Magnetoencephalography (MEG) is a functional neuroimaging technique for mapping brain activity by recording magnetic fields produced by electrical currents occurring naturally in the brain, using very sensitive magnetometers. Arrays of SQUIDs (superconducting quantum interference devices) are currently the most common magnetometer, while the SERF (spin exchange relaxation-free) magnetometer is being investigated for future machines.[1][2] Applications of MEG include basic research into perceptual and cognitive brain processes, localizing regions affected by pathology before surgical removal, determining the function of various parts of the brain, and neurofeedback. This can be applied in a clinical setting to find locations of abnormalities as well as in an experimental setting to simply measure brain activity.[3]

Magnetoencephalography
Person undergoing a MEG
MeSHD015225

History edit

 
Dr. Cohen's shielded room at MIT, in which first MEG was measured with a SQUID
 
First MEG measured with SQUID, in Dr. Cohen's room at MIT

MEG signals were first measured by University of Illinois physicist David Cohen in 1968,[4] before the availability of the SQUID, using a copper induction coil as the detector. To reduce the magnetic background noise, the measurements were made in a magnetically shielded room. The coil detector was barely sensitive enough, resulting in poor, noisy MEG measurements that were difficult to use. Later, Cohen built a much better shielded room at MIT, and used one of the first SQUID detectors, just developed by James E. Zimmerman, a researcher at Ford Motor Company,[5] to again measure MEG signals.[6] This time the signals were almost as clear as those of EEG. This stimulated the interest of physicists who had been looking for uses of SQUIDs. Subsequent to this, various types of spontaneous and evoked MEGs began to be measured.

At first, a single SQUID detector was used to successively measure the magnetic field at a number of points around the subject's head. This was cumbersome, and, in the 1980s, MEG manufacturers began to arrange multiple sensors into arrays to cover a larger area of the head.[7] Present-day MEG arrays are set in a helmet-shaped vacuum flask that typically contain 300 sensors, covering most of the head. In this way, MEGs of a subject or patient can now be accumulated rapidly and efficiently.

Recent developments attempt to increase portability of MEG scanners by using spin exchange relaxation-free (SERF) magnetometers. SERF magnetometers are relatively small, as they do not require bulky cooling systems to operate. At the same time, they feature sensitivity equivalent to that of SQUIDs. In 2012, it was demonstrated that MEG could work with a chip-scale atomic magnetometer (CSAM, type of SERF).[8] More recently, in 2017, researchers built a working prototype that uses SERF magnetometers installed into portable individually 3D-printed helmets,[2] which they noted in interviews could be replaced with something easier to use in future, such as a bike helmet.

The basis of the MEG signal edit

Synchronized neuronal currents induce weak magnetic fields. The brain's magnetic field, measuring at 10 femtotesla (fT) for cortical activity and 103 fT for the human alpha rhythm, is considerably smaller than the ambient magnetic noise in an urban environment, which is on the order of 108 fT or 0.1 μT. The essential problem of biomagnetism is, thus, the weakness of the signal relative to the sensitivity of the detectors, and to the competing environmental noise.

 
Origin of the brain's magnetic field. The electric current also produces the EEG signal.

The MEG (and EEG) signals derive from the net effect of ionic currents flowing in the dendrites of neurons during synaptic transmission. In accordance with Maxwell's equations, any electrical current will produce a magnetic field, and it is this field that is measured. The net currents can be thought of as current dipoles,[9] i.e. currents with a position, orientation, and magnitude, but no spatial extent.[dubious ] According to the right-hand rule, a current dipole gives rise to a magnetic field that points around the axis of its vector component.

To generate a signal that is detectable, approximately 50,000 active neurons are needed.[10] Since current dipoles must have similar orientations to generate magnetic fields that reinforce each other, it is often the layer of pyramidal cells, which are situated perpendicular to the cortical surface, that gives rise to measurable magnetic fields. Bundles of these neurons that are orientated tangentially to the scalp surface project measurable portions of their magnetic fields outside of the head, and these bundles are typically located in the sulci. Researchers are experimenting with various signal processing methods in the search for methods that detect deep brain (i.e., non-cortical) signal, but no clinically useful method is currently available.

It is worth noting that action potentials do not usually produce an observable field, mainly because the currents associated with action potentials flow in opposite directions and the magnetic fields cancel out. However, action fields have been measured from peripheral nerve system.

Magnetic shielding edit

Since the magnetic signals emitted by the brain are on the order of a few femtoteslas, shielding from external magnetic signals, including the Earth's magnetic field, is necessary. Appropriate magnetic shielding can be obtained by constructing rooms made of aluminium and mu-metal for reducing high-frequency and low-frequency noise, respectively.

 
Entrance to MSR, showing the separate shielding layers

Magnetically shielded room (MSR) edit

A magnetically shielded room (MSR) model consists of three nested main layers. Each of these layers is made of a pure aluminium layer plus a high-permeability ferromagnetic layer, similar in composition to molybdenum permalloy. The ferromagnetic layer is supplied as 1 mm sheets, while the innermost layer is composed of four sheets in close contact, and the outer two layers are composed of three sheets each. Magnetic continuity is maintained by overlay strips. Insulating washers are used in the screw assemblies to ensure that each main layer is electrically isolated. This helps eliminate radio frequency radiation, which would degrade SQUID performance. Electrical continuity of the aluminium is also maintained by aluminium overlay strips to ensure AC eddy current shielding, which is important at frequencies greater than 1 Hz. The junctions of the inner layer are often electroplated with silver or gold to improve conductivity of the aluminium layers.[11]

Active shielding system edit

Active systems are designed for three-dimensional noise cancellation. To implement an active system, low-noise fluxgate magnetometers are mounted at the center of each surface and oriented orthogonally to it. This negatively feeds a DC amplifier through a low-pass network with a slow falloff to minimize positive feedback and oscillation. Built into the system are shaking and degaussing wires. Shaking wires increase the magnetic permeability, while the permanent degaussing wires are applied to all surfaces of the inner main layer to degauss the surfaces.[4] Moreover, noise cancellation algorithms can reduce both low-frequency and high-frequency noise. Modern systems have a noise floor of around 2–3 fT/Hz0.5 above 1 Hz.

Source localization edit

The inverse problem edit

The challenge posed by MEG is to determine the location of electric activity within the brain from the induced magnetic fields outside the head. Problems such as this, where model parameters (the location of the activity) have to be estimated from measured data (the SQUID signals) are referred to as inverse problems (in contrast to forward problems[12] where the model parameters (e.g. source location) are known and the data (e.g. the field at a given distance) is to be estimated.) The primary difficulty is that the inverse problem does not have a unique solution (i.e., there are infinite possible "correct" answers), and the problem of defining the "best" solution is itself the subject of intensive research.[13] Possible solutions can be derived using models involving prior knowledge of brain activity.

The source models can be either over-determined or under-determined. An over-determined model may consist of a few point-like sources ("equivalent dipoles"), whose locations are then estimated from the data. Under-determined models may be used in cases where many different distributed areas are activated ("distributed source solutions"): there are infinitely many possible current distributions explaining the measurement results, but the most likely is selected. Localization algorithms make use of given source and head models to find a likely location for an underlying focal field generator.

One type of localization algorithm for overdetermined models operates by expectation-maximization: the system is initialized with a first guess. A loop is started, in which a forward model is used to simulate the magnetic field that would result from the current guess. The guess is adjusted to reduce the discrepancy between the simulated field and the measured field. This process is iterated until convergence.

Another common technique is beamforming, wherein a theoretical model of the magnetic field produced by a given current dipole is used as a prior, along with second-order statistics of the data in the form of a covariance matrix, to calculate a linear weighting of the sensor array (the beamformer) via the Backus-Gilbert inverse. This is also known as a linearly constrained minimum variance (LCMV) beamformer. When the beamformer is applied to the data, it produces an estimate of the power in a "virtual channel" at the source location.

