fbpx
Wikipedia

4Pi microscope

A 4Pi microscope is a laser scanning fluorescence microscope with an improved axial resolution. With it the typical range of the axial resolution of 500–700 nm can be improved to 100–150 nm, which corresponds to an almost spherical focal spot with 5–7 times less volume than that of standard confocal microscopy.[1]

Working principle edit

The improvement in resolution is achieved by using two opposing objective lenses, which both are focused to the same geometrical location. Also the difference in optical path length through each of the two objective lenses is carefully aligned to be minimal. By this method, molecules residing in the common focal area of both objectives can be illuminated coherently from both sides and the reflected or emitted light can also be collected coherently, i.e. coherent superposition of emitted light on the detector is possible. The solid angle   that is used for illumination and detection is increased and approaches its maximum. In this case the sample is illuminated and detected from all sides simultaneously.

 
Optical Scheme of 4Pi Microscope

The operation mode of a 4Pi microscope is shown in the figure. The laser light is divided by a beam splitter and directed by mirrors towards the two opposing objective lenses. At the common focal point superposition of both focused light beams occurs. Excited molecules at this position emit fluorescence light, which is collected by both objective lenses, combined by the same beam splitter and deflected by a dichroic mirror onto a detector. There superposition of both emitted light pathways can take place again.

In the ideal case each objective lens can collect light from a solid angle of  . With two objective lenses one can collect from every direction (solid angle  ). The name of this type of microscopy is derived from the maximal possible solid angle for excitation and detection. Practically, one can achieve only aperture angles of about 140° for an objective lens, which corresponds to  .

The microscope can be operated in three different ways: In a 4Pi microscope of type A, the coherent superposition of excitation light is used to generate the increased resolution. The emission light is either detected from one side only or in an incoherent superposition from both sides. In a 4Pi microscope of type B, only the emission light is interfering. When operated in the type C mode, both excitation and emission light are allowed to interfere, leading to the highest possible resolution increase (~7-fold along the optical axis as compared to confocal microscopy).

In a real 4Pi microscope light cannot be applied or collected from all directions equally, leading to so-called side lobes in the point spread function. Typically (but not always) two-photon excitation microscopy is used in a 4Pi microscope in combination with an emission pinhole to lower these side lobes to a tolerable level.

History edit

In 1971, Christoph Cremer and Thomas Cremer proposed the creation of a perfect hologram, i.e. one that carries the whole field information of the emission of a point source in all directions, a so-called   hologram.[2][3] However the publication from 1978 [4] had drawn an improper physical conclusion (i.e. a point-like spot of light) and had completely missed the axial resolution increase as the actual benefit of adding the other side of the solid angle.[5] The first description of a practicable system of 4Pi microscopy, i.e. the setup with two opposing, interfering lenses, was invented by Stefan Hell in 1991.[6] He demonstrated it experimentally in 1994.[7]

In the following years, the number of applications for this microscope has grown. For example, parallel excitation and detection with 64 spots in the sample simultaneously combined with the improved spatial resolution resulted in the successful recording of the dynamics of mitochondria in yeast cells with a 4Pi microscope in 2002.[8] A commercial version was launched by microscope manufacturer Leica Microsystems in 2004[9] and later discontinued.

Up to now, the best quality in a 4Pi microscope was reached in conjunction with super-resolution techniques like the stimulated emission depletion (STED) principle.[10] Using a 4Pi microscope with appropriate excitation and de-excitation beams, it was possible to create a uniformly 50 nm sized spot, which corresponds to a decreased focal volume compared to confocal microscopy by a factor of 150–200 in fixed cells. With the combination of 4Pi microscopy and RESOLFT microscopy with switchable proteins, it is now possible to take images of living cells at low light levels with isotropic resolutions below 40 nm.[11]

