fragmented it will then be detected usually by an electron multiplier which essentially turns the ionized mass fragment into an electrical signal that is then detected The ionization technique chosen is independent of using full scan or SIM Block diagram for gas chromatography using electron ionization for collecting mass spectrum Electron ionization Edit By far the most common and perhaps standard form of ionization is electron ionization EI The molecules enter into the MS the source is a quadrupole or the ion trap itself in an ion trap MS where they are bombarded with free electrons emitted from a filament not unlike the filament one would find in a standard light bulb The electrons bombard the molecules causing the molecule to fragment in a characteristic and reproducible way This hard ionization technique results in the creation of more fragments of low mass to charge ratio m z and few if any molecules approaching the molecular mass unit Hard ionization is considered by mass spectrometrists as the employ of molecular electron bombardment whereas soft ionization is charge by molecular collision with an introduced gas The molecular fragmentation pattern is dependent upon the electron energy applied to the system typically 70 eV electronvolts The use of 70 eV facilitates comparison of generated spectra with library spectra using manufacturer supplied software or software developed by the National Institute of Standards NIST USA Spectral library searches employ matching algorithms such as Probability Based Matching 12 and dot product 13 matching that are used with methods of analysis written by many method standardization agencies Sources of libraries include NIST 14 Wiley 15 the AAFS 16 and instrument manufacturers Cold electron ionization Edit The hard ionization process of electron ionization can be softened by the cooling of the molecules before their ionization resulting in mass spectra that are richer in information 17 18 In this method named cold electron ionization cold EI the molecules exit the GC column mixed with added helium make up gas and expand into vacuum through a specially designed supersonic nozzle forming a supersonic molecular beam SMB Collisions with the make up gas at the expanding supersonic jet reduce the internal vibrational and rotational energy of the analyte molecules hence reducing the degree of fragmentation caused by the electrons during the ionization process 17 18 Cold EI mass spectra are characterized by an abundant molecular ion while the usual fragmentation pattern is retained thus making cold EI mass spectra compatible with library search identification techniques The enhanced molecular ions increase the identification probabilities of both known and unknown compounds amplify isomer mass spectral effects and enable the use of isotope abundance analysis for the elucidation of elemental formulas 19 Chemical ionization Edit Main article Chemical ionization In chemical ionization CI a reagent gas typically methane or ammonia is introduced into the mass spectrometer Depending on the technique positive CI or negative CI chosen this reagent gas will interact with the electrons and analyte and cause a soft ionization of the molecule of interest A softer ionization fragments the molecule to a lower degree than the hard ionization of EI One of the main benefits of using chemical ionization is that a mass fragment closely corresponding to the molecular weight of the analyte of interest is produced In positive chemical ionization PCI the reagent gas interacts with the target molecule most often with a proton exchange This produces the species in relatively high amounts In negative chemical ionization NCI the reagent gas decreases the impact of the free electrons on the target analyte This decreased energy typically leaves the fragment in great supply Analysis EditA mass spectrometer is typically utilized in one of two ways full scan or selective ion monitoring SIM The typical GC MS instrument is capable of performing both functions either individually or concomitantly depending on the setup of the particular instrument The primary goal of instrument analysis is to quantify an amount of substance This is done by comparing the relative concentrations among the atomic masses in the generated spectrum Two kinds of analysis are possible comparative and original Comparative analysis essentially compares the given spectrum