A semiconductor detector in ionizing radiation detection physics is a device that uses a semiconductor (usually silicon or germanium) to measure the effect of incident charged particles or photons.

Semiconductor detectors find broad application for radiation protection, gamma and X-ray spectrometry, and as particle detectors.

Detection mechanism edit

In semiconductor detectors, ionizing radiation is measured by the number of charge carriers set free in the detector material which is arranged between two electrodes, by the radiation. Ionizing radiation produces free electrons and electron holes. The number of electron-hole pairs is proportional to the energy of the radiation to the semiconductor. As a result, a number of electrons are transferred from the valence band to the conduction band, and an equal number of holes are created in the valence band. Under the influence of an electric field, electrons and holes travel to the electrodes, where they result in a pulse that can be measured in an outer circuit, as described by the Shockley-Ramo theorem. The holes travel in the opposite direction and can also be measured. As the amount of energy required to create an electron-hole pair is known, and is independent of the energy of the incident radiation, measuring the number of electron-hole pairs allows the energy of the incident radiation to be determined.[1]

The energy required to produce electron-hole-pairs is very low compared to the energy required to produce paired ions in a gas detector. Consequently, in semiconductor detectors the statistical variation of the pulse height is smaller and the energy resolution is higher. As the electrons travel fast, the time resolution is also very good, and is dependent upon rise time.[2] Compared with gaseous ionization detectors, the density of a semiconductor detector is very high, and charged particles of high energy can give off their energy in a semiconductor of relatively small dimensions.[citation needed]

Detector types edit

Silicon detectors edit

 
A Forward Silicon Vertex Detector (FVTX) sensor of PHENIX detector on a microscope showing silicon strips spacing at 75 microns.[3]

Most silicon particle detectors work, in principle, by doping narrow (usually around 100 micrometers wide) silicon strips to turn them into diodes, which are then reverse biased. As charged particles pass through these strips, they cause small ionization currents that can be detected and measured. Arranging thousands of these detectors around a collision point in a particle accelerator can yield an accurate picture of what paths particles take. Silicon detectors have a much higher resolution in tracking charged particles than older technologies such as cloud chambers or wire chambers. The drawback is that silicon detectors are much more expensive than these older technologies and require sophisticated cooling to reduce leakage currents (noise source). They also suffer degradation over time from radiation, however, this can be greatly reduced thanks to the Lazarus effect.

Diamond detectors edit

Diamond detectors have many similarities with silicon detectors but are expected to offer significant advantages – in particular a high radiation hardness and very low drift currents. They are also suited to neutron detection. At present, however, they are much more expensive and more difficult to manufacture.

Germanium detectors edit

 
High-purity germanium detector (disconnected from liquid nitrogen dewar)

Germanium detectors are mostly used for gamma spectroscopy in nuclear physics, as well as x-ray spectroscopy. While silicon detectors cannot be thicker than a few millimeters, germanium can have a sensitive layer (depletion region) thickness of centimeters, and therefore can be used as a total absorption detector for gamma rays up to a few MeV. These detectors are also called high-purity germanium detectors (HPGe) or hyperpure germanium detectors. Before current purification techniques were refined, germanium crystals could not be produced with purity sufficient to enable their use as spectroscopy detectors. Impurities in the crystals trap electrons and holes, ruining the performance of the detectors. Consequently, germanium crystals were doped with lithium ions (Ge(Li)), in order to produce an intrinsic region in which the electrons and holes would be able to reach the contacts and produce a signal.

When germanium detectors were first developed, only very small crystals were available. Low efficiency was the result, and germanium detector efficiency is still often quoted in relative terms to a "standard" 3″ x 3″ NaI(Tl) scintillation detector. Crystal growth techniques have since improved, allowing detectors to be manufactured that are as large as or larger than commonly available NaI crystals, although such detectors cost more than €100,000 (US$113,000).

