Liver support system

A liver support system or diachysis is a type of therapeutic device to assist in performing the functions of the liver. Such systems focus either on removing the accumulating toxins (liver dialysis), or providing additional replacement of the metabolic functions of the liver through the inclusion of hepatocytes to the device (bioartificial liver device). This system is in trial to help people with acute liver failure (ALF) or acute-on-chronic liver failure.[1]

Liver support system
Specialtyhepatology
[edit on Wikidata]

The primary functions of the liver include removing toxic substances from the blood, manufacturing blood proteins, storing energy in the form of glycogen, and secreting bile. The hepatocytes that perform these tasks can be killed or impaired by disease, resulting in hepatic insufficiency. A sudden onset of life-threatening hepatic insufficiency is known as acute liver failure (ALF).

Etymology

  • The word diachysis derives from the Greek word, διάχυσησ, which means "Diffusion"
  • The word dialysis derives from the Greek word, διάλυσις, which means "Dissolution"

Liver failure

 

In hyperacute and acute liver failure the clinical picture develops rapidly with progressive encephalopathy and multiorgan dysfunction such as hyperdynamic circulation, coagulopathy, acute kidney injury and respiratory insufficiency, severe metabolic alterations and cerebral edema that can lead to brain death.[2][3] In these cases the mortality without liver transplantation (LTx) ranges between 40-80%.[4][5] LTx is the only effective treatment for these patients although it requires a precise indication and timing to achieve good results. Nevertheless, due to the scarcity of organs to carry out liver transplantations, it is estimated that one third of patients with ALF die while waiting to be transplanted.[6] On the other hand, a patient with a chronic hepatic disease can suffer an acute decompensation of liver function following a precipitating event such as variceal bleeding, sepsis and excessive alcohol intake among others that can lead to a condition referred to as acute-on-chronic liver failure (ACLF). Both types of hepatic insufficiency, ALF and ACLF, can potentially be reversible and liver functionality can return to a level similar to that prior to the insult or precipitating event. LTx is the only treatment that has shown an improvement in the prognosis and survival with most severe cases of ALF. Nevertheless, cost and donor scarcity have prompted researchers to look for new supportive treatments that can act as “bridge” to the transplant procedure. By stabilizing the patient's clinical state, or by creating the right conditions that could allow the recovery of native liver functions, both detoxification and synthesis can improve, after an episode of ALF or ACLF.[7] Basically, three different types of supportive therapies have been developed: bio-artificial, artificial and hybrid liver support systems (Table 2).

Table 2: Liver Support Systems
Bio-artificial Artificial Hybrids
ELAD[8]

Extracorporeal liver assist device

MARS[9]

Molecular adsorbent recirculating system

Hepat-Assist[10]
BLSS[11]

Bioartificial Liver Support System

Prometheus FPSA[12]

Fractionated plasma separation and adsorption system

TECLA-HALSS[13]

TECA-Hybrid Artificial Liver Support System

RFB[14]

Radial Flow Bioreactor

SPAD[15]

Single-pass albumin dialysis

MELS[16]

Modular Extracorporeal Liver Support

AMC-BAL[17]

Bioartificial Liver

SEPET[18]

Selective plasma filtration therapy

-

Bioartificial liver devices

Liver support system
Specialtyinternal medicine
[edit on Wikidata]

Bioartificial liver devices are experimental extracorporeal devices that use living cell lines to provide detoxification and synthesis support to the failing liver. Bio-artificial liver (BAL) Hepatassist 2000 uses porcine hepatocytes whereas ELAD system employs hepatocytes derived from human hepatoblastoma C3A cell lines.[19][20] Both techniques can produce, in fulminant hepatic failure (FHF), an improvement of hepatic encephalopathy grade and biochemical parameters. Nevertheless, they are therapies with high complexity that require a complex logistic approach for implementation; a very high cost and possible inducement of important side effects such as immunological issues (porcine endogenous retrovirus transmission), infectious complications and tumor transmigration have been documented. Other biological hepatic systems are Bioartificial Liver Support (BLSS) and Radial Flow Bioreactor (RFB). Detoxification capacity of these systems is poor and therefore they must be used combined with other systems to mitigate this deficiency. Today its use is limited to centers with high experience in their application.[21]

A bioartificial liver device (BAL) is an artificial extracorporeal liver support (ELS) system for an individual who is suffering from acute liver failure (ALF) or acute-on-chronic liver failure (ACLF). The fundamental difference between artificial and BAL systems lies in the inclusion of hepatocytes into the reactor, often operating alongside the purification circuits used in artificial ELS systems. The overall design varies between different BAL systems, but they largely follow the same basic structure, with patient blood or plasma flow through an artificial matrix housing hepatocytes. Plasma is often separated from the patient’s blood to improve efficiency of the system, and the device can be connected to artificial liver dialysis devices in order to further increase the effectiveness of the device in filtration of toxins. The inclusion of functioning hepatocytes in the reactor allows the restoration of some of the synthetic functions that the patient’s liver is lacking.[22]

History

Early history

The first bioartificial liver device was developed in 1993 by Dr. Achilles A. Demetriou at Cedars-Sinai Medical Center. The bioartificial liver helped an 18-year-old southern California woman survive without her own liver for 14 hours until she received a human liver using a 20-inch-long, 4-inch-wide plastic cylinder filled with cellulose fibers and pig liver cells. Blood was routed outside the patient's body and through the artificial liver before being returned to the body.[23][24]

Dr Kenneth Matsumara's work on the BAL led it to be named an invention of the year by Time magazine in 2001.[25] Liver cells obtained from an animal were used instead of developing a piece of equipment for each function of the liver. The structure and function of the first device also resembles that of today's BALs. Animal liver cells are suspended in a solution and a patient's blood is processed by a semipermeable membrane that allow toxins and blood proteins to pass but restricts an immunological response.[25]

Development

Advancements in bioengineering techniques in the decade after Matsumara's work have led to improved membranes and hepatocyte attachment systems.[26] As time has progressed the sources of hepatocytes have increased. Cell sources now include primary porcine hepatocytes, primary human hepatocytes, human hepatoblastoma (C3A), immortalized human cell lines and stem cells.[26]

Use

The purpose of BAL-type devices, currently, is not to permanently replace liver functions, but to serve as a supportive device,[27] either allowing the liver to regenerate properly upon acute liver failure, or to bridge the individual's liver functions until a transplant is possible.

Function

BALs are essentially bioreactors, with embedded hepatocytes (liver cells) that perform the functions of a normal [liver]. They process oxygenated blood plasma, which is separated from the other blood constituents.[28] Several types of BALs are being developed, including hollow fiber systems and flat membrane sheet systems.[29]

There are various types of hepatocytes that are used in these devices. Porcine hepatocytes are often used due to ease of acquisition and cost; however, they are relatively unstable and carry the risk of cross-species disease transmission.[30] Primary human hepatocytes sourced from donor organs are the most suitable for use but present several problems in their cost and difficulty to obtain, especially with the current lack in transplantable tissue.[30] In addition, questions have been raised about tissue collected from patients transmitting malignancy or infection via the BAL device. Several lines of human hepatocytes are also used in BAL devices, including C3A and HepG2 tumour cell lines, but due to their origin from hepatomas, they possess the potential to pass on malignancy to the patient.[31] There is ongoing research into the cultivation of new types of human hepatocytes capable of improved longevity and efficacy in a bioreactor over currently used cell types, that do not pose the risk of transfer of malignancy or infection, such as the HepZ cell line created by Werner et al..[32]

Hollow fibre systems

One type of BAL is similar to kidney dialysis systems that employ a hollow fiber cartridge. Hepatocytes are suspended in a gel solution such as collagen, which is injected into a series of hollow fibers. In the case of collagen, the suspension is then gelled within the fibers, usually by a temperature change. The hepatocytes then contract the gel by their attachment to the collagen matrix, reducing the volume of the suspension and creating a flow space within the fibers. Nutrient media is circulated through the fibers to sustain the cells. During use, plasma is removed from the patients blood. The patient's plasma is fed into the space surrounding the fibers. The fibers, which are composed of a semi-permeable membrane, facilitate transfer of toxins, nutrients and other chemicals between the blood and the suspended cells. The membrane also keeps immune bodies, such as immunoglobulins, from passing to the cells to prevent an immune system rejection.[33]

Cryogel-Based Systems

Currently, hollow-fibre bioreactors are the most commonly accepted design for clinical use due to their capillary-network allowing for easy perfusion of plasma across cell populations.[34] However, these structures have their limitations, with convectional transport issues, nutritional gradients, non-uniform seeding, inefficient immobilisation of cells, and reduced hepatocyte growth restricting their effectiveness in BAL designs.[35] Researchers are now investigating the use of cryogels to replace hollow-fibres as the cell carrier components in BAL systems.

Cryogels are super-macroporous three-dimensional polymers prepared at sub-zero temperatures, by the freezing of a solution of cryogel precursors and solvent. The pores develop during this freezing process – as the cryogel solution cools, the solvent begins to form crystals. This causes the concentration of the cryogel precursors in the solution to increase, initiating the cryogelation process and forming the polymer walls. As the cryogel warms, the solvent crystals thaw, leaving cavities that form the pores.[36] Cryogel pores range in size from 10-100 µm in size, forming an interconnected network that mimics a capillary system with a very large surface area to volume ratio, supporting large numbers of immobilised cells. Convection mediated transport is also supported by cryogels, enabling even distribution of nutrients and metabolite elimination, overcoming some of the shortcomings of hollow-fibre systems.[35] Most importantly, cryogel scaffolds demonstrate good mechanical strength and biocompatibility without triggering an immune response, improving their potential for long-term inclusion in BAL devices or in-vitro use.[37] Another advantage of cryogels is their flexibility for use in a variety of tasks, including separation and purification of substances, along with acting as extracellular matrix for cell growth and proliferation. Immobilisation of specific ligands onto cryogels enables adsorption of specific substances, supporting their use as treatment options for toxins,[38] for separation of haemoglobin from blood,[39] and as a localised and sustained method for drug delivery.[40]

Jain et al.[35] investigated poly(AN-co-NVP) and poly(NiPAAm)-chitosan cryogels as hepatocyte carriers in a BAL system. These cryogel polymers were chosen as they maintain a good balance of hydrophilicity/hydrophobicity to promote hepatocyte binding. Chitosan was added to cryogels as it promotes spheroid formation in hepatocytes, an indicator of healthy growth.[41] The researchers were able to demonstrate biocompatibility with human HepG2 cells after high-density growth for 7 days, and with surrounding tissue after implantation into mice. They also evaluated ureagenesis and drug detoxification by hepatocytes in the scaffold, finding that over time the hepatocytes were able to process ammonia into urea, and that CYP450 activity in the cells after 96 hours culturing was equal to that of freshly isolated hepatocytes. The researchers then seeded HepG2 cells onto cryogels and incubated for 48 hours, before incorporating the cryogels into a BAL bioreactor. They circulated plasma isolated from alcoholic ACLF patients for 3 hours through the bioreactor and ran control experiments with cryogels incubated in media only. After the experiment, they found that the cell-seeded bioreactors had reduced the levels of bilirubin and ammonia in the plasma (by 58.9 ± 6.7% and 61.2 ± 7%, respectively) and increased the level of albumin by 31.1 ± 28% compared to the control group. This demonstrated that the HepG2 cells in the bioreactor were able to carry out detoxifying and synthetic functions of normal liver cells. However, after 3 hours of plasma circulation, hepatocyte function deteriorated rapidly. This was likely due to the toxic effect of ACLF plasma on hepatocytes,[42] or nutrient and oxygen depletion. The study clearly demonstrated the ability of HepG2 seeded cryogels to detoxify plasma and support liver function, however further research is needed into what caused the reduction in function after 3 hours.

Damania et al.[43] took this system a step further, utilising HepG2 seeded poly(NiPAAm)-chitosan cryogels in a bioreactor, along with an activated charcoal cloth used as a filter. As seen with the artificial BAL devices, activated charcoal is useful in the filtration of liver toxins, including ammonia and bilirubin. The researchers then incorporated the bioreactor into a rat model, connecting rats with induced liver failure to the system and passing their blood across a plasma-separation membrane, with the plasma running through the seeded cryogel and charcoal filter. They compared results against rats with induced liver failure who were not connected to a bioreactor. Measuring levels of ammonia, urea, bilirubin, albumin, and AST over a 3-hour time period, they found a reduction in bilirubin, AST, and urea levels and an increase in albumin levels, suggesting the bioreactor is functioning in its removal of toxins and synthesis of new proteins. In addition, they ran the same experiment using an acellular bioreactor, and found that bilirubin and AST decreased over time, but to a lesser degree than the seeded bioreactor, showing that the detoxification results of the seeded reactor were due to a combination of both the HepG2 cells, and the activated charcoal cloth. However, in the cell-seeded reactor there was an increase in the levels of ammonia, which the researchers stated could be attributed to these toxins having a detrimental effect on the cells over time, reducing their function.

Clinical Studies

There have been numerous clinical studies involving hollow-fibre bioreactors. Overall, they show promise but do not provide statistically significant evidence supporting their effectiveness. This is generally due to inherent design limitations, causing convectional transport issues, nutritional gradients, non-uniform seeding, inefficient immobilisation of cells, and reduced hepatocyte growth.[35] As of writing, no cryogel-based devices have entered clinical trials. However, laboratory results have been promising,[35][43] and hopefully trials will begin soon.

HepatAssist

The HepatAssist, developed at the Cedars-Sinai Medical Center, is a BAL device containing porcine hepatocytes within a hollow-fibre bioreactor. These semi-permeable fibres act as capillaries, allowing the perfusion of plasma through the device, and across the hepatocytes surrounding the fibres. The system incorporates a charcoal column to act as a filter, removing additional toxins from the plasma.[44]

Demetriou et al.[44] carried out a large, randomised, multicentre, controlled trial on the safety and efficacy of the HepatAssist device. 171 patients with ALF stemming from viral hepatitis, paracetamol overdose or other drug complications, primary non-function (PNF), or of indeterminate aetiology, were involved in the study and were randomly assigned to either the experimental or control groups. Both groups were well balanced in terms of age, gender, race, and aetiology. The study found that at the primary end-point 30-day post admission mark, there was an increased survival rate in BAL patients over control patients (71% vs 62%), but the difference was not significant. However, when patients with PNF are excluded from the results there is a 44% reduction in mortality for BAL treated patients, a statistically significant advantage. The investigators noted that exclusion of PNF patients is justifiable due to early retransplantation and lack of intercranial hypertension, so HepatAssist would give little benefit to this group. For the secondary end-point of time-to-death, in patients with ALF of known aetiology there was a significant difference between BAL and control groups, with BAL patients surviving for longer. There was no significant difference for patients of unknown aetiology, however.

The conclusions of the study suggest that such a device has potentially significant importance when used as a treatment measure. While the overall findings were not statistically significant, when the aetiology of the patients was taken into account the BAL group gained a statistically significant reduction in mortality over the control group. This suggests that while the device may not be applicable to patients as an overall treatment for liver dysfunction, it can provide an advantage when the heterogeneity of patients is considered and is used with patients of specific aetiology.

Extracorporeal Liver Assist Device

The Extracorporeal Liver Assist Device (ELAD) is a human-cell based treatment system. A catheter removes blood from the patient, and an ultrafiltrate generator separates the plasma from the rest of the blood. This plasma is then run through a separate circuit containing cartridges filled with C3A cells, before being returned to the main circuit and re-entering the patient. C3A cells were chosen for this device due to their possession of anti-inflammatory proteins, such as IL-1 receptor antagonist, along with anti-apoptotic and anti-oxidative mechanisms that could aid in the reduction of further liver injury from the patients’ diseased state.[45]

Thompson et al.[45] performed a large open-label trial, measuring the effectiveness of ELAD on patients with severe alcoholic hepatitis resulting in ACLF. Their study involved patients screened at 40 sites across the US, UK, and Australia, and enrolled a total of 203 patients. Patients were then randomised into either ELAD (n=96) or standard medical care (n=107) groups, with even distribution for patients in terms of sex, MELD score, and bilirubin levels. Of the 96 patients in the ELAD group, 45 completed the full 120 hours of treatment – the rest were unable to complete the full regimen due to a variety of reasons, including withdrawal of consent or severe adverse events, though 37 completed >72 hours of treatment, with results showing minimal difference in mortality between those receiving either >72 hours or the full 120 hours of treatment. The study was unable to complete its goal, finding no statistically significant improvement in mortality rates for patients that received ELAD treatment over those receiving standard care at 28 and 91 days (76.0% versus 80.4% and 59.4% versus 61.7%, respectively). Biomarker measurements showed a significantly reduced level of bilirubin and alkaline phosphatase in ELAD patients, though neither improvement translated into increased survivability rates. Outcomes for patients with MELD score <28 showed trends towards improved survival on ELAD, whereas those with MELD >28 had decreased survivability on ELAD. These patients presented with raised creatinine from kidney failure, suggesting a reason why ELAD decreased survival chance over standard care. Unlike artificial ELS devices and HepatAssist, ELAD does not incorporate any filtration devices, such as charcoal columns and exchange resins. Therefore, it cannot replace the filtration capability of the kidneys and cannot compensate for multi-organ failure from more severe presentations of ACLF, resulting in increased mortality rates.

While the results of the study cannot provide conclusive evidence to suggest that a BAL device like ELAD improves the outcome of severe ACLF, it does suggest that it can aid the survival of patients with a less severe form of the disease. In those patients with a MELD <28, beneficial effects were seen 2–3 weeks post treatment, suggesting that while C3A incorporating BAL devices are unable to provide short-term aid like artificial albumin filtration devices, they instead provide more long-term aid in recovery of the patient’s liver.[45]

A randomized, phase 3 trial of the ELAD device in patients with severe alcoholic hepatitis failed to show benefit on overall survival and development was discontinued.[46]

Liver dialysis

Artificial liver support systems are aimed to temporally replace native liver detoxification functions and they use albumin as scavenger molecule to clear the toxins involved in the physiopathology of the failing liver. Most of the toxins that accumulate in the plasma of patients with liver insufficiency are protein bound, and therefore conventional renal dialysis techniques, such as hemofiltration, hemodialysis or hemodiafiltration are not able to adequately eliminate them.

Liver dialysis has shown promise for patients with hepatorenal syndrome. It is similar to hemodialysis and based on the same principles, but hemodialysis does not remove toxins bound to albumin that accumulate in liver failure. Like a bioartificial liver device, it is a form of artificial extracorporeal liver support.[47][48]

A critical issue of the clinical syndrome in liver failure is the accumulation of toxins not cleared by the failing liver. Based on this hypothesis, the removal of lipophilic, albumin-bound substances such as bilirubin, bile acids, metabolites of aromatic amino acids, medium-chain fatty acids and cytokines should be beneficial to the clinical course of a patient in liver failure. This led to the development of artificial filtration and absorption devices.

Liver dialysis is performed by physicians and surgeons and specialized nurses with training in gastroenterological medicine and surgery, namely, in hepatology, the study of the liver, alongside their colleagues in the intensive or critical care unit and the transplantation department, which is responsible for procuring and implanting a new liver, or a part (lobe) of one, if and when it becomes available in time and the patient is eligible. Because of the need for these experts, as well as the relative newness of the procedure in certain areas, it is usually available only in larger hospitals, such as level I trauma center teaching hospitals connected with medical schools.

Between the different albumin dialysis modalities, single pass albumin dialysis (SPAD) has shown some positive results at a very high cost;[49] it has been proposed that lowering the concentration of albumin in the dialysate does not seem to affect the detoxification capability of the procedure.[50] Nevertheless, the most widely used systems today are based on hemodialysis and adsorption. These systems use conventional dialysis methods with an albumin containing dialysate that is later regenerated by means of adsorption columns, filled with activated charcoal and ion exchange resins. At present, there are two artificial extracorporeal liver support systems: the Molecular Adsorbents Recirculating System (MARS) from Gambro and Fractionated Plasma Separation and Adsorption (FPSA), commercialised as Prometheus (PROM) from Fresenius Medical Care. Of the two therapies, MARS is the most frequently studied, and clinically used system to date.

Prognosis/survival

While the technique is in its infancy, the prognosis of patients with liver failure remains guarded. Liver dialysis, currently, is only considered to be a bridge to transplantation or liver regeneration (in the case of acute liver failure)[51][52][53] and, unlike kidney dialysis (for kidney failure), cannot support a patient for an extended period of time (months to years).

Devices

Artificial detoxification devices currently under clinical evaluation include the Single Pass Albumin Dialysis (SPAD), Molecular Adsorbent Recirculating System (MARS)®, Prometheus system, and Dialive.

Single Pass Albumin Dialysis (SPAD)

Single pass albumin dialysis (SPAD) is a simple method of albumin dialysis using standard renal replacement therapy machines without an additional perfusion pump system: The patient's blood flows through a circuit with a high-flux hollow fiber hemodiafilter, identical to that used in the MARS system. The other side of this membrane is cleansed with an albumin solution in counter-directional flow, which is discarded after passing the filter. Hemodialysis can be performed in the first circuit via the same high-flux hollow fibers.

Molecular adsorbents recirculation system

The Molecular Adsorbents Recirculation System (MARS) is the best known extracorporal liver dialysis system and has existed for approximately fifteen years. It consists of two separate dialysis circuits. The first circuit consists of human serum albumin, is in contact with the patient's blood through a semipermeable membrane and has two filters to clean the albumin after it has absorbed toxins from the patient's blood. The second circuit consists of a hemodialysis machine and is used to clean the albumin in the first circuit, before it is recirculated to the semipermeable membrane in contact with the patient's blood.

Comparing SPAD, MARS and CVVHDF

SPAD, MARS and continuous veno-venous haemodiafiltration (CVVHDF) were compared in vitro with regard to detoxification capacity.[54] SPAD and CVVHDF showed a significantly greater reduction of ammonia compared with MARS. No significant differences could be observed between SPAD, MARS and CVVHDF concerning other water-soluble substances. However, SPAD enabled a significantly greater bilirubin reduction than MARS. Bilirubin serves as an important marker substance for albumin-bound (non-water-soluble) substances. Concerning the reduction of bile acids no significant differences between SPAD and MARS were seen. It was concluded that the detoxification capacity of SPAD is similar or even higher when compared with the more sophisticated, more complex and hence more expensive MARS.

As albumin dialysis is a costly procedure, financial aspects are important: For a seven-hour treatment with MARS, approximately €300 for 600 mL human serum albumin solution (20%), €1740 for a MARS treatment kit, and €125 for disposables used by the dialysis machine have to be spent. The cost of this therapy adds up to approximately €2165. Performing SPAD according to the protocol by Sauer et al., however, requires 1000 mL of human albumin solution (20%) at a cost of €500. A high-flux dialyzer costing approximately €40 and the tubings (€125) must also be purchased. The overall costs of a SPAD treatment is approximately €656—30% of the costs of an equally efficient MARS therapy session. The expenditure for the MARS monitor necessary to operate the MARS disposables is not included in this calculation.

Prometheus

The Prometheus system (Fresenius Medical Care, Bad Homburg, Germany) is a device based on the combination of albumin adsorption with high-flux hemodialysis after selective filtration of the albumin fraction through a specific polysulfon filter (AlbuFlow). It has been studied[55] in a group of eleven patients with hepatorenal syndrome (acute-on-chronic liver failure and accompanying kidney failure). The treatment for two consecutive days for more than four hours significantly improved serum levels of conjugated bilirubin, bile acids, ammonia, cholinesterase, creatinine, urea and blood pH. Prometheus was proven to be a safe supportive therapy for patients with liver failure.

Dialive

Dialive (Yaqrit Limited, London, UK) incorporates albumin removal and replacement and, endotoxin removal. It is at "Technology readiness level" (TRL) 5, which means it is validated in the disease environment.[56][57]

The MARS System

MARS was developed by a group of researchers at the University of Rostock (Germany), in 1993 and later commercialized for its clinical use in 1999.[58] The system is able to replace the detoxification function of the liver while minimizing the inconvenience and drawbacks of previously used devices.[59][60][61]

In vivo preliminary investigations indicated the ability of the system to effectively remove bilirubin, biliary salts, free fatty acids and tryptophan while important physiological proteins such as albumin, alpha-1-glicoproteine, alpha 1 antitrypsin, alpha-2-macroglobulin, transferrin, globulin tyrosine, and hormonal systems are unaffected.[62] Also, MARS therapy in conjunction with CRRT/HDF can help clear cytokines acting as inflammatory and immunological mediators in hepatocellular damage, and therefore can create the right environment to favour hepatocellular regeneration and recovery of native liver function.

MARS System Components

 
Combined MARS and PrismaFlex monitors

MARS is an extracorporeal hemodialysis system composed of three different circuits: blood, albumin and low-flux dialysis. The blood circuit uses a double lumen catheter and a conventional hemodialysis device to pump the patient's blood into the MARS FLUX, a biocompatible polysulfone high-flux dialyser. With a membrane surface area of 2.1 m2, 100 nm of thickness and a cut-off of 50 KDa, the MARSFLUX is essential to retaining the albumin in the dialysate. Blood is dialysed against a human serum albumin (HSA) dialysate solution that allows blood detoxification of both water-soluble and protein-bound toxins, by means of the presence of albumin in the dialysate (albumin dialysis). The albumin dialysate is then regenerated in a close loop in the MARS circuit by passing through the fibres of the low-flux diaFLUX filter, to clear water-soluble toxins and provide electrolyte/acid-base balance, by a standard dialysis fluid. Next, the albumin dialysate passes through two different adsorption columns; protein-bound substances are removed by the diaMARS AC250, containing activated charcoal and anionic substances are removed by the diaMARS IE250, filled with cholestyramine, an anion-exchange resin. The albumin solution is then ready to initiate another detoxifying cycle of the patient's blood that can be sustained until both adsorption columns are saturated, eliminating the need to continuously infuse albumin into the system during treatment (Fig. 1).

 
Figure 1: The MARS system

Results published in the literature with the MARS system

A systematic review of the literature from 1999 to June 2011 was performed in the following databases:

  1. Specialized in systematic reviews: Cochrane Library Plus and NHS Centre database for Reviews and Dissemination (HTA, DARE and NHSEED).
  2. General databases: Medline, Pubmed and Embase.
  3. On-going clinical trials and research project databases: Clinical Trials Registry (National Institutes of Health, EE.UU.) and Health Services Research Projects in Progress.
  4. General web searching engines: Scholar Google.

Effects of MARS treatment on Hepatic Encephalopathy (HE)

Hepatic encephalopathy (HE) represents one of the more serious extrahepatic complications associated with liver dysfunction.[63][64] Neuro-psychiatric manifestations of HE affect consciousness and behaviour.

Evidence suggests that HE develops as some neurotoxins and neuro active substances, produced after hepatocellular breakdown, accumulates in the brain as a consequence of a portosystemic shunt and the limited detoxification capability of the liver. Substances involved are ammonia, manganese, aromatic aminoacids, mercaptans, phenols, medium chain fatty acids, bilirubin, endogenous benzodiazepines, etc.
The relationship between ammonia neurotoxicity and HE was first described in animal studies by Pavlov et al.[65]
Subsequently, several studies in either animals or humans have confirmed that, a ratio in ammonia concentration higher than 2 mM between the brain and blood stream, causes HE, and even a comatose state when the value is greater than 5 mM. Some investigators have also reported a decrease in serum ammonia following a MARS treatment (Table 3).