The extent to which the constraint-free MEG inverse problem is ill-posed cannot be overemphasized. If one's goal is to estimate the current density within the human brain with say a 5mm resolution then it is well established that the vast majority of the information needed to perform a unique inversion must come not from the magnetic field measurement but rather from the constraints applied to the problem. Furthermore, even when a unique inversion is possible in the presence of such constraints said inversion can be unstable. These conclusions are easily deduced from published works.[14]

Magnetic source imaging edit

The source locations can be combined with magnetic resonance imaging (MRI) images to create magnetic source images (MSI). The two sets of data are combined by measuring the location of a common set of fiducial points marked during MRI with lipid markers and marked during MEG with electrified coils of wire that give off magnetic fields. The locations of the fiducial points in each data set are then used to define a common coordinate system so that superimposing the functional MEG data onto the structural MRI data ("coregistration") is possible.

A criticism of the use of this technique in clinical practice is that it produces colored areas with definite boundaries superimposed upon an MRI scan: the untrained viewer may not realize that the colors do not represent a physiological certainty, not because of the relatively low spatial resolution of MEG, but rather some inherent uncertainty in the probability cloud derived from statistical processes. However, when the magnetic source image corroborates other data, it can be of clinical utility.

Dipole model source localization edit

A widely accepted source-modeling technique for MEG involves calculating a set of equivalent current dipoles (ECDs), which assumes the underlying neuronal sources to be focal. This dipole fitting procedure is non-linear and over-determined, since the number of unknown dipole parameters is smaller than the number of MEG measurements.[15] Automated multiple dipole model algorithms such as multiple signal classification (MUSIC) and multi-start spatial and temporal modeling (MSST) are applied to the analysis of MEG responses. The limitations of dipole models for characterizing neuronal responses are (1) difficulties in localizing extended sources with ECDs, (2) problems with accurately estimating the total number of dipoles in advance, and (3) dependency on dipole location, especially depth in the brain.

Distributed source models edit

Unlike multiple-dipole modeling, distributed source models divide the source space into a grid containing a large number of dipoles. The inverse problem is to obtain the dipole moments for the grid nodes.[16] As the number of unknown dipole moments is much greater than the number of MEG sensors, the inverse solution is highly underdetermined, so additional constraints are needed to reduce ambiguity of the solution. The primary advantage of this approach is that no prior specification of the source model is necessary. However, the resulting distributions may be difficult to interpret, because they only reflect a "blurred" (or even distorted) image of the true neuronal source distribution. The matter is complicated by the fact that spatial resolution depends strongly on various parameters such as brain area, depth, orientation, number of sensors etc.[17]

Independent component analysis (ICA) edit

Independent component analysis (ICA) is another signal processing solution that separates different signals that are statistically independent in time. It is primarily used to remove artifacts such as blinking, eye muscle movement, facial muscle artifacts, cardiac artifacts, etc. from MEG and EEG signals that may be contaminated with outside noise.[18] However, ICA has poor resolution of highly correlated brain sources.

Use in the field edit

 
Over 100 MEG systems are known to operate worldwide, with Japan possessing the greatest number of MEG systems per capita and the United States possessing the greatest overall number of MEG systems. A very small number of systems worldwide are designed for infant and/or fetal recordings.

In research, MEG's primary use is the measurement of time courses of activity. MEG can resolve events with a precision of 10 milliseconds or faster, while functional magnetic resonance imaging (fMRI), which depends on changes in blood flow, can at best resolve events with a precision of several hundred milliseconds. MEG also accurately pinpoints sources in primary auditory, somatosensory, and motor areas. For creating functional maps of human cortex during more complex cognitive tasks, MEG is most often combined with fMRI, as the methods complement each other. Neuronal (MEG) and hemodynamic fMRI data do not necessarily agree, in spite of the tight relationship between local field potentials (LFP) and blood oxygenation level-dependent (BOLD) signals. MEG and BOLD signals may originate from the same source (though the BOLD signals are filtered through the hemodynamic response).

MEG is also being used to better localize responses in the brain. The openness of the MEG setup allows external auditory and visual stimuli to be easily introduced. Some movement by the subject is also possible as long as it does not jar the subject's head. The responses in the brain before, during, and after the introduction of such stimuli/movement can then be mapped with greater spatial resolution than was previously possible with EEG.[19] Psychologists are also taking advantage of MEG neuroimaging to better understand relationships between brain function and behavior. For example, a number of studies have been done comparing the MEG responses of patients with psychological troubles to control patients. There has been great success isolating unique responses in patients with schizophrenia, such as auditory gating deficits to human voices.[20] MEG is also being used to correlate standard psychological responses, such as the emotional dependence of language comprehension.[21]

Recent studies have reported successful classification of patients with multiple sclerosis, Alzheimer's disease, schizophrenia, Sjögren's syndrome, chronic alcoholism, facial pain and thalamocortical dysrhythmias. MEG can be used to distinguish these patients from healthy control subjects, suggesting a future role of MEG in diagnostics.[22][23]

A large part of the difficulty and cost of using MEG is the need for manual analysis of the data. Progress has been made in analysis by computer, comparing a patient's scans with those drawn from a large database of normal scans, with the potential to reduce cost greatly.[24]

Brain connectivity and neural oscillations edit

Based on its perfect temporal resolution, magnetoencephalography (MEG) is now heavily used to study oscillatory activity in the brain, both in terms of local neural synchrony and cross-area synchronisation. As an example for local neural synchrony, MEG has been used to investigate alpha rhythms in various targeted brain regions, such as in visual[25][26] or auditory cortex.[27] Other studies have used MEG to study the neural interactions between different brain regions (e.g., between frontal cortex and visual cortex).[28] Magnetoencephalography can also be used to study changes in neural oscillations across different stages of consciousness, such as in sleep.[29]

Focal epilepsy edit

The clinical uses of MEG are in detecting and localizing pathological activity in patients with epilepsy, and in localizing eloquent cortex for surgical planning in patients with brain tumors or intractable epilepsy. The goal of epilepsy surgery is to remove the epileptogenic tissue while sparing healthy brain areas.[30] Knowing the exact position of essential brain regions (such as the primary motor cortex and primary sensory cortex, visual cortex, and areas involved in speech production and comprehension) helps to avoid surgically induced neurological deficits. Direct cortical stimulation and somatosensory evoked potentials recorded on electrocorticography (ECoG) are considered the gold standard for localizing essential brain regions. These procedures can be performed either intraoperatively or from chronically indwelling subdural grid electrodes. Both are invasive.

Noninvasive MEG localizations of the central sulcus obtained from somatosensory evoked magnetic fields show strong agreement with these invasive recordings.[31][32][33] MEG studies assist in clarification of the functional organization of primary somatosensory cortex and to delineate the spatial extent of hand somatosensory cortex by stimulation of the individual digits. This agreement between invasive localization of cortical tissue and MEG recordings shows the effectiveness of MEG analysis and indicates that MEG may substitute invasive procedures in the future.

Fetal edit

MEG has been used to study cognitive processes such as vision, audition, and language processing in fetuses and newborns.[34] Only two bespoke MEG systems, designed specifically for fetal recordings, operate worldwide.[35] The first was installed at the University of Arkansas in 2000, and the second was installed at the University of Tübingen in 2008. Both devices are referred to as SQUID arrays for reproductive assessment (SARA) and utilize a concave sensor array whose shape compliments the abdomen of a pregnant woman. Fetal recordings of cortical activity are feasible with a SARA device from a gestational age of approximately 25 weeks onward until birth. Although built for fetal recordings, SARA systems can also record from infants placed in a cradle head-first toward the sensory array.[35] While only a small number of devices worldwide are capable of fetal MEG recordings as of 2023, the proliferation of optically pumped magnetometers for MEG in neuroscience research[36] will likely result in a greater number of research centers capable of recording and publishing fetal MEG data in the near future.[35]

Traumatic brain injury edit

MEG can be used to identify traumatic brain injury, which is particularly common among soldiers exposed to explosions. Such injuries are not easily diagnosed by other methods, and are often misdiagnosed as post-traumatic stress disorder (PTSD).[24]

Comparison with related techniques edit

MEG has been in development since the 1960s but has been greatly aided by recent advances in computing algorithms and hardware, and promises improved spatial resolution coupled with extremely high temporal resolution (better than 1 ms). Since the MEG signal is a direct measure of neuronal activity, its temporal resolution is comparable with that of intracranial electrodes.

