fbpx
Wikipedia

Intraocular lens power calculation

The aim of an accurate intraocular lens power calculation is to provide an intraocular lens (IOL) that fits the specific needs and desires of the individual patient. The development of better instrumentation for measuring the eye's axial length (AL) and the use of more precise mathematical formulas to perform the appropriate calculations have significantly improved the accuracy with which the surgeon determines the IOL power.

In order to determine the power of intraocular lens, several values need to be known:

  • Eye's axial length (AL)
  • Corneal power (K)
  • Postoperative IOL position within the eye known as estimated lens position (ELP)
  • The anterior chamber constant: A-constant or another lens related constant

Of these parameters, the first two are measured before the implantation, the third parameter, the ELP, need to be estimated mathematically before the implantation and the last parameter is provided by the manufacturer of the intraocular lens.

Axial Length edit

The axial length (AL) is the distance between the anterior surface of the cornea and the fovea and usually measured by A-scan ultrasonography or optical coherence biometry. The AL is the most important factor in IOL calculation: A 1-mm error in AL measurement results in a refractive error of approximately 2.88 D or about 3.0-3.5 D error of IOL power in an average eye. A mean shortening of 0.25–0.33mm can translate into an error of IOL power by approximately 1 D[1]

Ultrasonography edit

In A-scan ultrasound biometry, a crystal oscillates to generate a high-frequency sound wave that penetrates into the eye. When the sound wave encounters a media interface, part of the sound wave is reflected back toward the probe. These echoes allow us to calculate the distance between the probe and various structures in the eye. Ultrasonography does not measure the distance but rather the time required for a sound pulse to travel from the cornea to the retina. The speed of sound varies in different parts of the eye. The eye is divided ultrasonographically into four components:Cornea, Anterior chamber, Lens thickness and Vitreous cavity. The velocity of sound in these compartments are 1620, 1532, 1641, 1532 m/s respectively.[2] Through normal eyes an average velocity of 1555 m/s is accepted for calculation. Modern instruments use separate sound velocities for the different eye components to obtain the total axial length. The measured transit time is converted to a distance using the formula d=t/v Where d is the distance, t is the time and v is the velocity.[1]

Two types of A-scan ultrasound biometry are currently in use. The first is contact applanation biometry. This technique requires placing an ultrasound probe on the central cornea. While this is a convenient way to determine the axial length for most normal eyes, errors in measurement almost invariably result from the probe indenting the cornea and shallowing the anterior chamber. Since the compression error is variable, it cannot be compensated for by a constant. IOL power calculations using these measurements will lead to an overestimation of the IOL power. In shorter eyes, this effect is amplified. The second type is immersion A-scan biometry, which requires placing a saline filled scleral shell between the probe and the eye. Since the probe does not exert direct pressure on the cornea, compression of the anterior chamber is avoided. A mean shortening of 0.25–0.33mm has been reported between applanation and immersion axial length measurements, which can translate into an error of IOL power by approximately 1 D. In general, immersion biometry has been shown to be more accurate than contact applanation biometry in several studies. The main limitation with the A-scan ultrasound is the poor image resolution due to the use of a relatively long, low-resolution wavelength (10 MHz) to measure a relatively short distance. In addition, variations in retinal thickness surrounding the fovea contribute to inconsistency in the final measurement.[3]

Partial coherence interferometry edit

The technique of partial coherence interferometry measures the time required for infrared light to travel to the retina. Because light travels at too high a speed to be measured directly, light interference methodology is used to determine the transit time and thus the AL. This technique does not require contact with the globe, so corneal compression artifacts are eliminated. Compared with ultrasonography, the partial coherence interferometry provides more accurate, reproducible AL measurement. However, it is difficult to obtain a measurement in the presence of a dense cataract or other media opacities, which limits the use of this technique.

Another advantage of PCI over ultrasound biometry is that the axial length measurement is performed through the visual axis since the patient is asked to fixate into the laser spot. In highly myopic or staphylomatous eyes, this can be particularly advantageous since it can sometimes be difficult to measure the true axial length through the visual axis with an ultrasound probe. PCI is also superior to ultrasound in the measurement of pseudophakic and silicone oil-filled eyes. For optical biometry, it is not as critical how the media change because the correction factor that must be applied is much smaller than in ultrasound biometry.[3] The axial length obtained from PCI may be slightly longer than that obtained from ultrasound. This is due to PCI measuring the distance from the corneal surface to the RPE while ultrasound measures to the anterior retinal surface. Therefore, many IOL measurement machines require refined IOL constants unique to their mechanism.