Table 3. Clinical studies showing some improvement in the treatment of HE following a MARS treatment
References Nº patients Age (years)
[aver. ±SD]
Treatment

Hours/patient

Ammonia Pre-MARS

(μg/dl)

Ammonia Post-MARS

(μg/dl)

p
Awad et al.[66] (1999) 9 38±5 73.2 130 64 <0.05
Novelli et al.[67] (2002) 10 42±12 51.2 247 126 <0.003
Schmidt et al.[68] (2001) 8 43±5 10.0 150 121 <0.05
Sorkyne et al.[69] (2001) 8 47±16 28.1 280 65 <0.005

Manganese and copper serum levels are increased in patients with either acute or acute on chronic liver failure. Nevertheless, only in those patients with chronic hepatic dysfunction, a bilateral magnetic resonance alteration on Globos Pallidus is observed,[70] probably because this type of patients selectively shows higher cerebral membrane permeability. Imbalance between aromatic and branched chain aminoacids (Fischer index), traditionally involved in HE genesis,[71][72][73] can be normalized following a MARS treatment. The effects are noticeable even after 3 hours of treatment and this reduction in the Fisher index is accompanied with an improvement in the HE.[74]

Novelli G et al.[75] published their three years experience on MARS analyzing the impact of the treatment in the cerebral level for 63 patients reporting an improvement in Glasgow Coma Score (GCS) for all observed in all patients. In the last 22 patients, cerebral perfusion pressure was monitored by Doppler (mean flow velocity in middle cerebral artery), establishing a clear relationship between a clinical improvement (especially neurological) and an improvement in arterial cerebral perfusion. This study confirms other results showing similar increments in cerebral perfusion in patients treated with MARS.[68]

More recently, several studies have shown a significant improvement of HE in patients treated with MARS. In the studies by Heemann et al.[76] and Sen et al.[77] an improvement in HE was considered when encephalopathy grade was reduced by one or more grades vs. basal values; for Hassenein et al., in their randomized controlled trial, improvement was considered when a decrease of two grades was observed.[78] In the latter, 70 patients with acute on chronic liver failure and encephalopathy grade III and IV were included. Likewise, Kramer et al.[79] estimated an HE improvement when an improvement in peak N70 latency in electroencephalograms was observed. Sen et al.44 observed a significant reduction in Child-Pugh Score (p<0,01) at 7 days following a MARS treatment, without any significant change in the controls. Nevertheless, when they looked at the Model for End-Stage Liver Disease Score (MELD), a significant reduction in both groups, MARS and controls, was recorded (p<0,01 y p<0,05, respectively). Likewise, in several case series, an improvement in HE grade with MARS therapy is also reported.[80][81][82][83][84][85][86][87][88]

Effects of MARS Treatment on Unstable Hemodynamics

Hemodynamic instability is often associated with acute liver insufficiency, as a consequence of endogenous accumulation of vasoactive agents in the blood. This is characterized by a systemic vasodilatation, a decrease of systemic vascular resistance, arterial hypotension, and an increase of cardiac output that gives rise to a hyperdynamic circulation.
During MARS therapy, systemic vascular resistance index and mean arterial pressure have been shown to increase and show improvement.[80][82][84][89][90]
Schmidt et al.[91] reported the treatment of 8 patients, diagnosed with acute hepatic failure, that were treated with MARS for 6 hours, and were compared with a control group of 5 patients to whom ice pads were applied to match the heat loss produced in the treatment group during the extracorporeal therapy. They analyzed hemodynamic parameters in both groups hourly. In the MARS group, a statistically significant increase of 46% on systemic vascular resistance was observed (1215 ± 437 to 1778 ± 710 dinas x s x cm−5 x m−2) compared with a 6% increase in the controls. Mean arterial pressure also increased (69 ± 5 to 83 ± 11 mmHg, p< 0.0001) in the MARS group, whereas no difference was observed in the controls. Cardiac output and heart rate also decreased in the MARS group as a consequence of an improvement in the hyperdynamic circulation. Therefore, it was shown that a statistically significant improvement was obtained with MARS when compared with the SMT.

Catalina et al.[92] have also evaluated systemic and hepatic hemodynamic changes produced by MARS therapy. In 4 patients with acute decompensation of chronic liver disease, they observed after MARS therapy, an attenuation of hyperdynamic circulation and a reduction in the portal pressure gradient was measured. Results are summarized in table 4.

Table 4: Hemodinamic parameters during MARS treatment

PARAMETER BASE LINE AFTER 1st TREAT. AFTER 2nd TREAT.
MAP(mmHg) 77.8 ± 11.7 82.7 ± 11.7 84.2 ± 8
WHVP(mmHg) 40.7 ± 5.6 34 ± 9.6 37.3 ± 5.5
FHVP(mmHg) 17.7 ± 7.4 16.7 ± 7.5 17 ± 3.6
HVPG(mmHg) 23 ± 7.0 17.3 ± 9.9 20.3 ± 5.5
PAP(mmHg) 23.7 ± 7.3 22 ± 4.8 15.7 ± 4.1
PCP(mmHg) 17.2 ± 8.3 14.5 ± 2.9 9.7 ± 7.0
rPAP (mmHg) 11.2 ± 4.6 10 ± 4.2 7.3 ± 7.6
CO(l/min) 11.2 ± 1.6 10 ± 2.8 9.4 ± 2.1
SVRI (dinas x seg/cm5) 478.5 ± 105 514 ± 104.7 622 ± 198
Abbreviations: MAP= mean arterial pressure; WHVP= wedged hepatic venous pressure;FHVP= free hepatic venous pressure; HVPG= hepatic venous pressure gradient; PAP= pulmonary arterial pressure; PCP= pulmonar capillary pressure; rPAP= right pulmonary arterial pressure; CO= cardiac output; SVRI= systemic vascular resistance index.


There are other studies also worth mentioning with similar results: Heemann et al.[76] and Parés et al.[93] among others. Dethloff T et al.[94] concluded that there is a statistically significant improvement favourable to MARS in comparison with Prometheus system (Table 5).

Table 5. Changes in hemodynamic parameters with MARS therapy
MARS System MAP SVRI CO BR
Mitzner et al.[95] (2000) ↑NS - - -
Heemann et al.[76] (2002) ↑S - - -
Schmidt et al.[91] (2003) ↑S ↑S ↓S ↓S
Laleman et al.[90] (2006) ↑S ↑S
Dethloff et al.[94] (2008) ↑NS
Abbreviations: MAP: Mean arterial pressure; SVRI: Systemic vascular resistance index; CO: Cardiac output; BR: Beat rate; S: Statistically significant, p< 0.05; NS: not statistically significant. ↑: Increase; ↓:Decrease; ↔: No changes.

Effects of MARS Treatment on Renal Function

Hepatorenal syndrome is one of the more serious complications in patients with an acute decompensation of cirrhosis and increased portal hypertension. It is characterized by hemodynamic changes in splanchnic, systemic and renal circulation. Splanchnic vasodilatation triggers the production of endogenous vasoactive substances that produce renal vasoconstriction and low glomerular filtration rate, leading to oliguria with a concomitant reduction in creatinine clearance. Renal insufficiency is always progressive with a very poor prognosis,[95][96] with survival at 1 and 2 months of 20 and 10% respectively.

Pierre Versin[97] is one of the pioneers in the study of hepatorenal syndrome in patients with liver impairment. Great efforts have been made trying to improve the prognosis of this type of patient; however, few have solved the problem. Orthotopic liver transplantation is the only treatment that has shown to improve acute and chronic complications derived from severe liver insufficiency. Today it is possible to combine albumin dialysis with continuous veno-venous hemodialfiltration, which provides a greater expectation for these patients[98] by optimization of their clinical status.

MARS treatment lowers serum urea and creatinine levels improving their clearance,[90][91][92][94] and even favors resolution of hepatorenal syndrome.[76][83][84][89][99] Results are confirmed in a randomized controlled trial published by Mitzner et al..[95] in which 13 patients diagnosed with hepatorenal syndrome type I were treated with MARS therapy. Mean survival was 25,2±34,6 days in the MARS group compared to 4,6±1,8 days observed in the controls in whom hemodiafiltration and standard care (SMT) was applied. This resulted in a statistically significance difference in survival at 7 and 30 days (p<0.05). Authors concluded that MARS therapy, applied to liver failure patients (Child-Pugh C and UNOS 2A scores) who develop hepatorenal syndrome type I, prolonged survival compared to patients treated with SMT.

Although mechanisms explaining previous findings are not yet fully understood, it has been reported that there was a decrease in plasma rennin concentrations in patients diagnosed with acute on chronic liver failure with renal impairment that were treated with MARS. Likewise, other studies have suggested some efficacy for MARS in the treatment of hepatorenal syndrome.[100][101][102]
However, other references have been published that do not show efficacy in the treatment of these types of patients with MARS therapy. Khuroo et al..[103] published a metaanalysis based in 4 small RCT's and 2 non RCT's in patients diagnosed with ACLF, concluding that MARS therapy would not bring any significant increment on survival compared with SMT. Another observational study in 6 patients with cirrhosis, refractory ascitis and hepatorenal syndrome type I, not responding to vasoconstrictor therapy, showed no impact on hemodynamics following MARS therapy; however authors concluded that MARS therapy could effectively serve as bridge to liver transplantation.[85][104]

Effects of MARS Treatment on Biochemical Parameters

Total bilirubin was the only parameter analyzed in all trials that was always reduced in the groups of patients treated with MARS; Banayosy et al..[105] measured bilirubin levels 14 days after since MARS therapy was terminated and observed a consistent, significant decrease not only for bilirubin but also for creatinine and urea (Table 6).

Table 6: Biochemical parameters analysis with MARS therapy
MARS study Controls (n) MARS Group (n) Pathology MARS Intervention Bilirubin Creatinine Albumin Urea ALT AST Ammonia BUN
Mitzner et al. (2000)[95] 8 5 ACLF (OH) 6 h x 3 d ↓S ↓S ↑NS - - - - -
Heemann et al. (2002)[76] 12 12 ACLF 6 h x 3 d ↓* ↓S - - - - - -
Sen et al. (2004)[77] 9 9 ACLF+HE (OH) 8 h x 7 d

4 sessions

↓S ↓S - - - -
Laleman et al.. (2006)[90] 6 6 SHR 6-8 h x 10 d ↓S
Hassanein et al.. (2007)[78] 31 39 ACLF+HE (III / IV) 6 h x 5 d ↓S ↓S - - - - ↓S ↓S
Schmidt et al.. (2003)[91] 5 8 ALF 6 h ↓S ↓S ↓S ↓S -
El Banayosy et al.. (2004)[105] 13 14 ALF 8 h x 3 d ↓S ↓S - - - ↓S
Dethloff et al.. (2008)[94] 8:8 Prometheus 8 ACLF - ↓S ↓S - - - - - -
Montejo et al. (2009)[106] 26 19 IHA/GD/ACLF 8 h x 3 d ↓S ↓S ↓S ↓S ↓S - -
Abbreviations; ALF= acute liver failure; ACLF= acute on chronic liver failure; GD= graft dysfunction; HE= hepatic encephalopathy; AST: aspartate amino transferase; BUN: blood urea nitrogen; NS: not significant; S: statistically significant (p<0,05); ↓decrease; ↑ increase; ↔ no change; ALT: alanine amino transferase; h: hours; d: days
  • Primary goal in the Heemann trial was to achieve a bilirubin level beloiw 15 mg/dl during three consecutive days, which was observed in 42% of the patients treated with MARS® compared to 17% in the control group.


Impact of MARS therapy on plasma biliary acids levels was evaluated in 3 studies. In the study from Stadbauer et al..,[107] that was specifically addressing the topic, it is reported that MARS and Prometheus systems lower to the same extent biliary acids plasma concentration. Heemann et al..[76] and Laleman et al..[90] have also published a significant improvement for these organic ions.

Effects of MARS Treatment on Pruritus

Pruritus is one of the most common clinical manifestations in cholestasis liver diseases and one of the most distressing symptoms in patients with chronic liver disease caused by viral hepatitis C. Many hypothesis have been formulated to explain physio pathogenesis of such manifestation, including incremental plasma concentration of biliary acids, abnormalities in the bile ducts,[108] increased central neurotransmitters coupling opioid receptors,[109][110] etc..... Despite the number of historical drugs used, individually or combined (exchange resins, hydrophilic biliary acids, antihistamines, antibiotics, anticonvulsants, opioid antagonists), there are reported cases of intractable or refractory pruritus with a dramatic reduction in patients’ quality of life (i.e. sleep disorders, depression, suicide attempts...).[111][112] Intractable pruritus can be an indication for liver transplantation.

The MARS indication for intractable pruritus is therapeutically an option that has shown to be beneficial for patients in desperate cases, although at high cost.[113][114][115][116] In several studies, it was confirmed that after MARS treatments, patients remain free from pruritus for a period of time ranging from 6 to 9 months.[116] Nevertheless, some authors have concluded that besides the good results found in the literature, application of MARS therapy in refractory pruritus requires larger evidence.[114]

Effects of MARS Treatment on Drugs and Poisons clearance

Pharmacokinetics and pharmacodynamics for a majority of drugs can be significantly be modified with liver failure, affecting the therapeutic approach and potential toxicity of the drugs. In these type of patients, Child-Pugh score represents a poor prognostic factor to assess the metabolic capacity of the failing liver.

  • Metabolic performance of the liver depends on several factors:
  • Hepatic flow rate
  • Cytochrome P-450 enzimatic activity
  • Albumin affinity for the drug
  • Extrahepatic clearance for the drug

In patients with hepatic failure, drugs that are only metabolized in the liver, accumulate in the plasma right after they are administered, and therefore it is needed to modify drug dosing in both, concentration and time intervals, to lower the risk of toxicity. It is also necessary to adjust the dosing for those drugs that are exclusively metabolized by the liver, and have low affinity for prioteins and high distribution volume, such as fluoroquinolones (Levofloxacin and Ciprofloxacin).[117][118][119][120]

Extracorporeal detoxification with albumin dialysis increases the clearance of drugs that are bound to plasmatic proteins (Table 7).

Table 7: Drugs Highly Bound to Proteins
NSAIDs Antibiotics Anti-H2 Statins Antipsychotics
Diclofenac Cephazoline Omeprazole Clofibrate Chlorpromazine
Ibuprofen Cefoperazone Antifungals Lovastatin Haloperidol
Indomethacin Ceftriaxone Amphotericine B Simvastatine Chemotherapy Drugs
Ketoprofen Cloxacilin Itraconazole Fluvastatine Chlorambucil
Naproxen Oxacilin Ketoconazole Hypotensors Etoposide
Phenylbutazone Clindamycin Barbiturates Hydralacine Melphalan
Piroxicam Erythromycin Thiopental Irbesartan D-penicillamine
Anaesthetics Teicoplanin Benzodiazepines Losartan Tamoxifen
Bupivacaine Rifampicin Clonazepam Misoprostol Opioids
Calcium Anthagonists Rifabutin Diazepam Prazosin Fentanyl
Nimodipine Antidepresants Flunitrazepam Valsartan Methadone
Nifedipine Amitriptyline Flurazepam ACE's Antipsychotics
Nitrendipine Desipramine Lorazepam Fosinopropil Clozapine
Antiarrhythmics Imipramine Midazolam Quinapril Fluoxetine
Amiodarone Nortriptyline Nitrazepam Immunosuppressors Risperidone
Quinidine Anticoagulants Oxacepam Ciclosporin Sertraline
Propafenone Warfarin Temazepam Tacrolimus Zolpidem
Lidocaine Antiemetics Beta Blockers Methylprednisolone Others
Digoxin Ondansetron Carvedilol Prednisone Celecoxib
Verapamil Antiepilectics Propanolol Budesonide Diphenhydramine
Oral Antidiabetics Carbamazepine Diuretics - Ethinylestradiol
Glibenclamide Valproic acid Furosemide - Mefloquine
- - Spironolactone - Paclitaxel
- - Torsemide - -

Effects of MARS on Survival

In the meta-analysis published by Khuroo et al..[103] which included 4 randomized trials[76][91][95][105] an improvement in survival for the patients with liver failure treated with MARS, compared with SMT, was not observed.

However, neither in the extracorporeal liver support systems review by the Cochrane[121] (published in 2004), nor the meta-analysis by Kjaergard et al..[122] was a significance difference in survival found for patients diagnosed with ALF treated with extracorporeal liver support systems. Nevertheless, these reviews included all kind of liver support systems and used a heterogeneous type of publication ( abstracts, clinical trials, cohort, etc.).

There is literature showing favorable results in survival for patients diagnosed with ALF, and treated with MARS., In a randomized controlled trial, Salibà et al..[123] studied the impact on survival of MARS therapy for patients with ALF, waiting on the liver transplant list. Forty-nine patients received SMT and 53 were treated with MARS. They observed that patients that received 3 or more MARS sessions showed a statistically significance increase in transplant-free survival compared with the others patients of the study. Notably, 75% of the patients underwent liver transplantation in the first 24 hours after inclusion in the waiting list, and besides the short exposure to MARS therapy, some patients showed a better survival trend compared to controls, when they were treated with MARS prior to the transplant.

In a case-controlled study by Montejo et al..[106] it was reported that MARS treatment do not decrease mortality directly; however, the treatment contributed to significantly improve survival in patients that were transplanted. In studies by Mitzner et al..[95] and Heemann et al..[76] they were able to show a significance statistical difference in 30-day survival for patients in the MARS group. However, El Banayosy et al..[105] and Hassanein et al..[78] noticed a non significant improvement in survival, probably because of the short number of patients included in the trials. In the majority of available MARS studies published with patients diagnosed with ALF, either transplanted or not, survival was greater in the MARS group with some variations according to the type of trial, ranging from 20-30%,[124][125] and 60-80%.[85][126][127][128] Data is summarized in Tables 8, 9 and 10.

Table 8: Studies of MARS Therapy
MARS Study Pathology Nº Patients Type of Study Level of Evidence Study Groups Analysed Variables
MARS Controls
Mitzner et al.[95] (2000) ACLF 13 RCT (LN) III 18 5 HD 1, 3, 4
Heemann et al.[76] (2002) ACLF 24 RCT (LN) III 12 12 SMT 1, 2, 3, 4, 5
Sen et al.[77] (2004) ACLF 18 RCT (LN) III 9 9 SMT 1, 3, 4, 5
Hassanein et al.[78] (2007) ACLF 70 RCT (HN) II 39 31 SMT 1, 2, 3, 5
Schmidt et al.[91] (2003) ALF 13 RCT (LN) III 8 5 SMT 1, 3, 4
El Banayosy et al.[105] (2004) ALF 27 RCT (LN) III 14 13 HD 1, 3
Montejo et al.[106] (2009) ACLF / ALF 45 Case-control - 19 26 SMT 1, 2, 3, 4, 5
MARS-Prometheus Studies
Evenepoel et al..[129] (2006) ACLF 18 Case serie VIII Comparative series 1, 2, 3
Faenza et al..[130] (2008) ACLF 57 Case serie VIII Comparative series 1, 2, 5
Krisper et al..[131] (2005) ACLF 10 RCT (LN) III Crossover 1, 2, 3
Laleman et al..[90] (2006) ACLF 18 RCT (LN) III 6 6 2, 3, 4
Stadlbauer et al..[132] (2006) ACLF 8 RCT (LN) III Crossover 1, 2, 3
Dethloff et al..[94] (2008) ESLD 24 RCT (LN) III 8 8 1, 2, 3, 4
Abbreviations; ACLF: Acute on chronic liver failure; ALF: Acute liver failure; ESLD: End-stage liver disease; RCT: Randomized controlled trial; RCT(LN): RCT, low number of patients; RCT (HN): RCT, high number of patients; HD: Hemodialysis; SMT: Standard medical treatment; 1: Survival/mortality; 2: Security; 3: Biochemical parameters; 4: hemodynamic parameters; 5: Clinical parameters.


Table 9: Survival with MARS Therapy
Study Pathology Controlled Study Follow-up Survival (%)
Faenza et al.[130] (2008) ACLF No Bridging to LTx 7/10 (70)
3 months 2/6 (33)
Krisper et al.[131](2005) ACLF Yes (crossover) 30 days 4/9 (44)
Stadlbauer et al.[132] (2006) ACLF Yes (crossover) 30 days 4/8 (50)
Laleman et al.[90] (2006) ACLF Yes 7 days 6/6 (100)
Dethloff et al.[94] (2008) ESLD Yes 6 months 5/8 (63)
Abbreviations; ALF: Acute liver failure; ACLF: Acute on chronic liver failure; ESLD: End-stage liver disease; Not statistically significance.


Table 10: Mortality in Selected Studies
Study N Mortality

MARS (%)

Mortality

Controls

Time Odds Ratio (OR)
Acute on Chronic Liver Failure
Mitzner et al.[95] 13 63 100 7 days 0.63
Heemann et al.[76] 24 50 67 6 months 0.75
Sen et al.[77] 18 56 56 3 months 1
Hassanein et al.[78] 70 49 55 NR 0.89
Dethloff et al.[94] 24 25 50 6 months 0.50
Acute Liver Failure
Schmidt et al.[91] 13 38 40 NR 0.94
El Banayosy et al.[105] 27 50 69 NR 0.72
Abbreviations; NR: Not reported.

For patients diagnosed with acute on chronic liver failure and treated with MARS therapy, clinical trial results showed a not statistically significant reduction in mortality (odds ratio [OR] =0,78; confident interval [CI] =95%: 0,58 – 1,03; p= 0,1059, Figure 3)

 
Figure 3: Meta-analysis showing the effect on survival of patients with ACLF treated with MARS therapy


A non-statistically significant reduction of mortality was shown in patients with ALF treated with MARS (OR = 0,75 [CI= 95%, 0,42 – 1,35]; p= 0,3427). (Figure 4)

 
Figure 4. Meta-analysis showing the effect on survival of patients with ALF treated with MARS therapy.


Combined results yielded a non-significant reduction on mortality in patients treated with MARS therapy. However, the low number of patients included in each of the studies may be responsible for not being able to achieve enough statistical power to show differences between both treatment groups. Moreover, heterogeneity for the number of MARS sessions and severity of liver disease of the patients included, make it very difficult for the evaluation of MARS impact on survival.

Recently, a meta-analysis on survival in patients treated with an extra-hepatic therapy has been published.[133] Searching strategies yielded 74 clinical trials: 17 randomized controlled trials, 5 case control and 52 cohort studies. Eight studies were included in the meta-analysis: three addressing acute liver failure, one with MARS therapy[105] and five addressing acute on chronic, being four MARS related.[76][77][78][95] Authors concluded that extra-hepatic detoxifying systems improve survival for acute liver insufficiency, whereas results for acute decompensation of chronic liver diseases suggested a non significant survival benefit. Also, due to an increased demand for liver transplantation together with an augmented risk of liver failure following large resections, development of detoxifying extrahepatic systems are necessary.

Safety Aspects

Safety, defined as presence of adverse events, is evaluated in few trials. Adverse events in patients receiving MARS therapy are similar to those in the controls with the exception of thrombocytopenia and hemorrhage that seems to occur more frequently with the MARS system.[134]

Heemann et al.[76] reported two adverse events most probably MARS related: fever and sepsis, presumably originated at the catheter.

In the study by Hassanein et al.,[78] two patients in the MARS group abandoned the study owing to hemodynamic instability, three patients required larger than average platelets transfusion and three more patients presented gastrointestinal bleeding.

Laleman et al..[90] detected one patient with thrombocytopenia in both the MARS and Prometheus treatment groups, and an additional patient with clotting of the dialysis circuit and hypotension, only in the Prometheus group.

Kramer et al.. (Biologic-DT)[79] wrote about 3 cases with disseminated intravascular coagulation in the interventional group, two of them with fatal outcomes.

Mitzner et al..[95] described, among patients treated with MARS, a thrombocytopenia case and a second patient with chronic hepatitis B, who underwent TIPS placement on day 44 after randomization and died on day 105 of multiorgan failure, as a consequence of complications related to the TIPS procedure.

Montejo et al..[106] showed that MARS is an easy technique, without serious adverse events related to the procedure, and also easy to implement in ICU settings that are used to renal extracorporeal therapies.

The MARS International Registry, with data from more than 500 patients (although sponsored by the manufacturer), shows that the adverse effects observed are similar to the control group. However, in these severely ill patients it is difficult to distinguish between complications of the disease itself and side effects attributable to the technique.

Health Economics

Only three Studies addressing cost-effectivenenss of MARS therapy have been found. Hassanein et al.[135] analysed costs of randomized patients with ACLF receiving MARS therapy or standard medical care. They used the study published in 2001 by Kim et al.[136] describing the impact of complications in hospitalization costs in patients diagnosed with alcoholic liver failure. Cost of 11 patients treated with standard medical care (SMT) were compared to those that received MARS, in addition to SMT (12 patients). In the MARS group, there was less in-hospital mortality and complications related to the disease, with a remarkable reduction in cost which compensated the MARS related expenditure (Table 11).

Table 11. Analysis of Complications According to the Modality of treatment.
MARS GROUP

n=12

CONTROL GROUP

n=11

In-hospital mortality 1/12 6/11
Worsening of hepatic encephalopathy (grade 4) 0/12 3/11
Worsening of renal function (Hepatorenal Syndrome) 1/12 7/11
Ascites 0/12 1/11
Variceal bleeding 0/12 1/11
Severe hypotension 2/12 3/11
Electrolyte disorders 4/12 10/11
Coagulopathy 4/12 3/11

There were 5 survivors in the control group, with a cost per patient of $35.904, whereas in the MARS group, 11 patients out of 12 survived with a cost per patient of $32.036 which represents a $4000 savings per patient in favors of the MARS group. Hessel et al.[137] published a 3-year follow-up of a cohort of 79 patients with ACLF, of whom 33 received MARS treatments and 46 received SMT. Survival was 67% for the MARS group and 63% for the controls, that was reduced to 58 and 35% respectively at one year follow-up, and then 52 and 17% at three years.

Hospitalization costs for the MARS treated group were greater than that for the controls (€31,539 vs. €7,543) and similarly direct cost at 3-year follow-up (€8,493 vs. €5,194). Nevertheless, after adjusting mortality rate, the annual cost per patient was €12,092 for controls and €5,827 for MARS group; also in the latter, they found an incremental cost-effectiveness ratio of 31.448 € per life-year gained (LYG) and an incremental costs per QALY gained of 47171 €.

Two years later, same authors published the results of 149 patients diagnosed with ACLF.[138] There were 67 patients (44,9%) treated with MARS and 82 patients (55,1%) were allocated to receive SMT. Mean survival time was 692 days in the MARS group (33% at 3 years) and 453 days in the controls (15% at 3 years); the results were significant (p=0,022). Differences in average cost was €19,853 (95% IC: 13.308-25.429): 35.639 € for MARS patients and 15.804 € for the control group. Incremental cost per LYG was 29.985 € (95% IC: 9.441-321.761) and €43,040 (95% IC: 13.551-461.856) per quality-adjusted life years (QALY).

Liver support systems, such as MARS, are very important to stabilize patients with acute or acute on chronic liver failure and avoid organ dysfunction, as well as a bridge-to-transplant. Although initial in-hospital costs are high, they are worth for the favorable outcome.

MARS Therapy Indications

Acute on Chronic Liver Failure

Etiology:

  • Chronic viral hepatitis[139]
  • Alcoholic liver disease[140][141]
  • Autoimmune disease[142]
  • Metabolic disease such as hemochromatosis
  • Idiopathic Cirrhosis

Goals of MARS Therapy

  • Re-compensation of previous chronic state.
  • Prolong survival time and bridge to urgent or elective transplant
  • Pre-transplant optimization of the patient

MARS Therapy Indication

  • Bilirubin > 15 mg/dl (255 μmol/L), not responding to standard medical care alter 3 days
  • Renal dysfunction or hepatorenal syndrome.
  • Hepatic encephalopathy ≥ II

Treatment Schedule:

  • 3 to 5 eight-hour treatment sessions on consecutive days
  • Continuous treatment with hemodynamic instability (in any case, treatment kit must be replaced every 24 hours)

Acute Liver failure

Etiology:

  • Viral infection[67][91][143]
  • Poisoning (paracetamol overdose, mushrooms)[126][144][145][146][147]
  • Multiorgan dysfunction (severe sepsis)
  • Vascular diseases (Budd Chiari syndrome)
  • Hypoxic hepatitis[148]
  • Liver failure during pregnancy or Reye syndrome
  • Unknown etiology

Goals of MARS Therapy

  • Native liver recovery.
  • Bridging to liver transplant
  • Pre-transplant optimization of the patient.