to a spectrum library to see if its characteristics are present for some sample in the library This is best performed by a computer because there are a myriad of visual distortions that can take place due to variations in scale Computers can also simultaneously correlate more data such as the retention times identified by GC to more accurately relate certain data Deep learning was shown to lead to promising results in the identification of VOCs from raw GC MS data 20 Another method of analysis measures the peaks in relation to one another In this method the tallest peak is assigned 100 of the value and the other peaks being assigned proportionate values All values above 3 are assigned The total mass of the unknown compound is normally indicated by the parent peak The value of this parent peak can be used to fit with a chemical formula containing the various elements which are believed to be in the compound The isotope pattern in the spectrum which is unique for elements that have many natural isotopes can also be used to identify the various elements present Once a chemical formula has been matched to the spectrum the molecular structure and bonding can be identified and must be consistent with the characteristics recorded by GC MS Typically this identification is done automatically by programs which come with the instrument given a list of the elements which could be present in the sample A full spectrum analysis considers all the peaks within a spectrum Conversely selective ion monitoring SIM only monitors selected ions associated with a specific substance This is done on the assumption that at a given retention time a set of ions is characteristic of a certain compound This is a fast and efficient analysis especially if the analyst has previous information about a sample or is only looking for a few specific substances When the amount of information collected about the ions in a given gas chromatographic peak decreases the sensitivity of the analysis increases So SIM analysis allows for a smaller quantity of a compound to be detected and measured but the degree of certainty about the identity of that compound is reduced Full scan MS Edit When collecting data in the full scan mode a target range of mass fragments is determined and put into the instrument s method An example of a typical broad range of mass fragments to monitor would be m z 50 to m z 400 The determination of what range to use is largely dictated by what one anticipates being in the sample while being cognizant of the solvent and other possible interferences A MS should not be set to look for mass fragments too low or else one may detect air found as m z 28 due to nitrogen carbon dioxide m z 44 or other possible interference Additionally if one is to use a large scan range then sensitivity of the instrument is decreased due to performing fewer scans per second since each scan will have to detect a wide range of mass fragments Full scan is useful in determining unknown compounds in a sample It provides more information than SIM when it comes to confirming or resolving compounds in a sample During instrument method development it may be common to first analyze test solutions in full scan mode to determine the retention time and the mass fragment fingerprint before moving to a SIM instrument method Selective ion monitoring Edit In selective ion monitoring SIM certain ion fragments are entered into the instrument method and only those mass fragments are detected by the mass spectrometer The advantages of SIM are that the detection limit is lower since the instrument is only looking at a small number of fragments e g three fragments during each scan More scans can take place each second Since only a few mass fragments of interest are being monitored matrix interferences are typically lower To additionally confirm the likelihood of a potentially positive result it is relatively important to be sure that the ion ratios of the various mass fragments are comparable to a known reference standard Applications EditEnvironmental monitoring and cleanup Edit GC MS is becoming the tool of choice for tracking organic pollutants in the environment The cost of GC MS equipment has decreased significantly and the reliability has increased at the same time which has contributed to its increased adoption in environmental studies Criminal forensics Edit GC MS can analyze the particles from a human