Fluorescence-lifetime imaging microscopy or FLIM is an imaging technique based on the differences in the exponential decay rate of the photon emission of a fluorophore from a sample. It can be used as an imaging technique in confocal microscopy, two-photon excitation microscopy, and multiphoton tomography.

The fluorescence lifetime (FLT) of the fluorophore, rather than its intensity, is used to create the image in FLIM. Fluorescence lifetime depends on the local micro-environment of the fluorophore, thus precluding any erroneous measurements in fluorescence intensity due to change in brightness of the light source, background light intensity or limited photo-bleaching. This technique also has the advantage of minimizing the effect of photon scattering in thick layers of sample. Being dependent on the micro-environment, lifetime measurements have been used as an indicator for pH,[1] viscosity[2] and chemical species concentration.[3][4]

Fluorescence lifetimes edit

A fluorophore which is excited by a photon will drop to the ground state with a certain probability based on the decay rates through a number of different (radiative and/or nonradiative) decay pathways. To observe fluorescence, one of these pathways must be by spontaneous emission of a photon. In the ensemble description, the fluorescence emitted will decay with time according to

 

where

 .

In the above,   is time,   is the fluorescence lifetime,   is the initial fluorescence at  , and   are the rates for each decay pathway, at least one of which must be the fluorescence decay rate  . More importantly, the lifetime,   is independent of the initial intensity and of the emitted light. This can be utilized for making non-intensity based measurements in chemical sensing.[5]

Measurement edit

Fluorescence-lifetime imaging yields images with the intensity of each pixel determined by  , which allows one to view contrast between materials with different fluorescence decay rates (even if those materials fluoresce at exactly the same wavelength), and also produces images which show changes in other decay pathways, such as in FRET imaging.

Pulsed illumination edit

Fluorescence lifetimes can be determined in the time domain by using a pulsed source. When a population of fluorophores is excited by an ultrashort or delta pulse of light, the time-resolved fluorescence will decay exponentially as described above. However, if the excitation pulse or detection response is wide, the measured fluorescence, d(t), will not be purely exponential. The instrumental response function, IRF(t) will be convolved or blended with the decay function, F(t).

 

The instrumental response of the source, detector, and electronics can be measured, usually from scattered excitation light. Recovering the decay function (and corresponding lifetimes) poses additional challenges as division in the frequency domain tends to produce high noise when the denominator is close to zero.

TCSPC edit

Time-correlated single-photon counting (TCSPC) is usually employed because it compensates for variations in source intensity and single photon pulse amplitudes. Using commercial TCSPC equipment a fluorescence decay curve can be recorded with a time resolution down to 405 fs.[citation needed] [6] The recorded fluorescence decay histogram obeys Poisson statistics which is considered in determining goodness of fit during fitting. More specifically, TCSPC records times at which individual photons are detected by a fast single-photon detector (typically a photo-multiplier tube (PMT) or a single photon avalanche photo diode (SPAD)) with respect to the excitation laser pulse. The recordings are repeated for multiple laser pulses and after enough recorded events, one is able to build a histogram of the number of events across all of these recorded time points. This histogram can then be fit to an exponential function that contains the exponential lifetime decay function of interest, and the lifetime parameter can accordingly be extracted. Multi-channel PMT systems with 16[7] to 64 elements have been commercially available, whereas the recently demonstrated CMOS single-photon avalanche diode (SPAD)-TCSPC FLIM systems can offer even higher number of detection channels and additional low-cost options.[8]

Gating method edit

Pulse excitation is still used in this method. Before the pulse reaches the sample, some of the light is reflected by a dichroic mirror and gets detected by a photodiode that activates a delay generator controlling a gated optical intensifier (GOI) that sits in front of the CCD detector. The GOI only allows for detection for the fraction of time when it is open after the delay. Thus, with an adjustable delay generator, one is able to collect fluorescence emission after multiple delay times encompassing the time range of the fluorescence decay of the sample.[9][10] In recent years integrated intensified CCD cameras entered the market. These cameras consist of an image intensifier, CCD sensor and an integrated delay generator. ICCD cameras with shortest gating times of down to 200ps and delay steps of 10ps allow sub-nanosecond resolution FLIM. In combination with an endoscope this technique is used for intraoperative diagnosis of brain tumors.[11]

Phase modulation edit

Fluorescence lifetimes can be determined in the frequency domain by a phase-modulation method. The method uses a light source that is pulsed or modulated at high frequency (up to 500 MHz) such as an LED, diode laser or a continuous wave source combined with an electro-optic modulator or an acousto-optic modulator. The fluorescence is (a.) demodulated and (b.) phase shifted; both quantities are related to the characteristic decay times of the fluorophore. Also, y-components to the excitation and fluorescence sine waves will be modulated, and lifetime can be determined from the modulation ratio of these y-components. Hence, 2 values for the lifetime can be determined from the phase-modulation method. The lifetimes are determined through a fitting procedures of these experimental parameters. An advantage of PMT-based or camera-based frequency domain FLIM is its fast lifetime image acquisition making it suitable for applications such as live cell research.[12]