MARS Therapy Indication

  • King's College or Clichy criteria for liver transplantation
  • Hepatic encephalopathy ≥ II
  • Increased intracraneal pressure
  • Acute hypoxic hepatitis with bilirubin > 8 mg/dl (100 μmol/L)
  • Renal dysfunction or hepatorenal syndrome
  • Progressive intrahepatic cholestasis
  • Fulminant Wilson disease
  • Acute liver dysfunction following paracetamol overdose

Treatment Schedule:

  • 3 to 5 eight-hour treatment sessions in consecutives days
  • Hypoxic hepatitis. 3 eight-hour treatment sessions in consecutives days
  • Paracetamol overdose: 3 to 5 twenty four-hour treatment sessions
  • Mushroom poisoning: 3 to 5 twenty four-hour treatment sessions
  • Fulminant Wilson: minimum 5 twenty four-hour treatment sessions owing to copper saturation of the treatment kit
  • Drug overdose: 3 to 5 eight-hour treatment sessions in consecutives days

MARS in Graft Dysfucntion After Liver Transplant

[149][150]

Etiology:

  • Graft damage during preparation and transportation
  • Infection
  • Hepatotoxic drugs
  • Graft rejection
  • Technical complications (vascular, biliary)
  • Recurrence of primary disease

Goals of MARS Therapy

  • Recovery and prevention of re-transplantation
  • • Prolong survival time and stabilize the patient to receive a re-transplant if the above goal is not achieved

MARS Therapy Indication

  • Primary graft dysfunction
  • Hepatic encephalopathy ≥ II
  • Increased intracranial pressure
  • Renal dysfunction or hepatorenal syndrome.
  • Progressive intrahepatic cholestasis

Treatment Schedule:

  • 3 to 5 eight-hour treatment sessions on consecutive days
  • Continuous treatment with hemodynamic instability (in any case, treatment kit must be replaced every 24 hours)

MARS in liver Failure after Liver Surgery

[151]

Etiology:

  • Liver Resection in hepatocellular carcinoma
  • Transarterial Chemoembolization (TACE)
  • Partial resection in living donor transplantation
  • Other surgical interventions

Goals of MARS Therapy

  • Recovery until hepatic regeneration

MARS Therapy Indication

  • Hepatic encephalopathy ≥ II
  • Renal dysfunction or hepatorenal syndrome.
  • Progressive intrahepatic cholestasis

Treatment Schedule:

  • 3 to 5 eight-hour treatment sessions on consecutive days
  • Continuous treatment with hemodynamic instability (in any case, treatment kit must be replaced every 24 hours)

MARS for intractable pruritus in Cholestasis

[99][139][142][152]

Etiology:

  • Primary biliary cirrhosis (PBC), primary sclerosing cholangitis (PSC)
  • Benign intrahepatic cholestasis (BIC)
  • Biliary Atresia

Goals of MARS Therapy

  • Attenuate pruritus symptoms and improve patients’ quality of life

MARS Therapy Indication

  • Pruritus not responding to SMT

Treatment Schedule:

  • 3 to 5 eight-hour treatment sessions in consecutives days
  • Repeat treatment when symptoms reoccur

MARS Therapy Contraindications

Same contraindications as with any other extracorporeal treatment may be applied to MARS therapy.

  • Unstable hemodynamics with mean arterial pressure (MAP)< 55 mmHg despite vasoconstrictors administration
  • Uncontrolled hemorrhage
  • Severe coagulopathy
  • Severe thrombocytopenia

Treatment Parameters

Blood Flow

The trend is to use high flow rates, although it is determined by the technical specifications of the combined machine and catheters’ size

Intermittent treatments:

  • Without renal dysfunction, it is recommended a blood and albumin flow rates ranging from 150 to 250 mL/min

Continuous treatments:

  • With or without renal impairment it is recommended to use flow rates from 100 to 150 mL/min.

Dyalisate Flow Rate

Intermittent treatments:

  • Without renal impairment: 1800 a 3000 mL/hour
  • With renal impairment: 3000 a 6000 mL/hour

Continuous treatments:

  • Recommended flow rate: 1000 to 2000 mL/hour.

Replacement Flow Rate

  • According to medical criteria and same as in CVVHD

Heparin Anticoagulation

Similarly to CVVHD, it depends on previous patient's coagulation status. In many cases it will not be needed, unless the patient presents a PTT inferior to 160 seconds. In patients with normal values, a bolus of 5000 to 10000 IU of heparin could be administered at the commencement of the treatment, followed by a continuous perfusion, to keep PTT in ratios from 1,5 to 2,5 or 160 to 180 seconds.

Monitoring

A biochemical analysis is recommended (liver and kidney profile, ionic, glucose) together with a hemogram at the end of first session and before starting the following one.

Coagulation analysis must be also performed before starting the session to adjusting heparin dose.

In case that medication susceptible to be eliminated by MARS is being administered, it is also recommended to monitor their levels in blood

End of the Session

  • Once the treatment is finalized, blood should be returned following the unit procedure,

and both catheter's lumens heparinized

  • For the next session a new kit must be used
  • For continuous treatments, kit must be changed by a new one every 24 hours
  • Treatment must be stopped before schedule owing to the particular circumstances listed below:
  1. MAP inferior to 40 mmHg at least for 10 minutes
  2. Air embolism of the extracorporeal circuit
  3. Transmembrane pressure (TMP) greater than 600 mmHg.
  4. Blood leak detection in the albumin circuit
  5. Disseminated intravascular coagulation (DIC)
  6. Severe active hemorrhage

FDA Clearance (US only)

Federal Drug Administration (FDA) cleared, in a document dated on May 27, 2005, MARS therapy for the treatment of drug overdose and poisoning. The only requirement is that the drug or poison must be susceptible to be dialysed and removed by activated charcoal or anionic exchange resins.

More recently, on December 17, 2012, MARS therapy has been cleared by the FDA for the treatment of hepatic encephalopathy due to a decompensation of a chronic liver disease Clinical trials conducted with MARS treatment in HE patients having a decompensation of chronic liver disease demonstrated a transient effect from MARS treatments to significantly decrease their hepatic encephalopathy scores by at least 2 grades compared to standard medical therapy (SMT).

The MARS is not indicated as a bridge to liver transplant. Safety and efficacy has not been demonstrated in controlled, randomized clinical trials.

The effectiveness of the MARS device in patients that are sedated could not be established in clinical studies and therefore cannot be predicted in sedated patients

LiverNet

The LiverNet is a database dedicated to the liver diseases treated with the support of extracorporeal therapies. To date, the most currently used system is the Molecular Adsorbent Recirculating System (MARS), which is based on the selective removal of albumin bound molecules and toxins from the blood in patients with acute and acute-on-chronic liver failure. The purpose is to register prospectively all patients treated worldwide with the MARS system in order to:

  1. Improve our understanding of the clinical course, pathophysiology and treatment of these diseases
  2. Evaluate the clinical impact of MARS therapy on the course of the disease in different specific indication
  3. Increase the knowledge in this extremely innovative area, a basis for an improvement of liver support devices and the treatment of these patients in the next future

The liverNet is an eCRF database (www.livernet.net) using a SAS platform that allows major advantages for the centres including the automatic calculations of most liver rand ICU scoring systems, instant queries online, instant export of all patients included in the database of each centre to an Excel file for direct statistical analysis and finally instant online statistical analysis of selective data decided by the scientific committee. Therefore, the LiverNet is an important tool to progress in the knowledge of liver support therapies.

