Electrocardiography

Electrocardiography is the process of producing an electrocardiogram (ECG or EKG[a]), a recording of the heart's electrical activity through repeated cardiac cycles.[4] It is an electrogram of the heart which is a graph of voltage versus time of the electrical activity of the heart[5] using electrodes placed on the skin. These electrodes detect the small electrical changes that are a consequence of cardiac muscle depolarization followed by repolarization during each cardiac cycle (heartbeat). Changes in the normal ECG pattern occur in numerous cardiac abnormalities, including:

Electrocardiography
ECG of a heart in normal sinus rhythm
ICD-10-PCSR94.31
ICD-9-CM89.52
MeSHD004562
MedlinePlus003868
[edit on Wikidata]
Use of real time monitoring of the heart in an intensive care unit in a German hospital (2015), the monitoring screen above the patient displaying an electrocardiogram and various values of parameters of the heart like heart rate and blood pressure

Traditionally, "ECG" usually means a 12-lead ECG taken while lying down as discussed below. However, other devices can record the electrical activity of the heart such as a Holter monitor but also some models of smartwatch are capable of recording an ECG. ECG signals can be recorded in other contexts with other devices.

In a conventional 12-lead ECG, ten electrodes are placed on the patient's limbs and on the surface of the chest. The overall magnitude of the heart's electrical potential is then measured from twelve different angles ("leads") and is recorded over a period of time (usually ten seconds). In this way, the overall magnitude and direction of the heart's electrical depolarization is captured at each moment throughout the cardiac cycle.[11]

There are three main components to an ECG:[12]

  • The P wave, which represents depolarization of the atria.
  • The QRS complex, which represents depolarization of the ventricles.
  • The T wave, which represents repolarization of the ventricles.

During each heartbeat, a healthy heart has an orderly progression of depolarization that starts with pacemaker cells in the sinoatrial node, spreads throughout the atrium, and passes through the atrioventricular node down into the bundle of His and into the Purkinje fibers, spreading down and to the left throughout the ventricles.[12] This orderly pattern of depolarization gives rise to the characteristic ECG tracing. To the trained clinician, an ECG conveys a large amount of information about the structure of the heart and the function of its electrical conduction system.[13] Among other things, an ECG can be used to measure the rate and rhythm of heartbeats, the size and position of the heart chambers, the presence of any damage to the heart's muscle cells or conduction system, the effects of heart drugs, and the function of implanted pacemakers.[14]

Medical uses edit

 
Normal 12-lead ECG
 
A 12-lead ECG of a 26-year-old male with an incomplete right bundle branch block (RBBB)

The overall goal of performing an ECG is to obtain information about the electrical functioning of the heart. Medical uses for this information are varied and often need to be combined with knowledge of the structure of the heart and physical examination signs to be interpreted. Some indications for performing an ECG include the following:

ECGs can be recorded as short intermittent tracings or continuous ECG monitoring. Continuous monitoring is used for critically ill patients, patients undergoing general anesthesia,[18][17] and patients who have an infrequently occurring cardiac arrhythmia that would unlikely be seen on a conventional ten-second ECG. Continuous monitoring can be conducted by using Holter monitors, internal and external defibrillators and pacemakers, and/or biotelemetry.[19]

Screening edit

 
A patient undergoing an ECG

For adults, evidence does not support the use of ECGs among those without symptoms or at low risk of cardiovascular disease as an effort for prevention.[20][21][22] This is because an ECG may falsely indicate the existence of a problem, leading to misdiagnosis, the recommendation of invasive procedures, and overtreatment. However, persons employed in certain critical occupations, such as aircraft pilots,[23] may be required to have an ECG as part of their routine health evaluations. Hypertrophic cardiomyopathy screening may also be considered in adolescents as part of a sports physical out of concern for sudden cardiac death.[24]

Electrocardiograph machines edit

 
An EKG electrode

Electrocardiograms are recorded by machines that consist of a set of electrodes connected to a central unit.[25] Early ECG machines were constructed with analog electronics, where the signal drove a motor to print out the signal onto paper. Today, electrocardiographs use analog-to-digital converters to convert the electrical activity of the heart to a digital signal. Many ECG machines are now portable and commonly include a screen, keyboard, and printer on a small wheeled cart. Recent advancements in electrocardiography include developing even smaller devices for inclusion in fitness trackers and smart watches.[26] These smaller devices often rely on only two electrodes to deliver a single lead I.[27] Portable twelve-lead devices powered by batteries are also available.

Recording an ECG is a safe and painless procedure.[28] The machines are powered by mains power but they are designed with several safety features including an earthed (ground) lead. Other features include:

  • Defibrillation protection: any ECG used in healthcare may be attached to a person who requires defibrillation and the ECG needs to protect itself from this source of energy.
  • Electrostatic discharge is similar to defibrillation discharge and requires voltage protection up to 18,000 volts.
  • Additionally, circuitry called the right leg driver can be used to reduce common-mode interference (typically the 50 or 60 Hz mains power).
  • ECG voltages measured across the body are very small. This low voltage necessitates a low noise circuit, instrumentation amplifiers, and electromagnetic shielding.
  • Simultaneous lead recordings: earlier designs recorded each lead sequentially, but current models record multiple leads simultaneously.

Most modern ECG machines include automated interpretation algorithms. This analysis calculates features such as the PR interval, QT interval, corrected QT (QTc) interval, PR axis, QRS axis, rhythm and more. The results from these automated algorithms are considered "preliminary" until verified and/or modified by expert interpretation. Despite recent advances, computer misinterpretation remains a significant problem and can result in clinical mismanagement.[29]

Cardiac monitors edit

Besides the standard electrocardiograph machine, there are other devices capable of recording ECG signals. Portable devices have existed since the Holter monitor was produced in 1962. Traditionally, these monitors have used electrodes with patches on the skin to record the ECG, but new devices can stick to the chest as a single patch without need for wires, developed by Zio (Zio XT), TZ Medical (Trident), Philips (BioTel) and BardyDx (CAM) among many others. Implantable devices such as the artificial cardiac pacemaker and implantable cardioverter-defibrillator are capable of measuring a "far field" signal between the leads in the heart and the implanted battery/generator that resembles an ECG signal (technically, the signal recorded in the heart is called an electrogram, which is interpreted differently). Advancement of the Holter monitor became the implantable loop recorder that performs the same function but in an implantable device with batteries that last on the order of years.

Additionally, there are available various Arduino kits with ECG sensor modules and smartwatch devices that are capable of recording an ECG signal as well, such as with the 4th generation Apple Watch, Samsung Galaxy Watch 4 and newer devices.

Electrodes and leads edit

 
Proper placement of the limb electrodes. The limb electrodes can be far down on the limbs or close to the hips/shoulders as long as they are placed symmetrically.[30]
 
Placement of the precordial electrodes

Electrodes are the actual conductive pads attached to the body surface.[31] Any pair of electrodes can measure the electrical potential difference between the two corresponding locations of attachment. Such a pair forms a lead. However, "leads" can also be formed between a physical electrode and a virtual electrode, known as Wilson's central terminal (WCT), whose potential is defined as the average potential measured by three limb electrodes that are attached to the right arm, the left arm, and the left foot, respectively.[32]

Commonly, 10 electrodes attached to the body are used to form 12 ECG leads, with each lead measuring a specific electrical potential difference (as listed in the table below).[33]

Leads are broken down into three types: limb; augmented limb; and precordial or chest. The 12-lead ECG has a total of three limb leads and three augmented limb leads arranged like spokes of a wheel in the coronal plane (vertical), and six precordial leads or chest leads that lie on the perpendicular transverse plane (horizontal).[34]

In medical settings, the term leads is also sometimes used to refer to the electrodes themselves, although this is technically incorrect.[35]

The 10 electrodes in a 12-lead ECG are listed below.[36]

Electrode name Electrode placement
RA On the right arm, avoiding thick muscle.
LA In the same location where RA was placed, but on the left arm.
RL On the right leg, lower end of inner aspect of calf muscle. (Avoid bony prominences)
LL In the same location where RL was placed, but on the left leg.
V1 In the fourth intercostal space (between ribs 4 and 5) just to the right of the sternum (breastbone)
V2 In the fourth intercostal space (between ribs 4 and 5) just to the left of the sternum.
V3 Between leads V2 and V4.
V4 In the fifth intercostal space (between ribs 5 and 6) in the mid-clavicular line.
V5 Horizontally even with V4, in the left anterior axillary line.
V6 Horizontally even with V4 and V5 in the mid-axillary line.

Two types of electrodes in common use are a flat paper-thin sticker and a self-adhesive circular pad. The former are typically used in a single ECG recording while the latter are for continuous recordings as they stick longer. Each electrode consists of an electrically conductive electrolyte gel and a silver/silver chloride conductor.[37] The gel typically contains potassium chloride – sometimes silver chloride as well – to permit electron conduction from the skin to the wire and to the electrocardiogram.[38]

The common virtual electrode, known as Wilson's central terminal (VW), is produced by averaging the measurements from the electrodes RA, LA, and LL to give an average potential of the body:

 

In a 12-lead ECG, all leads except the limb leads are assumed to be unipolar (aVR, aVL, aVF, V1, V2, V3, V4, V5, and V6). The measurement of a voltage requires two contacts and so, electrically, the unipolar leads are measured from the common lead (negative) and the unipolar lead (positive). This averaging for the common lead and the abstract unipolar lead concept makes for a more challenging understanding and is complicated by sloppy usage of "lead" and "electrode". In fact, instead of being a constant reference, VW has a value that fluctuates throughout the heart cycle. It also does not truly represent the center-of-heart potential due to the body parts the signals travel through.[39]

Limb leads edit

 
The limb leads and augmented limb leads (Wilson's central terminal is used as the negative pole for the latter in this representation)
 

Leads I, II and III are called the limb leads. The electrodes that form these signals are located on the limbs – one on each arm and one on the left leg.[40][41][42] The limb leads form the points of what is known as Einthoven's triangle.[43]

  • Lead I is the voltage between the (positive) left arm (LA) electrode and right arm (RA) electrode:
 
  • Lead II is the voltage between the (positive) left leg (LL) electrode and the right arm (RA) electrode:
 
  • Lead III is the voltage between the (positive) left leg (LL) electrode and the left arm (LA) electrode:
 

Augmented limb leads edit

Leads aVR, aVL, and aVF are the augmented limb leads. They are derived from the same three electrodes as leads I, II, and III, but they use Goldberger's central terminal as their negative pole. Goldberger's central terminal is a combination of inputs from two limb electrodes, with a different combination for each augmented lead. It is referred to immediately below as "the negative pole".

  • Lead augmented vector right (aVR) has the positive electrode on the right arm. The negative pole is a combination of the left arm electrode and the left leg electrode:
 
  • Lead augmented vector left (aVL) has the positive electrode on the left arm. The negative pole is a combination of the right arm electrode and the left leg electrode:
 
  • Lead augmented vector foot (aVF) has the positive electrode on the left leg. The negative pole is a combination of the right arm electrode and the left arm electrode:
 

Together with leads I, II, and III, augmented limb leads aVR, aVL, and aVF form the basis of the hexaxial reference system, which is used to calculate the heart's electrical axis in the frontal plane.[44]

Older versions of the nodes (VR, VL, VF) use Wilson's central terminal as the negative pole, but the amplitude is too small for the thick lines of old ECG machines. The Goldberger terminals scale up (augments) the Wilson results by 50%, at the cost of sacrificing physical correctness by not having the same negative pole for all three.[45]

Precordial leads edit

The precordial leads lie in the transverse (horizontal) plane, perpendicular to the other six leads. The six precordial electrodes act as the positive poles for the six corresponding precordial leads: (V1, V2, V3, V4, V5, and V6). Wilson's central terminal is used as the negative pole. Recently, unipolar precordial leads have been used to create bipolar precordial leads that explore the right to left axis in the horizontal plane.[46]

Specialized leads edit

Additional electrodes may rarely be placed to generate other leads for specific diagnostic purposes. Right-sided precordial leads may be used to better study pathology of the right ventricle or for dextrocardia (and are denoted with an R (e.g., V5R). Posterior leads (V7 to V9) may be used to demonstrate the presence of a posterior myocardial infarction. The Lewis lead or S5-lead (requiring an electrode at the right sternal border in the second intercostal space) can be used to better detect atrial activity in relation to that of the ventricles.[47]

An esophageal lead can be inserted to a part of the esophagus where the distance to the posterior wall of the left atrium is only approximately 5–6 mm (remaining constant in people of different age and weight).[48] An esophageal lead avails for a more accurate differentiation between certain cardiac arrhythmias, particularly atrial flutter, AV nodal reentrant tachycardia and orthodromic atrioventricular reentrant tachycardia.[49] It can also evaluate the risk in people with Wolff-Parkinson-White syndrome, as well as terminate supraventricular tachycardia caused by re-entry.[49]

An intracardiac electrogram (ICEG) is essentially an ECG with some added intracardiac leads (that is, inside the heart). The standard ECG leads (external leads) are I, II, III, aVL, V1, and V6. Two to four intracardiac leads are added via cardiac catheterization. The word "electrogram" (EGM) without further specification usually means an intracardiac electrogram.[50]

Lead locations on an ECG report edit

A standard 12-lead ECG report (an electrocardiograph) shows a 2.5 second tracing of each of the twelve leads. The tracings are most commonly arranged in a grid of four columns and three rows. The first column is the limb leads (I, II, and III), the second column is the augmented limb leads (aVR, aVL, and aVF), and the last two columns are the precordial leads (V1 to V6). Additionally, a rhythm strip may be included as a fourth or fifth row.[44]

The timing across the page is continuous and notes tracings of the 12 leads for the same time period. In other words, if the output were traced by needles on paper, each row would switch which leads as the paper is pulled under the needle. For example, the top row would first trace lead I, then switch to lead aVR, then switch to V1, and then switch to V4, and so none of these four tracings of the leads are from the same time period as they are traced in sequence through time.[51]

Contiguity of leads edit

 
Diagram showing the contiguous leads in the same color in the standard 12-lead layout

Each of the 12 ECG leads records the electrical activity of the heart from a different angle, and therefore align with different anatomical areas of the heart. Two leads that look at neighboring anatomical areas are said to be contiguous.[44]

Category Leads Activity
Inferior leads Leads II, III and aVF Look at electrical activity from the vantage point of the inferior surface (diaphragmatic surface of heart)
Lateral leads I, aVL, V5 and V6 Look at the electrical activity from the vantage point of the lateral wall of left ventricle
Septal leads V1 and V2 Look at electrical activity from the vantage point of the septal surface of the heart (interventricular septum)
Anterior leads V3 and V4 Look at electrical activity from the vantage point of the anterior wall of the right and left ventricles (Sternocostal surface of heart)

In addition, any two precordial leads next to one another are considered to be contiguous. For example, though V4 is an anterior lead and V5 is a lateral lead, they are contiguous because they are next to one another.