Analysis edit

The goal of the analysis algorithm is to extract the pure decay curve from the measured decay and to estimate the lifetime(s). The latter is usually accomplished by fitting single or multi exponential functions. A variety of methods have been developed to solve this problem. The most widely used technique is the least square iterative re-convolution which is based on the minimization of the weighted sum of the residuals. In this technique theoretical exponential decay curves are convoluted with the instrument response function, which is measured separately, and the best fit is found by iterative calculation of the residuals for different inputs until a minimum is found. For a set of observations   of the fluorescence signal in time bin i, the lifetime estimation is carried out by minimization of:

 

Besides experimental difficulties, including the wavelength dependent instrument response function, mathematical treatment of the iterative de-convolution problem is not straight forward and it is a slow process which in the early days of FLIM made it impractical for a pixel-by-pixel analysis. Non fitting methods are attractive because they offer a very fast solution to lifetime estimation. One of the major and straightforward techniques in this category is the rapid lifetime determination (RLD) method. RLD calculates the lifetimes and their amplitudes directly by dividing the decay curve into two parts of equal width  t. The analysis is performed by integrating the decay curve in equal time intervals  t:

 

Ii is the recorded signal in the i-th channel and K is the number of channels. The lifetime can be estimated using:

 

For multi exponential decays this equation provides the average lifetime. This method can be extended to analyze bi-exponential decays. One major drawback of this method is that it cannot take into account the instrument response effect and for this reason the early part of the measured decay curves should be ignored in the analyses. This means that part of the signal is discarded and the accuracy for estimating short lifetimes goes down.

One of the interesting features of the convolution theorem is that the integral of the convolution is the product of the factors that make up the integral. There are a few techniques which work in transformed space that exploit this property to recover the pure decay curve from the measured curve. Laplace and Fourier transformation along with Laguerre gauss expansion have been used to estimate the lifetime in transformed space. These approaches are faster than the deconvolution based methods but they suffer from truncation and sampling problems. Moreover, application of methods like Laguerre gauss expansion is mathematically complicated. In Fourier methods the lifetime of a single exponential decay curve is given by:

 

Where:

 

and n is the harmonic number and T is the total time range of detection.

Applications edit

FLIM has primarily been used in biology as a method to detect photosensitizers in cells and tumors as well as FRET in instances where ratiometric imaging is difficult. The technique was developed in the late 1980s and early 1990s (Gating method: Bugiel et al. 1989. König 1989,[13] Phase modulation: Lakowicz at al. 1992,[14][15]) before being more widely applied in the late 1990s. In cell culture, it has been used to study EGF receptor signaling[16] and trafficking.[17] Time domain FLIM (tdFLIM) has also been used to show the interaction of both types of nuclear intermediate filament proteins lamins A and B1 in distinct homopolymers at the nuclear envelope, which further interact with each other in higher order structures.[18] FLIM imaging is particularly useful in neurons, where light scattering by brain tissue is problematic for ratiometric imaging.[19] In neurons, FLIM imaging using pulsed illumination has been used to study Ras,[20] CaMKII, Rac, and Ran[21] family proteins. FLIM has been used in clinical multiphoton tomography to detect intradermal cancer cells as well as pharmaceutical and cosmetic compounds.

More recently FLIM has also been used to detect flavanols in plant cells.[22]

Autofluorescent coenzymes NAD(P)H and FAD[23] edit

Multi-photon FLIM is increasingly used to detect auto-fluorescence from coenzymes as markers for changes in mammalian metabolism.[24]

FRET imaging edit

Since the fluorescence lifetime of a fluorophore depends on both radiative (i.e. fluorescence) and non-radiative (i.e. quenching, FRET) processes, energy transfer from the donor molecule to the acceptor molecule will decrease the lifetime of the donor. Thus, FRET measurements using FLIM can provide a method to discriminate between the states/environments of the fluorophore.[25] In contrast to intensity-based FRET measurements, the FLIM-based FRET measurements are also insensitive to the concentration of fluorophores and can thus filter out artifacts introduced by variations in the concentration and emission intensity across the sample.