Axial length: Ultrasound versus Optical Biometry edit

For ultrasonography AL corresponds to the anterior retinal surface, whereas for optical biometry it corresponds to the retinal pigment epithelium (RPE)/Bruch's membrane (AL-RPE). AL can also be defined as the linear distance between the corneal surface and the inner limiting membrane (ILM) or AL-ILM. Since IOL Power Calculation formulas were developed earlier using ultrasound, the AL-ILM, each AL-RPE optical biometric measurement is converted to an AL-ILM by subtracting the retinal thickness, which is assumed to be 300 μm in all eyes.[4]

Corneal Power edit

The central corneal power is the second important factor in the calculation formula. To simplify the calculation, the cornea is assumed to be a thin spherical lens with a fixed anterior to posterior corneal curvature ratio and an index of refraction of 1.3375. Central corneal power can be measured by keratometry or corneal topography. Corneal radius of curvature relates to corneal power with the equation: r = 337.5/K.[3]

IOL power calculation formulas edit

Intraocular lens power calculation formulas fall into two major categories: regression formulas and theoretical formulas. Regression formulas are now obsolete and modern theoretic formulas are used instead.[5] The regression formulas are empiric formulas generated by averaging large numbers of postoperative clinical results (i.e. from retrospective computer analysis of data obtained from a great many patients who have undergone surgery). The most common regression formulas are the SRK and SRK II. In the 1980s SRK and SRK II were popular because they were simple to use. However, power error often resulted from the use of these formulas.

The SRK formula is calculated easily by hand as  , where   is the IOL power to be used for emmetropia,   is the IOL specific A constant,   is the average corneal refractive power (diopters), and   is the length of the eye (mm). The SRK II formula adjusts the A constant utilized depending on the axial length: increasing the A constant for short eyes and decreasing the A constant for long eyes.

Theoretical formulas are based on geometrical optics. The eye is considered a two lens system (i.e. IOL and cornea) and the predicted distance between them which is called the estimated lens position (ELP) is used to calculate the power of the IOL. All formulas require an estimation of the position that the IOL will sit in the eye, a factor known as the ELP, which is defined as the distance between the cornea and the IOL. ELP correlates with the placement of the IOL inside the eye, whether it is in the anterior chamber in the sulcus or in the capsular bag. It also varies with the implant's configuration and the location of its optical center. For example, the use of a meniscus lens calls for a smaller ELP value than a biconvex IOL.

IOL calculation formulas differ in the way they calculate ELP. In the original theoretical formula the ELP is considered a constant value of 4 mm for every lens in every patient.[5] Better results are obtained by relating the expected ELP to the axial length and corneal curvature. Modern theoretical formulas predict ELP differently based on axial length and corneal power: ELP decreases in the shorter eyes and flat corneas and increases in the longer eyes and steeper corneas. The improvements in IOL power calculation are the result of improvements in the predictability of the ELP.[2]

The best known modern formulas are SRK-T, Holladay 1, Holladay 2, Hoffer-Q and Haigis. These formulas are programmed into the IOLMaster, Lenstar and most modern ultrasonographic instruments, thus eliminating any need for regression formulas.[1]

A-constant edit

The A-constant was originally designed for the SRK equation and depends on multiple variables including IOL manufacturer, refraction index, style and placement within the eye. Because of its simplicity, the A-constant became the value used to characterize intraocular implants.

A-constants are used directly in SRK II and SRK/T formulas. The constant is a theoretical value that relates the lens power to AL and keratometry, it is not expressed in units and is specific to the design of the IOL and its intended location and orientation within the eye.

