Conductivity (or specific conductance) of an electrolyte solution is a measure of its ability to conduct electricity. The SI unit of conductivity is siemens per meter (S/m).

Conductivity measurements are used routinely in many industrial and environmental applications as a fast, inexpensive and reliable way of measuring the ionic content in a solution.[1] For example, the measurement of product conductivity is a typical way to monitor and continuously trend the performance of water purification systems.

The electrolytic conductivity of ultra-high purity water increases as a function of temperature (T) due to the higher dissociation of H2O in H+ and OH with T.

In many cases, conductivity is linked directly to the total dissolved solids (TDS).

High quality deionized water has a conductivity of
at 25 °C.
This corresponds to a specific resistivity of
.[2]
The preparation of salt solutions often takes place in unsealed beakers. In this case the conductivity of purified water often is 10 to 20 times higher. A discussion can be found below.


Typical drinking water is in the range of 200–800 μS/cm, while sea water is about 50 mS/cm[3] (or 0.05 S/cm).

Conductivity is traditionally determined by connecting the electrolyte in a Wheatstone bridge. Dilute solutions follow Kohlrausch's Laws of concentration dependence and additivity of ionic contributions. Lars Onsager gave a theoretical explanation of Kohlrausch's law by extending Debye–Hückel theory.

Units edit

The SI unit of conductivity is S/m and, unless otherwise qualified, it refers to 25 °C. More generally encountered is the traditional unit of μS/cm.

The commonly used standard cell has a width of 1 cm, and thus for very pure water in equilibrium with air would have a resistance of about 106 ohms, known as a megohm. Ultra-pure water could achieve 18 megohms or more. Thus in the past, megohm-cm was used, sometimes abbreviated to "megohm". Sometimes, conductivity is given in "microsiemens" (omitting the distance term in the unit). While this is an error, it can often be assumed to be equal to the traditional μS/cm. Often, by typographic limitations μS/cm is expressed as uS/cm.

The conversion of conductivity to the total dissolved solids depends on the chemical composition of the sample and can vary between 0.54 and 0.96. Typically, the conversion is done assuming that the solid is sodium chloride; 1 μS/cm is then equivalent to about 0.64 mg of NaCl per kg of water.

Molar conductivity has the SI unit S m2 mol−1. Older publications use the unit Ω−1 cm2 mol−1.

Measurement edit

 
Principle of the measurement

The electrical conductivity of a solution of an electrolyte is measured by determining the resistance of the solution between two flat or cylindrical electrodes separated by a fixed distance.[4] An alternating voltage is generally used in order to minimize water electrolysis.[citation needed] The resistance is measured by a conductivity meter. Typical frequencies used are in the range 1–3 kHz. The dependence on the frequency is usually small,[5] but may become appreciable at very high frequencies, an effect known as the Debye–Falkenhagen effect.

A wide variety of instrumentation is commercially available.[6] Most commonly, two types of electrode sensors are used, electrode-based sensors and inductive sensors. Electrode sensors with a static design are suitable for low and moderate conductivities, and exist in various types, having either two or four electrodes, where electrodes can be arrange oppositely, flat or in a cylinder.[7] Electrode cells with a flexible design, where the distance between two oppositely arranged electrodes can be varied, offer high accuracy and can also be used for the measurement of highly conductive media.[8] Inductive sensors are suitable for harsh chemical conditions but require larger sample volumes than electrode sensors.[9] Conductivity sensors are typically calibrated with KCl solutions of known conductivity. Electrolytic conductivity is highly temperature dependent but many commercial systems offer automatic temperature correction. Tables of reference conductivities are available for many common solutions.[10]

Definitions edit

Resistance, R, is proportional to the distance, l, between the electrodes and is inversely proportional to the cross-sectional area of the sample, A (noted S on the Figure above). Writing ρ (rho) for the specific resistance, or resistivity.

 

In practice the conductivity cell is calibrated by using solutions of known specific resistance, ρ*, so the individual quantities l and A need not be known precisely, but only their ratio.[11] If the resistance of the calibration solution is R*, a cell-constant, defined as the ratio of l and A (C = lA), is derived.

 

The specific conductance (conductivity), κ (kappa) is the reciprocal of the specific resistance.

 

Conductivity is also temperature-dependent. Sometimes the conductance (reciprocical of the resistance) is denoted as G = 1R. Then the specific conductance κ (kappa) is:

 

Theory edit

The specific conductance of a solution containing one electrolyte depends on the concentration of the electrolyte. Therefore, it is convenient to divide the specific conductance by concentration. This quotient, termed molar conductivity, is denoted by Λm

 

Strong electrolytes edit

Strong electrolytes are hypothesized to dissociate completely in solution. The conductivity of a solution of a strong electrolyte at low concentration follows Kohlrausch's Law

 

where Λ0
m
is known as the limiting molar conductivity, K is an empirical constant and c is the electrolyte concentration. (Limiting here means "at the limit of the infinite dilution".) In effect, the observed conductivity of a strong electrolyte becomes directly proportional to concentration, at sufficiently low concentrations i.e. when

 

As the concentration is increased however, the conductivity no longer rises in proportion. Moreover, Kohlrausch also found that the limiting conductivity of an electrolyte;

λ0
+
and λ0
are the limiting molar conductivities of the individual ions.