Electrophysiology edit

The study of the conduction system of the heart is called cardiac electrophysiology (EP). An EP study is performed via a right-sided cardiac catheterization: a wire with an electrode at its tip is inserted into the right heart chambers from a peripheral vein, and placed in various positions in close proximity to the conduction system so that the electrical activity of that system can be recorded.[52] Standard catheter positions for an EP study include "high right atrium" or hRA near the sinus node, a "His" across the septal wall of the tricuspid valve to measure bundle of His, a "coronary sinus" into the coronary sinus, and a "right ventricle" in the apex of the right ventricle.[53]

Interpretation edit

Interpretation of the ECG is fundamentally about understanding the electrical conduction system of the heart. Normal conduction starts and propagates in a predictable pattern, and deviation from this pattern can be a normal variation or be pathological. An ECG does not equate with mechanical pumping activity of the heart; for example, pulseless electrical activity produces an ECG that should pump blood but no pulses are felt (and constitutes a medical emergency and CPR should be performed). Ventricular fibrillation produces an ECG but is too dysfunctional to produce a life-sustaining cardiac output. Certain rhythms are known to have good cardiac output and some are known to have bad cardiac output. Ultimately, an echocardiogram or other anatomical imaging modality is useful in assessing the mechanical function of the heart.[54]

Like all medical tests, what constitutes "normal" is based on population studies. The heartrate range of between 60 and 100 beats per minute (bpm) is considered normal since data shows this to be the usual resting heart rate.[55]

Theory edit

 
QRS is upright in a lead when its axis is aligned with that lead's vector
 
Schematic representation of a normal ECG

Interpretation of the ECG is ultimately that of pattern recognition. In order to understand the patterns found, it is helpful to understand the theory of what ECGs represent. The theory is rooted in electromagnetics and boils down to the four following points:[56]

  • depolarization of the heart toward the positive electrode produces a positive deflection
  • depolarization of the heart away from the positive electrode produces a negative deflection
  • repolarization of the heart toward the positive electrode produces a negative deflection
  • repolarization of the heart away from the positive electrode produces a positive deflection

Thus, the overall direction of depolarization and repolarization produces positive or negative deflection on each lead's trace. For example, depolarizing from right to left would produce a positive deflection in lead I because the two vectors point in the same direction. In contrast, that same depolarization would produce minimal deflection in V1 and V2 because the vectors are perpendicular, and this phenomenon is called isoelectric.

Normal rhythm produces four entities – a P wave, a QRS complex, a T wave, and a U wave – that each have a fairly unique pattern.

  • The P wave represents atrial depolarization.
  • The QRS complex represents ventricular depolarization.
  • The T wave represents ventricular repolarization.
  • The U wave represents papillary muscle repolarization.

Changes in the structure of the heart and its surroundings (including blood composition) change the patterns of these four entities.

The U wave is not typically seen and its absence is generally ignored. Atrial repolarization is typically hidden in the much more prominent QRS complex and normally cannot be seen without additional, specialized electrodes.

Background grid edit

ECGs are normally printed on a grid. The horizontal axis represents time and the vertical axis represents voltage. The standard values on this grid are shown in the adjacent image at 25mm/sec:[57]

  • A small box is 1 mm × 1 mm and represents 0.1 mV × 0.04 seconds.
  • A large box is 5 mm × 5 mm and represents 0.5 mV × 0.20 seconds.

The "large" box is represented by a heavier line weight than the small boxes.

 
Measuring time and voltage with ECG graph paper

The standard printing speed in the United States is 25 mm per sec (5 big boxes per second), but in other countries it can be 50 mm per sec. Faster speeds such as 100 and 200 mm per sec are used during electrophysiology studies.

Not all aspects of an ECG rely on precise recordings or having a known scaling of amplitude or time. For example, determining if the tracing is a sinus rhythm only requires feature recognition and matching, and not measurement of amplitudes or times (i.e., the scale of the grids are irrelevant). An example to the contrary, the voltage requirements of left ventricular hypertrophy require knowing the grid scale.

Rate and rhythm edit

In a normal heart, the heart rate is the rate at which the sinoatrial node depolarizes since it is the source of depolarization of the heart. Heart rate, like other vital signs such as blood pressure and respiratory rate, change with age. In adults, a normal heart rate is between 60 and 100 bpm (normocardic), whereas it is higher in children.[58] A heart rate below normal is called "bradycardia" (<60 in adults) and above normal is called "tachycardia" (>100 in adults). A complication of this is when the atria and ventricles are not in synchrony and the "heart rate" must be specified as atrial or ventricular (e.g., the ventricular rate in ventricular fibrillation is 300–600 bpm, whereas the atrial rate can be normal [60–100] or faster [100–150]).[59]

In normal resting hearts, the physiologic rhythm of the heart is normal sinus rhythm (NSR). Normal sinus rhythm produces the prototypical pattern of P wave, QRS complex, and T wave. Generally, deviation from normal sinus rhythm is considered a cardiac arrhythmia. Thus, the first question in interpreting an ECG is whether or not there is a sinus rhythm. A criterion for sinus rhythm is that P waves and QRS complexes appear 1-to-1, thus implying that the P wave causes the QRS complex.[51]

Once sinus rhythm is established, or not, the second question is the rate. For a sinus rhythm, this is either the rate of P waves or QRS complexes since they are 1-to-1. If the rate is too fast, then it is sinus tachycardia, and if it is too slow, then it is sinus bradycardia.

If it is not a sinus rhythm, then determining the rhythm is necessary before proceeding with further interpretation. Some arrhythmias with characteristic findings:

Determination of rate and rhythm is necessary in order to make sense of further interpretation.

Axis edit

 
Diagram showing how the polarity of the QRS complex in leads I, II, and III can be used to estimate the heart's electrical axis in the frontal plane.

The heart has several axes, but the most common by far is the axis of the QRS complex (references to "the axis" imply the QRS axis). Each axis can be computationally determined to result in a number representing degrees of deviation from zero, or it can be categorized into a few types.[60]

The QRS axis is the general direction of the ventricular depolarization wavefront (or mean electrical vector) in the frontal plane. It is often sufficient to classify the axis as one of three types: normal, left deviated, or right deviated. Population data shows that a normal QRS axis is from −30° to 105°, with 0° being along lead I and positive being inferior and negative being superior (best understood graphically as the hexaxial reference system).[61] Beyond +105° is right axis deviation and beyond −30° is left axis deviation (the third quadrant of −90° to −180° is very rare and is an indeterminate axis). A shortcut for determining if the QRS axis is normal is if the QRS complex is mostly positive in lead I and lead II (or lead I and aVF if +90° is the upper limit of normal).[62]

The normal QRS axis is generally down and to the left, following the anatomical orientation of the heart within the chest. An abnormal axis suggests a change in the physical shape and orientation of the heart or a defect in its conduction system that causes the ventricles to depolarize in an abnormal way.[51]

Classification Angle Notes
Normal −30° to 105° Normal
Left axis deviation −30° to −90° May indicate left ventricular hypertrophy, left anterior fascicular block, or an old inferior STEMI
Right axis deviation +105° to +180° May indicate right ventricular hypertrophy, left posterior fascicular block, or an old lateral STEMI
Indeterminate axis +180° to −90° Rarely seen; considered an 'electrical no-man's land'

The extent of a normal axis can be +90° or 105° depending on the source.

Amplitudes and intervals edit

 
Animation of a normal ECG wave

All of the waves on an ECG tracing and the intervals between them have a predictable time duration, a range of acceptable amplitudes (voltages), and a typical morphology. Any deviation from the normal tracing is potentially pathological and therefore of clinical significance.[63]

For ease of measuring the amplitudes and intervals, an ECG is printed on graph paper at a standard scale: each 1 mm (one small box on the standard 25mm/s ECG paper) represents 40 milliseconds of time on the x-axis, and 0.1 millivolts on the y-axis.[64]

Feature Description Pathology Duration
P wave The P wave represents depolarization of the atria. Atrial depolarization spreads from the SA node towards the AV node, and from the right atrium to the left atrium. The P wave is typically upright in most leads except for aVR; an unusual P wave axis (inverted in other leads) can indicate an ectopic atrial pacemaker. If the P wave is of unusually long duration, it may represent atrial enlargement. Typically a large right atrium gives a tall, peaked P wave while a large left atrium gives a two-humped bifid P wave. <80 ms
PR interval The PR interval is measured from the beginning of the P wave to the beginning of the QRS complex. This interval reflects the time the electrical impulse takes to travel from the sinus node through the AV node. A PR interval shorter than 120 ms suggests that the electrical impulse is bypassing the AV node, as in Wolff-Parkinson-White syndrome. A PR interval consistently longer than 200 ms diagnoses first degree atrioventricular block. The PR segment (the portion of the tracing after the P wave and before the QRS complex) is typically completely flat, but may be depressed in pericarditis. 120 to 200 ms
QRS complex The QRS complex represents the rapid depolarization of the right and left ventricles. The ventricles have a large muscle mass compared to the atria, so the QRS complex usually has a much larger amplitude than the P wave. If the QRS complex is wide (longer than 120 ms) it suggests disruption of the heart's conduction system, such as in LBBB, RBBB, or ventricular rhythms such as ventricular tachycardia. Metabolic issues such as severe hyperkalemia, or tricyclic antidepressant overdose can also widen the QRS complex. An unusually tall QRS complex may represent left ventricular hypertrophy while a very low-amplitude QRS complex may represent a pericardial effusion or infiltrative myocardial disease. 80 to 100 ms
J-point The J-point is the point at which the QRS complex finishes and the ST segment begins. The J-point may be elevated as a normal variant. The appearance of a separate J wave or Osborn wave at the J-point is pathognomonic of hypothermia or hypercalcemia.[65]
ST segment The ST segment connects the QRS complex and the T wave; it represents the period when the ventricles are depolarized. It is usually isoelectric, but may be depressed or elevated with myocardial infarction or ischemia. ST depression can also be caused by LVH or digoxin. ST elevation can also be caused by pericarditis, Brugada syndrome, or can be a normal variant (J-point elevation).
T wave The T wave represents the repolarization of the ventricles. It is generally upright in all leads except aVR and lead V1. Inverted T waves can be a sign of myocardial ischemia, left ventricular hypertrophy, high intracranial pressure, or metabolic abnormalities. Peaked T waves can be a sign of hyperkalemia or very early myocardial infarction. 160 ms
Corrected QT interval (QTc) The QT interval is measured from the beginning of the QRS complex to the end of the T wave. Acceptable ranges vary with heart rate, so it must be corrected to the QTc by dividing by the square root of the RR interval. A prolonged QTc interval is a risk factor for ventricular tachyarrhythmias and sudden death. Long QT can arise as a genetic syndrome, or as a side effect of certain medications. An unusually short QTc can be seen in severe hypercalcemia. <440 ms
U wave The U wave is hypothesized to be caused by the repolarization of the interventricular septum. It normally has a low amplitude, and even more often is completely absent. A very prominent U wave can be a sign of hypokalemia, hypercalcemia or hyperthyroidism.[66]

Limb leads and electrical conduction through the heart edit

 
Formation of limb waveforms during a pulse

The animation shown to the right illustrates how the path of electrical conduction gives rise to the ECG waves in the limb leads. Recall that a positive current (as created by depolarization of cardiac cells) traveling towards the positive electrode and away from the negative electrode creates a positive deflection on the ECG. Likewise, a positive current traveling away from the positive electrode and towards the negative electrode creates a negative deflection on the ECG.[67][68] The red arrow represents the overall direction of travel of the depolarization. The magnitude of the red arrow is proportional to the amount of tissue being depolarized at that instance. The red arrow is simultaneously shown on the axis of each of the 3 limb leads. Both the direction and the magnitude of the red arrow's projection onto the axis of each limb lead is shown with blue arrows. Then, the direction and magnitude of the blue arrows are what theoretically determine the deflections on the ECG. For example, as a blue arrow on the axis for Lead I moves from the negative electrode, to the right, towards the positive electrode, the ECG line rises, creating an upward wave. As the blue arrow on the axis for Lead I moves to the left, a downward wave is created. The greater the magnitude of the blue arrow, the greater the deflection on the ECG for that particular limb lead.[69]

Frames 1–3 depict the depolarization being generated in and spreading through the Sinoatrial node. The SA node is too small for its depolarization to be detected on most ECGs. Frames 4–10 depict the depolarization traveling through the atria, towards the Atrioventricular node. During frame 7, the depolarization is traveling through the largest amount of tissue in the atria, which creates the highest point in the P wave. Frames 11–12 depict the depolarization traveling through the AV node. Like the SA node, the AV node is too small for the depolarization of its tissue to be detected on most ECGs. This creates the flat PR segment.[70]

Frame 13 depicts an interesting phenomenon in an over-simplified fashion. It depicts the depolarization as it starts to travel down the interventricular septum, through the Bundle of His and Bundle branches. After the Bundle of His, the conduction system splits into the left bundle branch and the right bundle branch. Both branches conduct action potentials at about 1 m/s. Interestingly, however, the action potential starts traveling down the left bundle branch about 5 milliseconds before it starts traveling down the right bundle branch, as depicted by frame 13. This causes the depolarization of the interventricular septum tissue to spread from left to right, as depicted by the red arrow in frame 14. In some cases, this gives rise to a negative deflection after the PR interval, creating a Q wave such as the one seen in lead I in the animation to the right. Depending on the mean electrical axis of the heart, this phenomenon can result in a Q wave in lead II as well.[71][72]

Following depolarization of the interventricular septum, the depolarization travels towards the apex of the heart. This is depicted by frames 15–17 and results in a positive deflection on all three limb leads, which creates the R wave. Frames 18–21 then depict the depolarization as it travels throughout both ventricles from the apex of the heart, following the action potential in the Purkinje fibers. This phenomenon creates a negative deflection in all three limb leads, forming the S wave on the ECG. Repolarization of the atria occurs at the same time as the generation of the QRS complex, but it is not detected by the ECG since the tissue mass of the ventricles is so much larger than that of the atria. Ventricular contraction occurs between ventricular depolarization and repolarization. During this time, there is no movement of charge, so no deflection is created on the ECG. This results in the flat ST segment after the S wave.[73]

Frames 24–28 in the animation depict repolarization of the ventricles. The epicardium is the first layer of the ventricles to repolarize, followed by the myocardium. The endocardium is the last layer to repolarize. The plateau phase of depolarization has been shown to last longer in endocardial cells than in epicardial cells. This causes repolarization to start from the apex of the heart and move upwards. Since repolarization is the spread of negative current as membrane potentials decrease back down to the resting membrane potential, the red arrow in the animation is pointing in the direction opposite of the repolarization. This therefore creates a positive deflection in the ECG, and creates the T wave.[74]

Ischemia and infarction edit

Ischemia or non-ST elevation myocardial infarctions (non-STEMIs) may manifest as ST depression or inversion of T waves. It may also affect the high frequency band of the QRS.

ST elevation myocardial infarctions (STEMIs) have different characteristic ECG findings based on the amount of time elapsed since the MI first occurred. The earliest sign is hyperacute T waves, peaked T waves due to local hyperkalemia in ischemic myocardium. This then progresses over a period of minutes to elevations of the ST segment by at least 1 mm. Over a period of hours, a pathologic Q wave may appear and the T wave will invert. Over a period of days the ST elevation will resolve. Pathologic Q waves generally will remain permanently.[75]

The coronary artery that has been occluded can be identified in an STEMI based on the location of ST elevation. The left anterior descending (LAD) artery supplies the anterior wall of the heart, and therefore causes ST elevations in anterior leads (V1 and V2). The LCx supplies the lateral aspect of the heart and therefore causes ST elevations in lateral leads (I, aVL and V6). The right coronary artery (RCA) usually supplies the inferior aspect of the heart, and therefore causes ST elevations in inferior leads (II, III and aVF).[76]

Artifacts edit

An ECG tracing is affected by patient motion. Some rhythmic motions (such as shivering or tremors) can create the illusion of cardiac arrhythmia.[77] Artifacts are distorted signals caused by a secondary internal or external sources, such as muscle movement or interference from an electrical device.[78][79]

Distortion poses significant challenges to healthcare providers,[78] who employ various techniques[80] and strategies to safely recognize[81] these false signals.[medical citation needed] Accurately separating the ECG artifact from the true ECG signal can have a significant impact on patient outcomes and legal liabilities.[82][unreliable medical source?]

Improper lead placement (for example, reversing two of the limb leads) has been estimated to occur in 0.4% to 4% of all ECG recordings,[83] and has resulted in improper diagnosis and treatment including unnecessary use of thrombolytic therapy.[84][85]

A Method for Interpretation edit

Whitbread, consultant nurse and paramedic, suggests ten rules of the normal ECG, deviation from which is likely to indicate pathology.[86] These have been added to, creating the 15 rules for 12-lead (and 15- or 18-lead) interpretation.[87]

Rule 1: All waves in aVR are negative.