body in order to help link a criminal to a crime The analysis of fire debris using GC MS is well established and there is even an established American Society for Testing and Materials ASTM standard for fire debris analysis GCMS MS is especially useful here as samples often contain very complex matrices and results used in court need to be highly accurate Law enforcement Edit GC MS is increasingly used for detection of illegal narcotics and may eventually supplant drug sniffing dogs 1 A simple and selective GC MS method for detecting marijuana usage was recently developed by the Robert Koch Institute in Germany It involves identifying an acid metabolite of tetrahydrocannabinol THC the active ingredient in marijuana in urine samples by employing derivatization in the sample preparation 21 GC MS is also commonly used in forensic toxicology to find drugs and or poisons in biological specimens of suspects victims or the deceased In drug screening GC MS methods frequently utilize liquid liquid extraction as a part of sample preparation in which target compounds are extracted from blood plasma 22 Sports anti doping analysis Edit GC MS is the main tool used in sports anti doping laboratories to test athletes urine samples for prohibited performance enhancing drugs for example anabolic steroids 23 Security Edit A post September 11 development explosive detection systems have become a part of all US airports These systems run on a host of technologies many of them based on GC MS There are only three manufacturers certified by the FAA to provide these systems citation needed one of which is Thermo Detection formerly Thermedics which produces the EGIS a GC MS based line of explosives detectors The other two manufacturers are Barringer Technologies now owned by Smith s Detection Systems and Ion Track Instruments part of General Electric Infrastructure Security Systems Chemical warfare agent detection Edit As part of the post September 11 drive towards increased capability in homeland security and public health preparedness traditional GC MS units with transmission quadrupole mass spectrometers as well as those with cylindrical ion trap CIT MS and toroidal ion trap T ITMS mass spectrometers have been modified for field portability and near real time detection of chemical warfare agents CWA such as sarin soman and VX 24 These complex and large GC MS systems have been modified and configured with resistively heated low thermal mass LTM gas chromatographs that reduce analysis time to less than ten percent of the time required in traditional laboratory systems 25 Additionally the systems are smaller and more mobile including units that are mounted in mobile analytical laboratories MAL such as those used by the United States Marine Corps Chemical and Biological Incident Response Force MAL and other similar laboratories and systems that are hand carried by two person teams or individuals much ado to the smaller mass detectors 26 Depending on the system the analytes can be introduced via liquid injection desorbed from sorbent tubes through a thermal desorption process or with solid phase micro extraction SPME Chemical engineering Edit GC MS is used for the analysis of unknown organic compound mixtures One critical use of this technology is the use of GC MS to determine the composition of bio oils processed from raw biomass 27 GC MS is also utilized in the identification of continuous phase component in a smart material Magnetorheological MR fluid 28 Food beverage and perfume analysis Edit Foods and beverages contain numerous aromatic compounds some naturally present in the raw materials and some forming during processing GC MS is extensively used for the analysis of these compounds which include esters fatty acids alcohols aldehydes terpenes etc It is also used to detect and measure contaminants from spoilage or adulteration which may be harmful and which is often controlled by governmental agencies for example pesticides Astrochemistry Edit Several GC MS have left earth Two were brought to Mars by the Viking program 29 Venera 11 and 12 and Pioneer Venus analysed the atmosphere of Venus with GC MS 30 The Huygens probe of the Cassini Huygens mission landed one GC MS on Saturn s largest moon Titan 31 The MSL Curiosity rover s Sample analysis at Mars SAM instrument contains both a gas chromatograph and quadrupole mass spectrometer that can be used in tandem as a GC