See also

References

  1. ^ Pless, G. (2007). "Artificial and bioartificial liver support". Organogenesis. 3 (1): 20–24. doi:10.4161/org.3.1.3635. PMC 2649611. PMID 19279696.
  2. ^ Shakil, OA; Kramer D; Mazariegos GV; Fung JJ; Rakela J (2000). "Acute liver failure: clinical features, outcome analysis, and applicability of prognostic criteria". Liver Transpl. 6 (2): 163–169. doi:10.1002/lt.500060218. PMID 10719014. S2CID 33789452.
  3. ^ Jalan, R; Williams R (2002). "Acute on chronic liver failure. Pathophysiological basis of therapeutic options". Blood Purif. 20 (3): 252–261. doi:10.1159/000047017. PMID 11867872. S2CID 36924439.
  4. ^ Stravitz, RT (2008). "Critical management decisions in patients with acute liver failure". Chest. 134 (5): 1092–1102. doi:10.1378/chest.08-1071. PMID 18988787.
  5. ^ Auzinger, G; Wendon, J (April 2008). "Intensive care management of acute liver failure". Current Opinion in Critical Care. 14 (2): 179–88. doi:10.1097/MCC.0b013e3282f6a450. PMID 18388681. S2CID 1289311.
  6. ^ Sen, S; Williams R; Jalan R (2005). "Emerging indications for albumin dialysis". Am. J. Gastroenterol. 100 (2): 468–475. doi:10.1111/j.1572-0241.2005.40864.x. PMID 15667509. S2CID 20937240.
  7. ^ Allen, JW; Hassanein, T; Bhatia, SN (September 2001). "Advances in bioartificial liver devices". Hepatology. 34 (3): 447–55. doi:10.1053/jhep.2001.26753. PMID 11526528. S2CID 6852149.
  8. ^ Sussman, NL; Chong, MG; Koussayer, T; He, DE; Shang, TA; Whisennand, HH; Kelly, JH (July 1992). "Reversal of fulminant hepatic failure using an extracorporeal liver assist device". Hepatology. 16 (1): 60–5. doi:10.1002/hep.1840160112. PMID 1618484. S2CID 45920140.
  9. ^ Stange, J; Ramlow, W; Mitzner, S; Schmidt, R; Klinkmann, H (September 1993). "Dialysis against a recycled albumin solution enables the removal of albumin-bound toxins". Artificial Organs. 17 (9): 809–13. doi:10.1111/j.1525-1594.1993.tb00635.x. PMID 8240075.
  10. ^ Demetriou AA, Brown RS Jr, Busuttil RW, Fair J, McGuire BM, Rosenthal P, Am Esch JS 2nd, Lerut J, Nyberg SL, Salizzoni M, Fagan EA, de Hemptinne B, Broelsch CE, Muraca M, Salmeron JM, Rabkin JM, Metselaar HJ, Pratt D, De La Mata M, McChesney LP, Everson GT, Lavin PT, Stevens AC, Pitkin Z, Solomon BA (May 2004). "Prospective, randomized, multicenter, controlled trial of a bioartificial liver in treating acute liver failure". Annals of Surgery. 239 (5): 660–7, discussion 667–70. doi:10.1097/01.sla.0000124298.74199.e5. PMC 1356274. PMID 15082970.
  11. ^ Mazariegos GV, Patzer JF 2nd, Lopez RC, Giraldo M, Devera ME, Grogan TA, Zhu Y, Fulmer ML, Amiot BP, Kramer DJ (March 2002). "First clinical use of a novel bioartificial liver support system (BLSS)". American Journal of Transplantation. 2 (3): 260–6. doi:10.1034/j.1600-6143.2002.20311.x. PMID 12096789.
  12. ^ Falkenhagen, D; Strobl, W; Vogt, G; Schrefl, A; Linsberger, I; Gerner, FJ; Schoenhofen, M (January 1999). "Fractionated plasma separation and adsorption system: a novel system for blood purification to remove albumin bound substances". Artificial Organs. 23 (1): 81–6. doi:10.1046/j.1525-1594.1999.06292.x. PMID 9950184.
  13. ^ Xue, YL; Zhao, SF; Luo, Y; Li, XJ; Duan, ZP; Chen, XP; Li, WG; Huang, XQ; Li, YL; Cui, X; Zhong, DG; Zhang, ZY; Huang, ZQ (December 2001). "TECA hybrid artificial liver support system in treatment of acute liver failure". World Journal of Gastroenterology. 7 (6): 826–9. doi:10.3748/wjg.v7.i6.826. PMC 4695603. PMID 11854910.
  14. ^ Morsiani E, Pazzi P, Puviani AC, Brogli M, Valieri L, Gorini P, Scoletta P, Marangoni E, Ragazzi R, Azzena G, Frazzoli E, Di Luca D, Cassai E, Lombardi G, Cavallari A, Faenza S, Pasetto A, Girardis M, Jovine E, Pinna AD (March 2002). "Early experiences with a porcine hepatocyte-based bioartificial liver in acute hepatic failure patients". The International Journal of Artificial Organs. 25 (3): 192–202. doi:10.1177/039139880202500305. PMID 11999191. S2CID 196422022.
  15. ^ Sauer, IM; Goetz, M; Steffen, I; Walter, G; Kehr, DC; Schwartlander, R; Hwang, YJ; Pascher, A; Gerlach, JC; Neuhaus, P (May 2004). "In vitro comparison of the molecular adsorbent recirculation system (MARS) and single-pass albumin dialysis (SPAD)". Hepatology. 39 (5): 1408–14. doi:10.1002/hep.20195. PMID 15122770.
  16. ^ Mundt, A; Puhl, G; Müller, A; Sauer, I; Müller, C; Richard, R; Fotopoulou, C; Doll, R; Gäbelein, G; Höhn, W; Hofbauer, R; Neuhaus, P; Gerlach, J (June 2002). "A method to assess biochemical activity of liver cells during clinical application of extracorporeal hybrid liver support". The International Journal of Artificial Organs. 25 (6): 542–8. doi:10.1177/039139880202500609. PMID 12117294. S2CID 38686470.
  17. ^ van de Kerkhove MP, Di Florio E, Scuderi V, Mancini A, Belli A, Bracco A, Dauri M, Tisone G, Di Nicuolo G, Amoroso P, Spadari A, Lombardi G, Hoekstra R, Calise F, Chamuleau RA (October 2002). "Phase I clinical trial with the AMC-bioartificial liver". The International Journal of Artificial Organs. 25 (10): 950–9. doi:10.1177/039139880202501009. PMID 12456036. S2CID 196427668.
  18. ^ Rozga, J (September 2006). "Liver support technology--an update". Xenotransplantation. 13 (5): 380–9. doi:10.1111/j.1399-3089.2006.00323.x. PMID 16925661. S2CID 7562668.
  19. ^ Sussman, NL; Gislason, GT; Conlin, CA; Kelly, JH (May 1994). "The Hepatix extracorporeal liver assist device: initial clinical experience". Artificial Organs. 18 (5): 390–6. doi:10.1111/j.1525-1594.1994.tb02221.x. PMID 8037614.
  20. ^ Ellis, AJ; Hughes, RD; Wendon, JA; Dunne, J; Langley, PG; Kelly, JH; Gislason, GT; Sussman, NL; Williams, R (December 1996). "Pilot-controlled trial of the extracorporeal liver assist device in acute liver failure". Hepatology. 24 (6): 1446–51. doi:10.1002/hep.510240625. PMID 8938179. S2CID 22557163.
  21. ^ Phua, J; Lee, KH (April 2008). "Liver support devices". Current Opinion in Critical Care. 14 (2): 208–15. doi:10.1097/MCC.0b013e3282f70057. PMID 18388685. S2CID 30297978.
  22. ^ Aron, Jonathan; Agarwal, Banwari; Davenport, Andrew (2016-04-02). "Extracorporeal support for patients with acute and acute on chronic liver failure". Expert Review of Medical Devices. 13 (4): 367–380. doi:10.1586/17434440.2016.1154455. ISSN 1743-4440. PMID 26894968. S2CID 13059506.
  23. ^ "Artificial Liver Used After Removal of Organ". The New York Times. Associated Press. 1993-05-19. ISSN 0362-4331. Retrieved 2020-02-06.
  24. ^ "Medical Miracles". PEOPLE.com. Retrieved 2020-02-06.
  25. ^ a b "Best Inventions of 2001". Time. 2001-11-19.
  26. ^ a b Tilles A, Berthiaume F, Yarmush M, Tompkins R, Toner M (2002). "Bioengineering of liver assist devices". Hepatobiliary Pancreat Surg. 9 (6): 686–696. doi:10.1007/s005340200095. PMID 12658402.
  27. ^ Allen J, Hassanein T, Bhatia S (2001). "Advances in bioartificial liver devices". Hepatology. 34 (3): 447–55. doi:10.1053/jhep.2001.26753. PMID 11526528. S2CID 6852149. Free Full Text.
  28. ^ Strain A, Neuberger J (2002). "A bioartificial liver--state of the art". Science. 295 (5557): 1005–9. Bibcode:2002Sci...295.1005S. doi:10.1126/science.1068660. PMID 11834813. S2CID 30433745.
  29. ^ "Current Work on the Bioartificial Liver".
  30. ^ a b He, Yu-Ting; Qi, Ya-Na; Zhang, Bing-Qi; Li, Jian-Bo; Bao, Ji (2019-07-21). "Bioartificial liver support systems for acute liver failure: A systematic review and meta-analysis of the clinical and preclinical literature". World Journal of Gastroenterology. 25 (27): 3634–3648. doi:10.3748/wjg.v25.i27.3634. ISSN 1007-9327. PMC 6658398. PMID 31367162.
  31. ^ Tsiaoussis, J (January 2001). "Which hepatocyte will it be? Hepatocyte choice for bioartificial liver support systems". Liver Transplantation. 7 (1): 2–10. doi:10.1053/jlts.2001.20845. PMID 11150414.
  32. ^ Werner, Andreas; Duvar, Sevim; Müthing, Johannes; Büntemeyer, Heino; Lünsdorf, Heinrich; Strauss, Michael; Lehmann, Jürgen (2000). "Cultivation of immortalized human hepatocytes HepZ on macroporous CultiSpher G microcarriers". Biotechnology and Bioengineering. 68 (1): 59–70. doi:10.1002/(SICI)1097-0290(20000405)68:1<59::AID-BIT7>3.0.CO;2-N. ISSN 1097-0290. PMID 10699872.
  33. ^ "University of Minnesota Bioartificial Liver: How it Works".
  34. ^ Wung, Nelly; Acott, Samuel M.; Tosh, David; Ellis, Marianne J. (December 2014). "Hollow fibre membrane bioreactors for tissue engineering applications". Biotechnology Letters. 36 (12): 2357–2366. doi:10.1007/s10529-014-1619-x. ISSN 0141-5492. PMID 25064452. S2CID 10046204.
  35. ^ a b c d e Jain, Era; Damania, Apeksha; Shakya, Akhilesh Kumar; Kumar, Anupam; Sarin, Shiv K.; Kumar, Ashok (December 2015). "Fabrication of macroporous cryogels as potential hepatocyte carriers for bioartificial liver support". Colloids and Surfaces B: Biointerfaces. 136: 761–771. doi:10.1016/j.colsurfb.2015.10.012. PMID 26519938.
  36. ^ Lozinsky, V. I. (May 2008). "Polymeric cryogels as a new family of macroporous and supermacroporous materials for biotechnological purposes". Russian Chemical Bulletin. 57 (5): 1015–1032. doi:10.1007/s11172-008-0131-7. ISSN 1066-5285. S2CID 95352668.
  37. ^ Erdem, Ahmet; Ngwabebhoh, Fahanwi Asabuwa; Yildiz, Ufuk (2016). "Fabrication and characterization of soft macroporous Jeffamine cryogels as potential materials for tissue applications". RSC Advances. 6 (113): 111872–111881. Bibcode:2016RSCAd...6k1872E. doi:10.1039/C6RA22523C. ISSN 2046-2069.
  38. ^ Ingavle, Ganesh C.; Baillie, Les W.J.; Zheng, Yishan; Lis, Elzbieta K.; Savina, Irina N.; Howell, Carol A.; Mikhalovsky, Sergey V.; Sandeman, Susan R. (May 2015). "Affinity binding of antibodies to supermacroporous cryogel adsorbents with immobilized protein A for removal of anthrax toxin protective antigen". Biomaterials. 50: 140–153. doi:10.1016/j.biomaterials.2015.01.039. PMID 25736504.
  39. ^ Derazshamshir, Ali; Baydemir, Gözde; Andac, Müge; Say, Rıdvan; Galaev, Igor Yu; Denizli, Adil (2010-03-15). "Molecularly Imprinted PHEMA-Based Cryogel for Depletion of Hemoglobin from Human Blood". Macromolecular Chemistry and Physics. 211 (6): 657–668. doi:10.1002/macp.200900425.
  40. ^ Koshy, Sandeep T.; Zhang, David K.Y.; Grolman, Joshua M.; Stafford, Alexander G.; Mooney, David J. (January 2018). "Injectable nanocomposite cryogels for versatile protein drug delivery". Acta Biomaterialia. 65: 36–43. doi:10.1016/j.actbio.2017.11.024. PMC 5716876. PMID 29128539.
  41. ^ Li, Jieliang; Pan, Jilun; Zhang, Liguo; Yu, Yaoting (June 2003). "Culture of hepatocytes on fructose-modified chitosan scaffolds". Biomaterials. 24 (13): 2317–2322. doi:10.1016/S0142-9612(03)00048-6. PMID 12699669.
  42. ^ Ito, Y.; Eguchi, S.; Kamohara, Y.; Inuo, H.; Yamanouchi, K.; Okudaira, S.; Yanaga, K.; Furui, J.; Kanematsu, T. (April 2004). "Influence of Serum from Rats with Fulminant Hepatic Failure on Hepatocytes in a Bioartificial Liver System". The International Journal of Artificial Organs. 27 (4): 303–310. doi:10.1177/039139880402700406. ISSN 0391-3988. PMID 15163064. S2CID 35856518.
  43. ^ a b Damania, Apeksha; Hassan, Mohsin; Shirakigawa, Nana; Mizumoto, Hiroshi; Kumar, Anupam; Sarin, Shiv K.; Ijima, Hiroyuki; Kamihira, Masamichi; Kumar, Ashok (February 2017). "Alleviating liver failure conditions using an integrated hybrid cryogel based cellular bioreactor as a bioartificial liver support". Scientific Reports. 7 (1): 40323. Bibcode:2017NatSR...740323D. doi:10.1038/srep40323. ISSN 2045-2322. PMC 5227920. PMID 28079174.
  44. ^ a b Demetriou, Achilles A.; Brown, Robert S.; Busuttil, Ronald W.; Fair, Jeffrey; McGuire, Brendan M.; Rosenthal, Philip; Am Esch, Jan Schulte; Lerut, Jan; Nyberg, Scott L.; Salizzoni, Mauro; Fagan, Elizabeth A. (May 2004). "Prospective, Randomized, Multicenter, Controlled Trial of a Bioartificial Liver in Treating Acute Liver Failure". Annals of Surgery. 239 (5): 660–670. doi:10.1097/01.sla.0000124298.74199.e5. ISSN 0003-4932. PMC 1356274. PMID 15082970.
  45. ^ a b c Thompson, Julie; Jones, Natasha; Al‐Khafaji, Ali; Malik, Shahid; Reich, David; Munoz, Santiago; MacNicholas, Ross; Hassanein, Tarek; Teperman, Lewis; Stein, Lance; Duarte‐Rojo, Andrés (March 2018). "Extracorporeal cellular therapy (ELAD) in severe alcoholic hepatitis: A multinational, prospective, controlled, randomized trial". Liver Transplantation. 24 (3): 380–393. doi:10.1002/lt.24986. ISSN 1527-6465. PMC 5873437. PMID 29171941.
  46. ^ "Vital Therapies falls 88%, ditches liver treatment after phase 3 fail". FierceBiotech. 12 September 2018. Retrieved 2021-08-24.
  47. ^ "Google Scholar". scholar.google.com.
  48. ^ "Advanced Search | Cochrane Library".
  49. ^ Peszynski, P; Klammt, S; Peters, E; Mitzner, S; Stange, J; Schmidt, R (2002). "Albumin dialysis: single pass vs. recirculation (MARS)". Liver. 22 Suppl 2: 40–2. doi:10.1034/j.1600-0676.2002.00007.x. PMID 12220302.
  50. ^ Chawla, LS; Georgescu, F; Abell, B; Seneff, MG; Kimmel, PL (March 2005). "Modification of continuous venovenous hemodiafiltration with single-pass albumin dialysate allows for removal of serum bilirubin". American Journal of Kidney Diseases. 45 (3): e51–6. doi:10.1053/j.ajkd.2004.11.023. PMID 15754264.
  51. ^ O'Grady J (June 2006). "Personal view: current role of artificial liver support devices". Aliment. Pharmacol. Ther. 23 (11): 1549–57. doi:10.1111/j.1365-2036.2006.02931.x. PMID 16696802. S2CID 31513268.
  52. ^ van de Kerkhove MP, Hoekstra R, Chamuleau RA, van Gulik TM (August 2004). "Clinical application of bioartificial liver support systems". Ann. Surg. 240 (2): 216–30. doi:10.1097/01.sla.0000132986.75257.19. PMC 1356396. PMID 15273544.
  53. ^ Neuberger J (January 2005). "Prediction of survival for patients with fulminant hepatic failure". Hepatology. 41 (1): 19–22. doi:10.1002/hep.20562. PMID 15690476.
  54. ^ Sauer IM, Goetz M, Steffen I, et al. (May 2004). "In vitro comparison of the molecular adsorbent recirculation system (MARS) and single-pass albumin dialysis (SPAD)". Hepatology. 39 (5): 1408–14. doi:10.1002/hep.20195. PMID 15122770.
  55. ^ Rifai K, Ernst T, Kretschmer U, et al. (December 2003). "Prometheus—a new extracorporeal system for the treatment of liver failure". J. Hepatol. 39 (6): 984–90. doi:10.1016/S0168-8278(03)00468-9. PMID 14642616.
  56. ^ "DIALIVE Concept".
  57. ^ "Yaqrit". yaqrit.com. Retrieved 2019-11-04.
  58. ^ Stange J, Mitzner SR, Risler T, Erley CM, Lauchart W, Goehl H, Klammt S, Peszynski P, Freytag J, Hickstein H, Löhr M, Liebe S, Schareck W, Hopt UT, Schmidt R (April 1999). "Molecular adsorbent recycling system (MARS): clinical results of a new membrane-based blood purification system for bioartificial liver support". Artificial Organs. 23 (4): 319–30. doi:10.1046/j.1525-1594.1999.06122.x. PMID 10226696.
  59. ^ Kapoor, D (December 2002). "Molecular adsorbent recirculating system: albumin dialysis-based extracorporeal liver assist device". Journal of Gastroenterology and Hepatology. 17 Suppl 3: S280–6. doi:10.1046/j.1440-1746.17.s3.14.x. PMID 12472950.
  60. ^ Davenport, A (Jun 1, 2003). "Extracorporeal support for patients with hepatic failure". Hemodialysis International. 7 (3): 256–63. doi:10.1046/j.1492-7535.2003.00046.x. PMID 19379373. S2CID 23916381.
  61. ^ Sechser, A; Osorio, J; Freise, C; Osorio, RW (May 2001). "Artificial liver support devices for fulminant liver failure". Clinics in Liver Disease. 5 (2): 415–30. doi:10.1016/s1089-3261(05)70172-0. PMID 11385970.
  62. ^ Stange, J; Mitzner, S (November 1996). "A carrier-mediated transport of toxins in a hybrid membrane. Safety barrier between a patients blood and a bioartificial liver". The International Journal of Artificial Organs. 19 (11): 677–91. doi:10.1177/039139889601901109. PMID 8970836.
  63. ^ Mitzner, S; Loock, J; Peszynski, P; Klammt, S; Majcher-Peszynska, J; Gramowski, A; Stange, J; Schmidt, R (December 2002). "Improvement in central nervous system functions during treatment of liver failure with albumin dialysis MARS--a review of clinical, biochemical, and electrophysiological data". Metabolic Brain Disease. 17 (4): 463–75. doi:10.1023/A:1021986624600. PMID 12602522. S2CID 19491464.
  64. ^ Butterworth, RF (2003). "Role of circulating neurotoxins in the pathogenesis of hepatic encephalopathy: potential for improvement following their removal by liver assist devices". Liver International. 23 Suppl 3: 5–9. doi:10.1034/j.1478-3231.23.s.3.1.x. PMID 12950954. S2CID 25292258.
  65. ^ Matthews, SA (1922). "Ammonia, a causative factor in meat poisoning in Eck fistula dos". Am J Physiol. 59: 459–460.
  66. ^ Awad, SS; Sawada, S; Soldes, OS; Rich, PB; Klein, R; Alarcon, WH; Wang, SC; Bartlett, RH (Jan–Feb 1999). "Can the clearance of tumor necrosis factor alpha and interleukin 6 be enhanced using an albumin dialysate hemodiafiltration system?". ASAIO Journal. 45 (1): 47–9. doi:10.1097/00002480-199901000-00011. PMID 9952006.
  67. ^ a b Novelli, G; Rossi, M; Pretagostini, R; Poli, L; Novelli, L; Berloco, P; Ferretti, G; Iappelli, M; Cortesini, R (2002). "MARS (Molecular Adsorbent Recirculating System): experience in 34 cases of acute liver failure". Liver. 22 Suppl 2: 43–7. doi:10.1034/j.1600-0676.2002.00008.x. PMID 12220303.
  68. ^ a b Schmidt, LE; Svendsen, LB; Sørensen, VR; Hansen, BA; Larsen, FS (August 2001). "Cerebral blood flow velocity increases during a single treatment with the molecular adsorbents recirculating system in patients with acute on chronic liver failure". Liver Transplantation. 7 (8): 709–12. doi:10.1053/jlts.2001.26059. PMID 11510016.
  69. ^ Sorkine P, Ben Abraham R, Szold O, Biderman P, Kidron A, Merchav H, Brill S, Oren R (July 2001). "Role of the molecular adsorbent recycling system (MARS) in the treatment of patients with acute exacerbation of chronic liver failure". Critical Care Medicine. 29 (7): 1332–6. doi:10.1097/00003246-200107000-00006. PMID 11445681. S2CID 42225349.
  70. ^ Spahr, L; Butterworth, RF; Fontaine, S; Bui, L; Therrien, G; Milette, PC; Lebrun, LH; Zayed, J; Leblanc, A; Pomier-Layrargues, G (November 1996). "Increased blood manganese in cirrhotic patients: relationship to pallidal magnetic resonance signal hyperintensity and neurological symptoms". Hepatology. 24 (5): 1116–20. doi:10.1002/hep.510240523. PMID 8903385.
  71. ^ Bergeron, M; Layrargues, GP; Butterworth, RF (September 1989). "Aromatic and branched-chain amino acids in autopsied brain tissue from cirrhotic patients with hepatic encephalopathy". Metabolic Brain Disease. 4 (3): 169–76. doi:10.1007/bf01000293. PMID 2796871. S2CID 24524207.
  72. ^ Bergeron, M; Swain, MS; Reader, TA; Grondin, L; Butterworth, RF (July 1990). "Effect of ammonia on brain serotonin metabolism in relation to function in the portacaval shunted rat". Journal of Neurochemistry. 55 (1): 222–9. doi:10.1111/j.1471-4159.1990.tb08842.x. PMID 1693945. S2CID 2219787.
  73. ^ Fischer, JE; Funovics, JM; Aguirre, A; James, JH; Keane, JM; Wesdorp, RI; Yoshimura, N; Westman, T (September 1975). "The role of plasma amino acids in hepatic encephalopathy". Surgery. 78 (3): 276–90. PMID 807982.
  74. ^ Loock, J; Stange, J; Mitzner, S; Schmidt, R; Gramowski, A; Schiffmann, D; Weiss, D; Keefer, EW; Gross, GW (June 2001). "Influence of albumin dialysis (MARS) on neuronal network activity in vitro--early results". Zeitschrift für Gastroenterologie. 39 Suppl 2: 28–32. doi:10.1055/s-2001-919055. PMID 16215892.
  75. ^ Novelli, G; Rossi, M; Pretagostini, R; Novelli, L; Poli, L; Ferretti, G; Iappelli, M; Berloco, P; Cortesini, R (2003). "A 3-year experience with Molecular Adsorbent Recirculating System (MARS): our results on 63 patients with hepatic failure and color Doppler US evaluation of cerebral perfusion". Liver International. 23 Suppl 3: 10–5. doi:10.1034/j.1478-3231.23.s.3.4.x. PMID 12950955. S2CID 25173997.
  76. ^ a b c d e f g h i j k l Heemann, U; Treichel, U; Loock, J; Philipp, T; Gerken, G; Malago, M; Klammt, S; Loehr, M; Liebe, S; Mitzner, S; Schmidt, R; Stange, J (October 2002). "Albumin dialysis in cirrhosis with superimposed acute liver injury: a prospective, controlled study". Hepatology. 36 (4 Pt 1): 949–58. doi:10.1053/jhep.2002.36130. PMID 12297843.
  77. ^ a b c d e Sen, S; Davies, NA; Mookerjee, RP; Cheshire, LM; Hodges, SJ; Williams, R; Jalan, R (September 2004). "Pathophysiological effects of albumin dialysis in acute-on-chronic liver failure: a randomized controlled study". Liver Transplantation. 10 (9): 1109–19. doi:10.1002/lt.20236. PMID 15350001.
  78. ^ a b c d e f g Hassanein TI, Tofteng F, Brown RS Jr, McGuire B, Lynch P, Mehta R, Larsen FS, Gornbein J, Stange J, Blei AT (December 2007). "Randomized controlled study of extracorporeal albumin dialysis for hepatic encephalopathy in advanced cirrhosis". Hepatology. 46 (6): 1853–62. doi:10.1002/hep.21930. PMID 17975845.
  79. ^ a b Kramer, L; Gendo, A; Madl, C; Mullen, KD; Kaminski-Russ, K; Sunder-Plassmann, G; Schaffer, A; Bauer, E; Roth, E; Ferenci, P (July 2001). "A controlled study of sorbent suspension dialysis in chronic liver disease and hepatic encephalopathy". The International Journal of Artificial Organs. 24 (7): 434–42. doi:10.1177/039139880102400707. PMID 11510914. S2CID 24205200.
  80. ^ a b Yuan, JZ; Ye, QF; Zhao, LL; Ming, YZ; Sun, H; Zhu, SH; Huang, ZF; Wang, MM (Aug 21, 2006). "Preoperative risk factor analysis in orthotopic liver transplantation with pretransplant artificial liver support therapy". World Journal of Gastroenterology. 12 (31): 5055–9. doi:10.3748/wjg.v12.i31.5055. PMC 4087413. PMID 16937506.
  81. ^ Gaspari, R; Cavaliere, F; Sollazzi, L; Perilli, V; Melchionda, I; Agnes, S; Gasbarrini, A; Avolio, AW (Jan–Feb 2009). "Molecular adsorbent recirculating system (Mars) in patients with primary nonfunction and other causes of graft dysfunction after liver transplantation in the era of extended criteria donor organs". Transplantation Proceedings. 41 (1): 253–8. doi:10.1016/j.transproceed.2008.10.066. hdl:10807/141915. PMID 19249528.