MEG complements other brain activity measurement techniques such as electroencephalography (EEG), positron emission tomography (PET), and fMRI. Its strengths consist in independence of head geometry compared to EEG (unless ferromagnetic implants are present), non-invasiveness, use of no ionizing radiation, as opposed to PET and high temporal resolution as opposed to fMRI.

MEG in comparison to EEG edit

Although EEG and MEG signals originate from the same neurophysiological processes, there are important differences.[37] Magnetic fields are less distorted than electric fields by the skull and scalp, which results in a better spatial resolution of the MEG. Whereas scalp EEG is sensitive to both tangential and radial components of a current source in a spherical volume conductor, MEG detects only its tangential components. Scalp EEG can, therefore, detect activity both in the sulci and at the top of the cortical gyri, whereas MEG is most sensitive to activity originating in sulci. EEG is, therefore, sensitive to activity in more brain areas, but activity that is visible in MEG can also be localized with more accuracy.

Scalp EEG is sensitive to extracellular volume currents produced by postsynaptic potentials. MEG detects intracellular currents associated primarily with these synaptic potentials because the field components generated by volume currents tend to cancel out in a spherical volume conductor.[38] The decay of magnetic fields as a function of distance is more pronounced than for electric fields. Therefore, MEG is more sensitive to superficial cortical activity, which makes it useful for the study of neocortical epilepsy. Finally, MEG is reference-free, while scalp EEG relies on a reference that, when active, makes interpretation of the data difficult.

