fbpx
Wikipedia

Detection limit

The limit of detection (LOD or LoD) is the lowest signal, or the lowest corresponding quantity to be determined (or extracted) from the signal, that can be observed with a sufficient degree of confidence or statistical significance. However, the exact threshold (level of decision) used to decide when a signal significantly emerges above the continuously fluctuating background noise remains arbitrary and is a matter of policy and often of debate among scientists, statisticians and regulators depending on the stakes in different fields.

Significance in analytical chemistry edit

In analytical chemistry, the detection limit, lower limit of detection, also termed LOD for limit of detection or analytical sensitivity (not to be confused with statistical sensitivity), is the lowest quantity of a substance that can be distinguished from the absence of that substance (a blank value) with a stated confidence level (generally 99%).[1][2][3] The detection limit is estimated from the mean of the blank, the standard deviation of the blank, the slope (analytical sensitivity) of the calibration plot and a defined confidence factor (e.g. 3.2 being the most accepted value for this arbitrary value).[4] Another consideration that affects the detection limit is the adequacy and the accuracy of the model used to predict concentration from the raw analytical signal.[5]

As a typical example, from a calibration plot following a linear equation taken here as the simplest possible model:

 

where,   corresponds to the signal measured (e.g. voltage, luminescence, energy, etc.), "b" the value in which the straight line cuts the ordinates axis, "a" the sensitivity of the system (i.e., the slope of the line, or the function relating the measured signal to the quantity to be determined) and "x" the value of the quantity (e.g. temperature, concentration, pH, etc.) to be determined from the signal  ,[6] the LOD for "x" is calculated as the "x" value in which   equals to the average value of blanks "y" plus "t" times its standard deviation "s" (or, if zero, the standard deviation corresponding to the lowest value measured) where "t" is the chosen confidence value (e.g. for a confidence of 95% it can be considered t = 3.2, determined from the limit of blank).[4]

Thus, in this didactic example:

 

There are a number of concepts derived from the detection limit that are commonly used. These include the instrument detection limit (IDL), the method detection limit (MDL), the practical quantitation limit (PQL), and the limit of quantitation (LOQ). Even when the same terminology is used, there can be differences in the LOD according to nuances of what definition is used and what type of noise contributes to the measurement and calibration.[7]

The figure below illustrates the relationship between the blank, the limit of detection (LOD), and the limit of quantitation (LOQ) by showing the probability density function for normally distributed measurements at the blank, at the LOD defined as 3 × standard deviation of the blank, and at the LOQ defined as 10 × standard deviation of the blank. (The identical spread along Abscissa of these two functions is problematic.) For a signal at the LOD, the alpha error (probability of false positive) is small (1%). However, the beta error (probability of a false negative) is 50% for a sample that has a concentration at the LOD (red line). This means a sample could contain an impurity at the LOD, but there is a 50% chance that a measurement would give a result less than the LOD. At the LOQ (blue line), there is minimal chance of a false negative.

 
Illustration of the concept of detection limit and quantitation limit by showing the theoretical normal distributions associated with blank, detection limit (LOD), and quantitation limit (LOQ) level samples.

Instrument detection limit edit

Most analytical instruments produce a signal even when a blank (matrix without analyte) is analyzed. This signal is referred to as the noise level. The instrument detection limit (IDL) is the analyte concentration that is required to produce a signal greater than three times the standard deviation of the noise level. This may be practically measured by analyzing 8 or more standards at the estimated IDL then calculating the standard deviation from the measured concentrations of those standards.

The detection limit (according to IUPAC) is the smallest concentration, or the smallest absolute amount, of analyte that has a signal statistically significantly larger than the signal arising from the repeated measurements of a reagent blank.

Mathematically, the analyte's signal at the detection limit ( ) is given by:

 

where,   is the mean value of the signal for a reagent blank measured multiple times, and   is the known standard deviation for the reagent blank's signal.