  82. ^ a b Stefoni S, Colì L, Bolondi L, Donati G, Ruggeri G, Feliciangeli G, Piscaglia F, Silvagni E, Sirri M, Donati G, Baraldi O, Soverini ML, Cianciolo G, Boni P, Patrono D, Ramazzotti E, Motta R, Roda A, Simoni P, Magliulo M, Borgnino LC, Ricci D, Mezzopane D, Cappuccilli ML (February 2006). "Molecular adsorbent recirculating system (MARS) application in liver failure: clinical and hemodepurative results in 22 patients". The International Journal of Artificial Organs. 29 (2): 207–18. doi:10.1177/039139880602900207. hdl:11585/24619. PMID 16552668. S2CID 22591855.
  83. ^ a b Di Campli C, Santoro MC, Gaspari R, Merra G, Zileri Dal Verme L, Zocco MA, Piscaglia AC, Di Gioacchino G, Novi M, Santoliquido A, Flore R, Tondi P, Proietti R, Gasbarrini G, Pola P, Gasbarrini A (Jul–Aug 2005). "Catholic university experience with molecular adsorbent recycling system in patients with severe liver failure". Transplantation Proceedings. 37 (6): 2547–50. doi:10.1016/j.transproceed.2005.06.048. PMID 16182739.
  84. ^ a b c Hetz, H; Faybik, P; Berlakovich, G; Baker, A; Bacher, A; Burghuber, C; Sandner, SE; Steltzer, H; Krenn, CG (September 2006). "Molecular adsorbent recirculating system in patients with early allograft dysfunction after liver transplantation: a pilot study". Liver Transplantation. 12 (9): 1357–64. doi:10.1002/lt.20804. PMID 16741899. S2CID 19946716.
  85. ^ a b c Camus, C; Lavoué, S; Gacouin, A; Compagnon, P; Boudjéma, K; Jacquelinet, C; Thomas, R; Le Tulzo, Y (December 2009). "Liver transplantation avoided in patients with fulminant hepatic failure who received albumin dialysis with the molecular adsorbent recirculating system while on the waiting list: impact of the duration of therapy". Therapeutic Apheresis and Dialysis. 13 (6): 549–55. doi:10.1111/j.1744-9987.2009.00708.x. PMID 19954480. S2CID 24898728.
  86. ^ Steiner, C; Mitzner, S (2002). "Experiences with MARS liver support therapy in liver failure: analysis of 176 patients of the International MARS Registry". Liver. 22 Suppl 2: 20–5. doi:10.1034/j.1600-0676.2002.00003.x. PMID 12220298.
  87. ^ Camus C, Lavoué S, Gacouin A, Le Tulzo Y, Lorho R, Boudjéma K, Jacquelinet C, Thomas R (November 2006). "Molecular adsorbent recirculating system dialysis in patients with acute liver failure who are assessed for liver transplantation". Intensive Care Medicine. 32 (11): 1817–25. doi:10.1007/s00134-006-0340-1. PMID 16941171. S2CID 8003937.
  88. ^ Parés, A; Deulofeu, R; Cisneros, L; Escorsell, A; Salmerón, JM; Caballería, J; Mas, A (2009). "Albumin dialysis improves hepatic encephalopathy and decreases circulating phenolic aromatic amino acids in patients with alcoholic hepatitis and severe liver failure". Critical Care. 13 (1): R8. doi:10.1186/cc7697. PMC 2688120. PMID 19175915.
  89. ^ a b Mitzner, SR; Stange, J; Klammt, S; Peszynski, P; Schmidt, R; Nöldge-Schomburg, G (February 2001). "Extracorporeal detoxification using the molecular adsorbent recirculating system for critically ill patients with liver failure". Journal of the American Society of Nephrology. 12 Suppl 17: S75–82. doi:10.1681/ASN.V12suppl_1s75. PMID 11251037.
  90. ^ a b c d e f g h Laleman, W; Wilmer, A; Evenepoel, P; Elst, IV; Zeegers, M; Zaman, Z; Verslype, C; Fevery, J; Nevens, F (2006). "Effect of the molecular adsorbent recirculating system and Prometheus devices on systemic haemodynamics and vasoactive agents in patients with acute-on-chronic alcoholic liver failure". Critical Care. 10 (4): R108. doi:10.1186/cc4985. PMC 1751025. PMID 16859530.
  91. ^ a b c d e f g h Schmidt, LE; Wang, LP; Hansen, BA; Larsen, FS (March 2003). "Systemic hemodynamic effects of treatment with the molecular adsorbents recirculating system in patients with hyperacute liver failure: a prospective controlled trial". Liver Transplantation. 9 (3): 290–7. doi:10.1053/jlts.2003.50051. PMID 12619027.
  92. ^ a b Catalina, MV; Barrio, J; Anaya, F; Salcedo, M; Rincón, D; Clemente, G; Bañares, R (2003). "Hepatic and systemic haemodynamic changes after MARS in patients with acute on chronic liver failure". Liver International. 23 Suppl 3: 39–43. doi:10.1034/j.1478-3231.23.s.3.10.x. PMID 12950960. S2CID 2200601.
  93. ^ Parés, A; Escorsell, A; Cisneros, L (2002). "Effects of the molecular adsorbent recirculating system (MARS) on vasoactive agents and systemic hemodynamics in patients with severe alcoholic hepatitis". 4th ISAD 2002 Abstrac.
  94. ^ a b c d e f g Dethloff, T; Tofteng, F; Frederiksen, HJ; Hojskov, M; Hansen, BA; Larsen, FS (Apr 7, 2008). "Effect of Prometheus liver assist system on systemic hemodynamics in patients with cirrhosis: a randomized controlled study". World Journal of Gastroenterology. 14 (13): 2065–71. doi:10.3748/wjg.14.2065. PMC 2701529. PMID 18395908.
  95. ^ a b c d e f g h i j Mitzner SR, Stange J, Klammt S, Risler T, Erley CM, Bader BD, Berger ED, Lauchart W, Peszynski P, Freytag J, Hickstein H, Loock J, Löhr JM, Liebe S, Emmrich J, Korten G, Schmidt R (May 2000). "Improvement of hepatorenal syndrome with extracorporeal albumin dialysis MARS: results of a prospective, randomized, controlled clinical trial". Liver Transplantation. 6 (3): 277–86. doi:10.1002/lt.500060326. PMID 10827226.
  96. ^ Arroyo, V (May 2000). "New treatments for hepatorenal syndrome". Liver Transplantation. 6 (3): 287–9. doi:10.1053/lv.2000.7569. PMID 10827227. S2CID 32465310.
  97. ^ Versin P, Hirsch-Marie H, Catan R (1962). "Línsufissance rénale circulatoire spntané du cirrhotique. Son évolution". Sem Hôp Paris. 38: 3598–3602.
  98. ^ McIntyre, CW; Fluck, RJ; Freeman, JG; Lambie, SH (November 2002). "Characterization of treatment dose delivered by albumin dialysis in the treatment of acute renal failure associated with severe hepatic dysfunction". Clinical Nephrology. 58 (5): 376–83. doi:10.5414/CNP58376. PMID 12425489.
  99. ^ a b Saich, R; Collins, P; Ala, A; Standish, R; Hodgson, H (May 2005). "Benign recurrent intrahepatic cholestasis with secondary renal impairment treated with extracorporeal albumin dialysis". European Journal of Gastroenterology & Hepatology. 17 (5): 585–8. doi:10.1097/00042737-200505000-00018. PMID 15827452. S2CID 23033545.
  100. ^ Cárdenas, A; Ginès, P (June 2006). "Therapy insight: Management of hepatorenal syndrome". Nature Clinical Practice Gastroenterology & Hepatology. 3 (6): 338–48. doi:10.1038/ncpgasthep0517. PMID 16741553. S2CID 12302583.
  101. ^ Moreau, R; Lebrec, D (2007). "Diagnosis and treatment of acute renal failure in patients with cirrhosis". Best Practice & Research. Clinical Gastroenterology. 21 (1): 111–23. doi:10.1016/j.bpg.2006.10.004. PMID 17223500.
  102. ^ Wong, F (January 2007). "Drug insight: the role of albumin in the management of chronic liver disease". Nature Clinical Practice Gastroenterology & Hepatology. 4 (1): 43–51. doi:10.1038/ncpgasthep0680. PMID 17203088. S2CID 23729017.
  103. ^ a b Khuroo, MS; Khuroo, MS; Farahat, KL (September 2004). "Molecular adsorbent recirculating system for acute and acute-on-chronic liver failure: a meta-analysis". Liver Transplantation. 10 (9): 1099–106. doi:10.1002/lt.20139. PMID 15349999.
  104. ^ Wong, F; Raina, N; Richardson, R (March 2010). "Molecular adsorbent recirculating system is ineffective in the management of type 1 hepatorenal syndrome in patients with cirrhosis with ascites who have failed vasoconstrictor treatment". Gut. 59 (3): 381–6. doi:10.1136/gut.2008.174615. PMID 19710033. S2CID 13455917.
  105. ^ a b c d e f g El Banayosy, A; Kizner, L; Schueler, V; Bergmeier, S; Cobaugh, D; Koerfer, R (Jul–Aug 2004). "First use of the Molecular Adsorbent Recirculating System technique on patients with hypoxic liver failure after cardiogenic shock". ASAIO Journal. 50 (4): 332–7. doi:10.1097/01.MAT.0000131251.88146.CD. PMID 15307543. S2CID 20982595.
  106. ^ a b c d Montejo González JC, Catalán González M, Meneu Díaz JC, Moreno Elola-Olaso A, De la Cruz J, Moreno González E (Mar–Apr 2009). "Artificial liver support system in acute liver failure patients waiting liver transplantation". Hepato-gastroenterology. 56 (90): 456–61. PMID 19579620.
  107. ^ Stadlbauer, V; Krisper, P; Beuers, U; Haditsch, B; Schneditz, D; Jung, A; Putz-Bankuti, C; Holzer, H; Trauner, M; Stauber, RE (Mar–Apr 2007). "Removal of bile acids by two different extracorporeal liver support systems in acute-on-chronic liver failure". ASAIO Journal. 53 (2): 187–93. doi:10.1097/01.mat.0000249852.71634.6c. PMID 17413559. S2CID 8630554.
  108. ^ Lebovics, E; Seif, F; Kim, D; Elhosseiny, A; Dworkin, BM; Casellas, A; Clark, S; Rosenthal, WS (May 1997). "Pruritus in chronic hepatitis C: association with high serum bile acids, advanced pathology, and bile duct abnormalities". Digestive Diseases and Sciences. 42 (5): 1094–9. doi:10.1023/A:1018865809556. PMID 9149069. S2CID 22902185.
  109. ^ Jones, EA; Bergasa, NV (Dec 16, 1992). "The pruritus of cholestasis and the opioid system". JAMA. 268 (23): 3359–62. doi:10.1001/jama.1992.03490230089034. PMID 1333541.
  110. ^ Bergasa, NV; Thomas, DA; Vergalla, J; Turner, ML; Jones, EA (1993). "Plasma from patients with the pruritus of cholestasis induces opioid receptor-mediated scratching in monkeys". Life Sciences. 53 (16): 1253–7. doi:10.1016/0024-3205(93)90569-o. PMID 8412484.
  111. ^ Marchesini G, Bianchi G, Amodio P, Salerno F, Merli M, Panella C, Loguercio C, Apolone G, Niero M, Abbiati R, Italian Study Group for quality of life in cirrhosis (January 2001). "Factors associated with poor health-related quality of life of patients with cirrhosis". Gastroenterology. 120 (1): 170–8. doi:10.1053/gast.2001.21193. PMID 11208726.
  112. ^ Younossi, ZM; Kiwi, ML; Boparai, N; Price, LL; Guyatt, G (February 2000). "Cholestatic liver diseases and health-related quality of life". The American Journal of Gastroenterology. 95 (2): 497–502. doi:10.1111/j.1572-0241.2000.01774.x. PMID 10685757. S2CID 1584814.
  113. ^ Huster, D; Schubert, C; Achenbach, H; Caca, K; Mössner, J; Berr, F (June 2001). "Successful clinical application of extracorporal albumin dialysis in a patient with benign recurrent intrahepatic cholestasis (BRIC)". Zeitschrift für Gastroenterologie. 39 Suppl 2: 13–4. doi:10.1055/s-2001-919024. PMID 16215886.
  114. ^ a b De Simone, P; Van Nuffelen, M; Donckier, V (September 2003). "Use of molecular adsorbent recirculating system for treatment of refractory pruritus". Liver Transplantation. 9 (9): 997–8. doi:10.1002/lt.500090918. PMID 12942466.
  115. ^ Mullhaupt, B; Kullak-Ublick, GA; Ambühl, PM; Stocker, R; Renner, EL (April 2003). "Successful use of the Molecular Adsorbent Recirculating System (MARS) in a patient with primary biliary cirrhosis (PBC) and treatment refractory pruritus". Hepatology Research. 25 (4): 442–446. doi:10.1016/s1386-6346(02)00310-8. PMID 12699855.
  116. ^ a b Macia, M; Avilés, J; Navarro, J; Morales, S; García, J (January 2003). "Efficacy of molecular adsorbent recirculating system for the treatment of intractable pruritus in cholestasis". The American Journal of Medicine. 114 (1): 62–4. doi:10.1016/s0002-9343(02)01354-2. PMID 12543292.
  117. ^ Majcher-Peszynska, J; P. Peszynki; SC Müller (2001). "Drugs in Liver Disease and During Albumin Dialysis-MARS". Z Gastroenterol. 39: 33–35. doi:10.1055/s-2001-919048. PMID 16215894.
  118. ^ Majcher-Peszynska, J; T. Schneider; S. Müller; P. Peszynski; R. Mundowski; A. Berg; S. Klammt; B. Drewelow (2002). "Elimination of Fluoroquinolones during Extracorporeal Albumin Dialysis (MARS)". 4th ISAD (Abstracts).
  119. ^ Majcher-Peszynska, J; S. Klammt; EM. Hehk (2000). "Removal of Albumin Bound Drugs in Albumin Dialysis (MARS)- A New Liver Support System". J. Hepatol. Suppl. 2. 32: 60. doi:10.1016/s0168-8278(00)80562-0.
  120. ^ Pszynski, PM; J. Majcher-Pszynska; J. Stangeet (2000). "Kinetics of Drugs During MARS Treatment". Hepatology. 32: 484A. doi:10.1002/hep.1840380507.
  121. ^ Liu, JP; Gluud, LL; Als-Nielsen, B; Gluud, C (2004). "Artificial and bioartificial support systems for liver failure". Cochrane Database Syst Rev. 1 (1): CD003628. doi:10.1002/14651858.CD003628.pub2. PMC 6991941. PMID 14974025.
  122. ^ Kjaergard, LL; Liu, J; Als-Nielsen, B; Gluud, C (2003). "Artificial and bioartificial support systems for acute and acute-on-chronic liver failure: a systematic review". JAMA. 289 (2): 217–222. doi:10.1001/jama.289.2.217. PMID 12517233.
  123. ^ Saliba, F; Camus, C; Durand, F; et al. (2008). "Randomized controlled multicenter trial evaluating the efficacy and safety of albumin dialysis with MARS in patients with fulminant and subfulminant hepatic failure". Hepatology. 48 (4 (Suppl 1)): 377A. doi:10.1002/hep.22615.
  124. ^ Lai, WK; Haydon, G; Mutimer, D; et al. (2005). "The effect of molecular adsorbent recirculating system on pathophysiological parameters in patients with acute liver failure". Intensive Care Med. 31 (11): 1544–1549. doi:10.1007/s00134-005-2786-y. PMID 16155752. S2CID 6642381.
  125. ^ Lee, KH; Lee, MK; Sutedja, DS; et al. (2005). "Outcome from molecular adsorbent recycling system (MARS) liver dialysis following drug-induced liver failure". Liver Int. 25 (5): 973–977. doi:10.1111/j.1478-3231.2005.01091.x. PMID 16162155. S2CID 44868151.
  126. ^ a b Zhou, XM; Miao, JY; Yang, Y; et al. (2004). "Clinical experience with molecular adsorbent recirculating system (MARS) in patients with drug-induced liver failure". Artificial Organs. 28 (5): 483–486. doi:10.1046/j.1525-1594.2003.00953.x-i1. PMID 15113343.
  127. ^ Kantola, T; Koivusalo, A; Höckerstedt, K; et al. (2008). "The effect of molecular adsorbent recirculating system on survival, native liver recovery, and need for liver transplantation in acute liver failure patients". Transpl Int. 21 (9): 857–866. doi:10.1111/j.1432-2277.2008.00698.x. PMID 18510596.
  128. ^ Camus, C; Lavoue, S; Gacouin, A; et al. (2009). "Liver transplantation avoided in patients with fulminant hepatic failure who received albumin dialysis with the molecular adsorbent recirculating system while on the waiting list: impact of the duration of therapy". Ther Apher Dial. 13 (6): 549–555. doi:10.1111/j.1744-9987.2009.00708.x. PMID 19954480. S2CID 24898728.
  129. ^ Evenepoel, P; Laleman, W; Wilmer, A; Claes, K; Kuypers, D; Bammens, B; Nevens, F; Vanrenterghem, Y (Apr 2006). "Prometheus versus molecular adsorbents recirculating system: comparison of efficiency in two different liver detoxification devices". Artificial Organs. 30 (4): 276–84. doi:10.1111/j.1525-1594.2006.00215.x. PMID 16643386.
  130. ^ a b Faenza S, Baraldi O, Bernardi M, Bolondi L, Coli L, Cucchetti A, Donati G, Gozzetti F, Lauro A, Mancini E, Pinna AD, Piscaglia F, Rasciti L, Ravaioli M, Ruggeri G, Santoro A, Stefoni S (May 2008). "Mars and Prometheus: our clinical experience in acute chronic liver failure". Transplantation Proceedings. 40 (4): 1169–71. doi:10.1016/j.transproceed.2008.03.069. PMID 18555140.
  131. ^ a b Krisper, P; Haditsch, B; Stauber, R; Jung, A; Stadlbauer, V; Trauner, M; Holzer, H; Schneditz, D (Sep 2005). "In vivo quantification of liver dialysis: comparison of albumin dialysis and fractionated plasma separation". Journal of Hepatology. 43 (3): 451–7. doi:10.1016/j.jhep.2005.02.038. PMID 16023249.
  132. ^ a b Stadlbauer, V; Krisper, P; Aigner, R; Haditsch, B; Jung, A; Lackner, C; Stauber, RE (2006). "Effect of extracorporeal liver support by MARS and Prometheus on serum cytokines in acute-on-chronic liver failure". Critical Care. 10 (6): R169. doi:10.1186/cc5119. PMC 1794485. PMID 17156425.
  133. ^ Stutchfield, BM; Simpson, K; Wigmore, SJ (May 2011). "Systematic review and meta-analysis of survival following extracorporeal liver support". The British Journal of Surgery. 98 (5): 623–31. doi:10.1002/bjs.7418. PMID 21462172. S2CID 23318049.
  134. ^ Mitzner, SR; Stange, J; Klammt, S; Koball, S; Hickstein, H; Reisinger, EC (Sep–Oct 2009). "Albumin dialysis MARS: knowledge from 10 years of clinical investigation". ASAIO Journal. 55 (5): 498–502. doi:10.1097/mat.0b013e3181b37d86. PMID 19730006. S2CID 40839375.
  135. ^ Hassanein, T; Oliver, D; Stange, J; Steiner, C (2003). "Albumin dialysis in cirrhosis with superimposed acute liver injury: possible impact of albumin dialysis on hospitalization costs". Liver International. 23 Suppl 3: 61–5. doi:10.1034/j.1478-3231.23.s.3.6.x. PMID 12950963. S2CID 23585670.
  136. ^ Kim WR, Gross JB Jr, Poterucha JJ, Locke GR 3rd, Dickson ER (Jan 2001). "Outcome of hospital care of liver disease associated with hepatitis C in the United States". Hepatology. 33 (1): 201–6. doi:10.1053/jhep.2001.20798. PMID 11124837. S2CID 21550617.
  137. ^ Hessel, FP (Oct 5, 2006). "Economic evaluation of the artificial liver support system MARS in patients with acute-on-chronic liver failure". Cost Effectiveness and Resource Allocation. 4: 16. doi:10.1186/1478-7547-4-16. PMC 1601969. PMID 17022815.
  138. ^ Hessel, FP; Bramlage, P; Wasem, J; Mitzner, SR (Feb 2010). "Cost-effectiveness of the artificial liver support system MARS in patients with acute-on-chronic liver failure". European Journal of Gastroenterology & Hepatology. 22 (2): 213–20. doi:10.1097/meg.0b013e3283314e48. PMID 19773666. S2CID 21972813.
  139. ^ a b Doria, C; Mandalá, L; Smith, J; Vitale, CH; Lauro, A; Gruttadauria, S; Marino, IR; Foglieni, CS; Magnone, M; Scott, VL (Apr 2003). "Effect of molecular adsorbent recirculating system in hepatitis C virus-related intractable pruritus". Liver Transplantation. 9 (4): 437–43. doi:10.1053/jlts.2003.50055. PMID 12682899. S2CID 28738261.
  140. ^ Sen, S; Mookerjee, RP; Cheshire, LM; Davies, NA; Williams, R; Jalan, R (Jul 2005). "Albumin dialysis reduces portal pressure acutely in patients with severe alcoholic hepatitis". Journal of Hepatology. 43 (1): 142–8. doi:10.1016/j.jhep.2005.01.032. PMID 15878216.
  141. ^ Jalan, R; Sen, S; Steiner, C; Kapoor, D; Alisa, A; Williams, R (Jan 2003). "Extracorporeal liver support with molecular adsorbents recirculating system in patients with severe acute alcoholic hepatitis". Journal of Hepatology. 38 (1): 24–31. doi:10.1016/s0168-8278(02)00334-3. PMID 12480556.
  142. ^ a b Manz, T; Ochs, A; Bisse, E; Strey, C; Grotz, W (2003). "Liver support--a task for nephrologists? Extracorporeal treatment of a patient with fulminant Wilson crisis". Blood Purification. 21 (3): 232–6. doi:10.1159/000070695. PMID 12784049. S2CID 45543585.
  143. ^ Chen, S; Zhang, L; Shi, Y; Yang, X; Wang, M (2002). "Molecular Adsorbent Recirculating System: clinical experience in patients with liver failure based on hepatitis B in China". Liver. 22 Suppl 2: 48–51. doi:10.1034/j.1600-0676.2002.00009.x. PMID 12220304.
  144. ^ Sen, S; Ytrebø, LM; Rose, C; Fuskevaag, OM; Davies, NA; Nedredal, GI; Williams, R; Revhaug, A; Jalan, R (Mar 2004). "Albumin dialysis: a new therapeutic strategy for intoxication from protein-bound drugs". Intensive Care Medicine. 30 (3): 496–501. doi:10.1007/s00134-003-2141-0. hdl:1866/24039. PMID 14735236. S2CID 30840312.
  145. ^ Koivusalo, AM; Yildirim, Y; Vakkuri, A; Lindgren, L; Höckerstedt, K; Isoniemi, H (Oct 2003). "Experience with albumin dialysis in five patients with severe overdoses of paracetamol". Acta Anaesthesiologica Scandinavica. 47 (9): 1145–50. doi:10.1034/j.1399-6576.2003.00190.x. PMID 12969110. S2CID 12576160.
  146. ^ Rubik, J; Pietraszek-Jezierska, E; Kamiński, A; Skarzynska, A; Jóźwiak, S; Pawłowska, J; Drewniak, T; Prokurat, S; Grenda, R; Kaliciński, P (Jun 2004). "Successful treatment of a child with fulminant liver failure and coma caused by Amanita phalloides intoxication with albumin dialysis without liver transplantation". Pediatric Transplantation. 8 (3): 295–300. doi:10.1111/j.1399-3046.2004.00170.x. PMID 15176968. S2CID 38538438.
  147. ^ Covic, A; Goldsmith, DJ; Gusbeth-Tatomir, P; Volovat, C; Dimitriu, AG; Cristogel, F; Bizo, A (2003). "Successful use of Molecular Absorbent Regenerating System (MARS) dialysis for the treatment of fulminant hepatic failure in children accidentally poisoned by toxic mushroom ingestion". Liver International. 23 Suppl 3: 21–7. doi:10.1034/j.1478-3231.23.s.3.9.x. PMID 12950957. S2CID 39904706.
  148. ^ Faybik, P; Hetz, H; Krenn, CG; Baker, A; Germann, P; Berlakovich, G; Steininger, R; Steltzer, H (Sep 15, 2003). "Liver support in fulminant liver failure after hemorrhagic shock". Wiener Klinische Wochenschrift. 115 (15–16): 595–8. doi:10.1007/bf03040455. PMID 14531174. S2CID 19746467.
  149. ^ Lahdenperä, A; Koivusalo, AM; Vakkuri, A; Höckerstedt, K; Isoniemi, H (Jan 2005). "Value of albumin dialysis therapy in severe liver insufficiency". Transplant International. 17 (11): 717–23. doi:10.1111/j.1432-2277.2004.tb00500.x. PMID 15580335.
  150. ^ Hommann, M; Kasakow, LB; Geoghegan, J; Kornberg, A; Schotte, U; Fuchs, D; Hermann, J; Zintl, F; Scheele, J (Aug 2002). "Application of MARS artificial liver support as bridging therapy before split liver retransplantation in a 15-month-old child". Pediatric Transplantation. 6 (4): 340–3. doi:10.1034/j.1399-3046.2002.02007.x. PMID 12234277. S2CID 24212977.
  151. ^ van de Kerkhove MP, de Jong KP, Rijken AM, de Pont AC, van Gulik TM (2003). "MARS treatment in posthepatectomy liver failure". Liver International. 23 Suppl 3: 44–51. doi:10.1034/j.1478-3231.23.s.3.2.x. PMID 12950961. S2CID 39425523.
  152. ^ Parés, A; Cisneros, L; Salmerón, JM; Caballería, L; Mas, A; Torras, A; Rodés, J (Jun 2004). "Extracorporeal albumin dialysis: a procedure for prolonged relief of intractable pruritus in patients with primary biliary cirrhosis". The American Journal of Gastroenterology. 99 (6): 1105–10. doi:10.1111/j.1572-0241.2004.30204.x. PMID 15180733. S2CID 22709637.