MS 32 The material in the comet 67P Churyumov Gerasimenko was analysed by the Rosetta mission with a chiral GC MS in 2014 33 Medicine Edit Dozens of congenital metabolic diseases also known as inborn errors of metabolism IEM are now detectable by newborn screening tests especially the testing using gas chromatography mass spectrometry GC MS can determine compounds in urine even in minor concentration These compounds are normally not present but appear in individuals suffering with metabolic disorders This is increasingly becoming a common way to diagnose IEM for earlier diagnosis and institution of treatment eventually leading to a better outcome It is now possible to test a newborn for over 100 genetic metabolic disorders by a urine test at birth based on GC MS In combination with isotopic labeling of metabolic compounds the GC MS is used for determining metabolic activity Most applications are based on the use of 13C as the labeling and the measurement of 13C 12C ratios with an isotope ratio mass spectrometer IRMS an MS with a detector designed to measure a few select ions and return values as ratios See also EditCapillary electrophoresis mass spectrometry Ion mobility spectrometry mass spectrometry Liquid chromatography mass spectrometry Prolate trochoidal mass spectrometer Pyrolysis gas chromatography mass spectrometryReferences Edit Sparkman DO Penton Z Kitson FG 17 May 2011 Gas Chromatography and Mass Spectrometry A Practical Guide Academic Press ISBN 978 0 08 092015 3 Jones M Gas Chromatography Mass Spectrometry American Chemical Society Retrieved 19 Nov 2019 Fang M Ivanisevic J Benton HP Johnson CH Patti GJ Hoang LT et al November 2015 Thermal Degradation of Small Molecules A Global Metabolomic Investigation Analytical Chemistry 87 21 10935 41 doi 10 1021 acs analchem 5b03003 PMC 4633772 PMID 26434689 Holmes JC Morrell FA 1957 Oscillographic Mass Spectrometric Monitoring of Gas Chromatography Applied Spectroscopy 11 2 86 87 doi 10 1366 000370257774633394 ISSN 0003 7028 Gohlke RS 1959 Time of Flight Mass Spectrometry and Gas Liquid Partition Chromatography Analytical Chemistry 31 4 535 541 doi 10 1021 ac50164a024 ISSN 0003 2700 Patton HW Lewis JS Kaye WI 1955 Separation and Analysis of Gases and Volatile Liquids by Gas Chromatography Analytical Chemistry 27 2 170 174 doi 10 1021 ac60098a002 a b c Brock DC 2011 A Measure of Success Chemical Heritage Magazine 29 1 Retrieved 22 March 2018 Webb Halpern L 2008 Detecting Success Chemical Heritage Magazine 26 2 31 Thermo Instrument Systems Inc History International Directory of Company Histories Volume 11 ed St James Press 1995 pp 513 514 Retrieved 23 January 2015 Optimizing the Analysis of Volatile Organic Compounds Technical Guide Restek Corporation Lit Cat 59887A Wang T Lenahan R April 1984 Determination of volatile halocarbons in water by purge closed loop gas chromatography Bulletin of Environmental Contamination and Toxicology 32 4 429 38 doi 10 1007 BF01607519 PMID 6713137 S2CID 992748 Stauffer DB McLafferty FW Ellis RD Peterson DW 1974 Probability based matching of mass spectra Rapid identification of specific compounds in mixtures Organic Mass Spectrometry 9 4 690 702 doi 10 1002 oms 1210090710 Stein SE Scott DR September 1994 Optimization and testing of mass spectral library search algorithms for compound identification Journal of the American Society for Mass Spectrometry 5 9 859 66 doi 10 1016 1044 0305 94 87009 8 PMID 24222034 Standard Reference Data nist gov Wiley s Scientific Technical and Medical Databases Home wiley com Mass Spectrometry Database Committee ualberta ca a b Amirav A Gordin A Poliak M Fialkov AB February 2008 Gas chromatography mass spectrometry with supersonic molecular beams Journal of Mass Spectrometry 43 2 141 63 Bibcode 2008JMSp 43 141A doi 10 1002 jms 1380 PMID 18225851 a b SMB MS Supersonic GC MS tau ac il Alon T Amirav A 2006 Isotope abundance analysis methods and software for improved sample identification with supersonic gas chromatography mass spectrometry Rapid Communications in Mass Spectrometry 20 17 2579 88 Bibcode 2006RCMS 20 2579A doi 10 1002 rcm 2637 PMID 16897787 Skarysz A July 2018 Convolutional neural networks for automated targeted analysis of raw gas chromatography mass spectrometry data International Joint Conferences on Neural Networks 2018 Rio de Janeiro Brazil 1 8 doi 10 1109 IJCNN 2018 8489539 ISBN 978 1 5090 6014 6 S2CID 52989098 