See also edit

References edit

  1. ^ J. Bewersdorf; A. Egner; S.W. Hell (2004). "4Pi-Confocal Microscopy is Coming of Age" (PDF). GIT Imaging & Microscopy (4): 24–25.
  2. ^ Cremer C., Cremer T. (1971)   Punkthologramme: Physikalische Grundlagen und mögliche Anwendungen. Enclosure to Patent application DE 2116521 „Verfahren zur Darstellung bzw. Modifikation von Objekt-Details, deren Abmessungen außerhalb der sichtbaren Wellenlängen liegen" (Procedure for the imaging and modification of object details with dimensions beyond the visible wavelengths). Filed April 5, 1971; publication date October 12, 1972. Deutsches Patentamt, Berlin. http://depatisnet.dpma.de/DepatisNet/depatisnet?action=pdf&docid=DE000002116521A
  3. ^ Considerations on a laser-scanning-microscope with high resolution and depth of field: C. Cremer and T. Cremer in MICROSCOPICA ACTA VOL. 81 NUMBER 1 September, p. 31–44 (1978). Basic design of a confocal laser scanning fluorescence microscope & principle of a confocal laser scanning 4Pi fluorescence microscope, 1978 2016-03-04 at the Wayback Machine.
  4. ^ C. Cremer and T. Cremer (1978): Considerations on a laser-scanning-microscope with high resolution and depth of field Microscopica Acta VOL. 81 NUMBER 1 September, pp. 31—44 (1978)
  5. ^ The Nobel Prize in Chemistry 2014 https://www.nobelprize.org/prizes/chemistry/2014/hell/biographical/
  6. ^ European patent EP 0491289.
  7. ^ S. W. Hell; E. H. K. Stelzer; S. Lindek; C. Cremer (1994). "Confocal microscopy with an increased detection aperture: type-B 4Pi confocal microscopy". Optics Letters. 19 (3): 222–224. Bibcode:1994OptL...19..222H. CiteSeerX 10.1.1.501.598. doi:10.1364/OL.19.000222. PMID 19829598.
  8. ^ A. Egner; S. Jakobs; S. W. Hell (2002). "Fast 100-nm resolution three-dimensional microscope reveals structural plasticity of mitochondria in live yeast" (PDF). PNAS. 99 (6): 3370–3375. Bibcode:2002PNAS...99.3370E. doi:10.1073/pnas.052545099. PMC 122530. PMID 11904401.
  9. ^ Review article 4Pi microscopy.
  10. ^ R. Schmidt; C. A. Wurm; S. Jakobs; J. Engelhardt; A. Egner; S. W. Hell (2008). "Spherical nanosized focal spot unravels the interior of cells". Nature Methods. 5 (6): 539–544. doi:10.1038/nmeth.1214. hdl:11858/00-001M-0000-0012-DBBB-8. PMID 18488034. S2CID 16580036.
  11. ^ U. Böhm; S. W. Hell; R. Schmidt (2016). "4Pi-RESOLFT nanoscopy". Nature Communications. 7 (10504): 1–8. Bibcode:2016NatCo...710504B. doi:10.1038/ncomms10504. PMC 4740410. PMID 26833381.

