The hemocytometer (or haemocytometer) is a counting-chamber device originally designed and usually used for counting blood cells.[1]

A hemocytometer. The two semi-reflective rectangles are the counting chambers.
Loading a chamber
Hemocytometer grid (see table)

The hemocytometer was invented by Louis-Charles Malassez and consists of a thick glass microscope slide with a rectangular indentation that creates a precision volume chamber. This chamber is engraved with a laser-etched grid of perpendicular lines. The device is carefully crafted so that the area bounded by the lines is known, and the depth of the chamber is also known. By observing a defined area of the grid, it is therefore possible to count the number of cells or particles in a specific volume of fluid, and thereby calculate the concentration of cells in the fluid overall. A well used type of hemocytometer is the Neubauer counting chamber.[2]

Other types of hemocytometers with different rulings are in use for different applications. Fuchs-Rosenthal rulings, commonly used for spinal fluid counting, Howard Mold rulings used for mold on food and food packaging products, McMaster Egg Slide ruling used for counting microbial eggs in fecal material, Nageotte Chamber ruling for counting low levels of white cells in white cell-reduced platelet components, Palmer Nanoplankton ruling for counting smaller plankters. Petroff-Hausser counter using Improved Neubauer rulings is used for bacteria or sperm counts, and is offered with varying chamber depths. The Sedgwick-Rafter Cell ruling in a hemocytometer is primarily designed for use in the microscopy of drinking water.

Principles edit

The gridded area of the Improved Neubauer ruled hemocytometer consists of nine 1 x 1 mm (1 mm2) squares. These are subdivided in three directions; 0.25 x 0.25 mm (0.0625 mm2), 0.25 x 0.20 mm (0.05 mm2) and 0.20 x 0.20 mm (0.04 mm2). The central square is further subdivided into 0.05 x 0.05 mm (0.0025 mm2) squares. The raised edges of the hemocytometer hold the coverslip 0.1 mm off the marked grid, giving each square a defined volume (see figure on the right).[3]

Dimensions Area Volume at 0.1 mm depth
1 x 1 mm 1 mm2 100 nL
0.25 x 0.25 mm 0.0625 mm2 6.25 nL
0.25 x 0.20 mm 0.05 mm2 5 nL
0.20 x 0.20 mm 0.04 mm2 4 nL
0.05 x 0.05 mm 0.0025 mm2 0.25 nL

Usage edit

To use the hemocytometer, first make sure that the special coverslip provided with the counting chamber is properly positioned on the surface of the counting chamber. When the two glass surfaces are in proper contact Newton's rings can be observed. If so, the cell suspension is applied to the edge of the coverslip to be sucked into the void by capillary action which completely fills the chamber with the sample. The number of cells in the chamber can be determined by direct counting using a microscope, and visually distinguishable cells can be differentially counted. The number of cells in the chamber is used to calculate the concentration or density of the cells in the mixture the sample comes from. It is the number of cells in the chamber divided by the chamber's volume, which is known from the start, taking account of any dilutions and counting shortcuts:

  [4]

 

where the volume of the diluted sample (after dilution) divided by the volume of the original mixture in the sample (before dilution) is the dilution factor. For example, if the volume of the original mixture was 20μL and it was diluted once (by adding 20μL dilutant), then the second term in parentheses is 40μL/20μL. The volume of the squares counted is the one shown in the table at the top, depending on the size (see figure on the right). The number of cells counted is the sum of all cells counted across squares in one chamber. The proportion of the cells counted applies if not all inner squares within a set square are counted (i.e., if only 4 out of the 20 in a corner square are counted, then this term will equal 0.2). When counting large squares with a volume of 100 nanoliter (nL), a multiplication by 10000 leads to the desired cell count per milliliter.

 
The parts of the hemocytometer (as viewed from the side) are identified.

For most applications, the four large corner squares are only used. The cells that are on or touching the top and left lines are counted, but the ones on or touching the right or bottom lines are ignored.[5]

Requirements edit

 
Empty hemocytometer grid at 100x power.

The original suspension must be mixed thoroughly before taking a sample. This ensures the sample is representative, and not just an artifact of the particular region of the original mixture it was drawn from.

An appropriate dilution of the mixture with regard to the number of cells to be counted should be used. If the sample is not diluted enough, the cells will be too crowded and difficult to count. If it is too dilute, the sample size will not be enough to make strong inferences about the concentration in the original mixture.

By performing a redundant test on a second chamber, the results can be compared. If they differ more than 2 times the counting error (square root of the count[6]), the method of taking the sample may be unreliable (e.g., the original mixture is not mixed thoroughly).

The counting chamber should be filled by capillary action after the chamber lid (special cover-slip with certified thickness and flatness) has been put in place. This avoids the risk that the cells could sediment/stick to the glass or some volume to evaporate before the cover-slip is placed on top and resulting in an overestimation of the cell concentration. Sedimentation is less of a problem with bacteria but evaporation, more prevalent in low-humidity air-conditioned laboratories, still has to be minimized.

