Resonance frequency analysis

Resonance frequency analysis (RFA) is a method used to determine stability (the level of osseointegration) in dental implants.[1][2] The stability is presented as an implant stability quotient (ISQ) value. The higher the ISQ value the higher the stability.

Utilizing RFA involves sending magnetic pulses to a small metal rod temporarily attached to the implant. As the rod vibrates, the probe reads its resonance frequency and translates it into an ISQ value.

RFA measurements are used to assess the stability of the implant immediately after placement, as well as to measure the stability during the healing time. This helps the dentist determine if further healing time (osseointegration) is needed before the prosthetic tooth is attached, as well as to identify at-risk patients with compromised bone tissue, or other risk factors.[3]

History edit

Resonance frequency analysis was first suggested as an alternative method of analyzing peri-implant bone in a scientific paper by Meredith N et al in 1996.[4] As stated in the paper’s abstract, in measuring implant stability and osseointegration, “radiographs are of value, but a standardised technique is necessary to ensure repeatability.”[4] The new technique tested involved connecting a small transducer (aluminum rod) to implants. Measurements showed that the resonance frequency increased in direct relation to the increase in the stiffness of the bone-implant interface,[4] thus demonstrating a repeatable and quantitative method of assessing stability.

The method underwent further research and in 1999 the Sweden-based company Osstell AB formed to commercialize the new technique.[5] Osstell developed a device that transmitted vibrational frequencies to a metal peg inserted in the implant and measured the frequency at which resonance was reached. Whereas Meredith et al. measured in the range of 3500–8500 kHz, Osstell developed the Implant Stability Quotient (ISQ) that translated this kHz range to a score 1–100.[6]

Resonance frequency analysis has considerable scientific interest since its advent, largely owing to the increasing number of patients demanding dental implants as the technology improves.[7] As it is a non-invasive and objective way to evaluate short- and long-term implant viability, RFA is an increasingly utilized method.[8]

Scientific foundations edit

The method that preceded RFA, percussion or “tapping,” may be used to understand the underlying functionality of RFA devices as the same principles are at work. When an implant was percussed with a blunt instrument, the nature of the sound elicited would qualitatively indicate the level of the implant’s stability.[9] A low pitched, dull sound (low frequency) indicated a loose bond with the bone, as the vibrations moved slower across the distance between the implant and surrounding tissue.[8] A high pitched, crystalline sound indicated a tight connection along the implant-bone interface, with vibrations moving quicker across a more restricted area. The dentist would make a qualitative assessment of the level of stability based on the sound heard.[10]

With RFA, vibrations are being used to determine stability, but on a micro scale and in a non-invasive manner. A metal peg (transducer) with a magnet top is attached to the implant. Magnetic pulses (alternating sine waves of uniform amplitude) cause the peg to vibrate, increasing steadily in pitch until the implant resonates.[11] The higher the resonant frequency, the more stable the implant.

Measurement edit

The frequency readings, translated to an Implant Stability Quotient (ISQ), are used as an assessment and ongoing monitoring tool.[12] Medical interpretations of ISQ values may then be used to inform treatment plans, as analyzed and documented in hundreds of clinical studies.[13]

A reading of 55 or below indicates that too much lateral movement is possible, and the implant needs to reach secondary stability (greater bonding with the bone) before the prosthesis may be attached.[14] If the resonance frequency reading increases, it signals that osseointegration is occurring.[14] Along with other diagnostic tools, measurements over time can be used to indicate the rate of osseointegration, and treatment plans may be assigned accordingly. If the rate is initially low and does not increase, it signals the implant is not viable.[10]