Other approaches for defining the detection limit have also been developed. In atomic absorption spectrometry usually the detection limit is determined for a certain element by analyzing a diluted solution of this element and recording the corresponding absorbance at a given wavelength. The measurement is repeated 10 times. The 3σ of the recorded absorbance signal can be considered as the detection limit for the specific element under the experimental conditions: selected wavelength, type of flame or graphite oven, chemical matrix, presence of interfering substances, instrument... .

Method detection limit edit

Often there is more to the analytical method than just performing a reaction or submitting the analyte to direct analysis. Many analytical methods developed in the laboratory, especially these involving the use of a delicate scientific instrument, require a sample preparation, or a pretreatment of the samples prior to being analysed. For example, it might be necessary to heat a sample that is to be analyzed for a particular metal with the addition of acid first (digestion process). The sample may also be diluted or concentrated prior to analysis by means of a given instrument. Additional steps in an analysis method add additional opportunities for errors. Since detection limits are defined in terms of errors, this will naturally increase the measured detection limit. This "global" detection limit (including all the steps of the analysis method) is called the method detection limit (MDL). The practical way for determining the MDL is to analyze seven samples of concentration near the expected limit of detection. The standard deviation is then determined. The one-sided Student's t-distribution is determined and multiplied versus the determined standard deviation. For seven samples (with six degrees of freedom) the t value for a 99% confidence level is 3.14. Rather than performing the complete analysis of seven identical samples, if the Instrument Detection Limit is known, the MDL may be estimated by multiplying the Instrument Detection Limit, or Lower Level of Detection, by the dilution prior to analyzing the sample solution with the instrument. This estimation, however, ignores any uncertainty that arises from performing the sample preparation and will therefore probably underestimate the true MDL.

Limit of each model edit

The issue of limit of detection, or limit of quantification, is encountered in all scientific disciplines. This explains the variety of definitions and the diversity of juridiction specific solutions developed to address preferences. In the simplest cases as in nuclear and chemical measurements, definitions and approaches have probably received the clearer and the simplest solutions. In biochemical tests and in biological experiments depending on many more intricate factors, the situation involving false positive and false negative responses is more delicate to handle. In many other disciplines such as geochemistry, seismology, astronomy, dendrochronology, climatology, life sciences in general, and in many other fields impossible to enumerate extensively, the problem is wider and deals with signal extraction out of a background of noise. It involves complex statistical analysis procedures and therefore it also depends on the models used,[5] the hypotheses and the simplifications or approximations to be made to handle and manage uncertainties. When the data resolution is poor and different signals overlap, different deconvolution procedures are applied to extract parameters. The use of different phenomenological, mathematical and statistical models may also complicate the exact mathematical definition of limit of detection and how it is calculated. This explains why it is not easy to come to a general consensus, if any, about the precise mathematical definition of the expression of limit of detection. However, one thing is clear: it always requires a sufficient number of data (or accumulated data) and a rigorous statistical analysis to render better signification statistically.

Limit of quantification edit

The limit of quantification (LoQ, or LOQ) is the lowest value of a signal (or concentration, activity, response...) that can be quantified with acceptable precision and accuracy.

The LoQ is the limit at which the difference between two distinct signals / values can be discerned with a reasonable certainty, i.e., when the signal is statistically different from the background. The LoQ may be drastically different between laboratories, so another detection limit is commonly used that is referred to as the Practical Quantification Limit (PQL).