Rule 2: The ST segment (J point) starts on the isoelectric line (except in V1 & V2 where it may be elevated by not greater than 1 mm).

Rule 3: The PR interval should be 0.12–0.2 seconds long.

Rule 4: The QRS complex should not exceed 0.11–0.12 seconds.

Rule 5: The QRS and T waves tend to have the same general direction in the limb leads.

Rule 6: The R wave in the precordial (chest) leads grows from V1 to at least V4 where it may or may not decline again.

Rule 7: The QRS is mainly upright in I and II.

Rule 8: The P wave is upright in I II and V2 to V6.

Rule 9: There is no Q wave or only a small q (<0.04 seconds in width) in I, II and V2 to V6.

Rule 10: The T wave is upright in I II and V2 to V6. The end of the T wave should not drop below the isoelectric baseline.

Rule 11: Does the deepest S wave in V1 plus the tallest R wave in V5 or V6 equal >35 mm?

Rule 12: Is there an Epsilon wave?

Rule 13: Is there an J wave?

Rule 14: Is there a Delta wave?

Rule 15: Are there any patterns representing an occlusive myocardial infarction (OMI)?

Diagnosis edit

Numerous diagnoses and findings can be made based upon electrocardiography, and many are discussed above. Overall, the diagnoses are made based on the patterns. For example, an "irregularly irregular" QRS complex without P waves is the hallmark of atrial fibrillation; however, other findings can be present as well, such as a bundle branch block that alters the shape of the QRS complexes. ECGs can be interpreted in isolation but should be applied – like all diagnostic tests – in the context of the patient. For example, an observation of peaked T waves is not sufficient to diagnose hyperkalemia; such a diagnosis should be verified by measuring the blood potassium level. Conversely, a discovery of hyperkalemia should be followed by an ECG for manifestations such as peaked T waves, widened QRS complexes, and loss of P waves. The following is an organized list of possible ECG-based diagnoses.[88]

Rhythm disturbances or arrhythmias:[89]

Heart block and conduction problems:

Electrolytes disturbances and intoxication:

Ischemia and infarction:

Structural:

Other phenomena:

History edit

 
An early commercial ECG device (1911)
 
ECG from 1957
  • In 1872, Alexander Muirhead is reported to have attached wires to the wrist of a patient with fever to obtain an electronic record of their heartbeat.[91]
  • In 1882, John Burdon-Sanderson working with frogs, was the first to appreciate that the interval between variations in potential was not electrically quiescent and coined the term "isoelectric interval" for this period.[92]
  • In 1887, Augustus Waller[93] invented an ECG machine consisting of a Lippmann capillary electrometer fixed to a projector. The trace from the heartbeat was projected onto a photographic plate that was itself fixed to a toy train. This allowed a heartbeat to be recorded in real time.
  • In 1895, Willem Einthoven assigned the letters P, Q, R, S, and T to the deflections in the theoretical waveform he created using equations which corrected the actual waveform obtained by the capillary electrometer to compensate for the imprecision of that instrument. Using letters different from A, B, C, and D (the letters used for the capillary electrometer's waveform) facilitated comparison when the uncorrected and corrected lines were drawn on the same graph.[94] Einthoven probably chose the initial letter P to follow the example set by Descartes in geometry.[94] When a more precise waveform was obtained using the string galvanometer, which matched the corrected capillary electrometer waveform, he continued to use the letters P, Q, R, S, and T,[94] and these letters are still in use today. Einthoven also described the electrocardiographic features of a number of cardiovascular disorders.
  • In 1897, the string galvanometer was invented by the French engineer Clément Ader.[95]
  • In 1901, Einthoven, working in Leiden, the Netherlands, used the string galvanometer: the first practical ECG.[96] This device was much more sensitive than the capillary electrometer Waller used.
  • In 1924, Einthoven was awarded the Nobel Prize in Medicine for his pioneering work in developing the ECG.[97]
  • By 1927, General Electric had developed a portable apparatus that could produce electrocardiograms without the use of the string galvanometer. This device instead combined amplifier tubes similar to those used in a radio with an internal lamp and a moving mirror that directed the tracing of the electric pulses onto film.[98]
  • In 1937, Taro Takemi invented a new portable electrocardiograph machine.[99]
  • In 1942, Emanuel Goldberger increases the voltage of Wilson's unipolar leads by 50% and creates the augmented limb leads aVR, aVL and aVF. When added to Einthoven's three limb leads and the six chest leads we arrive at the 12-lead electrocardiogram that is used today.[100]
  • In the late 1940s, Rune Elmqvist invented an inkjet printer involving thin jets of ink deflected by electrical potentials from the heart, with good frequency response and direct recording of ECG on paper. The device, called the Mingograf, was sold by Siemens Elema until the 1990s.[101]

Etymology edit

The word is derived from the Greek electro, meaning related to electrical activity; kardia, meaning heart; and graph, meaning "to write".[102]

See also edit

Notes edit

  1. ^ The version with '-K-', more commonly used in American English than in British English, is an early-20th-century loanword from the German acronym EKG for Elektrokardiogramm (electrocardiogram),[1] which reflects that German physicians were pioneers in the field at the time. Today, AMA style and – under its stylistic influence – most American medical publications use ECG instead of EKG.[2] The German term Elektrokardiogramm as well as the English equivalent, electrocardiogram, consist of the Neo-Latin/international scientific vocabulary elements elektro- (cognate electro-) and kardi- (cognate 'cardi-'), the latter from Greek kardia (heart).[3] The '-K-' version is more often retained under circumstances where there may be verbal confusion between ECG and EEG (electroencephalography) due to similar pronunciation.