Applications edit

  • Blood counts: for patients with abnormal blood cells, where automated counters don't perform well.
  • Sperm counts
  • Urine microscopy
  • Cell culture: when subculturing or recording cell growth over time.
  • Beer brewing: for the preparation of the yeast.
  • Phytoplankton cell counting
  • Cell processing for downstream analysis: accurate cell numbers are needed in many tests (PCR, flow cytometry), while some others require high cell viability.
  • Measurement of cell size: in a micrograph, the real cell size can be inferred by scaling it to the width of a hemocytometer square, which is known.[7]

References edit

  1. ^ Absher, Marlene (1973). "Hemocytometer Counting". Tissue Culture. pp. 395–397. doi:10.1016/B978-0-12-427150-0.50098-X. ISBN 9780124271500.
  2. ^ "Fig. 8. Views of an improved Neubauer ruling hemocytometer slide. (A) Top view..." ResearchGate. Retrieved 29 March 2018.
  3. ^ "Hemocytometer square size • Hemocytometer". April 11, 2013.
  4. ^ "Cell Counting with Neubauer Chamber - Basic Hemocytometer Usage, Óscar Bastidas" (PDF).
  5. ^ Strober W (2001). "Monitoring cell growth". In Coligan JE, Bierer BE, Margulies DH, Sherach EM, Strober W (eds.). Current Protocols in Immunology. Vol. 5. USA: John Wiley & Sons. p. A.2A.1. doi:10.1002/0471142735.ima03as21. ISBN 0471142735. PMID 18432653.
  6. ^ Shapiro, Howard (31 January 2001). "very rare events: how low can you go?".
  7. ^ . Archived from the original on 2012-06-29. Retrieved 2013-06-02.

External links edit

  • Online exercise - Yeast counts in Neubauer improved or Thoma counting chambers
  • Hemocytometer calculator online
  • Types of Hemocytometers