See also edit

References edit

  1. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "detection limit". doi:10.1351/goldbook.L03540
  2. ^ MacDougall D, Crummett WB, et al. (1980). "Guidelines for Data Acquisition and Data Quality Evaluation in Environmental Chemistry". Analytical Chemistry. 52 (14): 2242–49. doi:10.1021/ac50064a004.
  3. ^ Saah AJ, Hoover DR (1998). "[Sensitivity and specificity revisited: significance of the terms in analytic and diagnostic language]". Ann Dermatol Venereol. 125 (4): 291–4. PMID 9747274.
  4. ^ a b Armbruster DA, Pry T (August 2008). "Limit of blank, limit of detection and limit of quantitation". The Clinical Biochemist. Reviews. 29 Suppl 1 (1): S49–S52. PMC 2556583. PMID 18852857.
  5. ^ a b "R: "Detection" limit for each model". search.r-project.org. Retrieved 2022-01-04.
  6. ^ Quesada-González D, Stefani C, González I, de la Escosura-Muñiz A, Domingo N, Mutjé P, Merkoçi A (September 2019). "Signal enhancement on gold nanoparticle-based lateral flow tests using cellulose nanofibers". Biosensors & Bioelectronics. 141: 111407. doi:10.1016/j.bios.2019.111407. hdl:10261/201014. PMID 31207571. S2CID 190531742.
  7. ^ Long, Gary L.; Winefordner, J. D. (1983), "Limit of detection: a closer look at the IUPAC definition", Anal. Chem., 55 (7): 712A–724A, doi:10.1021/ac00258a724

Further reading edit

  • Altshuler B, Pasternack B (1963-03-01). "Statistical measures of the lower limit of detection of a radioactivity counter". Health Physics. 9 (3): 293–298. doi:10.1097/00004032-196303000-00005. ISSN 0017-9078. PMID 14040764. Retrieved 2022-01-03.
  • Currie LA (1968). "Limits for qualitative detection and quantitative determination. Application to radiochemistry". Analytical Chemistry. 40 (3): 586–593. doi:10.1021/ac60259a007. ISSN 0003-2700.
  • Long GL, Winefordner JD (1983). "Limit of detection. A closer look at the IUPAC definition". Analytical Chemistry. 55 (7): 712A–724A. doi:10.1021/ac00258a001. ISSN 0003-2700.
  • Armbruster DA, Pry T (August 2008). "Limit of blank, limit of detection and limit of quantitation". The Clinical Biochemist. Reviews. 29 (Suppl 1): S49–S52. PMC 2556583. PMID 18852857.
  • European Commission. Joint Research Centre. (2016). Guidance document on the estimation of LOD and LOQ for measurements in the field of contaminants in feed and food. Luxembourg: Publications Office. doi:10.2787/8931. ISBN 9789279617683. Retrieved 2022-01-03.
  • "DIN 32645 – Chemical analysis – Decision limit, detection limit and determination limit under repeatability conditions – Terms, methods, evaluation. Technical standard. Deutsches Institut für Normung, Berlin (DIN 32645:2008-11) | Via Engineering360" (in German). Published by Beuth Verlag, a subsidiary of the DIN Group. doi:10.31030/1465413. Retrieved 2022-01-03. {{cite journal}}: Cite journal requires |journal= (help)

External links edit

  • Evans WC (21 February 2019). "Limit of Detection – Interactive Java applet to illustrate some basic ideas of the limit of detection problem". GeoGebra. Retrieved 2022-01-04.
  • "The R Language". search.r-project.org. Retrieved 2022-01-04.
  • Garrett RG (2013-11-01). "The 'rgr' package for the R Open Source statistical computing and graphics environment – a tool to support geochemical data interpretation". Geochemistry: Exploration, Environment, Analysis. 13 (4): 355–378. Bibcode:2013GEEA...13..355G. doi:10.1144/geochem2011-106. ISSN 1467-7873. S2CID 129059022. Retrieved 2022-01-04.
  • "R: "Detection" limit for each model". search.r-project.org. Retrieved 2022-01-04.
  • Deutsches Institut für Normung. "R: Calibration data from DIN 32645 (Package envalysis version 0.5.1)". search.r-project.org. Retrieved 2022-01-04.
  • Downloads of articles (a.o. harmonization of concepts by ISO and IUPAC) and an extensive list of references 2009-04-21 at the Wayback Machine