Using A-constants is practical when a decision on the implant power has to be made during surgery because the power of the lens varies in a 1:1 relationship with the A-constants: if A decreases by 1 diopter, IOL power decreases by 1 diopter also. This straight relationship adds to the simplicity and popularity of the A-constant. Other constants used in modern IOL formulas include the ACD value in Binkhorst and Hoffer-Q formulas, the a0, a1, and a2 constants of the Haigis formula, and the Surgeon factor (SF) in Holladay formulas. True anterior chamber depth (ACD) is measured between the posterior corneal surface and the anterior lens surface. This measure is not to be confused with the anterior chamber constant (ACD constant) used in IOL power calculation formulas.[2]

All lens constants are estimates, to begin with. To obtain the best possible results, it is imperative that these constants be optimized. Optimization is a process that is user-specific and incorporates the various systematic errors attributable to measurement of ocular parameters. In order to optimize a lens-constant, the user must back calculate the formula so that the actual post-operative refractive error is included. This means that one must calculate the constant so that a recalculation of the formula would predict exactly the same refractive error as actually observed.

Intraocular lens power calculation after refractive surgery edit

Cataract extraction following refractive surgery poses special problems for the patient and the surgeon because the corneal change as a result of refractive surgery complicates accurate keratometry, a key element of lens implant power calculation. After laser refractive surgery for myopia, this could result in overestimation of corneal power, underestimation of the IOL power required, and hyperopic outcomes after cataract surgery.

The difficulty arises from several factors:[1]

  • The instruments used by ophthalmologists to measure the corneal power (keratometers, corneal topographers) cannot obtain accurate measurements in eyes that have undergone corneal refractive surgery. Most manual keratometers measure at the 3-mm zone of the central cornea, which often misses the central flatter zone of effective corneal power.
  • The assumed index of refraction of the normal cornea is based on the relationship between the anterior and posterior corneal curvatures. This relationship is changed in LASIK eyes.
  • Most IOL power formulas use the axial length and keratometric reading (K) to predict the position of the IOL postoperatively (ELP). In post-LASIK eyes this causes an error in this prediction because the anterior chamber dimensions do not really change in these eyes commensurately with the much flatter K. In order to address this problem the double-K method was developed, which uses the pre-LASIK corneal power for the calculation of the ELP, and the post-LASIK corneal power for the calculation of the Vergence component of the formula.

Auditing of Results edit

Auditing of results helps to compare formulae and optimisation strategies amongst each other. Due to considerable confusion in the past, a clear set of guidelines now exists to report IOL power related data. There are six key measures that are to be reported. In recognition of the fact that comparison of ideal IOL powers is likely to be error-prone, all comparisons are done for actual or predicted refractive errors.

1.Mean Error (ME) and standard deviation (SD) in prediction.

2.Mean Absolute Error (MAE) and standard deviation (SD) in prediction.

3.The percentage of eyes ± 0.5 D from the predicted target refraction.

4.The percentage of eyes ± 1.0 D from the predicted target refraction.

5.The percentage of eyes > 2.0 D from the predicted target refraction.

6.Range of errors from maximum plus error to maximum minus error.

Software tools can be used in order to perform an audit.

See also edit

References edit

  1. ^ a b c d Basic and Clinical Science Course, Section 3: Clinical Optics (2011-2012 ed.). American Academy of Ophthalmology. 2011. pp. 211–223. ISBN 978-1615251100.
  2. ^ a b c Roger F. Steinert; David F. Chang (2010). Cataract surgery (3rd ed.). Saunders. ISBN 9781416032250.
  3. ^ a b c Lee, AC; Qazi, MA; Pepose, JS (January 2008). "Biometry and intraocular lens power calculation". Current Opinion in Ophthalmology. 19 (1): 13–7. doi:10.1097/ICU.0b013e3282f1c5ad. PMID 18090891. S2CID 11807190.
  4. ^ Pardianto G, editor. The Intraocular Lens Power Calculation. Medan: Anak Sudarti Foundation. 2022. ISBN 9786236763506 Available on: https://rsmatasmec.com/perpustakaan-digital
  5. ^ a b Myron Yanoff; Jay S. Duker (2009). Ophthalmology (3rd ed.). Mosby Elsevier. pp. 416–419. ISBN 978-0-323-04332-8.