Further reading

  • Tandon R, Froghi S (September 2020). "Artificial liver support systems". J. Gastroenterol. Hepatol. 36 (5): 1164–1179. doi:10.1111/jgh.15255. PMID 32918840.
  • Sen S, Williams R, Jalan R (February 2005). "Emerging indications for albumin dialysis". Am. J. Gastroenterol. 100 (2): 468–75. doi:10.1111/j.1572-0241.2005.40864.x. PMID 15667509. S2CID 20937240.
  • Evenepoel P, Maes B, Wilmer A, et al. (2003). "Detoxifying capacity and kinetics of the molecular adsorbent recycling system. Contribution of the different inbuilt filters". Blood Purif. 21 (3): 244–52. doi:10.1159/000070697. PMID 12784051. S2CID 12355239.
  • Mitzner S, Klammt S, Stange J, Nöldge-Schomburg GF, Schmidt R (April 2005). "[Extracorporeal blood purification in severe liver failure with the albumin dialysis MARS – impact on relevant intensive care parameters]". Anasthesiol Intensivmed Notfallmed Schmerzther (in German). 40 (4): 199–206. doi:10.1055/s-2004-826116. PMID 15832238.

liver, support, system, this, article, contains, content, that, written, like, advertisement, please, help, improve, removing, promotional, content, inappropriate, external, links, adding, encyclopedic, content, written, from, neutral, point, view, august, 202. This article contains content that is written like an advertisement Please help improve it by removing promotional content and inappropriate external links and by adding encyclopedic content written from a neutral point of view August 2021 Learn how and when to remove this template message A liver support system or diachysis is a type of therapeutic device to assist in performing the functions of the liver Such systems focus either on removing the accumulating toxins liver dialysis or providing additional replacement of the metabolic functions of the liver through the inclusion of hepatocytes to the device bioartificial liver device This system is in trial to help people with acute liver failure ALF or acute on chronic liver failure 1 Liver support systemSpecialtyhepatology edit on Wikidata The primary functions of the liver include removing toxic substances from the blood manufacturing blood proteins storing energy in the form of glycogen and secreting bile The hepatocytes that perform these tasks can be killed or impaired by disease resulting in hepatic insufficiency A sudden onset of life threatening hepatic insufficiency is known as acute liver failure ALF Contents 1 Etymology 2 Liver failure 3 Bioartificial liver devices 3 1 History 3 1 1 Early history 3 1 2 Development 3 2 Use 3 3 Function 3 3 1 Hollow fibre systems 3 3 2 Cryogel Based Systems 3 4 Clinical Studies 3 4 1 HepatAssist 3 4 2 Extracorporeal Liver Assist Device 4 Liver dialysis 4 1 Prognosis survival 4 2 Devices 4 2 1 Single Pass Albumin Dialysis SPAD 4 2 2 Molecular adsorbents recirculation system 4 2 3 Comparing SPAD MARS and CVVHDF 4 2 4 Prometheus 4 2 5 Dialive 5 The MARS System 5 1 MARS System Components 5 2 Results published in the literature with the MARS system 5 3 Effects of MARS treatment on Hepatic Encephalopathy HE 5 4 Effects of MARS Treatment on Unstable Hemodynamics 5 5 Effects of MARS Treatment on Renal Function 5 6 Effects of MARS Treatment on Biochemical Parameters 5 7 Effects of MARS Treatment on Pruritus 5 8 Effects of MARS Treatment on Drugs and Poisons clearance 5 9 Effects of MARS on Survival 5 10 Safety Aspects 5 11 Health Economics 5 12 MARS Therapy Indications 5 12 1 Acute on Chronic Liver Failure 5 12 2 Acute Liver failure 5 12 3 MARS in Graft Dysfucntion After Liver Transplant 5 12 4 MARS in liver Failure after Liver Surgery 5 12 5 MARS for intractable pruritus in Cholestasis 5 13 MARS Therapy Contraindications 5 14 Treatment Parameters 5 15 FDA Clearance US only 6 LiverNet 7 See also 8 References 9 Further readingEtymology EditThe word diachysis derives from the Greek word diaxyshs which means Diffusion The word dialysis derives from the Greek word dialysis which means Dissolution Liver failure Edit In hyperacute and acute liver failure the clinical picture develops rapidly with progressive encephalopathy and multiorgan dysfunction such as hyperdynamic circulation coagulopathy acute kidney injury and respiratory insufficiency severe metabolic alterations and cerebral edema that can lead to brain death 2 3 In these cases the mortality without liver transplantation LTx ranges between 40 80 4 5 LTx is the only effective treatment for these patients although it requires a precise indication and timing to achieve good results Nevertheless due to the scarcity of organs to carry out liver transplantations it is estimated that one third of patients with ALF die while waiting to be transplanted 6 On the other hand a patient with a chronic hepatic disease can suffer an acute decompensation of liver function following a precipitating event such as variceal bleeding sepsis and excessive alcohol intake among others that can lead to a condition referred to as acute on chronic liver failure ACLF Both types of hepatic insufficiency ALF and ACLF can potentially be reversible and liver functionality can return to a level similar to that prior to the insult or precipitating event LTx is the only treatment that has shown an improvement in the prognosis and survival with most severe cases of ALF Nevertheless cost and donor scarcity have prompted researchers to look for new supportive treatments that can act as bridge to the transplant procedure By stabilizing the patient s clinical state or by creating the right conditions that could allow the recovery of native liver functions both detoxification and synthesis can improve after an episode of ALF or ACLF 7 Basically three different types of supportive therapies have been developed bio artificial artificial and hybrid liver support systems Table 2 Table 2 Liver Support Systems Bio artificial Artificial HybridsELAD 8 Extracorporeal liver assist device MARS 9 Molecular adsorbent recirculating system Hepat Assist 10 BLSS 11 Bioartificial Liver Support System Prometheus FPSA 12 Fractionated plasma separation and adsorption system TECLA HALSS 13 TECA Hybrid Artificial Liver Support SystemRFB 14 Radial Flow Bioreactor SPAD 15 Single pass albumin dialysis MELS 16 Modular Extracorporeal Liver SupportAMC BAL 17 Bioartificial Liver SEPET 18 Selective plasma filtration therapy Bioartificial liver devices EditLiver support systemSpecialtyinternal medicine edit on Wikidata Bioartificial liver devices are experimental extracorporeal devices that use living cell lines to provide detoxification and synthesis support to the failing liver Bio artificial liver BAL Hepatassist 2000 uses porcine hepatocytes whereas ELAD system employs hepatocytes derived from human hepatoblastoma C3A cell lines 19 20 Both techniques can produce in fulminant hepatic failure FHF an improvement of hepatic encephalopathy grade and biochemical parameters Nevertheless they are therapies with high complexity that require a complex logistic approach for implementation a very high cost and possible inducement of important side effects such as immunological issues porcine endogenous retrovirus transmission infectious complications and tumor transmigration have been documented Other biological hepatic systems are Bioartificial Liver Support BLSS and Radial Flow Bioreactor RFB Detoxification capacity of these systems is poor and therefore they must be used combined with other systems to mitigate this deficiency Today its use is limited to centers with high experience in their application 21 A bioartificial liver device BAL is an artificial extracorporeal liver support ELS system for an individual who is suffering from acute liver failure ALF or acute on chronic liver failure ACLF The fundamental difference between artificial and BAL systems lies in the inclusion of hepatocytes into the reactor often operating alongside the purification circuits used in artificial ELS systems The overall design varies between different BAL systems but they largely follow the same basic structure with patient blood or plasma flow through an artificial matrix housing hepatocytes Plasma is often separated from the patient s blood to improve efficiency of the system and the device can be connected to artificial liver dialysis devices in order to further increase the effectiveness of the device in filtration of toxins The inclusion of functioning hepatocytes in the reactor allows the restoration of some of the synthetic functions that the patient s liver is lacking 22 History Edit Early history Edit The first bioartificial liver device was developed in 1993 by Dr Achilles A Demetriou at Cedars Sinai Medical Center The bioartificial liver helped an 18 year old southern California woman survive without her own liver for 14 hours until she received a human liver using a 20 inch long 4 inch wide plastic cylinder filled with cellulose fibers and pig liver cells Blood was routed outside the patient s body and through the artificial liver before being returned to the body 23 24 Dr Kenneth Matsumara s work on the BAL led it to be named an invention of the year by Time magazine in 2001 25 Liver cells obtained from an animal were used instead of developing a piece of equipment for each function of the liver The structure and function of the first device also resembles that of today s BALs Animal liver cells are suspended in a solution and a patient s blood is processed by a semipermeable membrane that allow toxins and blood proteins to pass but restricts an immunological response 25 Development Edit Advancements in bioengineering techniques in the decade after Matsumara s work have led to improved membranes and hepatocyte attachment systems 26 As time has progressed the sources of hepatocytes have increased Cell sources now include primary porcine hepatocytes primary human hepatocytes human hepatoblastoma C3A immortalized human cell lines and stem cells 26 Use Edit The purpose of BAL type devices currently is not to permanently replace liver functions but to serve as a supportive device 27 either allowing the liver to regenerate properly upon acute liver failure or to bridge the individual s liver functions until a transplant is possible Function Edit BALs are essentially bioreactors with embedded hepatocytes liver cells that perform the functions of a normal liver They process oxygenated blood plasma which is separated from the other blood constituents 28 Several types of BALs are being developed including hollow fiber systems and flat membrane sheet systems 29 There are various types of hepatocytes that are used in these devices Porcine hepatocytes are often used due to ease of acquisition and cost however they are relatively unstable and carry the risk of cross species disease transmission 30 Primary human hepatocytes sourced from donor organs are the most suitable for use but present several problems in their cost and difficulty to obtain especially with the current lack in transplantable tissue 30 In addition questions have been raised about tissue collected from patients transmitting malignancy or infection via the BAL device Several lines of human hepatocytes are also used in BAL devices including C3A and HepG2 tumour cell lines but due to their origin from hepatomas they possess the potential to pass on malignancy to the patient 31 There is ongoing research into the cultivation of new types of human hepatocytes capable of improved longevity and efficacy in a bioreactor over currently used cell types that do not pose the risk of transfer of malignancy or infection such as the HepZ cell line created by Werner et al 32 Hollow fibre systems Edit One type of BAL is similar to kidney dialysis systems that employ a hollow fiber cartridge Hepatocytes are suspended in a gel solution such as collagen which is injected into a series of hollow fibers In the case of collagen the suspension is then gelled within the fibers usually by a temperature change The hepatocytes then contract the gel by their attachment to the collagen matrix reducing the volume of the suspension and creating a flow space within the fibers Nutrient media is circulated through the fibers to sustain the cells During use plasma is removed from the patients blood The patient s plasma is fed into the space surrounding the fibers The fibers which are composed of a semi permeable membrane facilitate transfer of toxins nutrients and other chemicals between the blood and the suspended cells The membrane also keeps immune bodies such as immunoglobulins from passing to the cells to prevent an immune system rejection 33 Cryogel Based Systems Edit Currently hollow fibre bioreactors are the most commonly accepted design for clinical use due to their capillary network allowing for easy perfusion of plasma across cell populations 34 However these structures have their limitations with convectional transport issues nutritional gradients non uniform seeding inefficient immobilisation of cells and reduced hepatocyte growth restricting their effectiveness in BAL designs 35 Researchers are now investigating the use of cryogels to replace hollow fibres as the cell carrier components in BAL systems Cryogels are super macroporous three dimensional polymers prepared at sub zero temperatures by the freezing of a solution of cryogel precursors and solvent The pores develop during this freezing process as the cryogel solution cools the solvent begins to form crystals This causes the concentration of the cryogel precursors in the solution to increase initiating the cryogelation process and forming the polymer walls As the cryogel warms the solvent crystals thaw leaving cavities that form the pores 36 Cryogel pores range in size from 10 100 µm in size forming an interconnected network that mimics a capillary system with a very large surface area to volume ratio supporting large numbers of immobilised cells Convection mediated transport is also supported by cryogels enabling even distribution of nutrients and metabolite elimination overcoming some of the shortcomings of hollow fibre systems 35 Most importantly cryogel scaffolds demonstrate good mechanical strength and biocompatibility without triggering an immune response improving their potential for long term inclusion in BAL devices or in vitro use 37 Another advantage of cryogels is their flexibility for use in a variety of tasks including separation and purification of substances along with acting as extracellular matrix for cell growth and proliferation Immobilisation of specific ligands onto cryogels enables adsorption of specific substances supporting their use as treatment options for toxins 38 for separation of haemoglobin from blood 39 and as a localised and sustained method for drug delivery 40 Jain et al 35 investigated poly AN co NVP and poly NiPAAm chitosan cryogels as hepatocyte carriers in a BAL system These cryogel polymers were chosen as they maintain a good balance of hydrophilicity hydrophobicity to promote hepatocyte binding Chitosan was added to cryogels as it promotes spheroid formation in hepatocytes an indicator of healthy growth 41 The researchers were able to demonstrate biocompatibility with human HepG2 cells after high density growth for 7 days and with surrounding tissue after implantation into mice They also evaluated ureagenesis and drug detoxification by hepatocytes in the scaffold finding that over time the hepatocytes were able to process ammonia into urea and that CYP450 activity in the cells after 96 hours culturing was equal to that of freshly isolated hepatocytes The researchers then seeded HepG2 cells onto cryogels and incubated for 48 hours before incorporating the cryogels into a BAL bioreactor They circulated plasma isolated from alcoholic ACLF patients for 3 hours through the bioreactor and ran control experiments with cryogels incubated in media only After the experiment they found that the cell seeded bioreactors had reduced the levels of bilirubin and ammonia in the plasma by 58 9 6 7 and 61 2 7 respectively and increased the level of albumin by 31 1 28 compared to the control group This demonstrated that the HepG2 cells in the bioreactor were able to carry out detoxifying and synthetic functions of normal liver cells However after 3 hours of plasma circulation hepatocyte function deteriorated rapidly This was likely due to the toxic effect of ACLF plasma on hepatocytes 42 or nutrient and oxygen depletion The study clearly demonstrated the ability of HepG2 seeded cryogels to detoxify plasma and support liver function however further research is needed into what caused the reduction in function after 3 hours Damania et al 43 took this system a step further utilising HepG2 seeded poly NiPAAm chitosan cryogels in a bioreactor along with an activated charcoal cloth used as a filter As seen with the artificial BAL devices activated charcoal is useful in the filtration of liver toxins including ammonia and bilirubin The researchers then incorporated the bioreactor into a rat model connecting rats with induced liver failure to the system and passing their blood across a plasma separation membrane with the plasma running through the seeded cryogel and charcoal filter They compared results against rats with induced liver failure who were not connected to a bioreactor Measuring levels of ammonia urea bilirubin albumin and AST over a 3 hour time period they found a reduction in bilirubin AST and urea levels and an increase in albumin levels suggesting the bioreactor is functioning in its removal of toxins and synthesis of new proteins In addition they ran the same experiment using an acellular bioreactor and found that bilirubin and AST decreased over time but to a lesser degree than the seeded bioreactor showing that the detoxification results of the seeded reactor were due to a combination of both the HepG2 cells and the activated charcoal cloth However in the cell seeded reactor there was an increase in the levels of ammonia which the researchers stated could be attributed to these toxins having a detrimental effect on the cells over time reducing their function Clinical Studies Edit There have been numerous clinical studies involving hollow fibre bioreactors Overall they show promise but do not provide statistically significant evidence supporting their effectiveness This is generally due to inherent design limitations causing convectional transport issues nutritional gradients non uniform seeding inefficient immobilisation of cells and reduced hepatocyte growth 35 As of writing no cryogel based devices have entered clinical trials However laboratory results have been promising 35 43 and hopefully trials will begin soon HepatAssist Edit The HepatAssist developed at the Cedars Sinai Medical Center is a BAL device containing porcine hepatocytes within a hollow fibre bioreactor These semi permeable fibres act as capillaries allowing the perfusion of plasma through the device and across the hepatocytes surrounding the fibres The system incorporates a charcoal column to act as a filter removing additional toxins from the plasma 44 Demetriou et al 44 carried out a large randomised multicentre controlled trial on the safety and efficacy of the HepatAssist device 171 patients with ALF stemming from viral hepatitis paracetamol overdose or other drug complications primary non function PNF or of indeterminate aetiology were involved in the study and were randomly assigned to either the experimental or control groups Both groups were well balanced in terms of age gender race and aetiology The study found that at the primary end point 30 day post admission mark there was an increased survival rate in BAL patients over control patients 71 vs 62 but the difference was not significant However when patients with PNF are excluded from the results there is a 44 reduction in mortality for BAL treated patients a statistically significant advantage The investigators noted that exclusion of PNF patients is justifiable due to early retransplantation and lack of intercranial hypertension so HepatAssist would give little benefit to this group For the secondary end point of time to death in patients with ALF of known aetiology there was a significant difference between BAL and control groups with BAL patients surviving for longer There was no significant difference for patients of unknown aetiology however The conclusions of the study suggest that such a device has potentially significant importance when used as a treatment measure While the overall findings were not statistically significant when the aetiology of the patients was taken into account the BAL group gained a statistically significant reduction in mortality over the control group This suggests that while the device may not be applicable to patients as an overall treatment for liver dysfunction it can provide an advantage when the heterogeneity of patients is considered and is used with patients of specific aetiology Extracorporeal Liver Assist Device Edit The Extracorporeal Liver Assist Device ELAD is a human cell based treatment system A catheter removes blood from the patient and an ultrafiltrate generator separates the plasma from the rest of the blood This plasma is then run through a separate circuit containing cartridges filled with C3A cells before being returned to the main circuit and re entering the patient C3A cells were chosen for this device due to their possession of anti inflammatory proteins such as IL 1 receptor antagonist along with anti apoptotic and anti oxidative mechanisms that could aid in the reduction of further liver injury from the patients diseased state 45 Thompson et al 45 performed a large open label trial measuring the effectiveness of ELAD on patients with severe alcoholic hepatitis resulting in ACLF Their study involved patients screened at 40 sites across the US UK and Australia and enrolled a total of 203 patients Patients were then randomised into either ELAD n 96 or standard medical care n 107 groups with even distribution for patients in terms of sex MELD score and bilirubin levels Of the 96 patients in the ELAD group 45 completed the full 120 hours of treatment the rest were unable to complete the full regimen due to a variety of reasons including withdrawal of consent or severe adverse events though 37 completed gt 72 hours of treatment with results showing minimal difference in mortality between those receiving either gt 72 hours or the full 120 hours of treatment The study was unable to complete its goal finding no statistically significant improvement in mortality rates for patients that received ELAD treatment over those receiving standard care at 28 and 91 days 76 0 versus 80 4 and 59 4 versus 61 7 respectively Biomarker measurements showed a significantly reduced level of bilirubin and alkaline phosphatase in ELAD patients though neither improvement translated into increased survivability rates Outcomes for patients with MELD score lt 28 showed trends towards improved survival on ELAD whereas those with MELD gt 28 had decreased survivability on ELAD These patients presented with raised creatinine from kidney failure suggesting a reason why ELAD decreased survival chance over standard care Unlike artificial ELS devices and HepatAssist ELAD does not incorporate any filtration devices such as charcoal columns and exchange resins Therefore it cannot replace the filtration capability of the kidneys and cannot compensate for multi organ failure from more severe presentations of ACLF resulting in increased mortality rates While the results of the study cannot provide conclusive evidence to suggest that a BAL device like ELAD improves the outcome of severe ACLF it does suggest that it can aid the survival of patients with a less severe form of the disease In those patients with a MELD lt 28 beneficial effects were seen 2 3 weeks post treatment suggesting that while C3A incorporating BAL devices are unable to provide short term aid like artificial albumin filtration devices they instead provide more long term aid in recovery of the patient s liver 45 A randomized phase 3 trial of the ELAD device in patients with severe alcoholic hepatitis failed to show benefit on overall survival and development was discontinued 46 Liver dialysis EditArtificial liver support systems are aimed to temporally replace native liver detoxification functions and they use albumin as scavenger molecule to clear the toxins involved in the physiopathology of the failing liver Most of the toxins that accumulate in the plasma of patients with liver insufficiency are protein bound and therefore conventional renal dialysis techniques such as hemofiltration hemodialysis or hemodiafiltration are not able to adequately eliminate them Liver dialysis has shown promise for patients with hepatorenal syndrome It is similar to hemodialysis and based on the same principles but hemodialysis does not remove toxins bound to albumin that accumulate in liver failure Like a bioartificial liver device it is a form of artificial extracorporeal liver support 47 48 A critical issue of the clinical syndrome in liver failure is the accumulation of toxins not cleared by the failing liver Based on this hypothesis the removal of lipophilic albumin bound substances such as bilirubin bile acids metabolites of aromatic amino acids medium chain fatty acids and cytokines should be beneficial to the clinical course of a patient in liver failure This led to the development of artificial filtration and absorption devices Liver dialysis is performed by physicians and surgeons and specialized nurses with training in gastroenterological medicine and surgery namely in hepatology the study of the liver alongside their colleagues in the intensive or critical care unit and the transplantation department which is responsible for procuring and implanting a new liver or a part lobe of one if and when it becomes available in time and the patient is eligible Because of the need for these experts as well as the relative newness of the procedure in certain areas it is usually available only in larger hospitals such as level I trauma center teaching hospitals connected with medical schools Between the different albumin dialysis modalities single pass albumin dialysis SPAD has shown some positive results at a very high cost 49 it has been proposed that lowering the concentration of albumin in the dialysate does not seem to affect the detoxification capability of the procedure 50 Nevertheless the most widely used systems today are based on hemodialysis and adsorption These systems use conventional dialysis methods with an albumin containing dialysate that is later regenerated by means of adsorption columns filled with activated charcoal and ion exchange resins At present there are two artificial extracorporeal liver support systems the Molecular Adsorbents Recirculating System MARS from Gambro and Fractionated Plasma Separation and Adsorption FPSA commercialised as Prometheus PROM from Fresenius Medical Care Of the two therapies MARS is the most frequently studied and clinically used system to date Prognosis survival Edit While the technique is in its infancy the prognosis of patients with liver failure remains guarded Liver dialysis currently is only considered to be a bridge to transplantation or liver regeneration in the case of acute liver failure 51 52 53 and unlike kidney dialysis for kidney failure cannot support a patient for an extended period of time months to years Devices Edit Artificial detoxification devices currently under clinical evaluation include the Single Pass Albumin Dialysis SPAD Molecular Adsorbent Recirculating System MARS Prometheus system and Dialive Single Pass Albumin Dialysis SPAD Edit Single pass albumin dialysis SPAD is a simple method of albumin dialysis using standard renal replacement therapy machines without an additional perfusion pump system The patient s blood flows through a circuit with a high flux hollow fiber hemodiafilter identical to that used in the MARS system The other side of this membrane is cleansed with an albumin solution in counter directional flow which is discarded after passing the filter Hemodialysis can be performed in the first circuit via the same high flux hollow fibers Molecular adsorbents recirculation system Edit The Molecular Adsorbents Recirculation System MARS is the best known extracorporal liver dialysis system and has existed for approximately fifteen years It consists of two separate dialysis circuits The first circuit consists of human serum albumin is in contact with the patient s blood through a semipermeable membrane and has two filters to clean the albumin after it has absorbed toxins from the patient s blood The second circuit consists of a hemodialysis machine and is used to clean the albumin in the first circuit before it is recirculated to the semipermeable membrane in contact with the patient s blood Comparing SPAD MARS and CVVHDF Edit SPAD MARS and continuous veno venous haemodiafiltration CVVHDF were compared in vitro with regard to detoxification capacity 54 SPAD and CVVHDF showed a significantly greater reduction of ammonia compared with MARS No significant differences could be observed between SPAD MARS and CVVHDF concerning other water soluble substances However SPAD enabled a significantly greater bilirubin reduction than MARS Bilirubin serves as an important marker substance for albumin bound non water soluble substances Concerning the reduction of bile acids no significant differences between SPAD and MARS were seen It was concluded that the detoxification capacity of SPAD is similar or even higher when compared with the more sophisticated more complex and hence more expensive MARS As albumin dialysis is a costly procedure financial aspects are important For a seven hour treatment with MARS approximately 300 for 600 mL human serum albumin solution 20 1740 for a MARS treatment kit and 125 for disposables used by the dialysis machine have to be spent The cost of this therapy adds up to approximately 2165 Performing SPAD according to the protocol by Sauer et al however requires 1000 mL of human albumin solution 20 at a cost of 500 A high flux dialyzer costing approximately 40 and the tubings 125 must also be purchased The overall costs of a SPAD treatment is approximately 656 30 of the costs of an equally efficient MARS therapy session The expenditure for the MARS monitor necessary to operate the MARS disposables is not included in this calculation Prometheus Edit The Prometheus system Fresenius Medical Care Bad Homburg Germany is a device based on the combination of albumin adsorption with high flux hemodialysis after selective filtration of the albumin fraction through a specific polysulfon filter AlbuFlow It has been studied 55 in a group of eleven patients with hepatorenal syndrome acute on chronic liver failure and accompanying kidney failure The treatment for two consecutive days for more than four hours significantly improved serum levels of conjugated bilirubin bile acids ammonia cholinesterase creatinine urea and blood pH Prometheus was proven to be a safe supportive therapy for patients with liver failure Dialive Edit Dialive Yaqrit Limited London UK incorporates albumin removal and replacement and endotoxin removal It is at Technology readiness level TRL 5 which means it is validated in the disease environment 56 57 The MARS System EditMARS was developed by a group of researchers at the University of Rostock Germany in 1993 and later commercialized for its clinical use in 1999 58 The system is able to replace the detoxification function of the liver while minimizing the inconvenience and drawbacks of previously used devices 59 60 61 In vivo preliminary investigations indicated the ability of the system to effectively remove bilirubin biliary salts free fatty acids and tryptophan while important physiological proteins such as albumin alpha 1 glicoproteine alpha 1 antitrypsin alpha 2 macroglobulin transferrin globulin tyrosine and hormonal systems are unaffected 62 Also MARS therapy in conjunction with CRRT HDF can help clear cytokines acting as inflammatory and immunological mediators in hepatocellular damage and therefore can create the right environment to favour hepatocellular regeneration and recovery of native liver function MARS System Components Edit Combined MARS and PrismaFlex monitors MARS is an extracorporeal hemodialysis system composed of three different circuits blood albumin and low flux dialysis The blood circuit uses a double lumen catheter and a conventional hemodialysis device to pump the patient s blood into the MARS FLUX a biocompatible polysulfone high flux dialyser With a membrane surface area of 2 1 m2 100 nm of thickness and a cut off of 50 KDa the MARSFLUX is essential to retaining the albumin in the dialysate Blood is dialysed against a human serum albumin HSA dialysate solution that allows blood detoxification of both water soluble and protein bound toxins by means of the presence of albumin in the dialysate albumin dialysis The albumin dialysate is then regenerated in a close loop in the MARS circuit by passing through the fibres of the low flux diaFLUX filter to clear water soluble toxins and provide electrolyte acid base balance by a standard dialysis fluid Next the albumin dialysate passes through two different adsorption columns protein bound substances are removed by the diaMARS AC250 containing activated charcoal and anionic substances are removed by the diaMARS IE250 filled with cholestyramine an anion exchange resin The albumin solution is then ready to initiate another detoxifying cycle of the patient s blood that can be sustained until both adsorption columns are saturated eliminating the need to continuously infuse albumin into the system during treatment Fig 1 Figure 1 The MARS system Results published in the literature with the MARS system Edit A systematic review of the literature from 1999 to June 2011 was performed in the following databases Specialized in systematic reviews Cochrane Library Plus and NHS Centre database for Reviews and Dissemination HTA DARE and NHSEED General databases Medline Pubmed and Embase On going clinical trials and research project databases Clinical Trials Registry National Institutes of Health EE UU and Health Services Research Projects in Progress General web searching engines Scholar Google Effects of MARS treatment on Hepatic Encephalopathy HE Edit Hepatic encephalopathy HE represents one of the more serious extrahepatic complications associated with liver dysfunction 63 64 Neuro psychiatric manifestations of HE affect consciousness and behaviour Evidence suggests that HE develops as some neurotoxins and neuro active substances produced after hepatocellular breakdown accumulates in the brain as a consequence of a portosystemic shunt and the limited detoxification capability of the liver Substances involved are ammonia manganese aromatic aminoacids mercaptans phenols medium chain fatty acids bilirubin endogenous benzodiazepines etc The relationship between ammonia neurotoxicity and HE was first described in animal studies by Pavlov et al 65 Subsequently several studies in either animals or humans have confirmed that a ratio in ammonia concentration higher than 2 mM between the brain and blood stream causes HE and even a comatose state when the value is greater than 5 mM Some investigators have also reported a decrease in serum ammonia following a MARS treatment Table 3 Table 3 Clinical studies showing some improvement in the treatment of HE following a MARS treatment References Nº patients Age years aver SD TreatmentHours patient Ammonia Pre MARS mg dl Ammonia Post MARS mg dl pAwad et al 66 1999 9 38 5 73 2 130 64 lt 0 05Novelli et al 67 2002 10 42 12 51 2 247 126 lt 0 003Schmidt et al 68 2001 8 43 5 10 0 150 121 lt 0 05Sorkyne et al 69 2001 8 47 16 28 1 280 65 lt 0 005Manganese and copper serum levels are increased in patients with either acute or acute on chronic liver failure Nevertheless only in those patients with chronic hepatic dysfunction a bilateral magnetic resonance alteration on Globos Pallidus is observed 70 probably because this type of patients selectively shows higher cerebral membrane permeability Imbalance between aromatic and branched chain aminoacids Fischer index traditionally involved in HE genesis 71 72 73 can be normalized following a MARS treatment The effects are noticeable even after 3 hours of treatment and this reduction in the Fisher index is accompanied with an improvement in the HE 74 Novelli G et al 75 published their three years experience on MARS analyzing the impact of the treatment in the cerebral level for 63 patients reporting an improvement in Glasgow Coma Score