As of 2012, HPGe detectors commonly use lithium diffusion to make an n+ ohmic contact, and boron implantation to make a p+ contact. Coaxial detectors with a central n+ contact are referred to as n-type detectors, while p-type detectors have a p+ central contact. The thickness of these contacts represents a dead layer around the surface of the crystal within which energy depositions do not result in detector signals. The central contact in these detectors is opposite to the surface contact, making the dead layer in n-type detectors smaller than the dead layer in p-type detectors. Typical dead layer thicknesses are several hundred micrometers for a Li diffusion layer and a few tenths of a micrometer for a B implantation layer.

The major drawback of germanium detectors is that they must be cooled to liquid nitrogen temperatures to produce spectroscopic data. At higher temperatures, the electrons can easily cross the band gap in the crystal and reach the conduction band, where they are free to respond to the electric field, producing too much electrical noise to be useful as a spectrometer. Cooling to liquid nitrogen temperature (77K) reduces thermal excitations of valence electrons so that only a gamma ray interaction can give an electron the energy necessary to cross the band gap and reach the conduction band. Cooling with liquid nitrogen is inconvenient, as the detector requires hours to cool down to operating temperature before it can be used, and cannot be allowed to warm up during use. Ge(Li) crystals could never be allowed to warm up, as the lithium would drift out of the crystal, ruining the detector. HPGe detectors can be allowed to warm up to room temperature when not in use.

Commercial systems became available that use advanced refrigeration techniques (for example pulse tube refrigerator) to eliminate the need for liquid nitrogen cooling.

Germanium detectors with multi-strip electrodes, orthogonal on opposing faces, can indicate the 2-D location of the ionization trail within a large single crystal of Ge. Detectors like this have been used in COSI balloon-born astronomy missions (NASA, 2016) and will be used in an orbital observatory (NASA, 2025) Compton Spectrometer and Imager (COSI).

Cadmium telluride and cadmium zinc telluride detectors edit

Cadmium telluride (CdTe) and cadmium zinc telluride (CZT) detectors have been developed for use in X-ray spectroscopy and gamma spectroscopy. The high density of these materials means they can effectively attenuate X-rays and gamma-rays with energies of greater than 20 keV that traditional silicon-based sensors are unable to detect. The wide band gap of these materials also means they have high resistivity and are able to operate at, or close to, room temperature (~295K) unlike germanium-based sensors. These detector materials can be used to produce sensors with different electrode structures for imaging and high-resolution spectroscopy. However, CZT detectors are generally unable to match the resolution of germanium detectors, with some of this difference being attributable to poor positive charge-carrier transport to the electrode. Efforts to mitigate this effect have included the development of novel electrodes to negate the need for both polarities of carriers to be collected.[4][5]

Integrated Systems edit

Semiconductor detectors are often commercially integrated into larger systems for various radiation measurement applications.

Automated Sample Changing for Germanium Detectors edit

 
HPGe automated with a low-cost, open-source autosampler.

Gamma spectrometers using HPGe detectors are often used for measurement of low levels of gamma-emitting radionuclides in environmental samples, which requires a low background environment, usually achieved by enclosing the sample and detector in a lead shield known as a 'lead castle'. Automated systems have been developed [6] to sequentially move a number of samples into and out of the lead castle for measurement. Due to the complexities of opening the shield and moving the samples, this automation has traditionally been expensive, but lower-cost autosamplers have recently been introduced.[7]

Radioactive Waste Assay Machines edit

Semiconductor detectors especially HPGe are often integrated into devices for characterising packaged radioactive waste. This can be as simple as detectors being mounted on a moveable platform to be brought to an area for in-situ measurements and paired with shielding to restrict the field-of-view of the detector to the area of interest for one-shot "open detector geometry" measurements,[8] or for waste in drums, systems such as the Segmented Gamma Scanner (SGS) combine a semiconductor detector with integrated mechatronics to rotate the item and scan the detector across different sections.[9] If the detector field of view is scanned across small areas of the item in multiple axes as is done with a Tomographic Gamma Scanner (TGS), Tomography can be used to extract 3D information about the density and gamma emissions of the item.[10]

