In radiotherapy, a percentage depth dose curve (PDD) (sometimes percent depth dose curve) relates the absorbed dose deposited by a radiation beam into a medium as it varies with depth along the axis of the beam. The dose values are divided by the maximum dose, referred to as dmax, yielding a plot in terms of percentage of the maximum dose. Dose measurements are generally made in water or "water equivalent" plastic with an ionization chamber, since water is very similar to human tissue with regard to radiation scattering and absorption.

Percent depth dose (PDD), which reflects the overall percentage of dose deposited as compared to the depth of maximum dose, depends on the depth of interest, beam energy, field size, and SSD (source to surface distance) as follows. Of note, PDD generally refers to depths greater than the depth of maximum dose

  • PDD decreases with increasing depth due to the inverse square law and due to attenuation of the radiation beam
  • PDD increases with increasing radiation field size due to greater primary and scattered photons from the irradiated medium
  • PDD increases with increasing SSD because inverse square variations over a fixed distance interval are smaller at large total distance than small total distance [1]


See also edit

References edit

  1. ^ McDermott, Patrick; Orton, Colin G. (2018). The physics & technology of radiation therapy (Second ed.). p. 10-7. ISBN 978-1930524989.
  • [1] Radiation Therapy Physics, Hendee W., Ibbott G. and Hendee E. (2005) Wiley-Liss Publ., ISBN 0-471-39493-9

percentage, depth, dose, curve, radiotherapy, percentage, depth, dose, curve, sometimes, percent, depth, dose, curve, relates, absorbed, dose, deposited, radiation, beam, into, medium, varies, with, depth, along, axis, beam, dose, values, divided, maximum, dos. In radiotherapy a percentage depth dose curve PDD sometimes percent depth dose curve relates the absorbed dose deposited by a radiation beam into a medium as it varies with depth along the axis of the beam The dose values are divided by the maximum dose referred to as dmax yielding a plot in terms of percentage of the maximum dose Dose measurements are generally made in water or water equivalent plastic with an ionization chamber since water is very similar to human tissue with regard to radiation scattering and absorption Percent depth dose PDD which reflects the overall percentage of dose deposited as compared to the depth of maximum dose depends on the depth of interest beam energy field size and SSD source to surface distance as follows Of note PDD generally refers to depths greater than the depth of maximum dose PDD decreases with increasing depth due to the inverse square law and due to attenuation of the radiation beam PDD increases with increasing radiation field size due to greater primary and scattered photons from the irradiated medium PDD increases with increasing SSD because inverse square variations over a fixed distance interval are smaller at large total distance than small total distance 1 See also editDosimetry Dose profileReferences edit McDermott Patrick Orton Colin G 2018 The physics amp technology of radiation therapy Second ed p 10 7 ISBN 978 1930524989 1 Radiation Therapy Physics Hendee W Ibbott G and Hendee E 2005 Wiley Liss Publ ISBN 0 471 39493 9 Retrieved from https en wikipedia org w index php title Percentage depth dose curve amp oldid 1170304512, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.