GCS for all observed in all patients In the last 22 patients cerebral perfusion pressure was monitored by Doppler mean flow velocity in middle cerebral artery establishing a clear relationship between a clinical improvement especially neurological and an improvement in arterial cerebral perfusion This study confirms other results showing similar increments in cerebral perfusion in patients treated with MARS 68 More recently several studies have shown a significant improvement of HE in patients treated with MARS In the studies by Heemann et al 76 and Sen et al 77 an improvement in HE was considered when encephalopathy grade was reduced by one or more grades vs basal values for Hassenein et al in their randomized controlled trial improvement was considered when a decrease of two grades was observed 78 In the latter 70 patients with acute on chronic liver failure and encephalopathy grade III and IV were included Likewise Kramer et al 79 estimated an HE improvement when an improvement in peak N70 latency in electroencephalograms was observed Sen et al 44 observed a significant reduction in Child Pugh Score p lt 0 01 at 7 days following a MARS treatment without any significant change in the controls Nevertheless when they looked at the Model for End Stage Liver Disease Score MELD a significant reduction in both groups MARS and controls was recorded p lt 0 01 y p lt 0 05 respectively Likewise in several case series an improvement in HE grade with MARS therapy is also reported 80 81 82 83 84 85 86 87 88 Effects of MARS Treatment on Unstable Hemodynamics Edit Hemodynamic instability is often associated with acute liver insufficiency as a consequence of endogenous accumulation of vasoactive agents in the blood This is characterized by a systemic vasodilatation a decrease of systemic vascular resistance arterial hypotension and an increase of cardiac output that gives rise to a hyperdynamic circulation During MARS therapy systemic vascular resistance index and mean arterial pressure have been shown to increase and show improvement 80 82 84 89 90 Schmidt et al 91 reported the treatment of 8 patients diagnosed with acute hepatic failure that were treated with MARS for 6 hours and were compared with a control group of 5 patients to whom ice pads were applied to match the heat loss produced in the treatment group during the extracorporeal therapy They analyzed hemodynamic parameters in both groups hourly In the MARS group a statistically significant increase of 46 on systemic vascular resistance was observed 1215 437 to 1778 710 dinas x s x cm 5 x m 2 compared with a 6 increase in the controls Mean arterial pressure also increased 69 5 to 83 11 mmHg p lt 0 0001 in the MARS group whereas no difference was observed in the controls Cardiac output and heart rate also decreased in the MARS group as a consequence of an improvement in the hyperdynamic circulation Therefore it was shown that a statistically significant improvement was obtained with MARS when compared with the SMT Catalina et al 92 have also evaluated systemic and hepatic hemodynamic changes produced by MARS therapy In 4 patients with acute decompensation of chronic liver disease they observed after MARS therapy an attenuation of hyperdynamic circulation and a reduction in the portal pressure gradient was measured Results are summarized in table 4 Table 4 Hemodinamic parameters during MARS treatment PARAMETER BASE LINE AFTER 1st TREAT AFTER 2nd TREAT MAP mmHg 77 8 11 7 82 7 11 7 84 2 8WHVP mmHg 40 7 5 6 34 9 6 37 3 5 5FHVP mmHg 17 7 7 4 16 7 7 5 17 3 6HVPG mmHg 23 7 0 17 3 9 9 20 3 5 5PAP mmHg 23 7 7 3 22 4 8 15 7 4 1PCP mmHg 17 2 8 3 14 5 2 9 9 7 7 0rPAP mmHg 11 2 4 6 10 4 2 7 3 7 6CO l min 11 2 1 6 10 2 8 9 4 2 1SVRI dinas x seg cm5 478 5 105 514 104 7 622 198Abbreviations MAP mean arterial pressure WHVP wedged hepatic venous pressure FHVP free hepatic venous pressure HVPG hepatic venous pressure gradient PAP pulmonary arterial pressure PCP pulmonar capillary pressure rPAP right pulmonary arterial pressure CO cardiac output SVRI systemic vascular resistance index There are other studies also worth mentioning with similar results Heemann et al 76 and Pares et al 93 among others Dethloff T et al 94 concluded that there is a statistically significant improvement favourable to MARS in comparison with Prometheus system Table 5 Table 5 Changes in hemodynamic parameters with MARS therapy MARS System MAP SVRI CO BRMitzner et al 95 2000 NS Heemann et al 76 2002 S Schmidt et al 91 2003 S S S SLaleman et al 90 2006 S S Dethloff et al 94 2008 NS Abbreviations MAP Mean arterial pressure SVRI Systemic vascular resistance index CO Cardiac output BR Beat rate S Statistically significant p lt 0 05 NS not statistically significant Increase Decrease No changes Effects of MARS Treatment on Renal Function Edit Hepatorenal syndrome is one of the more serious complications in patients with an acute decompensation of cirrhosis and increased portal hypertension It is characterized by hemodynamic changes in splanchnic systemic and renal circulation Splanchnic vasodilatation triggers the production of endogenous vasoactive substances that produce renal vasoconstriction and low glomerular filtration rate leading to oliguria with a concomitant reduction in creatinine clearance Renal insufficiency is always progressive with a very poor prognosis 95 96 with survival at 1 and 2 months of 20 and 10 respectively Pierre Versin 97 is one of the pioneers in the study of hepatorenal syndrome in patients with liver impairment Great efforts have been made trying to improve the prognosis of this type of patient however few have solved the problem Orthotopic liver transplantation is the only treatment that has shown to improve acute and chronic complications derived from severe liver insufficiency Today it is possible to combine albumin dialysis with continuous veno venous hemodialfiltration which provides a greater expectation for these patients 98 by optimization of their clinical status MARS treatment lowers serum urea and creatinine levels improving their clearance 90 91 92 94 and even favors resolution of hepatorenal syndrome 76 83 84 89 99 Results are confirmed in a randomized controlled trial published by Mitzner et al 95 in which 13 patients diagnosed with hepatorenal syndrome type I were treated with MARS therapy Mean survival was 25 2 34 6 days in the MARS group compared to 4 6 1 8 days observed in the controls in whom hemodiafiltration and standard care SMT was applied This resulted in a statistically significance difference in survival at 7 and 30 days p lt 0 05 Authors concluded that MARS therapy applied to liver failure patients Child Pugh C and UNOS 2A scores who develop hepatorenal syndrome type I prolonged survival compared to patients treated with SMT Although mechanisms explaining previous findings are not yet fully understood it has been reported that there was a decrease in plasma rennin concentrations in patients diagnosed with acute on chronic liver failure with renal impairment that were treated with MARS Likewise other studies have suggested some efficacy for MARS in the treatment of hepatorenal syndrome 100 101 102 However other references have been published that do not show efficacy in the treatment of these types of patients with MARS therapy Khuroo et al 103 published a metaanalysis based in 4 small RCT s and 2 non RCT s in patients diagnosed with ACLF concluding that MARS therapy would not bring any significant increment on survival compared with SMT Another observational study in 6 patients with cirrhosis refractory ascitis and hepatorenal syndrome type I not responding to vasoconstrictor therapy showed no impact on hemodynamics following MARS therapy however authors concluded that MARS therapy could effectively serve as bridge to liver transplantation 85 104 Effects of MARS Treatment on Biochemical Parameters Edit Total bilirubin was the only parameter analyzed in all trials that was always reduced in the groups of patients treated with MARS Banayosy et al 105 measured bilirubin levels 14 days after since MARS therapy was terminated and observed a consistent significant decrease not only for bilirubin but also for creatinine and urea Table 6 Table 6 Biochemical parameters analysis with MARS therapy MARS study Controls n MARS Group n Pathology MARS Intervention Bilirubin Creatinine Albumin Urea ALT AST Ammonia BUNMitzner et al 2000 95 8 5 ACLF OH 6 h x 3 d S S NS Heemann et al 2002 76 12 12 ACLF 6 h x 3 d S Sen et al 2004 77 9 9 ACLF HE OH 8 h x 7 d 4 sessions S S Laleman et al 2006 90 6 6 SHR 6 8 h x 10 d S Hassanein et al 2007 78 31 39 ACLF HE III IV 6 h x 5 d S S S SSchmidt et al 2003 91 5 8 ALF 6 h S S S S El Banayosy et al 2004 105 13 14 ALF 8 h x 3 d S S SDethloff et al 2008 94 8 8 Prometheus 8 ACLF S S Montejo et al 2009 106 26 19 IHA GD ACLF 8 h x 3 d S S S S S Abbreviations ALF acute liver failure ACLF acute on chronic liver failure GD graft dysfunction HE hepatic encephalopathy AST aspartate amino transferase BUN blood urea nitrogen NS not significant S statistically significant p lt 0 05 decrease increase no change ALT alanine amino transferase h hours d days Primary goal in the Heemann trial was to achieve a bilirubin level beloiw 15 mg dl during three consecutive days which was observed in 42 of the patients treated with MARS compared to 17 in the control group Impact of MARS therapy on plasma biliary acids levels was evaluated in 3 studies In the study from Stadbauer et al 107 that was specifically addressing the topic it is reported that MARS and Prometheus systems lower to the same extent biliary acids plasma concentration Heemann et al 76 and Laleman et al 90 have also published a significant improvement for these organic ions Effects of MARS Treatment on Pruritus Edit Pruritus is one of the most common clinical manifestations in cholestasis liver diseases and one of the most distressing symptoms in patients with chronic liver disease caused by viral hepatitis C Many hypothesis have been formulated to explain physio pathogenesis of such manifestation including incremental plasma concentration of biliary acids abnormalities in the bile ducts 108 increased central neurotransmitters coupling opioid receptors 109 110 etc Despite the number of historical drugs used individually or combined exchange resins hydrophilic biliary acids antihistamines antibiotics anticonvulsants opioid antagonists there are reported cases of intractable or refractory pruritus with a dramatic reduction in patients quality of life i e sleep disorders depression suicide attempts 111 112 Intractable pruritus can be an indication for liver transplantation The MARS indication for intractable pruritus is therapeutically an option that has shown to be beneficial for patients in desperate cases although at high cost 113 114 115 116 In several studies it was confirmed that after MARS treatments patients remain free from pruritus for a period of time ranging from 6 to 9 months 116 Nevertheless some authors have concluded that besides the good results found in the literature application of MARS therapy in refractory pruritus requires larger evidence 114 Effects of MARS Treatment on Drugs and Poisons clearance Edit Pharmacokinetics and pharmacodynamics for a majority of drugs can be significantly be modified with liver failure affecting the therapeutic approach and potential toxicity of the drugs In these type of patients Child Pugh score represents a poor prognostic factor to assess the metabolic capacity of the failing liver Metabolic performance of the liver depends on several factors Hepatic flow rate Cytochrome P 450 enzimatic activity Albumin affinity for the drug Extrahepatic clearance for the drugIn patients with hepatic failure drugs that are only metabolized in the liver accumulate in the plasma right after they are administered and therefore it is needed to modify drug dosing in both concentration and time intervals to lower the risk of toxicity It is also necessary to adjust the dosing for those drugs that are exclusively metabolized by the liver and have low affinity for prioteins and high distribution volume such as fluoroquinolones Levofloxacin and Ciprofloxacin 117 118 119 120 Extracorporeal detoxification with albumin dialysis increases the clearance of drugs that are bound to plasmatic proteins Table 7 Table 7 Drugs Highly Bound to Proteins NSAIDs Antibiotics Anti H2 Statins AntipsychoticsDiclofenac Cephazoline Omeprazole Clofibrate ChlorpromazineIbuprofen Cefoperazone Antifungals Lovastatin HaloperidolIndomethacin Ceftriaxone Amphotericine B Simvastatine Chemotherapy DrugsKetoprofen Cloxacilin Itraconazole Fluvastatine ChlorambucilNaproxen Oxacilin Ketoconazole Hypotensors EtoposidePhenylbutazone Clindamycin Barbiturates Hydralacine MelphalanPiroxicam Erythromycin Thiopental Irbesartan D penicillamineAnaesthetics Teicoplanin Benzodiazepines Losartan TamoxifenBupivacaine Rifampicin Clonazepam Misoprostol OpioidsCalcium Anthagonists Rifabutin Diazepam Prazosin FentanylNimodipine Antidepresants Flunitrazepam Valsartan MethadoneNifedipine Amitriptyline Flurazepam ACE s AntipsychoticsNitrendipine Desipramine Lorazepam Fosinopropil ClozapineAntiarrhythmics Imipramine Midazolam Quinapril FluoxetineAmiodarone Nortriptyline Nitrazepam Immunosuppressors RisperidoneQuinidine Anticoagulants Oxacepam Ciclosporin SertralinePropafenone Warfarin Temazepam Tacrolimus ZolpidemLidocaine Antiemetics Beta Blockers Methylprednisolone OthersDigoxin Ondansetron Carvedilol Prednisone CelecoxibVerapamil Antiepilectics Propanolol Budesonide DiphenhydramineOral Antidiabetics Carbamazepine Diuretics EthinylestradiolGlibenclamide Valproic acid Furosemide Mefloquine Spironolactone Paclitaxel Torsemide Effects of MARS on Survival Edit In the meta analysis published by Khuroo et al 103 which included 4 randomized trials 76 91 95 105 an improvement in survival for the patients with liver failure treated with MARS compared with SMT was not observed However neither in the extracorporeal liver support systems review by the Cochrane 121 published in 2004 nor the meta analysis by Kjaergard et al 122 was a significance difference in survival found for patients diagnosed with ALF treated with extracorporeal liver support systems Nevertheless these reviews included all kind of liver support systems and used a heterogeneous type of publication abstracts clinical trials cohort etc There is literature showing favorable results in survival for patients diagnosed with ALF and treated with MARS In a randomized controlled trial Saliba et al 123 studied the impact on survival of MARS therapy for patients with ALF waiting on the liver transplant list Forty nine patients received SMT and 53 were treated with MARS They observed that patients that received 3 or more MARS sessions showed a statistically significance increase in transplant free survival compared with the others patients of the study Notably 75 of the patients underwent liver transplantation in the first 24 hours after inclusion in the waiting list and besides the short exposure to MARS therapy some patients showed a better survival trend compared to controls when they were treated with MARS prior to the transplant In a case controlled study by Montejo et al 106 it was reported that MARS treatment do not decrease mortality directly however the treatment contributed to significantly improve survival in patients that were transplanted In studies by Mitzner et al 95 and Heemann et al 76 they were able to show a significance statistical difference in 30 day survival for patients in the MARS group However El Banayosy et al 105 and Hassanein et al 78 noticed a non significant improvement in survival probably because of the short number of patients included in the trials In the majority of available MARS studies published with patients diagnosed with ALF either transplanted or not survival was greater in the MARS group with some variations according to the type of trial ranging from 20 30 124 125 and 60 80 85 126 127 128 Data is summarized in Tables 8 9 and 10 Table 8 Studies of MARS Therapy MARS Study Pathology Nº Patients Type of Study Level of Evidence Study Groups Analysed VariablesMARS ControlsMitzner et al 95 2000 ACLF 13 RCT LN III 18 5 HD 1 3 4Heemann et al 76 2002 ACLF 24 RCT LN III 12 12 SMT 1 2 3 4 5Sen et al 77 2004 ACLF 18 RCT LN III 9 9 SMT 1 3 4 5Hassanein et al 78 2007 ACLF 70 RCT HN II 39 31 SMT 1 2 3 5Schmidt et al 91 2003 ALF 13 RCT LN III 8 5 SMT 1 3 4El Banayosy et al 105 2004 ALF 27 RCT LN III 14 13 HD 1 3Montejo et al 106 2009 ACLF ALF 45 Case control 19 26 SMT 1 2 3 4 5MARS Prometheus StudiesEvenepoel et al 129 2006 ACLF 18 Case serie VIII Comparative series 1 2 3Faenza et al 130 2008 ACLF 57 Case serie VIII Comparative series 1 2 5Krisper et al 131 2005 ACLF 10 RCT LN III Crossover 1 2 3Laleman et al 90 2006 ACLF 18 RCT LN III 6 6 2 3 4Stadlbauer et al 132 2006 ACLF 8 RCT LN III Crossover 1 2 3Dethloff et al 94 2008 ESLD 24 RCT LN III 8 8 1 2 3 4Abbreviations ACLF Acute on chronic liver failure ALF Acute liver failure ESLD End stage liver disease RCT Randomized controlled trial RCT LN RCT low number of patients RCT HN RCT high number of patients HD Hemodialysis SMT Standard medical treatment 1 Survival mortality 2 Security 3 Biochemical parameters 4 hemodynamic parameters 5 Clinical parameters Table 9 Survival with MARS Therapy Study Pathology Controlled Study Follow up Survival Faenza et al 130 2008 ACLF No Bridging to LTx 7 10 70 3 months 2 6 33 Krisper et al 131 2005 ACLF Yes crossover 30 days 4 9 44 Stadlbauer et al 132 2006 ACLF Yes crossover 30 days 4 8 50 Laleman et al 90 2006 ACLF Yes 7 days 6 6 100 Dethloff et al 94 2008 ESLD Yes 6 months 5 8 63 Abbreviations ALF Acute liver failure ACLF Acute on chronic liver failure ESLD End stage liver disease Not statistically significance Table 10 Mortality in Selected Studies Study N Mortality MARS Mortality Controls Time Odds Ratio OR Acute on Chronic Liver FailureMitzner et al 95 13 63 100 7 days 0 63Heemann et al 76 24 50 67 6 months 0 75Sen et al 77 18 56 56 3 months 1Hassanein et al 78 70 49 55 NR 0 89Dethloff et al 94 24 25 50 6 months 0 50Acute Liver FailureSchmidt et al 91 13 38 40 NR 0 94El Banayosy et al 105 27 50 69 NR 0 72Abbreviations NR Not reported For patients diagnosed with acute on chronic liver failure and treated with MARS therapy clinical trial results showed a not statistically significant reduction in mortality odds ratio OR 0 78 confident interval CI 95 0 58 1 03 p 0 1059 Figure 3 Figure 3 Meta analysis showing the effect on survival of patients with ACLF treated with MARS therapy A non statistically significant reduction of mortality was shown in patients with ALF treated with MARS OR 0 75 CI 95 0 42 1 35 p 0 3427 Figure 4 Figure 4 Meta analysis showing the effect on survival of patients with ALF treated with MARS therapy Combined results yielded a non significant reduction on mortality in patients treated with MARS therapy However the low number of patients included in each of the studies may be responsible for not being able to achieve enough statistical power to show differences between both treatment groups Moreover heterogeneity for the number of MARS sessions and severity of liver disease of the patients included make it very difficult for the evaluation of MARS impact on survival Recently a meta analysis on survival in patients treated with an extra hepatic therapy has been published 133 Searching strategies yielded 74 clinical trials 17 randomized controlled trials 5 case control and 52 cohort studies Eight studies were included in the meta analysis three addressing acute liver failure one with MARS therapy 105 and five addressing acute on chronic being four MARS related 76 77 78 95 Authors concluded that extra hepatic detoxifying systems improve survival for acute liver insufficiency whereas results for acute decompensation of chronic liver diseases suggested a non significant survival benefit Also due to an increased demand for liver transplantation together with an augmented risk of liver failure following large resections development of detoxifying extrahepatic systems are necessary Safety Aspects Edit Safety defined as presence of adverse events is evaluated in few trials Adverse events in patients receiving MARS therapy are similar to those in the controls with the exception of thrombocytopenia and hemorrhage that seems to occur more frequently with the MARS system 134 Heemann et al 76 reported two adverse events most probably MARS related fever and sepsis presumably originated at the catheter In the study by Hassanein et al 78 two patients in the MARS group abandoned the study owing tohemodynamic instability three patients required larger than average platelets transfusion and three more patients presented gastrointestinal bleeding Laleman et al 90 detected one patient with thrombocytopenia in both the MARS and Prometheus treatment groups and an additional patient with clotting of the dialysis circuit and hypotension only in the Prometheus group Kramer et al Biologic DT 79 wrote about 3 cases with disseminated intravascular coagulation in the interventional group two of them with fatal outcomes Mitzner et al 95 described among patients treated with MARS a thrombocytopenia case and a second patient with chronic hepatitis B who underwent TIPS placement on day 44 after randomization and died on day 105 of multiorgan failure as a consequence of complications related to the TIPS procedure Montejo et al 106 showed that MARS is an easy technique without serious adverse events related to the procedure and also easy to implement in ICU settings that are used to renal extracorporeal therapies The MARS International Registry with data from more than 500 patients although sponsored by the manufacturer shows that the adverse effects observed are similar to the control group However in these severely ill patients it is difficult to distinguish between complications of the disease itself and side effects attributable to the technique Health Economics Edit Only three Studies addressing cost effectivenenss of MARS therapy have been found Hassanein et al 135 analysed costs of randomized patients with ACLF receiving MARS therapy or standard medical care They used the study published in 2001 by Kim et al 136 describing the impact of complications in hospitalization costs in patients diagnosed with alcoholic liver failure Cost of 11 patients treated with standard medical care SMT were compared to those that received MARS in addition to SMT 12 patients In the MARS group there was less in hospital mortality and complications related to the disease with a remarkable reduction in cost which compensated the MARS related expenditure Table 11 Table 11 Analysis of Complications According to the Modality of treatment MARS GROUP n 12 CONTROL GROUP n 11In hospital mortality 1 12 6 11Worsening of hepatic encephalopathy grade 4 0 12 3 11Worsening of renal function Hepatorenal Syndrome 1 12 7 11Ascites 0 12 1 11Variceal bleeding 0 12 1 11Severe hypotension 2 12 3 11Electrolyte disorders 4 12 10 11Coagulopathy 4 12 3 11There were 5 survivors in the control group with a cost per patient of 35 904 whereas in the MARS group 11 patients out of 12 survived with a cost per patient of 32 036 which represents a 4000 savings per patient in favors of the MARS group Hessel et al 137 published a 3 year follow up of a cohort of 79 patients with ACLF of whom 33 received MARS treatments and 46 received SMT Survival was 67 for the MARS group and 63 for the controls that was reduced to 58 and 35 respectively at one year follow up and then 52 and 17 at three years Hospitalization costs for the MARS treated group were greater than that for the controls 31 539 vs 7 543 and similarly direct cost at 3 year follow up 8 493 vs 5 194 Nevertheless after adjusting mortality rate the annual cost per patient was 12 092 for controls and 5 827 for MARS group also in the latter they found an incremental cost effectiveness ratio of 31 448 per life year gained LYG and an incremental costs per QALY gained of 47171 Two years later same authors published the results of 149 patients diagnosed with ACLF 138 There were 67 patients 44 9 treated with MARS and 82 patients 55 1 were allocated to receive SMT Mean survival time was 692 days in the MARS group 33 at 3 years and 453 days in the controls 15 at 3 years the results were significant p 0 022 Differences in average cost was 19 853 95 IC 13 308 25 429 35 639 for MARS patients and 15 804 for the control group Incremental cost per LYG was 29 985 95 IC 9 441 321 761 and 43 040 95 IC 13 551 461 856 per quality adjusted life years QALY Liver support systems such as MARS are very important to stabilize patients with acute or acute on chronic liver failure and avoid organ dysfunction as well as a bridge to transplant Although initial in hospital costs are high they are worth for the favorable outcome MARS Therapy Indications Edit Acute on Chronic Liver Failure Edit Etiology Chronic viral hepatitis 139 Alcoholic liver disease 140 141 Autoimmune disease 142 Metabolic disease such as hemochromatosis Idiopathic CirrhosisGoals of MARS Therapy Re compensation of previous chronic state Prolong survival time and bridge to urgent or elective transplant Pre transplant optimization of the patientMARS Therapy Indication Bilirubin gt 15 mg dl 255 mmol L not responding to standard medical care alter 3 days Renal dysfunction or hepatorenal syndrome Hepatic encephalopathy IITreatment Schedule 3 to 5 eight hour treatment sessions on consecutive days Continuous treatment with hemodynamic instability in any case treatment kit must be replaced every 24 hours Acute Liver failure Edit Etiology Viral infection 67 91 143 Poisoning paracetamol overdose mushrooms 126 144 145 146 147 Multiorgan dysfunction severe sepsis Vascular diseases Budd Chiari syndrome Hypoxic hepatitis 148 Liver failure during pregnancy or Reye syndrome Unknown etiologyGoals of MARS Therapy Native liver recovery Bridging to liver transplant Pre transplant optimization of the patient MARS Therapy Indication King s College or Clichy criteria for liver transplantation Hepatic encephalopathy II Increased intracraneal pressure Acute hypoxic hepatitis with bilirubin gt 8 mg dl 100 mmol L Renal dysfunction or hepatorenal syndrome Progressive intrahepatic cholestasis Fulminant Wilson disease Acute liver dysfunction following paracetamol overdoseTreatment Schedule 3 to 5 eight hour treatment sessions in consecutives days Hypoxic hepatitis 3 eight hour treatment sessions in consecutives days Paracetamol overdose 3 to 5 twenty four hour treatment sessions Mushroom poisoning 3 to 5 twenty four hour treatment sessions Fulminant Wilson minimum 5 twenty four hour treatment sessions owing to copper saturation of the treatment kit Drug overdose 3 to 5 eight hour treatment sessions in consecutives daysMARS in Graft Dysfucntion After Liver Transplant Edit 149 150 Etiology Graft damage during preparation and transportation Infection Hepatotoxic drugs Graft rejection Technical complications vascular biliary Recurrence of primary diseaseGoals of MARS Therapy Recovery and prevention of re transplantation Prolong survival time and stabilize the patient to receive a re transplant if the above goal is not achievedMARS Therapy Indication Primary graft dysfunction Hepatic encephalopathy II Increased intracranial pressure Renal dysfunction or hepatorenal syndrome Progressive intrahepatic cholestasisTreatment Schedule 3 to 5 eight hour treatment sessions on consecutive days Continuous treatment with hemodynamic instability in any case treatment kit must be replaced every 24 hours MARS in liver Failure after Liver Surgery Edit 151 Etiology Liver Resection in hepatocellular carcinoma Transarterial Chemoembolization TACE Partial resection in living donor transplantation Other surgical interventionsGoals of MARS Therapy Recovery until hepatic regenerationMARS Therapy Indication Hepatic encephalopathy II Renal dysfunction or hepatorenal syndrome Progressive intrahepatic cholestasisTreatment Schedule 3 to 5 eight hour treatment sessions on consecutive days Continuous treatment with hemodynamic instability in any case treatment kit must be replaced every 24 hours MARS for intractable pruritus in Cholestasis Edit 99 139 142 152 Etiology Primary biliary cirrhosis PBC primary sclerosing cholangitis PSC Benign intrahepatic cholestasis BIC Biliary AtresiaGoals of MARS Therapy Attenuate pruritus symptoms and improve patients quality of lifeMARS Therapy Indication Pruritus not responding to SMTTreatment Schedule 3 to 5 eight hour treatment sessions in consecutives days Repeat treatment when symptoms reoccurMARS Therapy Contraindications Edit Same contraindications as with any other extracorporeal treatment may be applied to MARS therapy Unstable hemodynamics with mean arterial pressure MAP lt 55 mmHg despite vasoconstrictors administration Uncontrolled hemorrhage Severe coagulopathy Severe thrombocytopeniaTreatment Parameters Edit Blood FlowThe trend is to use high flow rates although it is determined by the technical specifications of the combined machine and catheters sizeIntermittent treatments Without renal dysfunction it is recommended a blood and albumin flow rates ranging from 150 to 250 mL minContinuous treatments With or without renal impairment it is recommended to use flow rates from 100 to 150 mL min Dyalisate Flow RateIntermittent treatments Without renal impairment 1800 a 3000 mL hour With renal impairment 3000 a 6000 mL hourContinuous treatments Recommended flow rate 1000 to 2000 mL hour Replacement Flow Rate According to medical criteria and same as in CVVHDHeparin AnticoagulationSimilarly to CVVHD it depends on previous patient s coagulation status In many cases it will not be needed unless the patient presents a PTT inferior to 160 seconds In patients with normal values a bolus of 5000 to 10000 IU of heparin could be administered at the commencement of the treatment followed by a continuous perfusion to keep PTT in ratios from 1 5 to 2 5 or 160 to 180 seconds MonitoringA biochemical analysis is recommended liver and kidney profile ionic glucose together with a hemogram at the end of first session and before starting the following one Coagulation analysis must be also performed before starting the session to adjusting heparin dose In case that medication susceptible to be eliminated by MARS is being administered it is also recommended to monitor their levels in bloodEnd of the Session Once the treatment is finalized blood should be returned following the unit procedure and both catheter s lumens heparinized For the next session a new kit must be used For continuous treatments kit must be changed by a new one every 24 hours Treatment must be stopped before schedule owing to the particular circumstances listed below MAP inferior to 40 mmHg at least for 10 minutes Air embolism of the extracorporeal circuit Transmembrane pressure TMP greater than 600 mmHg Blood leak detection in the albumin circuit Disseminated intravascular coagulation DIC Severe active hemorrhageFDA Clearance US only Edit Federal Drug Administration FDA cleared in a document dated on May 27 2005 MARS therapy for the treatment of drug overdose and poisoning The only requirement is that the drug or poison must be susceptible to be dialysed and removed by activated charcoal or anionic exchange resins More recently on December 17 2012 MARS therapy has been cleared by the FDA for the treatment of hepatic encephalopathy due to a decompensation of a chronic liver disease Clinical trials conducted with MARS treatment in HE patients having a decompensation of chronic liver disease demonstrated a transient effect from MARS treatments to significantly decrease their hepatic encephalopathy scores by at least 2 grades compared to standard medical therapy SMT The MARS is not indicated as a bridge to liver transplant Safety and efficacy has not been demonstrated in controlled randomized clinical trials The effectiveness of the MARS device in patients that are sedated could not be established in clinical studies and therefore cannot be predicted in sedated patientsLiverNet EditThe LiverNet is a database dedicated to the liver diseases treated with the support of extracorporeal therapies To date the most currently used system is the Molecular Adsorbent Recirculating System MARS which is based on the selective removal of albumin bound molecules and toxins from the blood in patients with acute and acute on chronic liver failure The purpose is to register prospectively all patients treated worldwide with the MARS system in order to Improve our understanding of the clinical course pathophysiology and treatment of these diseases Evaluate the clinical impact of MARS therapy on the course of the disease in different specific indication Increase the knowledge in this extremely innovative area a basis for an improvement of liver support devices and the treatment of these patients in the next futureThe liverNet is an eCRF database www livernet net using a SAS platform that allows major advantages for the centres including the automatic calculations of most liver rand ICU scoring systems instant queries online instant export of all patients included in the database of each centre to an Excel file for direct statistical analysis and finally instant online statistical analysis of selective data decided by the scientific committee Therefore the LiverNet is an important tool to progress in the knowledge of liver support therapies See also EditAmerican Society for Artificial Internal Organs Tissue engineeringReferences Edit Pless G 2007 Artificial and bioartificial liver support Organogenesis 3 1 20 24 doi 10 4161 org 3 1 3635 PMC 2649611 PMID 19279696 Shakil OA Kramer D Mazariegos GV Fung JJ Rakela J 2000 Acute liver failure clinical features outcome analysis and applicability of prognostic criteria Liver Transpl 6 2 163 169 doi 10 1002 lt 500060218 PMID 10719014 S2CID 33789452 Jalan R Williams R 2002 Acute on chronic liver failure Pathophysiological basis of therapeutic options Blood Purif 20 3 252 261 doi 10 1159 000047017 PMID 11867872 S2CID 36924439 Stravitz RT 2008 Critical management decisions in patients with acute liver failure Chest 134 5 1092 1102 doi 10 1378 chest 08 1071 PMID 18988787 Auzinger G Wendon J April 2008 Intensive care management of acute liver failure Current Opinion in Critical Care 14 2 179 88 doi 10 1097 MCC 0b013e3282f6a450 PMID 18388681 S2CID 1289311 Sen S Williams R Jalan R 2005 Emerging indications for albumin dialysis Am J Gastroenterol 100 2 468 475 doi 10 1111 j 1572 0241 2005 40864 x PMID 15667509 S2CID 20937240 Allen JW Hassanein T Bhatia SN September 2001 Advances in bioartificial liver devices Hepatology 34 3 447 55 doi 10 1053 jhep 2001 26753 PMID 11526528 S2CID 6852149 Sussman NL Chong MG Koussayer T He DE Shang TA Whisennand HH Kelly JH July 1992 Reversal of fulminant hepatic failure using an extracorporeal liver assist device Hepatology 16 1 60 5 doi 10 1002 hep 1840160112 PMID 1618484 S2CID 45920140 Stange J Ramlow W Mitzner S Schmidt R Klinkmann H September 1993 Dialysis against a recycled albumin solution enables the removal of albumin bound toxins Artificial Organs 17 9 809 13 doi 10 1111 j 1525 1594 1993 tb00635 x PMID 8240075 Demetriou AA Brown RS Jr Busuttil RW Fair J McGuire BM Rosenthal P Am Esch JS 2nd Lerut J Nyberg SL Salizzoni M Fagan EA de Hemptinne B Broelsch CE Muraca M Salmeron JM Rabkin JM Metselaar HJ Pratt D De La Mata M McChesney LP Everson GT Lavin PT Stevens AC Pitkin Z Solomon BA May 2004 Prospective randomized multicenter controlled trial of a bioartificial liver in treating acute liver failure Annals of Surgery 239 5 660 7 discussion 667 70 doi 10 1097 01 sla 0000124298 74199 e5 PMC 1356274 PMID 15082970 Mazariegos GV Patzer JF 2nd Lopez RC Giraldo M Devera ME Grogan TA Zhu Y Fulmer ML Amiot BP Kramer DJ March 2002 First clinical use of a novel bioartificial liver support system BLSS American Journal of Transplantation 2 3 260 6 doi 10 1034 j 1600 6143 2002 20311 x PMID 12096789 Falkenhagen D Strobl W Vogt G Schrefl A Linsberger I Gerner FJ Schoenhofen M January 1999 Fractionated plasma separation and adsorption system a novel system for blood purification to remove albumin bound substances Artificial Organs 23 1 81 6 doi 10 1046 j 1525 1594 1999 06292 x PMID 9950184 Xue YL Zhao SF Luo Y Li XJ Duan ZP Chen XP Li WG Huang XQ Li YL Cui X Zhong DG Zhang ZY Huang ZQ December 2001 TECA hybrid artificial liver support system in treatment of acute liver failure World Journal of Gastroenterology 7 6 826 9 doi 10 3748 wjg