References edit

  1. ^ Bilbao A et al. Assessment of dental implant stability in osseodistraction-generated bone: a resonance frequency analysis. Clin Oral Implants Res Apr 28 2009
  2. ^ Veltri M et al. Soft bone primary stability of 3 different miniscrews for orthodontic anchorage: a resonance frequency investigation. Am J Orthod Dentofacial Orthop. 2009 May;135(5):642-8.
  3. ^ Valderrama, Pilar (2007). "Evaluation of Two Different Resonance Frequency Devices to Detect Implant Stability: A Clinical Trial" (PDF). Journal of Periodontology. 78 (2): 262–272. doi:10.1902/jop.2007.060143. PMID 17274715.
  4. ^ a b c Meredith, N.; Alleyne, D.; Cawley, P. (September 1996). "Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis". Clinical Oral Implants Research. 7 (3): 261–267. doi:10.1034/j.1600-0501.1996.070308.x. ISSN 0905-7161. PMID 9151590.
  5. ^ www.osstell.com, Osstell -. "About - Osstell – Implant Stability". Osstell – Implant Stability. Retrieved 2016-12-13.
  6. ^ Konstantinovic Vitomir (2015). "Assessment of implant stability by resonant frequency analysis" (PDF). Vojnosanit Pregl. 72 (2): 169–174. doi:10.2298/vsp130801063k. PMID 25831910.
  7. ^ Group, Alexis Martino, Market Research Analyst, The Key (2012-03-21). "The continuing evolution of implant production". staging.dentalproductsreport.advanstar.com. Retrieved 2016-12-13.{{cite web}}: CS1 maint: multiple names: authors list (link)
  8. ^ a b Swami V, Vijayaraghavan V, Swami V (2016). "Current trends to measure implant stability". Journal of Indian Prosthodontic Society. 16 (2): 124–130. doi:10.4103/0972-4052.176539. PMC 4837777. PMID 27141160.{{cite journal}}: CS1 maint: multiple names: authors list (link)
  9. ^ Mistry Gaurang (2014). "Measuring implant stability: A review of different methods". Journal of Dental Implants. 4 (2): 165–169. doi:10.4103/0974-6781.140891.
  10. ^ a b Sennerby Lars (2008). "Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications" (PDF). Periodontology 2000. 47: 51–66. doi:10.1111/j.1600-0757.2008.00267.x. PMID 18412573.
  11. ^ Satwalekar, Parth; Nalla, Sandeep; Reddy, Ramaswamy; Chowdary, Sheeba Glory (2015-01-01). "Clinical evaluation of osseointegration using resonance frequency analysis". Journal of Indian Prosthodontic Society. 15 (3): 192–199. doi:10.4103/0972-4052.165171. ISSN 0972-4052. PMC 4762340. PMID 26929512.
  12. ^ Implant Dentistry - A Rapidly Evolving Practice. InTech. 2011. pp. 111–126. ISBN 978-953-307-658-4.
  13. ^ www.osstell.com, Osstell -. "Scientific Database - Osstell – Implant Stability". Osstell – Implant Stability. Retrieved 2016-12-13.
  14. ^ a b Suzuki, Senichi; Kobayashi, Hiroyuki; Ogawa, Takahiro (2013-10-01). "Implant stability change and osseointegration speed of immediately loaded photofunctionalized implants" (PDF). Implant Dentistry. 22 (5): 481–490. doi:10.1097/ID.0b013e31829deb62. ISSN 1538-2982. PMID 24021973. S2CID 25708862.

