Mass chromatogram

A mass chromatogram is a representation of mass spectrometry data as a chromatogram, where the x-axis represents time and the y-axis represents signal intensity.[1] The source data contains mass information; however, it is not graphically represented in a mass chromatogram in favor of visualizing signal intensity versus time. The most common use of this data representation is when mass spectrometry is used in conjunction with some form of chromatography, such as in liquid chromatography–mass spectrometry or gas chromatography–mass spectrometry. In this case, the x-axis represents retention time, analogous to any other chromatogram. The y-axis represents signal intensity or relative signal intensity. There are many different types of metrics that this intensity may represent, depending on what information is extracted from each mass spectrum.

Total ion current (TIC) chromatogram edit

 
An example total ion current chromatogram from an LC-MS analysis.

The total ion current (TIC) chromatogram represents the summed intensity across the entire range of masses being detected at every point in the analysis.[2][3] The range is typically several hundred mass-to-charge units or more. In complex samples, the TIC chromatogram often provides limited information as multiple analytes elute simultaneously, obscuring individual species.

Base peak chromatogram edit

 
An example base peak chromatogram from an LC-MS analysis.

The base peak chromatogram is similar to the TIC chromatogram, however it monitors only the most intense peak in each spectrum.[4][3] This means that the base peak chromatogram represents the intensity of the most intense peak at every point in the analysis. Base peak chromatograms often have a cleaner look and thus are more informative than TIC chromatograms because the background is reduced by focusing on a single analyte at every point.

Extracted-ion chromatogram (EIC or XIC) edit

 
An example extracted-ion chromatogram from an LC-MS analysis.

In an extracted-ion chromatogram (EIC or XIC), also called a reconstructed-ion chromatogram (RIC), one or more m/z values representing one or more analytes of interest are recovered ('extracted') from the entire data set for a chromatographic run.[3] The total intensity or base peak intensity within a mass tolerance window around a particular analyte's mass-to-charge ratio is plotted at every point in the analysis. The size of the mass tolerance window typically depends on the mass accuracy and mass resolution of the instrument collecting the data. This is useful for re-examining data to detect previously-unsuspected analytes, to highlight potential isomers, resolve suspected co-eluting substances, or to provide clean chromatograms of compounds of interest. An extracted-ion chromatogram is generated by separating the ions of interest from a data file containing the full mass spectrum over time after the fact; this is different from selected-ion chromatograms, discussed below, in which data is collected only for specific m/z values. A closely related term is extracted-compound chromatogram (ECC).

Selected-ion monitoring chromatogram (SIM) edit

A selected-ion monitoring (SIM) chromatogram is similar to an EIC/XIC, with the exception that the mass spectrometer is operated in SIM mode, such that only preselected m/z values are detected in the analysis.[5] SIM experiments can be performed using mass spectrometry (MS) or tandem mass spectrometry (MS/MS) instruments. They are more common on MS instruments. This differs significantly from the extracted-ion chromatogram mentioned above in that only data for the ion(s) of interest are collected in a SIM experiment; for extracted-ion chromatograms (EIC or XIC), data for an entire mass range are collected during the run and then examined for analytes of interest after the completion of the run.

Selected-reaction monitoring chromatogram (SRM, MRM) edit

The selected-reaction monitoring (SRM) experiment is very similar to the SIM experiment except that tandem mass spectrometry is used and a specific product ion of a specific parent ion is detected.[3] The mass of the parent analyte is first selected while other ions are filtered away. The parent analyte ion is then fragmented in the gas phase and a specific fragment ion is monitored. This experiment has very high specificity because the SRM chromatogram represents only ions of a particular mass that fragment in a manner that produce a very specific product mass. This type of experiment can only be performed using tandem mass spectrometry. The technology progress in the MS/MS area lead to the development of MRM, Multiple Reaction Monitoring, which allows simultaneous detection of several coeluting analytes with different parent and/or product ions.

See also edit

References edit

  1. ^ Hites, Ronald A.; Biemann, Klaus (July 1970). "Computer evaluation of continuously scanned mass spectra of gas chromatographic effluents". Analytical Chemistry. 42 (8): 855–860. doi:10.1021/ac60290a009.
  2. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "total ion current". doi:10.1351/goldbook.T06410
  3. ^ a b c d Murray, Kermit K.; Boyd, Robert K.; Eberlin, Marcos N.; Langley, G. John; Li, Liang; Naito, Yasuhide (2013). "Definitions of terms relating to mass spectrometry (IUPAC Recommendations 2013)". Pure and Applied Chemistry. 85 (7): 1515–1609. doi:10.1351/PAC-REC-06-04-06. ISSN 0033-4545.
  4. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "base peak". doi:10.1351/goldbook.B00608
  5. ^ IUPAC, Compendium of Chemical Terminology, 2nd ed. (the "Gold Book") (1997). Online corrected version: (2006–) "selected ion monitoring". doi:10.1351/goldbook.S05547