See also edit

References edit

  1. ^ Hämäläinen M, Hari R, Ilmoniemi RJ, Knuutila J, Lounasmaa OV (1993). "Magnetoencephalography—theory, instrumentation, and applications to noninvasive studies of the working human brain" (PDF). Reviews of Modern Physics. 65 (2): 413–497. Bibcode:1993RvMP...65..413H. doi:10.1103/RevModPhys.65.413. ISSN 0034-6861.
  2. ^ a b Boto E, Holmes N, Leggett J, Roberts G, Shah V, Meyer SS, Muñoz LD, Mullinger KJ, Tierney TM (March 2018). "Moving magnetoencephalography towards real-world applications with a wearable system". Nature. 555 (7698): 657–661. Bibcode:2018Natur.555..657B. doi:10.1038/nature26147. ISSN 1476-4687. PMC 6063354. PMID 29562238.
  3. ^ Carlson NR (2013). Physiology of Behavior. Upper Saddle River, NJ: Pearson Education Inc. pp. 152–153. ISBN 978-0-205-23939-9.
  4. ^ a b Cohen D (August 1968). "Magnetoencephalography: evidence of magnetic fields produced by alpha-rhythm currents". Science. 161 (3843): 784–6. Bibcode:1968Sci...161..784C. doi:10.1126/science.161.3843.784. PMID 5663803. S2CID 34001253.
  5. ^ Zimmerman JE, Theine P, Harding JT (1970). "Design and operation of stable rf-biased superconducting point-contact quantum devices, etc". Journal of Applied Physics. 41 (4): 1572–1580. Bibcode:1970JAP....41.1572Z. doi:10.1063/1.1659074.
  6. ^ Cohen D (February 1972). "Magnetoencephalography: detection of the brain's electrical activity with a superconducting magnetometer" (PDF). Science. 175 (4022): 664–6. Bibcode:1972Sci...175..664C. doi:10.1126/science.175.4022.664. PMID 5009769. S2CID 29638065.
  7. ^ Yamamoto T, Williamson SJ, Kaufman L, Nicholson C, Llinas R: Magnetic localization of neuronal activity in the human brain. Proc Natl Acad Sci 85:8732-8736,1988[9]
  8. ^ Sander TH, Preusser J, Mhaskar R, Kitching J, Trahms L, Knappe S (May 2012). "Magnetoencephalography with a chip-scale atomic magnetometer". Biomedical Optics Express. 3 (5): 981–90. doi:10.1364/BOE.3.000981. PMC 3342203. PMID 22567591.
  9. ^ Hämäläinen M, Hari R, Ilmoniemi RJ, Knuutila J, Lounasmaa OV (1993-04-01). "Magnetoencephalography---theory, instrumentation, and applications to noninvasive studies of the working human brain" (PDF). Reviews of Modern Physics. 65 (2): 413–497. Bibcode:1993RvMP...65..413H. doi:10.1103/RevModPhys.65.413.
  10. ^ Okada Y (1983). "Neurogenesis of evoked magnetic fields". In Williamson SH, Romani GL, Kaufman L, Modena I (eds.). Biomagnetism: an Interdisciplinary Approach. New York: Plenum Press. pp. 399–408. ISBN 978-1-4757-1785-3.
  11. ^ Cohen D, Schläpfer U, Ahlfors S, Hämäläinen M, Halgren E. (PDF). Charlestown, Massachusetts: Athinoula A. Martinos Center for Biomedical Imaging, Massachusetts General Hospital. S2CID 27016664. Archived from the original (PDF) on 2020-08-03. {{cite journal}}: Cite journal requires |journal= (help)
  12. ^ Tanzer IO (2006). Numerical Modeling in Electro- and Magnetoencephalography (Ph.D. thesis). Finland: Helsinki University of Technology.
  13. ^ Hauk O, Wakeman DG, Henson R (February 2011). "Comparison of noise-normalized minimum norm estimates for MEG analysis using multiple resolution metrics". NeuroImage. 54 (3): 1966–74. doi:10.1016/j.neuroimage.2010.09.053. PMC 3018574. PMID 20884360.
  14. ^ Sheltraw D, Coutsias E (2003). "Invertibility of current density from near-field electromagnetic data" (PDF). Journal of Applied Physics. 94 (8): 5307–5315. Bibcode:2003JAP....94.5307S. doi:10.1063/1.1611262.
  15. ^ Huang MX, Dale AM, Song T, Halgren E, Harrington DL, Podgorny I, Canive JM, Lewis S, Lee RR (July 2006). "Vector-based spatial-temporal minimum L1-norm solution for MEG". NeuroImage. 31 (3): 1025–37. doi:10.1016/j.neuroimage.2006.01.029. PMID 16542857. S2CID 9607000.
  16. ^ Hämäläinen MS, Ilmoniemi RJ (January 1994). "Interpreting magnetic fields of the brain: minimum norm estimates". Medical & Biological Engineering & Computing. 32 (1): 35–42. doi:10.1007/BF02512476. PMID 8182960. S2CID 6796187.
  17. ^ Molins A, Stufflebeam SM, Brown EN, Hämäläinen MS (September 2008). "Quantification of the benefit from integrating MEG and EEG data in minimum l2-norm estimation". NeuroImage. 42 (3): 1069–77. doi:10.1016/j.neuroimage.2008.05.064. PMID 18602485. S2CID 6462818.
  18. ^ Jung TP, Makeig S, Westerfield M, Townsend J, Courchesne E, Sejnowski TJ (October 2000). "Removal of eye activity artifacts from visual event-related potentials in normal and clinical subjects". Clinical Neurophysiology. 111 (10): 1745–58. CiteSeerX 10.1.1.164.9941. doi:10.1016/S1388-2457(00)00386-2. PMID 11018488. S2CID 11044416.
  19. ^ Cui R, Cunnington R, Beisteiner R, Deecke L (2012). "Effects of force-load on cortical activity preceding voluntary finger movement". Neurology, Psychiatry and Brain Research. 18 (3): 97–104. doi:10.1016/j.npbr.2012.03.001.
  20. ^ Hirano Y, Hirano S, Maekawa T, Obayashi C, Oribe N, Monji A, Kasai K, Kanba S, Onitsuka T (March 2010). "Auditory gating deficit to human voices in schizophrenia: a MEG study". Schizophrenia Research. 117 (1): 61–7. doi:10.1016/j.schres.2009.09.003. PMID 19783406. S2CID 7845180.
  21. ^ Ihara A, Wei Q, Matani A, Fujimaki N, Yagura H, Nogai T, Umehara H, Murata T (January 2012). "Language comprehension dependent on emotional context: a magnetoencephalography study". Neuroscience Research. 72 (1): 50–8. doi:10.1016/j.neures.2011.09.011. PMID 22001763. S2CID 836242.
  22. ^ Georgopoulos AP, Karageorgiou E, Leuthold AC, Lewis SM, Lynch JK, Alonso AA, Aslam Z, Carpenter AF, Georgopoulos A, Hemmy LS, Koutlas IG, Langheim FJ, McCarten JR, McPherson SE, Pardo JV, Pardo PJ, Parry GJ, Rottunda SJ, Segal BM, Sponheim SR, Stanwyck JJ, Stephane M, Westermeyer JJ (December 2007). "Synchronous neural interactions assessed by magnetoencephalography: a functional biomarker for brain disorders". Journal of Neural Engineering. 4 (4): 349–55. Bibcode:2007JNEng...4..349G. doi:10.1088/1741-2560/4/4/001. hdl:10161/12446. PMID 18057502. S2CID 2836220.