References edit

  1. ^ . Lexico Dictionaries. Archived from the original on 15 February 2020. Retrieved 20 January 2020.
  2. ^ "15.3.1 Electrocardiographic Terms", AMA Manual of Style, American Medical Association
  3. ^ "Merriam-Webster's Collegiate Dictionary". Merriam-Webster.
  4. ^ Bunce, Nicholas H.; Ray, Robin; Patel, Hitesh (2020). "30. Cardiology". In Feather, Adam; Randall, David; Waterhouse, Mona (eds.). Kumar and Clark's Clinical Medicine (10th ed.). Elsevier. pp. 1033–1038. ISBN 978-0-7020-7870-5.
  5. ^ Lilly, Leonard S. (2016). Pathophysiology of Heart Disease: A Collaborative Project of Medical Students and Faculty, 6th Edition. Lippincott Williams & Wilkins. pp. 70–78. ISBN 978-1-4698-9758-5. OCLC 1229852550.
  6. ^ Lyakhov, Pavel; Kiladze, Mariya; Lyakhova, Ulyana (January 2021). "System for Neural Network Determination of Atrial Fibrillation on ECG Signals with Wavelet-Based Preprocessing". Applied Sciences. 11 (16): 7213. doi:10.3390/app11167213.
  7. ^ Hoyland, Philip; Hammache, Néfissa; Battaglia, Alberto; Oster, Julien; Felblinger, Jacques; de Chillou, Christian; Odille, Freddy (2020). "A Paced-ECG Detector and Delineator for Automatic Multi-Parametric Catheter Mapping of Ventricular Tachycardia". IEEE Access. 8: 223952–223960. Bibcode:2020IEEEA...8v3952H. doi:10.1109/ACCESS.2020.3043542. ISSN 2169-3536.
  8. ^ Bigler, Marius Reto; Zimmermann, Patrick; Papadis, Athanasios; Seiler, Christian (1 January 2021). "Accuracy of intracoronary ECG parameters for myocardial ischemia detection". Journal of Electrocardiology. 64: 50–57. doi:10.1016/j.jelectrocard.2020.11.018. ISSN 0022-0736. PMID 33316551. S2CID 229173576.
  9. ^ Prabhakararao, Eedara; Dandapat, Samarendra (August 2020). "Myocardial Infarction Severity Stages Classification From ECG Signals Using Attentional Recurrent Neural Network". IEEE Sensors Journal. 20 (15): 8711–8720. Bibcode:2020ISenJ..20.8711P. doi:10.1109/JSEN.2020.2984493. ISSN 1558-1748. S2CID 216310175.
  10. ^ Carrizales-Sepúlveda, Edgar Francisco; Vera-Pineda, Raymundo; Jiménez-Castillo, Raúl Alberto; Treviño-García, Karla Belén; Ordaz-Farías, Alejandro (1 November 2019). "Toluene toxicity presenting with hypokalemia, profound weakness and U waves in the electrocardiogram". The American Journal of Emergency Medicine. 37 (11): 2120.e1–2120.e3. doi:10.1016/j.ajem.2019.158417. ISSN 0735-6757. PMID 31477355. S2CID 201804610.
  11. ^ Aswini Kumar MD. . LifeHugger. Archived from the original on 2 October 2017. Retrieved 11 February 2010.
  12. ^ a b Lilly 2016, pp. 80.
  13. ^ Walraven, Gail (2011). Basic arrhythmias (7th ed.). Boston: Brady/Pearson. pp. 1–11. ISBN 978-0-13-500238-4. OCLC 505018241.
  14. ^ Braunwald, Eugene, ed. (1997). Heart Disease: A Textbook of Cardiovascular Medicine (5th ed.). Philadelphia: Saunders. p. 118. ISBN 0-7216-5666-8. OCLC 32970742.
  15. ^ "What is a STEMI? - ECG Medical Training". ECG Medical Training. 24 June 2015. Retrieved 24 June 2018.
  16. ^ "What is NSTEMI? What You NEED to Know". MyHeart. 30 April 2015. Retrieved 24 June 2018.
  17. ^ a b Masters, Jo; Bowden, Carole; Martin, Carole; Chandler, Sharon (2003). Textbook of veterinary medical nursing (in Spanish). New York: Butterworth-Heinemann. p. 244. ISBN 978-0-7506-5171-4. OCLC 53094318.
  18. ^ Drew, B. J.; Califf, R. M.; Funk, M.; Kaufman, E. S.; Krucoff, M. W.; Laks, M. M.; Macfarlane, P. W.; Sommargren, C.; Swiryn, S.; Van Hare, G. F. (26 October 2004). "Practice Standards for Electrocardiographic Monitoring in Hospital Settings". Circulation. 110 (17): 2721–2746. doi:10.1161/01.CIR.0000145144.56673.59. PMID 15505110. S2CID 220573469.
  19. ^ Galli, Alessio; Ambrosini, Francesco; Lombardi, Federico (August 2016). "Holter Monitoring and Loop Recorders: From Research to Clinical Practice". Arrhythmia & Electrophysiology Review. 5 (2): 136–143. doi:10.15420/AER.2016.17.2. ISSN 2050-3369. PMC 5013174. PMID 27617093.
  20. ^ US Preventive Services Task, Force.; Curry, SJ; Krist, AH; Owens, DK; Barry, MJ; Caughey, AB; Davidson, KW; Doubeni, CA; Epling JW, Jr; Kemper, AR; Kubik, M; Landefeld, CS; Mangione, CM; Silverstein, M; Simon, MA; Tseng, CW; Wong, JB (12 June 2018). "Screening for Cardiovascular Disease Risk With Electrocardiography: US Preventive Services Task Force Recommendation Statement". JAMA. 319 (22): 2308–2314. doi:10.1001/jama.2018.6848. PMID 29896632.
  21. ^ Moyer VA (2 October 2012). "Screening for coronary heart disease with electrocardiography: U.S. Preventive Services Task Force recommendation statement". Annals of Internal Medicine. 157 (7): 512–518. doi:10.7326/0003-4819-157-7-201210020-00514. PMID 22847227.
  22. ^ Consumer Reports; American Academy of Family Physicians; ABIM Foundation (April 2012), (PDF), Choosing Wisely, Consumer Reports, archived from the original (PDF) on 20 December 2013, retrieved 14 August 2012
  23. ^ "Summary of Medical Standards" (PDF). U.S. Federal Aviation Administration. 2006. Retrieved 27 December 2013.
  24. ^ Corrado, D.; Basso, C.; Schiavon, M.; Thiene, G. (6 August 1998). "Screening for hypertrophic cardiomyopathy in young athletes". The New England Journal of Medicine. 339 (6): 364–369. doi:10.1056/NEJM199808063390602. ISSN 0028-4793. PMID 9691102.
  25. ^ "Electrocardiograph, ECG" (PDF). World Health Organization. Retrieved 1 August 2020.
  26. ^ "How we'll invent the future, by Bill Gates". MIT Technology Review. Retrieved 1 April 2019.
  27. ^ "FDA approves AliveCor heart monitor". Techcrunch. Retrieved 25 August 2018.
  28. ^ "EKG Risks". Stanford Health Care. Retrieved 1 April 2019.
  29. ^ Schläpfer, J; Wellens, HJ (29 August 2017). "Computer-Interpreted Electrocardiograms: Benefits and Limitations". Journal of the American College of Cardiology. 70 (9): 1183–1192. doi:10.1016/j.jacc.2017.07.723. PMID 28838369.
  30. ^ Macfarlane, P.W.; Coleman (1995). (PDF). Society for Cardiological Science and Technology. Archived from the original (PDF) on 19 February 2018. Retrieved 21 October 2017.
  31. ^ . www.emtresource.com. 27 April 2019. Archived from the original on 19 January 2022. Retrieved 24 May 2019.
  32. ^ "ECG Leads - an overview | ScienceDirect Topics". www.sciencedirect.com. Retrieved 28 October 2022.
  33. ^ . www.emtresource.com. 27 April 2014. Archived from the original on 19 January 2022. Retrieved 27 May 2019.
  34. ^ "EKG Interpretation". Nurses Learning Network. Retrieved 27 May 2019.
  35. ^ Jowett, N. I.; Turner, A. M.; Cole, A.; Jones, P. A. (1 February 2005). "Modified electrode placement must be recorded when performing 12-lead electrocardiograms". Postgraduate Medical Journal. 81 (952): 122–125. doi:10.1136/pgmj.2004.021204. ISSN 0032-5473. PMC 1743200. PMID 15701746.
  36. ^ "12-Lead ECG Placement Guide with Illustrations". Cables and Sensors. Retrieved 11 July 2017.
  37. ^ Kavuru, Madhav S.; Vesselle, Hubert; Thomas, Cecil W. (1987). Advances in Body Surface Potential Mapping (BSPM) Instrumentation; In: Pediatric and Fundamental Electrocardiography. Developments in Cardiovascular Medicine. Vol. 56. pp. 315–327. doi:10.1007/978-1-4613-2323-5_15. ISBN 978-1-4612-9428-3. ISSN 0166-9842.
  38. ^ Tsukada, Yayoi Tetsuou; Tokita, Miwa; Murata, Hiroshige; Hirasawa, Yasuhiro; Yodogawa, Kenji; Iwasaki, Yu-ki; Asai, Kuniya; Shimizu, Wataru; Kasai, Nahoko; Nakashima, Hiroshi; Tsukada, Shingo (24 January 2019). "Validation of wearable textile electrodes for ECG monitoring". Heart and Vessels. 34 (7): 1203–1211. doi:10.1007/s00380-019-01347-8. ISSN 0910-8327. PMC 6556171. PMID 30680493.
  39. ^ Gargiulo, GD (2015). "True unipolar ECG machine for Wilson Central Terminal measurements". BioMed Research International. 2015: 586397. doi:10.1155/2015/586397. PMC 460614. PMID 26495303.
  40. ^ Sensors, Cables and. "12-Lead ECG Placement Guide with Illustrations | Cables and Sensors". Cables and Sensors. Retrieved 21 October 2017.
  41. ^ "Limb Leads – ECG Lead Placement – Normal Function of the Heart – Cardiology Teaching Package – Practice Learning – Division of Nursing – The University of Nottingham". Nottingham.ac.uk. Retrieved 15 August 2009.
  42. ^ . Library.med.utah.edu. Archived from the original on 22 March 2009. Retrieved 15 August 2009.
  43. ^ Jin, Benjamin E.; Wulff, Heike; Widdicombe, Jonathan H.; Zheng, Jie; Bers, Donald M.; Puglisi, Jose L. (December 2012). "A simple device to illustrate the Einthoven triangle". Advances in Physiology Education. 36 (4): 319–324. Bibcode:2012BpJ...102..211J. doi:10.1152/advan.00029.2012. ISSN 1043-4046. PMC 3776430. PMID 23209014.
  44. ^ a b c Meek, S. (16 February 2002). "ABC of clinical electrocardiography: Introduction. I---Leads, rate, rhythm, and cardiac axis". BMJ. 324 (7334): 415–418. doi:10.1136/bmj.324.7334.415. ISSN 0959-8138. PMC 1122339. PMID 11850377.
  45. ^ Madias, JE (2008). "On recording the unipolar ECG limb leads via the Wilson's vs the Goldberger's terminals: aVR, aVL, and aVF revisited". Indian Pacing and Electrophysiology Journal. 8 (4): 292–297. PMC 2572021. PMID 18982138.
  46. ^ Mc Loughlin, MJ (2020). "Precordial bipolar leads: A new method to study anterior acute myocardial infarction". J Electrocardiol. 59 (2): 45–64. doi:10.1016/j.jelectrocard.2019.12.017. PMID 31986362. S2CID 210935474.
  47. ^ Buttner, Robert; Cadogan, Mike (29 January 2022). "Lewis lead". Life in the Fast Lane. Retrieved 2 February 2022.
  48. ^ Meigas, K; Kaik, J; Anier, A (2008). "Device and methods for performing transesophageal stimulation at reduced pacing current threshold". Estonian Journal of Engineering. 57 (2): 154. doi:10.3176/eng.2008.2.05. S2CID 42055085.
  49. ^ a b Pehrson, Steen M.; Blomströ-Lundqvist, Carina; Ljungströ, Erik; Blomströ, Per (1994). "Clinical value of transesophageal atrial stimulation and recording in patients with arrhythmia-related symptoms or documented supraventricular tachycardia-correlation to clinical history and invasive studies". Clinical Cardiology. 17 (10): 528–534. doi:10.1002/clc.4960171004. PMID 8001299.
  50. ^ Zhang, Yongan; Banta, Anton; Fu, Yonggan; John, Mathews M.; Post, Allison; Razavi, Mehdi; Cavallaro, Joseph; Aazhang, Behnaam; Lin, Yingyan (30 April 2022). "RT-RCG: Neural Network and Accelerator Search Towards Effective and Real-time ECG Reconstruction from Intracardiac Electrograms". ACM Journal on Emerging Technologies in Computing Systems. 18 (2): 29. doi:10.1145/3465372. ISSN 1550-4832. PMC 9236221. PMID 35765469.
  51. ^ a b c Ashley, Euan A.; Niebauer, Josef (2004). Conquering the ECG. Remedica.
  52. ^ "Electrode Catheter - an overview | ScienceDirect Topics". www.sciencedirect.com. Retrieved 28 October 2022.
  53. ^ Pennoyer, James; Bykhovsky, Michael; Sohinki, Daniel; Mallard, Rachel; Berman, Adam (1 October 2020). "Successful Catheter Ablation of Two Macro-reentrant Atrial Tachycardias in a Patient with Congenitally Corrected Transposition of the Great Arteries: A Case Report". Journal of Innovations in Cardiac Rhythm Management. 11 (10): 4273–4280. doi:10.19102/icrm.2020.111005. ISSN 2156-3977. PMC 7588239. PMID 33123416.
  54. ^ Ewy, G. A. (1984). "Defining electromechanical dissociation". Annals of Emergency Medicine. 13 (9 Pt 2): 830–832. doi:10.1016/s0196-0644(84)80452-7. ISSN 0196-0644. PMID 6476549.
  55. ^ Avram, Robert; Tison, Geoffrey H.; Aschbacher, Kirstin; Kuhar, Peter; Vittinghoff, Eric; Butzner, Michael; Runge, Ryan; Wu, Nancy; Pletcher, Mark J.; Marcus, Gregory M.; Olgin, Jeffrey (25 June 2019). "Real-world heart rate norms in the Health eHeart study". npj Digital Medicine. 2 (1): 58. doi:10.1038/s41746-019-0134-9. ISSN 2398-6352. PMC 6592896. PMID 31304404.
  56. ^ Schrepel, Caitlin; Amick, Ashley E.; Sayed, Madeline; Chipman, Anne K. (7 September 2021). "Ischemic ECG Pattern Recognition to Facilitate Interpretation While Task Switching: A Parallel Curriculum". MedEdPORTAL. 17: 11182. doi:10.15766/mep_2374-8265.11182. ISSN 2374-8265. PMC 8421424. PMID 34557588.
  57. ^ Becker, Daniel E. (2006). "Fundamentals of Electrocardiography Interpretation". Anesthesia Progress. 53 (2): 53–64. doi:10.2344/0003-3006(2006)53[53:FOEI]2.0.CO;2. ISSN 0003-3006. PMC 1614214. PMID 16863387.
  58. ^ Fleming, Susannah; Thompson, Matthew; Stevens, Richard; Heneghan, Carl; Plüddemann, Annette; Maconochie, Ian; Tarassenko, Lionel; Mant, David (19 March 2011). "Normal ranges of heart rate and respiratory rate in children from birth to 18 years of age: a systematic review of observational studies". Lancet. 377 (9770): 1011–1018. doi:10.1016/S0140-6736(10)62226-X. ISSN 1474-547X. PMC 3789232. PMID 21411136.
  59. ^ "Bradycardia - an overview | ScienceDirect Topics". www.sciencedirect.com. Retrieved 28 October 2022.
  60. ^ "Sample records for qrs complex relationship".
  61. ^ Surawicz, Borys; Knillans, Timothy (2008). Chou's electrocardiography in clinical practice : adult and pediatric (6th ed.). Philadelphia, PA: Saunders/Elsevier. p. 12. ISBN 978-1416037743.
  62. ^ Kashou, Anthony H.; Basit, Hajira; Chhabra, Lovely (2022), "Electrical Right and Left Axis Deviation", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 29262101, retrieved 28 October 2022
  63. ^ Publishing, M. D. K. (28 April 2015). EKGS and ECGS (Speedy Study Guides). Speedy Publishing LLC. ISBN 978-1-68185-011-5.
  64. ^ "ECG Study Guide".
  65. ^ Otero J, Lenihan DJ (2000). "The "normothermic" Osborn wave induced by severe hypercalcemia". Tex Heart Inst J. 27 (3): 316–317. PMC 101092. PMID 11093425.
  66. ^ Houghton, Andrew R; Gray, David (2012). Making Sense of the ECG, Third Edition. Hodder Education. p. 214. ISBN 978-1-4441-6654-5.
  67. ^ Cardio-online (12 December 2012). "ECG (EKG) Paper". Simple Cardiology. Retrieved 20 October 2019.
  68. ^ "Volume Conductor Principles and ECG Rules of Interpretation". CV Physiology. Retrieved 22 October 2019.
  69. ^ Sattar, Yasar; Chhabra, Lovely (2022), "Electrocardiogram", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 31747210, retrieved 28 October 2022
  70. ^ Noble, R. Joe; Hillis, J. Stanley; Rothbaum, Donald A. (1990), Walker, H. Kenneth; Hall, W. Dallas; Hurst, J. Willis (eds.), "Electrocardiography", Clinical Methods: The History, Physical, and Laboratory Examinations (3rd ed.), Butterworths, ISBN 9780409900774, PMID 21250195, retrieved 22 October 2019
  71. ^ Scher, Allen M.; Young, Allan C.; Malmgren, Arthur L.; Erickson, Robert V. (January 1955). "Activation of the Interventricular Septum". Circulation Research. 3 (1): 56–64. doi:10.1161/01.RES.3.1.56. ISSN 0009-7330. PMID 13231277.
  72. ^ "Ventricular Depolarization and the Mean Electrical Axis". CV Physiology. Retrieved 22 October 2019.
  73. ^ Kashou, Anthony H.; Basit, Hajira; Malik, Ahmad (2022), "ST Segment", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 29083566, retrieved 28 October 2022
  74. ^ Lukas, Anton (29 June 2016). "Electrophysiology of Myocardial Cells in the Epicardial, Midmyocardial, and Endocardial Layers of the Ventricle". Journal of Cardiovascular Pharmacology and Therapeutics. 2 (1): 61–72. doi:10.1177/107424849700200108. PMID 10684443. S2CID 44968291.
  75. ^ Alpert JS, Thygesen K, Antman E, Bassand JP (2000). "Myocardial infarction redefined – a consensus document of The Joint European Society of Cardiology/American College of Cardiology Committee for the redefinition of myocardial infarction". J Am Coll Cardiol. 36 (3): 959–969. doi:10.1016/S0735-1097(00)00804-4. PMID 10987628.
  76. ^ Warner, Matthew J.; Tivakaran, Vijai S. (2022), "Inferior Myocardial Infarction", StatPearls, Treasure Island (FL): StatPearls Publishing, PMID 29262146, retrieved 28 October 2022
  77. ^ Segura-Sampedro, Juan José; Parra-López, Loreto; Sampedro-Abascal, Consuelo; Muñoz-Rodríguez, Juan Carlos (2015). "Atrial Flutter EKG can be useful with the proper Electrophysiological Basis". International Journal of Cardiology. 179: 68–69. doi:10.1016/j.ijcard.2014.10.076. PMID 25464416.
  78. ^ a b Takla, George; Petre, John H.; Doyle, D John; Horibe, Mayumi; Gopakumaran, Bala (2006). "The Problem of Artifacts in Patient Monitor Data During Surgery: A Clinical and Methodological Review". Anesthesia & Analgesia. 103 (5): 1196–1204. doi:10.1213/01.ane.0000247964.47706.5d. PMID 17056954. S2CID 10614183.
  79. ^ Kligfield, Paul; Gettes, Leonard S.; Bailey, James J.; Childers, Rory; Deal, Barbara J.; Hancock, E. William; van Herpen, Gerard; Kors, Jan A.; Macfarlane, Peter (13 March 2007). "Recommendations for the standardization and interpretation of the electrocardiogram: part I: The electrocardiogram and its technology: a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society: endorsed by the International Society for Computerized Electrocardiology". Circulation. 115 (10): 1306–1324. doi:10.1161/CIRCULATIONAHA.106.180200. PMID 17322457.
  80. ^ "Minimizing ECG Artifact" (PDF). Physio-Control. Physio-Control, Inc., Redmond WA. 2015. Retrieved 21 October 2017.
  81. ^ Jafary, Fahim H (2007). "The "incidental" episode of ventricular fibrillation: A case report". Journal of Medical Case Reports. 1 (1): 72. doi:10.1186/1752-1947-1-72. PMC 2000884. PMID 17760955.
  82. ^ Mangalmurti, Sandeep; Seabury, Seth A.; Chandra, Amitabh; Lakdawalla, Darius; Oetgen, William J.; Jena, Anupam B. (2014). "Medical professional liability risk among US cardiologists". American Heart Journal. 167 (5): 690–696. doi:10.1016/j.ahj.2014.02.007. PMC 4153384. PMID 24766979.
  83. ^ Batchvarov, Velislav N.; Malik, Marek; Camm, A. John (November 2007). "Incorrect electrode cable connection during electrocardiographic recording". Europace. 9 (11): 1081–1090. doi:10.1093/europace/eum198. ISSN 1532-2092. PMID 17932025.
  84. ^ Chanarin N., Caplin J., Peacock A. (1990). ""Pseudo reinfarction": a consequence of electrocardiogram lead transposition following myocardial infarction". Clinical Cardiology. 13 (9): 668–669. doi:10.1002/clc.4960130916. PMID 2208827.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  85. ^ Guijarro-Morales A., Gil-Extremera B., Maldonado-Martín A. (1991). "ECG diagnostic errors due to improper connection of the right arm and leg cables". International Journal of Cardiology. 30 (2): 233–235. doi:10.1016/0167-5273(91)90103-v. PMID 2010249.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  86. ^ Whitbread, Mark (January 2006). "Reading a normal ECG". British Journal of Cardiac Nursing. 1 (1): 32–33. doi:10.12968/bjca.2006.1.1.20382. ISSN 1749-6403.
  87. ^ Mallinson, Tom (2 March 2023). "Additional rules for reading an electrocardiogram". Journal of Paramedic Practice. 15 (3): 95–97. doi:10.12968/jpar.2023.15.3.95. ISSN 1759-1376. S2CID 257390198.
  88. ^ Montague, Brian T.; Ouellette, Jason R.; Buller, Gregory K. (30 January 2008). "Retrospective Review of the Frequency of ECG Changes in Hyperkalemia". Clinical Journal of the American Society of Nephrology. 3 (2): 324–330. doi:10.2215/CJN.04611007. ISSN 1555-9041. PMC 2390954. PMID 18235147.
  89. ^ "Arrhythmia". nhs.uk. 19 February 2018. Retrieved 28 October 2022.
  90. ^ de Winter, Robert (6 November 2008). "A New ECG Sign of Proximal LAD Occlusion". NEJM. 359 (19): 2071–2073. doi:10.1056/NEJMc0804737. PMID 18987380. S2CID 205040240.
  91. ^ Birse, Ronald M. (23 September 2004). "Muirhead, Alexander (1848–1920), electrical engineer". In Knowlden, Patricia E. (ed.). Oxford Dictionary of National Biography (online ed.). Oxford University Press. doi:10.1093/ref:odnb/37794. Retrieved 20 January 2020. (Subscription or UK public library membership required.)
  92. ^ Rogers, Mark C. (1969). "Historical Annotation: Sir John Scott Burdon-Sanderson (1828-1905) A Pioneer in Electrophysiology". Circulation. 40 (1): 1–2. doi:10.1161/01.CIR.40.1.1. ISSN 0009-7322. PMID 4893441.
  93. ^ Waller AD (1887). "A demonstration on man of electromotive changes accompanying the heart's beat". J Physiol. 8 (5): 229–34. doi:10.1113/jphysiol.1887.sp000257. PMC 1485094. PMID 16991463.
  94. ^ a b c Hurst JW (3 November 1998). "Naming of the Waves in the ECG, With a Brief Account of Their Genesis". Circulation. 98 (18): 1937–42. doi:10.1161/01.CIR.98.18.1937. PMID 9799216.
  95. ^ Interwoven W (1901). "Un nouveau galvanometre". Arch Neerl Sc Ex Nat. 6: 625.
  96. ^ Rivera-Ruiz M, Cajavilca C, Varon J (29 September 1927). "Einthoven's String Galvanometer: The First Electrocardiograph". Texas Heart Institute Journal. 35 (2): 174–78. PMC 2435435. PMID 18612490.
  97. ^ Cooper JK (1986). "Electrocardiography 100 years ago. Origins, pioneers, and contributors". N Engl J Med. 315 (7): 461–64. doi:10.1056/NEJM198608143150721. PMID 3526152.
  98. ^ Blackford, John M., MD (1 May 1927). "Electrocardiography: A Short Talk Before the Staff of the Hospital". Clinics of the Virginia Mason Hospital. 6 (1): 28–34.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  99. ^ "Dr. Taro Takemi". Takemi Program in International Health. 27 August 2012. Retrieved 21 October 2017.
  100. ^ "A (not so) brief history of electrocardiography". 2009.
  101. ^ "A (not so) brief history of electrocardiography". ECG Library. 3 January 2006. from the original on 2 February 2012. Retrieved 11 January 2021.
  102. ^ "The Interesting History of EKGs". info.nhanow.com. Retrieved 21 January 2024.