detection, limit, redirects, here, company, listed, london, stock, exchange, airport, botswana, with, iata, code, lobatse, airport, limit, detection, lowest, signal, lowest, corresponding, quantity, determined, extracted, from, signal, that, observed, with, su. LOQ redirects here For the company listed as LOQ on the London Stock Exchange see Lo Q For the airport in Botswana with IATA code LOQ see Lobatse Airport The limit of detection LOD or LoD is the lowest signal or the lowest corresponding quantity to be determined or extracted from the signal that can be observed with a sufficient degree of confidence or statistical significance However the exact threshold level of decision used to decide when a signal significantly emerges above the continuously fluctuating background noise remains arbitrary and is a matter of policy and often of debate among scientists statisticians and regulators depending on the stakes in different fields Contents 1 Significance in analytical chemistry 2 Instrument detection limit 3 Method detection limit 4 Limit of each model 5 Limit of quantification 6 See also 7 References 8 Further reading 9 External linksSignificance in analytical chemistry editIn analytical chemistry the detection limit lower limit of detection also termed LOD for limit of detection or analytical sensitivity not to be confused with statistical sensitivity is the lowest quantity of a substance that can be distinguished from the absence of that substance a blank value with a stated confidence level generally 99 1 2 3 The detection limit is estimated from the mean of the blank the standard deviation of the blank the slope analytical sensitivity of the calibration plot and a defined confidence factor e g 3 2 being the most accepted value for this arbitrary value 4 Another consideration that affects the detection limit is the adequacy and the accuracy of the model used to predict concentration from the raw analytical signal 5 As a typical example from a calibration plot following a linear equation taken here as the simplest possible model f x a x b displaystyle f x ax b nbsp where f x displaystyle f x nbsp corresponds to the signal measured e g voltage luminescence energy etc b the value in which the straight line cuts the ordinates axis a the sensitivity of the system i e the slope of the line or the function relating the measured signal to the quantity to be determined and x the value of the quantity e g temperature concentration pH etc to be determined from the signal f x displaystyle f x nbsp 6 the LOD for x is calculated as the x value in which f x displaystyle f x nbsp equals to the average value of blanks y plus t times its standard deviation s or if zero the standard deviation corresponding to the lowest value measured where t is the chosen confidence value e g for a confidence of 95 it can be considered t 3 2 determined from the limit of blank 4 Thus in this didactic example LOD for x f x b a y 3 2 s b a displaystyle text LOD for x frac left f x b right a frac left y 3 2s b right a nbsp There are a number of concepts derived from the detection limit that are commonly used These include the instrument detection limit IDL the method detection limit MDL the practical quantitation limit PQL and the limit of quantitation LOQ Even when the same terminology is used there can be differences in the LOD according to nuances of what definition is used and what type of noise contributes to the measurement and calibration 7 The figure below illustrates the relationship between the blank the limit of detection LOD and the limit of quantitation LOQ by showing the probability density function for normally distributed measurements at the blank at the LOD defined as 3 standard deviation of the blank and at the LOQ defined as 10 standard deviation of the blank The identical spread along Abscissa of these two functions is problematic For a signal at the LOD the alpha error probability of false positive is small 1 However the beta error probability of a false negative is 50 for a sample that has a concentration at the LOD red line This means a sample could contain an impurity at the LOD but there is a 50 chance that a measurement would give a result less than the LOD At the LOQ blue line there is minimal chance of a false negative nbsp Illustration of the concept of detection limit and quantitation limit by showing the theoretical normal distributions associated with blank detection limit LOD and quantitation limit LOQ level samples Instrument detection limit editMost analytical instruments produce a signal even when a blank matrix without analyte is analyzed This signal is referred to as the noise level The instrument detection limit IDL is the analyte concentration that is required to produce a signal greater than three times the standard deviation of the noise level This