The following table gives values for the limiting molar conductivities for some selected ions.[12]

Table of limiting ion conductivity in water at 298 K (approx. 25 °C)[12]
Cations λ0
+
/ mS m2 mol−1
Cations λ0
+
/ mS m2 mol−1
Anions λ0
/ mS m2 mol−1
Anions λ0
/ mS m2 mol−1
H+ 34.982 Ba2+ 12.728 OH 19.8 SO2−
4
15.96
Li+ 3.869 Mg2+ 10.612 Cl 7.634 C
2
O2−
4
7.4
Na+ 5.011 La3+ 20.88 Br 7.84 HC
2
O
4
4.306[13]
K+ 7.352 Rb+ 7.64 I 7.68 HCOO 5.6
NH+
4
7.34 Cs+ 7.68 NO
3
7.144 CO2−
3
7.2
Ag+ 6.192 Be2+ 4.50 CH3COO 4.09 HSO2−
3
5.0
Ca2+ 11.90 ClO
4
6.80 SO2−
3
7.2
Co(NH
3
)3+
6
10.2 F 5.50

An interpretation of these results was based on the theory of Debye and Hückel, yielding the Debye–Hückel–Onsager theory:[14]

 

where A and B are constants that depend only on known quantities such as temperature, the charges on the ions and the dielectric constant and viscosity of the solvent. As the name suggests, this is an extension of the Debye–Hückel theory, due to Onsager. It is very successful for solutions at low concentration.

Weak electrolytes edit

A weak electrolyte is one that is never fully dissociated (there are a mixture of ions and complete molecules in equilibrium). In this case there is no limit of dilution below which the relationship between conductivity and concentration becomes linear. Instead, the solution becomes ever more fully dissociated at weaker concentrations, and for low concentrations of "well behaved" weak electrolytes, the degree of dissociation of the weak electrolyte becomes proportional to the inverse square root of the concentration.

Typical weak electrolytes are weak acids and weak bases. The concentration of ions in a solution of a weak electrolyte is less than the concentration of the electrolyte itself. For acids and bases the concentrations can be calculated when the value or values of the acid dissociation constant are known.

For a monoprotic acid, HA, obeying the inverse square root law, with a dissociation constant Ka, an explicit expression for the conductivity as a function of concentration, c, known as Ostwald's dilution law, can be obtained.

 

Various solvents exhibit the same dissociation if the ratio of relative permittivities equals the ratio cubic roots of concentrations of the electrolytes (Walden's rule).

Higher concentrations edit

Both Kohlrausch's law and the Debye–Hückel–Onsager equation break down as the concentration of the electrolyte increases above a certain value. The reason for this is that as concentration increases the average distance between cation and anion decreases, so that there is more interactions between close ions. Whether this constitutes ion association is a moot point. However, it has often been assumed that cation and anion interact to form an ion pair. So, an "ion-association" constant K, can be derived for the association equilibrium between ions A+ and B:

A+ + B ⇌ A+B   with   K = [A+B]/[A+] [B]

Davies describes the results of such calculations in great detail, but states that K should not necessarily be thought of as a true equilibrium constant, rather, the inclusion of an "ion-association" term is useful in extending the range of good agreement between theory and experimental conductivity data.[15] Various attempts have been made to extend Onsager's treatment to more concentrated solutions.[16]

The existence of a so-called conductance minimum in solvents having the relative permittivity under 60 has proved to be a controversial subject as regards interpretation. Fuoss and Kraus suggested that it is caused by the formation of ion triplets,[17] and this suggestion has received some support recently.[18][19]

Other developments on this topic have been done by Theodore Shedlovsky,[20] E. Pitts,[21] R. M. Fuoss,[22][23] Fuoss and Shedlovsky,[24] Fuoss and Onsager.[25][26]

Mixed solvents systems edit

The limiting equivalent conductivity of solutions based on mixed solvents like water alcohol has minima depending on the nature of alcohol. For methanol the minimum is at 15 molar % water,[20][27][28] and for the ethanol at 6 molar % water.[29]

Conductivity versus temperature edit

Generally the conductivity of a solution increases with temperature, as the mobility of the ions increases. For comparison purposes reference values are reported at an agreed temperature, usually 298 K (≈ 25 °C or 77 °F), although occasionally 20 °C (68 °F) is used. So called 'compensated' measurements are made at a convenient temperature but the value reported is a calculated value of the expected value of conductivity of the solution, as if it had been measured at the reference temperature. Basic compensation is normally done by assuming a linear increase of conductivity versus temperature of typically 2% per kelvin.[30][31] This value is broadly applicable for most salts at room temperature. Determination of the precise temperature coefficient for a specific solution is simple and instruments are typically capable of applying the derived coefficient (i.e. other than 2%).

Measurements of conductivity   versus temperature can be used to determine the activation energy  , using the Arrhenius equation:[32]

 

where   is the exponential prefactor, R the gas constant, and T the absolute temperature in Kelvin.

Solvent isotopic effect edit

The change in conductivity due to the isotope effect for deuterated electrolytes is sizable.[33]

Applications edit

Despite the difficulty of theoretical interpretation, measured conductivity is a good indicator of the presence or absence of conductive ions in solution, and measurements are used extensively in many industries.[34] For example, conductivity measurements are used to monitor quality in public water supplies, in hospitals, in boiler water and industries that depend on water quality such as brewing. This type of measurement is not ion-specific; it can sometimes be used to determine the amount of total dissolved solids (TDS) if the composition of the solution and its conductivity behavior are known.[1] Conductivity measurements made to determine water purity will not respond to non conductive contaminants (many organic compounds fall into this category), therefore additional purity tests may be required depending on application.

Applications of TDS measurements are not limited to industrial use; many people use TDS as an indicator of the purity of their drinking water. Additionally, aquarium enthusiasts are concerned with TDS, both for freshwater and salt water aquariums. Many fish and invertebrates require quite narrow parameters for dissolved solids. Especially for successful breeding of some invertebrates normally kept in freshwater aquariums—snails and shrimp primarily—brackish water with higher TDS, specifically higher salinity, water is required. While the adults of a given species may thrive in freshwater, this is not always true for the young and some species will not breed at all in non-brackish water.