Gamma Cameras edit

Semiconductor detectors are used in some Gamma Cameras and Gamma imaging systems [11]

See also edit

References edit

  1. ^ Knoll, G.F. (1999). Radiation Detection and Measurement (3rd ed.). Wiley. p. 365. ISBN 978-0-471-07338-3.
  2. ^ Knoll, p119
  3. ^ Kapustinsky, Jon S. (17 November 2010). "Sensors/FPHX Readout Chip WBS 1.4.1/1.4.2" (PDF). Retrieved 7 August 2017. {{cite journal}}: Cite journal requires |journal= (help)
  4. ^ Luke, P. N. (1 November 1994). "Unipolar charge sensing with coplanar electrodes -- Application to semiconductor detectors". doi:10.2172/34411. OSTI 34411. {{cite journal}}: Cite journal requires |journal= (help)
  5. ^ J. S. Kapustinsky, Nucl. Instrum. Methods A 617 (2010) 546 – 548.
  6. ^ "Robotic Gamma Spectrometer with sample changer – Nuclear System".
  7. ^ Carvalho, Matheus (2018). "Auto-HPGe, an autosampler for gamma-ray spectroscopy using high-purity germanium (HPGe) detectors and heavy shields". HardwareX. 4: e00040. doi:10.1016/j.ohx.2018.e00040.
  8. ^ ISO 19017:2015
  9. ^ LALP-91-011, Application Note, Segmented Gamma-Ray Scanner
  10. ^ LA-UR-93-1637, Tomographic gamma scanning (TGS) to measure inhomogeneous nuclear material matrices from future fuel cycles, LANL, 1993
  11. ^ "H100 - H3D, Inc".

External links edit

  • Silicon Detector powerpoint delivered for EDIT (Excellence in Detectors and Instrumentation Technologies) 2011 at CERN, M. Krammer, F. Hartmann.