may be practically measured by analyzing 8 or more standards at the estimated IDL then calculating the standard deviation from the measured concentrations of those standards The detection limit according to IUPAC is the smallest concentration or the smallest absolute amount of analyte that has a signal statistically significantly larger than the signal arising from the repeated measurements of a reagent blank Mathematically the analyte s signal at the detection limit S d l displaystyle S dl nbsp is given by S d l S r e a g 3 s r e a g displaystyle S dl S reag 3 sigma reag nbsp where S r e a g displaystyle S reag nbsp is the mean value of the signal for a reagent blank measured multiple times and s r e a g displaystyle sigma reag nbsp is the known standard deviation for the reagent blank s signal Other approaches for defining the detection limit have also been developed In atomic absorption spectrometry usually the detection limit is determined for a certain element by analyzing a diluted solution of this element and recording the corresponding absorbance at a given wavelength The measurement is repeated 10 times The 3s of the recorded absorbance signal can be considered as the detection limit for the specific element under the experimental conditions selected wavelength type of flame or graphite oven chemical matrix presence of interfering substances instrument Method detection limit editOften there is more to the analytical method than just performing a reaction or submitting the analyte to direct analysis Many analytical methods developed in the laboratory especially these involving the use of a delicate scientific instrument require a sample preparation or a pretreatment of the samples prior to being analysed For example it might be necessary to heat a sample that is to be analyzed for a particular metal with the addition of acid first digestion process The sample may also be diluted or concentrated prior to analysis by means of a given instrument Additional steps in an analysis method add additional opportunities for errors Since detection limits are defined in terms of errors this will naturally increase the measured detection limit This global detection limit including all the steps of the analysis method is called the method detection limit MDL The practical way for determining the MDL is to analyze seven samples of concentration near the expected limit of detection The standard deviation is then determined The one sided Student s t distribution is determined and multiplied versus the determined standard deviation For seven samples with six degrees of freedom the t value for a 99 confidence level is 3 14 Rather than performing the complete analysis of seven identical samples if the Instrument Detection Limit is known the MDL may be estimated by multiplying the Instrument Detection Limit or Lower Level of Detection by the dilution prior to analyzing the sample solution with the instrument This estimation however ignores any uncertainty that arises from performing the sample preparation and will therefore probably underestimate the true MDL Limit of each model editThe issue of limit of detection or limit of quantification is encountered in all scientific disciplines This explains the variety of definitions and the diversity of juridiction specific solutions developed to address preferences In the simplest cases as in nuclear and chemical measurements definitions and approaches have probably received the clearer and the simplest solutions In biochemical tests and in biological experiments depending on many more intricate factors the situation involving false positive and false negative responses is more delicate to handle In many other disciplines such as geochemistry seismology astronomy dendrochronology climatology life sciences in general and in many other fields impossible to enumerate extensively the problem is wider and deals with signal extraction out of a background of noise It involves complex statistical analysis procedures and therefore it also depends on the models used 5 the hypotheses and the simplifications or approximations to be made to handle and manage uncertainties When the data resolution is poor and different signals overlap different deconvolution procedures are applied to extract parameters The use of different phenomenological mathematical and statistical models may also complicate the exact mathematical definition of limit of detection and how it is calculated This explains why it is not easy to come to a general consensus if any about the precise mathematical definition of the expression of limit of detection However one thing is clear it always requires a sufficient number of data or accumulated data and a rigorous statistical analysis to render better signification statistically Limit of quantification editThe limit of