microscope, laser, scanning, fluorescence, microscope, with, improved, axial, resolution, with, typical, range, axial, resolution, improved, which, corresponds, almost, spherical, focal, spot, with, times, less, volume, than, that, standard, confocal, microsco. A 4Pi microscope is a laser scanning fluorescence microscope with an improved axial resolution With it the typical range of the axial resolution of 500 700 nm can be improved to 100 150 nm which corresponds to an almost spherical focal spot with 5 7 times less volume than that of standard confocal microscopy 1 Contents 1 Working principle 2 History 3 See also 4 ReferencesWorking principle editThe improvement in resolution is achieved by using two opposing objective lenses which both are focused to the same geometrical location Also the difference in optical path length through each of the two objective lenses is carefully aligned to be minimal By this method molecules residing in the common focal area of both objectives can be illuminated coherently from both sides and the reflected or emitted light can also be collected coherently i e coherent superposition of emitted light on the detector is possible The solid angle W displaystyle Omega nbsp that is used for illumination and detection is increased and approaches its maximum In this case the sample is illuminated and detected from all sides simultaneously nbsp Optical Scheme of 4Pi Microscope The operation mode of a 4Pi microscope is shown in the figure The laser light is divided by a beam splitter and directed by mirrors towards the two opposing objective lenses At the common focal point superposition of both focused light beams occurs Excited molecules at this position emit fluorescence light which is collected by both objective lenses combined by the same beam splitter and deflected by a dichroic mirror onto a detector There superposition of both emitted light pathways can take place again In the ideal case each objective lens can collect light from a solid angle of W 2 p displaystyle Omega 2 pi nbsp With two objective lenses one can collect from every direction solid angle W 4 p displaystyle Omega 4 pi nbsp The name of this type of microscopy is derived from the maximal possible solid angle for excitation and detection Practically one can achieve only aperture angles of about 140 for an objective lens which corresponds to W 1 3 p displaystyle Omega approx 1 3 pi nbsp The microscope can be operated in three different ways In a 4Pi microscope of type A the coherent superposition of excitation light is used to generate the increased resolution The emission light is either detected from one side only or in an incoherent superposition from both sides In a 4Pi microscope of type B only the emission light is interfering When operated in the type C mode both excitation and emission light are allowed to interfere leading to the highest possible resolution increase 7 fold along the optical axis as compared to confocal microscopy In a real 4Pi microscope light cannot be applied or collected from all directions equally leading to so called side lobes in the point spread function Typically but not always two photon excitation microscopy is used in a 4Pi microscope in combination with an emission pinhole to lower these side lobes to a tolerable level History editIn 1971 Christoph Cremer and Thomas Cremer proposed the creation of a perfect hologram i e one that carries the whole field information of the emission of a point source in all directions a so called 4 p displaystyle 4 pi nbsp hologram 2 3 However the publication from 1978 4 had drawn an improper physical conclusion i e a point like spot of light and had completely missed the axial resolution increase as the actual benefit of adding the other side of the solid angle 5 The first description of a practicable system of 4Pi microscopy i e the setup with two opposing interfering lenses was invented by Stefan Hell in 1991 6 He demonstrated it experimentally in 1994 7 In the following years the number of applications for this microscope has grown For example parallel excitation and detection with 64 spots in the sample simultaneously combined with the improved spatial resolution resulted in the successful recording of the dynamics of mitochondria in yeast cells with a 4Pi microscope in 2002 8 A commercial version was launched by microscope manufacturer Leica Microsystems in 2004 9 and later discontinued Up to now the best quality in a 4Pi microscope was reached in conjunction with super resolution techniques like the stimulated emission depletion STED principle 10 Using a 4Pi microscope with appropriate excitation and de excitation beams it was possible to create a uniformly 50 nm sized spot which corresponds to a decreased focal volume compared to confocal microscopy by a factor of 150 200 in fixed cells With the combination of 4Pi microscopy and RESOLFT microscopy with switchable proteins it is now possible to take images of living cells at low light levels with isotropic resolutions below 40 nm 11 See also editStimulated emission depletion microscope STED Multifocal plane microscopy MUM References edit J Bewersdorf A Egner S W Hell 2004 4Pi Confocal Microscopy is Coming of Age PDF GIT Imaging amp Microscopy 4 24 25 Cremer C Cremer T 1971 4 p displaystyle 4 pi nbsp Punkthologramme Physikalische Grundlagen und mogliche Anwendungen Enclosure to Patent application DE 2116521 Verfahren zur Darstellung bzw Modifikation von Objekt Details deren Abmessungen ausserhalb der sichtbaren Wellenlangen liegen Procedure for the imaging and modification of object details with dimensions beyond the visible wavelengths Filed April 5 1971 publication date October 12 1972 Deutsches Patentamt Berlin http depatisnet dpma de DepatisNet depatisnet action pdf amp docid DE000002116521A Considerations on a laser scanning microscope with high resolution and depth of field C Cremer and T Cremer in MICROSCOPICA ACTA VOL 81 NUMBER 1 September p 31 44 1978 Basic design of a confocal laser scanning fluorescence microscope amp principle of a confocal laser scanning 4Pi fluorescence microscope 1978 Archived 2016 03 04 at the Wayback Machine C Cremer and T Cremer 1978 Considerations on a laser scanning microscope with high resolution and depth of field Microscopica Acta VOL 81 NUMBER 1 September pp 31 44 1978 The Nobel Prize in Chemistry 2014 https www nobelprize org prizes chemistry 2014 hell biographical European patent EP 0491289 S W Hell E H K Stelzer S Lindek C Cremer 1994 Confocal microscopy with an increased detection aperture type B 4Pi confocal microscopy Optics Letters 19 3 222 224 Bibcode 1994OptL 19 222H CiteSeerX 10 1 1 501 598 doi 10 1364 OL 19 000222 PMID 19829598 A Egner S Jakobs S W Hell 2002 Fast 100 nm resolution three dimensional microscope reveals structural plasticity of mitochondria in live yeast PDF PNAS 99 6 3370 3375 Bibcode 2002PNAS 99 3370E doi 10 1073 pnas 052545099 PMC 122530 PMID 11904401 Review article 4Pi microscopy R Schmidt C A Wurm S Jakobs J Engelhardt A Egner S W Hell 2008 Spherical nanosized focal spot unravels the interior of cells Nature Methods 5 6 539 544 doi 10 1038 nmeth 1214 hdl 11858 00 001M 0000 0012 DBBB 8 PMID 18488034 S2CID 16580036 U Bohm S W Hell R Schmidt 2016 4Pi RESOLFT nanoscopy Nature Communications 7 10504 1 8 Bibcode 2016NatCo 710504B doi 10 1038 ncomms10504 PMC 4740410 PMID 26833381 Retrieved from https en wikipedia org w index php title 4Pi microscope amp oldid 1093425778, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.