See also edit

References edit

  1. ^ Nakabayashi, Takakazu; Wang, Hui-Ping; Kinjo, Masataka; Ohta, Nobuhiro (4 June 2008). "Application of fluorescence lifetime imaging of enhanced green fluorescent protein to intracellular pH measurements". Photochemical & Photobiological Sciences. 7 (6): 668–670. doi:10.1039/B800391B. ISSN 1474-9092. PMID 18528549. S2CID 42881416.
  2. ^ Levitt, James A.; Kuimova, Marina K.; Yahioglu, Gokhan; Chung, Pei-Hua; Suhling, Klaus; Phillips, David (9 July 2009). "Membrane-Bound Molecular Rotors Measure Viscosity in Live Cells via Fluorescence Lifetime Imaging". The Journal of Physical Chemistry C. 113 (27): 11634–11642. doi:10.1021/jp9013493. hdl:10044/1/15590. ISSN 1932-7447. S2CID 96097931.
  3. ^ Ruedas-Rama, Maria J.; Orte, Angel; Hall, Elizabeth A. H.; Alvarez-Pez, Jose M.; Talavera, Eva M. (20 February 2012). "A chloride ion nanosensor for time-resolved fluorimetry and fluorescence lifetime imaging". Analyst. 137 (6): 1500–1508. Bibcode:2012Ana...137.1500R. doi:10.1039/C2AN15851E. ISSN 1364-5528. PMID 22324050.[permanent dead link]
  4. ^ Agronskaia, Alexandra V.; Tertoolen, L.; Gerritsen, Hans C. (November 2004). "Fast fluorescence lifetime imaging of calcium in living cells". Journal of Biomedical Optics. 9 (6): 1230–1237. Bibcode:2004JBO.....9.1230A. doi:10.1117/1.1806472. ISSN 1083-3668. PMID 15568944.
  5. ^ Joseph R. Lakowicz. Principles of Fluorescence Spectroscopy 3rd edition. Springer (2006). ISBN 978-0387-31278-1.[page needed]
  6. ^ "SPC-150NX, Product description". Becker & Hickl. Becker & Hickl GmbH. April 26, 2017. Retrieved April 26, 2017.
  7. ^ . Becker & Hickl. Becker & Hickl GmbH. April 26, 2017. Archived from the original on March 3, 2018. Retrieved April 26, 2017.
  8. ^ Li, Day-Uei; Arlt, Jochen; Richardson, Justin; Walker, Richard; Buts, Alex; Stoppa, David; Charbon, Edoardo; Henderson, Robert (2010). "Real-time fluorescence lifetime imaging system with a 32 × 32 0.13μm CMOS low dark-count single-photon avalanche diode array". Optics Express. 18 (10): 10257–69. Bibcode:2010OExpr..1810257L. doi:10.1364/OE.18.010257. PMID 20588879.
  9. ^ Chang, CW; Sud, D; Mycek, MA (2007). "Fluorescence Lifetime Imaging Microscopy". Digital Microscopy, 3rd Edition. Methods in Cell Biology. Vol. 81. pp. 495–524. doi:10.1016/S0091-679X(06)81024-1. ISBN 9780123740250. PMID 17519182.
  10. ^ Elson, D. S.; Munro, I; Requejo-Isidro, J; McGinty, J; Dunsby, C; Galletly, N; Stamp, G W; Neil, M A A; Lever, M J; Kellett, P A; Dymoke-Bradshaw, A; Hares, J; French, P M W (2004). "Real-time time-domain fluorescence lifetime imaging including single-shot acquisition with a segmented optical image intensifier". New Journal of Physics. 6 (1): 180. Bibcode:2004NJPh....6..180E. doi:10.1088/1367-2630/6/1/180.
  11. ^ Sun, Yinghua; Hatami, Nisa; Yee, Matthew; Marcu, Jennifer; Elson, Daniel S.; Gorin, Fredric; Schrot, Rudolph J.; Phipps, Laura (2010). "Fluorescence lifetime imaging microscopy for brain tumor image-guided surgery" (PDF). Journal of Biomedical Optics. 15 (5): 056022–056022–5. Bibcode:2010JBO....15e6022S. doi:10.1117/1.3486612. PMC 2966493. PMID 21054116.
  12. ^ Gadella, T.W.J., editor, FRET and FLIM techniques. Elsevier, 2009 https://books.google.com/books/about/FRET_and_FLIM_Techniques.html?id=uHvqu4hLhH8C&redir_esc=y[page needed]
  13. ^ Oida, T.; Sako, Y; Kusumi, A (1993). "Fluorescence lifetime imaging microscopy (flimscopy). Methodology development and application to studies of endosome fusion in single cells". Biophysical Journal. 64 (3): 676–85. Bibcode:1993BpJ....64..676O. doi:10.1016/S0006-3495(93)81427-9. PMC 1262380. PMID 8471720.