Further reading edit

External links edit

  • "Toric Intraocular lens power calculator".
  • "Dr. Hill lens calculation materials".
  • "Online calculation of intraocular lenses using different IOL formulas".

intraocular, lens, power, calculation, accurate, intraocular, lens, power, calculation, provide, intraocular, lens, that, fits, specific, needs, desires, individual, patient, development, better, instrumentation, measuring, axial, length, more, precise, mathem. The aim of an accurate intraocular lens power calculation is to provide an intraocular lens IOL that fits the specific needs and desires of the individual patient The development of better instrumentation for measuring the eye s axial length AL and the use of more precise mathematical formulas to perform the appropriate calculations have significantly improved the accuracy with which the surgeon determines the IOL power In order to determine the power of intraocular lens several values need to be known Eye s axial length AL Corneal power K Postoperative IOL position within the eye known as estimated lens position ELP The anterior chamber constant A constant or another lens related constant Of these parameters the first two are measured before the implantation the third parameter the ELP need to be estimated mathematically before the implantation and the last parameter is provided by the manufacturer of the intraocular lens Contents 1 Axial Length 1 1 Ultrasonography 1 2 Partial coherence interferometry 1 3 Axial length Ultrasound versus Optical Biometry 2 Corneal Power 3 IOL power calculation formulas 4 A constant 5 Intraocular lens power calculation after refractive surgery 6 Auditing of Results 7 See also 8 References 9 Further reading 10 External linksAxial Length editThe axial length AL is the distance between the anterior surface of the cornea and the fovea and usually measured by A scan ultrasonography or optical coherence biometry The AL is the most important factor in IOL calculation A 1 mm error in AL measurement results in a refractive error of approximately 2 88 D or about 3 0 3 5 D error of IOL power in an average eye A mean shortening of 0 25 0 33mm can translate into an error of IOL power by approximately 1 D 1 Ultrasonography edit In A scan ultrasound biometry a crystal oscillates to generate a high frequency sound wave that penetrates into the eye When the sound wave encounters a media interface part of the sound wave is reflected back toward the probe These echoes allow us to calculate the distance between the probe and various structures in the eye Ultrasonography does not measure the distance but rather the time required for a sound pulse to travel from the cornea to the retina The speed of sound varies in different parts of the eye The eye is divided ultrasonographically into four components Cornea Anterior chamber Lens thickness and Vitreous cavity The velocity of sound in these compartments are 1620 1532 1641 1532 m s respectively 2 Through normal eyes an average velocity of 1555 m s is accepted for calculation Modern instruments use separate sound velocities for the different eye components to obtain the total axial length The measured transit time is converted to a distance using the formula d t v Where d is the distance t is the time and v is the velocity 1 Two types of A scan ultrasound biometry are currently in use The first is contact applanation biometry This technique requires placing an ultrasound probe on the central cornea While this is a convenient way to determine the axial length for most normal eyes errors in measurement almost invariably result from the probe indenting the cornea and shallowing the anterior chamber Since the compression error is variable it cannot be compensated for by a constant IOL power calculations using these measurements will lead to an overestimation of the IOL power In shorter eyes this effect is amplified The second type is immersion A scan biometry which requires placing a saline filled scleral shell between the probe and the eye Since the probe does not exert direct pressure on the cornea compression of the anterior chamber is avoided A mean shortening of 0 25 0 33mm has been reported between applanation and immersion axial length measurements which can translate into an error of IOL power by approximately 1 D In general immersion biometry has been shown to be more accurate than contact applanation biometry in several studies The main limitation with the A scan ultrasound is the poor image resolution due to the use of a relatively long low resolution wavelength 10 MHz to measure a relatively short distance In addition variations in retinal thickness surrounding the fovea contribute to inconsistency in the final measurement 3 Partial coherence interferometry edit The technique of partial coherence interferometry measures the time required for infrared light to travel to the retina Because light travels at too high a speed to be measured directly light interference methodology is used to determine the transit time and thus the AL This technique does not require contact with the globe so corneal compression artifacts are eliminated Compared with ultrasonography the partial coherence interferometry provides more accurate