  23. ^ Montez T, Poil SS, Jones BF, Manshanden I, Verbunt JP, van Dijk BW, Brussaard AB, van Ooyen A, Stam CJ, Scheltens P, Linkenkaer-Hansen K (February 2009). "Altered temporal correlations in parietal alpha and prefrontal theta oscillations in early-stage Alzheimer disease". Proceedings of the National Academy of Sciences of the United States of America. 106 (5): 1614–9. Bibcode:2009PNAS..106.1614M. doi:10.1073/pnas.0811699106. PMC 2635782. PMID 19164579.
  24. ^ a b Rose D (20 February 2022). "British army veterans denied treatment for brain injuries". The Observer.
  25. ^ Bagherzadeh Y, Baldauf D, Pantazis D, Desimone R (February 2020). "Alpha Synchrony and the Neurofeedback Control of Spatial Attention". Neuron. 105 (3): 577–587.e5. doi:10.1016/j.neuron.2019.11.001. hdl:11572/252726. ISSN 0896-6273. PMID 31812515. S2CID 208614924.
  26. ^ de Vries E, Baldauf D (2019-10-01). "Attentional Weighting in the Face Processing Network: A Magnetic Response Image-guided Magnetoencephalography Study Using Multiple Cyclic Entrainments". Journal of Cognitive Neuroscience. 31 (10): 1573–1588. doi:10.1162/jocn_a_01428. hdl:11572/252722. ISSN 0898-929X. PMID 31112470. S2CID 160012572.
  27. ^ Vries IE, Marinato G, Baldauf D (2021-08-24). "Decoding object-based auditory attention from source-reconstructed MEG alpha oscillations". Journal of Neuroscience. 41 (41): 8603–8617. doi:10.1523/JNEUROSCI.0583-21.2021. ISSN 0270-6474. PMC 8513695. PMID 34429378.
  28. ^ Baldauf D, Desimone R (2014-04-25). "Neural Mechanisms of Object-Based Attention". Science. 344 (6182): 424–427. Bibcode:2014Sci...344..424B. doi:10.1126/science.1247003. ISSN 0036-8075. PMID 24763592. S2CID 34728448.
  29. ^ Brancaccio A, Tabarelli D, Bigica M, Baldauf D (2020-04-24). "Cortical source localization of sleep-stage specific oscillatory activity". Scientific Reports. 10 (1): 6976. Bibcode:2020NatSR..10.6976B. doi:10.1038/s41598-020-63933-5. ISSN 2045-2322. PMC 7181624. PMID 32332806.
  30. ^ Luders HO (1992). Epilepsy Surgery. New York Raven Press.
  31. ^ Sutherling WW, Crandall PH, Darcey TM, Becker DP, Levesque MF, Barth DS (November 1988). "The magnetic and electric fields agree with intracranial localizations of somatosensory cortex". Neurology. 38 (11): 1705–14. doi:10.1212/WNL.38.11.1705. PMID 3185905. S2CID 8828767.
  32. ^ Rowley HA, Roberts TP (November 1995). "Functional localization by magnetoencephalography". Neuroimaging Clinics of North America. 5 (4): 695–710. PMID 8564291.
  33. ^ Gallen CC, Hirschkoff EC, Buchanan DS (May 1995). "Magnetoencephalography and magnetic source imaging. Capabilities and limitations". Neuroimaging Clinics of North America. 5 (2): 227–49. PMID 7640886.
  34. ^ Sheridan CJ, Matuz T, Draganova R, Eswaran H, Preissl H (2010). "Fetal Magnetoencephalography - Achievements and Challenges in the Study of Prenatal and Early Postnatal Brain Responses: A Review". Infant and Child Development. 19 (1): 80–93. doi:10.1002/icd.657. PMC 2830651. PMID 20209112.
  35. ^ a b c Frohlich J, Bayne T, Crone JS, DallaVecchia A, Kirkeby-Hinrup A, Mediano PA, Moser J, Talar K, Gharabaghi A, Preissl H (2023-06-01). "Not with a "zap" but with a "beep": Measuring the origins of perinatal experience". NeuroImage. 273: 120057. doi:10.1016/j.neuroimage.2023.120057. ISSN 1053-8119. PMID 37001834. S2CID 257807321.
  36. ^ Brookes MJ, Leggett J, Rea M, Hill RM, Holmes N, Boto E, Bowtell R (August 2022). "Magnetoencephalography with optically pumped magnetometers (OPM-MEG): the next generation of functional neuroimaging". Trends in Neurosciences. 45 (8): 621–634. doi:10.1016/j.tins.2022.05.008. ISSN 0166-2236. PMC 10465236. PMID 35779970. S2CID 250122240.
  37. ^ Cohen D, Cuffin BN (July 1983). "Demonstration of useful differences between magnetoencephalogram and electroencephalogram". Electroencephalography and Clinical Neurophysiology. 56 (1): 38–51. doi:10.1016/0013-4694(83)90005-6. PMID 6190632.
  38. ^ Barth DS, Sutherling W, Beatty J (March 1986). "Intracellular currents of interictal penicillin spikes: evidence from neuromagnetic mapping". Brain Research. 368 (1): 36–48. doi:10.1016/0006-8993(86)91040-1. PMID 3955364. S2CID 3078690.

Further reading edit

  • Baillet S, Mosher JC, Leahy RM (November 2001). "Electromagnetic Brain Mapping". IEEE Signal Processing Magazine. 18 (6): 14–30. Bibcode:2001ISPM...18...14B. doi:10.1109/79.962275.
  • Cohen D (2004). "Boston and the history of biomagnetism". Neurology and Clinical Neurophysiology. 30 (1): 114. PMID 16012683.
  • Cohen D, Halgren E (2004). "Magnetoencephalography". In Adelman G, Smith B (eds.). Encyclopedia of Neuroscience. Elsevier.
  • Hämäläinen M, Hari R, Ilmoniemi R, Knuutila J, Lounasmaa OV (1993). "Magnetoencephalography – theory, instrumentation, and applications to noninvasive studies of signal processing in the human brain" (PDF). Reviews of Modern Physics. 65 (2): 413–497. Bibcode:1993RvMP...65..413H. doi:10.1103/revmodphys.65.413.
  • Hansen PC, Kringelbach ML, Salmelin R (2010). MEG: An Introduction to Methods. New York: Oxford University Press Inc.
  • Murakami S, Okada Y (September 2006). "Contributions of principal neocortical neurons to magnetoencephalography and electroencephalography signals". The Journal of Physiology. 575 (Pt 3): 925–36. doi:10.1113/jphysiol.2006.105379. PMC 1995687. PMID 16613883.
  • Suk J, Ribary U, Cappell J, Yamamoto T, Llinás R (March 1991). "Anatomical localization revealed by MEG recordings of the human somatosensory system". Electroencephalography and Clinical Neurophysiology. 78 (3): 185–96. doi:10.1016/0013-4694(91)90032-y. PMID 1707790.
  • Tanzer OI (2006). Numerical Modeling in Electro- and Magnetoencephalography (Ph.D. thesis). Finland: Helsinki University of Technology.
  • Wolters CH, Anwander A, Tricoche X, Weinstein D, Koch MA, MacLeod RS (2006). "Influence of Tissue Conductivity Anisotropy on EEG/MEG Field and Return Current Computation in a realistic Head Model: A Simulation and Visualization Study using High-Resolution Finite Element Modeling". NeuroImage. 30 (3): 813–826. doi:10.1016/j.neuroimage.2005.10.014. hdl:11858/00-001M-0000-0010-BD20-9. PMID 16364662. S2CID 5578998.