mass, chromatogram, mass, chromatogram, representation, mass, spectrometry, data, chromatogram, where, axis, represents, time, axis, represents, signal, intensity, source, data, contains, mass, information, however, graphically, represented, mass, chromatogram. A mass chromatogram is a representation of mass spectrometry data as a chromatogram where the x axis represents time and the y axis represents signal intensity 1 The source data contains mass information however it is not graphically represented in a mass chromatogram in favor of visualizing signal intensity versus time The most common use of this data representation is when mass spectrometry is used in conjunction with some form of chromatography such as in liquid chromatography mass spectrometry or gas chromatography mass spectrometry In this case the x axis represents retention time analogous to any other chromatogram The y axis represents signal intensity or relative signal intensity There are many different types of metrics that this intensity may represent depending on what information is extracted from each mass spectrum Contents 1 Total ion current TIC chromatogram 2 Base peak chromatogram 3 Extracted ion chromatogram EIC or XIC 4 Selected ion monitoring chromatogram SIM 5 Selected reaction monitoring chromatogram SRM MRM 6 See also 7 ReferencesTotal ion current TIC chromatogram edit nbsp An example total ion current chromatogram from an LC MS analysis The total ion current TIC chromatogram represents the summed intensity across the entire range of masses being detected at every point in the analysis 2 3 The range is typically several hundred mass to charge units or more In complex samples the TIC chromatogram often provides limited information as multiple analytes elute simultaneously obscuring individual species Base peak chromatogram edit nbsp An example base peak chromatogram from an LC MS analysis The base peak chromatogram is similar to the TIC chromatogram however it monitors only the most intense peak in each spectrum 4 3 This means that the base peak chromatogram represents the intensity of the most intense peak at every point in the analysis Base peak chromatograms often have a cleaner look and thus are more informative than TIC chromatograms because the background is reduced by focusing on a single analyte at every point Extracted ion chromatogram EIC or XIC edit nbsp An example extracted ion chromatogram from an LC MS analysis In an extracted ion chromatogram EIC or XIC also called a reconstructed ion chromatogram RIC one or more m z values representing one or more analytes of interest are recovered extracted from the entire data set for a chromatographic run 3 The total intensity or base peak intensity within a mass tolerance window around a particular analyte s mass to charge ratio is plotted at every point in the analysis The size of the mass tolerance window typically depends on the mass accuracy and mass resolution of the instrument collecting the data This is useful for re examining data to detect previously unsuspected analytes to highlight potential isomers resolve suspected co eluting substances or to provide clean chromatograms of compounds of interest An extracted ion chromatogram is generated by separating the ions of interest from a data file containing the full mass spectrum over time after the fact this is different from selected ion chromatograms discussed below in which data is collected only for specific m z values A closely related term is extracted compound chromatogram ECC Selected ion monitoring chromatogram SIM editA selected ion monitoring SIM chromatogram is similar to an EIC XIC with the exception that the mass spectrometer is operated in SIM mode such that only preselected m z values are detected in the analysis 5 SIM experiments can be performed using mass spectrometry MS or tandem mass spectrometry MS MS instruments They are more common on MS instruments This differs significantly from the extracted ion chromatogram mentioned above in that only data for the ion s of interest are collected in a SIM experiment for extracted ion chromatograms EIC or XIC data for an entire mass range are collected during the run and then examined for analytes of interest after the completion of the run Selected reaction monitoring chromatogram SRM MRM editThe selected reaction monitoring SRM experiment is very similar to the SIM experiment except that tandem mass spectrometry is used and a specific product ion of a specific parent ion is detected 3 The mass of the parent analyte is first selected while other ions are filtered away The parent analyte ion is then fragmented in the gas phase and a specific fragment ion is monitored This experiment has very high specificity because the SRM chromatogram represents only ions of a particular mass that fragment in a manner that produce a very specific product mass This type of experiment can only be performed using tandem mass spectrometry The technology progress in the MS MS area lead to the development of MRM Multiple Reaction Monitoring which allows simultaneous detection of several coeluting analytes with different parent and or product ions See also editMass spectrumReferences edit Hites Ronald A Biemann Klaus July 1970 Computer evaluation of continuously scanned mass spectra of gas chromatographic effluents Analytical Chemistry 42 8 855 860 doi 10 1021 ac60290a009 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 total ion current doi 10 1351 goldbook T06410 a b c d Murray Kermit K Boyd Robert K Eberlin Marcos N Langley G John Li Liang Naito Yasuhide 2013 Definitions of terms relating to mass spectrometry IUPAC Recommendations 2013 Pure and Applied Chemistry 85 7 1515 1609 doi 10 1351 PAC REC 06 04 06 ISSN 0033 4545 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 base peak doi 10 1351 goldbook B00608 IUPAC Compendium of Chemical Terminology 2nd ed the Gold Book 1997 Online corrected version 2006 selected ion monitoring doi 10 1351 goldbook S05547 Retrieved from https en wikipedia org w index php title Mass chromatogram amp oldid 1127607435, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.