Sometimes, conductivity measurements are linked with other methods to increase the sensitivity of detection of specific types of ions. For example, in the boiler water technology, the boiler blowdown is continuously monitored for "cation conductivity", which is the conductivity of the water after it has been passed through a cation exchange resin. This is a sensitive method of monitoring anion impurities in the boiler water in the presence of excess cations (those of the alkalizing agent usually used for water treatment). The sensitivity of this method relies on the high mobility of H+ in comparison with the mobility of other cations or anions. Beyond cation conductivity, there are analytical instruments designed to measure Degas conductivity, where conductivity is measured after dissolved carbon dioxide has been removed from the sample, either through reboiling or dynamic degassing.

Conductivity detectors are commonly used with ion chromatography.[35]

Conductivity of purified water in electrochemical experiments edit

The electronic conductivity of purified distilled water in electrochemical laboratory settings at room temperature is often between 0.05 and 1 μS/cm. Environmental influences during the peparation of salt solutions as gas absorption due to storing the water in an unsealed beaker may immediately increase the conductivity from   and lead to values between 0.5 and 1  .
When distilled water is heated during the preparation of salt solutions, the conductivity increases even without adding salt. This is often not taken into account.


 
Temperature dependence of the electronic conductivity of purified distilled water. The gray area indicates the margin of error in the measurements.

In a typical experiment under the fume hood in an unsealed beaker the conductivity of purified water increases typically non linearly from values below 1 μS/cm to values close 3.5 μS/cm at   (data on GitHub). This temperature dependence has to be taken into account particularly in dilute salt solutions.