v7 i6 826 PMC 4695603 PMID 11854910 Morsiani E Pazzi P Puviani AC Brogli M Valieri L Gorini P Scoletta P Marangoni E Ragazzi R Azzena G Frazzoli E Di Luca D Cassai E Lombardi G Cavallari A Faenza S Pasetto A Girardis M Jovine E Pinna AD March 2002 Early experiences with a porcine hepatocyte based bioartificial liver in acute hepatic failure patients The International Journal of Artificial Organs 25 3 192 202 doi 10 1177 039139880202500305 PMID 11999191 S2CID 196422022 Sauer IM Goetz M Steffen I Walter G Kehr DC Schwartlander R Hwang YJ Pascher A Gerlach JC Neuhaus P May 2004 In vitro comparison of the molecular adsorbent recirculation system MARS and single pass albumin dialysis SPAD Hepatology 39 5 1408 14 doi 10 1002 hep 20195 PMID 15122770 Mundt A Puhl G Muller A Sauer I Muller C Richard R Fotopoulou C Doll R Gabelein G Hohn W Hofbauer R Neuhaus P Gerlach J June 2002 A method to assess biochemical activity of liver cells during clinical application of extracorporeal hybrid liver support The International Journal of Artificial Organs 25 6 542 8 doi 10 1177 039139880202500609 PMID 12117294 S2CID 38686470 van de Kerkhove MP Di Florio E Scuderi V Mancini A Belli A Bracco A Dauri M Tisone G Di Nicuolo G Amoroso P Spadari A Lombardi G Hoekstra R Calise F Chamuleau RA October 2002 Phase I clinical trial with the AMC bioartificial liver The International Journal of Artificial Organs 25 10 950 9 doi 10 1177 039139880202501009 PMID 12456036 S2CID 196427668 Rozga J September 2006 Liver support technology an update Xenotransplantation 13 5 380 9 doi 10 1111 j 1399 3089 2006 00323 x PMID 16925661 S2CID 7562668 Sussman NL Gislason GT Conlin CA Kelly JH May 1994 The Hepatix extracorporeal liver assist device initial clinical experience Artificial Organs 18 5 390 6 doi 10 1111 j 1525 1594 1994 tb02221 x PMID 8037614 Ellis AJ Hughes RD Wendon JA Dunne J Langley PG Kelly JH Gislason GT Sussman NL Williams R December 1996 Pilot controlled trial of the extracorporeal liver assist device in acute liver failure Hepatology 24 6 1446 51 doi 10 1002 hep 510240625 PMID 8938179 S2CID 22557163 Phua J Lee KH April 2008 Liver support devices Current Opinion in Critical Care 14 2 208 15 doi 10 1097 MCC 0b013e3282f70057 PMID 18388685 S2CID 30297978 Aron Jonathan Agarwal Banwari Davenport Andrew 2016 04 02 Extracorporeal support for patients with acute and acute on chronic liver failure Expert Review of Medical Devices 13 4 367 380 doi 10 1586 17434440 2016 1154455 ISSN 1743 4440 PMID 26894968 S2CID 13059506 Artificial Liver Used After Removal of Organ The New York Times Associated Press 1993 05 19 ISSN 0362 4331 Retrieved 2020 02 06 Medical Miracles PEOPLE com Retrieved 2020 02 06 a b Best Inventions of 2001 Time 2001 11 19 a b Tilles A Berthiaume F Yarmush M Tompkins R Toner M 2002 Bioengineering of liver assist devices Hepatobiliary Pancreat Surg 9 6 686 696 doi 10 1007 s005340200095 PMID 12658402 Allen J Hassanein T Bhatia S 2001 Advances in bioartificial liver devices Hepatology 34 3 447 55 doi 10 1053 jhep 2001 26753 PMID 11526528 S2CID 6852149 Free Full Text Strain A Neuberger J 2002 A bioartificial liver state of the art Science 295 5557 1005 9 Bibcode 2002Sci 295 1005S doi 10 1126 science 1068660 PMID 11834813 S2CID 30433745 Current Work on the Bioartificial Liver a b He Yu Ting Qi Ya Na Zhang Bing Qi Li Jian Bo Bao Ji 2019 07 21 Bioartificial liver support systems for acute liver failure A systematic review and meta analysis of the clinical and preclinical literature World Journal of Gastroenterology 25 27 3634 3648 doi 10 3748 wjg v25 i27 3634 ISSN 1007 9327 PMC 6658398 PMID 31367162 Tsiaoussis J January 2001 Which hepatocyte will it be Hepatocyte choice for bioartificial liver support systems Liver Transplantation 7 1 2 10 doi 10 1053 jlts 2001 20845 PMID 11150414 Werner Andreas Duvar Sevim Muthing Johannes Buntemeyer Heino Lunsdorf Heinrich Strauss Michael Lehmann Jurgen 2000 Cultivation of immortalized human hepatocytes HepZ on macroporous CultiSpher G microcarriers Biotechnology and Bioengineering 68 1 59 70 doi 10 1002 SICI 1097 0290 20000405 68 1 lt 59 AID BIT7 gt 3 0 CO 2 N ISSN 1097 0290 PMID 10699872 University of Minnesota Bioartificial Liver How it Works Wung Nelly Acott Samuel M Tosh David Ellis Marianne J December 2014 Hollow fibre membrane bioreactors for tissue engineering applications Biotechnology Letters 36 12 2357 2366 doi 10 1007 s10529 014 1619 x ISSN 0141 5492 PMID 25064452 S2CID 10046204 a b c d e Jain Era Damania Apeksha Shakya Akhilesh Kumar Kumar Anupam Sarin Shiv K Kumar Ashok December 2015 Fabrication of macroporous cryogels as potential hepatocyte carriers for bioartificial liver support Colloids and Surfaces B Biointerfaces 136 761 771 doi 10 1016 j colsurfb 2015 10 012 PMID 26519938 Lozinsky V I May 2008 Polymeric cryogels as a new family of macroporous and supermacroporous materials for biotechnological purposes Russian Chemical Bulletin 57 5 1015 1032 doi 10 1007 s11172 008 0131 7 ISSN 1066 5285 S2CID 95352668 Erdem Ahmet Ngwabebhoh Fahanwi Asabuwa Yildiz Ufuk 2016 Fabrication and characterization of soft macroporous Jeffamine cryogels as potential materials for tissue applications RSC Advances 6 113 111872 111881 Bibcode 2016RSCAd 6k1872E doi 10 1039 C6RA22523C ISSN 2046 2069 Ingavle Ganesh C Baillie Les W J Zheng Yishan Lis Elzbieta K Savina Irina N Howell Carol A Mikhalovsky Sergey V Sandeman Susan R May 2015 Affinity binding of antibodies to supermacroporous cryogel adsorbents with immobilized protein A for removal of anthrax toxin protective antigen Biomaterials 50 140 153 doi 10 1016 j biomaterials 2015 01 039 PMID 25736504 Derazshamshir Ali Baydemir Gozde Andac Muge Say Ridvan Galaev Igor Yu Denizli Adil 2010 03 15 Molecularly Imprinted PHEMA Based Cryogel for Depletion of Hemoglobin from Human Blood Macromolecular Chemistry and Physics 211 6 657 668 doi 10 1002 macp 200900425 Koshy Sandeep T Zhang David K Y Grolman Joshua M Stafford Alexander G Mooney David J January 2018 Injectable nanocomposite cryogels for versatile protein drug delivery Acta Biomaterialia 65 36 43 doi 10 1016 j actbio 2017 11 024 PMC 5716876 PMID 29128539 Li Jieliang Pan Jilun Zhang Liguo Yu Yaoting June 2003 Culture of hepatocytes on fructose modified chitosan scaffolds Biomaterials 24 13 2317 2322 doi 10 1016 S0142 9612 03 00048 6 PMID 12699669 Ito Y Eguchi S Kamohara Y Inuo H Yamanouchi K Okudaira S Yanaga K Furui J Kanematsu T April 2004 Influence of Serum from Rats with Fulminant Hepatic Failure on Hepatocytes in a Bioartificial Liver System The International Journal of Artificial Organs 27 4 303 310 doi 10 1177 039139880402700406 ISSN 0391 3988 PMID 15163064 S2CID 35856518 a b Damania Apeksha Hassan Mohsin Shirakigawa Nana Mizumoto Hiroshi Kumar Anupam Sarin Shiv K Ijima Hiroyuki Kamihira Masamichi Kumar Ashok February 2017 Alleviating liver failure conditions using an integrated hybrid cryogel based cellular bioreactor as a bioartificial liver support Scientific Reports 7 1 40323 Bibcode 2017NatSR 740323D doi 10 1038 srep40323 ISSN 2045 2322 PMC 5227920 PMID 28079174 a b Demetriou Achilles A Brown Robert S Busuttil Ronald W Fair Jeffrey McGuire Brendan M Rosenthal Philip Am Esch Jan Schulte Lerut Jan Nyberg Scott L Salizzoni Mauro Fagan Elizabeth A May 2004 Prospective Randomized Multicenter Controlled Trial of a Bioartificial Liver in Treating Acute Liver Failure Annals of Surgery 239 5 660 670 doi 10 1097 01 sla 0000124298 74199 e5 ISSN 0003 4932 PMC 1356274 PMID 15082970 a b c Thompson Julie Jones Natasha Al Khafaji Ali Malik Shahid Reich David Munoz Santiago MacNicholas Ross Hassanein Tarek Teperman Lewis Stein Lance Duarte Rojo Andres March 2018 Extracorporeal cellular therapy ELAD in severe alcoholic hepatitis A multinational prospective controlled randomized trial Liver Transplantation 24 3 380 393 doi 10 1002 lt 24986 ISSN 1527 6465 PMC 5873437 PMID 29171941 Vital Therapies falls 88 ditches liver treatment after phase 3 fail FierceBiotech 12 September 2018 Retrieved 2021 08 24 Google Scholar scholar google com Advanced Search Cochrane Library Peszynski P Klammt S Peters E Mitzner S Stange J Schmidt R 2002 Albumin dialysis single pass vs recirculation MARS Liver 22 Suppl 2 40 2 doi 10 1034 j 1600 0676 2002 00007 x PMID 12220302 Chawla LS Georgescu F Abell B Seneff MG Kimmel PL March 2005 Modification of continuous venovenous hemodiafiltration with single pass albumin dialysate allows for removal of serum bilirubin American Journal of Kidney Diseases 45 3 e51 6 doi 10 1053 j ajkd 2004 11 023 PMID 15754264 O Grady J June 2006 Personal view current role of artificial liver support devices Aliment Pharmacol Ther 23 11 1549 57 doi 10 1111 j 1365 2036 2006 02931 x PMID 16696802 S2CID 31513268 van de Kerkhove MP Hoekstra R Chamuleau RA van Gulik TM August 2004 Clinical application of bioartificial liver support systems Ann Surg 240 2 216 30 doi 10 1097 01 sla 0000132986 75257 19 PMC 1356396 PMID 15273544 Neuberger J January 2005 Prediction of survival for patients with fulminant hepatic failure Hepatology 41 1 19 22 doi 10 1002 hep 20562 PMID 15690476 Sauer IM Goetz M Steffen I et al May 2004 In vitro comparison of the molecular adsorbent recirculation system MARS and single pass albumin dialysis SPAD Hepatology 39 5 1408 14 doi 10 1002 hep 20195 PMID 15122770 Rifai K Ernst T Kretschmer U et al December 2003 Prometheus a new extracorporeal system for the treatment of liver failure J Hepatol 39 6 984 90 doi 10 1016 S0168 8278 03 00468 9 PMID 14642616 DIALIVE Concept Yaqrit yaqrit com Retrieved 2019 11 04 Stange J Mitzner SR Risler T Erley CM Lauchart W Goehl H Klammt S Peszynski P Freytag J Hickstein H Lohr M Liebe S Schareck W Hopt UT Schmidt R April 1999 Molecular adsorbent recycling system MARS clinical results of a new membrane based blood purification system for bioartificial liver support Artificial Organs 23 4 319 30 doi 10 1046 j 1525 1594 1999 06122 x PMID 10226696 Kapoor D December 2002 Molecular adsorbent recirculating system albumin dialysis based extracorporeal liver assist device Journal of Gastroenterology and Hepatology 17 Suppl 3 S280 6 doi 10 1046 j 1440 1746 17 s3 14 x PMID 12472950 Davenport A Jun 1 2003 Extracorporeal support for patients with hepatic failure Hemodialysis International 7 3 256 63 doi 10 1046 j 1492 7535 2003 00046 x PMID 19379373 S2CID 23916381 Sechser A Osorio J Freise C Osorio RW May 2001 Artificial liver support devices for fulminant liver failure Clinics in Liver Disease 5 2 415 30 doi 10 1016 s1089 3261 05 70172 0 PMID 11385970 Stange J Mitzner S November 1996 A carrier mediated transport of toxins in a hybrid membrane Safety barrier between a patients blood and a bioartificial liver The International Journal of Artificial Organs 19 11 677 91 doi 10 1177 039139889601901109 PMID 8970836 Mitzner S Loock J Peszynski P Klammt S Majcher Peszynska J Gramowski A Stange J Schmidt R December 2002 Improvement in central nervous system functions during treatment of liver failure with albumin dialysis MARS a review of clinical biochemical and electrophysiological data Metabolic Brain Disease 17 4 463 75 doi 10 1023 A 1021986624600 PMID 12602522 S2CID 19491464 Butterworth RF 2003 Role of circulating neurotoxins in the pathogenesis of hepatic encephalopathy potential for improvement following their removal by liver assist devices Liver International 23 Suppl 3 5 9 doi 10 1034 j 1478 3231 23 s 3 1 x PMID 12950954 S2CID 25292258 Matthews SA 1922 Ammonia a causative factor in meat poisoning in Eck fistula dos Am J Physiol 59 459 460 Awad SS Sawada S Soldes OS Rich PB Klein R Alarcon WH Wang SC Bartlett RH Jan Feb 1999 Can the clearance of tumor necrosis factor alpha and interleukin 6 be enhanced using an albumin dialysate hemodiafiltration system ASAIO Journal 45 1 47 9 doi 10 1097 00002480 199901000 00011 PMID 9952006 a b Novelli G Rossi M Pretagostini R Poli L Novelli L Berloco P Ferretti G Iappelli M Cortesini R 2002 MARS Molecular Adsorbent Recirculating System experience in 34 cases of acute liver failure Liver 22 Suppl 2 43 7 doi 10 1034 j 1600 0676 2002 00008 x PMID 12220303 a b Schmidt LE Svendsen LB Sorensen VR Hansen BA Larsen FS August 2001 Cerebral blood flow velocity increases during a single treatment with the molecular adsorbents recirculating system in patients with acute on chronic liver failure Liver Transplantation 7 8 709 12 doi 10 1053 jlts 2001 26059 PMID 11510016 Sorkine P Ben Abraham R Szold O Biderman P Kidron A Merchav H Brill S Oren R July 2001 Role of the molecular adsorbent recycling system MARS in the treatment of patients with acute exacerbation of chronic liver failure Critical Care Medicine 29 7 1332 6 doi 10 1097 00003246 200107000 00006 PMID 11445681 S2CID 42225349 Spahr L Butterworth RF Fontaine S Bui L Therrien G Milette PC Lebrun LH Zayed J Leblanc A Pomier Layrargues G November 1996 Increased blood manganese in cirrhotic patients relationship to pallidal magnetic resonance signal hyperintensity and neurological symptoms Hepatology 24 5 1116 20 doi 10 1002 hep 510240523 PMID 8903385 Bergeron M Layrargues GP Butterworth RF September 1989 Aromatic and branched chain amino acids in autopsied brain tissue from cirrhotic patients with hepatic encephalopathy Metabolic Brain Disease 4 3 169 76 doi 10 1007 bf01000293 PMID 2796871 S2CID 24524207 Bergeron M Swain MS Reader TA Grondin L Butterworth RF July 1990 Effect of ammonia on brain serotonin metabolism in relation to function in the portacaval shunted rat Journal of Neurochemistry 55 1 222 9 doi 10 1111 j 1471 4159 1990 tb08842 x PMID 1693945 S2CID 2219787 Fischer JE Funovics JM Aguirre A James JH Keane JM Wesdorp RI Yoshimura N Westman T September 1975 The role of plasma amino acids in hepatic encephalopathy Surgery 78 3 276 90 PMID 807982 Loock J Stange J Mitzner S Schmidt R Gramowski A Schiffmann D Weiss D Keefer EW Gross GW June 2001 Influence of albumin dialysis MARS on neuronal network activity in vitro early results Zeitschrift fur Gastroenterologie 39 Suppl 2 28 32 doi 10 1055 s 2001 919055 PMID 16215892 Novelli G Rossi M Pretagostini R Novelli L Poli L Ferretti G Iappelli M Berloco P Cortesini R 2003 A 3 year experience with Molecular Adsorbent Recirculating System MARS our results on 63 patients with hepatic failure and color Doppler US evaluation of cerebral perfusion Liver International 23 Suppl 3 10 5 doi 10 1034 j 1478 3231 23 s 3 4 x PMID 12950955 S2CID 25173997 a b c d e f g h i j k l Heemann U Treichel U Loock J Philipp T Gerken G Malago M Klammt S Loehr M Liebe S Mitzner S Schmidt R Stange J October 2002 Albumin dialysis in cirrhosis with superimposed acute liver injury a prospective controlled study Hepatology 36 4 Pt 1 949 58 doi 10 1053 jhep 2002 36130 PMID 12297843 a b c d e Sen S Davies NA Mookerjee RP Cheshire LM Hodges SJ Williams R Jalan R September 2004 Pathophysiological effects of albumin dialysis in acute on chronic liver failure a randomized controlled study Liver Transplantation 10 9 1109 19 doi 10 1002 lt 20236 PMID 15350001 a b c d e f g Hassanein TI Tofteng F Brown RS Jr McGuire B Lynch P Mehta R Larsen FS Gornbein J Stange J Blei AT December 2007 Randomized controlled study of extracorporeal albumin dialysis for hepatic encephalopathy in advanced cirrhosis Hepatology 46 6 1853 62 doi 10 1002 hep 21930 PMID 17975845 a b Kramer L Gendo A Madl C Mullen KD Kaminski Russ K Sunder Plassmann G Schaffer A Bauer E Roth E Ferenci P July 2001 A controlled study of sorbent suspension dialysis in chronic liver disease and hepatic encephalopathy The International Journal of Artificial Organs 24 7 434 42 doi 10 1177 039139880102400707 PMID 11510914 S2CID 24205200 a b Yuan JZ Ye QF Zhao LL Ming YZ Sun H Zhu SH Huang ZF Wang MM Aug 21 2006 Preoperative risk factor analysis in orthotopic liver transplantation with pretransplant artificial liver support therapy World Journal of Gastroenterology 12 31 5055 9 doi 10 3748 wjg v12 i31 5055 PMC 4087413 PMID 16937506 Gaspari R Cavaliere F Sollazzi L Perilli V Melchionda I Agnes S Gasbarrini A Avolio AW Jan Feb 2009 Molecular adsorbent recirculating system Mars in patients with primary nonfunction and other causes of graft dysfunction after liver transplantation in the era of extended criteria donor organs Transplantation Proceedings 41 1 253 8 doi 10 1016 j transproceed 2008 10 066 hdl 10807 141915 PMID 19249528 a b Stefoni S Coli L Bolondi L Donati G Ruggeri G Feliciangeli G Piscaglia F Silvagni E Sirri M Donati G Baraldi O Soverini ML Cianciolo G Boni P Patrono D Ramazzotti E Motta R Roda A Simoni P Magliulo M Borgnino LC Ricci D Mezzopane D Cappuccilli ML February 2006 Molecular adsorbent recirculating system MARS application in liver failure clinical and hemodepurative results in 22 patients The International Journal of Artificial Organs 29 2 207 18 doi 10 1177 039139880602900207 hdl 11585 24619 PMID 16552668 S2CID 22591855 a b Di Campli C Santoro MC Gaspari R Merra G Zileri Dal Verme L Zocco MA Piscaglia AC Di Gioacchino G Novi M Santoliquido A Flore R Tondi P Proietti R Gasbarrini G Pola P Gasbarrini A Jul Aug 2005 Catholic university experience with molecular adsorbent recycling system in patients with severe liver failure Transplantation Proceedings 37 6 2547 50 doi 10 1016 j transproceed 2005 06 048 PMID 16182739 a b c Hetz H Faybik P Berlakovich G Baker A Bacher A Burghuber C Sandner SE Steltzer H Krenn CG September 2006 Molecular adsorbent recirculating system in patients with early allograft dysfunction after liver transplantation a pilot study Liver Transplantation 12 9 1357 64 doi 10 1002 lt 20804 PMID 16741899 S2CID 19946716 a b c Camus C Lavoue S Gacouin A Compagnon P Boudjema K Jacquelinet C Thomas R Le Tulzo Y December 2009 Liver transplantation avoided in patients with fulminant hepatic failure who received albumin dialysis with the molecular adsorbent recirculating system while on the waiting list impact of the duration of therapy Therapeutic Apheresis and Dialysis 13 6 549 55 doi 10 1111 j 1744 9987 2009 00708 x PMID 19954480 S2CID 24898728 Steiner C Mitzner S 2002 Experiences with MARS liver support therapy in liver failure analysis of 176 patients of the International MARS Registry Liver 22 Suppl 2 20 5 doi 10 1034 j 1600 0676 2002 00003 x PMID 12220298 Camus C Lavoue S Gacouin A Le Tulzo Y Lorho R Boudjema K Jacquelinet C Thomas R November 2006 Molecular adsorbent recirculating system dialysis in patients with acute liver failure who are assessed for liver transplantation Intensive Care Medicine 32 11 1817 25 doi 10 1007 s00134 006 0340 1 PMID 16941171 S2CID 8003937 Pares A Deulofeu R Cisneros L Escorsell A Salmeron JM Caballeria J Mas A 2009 Albumin dialysis improves hepatic encephalopathy and decreases circulating phenolic aromatic amino acids in patients with alcoholic hepatitis and severe liver failure Critical Care 13 1 R8 doi 10 1186 cc7697 PMC 2688120 PMID 19175915 a b Mitzner SR Stange J Klammt S Peszynski P Schmidt R Noldge Schomburg G February 2001 Extracorporeal detoxification using the molecular adsorbent recirculating system for critically ill patients with liver failure Journal of the American Society of Nephrology 12 Suppl 17 S75 82 doi 10 1681 ASN V12suppl 1s75 PMID 11251037 a b c d e f g h Laleman W Wilmer A Evenepoel P Elst IV Zeegers M Zaman Z Verslype C Fevery J Nevens F 2006 Effect of the molecular adsorbent recirculating system and Prometheus devices on systemic haemodynamics and vasoactive agents in patients with acute on chronic alcoholic liver failure Critical Care 10 4 R108 doi 10 1186 cc4985 PMC 1751025 PMID 16859530 a b c d e f g h Schmidt LE Wang LP Hansen BA Larsen FS March 2003 Systemic hemodynamic effects of treatment with the molecular adsorbents recirculating system in patients with hyperacute liver failure a prospective controlled trial Liver Transplantation 9 3 290 7 doi 10 1053 jlts 2003 50051 PMID 12619027 a b Catalina MV Barrio J Anaya F Salcedo M Rincon D Clemente G Banares R 2003 Hepatic and systemic haemodynamic changes after MARS in patients with acute on chronic liver failure Liver International 23 Suppl 3 39 43 doi 10 1034 j 1478 3231 23 s 3 10 x PMID 12950960 S2CID 2200601 Pares A Escorsell A Cisneros L 2002 Effects of the molecular adsorbent recirculating system MARS on vasoactive agents and systemic hemodynamics in patients with severe alcoholic hepatitis 4th ISAD 2002 Abstrac a b c d e f g Dethloff T Tofteng F Frederiksen HJ Hojskov M Hansen BA Larsen FS Apr 7 2008 Effect of Prometheus liver assist system on systemic hemodynamics in patients with cirrhosis a randomized controlled study World Journal of Gastroenterology 14 13 2065 71 doi 10 3748 wjg 14 2065 PMC 2701529 PMID 18395908 a b c d e f g h i j Mitzner SR Stange J Klammt S Risler T Erley CM Bader BD Berger ED Lauchart W Peszynski P Freytag J Hickstein H Loock J Lohr JM Liebe S Emmrich J Korten G Schmidt R May 2000 Improvement of hepatorenal syndrome with extracorporeal albumin dialysis MARS results of a prospective randomized controlled clinical trial Liver Transplantation 6 3 277 86 doi 10 1002 lt 500060326 PMID 10827226 Arroyo V May 2000 New treatments for hepatorenal syndrome Liver Transplantation 6 3 287 9 doi 10 1053 lv 2000 7569 PMID 10827227 S2CID 32465310 Versin P Hirsch Marie H Catan R 1962 Linsufissance renale circulatoire spntane du cirrhotique Son evolution Sem Hop Paris 38 3598 3602 McIntyre CW Fluck RJ Freeman JG Lambie SH November 2002 Characterization of treatment dose delivered by albumin dialysis in the treatment of acute renal failure associated with severe hepatic dysfunction Clinical Nephrology 58 5 376 83 doi 10 5414 CNP58376 PMID 12425489 a b Saich R Collins P Ala A Standish R Hodgson H May 2005 Benign recurrent intrahepatic cholestasis with secondary renal impairment treated with extracorporeal albumin dialysis European Journal of Gastroenterology amp Hepatology 17 5 585 8 doi 10 1097 00042737 200505000 00018 PMID 15827452 S2CID 23033545 Cardenas A Gines P June 2006 Therapy insight Management of hepatorenal syndrome Nature Clinical Practice Gastroenterology amp Hepatology 3 6 338 48 doi 10 1038 ncpgasthep0517 PMID 16741553 S2CID 12302583 Moreau R Lebrec D 2007 Diagnosis and treatment of acute renal failure in patients with cirrhosis Best Practice amp Research Clinical Gastroenterology 21 1 111 23 doi 10 1016 j bpg 2006 10 004 PMID 17223500 Wong F January 2007 Drug insight the role of albumin in the management of chronic liver disease Nature Clinical Practice Gastroenterology amp Hepatology 4 1 43 51 doi 10 1038 ncpgasthep0680 PMID 17203088 S2CID 23729017 a b Khuroo MS Khuroo MS Farahat KL September 2004 Molecular adsorbent recirculating system for acute and acute on chronic liver failure a meta analysis Liver Transplantation 10 9 1099 106 doi 10 1002 lt 20139 PMID 15349999 Wong F Raina N Richardson R March 2010 Molecular adsorbent recirculating system is ineffective in the management of type 1 hepatorenal syndrome in patients with cirrhosis with ascites who have failed vasoconstrictor treatment Gut 59 3 381 6 doi 10 1136 gut 2008 174615 PMID 19710033 S2CID 13455917 a b c d e f g El Banayosy A Kizner L Schueler V Bergmeier S Cobaugh D Koerfer R Jul Aug 2004 First use of the Molecular Adsorbent Recirculating System technique on patients with hypoxic liver failure after cardiogenic shock ASAIO Journal 50 4 332 7 doi 10 1097 01 MAT 0000131251 88146 CD PMID 15307543 S2CID 20982595 a b c d Montejo Gonzalez JC Catalan Gonzalez M Meneu Diaz JC Moreno Elola Olaso A De la Cruz J Moreno Gonzalez E Mar Apr 2009 Artificial liver support system in acute liver failure patients waiting liver transplantation Hepato gastroenterology 56 90 456 61 PMID 19579620 Stadlbauer V Krisper P Beuers U Haditsch B Schneditz D Jung A Putz Bankuti C Holzer H Trauner M Stauber RE Mar Apr 2007 Removal of bile acids by two different extracorporeal liver support systems in acute on chronic liver failure ASAIO Journal 53 2 187 93 doi 10 1097 01 mat 0000249852 71634 6c PMID 17413559 S2CID 8630554 Lebovics E Seif F Kim D Elhosseiny A Dworkin BM Casellas A Clark S Rosenthal WS May 1997 Pruritus in chronic hepatitis C association with high serum bile acids advanced pathology and bile duct abnormalities Digestive Diseases and Sciences 42 5 1094 9 doi 10 1023 A 1018865809556 PMID 9149069 S2CID 22902185 Jones EA Bergasa NV Dec 16 1992 The pruritus of cholestasis and the opioid system JAMA 268 23 3359 62 doi 10 1001 jama 1992 03490230089034 PMID 1333541 Bergasa NV Thomas DA Vergalla J Turner ML Jones EA 1993 Plasma from patients with the pruritus of cholestasis induces opioid receptor mediated scratching in monkeys Life Sciences 53 16 1253 7 doi 10 1016 0024 3205 93 90569 o PMID 8412484 Marchesini G Bianchi G Amodio P Salerno F Merli M Panella C Loguercio C Apolone G Niero M Abbiati R Italian Study Group for quality of life in cirrhosis January 2001 Factors associated with poor health related quality of life of patients with cirrhosis Gastroenterology 120 1 170 8 doi 10 1053 gast 2001 21193 PMID 11208726 Younossi ZM Kiwi ML Boparai N Price LL Guyatt G February 2000 Cholestatic liver diseases and health related quality of life The American Journal of Gastroenterology 95 2 497 502 doi 10 1111 j 1572 0241 2000 01774 x PMID 10685757 S2CID 1584814 Huster D Schubert C Achenbach H Caca K Mossner J Berr F June 2001 Successful clinical application of extracorporal albumin dialysis in a patient with benign recurrent intrahepatic cholestasis BRIC Zeitschrift fur Gastroenterologie 39 Suppl 2 13 4 doi 10 1055 s 2001 919024 PMID 16215886 a b De Simone P Van Nuffelen M Donckier V September 2003 Use of molecular adsorbent recirculating system for treatment of refractory pruritus Liver Transplantation 9 9 997 8 doi 10 1002 lt 500090918 PMID 12942466 Mullhaupt B Kullak Ublick GA Ambuhl PM Stocker R Renner EL April 2003 Successful use of the Molecular Adsorbent Recirculating System MARS in a patient with primary biliary cirrhosis PBC and treatment refractory pruritus Hepatology Research 25 4 442 446 doi 10 1016 s1386 6346 02 00310 8 PMID 12699855 a b Macia M Aviles J Navarro J Morales S Garcia J January 2003 Efficacy of molecular adsorbent recirculating system for the treatment of intractable pruritus in cholestasis The American Journal of Medicine 114 1 62 4 doi 10 1016 s0002 9343 02 01354 2 PMID 12543292 Majcher Peszynska J P Peszynki SC Muller 2001 Drugs in Liver Disease and During Albumin Dialysis MARS Z Gastroenterol 39 33 35 doi 10 1055 s 2001 919048 PMID 16215894 Majcher Peszynska J T Schneider S Muller P Peszynski R Mundowski A Berg S Klammt B Drewelow 2002 Elimination of Fluoroquinolones during Extracorporeal Albumin Dialysis MARS 4th ISAD Abstracts Majcher Peszynska J S Klammt EM Hehk 2000 Removal of Albumin Bound Drugs in Albumin Dialysis MARS A New Liver Support System J Hepatol Suppl 2 32 60 doi 10 1016 s0168 8278 00 80562 0 Pszynski PM J Majcher Pszynska J Stangeet 2000 Kinetics of Drugs During MARS Treatment Hepatology 32 484A doi 10 1002 hep 1840380507 Liu JP Gluud LL Als Nielsen B Gluud C 2004 Artificial and bioartificial support systems for liver failure Cochrane Database Syst Rev 1 1 CD003628 doi 10 1002 14651858 CD003628 pub2 PMC 6991941 PMID 14974025 Kjaergard LL Liu J Als Nielsen B Gluud C 2003 Artificial and bioartificial support systems for acute and acute on chronic liver failure a systematic review JAMA 289 2 217 222 doi 10 1001 jama 289 2 217 PMID 12517233 Saliba F Camus C Durand F et al 2008 Randomized controlled multicenter trial evaluating the efficacy and safety of albumin dialysis with MARS in patients with fulminant and subfulminant hepatic failure Hepatology 48 4 Suppl 1 377A doi 10 1002 hep 22615 Lai WK Haydon G Mutimer D et al 2005 The effect of molecular adsorbent recirculating system on pathophysiological parameters in patients with acute liver failure Intensive Care Med 31 11 1544 1549 doi 10 1007 s00134 005 2786 y PMID 16155752 S2CID 6642381 Lee KH Lee MK Sutedja DS et al 2005 Outcome from molecular adsorbent recycling system MARS liver dialysis following drug induced liver failure Liver Int 25 5 973 977 doi 10 1111 j 1478 3231 2005 01091 x PMID 16162155 S2CID 44868151 a b Zhou XM Miao JY Yang Y et al 2004 Clinical experience with molecular adsorbent recirculating system MARS in patients with drug induced liver failure Artificial Organs 28 5 483 486 doi 10 1046 j 1525 1594 2003 00953 x i1 PMID 15113343 Kantola T Koivusalo A Hockerstedt K et al 2008 The effect of molecular adsorbent recirculating system on survival native liver recovery and need for liver transplantation in acute liver failure patients Transpl Int 21 9 857 866 doi 10 1111 j 1432 2277 2008 00698 x PMID 18510596 Camus C Lavoue S Gacouin A et al 2009 Liver transplantation avoided in patients with fulminant hepatic failure who received albumin dialysis with the molecular adsorbent recirculating system while on the waiting list impact of the duration of therapy Ther Apher Dial 13 6 549 555 doi 10 1111 j 1744 9987 2009 00708 x PMID 19954480 S2CID 24898728 Evenepoel P Laleman W Wilmer A Claes K Kuypers D Bammens B Nevens F Vanrenterghem Y Apr 2006 Prometheus versus molecular adsorbents recirculating system comparison of efficiency in two different liver detoxification devices Artificial Organs 30 4 276 84 doi 10 1111 j 1525 1594 2006 00215 x PMID 16643386 a b Faenza S Baraldi O Bernardi M Bolondi L Coli L Cucchetti A Donati G Gozzetti F Lauro A Mancini E Pinna AD Piscaglia F Rasciti L Ravaioli M Ruggeri G Santoro A Stefoni S May 2008 Mars and Prometheus our clinical experience in acute chronic liver failure Transplantation Proceedings 40 4 1169 71 doi 10 1016 j transproceed 2008 03 069 PMID 18555140 a b Krisper P Haditsch B Stauber R Jung A Stadlbauer V Trauner M Holzer H Schneditz D Sep 2005 In vivo quantification of liver dialysis comparison of albumin dialysis and fractionated plasma separation Journal of Hepatology 43 3 451 7 doi 10 1016 j jhep 2005 02 038 PMID 16023249 a b Stadlbauer V Krisper P Aigner R Haditsch B Jung A Lackner C Stauber RE 2006 Effect of extracorporeal liver support by MARS and Prometheus on serum cytokines in acute on chronic liver failure Critical Care 10 6 R169 doi 10 1186 cc5119 PMC 1794485 PMID 17156425 Stutchfield BM Simpson K Wigmore SJ May 2011 Systematic review and meta analysis of survival following extracorporeal liver support The British Journal of Surgery 98 5 623 31 doi 10 1002 bjs 7418 PMID 21462172 S2CID 23318049 Mitzner SR Stange J Klammt S Koball S Hickstein H Reisinger EC Sep Oct 2009 Albumin dialysis MARS knowledge from 10 years of clinical investigation ASAIO Journal 55 5 498 502 doi 10 1097 mat 0b013e3181b37d86 PMID 19730006 S2CID 40839375 Hassanein T Oliver D Stange J Steiner C 2003 Albumin dialysis in cirrhosis with superimposed acute liver injury possible impact of albumin dialysis on hospitalization costs Liver International 23 Suppl 3 61 5 doi 10 1034 j 1478 3231 23 s 3 6 x PMID 12950963 S2CID 23585670 Kim WR Gross JB Jr Poterucha JJ Locke GR 3rd Dickson ER Jan 2001 Outcome of hospital care of liver disease associated with hepatitis C in the United States Hepatology 33 1 201 6 doi 10 1053 jhep 2001 20798 PMID 11124837 S2CID 21550617 Hessel FP Oct 5 2006 Economic evaluation of the artificial liver support system MARS in patients with acute on chronic liver failure Cost Effectiveness and Resource Allocation 4 16 doi 10 1186 1478 7547 4 16 PMC 1601969 PMID 17022815 Hessel FP Bramlage P Wasem J Mitzner SR Feb 2010 Cost effectiveness of the artificial liver support system MARS in patients with acute on chronic liver failure European Journal of Gastroenterology amp Hepatology 22 2 213 20 doi 10 1097 meg 0b013e3283314e48 PMID 19773666 S2CID 21972813 a b Doria C Mandala L Smith J Vitale CH Lauro A Gruttadauria S Marino IR Foglieni CS Magnone M Scott VL Apr 2003 Effect of molecular adsorbent recirculating system in hepatitis C virus related intractable pruritus Liver Transplantation 9 4 437 43 doi 10 1053 jlts 2003 50055 PMID 12682899 S2CID 28738261 Sen S Mookerjee RP Cheshire LM Davies NA Williams R Jalan R Jul 2005 Albumin dialysis reduces portal pressure acutely in patients with severe alcoholic hepatitis Journal of Hepatology 43 1 142 8 doi 10 1016 j jhep 2005 01 032 PMID 15878216 Jalan R Sen S Steiner C Kapoor D Alisa A Williams R Jan 2003 Extracorporeal liver support with molecular adsorbents recirculating system in patients with severe acute alcoholic hepatitis Journal of Hepatology 38 1 24 31 doi 10 1016 s0168 8278 02 00334 3 PMID 12480556 a b Manz T Ochs A Bisse E Strey C Grotz W 2003 Liver support a task for nephrologists Extracorporeal treatment of a patient with fulminant Wilson crisis Blood Purification 21 3 232 6 doi 10 1159 000070695 PMID 12784049 S2CID 45543585 Chen S Zhang L Shi Y Yang X Wang M 2002 Molecular Adsorbent Recirculating System clinical experience in patients with liver failure based on hepatitis B in China Liver 22 Suppl 2 48 51 doi 10 1034 j 1600 0676 2002 00009 x PMID 12220304 Sen S Ytrebo LM Rose C Fuskevaag OM Davies NA Nedredal GI Williams R Revhaug A Jalan R Mar 2004 Albumin dialysis a new therapeutic strategy for intoxication from protein bound drugs Intensive Care Medicine 30 3 496 501 doi 10 1007 s00134 003 2141 0 hdl 1866 24039 PMID 14735236 S2CID 30840312 Koivusalo AM Yildirim Y Vakkuri A Lindgren L Hockerstedt K Isoniemi H Oct 2003 Experience with albumin dialysis in five patients with severe overdoses of paracetamol Acta Anaesthesiologica Scandinavica 47 9 1145 50 doi 10 1034 j 1399 6576 2003 00190 x PMID 12969110 S2CID 12576160 Rubik J Pietraszek Jezierska E Kaminski A Skarzynska A Jozwiak S Pawlowska J Drewniak T Prokurat S Grenda R Kalicinski P Jun 2004 Successful treatment of a child with fulminant liver failure and coma caused by Amanita phalloides intoxication with albumin dialysis without liver transplantation Pediatric Transplantation 8 3 295 300 doi 10 1111 j 1399 3046 2004 00170 x PMID 15176968 S2CID 38538438 Covic A Goldsmith DJ Gusbeth Tatomir P Volovat C Dimitriu AG Cristogel F Bizo A 2003 Successful use of Molecular Absorbent Regenerating System MARS dialysis for the treatment of fulminant hepatic failure in children accidentally poisoned by toxic mushroom ingestion Liver International 23 Suppl 3 21 7 doi 10 1034 j 1478 3231 23 s 3 9 x PMID 12950957 S2CID 39904706 Faybik P Hetz H Krenn CG Baker A Germann P Berlakovich G Steininger R Steltzer H Sep 15 2003 Liver support in fulminant liver failure after hemorrhagic shock Wiener Klinische Wochenschrift 115 15 16 595 8 doi 10 1007 bf03040455 PMID 14531174 S2CID 19746467 Lahdenpera A Koivusalo AM Vakkuri A Hockerstedt K Isoniemi H Jan 2005 Value of albumin dialysis therapy in severe liver insufficiency Transplant International 17 11 717 23 doi 10 1111 j 1432 2277 2004 tb00500 x PMID 15580335 Hommann M Kasakow LB Geoghegan J Kornberg A Schotte U Fuchs D Hermann J Zintl F Scheele J Aug 2002 Application of MARS artificial liver support as bridging therapy before split liver retransplantation in a 15 month old child Pediatric Transplantation 6 4 340 3 doi 10 1034 j 1399 3046 2002 02007 x PMID 12234277 S2CID 24212977 van de Kerkhove MP de Jong KP Rijken AM de Pont AC van Gulik TM 2003 MARS treatment in posthepatectomy liver failure Liver International 23 Suppl 3 44 51 doi 10 1034 j 1478 3231 23 s 3 2 x PMID 12950961 S2CID 39425523 Pares A Cisneros L Salmeron JM Caballeria L Mas A Torras A Rodes J Jun 2004 Extracorporeal albumin dialysis a procedure for prolonged relief of intractable pruritus in patients with primary biliary cirrhosis The American Journal of Gastroenterology 99 6 1105 10 doi 10 1111 j 1572 0241 2004 30204 x PMID 15180733 S2CID 22709637 Further reading EditTandon R Froghi S September 2020 Artificial liver support systems J Gastroenterol Hepatol 36 5 1164 1179 doi 10 1111 jgh 15255 PMID 32918840 Sen S Williams R Jalan R February 2005 Emerging indications for albumin dialysis Am J Gastroenterol 100 2 468 75 doi 10 1111 j 1572 0241 2005 40864 x PMID 15667509 S2CID 20937240 Evenepoel P Maes B Wilmer A et al 2003 Detoxifying capacity and kinetics of the molecular adsorbent recycling system Contribution of the different inbuilt filters Blood Purif 21 3 244 52 doi 10 1159 000070697 PMID 12784051 S2CID 12355239 Mitzner S Klammt S Stange J Noldge Schomburg GF Schmidt R April 2005 Extracorporeal blood purification in severe liver failure with the albumin dialysis MARS impact on relevant intensive care parameters Anasthesiol Intensivmed Notfallmed Schmerzther in German 40 4 199 206 doi 10 1055 s 2004 826116 PMID 15832238 Retrieved from https en wikipedia org w index php title Liver support system amp oldid 1133617336, wikipedia, wiki, book, books, library,