semiconductor, detector, this, article, about, ionizing, radiation, detectors, semiconductor, detectors, radio, detector, radio, crystal, detector, semiconductor, diodes, rectifier, semiconductor, detector, ionizing, radiation, detection, physics, device, that. This article is about ionizing radiation detectors For semiconductor detectors in radio see Detector radio Crystal detector Semiconductor diodes and Rectifier A semiconductor detector in ionizing radiation detection physics is a device that uses a semiconductor usually silicon or germanium to measure the effect of incident charged particles or photons Semiconductor detectors find broad application for radiation protection gamma and X ray spectrometry and as particle detectors Contents 1 Detection mechanism 2 Detector types 2 1 Silicon detectors 2 2 Diamond detectors 2 3 Germanium detectors 2 4 Cadmium telluride and cadmium zinc telluride detectors 3 Integrated Systems 3 1 Automated Sample Changing for Germanium Detectors 3 2 Radioactive Waste Assay Machines 3 3 Gamma Cameras 4 See also 5 References 6 External linksDetection mechanism editIn semiconductor detectors ionizing radiation is measured by the number of charge carriers set free in the detector material which is arranged between two electrodes by the radiation Ionizing radiation produces free electrons and electron holes The number of electron hole pairs is proportional to the energy of the radiation to the semiconductor As a result a number of electrons are transferred from the valence band to the conduction band and an equal number of holes are created in the valence band Under the influence of an electric field electrons and holes travel to the electrodes where they result in a pulse that can be measured in an outer circuit as described by the Shockley Ramo theorem The holes travel in the opposite direction and can also be measured As the amount of energy required to create an electron hole pair is known and is independent of the energy of the incident radiation measuring the number of electron hole pairs allows the energy of the incident radiation to be determined 1 The energy required to produce electron hole pairs is very low compared to the energy required to produce paired ions in a gas detector Consequently in semiconductor detectors the statistical variation of the pulse height is smaller and the energy resolution is higher As the electrons travel fast the time resolution is also very good and is dependent upon rise time 2 Compared with gaseous ionization detectors the density of a semiconductor detector is very high and charged particles of high energy can give off their energy in a semiconductor of relatively small dimensions citation needed Detector types editThis section does not cite any sources Please help improve this section by adding citations to reliable sources Unsourced material may be challenged and removed November 2020 Learn how and when to remove this template message Silicon detectors edit nbsp A Forward Silicon Vertex Detector FVTX sensor of PHENIX detector on a microscope showing silicon strips spacing at 75 microns 3 Most silicon particle detectors work in principle by doping narrow usually around 100 micrometers wide silicon strips to turn them into diodes which are then reverse biased As charged particles pass through these strips they cause small ionization currents that can be detected and measured Arranging thousands of these detectors around a collision point in a particle accelerator can yield an accurate picture of what paths particles take Silicon detectors have a much higher resolution in tracking charged particles than older technologies such as cloud chambers or wire chambers The drawback is that silicon detectors are much more expensive than these older technologies and require sophisticated cooling to reduce leakage currents noise source They also suffer degradation over time from radiation however this can be greatly reduced thanks to the Lazarus effect Diamond detectors edit Diamond detectors have many similarities with silicon detectors but are expected to offer significant advantages in particular a high radiation hardness and very low drift currents They are also suited to neutron detection At present however they are much more expensive and more difficult to manufacture Germanium detectors edit nbsp High purity germanium detector disconnected from liquid nitrogen dewar Germanium detectors are mostly used for gamma spectroscopy in nuclear physics as well as x ray spectroscopy While silicon detectors cannot be thicker than a few millimeters germanium can have a sensitive layer depletion region thickness of centimeters and therefore can be used as a total absorption detector for gamma rays up to a few MeV These detectors are also called high purity germanium detectors HPGe or hyperpure germanium detectors Before current purification techniques were refined germanium crystals could not be produced with purity sufficient to enable their use as spectroscopy detectors Impurities in the crystals trap electrons and holes ruining the performance of the detectors Consequently germanium crystals were doped with lithium ions Ge Li in order to produce an intrinsic region in which the electrons and holes would be able to reach the contacts and produce a signal When germanium detectors were first developed only very small crystals were available Low efficiency was the result and germanium detector efficiency is still often quoted in relative terms to a standard 3 x 3 NaI Tl scintillation detector Crystal growth techniques have since improved allowing detectors to be manufactured that are as large as or larger than commonly available NaI crystals although such detectors cost more than 100 000 US 113 000 As of 2012 update HPGe detectors commonly use lithium diffusion to make an n ohmic contact and boron implantation to make a p contact Coaxial detectors with a central n contact are referred to as n type detectors while p type detectors have a p central contact The thickness of these contacts represents a dead layer around the surface of the crystal within which energy depositions do not