  14. ^ Lakowicz, Joseph R.; Szmacinski, Henryk; Nowaczyk, Kazimierz; Berndt, Klaus W.; Johnson, Michael (1992). "Fluorescence lifetime imaging". Analytical Biochemistry. 202 (2): 316–30. doi:10.1016/0003-2697(92)90112-K. PMC 6986422. PMID 1519759.
  15. ^ Lakowicz, Joseph R.; Szmacinski, H; Nowaczyk, K; Johnson, ML (1992). "Fluorescence Lifetime Imaging of Free and Protein-Bound NADH". Proceedings of the National Academy of Sciences of the United States of America. 89 (4): 1271–5. Bibcode:1992PNAS...89.1271L. doi:10.1073/pnas.89.4.1271. PMC 48431. PMID 1741380.
  16. ^ Wouters, Fred S.; Bastiaens, Philippe I.H. (1999). "Fluorescence lifetime imaging of receptor tyrosine kinase activity in cells". Current Biology. 9 (19): 1127–30. doi:10.1016/S0960-9822(99)80484-9. PMID 10531012. S2CID 7640970.
  17. ^ Verveer, Peter J.; Wouters, FS; Reynolds, AR; Bastiaens, PI (2000). "Quantitative Imaging of Lateral ErbB1 Receptor Signal Propagation in the Plasma Membrane". Science. 290 (5496): 1567–70. Bibcode:2000Sci...290.1567V. doi:10.1126/science.290.5496.1567. PMID 11090353.
  18. ^ Delbarre, Erwan; Tramier, Marc; Coppey-Moisan, Maïté; Gaillard, Claire; Courvalin, Jean-Claude; Buendia, Brigitte (2006). "The truncated prelamin A in Hutchinson–Gilford progeria syndrome alters segregation of A-type and B-type lamin homopolymers" (PDF). Human Molecular Genetics. 15 (7): 1113–1122. doi:10.1093/hmg/ddl026. PMID 16481358.
  19. ^ Yasuda, Ryohei (2006). "Imaging spatiotemporal dynamics of neuronal signaling using fluorescence resonance energy transfer and fluorescence lifetime imaging microscopy". Current Opinion in Neurobiology. 16 (5): 551–61. doi:10.1016/j.conb.2006.08.012. PMID 16971112. S2CID 54398436.
  20. ^ Harvey, Christopher D.; Yasuda, R; Zhong, H; Svoboda, K (2008). "The Spread of Ras Activity Triggered by Activation of a Single Dendritic Spine". Science. 321 (5885): 136–40. Bibcode:2008Sci...321..136H. doi:10.1126/science.1159675. PMC 2745709. PMID 18556515.
  21. ^ Kaláb, Petr; Soderholm, Jon (2010). "The design of Förster (fluorescence) resonance energy transfer (FRET)-based molecular sensors for Ran GTPase". Methods. 51 (2): 220–32. doi:10.1016/j.ymeth.2010.01.022. PMC 2884063. PMID 20096786.
  22. ^ Mueller-Harvey, Irene; Feucht, Walter; Polster, Juergen; Trnková, Lucie; Burgos, Pierre; Parker, Anthony W.; Botchway, Stanley W. (2012). "Two-photon excitation with pico-second fluorescence lifetime imaging to detect nuclear association of flavanols". Analytica Chimica Acta. 719: 68–75. doi:10.1016/j.aca.2011.12.068. PMID 22340533. S2CID 24094780.
  23. ^ Cao, Ruofan; Wallrabe, Horst; Siller, Karsten; Periasamy, Ammasi (2020-02-05). "Optimization of FLIM imaging, fitting and analysis for auto-fluorescent NAD(P)H and FAD in cells and tissues". Methods and Applications in Fluorescence. 8 (2): 024001. Bibcode:2020MApFl...8b4001C. doi:10.1088/2050-6120/ab6f25. ISSN 2050-6120. PMID 31972557. S2CID 210883495.
  24. ^ Datta, Rupsa; Alfonso-García, Alba; Cinco, Rachel; Gratton, Enrico (2015-05-20). "Fluorescence lifetime imaging of endogenous biomarker of oxidative stress". Scientific Reports. 5 (1): 9848. doi:10.1038/srep09848. ISSN 2045-2322. PMC 4438616. PMID 25993434.
  25. ^ Becker, Wolfgang; Bergmann, Axel (2003). "Lifetime Imaging Techniques for Optical Microscopy" (PDF). p. 4.