Hubschmann HJ 22 April 2015 Handbook of GC MS Fundamentals and Applications 3 ed John Wiley amp Sons Incorporated p 735 ISBN 9783527674336 Retrieved 22 January 2018 Hubschmann HJ 22 April 2015 Handbook of GC MS Fundamentals and Applications 3 ed John Wiley amp Sons Incorporated p 731 ISBN 9783527674336 Retrieved 22 January 2018 Tsivou M Kioukia Fougia N Lyris E Aggelis Y Fragkaki A Kiousi X et al 2006 An overview of the doping control analysis during the Olympic Games of 2004 in Athens Greece Analytica Chimica Acta 555 1 13 doi 10 1016 j aca 2005 08 068 Smith PA Lepage CJ Lukacs M Martin N Shufutinsky A Savage PB 2010 Field portable gas chromatography with transmission quadrupole and cylindrical ion trap mass spectrometric detection Chromatographic retention index data and ion molecule interactions for chemical warfare agent identification International Journal of Mass Spectrometry 295 3 113 118 Bibcode 2010IJMSp 295 113S doi 10 1016 j ijms 2010 03 001 Sloan KM Mustacich RV Eckenrode BA 2001 Development and evaluation of a low thermal mass gas chromatograph for rapid forensic GC MS analyses Field Analytical Chemistry amp Technology 5 6 288 301 doi 10 1002 fact 10011 Patterson GE Guymon AJ Riter LS Everly M Griep Raming J Laughlin BC et al December 2002 Miniature cylindrical ion trap mass spectrometer Analytical Chemistry 74 24 6145 53 doi 10 1021 ac020494d PMID 12510732 Tekin K Karagoz S Bektas S 2014 12 01 A review of hydrothermal biomass processing Renewable and Sustainable Energy Reviews 40 673 687 doi 10 1016 j rser 2014 07 216 Unuh MH Muhamad P Waziralilah NF Amran MH 2019 Characterization of Vehicle Smart Fluid using Gas Chromatography Mass Spectrometry GCMS PDF Journal of Advanced Research in Fluid Mechanics and Thermal Sciences 55 2 240 248 SEARCHING FOR LIFE ON MARS The Development of the Viking GCMS NASA Krasnopolsky VA Parshev VA 1981 Chemical composition of the atmosphere of Venus Nature 292 5824 610 613 Bibcode 1981Natur 292 610K doi 10 1038 292610a0 S2CID 4369293 Niemann HB Atreya SK Bauer SJ Carignan GR Demick JE Frost RL et al December 2005 The abundances of constituents of Titan s atmosphere from the GCMS instrument on the Huygens probe PDF Nature 438 7069 779 84 Bibcode 2005Natur 438 779N doi 10 1038 nature04122 hdl 2027 42 62703 PMID 16319830 S2CID 4344046 MSL Science Corner Sample Analysis at Mars SAM msl scicorner jpl nasa gov Archived from the original on 2009 03 20 Retrieved 2019 06 25 Gosmann F Rosenbauer H Roll R Bohnhardt H October 2005 COSAC onboard Rosetta a bioastronomy experiment for the short period comet 67P Churyumov Gerasimenko Astrobiology 5 5 622 31 Bibcode 2005AsBio 5 622G doi 10 1089 ast 2005 5 622 PMID 16225435 Bibliography EditAdams RP 2007 Identification of Essential Oil Components By Gas Chromatography Mass Spectrometry Allured Pub Corp ISBN 978 1 932633 21 4 Adlard ER Handley AJ 2001 Gas chromatographic techniques and applications London Sheffield Academic ISBN 978 0 8493 0521 4 Barry EF Grob RE 2004 Modern practice of gas chromatography New York Wiley Interscience ISBN 978 0 471 22983 4 Eiceman GA 2000 Gas Chromatography In Meyers RA ed Encyclopedia of Analytical Chemistry Applications Theory and Instrumentation Chichester Wiley p 10627 ISBN 0 471 97670 9 Giannelli PC Imwinkelried EJ 1999 Drug Identification Gas Chromatography Scientific Evidence Vol 2 Charlottesville Lexis Law Publishing p 362 ISBN 0 327 04985 5 McEwen CN Kitson FG Larsen BS 1996 Gas chromatography and mass spectrometry a practical guide Boston Academic Press ISBN 978 0 12 483385 2 McMaster C McMaster MC 1998 GC MS a practical user s guide New York Wiley ISBN 978 0 471 24826 2 Message GM 1984 Practical aspects of gas chromatography mass spectrometry New York Wiley ISBN 978 0 471 06277 6 Niessen WM 2001 Current practice of gas chromatography mass spectrometry New York N Y Marcel Dekker ISBN 978 0 8247 0473 5 Weber A Maurer HW Pfleger K 2007 Mass Spectral and GC Data of Drugs Poisons Pesticides Pollutants and Their Metabolites Weinheim Wiley VCH ISBN 978 3 527 31538 3 External links EditGas chromatography mass spectrometry at the US National Library of Medicine Medical Subject Headings MeSH Golm Metabolome Database a mass spectral reference database of plant metabolites Retrieved from https en wikipedia org w index php title Gas chromatography mass spectrometry amp oldid 1131387091, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.