result in detector signals The central contact in these detectors is opposite to the surface contact making the dead layer in n type detectors smaller than the dead layer in p type detectors Typical dead layer thicknesses are several hundred micrometers for a Li diffusion layer and a few tenths of a micrometer for a B implantation layer The major drawback of germanium detectors is that they must be cooled to liquid nitrogen temperatures to produce spectroscopic data At higher temperatures the electrons can easily cross the band gap in the crystal and reach the conduction band where they are free to respond to the electric field producing too much electrical noise to be useful as a spectrometer Cooling to liquid nitrogen temperature 77K reduces thermal excitations of valence electrons so that only a gamma ray interaction can give an electron the energy necessary to cross the band gap and reach the conduction band Cooling with liquid nitrogen is inconvenient as the detector requires hours to cool down to operating temperature before it can be used and cannot be allowed to warm up during use Ge Li crystals could never be allowed to warm up as the lithium would drift out of the crystal ruining the detector HPGe detectors can be allowed to warm up to room temperature when not in use Commercial systems became available that use advanced refrigeration techniques for example pulse tube refrigerator to eliminate the need for liquid nitrogen cooling Germanium detectors with multi strip electrodes orthogonal on opposing faces can indicate the 2 D location of the ionization trail within a large single crystal of Ge Detectors like this have been used in COSI balloon born astronomy missions NASA 2016 and will be used in an orbital observatory NASA 2025 Compton Spectrometer and Imager COSI Cadmium telluride and cadmium zinc telluride detectors edit Cadmium telluride CdTe and cadmium zinc telluride CZT detectors have been developed for use in X ray spectroscopy and gamma spectroscopy The high density of these materials means they can effectively attenuate X rays and gamma rays with energies of greater than 20 keV that traditional silicon based sensors are unable to detect The wide band gap of these materials also means they have high resistivity and are able to operate at or close to room temperature 295K unlike germanium based sensors These detector materials can be used to produce sensors with different electrode structures for imaging and high resolution spectroscopy However CZT detectors are generally unable to match the resolution of germanium detectors with some of this difference being attributable to poor positive charge carrier transport to the electrode Efforts to mitigate this effect have included the development of novel electrodes to negate the need for both polarities of carriers to be collected 4 5 Integrated Systems editSemiconductor detectors are often commercially integrated into larger systems for various radiation measurement applications Automated Sample Changing for Germanium Detectors edit nbsp HPGe automated with a low cost open source autosampler Gamma spectrometers using HPGe detectors are often used for measurement of low levels of gamma emitting radionuclides in environmental samples which requires a low background environment usually achieved by enclosing the sample and detector in a lead shield known as a lead castle Automated systems have been developed 6 to sequentially move a number of samples into and out of the lead castle for measurement Due to the complexities of opening the shield and moving the samples this automation has traditionally been expensive but lower cost autosamplers have recently been introduced 7 Radioactive Waste Assay Machines edit Semiconductor detectors especially HPGe are often integrated into devices for characterising packaged radioactive waste This can be as simple as detectors being mounted on a moveable platform to be brought to an area for in situ measurements and paired with shielding to restrict the field of view of the detector to the area of interest for one shot open detector geometry measurements 8 or for waste in drums systems such as the Segmented Gamma Scanner SGS combine a semiconductor detector with integrated mechatronics to rotate the item and scan the detector across different sections 9 If the detector field of view is scanned across small areas of the item in multiple axes as is done with a Tomographic Gamma Scanner TGS Tomography can be used to extract 3D information about the density and gamma emissions of the item 10 Gamma Cameras edit Semiconductor detectors are used in some Gamma Cameras and Gamma imaging systems 11 See also editLazarus effect Pandemonium effect Synthetic diamonds Total absorption spectroscopy X ray spectroscopy Microstrip detector Hybrid pixel detector Liulin type instrumentsReferences edit Knoll G F 1999 Radiation Detection and Measurement 3rd ed Wiley p 365 ISBN 978 0 471 07338 3 Knoll p119 Kapustinsky Jon S 17 November 2010 Sensors FPHX Readout Chip WBS 1 4 1 1 4 2 PDF Retrieved 7 August 2017 a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help Luke P N 1 November 1994 Unipolar charge sensing with coplanar electrodes Application to semiconductor detectors doi 10 2172 34411 OSTI 34411 a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help J S Kapustinsky Nucl Instrum Methods A 617 2010 546 548 Robotic Gamma Spectrometer with sample changer Nuclear System Carvalho Matheus 2018 Auto HPGe an autosampler for gamma ray spectroscopy using high purity germanium HPGe detectors and heavy shields HardwareX 4 e00040 doi 10 1016 j ohx 2018 e00040 ISO 19017 2015 LALP 91 011 Application Note Segmented Gamma Ray Scanner LA UR 93 1637 Tomographic gamma scanning TGS to measure inhomogeneous nuclear material matrices from future fuel cycles LANL 1993 H100 H3D Inc External links editSilicon Detector powerpoint delivered for EDIT Excellence in Detectors and Instrumentation Technologies 2011 at CERN M Krammer F Hartmann Retrieved from https en wikipedia org w index php title Semiconductor detector amp oldid 1189599930, wikipedia, wiki, book, books, library,