External links edit

  • The whole ECG course on 1 A4 paper from ECGpedia, a wiki encyclopedia for a course on interpretation of ECG
  • Wave Maven – a large database of practice ECG questions provided by Beth Israel Deaconess Medical Center
  • PysioBank – a free scientific database with physiologic signals (here ecg)
  • EKG Academy – free EKG lectures, drills and quizzes
  • ECG Learning Center created by Eccles Health Sciences Library at University of Utah

electrocardiography, redirect, here, other, uses, disambiguation, disambiguation, confused, with, other, types, electrography, with, echocardiography, process, producing, electrocardiogram, recording, heart, electrical, activity, through, repeated, cardiac, cy. ECG and EKG redirect here For other uses see ECG disambiguation and EKG disambiguation Not to be confused with other types of electrography or with echocardiography Electrocardiography is the process of producing an electrocardiogram ECG or EKG a a recording of the heart s electrical activity through repeated cardiac cycles 4 It is an electrogram of the heart which is a graph of voltage versus time of the electrical activity of the heart 5 using electrodes placed on the skin These electrodes detect the small electrical changes that are a consequence of cardiac muscle depolarization followed by repolarization during each cardiac cycle heartbeat Changes in the normal ECG pattern occur in numerous cardiac abnormalities including Cardiac rhythm disturbances such as atrial fibrillation 6 and ventricular tachycardia 7 Inadequate coronary artery blood flow such as myocardial ischemia 8 and myocardial infarction 9 ElectrocardiographyECG of a heart in normal sinus rhythmICD 10 PCSR94 31ICD 9 CM89 52MeSHD004562MedlinePlus003868 edit on Wikidata Use of real time monitoring of the heart in an intensive care unit in a German hospital 2015 the monitoring screen above the patient displaying an electrocardiogram and various values of parameters of the heart like heart rate and blood pressure and electrolyte disturbances such as hypokalemia 10 Traditionally ECG usually means a 12 lead ECG taken while lying down as discussed below However other devices can record the electrical activity of the heart such as a Holter monitor but also some models of smartwatch are capable of recording an ECG ECG signals can be recorded in other contexts with other devices In a conventional 12 lead ECG ten electrodes are placed on the patient s limbs and on the surface of the chest The overall magnitude of the heart s electrical potential is then measured from twelve different angles leads and is recorded over a period of time usually ten seconds In this way the overall magnitude and direction of the heart s electrical depolarization is captured at each moment throughout the cardiac cycle 11 There are three main components to an ECG 12 The P wave which represents depolarization of the atria The QRS complex which represents depolarization of the ventricles The T wave which represents repolarization of the ventricles During each heartbeat a healthy heart has an orderly progression of depolarization that starts with pacemaker cells in the sinoatrial node spreads throughout the atrium and passes through the atrioventricular node down into the bundle of His and into the Purkinje fibers spreading down and to the left throughout the ventricles 12 This orderly pattern of depolarization gives rise to the characteristic ECG tracing To the trained clinician an ECG conveys a large amount of information about the structure of the heart and the function of its electrical conduction system 13 Among other things an ECG can be used to measure the rate and rhythm of heartbeats the size and position of the heart chambers the presence of any damage to the heart s muscle cells or conduction system the effects of heart drugs and the function of implanted pacemakers 14 Contents 1 Medical uses 1 1 Screening 2 Electrocardiograph machines 2 1 Cardiac monitors 3 Electrodes and leads 3 1 Limb leads 3 2 Augmented limb leads 3 3 Precordial leads 3 4 Specialized leads 3 5 Lead locations on an ECG report 3 6 Contiguity of leads 4 Electrophysiology 5 Interpretation 5 1 Theory 5 2 Background grid 5 3 Rate and rhythm 5 4 Axis 5 5 Amplitudes and intervals 5 6 Limb leads and electrical conduction through the heart 5 7 Ischemia and infarction 5 8 Artifacts 5 9 A Method for Interpretation 6 Diagnosis 7 History 7 1 Etymology 8 See also 9 Notes 10 References 11 External linksMedical uses edit nbsp Normal 12 lead ECG nbsp A 12 lead ECG of a 26 year old male with an incomplete right bundle branch block RBBB The overall goal of performing an ECG is to obtain information about the electrical functioning of the heart Medical uses for this information are varied and often need to be combined with knowledge of the structure of the heart and physical examination signs to be interpreted Some indications for performing an ECG include the following Chest pain or suspected myocardial infarction heart attack such as ST elevated myocardial infarction STEMI 15 or non ST elevated myocardial infarction NSTEMI 16 Symptoms such as shortness of breath murmurs 17 fainting seizures funny turns or arrhythmias including new onset palpitations or monitoring of known cardiac arrhythmias Medication monitoring e g drug induced QT prolongation Digoxin toxicity and management of overdose e g tricyclic overdose Electrolyte abnormalities such as hyperkalemia Perioperative monitoring in which any form of anesthesia is involved e g monitored anesthesia care general anesthesia This includes preoperative assessment and intraoperative and postoperative monitoring Cardiac stress testing Computed tomography angiography CTA and magnetic resonance angiography MRA of the heart ECG is used to gate the scanning so that the anatomical position of the heart is steady Clinical cardiac electrophysiology in which a catheter is inserted through the femoral vein and can have several electrodes along its length to record the direction of electrical activity from within the heart ECGs can be recorded as short intermittent tracings or continuous ECG monitoring Continuous monitoring is used for critically ill patients patients undergoing general anesthesia 18 17 and patients who have an infrequently occurring cardiac arrhythmia that would unlikely be seen on a conventional ten second ECG Continuous monitoring can be conducted by using Holter monitors internal and external defibrillators and pacemakers and or biotelemetry 19 Screening edit nbsp A patient undergoing an ECG For adults evidence does not support the use of ECGs among those without symptoms or at low risk of cardiovascular disease as an effort for prevention 20 21 22 This is because an ECG may falsely indicate the existence of a problem leading to misdiagnosis the recommendation of invasive procedures and overtreatment However persons employed in certain critical occupations such as aircraft pilots 23 may be required to have an ECG as part of their routine health evaluations Hypertrophic cardiomyopathy screening may also be considered in adolescents as part of a sports physical out of concern for sudden cardiac death 24 Electrocardiograph machines edit nbsp An EKG electrode Electrocardiograms are recorded by machines that consist of a set of electrodes connected to a central unit 25 Early ECG machines were constructed with analog electronics where the signal drove a motor to print out the signal onto paper Today electrocardiographs use analog to digital converters to convert the electrical activity of the heart to a digital signal Many ECG machines are now portable and commonly include a screen keyboard and printer on a small wheeled cart Recent advancements in electrocardiography include developing even smaller devices for inclusion in fitness trackers and smart watches 26 These smaller devices often rely on only two electrodes to deliver a single lead I 27 Portable twelve lead devices powered by batteries are also available Recording an ECG is a safe and painless procedure 28 The machines are powered by mains power but they are designed with several safety features including an earthed ground lead Other features include Defibrillation protection any ECG used in healthcare may be attached to a person who requires defibrillation and the ECG needs to protect itself from this source of energy Electrostatic discharge is similar to defibrillation discharge and requires voltage protection up to 18 000 volts Additionally circuitry called the right leg driver can be used to reduce common mode interference typically the 50 or 60 Hz mains power ECG voltages measured across the body are very small This low voltage necessitates a low noise circuit instrumentation amplifiers and electromagnetic shielding Simultaneous lead recordings earlier designs recorded each lead sequentially but current models record multiple leads simultaneously Most modern ECG machines include automated interpretation algorithms This analysis calculates features such as the PR interval QT interval corrected QT QTc interval PR axis QRS axis rhythm and more The results from these automated algorithms are considered preliminary until verified and or modified by expert interpretation Despite recent advances computer misinterpretation remains a significant problem and can result in clinical mismanagement 29 Cardiac monitors edit Besides the standard electrocardiograph machine there are other devices capable of recording ECG signals Portable devices have existed since the Holter monitor was produced in 1962 Traditionally these monitors have used electrodes with patches on the skin to record the ECG but new devices can stick to the chest as a single patch without need for wires developed by Zio Zio XT TZ Medical Trident Philips BioTel and BardyDx CAM among many others Implantable devices such as the artificial cardiac pacemaker and implantable cardioverter defibrillator are capable of measuring a far field signal between the leads in the heart and the implanted battery generator that resembles an ECG signal technically the signal recorded in the heart is called an electrogram which is interpreted differently Advancement of the Holter monitor became the implantable loop recorder that performs the same function but in an implantable device with batteries that last on the order of years Additionally there are available various Arduino kits with ECG sensor modules and smartwatch devices that are capable of recording an ECG signal as well such as with the 4th generation Apple Watch Samsung Galaxy Watch 4 and newer devices Electrodes and leads edit nbsp Proper placement of the limb electrodes The limb electrodes can be far down on the limbs or close to the hips shoulders as long as they are placed symmetrically 30 nbsp Placement of the precordial electrodes Electrodes are the actual conductive pads attached to the body surface 31 Any pair of electrodes can measure the electrical potential difference between the two corresponding locations of attachment Such a pair forms a lead However leads can also be formed between a physical electrode and a virtual electrode known as Wilson s central terminal WCT whose potential is defined as the average potential measured by three limb electrodes that are attached to the right arm the left arm and the left foot respectively 32 Commonly 10 electrodes attached to the body are used to form 12 ECG leads with each lead measuring a specific electrical potential difference as listed in the table below 33 Leads are broken down into three types limb augmented limb and precordial or chest The 12 lead ECG has a total of three limb leads and three augmented limb leads arranged like spokes of a wheel in the coronal plane vertical and six precordial leads or chest leads that lie on the perpendicular transverse plane horizontal 34 In medical settings the term leads is also sometimes used to refer to the electrodes themselves although this is technically incorrect 35 The 10 electrodes in a 12 lead ECG are listed below 36 Electrode name Electrode placement RA On the right arm avoiding thick muscle LA In the same location where RA was placed but on the left arm RL On the right leg lower end of inner aspect of calf muscle Avoid bony prominences LL In the same location where RL was placed but on the left leg V1 In the fourth intercostal space between ribs 4 and 5 just to the right of the sternum breastbone V2 In the fourth intercostal space between ribs 4 and 5 just to the left of the sternum V3 Between leads V2 and V4 V4 In the fifth intercostal space between ribs 5 and 6 in the mid clavicular line V5 Horizontally even with V4 in the left anterior axillary line V6 Horizontally even with V4 and V5 in the mid axillary line Two types of electrodes in common use are a flat paper thin sticker and a self adhesive circular pad The former are typically used in a single ECG recording while the latter are for continuous recordings as they stick longer Each electrode consists of an electrically conductive electrolyte gel and a silver silver chloride conductor 37 The gel typically contains potassium chloride sometimes silver chloride as well to permit electron conduction from the skin to the wire and to the electrocardiogram 38 The common virtual electrode known as Wilson s central terminal VW is produced by averaging the measurements from the electrodes RA LA and LL to give an average potential of the body V W 1 3 R A L A L L displaystyle V W frac 1 3 RA LA LL nbsp In a 12 lead ECG all leads except the limb leads are assumed to be unipolar aVR aVL aVF V1 V2 V3 V4 V5 and V6 The measurement of a voltage requires two contacts and so electrically the unipolar leads are measured from the common lead negative and the unipolar lead positive This averaging for the common lead and the abstract unipolar lead concept makes for a more challenging understanding and is complicated by sloppy usage of lead and electrode In fact instead of being a constant reference VW has a value that fluctuates throughout the heart cycle It also does not truly represent the center of heart potential due to the body parts the signals travel through 39 Limb leads edit nbsp The limb leads and augmented limb leads Wilson s central terminal is used as the negative pole for the latter in this representation nbsp Leads I II and III are called the limb leads The electrodes that form these signals are located on the limbs one on each arm and one on the left leg 40 41 42 The limb leads form the points of what is known as Einthoven s triangle 43 Lead I is the voltage between the positive left arm LA electrode and right arm RA electrode I L A R A displaystyle I LA RA nbsp Lead II is the voltage between the positive left leg LL electrode and the right arm RA electrode I I L L R A displaystyle II LL RA nbsp Lead III is the voltage between the positive left leg LL electrode and the left arm LA electrode I I I L L L A displaystyle III LL LA nbsp Augmented limb leads edit Leads aVR aVL and aVF are the augmented limb leads They are derived from the same three electrodes as leads I II and III but they use Goldberger s central terminal as their negative pole Goldberger s central terminal is a combination of inputs from two limb electrodes with a different combination for each augmented lead It is referred to immediately below as the negative pole Lead augmented vector right aVR has the positive electrode on the right arm The negative pole is a combination of the left arm electrode and the left leg electrode a V R R A 1 2 L A L L 3 2 R A V W displaystyle aVR RA frac 1 2 LA LL frac 3 2 RA V W nbsp Lead augmented vector left aVL has the positive electrode on the left arm The negative pole is a combination of the right arm electrode and the left leg electrode a V L L A 1 2 R A L L 3 2 L A V W displaystyle aVL LA frac 1 2 RA LL frac 3 2 LA V W nbsp Lead augmented vector foot aVF has the positive electrode on the left leg The negative pole is a combination of the right arm electrode and the left arm electrode a V F L L 1 2 R A L A 3 2 L L V W displaystyle aVF LL frac 1 2 RA LA frac 3 2 LL V W nbsp Together with leads I II and III augmented limb leads aVR aVL and aVF form the basis of the hexaxial reference system which is used to calculate the heart s electrical axis in the frontal plane 44 Older versions of the nodes VR VL VF use Wilson s central terminal as the negative pole but the amplitude is too small for the thick lines of old ECG machines The Goldberger terminals scale up augments the Wilson results by 50 at the cost of sacrificing physical correctness by not having the same negative pole for all three 45 Precordial leads edit The precordial leads lie in the transverse horizontal plane perpendicular to the other six leads The six precordial electrodes act as the positive poles for the six corresponding precordial leads V1 V2 V3 V4 V5 and V6 Wilson s central terminal is used as the negative pole Recently unipolar precordial leads have been used to create bipolar precordial leads that explore the right to left axis in the horizontal plane 46 Specialized leads edit Additional electrodes may rarely be placed to generate other leads for specific diagnostic purposes Right sided precordial leads may be used to