External links edit

  • Lifetime and spectral analysis tools in ImageJ: http://spechron.com 2013-03-11 at the Wayback Machine
  • Fluorescence Lifetime Imaging Microscopy
  • Principle of TCSPC FLIM (Becker&Hickl GmbH)

fluorescence, lifetime, imaging, microscopy, flim, redirects, here, other, uses, flim, flim, imaging, technique, based, differences, exponential, decay, rate, photon, emission, fluorophore, from, sample, used, imaging, technique, confocal, microscopy, photon, . FLIM redirects here For other uses see Flim Fluorescence lifetime imaging microscopy or FLIM is an imaging technique based on the differences in the exponential decay rate of the photon emission of a fluorophore from a sample It can be used as an imaging technique in confocal microscopy two photon excitation microscopy and multiphoton tomography The fluorescence lifetime FLT of the fluorophore rather than its intensity is used to create the image in FLIM Fluorescence lifetime depends on the local micro environment of the fluorophore thus precluding any erroneous measurements in fluorescence intensity due to change in brightness of the light source background light intensity or limited photo bleaching This technique also has the advantage of minimizing the effect of photon scattering in thick layers of sample Being dependent on the micro environment lifetime measurements have been used as an indicator for pH 1 viscosity 2 and chemical species concentration 3 4 Contents 1 Fluorescence lifetimes 2 Measurement 2 1 Pulsed illumination 2 1 1 TCSPC 2 1 2 Gating method 2 2 Phase modulation 3 Analysis 4 Applications 4 1 Autofluorescent coenzymes NAD P H and FAD 23 4 2 FRET imaging 5 See also 6 References 7 External linksFluorescence lifetimes editA fluorophore which is excited by a photon will drop to the ground state with a certain probability based on the decay rates through a number of different radiative and or nonradiative decay pathways To observe fluorescence one of these pathways must be by spontaneous emission of a photon In the ensemble description the fluorescence emitted will decay with time according to I t I 0 e t t displaystyle I t I 0 e t tau nbsp where 1 t k i displaystyle frac 1 tau sum k i nbsp In the above t displaystyle t nbsp is time t displaystyle tau nbsp is the fluorescence lifetime I 0 displaystyle I 0 nbsp is the initial fluorescence at t 0 displaystyle t 0 nbsp and k i displaystyle k i nbsp are the rates for each decay pathway at least one of which must be the fluorescence decay rate k f displaystyle k f nbsp More importantly the lifetime t displaystyle tau nbsp is independent of the initial intensity and of the emitted light This can be utilized for making non intensity based measurements in chemical sensing 5 Measurement editFluorescence lifetime imaging yields images with the intensity of each pixel determined by t displaystyle tau nbsp which allows one to view contrast between materials with different fluorescence decay rates even if those materials fluoresce at exactly the same wavelength and also produces images which show changes in other decay pathways such as in FRET imaging Pulsed illumination edit Fluorescence lifetimes can be determined in the time domain by using a pulsed source When a population of fluorophores is excited by an ultrashort or delta pulse of light the time resolved fluorescence will decay exponentially as described above However if the excitation pulse or detection response is wide the measured fluorescence d t will not be purely exponential The instrumental response function IRF t will be convolved or blended with the decay function F t d t I R F t F t displaystyle d t IRF t otimes F t nbsp The instrumental response of the source detector and electronics can be measured usually from scattered excitation light Recovering the decay function and corresponding lifetimes poses additional challenges as division in the frequency domain tends to produce high noise when the denominator is close to zero TCSPC edit Time correlated single photon counting TCSPC is usually employed because it compensates for variations in source intensity and single photon pulse amplitudes Using commercial TCSPC equipment a fluorescence decay curve can be recorded with a time resolution down to 405 fs citation needed 6 The recorded fluorescence decay histogram obeys Poisson statistics which is considered in determining goodness of fit during fitting More specifically TCSPC records times at which individual photons are detected by a fast single photon detector typically a photo multiplier tube PMT or a single photon avalanche photo diode SPAD with respect to the excitation laser pulse The recordings are repeated for multiple laser pulses and after enough recorded events one is able to build a histogram of the number of events across all of these recorded time points This histogram can then be fit to an exponential function that contains the exponential lifetime decay function of interest and the lifetime parameter can accordingly be extracted