resonance, frequency, analysis, method, used, determine, stability, level, osseointegration, dental, implants, stability, presented, implant, stability, quotient, value, higher, value, higher, stability, utilizing, involves, sending, magnetic, pulses, small, m. Resonance frequency analysis RFA is a method used to determine stability the level of osseointegration in dental implants 1 2 The stability is presented as an implant stability quotient ISQ value The higher the ISQ value the higher the stability Utilizing RFA involves sending magnetic pulses to a small metal rod temporarily attached to the implant As the rod vibrates the probe reads its resonance frequency and translates it into an ISQ value RFA measurements are used to assess the stability of the implant immediately after placement as well as to measure the stability during the healing time This helps the dentist determine if further healing time osseointegration is needed before the prosthetic tooth is attached as well as to identify at risk patients with compromised bone tissue or other risk factors 3 Contents 1 History 2 Scientific foundations 3 Measurement 4 ReferencesHistory editResonance frequency analysis was first suggested as an alternative method of analyzing peri implant bone in a scientific paper by Meredith N et al in 1996 4 As stated in the paper s abstract in measuring implant stability and osseointegration radiographs are of value but a standardised technique is necessary to ensure repeatability 4 The new technique tested involved connecting a small transducer aluminum rod to implants Measurements showed that the resonance frequency increased in direct relation to the increase in the stiffness of the bone implant interface 4 thus demonstrating a repeatable and quantitative method of assessing stability The method underwent further research and in 1999 the Sweden based company Osstell AB formed to commercialize the new technique 5 Osstell developed a device that transmitted vibrational frequencies to a metal peg inserted in the implant and measured the frequency at which resonance was reached Whereas Meredith et al measured in the range of 3500 8500 kHz Osstell developed the Implant Stability Quotient ISQ that translated this kHz range to a score 1 100 6 Resonance frequency analysis has considerable scientific interest since its advent largely owing to the increasing number of patients demanding dental implants as the technology improves 7 As it is a non invasive and objective way to evaluate short and long term implant viability RFA is an increasingly utilized method 8 Scientific foundations editThe method that preceded RFA percussion or tapping may be used to understand the underlying functionality of RFA devices as the same principles are at work When an implant was percussed with a blunt instrument the nature of the sound elicited would qualitatively indicate the level of the implant s stability 9 A low pitched dull sound low frequency indicated a loose bond with the bone as the vibrations moved slower across the distance between the implant and surrounding tissue 8 A high pitched crystalline sound indicated a tight connection along the implant bone interface with vibrations moving quicker across a more restricted area The dentist would make a qualitative assessment of the level of stability based on the sound heard 10 With RFA vibrations are being used to determine stability but on a micro scale and in a non invasive manner A metal peg transducer with a magnet top is attached to the implant Magnetic pulses alternating sine waves of uniform amplitude cause the peg to vibrate increasing steadily in pitch until the implant resonates 11 The higher the resonant frequency the more stable the implant Measurement editThe frequency readings translated to an Implant Stability Quotient ISQ are used as an assessment and ongoing monitoring tool 12 Medical interpretations of ISQ values may then be used to inform treatment plans as analyzed and documented in hundreds of clinical studies 13 A reading of 55 or below indicates that too much lateral movement is possible and the implant needs to reach secondary stability greater bonding with the bone before the prosthesis may be attached 14 If the resonance frequency reading increases it signals that osseointegration is occurring 14 Along with other diagnostic tools measurements over time can be used to indicate the rate of osseointegration and treatment plans may be assigned accordingly If the rate is initially low and does not increase it signals the implant is not viable 10 References edit Bilbao A et al Assessment of dental implant stability in osseodistraction generated bone a resonance frequency analysis Clin Oral Implants Res Apr 28 2009 Veltri M et al Soft bone primary stability of 3 different miniscrews for orthodontic anchorage a resonance frequency investigation Am J Orthod Dentofacial Orthop 2009 May 135 5 642 8 Valderrama Pilar 2007 Evaluation of Two Different Resonance Frequency Devices to Detect Implant Stability A Clinical Trial PDF Journal of Periodontology 78 2 262 272 doi 10 1902 jop 2007 060143 PMID 17274715 a b c Meredith N Alleyne D Cawley P September 1996 Quantitative determination of the stability of the implant tissue interface using resonance frequency analysis Clinical Oral Implants Research 7 3 261 267 doi 10 1034 j 1600 0501 1996 070308 x ISSN 0905 7161 PMID 9151590 www osstell com Osstell About Osstell Implant Stability Osstell Implant Stability Retrieved 2016 12 13 Konstantinovic Vitomir 2015 Assessment of implant stability by resonant frequency analysis PDF Vojnosanit Pregl 72 2 169 174 doi 10 2298 vsp130801063k PMID 25831910 Group Alexis Martino Market Research Analyst The Key 2012 03 21 The continuing evolution of implant production staging dentalproductsreport advanstar com Retrieved 2016 12 13 a href Template Cite web html title Template Cite web cite web a CS1 maint multiple names authors list link a b Swami V Vijayaraghavan V Swami V 2016 Current trends to measure implant stability Journal of Indian Prosthodontic Society 16 2 124 130 doi 10 4103 0972 4052 176539 PMC 4837777 PMID 27141160 a href Template Cite journal html title Template Cite journal cite journal a CS1 maint multiple names authors list link Mistry Gaurang 2014 Measuring implant stability A review of different methods Journal of Dental Implants 4 2 165 169 doi 10 4103 0974 6781 140891 a b Sennerby Lars 2008 Implant stability measurements using resonance frequency analysis biological and biomechanical aspects and clinical implications PDF Periodontology 2000 47 51 66 doi 10 1111 j 1600 0757 2008 00267 x PMID 18412573 Satwalekar Parth Nalla Sandeep Reddy Ramaswamy Chowdary Sheeba Glory 2015 01 01 Clinical evaluation of osseointegration using resonance frequency analysis Journal of Indian Prosthodontic Society 15 3 192 199 doi 10 4103 0972 4052 165171 ISSN 0972 4052 PMC 4762340 PMID 26929512 Implant Dentistry A Rapidly Evolving Practice InTech 2011 pp 111 126 ISBN 978 953 307 658 4 www osstell com Osstell Scientific Database Osstell Implant Stability Osstell Implant Stability Retrieved 2016 12 13 a b Suzuki Senichi Kobayashi Hiroyuki Ogawa Takahiro 2013 10 01 Implant stability change and osseointegration speed of immediately loaded photofunctionalized implants PDF Implant Dentistry 22 5 481 490 doi 10 1097 ID 0b013e31829deb62 ISSN 1538 2982 PMID 24021973 S2CID 25708862 Retrieved from https en wikipedia org w index php title Resonance frequency analysis amp oldid 1188025163, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.