better study pathology of the right ventricle or for dextrocardia and are denoted with an R e g V5R Posterior leads V7 to V9 may be used to demonstrate the presence of a posterior myocardial infarction The Lewis lead or S5 lead requiring an electrode at the right sternal border in the second intercostal space can be used to better detect atrial activity in relation to that of the ventricles 47 An esophageal lead can be inserted to a part of the esophagus where the distance to the posterior wall of the left atrium is only approximately 5 6 mm remaining constant in people of different age and weight 48 An esophageal lead avails for a more accurate differentiation between certain cardiac arrhythmias particularly atrial flutter AV nodal reentrant tachycardia and orthodromic atrioventricular reentrant tachycardia 49 It can also evaluate the risk in people with Wolff Parkinson White syndrome as well as terminate supraventricular tachycardia caused by re entry 49 An intracardiac electrogram ICEG is essentially an ECG with some added intracardiac leads that is inside the heart The standard ECG leads external leads are I II III aVL V1 and V6 Two to four intracardiac leads are added via cardiac catheterization The word electrogram EGM without further specification usually means an intracardiac electrogram 50 Lead locations on an ECG report edit A standard 12 lead ECG report an electrocardiograph shows a 2 5 second tracing of each of the twelve leads The tracings are most commonly arranged in a grid of four columns and three rows The first column is the limb leads I II and III the second column is the augmented limb leads aVR aVL and aVF and the last two columns are the precordial leads V1 to V6 Additionally a rhythm strip may be included as a fourth or fifth row 44 The timing across the page is continuous and notes tracings of the 12 leads for the same time period In other words if the output were traced by needles on paper each row would switch which leads as the paper is pulled under the needle For example the top row would first trace lead I then switch to lead aVR then switch to V1 and then switch to V4 and so none of these four tracings of the leads are from the same time period as they are traced in sequence through time 51 Contiguity of leads edit nbsp Diagram showing the contiguous leads in the same color in the standard 12 lead layout Each of the 12 ECG leads records the electrical activity of the heart from a different angle and therefore align with different anatomical areas of the heart Two leads that look at neighboring anatomical areas are said to be contiguous 44 Category Leads Activity Inferior leads Leads II III and aVF Look at electrical activity from the vantage point of the inferior surface diaphragmatic surface of heart Lateral leads I aVL V5 and V6 Look at the electrical activity from the vantage point of the lateral wall of left ventricle Septal leads V1 and V2 Look at electrical activity from the vantage point of the septal surface of the heart interventricular septum Anterior leads V3 and V4 Look at electrical activity from the vantage point of the anterior wall of the right and left ventricles Sternocostal surface of heart In addition any two precordial leads next to one another are considered to be contiguous For example though V4 is an anterior lead and V5 is a lateral lead they are contiguous because they are next to one another Electrophysiology editMain article Cardiac electrophysiology The study of the conduction system of the heart is called cardiac electrophysiology EP An EP study is performed via a right sided cardiac catheterization a wire with an electrode at its tip is inserted into the right heart chambers from a peripheral vein and placed in various positions in close proximity to the conduction system so that the electrical activity of that system can be recorded 52 Standard catheter positions for an EP study include high right atrium or hRA near the sinus node a His across the septal wall of the tricuspid valve to measure bundle of His a coronary sinus into the coronary sinus and a right ventricle in the apex of the right ventricle 53 Interpretation editInterpretation of the ECG is fundamentally about understanding the electrical conduction system of the heart Normal conduction starts and propagates in a predictable pattern and deviation from this pattern can be a normal variation or be pathological An ECG does not equate with mechanical pumping activity of the heart for example pulseless electrical activity produces an ECG that should pump blood but no pulses are felt and constitutes a medical emergency and CPR should be performed Ventricular fibrillation produces an ECG but is too dysfunctional to produce a life sustaining cardiac output Certain rhythms are known to have good cardiac output and some are known to have bad cardiac output Ultimately an echocardiogram or other anatomical imaging modality is useful in assessing the mechanical function of the heart 54 Like all medical tests what constitutes normal is based on population studies The heartrate range of between 60 and 100 beats per minute bpm is considered normal since data shows this to be the usual resting heart rate 55 Theory edit nbsp QRS is upright in a lead when its axis is aligned with that lead s vector nbsp Schematic representation of a normal ECG Interpretation of the ECG is ultimately that of pattern recognition In order to understand the patterns found it is helpful to understand the theory of what ECGs represent The theory is rooted in electromagnetics and boils down to the four following points 56 depolarization of the heart toward the positive electrode produces a positive deflection depolarization of the heart away from the positive electrode produces a negative deflection repolarization of the heart toward the positive electrode produces a negative deflection repolarization of the heart away from the positive electrode produces a positive deflection Thus the overall direction of depolarization and repolarization produces positive or negative deflection on each lead s trace For example depolarizing from right to left would produce a positive deflection in lead I because the two vectors point in the same direction In contrast that same depolarization would produce minimal deflection in V1 and V2 because the vectors are perpendicular and this phenomenon is called isoelectric Normal rhythm produces four entities a P wave a QRS complex a T wave and a U wave that each have a fairly unique pattern The P wave represents atrial depolarization The QRS complex represents ventricular depolarization The T wave represents ventricular repolarization The U wave represents papillary muscle repolarization Changes in the structure of the heart and its surroundings including blood composition change the patterns of these four entities The U wave is not typically seen and its absence is generally ignored Atrial repolarization is typically hidden in the much more prominent QRS complex and normally cannot be seen without additional specialized electrodes Background grid edit ECGs are normally printed on a grid The horizontal axis represents time and the vertical axis represents voltage The standard values on this grid are shown in the adjacent image at 25mm sec 57 A small box is 1 mm 1 mm and represents 0 1 mV 0 04 seconds A large box is 5 mm 5 mm and represents 0 5 mV 0 20 seconds The large box is represented by a heavier line weight than the small boxes nbsp Measuring time and voltage with ECG graph paper The standard printing speed in the United States is 25 mm per sec 5 big boxes per second but in other countries it can be 50 mm per sec Faster speeds such as 100 and 200 mm per sec are used during electrophysiology studies Not all aspects of an ECG rely on precise recordings or having a known scaling of amplitude or time For example determining if the tracing is a sinus rhythm only requires feature recognition and matching and not measurement of amplitudes or times i e the scale of the grids are irrelevant An example to the contrary the voltage requirements of left ventricular hypertrophy require knowing the grid scale Rate and rhythm edit In a normal heart the heart rate is the rate at which the sinoatrial node depolarizes since it is the source of depolarization of the heart Heart rate like other vital signs such as blood pressure and respiratory rate change with age In adults a normal heart rate is between 60 and 100 bpm normocardic whereas it is higher in children 58 A heart rate below normal is called bradycardia lt 60 in adults and above normal is called tachycardia gt 100 in adults A complication of this is when the atria and ventricles are not in synchrony and the heart rate must be specified as atrial or ventricular e g the ventricular rate in ventricular fibrillation is 300 600 bpm whereas the atrial rate can be normal 60 100 or faster 100 150 59 In normal resting hearts the physiologic rhythm of the heart is normal sinus rhythm NSR Normal sinus rhythm produces the prototypical pattern of P wave QRS complex and T wave Generally deviation from normal sinus rhythm is considered a cardiac arrhythmia Thus the first question in interpreting an ECG is whether or not there is a sinus rhythm A criterion for sinus rhythm is that P waves and QRS complexes appear 1 to 1 thus implying that the P wave causes the QRS complex 51 Once sinus rhythm is established or not the second question is the rate For a sinus rhythm this is either the rate of P waves or QRS complexes since they are 1 to 1 If the rate is too fast then it is sinus tachycardia and if it is too slow then it is sinus bradycardia If it is not a sinus rhythm then determining the rhythm is necessary before proceeding with further interpretation Some arrhythmias with characteristic findings Absent P waves with irregularly irregular QRS complexes are the hallmark of atrial fibrillation A saw tooth pattern with QRS complexes is the hallmark of atrial flutter A sine wave pattern is the hallmark of ventricular flutter Absent P waves with wide QRS complexes and a fast heart rate are ventricular tachycardia Determination of rate and rhythm is necessary in order to make sense of further interpretation Axis edit nbsp Diagram showing how the polarity of the QRS complex in leads I II and III can be used to estimate the heart s electrical axis in the frontal plane The heart has several axes but the most common by far is the axis of the QRS complex references to the axis imply the QRS axis Each axis can be computationally determined to result in a number representing degrees of deviation from zero or it can be categorized into a few types 60 The QRS axis is the general direction of the ventricular depolarization wavefront or mean electrical vector in the frontal plane It is often sufficient to classify the axis as one of three types normal left deviated or right deviated Population data shows that a normal QRS axis is from 30 to 105 with 0 being along lead I and positive being inferior and negative being superior best understood graphically as the hexaxial reference system 61 Beyond 105 is right axis deviation and beyond 30 is left axis deviation the third quadrant of 90 to 180 is very rare and is an indeterminate axis A shortcut for determining if the QRS axis is normal is if the QRS complex is mostly positive in lead I and lead II or lead I and aVF if 90 is the upper limit of normal 62 The normal QRS axis is generally down and to the left following the anatomical orientation of the heart within the chest An abnormal axis suggests a change in the physical shape and orientation of the heart or a defect in its conduction system that causes the ventricles to depolarize in an abnormal way 51 Classification Angle Notes Normal 30 to 105 Normal Left axis deviation 30 to 90 May indicate left ventricular hypertrophy left anterior fascicular block or an old inferior STEMI Right axis deviation 105 to 180 May indicate right ventricular hypertrophy left posterior fascicular block or an old lateral STEMI Indeterminate axis 180 to 90 Rarely seen considered an electrical no man s land The extent of a normal axis can be 90 or 105 depending on the source Amplitudes and intervals edit nbsp Animation of a normal ECG wave All of the waves on an ECG tracing and the intervals between them have a predictable time duration a range of acceptable amplitudes voltages and a typical morphology Any deviation from the normal tracing is potentially pathological and therefore of clinical significance 63 For ease of measuring the amplitudes and intervals an ECG is printed on graph paper at a standard scale each 1 mm one small box on the standard 25mm s ECG paper represents 40 milliseconds of time on the x axis and 0 1 millivolts on the y axis 64 Feature Description Pathology Duration P wave The P wave represents depolarization of the atria Atrial depolarization spreads from the SA node towards the AV node and from the right atrium to the left atrium The P wave is typically upright in most leads except for aVR an unusual P wave axis inverted in other leads can indicate an ectopic atrial pacemaker If the P wave is of unusually long duration it may represent atrial enlargement Typically a large right atrium gives a tall peaked P wave while a large left atrium gives a two humped bifid P wave lt 80 ms PR interval The PR interval is measured from the beginning of the P wave to the beginning of the QRS complex This interval reflects the time the electrical impulse takes to travel from the sinus node through the AV node A PR interval shorter than 120 ms suggests that the electrical impulse is bypassing the AV node as in Wolff Parkinson White syndrome A PR interval consistently longer than 200 ms diagnoses first degree atrioventricular block The PR segment the portion of the tracing after the P wave and before the QRS complex is typically completely flat but may be depressed in pericarditis 120 to 200 ms QRS complex The QRS complex represents the rapid depolarization of the right and left ventricles The ventricles have a large muscle mass compared to the atria so the QRS complex usually has a much larger amplitude than the P wave If the QRS complex is wide longer than 120 ms it suggests disruption of the heart s conduction system such as in LBBB RBBB or ventricular rhythms such as ventricular tachycardia Metabolic issues such as severe hyperkalemia or tricyclic antidepressant overdose can also widen the QRS complex An unusually tall QRS complex may represent left ventricular hypertrophy while a very low amplitude QRS complex may represent a pericardial effusion or infiltrative myocardial disease 80 to 100 ms J point The J point is the point at which the QRS complex finishes and the ST segment begins The J point may be elevated as a normal variant The appearance of a separate J wave or Osborn wave at the J point is pathognomonic of hypothermia or hypercalcemia 65 ST segment The ST segment connects the QRS complex and the T wave it represents the period when the ventricles are depolarized It is usually isoelectric but may be depressed or elevated with myocardial infarction or ischemia ST depression can also be caused by LVH or digoxin ST elevation can also be caused by pericarditis Brugada syndrome or can be a normal variant J point elevation T wave The T wave represents the repolarization of the ventricles It is generally upright in all leads except aVR and lead V1 Inverted T waves can be a sign of myocardial ischemia left ventricular hypertrophy high intracranial pressure or metabolic abnormalities Peaked T waves can be a sign of hyperkalemia or very early myocardial infarction 160 ms Corrected QT interval QTc The QT interval is measured from the beginning of the QRS complex to the end of the T wave Acceptable ranges vary with heart rate so it must be corrected to the QTc by dividing by the square root of the RR interval A prolonged QTc interval is a risk factor for ventricular tachyarrhythmias and sudden death Long QT can arise as a genetic syndrome or as a side effect of certain medications An unusually short QTc can be seen in severe hypercalcemia lt 440 ms U wave The U wave is hypothesized to be caused by the repolarization of the interventricular septum It normally has a low amplitude and even more often is completely absent A very prominent U wave can be a sign of hypokalemia hypercalcemia or hyperthyroidism 66 Limb leads and electrical conduction through the heart edit nbsp Formation of limb waveforms during a pulse The animation shown to the right illustrates how the path of electrical conduction gives rise to the ECG waves in the limb leads Recall that a positive current as created by depolarization of cardiac cells traveling towards the positive electrode and away from the negative electrode creates a positive deflection on the ECG Likewise a positive current traveling away from the positive electrode and towards the negative electrode creates a negative deflection on the ECG 67 68 The red arrow represents the overall direction of travel of the depolarization The magnitude of the red arrow is proportional to the amount of tissue being depolarized at that instance The red arrow is simultaneously shown on the axis of each of the 3 limb leads Both the direction