quantification LoQ or LOQ is the lowest value of a signal or concentration activity response that can be quantified with acceptable precision and accuracy The LoQ is the limit at which the difference between two distinct signals values can be discerned with a reasonable certainty i e when the signal is statistically different from the background The LoQ may be drastically different between laboratories so another detection limit is commonly used that is referred to as the Practical Quantification Limit PQL See also editBackground noise Sound other than the sound being monitored primary sound Background radiation Measure of ionizing radiation in the environment Electronic noise Random fluctuation in an electrical signal Noise spectral phenomenon Types of noise Chemometrics Science of extracting information from chemical systems by data driven means Gamma spectroscopy Calibration and background radiation Quantitative study of the energy spectra of gamma ray sources Malmquist bias Sampling bias in astronomy p value Function of the observed sample results Misuse of p values Misinterpretation of statistical significance Statistical significance Concept in inferential statisticsReferences edit IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 detection limit doi 10 1351 goldbook L03540 MacDougall D Crummett WB et al 1980 Guidelines for Data Acquisition and Data Quality Evaluation in Environmental Chemistry Analytical Chemistry 52 14 2242 49 doi 10 1021 ac50064a004 Saah AJ Hoover DR 1998 Sensitivity and specificity revisited significance of the terms in analytic and diagnostic language Ann Dermatol Venereol 125 4 291 4 PMID 9747274 a b Armbruster DA Pry T August 2008 Limit of blank limit of detection and limit of quantitation The Clinical Biochemist Reviews 29 Suppl 1 1 S49 S52 PMC 2556583 PMID 18852857 a b R Detection limit for each model search r project org Retrieved 2022 01 04 Quesada Gonzalez D Stefani C Gonzalez I de la Escosura Muniz A Domingo N Mutje P Merkoci A September 2019 Signal enhancement on gold nanoparticle based lateral flow tests using cellulose nanofibers Biosensors amp Bioelectronics 141 111407 doi 10 1016 j bios 2019 111407 hdl 10261 201014 PMID 31207571 S2CID 190531742 Long Gary L Winefordner J D 1983 Limit of detection a closer look at the IUPAC definition Anal Chem 55 7 712A 724A doi 10 1021 ac00258a724Further reading editAltshuler B Pasternack B 1963 03 01 Statistical measures of the lower limit of detection of a radioactivity counter Health Physics 9 3 293 298 doi 10 1097 00004032 196303000 00005 ISSN 0017 9078 PMID 14040764 Retrieved 2022 01 03 Currie LA 1968 Limits for qualitative detection and quantitative determination Application to radiochemistry Analytical Chemistry 40 3 586 593 doi 10 1021 ac60259a007 ISSN 0003 2700 Long GL Winefordner JD 1983 Limit of detection A closer look at the IUPAC definition Analytical Chemistry 55 7 712A 724A doi 10 1021 ac00258a001 ISSN 0003 2700 Armbruster DA Pry T August 2008 Limit of blank limit of detection and limit of quantitation The Clinical Biochemist Reviews 29 Suppl 1 S49 S52 PMC 2556583 PMID 18852857 European Commission Joint Research Centre 2016 Guidance document on the estimation of LOD and LOQ for measurements in the field of contaminants in feed and food Luxembourg Publications Office doi 10 2787 8931 ISBN 9789279617683 Retrieved 2022 01 03 DIN 32645 Chemical analysis Decision limit detection limit and determination limit under repeatability conditions Terms methods evaluation Technical standard Deutsches Institut fur Normung Berlin DIN 32645 2008 11 Via Engineering360 in German Published by Beuth Verlag a subsidiary of the DIN Group doi 10 31030 1465413 Retrieved 2022 01 03 a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help External links editEvans WC 21 February 2019 Limit of Detection Interactive Java applet to illustrate some basic ideas of the limit of detection problem GeoGebra Retrieved 2022 01 04 The R Language search r project org Retrieved 2022 01 04 Garrett RG 2013 11 01 The rgr package for the R Open Source statistical computing and graphics environment a tool to support geochemical data interpretation Geochemistry Exploration Environment Analysis 13 4 355 378 Bibcode 2013GEEA 13 355G doi 10 1144 geochem2011 106 ISSN 1467 7873 S2CID 129059022 Retrieved 2022 01 04 R Detection limit for each model search r project org Retrieved 2022 01 04 Deutsches Institut fur Normung R Calibration data from DIN 32645 Package envalysis version 0 5 1 search r project org Retrieved 2022 01 04 Downloads of articles a o harmonization of concepts by ISO and IUPAC and an extensive list of references Archived 2009 04 21 at the Wayback Machine Retrieved from https en wikipedia org w index php title Detection limit amp oldid 1222372247, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.