Multi channel PMT systems with 16 7 to 64 elements have been commercially available whereas the recently demonstrated CMOS single photon avalanche diode SPAD TCSPC FLIM systems can offer even higher number of detection channels and additional low cost options 8 Gating method edit Pulse excitation is still used in this method Before the pulse reaches the sample some of the light is reflected by a dichroic mirror and gets detected by a photodiode that activates a delay generator controlling a gated optical intensifier GOI that sits in front of the CCD detector The GOI only allows for detection for the fraction of time when it is open after the delay Thus with an adjustable delay generator one is able to collect fluorescence emission after multiple delay times encompassing the time range of the fluorescence decay of the sample 9 10 In recent years integrated intensified CCD cameras entered the market These cameras consist of an image intensifier CCD sensor and an integrated delay generator ICCD cameras with shortest gating times of down to 200ps and delay steps of 10ps allow sub nanosecond resolution FLIM In combination with an endoscope this technique is used for intraoperative diagnosis of brain tumors 11 Phase modulation edit Fluorescence lifetimes can be determined in the frequency domain by a phase modulation method The method uses a light source that is pulsed or modulated at high frequency up to 500 MHz such as an LED diode laser or a continuous wave source combined with an electro optic modulator or an acousto optic modulator The fluorescence is a demodulated and b phase shifted both quantities are related to the characteristic decay times of the fluorophore Also y components to the excitation and fluorescence sine waves will be modulated and lifetime can be determined from the modulation ratio of these y components Hence 2 values for the lifetime can be determined from the phase modulation method The lifetimes are determined through a fitting procedures of these experimental parameters An advantage of PMT based or camera based frequency domain FLIM is its fast lifetime image acquisition making it suitable for applications such as live cell research 12 Analysis editThis article needs additional citations for verification Please help improve this article by adding citations to reliable sources Unsourced material may be challenged and removed Find sources Fluorescence lifetime imaging microscopy news newspapers books scholar JSTOR May 2014 template removal help The goal of the analysis algorithm is to extract the pure decay curve from the measured decay and to estimate the lifetime s The latter is usually accomplished by fitting single or multi exponential functions A variety of methods have been developed to solve this problem The most widely used technique is the least square iterative re convolution which is based on the minimization of the weighted sum of the residuals In this technique theoretical exponential decay curves are convoluted with the instrument response function which is measured separately and the best fit is found by iterative calculation of the residuals for different inputs until a minimum is found For a set of observations d t i displaystyle d t i nbsp of the fluorescence signal in time bin i the lifetime estimation is carried out by minimization of x 2 i d i t i d 0 i t i a t 2 displaystyle chi 2 sum limits i left d i t i d 0i t i a tau right 2 nbsp Besides experimental difficulties including the wavelength dependent instrument response function mathematical treatment of the iterative de convolution problem is not straight forward and it is a slow process which in the early days of FLIM made it impractical for a pixel by pixel analysis Non fitting methods are attractive because they offer a very fast solution to lifetime estimation One of the major and straightforward techniques in this category is the rapid lifetime determination RLD method RLD calculates the lifetimes and their amplitudes directly by dividing the decay curve into two parts of equal width d displaystyle delta nbsp t The analysis is performed by integrating the decay curve in equal time intervals d displaystyle delta nbsp t D 0 i 1 K 2 I i d t D 1 i K 2 K I i d t displaystyle begin matrix D 0 sum limits i 1 K 2 I i delta t amp D 1 sum limits i K 2 K I i delta t end matrix nbsp Ii is the recorded signal in the i th channel and K is the number of channels The lifetime can be estimated using t d t ln D 0 D 1 displaystyle tau delta t ln D 0 D 1 nbsp For multi exponential decays this equation provides the average lifetime This method can be extended to analyze bi exponential decays One major drawback of this method is that it cannot take into account the instrument response effect and for this reason the early part of the measured decay curves should be ignored in the analyses This means that part of the signal is discarded and the accuracy for estimating short lifetimes goes down One of the interesting features of the convolution theorem is that the