and the magnitude of the red arrow s projection onto the axis of each limb lead is shown with blue arrows Then the direction and magnitude of the blue arrows are what theoretically determine the deflections on the ECG For example as a blue arrow on the axis for Lead I moves from the negative electrode to the right towards the positive electrode the ECG line rises creating an upward wave As the blue arrow on the axis for Lead I moves to the left a downward wave is created The greater the magnitude of the blue arrow the greater the deflection on the ECG for that particular limb lead 69 Frames 1 3 depict the depolarization being generated in and spreading through the Sinoatrial node The SA node is too small for its depolarization to be detected on most ECGs Frames 4 10 depict the depolarization traveling through the atria towards the Atrioventricular node During frame 7 the depolarization is traveling through the largest amount of tissue in the atria which creates the highest point in the P wave Frames 11 12 depict the depolarization traveling through the AV node Like the SA node the AV node is too small for the depolarization of its tissue to be detected on most ECGs This creates the flat PR segment 70 Frame 13 depicts an interesting phenomenon in an over simplified fashion It depicts the depolarization as it starts to travel down the interventricular septum through the Bundle of His and Bundle branches After the Bundle of His the conduction system splits into the left bundle branch and the right bundle branch Both branches conduct action potentials at about 1 m s Interestingly however the action potential starts traveling down the left bundle branch about 5 milliseconds before it starts traveling down the right bundle branch as depicted by frame 13 This causes the depolarization of the interventricular septum tissue to spread from left to right as depicted by the red arrow in frame 14 In some cases this gives rise to a negative deflection after the PR interval creating a Q wave such as the one seen in lead I in the animation to the right Depending on the mean electrical axis of the heart this phenomenon can result in a Q wave in lead II as well 71 72 Following depolarization of the interventricular septum the depolarization travels towards the apex of the heart This is depicted by frames 15 17 and results in a positive deflection on all three limb leads which creates the R wave Frames 18 21 then depict the depolarization as it travels throughout both ventricles from the apex of the heart following the action potential in the Purkinje fibers This phenomenon creates a negative deflection in all three limb leads forming the S wave on the ECG Repolarization of the atria occurs at the same time as the generation of the QRS complex but it is not detected by the ECG since the tissue mass of the ventricles is so much larger than that of the atria Ventricular contraction occurs between ventricular depolarization and repolarization During this time there is no movement of charge so no deflection is created on the ECG This results in the flat ST segment after the S wave 73 Frames 24 28 in the animation depict repolarization of the ventricles The epicardium is the first layer of the ventricles to repolarize followed by the myocardium The endocardium is the last layer to repolarize The plateau phase of depolarization has been shown to last longer in endocardial cells than in epicardial cells This causes repolarization to start from the apex of the heart and move upwards Since repolarization is the spread of negative current as membrane potentials decrease back down to the resting membrane potential the red arrow in the animation is pointing in the direction opposite of the repolarization This therefore creates a positive deflection in the ECG and creates the T wave 74 Ischemia and infarction edit Main article Electrocardiography in myocardial infarction Ischemia or non ST elevation myocardial infarctions non STEMIs may manifest as ST depression or inversion of T waves It may also affect the high frequency band of the QRS ST elevation myocardial infarctions STEMIs have different characteristic ECG findings based on the amount of time elapsed since the MI first occurred The earliest sign is hyperacute T waves peaked T waves due to local hyperkalemia in ischemic myocardium This then progresses over a period of minutes to elevations of the ST segment by at least 1 mm Over a period of hours a pathologic Q wave may appear and the T wave will invert Over a period of days the ST elevation will resolve Pathologic Q waves generally will remain permanently 75 The coronary artery that has been occluded can be identified in an STEMI based on the location of ST elevation The left anterior descending LAD artery supplies the anterior wall of the heart and therefore causes ST elevations in anterior leads V1 and V2 The LCx supplies the lateral aspect of the heart and therefore causes ST elevations in lateral leads I aVL and V6 The right coronary artery RCA usually supplies the inferior aspect of the heart and therefore causes ST elevations in inferior leads II III and aVF 76 Artifacts edit An ECG tracing is affected by patient motion Some rhythmic motions such as shivering or tremors can create the illusion of cardiac arrhythmia 77 Artifacts are distorted signals caused by a secondary internal or external sources such as muscle movement or interference from an electrical device 78 79 Distortion poses significant challenges to healthcare providers 78 who employ various techniques 80 and strategies to safely recognize 81 these false signals medical citation needed Accurately separating the ECG artifact from the true ECG signal can have a significant impact on patient outcomes and legal liabilities 82 unreliable medical source Improper lead placement for example reversing two of the limb leads has been estimated to occur in 0 4 to 4 of all ECG recordings 83 and has resulted in improper diagnosis and treatment including unnecessary use of thrombolytic therapy 84 85 A Method for Interpretation edit Whitbread consultant nurse and paramedic suggests ten rules of the normal ECG deviation from which is likely to indicate pathology 86 These have been added to creating the 15 rules for 12 lead and 15 or 18 lead interpretation 87 Rule 1 All waves in aVR are negative Rule 2 The ST segment J point starts on the isoelectric line except in V1 amp V2 where it may be elevated by not greater than 1 mm Rule 3 The PR interval should be 0 12 0 2 seconds long Rule 4 The QRS complex should not exceed 0 11 0 12 seconds Rule 5 The QRS and T waves tend to have the same general direction in the limb leads Rule 6 The R wave in the precordial chest leads grows from V1 to at least V4 where it may or may not decline again Rule 7 The QRS is mainly upright in I and II Rule 8 The P wave is upright in I II and V2 to V6 Rule 9 There is no Q wave or only a small q lt 0 04 seconds in width in I II and V2 to V6 Rule 10 The T wave is upright in I II and V2 to V6 The end of the T wave should not drop below the isoelectric baseline Rule 11 Does the deepest S wave in V1 plus the tallest R wave in V5 or V6 equal gt 35 mm Rule 12 Is there an Epsilon wave Rule 13 Is there an J wave Rule 14 Is there a Delta wave Rule 15 Are there any patterns representing an occlusive myocardial infarction OMI Diagnosis editNumerous diagnoses and findings can be made based upon electrocardiography and many are discussed above Overall the diagnoses are made based on the patterns For example an irregularly irregular QRS complex without P waves is the hallmark of atrial fibrillation however other findings can be present as well such as a bundle branch block that alters the shape of the QRS complexes ECGs can be interpreted in isolation but should be applied like all diagnostic tests in the context of the patient For example an observation of peaked T waves is not sufficient to diagnose hyperkalemia such a diagnosis should be verified by measuring the blood potassium level Conversely a discovery of hyperkalemia should be followed by an ECG for manifestations such as peaked T waves widened QRS complexes and loss of P waves The following is an organized list of possible ECG based diagnoses 88 Rhythm disturbances or arrhythmias 89 Atrial fibrillation and atrial flutter without rapid ventricular response Premature atrial contraction PACs and premature ventricular contraction PVCs Sinus arrhythmia Sinus bradycardia and sinus tachycardia Sinus pause and sinoatrial arrest Sinus node dysfunction and bradycardia tachycardia syndrome Supraventricular tachycardia Atrial fibrillation with rapid ventricular response Atrial flutter with rapid ventricular response AV nodal reentrant tachycardia Atrioventricular reentrant tachycardia Junctional ectopic tachycardia Atrial tachycardia Ectopic atrial tachycardia unicentric Multifocal atrial tachycardia Paroxysmal atrial tachycardia Sinoatrial nodal reentrant tachycardia Wide complex tachycardia Ventricular flutter Ventricular fibrillation Ventricular tachycardia monomorphic ventricular tachycardia Torsades de pointes polymorphic ventricular tachycardia Pre excitation syndrome Lown Ganong Levine syndrome Wolff Parkinson White syndrome J wave Osborn wave Heart block and conduction problems Sinoatrial block first second and third degree AV node First degree AV block Second degree AV block Mobitz Wenckebach I and II Third degree AV block or complete AV block Right bundle Incomplete right bundle branch block IRBBB Complete right bundle branch block RBBB Left bundle Incomplete left bundle branch block ILBBB Complete left bundle branch block LBBB Left anterior fascicular block LAFB Left posterior fascicular block LPFB Bifascicular block LAFB plus LPFB Trifascicular block LAFP plus FPFB plus RBBB QT syndromes Brugada syndrome Short QT syndrome Long QT syndromes genetic and drug induced Right and left atrial abnormality Electrolytes disturbances and intoxication Digitalis intoxication Calcium hypocalcemia and hypercalcemia Potassium hypokalemia and hyperkalemia Serotonin Toxicity Ischemia and infarction Wellens syndrome LAD occlusion de Winter T waves LAD occlusion 90 ST elevation and ST depression High Frequency QRS changes Myocardial infarction heart attack Non Q wave myocardial infarction NSTEMI STEMI Sgarbossa s criteria for ischemia with a LBBB Structural Acute pericarditis Right and left ventricular hypertrophy Right ventricular strain or S1Q3T3 can be seen in pulmonary embolism Other phenomena Cardiac aberrancy Ashman phenomenon Concealed conduction Electrical alternansHistory edit nbsp An early commercial ECG device 1911 nbsp ECG from 1957 In 1872 Alexander Muirhead is reported to have attached wires to the wrist of a patient with fever to obtain an electronic record of their heartbeat 91 In 1882 John Burdon Sanderson working with frogs was the first to appreciate that the interval between variations in potential was not electrically quiescent and coined the term isoelectric interval for this period 92 In 1887 Augustus Waller 93 invented an ECG machine consisting of a Lippmann capillary electrometer fixed to a projector The trace from the heartbeat was projected onto a photographic plate that was itself fixed to a toy train This allowed a heartbeat to be recorded in real time In 1895 Willem Einthoven assigned the letters P Q R S and T to the deflections in the theoretical waveform he created using equations which corrected the actual waveform obtained by the capillary electrometer to compensate for the imprecision of that instrument Using letters different from A B C and D the letters used for the capillary electrometer s waveform facilitated comparison when the uncorrected and corrected lines were drawn on the same graph 94 Einthoven probably chose the initial letter P to follow the example set by Descartes in geometry 94 When a more precise waveform was obtained using the string galvanometer which matched the corrected capillary electrometer waveform he continued to use the letters P Q R S and T 94 and these letters are still in use today Einthoven also described the electrocardiographic features of a number of cardiovascular disorders In 1897 the string galvanometer was invented by the French engineer Clement Ader 95 In 1901 Einthoven working in Leiden the Netherlands used the string galvanometer the first practical ECG 96 This device was much more sensitive than the capillary electrometer Waller used In 1924 Einthoven was awarded the Nobel Prize in Medicine for his pioneering work in developing the ECG 97 By 1927 General Electric had developed a portable apparatus that could produce electrocardiograms without the use of the string galvanometer This device instead combined amplifier tubes similar to those used in a radio with an internal lamp and a moving mirror that directed the tracing of the electric pulses onto film 98 In 1937 Taro Takemi invented a new portable electrocardiograph machine 99 In 1942 Emanuel Goldberger increases the voltage of Wilson s unipolar leads by 50 and creates the augmented limb leads aVR aVL and aVF When added to Einthoven s three limb leads and the six chest leads we arrive at the 12 lead electrocardiogram that is used today 100 In the late 1940s Rune Elmqvist invented an inkjet printer involving thin jets of ink deflected by electrical potentials from the heart with good frequency response and direct recording of ECG on paper The device called the Mingograf was sold by Siemens Elema until the 1990s 101 Etymology edit The word is derived from the Greek electro meaning related to electrical activity kardia meaning heart and graph meaning to write 102 See also editSignal averaged electrocardiogram Electrical conduction system of the heart Electroencephalography Electrogastrogram Electropalatography Electroretinography Emergency medicine Forward problem of electrocardiology Heart rate Heart rate monitor KardiaMobile Wireless ambulatory ECGNotes edit The version with K more commonly used in American English than in British English is an early 20th century loanword from the German acronym EKG for Elektrokardiogramm electrocardiogram 1 which reflects that German physicians were pioneers in the field at the time Today AMA style and under its stylistic influence most American medical publications use ECG instead of EKG 2 The German term Elektrokardiogramm as well as the English equivalent electrocardiogram consist of the Neo Latin international scientific vocabulary elements elektro cognate electro and kardi cognate cardi the latter from Greek kardia heart 3 The K version is more often retained under circumstances where there may be verbal confusion between ECG and EEG electroencephalography due to similar pronunciation References edit Definition of EKG Lexico Dictionaries Archived from the original on 15 February 2020 Retrieved 20 January 2020 15 3 1 Electrocardiographic Terms AMA Manual of Style American Medical Association Merriam Webster s Collegiate Dictionary Merriam Webster Bunce Nicholas H Ray Robin Patel Hitesh 2020 30 Cardiology In Feather Adam Randall David Waterhouse Mona eds Kumar and Clark s Clinical Medicine 10th ed Elsevier pp 1033 1038 ISBN 978 0 7020 7870 5 Lilly Leonard S 2016 Pathophysiology of Heart Disease A Collaborative Project of Medical Students and Faculty 6th Edition Lippincott Williams amp Wilkins pp 70 78 ISBN 978 1 4698 9758 5 OCLC 1229852550 Lyakhov Pavel Kiladze Mariya Lyakhova Ulyana January 2021 System for Neural Network Determination of Atrial Fibrillation on ECG Signals with Wavelet Based Preprocessing Applied Sciences 11 16 7213 doi 10 3390 app11167213 Hoyland Philip Hammache Nefissa Battaglia Alberto Oster Julien Felblinger Jacques de Chillou Christian Odille Freddy 2020 A Paced ECG Detector and Delineator for Automatic Multi Parametric Catheter Mapping of Ventricular Tachycardia IEEE Access 8 223952 223960 Bibcode 2020IEEEA 8v3952H doi 10 1109 ACCESS 2020 3043542 ISSN 2169 3536 Bigler Marius Reto Zimmermann Patrick Papadis Athanasios Seiler Christian 1 January 2021 Accuracy of intracoronary ECG parameters for myocardial ischemia detection Journal of Electrocardiology 64 50 57 doi 10 1016 j jelectrocard 2020 11 018 ISSN 0022 0736 PMID 33316551 S2CID 229173576 Prabhakararao Eedara Dandapat Samarendra August 2020 Myocardial Infarction Severity Stages Classification From ECG Signals Using Attentional Recurrent Neural Network IEEE Sensors Journal 20 15 8711 8720 Bibcode 2020ISenJ 20 8711P doi 10 1109 JSEN 2020 2984493 ISSN 1558 1748 S2CID 216310175 Carrizales Sepulveda Edgar Francisco Vera Pineda Raymundo Jimenez Castillo Raul Alberto Trevino Garcia Karla Belen Ordaz Farias Alejandro 1 November 2019 Toluene toxicity presenting with hypokalemia profound weakness and U waves in the electrocardiogram The American Journal of Emergency Medicine 37 11 2120 e1 2120 e3 doi 10 1016 j ajem 2019 158417 ISSN 0735 6757 PMID 31477355 S2CID 201804610 Aswini Kumar MD ECG simplified LifeHugger Archived from the original on 2 October 2017 Retrieved 11 February 2010 a b Lilly 2016 pp 80 Walraven Gail 2011 Basic arrhythmias 7th ed Boston Brady Pearson pp 1 11 ISBN 978 0 13 500238 4 OCLC 505018241 Braunwald Eugene ed 1997 Heart Disease A Textbook of Cardiovascular Medicine 5th ed