magnetoencephalography, functional, neuroimaging, technique, mapping, brain, activity, recording, magnetic, fields, produced, electrical, currents, occurring, naturally, brain, using, very, sensitive, magnetometers, arrays, squids, superconducting, quantum, in. Magnetoencephalography MEG is a functional neuroimaging technique for mapping brain activity by recording magnetic fields produced by electrical currents occurring naturally in the brain using very sensitive magnetometers Arrays of SQUIDs superconducting quantum interference devices are currently the most common magnetometer while the SERF spin exchange relaxation free magnetometer is being investigated for future machines 1 2 Applications of MEG include basic research into perceptual and cognitive brain processes localizing regions affected by pathology before surgical removal determining the function of various parts of the brain and neurofeedback This can be applied in a clinical setting to find locations of abnormalities as well as in an experimental setting to simply measure brain activity 3 MagnetoencephalographyPerson undergoing a MEGMeSHD015225 Contents 1 History 2 The basis of the MEG signal 3 Magnetic shielding 3 1 Magnetically shielded room MSR 3 2 Active shielding system 4 Source localization 4 1 The inverse problem 4 2 Magnetic source imaging 4 3 Dipole model source localization 4 4 Distributed source models 4 5 Independent component analysis ICA 5 Use in the field 5 1 Brain connectivity and neural oscillations 5 2 Focal epilepsy 5 3 Fetal 5 4 Traumatic brain injury 6 Comparison with related techniques 6 1 MEG in comparison to EEG 7 See also 8 References 9 Further readingHistory edit nbsp Dr Cohen s shielded room at MIT in which first MEG was measured with a SQUID nbsp First MEG measured with SQUID in Dr Cohen s room at MITMEG signals were first measured by University of Illinois physicist David Cohen in 1968 4 before the availability of the SQUID using a copper induction coil as the detector To reduce the magnetic background noise the measurements were made in a magnetically shielded room The coil detector was barely sensitive enough resulting in poor noisy MEG measurements that were difficult to use Later Cohen built a much better shielded room at MIT and used one of the first SQUID detectors just developed by James E Zimmerman a researcher at Ford Motor Company 5 to again measure MEG signals 6 This time the signals were almost as clear as those of EEG This stimulated the interest of physicists who had been looking for uses of SQUIDs Subsequent to this various types of spontaneous and evoked MEGs began to be measured At first a single SQUID detector was used to successively measure the magnetic field at a number of points around the subject s head This was cumbersome and in the 1980s MEG manufacturers began to arrange multiple sensors into arrays to cover a larger area of the head 7 Present day MEG arrays are set in a helmet shaped vacuum flask that typically contain 300 sensors covering most of the head In this way MEGs of a subject or patient can now be accumulated rapidly and efficiently Recent developments attempt to increase portability of MEG scanners by using spin exchange relaxation free SERF magnetometers SERF magnetometers are relatively small as they do not require bulky cooling systems to operate At the same time they feature sensitivity equivalent to that of SQUIDs In 2012 it was demonstrated that MEG could work with a chip scale atomic magnetometer CSAM type of SERF 8 More recently in 2017 researchers built a working prototype that uses SERF magnetometers installed into portable individually 3D printed helmets 2 which they noted in interviews could be replaced with something easier to use in future such as a bike helmet The basis of the MEG signal editSynchronized neuronal currents induce weak magnetic fields The brain s magnetic field measuring at 10 femtotesla fT for cortical activity and 103 fT for the human alpha rhythm is considerably smaller than the ambient magnetic noise in an urban environment which is on the order of 108 fT or 0 1 mT The essential problem of biomagnetism is thus the weakness of the signal relative to the sensitivity of the detectors and to the competing environmental noise nbsp Origin of the brain s magnetic field The electric current also produces the EEG signal The MEG and EEG signals derive from the net effect of ionic currents flowing in the dendrites of neurons during synaptic transmission In accordance with Maxwell s equations any electrical current will produce a magnetic field and it is this field that is measured The net currents can be thought of as current dipoles 9 i e currents with a position orientation and magnitude but no spatial extent dubious discuss According to the right hand rule a current dipole gives rise to a magnetic field that points around the axis of its vector component To generate a signal that is detectable approximately 50 000 active neurons are needed 10 Since current dipoles must have similar orientations to generate magnetic fields that reinforce each other it is often the layer of pyramidal cells which are situated perpendicular to the cortical surface that gives rise to measurable magnetic fields Bundles of these neurons that are orientated tangentially to the scalp surface project measurable portions of their magnetic fields outside of the head and these bundles are typically located in the sulci Researchers are experimenting with various signal processing methods in the search for methods that detect deep brain i e non cortical signal but no clinically useful method is currently available It is worth noting that action potentials do not usually produce an observable field mainly because the currents associated with action potentials flow in opposite directions and the magnetic fields cancel out However action fields have been measured from peripheral nerve system Magnetic shielding editSince the magnetic signals emitted by the brain are on the order of a few femtoteslas shielding from external magnetic signals including the Earth s magnetic field is necessary Appropriate magnetic shielding can be obtained by constructing rooms made of aluminium and mu metal for reducing high frequency and low frequency noise respectively nbsp Entrance to MSR showing the separate shielding layersMagnetically shielded room MSR edit A magnetically shielded room MSR model consists of three nested main layers Each of these layers is made of a pure aluminium layer plus a high permeability ferromagnetic layer similar in composition to molybdenum permalloy The ferromagnetic layer is supplied as 1 mm sheets while the innermost layer is composed of four sheets in close contact and the outer two layers are composed of three sheets each Magnetic continuity is maintained by overlay strips Insulating washers are used in the screw assemblies to ensure that each main layer is electrically isolated This helps eliminate radio frequency radiation which would degrade SQUID performance Electrical continuity of the aluminium is also maintained by aluminium overlay strips to ensure AC eddy current shielding which is important at frequencies greater than 1 Hz The junctions of the inner layer are often electroplated with silver or gold to improve conductivity of the aluminium layers 11 Active shielding system edit Active systems are designed for three dimensional noise cancellation To implement an active system low noise fluxgate magnetometers are mounted at the center of each surface and oriented orthogonally to it This negatively feeds a DC amplifier through a low pass network with a slow falloff to minimize positive feedback and oscillation Built into the system are shaking and degaussing wires Shaking wires increase the magnetic permeability while the permanent degaussing wires are applied to all surfaces of the inner main layer to degauss the surfaces 4 Moreover noise cancellation algorithms can reduce both low frequency and high frequency noise Modern systems have a noise floor of around 2 3 fT Hz0 5 above 1 Hz Source localization editThe inverse problem edit Main article Inverse problem The challenge posed by MEG is to determine the location of electric activity within the brain from the induced magnetic fields outside the head Problems such as this where model parameters the location of the activity have to be estimated from measured data the SQUID signals are referred to as inverse problems in contrast to forward problems 12 where the model parameters e g source location are known and the data e g the field at a given distance is to be estimated The primary difficulty is that the inverse problem does not have a unique solution i e there are infinite possible correct answers and the problem of defining the best solution is itself the subject of intensive research 13 Possible solutions can be derived using models involving prior knowledge of brain activity The source models can be either over determined or under determined An over determined model may consist of a few point like sources equivalent dipoles whose locations are then estimated from the data Under determined models may be used in cases where many different distributed areas are activated distributed source solutions there are infinitely many possible current distributions explaining the measurement results but the most likely is selected Localization algorithms make use of given source and head models to find a likely location for an underlying focal field generator One type of localization algorithm for overdetermined models operates by expectation maximization the system is initialized with a first guess A loop is started in which a forward model is used to simulate the magnetic field that would result from the current guess The guess is adjusted to reduce the discrepancy between the simulated field and the measured field This process is iterated until convergence Another common technique is beamforming wherein a theoretical model of the magnetic field produced by a given current dipole is used as a prior along with second order statistics of the data in the form of a covariance matrix to calculate a linear weighting of the sensor array the beamformer via the Backus Gilbert inverse This is also known as a linearly constrained minimum variance LCMV beamformer When the beamformer is applied to the data it produces an estimate of the power in a virtual channel at the source location The extent to which the constraint free MEG inverse problem is ill posed cannot be overemphasized If one s goal is to estimate the current density within the human brain with say a 5mm resolution then it is well established that the vast majority of the information needed to perform a unique inversion must come not from the magnetic field measurement but rather from the constraints applied to the problem Furthermore even when a unique inversion is possible in the presence of such constraints said inversion can be unstable These conclusions are easily deduced from published works 14 Magnetic source imaging edit The source locations can be combined with magnetic resonance imaging MRI images to create magnetic source images MSI The two sets of data are combined by measuring the location of a common set of fiducial points marked during MRI with lipid markers and marked during MEG with electrified