integral of the convolution is the product of the factors that make up the integral There are a few techniques which work in transformed space that exploit this property to recover the pure decay curve from the measured curve Laplace and Fourier transformation along with Laguerre gauss expansion have been used to estimate the lifetime in transformed space These approaches are faster than the deconvolution based methods but they suffer from truncation and sampling problems Moreover application of methods like Laguerre gauss expansion is mathematically complicated In Fourier methods the lifetime of a single exponential decay curve is given by t 1 n w A n B n displaystyle tau frac 1 n omega frac A n B n nbsp Where A n t d t sin n w t t I R F t sin n w t w t 1 w 2 t 2 B n t d t cos n w t t I R F cos n w t 1 1 n w 2 t 2 w 2 p T displaystyle begin matrix A n frac sum limits t d t sin n omega t sum limits t IRF t sin n omega t frac omega tau 1 omega 2 tau 2 amp B n frac sum limits t d t cos n omega t sum limits t IRF cos n omega t frac 1 1 n omega 2 tau 2 amp omega frac 2 pi T end matrix nbsp and n is the harmonic number and T is the total time range of detection Applications editFLIM has primarily been used in biology as a method to detect photosensitizers in cells and tumors as well as FRET in instances where ratiometric imaging is difficult The technique was developed in the late 1980s and early 1990s Gating method Bugiel et al 1989 Konig 1989 13 Phase modulation Lakowicz at al 1992 14 15 before being more widely applied in the late 1990s In cell culture it has been used to study EGF receptor signaling 16 and trafficking 17 Time domain FLIM tdFLIM has also been used to show the interaction of both types of nuclear intermediate filament proteins lamins A and B1 in distinct homopolymers at the nuclear envelope which further interact with each other in higher order structures 18 FLIM imaging is particularly useful in neurons where light scattering by brain tissue is problematic for ratiometric imaging 19 In neurons FLIM imaging using pulsed illumination has been used to study Ras 20 CaMKII Rac and Ran 21 family proteins FLIM has been used in clinical multiphoton tomography to detect intradermal cancer cells as well as pharmaceutical and cosmetic compounds More recently FLIM has also been used to detect flavanols in plant cells 22 Autofluorescent coenzymes NAD P H and FAD 23 edit Multi photon FLIM is increasingly used to detect auto fluorescence from coenzymes as markers for changes in mammalian metabolism 24 FRET imaging edit Since the fluorescence lifetime of a fluorophore depends on both radiative i e fluorescence and non radiative i e quenching FRET processes energy transfer from the donor molecule to the acceptor molecule will decrease the lifetime of the donor Thus FRET measurements using FLIM can provide a method to discriminate between the states environments of the fluorophore 25 In contrast to intensity based FRET measurements the FLIM based FRET measurements are also insensitive to the concentration of fluorophores and can thus filter out artifacts introduced by variations in the concentration and emission intensity across the sample See also editPhasor approach to fluorescence lifetime and spectral imagingReferences edit Nakabayashi Takakazu Wang Hui Ping Kinjo Masataka Ohta Nobuhiro 4 June 2008 Application of fluorescence lifetime imaging of enhanced green fluorescent protein to intracellular pH measurements Photochemical amp Photobiological Sciences 7 6 668 670 doi 10 1039 B800391B ISSN 1474 9092 PMID 18528549 S2CID 42881416 Levitt James A Kuimova Marina K Yahioglu Gokhan Chung Pei Hua Suhling Klaus Phillips David 9 July 2009 Membrane Bound Molecular Rotors Measure Viscosity in Live Cells via Fluorescence Lifetime Imaging The Journal of Physical Chemistry C 113 27 11634 11642 doi 10 1021 jp9013493 hdl 10044 1 15590 ISSN 1932 7447 S2CID 96097931 Ruedas Rama Maria J Orte Angel Hall Elizabeth A H Alvarez Pez Jose M Talavera Eva M 20 February 2012 A chloride ion nanosensor for time resolved fluorimetry and fluorescence lifetime imaging Analyst 137 6 1500 1508 Bibcode 2012Ana 137 1500R doi 10 1039 C2AN15851E ISSN 1364 5528 PMID 22324050 permanent dead link Agronskaia Alexandra V Tertoolen L Gerritsen Hans C November 2004 Fast fluorescence lifetime imaging of calcium in living cells Journal of Biomedical Optics 9 6 1230 1237 Bibcode 2004JBO 9 1230A doi 10 1117 1 1806472 ISSN 1083 3668 PMID 15568944 Joseph R Lakowicz Principles of Fluorescence Spectroscopy 3rd edition Springer 2006 ISBN 978 0387 31278 1 page needed SPC 150NX Product description Becker amp Hickl Becker amp Hickl GmbH April 26 2017 Retrieved April 26 2017 PML 16 Product description Becker amp Hickl Becker amp Hickl GmbH April 26 2017 Archived from the original on March 3 2018 Retrieved April 26 2017 Li Day Uei Arlt Jochen Richardson Justin Walker Richard Buts Alex Stoppa David Charbon Edoardo Henderson Robert 2010 Real time fluorescence lifetime imaging system with a 32 