See also edit

References edit

  1. ^ a b Gray, James R. (2004). "Conductivity Analyzers and Their Application". In Down, R. D.; Lehr, J. H. (eds.). Environmental Instrumentation and Analysis Handbook. Wiley. pp. 491–510. ISBN 978-0-471-46354-2. Retrieved 10 May 2009.
  2. ^ Light, Truman; Licht, Stuart; Bevilaqua, Anthony; Morash, Kenneth (2004). "The Fundamental Conductivity and Resistivity of Water". Electrochemical and Solid-State Letters. 8 (1): E16–E19.
  3. ^ "Water Conductivity". Lenntech. Retrieved 5 January 2013.
  4. ^ Bockris, J. O'M.; Reddy, A.K.N; Gamboa-Aldeco, M. (1998). Modern Electrochemistry (2nd. ed.). Springer. ISBN 0-306-45555-2. Retrieved 10 May 2009.
  5. ^ Marija Bešter-Rogač and Dušan Habe, "Modern Advances in Electrical Conductivity Measurements of Solutions", Acta Chim. Slov. 2006, 53, 391–395 (pdf)
  6. ^ Boyes, W. (2002). Instrumentation Reference Book (3rd. ed.). Butterworth-Heinemann. ISBN 0-7506-7123-8. Retrieved 10 May 2009.
  7. ^ Gray, p 495
  8. ^ Doppelhammer, Nikolaus; Pellens, Nick; Martens, Johan; Kirschhock, Christine E. A.; Jakoby, Bernhard; Reichel, Erwin K. (27 October 2020). "Moving Electrode Impedance Spectroscopy for Accurate Conductivity Measurements of Corrosive Ionic Media". ACS Sensors. 5 (11): 3392–3397. doi:10.1021/acssensors.0c01465. PMC 7706010. PMID 33107724.
  9. ^ Ghosh, Arun K. (2013). Introduction to measurements and instrumentation (4th ed., Eastern economy ed.). Delhi: PH Learning. ISBN 978-81-203-4625-3. OCLC 900392417.
  10. ^ (PDF). EXW Foxboro. 3 October 1999. Archived from the original (PDF) on 7 September 2012. Retrieved 5 January 2013.
  11. ^ "ASTM D1125 - 95(2005) Standard Test Methods for Electrical Conductivity and Resistivity of Water". Retrieved 12 May 2009.
  12. ^ a b Adamson, Arthur W. (1973). Textbook of Physical Chemistry. London: Academic Press inc. p. 512.
  13. ^ Bešter-Rogač, M.; Tomšič, M.; Barthel, J.; Neueder, R.; Apelblat, A. (1 January 2002). "Conductivity Studies of Dilute Aqueous Solutions of Oxalic Acid and Neutral Oxalates of Sodium, Potassium, Cesium, and Ammonium from 5 to 35 °C". Journal of Solution Chemistry. 31 (1): 1–18. doi:10.1023/A:1014805417286. ISSN 1572-8927. S2CID 92641871.
  14. ^ Wright, M.R. (2007). An Introduction to Aqueous Electrolyte Solutions. Wiley. ISBN 978-0-470-84293-5.
  15. ^ Davies, C. W. (1962). Ion Association. London: Butterworths.
  16. ^ Miyoshi, K. (1973). "Comparison of the Conductance Equations of Fuoss–Onsager, Fuoss–Hsia and Pitts with the Data of Bis(2,9-dimethyl-1,10-phenanthroline)Cu(I) Perchlorate". Bull. Chem. Soc. Jpn. 46 (2): 426–430. doi:10.1246/bcsj.46.426.
  17. ^ Fuoss, R. M.; Kraus, C. A. (1935). "Properties of Electrolytic Solutions. XV. Thermodynamic Properties of Very Weak Electrolytes". J. Am. Chem. Soc. 57: 1–4. doi:10.1021/ja01304a001.
  18. ^ Weingärtner, H.; Weiss, V. C.; Schröer, W. (2000). "Ion association and electrical conductance minimum in Debye–Hückel-based theories of the hard sphere ionic fluid". J. Chem. Phys. 113 (2): 762–. Bibcode:2000JChPh.113..762W. doi:10.1063/1.481822.
  19. ^ Schröer, W.; Weingärtner, H. (2004). "Structure and criticality of ionic fluids" (PDF). Pure Appl. Chem. 76 (1): 19–27. doi:10.1351/pac200476010019. S2CID 39716065.
  20. ^ a b Shedlovsky, Theodore (1932). "The Electrolytic Conductivity of some Uni-Univalent Electrolytes in Water at 25°". Journal of the American Chemical Society. American Chemical Society (ACS). 54 (4): 1411–1428. doi:10.1021/ja01343a020. ISSN 0002-7863.
  21. ^ Pitts, E.; Coulson, Charles Alfred (1953). "An extension of the theory of the conductivity and viscosity of electrolyte solutions". Proc. R. Soc. A217 (1128): 43. Bibcode:1953RSPSA.217...43P. doi:10.1098/rspa.1953.0045. S2CID 123363978.
  22. ^ Fuoss, Raymond M. (1958). "Conductance of Ionophores". Journal of the American Chemical Society. American Chemical Society (ACS). 80 (12): 3163. doi:10.1021/ja01545a064. ISSN 0002-7863.
  23. ^ Fuoss, Raymond M. (1959). "Conductance of Dilute Solutions of 1-1 Electrolytes1". Journal of the American Chemical Society. American Chemical Society (ACS). 81 (11): 2659–2662. doi:10.1021/ja01520a016. ISSN 0002-7863.
  24. ^ Fuoss, Raymond M.; Shedlovsky, Theodore. (1949). "Extrapolation of Conductance Data for Weak Electrolytes". Journal of the American Chemical Society. American Chemical Society (ACS). 71 (4): 1496–1498. doi:10.1021/ja01172a507. ISSN 0002-7863.
  25. ^ Fuoss, Raymond M.; Onsager, Lars (1964). "The Conductance of Symmetrical Electrolytes.1aIV. Hydrodynamic and Osmotic Terms in the Relaxation Field". The Journal of Physical Chemistry. American Chemical Society (ACS). 68 (1): 1–8. doi:10.1021/j100783a001. ISSN 0022-3654.
  26. ^ Fuoss, Raymond M.; Onsager, Lars; Skinner, James F. (1965). "The Conductance of Symmetrical Electrolytes. V. The Conductance Equation1,2". The Journal of Physical Chemistry. American Chemical Society (ACS). 69 (8): 2581–2594. doi:10.1021/j100892a017. ISSN 0022-3654.
  27. ^ Shedlovsky, Theodore; Kay, Robert L. (1956). "The Ionization Constant of Acetic Acid in Water–Methanol Mixtures at 25° from Conductance Measurements". The Journal of Physical Chemistry. American Chemical Society (ACS). 60 (2): 151–155. doi:10.1021/j150536a003. ISSN 0022-3654.
  28. ^ Strehlow, H. (1960). "Der Einfluß von Wasser auf die Äquivalentleitfähigkeit von HCl in Methanol". Zeitschrift für Physikalische Chemie. Walter de Gruyter GmbH. 24 (3–4): 240–248. doi:10.1524/zpch.1960.24.3_4.240. ISSN 0942-9352.
  29. ^ Bezman, Irving I.; Verhoek, Frank H. (1945). "The Conductance of Hydrogen Chloride and Ammonium Chloride in Ethanol-Water Mixtures". Journal of the American Chemical Society. American Chemical Society (ACS). 67 (8): 1330–1334. doi:10.1021/ja01224a035. ISSN 0002-7863.
  30. ^ "NIST Special Publication 260-142 Primary Standards and Standard Reference Materials for Electrolytic Conductivity" (PDF). U.S. DEPARTMENT OF COMMERCE, Technology Administration, National Institute of Standards and Technology. p. 5.
  31. ^ R. A. Robinson and R. H. Stokes, Electrolyte Solutions, 3 ld ed., Butterworths, London (1959).
  32. ^ Petrowsky, Matt; Frech, Roger (30 April 2009). "Temperature Dependence of Ion Transport: The Compensated Arrhenius Equation". The Journal of Physical Chemistry B. 113 (17): 5996–6000. doi:10.1021/jp810095g. ISSN 1520-6106. PMID 19338318.
  33. ^ Biswas, Ranjit (1997). "Limiting Ionic Conductance of Symmetrical, Rigid Ions in Aqueous Solutions: Temperature Dependence and Solvent Isotope Effects". Journal of the American Chemical Society. 119 (25): 5946–5953. doi:10.1021/ja970118o.
  34. ^ (PDF). Aquarius Technologies Pty Ltd. Archived from the original (PDF) on 12 September 2009.
  35. ^ . Archived from the original on 20 August 2009. Retrieved 17 May 2009.