coils of wire that give off magnetic fields The locations of the fiducial points in each data set are then used to define a common coordinate system so that superimposing the functional MEG data onto the structural MRI data coregistration is possible A criticism of the use of this technique in clinical practice is that it produces colored areas with definite boundaries superimposed upon an MRI scan the untrained viewer may not realize that the colors do not represent a physiological certainty not because of the relatively low spatial resolution of MEG but rather some inherent uncertainty in the probability cloud derived from statistical processes However when the magnetic source image corroborates other data it can be of clinical utility Dipole model source localization edit A widely accepted source modeling technique for MEG involves calculating a set of equivalent current dipoles ECDs which assumes the underlying neuronal sources to be focal This dipole fitting procedure is non linear and over determined since the number of unknown dipole parameters is smaller than the number of MEG measurements 15 Automated multiple dipole model algorithms such as multiple signal classification MUSIC and multi start spatial and temporal modeling MSST are applied to the analysis of MEG responses The limitations of dipole models for characterizing neuronal responses are 1 difficulties in localizing extended sources with ECDs 2 problems with accurately estimating the total number of dipoles in advance and 3 dependency on dipole location especially depth in the brain Distributed source models edit Unlike multiple dipole modeling distributed source models divide the source space into a grid containing a large number of dipoles The inverse problem is to obtain the dipole moments for the grid nodes 16 As the number of unknown dipole moments is much greater than the number of MEG sensors the inverse solution is highly underdetermined so additional constraints are needed to reduce ambiguity of the solution The primary advantage of this approach is that no prior specification of the source model is necessary However the resulting distributions may be difficult to interpret because they only reflect a blurred or even distorted image of the true neuronal source distribution The matter is complicated by the fact that spatial resolution depends strongly on various parameters such as brain area depth orientation number of sensors etc 17 Independent component analysis ICA edit Independent component analysis ICA is another signal processing solution that separates different signals that are statistically independent in time It is primarily used to remove artifacts such as blinking eye muscle movement facial muscle artifacts cardiac artifacts etc from MEG and EEG signals that may be contaminated with outside noise 18 However ICA has poor resolution of highly correlated brain sources Use in the field edit nbsp Over 100 MEG systems are known to operate worldwide with Japan possessing the greatest number of MEG systems per capita and the United States possessing the greatest overall number of MEG systems A very small number of systems worldwide are designed for infant and or fetal recordings In research MEG s primary use is the measurement of time courses of activity MEG can resolve events with a precision of 10 milliseconds or faster while functional magnetic resonance imaging fMRI which depends on changes in blood flow can at best resolve events with a precision of several hundred milliseconds MEG also accurately pinpoints sources in primary auditory somatosensory and motor areas For creating functional maps of human cortex during more complex cognitive tasks MEG is most often combined with fMRI as the methods complement each other Neuronal MEG and hemodynamic fMRI data do not necessarily agree in spite of the tight relationship between local field potentials LFP and blood oxygenation level dependent BOLD signals MEG and BOLD signals may originate from the same source though the BOLD signals are filtered through the hemodynamic response MEG is also being used to better localize responses in the brain The openness of the MEG setup allows external auditory and visual stimuli to be easily introduced Some movement by the subject is also possible as long as it does not jar the subject s head The responses in the brain before during and after the introduction of such stimuli movement can then be mapped with greater spatial resolution than was previously possible with EEG 19 Psychologists are also taking advantage of MEG neuroimaging to better understand relationships between brain function and behavior For example a number of studies have been done comparing the MEG responses of patients with psychological troubles to control patients There has been great success isolating unique responses in patients with schizophrenia such as auditory gating deficits to human voices 20 MEG is also being used to correlate standard psychological responses such as the emotional dependence of language comprehension 21 Recent studies have reported successful classification of patients with multiple sclerosis Alzheimer s disease schizophrenia Sjogren s syndrome chronic alcoholism facial pain and thalamocortical dysrhythmias MEG can be used to distinguish these patients from healthy control subjects suggesting a future role of MEG in diagnostics 22 23 A large part of the difficulty and cost of using MEG is the need for manual analysis of the data Progress has been made in analysis by computer comparing a patient s scans with those drawn from a large database of normal scans with the potential to reduce cost greatly 24 Brain connectivity and neural oscillations edit Based on its perfect temporal resolution magnetoencephalography MEG is now heavily used to study oscillatory activity in the brain both in terms of local neural synchrony and cross area synchronisation As an example for local neural synchrony MEG has been used to investigate alpha rhythms in various targeted brain regions such as in visual 25 26 or auditory cortex 27 Other studies have used MEG to study the neural interactions between different brain regions e g between frontal cortex and visual cortex 28 Magnetoencephalography can also be used to study changes in neural oscillations across different stages of consciousness such as in sleep 29 Focal epilepsy edit The clinical uses of MEG are in detecting and localizing pathological activity in patients with epilepsy and in localizing eloquent cortex for surgical planning in patients with brain tumors or intractable epilepsy The goal of epilepsy surgery is to remove the epileptogenic tissue while sparing healthy brain areas 30 Knowing the exact position of essential brain regions such as the primary motor cortex and primary sensory cortex visual cortex and areas involved in speech production and comprehension helps to avoid surgically induced neurological deficits Direct cortical stimulation and somatosensory evoked potentials recorded on electrocorticography ECoG are considered the gold standard for localizing essential brain regions These procedures can be performed either intraoperatively or from chronically indwelling subdural grid electrodes Both are invasive Noninvasive MEG localizations of the central sulcus obtained from somatosensory evoked magnetic fields show strong agreement with these invasive recordings 31 32 33 MEG studies assist in clarification of the functional organization of primary somatosensory cortex and to delineate the spatial extent of hand somatosensory cortex by stimulation of the individual digits This agreement between invasive localization of cortical tissue and MEG recordings shows the effectiveness of MEG analysis and indicates that MEG may substitute invasive procedures in the future Fetal edit MEG has been used to study cognitive processes such as vision audition and language processing in fetuses and newborns 34 Only two bespoke MEG systems designed specifically for fetal recordings operate worldwide 35 The first was installed at the University of Arkansas in 2000 and the second was installed at the University of Tubingen in 2008 Both devices are referred to as SQUID arrays for reproductive assessment SARA and utilize a concave sensor array whose shape compliments the abdomen of a pregnant woman Fetal recordings of cortical activity are feasible with a SARA device from a gestational age of approximately 25 weeks onward until birth Although built for fetal recordings SARA systems can also record from infants placed in a cradle head first toward the sensory array 35 While only a small number of devices worldwide are capable of fetal MEG recordings as of 2023 the proliferation of optically pumped magnetometers for MEG in neuroscience research 36 will likely result in a greater number of research centers capable of recording and publishing fetal MEG data in the near future 35 Traumatic brain injury edit MEG can be used to identify traumatic brain injury which is particularly common among soldiers exposed to explosions Such injuries are not easily diagnosed by other methods and are often misdiagnosed as post traumatic stress disorder PTSD 24 Comparison with related techniques editMEG has been in development since the 1960s but has been greatly aided by recent advances in computing algorithms and hardware and promises improved spatial resolution coupled with extremely high temporal resolution better than 1 ms Since the MEG signal is a direct measure of neuronal activity its temporal resolution is comparable with that of intracranial electrodes MEG complements other brain activity measurement techniques such as electroencephalography EEG positron emission tomography PET and fMRI Its strengths consist in independence of head geometry compared to EEG unless ferromagnetic implants are present non invasiveness use of no ionizing radiation as opposed to PET and high temporal resolution as opposed to fMRI MEG in comparison to EEG edit Although EEG and MEG signals originate from the same neurophysiological processes there are important differences 37 Magnetic fields are less distorted than electric fields by the skull and scalp which results in a better spatial resolution of the MEG Whereas scalp EEG is sensitive to both tangential and radial components of a current source in a spherical volume conductor MEG detects only its tangential components Scalp EEG can therefore detect activity both in the sulci and at the top of the cortical gyri whereas MEG is most sensitive to activity originating in sulci EEG is therefore sensitive to activity in more brain areas but activity that is visible in MEG can also be localized with more accuracy Scalp EEG is sensitive to extracellular volume currents produced by postsynaptic potentials MEG detects intracellular currents associated primarily with these synaptic potentials because the field components generated by volume currents tend to cancel out in a spherical volume conductor 38 The decay of magnetic fields as a function of distance is more pronounced than for electric fields Therefore MEG is more sensitive to superficial cortical activity which makes it useful for the study of neocortical epilepsy Finally MEG is reference free while scalp EEG relies on a reference that when active makes interpretation of the data difficult See also