reproducible AL measurement However it is difficult to obtain a measurement in the presence of a dense cataract or other media opacities which limits the use of this technique Another advantage of PCI over ultrasound biometry is that the axial length measurement is performed through the visual axis since the patient is asked to fixate into the laser spot In highly myopic or staphylomatous eyes this can be particularly advantageous since it can sometimes be difficult to measure the true axial length through the visual axis with an ultrasound probe PCI is also superior to ultrasound in the measurement of pseudophakic and silicone oil filled eyes For optical biometry it is not as critical how the media change because the correction factor that must be applied is much smaller than in ultrasound biometry 3 The axial length obtained from PCI may be slightly longer than that obtained from ultrasound This is due to PCI measuring the distance from the corneal surface to the RPE while ultrasound measures to the anterior retinal surface Therefore many IOL measurement machines require refined IOL constants unique to their mechanism Axial length Ultrasound versus Optical Biometry edit For ultrasonography AL corresponds to the anterior retinal surface whereas for optical biometry it corresponds to the retinal pigment epithelium RPE Bruch s membrane AL RPE AL can also be defined as the linear distance between the corneal surface and the inner limiting membrane ILM or AL ILM Since IOL Power Calculation formulas were developed earlier using ultrasound the AL ILM each AL RPE optical biometric measurement is converted to an AL ILM by subtracting the retinal thickness which is assumed to be 300 mm in all eyes 4 Corneal Power editThe central corneal power is the second important factor in the calculation formula To simplify the calculation the cornea is assumed to be a thin spherical lens with a fixed anterior to posterior corneal curvature ratio and an index of refraction of 1 3375 Central corneal power can be measured by keratometry or corneal topography Corneal radius of curvature relates to corneal power with the equation r 337 5 K 3 IOL power calculation formulas editIntraocular lens power calculation formulas fall into two major categories regression formulas and theoretical formulas Regression formulas are now obsolete and modern theoretic formulas are used instead 5 The regression formulas are empiric formulas generated by averaging large numbers of postoperative clinical results i e from retrospective computer analysis of data obtained from a great many patients who have undergone surgery The most common regression formulas are the SRK and SRK II In the 1980s SRK and SRK II were popular because they were simple to use However power error often resulted from the use of these formulas The SRK formula is calculated easily by hand as P A 0 9 K 2 5 L displaystyle P A 0 9K 2 5L nbsp where P displaystyle P nbsp is the IOL power to be used for emmetropia A displaystyle A nbsp is the IOL specific A constant K displaystyle K nbsp is the average corneal refractive power diopters and L displaystyle L nbsp is the length of the eye mm The SRK II formula adjusts the A constant utilized depending on the axial length increasing the A constant for short eyes and decreasing the A constant for long eyes Theoretical formulas are based on geometrical optics The eye is considered a two lens system i e IOL and cornea and the predicted distance between them which is called the estimated lens position ELP is used to calculate the power of the IOL All formulas require an estimation of the position that the IOL will sit in the eye a factor known as the ELP which is defined as the distance between the cornea and the IOL ELP correlates with the placement of the IOL inside the eye whether it is in the anterior chamber in the sulcus or in the capsular bag It also varies with the implant s configuration and the location of its optical center For example the use of a meniscus lens calls for a smaller ELP value than a biconvex IOL IOL calculation formulas differ in the way they calculate ELP In the original theoretical formula the ELP is considered a constant value of 4 mm for every lens in every patient 5 Better results are obtained by relating the expected ELP to the axial length and corneal curvature Modern theoretical formulas predict ELP differently based on axial length and corneal power ELP decreases in the shorter eyes and flat corneas and increases in the longer eyes and steeper corneas The improvements in IOL power calculation are the result of improvements in the predictability of the ELP 2 The best known modern formulas are SRK T Holladay 1 Holladay 2 Hoffer Q and Haigis These formulas are programmed into the IOLMaster Lenstar and most modern ultrasonographic instruments thus eliminating any need for regression formulas 1 A constant editThe A constant was originally designed for the SRK equation and depends on multiple variables including IOL manufacturer refraction index style and placement within the eye Because of its simplicity the A constant became the value used to characterize intraocular implants A constants are used directly in SRK II and SRK T formulas The constant is a