Further reading edit

  • Hans Falkenhagen, Theorie der Elektrolyte, S. Hirzel Verlag, Leipzig, 1971
  • Friedman, Harold L. (1965). "Relaxation Term of the Limiting Law of the Conductance of Electrolyte Mixtures". The Journal of Chemical Physics. 42 (2): 462–469. Bibcode:1965JChPh..42..462F. doi:10.1063/1.1695956.
  • Conductivity of concentrated solutions of electrolytes in methyl and ethyl alcohols
  • Concentrated solutions and ionic cloud model
  • H. L. Friedman, F. Franks, Aqueous Simple Electrolytes Solutions

conductivity, electrolytic, conductivity, specific, conductance, electrolyte, solution, measure, ability, conduct, electricity, unit, conductivity, siemens, meter, conductivity, measurements, used, routinely, many, industrial, environmental, applications, fast. Conductivity or specific conductance of an electrolyte solution is a measure of its ability to conduct electricity The SI unit of conductivity is siemens per meter S m Conductivity measurements are used routinely in many industrial and environmental applications as a fast inexpensive and reliable way of measuring the ionic content in a solution 1 For example the measurement of product conductivity is a typical way to monitor and continuously trend the performance of water purification systems The electrolytic conductivity of ultra high purity water increases as a function of temperature T due to the higher dissociation of H2O in H and OH with T In many cases conductivity is linked directly to the total dissolved solids TDS High quality deionized water has a conductivity of k 0 05501 0 0001 m S c m displaystyle mathrm kappa 0 05501 pm 0 0001 frac mu S cm at 25 C This corresponds to a specific resistivity of r 18 18 0 03 M W c m displaystyle rho 18 18 pm 0 03 mathrm M Omega cdot cm 2 The preparation of salt solutions often takes place in unsealed beakers In this case the conductivity of purified water often is 10 to 20 times higher A discussion can be found below Typical drinking water is in the range of 200 800 mS cm while sea water is about 50 mS cm 3 or 0 05 S cm Conductivity is traditionally determined by connecting the electrolyte in a Wheatstone bridge Dilute solutions follow Kohlrausch s Laws of concentration dependence and additivity of ionic contributions Lars Onsager gave a theoretical explanation of Kohlrausch s law by extending Debye Huckel theory Contents 1 Units 2 Measurement 3 Definitions 4 Theory 4 1 Strong electrolytes 4 2 Weak electrolytes 4 3 Higher concentrations 4 4 Mixed solvents systems 4 5 Conductivity versus temperature 4 6 Solvent isotopic effect 5 Applications 6 Conductivity of purified water in electrochemical experiments 7 See also 8 References 9 Further readingUnits editThe SI unit of conductivity is S m and unless otherwise qualified it refers to 25 C More generally encountered is the traditional unit of mS cm The commonly used standard cell has a width of 1 cm and thus for very pure water in equilibrium with air would have a resistance of about 106 ohms known as a megohm Ultra pure water could achieve 18 megohms or more Thus in the past megohm cm was used sometimes abbreviated to megohm Sometimes conductivity is given in microsiemens omitting the distance term in the unit While this is an error it can often be assumed to be equal to the traditional mS cm Often by typographic limitations mS cm is expressed as uS cm The conversion of conductivity to the total dissolved solids depends on the chemical composition of the sample and can vary between 0 54 and 0 96 Typically the conversion is done assuming that the solid is sodium chloride 1 mS cm is then equivalent to about 0 64 mg of NaCl per kg of water Molar conductivity has the SI unit S m2 mol 1 Older publications use the unit W 1 cm2 mol 1 Measurement editMain article Electrical conductivity meter nbsp Principle of the measurementThe electrical conductivity of a solution of an electrolyte is measured by determining the resistance of the solution between two flat or cylindrical electrodes separated by a fixed distance 4 An alternating voltage is generally used in order to minimize water electrolysis citation needed The resistance is measured by a conductivity meter Typical frequencies used are in the range 1 3 kHz The dependence on the frequency is usually small 5 but may become appreciable at very high frequencies an effect known as the Debye Falkenhagen effect A wide variety of instrumentation is commercially available 6 Most commonly two types of electrode sensors are used electrode based sensors and inductive sensors Electrode sensors with a static design are suitable for low and moderate conductivities and exist in various types having either two or four electrodes where electrodes can be arrange oppositely flat or in a cylinder 7 Electrode cells with a flexible design where the distance between two oppositely arranged electrodes can be varied offer high accuracy and can also be used for the measurement of highly conductive media 8 Inductive sensors are suitable for harsh chemical conditions but require larger sample volumes than electrode sensors 9 Conductivity sensors are typically calibrated with KCl solutions of known conductivity Electrolytic conductivity is highly temperature dependent but many commercial systems offer automatic temperature correction Tables of reference conductivities are available for many common solutions 10 Definitions editResistance R is proportional to the distance l between the electrodes and is inversely proportional to the cross sectional area of the sample A noted S on the Figure above Writing r rho for the specific resistance or resistivity R r l A displaystyle R rho frac l A nbsp In practice the conductivity cell is calibrated by using solutions of known specific resistance r so the individual quantities l and A need not be known precisely but only their ratio 11 If the resistance of the calibration solution is R a cell constant defined as the ratio