hemocytometer, this, article, multiple, issues, please, help, improve, discuss, these, issues, talk, page, learn, when, remove, these, template, messages, this, article, needs, additional, citations, verification, please, help, improve, this, article, adding, . This article has multiple issues Please help improve it or discuss these issues on the talk page Learn how and when to remove these template messages This article needs additional citations for verification Please help improve this article by adding citations to reliable sources Unsourced material may be challenged and removed Find sources Hemocytometer news newspapers books scholar JSTOR December 2009 Learn how and when to remove this template message This article is written like a manual or guide Please help rewrite this article and remove advice or instruction February 2022 Learn how and when to remove this template message The hemocytometer or haemocytometer is a counting chamber device originally designed and usually used for counting blood cells 1 A hemocytometer The two semi reflective rectangles are the counting chambers Loading a chamberHemocytometer grid see table The hemocytometer was invented by Louis Charles Malassez and consists of a thick glass microscope slide with a rectangular indentation that creates a precision volume chamber This chamber is engraved with a laser etched grid of perpendicular lines The device is carefully crafted so that the area bounded by the lines is known and the depth of the chamber is also known By observing a defined area of the grid it is therefore possible to count the number of cells or particles in a specific volume of fluid and thereby calculate the concentration of cells in the fluid overall A well used type of hemocytometer is the Neubauer counting chamber 2 Other types of hemocytometers with different rulings are in use for different applications Fuchs Rosenthal rulings commonly used for spinal fluid counting Howard Mold rulings used for mold on food and food packaging products McMaster Egg Slide ruling used for counting microbial eggs in fecal material Nageotte Chamber ruling for counting low levels of white cells in white cell reduced platelet components Palmer Nanoplankton ruling for counting smaller plankters Petroff Hausser counter using Improved Neubauer rulings is used for bacteria or sperm counts and is offered with varying chamber depths The Sedgwick Rafter Cell ruling in a hemocytometer is primarily designed for use in the microscopy of drinking water Contents 1 Principles 2 Usage 3 Requirements 4 Applications 5 References 6 External linksPrinciples editThe gridded area of the Improved Neubauer ruled hemocytometer consists of nine 1 x 1 mm 1 mm2 squares These are subdivided in three directions 0 25 x 0 25 mm 0 0625 mm2 0 25 x 0 20 mm 0 05 mm2 and 0 20 x 0 20 mm 0 04 mm2 The central square is further subdivided into 0 05 x 0 05 mm 0 0025 mm2 squares The raised edges of the hemocytometer hold the coverslip 0 1 mm off the marked grid giving each square a defined volume see figure on the right 3 Dimensions Area Volume at 0 1 mm depth1 x 1 mm 1 mm2 100 nL0 25 x 0 25 mm 0 0625 mm2 6 25 nL0 25 x 0 20 mm 0 05 mm2 5 nL0 20 x 0 20 mm 0 04 mm2 4 nL0 05 x 0 05 mm 0 0025 mm2 0 25 nLUsage editTo use the hemocytometer first make sure that the special coverslip provided with the counting chamber is properly positioned on the surface of the counting chamber When the two glass surfaces are in proper contact Newton s rings can be observed If so the cell suspension is applied to the edge of the coverslip to be sucked into the void by capillary action which completely fills the chamber with the sample The number of cells in the chamber can be determined by direct counting using a microscope and visually distinguishable cells can be differentially counted The number of cells in the chamber is used to calculate the concentration or density of the cells in the mixture the sample comes from It is the number of cells in the chamber divided by the chamber s volume which is known from the start taking account of any dilutions and counting shortcuts concentration of cells in original mixture number of cells counted proportion of chamber counted volume of squares counted volume of diluted sample volume of original mixture in sample displaystyle mbox concentration of cells in original mixture left frac mbox number of cells counted mbox proportion of chamber counted mbox volume of squares counted right left frac mbox volume of diluted sample mbox volume of original mixture in sample right nbsp 4 cells mL number of cells counted dilution factor number of large squares counted 10000 displaystyle mbox cells mL left frac mbox number of cells counted mbox dilution factor mbox number of large squares counted right times 10000 nbsp where the volume of the diluted sample after dilution divided by the volume of the original mixture in the sample before dilution is the dilution factor For example if the volume of the original mixture was 20mL and it was diluted once by adding 20mL dilutant then the second term in parentheses is 40mL 20mL The volume of the squares counted is the one shown in the table at the top depending on the size see figure on the right The number of cells counted is the sum of all cells counted across squares in one chamber The proportion of the cells counted applies if not all inner squares within a set square are counted i e if only 4 out of the 20 in a corner square are counted then this term will equal 0 2 When counting large squares with a volume of 100 nanoliter nL a multiplication by 10000 leads to the desired cell count per milliliter nbsp The parts of the hemocytometer as viewed from the side are identified For most applications the four large corner squares are only used The cells that are on or touching the top and left lines are counted but the ones on or touching the right or bottom lines are ignored 5 Requirements edit nbsp Empty hemocytometer grid at 100x power The original suspension must be mixed thoroughly before taking a sample This ensures the sample is representative and not just an artifact of the particular region of the original mixture it was drawn from An appropriate dilution of the mixture with regard to the number of cells to be counted should be used If the sample is not diluted enough the cells will be too crowded and difficult to count If it is too dilute the sample size will not be enough to make strong inferences about the concentration in the original mixture By performing a redundant test on a second chamber the results can be compared If they differ more than 2 times the counting error square root of the count 6 the method of taking the sample may be unreliable e g the original mixture is not mixed thoroughly The counting chamber should be filled by capillary action after the chamber lid special cover slip with certified thickness and flatness has been put in place This avoids the risk that the cells could sediment stick to the glass or some volume to evaporate before the cover slip is placed on top and resulting in an overestimation of the cell concentration Sedimentation is less of a problem with bacteria but evaporation more prevalent in low humidity air conditioned laboratories still has to be minimized Applications editBlood counts for patients with abnormal blood cells where automated counters don t perform well Sperm counts Urine microscopy Cell culture when subculturing or recording cell growth over time Beer brewing for the preparation of the yeast Phytoplankton cell counting Cell processing for downstream analysis accurate cell numbers are needed in many tests PCR flow cytometry while some others require high cell viability Measurement of cell size in a micrograph the real cell size can be inferred by scaling it to the width of a hemocytometer square which is known 7 References edit Absher Marlene 1973 Hemocytometer Counting Tissue Culture pp 395 397 doi 10 1016 B978 0 12 427150 0 50098 X ISBN 9780124271500 Fig 8 Views of an improved Neubauer ruling hemocytometer slide A Top view ResearchGate Retrieved 29 March 2018 Hemocytometer square size Hemocytometer April 11 2013 Cell Counting with Neubauer Chamber Basic Hemocytometer Usage oscar Bastidas PDF Strober W 2001 Monitoring cell growth In Coligan JE Bierer BE Margulies DH Sherach EM Strober W eds Current Protocols in Immunology Vol 5 USA John Wiley amp Sons p A 2A 1 doi 10 1002 0471142735 ima03as21 ISBN 0471142735 PMID 18432653 Shapiro Howard 31 January 2001 very rare events how low can you go Measuring cell size with a hemocytometer Archived from the original on 2012 06 29 Retrieved 2013 06 02 External links editOnline exercise Yeast counts in Neubauer improved or Thoma counting chambers Hemocytometer calculator online Types of Hemocytometers Retrieved from https en wikipedia org w index php title Hemocytometer amp oldid 1148726326, wikipedia, wiki, book, books, library,