theoretical value that relates the lens power to AL and keratometry it is not expressed in units and is specific to the design of the IOL and its intended location and orientation within the eye Using A constants is practical when a decision on the implant power has to be made during surgery because the power of the lens varies in a 1 1 relationship with the A constants if A decreases by 1 diopter IOL power decreases by 1 diopter also This straight relationship adds to the simplicity and popularity of the A constant Other constants used in modern IOL formulas include the ACD value in Binkhorst and Hoffer Q formulas the a0 a1 and a2 constants of the Haigis formula and the Surgeon factor SF in Holladay formulas True anterior chamber depth ACD is measured between the posterior corneal surface and the anterior lens surface This measure is not to be confused with the anterior chamber constant ACD constant used in IOL power calculation formulas 2 All lens constants are estimates to begin with To obtain the best possible results it is imperative that these constants be optimized Optimization is a process that is user specific and incorporates the various systematic errors attributable to measurement of ocular parameters In order to optimize a lens constant the user must back calculate the formula so that the actual post operative refractive error is included This means that one must calculate the constant so that a recalculation of the formula would predict exactly the same refractive error as actually observed Intraocular lens power calculation after refractive surgery editCataract extraction following refractive surgery poses special problems for the patient and the surgeon because the corneal change as a result of refractive surgery complicates accurate keratometry a key element of lens implant power calculation After laser refractive surgery for myopia this could result in overestimation of corneal power underestimation of the IOL power required and hyperopic outcomes after cataract surgery The difficulty arises from several factors 1 The instruments used by ophthalmologists to measure the corneal power keratometers corneal topographers cannot obtain accurate measurements in eyes that have undergone corneal refractive surgery Most manual keratometers measure at the 3 mm zone of the central cornea which often misses the central flatter zone of effective corneal power The assumed index of refraction of the normal cornea is based on the relationship between the anterior and posterior corneal curvatures This relationship is changed in LASIK eyes Most IOL power formulas use the axial length and keratometric reading K to predict the position of the IOL postoperatively ELP In post LASIK eyes this causes an error in this prediction because the anterior chamber dimensions do not really change in these eyes commensurately with the much flatter K In order to address this problem the double K method was developed which uses the pre LASIK corneal power for the calculation of the ELP and the post LASIK corneal power for the calculation of the Vergence component of the formula Auditing of Results editAuditing of results helps to compare formulae and optimisation strategies amongst each other Due to considerable confusion in the past a clear set of guidelines now exists to report IOL power related data There are six key measures that are to be reported In recognition of the fact that comparison of ideal IOL powers is likely to be error prone all comparisons are done for actual or predicted refractive errors 1 Mean Error ME and standard deviation SD in prediction 2 Mean Absolute Error MAE and standard deviation SD in prediction 3 The percentage of eyes 0 5 D from the predicted target refraction 4 The percentage of eyes 1 0 D from the predicted target refraction 5 The percentage of eyes gt 2 0 D from the predicted target refraction 6 Range of errors from maximum plus error to maximum minus error Software tools can be used in order to perform an audit See also editIntraocular LensReferences edit a b c d Basic and Clinical Science Course Section 3 Clinical Optics 2011 2012 ed American Academy of Ophthalmology 2011 pp 211 223 ISBN 978 1615251100 a b c Roger F Steinert David F Chang 2010 Cataract surgery 3rd ed Saunders ISBN 9781416032250 a b c Lee AC Qazi MA Pepose JS January 2008 Biometry and intraocular lens power calculation Current Opinion in Ophthalmology 19 1 13 7 doi 10 1097 ICU 0b013e3282f1c5ad PMID 18090891 S2CID 11807190 Pardianto G editor The Intraocular Lens Power Calculation Medan Anak Sudarti Foundation 2022 ISBN 9786236763506 Available on https rsmatasmec com perpustakaan digital a b Myron Yanoff Jay S Duker 2009 Ophthalmology 3rd ed Mosby Elsevier pp 416 419 ISBN 978 0 323 04332 8 Further reading editPardianto G The Intraocular Power Calculation Medan Anak Sudarti Foundation 2022 ISBN 978 623 6763 50 6 e ISBN 978 623 6763 51 3 PDF External links edit Toric Intraocular lens power calculator Dr Hill lens calculation materials Online calculation of intraocular lenses using different IOL formulas Retrieved from https en wikipedia org w index php title Intraocular lens power calculation amp oldid 1160150790, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.