of l and A C l A is derived R r C displaystyle R rho times C nbsp The specific conductance conductivity k kappa is the reciprocal of the specific resistance k 1 r C R displaystyle kappa frac 1 rho frac C R nbsp Conductivity is also temperature dependent Sometimes the conductance reciprocical of the resistance is denoted as G 1 R Then the specific conductance k kappa is k C G displaystyle kappa C times G nbsp Theory editThe specific conductance of a solution containing one electrolyte depends on the concentration of the electrolyte Therefore it is convenient to divide the specific conductance by concentration This quotient termed molar conductivity is denoted by Lm L m k c displaystyle Lambda mathrm m frac kappa c nbsp Strong electrolytes edit Strong electrolytes are hypothesized to dissociate completely in solution The conductivity of a solution of a strong electrolyte at low concentration follows Kohlrausch s Law L m L m 0 K c displaystyle Lambda mathrm m Lambda mathrm m 0 K sqrt c nbsp where L0m is known as the limiting molar conductivity K is an empirical constant and c is the electrolyte concentration Limiting here means at the limit of the infinite dilution In effect the observed conductivity of a strong electrolyte becomes directly proportional to concentration at sufficiently low concentrations i e when L m 0 K c displaystyle Lambda mathrm m 0 gg K sqrt c nbsp As the concentration is increased however the conductivity no longer rises in proportion Moreover Kohlrausch also found that the limiting conductivity of an electrolyte l0 and l0 are the limiting molar conductivities of the individual ions The following table gives values for the limiting molar conductivities for some selected ions 12 Table of limiting ion conductivity in water at 298 K approx 25 C 12 Cations l0 mS m2 mol 1 Cations l0 mS m2 mol 1 Anions l0 mS m2 mol 1 Anions l0 mS m2 mol 1H 34 982 Ba2 12 728 OH 19 8 SO2 4 15 96Li 3 869 Mg2 10 612 Cl 7 634 C2 O2 4 7 4Na 5 011 La3 20 88 Br 7 84 HC2 O 4 4 306 13 K 7 352 Rb 7 64 I 7 68 HCOO 5 6NH 4 7 34 Cs 7 68 NO 3 7 144 CO2 3 7 2Ag 6 192 Be2 4 50 CH3COO 4 09 HSO2 3 5 0Ca2 11 90 ClO 4 6 80 SO2 3 7 2Co NH3 3 6 10 2 F 5 50An interpretation of these results was based on the theory of Debye and Huckel yielding the Debye Huckel Onsager theory 14 L m L m 0 A B L m 0 c displaystyle Lambda mathrm m Lambda mathrm m 0 left A B Lambda mathrm m 0 right sqrt c nbsp where A and B are constants that depend only on known quantities such as temperature the charges on the ions and the dielectric constant and viscosity of the solvent As the name suggests this is an extension of the Debye Huckel theory due to Onsager It is very successful for solutions at low concentration Weak electrolytes edit A weak electrolyte is one that is never fully dissociated there are a mixture of ions and complete molecules in equilibrium In this case there is no limit of dilution below which the relationship between conductivity and concentration becomes linear Instead the solution becomes ever more fully dissociated at weaker concentrations and for low concentrations of well behaved weak electrolytes the degree of dissociation of the weak electrolyte becomes proportional to the inverse square root of the concentration Typical weak electrolytes are weak acids and weak bases The concentration of ions in a solution of a weak electrolyte is less than the concentration of the electrolyte itself For acids and bases the concentrations can be calculated when the value or values of the acid dissociation constant are known For a monoprotic acid HA obeying the inverse square root law with a dissociation constant Ka an explicit expression for the conductivity as a function of concentration c known as Ostwald s dilution law can be obtained 1 L m 1 L m 0 L m c K a L m 0 2 displaystyle frac 1 Lambda mathrm m frac 1 Lambda mathrm m 0 frac Lambda mathrm m c K mathrm a left Lambda mathrm m 0 right 2 nbsp Various solvents exhibit the same dissociation if the ratio of relative permittivities equals the ratio cubic roots of concentrations of the electrolytes Walden s rule Higher concentrations edit Both Kohlrausch s law and the Debye Huckel Onsager equation break down as the concentration of the electrolyte increases above a certain value The reason for this is that as concentration increases the average distance between cation and anion decreases so that there is more interactions between close ions Whether this constitutes ion association is a moot point However it has often been assumed that cation and anion interact to form an ion pair So an ion association constant K can be derived for the association equilibrium between ions A and B A B A B with K A B A B Davies describes the results of such calculations in great detail but states that K should not necessarily be thought of as a true equilibrium constant rather the inclusion of an ion association term is useful in extending the range of good agreement between theory and experimental conductivity data 15 Various attempts have been made to extend Onsager s treatment to more concentrated solutions 16 The existence of a so called conductance minimum in solvents having the relative permittivity under 60 has proved to be a controversial subject as regards interpretation Fuoss and Kraus suggested that it is caused by the formation of ion triplets 17 and this suggestion has received some support recently 18 19 Other developments on this topic have been done by Theodore Shedlovsky 20 E Pitts 21 R M Fuoss 22 23 Fuoss and Shedlovsky 24 Fuoss and Onsager 25 26 Mixed solvents systems edit The limiting equivalent conductivity of solutions based on mixed solvents like water alcohol has minima depending on the nature of alcohol For methanol the minimum is at 15 molar water 20 27 28 and for the ethanol at 6 molar water 29 Conductivity versus temperature edit Generally the conductivity of a solution increases with temperature as the mobility of