Philadelphia Saunders p 118 ISBN 0 7216 5666 8 OCLC 32970742 What is a STEMI ECG Medical Training ECG Medical Training 24 June 2015 Retrieved 24 June 2018 What is NSTEMI What You NEED to Know MyHeart 30 April 2015 Retrieved 24 June 2018 a b Masters Jo Bowden Carole Martin Carole Chandler Sharon 2003 Textbook of veterinary medical nursing in Spanish New York Butterworth Heinemann p 244 ISBN 978 0 7506 5171 4 OCLC 53094318 Drew B J Califf R M Funk M Kaufman E S Krucoff M W Laks M M Macfarlane P W Sommargren C Swiryn S Van Hare G F 26 October 2004 Practice Standards for Electrocardiographic Monitoring in Hospital Settings Circulation 110 17 2721 2746 doi 10 1161 01 CIR 0000145144 56673 59 PMID 15505110 S2CID 220573469 Galli Alessio Ambrosini Francesco Lombardi Federico August 2016 Holter Monitoring and Loop Recorders From Research to Clinical Practice Arrhythmia amp Electrophysiology Review 5 2 136 143 doi 10 15420 AER 2016 17 2 ISSN 2050 3369 PMC 5013174 PMID 27617093 US Preventive Services Task Force Curry SJ Krist AH Owens DK Barry MJ Caughey AB Davidson KW Doubeni CA Epling JW Jr Kemper AR Kubik M Landefeld CS Mangione CM Silverstein M Simon MA Tseng CW Wong JB 12 June 2018 Screening for Cardiovascular Disease Risk With Electrocardiography US Preventive Services Task Force Recommendation Statement JAMA 319 22 2308 2314 doi 10 1001 jama 2018 6848 PMID 29896632 Moyer VA 2 October 2012 Screening for coronary heart disease with electrocardiography U S Preventive Services Task Force recommendation statement Annals of Internal Medicine 157 7 512 518 doi 10 7326 0003 4819 157 7 201210020 00514 PMID 22847227 Consumer Reports American Academy of Family Physicians ABIM Foundation April 2012 EKGs and exercise stress tests When you need them for heart disease and when you don t PDF Choosing Wisely Consumer Reports archived from the original PDF on 20 December 2013 retrieved 14 August 2012 Summary of Medical Standards PDF U S Federal Aviation Administration 2006 Retrieved 27 December 2013 Corrado D Basso C Schiavon M Thiene G 6 August 1998 Screening for hypertrophic cardiomyopathy in young athletes The New England Journal of Medicine 339 6 364 369 doi 10 1056 NEJM199808063390602 ISSN 0028 4793 PMID 9691102 Electrocardiograph ECG PDF World Health Organization Retrieved 1 August 2020 How we ll invent the future by Bill Gates MIT Technology Review Retrieved 1 April 2019 FDA approves AliveCor heart monitor Techcrunch Retrieved 25 August 2018 EKG Risks Stanford Health Care Retrieved 1 April 2019 Schlapfer J Wellens HJ 29 August 2017 Computer Interpreted Electrocardiograms Benefits and Limitations Journal of the American College of Cardiology 70 9 1183 1192 doi 10 1016 j jacc 2017 07 723 PMID 28838369 Macfarlane P W Coleman 1995 Resting 12 Lead Electrode PDF Society for Cardiological Science and Technology Archived from the original PDF on 19 February 2018 Retrieved 21 October 2017 12 Lead ECG Placement www emtresource com 27 April 2019 Archived from the original on 19 January 2022 Retrieved 24 May 2019 ECG Leads an overview ScienceDirect Topics www sciencedirect com Retrieved 28 October 2022 12 Lead ECG Placement www emtresource com 27 April 2014 Archived from the original on 19 January 2022 Retrieved 27 May 2019 EKG Interpretation Nurses Learning Network Retrieved 27 May 2019 Jowett N I Turner A M Cole A Jones P A 1 February 2005 Modified electrode placement must be recorded when performing 12 lead electrocardiograms Postgraduate Medical Journal 81 952 122 125 doi 10 1136 pgmj 2004 021204 ISSN 0032 5473 PMC 1743200 PMID 15701746 12 Lead ECG Placement Guide with Illustrations Cables and Sensors Retrieved 11 July 2017 Kavuru Madhav S Vesselle Hubert Thomas Cecil W 1987 Advances in Body Surface Potential Mapping BSPM Instrumentation In Pediatric and Fundamental Electrocardiography Developments in Cardiovascular Medicine Vol 56 pp 315 327 doi 10 1007 978 1 4613 2323 5 15 ISBN 978 1 4612 9428 3 ISSN 0166 9842 Tsukada Yayoi Tetsuou Tokita Miwa Murata Hiroshige Hirasawa Yasuhiro Yodogawa Kenji Iwasaki Yu ki Asai Kuniya Shimizu Wataru Kasai Nahoko Nakashima Hiroshi Tsukada Shingo 24 January 2019 Validation of wearable textile electrodes for ECG monitoring Heart and Vessels 34 7 1203 1211 doi 10 1007 s00380 019 01347 8 ISSN 0910 8327 PMC 6556171 PMID 30680493 Gargiulo GD 2015 True unipolar ECG machine for Wilson Central Terminal measurements BioMed Research International 2015 586397 doi 10 1155 2015 586397 PMC 460614 PMID 26495303 Sensors Cables and 12 Lead ECG Placement Guide with Illustrations Cables and Sensors Cables and Sensors Retrieved 21 October 2017 Limb Leads ECG Lead Placement Normal Function of the Heart Cardiology Teaching Package Practice Learning Division of Nursing The University of Nottingham Nottingham ac uk Retrieved 15 August 2009 Lesson 1 The Standard 12 Lead ECG Library med utah edu Archived from the original on 22 March 2009 Retrieved 15 August 2009 Jin Benjamin E Wulff Heike Widdicombe Jonathan H Zheng Jie Bers Donald M Puglisi Jose L December 2012 A simple device to illustrate the Einthoven triangle Advances in Physiology Education 36 4 319 324 Bibcode 2012BpJ 102 211J doi 10 1152 advan 00029 2012 ISSN 1043 4046 PMC 3776430 PMID 23209014 a b c Meek S 16 February 2002 ABC of clinical electrocardiography Introduction I Leads rate rhythm and cardiac axis BMJ 324 7334 415 418 doi 10 1136 bmj 324 7334 415 ISSN 0959 8138 PMC 1122339 PMID 11850377 Madias JE 2008 On recording the unipolar ECG limb leads via the Wilson s vs the Goldberger s terminals aVR aVL and aVF revisited Indian Pacing and Electrophysiology Journal 8 4 292 297 PMC 2572021 PMID 18982138 Mc Loughlin MJ 2020 Precordial bipolar leads A new method to study anterior acute myocardial infarction J Electrocardiol 59 2 45 64 doi 10 1016 j jelectrocard 2019 12 017 PMID 31986362 S2CID 210935474 Buttner Robert Cadogan Mike 29 January 2022 Lewis lead Life in the Fast Lane Retrieved 2 February 2022 Meigas K Kaik J Anier A 2008 Device and methods for performing transesophageal stimulation at reduced pacing current threshold Estonian Journal of Engineering 57 2 154 doi 10 3176 eng 2008 2 05 S2CID 42055085 a b Pehrson Steen M Blomstro Lundqvist Carina Ljungstro Erik Blomstro Per 1994 Clinical value of transesophageal atrial stimulation and recording in patients with arrhythmia related symptoms or documented supraventricular tachycardia correlation to clinical history and invasive studies Clinical Cardiology 17 10 528 534 doi 10 1002 clc 4960171004 PMID 8001299 Zhang Yongan Banta Anton Fu Yonggan John Mathews M Post Allison Razavi Mehdi Cavallaro Joseph Aazhang Behnaam Lin Yingyan 30 April 2022 RT RCG Neural Network and Accelerator Search Towards Effective and Real time ECG Reconstruction from Intracardiac Electrograms ACM Journal on Emerging Technologies in Computing Systems 18 2 29 doi 10 1145 3465372 ISSN 1550 4832 PMC 9236221 PMID 35765469 a b c Ashley Euan A Niebauer Josef 2004 Conquering the ECG Remedica Electrode Catheter an overview ScienceDirect Topics www sciencedirect com Retrieved 28 October 2022 Pennoyer James Bykhovsky Michael Sohinki Daniel Mallard Rachel Berman Adam 1 October 2020 Successful Catheter Ablation of Two Macro reentrant Atrial Tachycardias in a Patient with Congenitally Corrected Transposition of the Great Arteries A Case Report Journal of Innovations in Cardiac Rhythm Management 11 10 4273 4280 doi 10 19102 icrm 2020 111005 ISSN 2156 3977 PMC 7588239 PMID 33123416 Ewy G A 1984 Defining electromechanical dissociation Annals of Emergency Medicine 13 9 Pt 2 830 832 doi 10 1016 s0196 0644 84 80452 7 ISSN 0196 0644 PMID 6476549 Avram Robert Tison Geoffrey H Aschbacher Kirstin Kuhar Peter Vittinghoff Eric Butzner Michael Runge Ryan Wu Nancy Pletcher Mark J Marcus Gregory M Olgin Jeffrey 25 June 2019 Real world heart rate norms in the Health eHeart study npj Digital Medicine 2 1 58 doi 10 1038 s41746 019 0134 9 ISSN 2398 6352 PMC 6592896 PMID 31304404 Schrepel Caitlin Amick Ashley E Sayed Madeline Chipman Anne K 7 September 2021 Ischemic ECG Pattern Recognition to Facilitate Interpretation While Task Switching A Parallel Curriculum MedEdPORTAL 17 11182 doi 10 15766 mep 2374 8265 11182 ISSN 2374 8265 PMC 8421424 PMID 34557588 Becker Daniel E 2006 Fundamentals of Electrocardiography Interpretation Anesthesia Progress 53 2 53 64 doi 10 2344 0003 3006 2006 53 53 FOEI 2 0 CO 2 ISSN 0003 3006 PMC 1614214 PMID 16863387 Fleming Susannah Thompson Matthew Stevens Richard Heneghan Carl Pluddemann Annette Maconochie Ian Tarassenko Lionel Mant David 19 March 2011 Normal ranges of heart rate and respiratory rate in children from birth to 18 years of age a systematic review of observational studies Lancet 377 9770 1011 1018 doi 10 1016 S0140 6736 10 62226 X ISSN 1474 547X PMC 3789232 PMID 21411136 Bradycardia an overview ScienceDirect Topics www sciencedirect com Retrieved 28 October 2022 Sample records for qrs complex relationship Surawicz Borys Knillans Timothy 2008 Chou s electrocardiography in clinical practice adult and pediatric 6th ed Philadelphia PA Saunders Elsevier p 12 ISBN 978 1416037743 Kashou Anthony H Basit Hajira Chhabra Lovely 2022 Electrical Right and Left Axis Deviation StatPearls Treasure Island FL StatPearls Publishing PMID 29262101 retrieved 28 October 2022 Publishing M D K 28 April 2015 EKGS and ECGS Speedy Study Guides Speedy Publishing LLC ISBN 978 1 68185 011 5 ECG Study Guide Otero J Lenihan DJ 2000 The normothermic Osborn wave induced by severe hypercalcemia Tex Heart Inst J 27 3 316 317 PMC 101092 PMID 11093425 Houghton Andrew R Gray David 2012 Making Sense of the ECG Third Edition Hodder Education p 214 ISBN 978 1 4441 6654 5 Cardio online 12 December 2012 ECG EKG Paper Simple Cardiology Retrieved 20 October 2019 Volume Conductor Principles and ECG Rules of Interpretation CV Physiology Retrieved 22 October 2019 Sattar Yasar Chhabra Lovely 2022 Electrocardiogram StatPearls Treasure Island FL StatPearls Publishing PMID 31747210 retrieved 28 October 2022 Noble R Joe Hillis J Stanley Rothbaum Donald A 1990 Walker H Kenneth Hall W Dallas Hurst J Willis eds Electrocardiography Clinical Methods The History Physical and Laboratory Examinations 3rd ed Butterworths ISBN 9780409900774 PMID 21250195 retrieved 22 October 2019 Scher Allen M Young Allan C Malmgren Arthur L Erickson Robert V January 1955 Activation of the Interventricular Septum Circulation Research 3 1 56 64 doi 10 1161 01 RES 3 1 56 ISSN 0009 7330 PMID 13231277 Ventricular Depolarization and the Mean Electrical Axis CV Physiology Retrieved 22 October 2019 Kashou Anthony H Basit Hajira Malik Ahmad 2022 ST Segment StatPearls Treasure Island FL StatPearls Publishing PMID 29083566 retrieved 28 October 2022 Lukas Anton 29 June 2016 Electrophysiology of Myocardial Cells in the Epicardial Midmyocardial and Endocardial Layers of the Ventricle Journal of Cardiovascular Pharmacology and Therapeutics 2 1 61 72 doi 10 1177 107424849700200108 PMID 10684443 S2CID 44968291 Alpert JS Thygesen K Antman E Bassand JP 2000 Myocardial infarction redefined a consensus document of The Joint European Society of Cardiology American College of Cardiology Committee for the redefinition of myocardial infarction J Am Coll Cardiol 36 3 959 969 doi 10 1016 S0735 1097 00 00804 4 PMID 10987628 Warner Matthew J Tivakaran Vijai S 2022 Inferior Myocardial Infarction StatPearls Treasure Island FL StatPearls Publishing PMID 29262146 retrieved 28 October 2022 Segura Sampedro Juan Jose Parra Lopez Loreto Sampedro Abascal Consuelo Munoz Rodriguez Juan Carlos 2015 Atrial Flutter EKG can be useful with the proper Electrophysiological Basis International Journal of Cardiology 179 68 69 doi 10 1016 j ijcard 2014 10 076 PMID 25464416 a b Takla George Petre John H Doyle D John Horibe Mayumi Gopakumaran Bala 2006 The Problem of Artifacts in Patient Monitor Data During Surgery A Clinical and Methodological Review Anesthesia amp Analgesia 103 5 1196 1204 doi 10 1213 01 ane 0000247964 47706 5d PMID 17056954 S2CID 10614183 Kligfield Paul Gettes Leonard S Bailey James J Childers Rory Deal Barbara J Hancock E William van Herpen Gerard Kors Jan A Macfarlane Peter 13 March 2007 Recommendations for the standardization and interpretation of the electrocardiogram part I The electrocardiogram and its technology a scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee Council on Clinical Cardiology the American College of Cardiology Foundation and the Heart Rhythm Society endorsed by the International Society for Computerized Electrocardiology Circulation 115 10 1306 1324 doi 10 1161 CIRCULATIONAHA 106 180200 PMID 17322457 Minimizing ECG Artifact PDF Physio Control Physio Control Inc Redmond WA 2015 Retrieved 21 October 2017 Jafary Fahim H 2007 The incidental episode of ventricular fibrillation A case report Journal of Medical Case Reports 1 1 72 doi 10 1186 1752 1947 1 72 PMC 2000884 PMID 17760955 Mangalmurti Sandeep Seabury Seth A Chandra Amitabh Lakdawalla Darius Oetgen William J Jena Anupam B 2014 Medical professional liability risk among US cardiologists American Heart Journal 167 5 690 696 doi 10 1016 j ahj 2014 02 007 PMC 4153384 PMID 24766979 Batchvarov Velislav N Malik Marek Camm A John November 2007 Incorrect electrode cable connection during electrocardiographic recording Europace 9 11 1081 1090 doi 10 1093 europace eum198 ISSN 1532 2092 PMID 17932025 Chanarin N Caplin J Peacock A 1990 Pseudo reinfarction a consequence of electrocardiogram lead transposition following myocardial infarction Clinical Cardiology 13 9 668 669 doi 10 1002 clc 4960130916 PMID 2208827 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint multiple names authors list link Guijarro Morales A Gil Extremera B Maldonado Martin A 1991 ECG diagnostic errors due to improper connection of the right arm and leg cables International Journal of Cardiology 30 2 233 235 doi 10 1016 0167 5273 91 90103 v PMID 2010249 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint multiple names authors list link Whitbread Mark January 2006 Reading a normal ECG British Journal of Cardiac Nursing 1 1 32 33 doi 10 12968 bjca 2006 1 1 20382 ISSN 1749 6403 Mallinson Tom 2 March 2023 Additional rules for reading an electrocardiogram Journal of Paramedic Practice 15 3 95 97 doi 10 12968 jpar 2023 15 3 95 ISSN 1759 1376 S2CID 257390198 Montague Brian T Ouellette Jason R Buller Gregory K 30 January 2008 Retrospective Review of the Frequency of ECG Changes in Hyperkalemia Clinical Journal of the American Society of Nephrology 3 2 324 330 doi 10 2215 CJN 04611007 ISSN 1555 9041 PMC 2390954 PMID 18235147 Arrhythmia nhs uk 19 February 2018 Retrieved 28 October 2022 de Winter Robert 6 November 2008 A New ECG Sign of Proximal LAD Occlusion NEJM 359 19 2071 2073 doi 10 1056 NEJMc0804737 PMID 18987380 S2CID 205040240 Birse Ronald M 23 September 2004 Muirhead Alexander 1848 1920 electrical engineer In Knowlden Patricia E ed Oxford Dictionary of National Biography online ed Oxford University Press doi 10 1093 ref odnb 37794 Retrieved 20 January 2020 Subscription or UK public library membership required Rogers Mark C 1969 Historical Annotation Sir John Scott Burdon Sanderson 1828 1905 A Pioneer in Electrophysiology Circulation 40 1 1 2 doi 10 1161 01 CIR 40 1 1 ISSN 0009 7322 PMID 4893441 Waller AD 1887 A demonstration on man of electromotive changes accompanying the heart s beat J Physiol 8 5 229 34 doi 10 1113 jphysiol 1887 sp000257 PMC 1485094 PMID 16991463 a b c Hurst JW 3 November 1998 Naming of the Waves in the ECG With a Brief Account of Their Genesis Circulation 98 18 1937 42 doi 10 1161 01 CIR 98 18 1937 PMID 9799216 Interwoven W 1901 Un nouveau galvanometre Arch Neerl Sc Ex Nat 6 625 Rivera Ruiz M Cajavilca C Varon J 29 September 1927 Einthoven s String Galvanometer The First Electrocardiograph Texas Heart Institute Journal 35 2 174 78 PMC 2435435 PMID 18612490 Cooper JK 1986 Electrocardiography 100 years ago Origins pioneers and contributors N Engl J Med 315 7 461 64 doi 10 1056 NEJM198608143150721 PMID 3526152 Blackford John M MD 1 May 1927 Electrocardiography A Short Talk Before the Staff of the Hospital Clinics of the Virginia Mason Hospital 6 1 28 34 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint multiple names authors list link Dr Taro Takemi Takemi Program in International Health 27 August 2012 Retrieved 21 October 2017 A not so brief history of electrocardiography 2009 A not so brief history of electrocardiography ECG Library 3 January 2006 Archived from the original on 2 February 2012 Retrieved 11 January 2021 The Interesting History of EKGs info nhanow com Retrieved 21 January 2024 External links edit nbsp Wikimedia Commons has media related to ECG The whole ECG course on 1 A4 paper from ECGpedia a wiki encyclopedia for a course on interpretation of ECG Wave Maven a large database of practice ECG questions provided by Beth Israel Deaconess Medical Center PysioBank a free scientific database with physiologic signals here ecg EKG Academy free EKG lectures drills and quizzes ECG Learning Center created by Eccles Health Sciences Library at University of Utah Retrieved from https en wikipedia org w index php title Electrocardiography amp oldid 1220065958, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.