32 0 13mm CMOS low dark count single photon avalanche diode array Optics Express 18 10 10257 69 Bibcode 2010OExpr 1810257L doi 10 1364 OE 18 010257 PMID 20588879 Chang CW Sud D Mycek MA 2007 Fluorescence Lifetime Imaging Microscopy Digital Microscopy 3rd Edition Methods in Cell Biology Vol 81 pp 495 524 doi 10 1016 S0091 679X 06 81024 1 ISBN 9780123740250 PMID 17519182 Elson D S Munro I Requejo Isidro J McGinty J Dunsby C Galletly N Stamp G W Neil M A A Lever M J Kellett P A Dymoke Bradshaw A Hares J French P M W 2004 Real time time domain fluorescence lifetime imaging including single shot acquisition with a segmented optical image intensifier New Journal of Physics 6 1 180 Bibcode 2004NJPh 6 180E doi 10 1088 1367 2630 6 1 180 Sun Yinghua Hatami Nisa Yee Matthew Marcu Jennifer Elson Daniel S Gorin Fredric Schrot Rudolph J Phipps Laura 2010 Fluorescence lifetime imaging microscopy for brain tumor image guided surgery PDF Journal of Biomedical Optics 15 5 056022 056022 5 Bibcode 2010JBO 15e6022S doi 10 1117 1 3486612 PMC 2966493 PMID 21054116 Gadella T W J editor FRET and FLIM techniques Elsevier 2009 https books google com books about FRET and FLIM Techniques html id uHvqu4hLhH8C amp redir esc y page needed Oida T Sako Y Kusumi A 1993 Fluorescence lifetime imaging microscopy flimscopy Methodology development and application to studies of endosome fusion in single cells Biophysical Journal 64 3 676 85 Bibcode 1993BpJ 64 676O doi 10 1016 S0006 3495 93 81427 9 PMC 1262380 PMID 8471720 Lakowicz Joseph R Szmacinski Henryk Nowaczyk Kazimierz Berndt Klaus W Johnson Michael 1992 Fluorescence lifetime imaging Analytical Biochemistry 202 2 316 30 doi 10 1016 0003 2697 92 90112 K PMC 6986422 PMID 1519759 Lakowicz Joseph R Szmacinski H Nowaczyk K Johnson ML 1992 Fluorescence Lifetime Imaging of Free and Protein Bound NADH Proceedings of the National Academy of Sciences of the United States of America 89 4 1271 5 Bibcode 1992PNAS 89 1271L doi 10 1073 pnas 89 4 1271 PMC 48431 PMID 1741380 Wouters Fred S Bastiaens Philippe I H 1999 Fluorescence lifetime imaging of receptor tyrosine kinase activity in cells Current Biology 9 19 1127 30 doi 10 1016 S0960 9822 99 80484 9 PMID 10531012 S2CID 7640970 Verveer Peter J Wouters FS Reynolds AR Bastiaens PI 2000 Quantitative Imaging of Lateral ErbB1 Receptor Signal Propagation in the Plasma Membrane Science 290 5496 1567 70 Bibcode 2000Sci 290 1567V doi 10 1126 science 290 5496 1567 PMID 11090353 Delbarre Erwan Tramier Marc Coppey Moisan Maite Gaillard Claire Courvalin Jean Claude Buendia Brigitte 2006 The truncated prelamin A in Hutchinson Gilford progeria syndrome alters segregation of A type and B type lamin homopolymers PDF Human Molecular Genetics 15 7 1113 1122 doi 10 1093 hmg ddl026 PMID 16481358 Yasuda Ryohei 2006 Imaging spatiotemporal dynamics of neuronal signaling using fluorescence resonance energy transfer and fluorescence lifetime imaging microscopy Current Opinion in Neurobiology 16 5 551 61 doi 10 1016 j conb 2006 08 012 PMID 16971112 S2CID 54398436 Harvey Christopher D Yasuda R Zhong H Svoboda K 2008 The Spread of Ras Activity Triggered by Activation of a Single Dendritic Spine Science 321 5885 136 40 Bibcode 2008Sci 321 136H doi 10 1126 science 1159675 PMC 2745709 PMID 18556515 Kalab Petr Soderholm Jon 2010 The design of Forster fluorescence resonance energy transfer FRET based molecular sensors for Ran GTPase Methods 51 2 220 32 doi 10 1016 j ymeth 2010 01 022 PMC 2884063 PMID 20096786 Mueller Harvey Irene Feucht Walter Polster Juergen Trnkova Lucie Burgos Pierre Parker Anthony W Botchway Stanley W 2012 Two photon excitation with pico second fluorescence lifetime imaging to detect nuclear association of flavanols Analytica Chimica Acta 719 68 75 doi 10 1016 j aca 2011 12 068 PMID 22340533 S2CID 24094780 Cao Ruofan Wallrabe Horst Siller Karsten Periasamy Ammasi 2020 02 05 Optimization of FLIM imaging fitting and analysis for auto fluorescent NAD P H and FAD in cells and tissues Methods and Applications in Fluorescence 8 2 024001 Bibcode 2020MApFl 8b4001C doi 10 1088 2050 6120 ab6f25 ISSN 2050 6120 PMID 31972557 S2CID 210883495 Datta Rupsa Alfonso Garcia Alba Cinco Rachel Gratton Enrico 2015 05 20 Fluorescence lifetime imaging of endogenous biomarker of oxidative stress Scientific Reports 5 1 9848 doi 10 1038 srep09848 ISSN 2045 2322 PMC 4438616 PMID 25993434 Becker Wolfgang Bergmann Axel 2003 Lifetime Imaging Techniques for Optical Microscopy PDF p 4 External links editFluorescence Excited State Lifetime Imaging Lifetime and spectral analysis tools in ImageJ http spechron com Archived 2013 03 11 at the Wayback Machine Fluorescence Lifetime Imaging Microscopy Principle of TCSPC FLIM Becker amp Hickl GmbH Retrieved from https en wikipedia org w index php title Fluorescence lifetime imaging microscopy amp oldid 1183169216, wikipedia, wiki, book, books, library,

article

, read, download, free, free download, mp3, video, mp4, 3gp, jpg, jpeg, gif, png, picture, music, song, movie, book, game, games.