the ions increases For comparison purposes reference values are reported at an agreed temperature usually 298 K 25 C or 77 F although occasionally 20 C 68 F is used So called compensated measurements are made at a convenient temperature but the value reported is a calculated value of the expected value of conductivity of the solution as if it had been measured at the reference temperature Basic compensation is normally done by assuming a linear increase of conductivity versus temperature of typically 2 per kelvin 30 31 This value is broadly applicable for most salts at room temperature Determination of the precise temperature coefficient for a specific solution is simple and instruments are typically capable of applying the derived coefficient i e other than 2 Measurements of conductivity s displaystyle sigma nbsp versus temperature can be used to determine the activation energy E A displaystyle E A nbsp using the Arrhenius equation 32 s s 0 e E a R T displaystyle sigma sigma 0 e E a RT nbsp where s 0 displaystyle sigma 0 nbsp is the exponential prefactor R the gas constant and T the absolute temperature in Kelvin Solvent isotopic effect edit The change in conductivity due to the isotope effect for deuterated electrolytes is sizable 33 Applications editDespite the difficulty of theoretical interpretation measured conductivity is a good indicator of the presence or absence of conductive ions in solution and measurements are used extensively in many industries 34 For example conductivity measurements are used to monitor quality in public water supplies in hospitals in boiler water and industries that depend on water quality such as brewing This type of measurement is not ion specific it can sometimes be used to determine the amount of total dissolved solids TDS if the composition of the solution and its conductivity behavior are known 1 Conductivity measurements made to determine water purity will not respond to non conductive contaminants many organic compounds fall into this category therefore additional purity tests may be required depending on application Applications of TDS measurements are not limited to industrial use many people use TDS as an indicator of the purity of their drinking water Additionally aquarium enthusiasts are concerned with TDS both for freshwater and salt water aquariums Many fish and invertebrates require quite narrow parameters for dissolved solids Especially for successful breeding of some invertebrates normally kept in freshwater aquariums snails and shrimp primarily brackish water with higher TDS specifically higher salinity water is required While the adults of a given species may thrive in freshwater this is not always true for the young and some species will not breed at all in non brackish water Sometimes conductivity measurements are linked with other methods to increase the sensitivity of detection of specific types of ions For example in the boiler water technology the boiler blowdown is continuously monitored for cation conductivity which is the conductivity of the water after it has been passed through a cation exchange resin This is a sensitive method of monitoring anion impurities in the boiler water in the presence of excess cations those of the alkalizing agent usually used for water treatment The sensitivity of this method relies on the high mobility of H in comparison with the mobility of other cations or anions Beyond cation conductivity there are analytical instruments designed to measure Degas conductivity where conductivity is measured after dissolved carbon dioxide has been removed from the sample either through reboiling or dynamic degassing Conductivity detectors are commonly used with ion chromatography 35 Conductivity of purified water in electrochemical experiments editThe electronic conductivity of purified distilled water in electrochemical laboratory settings at room temperature is often between 0 05 and 1 mS cm Environmental influences during the peparation of salt solutions as gas absorption due to storing the water in an unsealed beaker may immediately increase the conductivity from 0 055 m S c m displaystyle mathrm 0 055 mu S cm nbsp and lead to values between 0 5 and 1 m S c m displaystyle mathrm mu S cm nbsp When distilled water is heated during the preparation of salt solutions the conductivity increases even without adding salt This is often not taken into account nbsp Temperature dependence of the electronic conductivity of purified distilled water The gray area indicates the margin of error in the measurements In a typical experiment under the fume hood in an unsealed beaker the conductivity of purified water increases typically non linearly from values below 1 mS cm to values close 3 5 mS cm at 95 0 C displaystyle mathrm 95 0 C nbsp data on GitHub This temperature dependence has to be taken into account particularly in dilute salt solutions See also editEinstein relation kinetic theory Born equation Debye Falkenhagen effect Law of dilution Ion transport number Ionic atmosphere Wien effect Conductimetric titration methods to determine the equivalence pointReferences edit a b Gray James R 2004 Conductivity Analyzers and Their Application In Down R D Lehr J H eds Environmental Instrumentation and Analysis Handbook Wiley pp 491 510 ISBN 978 0 471 46354 2 Retrieved 10 May 2009 Light Truman Licht Stuart Bevilaqua Anthony Morash Kenneth 2004 The Fundamental Conductivity and Resistivity of Water Electrochemical and Solid State Letters 8 1 E16 E19 Water Conductivity Lenntech Retrieved 5 January 2013 Bockris J O M Reddy A K N Gamboa Aldeco M 1998 Modern Electrochemistry 2nd ed Springer ISBN 0 306 45555 2 Retrieved 10 May 2009 Marija Bester Rogac and Dusan Habe Modern Advances in Electrical Conductivity Measurements of Solutions Acta Chim Slov 2006 53 391 395 pdf Boyes W 2002 Instrumentation Reference Book 3rd ed Butterworth Heinemann ISBN 0 7506 7123 8 Retrieved 10 May 2009 Gray