editAuditory evoked field Direct brain interfaces Electrophysiology Evoked field FieldTrip Magnetocardiography Magnetogastrography Magnetometer Magnetomyography SQUID Whole brain emulationReferences edit Hamalainen M Hari R Ilmoniemi RJ Knuutila J Lounasmaa OV 1993 Magnetoencephalography theory instrumentation and applications to noninvasive studies of the working human brain PDF Reviews of Modern Physics 65 2 413 497 Bibcode 1993RvMP 65 413H doi 10 1103 RevModPhys 65 413 ISSN 0034 6861 a b Boto E Holmes N Leggett J Roberts G Shah V Meyer SS Munoz LD Mullinger KJ Tierney TM March 2018 Moving magnetoencephalography towards real world applications with a wearable system Nature 555 7698 657 661 Bibcode 2018Natur 555 657B doi 10 1038 nature26147 ISSN 1476 4687 PMC 6063354 PMID 29562238 Carlson NR 2013 Physiology of Behavior Upper Saddle River NJ Pearson Education Inc pp 152 153 ISBN 978 0 205 23939 9 a b Cohen D August 1968 Magnetoencephalography evidence of magnetic fields produced by alpha rhythm currents Science 161 3843 784 6 Bibcode 1968Sci 161 784C doi 10 1126 science 161 3843 784 PMID 5663803 S2CID 34001253 Zimmerman JE Theine P Harding JT 1970 Design and operation of stable rf biased superconducting point contact quantum devices etc Journal of Applied Physics 41 4 1572 1580 Bibcode 1970JAP 41 1572Z doi 10 1063 1 1659074 Cohen D February 1972 Magnetoencephalography detection of the brain s electrical activity with a superconducting magnetometer PDF Science 175 4022 664 6 Bibcode 1972Sci 175 664C doi 10 1126 science 175 4022 664 PMID 5009769 S2CID 29638065 Yamamoto T Williamson SJ Kaufman L Nicholson C Llinas R Magnetic localization of neuronal activity in the human brain Proc Natl Acad Sci 85 8732 8736 1988 9 Sander TH Preusser J Mhaskar R Kitching J Trahms L Knappe S May 2012 Magnetoencephalography with a chip scale atomic magnetometer Biomedical Optics Express 3 5 981 90 doi 10 1364 BOE 3 000981 PMC 3342203 PMID 22567591 Hamalainen M Hari R Ilmoniemi RJ Knuutila J Lounasmaa OV 1993 04 01 Magnetoencephalography theory instrumentation and applications to noninvasive studies of the working human brain PDF Reviews of Modern Physics 65 2 413 497 Bibcode 1993RvMP 65 413H doi 10 1103 RevModPhys 65 413 Okada Y 1983 Neurogenesis of evoked magnetic fields In Williamson SH Romani GL Kaufman L Modena I eds Biomagnetism an Interdisciplinary Approach New York Plenum Press pp 399 408 ISBN 978 1 4757 1785 3 Cohen D Schlapfer U Ahlfors S Hamalainen M Halgren E New Six Layer Magnetically Shielded Room for MEG PDF Charlestown Massachusetts Athinoula A Martinos Center for Biomedical Imaging Massachusetts General Hospital S2CID 27016664 Archived from the original PDF on 2020 08 03 a href Template Cite journal html title Template Cite journal cite journal a Cite journal requires journal help Tanzer IO 2006 Numerical Modeling in Electro and Magnetoencephalography Ph D thesis Finland Helsinki University of Technology Hauk O Wakeman DG Henson R February 2011 Comparison of noise normalized minimum norm estimates for MEG analysis using multiple resolution metrics NeuroImage 54 3 1966 74 doi 10 1016 j neuroimage 2010 09 053 PMC 3018574 PMID 20884360 Sheltraw D Coutsias E 2003 Invertibility of current density from near field electromagnetic data PDF Journal of Applied Physics 94 8 5307 5315 Bibcode 2003JAP 94 5307S doi 10 1063 1 1611262 Huang MX Dale AM Song T Halgren E Harrington DL Podgorny I Canive JM Lewis S Lee RR July 2006 Vector based spatial temporal minimum L1 norm solution for MEG NeuroImage 31 3 1025 37 doi 10 1016 j neuroimage 2006 01 029 PMID 16542857 S2CID 9607000 Hamalainen MS Ilmoniemi RJ January 1994 Interpreting magnetic fields of the brain minimum norm estimates Medical amp Biological Engineering amp Computing 32 1 35 42 doi 10 1007 BF02512476 PMID 8182960 S2CID 6796187 Molins A Stufflebeam SM Brown EN Hamalainen MS September 2008 Quantification of the benefit from integrating MEG and EEG data in minimum l2 norm estimation NeuroImage 42 3 1069 77 doi 10 1016 j neuroimage 2008 05 064 PMID 18602485 S2CID 6462818 Jung TP Makeig S Westerfield M Townsend J Courchesne E Sejnowski TJ October 2000 Removal of eye activity artifacts from visual event related potentials in normal and clinical subjects Clinical Neurophysiology 111 10 1745 58 CiteSeerX 10 1 1 164 9941 doi 10 1016 S1388 2457 00 00386 2 PMID 11018488 S2CID 11044416 Cui R Cunnington R Beisteiner R Deecke L 2012 Effects of force load on cortical activity preceding voluntary finger movement Neurology Psychiatry and Brain Research 18 3 97 104 doi 10 1016 j npbr 2012 03 001 Hirano Y Hirano S Maekawa T Obayashi C Oribe N Monji A Kasai K Kanba S Onitsuka T March 2010 Auditory gating deficit to human voices in schizophrenia a MEG study Schizophrenia Research 117 1 61 7 doi 10 1016 j schres 2009 09 003 PMID 19783406 S2CID 7845180 Ihara A Wei Q Matani A Fujimaki N Yagura H Nogai T Umehara H Murata T January 2012 Language comprehension dependent on emotional context a magnetoencephalography study Neuroscience Research 72 1 50 8 doi 10 1016 j neures 2011 09 011 PMID 22001763 S2CID 836242 Georgopoulos AP Karageorgiou E Leuthold AC Lewis SM Lynch JK Alonso AA Aslam Z Carpenter AF Georgopoulos A Hemmy LS Koutlas IG Langheim FJ McCarten JR McPherson SE Pardo JV Pardo PJ Parry GJ Rottunda SJ Segal BM Sponheim SR Stanwyck JJ Stephane M Westermeyer JJ December 2007 Synchronous neural interactions assessed by magnetoencephalography a functional biomarker for brain disorders Journal of Neural Engineering 4 4 349 55 Bibcode 2007JNEng 4 349G doi 10 1088 1741 2560 4 4 001 hdl 10161 12446 PMID 18057502 S2CID 2836220 Montez T Poil SS Jones BF Manshanden I Verbunt JP van Dijk BW Brussaard AB van Ooyen A Stam CJ Scheltens P Linkenkaer Hansen K February 2009 Altered temporal correlations in parietal alpha and prefrontal theta oscillations in early stage Alzheimer disease Proceedings of the National Academy of Sciences of the United States of America 106 5 1614 9 Bibcode 2009PNAS 106 1614M doi 10 1073 pnas 0811699106 PMC 2635782 PMID 19164579 a b Rose D 20 February 2022 British army veterans denied treatment for brain injuries The Observer Bagherzadeh Y Baldauf D Pantazis D Desimone R February 2020 Alpha Synchrony and the Neurofeedback Control of Spatial Attention Neuron 105 3 577 587 e5 doi 10 1016 j neuron 2019 11 001 hdl 11572 252726 ISSN 0896 6273 PMID 31812515 S2CID 208614924 de Vries E Baldauf D 2019 10 01 Attentional Weighting in the Face Processing Network A Magnetic Response Image guided Magnetoencephalography Study Using Multiple Cyclic Entrainments Journal of Cognitive Neuroscience 31 10 1573 1588 doi 10 1162 jocn a 01428 hdl 11572 252722 ISSN 0898 929X PMID 31112470 S2CID 160012572 Vries IE Marinato G Baldauf D 2021 08 24 Decoding object based auditory attention from source reconstructed MEG alpha oscillations Journal of Neuroscience 41 41 8603 8617 doi 10 1523 JNEUROSCI 0583 21 2021 ISSN 0270 6474 PMC 8513695 PMID 34429378 Baldauf D Desimone R 2014 04 25 Neural Mechanisms of Object Based Attention Science 344 6182 424 427 Bibcode 2014Sci 344 424B doi 10 1126 science 1247003 ISSN 0036 8075 PMID 24763592 S2CID 34728448 Brancaccio A Tabarelli D Bigica M Baldauf D 2020 04 24 Cortical source localization of sleep stage specific oscillatory activity Scientific Reports 10 1 6976 Bibcode 2020NatSR 10 6976B doi 10 1038 s41598 020 63933 5 ISSN 2045 2322 PMC 7181624 PMID 32332806 Luders HO 1992 Epilepsy Surgery New York Raven Press Sutherling WW Crandall PH Darcey TM Becker DP Levesque MF Barth DS November 1988 The magnetic and electric fields agree with intracranial localizations of somatosensory cortex Neurology 38 11 1705 14 doi 10 1212 WNL 38 11 1705 PMID 3185905 S2CID 8828767 Rowley HA Roberts TP November 1995 Functional localization by magnetoencephalography Neuroimaging Clinics of North America 5 4 695 710 PMID 8564291 Gallen CC Hirschkoff EC Buchanan DS May 1995 Magnetoencephalography and magnetic source imaging Capabilities and limitations Neuroimaging Clinics of North America 5 2 227 49 PMID 7640886 Sheridan CJ Matuz T Draganova R Eswaran H Preissl H 2010 Fetal Magnetoencephalography Achievements and Challenges in the Study of Prenatal and Early Postnatal Brain Responses A Review Infant and Child Development 19 1 80 93 doi 10 1002 icd 657 PMC 2830651 PMID 20209112 a b c Frohlich J Bayne T Crone JS DallaVecchia A Kirkeby Hinrup A Mediano PA Moser J Talar K Gharabaghi A Preissl H 2023 06 01 Not with a zap but with a beep Measuring the origins of perinatal experience NeuroImage 273 120057 doi 10 1016 j neuroimage 2023 120057 ISSN 1053 8119 PMID 37001834 S2CID 257807321 Brookes MJ Leggett J Rea M Hill RM Holmes N Boto E Bowtell R August 2022 Magnetoencephalography with optically pumped magnetometers OPM MEG the next generation of functional neuroimaging Trends in Neurosciences 45 8 621 634 doi 10 1016 j tins 2022 05 008 ISSN 0166 2236 PMC 10465236 PMID 35779970 S2CID 250122240 Cohen D Cuffin BN July 1983 Demonstration of useful differences between magnetoencephalogram and electroencephalogram Electroencephalography and Clinical Neurophysiology 56 1 38 51 doi 10 1016 0013 4694 83 90005 6 PMID 6190632 Barth DS Sutherling W Beatty J March 1986 Intracellular currents of interictal penicillin spikes evidence from neuromagnetic mapping Brain Research 368 1 36 48 doi 10 1016 0006 8993 86 91040 1 PMID 3955364 S2CID 3078690 Further reading editBaillet S Mosher JC Leahy RM November 2001 Electromagnetic Brain Mapping IEEE Signal Processing Magazine 18 6 14 30 Bibcode 2001ISPM 18 14B doi 10 1109 79 962275 Cohen D 2004 Boston and the history of biomagnetism Neurology and Clinical Neurophysiology 30 1 114 PMID 16012683 Cohen D Halgren E 2004 Magnetoencephalography In Adelman G Smith B eds Encyclopedia of Neuroscience Elsevier Hamalainen M Hari R Ilmoniemi R Knuutila J Lounasmaa OV 1993 Magnetoencephalography theory instrumentation and applications to noninvasive studies of signal processing in the human brain PDF Reviews of Modern Physics 65 2 413 497 Bibcode 1993RvMP 65 413H doi 10 1103 revmodphys 65 413 Hansen PC Kringelbach ML Salmelin R 2010 MEG An Introduction to Methods New York Oxford University Press Inc Murakami S Okada Y September 2006 Contributions of principal neocortical neurons to magnetoencephalography and electroencephalography signals The Journal of Physiology 575 Pt 3 925 36 doi 10 1113 jphysiol 2006 105379 PMC 1995687 PMID 16613883 Suk J Ribary U Cappell J Yamamoto T Llinas R March 1991 Anatomical localization revealed by MEG recordings of the human somatosensory system Electroencephalography and Clinical Neurophysiology 78 3 185 96 doi 10 1016 0013 4694 91 90032 y PMID 1707790 Tanzer OI 2006 Numerical Modeling in Electro and Magnetoencephalography Ph D thesis Finland Helsinki University of Technology Wolters CH Anwander A Tricoche X Weinstein D Koch MA MacLeod RS 2006 Influence of Tissue Conductivity Anisotropy on EEG MEG Field and Return Current Computation in a realistic Head Model A Simulation and Visualization Study using High Resolution Finite Element Modeling NeuroImage 30 3 813 826 doi 10 1016 j neuroimage 2005 10 014 hdl 11858 00 001M 0000 0010 BD20 9 PMID 16364662 S2CID 5578998 Retrieved from https en wikipedia org w index php title Magnetoencephalography amp oldid 1199933312, wikipedia, wiki, book, books, library,