p 495 Doppelhammer Nikolaus Pellens Nick Martens Johan Kirschhock Christine E A Jakoby Bernhard Reichel Erwin K 27 October 2020 Moving Electrode Impedance Spectroscopy for Accurate Conductivity Measurements of Corrosive Ionic Media ACS Sensors 5 11 3392 3397 doi 10 1021 acssensors 0c01465 PMC 7706010 PMID 33107724 Ghosh Arun K 2013 Introduction to measurements and instrumentation 4th ed Eastern economy ed Delhi PH Learning ISBN 978 81 203 4625 3 OCLC 900392417 Conductivity ordering guide PDF EXW Foxboro 3 October 1999 Archived from the original PDF on 7 September 2012 Retrieved 5 January 2013 ASTM D1125 95 2005 Standard Test Methods for Electrical Conductivity and Resistivity of Water Retrieved 12 May 2009 a b Adamson Arthur W 1973 Textbook of Physical Chemistry London Academic Press inc p 512 Bester Rogac M Tomsic M Barthel J Neueder R Apelblat A 1 January 2002 Conductivity Studies of Dilute Aqueous Solutions of Oxalic Acid and Neutral Oxalates of Sodium Potassium Cesium and Ammonium from 5 to 35 C Journal of Solution Chemistry 31 1 1 18 doi 10 1023 A 1014805417286 ISSN 1572 8927 S2CID 92641871 Wright M R 2007 An Introduction to Aqueous Electrolyte Solutions Wiley ISBN 978 0 470 84293 5 Davies C W 1962 Ion Association London Butterworths Miyoshi K 1973 Comparison of the Conductance Equations of Fuoss Onsager Fuoss Hsia and Pitts with the Data of Bis 2 9 dimethyl 1 10 phenanthroline Cu I Perchlorate Bull Chem Soc Jpn 46 2 426 430 doi 10 1246 bcsj 46 426 Fuoss R M Kraus C A 1935 Properties of Electrolytic Solutions XV Thermodynamic Properties of Very Weak Electrolytes J Am Chem Soc 57 1 4 doi 10 1021 ja01304a001 Weingartner H Weiss V C Schroer W 2000 Ion association and electrical conductance minimum in Debye Huckel based theories of the hard sphere ionic fluid J Chem Phys 113 2 762 Bibcode 2000JChPh 113 762W doi 10 1063 1 481822 Schroer W Weingartner H 2004 Structure and criticality of ionic fluids PDF Pure Appl Chem 76 1 19 27 doi 10 1351 pac200476010019 S2CID 39716065 a b Shedlovsky Theodore 1932 The Electrolytic Conductivity of some Uni Univalent Electrolytes in Water at 25 Journal of the American Chemical Society American Chemical Society ACS 54 4 1411 1428 doi 10 1021 ja01343a020 ISSN 0002 7863 Pitts E Coulson Charles Alfred 1953 An extension of the theory of the conductivity and viscosity of electrolyte solutions Proc R Soc A217 1128 43 Bibcode 1953RSPSA 217 43P doi 10 1098 rspa 1953 0045 S2CID 123363978 Fuoss Raymond M 1958 Conductance of Ionophores Journal of the American Chemical Society American Chemical Society ACS 80 12 3163 doi 10 1021 ja01545a064 ISSN 0002 7863 Fuoss Raymond M 1959 Conductance of Dilute Solutions of 1 1 Electrolytes1 Journal of the American Chemical Society American Chemical Society ACS 81 11 2659 2662 doi 10 1021 ja01520a016 ISSN 0002 7863 Fuoss Raymond M Shedlovsky Theodore 1949 Extrapolation of Conductance Data for Weak Electrolytes Journal of the American Chemical Society American Chemical Society ACS 71 4 1496 1498 doi 10 1021 ja01172a507 ISSN 0002 7863 Fuoss Raymond M Onsager Lars 1964 The Conductance of Symmetrical Electrolytes 1aIV Hydrodynamic and Osmotic Terms in the Relaxation Field The Journal of Physical Chemistry American Chemical Society ACS 68 1 1 8 doi 10 1021 j100783a001 ISSN 0022 3654 Fuoss Raymond M Onsager Lars Skinner James F 1965 The Conductance of Symmetrical Electrolytes V The Conductance Equation1 2 The Journal of Physical Chemistry American Chemical Society ACS 69 8 2581 2594 doi 10 1021 j100892a017 ISSN 0022 3654 Shedlovsky Theodore Kay Robert L 1956 The Ionization Constant of Acetic Acid in Water Methanol Mixtures at 25 from Conductance Measurements The Journal of Physical Chemistry American Chemical Society ACS 60 2 151 155 doi 10 1021 j150536a003 ISSN 0022 3654 Strehlow H 1960 Der Einfluss von Wasser auf die Aquivalentleitfahigkeit von HCl in Methanol Zeitschrift fur Physikalische Chemie Walter de Gruyter GmbH 24 3 4 240 248 doi 10 1524 zpch 1960 24 3 4 240 ISSN 0942 9352 Bezman Irving I Verhoek Frank H 1945 The Conductance of Hydrogen Chloride and Ammonium Chloride in Ethanol Water Mixtures Journal of the American Chemical Society American Chemical Society ACS 67 8 1330 1334 doi 10 1021 ja01224a035 ISSN 0002 7863 NIST Special Publication 260 142 Primary Standards and Standard Reference Materials for Electrolytic Conductivity PDF U S DEPARTMENT OF COMMERCE Technology Administration National Institute of Standards and Technology p 5 R A Robinson and R H Stokes Electrolyte Solutions 3 ld ed Butterworths London 1959 Petrowsky Matt Frech Roger 30 April 2009 Temperature Dependence of Ion Transport The Compensated Arrhenius Equation The Journal of Physical Chemistry B 113 17 5996 6000 doi 10 1021 jp810095g ISSN 1520 6106 PMID 19338318 Biswas Ranjit 1997 Limiting Ionic Conductance of Symmetrical Rigid Ions in Aqueous Solutions Temperature Dependence and Solvent Isotope Effects Journal of the American Chemical Society 119 25 5946 5953 doi 10 1021 ja970118o Electrolytic conductivity measurement Theory and practice PDF Aquarius Technologies Pty Ltd Archived from the original PDF on 12 September 2009 Detectors for ion exchange chromatography Archived from the original on 20 August 2009 Retrieved 17 May 2009 Further reading edit nbsp Wikimedia Commons has media related to Conductometry Hans Falkenhagen Theorie der Elektrolyte S Hirzel Verlag Leipzig 1971 Friedman Harold L 1965 Relaxation Term of the Limiting Law of the Conductance of Electrolyte Mixtures The Journal of Chemical Physics 42 2 462 469 Bibcode 1965JChPh 42 462F doi 10 1063 1 1695956 Conductivity of concentrated solutions of electrolytes in methyl and ethyl alcohols Concentrated solutions and ionic cloud model H L Friedman F Franks Aqueous Simple Electrolytes Solutions Retrieved from https en wikipedia org w index php title Conductivity electrolytic amp oldid 1197046566, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.